No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
B1500-90004
Edition
Edition 1, July 2005
Edition 2, December 2005
Edition 3, April 2006
Edition 4, January 2007
Edition 5, June 2007
Agilent Technologies, Inc.
5301 Stevens Creek Blvd
Santa Clara, CA 95051 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Tec hn ol og y L i c e ns es
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or
contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
Page 3
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name:
Agilent Technologies International sarl
Manufacturer’s Address:
Supplier’s Address:
Rue de la Gare 29
CH - 1110 Morges
Switzerland
Declares under sole responsibility that the product as originally delivered
Product Name:
Semiconductor Device Analyzer
High Power Source/Monitor Unit Mo du le
Medium Power Source/Monitor Unit Module
High Resolution Source/Monitor Unit Module
Multi Frequency Capacitance Measurement Unit Module
This declaration covers all options of the above product(s)
complies with the essential requirements of the following applicable European Directives, and carries
the CE marking accordingly:
Low Voltage Directive (73/23/EEC, amended by 93/68/EEC)
EMC Directive (89/336/EEC, amended by 93/68/EEC)
and conforms with the following product standards
EMCStandard
IEC 61326:2002 / EN 61326:02
CISPR 11:1999+A2 / EN 55011:1998+A1
IEC 61000-4-2:2001 / EN 61000-4-2:1995+A1+A2
IEC 61000-4-3:2002 / EN 61000-4-3:1996+A1+A2
IEC 61000-4-4:1995+A1+A2 / EN 61000-4-4:1995+A1
IEC 61000-4-5:2001 / EN 61000-4-5:1995+A1
IEC 61000-4-6:2001 / EN 61000-4-6:1996+A1
IEC 61000-4-11:2001 / EN 61000-4-11:1994+A1
Supplementary Information:
This DoC applies to above-listed products placed on the EU market after:
Jun 14, 2005
Date
Toshiyuki Kawaji
QA Manager
Agilent Technologies
Page 4
•Herstellerbescheinigung
GEÄUSCHEMISSION
Lpa < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
•M
anufacturer’s Declaration
ACO
USTIC NOISE EMISSION
Lpa < 70dB
operator position
normal operation
per ISO 7779
NOTEThis ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme ?Hla norme NMB-001 du Canada.
This product complies with the WEEE Directive (2002/96/EC) marking requirements.
The affixed label indicates that you must not discard this electrical/ electronic product
in d
omestic household waste.
Product Category: With reference to the equipment types in the WEEE Directive
Ann
ex I, this product is classed as a “Monitoring and Control instrumentation”
product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or see
www
.agilent.com/environment/product/ for more information.
Microsoft and Windows are registered trademarks of Microsoft Corporation. All other trademarks are the property of their
respective owners. EasyEXPERT is a trademark of Agilent Technologies.
Page 5
Safety Summary
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual may impair the
protections provided by the equipment. In addition, it violates safety standards of
design, manufacture, and intended use of the instrument. Agilent Technologies, Inc.
assumes no liability for customer’s failure to comply with these requirements.
NOTEAgilent B1500 complies with INSTALLATION CATEGORY II for mains input and
INSTALLATION CATEGORY I for measurement input terminals, and POLLUTION
DEGREE 2 defined in IEC 1010-1.
Agilent B1500 is an INDOOR USE product.
NOTELED in Agilent B1500 is Class 1 in accordance with IEC 825-1. CLASS 1 LED PRODUCT.
•GROUND THE INSTRUMENT
T
his is Safety Class I instrument. To minimize shock hazard, the instrument chassis
and c
abinet must be connected to an electrical ground. The power terminal and the
power cable must meet International Electrotechnical Commission (IEC) safety
standards.
•DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
o not operate the instrument in the presence of flammable gases or fumes.
D
Ope
ration of any electrical instrument in such an environment constitutes a
definite safety hazard.
•KEEP AWAY FROM LIVE CIRCUITS
eration personnel must not remove instrument covers. Component replacement
Op
and in
ternal adjustments must be made by qualified maintenance personnel. Do
not replace components with power cable connected. Under certain conditions,
dangerous voltages may exist even with the power cable removed. To avoid
injuries, always disconnect power and discharge circuits before touching them.
•DO NOT SERVICE OR ADJUST ALONE
not attempt internal service or adjustment unless another person, capable of
Do
ren
dering first aid and resuscitation, is present.
Page 6
•DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute
pa
rts or perform any unauthorized modification to the instrument. Return the
instrument to a Agilent Technologies Sales and Service Office for services and
repair to ensure that safety features are maintained.
•DANGEROUS PROCEDURE WARNINGS
arnings, such as example below, precede potentially dangerous procedures
W
th
roughout this manual. Instructions contained in the warnings must be followed.
WARNINGDangerous Voltage, capable of causing death, are present in this instrument. Use
extreme caution when handling, testing, and adjusting.
Page 7
Safety Symbols
>
The general definitions of safety symbols used on equipment or in manuals are listed
below.
Instruction manual symbol: the product will be ma
necessary for the user to refer to the instruction manual in order to protect against
mage to the instrument.
da
Indicates dangerous voltage and potential for electrical shock. Do not touch terminals
that have this symbol when instrument is on.
Protective conductor terminal. For protection against electrical shock in case of a fault.
Used with field wiring terminals to indicate the terminal which must be connected to
d before operating equipment.
groun
Frame or chassis terminal. A connection to the frame (chassis) of the equipment which
normally includes all exposed metal structures.
Indicates earth (ground) terminal.
Alternating current.
`
Direct current.
STANDBY (Supply).
Means INSTALLATION CATEGORY I. Measurement terminals on the rear panel comply
&$7,
with INSTALLATION CATEGORY I.
WARNINGThe warning sign denotes a hazard. It calls attention to a procedure, practice, condition
or the like, which, if not correctly performed or adhered to, could result in injury or
death to personal.
rked with this symbol when it is
CAUTIONThe caution sign denotes a hazard. It calls attention to an operating procedure,
practice, condition or the like, which, if not correctly performed or adhered to, could
result in damage to or destruction of part or all of the product.
Page 8
HIGH VOLTAGE SHOCK HAZARD
Agilent B1500 can force dangerous voltages (200 V for HPSMU, and 100 V for
MPSMU/HR
SMU) at the force, guard, and sense terminals. To prevent electric
shock hazard, the following safety precautions must be observed during the use of
Agilent B1500.
•Use a three-conductor AC power cable to connect cab
inet (if used) and Agilent
B1500 to an electric ground (safety ground).
•If you do not use Agilent 16442 Test Fixture, make sure to connect the Interlock
termi
nal to a switch that turns off when the shielding box access door is opened.
•Confirm periodically that interlock function works normally.
•Before touching the connections of the force, guard, and sense terminals, turn
Ag
ilent B1500 off and discharge any capacitors whenever possible. If you do
not turn Agilent B1500 off, complete “all” of the following items, regardless of
any Agilent B1500 settings.
•T erminate measurem ent by pressing Stop k ey, confirm that the Measurement
s indicator is not lit.
statu
•Confirm that the
High Voltage indicator is not lit.
•Open the shielding box access door (open the Interlock terminal).
•Discharge any capacitors if the capacitance is connected to an SMU.
•Warn workers in the vicinity of Agilent B1500 about dangerous conditions.
Une tension dangereuse (max. ± pour HPSMU; 200 Vdc, max. ± pour MPSMU/
HRSM
U; 100 Vdc) émanant du dispositif Agilent B1500 peut être sortie aux bornes
de force, d'appareil de protection ou de détection. Les précautions suivantes doivent
être obserées contre commotion électrique accidentelle:
•Mettre à la terre le dispositif Agilent B1500 au moyen du câble d'alimentation
tripo
lair.
•En cas de hors service du disp ositif d'e ssai, FIXTUR E Agilent 1 6442, connecter
les bo
rnes de verrouillage (Interlock) de façon à ce que soit ouverte lorsque le
cle de la boîte de blindage est ouvert.
couver
•Essayer périodiquement le fonctionnement normal de verrouillage.
•Avant de toucher la partie connexion à partir des bornes de force, d'appareil de
pr
otection et de détection, mettre hor s tension le di spositif A gilent B1500 . Et, en
cas de condensateurs connectés au circuit de mesure, décharger ces
condensateurs. Lorsque l'alimentation n'est pas mise hors tension , les 4
t
ructions suivantes doivent être exécutées:
ins
•Finir la mésure en appuyant sur la touche “Stop”; verifier que l'indicateur
“Measurem
•S'assurer que soit allumé l'indicateur d'alarme de la haute tension.
•Ouvrir le convercle de la boîte de blindage (Ouvrir les bornes de
ver
rouillage).
ent” n'est pas allumé.
•En cas de condensateurs connectés à SMU, décharger les condensateurs.
•Alerter d'autres personnes autour de vous contre le danger de haute tension.
Page 11
Achtung! Gefährliche Spannung
Von den Geräten Agilent B1500 können Spannungen an den Anschlüssen “Force,
Gu
ard und Sense” von bis zu 200 V ausgehen. Um elektrischem Schlag vorzubeu-
gen, ist bei der Benützung der Geräte Agilent B1500 folgendes zu beachten:
•Erden Sie das Kabinett (falls verwendet) sowie die Geräte Agilent B1500 mittels
dre
iadriger Netzleitungen.
•Wenn die Meßfassung, Agilent 16442, zwar angeschlossen, jedoch nicht
ve
rwendet wird, schließen Sie die Verriegelungsklemme (Interlock) so an, daß
geöffnetem Deckel die Stromzuführung zu Agilent 16442 auf jeden Fall
bei
terbrochen wird.
un
•Vergewissern Sie sich regelmäßig daß die Verriegelungsfunktion korrekt
arbe
itet.
•Schalten Sie die Geräte Agilent B1500 aus, und entladen Sie alle Kapazitäten
be
vor Sie die Anschlüsse “Force, Guard und Sense” berühren. Falls Sie die
Geräte Agilent B1500 nicht ausschalten, führen Sie unabhängig von den
Geräteeinstellungen folgende Schritt durch:
•Beenden Sie die gegenwärtige Messung durch Drücken der Stop Taste (die
urem
Meas
•Vergewissern Sie sich daß die Hochspannungswarnlampe erloschen ist.
•Öffnen Sie den Deckel der Meßfassung Agilent 16442 (die
Ve
rriegelungsklemme (Interlock) öffnen).
ent Leuchtdiode erlischt).
•Entladen Sie alle an SMUs angeschlossenen Kondensatoren (falls
vor
handen).
•Informieren Sie Personen in unmittelbarer Nähe d er Geräte Ag ilent B1500 über
die
Gefährlichkeit der bestehenden Hochspannung, und sichern Sie den Zugang
zum Prüf
platz zum Schutz Dritter.
Page 12
Precautionary Statement
Agilent B1500 Semiconductor Device Analyzer operates in the Microsoft Windows XP
Professional environment. Agilent B1500 requires Agilent EasyEXPERT software, a
specially-designed Windows XP application program.
•About guarantee and support for Agilent B1500
ilent Technologies guarantees and supports the performance of Agilent B1500
Ag
for
the same condition as the preload condition when shipped from the factory.
•About updating Agilent EasyEXPERT and the Windows Update
Agilent Technologies confirms the operation of Agilent EasyEXPERT patch
prog
rams and important Windows security patches, and provides recommended
update information. Visit Agilent B1500 support site, download the patches, and
perform the software update.
•About Windows XP application programs and peripherals (including driver)
Using commercial products on Agilent B1500 is your responsibility. Agilent
chnologies cannot provide compatibility information for commercial products. If
Te
problems arise, perform Agilent B1500 system recovery.
•About servicing
Bench repair service is available at your
center. Be aware that the B1500 configuration might be updated to the latest one
without notice because of support issues.
The internal hard disk drive (HDD) might be initialized during servicing. If
ripherals are connected, they will be removed.
pe
When Agilent B1500 is returned, the internal HDD might be initialized. Peripherals
will
be returned separately.
•Other notes
•Back up the internal HDD to prevent loss of data by accident or failure.
•Protect Agilent B1500 from computer viruses.
•I
f you connect Agilent B1500 to the network, take care to protect it from
uter virus.
comp
nearest Agilent Technologies service
Page 13
Working in Comfort
To optimize your comfort and productivity, it is important that you set up your work area
correctly and use your instrument properly. With that in mind, we have developed some
set-up and use recommendations for you to follow based on established ergonomic
principles. Improper and prolonged use of keyboards and input devices are among
those tasks that have been associated with repetitive strain injury (RSI) to soft tissues
in the hands and arms. If you experience discomfort or pain while using the instrument,
discontinue use immediately and consult your physician as soon as possible. For more
information on RSI you may wish to consult the About Repetitive Strain Injury section.
Please study the recommendations described below. Included there are references to
relevant parts of international standards, regulations and guidelines, such as ISO 9241
and the European Community Display Screen Equipment directive. You may also wish
to consult your employer’s human resources department or other relevant departments
for guidance specific to your company.
About Repetitive Strain Injury
Because your comfort and safety are our primary concern, we strongly recommend
h
at you use the instrument in accordance with established ergonomic principles and
t
recommendations. Scientific literature suggests that there may be a relationship
between injury to soft tissues -especially in the hands and arms- and prolonged
improper use of keyboards or other equipment requiring repeated motions of the
hands and forearms. This literature also suggests that there are many other risk
factors that may increase the chance of such injury, commonly called Repetitive Strain
Injury.
What is RSI?
Repetitive Strain Injury (RSI -also known as cumulative trauma disorder or repetitive
tio
n injury) is a type of injury where soft tissues in the body, such as muscles,
mo
nerves, or tendons, become irritated or inflamed. RSI has been a reported problem for
those who perform repetitive tasks such as assembly line work, meatpacking, sewing,
playing musical instruments, and computer work. RSI also has been observed in those
who frequently engage in activities such as carpentry, knitting, housework, gardening,
tennis, windsurfing and lifting children.
What causes RSI?
The specific causes of RSI have not been established. Nevertheless, the incidence of
I h
as been associated with a variety of risk factors, including:
RS
•Too many uninterrupted repetitions of an activity or motion.
Page 14
•Performing an activity in an awkward or unnatural posture.
•Maintaining static posture for prolonged periods.
•Failing to take frequent short breaks.
•Other environmental and psychosocial factors.
In addition, there have been reports associating the occurrence of RSI with the use of
ke
boards, mice, and other input devices. Also, certain medical conditions, such as
y
rheumatoid arthritis, obesity and diabetes, may predispose some people to this type of
injury.
What if I experience discomfort?
If you are experiencing any discomfort, seek professional medical advice immediately.
pically, the earlier a problem is diagnosed and treated, the easier it is to resolve.
Ty
Mice and Other Input Devices
Various aspects of using mice and other inpu
discomfort or injury. Observing the following recommendations may reduce that risk.
•Try to keep your hand, wrist, and forearm in a neutral position while using your
e or other input device.
mous
•If you use your thumb to rotate the ball on a trackball or spaceball, keep it in a
relaxe
d, natural shape, and maintain a neutral posture in your hand, wrist, and
forearm.
•Hold the mouse gently by draping your fingers over it. Keep your hand relaxed and
ers loose. Do not grip the mouse tightly.
fing
•It takes very little pressure or force from your fingers to activate the buttons or
sc
roll wheel on your mouse, scrolling mouse, trackball, or other input device.
Using too much force can place unnecessary stress on the tendons and muscles
in your hands, wrists, and forearms.
•If you are using a scrolling mouse, be sure to keep your fingers and hand in a
xed, neutral position when activating the scroll wheel. Also, this type of mouse
rela
features software that can minimize the number of mouse movements or button
clicks.
•When using a mouse, trackball, or other input device, position it as close to the
yboard as possible, and keep it at the same level as you do not have to stretch
ke
while using it.
•Be sure to keep your mouse and trackball clean. Regular removal of accumulated
ust and dirt helps ensure proper tracking and reduces unnecessary hand and
d
wrist motions.
t devices may increase your risk of
Page 15
In This Manual
This manual describes the front panel operation, installation, and functions of Agilent
Technologies B1500. This manual consists of the following chapters:
•Chapter 1, “Getting Started”
This chapter describes the basic operations of Agilent B1500.
•Chapter 2, “Introduction”
This chapter describes overview, specifications, accessories and options of
Ag
nt B1500.
ile
•Chapter 3, “Installation”
This chapter explains how to install Agilent B1500, and how to connect the device
un
der test to a test fixture.
•Chapter 4, “Using EasyEXPERT”
This chapter provides the reference inform
•Chapter 5, “Classic Test Definition”
This chapter provides the reference inform
•Chapter 6, “Application Test Definition”
This chapter provides the reference information of the application test definition.
•Chapter 7, “Function Details”
This chapter explains the several functions and initial settings of Agilent B1500.
•Chapter 8, “Built-in Programming Tool”
This chapter provides information about the built-in mathematical functions and
r
e
ad out functions.
•Chapter 9, “If You Have a Problem”
This chapter explains how to solve a problem if you encounter any problem, and
de
scribes error codes. This chapter also describes how to perform the system
recovery and the data backup/recovery.
ation of Agilent EasyEXPERT software.
ation of the classic test definition.
Page 16
NOTETo get the latest firmware/software/electronic manuals/support information, visit
Agilent Technologies support site (http://www.home.agilent.com), and click
Additional Test & Measurement Products > Parametric Test, click Semiconductor
Parameter/Device Analyzer Series, and click B1500A Semiconductor Device Analyzer.
You can reach Agilent B1500A support site.
Page 17
Online Documents
The following electronic documentation files are stored in Agilent B1500’s internal
hard disk drive. The files provide the information of how to use Agilent B1500 and
Agilent EasyEXPERT software.
Folder: C:\Program Files\Agilent\B1500\Documents
•B1500_help.chm
A part of Agilent B1500A User’s Guide.
from the Help menu on the EasyEXPERT main screen.
•B1500_Self_Trng.mht
Agilent B1500A Self-paced Training Manual that
This is the online help that can be opened
can be opened by a browser
Page 18
Utilities
The following Windows execution files are stored in Agilent B1500’s internal hard disk
drive. For more information, see Chapter 7, “Function Details”.
This chapter describes the basic operations of Agilent B1500. Before learning the
product details, let’s try to use Agilent B1500 briefly. The operations need Agilent
B1500, power cable, and USB keyboard (USB mouse and stylus pen are optional).
During the operations, open the measurement terminals.
Basic Operations:
1. T o turn on Agilent B1500
2. To use Agilent EasyEXPERT
3. To create a workspace
4. To specify your workspace
5. To select a test definition (pre-defined test setup)
6. To change the test conditions
7. T o start measurement
8. To set flag/string to the test result record
9. To use the analysis tools
10. To turn off Agilent B1500A
1-2Agilent B1500 User’s Guide, Edition 5
Page 33
Getting Started
Agilent B1500 has the following front panel keys.
•Stop
Stops the present measurement or source output immediately.
•Rotary knob
Rotating the knob moves the marker on the graph window, or
increases/decreases/changes the value in the active entry field.
Pressing the knob sets or enters the value.
•Softkeys
Seven softkeys are available. Used to select one for the entry field specified or
the dialog box. They are also used to recall the test definitions (pre-defined test
setups).
•Touch Panel Off
Enables or disables the touch screen operation.
•Standby switch (lower right corner of the front panel)
Turns the B1500A on. Pressing the button in the ON state makes the B1500A in
the standby state.
•LCD Off
Enables or disables the LCD panel. LED lights when the LCD is disabled.
•Auto, -, +, Menu
Four keys are available for the display adjustment. Press Auto to perform
adjustment automatically. Use -, +, and Menu to adjust manually.
Agilent B1500 also requires Agilent EasyEXPERT software as its graphical user
interface. You can operate Agilent EasyEXPERT by using the touch panel. Use your
fingers, stylus pen, and so on for the touch panel operation. The USB keyboard and
the USB mouse are also available for operating the EasyEXPERT.
Agilent B1500 User’s Guide, Edition 51-3
Page 34
Getting Started
NOTEWhen turning the B1500A on
Open the measurement terminals at the device side when turning the B1500A on.
Also disconnect the device from the measurement terminals and open the
measurement terminals after the test. If you leave the connection with the device,
the device may be damaged by unexpected operations or char ge-u p of measuremen t
cables.
NOTEIf you use the B1500A at the first time
If this is the first time to turn Agilent B1500 on after the delivery, you need to
perform the initial setup of the B1500. See “Inspection and Installation” on page
3-6. After the initial setup and if no users are added, you can automatically logon as
“Agilent B1500 User” account without passw ord .
NOTEIf Start EasyEXPERT button is not displayed
Select All Programs > Start EasyEXPERT from the Start menu. The
Start EasyEXPERT button will be displayed.
Step 1. T o turn on Agilent B1500
1. Connect the power cable from Agilent B1500 to an AC power outlet.
2. Connect the USB keyboard to the B1500. Optionally , conn ect the USB mouse to
the B1500.
3. Press the Standby switch (lower right corner of the front panel). Windows,
measurement module initialization, and self-calibration will start. For the
Windows logon screen, log on Windows.
After logging on, click the Start EasyEXPERT button.
And wait until the EasyEXPERT is activated.
1-4Agilent B1500 User’s Guide, Edition 5
Page 35
Step 2. To use Agilent EasyEXPERT
1. If this is the first time to start the EasyEXPERT, or if there is no workspace, the
B1500 displays the screen as shown in Step 5. Go to Step 5. The B1500 will
create a workspace automatically.
Workspace means the space created in Agilent B1500A’s internal hard disk
drive, and is used to store the test setup, test result data, and so on. The
workspace can be created and allocated for each user.
2. If there is two or more workspaces, the B1500 displays the screen as shown in
Step 4. Go to Step 4.
3. If there is one workspace only, the following screen appears.
Getting Started
a. If you do not want to create a workspace, select the Yes radio button and
click OK. Then go to Step 5.
b. If you want to create a workspace, select the No radio button and click Next.
Then go to Step 3.
Agilent B1500 User’s Guide, Edition 51-5
Page 36
Getting Started
Step 3. To create a workspace
1. Enter the name of the new workspace into the above-entry field.
Check Allow other users to access this workspace box if you want to create a
public workspace that is opened for all users.
2. If you are the owner of the existing workspace, you can change the name of the
existing workspace.
If you want to rename the existing workspace, enter the name into the
below-ent ry field.
Check Allow other users to access this workspace box if you want to set it to a
public workspace that is opened for all users.
3. Click OK. Then go to Step 5.
To cancel creating workspace, click Prev.
1-6Agilent B1500 User’s Guide, Edition 5
Page 37
Step 4. To specify your workspace
Getting Started
For the screen as shown above, perform the following steps. This example selects
the workspace named as Demo Workspace that is a public workspace that is opened
for all users.
1. Select Public Workspaces owned by other users radio button.
2. Click Demo Workspace to select it.
3. Click Continue.
Agilent B1500 User’s Guide, Edition 51-7
Page 38
Getting Started
Step 5. To select a test definition (pre-defined test setup)
For the screen as shown above, perform the following steps.
1. Set the Category area to change the test definitions listed in the Library area.
For example, check the CMOS only.
2. Select a test definition from the Library area.
For example, click Id-Vg. Display will be changed as shown in Step 6.
NOTETest Definitions
Test definitions are the built-in test s etups that h a ve been defined and stor ed in th e
EasyEXPERT as the application library. The EasyEXPERT contains more than one
hundred test definitions. You can execute the test without modifications. Also you
can create your own test setup/definition by making any changes.
All furnished test definitions are just sample. If the samples damage your devices,
Agilent Technologies is NOT LIABLE for the damage.
1-8Agilent B1500 User’s Guide, Edition 5
Page 39
Step 6. To change the test conditions
Getting Started
Use the entry fields in the Test Parameters area to change the test conditions. For
example, the following steps change the sweep range of voltage output.
1. Set VgStart to 0 V and VgStep to 0.2 V.
2. VsubsStep to -1.0 V.
T o enter the value, us e the USB keyboard, or the scr een keyboard/num eric keyboard
opened by clicking the button to the right of the entry field.
Agilent B1500 User’s Guide, Edition 51-9
Page 40
Getting Started
Step 7. T o start measurement
1. Click the Single button (upper right on the screen) to start a single sweep
measurement.
This opens the Data Display window as shown below, and starts the single
measurement. The window updates the display and shows the measurement data in
real time.
1-10Agilent B1500 User’s Guide, Edition 5
Page 41
Step 8. To set flag/string to the test result record
After the measurement, the measurement data is displayed on the Data Display
window. For the measurement after doing the changes as shown in Step 6, you will
have the test result as sh own i n Step 7. And the following dialog box appears on t he
lower left area of the B1500 screen. This dialog box is called as Test Result Editor.
The Test Result Editor provides the following GUI to set a flag and remarks to the
test result record.
•Save button and Delete button
Divides test records into groups, Save and Delete.
Test records in Save-group are always listed in the lower area of the
EasyEXPERT main screen.
Test records in Delete-group can be listed when Results > Filter >
Show Deleted Data is checked.
Getting Started
•Flags buttons
The following buttons are available. You can set one of the following flags.
•Important ! button
Sets the important flag (!) to the test result record.
•Valid # button
Sets the valid flag (#) to the test result record.
•Questionable ? button
Sets the questionable flag (?) to the test result record.
•Remarks field
You can enter characters into this field. The characters will be recorded as the
Remarks value of the test result record.
•OK button
Applies the setup on the Test Result Editor, and closes this dialog box.
Agilent B1500 User’s Guide, Edition 51-11
Page 42
Getting Started
Step 9. To use the analysis tools
Measurement result graph is displayed on the Graph Plot area. This area also
displays parameters for the marker, cursor, and lines when they are effective.
•View menu: used to adjust the graph scale.
•Marker menu: used to display and control the marker.
•Cursor menu: used to display and control the cursor.
•Line menu: used to display and control the lines.
The List Display area lists the measurement data.
Step 10. To turn off Agilent B1500A
1. Perform W i nd ow s s hut do w n p roces s us i ng t he Start menu. Or press the Standby
switch (lower right corner of the front panel).
Windows will be shutdown and Agilent B1500A will become the standby state.
1-12Agilent B1500 User’s Guide, Edition 5
Page 43
2Introduction
Page 44
Introduction
This chapter describes the basic functions and features of Agilent B1500A
Semiconductor Device Analyzer, and consists of the following sections:
•“Overview”
•“Front View”
•“Rear View”
•“Measurement Units”
•“Specifications”
•“Accessories and Options”
NOTEApplication Library
Agilent B1500A has installed Agilent EasyEXPERT software as the user interface.
The EasyEXPERT contains an application library that is a set of test definitions. The
application test can be performed by selecting a test definition and setting the test
condition for the actual DUT (device under test).
All test definitions are just sample. If the samples damage your devices, Agilent
Technologies is NOT LIABLE for the damage.
NOTEUtility
The folder C:\Program Files\Agilent\B1500\EasyEXPERT\Utilities stores the
following sample programs.
•Prober control programs
•4155/4156 setup file converter
•sleep.exe program
For more details, see Chapter 7, “Function Details.”
2- 2Agilent B1500 User’s Guide, Edition 5
Page 45
Introduction
Overview
Overview
Agilent B1500A Semiconductor Device Analyzer is the new generation one box
solution for the semiconductor device DC/AC parametric measurement and analysis
application. You can perform measurement and analysis easily and effectively on
the Microsoft Windows environment with intuitive graphical user interface, touch
screen LCD panel, keyboard, and mouse by using Agilent B1500A.
Agilent B1500A provides the DC voltage/current output capability, the DC
ltage/current measurement capability, and the AC signal output and impedance
vo
measurement capability. So you can perform the current-voltage sweep
measurement and the capacitance-voltage sweep measurement for example by one
instrument. You can also analyze the measurement result characteristics graph by
using several tools such as markers, cursors, and lines.
Also, in the GPIB remote control mode, you can control the B1500A from an
t
ernal computer by using Agilent FLEX command set that is the common
ex
language for Agilent semi conductor DC measurement instruments. So you can reuse
the measurement program created for Agilent 4155/4156/E5 260 / E527 0.
Agilent B1500 User’s Guide, Edition 52- 3
Page 46
Introduction
Overview
•Agilent B1500A Semiconductor Device Analyzer
Mainframe that provides the ground unit (GNDU) and ten empty slots for the
measu
rement facilities, and Agilent EasyEXPERT software for the operating
environment. The EasyEXPERT is the GUI based measurement control and
analysis software runs on the Microsoft Windows XP Professional.
The B1500A is equipped with the 15 inch LCD with touch panel, hard disk
ive, DVD-ROM/CD-ROM/CD-RW drive, and USB/LAN/GP-IB interfaces.
dr
The USB keyboard, the USB mouse, and the stylus pen are available as option.
•Agilent B1510A HPSMU mod ul e
High power source/monitor unit module for the B1500A. Occupies two slots.
•Agilent B1511A MPSMU module
Medium power source/monitor unit module for the B1500A. Occupies one slot.
•Agilent B1517A HRSMU module
High resolution source/ mo nit or uni t mo dul e f or the B1500A. Occupies one slot.
Need
s Ag
ilent E5288A ASU to perform atto level current measurement.
•Agilent B1520A MFCMU module
Multi frequency capacitance measurement unit module for the B1500A.
Occupi
es one slot. Needs Agilent N1301A-100 SCUU to switch the CMU/SMU
connection automatically or to use Agilent N1301A-100 GSWU.
•Agilent N1301A-10 0 SCUU
SMU CMU unify unit (SCUU). Used to switch the CMU/SMU connection to
the D
UT interface. Attach or connect to the CMU and two SMUs, MPSMU
and/or HRSMU).
•Agilent N1301A-20 0 GS WU
Guard switch unit (GSWU). Used to connect th
and low near the DUT. The GSWU must be connected to the SCUU and the
guard lines of the DUT interface. Effective for the accurate impedance
measurements.
•Agilent E5288A ASU
Atto sense and switch unit (ASU). Used to enable the 1
level current measurement, also switch the SMU/AUX connection to the DUT
interface. Available for the HRSMU.
2- 4Agilent B1500 User’s Guide, Edition 5
e guard lines between CMU high
pA range for the atto
Page 47
Introduction
Overview
Agilent EasyEXPERT Software
Agilent EasyEXPERT software is a specially-designed Windows XP application
program for controlling Agilent B1500A. The EasyEXPERT provides the easy and
effective measurement and analysis environment with intuitive graphical user
interface (GUI), touch screen LCD panel, keyboard, and mouse. Some of the
functions are listed below.
•Single measurement, repeat measurement, and append measurement
•Switching matrix control (Agilent B2200A/B2201A)
•Measurement/setup data management by workspace
•Graph display and anal ysis with markers,
•Data import/export capability, data output by CSV/XML format, and graph
outp
ut by EMF/BMP/GIF/PNG format
•Maintenance; self-test and self-calibration
The EasyEXPERT has the following measurement execution environments.
•Application test
The EasyEXPERT contains an applicati
measurements for CMOS devices, TFT, B JT, memory, nanotechnology devices
such as CNT FET, and so on. The application library includes more than one
hundred test definitions.
You can perform measurements by choosing the best one for your device under
(DUT) from the application libraries, and modifying and executing it.
test
Modification is really simple. For example, it will be completed by changing the
output voltage only to DUT terminals. And the setup with your modifications
can be saved and recalled as your setup (My Favorite Setup).
•Classic test
You can perform measurements by using the u
semiconductor parameter analyzers such as Agilent 4156C. The setup can be
made by entering values into the setup tables used for the measurement module
control. And it can be saved an d recalled as your setup (My Favorite Setup). The
I-V, C-V, and sampling measurement modes are supported.
cursors, and lines; and auto analysis
on library that supports typical
ser interface similar to the
•Quick test
You can execute the test setups saved in a preset group (My Favo
sequentially.
Agilent B1500 User’s Guide, Edition 52- 5
rite Setup)
Page 48
Introduction
Front View
Front View
This section describes the front view of Agilent B1500A.
1. Standby switch
Turns the B1500A on. Pressing the button in the ON state makes the B1500A in
the s
tandby state. The green LED lights when the B1500A is in the ON state.
NOTEWhen turning the B1500A on
Open the measurement terminals at the device side when turning the B1500A on.
Also disconnect the device from the measurement terminals and open the
measurement terminals after the measurement. If you leave the connection with the
device, the device may be damaged by unexpected operations or charge-up of
measurement cables.
2- 6Agilent B1500 User’s Guide, Edition 5
Page 49
2. LCD panel
Introduction
Front View
15 inch TFT XGA display, 1024 × 768 r
screen, Agilent EasyEXPERT software window, and so on. Touch screen
operation is available when the Touch Pane l Off indicator does not light.
NOTEAbout screen adjustment
If you find swinging or flickering in screen while you use this instrument, perform
the screen adjustment. In most cases, the symptoms can be resolved by pressing the
Auto button below the LCD panel. The Auto button is used to start the automatic
screen adjustment. Also the manual screen adjustment is available by using the
adjustment menu displayed by pressing the Menu button.
3. Touch Panel Off key
Works on Agilent EasyEXPERT environment. Enables or disables the touch
sc
reen operation. The green LED lights when the touch screen is disabled.
NOTETo disable the touch panel
Use T ouch Panel Device Properties, and remo ve the check from the Enabled box on
the Advanced tab screen. The Touch Panel Device Properties dialog box can be
opened by selecting Gunze TPDD > Adjust Settings from Windows program menu.
4. Stop key
Stops the present measurement or source output immediately.
esolution. Displays the Windows
5. High vol t age status indicator
This red LED lights when a source channel applies more than ± 42
6. Measurement status indicator
This green LED lights when a measurement chan
7. Rotary knob
Works on Agilent EasyEXPERT environment. Rotating the knob moves the
mark
er on the graph window, or increases/decreases/changes the value in the
active entry field. Pressing the knob sets or enters the value.
8. Softkeys
Seven softkeys are available for Agilent EasyEXPERT environment. Used to
select an altern
Agilent B1500 User’s Guide, Edition 52- 7
ative for the entry field specified or the dialog box.
nel performs measurement.
V
.
Page 50
Introduction
Front View
9. DVD-ROM/CD-ROM/CD-RW drive
For data backup, B1500A update, or data transfer.
10. USB interfaces
USB, 2 ports. For keyboard, mouse, and so on. USB keyboard and USB mouse
are
available as
option.
11. OSD butto ns
LCD Off enables or disables the LCD pane
l. The green LED lights when the
LCD is disabled.
Four keys are available for the display adju
stment. Press Auto to perform
adjustment automatically. Use -, +, and Menu to adjust manually.
12. HDD access indicator
This green LED lights in the access status of HDD or DVD-ROM/CD-ROM/
CD-RW
drive. Do not turn the B1500A off during this LED lights.
2- 8Agilent B1500 User’s Guide, Edition 5
Page 51
Rear View
This section describes the rear view of Agilent B1500A.
Introduction
Rear View
1. Serial number
You need this se
assistance program.
2. GNDU terminal
0 V constant voltage source. Used for the reference of measurement ground.
Tr
iaxial connector .
3. Zero Check terminal
Ground reference point of the B1500A.
CAUTIONThe Zero Check terminal can be used for the service purpose only. For the B1500
normal operation, leave this terminal open and do not connect anything to this
terminal. Connecting anything can damage the B1500.
ial number when using Agilent Technologies telephone
r
4. Circuit Common () and Frame ground () terminals
For floating measurement, remove the shorting bar (Agilent part number
5000
-4206).
Agilent B1500 User’s Guide, Edition 52- 9
Page 52
Introduction
Rear View
CAUTIONDo not float the Circuit Common terminal at voltages more than ±42 V referenced
to frame ground. Failure to heed this caution may result in damage to Agilent
B1500A.
WARNINGIf the Circuit Common terminal is not connected to the frame ground terminal
(for floating measurement), a potentia l shock hazard may present. Do not
touch any of measurement circuit at any time while a floating measurement is
in progress.
5. SMU terminals
The source/monitor unit (SMU) has two triaxial connecto
the Kelvin connections.
The high power source/monitor unit (HPSMU) occupies two slots. And the
HPSM
Us are allowed to be installed in slots 1 to 2 (one HPSMU), 1 to 4 (two
HPSMUs), 1 to 6 (three HPSMUs), or 1 to 8 (four HPSMUs). In the GPIB
remote control mode, channel number is always lower slot number.
The medium power source/monitor unit (MPSMU) and the high resolution
urce/monitor unit (HRSMU) occupy one slot. In the GPIB remote control
so
mode, channel number is always the slot number that the module is installed.
The HRSMUs are installed between HPSMUs and MPSMUs.
In the EasyEXPERT operation mode, the SMUs
number. SMU1 is assigned to the SMU installed in the lowest module slot. And
the following SMU numbers are assigned to the upper SMUs in sequence.
NOTEThe B1500A shown above installs two HPSMUs in slots 1 to 4, two HRSMUs in
slots 5 to 6, two MPSMUs in slots 7 to 8, and one MFCMU in slot 9. The SMU
numbers 1 to 6 and the channel numbers 1, 3, and 5 to 8 are assigned to the SMUs.
6. CMU terminals
The multi frequency capacitance measurement unit (MFCMU) has four coaxial
connect
The Hcur and Hpot have to be connected together at a terminal of the device
under test (DUT). Also the Lcur a nd Lpot have to be connected together at the
other terminal of the DUT. Then, the guard lines of four-terminal pair must be
connected together.
ors, Lcur, Lpot, Hpot, and Hcur, for the four-terminal pair connection.
can be specified by the SMU
rs, force and sense, for
The MFCMU occupies one slot. And the MFCMU have to be installed in the
l
ot next to the last SMU. In the GPIB remote control mode, channel number is
s
always the slot number and the largest number available fo r the chan nel num ber.
2- 10Agilent B1500 User’s Guide, Edition 5
Page 53
Introduction
7. Blank panel
Blank pane ls are fixed to the empty slots.
8. GPIB interface
Use an Agilent 82357A USB/GPIB interface or Agilent 10833A/B/C/D GPIB
cab
e to connect to an external computer or equipment.
l
9. LINE input receptacle
AC power cable is connected to this receptacle.
10. LAN interface
RJ45 connector.
11. USB interfaces
USB, 2 ports. For keyboard, mouse, or peripherals. USB keyboard and USB
mo
us
e are available as option.
12. Ext Trig terminals
Two BNC connectors, one for trigger input, and one for trigger output. For
de
tails about the trigger function, see Programming Guide.
Rear View
13. Digital I/O terminal
DSUB 25 pin connector. Can be used for the trigger input/o utput t erminal s or an
interf
ace to control an external relay circuit and so on. For details, see
Programming Guide.
14. Interlock terminal
Used in conjunction with the interlock function of the B1500A. If the Interlock
termi
nal is open, maximum SMU output is limited to ± 42 V. Be sure to connect
this terminal to the 16442B test fixture or connector plate before performing
measurement. If you use a connector plate, you must install an interlock circuit.
For details on how to install the interlock circuit, see “To Make an Interlock
Circuit” on page 3-32.
WARNINGDangerous voltage of up to the maximum voltage of SMUs may be present at
force, guard, and sense terminals if the Interlock terminal is closed.
Agilent B1500 User’s Guide, Edition 52- 11
Page 54
Introduction
Measurement Units
Measurement Units
Agilent B1500A can be equipped with the following measurement modules.
•“GNDU - Ground Unit”
•“B1510A HPSMU”
•“B1511A MPSMU”
•“B1517A HRSMU”
•“B1520A MFCMU”
See “Using EasyEXPERT” on page 4-1 for the functions available for Agilent
EasyEXPERT software. See Programming Guide for the functions available for the
GPIB remote mode.
GNDU - Ground Unit
Agilent B1500A is equipped with the ground unit (GNDU). The GNDU is a 0 V
constant voltage source, and used for the reference of measurement ground. Also the
GNDU can sink up to ± 4.2 A, so it is effective for using the HPSMU (high power
SMU). Figure 2-1 is a simplified circuit diagram of GNDU.
Figure 2-1Simplified GNDU Circuit Diagram
9
&RPPRQ
2- 12Agilent B1500 User’s Guide, Edition 5
*1'8
)RUFH
Page 55
Introduction
Measurement Units
About SMU
Source/monitor unit (SMU) can apply DC voltage or current, and can measure DC
current or voltage. Figure 2-2 is a simplified SMU circuit diagram. The SMU can
perform the following operations:
•Apply voltage and measures current
•Apply current and measures voltage
•Apply voltage and measures voltage
•Apply current and measures current
NOTETo select the SMU operation mode in the GPIB remote condition , enter the CMM
command.
Figure 2-2Simplified SMU Circuit Diagram
$
608
)RUFH
9
&RPPRQ
Compliance
The SMU has a compliance feature that limits output voltage or current to prevent
damage to the device under test. When the SMU applies voltage, you can specify
current compliance. When the SMU applies current, you can specify voltage
compliance. For details about the compliance, refer to “SMU Compliance” on page
7-41.
Agilent B1500 User’s Guide, Edition 52- 13
Page 56
Introduction
Measurement Units
B1510A HPSMU
This section describes typical specification of the high power source/monitor unit
(HPSMU) for Agilent B1500A.
•Maximum voltage, current, output power: ± 200 V, ± 1 A, 20 W
•Minimum range: 2 V, 1 nA
•Output/measurement value and resolution: see Table 2-1 to Table 2-4.
Figure 2-3HPSMU Output and Meas u r em ent Ranges
&XUUHQWP$
2- 14Agilent B1500 User’s Guide, Edition 5
9ROWDJH9
Page 57
Table 2-1HPSMU Output Voltage Ranges and Resolu ti on s
Introduction
Measurement Units
Range Output Value
2 V0 ≤ |V| ≤ 2 V100 μV± 1000 mA
20 V0 ≤ |V| ≤ 20 V1 mV± 1000 mA
40 V0 ≤ |V| ≤ 40 V2 mV± 500 mA
100 V0 ≤ |V| ≤ 100 V5 mV± 125 mA
200 V0 ≤ |V| ≤ 200 V10 mV± 50 mA
Table 2-2HPSMU Output Current Ranges and Resolutions
Range Output Value
1 nA0 ≤ |I| ≤ 1.15 nA50 fA± 200 V
10 nA0 ≤ |I| ≤ 11.5 nA500 fA± 200 V
100 nA0 ≤ |I| ≤ 115 nA5 pA± 200 V
1 μA0 ≤ |I| ≤
10 μA0 ≤ |I| ≤ 11.5 μA500 pA± 200 V
1.15 μA50 pA± 200 V
Setting
Res
Setting
Reso
olution
lution
Maximum
Current
Maximum
Voltage
100 μA0 ≤ |I| ≤ 115 μA5 nA± 200 V
1 mA0 ≤ |I| ≤ 1.15 mA50 nA± 200 V
10 mA0 ≤ |I| ≤ 11.5 mA500 nA± 200 V
100 mA0 ≤ |I| ≤ 50 mA5 μA± 200 V
50 mA < |I| ≤ 115 mA5 μA± 100 V
1 A0 ≤ |I| ≤ 50 mA50 μA± 200 V
50 mA < |I| ≤ 125 mA50 μA± 100 V
125 mA < |I| ≤ 500 mA50 μA± 40 V
500 mA < |I|
Agilent B1500 User’s Guide, Edition 52- 15
≤ 1 A50 μA± 20 V
Page 58
Introduction
Measurement Units
Table 2-3HPSMU Measurement Voltage Values and Resolutions
Table 2-12HRSMU Measurement Current Values and Resolutions
Measurement Resolutions
Speed ADCHigh Resolution ADC
Range
b
1 pA
Measurement
a
Value
0 ≤ |I| ≤ 1.15 pA100 aA100 aA
High
10 pA0 ≤ |I| ≤ 11.5 pA1 fA
100 pA0 ≤ |I| ≤ 115 pA5 fA
1 nA0 ≤ |I| ≤ 1.15 nA50 fA10 fA
1 fA / 400 aA
2 fA / 500 aA
b
b
10 nA0 ≤ |I| ≤ 11.5 nA500 fA10 fA
100 nA0 ≤ |I| ≤ 115 nA5 pA100 fA
1 μA0 ≤ |I| ≤ 1.15 μA50 pA1 pA
10 μA0 ≤ |I| ≤ 11.5 μA500 pA10 pA
100 μA0 ≤ |I| ≤ 115 μA5 nA100 pA
1 mA
0 ≤ |I| ≤ 1.15 mA50 nA1 nA
10 mA0 ≤ |I| ≤ 11.5 mA500 nA10 nA
100 mA0 ≤ |I| ≤ 100 mA5 μA100 nA
a. This column is applied to the auto ranging or the limited auto ranging. For
fixed ranging, maximum measurement value is Range column value.
b. Available when the E5288A Atto Sense and
Switch Unit (ASU) is used.
2- 22Agilent B1500 User’s Guide, Edition 5
Page 65
NOTETo connect ASU
The specifications are satisfied and guaranteed for the exclusive combination of the
ASU and the HRSMU. So confirm the serial number of the ASU and connect it to
the dedicated HRSMU properly.
To connect the ASU, see “To Connect ASU” on page 3-21.
To confirm the serial number, see “ASU” on page 4-42.
NOTEBefore using 1 pA range
The measurement channel connected to the ASU suppo rts the 1 pA rang e. To use the
1 pA range, set the 1 pA fixed range or the 1 pA limited auto ranging.
Introduction
Measurement Units
The B1500A automatically performs the compens
1 pA range and returns the compensated data. You can use either the pre-stored
offset data or the pre-measured offset data.
To measure the offset data, perform the self-calibration before starting the
ement for a DUT. This offset data is temporarily memorized until the
measur
B1500A is turned off. See “SMU Calibration” on page 4-33.
ation of the data measured by the
Agilent B1500 User’s Guide, Edition 52- 23
Page 66
Introduction
Measurement Units
B1520A MFCMU
This section describes typical specification of the multi frequency capacitance
measurement unit (MFCMU) for Agilent B1500A. The MFCMU performs the
impedance measurement and returns the specified measurement data such as Cp-G.
•Measurement parameters: See Table 2-14.
All measurement parameters are supported by Agilent B1500A in the GPIB
te mode. However, Agilent EasyEXPERT software supports the Cp-G,
remo
Cp-D, Cs-Rs, and Z-θ measurements only.
•AC signal, output frequency: 1 kHz to 5 MHz
Setting resolution: 1 mHz (1 kHz to), 10 mHz (10 kH z to), 0. 1 H z (100 kH z to),
•DC bias:
0 ≤ DC bias ≤ ± 25 V, 1 mV resolution
± 25 V < DC bias ≤ ± 100 V, 5 mV resolution
DC bias more than ± 25 V is available when using the SMU CMU unify unit
Ls (series inductance, H)Rs (series resistance, Ω)
Ls (series inductance, H)D (dissipation factor)
Ls (series inductance, H)Q (quality factor)
Agilent B1500 User’s Guide, Edition 52- 25
Page 68
Introduction
Specifications
Specifications
This chapter lists specifications and supplemental information for Agilent B1500A
and its associated modules. The specifications are the standards against which the
B1500A and its associated modules are tested. When the B1500A or any of its
associated modules are shipped from the factory, they meet the specifications.
The “supplemental” information and “typical” entries in the following
ecifications are not warranted, but provide useful information about the functions
sp
and performance of the instruments.
Introduction
Agilent B1500A Semiconductor Device Analyzer, with EasyEXPERT software,
makes every user a parametric test expert. The Microsoft Windows based
EasyEXPERT interface is familiar, even to new engineers who are inexperienced
with parametric measurement instruments. Its unique task-based approach enables
the user to focus on their real task-at-hand (device characterization), without having
to become a specialist at using the instrument hardware. This new approach is
enhanced via a touch screen interface, which makes the instrument as easy to use
when racked as when on a benchtop.
The modular 10-slot con figu rat i on helps reduce the cost of test by ensuring that you
rchase only what you need up-front, leaving room to grow as your needs change.
u
p
This flexibility is advantageous for those needing capacitance measurement, as this
function is available via a single-slot capacitance measurement unit (CMU). An
innovative SMU CMU unify unit (SCUU) is available to eliminate cabling
confusion when connecting the SMUs and CMU to your positioner-based wafer
probing environment. This greatly improves efficiency and accuracy by resolving
cable swapping and measurement compensation issues.
ent performance, with available
The B1500A also possesses impressive meas
low-current measurement of 0.1 femtoamp and low-voltage measurement resolu tion
of 0.5 microvolts.
2- 26Agilent B1500 User’s Guide, Edition 5
urem
Page 69
Basic Features
•EasyEXPERT software resident on instrument
•Performs IV and CV measurements
•Ten module slots
•Multiple SMU types available: MPSMU, HPSMU, and HRSMU
Introduction
Specifications
•Multi-frequency capacitance measurement u
•High-resolution, analog-to-digital converter (ADC) available for all SMUs
•High-speed ADC present on each SMU
•4.2-Amp ground unit
•SMU/AUX path switching available on the Atto Sense and Switch Unit (ASU)
•Supported CMU accessories include SMU CMU unify unit (SCUU) and guard
swit
c
h unit (GSWU)
•Automatic identification of capacitance measur
•GPIB port for instrument control
•Self-test, self-calibration, diagnostics
it (1 kHz to 5 MHz) available
n
ement accessories (MFCMU)
Agilent B1500 User’s Guide, Edition 52- 27
Page 70
Introduction
Specifications
Hardware
•“Mainframe”
•“SMU Common Specifications”
•“MPSMU/HRSMU”
•“HPSMU”
•“SMU Supplemental Information”
•“MFCMU”
•“SCUU/GSWU”
•“ASU”
See “Using EasyEXPERT” on page 4-1 for the functions available for Agilent
EasyEXPERT software. See Programming Guide for the functions available for the
GP
IB remote mode.
pecification Conditions
S
The measurement and output accuracies are s
conditions. And the SMU measurement and output accuracies are specified at the
SMU connector terminals when referenced to the Zero Check terminal.
pecified under the following
1. Temperature: 23 ° C ± 5 ° C
2. Humidity: 20 % to 60 %
3. After 40 minutes warm-up
4. Ambient temperature changes less than ± 1 °C after s
not applicable for MFCMU
5. Measurement made within one hour af
not applicable for MFCMU
6. Calibration period: 1 year
7. MFCMU measurement time: 1 PLC
8. SMU integration time setting (high-resolution ADC must be u sed):
1 PLC (1 nA to 1 A range, voltage range)
20 PLC (100 pA range)
50 PLC (1 pA to 10 pA range)
9. SMU Filter: ON (for SMUs)
10. SMU measurement terminal connect
2- 28Agilent B1500 User’s Guide, Edition 5
ter self-calibration execution,
ion: Kelvin connection
elf-calibration execution,
Page 71
Mainframe
The B1500A supports eight slots for plug-in modules.
Introduction
Specifications
Module
Slots
occupi
ed
nge of operation
Ra
B1510A HPSMU2-200 V to 200 V,
-1A
to 1 A
B1511A MPSMU1-100 V to 100 V,
-1
00 mA to 100 mA
B1517A HRSMU1-100 V to 100 V,
-1
00 mA to 100 mA
E5288A ASU
a
--100
V to 100 V,
-1
00 mA to 100 mA
B1520A MFCMU11 kHz to 5 MHz
Measurement
solution
re
2 μV, 10 fA
0.5 μV, 10 f A
0.5 μV, 1 fA
0.5 μV, 0. 1 fA
0.035 fFrms
a. The E5288A Atto Sense and Switch Unit (ASU) is connected to the
B1517A High Resolution SMU (HRSMU).
b. Dispers ion of measurement values when connecting a DUT 10 pF to the
measurement terminals under the measurement condition of frequency
1 MHz, signal level 250 mVac, and measurement time 1 PLC.
•Maximum Output Power
The total power consumption of all modules
cannot exceed 82 W. Under this
rule, the B1500A can contain up to 4 dual-slot SMUs (HPSMUs) and 2
single-slot SMUs (MPSMUs and/or HRSMUs); it can contain up to 10
single-slot SMUs; and it can contain any combination of dual-slot and
single-slot SMUs between these two extremes. One single-slot MFCMU may
take the place of a single-slot SMU in any of these configurations.
b
•Maximum Voltage between Common and Ground
≤ ± 42 V
•Pulse Measurement
Width 500 μs to 2 s
Perio
d 5 ms to 5 s
Period ≥ Width+2 ms (when Width ≤ 100 ms)
Period ≥ Width+10 ms (when Width > 100 ms)
Resolution 100 μs
Agilent B1500 User’s Guide, Edition 52- 29
Page 72
Introduction
Specifications
•Ground Unit (GNDU)
The GNDU is furnished with the mainframe.
Output voltage 0 V ± 100 μV
Maximum sink current 4.2 A
Output terminal / connection Triaxial connector, Kelvin (remote sensing)
Load capacitance 1 μF
Cable r
esistance: innocent
Terminal
Sense
Force
≤ 10 Ω
< 1 Ω (Is ≤ 1.6 A)
< 0.7 Ω (
< 0.35 Ω (2.0 A < Is ≤ 4.
Is is the current being sunk by the GNDU.
•Voltage/Current Compliance (Limiting)
The SMU can limit output voltage or current to prevent damaging the device
under
test.
Voltage Compliance 0 V to ± 100 V (MPSMU, HRSMU)
0 V to ± 200 V (HPSMU)
Current Compliance ± 10 fA to ± 100 mA (HRSMU with ASU)
± 100 fA to ± 100 mA (HRSMU)
± 1 pA to ± 100 mA (MPSMU)
± 1 pA to ± 1 A (HPSMU)
Compliance Accuracy Same as the current or voltage set accuracy.
About Measurement Accuracy
•RF electromagnetic field and SMU measurement accuracy
Voltage and Current measurement accuracy o
electromagnetic field of strength over 3 V/m at frequencies from 80 MHz to
1 GHz. The frequency and degree of affection may vary with the installation
dition of the instrument.
con
•Conducted RF field noise and SMU measurement accuracy
Voltage and Current measurement accuracy of
conducted RF field noi se of streng th over 3 Vrms at frequencies from 1 50 kHz t o
80 MHz. The frequency and degree of affection may vary with the installation
condition of the instrument.
Agilent B1500 User’s Guide, Edition 52- 31
SMUs may be affected by RF
f
SMUs may be affected by
Page 74
Introduction
Specifications
MPSMU/HRSMU
MPSMU: Medium Power Source/Monitor Unit
HRSMU: High Resolution Source/Monitor Unit
•Voltage Range, Resolution, and Accuracy (High resolution ADC)
See Table 2-15.
•Current Range, Resolution, and Accuracy (High resolution ADC)
See Table 2-16. Specified measurement reso
tion is limited by fundamental
lu
noise limits. Minimum displayed resolution is 1 aA at 1 pA range by 6 digits.
Measurem
ents at lower range are affected strongly by vibrations and s hocks. Do
not place the environment of vibrations and shocks during measurements.
•Power Consumption
Voltage source mode:
Voltage Range
Power Consumption
a
0.5 VIc × 20 W
2 VIc × 20 W
5 VIc × 20 W
VIc × 20 W
20
40
VIc × 40 W
100 VIc × 100 W
a. Ic is the current compliance setting.
Current source mode:
V
oltage Compliance
a
Power Consumption
b
Vc ≤ 20 VIo × 20 W
20 V < Vc ≤ 40 VIo × 40 W
V < Vc ≤ 100 VIo × 100 W
40
a. Vc is the voltage compliance setting.
b. Io is the output current.
2- 32Agilent B1500 User’s Guide, Edition 5
Page 75
Table 2-15Voltage Range, Resolution, and Accuracy (High Resolution ADC)
a. ±(% of output/mea su re d va lu e + offset value in mV)
b. 100 mA (Vo ≤ 20 V), 50 mA (20 V < Vo ≤ 40 V), 20 mA (40 V < Vo), Vo is the output voltage.
Table 2-16Current Range, Resolution, and Accuracy (High Resolution ADC)
Range
±1 pA
±10 pA
±100 pA
Force
olution
res
b
1 fA0.1 fA
c
5 fA
c
5 fA
Measurement
resolution
0.4 fAb / 1 fA±(0.46+3E
b
/ 2 fA±(0.3+1E
0.5 fA
±1 nA50 fA10 fA
±10 nA500 fA10 fA
±100 nA5 pA100 fA
±1 μA50 pA1 pA
±10 μA500 pA10 pA
±100 μA5 nA100 pA
±1 mA50 nA1 nA
±10 mA500 nA10 nA
±100 mA5 μA100 nA
Force accuracy ±(%+A
-15
±(0.9+15E
-13
±(0.1+3E
±(0.1+3E
-13
-12
±(0.05+3E
±(0.05+3E
±(0.05+3E
±(0.035+15E
±(0.04+15E
±(0.04+15E
±(0.045+15E
)±(0.9+12E
-14
+Vo×1E
+Vo×1E
+Vo× 1E
+Vo× 1E
-11
+Vo× 1E
-10
+Vo× 1E
-9
+Vo× 1E
-9
+Vo× 1E
-8
+Vo× 1E-9)±(0.03+6E-8+Vo× 1E-9)
-7
+Vo× 1E-8)±(0.03+2E-7+Vo× 1E-8)
-6
+Vo× 1E-7)±(0.04+6E-6+Vo× 1E-7)
+A)
-17
-16
-15
-14
-13
-12
-11
-10
Measurement accuracy
a
±(%+A
)±(0.46+15E
)±(0.3+3E
)±(0.1+2E
)±(0.1+1E
)±(0.05+2E
)±(0.05+1E
-15
-14
+Vo× 1E
-13
+Vo× 1E
-12
+Vo× 1E
-11
+Vo× 1E
-10
+Vo× 1E
a
+A)
-15
)
+Vo× 1E
)±(0.04+2E-9+Vo× 1E
)±(0.03+3E-9+Vo× 1E
-16
-15
-14
-13
-12
-11
-10
Maximum
-17
)
)
)
)
)
)
)
)
current
b
Maximum
voltage
100 V
100 V
100 V
100 V
100 V
100 V
100 V
100 V
100 V
100 V
100 V
d
a. ±(% of output/measured value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.
b. Available when Agilent E5288A Atto Sense and Switch Unit (ASU) is used.
c. Available for HRSMU.
d. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io), Io is the output current.
Agilent B1500 User’s Guide, Edition 52- 33
Page 76
Introduction
Specifications
HPSMU
HPSMU: High Power Source/Monitor Unit
•Voltage Range, Resolution, and Accuracy
See Table 2-17.
•Current Range, Resolution, and Accuracy
See Table 2-18.
•Power Consumption
Voltage source mode:
Voltage Rang e
Power Consumption
2 VIc × 20 W
VIc × 20 W
20
40 VIc × 40 W
10
0 VIc × 100 W
20
0 VIc × 200 W
a. Ic is the current compliance setting.
Current source mode:
Voltage Compliance
a
Power Consumption
Vc ≤ 20 VIo × 20 W
20 V < Vc ≤ 40 VIo × 40 W
40
V < Vc ≤ 100 VIo × 100 W
1
00 V < Vc ≤ 200 VIo × 200 W
a. Vc is the voltage compliance setting.
b. Io is the output current.
a
b
2- 34Agilent B1500 User’s Guide, Edition 5
Page 77
Table 2-17Voltage Range, Resolution, and Accuracy (High Resolution ADC)
Introduction
Specifications
Range
Force
resolution
Measurement
resolution
±2 V100 μV2 μV±(
Force accuracy
V)
a
±(%+m
0.018 + 0.4)±
Measurement accuracy
±(%+mV)
a
(0.01 + 0.14)1 A
±20 V1 mV20 μV±(0.018 + 3)±(0.009 + 0.9)1 A
±40 V2 mV40 μV±(0.018 + 6)±(0.01 + 1)500 mA
±100 V5 mV100 μV±(0.018 + 15)±(0.012 + 2.5)125 mA
±200 V10 mV200 μV±(0.018 + 30)±(0.014 + 2. 8)50 mA
a. ±(% of output/mea sured value + offset value in mV)
Table 2-18Current Range, Resolution, and Accuracy (High Resolution ADC)
Range
res
Force
olution
Measurement
resolution
±1 nA50 fA10 fA
±10 nA500 fA10 fA
±100 nA5 pA100 fA
±1 μA50 pA1 pA
±10 μA500 pA10 pA
Force accuracy ±(%+A+A
±(0.1+3E
±(0.1+3E
±(0.05+3E
±(0.05+3E
-13
+Vo× 1E
-12
+Vo× 1E
-11
-10
±(0.05+3E-9+Vo× 1E
+Vo×1E
+Vo×1E
-15
-14
-13
-12
-11
)±(0.1+3E
)±(0.1+25E
)±(0.04+2E-9+Vo× 1E
Measurement accuracy
a
)
±(%+A+A)
)±(0.05+25E
)±(0.05+1E
-13
+Vo× 1E
-13
+Vo× 1E
-12
-10
+Vo× 1E
a
+Vo× 1E
-15
-11
-14
-12
-13
)
)
)
)
)
Maximum
current
Maximum
voltage
200 V
200 V
200 V
200 V
200 V
±100 μA5 nA100 pA
±1 mA50 nA1 nA
±10 mA500 nA10 nA
±100 mA5 μA100 nA
±1 A50 μA1 μA
-9
±(0.035+15E
±(0.04+15E
+Vo×1E
-8
+Vo× 1E-9)±(0.03+6E-8+Vo× 1E-9)
±(0.04+15E-7+Vo×1E-8)±(0.03+2E-7+Vo× 1E-8)
±(0.045+15E
±(0.4+3E
-6
-4
+Vo×1E-6)±(0.4+15E-5+Vo× 1E-6)
-10
)±(0.03+3E-9+Vo× 1E
-10
+Vo×1E-7)±(0.04+6E-6+Vo× 1E-7)
)
a. ±(% of output/measured value + fixed offset in A + proportional offset in A), Vo is the output
voltage in V.
b. 200 V (Io ≤ 50 mA), 100 V (50 mA < Io ≤ 125 mA), 40 V (125 mA < Io ≤ 500 mA),
20 V (500 mA < Io ≤ 1 A), Io is the output current.
•Trigger output d elay (pulsed measureme nts)
0 to 32.7 ms with 100 μs r
HRSMU)
e
0 μA)
solution
Trigger output delay must be less than pulse width.
•Vo ltage Measurement Range, Resolution, an
See Table 2-19.
•Current Measurement Range, Resolution, and Accuracy (High speed ADC)
See Table 2-20.
Specified measurement resolution is limited by fundamental noise limits.
Mi
mum displayed resolution is 1 aA at 1 pA range by 6 digits.
ni
Measurements at lower range are affected stro ngly by vibrations and s hocks. Do
no
t place the environment of vibrations and shocks during measurements.
Agilent B1500 User’s Guide, Edition 52- 37
d Accuracy (High speed ADC)
Page 80
Introduction
Specifications
Table 2-19Voltage Measurement Range, Resolution, and Accuracy (High Speed ADC),
supplemental information
Range
±0.5 V
Measurement
b
olution
res
25 μV±(0. 01 + 0.25)±40 V2 mV±(0.015 + 8)
±2 V100 μV±(0.01 + 0.7)±10
b
±5 V
0 V1 mV±(0.01 + 4)
±2
250 μV±(0.01 + 2)
Measure m ent a ccu ra cy
±(%+mV)
a
Range
Meas
0 V5 mV±(0.02 + 20)
c
±200 V
a. ±(% of reading + offset value in mV). Averaging is 128 samples in 1 PLC
b. Available for MPSMU/HRSMU.
c. Available for HPSMU.
Table 2-20Current Measurement Range, Resolution, an
supplemental information
nge
Ra
±1 pA
±10 pA
±100 pA
Measurement
res
b
c
c
olution
0.1 fA
1 fA
5 fA
Measurement accuracy
-15
+Vo× 1E
-14
+Vo× 1E
-15
a
)
-17
-16
±(
%+A+A)
±(1.8+12E
±(0.5+15E
±(0.3+3E
Range
±10 μA500 pA
±100 μA5 nA
)
±1 mA50 nA
)
Measu
resolution
urement
resolution
Measurement accuracy
±(%+mV)
a
10 mV±(0.035 + 40)
d Accuracy (High Speed ADC),
rement
Measurement accuracy
±(%+A+A)
a
±(0.04+2E-9+Vo× 1E
±(0.03+2E-8+Vo× 1E
±(0.03+2E-7+Vo× 1E-9)
-11
-10
)
)
±1 nA50 fA
±10 nA500 fA
±100 nA5 pA
±1 μA50 pA
.1+3E
±(0
±(0.1+2E
0.05+2E
±(
0.05+2E
±(
-13
+Vo× 1E
-12
+Vo× 1E
-11
-10
+Vo× 1E
+Vo× 1E
-15
)
-14
)
-13
)±1 A
-12
)
±10 mA500 nA
±100 mA5 μA
d
50 μA
a. ±(% of reading + fixed offset in A + proportional offset in A), Vo is the output volta ge in V.
b. Availa bl e when Agil ent E52 88A Atto Sen se and Sw itch Uni t (ASU) is used .
c. Available for HRSMU.
d. Available for HPSMU.
2- 38Agilent B1500 User’s Guide, Edition 5
±(0.03+2E-6+Vo× 1E-8)
±(0.04+2E-5+Vo× 1E-7)
±(0.4+3E-4+Vo× 1E-6)
Page 81
Introduction
Specifications
MFCMU
MFCMU: Multi Frequency Capacitance Measurement Unit
0.0 % + 1 mVrms) at four-terminal pair port of MFCMU
±(15.0 % + 1 mVrms) at measurement port of MFCMU cable (1.5 m or 3.0 m)
•
Output imp edance: 50 Ω, t
•Signal level monitor
ypical
Voltage range: Same as the voltage signal level setting range.
Voltage monitor accuracy (open load):
±(1
0.0 % of reading + 1 mVrms) at four-terminal pair port of MFCMU
±(15.0 % of reading + 1 mVrms) at measurement port of MFCMU cable (1.5 m
o
r 3.0 m)
Agilent B1500 User’s Guide, Edition 52- 39
Page 82
Introduction
Specifications
DC Bias Function.
•DC voltage bias
Range: 0 to ± 25 V
R
esolution: 1 mV
Accuracy (open load): ±(0.5 %
MFCMU measurement port or the MFCMU 1.5 m/3 m cable end
Maximum DC bias current (supplemental information):
Impedance measurement rangeMaximum DC bias current
50 Ω10 mA
100 Ω10 mA
300 Ω10 mA
1 kΩ1 mA
3 kΩ1 mA
10 kΩ100 μA
30 kΩ100 μA
100 kΩ10 μA
300 kΩ10 μA
•Output impedance 50 Ω, typi
•DC bias monitor
+ 5.0 mV) at the High and Low terminals of the
cal
Range: Same as the voltage signal level setting range.
Accuracy (open load): ±(0.2
terminals of the MFCMU measurement port or the MFCMU 1.5 m/3 m cable
end
reading + 10 mV) at the High and Low
% of
Sweep Characteristics.
•Available sweep parameters: Oscillator level, DC bias voltage, frequency
•Sweep type: Linear, Log
•Sweep mode: Single, Double
2- 40Agilent B1500 User’s Guide, Edition 5
Page 83
Introduction
Specifications
•Sweep direction: Up, Down
•Number of measurement points: maximum 1001
Measurement Accuracy.
The following parameters are used to express the impedance measur ement accuracy
at four-terminal pair port of MFCMU and measurement port of MFCMU cable.
Zx: Impedance measurement value (Ω)
Dx: Measurement value of D
E = E
’ + (ZS’ / |Zx| + YO’× |Zx| ) × 100 (%)
P
E
’ = EPL + E
P
Y
’ = YOL + Y
O
Z
’ = ZSL + Z
S
For the parameter value, see Table 2-21 to Table 2-23. Calculation example of the
measurement accuracy is shown in Table 2-24.
POSC
OSC
OSC
+ EP (%)
+ YO (S)
+ ZS (Ω)
•|Z| accuracy: ± E (%
•θ accuracy: ± E/
)
100 (radian)
•C accuracy:
± E (%) at Dx ≤ 0.1
± E
× (1+Dx
2)1/2
(%) at Dx > 0.1
•D accuracy:
± E/100 at Dx ≤ 0.
1
± E× (1+Dx)/100 at 0.1 < D x ≤ 1
•
G accuracy:
± E/Dx (%) at Dx ≤ 0.1
± E
× (1+Dx
2)1/2
/Dx (%) at Dx > 0.1
Agilent B1500 User’s Guide, Edition 52- 41
Page 84
Introduction
Specifications
Table 2-21Parameters E
Oscillator level
Vosc mV
125 < Vosc ≤ 250 0.03×(250/Vosc-1)5× (250/Vosc-1)
64
< Vosc ≤ 125 0.03× (125/Vosc-1)5× (125/Vosc-1)
32
< Vosc ≤ 64 0.03× (64/Vosc-1)5× (64/Vosc-1)
Vo
sc ≤ 32 0.03× (32/Vosc-1)5× (64/Vosc-1)
Vo sc is oscillator level in mV.
Table 2-22Parameters E
Cable length
1.5 m0.02+3× f/10075
3 m0.02+5×f/10015
f is frequency in MHz. If measurement cable is extended, open compensation, short
compensation, and load compensation must be performed.
Table 2-23Parameters Y
, Z
POSC
, YOL, Z
PL
, YO, EP, Z
OSC
OSC
%Z
E
POSC
SL
E
%YOL nSZSL mΩ
PL
0× f/1005.0
00× f/1005.0
S
OSC
mΩ
nSYO nSEP %ZS mΩ
Frequency f Hz
1k ≤ f ≤ 200k1×(125/Vosc-0.
200k < f ≤ 1M2× (125/Vosc-0.
1M < f ≤ 2M2× (125/Vosc-0.
2M < f20×(125/Vosc-0.
Y
OSC
5)1.50.0955.0
5)3.00.0955.0
5)3.00.285.0
5)30.00.285.0
Vo sc is oscillator level in mV.
2- 42Agilent B1500 User’s Guide, Edition 5
Page 85
Table 2-24Example of Calculated C/G Measurement Accuracy
Introduction
Specifications
Frequency
5 MHz10 pF≤ 31 μS± 0
Hz10 pF≤ 6 μS± 0.11 %± 68 nS
1 M
100 kHz10 pF≤ 628 nS± 0.18 %± 11 nS
10 kHz100 pF≤ 628 nS± 0.18 %± 11 nS
Measured
citance
Capa
10
0 pF≤ 314 μS± 0.29 %± 9
1 nF≤ 3 mS± 0.32 %± 99 μS
0 pF≤ 63 μS± 0.10 %± 620 nS
10
F≤ 628 μS± 0.10 %± 7
1 n
0 pF≤ 6 μS± 0.11 %± 66 nS
10
F≤ 63 μS± 0.10 %± 619 nS
1 n
nF≤ 628 μS± 0.10 %± 7
10
F≤ 6 μS± 0.11 %± 66 n S
1 n
nF≤ 63 μS± 0.10 %± 619 nS
10
0 nF≤ 628 μS
10
Measured
Conductance
C AccuracyG Accuracy
.32 %± 990 nS
± 0.10 %± 7 μS
μS
μS
μS
1 kHz100 pF≤ 63 nS± 0.92 %± 6 nS
F≤ 628 nS± 0.18 %± 11 nS
1 n
nF≤ 6 μS± 0.11 %± 66 nS
10
0 nF≤ 63 μS± 0.10 %± 619 nS
10
The capacitance and conductance measurement accuracy
following conditions:
D ≤ 0.1
Test signal level: 3 0 mVrms
Cable length: 0 m
Agilent B1500 User’s Guide, Edition 52- 43
is specified under the
Page 86
Introduction
Specifications
SCUU/GSWU
SCUU: SMU CMU Unify Unit
GSWU: Guard Switch Unit
The SCUU switches either two SMUs or CMU to the output port. The MPSMU and
HR
SMU are supported as a module. The SCUU has DC bias adapter funct ion by the
SMU as DC bias source. Also error compensation function is supported. To connect
the SCUU, the dedicated cable assembly is supplied. I t can help to co nnect corr ectly
and easily, and it eliminates cabling mistake. Turn the B1500A off before
connecting the SCUU.
The GSWU switches open when using SMUs and close when using CMU for the
u
ard return path automatically depending on which the SCUU output is used for
g
the capacitance measurement or not.
SCUU.
•Inputs
Four triaxial ports: for Force1, Sense1, Force2, and Sense2
Four BNC ports: for MFCMU
One SCUU control port
•Outputs
Four triaxial ports: for Force1/CMUH
One GSWU control port
Three LEDs, SMU/CMU output status indicator
•Docking mode
Direct and indirect mode
,
Sense1, Force2/CMUL, and Sense2
GSWU.
•Input
One control port from SCUU
Mini pin plug port: 2 ports (to Guard1, to Guard2)
•Maximum Voltage
100 V (AUX input to AUX common)
100 V (AUX input to circuit common)
42 V (AUX common to circuit common)
•Maximum Current
0.5 A (AUX input to Force output)
ASU Supplemental Information.
•Band width (at -3 dB):
< 30 MHz (AUX port)
NOTEConnect ASU to dedicated HRSMU
The specifications are satisfied and guaranteed for the exclusive combination of the
ASU and the HRSMU. So confirm the serial number of the ASU and connect it to
the dedicated HRSMU properly.
NOTEConnection path control
Disable source output of the SMU and other instrument connected to the ASU
inputs before controlling the ASU connection path.
2- 48Agilent B1500 User’s Guide, Edition 5
Page 91
Agilent EasyEXPERT Software
Function
•Operation mode
•Application test mode
•Classic test mode
•Key functions
Introduction
Specifications
•Categorized and predefin
•Device definition
•Measurement parameter settings
•Save/recall My Favorite setups
•Define/customize application library
•Execute measurement (single/repeat)
•Save/recall measurement data and settings
•Test result da ta m anagement
•Import/export device definition, measurement settings, my favorite setup,
meas
ement data, and application library
ur
•Graph plot display/analysis/printing
•Workspace management
•Self-test, self calibration, diagnostics
application library
ed
Application Library
Sample test definitions for the following applications. They subject to change
without notice.
Generic, Structure, CMOS, Bipolar, Memory, Mixed Signal Device, Analog RF,
Optical, TFT
, Discrete, Reliability, Nanotechnology
Measurement Mode Details
Agilent B1500A supports the following measurement modes:
•Staircase Sweep
Agilent B1500 User’s Guide, Edition 52- 49
Page 92
Introduction
Specifications
•Multi-Channel Sweep (EasyEXPERT supports VAR1 and VAR1’)
•Pulsed Sweep
•Staircase Sweep with Pulsed Bias
•Sampling
•C-V Sweep
Each SMU can sweep using VAR1 (primary sweep), VAR 2 (secondary sweep), or
V
A
R1’ (synchronous sweep).
•VAR1
Primary sweep controls the staircase (DC o
•VAR2
Subordinate linear staircase or linear puls
completed, the VAR2 unit output value is changed.
Maximum nu mber of steps: 128
•VAR1’
Staircase or pulse sweep synchronized with the VAR1 sweep. Sweep is made
with a u
VAR 1 ’ = a × VAR
“a” is the user specified ratio, and “b” i
•CONST
A source unit can be set as a constant vo ltage or cu rrent sourc e depend ing on the
un
er specified ratio and offset values.
s
1 + b
it.
r pulsed) voltage or current sweep.
ed sweep. After primary sweep is
s the user specified offset value.
Staircase Sweep Measurement Mode.
Forces swept voltage or current, and measures DC voltage or current. One channel
can sweep current or voltage while up to ten channels can measure current or
voltage. A second channel can be synchronized with the primary sweep channel as
an additional voltage or current sweep source.
Number of steps: 1 to 1001
Sweep type: Linear or logarithmic
Sweep direction: Single or double sweep
Hold time: 0 to 655.35 s, 10 ms resolution
2- 50Agilent B1500 User’s Guide, Edition 5
Page 93
Introduction
Specifications
Delay time: 0 to 65.535 s, 100 μs resolution
Pulsed Sweep Measurement Mode.
Forces pulsed swept voltage or current, and measures DC voltage or current. A
second channel can be programmed to output a staircase sweep voltage or current
synchronized with the pulsed sweep output.
Staircase Sweep with Pulsed Bias Measurement Mode.
Forces swept voltage or current, and measures DC voltage or current. A second
channel can be programmed to output a pulsed bias voltage or current. A third
channel can be synchronized with the primary sweep channel as an additional
voltage or current sweep source.
Sampling (Time Domain) Measurement Mode.
Displays the time sampled voltage/current data (by SMU) versus time.
Sampling points: 1 to 10,001
Sampling mode: linear
Sampling interval range: 2 ms to 65.535 s, 1 ms resolution
Hold time, initial wait time: 0 s to 655.35 s, 10 ms resolution
Measurement time resolution: 100 μs
Search Measurement Mode.
Forces and measures voltage or current by using linear search method or binary
search method.
Timestamp.
The B1500A supports a time stamp function utilizing an internal quartz clock.
Resolution: 100 μs
Other Measurement Characteristics.
•Measurement Control:
Single, repeat, and stop
•SMU Setting Capabilities:
Limited auto ranging, voltage/current compliance, power compliance, automatic
sweep
Agilent B1500 User’s Guide, Edition 52- 51
ort functions, self-test, and self-calibration
ab
Page 94
Introduction
Specifications
Arithmetic and Analysis Functions
•User Functions
Up to 20 user-defined functions can be defined using arithmetic expressions.
Measured data and pre-defined variables can be used in the computation. The
(con
(integration), LGT (logarithm, base 10), LOG (logarithm, base e), MAVG
(moving average), MAX, MIN, SQRT, trigonometric function, inverse
trigonometric function, and so on
•Physical Constants
Keyboard constants are stored in memory as follows:
•ε (e): Dielectric constant of vacuum, 8.854188E-12
•
Engineering Units
The following unit symbols are also available on the keyboard: a (10
f (10
T (10
-15
12
-12
), p (10
) , P (1015)
), n (10-9), u or μ (10-6), m (10-3), k (103), M (106), G (109),
-18
),
Analysis Capabilities
•Overlay Graph Comparison
A graphics plot can be stored and overlaid.
•Scale
Auto scale and zoom
•Marker
Marker to min/max, interpolation, d
•Cursor
Direct cursor
•Line
2- 52Agilent B1500 User’s Guide, Edition 5
i
rect marker, and marker skip
Page 95
Two lines, normal mode, grad mode, tangent mode, and regression mode
•Auto Analysis Function
Introduction
Specifications
On a graphics plot, the markers and lines can be auto
auto analysis setup. Parameters can be automatically determined using
automatic analysis, user function, and read out functions.
•Data Variable Display
Up to 20 user-defined parameters can be displayed on the graphics screen.
•Analysis Functions
Up to 20 user-defined analysis functions can be defined using arithmetic
ex
sions.
pres
Measured data, pre-defined variables, and read out functions can be used in the
comput
•Read Out Functions
The read out functions are built-in functions for reading various values related to
the marker
ation. The results can be displayed on the LCD.
, cursor, or line.
matically located using the
Graph Plot
•Display Mode
Data display window can be printed. Only X-Y graph can be printed.
•Graph Plot File
Graph plot can be stored as image data to clip board or mass storage device.
File type: bmp, gif, png, emf
Output
•Display Mode
Graph display, list display, and parameter display
•X-Y Graph Display
X-axis and up to eight Y-axis. Linear an
•List Display
Measurement data and calculated user function data are listed in conjunction
with V
AR1 step number or time domain sampling step number. Up to 20 data
sets can be displayed.
Agilent B1500 User’s Guide, Edition 52- 53
g scale. Real time graph plotting.
d lo
Page 96
Introduction
Specifications
Other Functions
•Import/Export files
File type: Agilent EasyEXPERT format, XML-SS format, CSV format
•Data Storage
Hard disk drive, DVD-ROM/CD-ROM/CD-RW drive
•Interfaces
GPIB port for instrument control, Interlock port, USB port (USB 2.0, front 2,
rea
r
2), LAN interface (100BASE-TX/10BASE-T), Trigger in/out, Generic
purpose digital I/O
•Trigger I/O
Only for GPIB remo te mode.
Trigger in/ou t synchronizatio n pulses bef ore and after setting and measuring DC
vo
ltage and current. Arbitrary trigger events can be masked or activated
independently.
Attached Software
A VXIplug&play d river for the B1500A is supplied .
Supported operating systems:
Microsoft Windows 95/NT/2000 Professional/XP Professional
2- 54Agilent B1500 User’s Guide, Edition 5
Page 97
General Specifications
•Temperature Range
Operating +5 °C to +40 ° C
Introduction
Specifications
torage -20 ° C to
S
•Humidity
Operating 20 % to 70 %RH, non-condensing
Storage 20 % to 90 %RH, at +40 ° C, n
•Altitude
Operating 0 to 2,000 m (6,561 ft.)
Storage 0 to 4,600 m (15,092 ft.)
•Power Requirement: 90 to 264 V, 47 to 63 Hz
•Maximum Volt-Amps (VA):
900 VA
•Regulatory Compliance
EMC IEC 61326-1:+A1 / EN 61326-1:+A1
A 1.5 kg, E5288A 0.5 kg, N1301A-100 0.8 kg, N1301A-200 0.1 kg
Agilent B1500 User’s Guide, Edition 52- 55
mm (H) × 575 mm (D)
H) × 70 mm (D)
mm (H) × 32.8 mm (D)
Page 98
Introduction
Accessories and Options
Accessories and Options
Agilent B1500A is furnished with the following accessories.
•Manual CD-ROM, 1 ea.
•Software CD-ROM, 1 ea.
•Desktop EasyEXPERT (CD-ROM), 1 ea.
•Power Cable, 1 ea.
•Label, SMU Number, 1 sheet
Table 2-25 lists the options and the available accessories for Agilent B1500A.
2- 56Agilent B1500 User’s Guide, Edition 5
Page 99
Table 2-25Options and Accessories
Introduction
Accessories and Options
Model
Number
B1500ASemiconductor Device Analyzer
B1510A
B1511AMedium Power Source/Monitor Unit module (MPSMU)
B1517AHigh Resolution Source/Monitor Unit module (HRSMU)
B1520AMulti Frequency Capacitance Measurement Unit module (MFCMU)
B1540AAgilent EasyEXPERT software with application libraries
E5288AAtto Sense and Switch Unit (ASU)
tion Item Description
Op
B1500A-05050 Hz power line frequency
B1500A-06060 Hz power line frequency
B1500A-A6JANSI Z540 compliant calibration
B1500A-UK6Commercial cal. certificate w/ test data
B1500A-ABAPaper manual se t, English
B1500A-ABJPaper manual set, Japanese
High Power Source/Monitor Unit module (HPSMU)
B1540A-001EasyEXPERT software with general application libraries
E5288A-001Add triaxial and D-sub cables, 1.5 m
E5288A-002Add triaxial and D-sub cables, 3 m
N1254A-100GNDU to Kelvin Adapter
N1254A-107Triax(m)-Triax(f) Adapter
N1254A-108AS U Magnetic Stand
N1300ACMU cable
N1300A-0011.5 m length
N1300A-0023 m length
Agilent B1500 User’s Guide, Edition 52- 57
Page 100
Introduction
Accessories and Options
Model
mber
Nu
N1301ACMU accessories
16435AInterlock cable adapter (B1500 type to BNC)
16442BTest fixture
O
ption Item Description
N1301A-100S M U CMU Un ify Unit (SCUU)
N1301A-102S CUU cable, 3 m
N1301A-110SCUU magnetic stand
N1301A-200Guard Switch Unit (GSWU)
N1301A-201GSWU cable, 1.5 m
N1301A-202GSWU cable, 3 m
16442B-010Add triaxial cables, 1.5 m, 4 ea.
16442B-011Add triaxial cables, 3 m, 4 ea.
16442B-800Ext ra blank tef l on board
16442B-801Universal socket module, 0.1 inch pitch, with 10 pins
16442B-802Universal socket module, 0.075 inch pitch, with 10 pins
16442B-803Universal socket module, 0.05 inch pitch, with 10 pins
16442B-810Extra pin set (for universal socket module, 10 pins.)
16442B-811Extra wire set (mini banana to pin plug, 6 ea.)
16442B-812Extra wire set (pin plug to pin plug, 6 ea.)
16442B-813Extra wire set (mini banana to mini clip, 6 ea.)
16442B-814Extra wire set (mini banana to mini banana, 6 ea.)
16442B-821Socket module, 4-pin TO package
16442B-822Socket module, 18-pin DIP package
16442B-823Extra socket module, 2
16442B-890Extra accessory case