Agilent 8960 Getting Started Guide

Page 1
Agilent Technologies 8960 Series 10 Wireless Communications Test Set
Manual Operation: Getting Started Guide
AMPS/136 Mobile Test Application E1961A Revision: A.05
cdma2000 Mobile Test Application E1962B Revision: B.02
GPRS Mobile Test Application E1964A Revision: A.02
GSM Mobile Test Application E1960A Revision: A.08
GSM_AMPS/136_GPRS Mobile Test Application (fast switching) E1985A Revision: A.02
Print Date: September 2001
© Copyright Agilent Technologies 2001
www.agilent.com/find/8960support/
s1
Page 2
Edition/Print Date
All Editions and Updates of this manual and their creation dates are listed below.
March 2001 - 5967-5124
July 2001 - 5967-5156
September 2001 - 5967-5172
Safety Summary
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies Inc. assumes no liability for the customer’s failure to comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs as per IEC 60825-1.
This product has been designed and tested in accordance with IEC Publication 1010, "Safety Requirements for Electronic Measuring Apparatus," and has been supplied in a safe condition. This instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
2
S:\Hp8960\Generic Documents\Manual Operation Getting S tarted Guide\Pi_manual operation gett ing started guide\chapters\titlepage.fm
Page 3
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
Ventilation Requirements: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by
4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety precautions are taken. Note the instrument's external markings described under Safety Symbols.
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. To do so could cause a shock or fire hazard.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE INSTRUMENT COVER
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\titlepage.fm
3
Page 4
WARNING The WARNING sign denotes a hazard. It calls attention to a
procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
CAUTION The CAUTION sign denotes a hazard. It calls attention to an operating
procedure, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.
Safety Symbols
Caution, refer to accompanying documents
Warning, risk of electric shock
Earth (ground) terminal
Alternating current
Frame or chassis terminal
Standby (supply). Units with this symbol are not completely disconnected from ac mains when this switch is off.
To completely disconnect the unit from ac mains, either disconnect the power cord, or have a qualified electrician install an external switch.
4
S:\Hp8960\Generic Documents\Manual Operation Getting S tarted Guide\Pi_manual operation gett ing started guide\chapters\titlepage.fm
Page 5
Product Markings
CE - the CE mark is a registered trademark of the European Community. A CE mark accompanied by a year indicated the year the design was proven.
CSA - the CSA mark is a registered trademark of the Canadian Standards Association.
CERTIFICATION
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\titlepage.fm
5
Page 6
Agilent Technologies Warranty Statement for Commercial Products
Agilent Technologies 8960 Series 10 Wireless Communications Test Set, AMPS/136 Mobile Test Application E1961A cdma2000 Mobile Test Application E1962B GPRS Mobile Test Application E1964A GSM Mobile Test Application E1960A IS-2000 Mobile Test Application E1962A
Duration of Warranty: 1 year
1. Agilent Technologies warrants Agilent Technologies hardware, accessories and supplies against defects in materials and workmanship for the period specified above. If Agilent Technologies receives notice of such defects during the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
2. Agilent Technologies warrants that Agilent Technologies software will not fail to execute its programming instructions, for the period specified above, due to defects in material and workmanship when properly installed and used. If Agilent Technologies receives notice of such defects during the warranty period, Agilent Technologies will replace software media which does not execute its programming instructions due to such defects.
3. Agilent Technologies does not warrant that the operation of Agilent Technologies products will be uninterrupted or error free. If Agilent Technologies is unable, within a reasonable time, to repair or replace any product to a condition as warranted, customer will be entitled to a refund of the purchase price upon prompt return of the product.
4. Agilent Technologies products may contain remanufactured parts equivalent to new in performance or may have been subject to incidental use.
5. The warranty period begins on the date of delivery or on the date of installation if installed by Agilent Technologies. If customer schedules or delays Agilent Technologies installation more than 30 days after delivery, warranty begins on the 31st day from delivery.
6
S:\Hp8960\Generic Documents\Manual Operation Getting S tarted Guide\Pi_manual operation gett ing started guide\chapters\titlepage.fm
Page 7
6. Warranty does not apply to defects resulting from (a) improper or inadequate maintenance or calibration, (b) software, interfacing, parts or supplies not supplied by Agilent Technologies, (c) unauthorized modification or misuse, (d) operation outside of the published environmental specifications for the product, or (e) improper site preparation or maintenance.
7. TO THE EXTENT ALLOWED BY LOCAL LAW, THE ABOVE WARRANTIES ARE EXCLUSIVE AND NO OTHER WARRANTY OR CONDITION, WHETHER WRITTEN OR ORAL IS EXPRESSED OR IMPLIED AND AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDITIONS OR MERCHANTABILITY, SATISFACTORY QUALITY, AND FITNESS FOR A PARTICULAR PURPOSE.
8. Agilent Technologies will be liable for damage to tangible property per incident up to the greater of $300,000 or the actual amount paid for the product that is the subject of the claim, and for damages for bodily injury or death, to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective Agilent Technologies product.
9. TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES IN THIS WARRANTY STATEMENT ARE CUSTOMERS SOLE AND EXCLUSIVE REMEDIES. EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL AGILENT TECHNOLOGIES OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL (INCLUDING LOST PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN CONTRACT, TORT, OR OTHERWISE.
FOR CONSUMER TRANSACTIONS IN AUSTRALIA AND NEW ZEALAND: THE WARRANTY TERMS CONTAINED IN THIS STATEMENT, EXCEPT TO THE EXTENT LAWFULLY PERMITTED, DO NOT EXCLUDE RESTRICT OR MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU.
ASSISTANCE
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\titlepage.fm
7
Page 8
Service and Support
Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Service Center. You can find a list of local service representatives on the Web at:
http://www.agilent-tech.com/services/English/index.html
If you do not have access to the Internet, one of these centers can direct you to your nearest representative:
United States Test and Measurement Call Center
(Toll free in US)
(800) 452-4844
Europe
(31 20) 547 9900
Canada
(905) 206-4725
Japan Measurement Assistance Center
(81) 426 56 7832 (81) 426 56 7840 (FAX)
Latin America
(305) 267 4288 (FAX)
Australia/New Zealand
1 800 629 485 (Australia) 0800 738 378 (New Zealand)
Asia-Pacific
(852) 2599 7777 (852) 2506 9285 (FAX)
8
S:\Hp8960\Generic Documents\Manual Operation Getting S tarted Guide\Pi_manual operation gett ing started guide\chapters\titlepage.fm
Page 9
Regional Sales Offices
United States of America:
Agilent Technologies (tel) 1 800 452 4844 Test and Measurement Call Center P.O. Box 4026
Englewood, CO 80155-4026
Canada:
Agilent Technologies Canada Inc. (tel) 1 877 894 4414 2660 Matheson Blvd. E Mississauga, Ontario L4W 5G1
Europe:
Agilent Technologies (tel) (3120) 547 9999 European Marketing Organization P.O . B o x 99 9 1180 AZ Amstelveen The Netherlands
Japan:
Agilent Technologies Japan Ltd. (tel) (81) 456-56-7832 Measurement Assistance Center (fax) (81) 426-56-7840 9-1 Takakura-Cho, Hachioji-Shi, Tokyo 192-8510, Japan
Latin America:
Agilent Technologies (tel) (305) 267 4245 Latin America Region Headquarters (fax) (305) 267 4286 5200 Blue Lagoon Drive, Suite #950 Miami, Florida 33126 U.S. A.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\titlepage.fm
9
Page 10
Australia/New Zealand:
Agilent Technologies Australia Pty Ltd. Australia New Zealand 347 Burwood Highway (tel) 1 800 629 485 (tel) 0 800 738 378 Forest Hill, Victoria 3131 (fax) (61 3) 9272 0749 (fax) (64 4) 802 6881
Asia Pacific:
Agilent Technologies(tel) (852) 3197 7777 24/F, Cityplaza One,(fax) (852) 2506 9233 111 Kings Road, Taikoo Shing, Hong Kong
10
S:\Hp8960\Generic Documents\Manual Operation Getting S tarted Guide\Pi_manual operation gett ing started guide\chapters\titlepage.fm
Page 11
Declaration of Conformity
According ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name Agilent Technologies UK Limited Agilent Technologies, Incorporated
Manufacturer’s Address Electronics Products & Solutions
Group - Queensferry South Queensferry West Lothian, EH30 9TG Scotland, United Kingdom
RF Communications PGU 24001 E. Mission Avenue Liberty Lake, Washington 99019-9599 USA
Declares, that the product Product Name:
Model Number: Product Options:
8960 Series 10 Wireless Communications Test Set
E5515B
This declaration covers all options of the above product.
Conforms with the following European Directives
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CD Marking accordingly.
Conforms with the following product standards:
EMC
Standard
IEC 61326-1:1997+A1:1998/EN 61326-1:1997+A1:1998 CISPR 11:1990 / EN 55011:1991 IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 IEC 61000-4-3:1995 / EN 61000-4-3:1995 IEC 61000-4-4:1995 / EN 61000-4-4:1995 IEC 61000-4-5:1995 / EN 61000-4-5:1995 IEC 61000-4-6:1996 / EN 61000-4-6:1996 IEC 61000-4-11:1994 / EN 61000-4-11:1994
Limit
Group 1 Class A [1] 4kV CD, 8kV AD
3 V/m, 80-1000 MHz
0.5V signal lines, 1kV power lines
0.5 kV line-line, 1 kV line-ground 3V, 0.15-80 MHz 1 cycle, 100%
Safety IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995
Canada CSA C22.2 No. 1010.1:1992
Supplemental Information:
[1] The product was tested in a typical configuration with Agilent Technologies test systems
14 December 2000
R.M. Evans / Quality Manager
14 December 2000
W. V. Roland / Reliability & Regulatory Engineering Manager
For further information, please c ontact your local Agilent Technologies sales office, agent or distributor. Authorized EU-repr esentative: Agilent Technologies Deutsch land, GmbH, Herrenberger Strabe 130, D 71034 Boblingen, Germany
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\titlepage.fm
11
Page 12
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturers Name Agilent Technologies UK Limited Agilent Technologies, Incorporated
Declaration of Conformity
Manufacturers Address Electronics Products & Solutions
Declares, that the product Product Name:
Model Number: Product Options:
Conforms with the following European Directives
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CD Marking accordingly.
Conforms with the following product standards:
EMC
Safety IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995
Group - Queensferry South Queensferry West Lothian, EH30 9TG Scotland, United Kingdom
8960 Series 10 Wireless Communications Tes t Se t
E5515C
This declaration covers all options of the above product.
Standard
IEC 61326-1:1997+A1:1998/EN 61326:1997/A1:1998 CISPR 11:1990 / EN 55011:1991 IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 IEC 61000-4-3:1995 / EN 61000-4-3:1995 IEC 61000-4-4:1995 / EN 61000-4-4:1995 IEC 61000-4-5:1995 / EN 61000-4-5:1995 IEC 61000-4-6:1996 / EN 61000-4-6:1996 IEC 61000-4-11:1994 / EN 61000-4-11:1994
Canada CSA C22.2 No. 1010.1:1992
RF Communications PGU 24001 E. Mission Avenue Liberty Lake, Washington 99019-9599 USA
Limit
Group 1 Class A [1] 4kV CD, 8kV AD
3 V/m, 80-1000 MHz
0.5kV signal lines, 1kV power lines
0.5kV line-line, 1kV line-ground 3V, 0.15-80 MHz 1 cycle, 100%
Supplemental Information:
[1] The product was tested in a typical configuration with Agilent Technologies test systems
01 May 2001
01 May 2001
For further information, please contact your local Agilent Technologies sales office, agent or distributor. Authorized EU-representative: Agilent Technologies Deutschland, GmbH, Herrenberger Strabe 130, D 71034 Boblingen, Germany
R.M. Evans / Quality Manager
W. V. Roland / Reliability & Regulatory Engineering Manager
12
S:\Hp8960\Generic Documents\Manual Operation Getting S tarted Guide\Pi_manual operation gett ing started guide\chapters\titlepage.fm
Page 13
Declaration of Conformity
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturers Name Agilent Technologies, Incorporated Agilent Technologies, Incorporated
Manufacturers Address Electronics Products & Solutions
Group - Queensferry South Queensferry West Lothian, EH30 9TG Scotland, United Kingdom
Declares, that the product Product Name:
Model Number: Product Options:
8960 Series 10 Wireless Communications Test Set
E5515T
This declaration covers all options of the above product.
Conforms with the following European Directives
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CD Marking accordingly.
Conforms with the following product standards:
EMC
Standard
IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998 CISPR 11:1990 / EN 55011:1991 IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 IEC 61000-4-3:1995 / EN 61000-4-3:1995 IEC 61000-4-4:1995 / EN 61000-4-4:1995 IEC 61000-4-5:1995 / EN 61000-4-5:1995 IEC 61000-4-6:1996 / EN 61000-4-6:1996 IEC 61000-4-11:1994 / EN 61000-4-11:1994
Limit
Group 1 Class A 4kV CD, 8kV AD
3 V/m, 80-1000 MHz
0.5V signal lines, 1kV power lines
[1]
0.5 kV line-line, 1 kV line-ground 3V, 0.15-80 MHz 1 cycle, 100%
Safety IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995
Canada CSA C22.2 No. 1010.1:1992
Supplemental Information:
[1]
The product was tested in a typical configuration with Agilent Technologies test systems
14 December 2000
W. V. Roland / Reliability & Regulatory Engineering Manager
For further information, please contact your local Agilent Technologies sales office, agent or distributor. Authorized EU-representative: Agilent Technologies Deutschland, GmbH, Herrenberger Strabe 130, D 71034 Boblingen, Germany
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\titlepage.fm
13
Page 14
Manufacturers Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB(A).
Sound Pressure Lp < 70 dB(A).
At Operator Position.
Normal Operation.
According to ISO 7779:1988/EN 27779:1991 (Type Test).
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der Maschinenlärminformationsverordnung vom 18 Januar 1991.
Schalldruckpegel Lp < 70 dB(A).
Am Arbeitsplatz.
Normaler Betrieb.
Nach ISO 7779:1988/EN 27779:1991 (Typprüfung).
14
S:\Hp8960\Generic Documents\Manual Operation Getting S tarted Guide\Pi_manual operation gett ing started guide\chapters\titlepage.fm
Page 15
Contents
AMPS/136 Mobile Test Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
How Do I Make Measurements on a Mobile? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
A. Establish a call. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
B. Select measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
How Do I Change Measurement Setup? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
A. Select a measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
B. Set up the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
How Do I Turn Off a Measurement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
How Do I Change Call Parameters? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
How Do I Change Cell Parameters? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
How Do I Change the MS TX Level? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
A. Change the digital MS TX level immediately. . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
B. Change the MS TX level during a handoff. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
How do I Perform a Handoff? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
How Do I Configure the Test Set for My Test System? . . . . . . . . . . . . . . . . . . . . . 32
A. Configure instrument information and setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
B. Set amplitude offsets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
C. Check the message log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
How Do I End a Call? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
AMPS/136 Mobile Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Measuring Digital Transmit Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Measuring Modulation Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Measuring Adjacent Channel Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Measuring Bit Error Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Measuring IQ Tuning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Measuring Analog Transmit Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Measuring Frequency Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Measuring Frequency Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
15
Page 16
Contents
cdma2000 Mobile Test Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
How Do I Make Measurements on a Mobile Station? . . . . . . . . . . . . . . . . . . . . . . . 48
A. Establish a call. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
B. Select measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
How Do I Change the Measurement Setup? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
A. Select a measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
B. Set up the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
How Do I Turn Off a Measurement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
How Do I Set Up a Call? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
How Do I Set Up an IS-2000 Call? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
How Do I Set Up an IS-95 Call? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
How Do I Change Call Parameters? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
How Do I Change Cell Information? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
A. Set cell parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
B. Set access parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
C. Set registration parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
How Do I Change Code Channel Levels? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
How Do I Perform a Handoff? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
How Do I Configure the Test Set for My Test System? . . . . . . . . . . . . . . . . . . . . . 65
A. Configure instrument information and setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
B. Set amplitude offsets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
C. Check message log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
How Do I End a Call? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
cdma2000 Measurement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Measuring Waveform Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Measuring Handoff Waveform Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Measuring Digital Average Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Measuring Channel Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Measuring Access Probe Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
16
Page 17
Contents
Measuring Frame Error Rate (FER) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Measuring TDSO (Test Data Service Option) Frame Error Rate (FER) . . . . . . . . . .77
Measuring Code Domain Power and Code Domain Power + Noise . . . . . . . . . . . . . .79
Measuring Code Channel Time/Phase Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
Measuring Gated Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Measuring TX Spurious Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
Related Topics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
GPRS Mobile Test Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
How Do I Make Measurements on a Mobile? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
A. Establish a data connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
B. Select measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
How Do I Change Measurement Setup? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
A. Select a measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
B. Set up the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
How Do I Turn Off a Measurement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
How Do I Change Call Parameters? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
How Do I Change Cell Parameters? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
A. Select the Cell Parameters menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94
B. Set a cell parameter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
How Do I Perform a Handover? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
How Do I Change the MS TX Level? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
A. Change the MS TX level immediately. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
B. Change the MS TX level during a handover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98
How Do I Configure the Test Set for My Test System? . . . . . . . . . . . . . . . . . . . . . 99
A. Configure instrument information and setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
B. Set amplitude offsets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
C. Check the message log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .101
How Do I End the Data Connection? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
17
Page 18
Contents
GPRS Mobile Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Measuring Transmit Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Measuring Power versus Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Measuring Phase and Frequency Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Measuring Output RF Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Measuring Bit Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
GSM Mobile Test Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
How Do I Make Measurements on a Mobile? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
A. Establish a call. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
B. Select measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
How Do I Change a Measurements Setup? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
A. Select a measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
B. Set up the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
How Do I Turn Off a Measurement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
How Do I Change Call Parameters? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
How Do I Change Cell Parameters? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
A. Select the Cell Parameters menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
B. Set a cell parameter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
How Do I Configure the Test Set for My Test System? . . . . . . . . . . . . . . . . . . . . 132
A. Configure instrument information and setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
B. Set amplitude offsets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
C. Check the message log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
How Do I End a Call? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
GSM Mobile Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Measuring Transmit Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Measuring Power versus Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Power versus Time Graph Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Measuring Phase and Frequency Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Phase and Frequency Error Graph Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
18
Page 19
Contents
Measuring Fast Bit Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .148
Measuring Bit Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .149
Measuring Output RF Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .150
Output RF Spectrum Graph Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .154
Measuring IQ Tuning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .155
IQ Tuning Graph Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .158
E1985A GSM_AMPS/136_GPRS Mobile Test Application (fast switching). 159
How Do I Switch Formats? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
A. Choose a format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
B. Switch formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .161
Switching Test Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
How Do Switch Test Applications? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
A. Choose a test application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
B. Switch test applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .165
19
Page 20
Contents
20
Page 21

AMPS/136 Mobile Test Application

AMPS/136 Mobile Test Application
21
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 22

How Do I Make Measurements on a Mobile?

How Do I Make Measurements on a Mobile?

A. Establish a call.

1
6
2
1. Press the blue SHIFT key.
2. Press the green
3. Connect the mobile to the RF IN/OUT port.
4. Turn the mobile on and wait for it to camp.
5. On the mobile, press 1, 2, 3, then press send.
6. Check for Connected in the Active Cell Status: field.
PRESET key.
3
22
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_gen_op_overview_make_meas.fm
Page 23

B. Select measurements.

The gray boxes indicate that the measurement is being made, but the results are not being
These black boxes indicate that adjacent channel power measurement results are being displayed in the lower measurement window, and frequency modulation measurement results are being displayed
How Do I Make Measurements on a Mobile?
2
1
1. Press the Measurement selection key.
2. Highlight a measurement and press the knob.
3. Repeat steps 1 and 2 to add measurements.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
23
Page 24

How Do I Change Measurement Setup?

How Do I Change Measurement Setup?

A. Select a measurement.

2
1
1. Press the Measurement selection key.
2. Highlight a measurement to setup (the measurement must already be enabled) and press the knob.

B. Set up the measurement.

1
2
3
4
1. Press the measurement’s setup key (F1).
24
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_gen_op_overview_meas_setup.fm
Page 25
How Do I Change Measurement Setup?
2. Highlight a parameter and press the knob.
3. Enter a value or selection and press the knob.
NOTE For statistical measurement results, change the Multi-Measurement
Count parameter from Off to a number >1.
4. Press the Close Menu (
F6) key.
25
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 26

How Do I Turn Off a Measurement?

How Do I Turn Off a Measurement?
2
3
4
1
1. Press the Measurement selection key.
2. Highlight the measurement you want to turn off.
3. Press the Close Measurement (
4. Press the Close Menu (
26
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_gen_op_overview_turn_off_meas.fm
F6) key.
F4) key.
Page 27

How Do I Change Call Parameters?

1. Press the CALL SETUP key.
How Do I Change Call Parameters?
1
2
4
2. On the Call Parms menu (1 of 2) press
F7, F9 or F12.
3. Enter a value or highlight a selection and press the knob.
4. Press the
More key for additional call parameters.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
27
Page 28

How Do I Change Cell Parameters?

How Do I Change Cell Parameters?
Cell parameters specify the information sent to the mobile to identify the cell and to tell the mobile to report mobile assisted handoff (MAHO) information.
1
4
5
3
1. Press the
2. Press the Operating Mode (
3. Press the
4. Press System ID, SOC, and MCC. Analog Control Channel uses System ID.
5. Press the MAHO Info (F6) key to access MAHO BER and RSSI features (only available during digital traffic channel (DTC) operation).
28
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_gen_op_overview_cell_parms.fm
CALL SETUP key.
F1) key and select Active Cell operating mode.
More key to select screen 2 of 2.
F1, F2, or F3 to change available cell parameters: digital control channel uses
Page 29

How Do I Change the MS TX Level?

How Do I Change the MS TX Level?
There are two ways to change the MS TX Level:
A. Change level immediately
B. Change level during a handoff
Both are explained below.

A. Change the digital MS TX level immediately.

1
2
4
1. Press the CALL SETUP key.
2. On Call Parms menu (1 of 2), press the Dig MS TX Level (
3. Select a new MS TX level and press the knob.
4. Check the current traffic channel info to see the new MS TX level.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F9) key.
29
Page 30
How Do I Change the MS TX Level?

B. Change the MS TX level during a handoff.

3
2
4
1. On Call Control menu, press the Call/Handoff Setup (F5) key.
2. Select and change the MS TX Level.
3. If you press the Execute Handoff (
F5) key, the digital MS TX level changes.
4. If you abort the handoff by pressing the Close Menu ( does not change.
30
F6) key, digital MS TX level
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_gen_op_overview_tx_level.fm
Page 31

How do I Perform a Handoff?

How do I Perform a Handoff?
3
4
5
1. Press the CALL SETUP key.
2. Press the Call/Handoff Setup (
3. Select the type of channel to hand off to (AVC or DTC).
4. Change the various parameters for the AVC or DTC.
5. Press the Execute Handoff ( Menu (
F6) key to abort the handoff.
6. Check for Connected in the Active Cell Status: field.
7. Note the change in the current traffic channel information.
F5) key.
F5) key to complete the handoff, or press the Close
6
31
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 32

How Do I Configure the Test Set for My Test System?

How Do I Configure the Test Set for My Test System?

A. Configure instrument information and setup.

Test
2
3
1. Press the SYSTEM CONFIG key.
Application Revision
1
2. Press the Instrument Setup (
3. Adjust an instrument setting and then press the Close Menu (
32
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_gen_op_overview_sys_config.fm
F1) key.
F6) key.
Page 33
How Do I Configure the Test Set for My Test System?

B. Set amplitude offsets.

2
4
1. On the Configuration Summary Screen, press the RF IN/OUT Amptd Offset
F5) key.
(
2. On the RF IN/OUT Amplitude Offset screen, press the RF IN/OUT Amptd Offset Setup (F2) key.
3. Enter the amplitude offset for the test frequencies you use.
4. Press the Close Menu (
5. Press the Return to Config Summary (
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F6) key.
F6) key.
33
Page 34
How Do I Configure the Test Set for My Test System?

C. Check the message log.

1
2
1. Press the Message Log (F7) key and view the message log.
2. Press theReturn to Utilities (
34
F12) key.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_gen_op_overview_sys_config.fm
Page 35

How Do I End a Call?

2
1. Press the CALL SETUP key.
How Do I End a Call?
2. Press the End Call (
F3) key, or end the call from the mobile.
3. Check for Idle in the Active Cell Status: field.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
35
Page 36

AMPS/136 Mobile Measurements

AMPS/136 Mobile Measurements
The following step-by-step procedures explain how to perform specific measurements available on the AMPS/136 Test Application. A detailed description of each measurement is given in the Reference information on the documentation CD-ROM and at the Agilent 8960 support website (www.agilent.com/find/8960support).
Setup parameters for each measurement are assumed to be set to their default (full preset) values unless otherwise stated.
The following measurement procedures are provided:
Measuring Digital Transmit Power” on page 37
Measuring Modulation Accuracy” on page 38
Measuring Adjacent Channel Power” on page 39
Measuring Bit Error Ratio” on page 40
Measuring IQ Tuning” on page 41
Measuring Analog Transmit Power” on page 43
Measuring Frequency Stability” on page 44
Measuring Frequency Modulation” on page 45
36
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_man_measurements.fm
Page 37
AMPS/136 Mobile Measurements

Measuring Digital Transmit Power

This measurement can only be performed with the mobile on a digital traffic channel.
1. Establish a call with the mobile on a digital traffic channel.
2. Press the Measurement selection key.
3. Select the Digital Transmit Power measurement.
4. Press the Digital Transmit Power Setup (
F1) key.
5. Set the measurement parameters as needed for your measurement situation.
Observe how sending a change power level command causes the mobile’s output level to change.
To see the effects of sending a change power level command to the mobile, press Dig MS TX Level (
Power display.
F9) and slowly change the setting while observing the Digital Transmit
37
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 38
AMPS/136 Mobile Measurements

Measuring Modulation Accuracy

This measurement can only be performed with the mobile on a digital traffic channel.
1. Establish a call with the mobile on a digital traffic channel.
2. Press the Measurement selection key.
3. Select the Modulation Accuracy measurement.
4. Press the Modulation Accuracy Setup (
F1) key.
5. Set the measurement parameters as needed for your measurement situation, including: Multi-Measurement Count = 20 (not required, but used in this example so you can see the effects of using multi-measurements)
Typical modulation accuracy results when using multi-measurements are shown above.
38
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_man_measurements.fm
Page 39
AMPS/136 Mobile Measurements

Measuring Adjacent Channel Power

This measurement can only be performed with the mobile on a digital traffic channel.
1. Establish a call with the mobile on a digital traffic channel.
2. Press the Measurement selection key.
3. Select the Adjacent Channel Power measurement.
4. Press the Adjacent Chan Power Setup (
F1) key.
5. Set the measurement parameters as needed for your measurement situation, including: Multi-Measurement Count = 10 (not required, but used in this example)
Typical adjacent channel power results when using multi-measurements are shown above.
39
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 40
AMPS/136 Mobile Measurements

Measuring Bit Error Ratio

This measurement can only be performed with the mobile on a digital traffic channel in DTC Test mode.
1. Establish a call on a digital traffic channel in test mode.
a. Set the test set’s Operating Mode (
F1) to DTC Test.
b. Put the mobile into digital test mode.
2. Press the
Measurement selection key.
3. Select the Loopback BER measurement.
4. Press the Loopback BER Setup (
F1) key.
5. Set the measurement parameters as needed for your measurement situation.
Observe how reducing the cell power to low levels causes the bit error ratio to increase.
To see how low levels of cell power into the mobile affect the bit error ratio, press Cell Power (
40
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_man_measurements.fm
F7) and slowly reduce the level.
Page 41
AMPS/136 Mobile Measurements

Measuring IQ Tuning

This measurement can only be performed with the mobile on a digital traffic channel.
1. Establish a call on a digital traffic channel. If the mobile is configured in test mode, which is typically the case for this measurement, use the following steps:
a. Set the test set’s Operating Mode (
F1) to DTC Test.
b. Put the mobile into digital test mode.
2. Press the
Measurement selection key.
3. Select the IQ Tuning measurement.
4. Press the IQ Tuning Setup (
5. Press the Measurement Setup (
F1) key.
F1) key.
6. Set the measurement parameters as needed for your measurement situation, including: Multi-Measurement Count = 50 (not required, but used in this example)
7. Press the IQ Tuning Offset Freqs (
F2) key and enter the desired offset
frequencies.
8. Set the mobile to transmit a pattern of 00 (00, 00, 00, ...).
The results are displayed on the IQ Tuning screen. Notice that when the mobile is transmitting a series of all zeros the carrier offset is at +3037.50 Hz.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
41
Page 42
AMPS/136 Mobile Measurements
9. Change the settings on the mobile to transmit a repeating pattern of 10 (10, 10, 10,...).
The results are displayed on the IQ Tuning screen. Notice that when the mobile is transmitting a repeating pattern of 10 the carrier offset is at -3037.50 Hz.
42
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_man_measurements.fm
Page 43
AMPS/136 Mobile Measurements

Measuring Analog Transmit Power

This measurement can only be performed with the mobile on an analog voice channel (AVC).
1. Establish a call with the mobile on an analog voice channel.
2. Press the
Measurement selection key.
3. Select the Analog Transmit Power measurement.
4. Press the Analog Transmit Power Setup (
F1) key.
5. Set the measurement parameters as needed for your measurement situation.
Observe how sending a change power level
command causes the mobile’s output level to change.
To see the effects of sending a power level change command to the mobile, press Anl MS TX Level (
F9) and slowly change the setting while observing the Analog Transmit
Power display.
43
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 44
AMPS/136 Mobile Measurements

Measuring Frequency Stability

This measurement can only be performed with the mobile on an analog voice channel (AVC).
1. Establish a call with the mobile on an analog voice channel.
2. Press the
Measurement selection key.
3. Select the Frequency Stability measurement.
4. Press the Frequency Stability Setup (
F1) key.
5. Set the measurement parameters as needed for your measurement situation.
A typical frequency stability measurement result is shown above.
44
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_man_measurements.fm
Page 45
AMPS/136 Mobile Measurements

Measuring Frequency Modulation

This measurement can only be performed with the mobile on an analog voice channel (AVC).
1. Establish a call with the mobile on an analog voice channel.
2. Press the
Measurement selection key.
3. Select the Frequency Modulation measurement.
4. Press the Frequency Mod Setup (
F1) key.
5. Set the measurement parameters as needed for your measurement situation, including these typcial settings:
Distortion State = On Modulation Frequency State = On Distortion Fundamental Frequency = the modulation frequency Filter Type = 100 Hz BW BPF (or as needed for the type of signal being
measured) 100 Hz BW BPF Center Frequency = the modulation frequency (if this filter is used)
A typical frequency modulation measurement result is shown above. In this case, the mobiles microphone was stimulated by a 1 kHz tone. The Distortion Fundamental
Frequency and 100 Hz BW BPF Center Frequency were therefore both set to 1 kHz.
45
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 46
AMPS/136 Mobile Measurements
46
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\is136_man_measurements.fm
Page 47

cdma2000 Mobile Test Application

cdma2000 Mobile Test Application
47
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 48

How Do I Make Measurements on a Mobile Station?

How Do I Make Measurements on a Mobile Station?

A. Establish a call.

1. Connect the mobile station to the Test Set.
2. Make a call (see “How Do I Set Up a Call?” on page 52).

B. Select measurements.

These black boxes indicate that access probe power measurement results are being displayed in the lower measurement window, and channel power results are being displayed in the upper measurement window.
The gray boxes indicate that the measurement is being made, but the results are not being displayed.
1
1. Press the Measurement selection key.
2. Highlight a measurement and press the knob.
3. Repeat steps 1 and 2 to select additional measurements.
For a complete list of measurement procedures, see “cdma2000 Measurement
Procedures on page 69
2
48
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_make_meas.fm
Page 49

How Do I Change the Measurement Setup?

How Do I Change the Measurement Setup?

A. Select a measurement.

2
1
1. Press the Measurement selection key.
2. Highlight a measurement to set up and press the knob.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
49
Page 50
How Do I Change the Measurement Setup?

B. Set up the measurement.

1
2
3
4
1. Press the MEASUREMENT key.
2. Press the Control (
F1) key to display the measurements setup menu.
3. Highlight a parameter and press the knob.
4. Enter a value or selection and press the knob.
NOTE For statistical measurement results, highlight the Multi-Measurement
Count parameter and press the ON key. The number of averages will
default to 10.
5. Press the Close Menu (
F6) key.
50
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_meas_setup.fm
Page 51

How Do I Turn Off a Measurement?

2
3
4
1
How Do I Turn Off a Measurement?
1. Press the Measurement selection key.
2. Use the knob to highlight the measurement you want to turn off.
3. Press Close Measurement (
4. Press Close Menu (
F6).
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F4).
51
Page 52

How Do I Set Up a Call?

How Do I Set Up a Call?
How Do I Set Up an IS-2000 Call? on page 52
How Do I Set Up an IS-95 Call? on page 54

How Do I Set Up an IS-2000 Call?

The Call Parms keys and Call Control keys provide access to the parameters needed to set up a call.
2 3 1
1
1. Press the
2. Press the Operating Mode (
3. Press the System Type (
4. Press the Cell Band ( up the call.
CALL SETUP key.
F2) key and select IS-2000.
F8) key and select the band in which you would like to bring
1
4
5 6
7
F1) key and set the operating mode to Active Cell.
5. Press the Radio Config (F11) key and select the radio configuration in which you would like to bring up the call.
52
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_setup_call.fm
Page 53
How Do I Set Up a Call?
6. Press the FCH Service Option Setup (F12) key and set the service option for the radio configurations you plan on testing.
NOTE SO55 (Service Option 55) is not supported by some mobile stations. If
your mobile station does not support SO55 calls cannot be connected, and you will need to select one of the other loopback or echo service options.
7. Check the rest of the Call Parms settings by pressing the
More key to display the 2
of 3 and 3 of 3 menus.
8. You may need to change cell settings such as SID (System Identification) to allow the mobile station to find service. See “How Do I Change Cell Information?” on
page 57.
9. Turn on power to the mobile station and wait for an indication that it has found service.
10.Make a mobile station originated call (by entering a phone number and pressing SEND on the mobile station handset); or, wait for the mobile station to perform a power up registration and press the Originate Call (
F3) key.
11.Verify that the call is connected.
53
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 54
How Do I Set Up a Call?

How Do I Set Up an IS-95 Call?

2 3 1
1
1. Press the CALL SETUP key.
2. Press the Operating Mode (
3. Press the System Type (
4. Press the Cell Band ( up the call.
F2) key and select IS-95.
F8) key and select the band in which you would like to bring
1
4 5
6
7
8
F1) key and set the operating mode to Active Cell.
5. Press the Channel (
6. Press the Protocol Rev ( mobile station.
7. Press the Service Option (
8. Check the rest of the Call Parms settings by pressing the of 3 and 3 of 3 menus.
9. You may need to change cell settings such as SID (System Identification) to allow the mobile station to find service. See “How Do I Change Cell Information?” on
page 57.
10.Turn on power to the mobile station and wait for an indication that it has found service.
54
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_setup_call.fm
F9) key and set the RF channel number.
F10) key and set the protocol revision supported by the
F12) key and select a service option for the call.
More key to display the 2
Page 55
How Do I Set Up a Call?
11.Make a mobile station originated call (by entering a phone number and pressing SEND on the mobile station handset), or wait for the mobile station to perform a power up registration and press the Originate Call (
F3) key.
12.Verify that the call is connected.
55
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 56

How Do I Change Call Parameters?

How Do I Change Call Parameters?
1. Press the CALL SETUP key.
2. On the Call Parms menu (1 of 2) press any key.
1
2
4
3. Highlight a selection and press the knob. Enter a value.
4. Press the
56
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_call_parms.fm
More key for additional call parameters.
Page 57

How Do I Change Cell Information?

How Do I Change Cell Information?
There are three types of cell information: cell parameters, access parameters, and registration parameters.

A. Set cell parameters.

1. Select the Cell Parameters menu.
1
3
2
1. Press the
2. Press the
3. Press the Cell Info (
CALL SETUP key.
More key.
F2) key.
4. Press the Cell Parameters (
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F2) key.
57
Page 58
How Do I Change Cell Information?
2. Set a cell parameter.
3
1. Turn the knob to highlight a parameter and then press the knob.
2. Enter a value or selection and press the knob.
2
3. Press the Close Menu (
F6) key.
58
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_cell_info.fm
Page 59

B. Set access parameters.

1. Select the Access Parameters menu.
3
2
How Do I Change Cell Information?
1
1. Press the
2. Press the
3. Press the Cell Info (
CALL SETUP key.
More key.
F2) key.
4. Press the Access Parameters (
F3) key.
59
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 60
How Do I Change Cell Information?
2. Set an access parameter.
3
1. Turn the knob to highlight a parameter and then press the knob.
2. Enter a value or selection and press the knob.
2
3. Press the Close Menu (
F6) key.
60
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_cell_info.fm
Page 61

C. Set registration parameters.

1. Select the Registration Parameters menu.
3
2
How Do I Change Cell Information?
1
1. Press the
2. Press the
3. Press the Cell Info (
CALL SETUP key.
More key.
F2) key.
4. Press the Registration Parameters (
F4) key.
61
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 62
How Do I Change Cell Information?
2. Set a registration parameter.
3
1. Turn the knob to highlight a parameter and then press the knob.
2. Enter a value or selection and press the knob.
1
3. Press the Close Menu (
F6) key.
62
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_cell_info.fm
Page 63

How Do I Change Code Channel Levels?

How Do I Change Code Channel Levels?
4
1
2
1. Press the CALL SETUP key.
2. Press the
3. Press the Generator Info (
4. Press the Code Channel Parameters
5. Set the level of the desired channel by using the knob and numeric keys.
More key.
F3) key.
(F2 ) key.
63
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 64

How Do I Perform a Handoff?

How Do I Perform a Handoff?
1
2
3
5
4
6
1. On the Call Setup screen, press the Call/Handoff Setup (F5) key.
2. Highlight the Handoff System Type parameter and select the desired system.
3. Highlight the Handoff Cell Band parameter and select the desired band.
4. Highlight the Handoff Channel parameter and select the desired channel.
5. Press the Execute Handoff
F6) key to abort the handoff.
(
6. Check for Connected in the Active Cell status field.
(F5) key to complete the handoff, or the Close Menu
64
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_handoff.fm
Page 65

How Do I Configure the Test Set for My Test System?

How Do I Configure the Test Set for My Test System?

A. Configure instrument information and setup.

2
3
4
1. Press the SYSTEM CONFIG key.
Test Application Revision
1
2. Press the Instrument Setup
3. Adjust an instrument setting.
4. Press the
Close Menu (F6) key.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
(F1) key.
65
Page 66
How Do I Configure the Test Set for My Test System?

B. Set amplitude offsets.

2
4
1. On the Configuration Summary screen (not shown), press the RF IN/OUT Amptd Offset (
2. On the RF IN/OUT Amplitude Offset screen, press the RF IN/OUT Amptd Offset Setup (F2) key.
F5) key.
3. Enter the amplitude offset for the test frequencies you use.
4. Press the Close Menu (
66
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_sys_config.fm
F6) key.
Page 67
How Do I Configure the Test Set for My Test System?

C. Check message log.

1
2
1. From any System Config screen, press the Message Log (F7) key and view the message log.
2. Press the Return (F12) key.
67
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 68

How Do I End a Call?

How Do I End a Call?
2
1. Press the CALL SETUP key.
2. Press the End Call (
F3) key, or end the call from the mobile.
3. Check for Idle in the Active Cell status field.
68
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_end_call.fm
Page 69

cdma2000 Measurement Procedures

cdma2000 Measurement Procedures
The following step-by-step procedures explain how to manually perform specific measurements available on the cdma2000 Test Application. A detailed description of each measurement is given in the Reference information on the documentation CD-ROM and at the Agilent 8960 support website (http://www.agilent.com/find/8960support/).
Setup parameters for each measurement are assumed to be set to their default (full preset) values unless otherwise stated.
NOTE To obtain measurement results a call must first be connected See “A.
Establish a call. on page 48
The following measurement procedures are provided:
Measuring Waveform Quality” on page 70
Measuring Handoff Waveform Quality”
Measuring Digital Average Power” on page 72
Measuring Channel Power” on page 73
Measuring Access Probe Power” on page 74
Measuring Frame Error Rate (FER)” on page 76
Measuring Code Domain Power and Code Domain Power + Noise” on page 79
Measuring Code Channel Time/Phase Error” on page 82
Measuring Gated Power” on page 83
Measuring TX Spurious Emissions” on page 84
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
69
Page 70
cdma2000 Measurement Procedures

Measuring Waveform Quality

Each waveform quality measurement returns a number of measurements which includes rho.
1. Select the Waveform Quality/Code Domain measurement (see “B. Select
measurements. on page 48). A typical display is shown below.
2. Set up the measurement parameters as needed for your measurement situation (see “B. Set up the measurement. on page 50.
Numeric rho is the screen currently displayed. In addition to rho, it also provides the following measurements, calculated from the same data:
Time Error
Carrier Feedthrough
Phase Error
Magnitude Error
EVM
There are graphical measurement results available as a result of performing this measurement. See “Measuring Code Domain Power and Code Domain Power + Noise
on page 79.
70
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
Page 71
cdma2000 Measurement Procedures

Measuring Handoff Waveform Quality

Each waveform quality measurement returns a number of measurements which includes rho.
These measurements can be made only when one of the following radio configurations (see “How Do I Set Up a Call?” on page 52) is selected:
(Fwd3, Rvs3)
(Fwd4, Rvs3)
1. Select the Handoff Waveform Quality measurement (see “B. Select
measurements. on page 48).
2. Set up the measurement parameters (Measurement Timeout is the only parameter provided with this measurement) as needed for your measurement situation (see “B. Set up the measurement. on page 50.)
3. Press the
START SINGLE key. A typical display is shown below.
In addition to rho, the waveform quality measurement also provides the following measurements, calculated from the same data:
Time Error
Carrier Feedthrough
Phase Error
Magnitude Error
71
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 72
cdma2000 Measurement Procedures

Measuring Digital Average Power

1. Select the Digital Average Power measurement (see “B. Select measurements.” on
page 48). A typical display is shown below.
2. Press the Digital Average Power Setup (
F1) key.
3. Set up the measurement parameters as needed for your measurement situation (see “B. Set up the measurement. on page 50.
72
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
Page 73
cdma2000 Measurement Procedures

Measuring Channel Power

1. Select the Channel Power measurement (see “B. Select measurements. on page
48). A typical display is shown below.
2. Set up the measurement parameters as needed for your measurement situation (see “B. Set up the measurement.” on page 50.
73
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 74
cdma2000 Measurement Procedures

Measuring Access Probe Power

1. Press the CALL SETUP key.
2. Press the End Call key (
F3).
3. Select the Access Probe Power measurement (see “B. Select measurements. on
page 48).
4. Press the Register Mobile key (
F6). A typical display is shown below.
NOTE The remainder of this procedure lists the steps required to make access
probe power measurements using call limit mode. Call limit mode provides testing of complete access probe sequences during a call origination.
5. Set up the measurement parameters as needed for your measurement situation (see “B. Set up the measurement. on page 50).
6. Press the displayed. Press the Cell Info key (
More key on the lower left side of the display (Call Control) until 2 of 2 is
F2) key, then the Registration
Parameters key.
7. Turn the Timer Based Registration State off.
8. Press the
9. Press the displayed. Press the Call Limit Mode key (
74
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
MEASUREMENT key.
More key on the lower right side of the display (Call Parms) until 2 of 3 is
F9) key and turn call limiting on.
Page 75
cdma2000 Measurement Procedures
10.Press the Originate Call key (F3). Wait for the Access Probe call status indication in the Active Cell field below the measurement window. A typical display is shown below.
11.Turn off call limiting.
75
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 76
cdma2000 Measurement Procedures

Measuring Frame Error Rate (FER)

1. Press the Measurement selection key.
2. Select the Frame Error Rate measurement.
3. Press the Frame Error Rate Setup (
F1) key. A typical display is shown below.
4. Set up the measurement parameters as needed for your measurement situation.
76
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
Page 77
cdma2000 Measurement Procedures

Measuring TDSO (Test Data Service Option) Frame Error Rate (FER)

1. Press the CALL SETUP key.
2. If a call is currently connected, press the End Call key (
3. Press the Radio Config key (
F11).
F3).
4. Select one of the following radio configurations:
(Fwd3, Rvs3)
(Fwd4, Rvs3)
5. Press the FCH Service Option Setup key (
F12).
6. Select the Service Option for the currently selected radio configuration, then select SO32 (+F-SCH).
7. Press the
Measurement selection key.
8. Select the TDSO Frame Error Rate measurement.
9. Press the TDSO FER Setup (
F1) key.
10.Set up the measurement parameters as needed for your measurement situation.
11.Press the
12.Press the Originate Call key (
13.Press the
CALL SETUP key.
F3).
MEASUREMENT key. Wait for the call to connect, then wait
approximately 10 seconds for the measurement to begin. A typical display is shown below.
77
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 78
cdma2000 Measurement Procedures
14.Press the TDSO FER Setup (F1) key.
15.Set up the measurement parameters as needed for your measurement situation.
78
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
Page 79
cdma2000 Measurement Procedures

Measuring Code Domain Power and Code Domain Power + Noise

These measurements can be made only when one of the following radio configurations (see “How Do I Set Up a Call?” on page 52) is selected:
(Fwd3, Rvs3)
(Fwd4, Rvs3)
(Fwd5, Rvs4)
1. Select the Waveform Quality/Code Domain measurement (see “B. Select
measurements. on page 48).
79
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 80
cdma2000 Measurement Procedures
2. Press the Code Domain Power (F3) key. A typical display is shown below.
The red bars indicate power levels
The yellow bars indicate noise levels
3. Press the Graph Control key (
F6). Change the value in the Limit field.
Note that both green lines, indicating the I and Q channel noise limits, move together.
4. Press the Return key (
5. Press the Code Domain Power + Noise (
F6).
F4) key. A typical display is shown
below.
80
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
Page 81
The red bars indicate power levels
cdma2000 Measurement Procedures
The yellow bars indicate noise levels
6. Return to the numeric Rho display at any time by pressing the Numeric Rho key
F2).
(
81
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 82
cdma2000 Measurement Procedures

Measuring Code Channel Time/Phase Error

These measurements can be made only when one of the following radio configurations (see “How Do I Set Up a Call? on page 52) is selected:
(Fwd3, Rvs3)
(Fwd4, Rvs3)
(Fwd5, Rvs4)
1. Select the Code Channel Time/Phase Error measurement (see “B. Select
measurements. on page 48). A typical display is shown below.
2. Set up the Time Limit and Phase Limit using the (
82
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
F3) and (F4) keys.
Page 83
cdma2000 Measurement Procedures

Measuring Gated Power

These measurements can be made only when one of the following radio configurations (see “How Do I Set Up a Call?” on page 52) is selected:
(Fwd1, Rvs1)
(Fwd2, Rvs2)
1. Select the Gated Power measurement (see “B. Select measurements.” on page 48).
2. Press the (
F4) key to select Zoom.
3. Turn the knob clockwise slowly to display the entire burst. Note the falling edge of the burst at the last zoom position on the right side of the display.
4. Press the (
F2) key to select the Rise/Fall View.
83
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 84
cdma2000 Measurement Procedures

Measuring TX Spurious Emissions

1. Select the Tx Spurious Emissions measurement (see “B. Select measurements. on
page 48). A typical display is shown below.
2. Press the Graph (
F3) key.
NOTE Press the Numeric (
Emissions measurement results displayed in tables.
84
F2) key if you want to return to Tx Spurious
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
Page 85
cdma2000 Measurement Procedures

Related Topics

Refer to Product Note 5988-1989EN Multi-Coded Waveform Quality and Code Domain Measurements for cdma2000 for more information about Code Domain Power measurements. This is located on the web or CDROM under Technical notes.
85
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
Page 86
cdma2000 Measurement Procedures
86
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\cdma2000_gen_op_overview_man_measurements.fm
Page 87

GPRS Mobile Test Application

GPRS Mobile Test Application
87
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\ch_GPRS.fm
Page 88

How Do I Make Measurements on a Mobile?

How Do I Make Measurements on a Mobile?

A. Establish a data connection.

5
4
2
1. Press the blue SHIFT key.
2. Press the green
3. Connect the mobile.
4. Turn the mobile on and wait for Attached in the Active Cell: field.
NOTE For mobiles that dont perform GPRS attach automatically, set the
mobile to data mode.
5. Press the Start Data Connection ( field changing to Transferring.
Preset key.
F3) key and watch for the Active Cell:
1
3
88
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\GPRS_gen_op_overview_make_meas.fm
Page 89

B. Select measurements.

The gray boxes indicate that the measurement is being made, but the results are not being displayed.
These black boxes indicate that phase and frequency error measurement results are being displayed in the lower measurement window, and bit error measurement results are being displayed in the upper measurement window.
How Do I Make Measurements on a Mobile?
1
1. Press the Measurement selection key.
2. Highlight a measurement and press the knob.
3. Repeat steps 1 and 2 to add measurements.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
89
Page 90

How Do I Change Measurement Setup?

How Do I Change Measurement Setup?

A. Select a measurement.

2
1
1. Press the Measurement selection key.
2. Highlight a measurement to set up and press the knob.
90
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\GPRS_gen_op_overview_meas_setup.fm
Page 91

B. Set up the measurement.

1
2
3
4
How Do I Change Measurement Setup?
1. Press the measurement’s setup (F1) key.
2. Highlight a parameter and press the knob.
3. Enter a value or make a selection and press the knob.
NOTE For Statistical measurement results, change the Multi-Measurement
Count parameter from Off to a number >1.
4. Press the Close Menu (
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F6) key.
91
Page 92

How Do I Turn Off a Measurement?

How Do I Turn Off a Measurement?
3
4
2
1
1. Press the Measurement selection key.
2. Highlight the measurement you want to turn off.
3. Press the Close Measurement (
4. Press the Close Menu (
92
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\GPRS_gen_op_overview_turn_off_meas.fm
F6) key.
F4) key.
Page 93

How Do I Change Call Parameters?

How Do I Change Call Parameters?
1
2
4
1. Press the CALL SETUP key.
NOTE When the mobile is transferring data (Active Cell: field is
Transferring) some call parameters cannot be changed.
2. On the Call Parms menu (1 of 4) press
F7, F8 or F9.
3. Enter a value or highlight a selection and press the knob.
4. Press the
More key for additional call parameters.
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
93
Page 94

How Do I Change Cell Parameters?

How Do I Change Cell Parameters?
NOTE You can only change two cell parameters - 3 Digit MNC for PCS1900
and Guard Period Length. Other cell parameters such as MCC, MNC, and LAC are fixed. To change the 3 digit MNC for PCS1900 the cell must be set to off by pressing the Mode (
F1), and then selecting Cell Off.

A. Select the Cell Parameters menu.

3
CALL SETUP key, then Operating
1
1. Press the CALL SETUP key.
2. Press the Cell Info (
3. Press the Cell Parameters (
94
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\GPRS_gen_op_overview_cell_parms.fm
F6) key.
F2) key.
Page 95

B. Set a cell parameter.

How Do I Change Cell Parameters?
3
1. Highlight a parameter and press the knob.
2
1
2. Enter a value or selection and press the knob.
3. Press the Close Menu (
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F6) key.
95
Page 96

How Do I Perform a Handover?

How Do I Perform a Handover?
1
3
4
1. Press the CALL SETUP key.
2. Press the Handover Setup (
3. Change the various parameters, for example Traffic Band.
4. Press the Handover Execute ( Close Menu (
F6) key to abort the handover.
5. Check for Transferring in the Active Cell: field.
6. Press the
More key to check that the changes have been implemented on the Call
Parms menus (2 of 4) and (3 of 4).
F5) key.
F5) key to complete the handover, or press the
5
6
96
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\GPRS_gen_op_overview_handover.fm
Page 97

How Do I Change the MS TX Level?

There are two ways to change the MS TX Level:
A. Change the level immediately
B. Change the level during a handover
Both are explained below.

A. Change the MS TX level immediately.

3
How Do I Change the MS TX Level?
1
4
1. Press the CALL SETUP key.
2. On the Call Parms menu (3 of 4) press the MS TX Level (
3. Set a new MS TX level and press the knob.
4. Press the Close Menu (
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F6) key.
2
F8) key.
97
Page 98
How Do I Change the MS TX Level?

B. Change the MS TX level during a handover.

4
1
3
5
1. Press the CALL SETUP key.
2. On Control menu, press the Handover Setup (
F5) key.
3. Select and change the MS TX Level.
4. Press the Handover Execute ( Close Menu (
5. Use the
More key to check that the MS TX level has been changed on the Call
F6) key to leave the level unchanged.
F5) key to change the MS TX level, or press the
Parms menu (3 of 4).
98
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\GPRS_gen_op_overview_chge_mstx.fm
Page 99

How Do I Configure the Test Set for My Test System?

How Do I Configure the Test Set for My Test System?

A. Configure instrument information and setup.

Test Application Revision
2
3
1
1. Press the SYSTEM CONFIG key.
2. Press the Instrument Setup (
3. Adjust an instrument setting and then press the Close Menu (
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
F1) key.
F6) key.
99
Page 100
How Do I Configure the Test Set for My Test System?

B. Set amplitude offsets.

2
1
4
5
1. On the Configuration Summary screen, press the RF IN/OUT Amptd Offset (F5) key.
2. On the RF IN/OUT Amplitude Offset screen, press the RF IN/OUT Amptd Offset Setup (
F2) key.
3. Enter the amplitude offsets for the test frequencies you use.
4. Press the Close Menu (
5. Press the Return (
100
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\GPRS_gen_op_overview_config_testset.fm
F6) key.
F6) key to return to the Configuration Summary screen.
Loading...