Agilent E9321A Operating Guide

Page 1
E-Series E9320 Peak
and Average Power
Sensors
Operating and Service Guide
Page 2
General
Information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpo se. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishings, performance, or use of this material. No part of this document may be ph otocopi ed, repr oduce d, or translated to another language without the prior written consent of Agilent Technologies.
Copyright 2003 Agilent Technologies
Station Road, South Queensferry, Scotland, EH30 9TG, UK.
Agilent Part No. E9321-90001 Printed in USA
June 2007
Legal Information
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measure ments are traceable to the United States National Inst itute of Sta ndards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Page 3
Safety Symbols
Safety Symbols
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
The Instruction Documentation Symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied documentation.
Safety
This guide uses warnings and cautions to denote hazards
Notices
WARNING A warning calls attention to a procedure, practice or the like,
which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning until the indicated conditions are fully understood and met.
CAUTION A caution calls attention to a procedure, practice or the like
which, if not correctly performed or adhered to, could result in damage to or the destruction of part or all of the equipment. Do not proceed beyond a caution until the indicated conditions are fully understood and met.
E-Series E9320 Operating and Service Guide 3
Page 4
General Safety Information
General Safety Information
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
WARNING BEFORE CONNECTING THE POWER SENSOR TO OTHER
INSTRUMENTS ensure that all instruments are connected to the protective (earth) ground. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in personal injury.
Sound Emission
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der Maschinenlarminformationsverordnung vom 18 Januar 1991.
• Sound Pressure LpA < 70 dB.
• Am Arbeitsplatz.
•Normaler Betrieb.
• Nach DIN 45635 T. 19 (Typprufung).
4 E-Series E9320 Operating and Service Guide
Page 5
General Safety Information
Manufacturers Declaration
This statement is provided to comply with the requirements of the German Sound DIN 45635 T. 19 (Typprufung).
• Sound Pressure LpA < 70 dB.
• At operator position.
•Normal operation.
• According to ISO 7779 (Type Test).
E-Series E9320 Operating and Service Guide 5
Page 6
General Safety Information
6 E-Series E9320 Operating and Service Guide
Page 7
Contents
Contents
Safety Symbols ........................................................................ 3
General Safety Information ..................................................... 4
Sound Emission.................................................................... 4
Contents .................................................................................. 7
Introduction...............................................................................9
General Information .............................................................. 10
Initial Inspection................................................................. 10
Power Meter and Sensor Cable Requirements....................11
Interconnections ................................................................ 11
Calibration.......................................................................... 11
The E-Series E9320 Power Sensors in Detail ........................ 13
Specifications and
Characteristics ......................................................................15
Introduction ........................................................................... 16
Specifications ....................................................................... 17
Service .......................................................................................43
General Information .............................................................. 44
Cleaning............................................................................. 44
Performance Test................................................................ 45
Replaceable Parts...............................................................47
Service .................................................................................. 51
Principles of Operation ....................................................... 51
Troubleshooting.................................................................. 54
Repair of Defective Sensor ................................................. 54
Disassembly Procedure...................................................... 55
Reassembly Procedure ....................................................... 55
E-Series E9320 Operating and Service Guide 7
Page 8
Contents
8 E-Series E9320 Operating and Service Guide
Page 9
1Introduction
What You’ll Find
In This Chapter
This Chapter introduces you to the E-series E9320 power sensors. It contains the following sections:
General Information on page 10
Initial Inspection on page 10
Power Meter and Sensor Cable Requirements on page 11
Interconnections on page 11
The E-Series E9320 Power Sensors in Detail on page 13
Page 10
General Information
General Information
Welcome to the E-series E9320 power sensors Operating and Service Guide! This guide contains information about the initial inspection, connection, and specifications of the E­series E9320 power sensors. You can also find a copy of this guide on the CD-ROM supplied with the EPM-P series peak and average power meters.
To make best use of your sensor refer to the chapter “Using E-Series E9320 Sensors” in the EPM-P Series Power Meters Operating and Service Guide.
Initial Inspection
Inspect the shipping container for damage. If the shipping container or packaging material is damaged, it should be kept until the contents of the shipment have been checked mechanically and electrically. If there is mechanical damage, notify the nearest Agilent office. Keep the damaged shipping materials (if any) for inspection by the carrier and an Agilent representative. If required, you can find a list of Agilent Sales and Service offices on page 56.
10 E-Series E9320 Operating and Service Guide
Page 11
General Information
Power Meter and Sensor Cable Requirements
The E-series E9320 power sensors are compatible ONLY with the EPM-P series power meters and with E9288 sensor cables. (The E9288 cables are color coded to help distinguish them from the 11730 series cables.)
Interconnections
Connect one end of an E9288 sensor cable to the E-series E9320 power sensor and connect the other end of the cable to the power meter’s channel input. Allow a few seconds for the power meter to download the data contained in the power sensor.
Ensure power sensors and cables are attached and removed in an indoor environment.
Calibration
To carry out a zero and calibration cycle as requested by the
power meter proceed as follows:
Ensure the E-series E9320 power sensor is disconnected
When the wait period is complete connect the E-series E9320 power sensor to the power meter’s POWER REF output.
Press (or , / ). The wait symbol is
from any signal source. On the power meter, press
Zero
, (or / ). During zeroing
Cal
the wait symbol is displayed.
again displayed during calibration.
Zero Zero A Zero B
Cal Cal Cal A Cal B
TIP You can reduce the steps required to carry out the zero and
calibration procedure as follows:
E-Series E9320 Operating and Service Guide 11
Page 12
General Information
Connect the power sensor to the POWER REF output.
Press and . (For dual channel
Zero Cal
meters, press , or
Zero + Cal B
Zero + Cal
Zero + Cal Zero + Cal A
as required.)
On completion, the power meter and sensor are ready to connect to the device under test (DUT).
WARNING BEFORE CONNECTING THE POWER SENSOR TO OTHER
INSTRUMENTS ensure that all instruments are connected to the protective (earth) ground. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in personal injury.
CAUTION The measurement connector (for connection to DUT) is
Type-N (male). A torque wrench should be used to tighten these connectors. Use a 3/4-inch open-end wrench and torque to 12 in-lb (135 Ncm) for the Type-N connector.
12 E-Series E9320 Operating and Service Guide
Page 13
The E-Series E9320 Power Sensors in Detail
The E-Series E9320 Power Sensors in Detail
The E-series E9320 power sensors have two frequency ranges. The E9325A, E9326A, and E9327A have a frequency range of 50 MHz to 18 GHz while the 50 MHz to 6 GHz range of the E9321A, E9322A, and E9323A covers most wireless communication applications.
The sensors have two independent measurement paths as shown in Figure 1.
Normal Path
Average Only Path
Chopper
Selectable Gain Differential Amplifier* (300 kHz, 1.5 MHz, 5 MHz)
PEAK AUTO-ZERO
Thermistor Bias
2
C Buffer
I Gain/Mode Control Sensor ID
2
E
PROM
CW Differential AMP
Switched Gain Preamp
GAIN SELECT
SERIAL BUS
Average Only Path
CHOP/CLOCK
CHOP/GAIN
Normal Path
RF IN 50ohm
Sensor D iode Bulkhead
* Bandwi dth is sens or dependent
Load Filter* (300 kHz, 1.5 MHz 5 MHz low pass)
Figure 1 Simplified Sensor Block Diagram
Use the default normal path for continuously sampled measurements of modulated signals and time gated measurements. For each frequency range there is a choice of sensors with three video (modulation) bandwidths.
E-Series E9320 Operating and Service Guide 13
Page 14
The E-Series E9320 Power Sensors in Detail
E9321A and E9325A sensors with 300 kHz bandwidth are suitable for measuring TDMA signals such as GSM.
E9322A and E9326A sensors with 1.5 MHz bandwidth are suitable for measuring IS-95 CDMA signals.
E9323A and E9327A sensors with 5 MHz bandwidth are suitable for measuring W-CDMA signals.
Note however, that the sensors with widest bandwidth have the smallest dynamic range (in normal mode). If dynamic range is an important factor, use the sensor model with just enough video bandwidth for the measurement you want to make.
The average-only path is suitable for average power measurements of Continuous Wave (CW) and constant amplitude signals between -65 dBm (sensor dependent) and +20 dBm. The average-only path can also be used to measure true average power of any complex modulated signal below ­20 dBm.
Calibration factors, linearity, temperature, and bandwidth compensation data are stored in the sensor EEPROM during the manufacturing process. All the compensation data is downloaded to the EPM-P series power meter at power-on or when the sensor is connected. You only need to enter the frequency of the RF signal you are measuring to achieve a high degree of accuracy.
14 E-Series E9320 Operating and Service Guide
Page 15
2 Specifications and
Characteristics
What You’ll Find
In This Chapter
This Chapter describes the specifications and characteristics of the E-series E9320 power sensors. It contains the following sections:
“Introduction” on page 16
“Specifications” on page 17
Page 16
Introduction
Introduction
E-series E9320 power sensors are designed for use with the
Agilent EPM series power meters. The E-series E9320 power
sensors have two measurement paths:
Normal path: (default mode) for peak, average and
time-related measurements.
Average-only path: is designed primarily for average power measurements on low level signals.
These specifications are valid ONLY after zero and calibration of the power meter and sensor.
Supplemental characteristics, which are shown in italics, are intended to provide information useful in applying the power sensors by giving typical, but nonwarranted performance parameters. These characteristics are shown in italics or denoted as typical”, “nominal or approximate”.
16 E-Series E9320 Operating and Service Guide
Page 17
Specifications
Specifications
Frequency, Bandwidth, and Power Range
Table 1 Frequency, Bandwidth, and Power Range
Maximum
Power Range
Video
Sensor
E9321A 300 kHz 50 MHz
E9325A 300 kHz 50 MHz
E9322A 1.5 MHz 50 MHz
E9326A 1.5 MHz 50 MHz
E9323A 5 MHz 50 MHz
E9327A 5 MHz 50 MHz
Bandwidth
* For average power measurements, Free Run acquisition mode.
Frequency Range
to
6GHz
to
18 GHz
to
6GHz
to
18 GHz
to
6GHz
to
18 GHz
Average only Mode
–65 dBm
to
+20 dBm
–65 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
–60 dBm
to
+20 dBm
Normal Mode*
–50 dBm
to
+20 dBm
–50 dBm
to
+20 dBm
–45 dBm
to
+20 dBm
–45 dBm
to
+20 dBm
–40 dBm
to
+20 dBm
–40 dBm
to
+20 dBm
E-Series E9320 Operating and Service Guide 17
Page 18
Maximum Power, RF Connector
Table 2 Maximum Power, RF Connector
Specifications
Sensor
E9321A E9322A E9323A E9325A E9326A E9327A
RF Connector
N-Type (m) +23 dBm average +30 dBm,
Maximum Average Power
Maximum Peak Power
<10µs duration
18 E-Series E9320 Operating and Service Guide
Page 19
Specifications
Measurement Ranges
The E-series E9320 power sensors have two measurement ranges (Lower and Upper) as shown in Table 3, Table 4, and Ta b le 5 .
Table 3 Lower and Upper Measurement Ranges
E9321A and E9325A
Lower Range (Min. Power)
Normal mode
50 dBm 65 dBm
Average only mode
Lower Range (Max. Power) Lower to Upper Auto Range
+0.5 dBm
Point
Upper to Lower Auto Range Point
Upper Range (Min. Power)
Upper Range (Max. Power
1
Applies to CW and constant amplitude signals only
-9.5dBm -18.5 dBm
35 dBm 50 dBm
+20 dBm
above 20 dBm.
17.5 dBm
+20 dBm
1
1
E-Series E9320 Operating and Service Guide 19
Page 20
Specifications
Table 4 Lower and Upper Measurement Ranges
E9322A and E9326A
Lower Range (Min. Power)
Normal mode
45 dBm 60 dBm
Average only mode
Lower Range (Max. Power) Lower to Upper Auto Range
5 dBm
13.5 dBm
Point
Upper to Lower Auto Range Point
15 dBm 14.5 dBm
Upper Range (Min. Power) 35 dBm 45 dBm
Upper Range (Max. Power
1
Applies to CW and constant amplitude signals only
+20 dBm
+20 dBm
above 20 dBm.
1
1
20 E-Series E9320 Operating and Service Guide
Page 21
Specifications
Table 5 Lower and Upper Measurement Ranges
E9323A and E9327A
Lower Range (Min. Power)
Lower Range (Max. Power)
Normal mode
40 dBm 60 dBm
5 dBm
Average only mode
Lower to Upper Auto Range Point
Upper to Lower Auto Range
15 dBm 11.5 dBm
Point
Upper Range (Min. Power)
30 dBm 35 dBm
Upper Range (Max. Power +20 dBm
1
Applies to CW and constant amplitude signals only above 20 dBm.
10.5 dBm
20 dBm
+
1
1
E-Series E9320 Operating and Service Guide 21
Page 22
Power Sensor Maximum SWR
Table 6 Power Sensor Maximum SWR
Sensor Maximum SWR ≤0dBm
Specifications
1.12
1.16
1.23
1.28
1.12
1.18
1.21
1.27
1.14
1.22
1.26
16GHz
17GHz
SWR
1.300
1.250
1.200
1.150
1.100
1.050
1.000
0.02GHz
E9321A E9325A
E9322A E9326A
E9323A E9327A
0.05GHz
50 MHz to 2 GHz:
2GHz to 10GHz: 10 GHz to 16 GHz: 16 GHz to 18 GHz:
50 MHz to 2 GHz:
2GHz to 12GHz: 12 GHz to 16 GHz: 16 GHz to 18 GHz:
50 MHz to 2 GHz:
2 GHz to 16 GHz: 16 GHz to 18 GHz:
1GHz
6GHz
9GHz
12GHz
14GHz
0.3GHz
0.5GHz
15GHz
Figure 2 Typicl SWR (50 MHz to 18 GHz) for the E9321A and E9325A
sensors at various power levels
-30
-20
-10
0
10
20
Freq
18GHz
22 E-Series E9320 Operating and Service Guide
Page 23
Specifications
1.300
SWR
1.250
1.200
1.150
1.100
1.050
1.000
1GHz
6GHz
9GHz
12GHz
14GHz
15GHz
16GHz
0.3GHz
0.02GHz
0.05GHz
0.5GHz
17GHz
Figure 3 Typical SWR (50 MHz to 18 GHz) for the E9322A and E9326A
sensors at various power levels
1.300
SWR
1.250
1.200
1.150
1.100
1.050
1.000
1GHz
6GHz
9GHz
12GHz
14GHz
15GHz
16GHz
0.3GHz
0.02GHz
0.5GHz
0.05GHz
17GHz
-30
-20
-10
0
10
20
Freq
18GHz
-30
-20
-10
0
10
20
Freq
18GHz
Figure 4 Typical SWR (50 MHz to 18 GHz) for the E9323A and E9327A
sensors at various power levels
E-Series E9320 Operating and Service Guide 23
Page 24
Sensor Linearity
Table 7 Power Sensor Linearity Normal Mode
(upper and lower range)
Specifications
Sensor
E9321A E9325A
E9322A E9326A
E9323A E9327A
Temperature (25 ±10°C)
±4.2% ±5.0%
±4.2% ±5.0%
±4.2% ±5.5%
Te mp e ra t u r e (0 to 55°C)
Table 8 Power Sensor Linearity Average-Only Mode
(upper and lower range)
Sensor
E9321A E9325A
E9322A E9326A
E9323A E9327A
Te mp e ra t u r e (25 ±10°C)
±3.7% ±4.5%
±3.7% ±4.5%
±3.7% ±5.0%
Temperature (0 to 55°C)
If the temperature changes after calibration and the meter and sensor are NOT recalibrated, the following additional linearity errors should be added to the Power Linearity figures in Table 7 and Table 8.
24 E-Series E9320 Operating and Service Guide
Page 25
Specifications
Table 9 Additional Linearity Error (normal and average-
only modes)
%
-0.50
-1.00
-1.50
-2.00
2.00
1.50
1.00
0.50
0.00
Temperature
Sensor
E9321A E9325A
E9322A E9326A
E9323A E9327A
1-7-3026
-30
-27-23
-19-15
1
-
(25 ±10°C)
±1.0% ±1.0%
±1.0% ±1.5%
±1.0% ±2.0%
4
18
10
1
dBm
20
Temperature (0 to 55°C)
Figure 5 Typical Power Linearity at 25°C for E9323A and E9327A
5 MHz bandwidth sensors after zero and calibration, with associated measurement uncertainty
Aver age Only
Normal (lower range)
Norm al (upper range)
E-Series E9320 Operating and Service Guide 25
Page 26
+20 dBm
0dBm
Specifications
Power Range Measurement Uncertainty
30 to 20 dBm ±0.9%
20 to 10 dBm ±0.8%
10 to 0 dBm ±0.65%
0 to +10 dBm ±0.55%
+10 to +20 dBm ±0.45%
±
±
4.0%
2.0%
±
2.0%
±
4.0%
60 dBm
65 dBm
65 dBm
60 dBm
0 dBm +20 dBm
Figure 6 Relative Mode Power Measurement Linearity with an
EPM-P series power meter, at 25°C (typical)
26 E-Series E9320 Operating and Service Guide
Page 27
Specifications
Figure 6 shows the typical uncertainty in making a relative power measurement, using the same power meter channel
and the same power sensor to obtain the reference and measured values. It also assumes that negligible change in
frequency and mismatch error occurs when transitioning from the power level used as the reference to the power level
measured.
E-Series E9320 Operating and Service Guide 27
Page 28
Peak Flatness
The peak flatness is the flatness of a Peak-to-Average ratio measurement for various tone separations for an equal magnitude two-tone RF input. Figure 7, Figure 8, and Figure 9 refer to the relative error in Peak-to-Average measurement as the tone separation is varied. The measurements were performed at 10 dBm average power using an E9288A sensor cable.
Error (dB)
0.5
0.0
-0.5
-1.0
-1.5
-2.0
-2.5
-3.0
-3.5
0.01.02.03.04.05.06.0
Input Tone Seperation Frequency (MHz)
Specifications
HIGH LOW MED OFF
Figure 7 E9321A and E9325A Error in Peak-to-Average
Measurements for a Two-tone Input (High, Medium, Low, and Off Filters)
28 E-Series E9320 Operating and Service Guide
Page 29
Specifications
Error (dB)
0.5
0
-0.5
-1
-1.5
-2
-2.5
-3
-3.5
0.0 1.0 2.0 3.0 4.0 5.0 6.0
Input Tone Seperation Frequency (MHz)
Figure 8 Filter responses for the E9322A and E9326A power sensors
(High, Medium, Low, and Off)
Error (dB)
0.5
0.0
-0.5
-1.0
-1.5
-2.0
-2.5
-3.0
-3.5
0.0 1.0 2.0 3.0 4.0 5.0 6.0
Input Tone Seperation Frequency (MHz)
HIGH LOW MED OFF
LOW
MED
OFF
HIGH
Figure 9 Filter responses for the E9323A and E9327A power sensors
(High, Medium, Low, and Off)
E-Series E9320 Operating and Service Guide 29
Page 30
Specifications
Calibration Factor (CF) and Reflection Coefficient (Rho)
Calibration Factor and Reflection Coefficient data are provided at frequency intervals on a data sheet included with the power sensor. This data is unique to each sensor. If you have more than one sensor, match the serial number on the data sheet with the serial number of the sensor you are using. The CF corrects for the frequency response of the sensor. The EPM-P Series power meter automatically reads the CF data stored in the sensor and uses it to make corrections.
For power levels greater than 0 dBm, add to the calibration factor uncertainty specification:
±0.1% per dB for E9321A and E9325A power sensors
±0.15% per dB for E9322A and E9326A power sensors
±0.2% per dB for E9323A and E9327A power sensors
Reflection coefficient (Rho) relates to the SWR according to the formula:
SWR=(1+Rho)/(1Rho)
Maximum uncertainties of the CF data are listed in Table 10. The uncertainty analysis for the calibration of the sensors was calculated in accordance with the ISO Guide. The uncertainty data, reported on the calibration certificate, is the expanded uncertainty with a 95% confidence level and a coverage factor of 2.
30 E-Series E9320 Operating and Service Guide
Page 31
Specifications
Table 10 Cal Factor Uncertainty at 0.1 mW (-10dBm
Frequency
50 MHz
100 MHz
300 MHz
500 MHz
800 MHz
1.0 GHz
1.2 GHz
1.5 GHz
2.0 GHz
3.0 GHz
4.0 GHz
5.0 GHz
6.0 GHz
7.0 GHz
8.0 GHz
9.0 GHz
11.0 GHz
12.0 GHz
13.0 GHz
14.0 GHz
15.0 GHz
16.0 GHz
17.0 GHz
18.0 GHz
Uncertainty (%) (25 ±10°C)
Reference Reference
±1.8 ±2.0
±1.8 ±2.0
±1.8 ±2.0
±1.8 ±2.0
±2.1 ±2.3
±2.1 ±2.3
±2.1 ±2.3
±2.1 ±2.3
±2.1 ±2.3
±2.1 ±2.3
±2.1 ±2.3
±2.1 ±2.3
±2.3 ±2.5
±2.3 ±2.5
±2.3 ±2.5
±2.3 ±2.5
±2.3 ±2.5
±2.3 ±2.5
±2.5 ±2.8
±2.5 ±2.8
±2.5 ±2.8
±2.5 ±2.8
±2.5 ±2.8
Uncertainty(%) (0 to 55°C)
)
E-Series E9320 Operating and Service Guide 31
Page 32
Specifications
Zero Set
This specification applies to a ZERO performed when the sensor input is not connected to the POWER REFERENCE.
Ta b le 1 1 Z er o S et
Sensor
E9321A E9325A
E9322A E9326A
E9323A E9327A
Zero Set (Normal mode)
5 nW 0.17 nW
19 nW 0.5 nW
60 nW 0.6 nW
Zero Set (Average-only Mode)
32 E-Series E9320 Operating and Service Guide
Page 33
Specifications
Zero Drift
Sensor
E9321A E9325A
E9322A E9326A
E9323A E9327A
Zero Drift and Measurement Noise
Table 12 Zero Drift and Measurement Noise
Normal
Mode
1
Average only
Mode
Measurement Noise
Normal
3
Mode
5 nW 60 pW <6 nW <75 nW <165 pW
5 nW 100 pW <12 nW <180 nW <330 pW
40 nW 100 pW <25 nW <550 nW <400 pW
1
Within 1 hour after zero set, at a constant temperature, after a
24 hour warm-up of the power meter.
2
Measured over a one minute interval, at a constant temperature,
two standard deviations, with averaging set to 1 (normal mode), 16 (for average-only mode, normal speed) and 32 (average-only mode, x2 speed).
3
In Free run acquisition mode.
4
Noise per sample, Video bandwidth set to OFF with no averaging
(i.e., averaging set to 1) - see Effect of Video Bandwidth Settings and Table 14.
2
Normal
Mode
Average
4
only
Mode
E-Series E9320 Operating and Service Guide 33
Page 34
Specifications
Effect of Averaging on Noise: Averaging over 1 to 1024 readings is available for reducing noise. Table 12 provides the
Measurement Noise for a particular sensor. Use the Noise Multipliers in Table 13 for the appropriate speed (normal or
x2), or measurement mode (normal and average-only), and the number of averages to determine the total Measurement
Noise value.
In addition, for x2 speed (in normal mode) the total Measurement Noise should be multiplied by 1.2, and for fast
speed (in normal mode) the multiplier is 3.4. Note that in fast speed no additional averaging is implemented.
34 E-Series E9320 Operating and Service Guide
Page 35
Specifications
Table 13 Noise Multipliers
Noise Multiplier
Average-only Normal
Number of Averages
Normal
Speed
X 2
Speed
Normal Speed,
Free run acquisition
1 5.5 6.5 1.0
2 3.89 4.6 0.94
4 2.75 3.25 0.88
8 1.94 2.3 0.82
16 1.0 1.63 0.76
32 0.85 1.0 0.70
64 0.61 0.72 0.64
128 0.49 0.57 0.58
256 0.34 0.41 0.52
512 0.24 0.29 0.46
1024 0.17 0.2 0.40
Example: E9321A power sensor, Number of Averages = 4,
Free Run acquisition, normal mode, x2 speed.
Measurement Noise calculation:
(<6 nW x 0.88 x 1.2) = <6.34 nW
E-Series E9320 Operating and Service Guide 35
Page 36
Specifications
Effect of Video Bandwidth Setting: The noise per sample is reduced by applying the meter video bandwidth reduction filter setting (High, Medium, or Low). If averaging is implemented, this will dominate any effect of changing the video bandwidth.
Table 14 Effect of Video Bandwidth on Noise per Sample
Noise Multipliers
Sensor
E9321A E9325A
E9322A E9326A
E9323A E9327A
Low Medium High
0.32 0.5 0.63
0.50 0.63 0.80
0.40 0.63 1.0
Example: E9322A power sensor, triggered acquisition, high
video bandwidth.
Noise per sample calculation:
(<180nW x 0.80) = <144 nW
Effect of Time-Gating on Measurement Noise: The measurement noise will depend on the time gate length over which measurements are made. Effectively 20 averages are carried out every 1 µs of gate length.
36 E-Series E9320 Operating and Service Guide
Page 37
Specifications
Settling Times
Average Only Mode:
In normal and x2 speed mode, manual filter, 10 dB decreasing
power step, refer to Table 15.
Table 15 Averages vs. Settling Time (Average-only mode)
Number of Averages
1024 57 27
In Fast mode, within the range 50 to +20 dBm, for a 10 dB
decreasing power step, the settling time is: E4416A: 10 ms
E4417A: 20 ms
When a power step crosses the power sensors auto-range switch point, add 25 ms.
Normal Speed Mode X 2 Speed Mode
1 0.08 0.07
2 0.13 0.09
4 0.24 0.15
8 0.45 0.24
16 1.1 0.45
32 1.9 1.1
64 3.5 1.9
128 6.7 3.5
256 14 6.7
512 27 14
Settling Time (s)
E-Series E9320 Operating and Service Guide 37
Page 38
Specifications
Normal Mode:
In Normal, free run acquisition mode, within the range 20 to +20 dBm, for a 10 dB decreasing power step, the settling time
is dominated by the measurement update rate and is listed in Table 16 for various filter settings.
Table 16 Settling Time vs. Averages
Settling Time
Number of Averages
Free Run Acquisition Mode
Normal Speed X 2 Speed
1 0.1 sec 0.08 sec
2 0.15 sec 0.1 sec
4 0.25 sec 0.15 sec
8 0.45 sec 0.25 sec
16 0.9 sec 0.45 sec
32 1.7 sec 0.9 sec
64 3.3 sec 1.7 sec
128 6.5 sec 3.3 sec
256 13.0 sec 6.5 sec
512 25.8 sec 13.0 sec
1024 51.5 sec 25.8 sec
38 E-Series E9320 Operating and Service Guide
Page 39
Specifications
In normal mode, measuring in continuous or single acquisition mode, the performance of rise times, fall times and 99% settled results are shown in Table 17. Rise and fall time specifications are for a 0.0 dBm pulse, with the rise time and fall time measured between 10% to 90% points and upper range selected.
Table 17 Rise and Fall Times vs. Sensor Bandwidth
1
Video Bandwidth Setting
Sensor
E9321A E9325A
E9322A E9326A
E9323A E9327A
1
Rise and fall time specifications are only valid when used
Parameter Low Medium High Off
Rise time < µs 2.6 1.5 0.9 0.3
Fall time
< µs 2.7 1.5 0.9 0.5
Settling time (rising)
< µs
Settling time
< µs
(falling)
Rise time < µs 1.5 0.9 0.4 0.2
Fall time
< µs 1.5 0.9 0.4 0.3
Settling time
< µs
(rising)
Settling time
< µs
(falling)
Rise time < µs 0.9 0.4 0.2 0.2
Fall time
< µs 0.9 0.4 0.2 0.2
Settling time (rising)
< µs
Settling time
< µs
(falling)
5.1 5.1 4.5 0.6
5.1 5.1 4.5 0.9
5.3 4.5 3.5 0.5
5.3 4.5 3.5 0.9
4.5 3.5 1.5 0.4
4.5 3.5 2 0.4
with the E9288A sensor cable (1.5 metres).
E-Series E9320 Operating and Service Guide 39
Page 40
Specifications
Overshoot in response to power steps with fast rise times, i.e., less than the sensor rise time is less than 10%. When a power step crosses the power sensors auto-range switch point, add 10 µs.
40 E-Series E9320 Operating and Service Guide
Page 41
Specifications
Physical Specifications
Table 18 Physical Specifications
Physical Characteristics
Net Weight
Shipping Weight
Dimensions
0.2 kg (0.45 lb)
0.55 kg (1.2 lb)
Length: 150 mm (5.9 in) Width: 38 mm (1.5 in) Height: 30 mm (1.2 in)
E-Series E9320 Operating and Service Guide 41
Page 42
Specifications
42 E-Series E9320 Operating and Service Guide
Page 43
3Service
What Youll Find
In This Chapter
This Chapter introduces you to the E-series E9320 power sensors. It contains the following sections:
General Information on page 44
Service on page 51
Sales and Service Offices on page 56
Page 44
General Information
General Information
This chapter contains information about general maintenance, performance tests, troubleshooting and repair of E-series E9320 power sensors.
Cleaning
Use a clean, damp cloth to clean the body of the E-series E9320 power sensor.
Connector Cleaning
The RF connector beads deteriorate when contacted by hydrocarbon compounds such as acetone, trichloroethylene, carbon tetrachloride, and benzene.
Clean the connector only at a static free workstation. Electrostatic discharge to the center pin of the connector will render the power sensor inoperative.
Keeping in mind its flammable nature; a solution of pure isopropyl or ethyl alcohol can be used to clean the connector.
Clean the connector face using a cotton swab dipped in isopropyl alcohol. If the swab is too big use a round wooden toothpick wrapped in a lint free cotton cloth dipped in isopropyl alcohol.
44 E-Series E9320 Operating and Service Guide
Page 45
General Information
Performance Test
Standing Wave Ratio (SWR) and Reflection Coefficient (Rho) Performance Test
This section does not establish preset SWR test procedures since there are several test methods and different equipment available for testing the SWR or reflection coefficient. Therefore, the actual accuracy of the test equipment must be accounted for when measuring against instrument specifications to determine a pass or fail condition. The test system used must not exceed the system Rho uncertainties shown in Table 19 when testing the E9325A, E9326A, and E9327A. Use Table 20 when testing the E9321A, E9322A, and E9323A.
Table 19 Reflection Coefficient for the E9321A and E9325A
Frequency
50 MHz to 2 GHz
2 GHz to 10 GHz
10 GHz to 16 GHz ±0.010 0.103
16 GHz to 18 GHz ±0.010 0.123
E-Series E9320 Operating and Service Guide 45
System Rho Uncertainty
±0.010 0.057
±0.010 0.074
Measured
value
Maximum
Rho
Page 46
General Information
Table 20 Reflection Coefficient for the E9322A and E9326A
System Rho
Frequency
50 MHz to 2 GHz ±0.010 0.057
2 GHz to 10 GHz
10 GHz to 16GHz ±0.010 0.095
16 GHz to 18GHz ±0.010 0.119
Uncertainty
±0.010 0.083
Measured
value
Maximum
Table 21 Reflection Coefficient for the E9323A and E9327A
Frequency
50 MHz to 2 GHz ±0.010 0.065
2 GHz to 16GHz ±0.010 0.099
16 GHz to 18GHz ±0.010 0.115
System Rho Uncertainty
Measured
value
Maximum
Rho
Rho
46 E-Series E9320 Operating and Service Guide
Page 47
General Information
Replaceable Parts
Table 22 is a list of replaceable parts. Figure 10 is the illustrated parts breakdown (IPB) that identifies all of the replaceable parts. To order a part, quote the Agilent part number, specify the quantity required, and address the order to the nearest Agilent office.
NOTE Within the USA, it is better to order directly from the Agilent
Parts Center in Roseville, California. Ask your nearest Agilent office for information and forms for the Direct Mail Order System. Also your nearest Agilent office can supply toll free telephone numbers for ordering parts and supplies
E-Series E9320 Operating and Service Guide 47
Page 48
General Information
.
Figure 10 Illustrated Parts Break down
48 E-Series E9320 Operating and Service Guide
Page 49
General Information
Table 22 Replaceable Parts List
Reference Designation
A1/A2 E9321A
A1/A2 E9321A
A1/A2 E9322A
A1/A2 E9322A
A1/A2 E9323A
A1/A2 E9323A
A1/A2 E9325A
A1/A2 E9325A
A1/A2 E9326A
A1/A2 E9326A
Part Number Qty Description
E9321-60011 1 SENSOR MODULE
E9321-69011 1 RESTORED SENSOR
MODULE
E9322-60004 1 SENSOR MODULE
E9322-69004 1 RESTORED SENSOR
MODULE
E9323-60002 1 SENSOR MODULE
E9323-69002 1 RESTORED SENSOR
MODULE
E9325-60002 1 SENSOR MODULE
E9325-69002 1 RESTORED SENSOR
MODULE
E9326-60002 1 SENSOR MODULE
E9326-69002 1 RESTORED SENSOR
MODULE
A1/A2 E9327A
A1/A2 E9327A
E9327-60002 1 SENSOR MODULE
E9327-69002 1 RESTORED SENSOR
MODULE
E-Series E9320 Operating and Service Guide 49
Page 50
Table 22 Replaceable Parts List
General Information
Reference Designation
MP1 and MP2 E9321-40001 2 SHELL-PLASTIC
MP3 and MP4 E9321-20002 2 CHASSIS
MP8 and MP9 E9321-00001 2 SHIELD
MP26 E9321-80001 1 LABEL, ID E9321A
MP26 E9322-80001 1 LABEL, ID E9322A
MP26 E9323-80001 1 LABEL, ID E9323A
MP26 E9325-80001 1 LABEL, ID E9325A
MP26 E9326-80001 1 LABEL, ID E9326A
MP26 E9327-80001 1 LABEL, ID E9327A
MP27 E9321-80002 2 LABEL, POWER
MP30 E9321-80003 1 LABEL, CAL/ESD
Part Number Qty Description
SENSOR
50 E-Series E9320 Operating and Service Guide
Page 51
Service
Service
Service instructions consist of principles of operation, troubleshooting, and repairs.
Principles of Operation
The Power Sensor ‘bulkhead assembly converts input RF to a low frequency voltage signal representing the RF power envelope. The input is AC coupled to a 3dB attenuator followed by a 50 ohm load resistor. Two diodes are connected to the load resistor, forming a pair of half-wave detectors with opposite polarity and complementary voltage output. The detected signal passes through a low-pass load filter. The cutoff frequency of the filter is 300kHz, 1.5MHz, or 5MHz, depending on the model/ bandwidth specification of the sensor.
The detected signal can now follow one of two paths. The average-only signal path is optimized for high sensitivity and low drift at the expense of detector video bandwidth. This path chops the signal to a carrier frequency around 440Hz to remove sensitivity to DC offsets, then amplifies the AC signal. Amplification and chopping parameters are much the same as in previous Agilent diode sensors, with typical dynamic power range of -65 to +20 dBm.
The chopper is a switch that connects the two balanced signals to the two inputs of a differential amplifier. Thus, the small DC signal from the detector is converted to an AC signal. The output of the differential amplifier is connected to a switched gain preamplifier.
E-Series E9320 Operating and Service Guide 51
Page 52
RF IN 50ohm
Sensor Di ode Bulkhead
Service
The dynamic range of the sensor is greater than 80dB in this mode, so the sensor has two power ranges. On the high power range the signal is attenuated before further amplification. The bandwidth of the chopped signal is limited to less than half the chop rate. So, this method cannot be used for wide (~5 MHz) bandwidth measurements.
The normal path is used to detect the instantaneous power of an RF signal and is optimized for a bandwidth of up to 5MHz. The peak path trade off includes reduced dynamic range and increased temperature sensitivity.
Average Only Path
Load Filter* (300 kHz, 1.5 MHz 5 MHz low pass)
Chopp er
CW Differential AMP
Switched Gain Preamp
Ave ra g e Only Path
CHOP/CLOCK
CHOP/GAIN
Selecta ble Gain Differential Amplifier* (300 kHz, 1.5 MHz, 5 MHz)
PEAK AUTO-ZERO
Thermistor Bias
2
I
C Buffe r Gain/Mode Control Sensor ID
2
PROM
E
GAIN SELECT
SERIAL BUS
Normal Path
* Bandwidth is sensor dependen t
Normal Path
Figure 11 Simplified Sensor Block Diagram
The output of the load filter is connected to a gain selectable amplifier with a bandwidth corresponding to the sensor model/ bandwidth spec. The differential configuration minimizes sensitivity to ground noise, dc offset and drift. In normal mode, the amplifier provides maximum bandwidths of 300 kHz, 1.5 MHz or 5 MHz, allowing the user to match the test signals modulation bandwidth to the sophisticated instrument data processing. This permits the meter to
52 E-Series E9320 Operating and Service Guide
Page 53
Service
measure burst average and peak power, to compute peak-to-average ratios, and display other time-gated power profiles on the power meter's large LCD screen.
The three dimensional Calibration data is stored in an EEPROM on the sensor PCA. This data is unique to each sensor and consists of frequency vs. input power vs. temperature. Upon power-up, or when the sensor cable is connected, these calibration factors are downloaded into the EPM-P (E4416A/17A) series power meters. This means that the operator is not required to enter any calibration information when changing sensors, simply entering the frequency of the input signal is all that is required
E-Series E9320 Operating and Service Guide 53
Page 54
Service
Troubleshooting
Troubleshooting information is intended to first isolate the power sensor, cable, or power meter as the defective component. When the power sensor is isolated, an appropriate Sensor Module must be used for repair. See Table 22 on page 49.
If error message 241 or 310 is displayed on the power meter, suspect a power sensor failure. Error 241 will occur if the sensor is missing. An E9288 cable must be used to connect an E-series 9320A sensor to an EPM-P Series power meter.
If no error message is displayed, but a problem occurs when making a measurement, try replacing the cable from the power meter to the power sensor. If the problem still exists, try using a different power sensor to determine if the problem is in the power meter or in the power sensor.
Electrostatic discharge will render the power sensor inoperative. Do not, under any circumstances, open the power sensor unless you and the power sensor are in a static free environment.
Repair of Defective Sensor
There are no serviceable parts inside the E-series E9320 power sensors. If the sensor is defective, replace the entire module with the appropriate Restored Sensor Module listed in Table 22 on page 49.
54 E-Series E9320 Operating and Service Guide
Page 55
Service
Disassembly Procedure
Disassemble the power sensor by performing the following steps:
Disassemble the power sensor only in a static free workstation. Electrostatic discharge renders the power sensor inoperative.
Figure 12 Removing Power Sensor Shell
At the rear of the power sensor, insert the blade of a screwdriver between the plastic shells (See Figure 12). To prevent damage to the plastic shells use a screwdriver blade as wide as the slot between the two shells.
Pry alternately at both sides of the connector J1 until the plastic shells are apart. Remove the shells and the magnetic
shields.
Reassembly Procedure
Replace the magnetic shields and the plastic shells. Snap the plastic shells together.
E-Series E9320 Operating and Service Guide 55
Page 56
Service
Sales and Service Offices
In any correspondence or telephone conversations, refer to the power sensor by its model number and full serial number. With this information, the Agilent representative can quickly determine whether your unit is still within its warranty period.
UNITED STATES Agilent Technologies
(tel) 1 800 829 4444
CANADA Agilent Technologies Canada Inc.
Tes t & Me as u r em en t (tel) 1 877 894 4414
EUROPE Agilent Technologies
Tes t & Me as u r em en t European Marketing Organization (tel) (31 20) 547 2000
JAPAN Agilent Technologies Japan Ltd.
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
LATIN AMERICA Agilent Technologies
Latin America Region Headquarters,USA (tel) (305) 267 4245 (fax) (305) 267 4286
AUSTRALIA and NEW ZEALAND
ASIA PACIFIC Agilent Technologies, Hong Kong
Agilent Technologies Australia Pty Ltd. (tel) 1-800 629 4852 (Australia) (fax) (61 3) 9272 0749 (Australia) (tel) 0-800 738 378 (New Zealand) (fax) (64 4) 802 6881 (New Zealand)
(tel) (852) 3197 7777 (fax) (852) 2506 9284
56 E-Series E9320 Operating and Service Guide
Loading...