No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation into
a foreign language) without prior agreement
and written consent from Agilent Technologies, Inc. as governed by United States and
international copyright laws.
Manual Part Number
5973-1528
Edition
Seventh edition, July 2020
Sixth edition, April 2020
Fifth edition, November 2019
Fourth edition, July 2019
Third edition, August 2013
Second edition, May 2013
First edition, August 2012
Printed in USA
Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA
Warranty
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future editions.
Further, to the maximum extent permitted by
applicable law, Agilent disclaims all warranties, either express or implied, with regard to
this manual and any information contained
herein, including but not limited to the implied
warranties of merchantability and fitness for
a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the
furnishing, use, or performance of this document or of any information contained herein.
Should Agilent and the user have a separate
written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in
the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software
and technical data rights granted to the federal government include only those rights
customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical
data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data -Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in damage to the product or
loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are
fully understood and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions
are fully understood and met.
Using the Gas Clean Filter System
About the Gas Clean Filter System 4
Components 4
Features 4
Filter types 5
Safety Information 9
General gas safety 9
Cleaning 9
Recycling the product 9
Installation 10
Before you begin 10
Prepare your instrument 10
Install the connecting unit 10
Connect the gas lines 21
Install the filter to the connecting unit 23
Install a two- or four-position connecting unit 34
Gas Flow in a Gas Clean Filter 36
Gas Clean Sensor 37
About the Gas Clean sensor 37
To install the Gas Clean sensor 37
Replacing the Gas Clean Filter 39
When to replace the Gas Clean Filter 39
To replace the Gas Clean Filter 40
Checking for Leaks 43
Indicator changes color from the top down 43
Indicator changes color from the bottom up 44
Replacement Parts 45
User Manual3
About the Gas Clean Filter System
FilterConnecting unit
Inlet connector
Outlet connector
About the Gas Clean Filter System
The Agilent Gas Clean Filter system delivers high purity gases to your analytical instruments,
reducing the risk of column damage, sensitivity loss, and instrument downtime. The filters are
designed for use with the GC, GC/MS, ICP-OES, ICP-MS, LC/MS, and any other analysis
instrument using carrier gas. Six filters are available, including CO
organics trap (charcoal).
Components
The Gas Clean Filter System consists of a connecting unit and filter (Figure 1). The connecting
unit is equipped with inlet and outlet connectors for the gas lines, and two needle-like valves
that automatically start the flow of gas once a filter is attached. The filters come filled with dry
nitrogen and are made from a heavy-walled polycarbonate sealed at the base with PTFE.
When installed onto the connecting unit, the needle-like valves puncture the PTFE seals and
allow gas to flow through the filter.
, oxygen, moisture, and
2
Figure 1.The components of the Gas Clean Filter System (connecting units vary in appearance but
perform the same functionality)
Features
The Gas Clean Filter System delivers the following features and benefits:
•Cleaner gas, longer instrument life, greater sensitivity, higher data accuracy, and overall
long-term stability for your instruments and analyses.
4User Manual
Filter types
•A high-flow connecting unit that handles flow rates up to 20 L/min for collision gas
applications and supply gas for ICP-MS, ICP-OES, and LC-MS.
•A fast-stabilizing and absorbent filter material.
•Filter housing constructed with a transparent and virtually unbreakable, heavy-walled
polycarbonate material.
•The ability to change the Gas Clean Filter without tools or gas shut-off.
Filter types
Each Gas Clean Filter type is designed to filter out a specific impurity that may exist in the gas
supply. The following filter types are available:
•Carbon Dioxide - Eliminates CO
analyzer, and improves sensitivity and accuracy in TOC analysis.
•Oxygen - Prevents oxidation of the GC column, septum, liner, and glass wool.
•Moisture - Delivers faster stabilization times for increased GC productivity, and prevents
hydrolization damage to the stationary phase, column, liner, glass wool, or septum in the
GC.
•Process Moisture - Prevents oxidation of GC components and is safe to use with acetylene
in process GC applications.
from supply gas to the Total Organic Carbon (TOC)
2
•Charcoal - Removes organic compounds and ensures correct performance of FID
detectors in the GC.
•Carrier Gas - Delivers faster stabilization times for increased GC productivity, removes
oxygen, moisture, and hydrocarbons from the carrier gas for MS applications, and provides
ultimate GC column protection.
•Sulfur - Removes water and sulfur specific compounds such as H
gas streams for the SCD.
Table 1 on page 6 shows recommended filter connection diagrams for common instrument
configurations.
S, COS, and SO2 from
2
User Manual5
Filter types
ECD
Electron Capture Detector
Carrier Gas
Nitrogen
Oxygen Filter
Oxygen Filter
Column
ECD
Make-Up Gas
Anode Purge Gas
Moisture Filter
Moisture Filter
FID
Flame Ionization Detector
(Carrier Gas = Make-Up Gas)
Carrier Gas
Hydrogen
Air
Oxygen FilterMoisture Filter
FID
Charcoal Filter
Charcoal Filter
Column
FID
Flame Ionization Detector
(Carrier Gas differs
from Make-Up Gas)
Carrier Gas
Make-Up Gas
Hydrogen
Air
GC/MS Filter
Charcoal Filter
FID
Charcoal Filter
Charcoal Filter
Column
FPD
Flame Photometric Detector
PFPD
Pulsed Flame Photometric
Detector
Carrier Gas
Hydrogen
Air
Oxygen FilterMoisture Filter
FPD
Column
Charcoal Filter
Charcoal Filter
Flame 2
Flame 1
DetectorConnection Diagram
Table 1Connection diagrams for common detectors
6User Manual
Filter types
MS (ITD, MSD)
Ion Trap Detector,
Mass Selective Detector
Carrier GasOxygen FilterMoisture FilterColumn
MS
Carrier GasGC/MS FilterColumn
MS
NPD, PND
Nitrogen-Phosphorous Detector
TID, TSD
Thermionic Detector
(Carrier Gas = Make-Up Gas)
Carrier Gas
Hydrogen
Air
Oxygen Filter
Charcoal Filter
Charcoal Filter
Column
TSD
Make-Up Gas
Moisture Filter
TCD
Thermal Conductivity Detector
Carrier GasOxygen FilterMoisture FilterColumn
TCD
Reference Channel
Make-Up Gas,
if necessary
SCD
Sulfur Chemiluminescence
Detector
Hydrogen
Oxygen/Air
Sulfur Filter
Sulfur Filter
SCD
DetectorConnection Diagram
Table 1Connection diagrams for common detectors (continued)
Table 1Connection diagrams for common detectors (continued)
Contact your local Agilent sales representative for the filter sets applicable to your instrument
configuration.
8User Manual
Safety Information
Safety Information
General gas safety
•Wear eye protection when using compressed gas to avoid eye injury.
•Fasten all compressed gas cylinders securely to an immovable structure or permanent
wall.
•Store and handle compressed gases in accordance with relevant safety codes.
•Do not put gas cylinders in the path of a hot air vent (including a GC oven exhaust).
•Perform periodic leak checks on supply lines, fittings, and pneumatic plumbing to prevent a
potentially hazardous condition.
Cleaning
Wipe away dust and residue using a lint-free cloth.
Once the filter is consumed, the filter and filter material cannot be replenished or reused.
Recycling the product
When filters are replaced, they must be treated as chemical waste and disposed of according
to local policy.
User Manual9
Installation
Installation
Before you begin
Be sure to choose a centralized installation location where all instruments are easily reachable
and where the filter’s indicating material is always visible.
Gas Clean Filter connecting units are available for both 1/4-inch and 1/8-inch gas lines. Be sure
you have selected the appropriate connecting unit to use with your lab setup.
Prepare your instrument
Prepare your instrument by doing the following:
1Lower any heated zone temperatures on your instrument to less than 100 °C.
2Set any purge flows on your instrument to 400 mL/min.
3If your system pressure is higher than 7 bar (100 psi), reduce the pressure.
Install the connecting unit
The connecting unit can be mounted on a laboratory bench top, affixed to the wall (with
optional wall-mounting bracket for the single filter connecting unit), or mounted to the back of
a 7890, 9000, 8860, or 8890 GC (with the respective Gas Clean Carrier Gas Kit).
•If you wish to install your connecting until to a laboratory bench top, see “Installing on a
laboratory bench top” below.
•If you wish to install your connecting unit to a wall, skip to “Installing on a wall” on page 11.
•If you have the Gas Clean Carrier Gas Kit for 7890 (CP17988) and wish to install the
connecting unit to the back of a 7890 GC, skip to “Installing the Gas Clean Carrier Gas Kit
for 7890” on page 17.
•If you have the Gas Clean Carrier Gas Kit for 9000 (CP17995) and wish to install the
connecting unit to the back of a 9000 GC, skip to “Installing the Gas Clean Carrier Gas Kit
for 9000” on page 19.
•If you have the Gas Clean Carrier Gas Kit for 8860 and 8890 (CP179880) and wish to install
the connecting unit to the back of a 8860 or 8890 GC, skip to “Installing the Gas Clean
Carrier Gas Kit for 8860 and 8890” on page 20.
Installing on a laboratory bench top
You can install the connecting unit onto the laboratory bench top using two screws (not
supplied).
Gather the following:
•Screws (2), at least 25 mm long and less than 5-mm od (not supplied)
•Screw driver or power drill
To install the connecting unit to the laboratory bench top:
10User Manual
Install the connecting unit
1Remove the two black plastic caps by pushing a thin, narrow object (such as a paper clip)
Figure 2.Removing the plastic caps from the connecting unit
from the bottom of the connecting unit into the underside of the caps until they pop out
(Figure 2).
2Determine the best location on your laboratory bench top to install the connecting unit and
filter.
• Make sure the filter’s indicator is clearly visible so that exhausted filters can be easily
identified.
• Make sure you have ample space above the bench top and connecting unit for
installation and removal of the filter.
• Make sure the gas lines reach the location and will not interfere with other activities in
the chosen installation location.
• Check that the mounting screws will not penetrate any dangerous objects such as gas
and power lines.
3Place the connecting unit in the chosen location, and use a screw driver or power drill to
install the screws into the bench top through the two openings on the top of the connecting
unit.
4Replace the black plastic caps on the connecting unit to cover the two screw hole
openings.
Installing on a wall
This option is only available for a single filter connecting unit using the optional wall mounting
bracket.
Gather the following:
•Wall mounting bracket kit (CP7981)
•Flathead screw driver
•Hex wrench (4 mm)
•Power drill with 5-mm od drill bit
•Pencil
User Manual11
Install the connecting unit
To install the wall mounting bracket:
1Determine the location on the wall where you wish to install the wall mounting bracket.
2Hold the wall mounting bracket against the wall in the location that you plan to install it,
• Make sure you have ample space above the mounting bracket with connecting unit
installed for installation/removal of the filter.
• Make sure the gas lines reach the location and will not interfere with other activities in
the chosen installation location.
• Check that the mounting screws will not penetrate any dangerous objects such as gas
and power lines once installed.
and use a pencil to mark the drilling location for the screw holes (Figure 3).
Figure 3.Mark the drilling locations on the wall using the screw holes in the wall mounting bracket
as a guide
Lay the mounting bracket down on a flat surface.
3Using a power drill with a 5-mm od drill bit, drill about 5 cm into the wall in both locations.
Wipe away any excess dust and residue.
4If required, install the supplied drywall anchors into the drilled holes, and make sure each
anchor is flush against the wall’s edge (Figure 4).
Figure 4.Plastic drywall anchors installed in wall
12User Manual
Install the connecting unit
5Hold the wall mounting bracket in place against the wall using one hand, then use your
other hand to install each screw into the wall (or drywall anchor, Figure 5). Finger tighten
the screws as far as possible, and then use a screwdriver to complete the screw
installation. The wall mounting bracket should rest flush against the wall.
Figure 5.Installing the wall mounting bracket
6The connecting unit supports gas connections on the base plate edge for surface
mounting and from below when mounting the connecting unit against a wall. The
connecting unit ships from the factory with the gas connections installed on the edge. In
this step, the standard connecting unit is shown. Steps are similar for all connecting unit
types.
User Manual13
Install the connecting unit
Plugs
Inlet fittingOutlet fitting
You can optionally change the gas inlet-outlet locations to the bottom of the connecting
unit as follows:
a Using a 5-mm hex wrench, remove the two plugs on the underside of the connecting
unit (Figure 6).
Figure 6.Plugs and inlet/outlet fittings on the connecting unit
b Using a 1/2-inch wrench, remove the inlet and outlet fittings on the end of the
connecting unit (Figure 6).
14User Manual
Install the connecting unit
Plugs
Inlet fittingOutlet fitting
c Install the plugs on the end of the connecting unit using a 5-mm hex wrench
(Figure 7).
Figure 7.Plugs installed on the end of the connecting unit
d Install the inlet and outlet fittings on the underside of the connecting unit using a
1/2-inch wrench (Figure 8).
Figure 8.Inlet and outlet fittings installed on the underside of the connecting unit
User Manual15
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.