Agilent Cary 100/300/4000/5000/6000i/7000 Spectrophotometer
Site Preparation Guide
Notices
Manual Part Number
WARNING
CAUTION
8510128000 Edition 12, 03/21
Copyright
© Agilent Technologies, Inc. 1996­2000, 2002, 2008, 2011-2012, 2016. 2021
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Agilent Technologies Australia (M) Pty Ltd
679 Springvale Road Mulgrave, Victoria, 3170 Australia
www.agilent.com
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met
.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data ­Commercial Items) and DFARS
227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation).

Request for Installation

All preparations have been completed. Please arrange for the installation to be completed as soon as possible. I understand that if the installation site is not prepared in accordance with the enclosed instructions, additional installation charges may apply.
Company name:
Company address:
Name:
Position:
Telephone:
Preferred installation date:
Signed:
Date:
Agilent Cary 100/300/4000/5000/6000i/7000 Spectrophotometer Site Preparation Guide 3
Requirements
The work area is prepared and meets requirements (see Chapter 3).
The power supply meets requirements (see Chapter 4).
A nitrogen gas supply for optics path purging is ready (Cary 100/300 needs the Extended Sample Compartment option) (see Chapter 5).
The Cary instrument and accessories are on site and packages have been inspected (see Chapter 6).
The computer meets requirements (see Chapter 7). Computer type: _________________________
The Microsoft Windows operating system is installed (see Chapter 7).
Operator training is required (see Chapter 8).

Site Preparation Checklist

Your site must meet all requirements before you request installation. Complete each requirement listed in the table. After completing each requirement, place a check in the appropriate checkbox. Ensure you compare each item inside the boxes with the packing list supplied with the boxes.
NOTE
Do not unpack the instrument. If unpacked incorrectly, equipment may become damaged. Your Agilent representative will do this for you.
Agilent Cary 100/300/4000/5000/6000i/7000 Spectrophotometer Site Preparation Guide 4

Contents

Contents
Request for Installation 3
Site Preparation Checklist 4
1 Introduction 7
2 Safety Practices and Hazards 9
Instrument ventilation 9
Electrical hazards 9
3 Work Area 11
Suitability 11
Environmental conditions 11
Workbench 12
4 Electrical Specifications 15
Mains supply 15
Fuses 16
External connections 16
Mains inlet coupler 16
Mains power cord 16
5 Optional Nitrogen Supply 17
6 Equipment On-Site 19
Insurance 19
In-house transit routes 19
Inspecting for transit damage 19
7 Computer System Requirements 21
Recommended configuration 21
Agilent Cary 100/300/4000/5000/6000i/7000 Spectrophotometer Site Preparation Guide 5
Contents
Recommended printer 22
IEEE 488.2 interface 22
Interconnecting cables 22
Setting up your computer 23
8 Operator Training 25
Agilent Cary 100/300/4000/5000/6000i/7000 Spectrophotometer Site Preparation Guide 6
Introduction

1 Introduction

The range of Agilent Cary spectrophotometers are complete analysis systems that have been fully tested and proven to specification before dispatch from the manufacturing plant.
This manual contains general information relevant to the preparation of an installation site, and details the facilities that must be provided to ensure that the system can be effective. Detailed operating procedures are provided in the operation manual that is supplied with the spectrophotometer.
You should complete the site preparation checklist on Page 4 and then send a copy of the checklist to the Agilent sales and service office or local Agilent agent dealing with the supply of the equipment. On receipt of this document, the Agilent representative will contact you and arrange a convenient time for installation.
As a guide to the time required for installation by an Agilent representative, a typical Cary system will be completed and ready for use within four hours. This time may be extended if the system includes accessories. The installation time can be kept to a minimum by ensuring proper preparation of the site and arrangement of all units to allow for easy access.
NOTE
The remainder of this manual contains information such as the environmental requirements and technical specifications for the Cary spectrophotometer, and it should not be discarded — keep this manual for future reference.
Agilent Cary 100/300/4000/5000/6000i/7000 Spectrophotometer Site Preparation Guide 7
Introduction
This page is intentionally left blank.
Agilent Cary 100/300/4000/5000/6000i/7000 Spectrophotometer Site Preparation Guide 8
Safety Practices and Hazards

2 Safety Practices and Hazards

Your Agilent Cary instrument and accessories have been carefully designed so that when used properly you have an accurate, fast, flexible and safe analytical system.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Information on safety practices appears throughout the documentation (both printed and online) provided with your instrument and accessories. Before using the instrument or accessories, you must thoroughly read these safety practices.
Observe all relevant safety practices at all times.

Instrument ventilation

Ozone can be generated by radiation from the source lamps. Exposure to ozone can result in severe irritation to the skin, eyes, and upper respiratory system. The maximum permissible exposure level is 0.1 parts per million (0.2 milligrams per cubic meter).
ALWAYS ventilate the area surrounding the spectrophotometer such that the concentration of ozone does not exceed the maximum permissible level. All venting must be to outside air, never within the building.

Electrical hazards

Good grounding/earthing is essential to avoid a potentially serious electric shock hazard. Ensure that there is an integral ground connection between the metal base of the spectrophotometer and accessories and the 3-pin earth-grounded receptacle. Consult the manuals or product labels supplied with your computer, monitor and printer for the relevant grounding requirements.