89410A Vector Signal Analyzer
Serial Numbers: 0000A00000 / 3346A00437
Analyzers may fail radiated emissions requirement due to missing
gaskets
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
Part No. Qty. Description
8160-0360 3 Rope Gasket
Situation:
The 3 rope gaskets which should be installed around the front bezel were found to be
missing. The purpose of the gasket is to improve the contact between the front frame casting and the covers. The analyzer may fail to meet the radiated emissions requirement without these gaskets.
1 of 2
AUTHOR: CMG
ENTITY: A100
© 1994 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION
CATEGORY:
LOCATION
CATEGORY:
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: 15 June 1996
ADDITIONAL INFORMATION:
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN
SCRAP
SEE TEXT
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
STANDARDS:
Labor 1.0 Hour
Continued

Page 2 Service Note 89410A-06
Solution/Action:
Install gaskets.
STEP 1. Remove top, side, and bottom covers.
STEP 2. Add rope gaskets around the front frame.
1. Press 1 rope gasket into the bottom of the front frame (MP402).
2. Press 1 rope gasket into the top of the front frame.
3. Cut 1 rope gasket in half.
4. Press 1/2 rope gasket into each side of the front frame.
STEP 3. Replace covers.