Cell-free DNA ScreenTape Assay for TapeStation Systems
Quick Guide
The Agilent 4150 (G2992AA) and 4200 (G2991AA and G2991BA) TapeStation systems are automated platforms
for scalable, flexible, fast, and reliable electrophoresis of nucleic acids.
This Quick Guide is intended for use with the Agilent 4150 and 4200 TapeStation systems only. The Agilent Cell-free
DNA ScreenTape assay is exclusively available for use with the Agilent 4150 or 4200 TapeStation systems.
The Cell-free DNA ScreenTape assay is designed for analyzing cell-free DNA samples and double-stranded DNA
molecules from 50 to 800 basepairs.
Specifications
Analytical specificationsCell-free DNA ScreenTape assay
Sizing range50 – 800 bp
Sensitivity
Sizing precision
Sizing accuracy
Quantitative precision
Quantitative accuracy
Quantitative range100 – 4000 pg/µL
%cfDNA functional range100 – 5000 pg/µL
Maximum buffer concentration in
sample
Physical specifications
Analysis time16 samples: <25 min, 96 samples: <150 min
Samples per consumable16
Sample volume required2 µL
Kit stability6 months
Kit size112 samples
1
2
3
1
2
2,3
2
2
Signal-to-noise >3 (single peak)
Determined using the Cell-free DNA Ladder as sample
Sizing Accuracy for analysis with electronic ladder: ±20 %
20 pg/µL
10 % CV
±15 %
15 % CV
±20 %
25 mM NaCl, 25 mM KCl, 3 mM EDTA, 0.1 % NaN
ethanol, 10 % 2-propanol
, 5 mM phosphate buffer, 10 %
3
Agilent Cell-free DNA ScreenTape Quick Guide for TapeStation Systems
Cell-free DNA ScreenTape Assay for TapeStation Systems
Storage Conditions
• Reagent vials and ScreenTape devices: 2 – 8 °C (36 – 46 °F).
• Store partially used ScreenTape devices upright at 2 – 8 °C (36 – 46 °F) for a maximum of 2 weeks.
• Never freeze ScreenTape devices. Discard any accidentally frozen ScreenTape devices.
Kit Components
Part NumberNameColorAmount
5067-5630Cell-free DNA ScreenTape7 ScreenTape devices
5067-5631Cell-free DNA Reagents
• Cell-free DNA Ladder
• Cell-free DNA Sample Buffer
5067-5632Cell-free DNA Ladder1 vial, 50 µL
5067-5633Cell-free DNA Sample Buffer1 vial, 300 µL
2 vials
50 µL
300 µL
Limited Use Label License
Some products within this assay contain SYBR™ Green I, which is licensed from Life Technologies Corporation for use
in research and development only. SYBR™ is a registered trademark of Life Technologies Corporation.
For Research Use Only
Not for use in Diagnostic Procedures.
Additional Material Required for Analysis with the TapeStation Systems
• Loading tips (5067-5598, 1 pk or 5067-5599, 10 pk)
• Optical Tube 8x Strip (401428) and Optical Tube Cap 8x Strip (401425)
• Vortex mixer IKA MS3 with 96-well sample plate adapter
• 96-well sample plates (5042-8502) and 96-well Plate Foil Seal (5067-5154) (4200 TapeStation systems only)
Additional Equipment Required (Not Supplied)
• Volumetric micropipettes for handling volumes from 1 to 15 µL
• Centrifuges for tube strips and 96-well sample plates
WAR NI NG
CA UTI ON
2
Toxic agents
Refer to product material safety datasheets for further information.
When working with the ScreenTape assay follow the appropriate safety procedures such as
wearing safety goggles, laboratory gloves and protective clothing.
Damage to the TapeStation systems
Only use the recommended consumables and reagents with the TapeStation systems.
Cell-free DNA ScreenTape Assay for TapeStation Systems
Essential Measurement Practices
Read about good measurement practices in the Agilent Information Center and/or in the System Manual.
Environmental
conditions
Steps before sample
preparation
Pipetting practice• Pipette reagents carefully against the side of the 96-well sample plate or sample tube
• Keep reagents during sample preparation at room temperature
• Allow Sample Buffer to equilibrate at room temperature for 30 min prior to use
• Vortex each vial and briefly spin down
• Flick ScreenTape device to eliminate bubbles in the buffer chamber
• Ensure that no sample or Sample Buffer remains within or on the outside of the tip
• Care must be taken due to viscosity of the Sample Buffer
• Apply foil seal to 96-well sample plate or cap the tube strips prior to mixing and centrifugation
• Centrifuge to collect liquid at the base; then vortex using the IKA MS3 vortexer and adaptor at
2000 rpm for 1 min
• Briefly centrifuge and visually confirm that all liquid is collected at the bottom of the 96-well
sample plate or tube strips and any air bubble is removed
• Run samples immediately after preparation
Ladder Considerations
• Ladder is exclusively loaded from location A1 on the tube strip holder.
• Always use a complete tube strip when running ladder or samples from the tube strip holder.
• For best sizing and molarity quantification results, a ladder per ScreenTape device is recommended. Alternatively,
an electronic ladder is available, which can be selected in the Agilent TapeStation Controller software.
3
Cell-free DNA ScreenTape Assay for TapeStation Systems
Sample
Buffer
DNA
2 µL
2 µL
Flick
Spin down 1 min
Samples, tips, and ScreenTape device
Vortex 1 min
Cell-free DNA
Cell-free DNA ScreenTape Assay Operating Procedure
1 Allow Cell-free DNA Reagents to equilibrate at room temperature for 30 minutes.
2 Launch the Agilent TapeStation Controller software.
3 Flick the Cell-free DNA ScreenTape device and insert it into the ScreenTape nest of the TapeStation instrument.
4 Select required sample positions in the TapeStation Controller software.
5 The required consumables (tips, further ScreenTape devices) are displayed in the TapeStation Controller software.
6 Vortex reagents and samples. Spin down before use.
7 Prepare ladder:
• For 1 ScreenTape device: pipette 2 µL Cell-free DNA Sample Buffer ( ) and 2 µL Cell-free DNA Ladder ( ) at
position A1 in a tube strip.
• For 2 ScreenTape devices: pipette 4 µL Cell-free DNA Sample Buffer ( ) and 4 µL Cell-free DNA Ladder ( ) at
position A1 in a tube strip.
• For more than 2 ScreenTape devices
Ladder ( ) at position A1 in a tube strip.
8 For each sample, pipette 2 µL Cell-free DNA Sample Buffer ( ) and 2 µL DNA sample in a tube strip or 96-well
sample plate
9 Apply caps to tube strips and/or foil seals to 96-well sample plates.
10 Mix liquids using the IKA MS3 vortexer at 2000 rpm for 1 min.
11 Spin down samples and ladder for 1 min.
1
.
1
: pipette 15 µL Cell-free DNA Sample Buffer ( ) and 15 µL Cell-free DNA
Sample Analysis
1 Load samples into the TapeStation instrument. Place ladder in position A1 on tube strip holder.
2 Carefully remove caps of tube strips. Visually confirm that liquid is positioned at the bottom.
3 Click Start.
4 The TapeStation Analysis software opens automatically after the run and displays results.
Technical Support and Further Information
For technical support, please visit www.agilent.com/chem/contactus. Visit Agilent Technologies’ web site. It offers
useful information, support and current developments about the products and technology:
www.agilent.com/genomics/tapestation.
1
Agilent 4200 TapeStation system only
www.agilent.com
Agilent Technologies Inc. 2018-2021
Printed in Germany, Edition: 01/2021
*G2991-90060 *
Part No: G2991-90060 Rev. C
Document No: D0000319 Rev. C
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.