THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER,
TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LA W , AGILENT DISCLAIMS ALL
WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.
SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT
WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL
CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or
subcontract, Software is delivered and licensed as “Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or
any equivalent agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard commercial license terms, and
non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users
will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Certification
Agilent Technologies, Inc. certifies that this product met its published specifications at the
time of shipment from the factory. Agilent Technologies, Inc. further certifies that its
calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institute's calibration facility, and
to the calibration facilities of other International Standards Organization members.
ii
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies, Inc. products. For information about these agreements and for
other assistance, contact Agilent. Refer to “Contacting Agilent Technologies” on page 2-2.
Technology Licenses
The hardware and / or software described in this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Trademarks
Microsoft is a U.S. registered trademark of Microsoft Corporation. Windows and MS
Windows are U.S. registered trademarks of Microsoft Corporation.
Safety Notes
The following safety notes are used throughout this manual. Familiarize yourself with
each of the notes and its meaning before operating this instrument. All pertinent safety
notes for using this product are located in Chapter 3 , “Safety and Regulatory
Information.”
WARNINGWarning denotes a hazard. It calls attention to a procedure which, if
not correctly performed or adhered to, could result in injury or loss
of life. Do not proceed beyond a warning note until the indicated
conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or destruction of
the instrument. Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met.
iii
Documentation Map
The online Help files are embedded in the analyzer, offering quick
reference to programming and user documentation. From the Help
drop-down menu, you can access the Help system in six different
languages. Also, you can view the Analyzer Product Overview
multimedia presentation and access the analyzer’s Web page.
The Installation and Quick Start Guide helps to quickly familiarize
yourself with the analyzer. Procedures are provided for installing,
configuring, and verifying the operation of the analyzer.
1. Unpack the contents of the shipping container. (Keep the packaging material in case the
analyzer needs to be returned to the factory.)
WARNINGThe network analyzer is heavy. Use proper lifting techniques.
2. Carefully inspect the analyzer to make sure that it was not damaged during shipment.
NOTEIf your analyzer was damaged during shipment, contact Agilent
Technologies. A list of contacts is provided on page 2-2.
1-1
Installing Your Analyzer
STEP 1. Check the Shipment
3. Locate the Contents List in the shipping container and verify the completeness of your
shipment. Although the Contents List is the most accurate, you can also use the following
illustration to verify the items in your shipment. If your shipment is not complete, contact
Agilent Technologies. A list of contacts is provided on page 2-2.
Item
Part NumberPart Description
Number
1E8356-90001Installation and Quick Start Guide (this document)
2E8356-10001PNA Series Network Analyzer Documentation CD-ROM
3Unique to
AC Power Cord
country
4
5
not applicable
not applicable
a
USB Keyboard (English only)
a
USB Mouse
65063-9229Front Handle Kit, RF-Frequency Model
5063-9230Front Handle Kit, Microwave-Frequency Model
7not applicableFerrite cores and cable ties (refer to “STEP 5. Install Ferrite
Cores” on page 1-7)
a. See the Contents List in the shipping container for the current part number.
1-2Chapter 1
Installing Your Analyzer
STEP 1. Check the Shipment
NOTEA parallel port adapter, part number 1253-4226, was included with early versions of
the PNA. This adapter is no longer necessary.
4. If you ordered any of the following options, check the parts. Option 1CM or 1CP (item
numbers 1 or 2, below) are shipped in a separate container.
Item
Number
11CMRack Mount Flange Kit (For use without handles)
2Rack Mount Rail Set
11CPRack Mount Flange Kit (For use with handles)
2Rack Mount Rail Set
382375AUSB/GPIB Interface (included with Option 083 only)
Option
Number
Part Descript ion
Chapter 11-3
Installing Your Analyzer
STEP 2. Meet Electrical and Environmental Requirements
STEP 2. Meet Electrical and Environmental Requirements
1. Ensure the available AC power source
meets the following requirements:
CAUTIONThis instrument has
autoranging line voltage
input. Be sure the supply
voltage is within the specified
range.
The specified range for the power
requirements is:
•90 − 132 V
•48 − 66 Hz / 400 Hz (single phase)
- or -
• 198 − 264 V
•48 − 66 Hz (single phase)
The instrument power consumption is
600 VA maximum.
3. Verify that the AC power cable is not damaged, and that the power-source outlet
provides a protective earth contact.
AC
AC
2. Ensure the operating environment meets
the following requirements:
•0 to 40 °C
• 5% to 95% relative humidity at 40 °C
(non-condensing)
• 0 to 4,500 meters (0 to 14,760 feet)
altitude
NOTESome analyzer performance
parameters are specified for
25 °C. Later, you can refer to
the Help system in the
analyzer for the complete
specifications.
CAUTIONAlways use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord may cause
product damage.
1-4Chapter 1
Installing Your Analyzer
STEP 3. Install the Analyzer on a Bench Top or in an Equipment Rack
STEP 3. Install the Analyzer on a Bench Top or in an
Equipment Rack
IMPORTANTPlease review all safety information located in Chapter 3 , “Safety and
Regulatory Information” before proceeding with installing and using the
analyzer.
The analyzer can be installed on a bench top or in an equipment rack. When installing the
analyzer in an equipment rack, the analyzer can be set up with handles (option 1CP) or
without handles (option 1CM).
In all installations, consider the following ventilation requirements when deciding where
to set up your analyzer.
CAUTIONVentilation Requirements: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C× for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
Refer to the section, listed below, that applies to your installation.
• For bench top installation, continue on page 1-5.
• For equipment rack installation, continue on page 1-5.
To Install on a Bench Top
When you install the analyzer for bench top use, you may wish to attach the supplied
handles to the analyzer.
If you wish to install the handles, follow the instructions provided with the handle kit.
To Install in an Equipment Rack
(Option 1CP - Rack Mounting with Handles)
(Option 1CM - Rack Mounting without Handles)
When you install the analyzer in an equipment rack, you must install rails in the rack to
support the weight of the analyzer, attach the rack mount flanges (with or without
handles) to the analyzer, and secure the analyzer to the equipment rack.
For installation, follow the instructions provided with the rack mount kit and the rail kit.
Chapter 11-5
Installing Your Analyzer
STEP 4. Connect the Accessories
STEP 4. Connect the Accessories
You can connect several USB accessories to the PNA. Some PNA models have a four-port
USB hub mounted internally and accessible through the rear panel. Other PNA models
have two USB ports: one on the front panel and one on the rear panel. You can also use an
optional USB Hub (model N4689) to increase the available USB ports on the PNA.
USB devices, including the optional hub, can be connected to any available USB port. The
PNA will recognize the USB device when it is connected, and run the appropriate software
driver to make the device operate correctly.
The following is a list of USB devices commonly used with the PNA. Some of these devices
may require additional installation. Refer to the Help that is embedded in the PNA for
more information.
• 82357A USB/GPIB Interface — supplied with Option 083
1-6Chapter 1
Installing Your Analyzer
STEP 5. Install Ferrite Cores
STEP 5. Install Ferrite Cores
CISPR 11 is an international regulation that specifies the maximum level of frequency
emissions that can originate from an electronic device. The regulation is designed to
minimize the interference that one electronic device may have with another.
Frequency emissions can occur when cables are connected to an electronic device. To
minimize this, try to use high-quality cables with built-in molded ferrite cores, with the
core closest to the PNA. If cables with built-in cores cannot be found, use one of the
clamp-on ferrite cores that are included with the PNA.
It is especially important to use these cores on cables that are connected to the following
PNA connectors:
• Parallel Port (rear panel)
• Serial Port (rear panel)
• VGA (rear panel)
• Probe power (front panel)
The core should snap closed to grip the cable firmly, but without damaging the cable.
Attach the core to the cable as close as possible to the PNA connector end. Cable ties are
included to help ensure that the ferrite cores remain in place.
Chapter 11-7
Installing Your Analyzer
STEP 6. Start Up the Analyzer
STEP 6. Start Up the Analyzer
CAUTIONInstall the instrument so that the detachable power cord is readily
identifiable and is easily reached by the operator. The detachable power cord
is the instrument disconnecting device. It disconnects the mains circuits from
the mains supply before other parts of the instrument. The front panel switch
is only a standby switch and is not a LINE switch. Alternatively, an
externally installed switch or circuit breaker (which is readily identifiable
and is easily reached by the operator) may be used as a disconnecting device.
1. Connect the AC power cable.
1-8Chapter 1
Installing Your Analyzer
STEP 7. Perform Administrative Tasks
2. Turn the analyzer on by pressing the front panel On/Off switch. When the analyzer is
turned on, the On/Off switch changes color from yellow to green.
NOTEThe Agilent PNA Series Network Analyzer is a network analyzer that operates on a
personal computer (PC) platform with Microsoft Windows as the operating
system.
The first time you start the PNA, the Microsoft Windows license agreement is
presented. Click I Accept, then Next
.
STEP 7. Perform Administrative Tasks
IMPORTANTThe basic Windows configuration and software shipped with the PNA should
not be modified. This includes the addition or deletion of any hard-disk drive
partitions, the deletion of the “Agilent” user profile, or the modification of any
of the Agilent registry entries. Problems that result from modifications will
not be covered under warranty. For more information, type the keywords
Windows in the index of the Help system in the analyzer.
We recommend that you consider performing the following administrative tasks shortly
after powering up the analyzer for the first time.
❏ Set up analyzer users
❏ Set up the administrator password
❏ Enter the computer name
❏ Set the time and date
❏ Select a local language for the help system (default is English)
Procedures for the above tasks are in the Help system in the analyzer. To view these
procedures:
• On the analyzer’s front panel, press the
Chapter 11-9
Help button. Or, on the PNA display, click Help,
Installing Your Analyzer
STEP 8. Run the Operator’s Check
Network Analyzer Help.
• Click
• Click
• Click the task you wish to perform.
Quick Start in PNA Help.
Perform Administrative Tasks.
STEP 8. Run the Operator’s Check
The operator’s check should be performed when you first receive your analyzer, and any
time you wish to have confidence that the analyzer is working properly. The operator’s
check does not verify performance to specifications, but should give you a high degree of
confidence that the instrument is performing properly.
On the System menu, point to Service, and then click Operator’s Check. Follow the
prompts.
For assistance, use the
Help button displayed on the screen of the Operator’s Check.
STEP 9. Turn Off the Analyzer
The following two modes are used to turn off the analyzer:
• hibernate
• shut down
To Hibernate the Analyzer
There are two methods to hibernate the analyzer:
• Method 1: Press the front panel On/Off switch momentarily.
— Do not hold down the switch.
— Do not repeatedly press the switch.
• Method 2: Click
Within a few seconds, two banners are displayed momentarily. (The first banner states
“Preparing to Hibernate”. The second banner states “Hibernating”.) The current analyzer
state is saved to the analyzer’s hard disk.
The On/Off switch color changes from green to yellow when the analyzer has finished
switching to hibernation state.
When the analyzer is in the hibernation state, pressing the On/Off switch wakes up the
analyzer. The analyzer application reads the hard disk to load the exact state that the
analyzer was in before it was hibernated. This saves time by avoiding a full system boot
up.
Powering-up the analyzer after Hibernation mode does not require a password. If security
is an important issue and you need password protection, choose the shut down mode to
turn off the analyzer.
Start, Shutdown. In the dialog box, select Hibernate.
For more hibernate information, type hibernate in the index of the Help system in the
1-10Chapter 1
Installing Your Analyzer
STEP 9. Turn Off the Analyzer
analyzer.
To Shut Down the Analyzer
IMPORTANTOnly shut down the analyzer when preparing to service it, or provide security
via password protection.
To shut down the analyzer, click
Start, Shutdown. In the dialog box, select Shut Down.
The On/Off switch color changes from green to yellow when the analyzer has finished
switching to shut down mode.
When turned on, the analyzer application powers-up in the preset settings after
performing a full system boot-up and requesting a password.
CAUTIONOnly in an emergency, should the power cord be removed from the analyzer
without first turning off the analyzer. Removing the power cord without the
analyzer in hibernation or shut down mode is likely to damage the analyzer’s
hard drive.
Chapter 11-11
Installing Your Analyzer
STEP 9. Turn Off the Analyzer
1-12Chapter 1
2Getting Help with Your Analyzer
Help System
Use on-line Help to quickly reference programming and user documentation. English,
Chinese, French, German, Japanese, and Spanish language versions are available through
the Help menu. (Programming information is only available in English.)
We recommend that you begin by accessing Help and reading the topics in the Quick Start
section. To access this information:
• On the analyzer’s front panel, press the
Network Analyzer Help.
• Click
Quick Start in PNA Help.
Help button. Or, on the PNA display, click Help,
PNA on the Internet
The PNA Series Network Analyzer System Web page is another resource for help with the
analyzer. This Web page has links to information about equipment and literature related
to PNA series network analyzers. You can visit the Agilent Technologies web site where
you can find product information, service notes, training information, application notes,
online manuals, and other information. To access the PNA Series Network Analyzer
System Web page:
To remove dirt or dust from the external case of the network analyzer, clean the case using
a dry or slightly-dampened cloth only.
WARNINGTo prevent electrical shock, disconnect analyzer from mains before
cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
Shipment for Service
Contact Agilent Technologies for instructions on where to ship the analyzer for service.
Refer to page 2-2 for a list of Agilent Technologies contacts.
Ship the analyzer using the original packaging materials. Shipping the analyzer in
anything other than the original packaging may result in non-warranted damage.
Safety Symbols
The following safety symbols are used throughout this manual. Familiarize yourself with
each of the symbols and its meaning before operating this instrument.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, would result in damage to or destruction of
the product. Do not proceed beyond a caution note until the indicated
conditions are fully understood and met.
WARNINGWarning denotes a hazard. It calls attention to a procedure which, if
not correctly performed or adhered to, could result in injury or loss
of life. Do not proceed beyond a warning note until the indicated
conditions are fully understood and met.
3-1
Safety and Regulatory Information
General Safety Considerations
General Safety Considerations
Safety Earth Ground
WARNINGThis is a Safety Class I Product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only
be inserted in a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor, inside or
outside the product, is likely to make the product dangerous.
Intentional interruption is prohibited.
CAUTIONAlways use the three-prong AC power cord supplied with this product. Failure
to ensure adequate earth grounding by not using this cord may cause product
damage.
Before Applying Power
WARNINGIf this product is not used as specified, the protection provided by
the equipment could be impaired. This product must be used in a
normal condition (in which all means for protection are intact) only.
WARNINGIf an instrument handle is damaged, you should replace it
immediately. Damaged handles can break while you are moving or
lifting the instrument and cause personal injury or damage to the
instrument.
CAUTIONThis instrument has autoranging line voltage input. Be sure the supply
voltage is within the specified range.
CAUTIONThis product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 61010 and 664 respectively.
CAUTIONVentilation Requirements: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C× for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
3-2
Safety and Regulatory Information
Regulatory Information
CAUTIONThe measuring terminals on this instrument are designed to be used with
external signals described in Measurement Category I, but NOT with
external signals described in Categories II, III, and IV. The input of this
instrument cannot be connected to the mains.
Servicing
WARNINGThese servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you
are qualified to do so.
WARNINGNo operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
WARNINGThe opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the product from all voltage sources
while it is being opened.
WARNINGThe power cord is connected to internal capacitors that may remain
live for 5 seconds after disconnecting the plug from its power supply.
Regulatory Information
This section contains information that is required by various government regulatory
agencies.
Instructions for Use
This product has been designed and tested in accordance with the standards listed on the
Manufacturer’s Declaration of Conformity, and has been supplied in a safe condition. The
documentation contains information and warnings that must be followed by the user to
ensure safe operation and to maintain the product in a safe condition.
Install the instrument so that the detachable power cord is readily identifiable and is
easily reached by the operator.
The detachable power cord is the instrument disconnecting device. It disconnects the
mains circuits from the mains supply before other parts of the instrument. The front
panel switch is only a standby switch and is not a LINE switch (disconnecting
3-3
Safety and Regulatory Information
Regulatory Information
device).
Alternatively, an externally installed switch or circuit breaker (which is readily
identifiable and is easily reached by the operator) may be used as a disconnecting device.
3-4
Instrument Markings
The instruction documentation symbol. The product is marked with this
symbol when it is necessary for the user to refer to the instructions in the
documentation.
The CE mark is a registered trademark of the European Community.
(If accompanied by a year, it is when the design was proven.)
The CSA mark is a registered trademark of the Canadian Standards
Association.
Safety and Regulatory Information
Regulatory Information
ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
This is a symbol of an Industrial Scientific and Medical Group 1
Class A product.
This is a required mark signifying compliance with an EMC requirement.
The C-Tick mark is a registered trademark of the Australian Spectrum
Management Agency.
The ON, OFF, and STANDBY symbols are used to mark the positions of the instrument
power line switch. The AC symbol is used to indicate the required nature of the line
module input power.
ON symbolON symbol
STANDBY symbolOFF symbol
OFF symbolAC (alternating current) symbol
Declaration of Conformity
A declaration of conformity is on file for the PNA models, and a copy is available upon
request.
3-5
Safety and Regulatory Information
Warranty
Compliance with German Noise Requirements
This is to declare that this instrument is in conformance with the German Regulation on
Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordung
−3.GSGV Deutschland).
Acoustic Noise Emission/Geraeuschemission
LpA<70 dBLpa<70 dB
Operator Positionam Arbeitsplatz
Normal Operationnormaler Betrieb
per ISO 7779nach DIN 45635 t. 19
Warranty
The actual warranty on your instrument depends on the date it was ordered as well as
whether or not any warranty options were purchased at that time. To determine the exact
warranty on your instrument, contact Agilent Technologies with the model and serial
number of your instrument. See page 2-2 for contact information.
3-6
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.