Information contained in this document is subject to
change without notice. Agilent Technologies makes
no warranty of any kind with regard to this material,
including, but not limited to, the implied warranties of
merchantability and fitness for a particular purpo se.
Agilent Technologies shall not be liable for errors
contained herein or for incidental or consequential
damages in connection with the furnishings,
performance, or use of this material. No part of this
document may be ph otocopi ed, repr oduce d, or
translated to another language without the prior
written consent of Agilent Technologies.
Copyright 2003 Agilent Technologies
Station Road, South Queensferry, Scotland, EH30 9TG,
UK.
Agilent Part No. E9321-90001
Printed in USA
June 2007
Legal Information
Certification
Agilent Technologies certifies that this product met its
published specifications at the time of shipment from
the factory. Agilent Technologies further certifies that
its calibration measure ments are traceable to the
United States National Inst itute of Sta ndards and
Technology, to the extent allowed by the Institute’s
calibration facility, and to the calibration facilities of
other International Standards Organization members.
2E-Series E9320 Operating and Service Guide
Page 3
Safety Symbols
Safety Symbols
The following symbols on the instrument and in the
documentation indicate precautions which must be taken to
maintain safe operation of the instrument.
The Instruction Documentation Symbol. The product
is marked with this symbol when it is necessary for
the user to refer to the instructions in the supplied
documentation.
Safety
This guide uses warnings and cautions to denote hazards
Notices
WARNINGA warning calls attention to a procedure, practice or the like,
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning until the
indicated conditions are fully understood and met.
CAUTIONA caution calls attention to a procedure, practice or the like
which, if not correctly performed or adhered to, could result in
damage to or the destruction of part or all of the equipment.
Do not proceed beyond a caution until the indicated
conditions are fully understood and met.
E-Series E9320 Operating and Service Guide3
Page 4
General Safety Information
General Safety Information
The following general safety precautions must be observed
during all phases of operation, service, and repair of this
instrument. Failure to comply with these precautions or with
specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the
instrument. Agilent Technologies assumes no liability for the
customer’s failure to comply with these requirements.
WARNINGBEFORE CONNECTING THE POWER SENSOR TO OTHER
INSTRUMENTS ensure that all instruments are connected to
the protective (earth) ground. Any interruption of the
protective earth grounding will cause a potential shock
hazard that could result in personal injury.
Sound Emission
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den
Anforderungen der Maschinenlarminformationsverordnung
vom 18 Januar 1991.
• Sound Pressure LpA < 70 dB.
• Am Arbeitsplatz.
•Normaler Betrieb.
• Nach DIN 45635 T. 19 (Typprufung).
4E-Series E9320 Operating and Service Guide
Page 5
General Safety Information
Manufacturers Declaration
This statement is provided to comply with the requirements of
the German Sound DIN 45635 T. 19 (Typprufung).
This Chapter introduces you to the E-series E9320 power
sensors. It contains the following sections:
•General Information on page 10
•Initial Inspection on page 10
•Power Meter and Sensor Cable Requirements on
page 11
•Interconnections on page 11
•The E-Series E9320 Power Sensors in Detail on page
13
Page 10
General Information
General Information
Welcome to the E-series E9320 power sensors Operating and Service Guide! This guide contains information about the
initial inspection, connection, and specifications of the Eseries E9320 power sensors. You can also find a copy of this
guide on the CD-ROM supplied with the EPM-P series peak
and average power meters.
To make best use of your sensor refer to the chapter “Using
E-Series E9320 Sensors” in the EPM-P Series Power Meters
Operating and Service Guide.
Initial Inspection
Inspect the shipping container for damage. If the shipping
container or packaging material is damaged, it should be kept
until the contents of the shipment have been checked
mechanically and electrically. If there is mechanical damage,
notify the nearest Agilent office. Keep the damaged shipping
materials (if any) for inspection by the carrier and an Agilent
representative. If required, you can find a list of Agilent Sales
and Service offices on page 56.
10E-Series E9320 Operating and Service Guide
Page 11
General Information
Power Meter and Sensor Cable
Requirements
The E-series E9320 power sensors are compatible ONLY with
the EPM-P series power meters and with E9288 sensor
cables. (The E9288 cables are color coded to help distinguish
them from the 11730 series cables.)
Interconnections
Connect one end of an E9288 sensor cable to the E-series
E9320 power sensor and connect the other end of the cable to
the power meter’s channel input. Allow a few seconds for the
power meter to download the data contained in the power
sensor.
Ensure power sensors and cables are attached and removed
in an indoor environment.
Calibration
To carry out a zero and calibration cycle as requested by the
power meter proceed as follows:
•Ensure the E-series E9320 power sensor is disconnected
When the wait period is complete connect the E-series E9320
power sensor to the power meter’s POWER REF output.
•Press (or , / ). The wait symbol is
from any signal source. On the power meter, press
Zero
, (or / ). During zeroing
Cal
the wait symbol is displayed.
again displayed during calibration.
ZeroZero AZero B
CalCal Cal ACal B
TIPYou can reduce the steps required to carry out the zero and
calibration procedure as follows:
E-Series E9320 Operating and Service Guide11
Page 12
General Information
Connect the power sensor to the POWER REF output.
•Press and . (For dual channel
Zero
Cal
meters, press , or
Zero + Cal B
Zero + Cal
Zero + CalZero + Cal A
as required.)
On completion, the power meter and sensor are ready to
connect to the device under test (DUT).
WARNINGBEFORE CONNECTING THE POWER SENSOR TO OTHER
INSTRUMENTS ensure that all instruments are connected to
the protective (earth) ground. Any interruption of the
protective earth grounding will cause a potential shock hazard
that could result in personal injury.
CAUTIONThe measurement connector (for connection to DUT) is
Type-N (male). A torque wrench should be used to tighten
these connectors. Use a 3/4-inch open-end wrench and
torque to 12 in-lb (135 Ncm) for the Type-N connector.
12E-Series E9320 Operating and Service Guide
Page 13
The E-Series E9320 Power Sensors in Detail
The E-Series E9320 Power
Sensors in Detail
The E-series E9320 power sensors have two frequency
ranges. The E9325A, E9326A, and E9327A have a frequency
range of 50 MHz to 18 GHz while the 50 MHz to 6 GHz range of
the E9321A, E9322A, and E9323A covers most wireless
communication applications.
The sensors have two independent measurement paths as
shown in Figure 1.
Normal
Path
Average Only Path
Chopper
Selectable Gain
Differential Amplifier*
(300 kHz, 1.5 MHz, 5 MHz)
PEAK AUTO-ZERO
Thermistor Bias
2
C Buffer
I
Gain/Mode Control
Sensor ID
2
E
PROM
CW Differential
AMP
Switched
Gain
Preamp
GAIN SELECT
SERIAL BUS
Average
Only Path
CHOP/CLOCK
CHOP/GAIN
Normal
Path
RF IN
50ohm
Sensor D iode
Bulkhead
* Bandwi dth is sens or dependent
Load Filter*
(300 kHz, 1.5 MHz
5 MHz low pass)
Figure 1 Simplified Sensor Block Diagram
Use the default normal path for continuously sampled
measurements of modulated signals and time gated
measurements. For each frequency range there is a choice of
sensors with three video (modulation) bandwidths.
E-Series E9320 Operating and Service Guide13
Page 14
The E-Series E9320 Power Sensors in Detail
•E9321A and E9325A sensors with 300 kHz bandwidth are
suitable for measuring TDMA signals such as GSM.
•E9322A and E9326A sensors with 1.5 MHz bandwidth are
suitable for measuring IS-95 CDMA signals.
•E9323A and E9327A sensors with 5 MHz bandwidth are
suitable for measuring W-CDMA signals.
Note however, that the sensors with widest bandwidth have
the smallest dynamic range (in normal mode). If dynamic
range is an important factor, use the sensor model with just
enough video bandwidth for the measurement you want to
make.
The average-only path is suitable for average power
measurements of Continuous Wave (CW) and constant
amplitude signals between -65 dBm (sensor dependent) and
+20 dBm. The average-only path can also be used to measure
true average power of any complex modulated signal below 20 dBm.
Calibration factors, linearity, temperature, and bandwidth
compensation data are stored in the sensor EEPROM during
the manufacturing process. All the compensation data is
downloaded to the EPM-P series power meter at power-on or
when the sensor is connected. You only need to enter the
frequency of the RF signal you are measuring to achieve a high
degree of accuracy.
14E-Series E9320 Operating and Service Guide
Page 15
2Specifications and
Characteristics
What You’ll Find
In This Chapter
This Chapter describes the specifications and
characteristics of the E-series E9320 power sensors. It
contains the following sections:
•“Introduction” on page 16
•“Specifications” on page 17
Page 16
Introduction
Introduction
E-series E9320 power sensors are designed for use with the
Agilent EPM series power meters. The E-series E9320 power
sensors have two measurement paths:
•Normal path: (default mode) for peak, average and
time-related measurements.
•Average-only path: is designed primarily for average
power measurements on low level signals.
These specifications are valid ONLY after zero and calibration
of the power meter and sensor.
Supplemental characteristics, which are shown in italics, are
intended to provide information useful in applying the power
sensors by giving typical, but nonwarranted performance
parameters. These characteristics are shown in italics or
denoted as “typical”, “nominal” or “approximate”.
16E-Series E9320 Operating and Service Guide
Page 17
Specifications
Specifications
Frequency, Bandwidth, and Power Range
Table 1 Frequency, Bandwidth, and Power Range
Maximum
Power Range
Video
Sensor
E9321A300 kHz50 MHz
E9325A300 kHz50 MHz
E9322A1.5 MHz50 MHz
E9326A1.5 MHz50 MHz
E9323A5 MHz50 MHz
E9327A5 MHz50 MHz
Bandwidth
* For average power measurements, Free Run acquisition mode.
Frequency
Range
to
6GHz
to
18 GHz
to
6GHz
to
18 GHz
to
6GHz
to
18 GHz
Average
only Mode
–65 dBm
to
+20 dBm
–65 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
Normal
Mode*
–50 dBm
to
+20 dBm
–50 dBm
to
+20 dBm
–45 dBm
to
+20 dBm
–45 dBm
to
+20 dBm
–40 dBm
to
+20 dBm
–40 dBm
to
+20 dBm
E-Series E9320 Operating and Service Guide17
Page 18
Maximum Power, RF Connector
Table 2 Maximum Power, RF Connector
Specifications
Sensor
E9321A
E9322A
E9323A
E9325A
E9326A
E9327A
RF
Connector
N-Type (m)+23 dBm average+30 dBm,
Maximum
Average Power
Maximum Peak
Power
<10µs duration
18E-Series E9320 Operating and Service Guide
Page 19
Specifications
Measurement Ranges
The E-series E9320 power sensors have two measurement
ranges (Lower and Upper) as shown in Table 3, Table 4, and
Ta b le 5 .
Table 3 Lower and Upper Measurement Ranges
E9321A and E9325A
Lower Range (Min. Power)
Normal
mode
−50 dBm −65 dBm
Average only
mode
Lower Range (Max. Power)
Lower to Upper Auto Range
+0.5 dBm
Point
Upper to Lower Auto Range
Point
Upper Range (Min. Power)
Upper Range (Max. Power
1
Applies to CW and constant amplitude signals only
-9.5dBm-18.5 dBm
−35 dBm−50 dBm
+20 dBm
above −20 dBm.
−17.5 dBm
+20 dBm
1
1
E-Series E9320 Operating and Service Guide19
Page 20
Specifications
Table 4 Lower and Upper Measurement Ranges
E9322A and E9326A
Lower Range (Min. Power)
Normal
mode
−45 dBm −60 dBm
Average only
mode
Lower Range (Max. Power)
Lower to Upper Auto Range
−5dBm
−13.5 dBm
Point
Upper to Lower Auto Range
Point
−15 dBm−14.5 dBm
Upper Range (Min. Power)−35 dBm−45 dBm
Upper Range (Max. Power
1
Applies to CW and constant amplitude signals only
+20 dBm
+20 dBm
above −20 dBm.
1
1
20E-Series E9320 Operating and Service Guide
Page 21
Specifications
Table 5 Lower and Upper Measurement Ranges
E9323A and E9327A
Lower Range (Min. Power)
Lower Range (Max. Power)
Normal
mode
−40 dBm −60 dBm
−5dBm
Average only
mode
Lower to Upper Auto Range
Point
Upper to Lower Auto Range
−15 dBm−11.5 dBm
Point
Upper Range (Min. Power)
−30 dBm−35 dBm
Upper Range (Max. Power+20 dBm
1
Applies to CW and constant amplitude signals only
above −20 dBm.
−10.5 dBm
20 dBm
+
1
1
E-Series E9320 Operating and Service Guide21
Page 22
Power Sensor Maximum SWR
Table 6 Power Sensor Maximum SWR
Sensor Maximum SWR ≤0dBm
Specifications
1.12
1.16
1.23
1.28
1.12
1.18
1.21
1.27
1.14
1.22
1.26
16GHz
17GHz
SWR
1.300
1.250
1.200
1.150
1.100
1.050
1.000
0.02GHz
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
0.05GHz
50 MHz to 2 GHz:
2GHz to 10GHz:
10 GHz to 16 GHz:
16 GHz to 18 GHz:
50 MHz to 2 GHz:
2GHz to 12GHz:
12 GHz to 16 GHz:
16 GHz to 18 GHz:
50 MHz to 2 GHz:
2 GHz to 16 GHz:
16 GHz to 18 GHz:
1GHz
6GHz
9GHz
12GHz
14GHz
0.3GHz
0.5GHz
15GHz
Figure 2 Typicl SWR (50 MHz to 18 GHz) for the E9321A and E9325A
sensors at various power levels
-30
-20
-10
0
10
20
Freq
18GHz
22E-Series E9320 Operating and Service Guide
Page 23
Specifications
1.300
SWR
1.250
1.200
1.150
1.100
1.050
1.000
1GHz
6GHz
9GHz
12GHz
14GHz
15GHz
16GHz
0.3GHz
0.02GHz
0.05GHz
0.5GHz
17GHz
Figure 3 Typical SWR (50 MHz to 18 GHz) for the E9322A and E9326A
sensors at various power levels
1.300
SWR
1.250
1.200
1.150
1.100
1.050
1.000
1GHz
6GHz
9GHz
12GHz
14GHz
15GHz
16GHz
0.3GHz
0.02GHz
0.5GHz
0.05GHz
17GHz
-30
-20
-10
0
10
20
Freq
18GHz
-30
-20
-10
0
10
20
Freq
18GHz
Figure 4 Typical SWR (50 MHz to 18 GHz) for the E9323A and E9327A
sensors at various power levels
E-Series E9320 Operating and Service Guide23
Page 24
Sensor Linearity
Table 7 Power Sensor Linearity Normal Mode
(upper and lower range)
Specifications
Sensor
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
Temperature
(25 ±10°C)
±4.2%±5.0%
±4.2%±5.0%
±4.2%±5.5%
Te mp e ra t u r e
(0 to 55°C)
Table 8 Power Sensor Linearity Average-Only Mode
(upper and lower range)
Sensor
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
Te mp e ra t u r e
(25 ±10°C)
±3.7%±4.5%
±3.7%±4.5%
±3.7%±5.0%
Temperature
(0 to 55°C)
If the temperature changes after calibration and the meter and
sensor are NOT recalibrated, the following additional linearity
errors should be added to the Power Linearity figures in Table
7 and Table 8.
24E-Series E9320 Operating and Service Guide
Page 25
Specifications
Table 9 Additional Linearity Error (normal and average-
only modes)
%
-0.50
-1.00
-1.50
-2.00
2.00
1.50
1.00
0.50
0.00
Temperature
Sensor
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
1-7-3026
-30
-27-23
-19-15
1
-
(25 ±10°C)
±1.0%±1.0%
±1.0%±1.5%
±1.0%±2.0%
4
18
10
1
dBm
20
Temperature
(0 to 55°C)
Figure 5 Typical Power Linearity at 25°C for E9323A and E9327A
5 MHz bandwidth sensors after zero and calibration, with
associated measurement uncertainty
Aver age Only
Normal (lower range)
Norm al (upper range)
E-Series E9320 Operating and Service Guide25
Page 26
+20 dBm
0dBm
Specifications
Power RangeMeasurement Uncertainty
−30 to −20 dBm±0.9%
−20 to −10 dBm±0.8%
−10 to 0 dBm±0.65%
0 to +10 dBm±0.55%
+10 to +20 dBm±0.45%
±
±
4.0%
2.0%
±
2.0%
±
4.0%
−
60 dBm
−
65 dBm
−
65 dBm
−
60 dBm
0 dBm+20 dBm
Figure 6 Relative Mode Power Measurement Linearity with an
EPM-P series power meter, at 25°C (typical)
26E-Series E9320 Operating and Service Guide
Page 27
Specifications
Figure 6 shows the typical uncertainty in making a relative
power measurement, using the same power meter channel
and the same power sensor to obtain the reference and
measured values. It also assumes that negligible change in
frequency and mismatch error occurs when transitioning from
the power level used as the reference to the power level
measured.
E-Series E9320 Operating and Service Guide27
Page 28
Peak Flatness
The peak flatness is the flatness of a Peak-to-Average ratio
measurement for various tone separations for an equal
magnitude two-tone RF input. Figure 7, Figure 8, and Figure 9
refer to the relative error in Peak-to-Average measurement as
the tone separation is varied. The measurements were
performed at −10 dBm average power using an E9288A sensor
cable.
Error (dB)
0.5
0.0
-0.5
-1.0
-1.5
-2.0
-2.5
-3.0
-3.5
0.01.02.03.04.05.06.0
Input Tone Seperation Frequency (MHz)
Specifications
HIGH
LOW
MED
OFF
Figure 7 E9321A and E9325A Error in Peak-to-Average
Measurements for a Two-tone Input
(High, Medium, Low, and Off Filters)
28E-Series E9320 Operating and Service Guide
Page 29
Specifications
Error (dB)
0.5
0
-0.5
-1
-1.5
-2
-2.5
-3
-3.5
0.01.02.03.04.05.06.0
Input Tone Seperation Frequency (MHz)
Figure 8 Filter responses for the E9322A and E9326A power sensors
(High, Medium, Low, and Off)
Error (dB)
0.5
0.0
-0.5
-1.0
-1.5
-2.0
-2.5
-3.0
-3.5
0.01.02.03.04.05.06.0
Input Tone Seperation Frequency (MHz)
HIGH
LOW
MED
OFF
LOW
MED
OFF
HIGH
Figure 9 Filter responses for the E9323A and E9327A power sensors
(High, Medium, Low, and Off)
E-Series E9320 Operating and Service Guide29
Page 30
Specifications
Calibration Factor (CF) and Reflection
Coefficient (Rho)
Calibration Factor and Reflection Coefficient data are provided
at frequency intervals on a data sheet included with the power
sensor. This data is unique to each sensor. If you have more
than one sensor, match the serial number on the data sheet
with the serial number of the sensor you are using. The CF
corrects for the frequency response of the sensor. The EPM-P
Series power meter automatically reads the CF data stored in
the sensor and uses it to make corrections.
For power levels greater than 0 dBm, add to the calibration
factor uncertainty specification:
±0.1% per dB for E9321A and E9325A power sensors
±0.15% per dB for E9322A and E9326A power sensors
±0.2% per dB for E9323A and E9327A power sensors
Reflection coefficient (Rho) relates to the SWR according to
the formula:
SWR=(1+Rho)/(1−Rho)
Maximum uncertainties of the CF data are listed in Table 10.
The uncertainty analysis for the calibration of the sensors was
calculated in accordance with the ISO Guide. The uncertainty
data, reported on the calibration certificate, is the expanded
uncertainty with a 95% confidence level and a coverage factor
of 2.
30E-Series E9320 Operating and Service Guide
Page 31
Specifications
Table 10 Cal Factor Uncertainty at 0.1 mW (-10dBm
Frequency
50 MHz
100 MHz
300 MHz
500 MHz
800 MHz
1.0 GHz
1.2 GHz
1.5 GHz
2.0 GHz
3.0 GHz
4.0 GHz
5.0 GHz
6.0 GHz
7.0 GHz
8.0 GHz
9.0 GHz
11.0 GHz
12.0 GHz
13.0 GHz
14.0 GHz
15.0 GHz
16.0 GHz
17.0 GHz
18.0 GHz
Uncertainty (%)
(25 ±10°C)
ReferenceReference
±1.8±2.0
±1.8±2.0
±1.8±2.0
±1.8±2.0
±2.1±2.3
±2.1±2.3
±2.1±2.3
±2.1±2.3
±2.1±2.3
±2.1±2.3
±2.1±2.3
±2.1±2.3
±2.3±2.5
±2.3±2.5
±2.3±2.5
±2.3±2.5
±2.3±2.5
±2.3±2.5
±2.5±2.8
±2.5±2.8
±2.5 ±2.8
±2.5±2.8
±2.5±2.8
Uncertainty(%)
(0 to 55°C)
)
E-Series E9320 Operating and Service Guide31
Page 32
Specifications
Zero Set
This specification applies to a ZERO performed when the
sensor input is not connected to the POWER REFERENCE.
Ta b le 1 1 Z er o S et
Sensor
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
Zero Set
(Normal mode)
5 nW0.17 nW
19 nW0.5 nW
60 nW0.6 nW
Zero Set
(Average-only Mode)
32E-Series E9320 Operating and Service Guide
Page 33
Specifications
Zero Drift
Sensor
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
Zero Drift and Measurement Noise
Table 12 Zero Drift and Measurement Noise
Normal
Mode
1
Average only
Mode
Measurement Noise
Normal
3
Mode
<±5 nW<±60 pW<6 nW<75 nW<165 pW
<±5 nW<±100 pW<12 nW<180 nW<330 pW
<±40 nW<±100 pW<25 nW<550 nW<400 pW
1
Within 1 hour after zero set, at a constant temperature, after a
24 hour warm-up of the power meter.
2
Measured over a one minute interval, at a constant temperature,
two standard deviations, with averaging set to 1 (normal mode), 16
(for average-only mode, normal speed) and 32 (average-only mode,
x2 speed).
3
In Free run acquisition mode.
4
Noise per sample, Video bandwidth set to OFF with no averaging
(i.e., averaging set to 1) - see “Effect of Video Bandwidth Settings”
and Table 14.
2
Normal
Mode
Average
4
only
Mode
E-Series E9320 Operating and Service Guide33
Page 34
Specifications
Effect of Averaging on Noise: Averaging over 1 to 1024
readings is available for reducing noise. Table 12 provides the
Measurement Noise for a particular sensor. Use the Noise
Multipliers in Table 13 for the appropriate speed (normal or
x2), or measurement mode (normal and average-only), and the
number of averages to determine the total Measurement
Noise value.
In addition, for x2 speed (in normal mode) the total
Measurement Noise should be multiplied by 1.2, and for fast
speed (in normal mode) the multiplier is 3.4. Note that in fast
speed no additional averaging is implemented.
34E-Series E9320 Operating and Service Guide
Page 35
Specifications
Table 13 Noise Multipliers
Noise Multiplier
Average-onlyNormal
Number of
Averages
Normal
Speed
X 2
Speed
Normal Speed,
Free run acquisition
15.56.51.0
23.894.60.94
42.753.250.88
81.942.30.82
161.01.630.76
320.851.00.70
640.610.720.64
1280.490.570.58
2560.340.410.52
5120.240.290.46
10240.170.20.40
Example: E9321A power sensor, Number of Averages = 4,
Free Run acquisition, normal mode, x2 speed.
Measurement Noise calculation:
(<6 nW x 0.88 x 1.2) = <6.34 nW
E-Series E9320 Operating and Service Guide35
Page 36
Specifications
Effect of Video Bandwidth Setting: The noise per sample is
reduced by applying the meter video bandwidth reduction
filter setting (High, Medium, or Low). If averaging is
implemented, this will dominate any effect of changing the
video bandwidth.
Table 14 Effect of Video Bandwidth on Noise per Sample
Noise Multipliers
Sensor
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
LowMediumHigh
0.320.50.63
0.500.630.80
0.400.631.0
Example: E9322A power sensor, triggered acquisition, high
video bandwidth.
Noise per sample calculation:
(<180nW x 0.80) = <144 nW
Effect of Time-Gating on Measurement Noise: The
measurement noise will depend on the time gate length over
which measurements are made. Effectively 20 averages are
carried out every 1 µs of gate length.
36E-Series E9320 Operating and Service Guide
Page 37
Specifications
Settling Times
Average Only Mode:
In normal and x2 speed mode, manual filter, 10 dB decreasing
power step, refer to Table 15.
Table 15 Averages vs. Settling Time (Average-only mode)
Number of
Averages
10245727
In Fast mode, within the range −50 to +20 dBm, for a 10 dB
decreasing power step, the settling time is:
E4416A: 10 ms
E4417A:20 ms
When a power step crosses the power sensor’s auto-range
switch point, add 25 ms.
Normal Speed ModeX 2 Speed Mode
10.080.07
20.130.09
40.240.15
80.450.24
161.10.45
321.91.1
643.51.9
1286.73.5
256146.7
5122714
Settling Time (s)
E-Series E9320 Operating and Service Guide37
Page 38
Specifications
Normal Mode:
In Normal, free run acquisition mode, within the range −20 to
+20 dBm, for a 10 dB decreasing power step, the settling time
is dominated by the measurement update rate and is listed in
Table 16 for various filter settings.
Table 16 Settling Time vs. Averages
Settling Time
Number of
Averages
Free Run Acquisition Mode
Normal SpeedX 2 Speed
10.1 sec0.08 sec
20.15 sec0.1 sec
40.25 sec0.15 sec
80.45 sec0.25 sec
160.9 sec0.45 sec
321.7 sec0.9 sec
643.3 sec1.7 sec
1286.5 sec3.3 sec
25613.0 sec6.5 sec
51225.8 sec13.0 sec
102451.5 sec25.8 sec
38E-Series E9320 Operating and Service Guide
Page 39
Specifications
In normal mode, measuring in continuous or single acquisition
mode, the performance of rise times, fall times and 99%
settled results are shown in Table 17. Rise and fall time
specifications are for a 0.0 dBm pulse, with the rise time and
fall time measured between 10% to 90% points and upper
range selected.
Table 17 Rise and Fall Times vs. Sensor Bandwidth
1
Video Bandwidth Setting
Sensor
E9321A
E9325A
E9322A
E9326A
E9323A
E9327A
1
Rise and fall time specifications are only valid when used
ParameterLow Medium High Off
Rise time < µs2.61.50.90.3
Fall time
< µs2.71.50.90.5
Settling time
(rising)
< µs
Settling time
< µs
(falling)
Rise time < µs1.50.90.40.2
Fall time
< µs1.50.90.40.3
Settling time
< µs
(rising)
Settling time
< µs
(falling)
Rise time < µs0.90.40.20.2
Fall time
< µs0.90.40.20.2
Settling time
(rising)
< µs
Settling time
< µs
(falling)
5.15.14.50.6
5.15.14.50.9
5.34.53.50.5
5.34.53.50.9
4.53.51.50.4
4.53.520.4
with the E9288A sensor cable (1.5 metres).
E-Series E9320 Operating and Service Guide39
Page 40
Specifications
Overshoot in response to power steps with fast rise times, i.e.,
less than the sensor rise time is less than 10%.
When a power step crosses the power sensor’s auto-range
switch point, add 10 µs.
40E-Series E9320 Operating and Service Guide
Page 41
Specifications
Physical Specifications
Table 18 Physical Specifications
Physical Characteristics
Net Weight
Shipping Weight
Dimensions
0.2 kg (0.45 lb)
0.55 kg (1.2 lb)
Length: 150 mm (5.9 in)
Width: 38 mm (1.5 in)
Height: 30 mm (1.2 in)
E-Series E9320 Operating and Service Guide41
Page 42
Specifications
42E-Series E9320 Operating and Service Guide
Page 43
3Service
What You’ll Find
In This Chapter
This Chapter introduces you to the E-series E9320 power
sensors. It contains the following sections:
•General Information on page 44
•Service on page 51
•Sales and Service Offices on page 56
Page 44
General Information
General Information
This chapter contains information about general maintenance,
performance tests, troubleshooting and repair of E-series
E9320 power sensors.
Cleaning
Use a clean, damp cloth to clean the body of the
E-series E9320 power sensor.
Connector Cleaning
The RF connector beads deteriorate when contacted by
hydrocarbon compounds such as acetone, trichloroethylene,
carbon tetrachloride, and benzene.
Clean the connector only at a static free workstation.
Electrostatic discharge to the center pin of the connector will
render the power sensor inoperative.
Keeping in mind its flammable nature; a solution of pure
isopropyl or ethyl alcohol can be used to clean the connector.
Clean the connector face using a cotton swab dipped in
isopropyl alcohol. If the swab is too big use a round wooden
toothpick wrapped in a lint free cotton cloth dipped in
isopropyl alcohol.
44E-Series E9320 Operating and Service Guide
Page 45
General Information
Performance Test
Standing Wave Ratio (SWR) and Reflection
Coefficient (Rho) Performance Test
This section does not establish preset SWR test procedures
since there are several test methods and different equipment
available for testing the SWR or reflection coefficient.
Therefore, the actual accuracy of the test equipment must be
accounted for when measuring against instrument
specifications to determine a pass or fail condition. The test
system used must not exceed the system Rho uncertainties
shown in Table 19 when testing the E9325A, E9326A, and
E9327A. Use Table 20 when testing the E9321A, E9322A, and
E9323A.
Table 19 Reflection Coefficient for the E9321A and E9325A
Frequency
50 MHz to 2 GHz
2 GHz to 10 GHz
10 GHz to 16 GHz±0.0100.103
16 GHz to 18 GHz±0.0100.123
E-Series E9320 Operating and Service Guide45
System Rho
Uncertainty
±0.0100.057
±0.0100.074
Measured
value
Maximum
Rho
Page 46
General Information
Table 20 Reflection Coefficient for the E9322A and E9326A
System Rho
Frequency
50 MHz to 2 GHz±0.0100.057
2 GHz to 10 GHz
10 GHz to 16GHz±0.0100.095
16 GHz to 18GHz±0.0100.119
Uncertainty
±0.0100.083
Measured
value
Maximum
Table 21 Reflection Coefficient for the E9323A and E9327A
Frequency
50 MHz to 2 GHz±0.0100.065
2 GHz to 16GHz±0.0100.099
16 GHz to 18GHz±0.0100.115
System Rho
Uncertainty
Measured
value
Maximum
Rho
Rho
46E-Series E9320 Operating and Service Guide
Page 47
General Information
Replaceable Parts
Table 22 is a list of replaceable parts. Figure 10 is the
illustrated parts breakdown (IPB) that identifies all of the
replaceable parts. To order a part, quote the Agilent part
number, specify the quantity required, and address the order
to the nearest Agilent office.
NOTEWithin the USA, it is better to order directly from the Agilent
Parts Center in Roseville, California. Ask your nearest Agilent
office for information and forms for the “Direct Mail Order
System.” Also your nearest Agilent office can supply toll free
telephone numbers for ordering parts and supplies
E-Series E9320 Operating and Service Guide47
Page 48
General Information
.
Figure 10 Illustrated Parts Break down
48E-Series E9320 Operating and Service Guide
Page 49
General Information
Table 22 Replaceable Parts List
Reference
Designation
A1/A2
E9321A
A1/A2
E9321A
A1/A2
E9322A
A1/A2
E9322A
A1/A2
E9323A
A1/A2
E9323A
A1/A2
E9325A
A1/A2
E9325A
A1/A2
E9326A
A1/A2
E9326A
Part
NumberQtyDescription
E9321-600111SENSOR MODULE
E9321-690111RESTORED SENSOR
MODULE
E9322-600041SENSOR MODULE
E9322-690041RESTORED SENSOR
MODULE
E9323-600021SENSOR MODULE
E9323-690021RESTORED SENSOR
MODULE
E9325-600021SENSOR MODULE
E9325-690021RESTORED SENSOR
MODULE
E9326-600021SENSOR MODULE
E9326-690021RESTORED SENSOR
MODULE
A1/A2
E9327A
A1/A2
E9327A
E9327-600021SENSOR MODULE
E9327-690021RESTORED SENSOR
MODULE
E-Series E9320 Operating and Service Guide49
Page 50
Table 22 Replaceable Parts List
General Information
Reference
Designation
MP1 and MP2E9321-400012SHELL-PLASTIC
MP3 and MP4E9321-200022CHASSIS
MP8 and MP9E9321-000012SHIELD
MP26E9321-800011LABEL, ID E9321A
MP26E9322-800011LABEL, ID E9322A
MP26E9323-800011LABEL, ID E9323A
MP26E9325-800011LABEL, ID E9325A
MP26E9326-800011LABEL, ID E9326A
MP26E9327-800011LABEL, ID E9327A
MP27E9321-800022LABEL, POWER
MP30E9321-800031LABEL, CAL/ESD
Part
NumberQtyDescription
SENSOR
50E-Series E9320 Operating and Service Guide
Page 51
Service
Service
Service instructions consist of principles of operation,
troubleshooting, and repairs.
Principles of Operation
The Power Sensor ‘bulkhead’ assembly converts input RF to a
low frequency voltage signal representing the RF power
envelope. The input is AC coupled to a 3dB attenuator
followed by a 50 ohm load resistor. Two diodes are connected
to the load resistor, forming a pair of half-wave detectors with
opposite polarity and complementary voltage output. The
detected signal passes through a low-pass load filter. The
cutoff frequency of the filter is 300kHz, 1.5MHz, or 5MHz,
depending on the model/ bandwidth specification of the
sensor.
The detected signal can now follow one of two paths. The
average-only signal path is optimized for high sensitivity and
low drift at the expense of detector video bandwidth. This path
chops the signal to a carrier frequency around 440Hz to
remove sensitivity to DC offsets, then amplifies the AC signal.
Amplification and chopping parameters are much the same as
in previous Agilent diode sensors, with typical dynamic power
range of -65 to +20 dBm.
The chopper is a switch that connects the two balanced
signals to the two inputs of a differential amplifier. Thus, the
small DC signal from the detector is converted to an AC signal.
The output of the differential amplifier is connected to a
switched gain preamplifier.
E-Series E9320 Operating and Service Guide51
Page 52
RF IN
50ohm
Sensor Di ode
Bulkhead
Service
The dynamic range of the sensor is greater than 80dB in this
mode, so the sensor has two power ranges. On the high power
range the signal is attenuated before further amplification.
The bandwidth of the chopped signal is limited to less than
half the chop rate. So, this method cannot be used for wide
(~5 MHz) bandwidth measurements.
The normal path is used to detect the instantaneous power of
an RF signal and is optimized for a bandwidth of up to 5MHz.
The peak path trade off includes reduced dynamic range and
increased temperature sensitivity.
Average Only Path
Load Filter*
(300 kHz, 1.5 MHz
5 MHz low pass)
Chopp er
CW Differential
AMP
Switched
Gain
Preamp
Ave ra g e
Only Path
CHOP/CLOCK
CHOP/GAIN
Selecta ble Gain
Differential Amplifier*
(300 kHz, 1.5 MHz, 5 MHz)
PEAK AUTO-ZERO
Thermistor Bias
2
I
C Buffe r
Gain/Mode Control
Sensor ID
2
PROM
E
GAIN SELECT
SERIAL BUS
Normal
Path
* Bandwidth is sensor dependen t
Normal
Path
Figure 11 Simplified Sensor Block Diagram
The output of the load filter is connected to a gain selectable
amplifier with a bandwidth corresponding to the sensor
model/ bandwidth spec. The differential configuration
minimizes sensitivity to ground noise, dc offset and drift. In
normal mode, the amplifier provides maximum bandwidths of
300 kHz, 1.5 MHz or 5 MHz, allowing the user to match the
test signal’s modulation bandwidth to the sophisticated
instrument data processing. This permits the meter to
52E-Series E9320 Operating and Service Guide
Page 53
Service
measure burst average and peak power, to compute
peak-to-average ratios, and display other time-gated power
profiles on the power meter's large LCD screen.
The three dimensional Calibration data is stored in an
EEPROM on the sensor PCA. This data is unique to each
sensor and consists of frequency vs. input power vs.
temperature. Upon power-up, or when the sensor cable is
connected, these calibration factors are downloaded into the
EPM-P (E4416A/17A) series power meters. This means that
the operator is not required to enter any calibration
information when changing sensors, simply entering the
frequency of the input signal is all that is required
E-Series E9320 Operating and Service Guide53
Page 54
Service
Troubleshooting
Troubleshooting information is intended to first isolate the
power sensor, cable, or power meter as the defective
component. When the power sensor is isolated, an
appropriate Sensor Module must be used for repair. See
Table 22 on page 49.
If error message 241 or 310 is displayed on the power meter,
suspect a power sensor failure. Error 241 will occur if the
sensor is missing. An E9288 cable must be used to connect an
E-series 9320A sensor to an EPM-P Series power meter.
If no error message is displayed, but a problem occurs when
making a measurement, try replacing the cable from the
power meter to the power sensor. If the problem still exists, try
using a different power sensor to determine if the problem is
in the power meter or in the power sensor.
Electrostatic discharge will render the power sensor
inoperative. Do not, under any circumstances, open the power
sensor unless you and the power sensor are in a static free
environment.
Repair of Defective Sensor
There are no serviceable parts inside the E-series E9320
power sensors. If the sensor is defective, replace the entire
“module” with the appropriate “Restored Sensor Module”
listed in Table 22 on page 49.
54E-Series E9320 Operating and Service Guide
Page 55
Service
Disassembly Procedure
Disassemble the power sensor by performing the following
steps:
Disassemble the power sensor only in a static free
workstation. Electrostatic discharge renders the power sensor
inoperative.
Figure 12 Removing Power Sensor Shell
At the rear of the power sensor, insert the blade of a
screwdriver between the plastic shells (See Figure 12). To
prevent damage to the plastic shells use a screwdriver blade
as wide as the slot between the two shells.
Pry alternately at both sides of the connector J1 until the
plastic shells are apart. Remove the shells and the magnetic
shields.
Reassembly Procedure
Replace the magnetic shields and the plastic shells. Snap the
plastic shells together.
E-Series E9320 Operating and Service Guide55
Page 56
Service
Sales and Service Offices
In any correspondence or telephone conversations, refer to
the power sensor by its model number and full serial number.
With this information, the Agilent representative can quickly
determine whether your unit is still within its warranty period.
UNITED STATES Agilent Technologies
(tel) 1 800 829 4444
CANADAAgilent Technologies Canada Inc.
Tes t & Me as u r em en t
(tel) 1 877 894 4414
EUROPEAgilent Technologies
Tes t & Me as u r em en t
European Marketing Organization
(tel) (31 20) 547 2000
JAPANAgilent Technologies Japan Ltd.
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
LATIN AMERICA Agilent Technologies
Latin America Region Headquarters,USA
(tel) (305) 267 4245
(fax) (305) 267 4286
AUSTRALIA and
NEW ZEALAND
ASIA PACIFICAgilent Technologies, Hong Kong
Agilent Technologies Australia Pty Ltd.
(tel) 1-800 629 4852 (Australia)
(fax) (61 3) 9272 0749 (Australia)
(tel) 0-800 738 378 (New Zealand)
(fax) (64 4) 802 6881 (New Zealand)
(tel) (852) 3197 7777
(fax) (852) 2506 9284
56E-Series E9320 Operating and Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.