Hewlett-Packard to Agilent Technologies Transition
This manual may contain references to HP or Hewlett-Packard. Please note that
Hewlett-Packard's former test and measurement, semiconductor products and chemical analysis
businesses are now part of Agilent Technologies. To reduce potential confusion, the only
change to product numbers and names has been in the company name prefix: where a product
number/name was HP XXXX the current name/number is now Agilent XXXX. For example,
model number HP478A is now model number Agilent 478A.
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT
DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO
THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT
NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR
ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY
INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE
MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the per formance of a U.S. Government prime contract or
subcontract, Software is delivered and licensed as “Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer sof tware” as defined in FAR 52.227-19 (June 1987) or
any equivalent agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard commercial l icense terms, and
non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987) . U.S. Government users
will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Agilent
web site:
•Go to http://www.agilent.com.
• Enter the document’s pa rt number (located on the title page) in the Quick Search box.
• Click GO.
• Click on the hyperlink for the document.
• Click the printer icon located in the tool bar.
Contacting Agilent
This information supersedes all prior HP contact information.
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this material.
@
Copyright 1996-1999 Hewlett-Packard Company
Certification
Hewlett-Packard Company certifies that this product met its published specifications at the
time of shipment from the factory. Hewlett-Packard further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration
facilities of other International Standards Organization members.
Warranty
Note
The actual warranty on your instrument depends on the date it was ordered
as well as whether or not any warranty options were purchased at that time.
lb
determine the exact warranty on your instrument, contact the nearest
Hewlett-Packard sales or service office with the model and serial number of
your instrument. See the table titled “Hewlett-Packard
Sales
and Service
Offices,” later in this section, for a list of sales and service offices.
This Hewlett-Packard instrument product is warranted against defects in material and
workmanship for the warranty period. During the warranty period, Hewlett-Packard Company
will, at its option, either repair or replace products which prove to be defective.
If the warranty covers repair or service to be performed at Buyer’s facility, then the service or
repair will be performed at the Buyer’s facility at no charge within HP service travel areas.
Outside HP service travel areas, warranty service will be performed at Buyer’s facility only
upon HP’s prior agreement, and Buyer
shaIl
pay HP’s round-trip travel expenses. In all other
areas, products must be returned to a service facility designated by HP
If the product is to be returned to Hewlett-Packard for service or repair, it must be returned
to a service facility designated by Hewlett-Packard. Buyer
shaIl
prepay shipping charges to
Hewlett-Packard and Hewlett-Packard shall pay shipping charges to return the product to
Buyer. However, Buyer
shaIl
pay
all
shipping charges, duties, and taxes for products returned
to Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and
use with an instrument will execute its progr
ilrmware designated by Hewlett-Packard for
amming
instructions when properly installed on
that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or
software, or
Iirmware will
be uninterrupted or error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or
misuse, operation outside of the environmental specifications for the product, or improper
site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY
DISCLAIMS THE IMPLIED
WmRANTIES
OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
E
XCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT,
OR ANY OTHER LEGAL THEORY
Hewlett-Packard Company
3495 Deer Creek Road
Palo Alto,
94304-1316
(416)
Support Center
Berkshire RG415DZ
California,
857-6027
USA
EUROPEAN FIELD OPERATIONS
Prance
Hewlett-Packard PranceHewlett-Packard
1 Avenue Du Canada
Zone
D’Activite
F-91947 Lea
France(49 6172)
(33 1) 69 82 60 60
INTERCONFIELD OPERATIONS
AnstralIa
Hewlett-Packard Australia Ltd.Hewlett-Packard (Canada) Ltd.
31-41
Joseph Street
Blackbum, Victoria 3130
(61 3) 895-2896
De Courtaboeuf
Ulis Cedex
h-Y
Hewlett-Packard
61352 Bad Homburg v.d.H
Germany
17500 South Service Road
TrausCanada Highway
Kirkland, Quebec
Canada
(614) 697-4232
GmbH
Strasse
16-O
HQJ 2X&?
1
China Hewlett-Packard Company Hewlett-Packard Japan, Ltd.
38BeiSanHuanXlRoad
shuaug Yu shu
Hai Dian District
Beijiq$
china
(86
1)
253-6888
l]iiWLWl
Hewlett-Packard
8th Flooq
337 Fu
Ihipei, T&wan
(886
H-P Building
I-king
2) 712-0404
‘Ihiwan
North Road
9-l
%kakura-Cho, Hachioji
lbkyo
192, Japan
(81426) 60-2111
Japan
Singapore
Hewlett-Packard Singapore
150 Beach Road
#29-00
Gateway West
Singapore 0718
(66) 291-9088
(Pte.)
Ltd.
Safety Symbols
The following safety symbols are used throughout this manual.
of the symbols and its meaning before operating this instrument.
Caution
Warning
Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, would result in damage to or destruction
of the instrument. Do not proceed beyond a caution note until the indicated
conditions are fully understood and met.
Wkning
correctly performed or adhered to, could result in injury or loss of life.
Do not proceed beyond a warning note until the indicated conditions are
fully understood and met.
denotes a hazard. It calls attention to a procedure which, if not
FamiIiarize
yourself with each
Instrument Markings
!
A
is necessary for the user to refer to the instructions in the documentation.
YE”
a year, it is when the design was proven.)
The instruction documentation symbol. The product is marked with this symbol when it
The CE mark is a registered trademark of the European Community. (If accompanied by
“ISMl-A”
“CSA” The CSA mark is a registered trademark of the Canadian Standards Association.
This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
vi
General Safety Considerations
Warning
Warning
Caution
Warning
Warning
This is a
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. Any
interruption of the protective conductor, inside or outside the instrument,
is likely to make the instrument dangerous. Intentional interruption is
prohibited.
No operator serviceable parts inside. Refer servicing to
personnel. To prevent electrical shock, do not remove covers.
Before switching on this instrument, make sure that the line voltage selector
switch is set to the voltage of the power supply and the correct fuse is
installed.
The opening of covers or removal of parts is likely to expose dangerous
voltages. Disconnect the instrument from
being opened.
The power cord is connected to internal capacitors that may remain live
for 10 seconds after disconnecting the plug from its power supply.
safety
Class I product (provided with a protective earthing
qualilled
all voltage sources while it is
Warning
Warning
Note
For continued protection against fire hazard, replace line fuse only with
same type and rating (F
prohibited.
If this instrument is used in a manner not specilled by
Co., the protection provided by the instrument may be impaired.
This instrument has been designed and tested in accordance with IEC
Publication 348, Safety Requirements for Electronics Measuring Apparatus, and
has been supplied in a safe condition. This instruction documentation contains
information and warnings which must be
operation and to maintain the instrument in a safe condition.
3A/250V).
The use of other fuses or material is
Hewlett-I%&a.rd
followed
by the user to ensure safe
vii
User’s Guide Overview
n Chapter 1, “HP
8719D/20D/22D
Description and Options,” describes features, functions, and
available options
n Chapter 2, “Making Measurements,” contains step-by-step procedures for making
measurements or using particular functions.
n
Chapter 3, “Making Mixer Measurements,”
contains step-by-step procedures for making
calibrated and error-corrected mixer measurements.
n
Chapter 4, “Printing, Plotting, and Saving Measurement Results, n contains instructions
for saving to disk or the analyzer internal memory, and printing and plotting displayed
measurements.
n
Chapter 5, “Optimizing Measurement Results,” describes techniques and functions for
achieving the best measurement results.
n
Chapter 6, “Application and Operation Concepts, n contains explanatory-style information
about many applications and analyzer operation.
n Chapter 7, “Specifications and Measurement Uncertainties,” defines the performance
capabilities of the analyzer.
n Chapter 8, “Menu
n Chapter 9, “Key Definitions,” describes all the front panel keys, softkeys, and their
Maps,m
shows softkey menu relationships.
corresponding HP-IB commands.
n Chapter 10, “Error Messages,” provides information for interpreting error messages.
n Chapter 11, “Compatible Peripherals,
n
lists measurement and system accessories, and
other applicable equipment compatible with the analyzer. Procedures for configuring the
peripherals, and an HP-IB progr
n
Chapter 12, “Preset State and Memory Allocation, n contains a discussion of memory
amming
overview are also included.
allocation, memory storage, instrument state definitions, and preset conditions.
n
Appendix A, “The
the
CITItile
data format as well as a list of
CITIiile
Data Format and Key Word Reference, n contains information on
CITIille
keywords.
VIII
. .
.
Network Analyzer Documentation Set
The Installation and Quick Start Guide
familiarizes you with the network analyzer’s
front and rear panels, electrical and
environmental operating requirements, as well
as procedures for installing, configuring, and
verifying the operation of the analyzer.
The User’s Guide
measurements, explains commonly-used
features, and tells you how to get the most
performance from your analyzer.
The Quick Reference Guide
summary of selected user features.
The
Progmmm
programming information including an HP-IB
programming and command reference as well
as programming examples.
shows how to make
provides a
er’s
Guide
provides
The Service Guide
needed to adjust, troubleshoot, repair, and
verify conformance to published
specifications. Available with Option OBW.
provides the information
lx
hanufacturer’s
DECLARATION OF CONFORMITY
Name:
accordq to ISWIEC Guide
Hewlett-Packard Co.
22
and
EN 45014
Manufacturer’s Address:
Microwave Instruments Division
1400 Fountaingrove Parkway
Santa Rosa, CA
USA
declares that the product
Product Name:
Model Number:
Product Options:
Network Analyzer
HP 87190, HP 87200, HP 87220
This declaration covers all options of the
above products.
conforms to the following Product specifications:
Safety: IEC
EMC:
lOlO-1:199O+Al
CANKSA-C22.2
/EN
No. 10
61010-1:1993
10.1-92
CISPR 11:199O/EN 55011:1991
IECEOl-2:1984/EN
50082-1:1992
Group 1, Class A
4 kVCD, 8
/ECEOl-3:1984/EN 50082-1:1992 3 V/m, 27-500 MHz
/EC801-4:1988IEN 50082-1:1992 0.5 kV Sig. Lines, 1 kV Power Lines
/EC 555-2: 1982 +A 1: 1985 / EN 60555-2: 1987
IEC
555-3:1982 +
A1:1990/
EN 60555-3:1987 +
95403-
kVAD
Al:1991
1799
Supplementary Information:
These products herewith comply with the requirements of the Low Voltage Directive
7U2ZYEEC
Santa Rosa, California, USA
Eumpean
and the EMC Directive 89/336/EEC
and
4 June 1996
Dixon Browder/Quality Manager
Contact: Your knxl Hewlett-Packard Sates and Service
HO-TRE, Hermnberger Stmse 130, D-71034 B6b@ten,
carry the CE-marking accordingly.
officeor
Hewfett-Packard GmL#l,
Gennany
(FAX AS-7031-14-3143)
Department
Contents
1.
EP 8719D/20D/22D
Where to Look for More Information
Analyzer Description
Front Panel Features
Analyzer Display
Rear Panel Features and Connectors
Analyzer Options Available
Option
lD5,
Option 007, Mechanical Transfer Switch
Option 085, High Power System
Option 089, Frequency Offset Mode
Option 012, Direct Access Receiver Configuration
Option 400, Four-Sampler
Option 010, Time Domain
Option
Option
lCM,
lCP,
Rack Mount Flange Kit With Handles
Service and Support Options
2.
BMcing
Measurements
Where to Look for More Information
Principles of Microwave Connector Care
Basic Measurement Sequence and Example
Basic Measurement Sequence
Basic Measurement Example
Step 1. Connect the device under test and any required test equipment.
Step 2. Choose the measurement parameters.
Setting the Frequency Range
Setting the Source
Setting the Measurement
Step 3. Perform and apply the appropriate error-correction.
Step 4. Measure the device under test.
Step 5. Output the measurement results.
Using the Display Functions
To View Both Primary Measurement Channels
To Save a Data Trace to the Display Memory
‘Ib
View the Measurement Data and Memory Trace
To Divide Measurement Data by the Memory Trace
To Subtract the Memory Trace from the Measurement Data Trace
Typical test setup
Tuned receiver mode in-depth description
Frequency Range
Compatible Sweep Types
External Source Requirements
Test Sequencing
Creating a Sequence
Running a Sequence
Stopping a Sequence
Editing a Sequence
Deleting Commands
Inserting a Command
Modifying a Command
Clearing a Sequence from Memory
Changing the Sequence Title
Naming Files Generated by a Sequence
Storing a Sequence on a Disk
Loading a Sequence from Disk
Purging a Sequence from Disk
Printing a Sequence
Cascading Multiple Example Sequences
Loop Counter Example Sequence
Generating Files in a Loop Counter Example Sequence
Limit Test Example Sequence
................................
................
.............................
..................
............................
.........................
......................
...............................
.............................
............................
............................
............................
...........................
..........................
..........................
.....................
........................
....................
........................
.......................
.......................
............................
...................
......................
............
........................
Measuring a Device in the Time Domain (Option 010 Only)
Transmission Response in Time Domain
Reflection Response in Time Domain
Non-coaxial Measurements
..........................
...................
....................
...........
2-62
2-64
2-64
2-64
2-64
2-64
2-65
2-65
2-66
2-67
2-67
2-68
2-68
2-68
2-69
2-69
2-70
2-70
2-71
2-72
2-72
2-72
2-73
2-74
2-75
2-77
2-79
2-79
2-83
2-86
3.Making Mixer Measurements (Option 089 Only)
Where to Look for More Information
Measurement Considerations
....
Mmumzmg
Source and Load Mismatches
.........................
Reducing the Effect of Spurious Responses
Eliminating Unwanted Mixing and Leakage
HowRFandIFAreDeflned
Frequency Offset Mode Operation
.....................
..................
.................
Signals.
........................
......................
Differences Between Internal and External R-Channel Inputs
Power Meter Calibration
Conversion Loss Using the Frequency Offset Mode
High Dynamic Range Swept RF/IF Conversion Loss
Fixed IF Mixer Measurements
Tuned Receiver Mode
Sequence 1 Setup
Sequence 2 Setup
Phase or Group Delay Measurements
Amplitude and Phase Tracking
..........................
...............
..............
........................
...........................
.............................
.............................
.....................
........................
Conversion Compression Using the Frequency Offset Mode
Isolation Example Measurements
LO to RF Isolation
RF Feedthrough
.............................
..............................
.......................
.............
.........
...........
3-1
3-2
3-2
3-2
3-2
3-2
3-4
3-4
3-6
3-7
3-12
3-17
3-17
3-17
3-21
3-24
3-27
3-28
3-33
3-33
3-35
Contents-3
4.
Printing, Plotting, and Saving Measurement Results
Where to Look for More Information
Printing or Plotting Your Measurement Results
Configuring a Print Function
DellningaPrintFunction
.........................
..........................
If You Are Using a Color Printer
To Reset the Printing Parameters to Default Values
Printing One Measurement Per Page
Printing Multiple Measurements Per Page
Configuring
If You Are Plotting to an
a Plot Function
.........................
HPGL/2
If You Are Plotting to a Pen Plotter
IfYouArePlottingtoaDiskDrive
Detlning
a Plot Function
...........................
To Reset the Plotting Parameters to Default Values
Plotting One Measurement Per Page Using a Pen Plotter
Plotting Multiple Measurements Per Page Using a Pen Plotter
If You Are Plotting to an HPGL Compatible Printer
Plotting a Measurement to Disk
‘IbOutputthePlotFiles
TlbViewPlotFilesonaPC
usingAmiPro
Using Freelance
...............................
..............................
........................
..........................
..........................
Outputting Plot Files from a PC to a Plotter
.....................
.................
............
:.:
...
.....................
...................
Compatible Printer
.............
.....................
.....................
..............
............
..........
.............
..................
Outputting Plot Files from a PC to an HPGL Compatible Printer
Step 1. Store the HPGL initialization sequence.
...............
Step 2. Store the exit HPGL mode and form feed sequence.
Step 3. Send the HPGL initialization sequence to the printer.
Step 4. Send the plot
6Ie
to the printer.
...................
Step 5. Send the exit HPGL mode and form feed sequence to the printer.
OutputtingSiiePagePlotsUsingaPrinter.
Outputting Multiple Plots to a Siie Page Using a Printer
Printing or Plotting the List Values or Operating Parameters
IfYouWantaSiiePageofValues
IfYouWanttheEntireListofValues
.....................
....................
Solving Problems with Printing or Plotting
Saving and Recalling Instrument States
Places Where You Can Save
........................
....................
What You Can Save to the Analyzer’s Internal Memory
WhatYouCanSavetoaFloppyDisk
What You Can
Saving an Instrument State
Save
to a Computer
..........................
Saving Measurement Results
ASCII Data Formats
CITIfile
S2P
.................................
Data Format.
............................
............................
Re-Saving an Instrument State
Deleting a File
‘Ib
Delete an Instrument State
................................
.........................
........................
File
....................
.....................
.....................
.................
...........
..............
................
..................
..............
..........
..................
............
....:..:....
:
........
........
........
...
:
4-2
4-3
4-3
4-5
4-5
4-6
4-6
4-6
4-8
4-8
4-10
4-11
4-12
4-16
4-16
4-17
4-18
4-19
4-19
4-20
4-21
4-22
4-22
4-23
4-23
4-24
4-24
4-24
4-24
4-24
4-25
4-26
4-26
4-28
4-29
4-30
4-30
4-30
4-30
4-31
4-32
4-33
4-33
4-33
4-33
4-34
4-35
4-36
4-39
4-39
4-39
4-41
4-41
4-41
Contents4
‘IbDeleteallFiles
RenamingaFile
RecaIlingaFile
Formatting a Disk
Solving Problems with Saving or Recalling Files
IfYouAreUsinganExtemalDiskDrive.
.............................
...............................
...............................
..............................
................
..................
5. Optimizing Measurement Results
Where to Look for More Information
Increasing Measurement Accuracy
Connector Repeatability
Interconnecting Cables
Temperature Drift
Frequency Drift
.............................
..............................
Performance Verification
..........................
...........................
..........................
Reference Plane and Port Extensions
Measurement Error-Correction
Conditions Where Error-Correction Is Suggested
Types of Error-Correction
.........................
Error-Correction Stimulus State
Calibration Standards
...........................
Choosing Calibration Load Standards
.....................
......................
....................
........................
...............
.......................
...................
Compensating for the Electrical Delay of Calibration Standards
Chi.rifying Type-N
Connector Sex
When to Use Interpolated Error-Correction
Procedures for Error-Correcting Your Measurements
Frequency Response Error-Corrections
Response Error-Correction for Reflection Measurements
Response Error-Correction for Transmission Measurements
Receiver Calibration
............................
Frequency Response and Isolation Error-Corrections
.....................
.................
..............
....................
...........
..........
..............
Response and Isolation Error-Correction for Reflection Measurements
Response and Isolation Error-Correction for Transmission Measurements
One-Port Reflection Error-Correction
Full Two-Port Error-Correction
TRL and TRM Error-Correction
TRL Error-Correction
TRM Error-Correction
...........................
...........................
Modifying Calibration Kit Standards.
Definitions
................................
Outline of Standard Modification
Modifying Standards
Modifying TRL Standards.
Modifying TRM Standards
............................
.........................
.........................
Power Meter Measurement Calibration
Entering the Power Sensor Calibration Data
Editing Frequency Segments
Deleting Frequency Segments
Compensating for Directional Coupler Response
Using Sample-and-Sweep Correction Mode
Using Continuous Correction Mode
‘RI
Calibrate the Analyzer Receiver to Measure Absolute Power
Calibrating for Noninsertable Devices
Adapter Removal
Perform the
.............................
2-port
Error Corrections
.....................
........................
........................
.....................
......................
....................
.................
.......................
......................
...............
..................
.....................
.....................
...................
.......
.....
...
.......
4-41
4-42
4-42
4-43
4-43
4-43
5-2
5-2
5-2
5-2
5-2
5-3
5-3
5-3
5-4
5-4
5-4
5-5
5-6
5-6
5-6
5-7
5-7
5-8
5-9
5-9
5-11
5-12
5-14
5-14
5-16
5-18
5-21
5-24
5-24
5-26
5-28
5-28
5-28
5-29
5-30
5-32
5-35
5-36
5-36
5-37
5-37
5-38
5-39
5-40
5-41
5-42
5-43
Contents-5
Remove the Adapter
Verify the Results
Example Program
Matched Adapters
Modify the Cal Kit Thru Definition
Maintaining Test Port Output Power During Sweep Retrace
...........................
............................
............................
.............................
.....................
...........
Making Accurate Measurements of Electrically Long Devices
The Cause of Measurement Problems
To Improve Measurement Results
Decreasing the Sweep Rate
Decreasing the Time Delay
Using Stepped Sweep Mode
Increasing Sweep Speed
lb
Decrease the Frequency Span
...........................
........................
........................
.......................
‘IbSettheAutoSweepTimeMode
‘Lb
Widen the System Bandwidth
‘Lb
Reduce the Averaging Factor
‘lb
Reduce the Number of Measurement Points
TbSettheSweepType
....
..........................
....................
......................
......................
.....................
......................
......................
................
To View a Single Measurement Channel
lb
Activate Chop Sweep Mode
To Use External Calibration
‘lb
Use Fast Z-Port Calibration
Increasing Dynamic Range
‘lb
Increase the Test Port Input Power
‘lb
Reduce the Receiver Noise Floor
..........................
Changing System Bandwidth
Changing Measurement Averaging
Reducing Trace Noise
To Activate Averaging
............................
...........................
To Change System Bandwidth
Reducing Receiver CrosstaIk
Reducing Recall Time
............................
.......................
........................
.......................
....................
.....................
.......................
....................
.......................
.........................
..........
:: : : : : :
:
5-44
5-46
5-47
5-48
5-49
5-50
5-51
5-51
5-51
5-51
5-52
5-52
5-53
5-53
5-53
5-54
5-54
5-54
5-55
5-55
5-56
5-56
5-57
5-58
5-58
5-58
5-58
5-58
5-59
5-59
5-59
5-59
5-60
6.
Application and Operation Concepts
Where to Look for More Information
System Operation
The Built-In Synthesized Source
TheBuilt-InTestSet
The Receiver Block
The Microprocessor
Required Peripheral Equipment
Data Processing
Processing Details
Contents-6
.....................
..............................
......................
The Source Step Attenuator
.......................
............................
............................
............................
.......................
...............................
.............................
TheADC
IF Detection
Ratio Calculations
Sampler/IF Correction
Sweep-lb-Sweep Averaging
Pre-Raw Data Arrays
Raw Arrays
Edit Limits Menu
Edit Segment Menu
Offset Limits Menu.
Knowing the Instrument Modes
Network Analyzer Mode
Tuned Receiver Mode
Frequency Offset Menu (Option 089)
Primary Applications.
Typical
Test Setup
..........................
............................
...........................
............................
...........................
...........................
.......................
..........................
...........................
....................
..........................
............................
Frequency Offset In-Depth Description
The Receiver Frequency
The Offset Frequency (LO)
Frequency
Frequency Ranges
Hierarchy
...........................
........................
.......................
..........................
Compatible Instrument Modes and Sweep Types
Receiver and Source Requirements
Display Annotations
Error Message
.............................
Spurious Signal
Time Domain Operation (Option 010)
The Transform Menu.
General Theory
..............................
Time Domain Bandpass
Adjusting
the Relative Velocity
..........................
Passband
kequencies
.....................
...........................
..........................
Factor
..................
................
..................
...................
..................
.............
...................
..................
...................
6-102
6-102
6-102
6-102
6-103
6-104
6-104
6-104
6-104
6-104
6-105
6-106
6-106
6-106
6-107
6-107
6-107
6-107
6-108
6-109
6-109
6-110
6-110
6-110
6-110
6-110
6-111
6-111
6-111
6-112
6-112
6-113
6-113
6-114
6-115
6-115
6-115
6-116
6-116
6-116
6-117
6-117
6-117
6-117
6-117
6-118
6-118
6-118
6-118
6-118
6-119
6-119
6-120
6-121
6-121
Contents-10
Reflection Measurements Using Bandpass Mode
..............
Interpreting the bandpass reflection response horizontal axis .......
Interpreting the bandpass reflection response vertical axis
Transmission Measurements Using Bandpass Mode
.............
Interpreting the bandpass transmission response horizontal axis
........
.....
Interpreting the bandpass transmission response vertical axis .......
TimeDomainLowPass.
Setting Frequency Range for Time Domain Low Pass
Minimum Allowable Stop Frequencies
Reflection Measurements In Time Domain Low Pass
Interpreting the low pass response horizontal axis
Interpreting the low pass response vertical axis
Fault Location Measurements Using Low Pass
Transmission Measurements In Time Domain Low Pass
Measuring
small
..........................
............
..................
............
............
.............
...............
...........
signal transient response using low pass step
......
Interpreting the low pass step transmission response horizontal axis
Interpreting the low pass step transmission response vertical axis
.....
Measuring separate transmission paths through the test device using low
pass impulse mode
Time Domain Concepts
Masking
Windowing
Range
Resolution
.................................
...............................
..................................
................................
Response resolution
Range resolution
Gating
.................................
Setting the gate
............................
............................
Selecting gate shape
Transforming CW Time Measurements into the Frequency Domain
Forward Transform Measurements
Interpreting the forward transform vertical axis
Interpreting the forward transform horizontal axis
Demodulating the results of the forward transform
Forward transform range
Test
Sequencing
...............................
In-Depth Sequencing Information
Features That Operate Differently When Executed In a Sequence
.........................
...........................
..........................
..........................
......
....................
.............
............
...........
........................
......................
......
Commands That Sequencing Completes Before the Next Sequence Command
Begins
Commands That Require a Clean Sweep
Forward Stepping In Edit Mode
Titles..
Sequence Size
Embedding the Value of the Loop Counter In a Title
Autostarting Sequences
The GPIO Mode
The Sequencing Menu
Gosub
Sequence Command
lTLI/OMenu
TI’L
Output for Controlling Peripherals
‘ITL
Input Decision Making
‘ITLOutMenu
Sequencing Special Functions Menu
Sequence Decision Making Menu
2-2. Example of Viewing Both Primary Channels with a Split Display .......
2-3. Example of Viewing Both Primary Channels on a Single Graticule
2-4. Example of a Display Title
2-5.
2-6.
2-7. Duplexer Measurement
2-8. Active Marker Control
2-9. Active and Inactive Markers
2-10. Marker Information Moved into the
2-11. Marker Information on the Graticules
2-12. Marker 1 as the Reference Marker
2-13. Example of a Fixed Reference Marker Using
2-14. Example of a Fixed Reference Marker Using m
2-15. Example of Coupled and Uncoupled Markers
2-16. Example of a Log Marker in Polar Format
2-17. Example of Impedance Smith Chart Markers
2-18. Example of Setting the Start Frequency Using a Marker...........
2-19. Example of Setting the Stop Frequency Using a Marker...........
2-20. Example of Setting the Center Frequency Using a Marker..........
2-21. Example of Setting the Frequency Span Using Markers ............
2-22. Example of Setting the Reference
2-23. Example of Setting the Electrical Delay Using a Marker...........
2-24. Example of Searching for the Maximum Amplitude Using a Marker ......
2-25. Example of Searching for the Minimum Amplitude Using a Marker
2-26. Example of Searching for a
2-27. Example of Searching for a Bandwidth Using Markers ............
2-28. Example Statistics of Measurement Data
2-29. Device Connections for Measuring a Magnitude Response
2-30. Example Magnitude Response Measurement Results
2-31. Example Insertion Phase Response Measurement
2-32. Phase Samples
2-33. Device Connections for Measuring Electrical Length
2-34. Linearly Changing Phase
2-35. Example Best Flat Line with Added Electrical Delay
2-36. Deviation from Linear Phase Example Measurement .............
2-37. Group Delay Example Measurement
2-38. Group Delay Example Measurement with Smoothing
2-39. Group Delay Example Measurement with Smoothing Aperture Increased
2-40. Connections for SAW Filter Example Measurement
2-41. Example Flat Limit Line
2-42. Example Flat Limit Lines
8719D/20D/22D
8719D/20D/22D
3-Channel
4-Channel
Display
Display
Front Panel
Rear Panel
.............................
.............................
...........................
...........................
...............................
..........................
..........................
..........................
.......................
.............
.......................
.........................
......
.........................
........................
Softkey
.....................
Value
‘Ihrget
Amplitude Using a Marker .........
Menu Area
....................
AKEF=AFIm l%KI%
ZERO
................
..................
................
Using a Marker ...........
..................
..............
....................
..............
...........
.......
...........
......
...........
.............
.............
.............
.............
...
l-4
l-6
l-10
2-3
2-5
2-6
2-8
2-10
2-11
2-15
2-17
2-17
2-18
2-19
2-20
2-21
2-22
2-23
2-24
2-25
2-26
2-26
2-27
2-28
2-28
2-29
2-30
2-30
2-31
2-32
2-33
2-34
2-35
2-35
2-36
2-37
2-38
2-39
2-40
2-41
2-41
2-42
2-43
2-45
2-46
Contents-16
2-43. Sloping Limit Lines
2-44. Example Single Points Limit Line
2-45. Example Stimulus Offset of Limit Lines
2-46. Diagram of Gain Compression
2-47. Gain Compression Using Linear Sweep and
2-48. Gain Compression Using Linear Sweep and
2-4g. Gain Compression using Power Sweep
............................
......................
...................
.......................
~~~A~~
.................................................
,D2&?2 Ax+: D2 O!l
“,“” “‘:.
......
.....
............
,““““.““““~.
..............
........
,’ . . . .
. . . . .
2-50. Gain and Reverse Isolation.........................
2-51. High Power Test Setup (Step 1)
2-52. High Power Test Setup (Step
2-53. High Power Test Setup (Step
2-54. Internal
2-55. High Power
2-56.
Typical Test
SiiaI
Paths of Analyzer......................
‘Pest
Setup (Step 3).......................
Setup for Tuned Receiver Mode.................
2-57. Test Sequencing Help Instructions
2-58. Device Connections for Time Domain Transmission Example Measurement
2-59. Time Domain Transmission Example Measurement
2-60. Gating in a Time Domain Transmission Example Measurement
.......................
2a)
......................
2b)
......................
......................
..............
........
. .
2-61. Gating Effects in a Frequency Domain Example Measurement ........
2-62. Device Connections for Reflection Time Domain Example Measurement
....
2-63. Device Response in the Frequency Domain .................
2-64. Device Response in the Time Domain
3-l.
Down Converter Port Connections .....................
....................
3-2. Up Converter Port Connections.......................
3-3. R-Channel External Connection
3-4. An Example Spectrum of RF, LO, and IF
.......................
Signals
Present in a Conversion Loss
Measurement..............................
3-5. Connections for R Channel and Source Calibration
..............
3-6. Connections for a One-Sweep Power Meter Calibration for Mixer Measurements
3-7. Diagram of Measurement Frequencies....................
3-8. Measurement Setup from Display
3-9. Conversion Loss Example Measurement
3-10. Connections for Broad Band Power Meter Calibration
3-l
1. Connections for Receiver Calibration
......................
...................
............
....................
3-12. Connections for a High Dynamic Range Swept IF Conversion Loss Measurement
3-13. Example of Swept IF Conversion Loss Measurement
3-14. Connections for a Response Calibration
3-15. Connections for a Conversion Loss Using the
...................
Tuned
.............
Receiver Mode
......
3-16. Example Fixed IF Mixer Measurement ...................
3-17. Connections for a Group Delay Measurement................
3-18. Group Delay Measurement.........................
3-19. Conversion Loss and Output Power as a Function of Input Power Level....
3-20. Connections for the First Portion of Conversion Compression Measurement
. .
3-21. Connections for the Second Portion of Conversion Compression Measurement .
3-22. Measurement Setup Diagram Shown on Analyzer Display
3-23. Example Swept Power Conversion Compression Measurement
3-24.
SiiaIFIowinaMixer
3-25. Connections for a Response Calibration
3-26. Connections for a Mixer Isolation Measurement
3-27. Example Mixer LO to RF Isolation Measurement
3-28. Connections for a Response Calibration
...........................
...................
...............
...............
...................
3-29. Connections for a Mixer RF Feedthrough Measurement
6-41. Open Circuit Termination
6-42. Measured
6-43.
Major
Sources of Error
6-44.
Transmission Coefficient
6-45. Load Match
6-46. Isolation
6-47.
FuIi
6-48.
FuII
EXF
Two-Port Error Model
Two-Port Error Model Equations
...............................
S11
..............................
ELF
...............................
..........................
...........................
..........................
6-49. Typical Responses of Calibration Standards after Calibration
6-50. Response versus
6-51. Response versus
6-52. Response versus
6-53. HP
8719D/20D/22D
Sll
l-Port Calibration on Log Magnitude Format
Sll
l-Port Calibration on Smith Chart
FuII
Two-Port Calibration
functional block diagram for a 2-port error-corrected
measurement& system.
6-54.
g-term TRL*
error model and generalized coefficients.
6-55. Comparison of Standard and Option 400 Instruments
6-56. Typical Measurement Setup
6-57.
Test
Setup for Continuous Sample Mode
6-58.
Test
Setup for Sample-and-Sweep Mode
6-59. Alternate and Chop Sweeps Overlaid
6-60. Instrument State Function Block
6-61. Typical Test Setup for a Frequency Offset Measurement
6-62. Device Frequency Domain and Time Domain Reflection Responses
6-63. A Reflection Measurement of Two Cables
6-64. Transmission Measurement in Time Domain Bandpass Mode
6-65. Time Domain Low Pass Measurements of an Unterminated Cable
6-66. Simulated Low Pass Step and Impulse Response Waveforms
6-67. Low Pass Step Measurements of Common Cable Faults
6-68. Time Domain Low Pass Measurement of an AmpIifier
Response
...............................
6-69. Transmission Measurements Using Low Pass Impulse Mode
6-70. Masking Example
.............................
6-71. Impulse Width, Sidelobes, and Windowing
6-72. The Effects of Windowing on the Time Domain Responses of a Short Circuit .
6-73. Response Resolution
............................
6-74. Range Resolution of a Siie Discontinuity
6-75. Sequence of Steps in Gating Operation
6-76. Gate Shape
................................
6-77. Amplifier Gain Measurement
6-78. Combined Effects of Amplitude and Phase Modulation
6-79. Separating the Amplitude and Phase Components of Test-Device-Induced
Modulation
...............................
6-80. Range of a Forward Transform Measurement
6-81.
ParaIIel
6-82. Amplifier Parameters
6-83. Diagram of Gain Compression
6-84. Gain Compression Using Power Sweep
6-85. Test Configuration for Setting RF Input using Automatic Power Meter Calibration
6-86. Mixer Parameters
6-87. Conversion Loss versus Output Frequency Without Attenuators at Mixer Ports
6-88. Example of Conversion Loss versus Output Frequency Without Correct IF
Port Input and Output Bus Pin Locations
............................
.......................
...................
.............................
Path Filtering
.............................
.m’GPIb mode
. 1 : : 1 : :
SiiaI
1
6-143
6-147
6-148
6-149
6-151
6-152
6-153
6-89. Example of Conversion Loss versus Output Frequency With Correct IF Signal
Path Filtering and Attenuation at
6-90. Examples of Up Converters and Down Converters
6-91. Down Converter Port Connections
6-92. Up Converter Port Connections
alI
Mixer Ports
.............
..............
.....................
.......................
6-153
6-154
6-155
6-156
6-93. Example Spectrum of RF, LO, and IF signaIs Present in a Conversion Loss
Measurement.
6-94. Main Isolation Terms
.............................
............................
6-95. Conversion Loss and Output Power as a Function of Input Power Level
....
6-157
6-157
6-159
6-96. Connections for an Amplitude and Phase Tracking Measurement Between Two
Mixers
6-97. Adapter Considerations
7-l. External Trigger Circuit
11-l.
Peripheral Connections to the Analyzer
11-2. HP-IB Bus Structure
11-3. Analyzer Siie Bus Concept
.................................
..........................
..........................
...................
............................
........................
6-160
6-162
11-10
11-18
11-21
6-138
6-139
6-148
7-24
Contents-20
‘Ihbles
2-l.
Connector Care Quick Reference
2-2. Gate Characteristics
............................
4-l. Default Values for Printing Parameters...................
4-2. Default Pen Numbers and Corresponding Colors
4-3. Default Pen Numbers for Plot Elements...................
4-4. Default Line Types for Plot Elements
4-5. Plotting Parameter Default Values
4-6. HPGL Initialization Commands
4-7. HPGL Test
5-l.
Differences between PORT EXTENSIONS and ELECTRICAL DELAY
File
Commands
5-2. Purpose and Use of Different Error-Correction Procedures
5-3. Frequency Cutoff Points of Load Standards
5-4. Typical Calibration Kit Standard and Corresponding Number
5-5. Characteristic Power Meter Calibration Speed and Accuracy
6-3. Display Colors with Maximum Viewing Angie
6-4. Calibration Standard Types and Expected Phase Shift
6-5. Standard Definitions
............................
6-6. Standard Class Assignments
6-7. Characteristic Power Meter Calibration Speed and Accuracy
6.8. Time Domain Reflection Formats
6-9. Minimum Frequency Ranges for Time Domain Low Pass
6-10. Impulse Width,
6-l
1. Gate Characteristics
7-l. HP
7-2. HP
8719D/8720D
8722D
Sidelobe
Level, and Windowing Values
............................
Characteristics Without Error-Correction
Characteristics Without Error-Correction
7-3. Instrument Specifications (1 of 4)
9-l. Cross Reference of Key Function toProgr
9-2. Softkey Locations
11-l.
Default Addresses for HP-IB Peripherals...................
11-2. Code Naming Convention
.............................
.........................
12-1. Memory Requirements of Calibration and Memory Trace Arrays
12-2.
Suffix
Character Definitions
12-3. Preset Conditions (1 of 5)
.........................
12-4. Power-on Conditions (versus Preset)
12-5. Results of Power Loss to Non-Volatile Memory
......................
...............
....................
.....................
.......................
.........................
.....
..........
.................
.........
.........
......................
................
............
........................
.........
......................
...........
............
..........
.............
......................
amming Command
.........
.......
........................
....................
................
2-2
2-82
4-6
4-13
4-13
4-14
4-16
4-23
4-24
5-3
5-5
5-6
5-29
5-35
5-53
5-60
6-19
6-44
6-49
6-73
6-84
6-87
6-105
6-123
6-124
6-131
6-136
7-3
7-3
7-4
9-53
9-74
11-13
11-24
12-3
12-5
12-8
12-12
12-12
Contents-21
HP 8719D/ZOD/ZZD Description and Options
This chapter contains information on the following topics:
n
Analyzer Overview
n
Analyzer Description
w
Front Panel Features
w
Analyzer Display
n
Rear Panel Features and Connectors
n
Analyzer Options Available
n
Service and Support Options
Where to Look for More Information
Additional information about many of the topics discussed in this chapter is located in the
following areas:
1
w
Chapter 2, “Making Measurements,” contains step-by-step procedures for making
measurements or using particular functions.
w
Chapter 4, “Printing, Plotting, and Saving Measurement Results,” contains instructions
for saving to disk or the analyzer internal memory, and printing and plotting displayed
measurements
w
Chapter 5, “Optimizing Measurement Results,” describes techniques and functions for
achieving the best measurement results.
n
Chapter 6, “Application and Operation Concepts, n contains explanatory-style information
about many applications and analyzer operation.
HP 8718D120D/22D Descriptionand Options
l-1
Analyzer Description
The HP
8719D/20D/22D
is a high performance vector network analyzer for laboratory or
production measurements of reflection and transmission parameters. It integrates a high
resolution synthesized RF source, an S-parameter test set, and a four-channel three-input
receiver (four-input receiver, Option 400) to measure and display magnitude, phase, and group
delay responses of active and passive RF networks.
Two independent primary channels, two auxiliary channels, and a large screen color
display
show the measured results of one or all channels, in cartesian or polar/Smith chart formats.
For information on options, refer to “Options Available” later in this chapter.
The analyzer has the additional following features:
n Measurement functions selection with front panel keys and softkey menus.
w
Simultaneous viewing of all four S-parameters by enabling the auxiliary channels 3 and 4.
n Direct print or plot output of displayed measurement results, with a time stamp if desired, to
a compatible peripheral with a serial, parallel, or HP-IB interface.
w
Instrument states storage in internal memory for the following times, or on disk
number of points, averaging mode, frequency range, and sweep type.
w
Built-in service diagnostics are available to simplify troubleshooting procedures.
w
Performance improvement and flexibility through trace math, data averaging, trace
smoothing, electrical delay, and accuracy enhancement.
n Accuracy enhancement methods that range from normalizing data to complete one or two
port vector error correction with up to 1601 measurement points, and
TRL*/LRM*.
(Vector
error correction reduces the effects of system directivity, frequency response, source and
load match, and crosstalk.)
w
‘True TRL” measurement capability with Option 400. This option includes a four-input
receiver to improve TRL calibration accuracy for in-fixture and on-wafer measurements.
w
Complete reflection and transmission measurements in a 50 ohm impedance environment.
w
Receiver/source frequency offset mode (Option 089) that allows you to set the analyzer’s
receiver and source with a
w
Power meter calibration that allows you to use an HP-IB compatible power meter to monitor
lixed
frequency offset for mixer test applications
and correct the analyzer’s output power at each data point. (The analyzer stores a power
correction table that contains the correction values.)
n Test system automation with the addition of an external controller. This allows all of the
analyzer’s measurement capabilities to be programmed over the Hewlett-Packard Interface
Bus (HP-IB). (Refer to the “Compatible Peripherals” chapter or the HP
Network
Analgzer Programmin9
Guide.)
87190/200/220
l-2
HP
8719DI20D/22D
Description and Options
w
External keyboard compatibility that
n
LIF/DOS
w
Integration of a high capacity micro-floppy disk drive.
w
Internal automation, using test sequencing to program analyzer measurements and control
disk formats for saving instrument states and measurement data.
allows
you to title files and control the analyzer.
other devices without an external controller.
w
A general purpose input/output (GPIO) bus that can control eight output bits and read five
input bits through test sequencing. This can be useful for interfacing to material handlers or
custom test sets.
HP 87 1
gD/200/22D
Description and Options
l-3
Front Panel Features
Figure l-l. HP
F’igure
These features are described in more detail later in this chapter, and in the “Key
2.
3.
4.
6.
l-l shows the location of the following front panel features and key function blocks.
chapter.
LINE
1.
5.
7.
switch. This switch controls ac power to the analyzer. 1 is on, 0 is off.
Display. This shows the measurement data traces, measurement annotation, and softkey
labels. The display is divided into specific information areas, illustrated in Figurel-2.Disk drive. This 3.5 inch drive allows you to store and recall instrument states and
measurement results for later analysis.
Softkeys. These keys provide access to menus that are shown on the display.
STIMULUS function block. The
source’s frequency, power, and other stimulus functions.
RESPONSE function block. The
and display functions of the active display channel.
ACTIVE
auxiliary channels. These keys allow you to select the active channel. Any function you
enter applies to the selected active channel.
CHANNEL keys, The analyzer has two independent primary channels and two
8719D/20D/22D
keys in this block allow you to control
keys in this block allow you to control the measurement
Front Panel
Definitions”
the
analyzer
14 HP87190/200/2200sscriptionand
Options
8.
The ENTRY block.
This block includes the knob, the step QD QD keys, the number pad,
and the backspace @ key. These allow you to enter numerical data and control the
markers.
You can use the numeric keypad to select digits, decimal points, and a minus sign for
numerical entries. You must also select a units terminator to complete value inputs.
The backspace key @ has two independent functions: it modifies or deletes entries, and
it turns off the softkey menu so that marker information can be
moveed
off of the grids
and into the softkey menu area.
9.
INSTRUMENT STATE function block.
These keys allow you to control
channel-independent system functions such as the following:
n
copying, save/recall, and HP-IB controller mode
n
limit testing
n
tuned receiver mode
n
frequency offset mode (Option 089)
n
test sequence function
w
time domain transform (Option 010)
HP-IB STATUS indicators are
10.
w
key. This key returns the instrument to either a known factory preset state, or a
user preset state that can be
also
included in this block.
defined.
Refer to the “Preset State and Memory Allocation”
chapter for a complete listing of the instrument preset condition.
11.
R CHANNEL
co~ectors.
These connectors allow you to apply an input signal to the
analyzer’s R channel, for frequency offset mode.
12.
PORT 1 and PORT 2.
These ports output a signal from the source and receive input
signals from a device under test. PORT 1 allows you to measure
The analyzer display shows various measurement information:
w
The grid where the analyzer plots the measurement data.
n The currently selected measurement parameters.
w
The measurement data traces.
Figure 1-2 illustrates the locations of the different information labels described below.
In addition to the full-screen display shown in Figure l-2, a split display is available, as
described in the “Making Measurements” chapter. In the split display mode, the analyzer
provides information labels for each half of the display.
Several display formats are available for different measurements, as described under
in the “Key
1.
Stimulus start value. This value could be any one of the following:
w
w
n The lower power value in power sweep
Deli&ions”
chapter.
The start frequency of the source in frequency domain measurements
The start time in CW mode (0 seconds) or time domain measurements
“I=)”
When the stimulus is in center/span mode, the center stimulus value is shown in this
space. The color of the stimulus display reflects the current active channel.
l-6
HP
87190/2001220
Description and Options
Stimulus stop Value. This value could be any one of the following:
2.
w
The stop frequency of the source in frequency domain measurements.
w
The stop time in time domain measurements or CW sweeps.
n
The upper limit of a power sweep.
When the stimulus is in center/span mode, the span
values can be blanked, as described under U
Deli&ions”
chapter.
~~~~,.;.?~~~
isshown
in this space. The stimulus
Key” in the “Key
(For CW time and power sweep measurements, the CW frequency is displayed centered
between the start and stop times or power values.)
3.
Status Notations. This area shows the current status of various functions for the active
channel.
The following notations are used:
Avg =
Cor =
Sweep-to-sweep averaging is on. The averaging count is shown immediately
below (See
“m
Key” in the “Key Definitions” chapter.)
Error correction is on. (For error-correction procedures, refer to the “Optimizing
Measurement Results” chapter. For error correction theory, refer to the
“Application and Operation Concepts” chapter.)
CA
=
Stimulus parameters have changed from the error-corrected state, or interpolated
error correction is on. (For error-correction procedures, refer to the “Optimizing
Measurement Results” chapter. For error correction theory, refer to the
*Application and Operation Concepts” chapter.)
c2
=
Full two-port error-correction is active and
is different (~coupled), or the ~~~~~,~~~~:~~~~~~i is a&iv&,ed. me mot&on
.:: T.. ..i
i:........
either..the
,.A. .,.,..,.,.
:...,.,.,.,,/
... .:
. . . . .
s..;;; ./ ii..;;
..
. . . .
.T i..z:.
power range for each port
occurs because the analyzer does not switch between the test ports every sweep
under these conditions. The measurement stays on the active port after an
initial cycling between the ports. (The active port is determined by the selected
measurement parameter.) You can update all the parameters by pressing
:.
,,, *. ..%
,” ) .<wc<c<..
;,<::=:-
~~~~~~~~
:.,::~~~~~.,.;:.~.~;;,.,.,~.;
.
......
. . . . . . . . . . .
s.2;
“‘:<:z. : :,‘,::.~..:.
or
. ...//....
9
..~........
IMEAS)
key.
LMenu)
Note
Del =
ext
Ofs =
Of?
Gat =
On instruments equipped with Option 007, C2 should be displayed at all times if
a full two-port error-correction is active.
Electrical delay has been added or subtracted, or port extensions are active. (See
the “Application and Operation Concepts” chapter and
“@iZZ)
Key” in the
“Key Definitions” chapter.)
Waiting for an external trigger.
=
Frequency offset mode is on (Option 089 only). (See “Frequency Offset
Operation” in the “Application and Operation Concepts” chapter.)
=
Frequency offset mode error (Option 089 only), the IF frequency is not within
10 MHz of expected frequency. LO inaccuracy is the most likely cause. (See
“Frequency Offset Operation” in the “Application and Operation Concepts”
chapter.)
Gating is on (time domain Option 010 only). (For time domain measurement
procedures, refer to the “Making Measurements” chapter. For time domain
theory, refer to the “Application and Operation Concepts” chapter.)
HP
87190/200/220
Description and Options
l-7
Hld =
Hold sweep. (See HOLD in the “Key Definitions” chapter.)
man=
PC =
PC? =
P? =
P1
=
PRm
=
Smo =
tsH
=
Waiting for manual trigger.
Power meter calibration is on. (For power meter calibration procedures, refer
to the “Optimizing Measurement Results” chapter. For power meter calibration
theory, refer to the “Application and Operation Concepts” chapter.)
The analyzer’s source could not be set to the desired level, following a power
meter calibration. (For power meter calibration procedures, refer to the
“Optimizing Measurement Results” chapter. For power meter calibration theory,
refer to the “Application and Operation Concepts” chapter.)
Source power is unleveled at start or stop of sweep. (Refer to the
HP
8719~/20~/2?2~
Network
Anulym
Semrice Guide
for troubleshooting.)
Source power has been automatically set to minimum, due to receiver overload.
(See
PB?ER
in the “Key
Delinitions”
chapter.)
Power range is in manual mode.
Trace smoothing is on. (See
“(XJ’
in the “Key Definitions” chapter.)
Indicates that the test set hold mode is engaged.
That is, a mode of operation is selected which would cause repeated switching of
the step attenuator or mechanical transfer switch (Option 007). This hold mode
may be overridden. See
MEASURE lV%TART
or
EBIE%R
OF
GROUFS
in the “Key
Definitions” chapter.
I=
Fast sweep indicator. This symbol is displayed in the status notation block when
sweep time is less than 1.0 second. When sweep time is greater than 1.0 second,
this symbol moves along the displayed trace.
*=
Source parameters changed: measured data in doubt until a complete fresh
sweep has been taken.
4.
Active Entry Area.
5.
Message Area.
6.
Title. This is a descriptive alpha-numeric string title that you
This displays the active function and its current value.
This displays prompts or error messages.
define
and enter through
attached keyboard or as described in the “Printing, Plotting, and Saving Measurement
Results” chapter.
7.
Active Channel.
@G-ily(Chan 3) and CchanJ(Ch
This is the label of the current active channel, selected with the
an 4) keys. The active channel is denoted by a rectangle
around the channel number. If multiple channels are overlaid, the labels will appear in
this area. For multiple-graticule displays, the channel information labels will be in the
same relative position for each graticule.
8.
Measured Input(s).
measured, as selected using the
This shows the S-parameter, input, or ratio of inputs currently
LMeas)
key. Also indicated in this area is the current
display memory status.
9.
Format.
10.
Scale/Div.
This is the display format that you selected using the
This is the scale that you selected using the
@iXZj
Cj)
key.
key, in units appropriate
to the current measurement.
a.n
Reference Level.
11.
This value is the reference line in Cartesian formats or the outer circle
in polar formats, whichever you selected using the
1-8HP 87 18D120DI22D Description and Options
&ZEZRef)
key. The reference level is
also indicated by a small triangle adjacent to the graticule, at the left for channel 1 and at
the right for channel 2 in Cartesian formats.
12.
Marker Values. These are the values of the active marker, in units appropriate to the
current measurement. (Refer to “Using Analyzer Display Markers” in the “Making
Measurements” chapter.)
13.
Marker Stats, Bandwidth. These are statistical marker values that the analyzer
calculates when you access the menus with the
(jMarkerj
key. (Refer to “Using Analyzer
Display Markers” in the “Making Measurements” chapter.)
14.
Softkey
Labels. These menu labels redehne the function of the softkeys that are located
to the right of the analyzer display.
15.
Pass l%il.
During limit testing, the result will be annunciated as PASS if the limits are not
exceeded, and FAIL if any points exceed the limits.
HP 87
190/200/220
Oessription and Options
1-g
Rear Panel Features and Connectors
Figure 1-3. HP
8719D/20D/22D
Rear Panel
Figure l-3 illustrates the features and connectors of the rear panel, described below.
Requirements for input signals to the rear panel connectors are provided in the “Specifications
and Measurement Uncertainties” chapter.
1.
HP-IB connector. This allows you to connect the analyzer to an external controller,
compatible peripherals, and other instruments for an automated system. Refer to the
“Compatible Peripherals” chapter in this document for HP-IB information, limitations, and
configurations.
2.
PARALLEL interface. This connector
allows
the analyzer to output to a peripheral with
a parallel input. Also included, is a general purpose input/output (GPIO) bus that can
control eight output bits and read five input bits through test sequencing. Refer to the
“Compatible Peripherals” chapter for information on
conllguring
a peripheral. Also refer
to “Application and Operation Concepts” for information on GPIO.
3.
NJ-232
RS-232
4.
KEYBOARD input
interface. This connector
(serial) input.
(DIN). This connector allows you to connect an external keyboard.
allows
the analyzer to output to a peripheral with an
This provides a more convenient means to enter a title for storage files, as well as
substitute for the analyzer’s front panel keyboard. The keyboard must be connected to
the analyzer before the power is switched on.
5.
Power cord receptacle, with fuse. For information on replacing the fuse, refer to the
HP
8719D~OD/ZZD
HP
8719D/2OD/ZZD
Network Anu&ze.r
Network Ana1gw.r
Ihstallation
Semrice
Guide.
and Quick
Sturt
Guideor the
6.
Line voltage selector switch. For more information refer to the HP
Network
7.
F&r.
8.
10 MHZ
l-10
HP 8719012001220 Oessriptionand Options
Analgwr
Installation and Quick Start Guide.
This fan provides forced-air cooling for the analyzer.
PRECISION REFEBENCE
OUTPUT. (Option lD5)
8719DL?OD/ZZD
9.
10 MHZ REFERENCE ADJUST. (Option
10.
EXTERNAL REFERENCE INPUT connector. This allows for a frequency reference signal
lD5)
input that can phase lock the analyzer to an external frequency standard for increased
frequency accuracy.
The analyzer automatically enables the external frequency reference feature when a
signal is connected to this input. When the signal is removed, the analyzer automatically
switches back to its internal frequency reference.
AUXILIARY INPUT connector. This allows for a dc or ac voltage input from an external
11.
signal source, such as a detector or function generator, which you can then measure, using
the S-parameter menu. (You can also use this connector as an analog output in service
routines, as described in the service manual.)
EXTERNAL AM connector. This allows for an external analog signal input that is
12.
applied to the ALC circuitry of the analyzer’s source. This input analog signal amplitude
modulates the RF output signal.
EXTERNAL TRIGGER connector. This allows connection of an external negative-going
13.
ll’L-compatible signal that will trigger a measurement sweep. The trigger can be set to
external through
softkey
functions.
TEST SEQUENCE. This outputs a
14.
to be high or low, or pulse (10
TTL
signal that can be programmed in a test sequence
/Iseconds)
high or low at the end of a sweep for robotic part
handler interface.
LIMIT
15.
n
n
MEASURE RESTART. This allows the connection of an optional foot switch. Using the
16.
foot switch will duplicate the key sequence
TEST SET
17.
BIASINPUTS AND FUSES. These connectors bias devices connected to port 1 and
18.
TEST. This outputs a ‘ITL signal of the limit test results as follows:
Pass:TrLhigh
Fail:
TI’L
low
(Meas) MEASURE AESTART
INTERCONNECI’.
Not Used
.
port 2. The fuses (1 A, 125 V) protect the port 1 and port 2 bias lines.
Serial number plate. The serial number of the instrument is located on this plate.
19.
RF IN/OUT. (Option 085) This allows the connection of an optional booster amplifier to
20.
increase the output power of the analyzer.
EXTERNAL MONITOB: VGA. VGA output connector provides analog red, green, and blue
21.
video signals which can drive a VGA monitor.
HP 97
19D/200/22D
Description and Options
1-l
1
Analyzer Options Available
Option
Option lD5 offers
lD6,
High Stability Frequency Reference
f0.05
ppm temperature stability from 0 to 55 OC (referenced to 25
“C).
Option 007, Mechanical Transfer Switch
This option replaces the solid state transfer switch with a mechanical switch in the test set,
providing the instrument with greater power handling capability. Because the mechanical
transfer switch has less loss than the standard switch, the output power of Option 007
instruments is 5 dB higher.
Option
This option is designed to permit the measurement of high power devices. With an external
power amplifier, this configuration will
ports. The maximum test port input power is 1 Watt (+ 30
front panel allow the insertion of high power attenuators or isolators. This allows test device
output levels up to the power limits of the inserted components Additionally, there is an
external reference input that allows the external amplifier’s frequency response and drift
to be ratioed out, and there are internally controlled step attenuators between the couplers
and samplers to prevent overload. A network analyzer with this option can be
to operate as a normal instrument (with slightly degraded output power level and accuracy)
or as an instrument capable of making single connection multiple measurements. Because of
high output power, Option 085 is only available with a mechanical transfer switch similar to
Option 007.
086,
High Power System
allow
up to 20 Watts (+ 43
dBm)
of output at the test
dBm)
CW, but jumpers on the
configured
Option 089, Frequency Offset Mode
This option adds the ability to offset the source and receiver frequencies for frequency
translated measurements This provides the instrument with mixer measurement capability. It
also provides a graphical setup that allows easy configuration of your mixer measurement.
Option 012, Direct Access Receiver Configuration
This option provides front panel access to the A and B samplers. This allows direct access to
the sampler inputs for improved sensitivity in applications such as antenna tests, or for the
insertion of attenuators between the couplers and samplers to allow measurements of up to 1
Watt (+ 30
configuration for the HP
Option 400, Four-Sampler
This option reconiigures the instrument’s test set to ratio out the characteristics of the test port
transfer switch, and to include a second reference channel that allows full accuracy with a TRL
measurement calibration.
dBm)
at the input of the test ports. Direct access to the B sampler provides a test
87221)
that gives increased dynamic range in the forward direction.
‘I&t
Set
l-12
HP
871901200/220
Description and Options
Option 010, Time Domain
This option allows the analyzer to display the time domain response of a network by computing
the inverse Fourier transform of the frequency domain response. The analyzer shows the
response of a test device as a function of time or distance. Displaying the reflection coefficient
of a network versus time determines the magnitude and location of each discontinuity.
Displaying the transmission coefficient of a network versus time determines the characteristics
of individual transmission paths. Time domain operation retains all accuracy inherent with the
active error correction.
Option
Option
mount the instrument, with handles detached, in an equipment rack with 482.6 mm (19 inches)
horizontal spacing.
Option
Option
mount the instrument with handles attached in an equipment rack with 482.6 mm (19 inches)
spacing.
lCM,
Rack Mount Flange Kit Without Handles
1CM
is a rack mount kit containing a pair of flanges and the necessary hardware to
lCP,
Rack Mount Flange Kit With Handles
1CP
is a rack mount kit containing a pair of flanges and the necessary hardware to
Service and Support Options
Hewlett-Packards offers many repair and calibration options for your analyzer. Contact the
nearest Hewlett-Packard sales or service office for information on options available for your
analyzer. See the table titled “Hewlett-Packard Sales and Service Offices” in the front of this
manual for a table of sales and service offices.
HP 8719D/20D122D Description and Options
l-13
Making Measurements
This chapter contains the following example procedures for making measurements or using
particular functions:
n
Basic Measurement Sequence and Example
0
Setting frequency range
0
Setting source power
w
Using the Analyzer Display Functions
n
Four-Parameter Display Mode
n
Using the Analyzer Display Markers
n
Measuring Magnitude and Insertion Phase Response
n
Measuring Electrical Length and Phase Distortion
q
Deviation from linear phase
q
Group delay
n
Testing a Device with Limit Lines
n
Measuring Gain Compression
n
Measuring Gain and Reverse Isolation Simultaneously
n
High Power Measurements (Option 085)
n
Measurments Using the Tuned Receiver Mode
n
Test Sequencing
n
Measuring a Device in the Time Domain
q
Transmission response in the time domain
q
Reflection response in the time domain
w
Non-coaxial Measurements
2
Where to Look for More Information
Additional information about many of the topics discussed in this chapter is located in the
following areas:
n
Chapter 4, “Printing, Plotting, and Saving Measurement Results,” contains instructions
for saving to disk or the analyzer internal memory, and printing and plotting displayed
measurements.
n
Chapter 5, “Optimizing Measurement Results,” describes techniques and functions for
achieving the best measurement results
n
Chapter 6, “Application and Operation Concepts,” contains explanatory-style information
about many applications and analyzer operation.
n Chapter 9, “Key
n Chapter 11, “Compatible Peripherals,” lists measurement and system accessories, and other
applicable equipment compatible with the analyzers.
DeWitions,”
describes all the front panel keys and
softkeys.
Making Measurements
2-l
Principles of Microwave Connector Care
Proper connector care and connection techniques are critical for accurate, repeatable
measurements.
Refer to the calibration kit documentation for connector care information. Prior to making
connections to the network analyzer, carefully review the information about inspecting,
cleaning and gaging connectors.
Having good connector care and connection techniques extends the life of these devices. In
addition, you obtain the most accurate measurements.
This type of information is typically located in Chapter 3 of the calibration kit manuals.
For additional connector care instruction, contact your local Hewlett-Packard Sales and Service
Office about course numbers HP
See the following table for quick reference tips about connector care.
8505OA
‘lhble
2-1. Connector Care Quick Reference
+
24A
and HP
8505OA
+
24D.
Extend sleeve or connector nut
Use plastic
Look for metal particles,
Uee
the correct
Use correct end of calibration block
Gage all connectors before
Align connectors carefully
Make
Turn only the connector nut
Use a torque wrench for
endcaps
preliminary conuection
during storage
scratches, and dents
@age
type
Do
fInal
fhst
use
Making
lightly
connect
Set connectors contactend down
Get liquid into plastic support beads
tinnectionf3
Do Not
Apply bending force to connection
Over tighten prelhuimuy connection
Twist or screw any connection
Tighten past torque wrench “break” point
2-2
Making Measurements
Basic Measurement Sequence and Example
Basic Measurement Sequence
There are five basic steps when you are making a measurement.
1. Connect the device under test and any required test equipment.
Caution
2. Choose the measurement parameters.
3. Perform and apply the appropriate error-correction.
4. Measure the device under test.
5. Output the measurement results.
Damage may result to the device under test if it is sensitive to analyzer’s
default output power level. lb avoid damaging a sensitive DUT, perform step 2
before step 1.
Basic Measurement Example
This example procedure shows you how to measure the transmission response of a bandpass
flter.
Step 1. Connect the device under test and any required test equipment.
1. Make the connections as shown in Figure 2-l.
Figure 2-1. Basic Measurement Setup
Step 2. Choose the measurement parameters.
2.
Press-.
lb set preset to
Setting
3. lb set the center frequency to 10.24
the
Frequency Range.
pig @sg m
“Factory
Preset,” press:
GHz,
press:
Making Measurements
2-3
4.
‘Ib
set the span to 3
(%jJ@(gg
GHz,
press:
Note
You could also press the
Istart_]
and
Lstoe)
keys and enter the frequency range
limits as start frequency and stop frequency values.
Setting the Source Power.
5. To change the power level to -5
Note
You co&f &o
press
dBm,
press:
.~~~~~;~G~..~~
. . . . . . . . . . . . . . . . . . . . . . .
,,.,..........,
:...<...
::..i,.....
~~~.~~~~.’
. . .
.:::.i
power ranges, to keep the power setting within the defined range.
Setting the Measurement.
6. To change the number of measurement data points to 101, press:
To
select the transmission measurement, press:
7.
8.
‘RI
view the data trace, press:
ad
. . . . . . . . . . . . . . .
1..;;
s&e&
one of
the
Step 3. Perform
tend
apply the appropriate error-correction.
9. Refer to the “Optimizing Measurement Results” chapter for procedures on correcting
measurement errors.
10.
To
save the instrument state and error-correction in the analyzer internal memory, press:
Step 4. Measure the device under test.
11. Replace any standard used for error-correction with the device under test.
12.
‘Ib
measure the insertion loss of the bandpass
hlter,
press:
Step 6. Output the measurement results.
13. To create a hardcopy of the measurement results, press:
Refer to the “Printing, Plotting, and Saving Measurement Results” for procedures on how
to define a print, plot, or save. For information on
confIguring
a peripheral, refer to the
“Compatible Peripherals” chapter.
24 Making Measurements
Using the Display Functions
To
View Both Primary Measurement Channels
In some cases, you may want to view more than one measured parameter at a time.
Simultaneous gain and phase measurements for example, are useful in evaluating stability in
negative feedback amplifiers. You can easily make such measurements using the dual channel
display.
1. To see channels 1 and 2 on two separate graticules, press:
The analyzer shows channel 1 on the upper half of the display and channel 2 on the lower
half of the display. The analyzer also defaults to measuring
S11
on channel 1 and
Sfl
channel 2.
Figure 2-2.
Example
of Viewing Both Primary Channels with a Split Display
on
Making Measurements
2-5
2. To view both primary channels on a single graticule, press:
,,.
Set
SFiZ~~ZKSP
,/ ,,,......... :...
..:
. . . . . . . .. ..f.
../
to
1X.
Example of Viewing Both Primary Channels on a Single Graticule
Note
You can control the stimulus functions of the two primary channels
independent of
each
other, by pressing
LMenu_) ;~~~~~~~~~~~~~~~~~
.<.. / ,:.. i,,;
. However,
auxiliary channels 3 and 4 are permanently coupled by stimulus to primary
channels 1 and 2 respectively.
To
Save a Data Trace to the Display Memory
.:&St:; .<z ..*: i. ET ,:.+ ‘9,;.
Press
C-1
~~~~~~~~~~:. to store the current active measurement data in the memory of
,,...
.._.........................
;:.:..a .. . . . . L/ .... . . . .
s..w
i/
the active channel.
The data trace is now also the memory trace. You can use a memory trace for subsequent math
manipulations.
!lb
View the Measurement Data and Memory Trace
The analyzer default setting shows you the
1.
‘lb
view a data trace that you have already stored to the active channel memory, press:
current
measurement data for the active channel.
This is the only memory display mode where you can change the smoothing and gating of
the memory trace.
2. To view both the memory trace and the current measurement data trace, press:
24
Making Measurements
To
Divide Measurement Data by the Memory
Trace
You can use this feature for ratio comparison of two traces, for example, measurements of gain
or attenuation.
1. You must have already stored a data trace to the active channel memory, as described in “To
Save a Data Trace to the Display Memory.
n
2. Press
(-1.~~~~~~.
to divide the data by the memory.
The analyzer normalizes the data to the memory, and shows the results
To
Subtract the Memory Trace from the Measurement Data Trace
You can use this feature for storing a measured vector error, for example, directivity. Then,
you can later subtract it from the device measurement.
1. You must have already stored a data trace to the active channel memory, as described in “To
Save a Data Trace to the Display Memory.”
_ _ _ .,. _
2.
press (jj)
;~~~~~~, to subtract the memory from the measurement
i
:,:,.
i.
data.
The analyzer performs a vector subtraction on the complex data.
To
Ratio Measurements in Channel 1 and 2
You may want to use this feature when making amplifier measurements to produce a trace that
represents gain compression. For example, with the channels uncoupled, you can increase the
power for channel 2 while channel 1 remains unchanged. This will
allow
you to observe the
gain compression on channel 2 .
1.
fiesm
~~~~~~~~~~~~ito
uncouple the
channel&
2. Make sure that both channels must have the same number of points.
to ON to ratio channels 1 and 2, and put the results in the channel 2 data
,.,,,,,,
::;;;;,y:.;:;:
..........
.:..::
set ,$&&j~,:
3
i..i,i
.........;
,,,,,.................. ...:..... . .../
to
ON, press ‘i&$&f&;. :~~~~ set
.,
’
array. This ratio is applied to the complex data.
4. Refer to “Measuring Gain Compression” for the procedure on identifying the 1
compression point.
Making Measurements 2-7
dl3
To
Title the Active Channel Display
l-
Press
2-
Press
(EjLJ :KlIH TXT’LE
...........................
H%A#$ .TXi”LE
q If you have a DIN keyboard attached to the analyzer, type the title you want from the
keyboard. Then press
to access the title menu.
.......
and enter the title you want for your measurement display.
@i?EE)
to enter the title into the analyzer. You can enter a title
that has a maximum of 50 characters.
q If you do not have a DIN keyboard attached to the analyzer, enter the title from the
analyzer front panel.
a.
Turn the front panel knob to move the arrow pointer to the first character of the title.
b.
C.
Repeat the previous two steps to enter the rest of the characters in your title. You can
enter a title that has a maximum of 50 characters.
..,
d.
Press
~~@JR
to complete the title entry.
2-g
Making Measurements
Using the Four-Parameter Display
All four S-parameters of a two-port device may be viewed simultaneously by enabling auxiliary
channels 3 and 4. Although independent of other channels in most variables, channels 3 and
4 are permanently coupled to channels 1 and 2 respectively by stimulus. That is, if channel 1
is set for a center frequency of 200 MHz and a span of 50 MHz, channel 3
stimulus values.
Channels 1 and 2 are referred to as primary channels, and channels 3 and 4 are referred to as
auxiliary channels.
Four-Parameter Display and Calibration
wiII
have the same
A full two-port calibration must be
active
before an
auxiliary
channel can be enabled. The
following measurement example uses a full two-port calibration covering the entire frequency
range of the analyzer, which is then narrowed to the range of the DUT (device under test)
using interpolated error correction. Refer to the “Full Two-Port Error-Correction” procedure in
Chapter 5.
Interpolated error correction is a useful feature which allows you to select stimulus parameters
which are subsets of a calibration which originally covered a wider frequency range.
Interpolation retains the calibration for the new stimulus parameters as long as they
fall
within
the range of the original calibration. Refer to Chapters 5 and 6 for a full description of error
correction.
The status notation CA will appear on the display if interpolated error correction is on. This is
normal. Refer to “Status Notations” in Chapter 2.
A full two-port calibration can also be recalled from a previously saved instrument state. See
“Recalling a File” in Chapter 4.
lb
View All Four S-Parameters of a Two-Port Device
The device used in this measurement example is a bandpass
filter
with a center frequency of
134 MHz.
1.
Press
w
2.
If a full two-port calibration has been performed or recalled from a previously saved
instrument state, go to step 5. If not, proceed to the next step.
3.
Set the stimulus values for the DUT. For this example, a center frequency of 134 MHz
and span of 45 MHz were selected, and the IF bandwidth was left at its default value of
3700
Hz.
4.
Perform a full two-port calibration on your analyzer. Refer to “Full Two-Port
Error-Correction” in Chapter 5.
Connect the DUT to the analyzer.
5.
Press
6.
[K]
to select the type of display of the data. This example uses the log mag
format.
If channel 1 is not active, make it active by pressing
7.
8.
c-1).
Making Measurements
2-9
.-.
The display will appear as shown in Figure 2-5. Channel 1 is in the upper left quadrant of the
display, channel 2 is in the upper right quadrant, and channel 3 is in the lower half of the
display.
-2
CHl
LOG
Sll
CENTR 134.000 MHz SPAN 45.000 MHz
CH3
.5dB/
REF
dB
17 Sep 1998
LOG
CH2
s21
PRm
Car
START 111.500 MHz STOP 156.500 MHz
10
11:13:
dB.‘REF
31
-50
dB
OUAL CHAN
ON off
-
AUX
CHhN
ON off
-
4
PARAM
DISPLAYS
SPLIT OISP
IX
2x
4x
-
CHANNEL
POSITION
CENTER
134.000 000 MHz
Figure 2-5.
3-Channel
RETURN
SPAN 45000 000 MHz
Display
2.10 Making Measurements
9. press I-),
set &&.~b.~~ to ON.
This enables channel 4 and the screen now displays four separate grids as shown in
Channel 4 is in the lower-right quadrant of the screen.
2
Sep
CHl
Sll
LOG
.
REF -2
dB
5
dE/
CA
t
CENTR 134.000 MHz SPAN 45.000 MHz
glz
LOG10
dB/
REF -50 dB
CH2
LOG
521
PRm
CA
t
CENTR
134.000 MHz SPAN
EZ
Lo8
1998
10 dB/
-
5
dB,‘REF
14:15:57
REF -50
45.000
-3 dB
dB
MHz
Figure
2-6.
DUAL
CHAN
ON off
-
AUX
CHAN
ON off
-
4
P~IRAM
0 ISPLAYS
SPLIT OISP
1X
2x
4x
-
t
CENTR 134.000 MHz SPAN 45.000 MHz
lb
Activate and
Gml3gure
Figure 2-6.
the
AnxWary
CENTR
134.000 MHz SPAN
4-Channel
Channels
Display
45.000
MHz
This procedure continues from the previous procedure.
Markers 1 and 2 appear on all four channel traces. Rotating the front panel control knob
moves marker 2 on ah four channel traces. Note that the active function, in this case the
marker frequency, is the same color and in the same grid as the active channel (channel 4.)
12.
Press
(than]
Observe that the LED adjacent to
Ichan
is constantly lit, indicating channel 1 is active.
CHANNEL
POSITION
RETURN
13.
Press
(Chanj
again.
Making Measurements
2-l
1
Observe that the LED is flashing, indicating that channel 3 is active.
14. Rotate the front panel control knob and notice that marker 2 still moves on all four
channel traces.
15. To independently control the channel markers:
Rotate the front panel control knob. Marker 2 moves only on the channel 3 trace.
Once made active, a channel can be configured independently of the other channels in most
variables except stimulus. For
S&h cha.rt by pre=ing Cm] ~~~~~,,,~,,~~,
example,once channel 3 is active, you can change its format to a
,....
.
Quick Four-Parameter Display
Steps 6 through 9 describe one way to create a four-parameter display. A quicker way is to use
one of the setups in the z4
After step
The
6,
press CDisplay~ ;#&Q’.; j ;&#j -SE+,
,..,.,....:.. ,,,,) ,...,. r.. .,... :.,.:.:.:.:.
~-~~~~~~~~~~~~~~
./...../ .i
. . . . . ..*>.......
..;<;...;+ ‘“i.B . . .
s .,.,.,.,.,,.,.,.,
..w;..;..i
i: . . . . . . .i.~.A::.:
.~~~~~~~~~~~
,.‘.. .:,
;+*
menu gives you a choice of standard channel
i.
;:.: ./:
.
...,
/......
_
sub-menu in the
4:
.ii
:,~~~.~~~~~~~
.:...;; . . . . . . . . ..i.....
(&GJ
..................... ............... .::: ..... i
menu.
confIgurations
and
parameter assignments.
):; : :
i
::<.
Press
:~~~~~
in place of step 6.
.” : ”
For more information about the
and Operation Concepts.
n
~:‘~~~~~~~~~~~
..::.:
.,.?./..ii .:.....L.‘../,L
menu, refer to chapter 6, “Application
,,,,
Characterizing a Duplexer
The following example demonstrates how to characterize a S-port device, in this case a
duplexer. This measurement
A duplexer’s three ports are:
n
Transmit (TX)
n
Receive (Rx)
w
Antenna (Ant)
utihzes
four-parameter display mode.
There are two signal paths through a duplexer: from TX to Ant, and from Ant to Rx. The
two signal paths are offset in frequency from each other and have the antenna (Ant) port in
common.
This example displays the transmission
duplexer in the top half of the display, and the
(‘Ix-to-Ant
reelection
and Ant-to-Rx) characteristics of the
characteristics
(TX
and Rx ports) in
the bottom half. Therefore, the stimulus is set up so that it is centered midway between the
transmit and receive frequencies of the duplexer, and the span is set to cover the combined
receive and transmit frequencies.
Other display configurations are possible. For example, the display can be
the transmission and reflection of the
TX-Ant
path is shown in the top half of the display, and
conhgured so that
the transmission and reflection of the Ant-Rx path is shown in the bottom half of the display.
2-12
Making Measurements
Required Equipment
Characterizing a duplexer requires that the test signals between the analyzer (a Z-port
instrument) and the duplexer (a
3-port
device) are routed correctly. This example uses one of
the following adapters to perform this function:
n HP 87533 Option
w
HP 87533 Option
You must also have a set of calibration standards for performing a full
K36
duplexer test adapter
K39 3-port
test adapter
2-port
calibration on
your test set up.
Procedure for Characterizing a Duplexer
1. Connect the test adapter to the analyzer according to the instructions for your particular
model. Connect any test fixture or cables to the duplexer test adapter.
2. Press
IPreset)
3. Set up the stimulus parameters for channel 1 (center/span frequencies, power level, IF
bandwidth). This example uses a span of 120 MHz centered at 860 MHz.
4. Uncouple the primary channels from each other:
Press
m,
set
CDUPLEB @I
to OFF.
(This is necessary in order to set the test set I/O independently for each channel.)
5. Set up control of the test adapter so that channel 1 is TX:
6. Perform a full
Press
Cal) $XLfBIkATE l#$NJ FTJLL 2-PORT
2-port
calibration on channel 1. (Refer to chapter 5, if necessary.)
, and follow the instructions to complete the
calibration.
Note
Make sure you connect the standards to the TX port of the test adapter (or a
cable attached to it) for the FORWARD calibration and to the Ant port for the
REVERSE calibration. The
LEDs
on the test adapter indicate the active ports:
a brightly lit LED indicates the source port; a dimly lit port indicates the input
port; an unlit LED indicates no connection.
7. When the calibration has been completed, save the instrument state:
Press
8.
@ave/Recall] SAVJJ
Press
@iGTj.
STATE
9. Set up channel 2 for the same stimulus parameters as channel 1.
Press
1Center) (%ZJ (iGJJ (%jZJ (ZGJ cM_II?)
10. Set up control of the test adapter so that channel 2 is measuring the receive path of the
duplexer: (Uncoupling the channels
Make sure you connect the calibration standards to the Rx port of the test
adapter (or a cable attached to it) for the FORWARD calibration, and the
Antenna port for the REVERSE calibration.
12. When the calibration has been completed, save this state in the analyzer:
,. ,,,
13. Connect the DUT to the test adapter.
14. Enable both auxiliary channels 3 and 4:
i
16. Set the measurement parameters (channel 1 should be active):
.y...
a.
Press
B
.g%~.
.i .:..: /
. . . .
This is the transmission of the Tx-to-Ant path.
,
..,.: :..
<.:.
b.Press(Chan] to activate channel 3, press
:@,%,;
This is the reflection at the TX port.
;
C.
Press
(m)
#&,
>>>>.>,.>>>..
.,,,., /
This is the transmission of the Ant-to-Rx path.
d-
Press@iGZ] to activate channel 4, press
This is the reflection at the Rx port.
The display will be similar to Figure 2-7.
I&$
2-14 Making Measurements
CHl
CH2
s21
s12
LOG
LOG
10 dB/REF
10
dB/REF
-40
-40
dB
dE
5 Aug 1998
13:lO:ll
Ref
1: FWO
Sll IA/R)
PRm
Cot-
PRm
Cot-
t
T
CHl/CH3
CH2/CH4
zi;
II
CENTER
CENTER
kE
10
dB/REF
10 dB/REF
860.000 000
860.000 000
I
Ih
I
0
dB
0
dR
N
MHz
MHzSPFIN
Figure
2-7. Duplexer Measurement
SP4N 120.000 000
120.000 000
MHz
MHz
Trans: FWD
521
Trans.: REV
512 <A/R)
Ref 1: REV
S22 <B/R)
ANALOG IN
Aux Input
CONVERSION
I
<B/R>
COFFI
INPUT
PORTS
Normally, a
the displayed trace. For faster
active display
2-port
calibration requires a forward and reverse sweep to finish before updating
(Sll
and
timing,
S21
for channel 1 in this case) to update more often than the unused
it is possible to set the number of sweeps for the
parameters. In this example we choose 8 updates of the forward parameters to 1 update of the
reverse in channel 1, and 8 updates of the reverse to 1 update of the forward in channel 2
(where the active parameters are
S22
and
S12).
Making Measurements
2-l
6
Using Analyzer Display Markers
The analyzer markers provide numerical readout of trace data. You can control the marker
search, the statistical functions, and the capability for quickly changing stimulus parameters
with markers, from the
Markers have a stimulus value (the x-axis value in a Cartesian format) and a response value
(the y-axis value in a Cartesian format). In a polar or Smith chart format, the second part of
a complex data pair is also provided as an auxiliary response value. When you switch on a
marker, and no other function is active, the analyzer shows the marker stimulus value in the
active entry area. You can control the marker with the front panel knob, the step keys, or the
front panel numeric keypad.
w
If you activate both data and memory traces, the marker values apply to the data trace.
n If you activate only the memory trace, the marker values apply to the memory trace.
n If you activate a memory math function (data/memory or data-memory), the marker values
apply to the trace resulting from the memory math function.
($GZZZJ
key.
The examples in this section are shown with
‘Ib
Use Continuous and Discrete Markers
Alter
measurement results
The analyzer can either place markers on discrete measured points, or move the markers
continuously along a trace by interpolating the data value between measured points.
Choose
,:; ;,,,:.;..;.;.::z
~~~~~~~~~~~~
.:.;.:.:..:::::::t.::.:::::.:::::::
;.:>y;;; ,<<<;;;. .“’ : ,,,, .,.;_ ,.,.....
AC%+%.::.2 . . . . . i . . . . . . . . . .
<.i:
..:.................:
if
:::..:..:.::
you want the analyzer to place markers
at my
p&t
i;;,;.;z;T;~,y,;;.’
on the trace, by interpolating between measured points. This default mode allows you to
conveniently obtain round numbers for the stimulus value.
. . . .
Choose ~,~~~~~~~~~~~
..
. . . . .
....~~~....
.
..........
. ..,.........
t.... :Y;.;.:r.::::
.._
z.:
:::;:..>.-:
if you
.....>.i
wit
the analyzer to place markers only on measured
trace points determined by the stimulus settings. This may be the best mode to use with
automated testing, using a computer or test sequencing because the analyzer does not
interpolate between measured points.
functions when the value entered in a search or positioning function does not
exist as a measurement point.
2-16
Making Measurements
‘lb
Activate Display Markers
To
switch on marker 1 and make it the active marker, press:
The active marker appears on the analyzer display as V. The active marker stimulus value is
displayed in the active entry area. You can modify the stimulus value of the active marker,
using the front panel knob or numerical keypad. All of the marker response and stimulus
values
are displayed in the upper right comer of the display.
Figure 2-8. Active Marker Control
‘#lb
switch on the corresponding marker and make it the active marker, press:
All of the markers, other than the active marker, become inactive and are represented on the
analyzer display as A.
Figure 2-9. Active and Inactive
To
switch off all of the markers, press:
Markers
Making Measurements
2-l
7
lb
Move Marker Information off of the Grids
If marker information obscures the display traces, you can turn off the softkey menu and move
the marker information off of the display traces and into the softkey menu area. Pressing the
backspace key @ performs this function. This is a toggle function of the backspace key. That
is, pressing &) alternately hides and restores the current softkey menu. The softkey menu is
also restored when you press any softkey or a hardkey which leads to a menu.
1. Set up a four-graticule display as described in “Using the Four-Parameter Display Mode.”
2. Activate four markers: Press
Note
Observe that the markers appear on all of the grids. F activate markers on
individual
UNCOUPLED.
press
grids, press 0))
@iGiG),
[jGi&J
Then, activate the channel:m~~~~~~~
1 2 3 4
~~~~~~..~~.,
i . ../. _:
then select the markers for that channel.
ad
set ;f#&&&
wish to :K;2AL.AAiers,
3. Turn off the softkey menu and move the marker information off of the grids:
Figure 2-10. Marker Information Moved into the Softkey Menu Area
CH4
Markers
S-1.7005
116.88200
2-3.8129
129.46850 MHz
3:-3.9114
139.97600 MHz
dB
MHz
dB
dB
4. Restore the softkey menu and move the marker information back onto the graticules: Press
@
The display will be similar to Figure 2- 11.
CHI
LO85 dB/’ REF -2
4:-1.4066 dB
Sll
151.509 500 MHz
dB
CHI
l:-1.0169 dB
‘9
CENTR 134.000 MHz SPAN 45.000 MHz
10 dB/
zi: LDG
4:-71.254
REF -50
dB 151.509 500 MHz
dB
I I I I I I I I I I
/-ICI3
Markers
.88200
MHz
:-3.1934
.46850
‘-3.1585
-97600 MHz
6.88200 MHz
.97600
dB
MHz
dB
Markers
MHz
MHz
2 Sep 1998
CH2
LOG 10
521
4:-69.313 dB 151.509 500
PRm
12:09:
dB/
REF -50 dB
CA
t
CENTR
134.000 MHz SPAN 45000 MHz
E
LDG4:-2.114i
5dB
dB/REF
43
MHz
151.509 -2.5 500 dB MHz
37 dB
0 MHz
36
dB
0 MHz
MARKER
1
I
2
I
3
I
4
5
I
all
OFF
A
MODE
MENU
CENTR 134.000 MHz SPAN 45.000 MHz
Figure 2-11.
You can also restore the
m)
or pressing a
softkey
softkey.
MKR
Marker
t
CENTR
134.000 MHz SPAN
45000
MHz
Information on the Graticules
menu by pressing a hardkey which opens a menu (such as
ZERO
Making Measurements
248
To
Use Delta (A) Markers
This is a relative mode, where the marker values show the position of the active marker
relative to the delta reference marker. You can switch on the delta mode by
dehning
one of the
five markers as the delta reference.
1.
Press LMarker_) :A. ~~~~~
:& I#F=:i to make marker 1 a reference marker.
2. To move marker 1 to any point that you want to reference:
q turn the front panel knob,
OR
q enter the frequency value (relative to the reference marker) on the numeric keypad.
2.
Press
~I$A&$#$
and move marker 2 to any position that you want to measure in reference
to marker 1.
Figure
2-12.
Marker
1 as the Reference Marker
4. lb change the reference marker to marker 2, press:
~~~~~~,,~~~~ ~~~~
L.. .,,, :.~.k
.: a:.:: T..Y :::.21..
To
Activate a Fixed Marker
When a reference marker is
position. The analyzer allows you to activate a
it does not rely on a current trace to maintain its fixed
fIxed
marker with one of the following key
sequences:
. . . .
.
B
_
&
.~~~~~
.:...i.:
i....... / /.
.,
,.
. . . . . . . . . .
..>.i i. ~.......-i.i~.;:...;::.:..:::.
.:...,.,
. . ;
..,.
. . . .
/.,;,
~~~~~~~~~~~~(
.<<<......
~:::<+::<....~.
..i ..:i
<<.:....:::..
.:.::<.:..
2-20 Making Measurements
Using
the
AREF=AFIXED MHR Key to Activate a
F’ixed
Reference Marker
1. To set the frequency value of a hxed marker that appears on the analyzer display, press:
(Marker_)
AMODE
MENU
AREF=AFIXED
MKR
AMODE
MENU FIXED MKR POSITION
FIXED MKRSTIMULUS and turn the front panel knob or enter a value from the front panel
keypad.
The marker is shown on the display as a small delta (A), smaller than the inactive marker
triangles.
2. To set the response value
(dB)
of a
fixed
marker, press:
FIXED MKR VALUE and turn the front panel knob or enter a value from the front panel
keypad.
In a Cartesian format, the setting is the y-axis value. In polar or Smith chart format, with a
magnitude/phase marker, a real/imaginary marker, an R + jX marker, or a G + jB marker, the
setting applies to the first part of the complex data pair. (Fixed marker response values are
always uncoupled in the two channels.)
3. To set the auxiliary response value of a fixed marker when you are viewing a polar or Smith
format, press:
FIXED
MKR !UX
VALUE and turn the front panel knob or enter a value from the front panel
keypad.
This value is the second part of complex data pair, and applies to a magnitude/phase marker,
a
real/iaginary
marker, an R+ jX marker, or a G+jB marker. (Fixed marker auxiliary
response values are always uncoupled in the two channels.)
Figure 2-13.
Example of a Fixed Reference Marker Using
&REF~AFXXED
MKR
Making Measurements 2-21
Using the
..
MK&.$$i!%Cl
:i
Key to Activate a Fixed Reference Marker
::9.:3
Marker zero enters the position of the active marker as the A reference position. Alternatively,
you can specify the
point with
.jZU$D HE~,.PfEUTXO~
. Marker zero is canceled by
fixed
switching delta mode off.
1. To place marker 1 at a point that you would like to reference, press:
(JGiG] and turn the front panel knob or enter a value from the front panel keypad.
2. lb measure values along the measurement data trace, relative to the reference point that
you set in the previous step, press:
i ,. ;”
:;#&&2Jj2$&
.,.,.,.,
and turn the front panel knob or enter a
value
from the front panel keypad.
3. To move the reference position, press:
. .
.:..
. . .
.,,, /,.., ,..’
~.~~~~.~;~,~~
. . . . . . .
: ..,... :; ..:.. :.:.:.:z ./
*,...
. . . . . . . . . . . . .
~~~~~~~;~~~~~~~~amr- $“xm m..c$TxaJJ$ ad turn the front panel knob
...“’
::i
:..,... :::....;>.I;::
.
...: ..:.......:...
,,,... :..:.:.:..:...p . ..i
. . .
i
..:...
. . . . . . . . . . . . .
. . .../
;.d:
or enter a value from the front panel keypad.
Figure 2-14. Example of a Fixed Reference
2-22 Making Measurements
CEIITER 10.230
000
CI~JO
GH:
SPAIJ
I.000 000 @On GHz
.~.~.:,~,,:~~~.~ ,.z
Marker
Using
&@J$I#&
.~~......,.;..,...
,....;..;:;
_;;;;;.i.i._
.
. . .
i
To
Couple and Uncouple Display Markers
At a preset state, the markers have the same stimulus values on each channel, but they can be
uncoupled so that each channel has independent
1.
F?ess (j-1
&&&$$#fJfi& $IEMlJ
i /i ..A..
>;;; ;:
and select from the following keys:
markers.
CHI
PRm
PRm
t
b-J
CENTER IO.240
Choose
~~~~~~~~~~~~
if you want the analyzer to couple the marker stimulus values
for the two display channels.
,.~., .,.,. ,. .,.,.
. . .
.,. .,.
Choose ~~~~.:~.~~~~~~~~
. .
..z... . ..i....
,..,..,. _ _ _ ,.
.,.
.;.........*
,. ..;
. . . . ..i./.
..,.., /,.., .,,.
if
you want
theanalyzer to uncouple
the marker sth&ls
values for the two display channels. This allows you to control the marker stimulus values
independently for each channel.
log MAG
I I
lcsg
MA6 10de/ REF
,-
10 de/ REF
I/” I
I” I
I II
000000 GHz
-30 de
II
-29.92
SPAN 2.500
:i
-: 77.54Q dE
.I
II
dEi
3X:
77.549
de log
-
y.-
V
000
00!1 GHz;TARTGHz
log MAG
F'
Rm
e
e
t I
MA6
PRK!
f
10 de/
,I
I
I I
10
REF
-30 de
I I I
de,/
REF
-29.92
dB
STUP 11.490
~69.4qij dE
x
I
I
3
~22.23
000GHz
dE_
Figure 2-15. Example of Coupled and Uncoupled
Biarkers
To Use Polar Format Markers
The analyzer can display the marker
. . . “:
;.&~#EJ!E% .,
.._..........................................
.,.
.z~:zc, .‘“.~:::~
:,
gives
emmagnitude and phase, ,;@i& m gives log magnitude md phw,
lin
gives the real valuehrst, then the imaginary
You can use these markers only when you are viewing a polar display format. (The format is
available from the
Note
(jj)
key.)
For greater accuracy when using markers in the polar format, it is
recommended to activate the discrete marker mode. Press
‘~~~~~~~~~, ~~~~~~~~:,
/.........&..
. . . . . . . . . .
. . . . . . . . . . . . .
. . . . . .
..::..z..i:
. .
..A
. . .
..~.....;..~......~;;~~~ ii :..:a ..A.. .A.. ii i i i: i
1. lb access the polar markers, press:
/ ,,,,, ,,
value
as magnitude and phase, or as a real/imaginary pair:
. . . . . . . . . . . . . .
..i i
value.
(jj’
.
i.:.::
/..
i
. . . .
_
~~~~~
.. ..A. i
. . . .
pb677d
i;..:.. ii
Making Measurements 2-23
2. Select the type of polar marker you want from the following choices:
;.
n Choose
LLW ..!!I&
if you want to view the magnitude and the phase of the active marker.
The magnitude values appear in units and the phase values appear in degrees.
n Choose
active marker. The magnitude values appear in
&JG.$LR
if you want to view the logarithmic magnitude and the phase of the
dE3
and the phase values appear in
degrees.
n Choose
.&/X&XK%
if you want to view the real and imaginary pair, where the complex
data is separated into its real part and imaginary part. The analyzer shows the
marker value the real part (M cos
0),
and the second value is the imaginary part
(M sin 8, where M=magnitude).
Figure
CENTER10240 ouo 000 GHZSPAN*500
2-16. Example of a Log
Marker
000 000 GHZ
in Polar Format
first
‘It,
Use Smith Chart Markers
The amount of power reflected from a device is directly related to the impedance of the device
and the measuring system. Each value of the reflection coefficient (I’) uniquely
defies
a device
impedance; r = 0 only occurs when the device and analyzer impedance are exactly the same.
The reflection coefficient for a short circuit is: I’ = 1 L
180“.
Every other value for I’ also
corresponds uniquely to a complex device impedance, according to the equation:
zL = [( 1 -t r) / (1 - qxzo
where ZL is your test device impedance and Z0 is the measuring system’s characteristic
impedance.
Note
1.
Press ~
2.
press (jj,
For greater accuracy when using markers in the Smith chart format, it
is recommended to activate the discrete marker mode. Press
~~~,~~~~
.~~~~~~.
~~~~~~~
~~~~~~~~.
.... )(? ..:‘<v~,v
~~~~~~~~~,
““::?y
..: ;
,& y
.S....““,
.<:grz.
md
turn the front panel knob or
(Marker4
enter a value from the front panel keypad to read the resistive and reactive components of
the complex impedance at any point along the trace.
!I’his
is the default Smith chart marker.
The marker annotation tells that the complex impedance is capacitive in the bottom half of
the Smith chart display and is inductive in the top half of the display.
2-24 Making Measurements
n
Choose
LIN MKR
if you want the analyzer to show the linear magnitude and
the
phase of
the reflection coefficient at the marker.
n
Choose LOG
MRR
if you want the analyzer to show the logarithmic magnitude and the
phase of the reflection coefficient at the active marker. This is useful as a fast method of
obtaining a reading of the log magnitude value without changing to log magnitude format.
. Choose
R&m MKR
if you want the analyzer to show the values of the reflection
coefficient at the marker as a real and imaginary pair.
n
Choose
RtjX MRR
to show the real and imaginary parts of the device impedance at
the marker. Also shown is the equivalent series inductance or capacitance (the series
resistance and reactance, in ohms).
. Choose
G+jB MKR
to show the complex admittance values of the active marker in
rectangular form. The active marker values are displayed in terms of conductance (in
Siemens), susceptance, and equivalent parallel circuit capacitance or inductance. Siemens
are the international unit of admittance and are equivalent to mhos (the inverse of ohms).
To
Set Measurement Parameters Using Markers
The analyzer allows you to set measurement parameters with the markers, without going
through the usual key sequence. You can change certain stimulus and response parameters to
make them equal to the current active marker value.
Setting the Start
1. Press
[jGiGEG~
Frequency
and turn the front panel knob or enter a value from the front panel
keypad to position the marker at the value that you want for the start frequency.
2.
Press
MARKER-tSTART
to change the start frequency value to the value of the active marker.
Making Measurements 2-25
CHl 521
lvg
MAG
10
de/ REF
30
de
lL~74.17 dB
iH
1
MAK(EK 1
8 653150079- ‘Hi:
Figure 2-18. Example of Setting the Start Frequency Using a Marker
Setting the Stop Frequency
1.
FVess (JGGX$
and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at the value that you want for the stop frequency.
2.
Press
, “‘,...”
‘~~~~.
;
to change the stop frequency value to the value of the active marker.
pb678d
CHl S21
PRm
CENTER
log MAG10 de/ REF
575 036
11.071
GHr
-30 dB
SPAN 4.836
l-1-74.426 de
849 927
GHr
CHl
521
PRm
CENTER 10.319
log MA6
928
10 de/
457
REF
6Hr
Figure 2-19. Example of Setting the Stop Frequency Using a
Setting the Center Frequency
1. Press
(JGZZZJ
and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at the value that you want for the center frequency.
. . . . . . . . . . . . . .
..A /.
2.
Press ~~~~:
..;....;.....................
~,..~;;~..,.;.~...~.~;::
:.: ,....,,:
./ .
. . . . . ii....
..., I:..bi
. . . . . s..w;..;
ii .A......... .
to change the center frequenw
value to the value of the active
marker.
-30 de
SPAN 3.333
Marker
I-:-eo.cio5
556 769
pb679d
de
GHz
2-26 Making Measurements
_-..
MAiKEK 1
10.21j95 (Hz
I
MAdKEK b
1 .21 95 G z
I
1
I
III /
1
'7
1
\
'!
CENTER e.265
9511 OO(i GHz
Figure
SPAN 6.500
Z-20.
Example of Setting the Center Frequency Using a Marker
000 iJO0
GHi
Setting the Frequency Span
You can the span equal to the spacing between two markers. If you set the center frequency
before you set the frequency span, you will have a better view of the area of interest.
2. Turn the front panel knob or enter a value from the front panel keypad to position the
markers where you want the frequency span.
Iterate
between
maker 1 ad marker
2
by pressing ~#&&&&
ii/:i ..A. i.i'.A.>>.%/
a&
. . . . . .
~~~
.
respedively,
and turning the front panel knob or entering values from the front panel keypad to position
the markers around the center frequency. When finished positioning the markers, make sure
that marker 2 is selected as the active marker.
Note
Step 2 cm &-J&
p&o~edusingS~.~~
using this method, it
; .,.,
.,.,
wiIl
not be possible to iterate between marker zero and
i i /r:;...
i .;; 'Z, ". "'Y: t..:<<<;
and ~~~~~,.
/ .,, . .
.*;.;
However,when
marker 1.
ptB80d
Making Measurements 2-27
3il;1e/
REF
44.44
dE
Z-:
13.124
dE
log
MAI,
311
de/REF
44.44
dE
PRm
CENTER
l(J.215 95rJ OOfl GHi
Figure
Z-21.
SPAN 6.500
0Oi1 000 GHr
Example of Setting the Frequency Span Using
e
I
CENTER10.443
II/I
4112
355 6HzSPAN
Markers
Setting the Display Reference Value
1. Press
@E&ZG)
and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at the value that you want for the analyzer display reference
value.
2. Press
,
.,...
~~~~~~..
:..::;;.. ..A
. .. ..i.
ii::.:.;
.._. z:...::
to change the reference value to the value of the active marker.
II
Z.809
204
4.31 I;Hr
pb5Bld
CHl S21CHl S21
PRmPRm
CENTER
CENTER 10.240
Log MkbLog Mkb
10.240
Oiir?
Oiir?
15 de/
REF 39
15 de/
REF 39
000
GHzSPANSPAN 2.500000 000 GHz
000GHz
l-:l-:
dE
dE
2.500
000 000
1.650 dB
1.650 dB
GHz
MAG
PR
/
Jn/~-J
CENTER
10.240 000 000
15
dE/
REF
2'
_,
GHr
Lbg
Figure 2-22. Example of Setting the Reference Value Using a Marker
1.702
SPAN
dE
l-:1.615
'.
2.500
.-.
' /
000 000
dB
i bG.i\
GHz
2-28 Making Measurements
Setting the Electrical
Delay
This feature adds phase delay to a variation in phase versus frequency, therefore it is only
applicable for
1.
Press
ratioed inputs.
(K)
FW33.
2. Press (MarkerFctn) and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at a point of interest.
3.
Press
~~~~~~~~.
.:
..i ,.,..,
. . . . . .
..A. :.,.
,,:.: :.:.
to automatically add or subtract enough line length to the receiver
..: ,......
input to compensate for the phase slope at the active marker position. This effectively
flattens the phase trace around the active marker. You can use this to measure the electrical
length or deviation from linear phase.
Additional electrical delay adjustments are required on devices without constant group delay
over the measured frequency span.
$21
CHI
PRm
c2
Phase
ea
MAF.KER
10.1
CENTER 10.240000000 6Hz SPAN 3.500000000 GHz
1
GHz
90
‘I
KEF0D1 :
11.100
M
300
n
163.0f1°
OCO GHz
Figure 2-23. Example of Setting the Electrical Delay Using a
Setting the CW Frequency
1.
‘Ib
place a marker at the desired CW frequency, press:
LHl 521
phase
m
PRm
c2
MARKER
Del
CENTER 10.240 000 000
1
10.1 2Hz
9Cl
GHz
“I
KFF0D
10.240 000 000
1
SPAN 7.500 000 000
BIarker
1 :173.57O
M:*
GHr
GHz
m
You can use this function to set the marker to a gain peak in an amplifier. After pressing
:~~~~~~~~~-, activate a CW frequency power sweep to look at the g& compre&on with
/. . . . . . . .
... .
and either turn the front panel knob or enter the value, followed by
These functions place the marker at an amplitude-related point on the trace. If you switch on
tracking, the analyzer searches every new trace for the target point.
Searching for the Maximum Amplitude
1.
press Ij) ~~~~~~~~
2-
Press
~~,~~~;~~~.~.
trace.
”
.”
to move the active marker to the
..r..:.:...::..:: ::.:...
to access the marker search menu.
maximum
point on the measurement
Figure 2-24.
Example of Searching for the Maximum Amplitude Using a
Searching for the Minimum Amplitude
Figure 2-25.
Example of Searching for the Minimum Amplitude Using a Marker
Blarker
Searching for a
2.30 Making
Measursments
Target
Amplitude
2.
Press
SEM%Il:. 2’~
,,...........
:.:
to move the active marker to the target point on the measurement
. . . . . . . . . .:
trace.
3.
If you want to change the target amplitude value (default is -3
enter the new value from the front panel keypad.
4.
If you want to search
.,, ,,,:.: ,/
‘SW ‘;Jm
and
for,,multiple
.:.., ;.
.$$J&&#f #$?&jm,.
responses at the target amplitude value, press
,:..::.., ;,..
.:..
dB),
press
X&S&%
i
and
CHI 521
FRm
leg MAI,
111 dE/REF
-30
de
FRm
lc,g
MA6
\
, ,, / -1.J
CENTER
11,.240
UOCJ O(J(I GHz
Example of Searching for a
Searching for a Bandwidth
The analyzer can automatically calculate and display the -3 dB bandwidth (BW:), center
frequency (CENT:), Q, and loss of the device under test at the center frequency. (Q stands for
“quality factor,”
dellned
values are shown in the marker data readout.
SPAN 2.500
as the ratio of a circuit’s resonant frequency to its bandwidth.) These
7,”
Y\
irO0 000 GHr
Figure 2-26.
‘beget
CENTER 10.240 OUO
000 GHr
SPAN 2.500
Amplitude Using a Marker
OOil 000 6Hz
pt533d
1. Press
@GiZ)
and turn the front panel knob, or enter a value from the front panel keypad to
place the marker at the center of the filter passband.
3. press ~~~~~~~~:~
,.,... ;. ,,.,.;.: / ., ,,,,
bandpass
or band reject shape on the measurement trace.
4. If you want to change the amplitude value (default is -3
rejectband, press ~~~~~~~~
. . . .
I
to
&date
thecentersthuh.ls
,::::..:~ ;:..:;:<:yv
.i
. . . . . . . . . . . . . . . ~.wu;;;.;;.;;~.~
.A.. i
. . . . . . . . . . . .
L ii ~.;...:~..>>:..::
ad
enter the
v&e,
new
v&e from the front panel keypad.
bandwidth, ad the Q of
dl3)
that
defines
the passband or
Making Measurements 231
a
Figure 2-27.
CE,,,EH 1” Z4il 0”” 00” OH:
Example of Searching for a Bandwidth
SPA,! 1
OClll b”,1
Trackiug the Amplitude that You Are Searching
‘OCiGHZ
using
Markers
1. Set up an amplitude search by following one of the previous procedures in
Specific Amplitude.
2.
Press [-FctnJ ~~~~~~ ~~~~~~~~;~~,, to track the spehfied mpfitude search
n
..:::.::...
~.~.T./.i .;;;./;:,....~.,....;;..././.
ii
..:
.ss....i
.:.._/
. . . . . . . .
~~.~.~.~.~.
/i..i.i
‘To
Search for a
with every new trace and put the active marker on that point.
When tracking is not activated, the analyzer
sweep and the marker remains at same
finds
stimulus
the specified amplitude on the current
value, regardless of changes in the trace
response value with subsequent sweeps.
232 Making Measurements
To
Calculate the Statistics of the Measurement Data
This function calculates the mean, standard deviation, and peak-to-peak values of the section
of the displayed trace between the active marker and the delta reference. If there is no delta
reference, the analyzer calculates the statistics for the entire trace.
2. Move marker 1 to any point that you want to reference:
w
Turn the front panel knob.
OR
w
Enter the frequency value on the numeric keypad.
3.
Press
$&I&&$;.&?
and move marker 2 to any position that you want to measure in reference
to marker 1.
4. Press
l$iXXG] ~~~~~~~~~;~ &lT$&&@~
to calculate and view the mean, standard
deviation, and peak-to-peak values of the section of the measurement data between the
active marker and the delta reference marker.
An application for this feature is to
Gnd
the peak-to-peak value of passband ripple without
searching separately for the maximum and minimum values
If you are viewing a measurement in the polar or Smith Chart format, the analyzer
calculates the statistics using the
first
value of the complex pair (magnitude, real part,
resistance, or conductance).
CHI 5-4
P R m
loa MAG
10 de/
REF -30 dE?
2 :
1.117
de.
CENTER
IO.240
000
000
GHzSPAti
1
.OOG
000 000
GHz
Figure 2-28. Example Statistics of Measurement Data
Making Measurements 233
.
_-.
Measuring Magnitude and Insertion Phase Response
The analyzer allows you to make two different measurements simultaneously. You can make
these measurements in different formats for the same parameter. For example, you could
measure both the magnitude and phase of transmission. You could also measure two different
parameters
This
measurement example shows you how to measure the
and then how to view the measurement data in the phase format, which provides information
about the phase response.
Measuring the Magnitude Response
1. Connect your test device as shown in Figure 2-29.
(SH
and
SZZ).
maximm
amplitude of a SAW
filter
DEVICE UNDER TEST
Figure 2-29. Device Connections for Measuring a Magnitude Response
2. Press
B
and choose the measurement settings. For this example, the measurement
parameters are set as follows:
You may also want to select settings for the number of data points, averaging, and IF
bandwidth.
234
Making
Measurements
4.Reconnect your test device.
5. To better view the measurement trace, press:
c-j
&&q:., Km
6. To locate the maximum amplitude of the device response, as shown in Figure 2-30, press:
[m);*.sw
1.:
&k&H: tix-
Example Magnitude Response Measurement
Measuring Insertion Phase Response
‘Ib
view both the magnitude and phase response of the device, as shown in Figure 2-31,
7.
press:
@iZ)
The channel 2 portion of Figure 2-31 shows the insertion phase response of the device under
test. The analyzer measures and displays phase over the range of
-
180 O to + 180 O. As phase
changes beyond these values, a sharp 360 O transition occurs in the displayed data.
CENTER IU.240 000 c’oo OH:
*PAP4 2 50” noo 000 GHZ
Figure 2-31. Example Insertion Phase Response Measurement
The phase response shown in Figure 2-32 is undersampled; that is, there is more than
MO0
phase delay between frequency points If the A4 = >
WOO,
incorrect phase and delay
Making Measurements
2-35
information may result. Figure 2-32 shows an example of phase samples with
and greater than
180°.
ACTUAL PHASE -
F,E<PONSE
\
AC#J
less than
HO0
Figure 2-32. Phase Samples
Undersampling may arise when measuring devices with long electrical length. To correct this
problem, the frequency span should be reduced, or the number of points increased until Ad is
less than
180°
per point. Electrical delay may also be used to compensate for this effect (as
shown in the next example procedure).
236 Making Measurements
Measuring Electrical Length and Phase Distortion
Electrical Length
The analyzers mathematically implement a function similar to the mechanical “line stretchers”
of earlier analyzers. This feature simulates a variable length
you can add to or remove from the analyzer’s receiver input to compensate for interconnecting
cables,
etc.
In this example, the electronic line stretcher measures the electrical length of a
SAW iilter.
Phase Distortion
The analyzers allow you to measure the linearity of the phase shift through a device over a
range of frequencies and the analyzers can express it in two different ways:
n deviation from linear phase
n
group delay
Measuring Electrical Length
1.
Connect your test device as shown in Figure 2-33.
NETWORK ANALYZER
lossless
transmission line, which
DEVICE UNDER TEST
Figure 2-33. Device Connections for Measuring Electrical Length
2.
Press
1Preset)
and choose the measurement settings. For this example, the measurement
settings include reducing the frequency span to eliminate under sampled phase response.
Press the following keys as shown:
(jj)(izJ(iiJJ
=c@i@
m
[K)
!$$#m. @ Lxl] (a @), HP
..~......... ::
.:. ..y,;
““.
jgg&Q.j
. . . . . . ;.... . . . . . . .
.-c.;.i .. . . ;...
...”
8722D)
You may also want to select settings for the number of data points, averaging, and IF
bandwidth.
Makine
Mearurrments
2-37
3.
Substitute a thru for the device and perform a response calibration and press:
0~~~~~~~~
4.
Reconnect your test device.
‘lb
better view the measurement trace, press:
5.
..z
,.,., ,.%.
:&f&l
.~~~~~~
‘z!EK
@Eiizz) asp@-. @Ii&E
Notice that in Figure 2-34 the SAW
iilter
under test has considerable phase shift within only
a 2 MHz span. Other filters may require a wider frequency span to see the effects of phase
shift.
The linearly changing phase is due to the device’s electrical length. You can measure this
changing phase by adding electrical length (electrical delay) to compensate for it.
6. To place a marker at the center of the band, press:
(j]
and turn the front panel knob or enter a value from the front panel keypad.
7. To activate the electrical delay function, press:
:: .:,,; ;;;;
_* . .
.;
cF==q
+ ~~~~~~~,
~~~:~~~~~~~~~~.,.,.~.,;,:
This function calculates and adds in the appropriate electrical delay by taking a
about the marker, measuring the
Ac$,
and computing the delay as
A+/Afrequency.
flO%
span
238
Makina
Measurements
3.
Press
[sCALE) &&TRK&, ~l&&Y
and turn the front panel knob to increase the
electrical length until you achieve the best flat line, as shown in Figure 2-35.
The measurement value that the analyzer displays represents the electrical length of your
device relative to the speed of light in free space. The physical length of your device is
related to this value by the propagation velocity of its medium.
Note
Velocity factor is the ratio of the velocity of wave propagation in a
coaxial cable to the velocity of wave propagation in free space. Most
cables have a relative velocity of about 0.66 the speed in free space. This
velocity depends on the relative permittivity of the cable dielectric
velocity factor =
116.
(cp)
as
You could change the velocity factor to compensate for propagation velocity by
pressing
Ical] $@I#& #&&%XiX FA@‘IXJR
.
,,, i
i
/.:;.i
(enter the value) (XJ This will help the
;;;
analyzer to accurately calculate the equivalent distance that corresponds to the
entered electrical delay.
Figure 2-35.
9.
‘lb
display the electrical length, press:
Example Best Flat Line with Added Electrical
Delay
In this example, there is a large amount of electrical delay due to the long electrical length
of the SAW
lUer
under test.
Measuring Phase Distortion
This portion of the example shows you how to measure the linearity of the phase shift over
a range of frequencies. The analyzer allows you to measure this linearity and read it in two
different ways: deviation from linear phase, or group delay.
Making Measurements 238
Deviation
From
Linear Phase
By adding electrical length to “flatten out” the phase response, you have removed the linear
phase shift through your device. The deviation from linear phase shift through your device is
all that remains.
1.
Follow the procedure in “Measuring Electrical Length.
2.
‘Ib
increase the scale resolution, press:
n
kz]
,&Kg: #y,
and turn the front panel knob or enter a value from the front panel
.
3.
To use the marker statistics to measure the maximum peak-to-peak deviation from linear
phase, press:
adjust
Activate and
4.
Note
the electrical delay to obtain a minimum peak-to-peak value.
It is possible to use
Amarkers to measure peak-to-peak deviation in only one
portion of the trace, see “lb Calculate the Statistics of the Measurement Data”
located earlier in this chapter.
Figure 2-36. Deviation from Linear Phase Example Measurement
Group Delay
The phase linearity of many devices is specified in terms of group or envelope delay. The
analyzers can translate this information into a related parameter, group delay. Group
delay is the transmission time through your device under test as a function of frequency.
Mathematically, it is the derivative of the phase response which can be approximated by the
following ratio.
A4/(360 * AF)
where Ad is the difference in phase at two frequencies separated by
commonly
caIled
the “aperture” of the measurement. The analyzer calculates group delay from
Al?
The quantity AF is
its phase response measurements
The default aperture is the total frequency span divided by the number of points across the
display (i.e. 201 points or 0.5% of the total span in this example).
1. Continue with the same instrument settings and measurements as in the previous procedure,
“Deviation from Linear Phase.”
240 Making Measurements
2. To view the measurement in delay format, as shown in Figure 2-37, press:
3. To activate a marker to measure the group delay at a particular frequency, press:
B
and turn the front panel knob or enter a value from the front panel keypad.
Figure 2-37. Group Delay Example Measurement
Group delay measurements may require a specific aperture
(Af)
or frequency spacing between
measurement points The phase shift between two adjacent frequency points must be less than
1800,
otherwise incorrect group delay information may result.
4. To vary the effective group delay aperture from
approximately 1% of the frequency spa,
press:
When you increase the aperture, the analyzer removes
minimum aperture (no smoothing) to
~'~~~~~~~~~:~~:~
.
..v.>..;;;;>
. . . . . . . . .
..i.// i i
fine
grain variations from the
/.
response. It is critical that you specify the group delay aperture when you compare group
delay measurements
CENTER
,134 00” 000 GHi
SPAN
cl”2 00” 000 WI:
Figure 2-38. Group Delay Example Measurement with Smoothing
Making Measurements 241
5.
‘lb
increase the effective group delay aperture, by increasing the number of measurement
points over which the analyzer calculates the group delay, press:
As the aperture is increased, the “smoothness” of the trace improves markedly, but at the
expense of
measurement
detail.
Group Delay Example Measurement with Smoothing Aperture Increased
242 Making Measurements
Tbsting
a Device with Limit Lines
Limit testing is a measurement technique that compares measurement data to constraints that
you define. Depending on the results of this comparison, the analyzer will indicate if your
device either passes or fails the test.
Limit testing is implemented by creating individual flat, sloping, and single point limit lines on
the analyzer display. When combined, these lines represent the performance parameters for
your device under test. The limit lines created on each measurement channel are independent
of each other.
This example measurement shows you how to test a
bandpassfilter using the following
procedures:
n
creating flat limit lines
n
creating sloping limit lines
n
creating single point limit lines
n
editing limit segments
n running a limit test
Setting Up the Measurement Parameters
1. Connect your test device as shown in Figure
Z-40.
DEVICE UNDER TEST
2. Press
Figure 2-40. Connections for
B
and choose the measurement settings For this example, the measurement
SAW
Filter Example Measurement
settings are as follows:
You may also want to select settings for the number of data points, power, averaging, and IF
bandwidth.
3. Substitute a thru for the device and perform a response calibration by pressing:
Making Measurements 243
pb69d
4. Reconnect your test device.
5.
lb
better view the measurement trace, press:
Creating Flat Limit Lines
In this example procedure, the following flat Iimit Iine values are set:
The analyzer generates a new segment that appears on the center of the display.
determining the value of the new segment, marker 1 also appears on the display.
‘ib
specify the stimuius value, test limits (upper and lower), and the iimit type of the iimit
3. To terminate the lines and create a sloping limit line, press:
4. To establish the start frequency and limits for a sloping limit line that tests the high side of
the
filter,
press:
Figure 2-43. Sloping Limit Lines
Making Measurements 247
Creating Single Point Limits
In this example procedure, the following
from +4 dB to -2 dB at 10.15
from +4 dB to -2 dB at 10.33
1. To access the
2. lb designate a
w
downward pointing, indicating the upper test Iimit
n
upward pointing, indicating the lower test Iimit
limits
menu and activate the Iimit
single
point Iimit
GHz
GHz
En-tits
line,
as shown in
are set:
lines,
F’igure
press:
2-44, you must
dellne
two pointers;
248
Makiq
Measurements
Figure 2-44. Example Single Points Limit Line
Editing Limit Segments
This example shows you how to edit the upper limit of a limit line.
1. To access the limits menu and activate the limit lines, press:
2. To move the pointer symbol
(>)
on the analyzer display to the segment you wish to modify,
press:
OR
%&&’
. . . . . .: . . . . . . . . . . . .
3. To change the upper limit (for example, -20) of a limit line, press:
:,&&g:,, ~~~~~:~~:~, L-20) Lxl] g#
and enter the segment number followed by
..-..i..~...
Lxl]
;.,, ,,,, ,...T.
Deleting Limit Segments
1. To access the limits menu and activate the limit lines, press:
2.
‘lb
move the pointer symbol
press:
,&$@~or@repeate~y
OR .;;;;/; fl,:
~$@%&J@
.
. .
,..........
I::...:.~~.:.~;.;;;.-i..r
.<:.<<.
and enter the segment number followed by
. . . . . .
3. lb delete the segment that you have selected with the pointer symbol, press:
~~~
.........
. . .
. . .
..~~~~..~.....~.;;..;;~.;;..~
(>)
on the analyzer display to the segment you wish to delete,
@
Making Measurements 248
Running a Limit Test
1. To access the limits menu and activate the limit lines, press:
Reviewing the Limit Line Segments
The limit table data that you have previously entered is shown on the analyzer display.
2. To verify that each segment in your
3. To modify an incorrect entry, refer to the “Editing
limits
table is correct, review the entries by pressing:
Limit
Segments” procedure, located
earlier in this section.
Activating the Limit
l&t
4. To activate the limit test and the beep fall indicator, press:
Note
&lecting
the beep fail ~&&or
:6&J; ~~~~~~~~. is optional
and a add
approximately 50 ms of sweep cycle time. Because the limit test will still work
if the me lines ue off, selecting ~~~~~~~~~~~~~:~~ is alSo optional.
..:.:.
.::..v
. . . . . . . .
.;.;z:
,...i .
..m 2
The limit test results appear on the right side on the analyzer display. The analyzer indicates
whether the filter passes or fails the dellned limit test:
o
The message FAIL will appear on the right side of the display if the limit test fails
0 me analyzer
n
The analyzer alternates a red trace where the measurement trace is out of
q
A
Tl’L
beeps if the mt test fails ad if ~~~~~~~~~.~~I~~~
signal on the rear panel BNC connector
.T:.::,.y,.'.:.~ ') / "' ...
“LIMIT
. . . . . . . . . . . .
.._...
ha
been s&&d.
limits.
TEST” provides a pass/fail (5 V/O
V)
indication of the limit test results
260
Making Measurements
Offsetting Limit Lines
The limit offset functions allow you to adjust the limit lines to the frequency and output level
of your device. For example, you could apply the stimulus offset feature for testing tunable
filters. Or, you could apply the amplitude offset feature for testing variable attenuators, or
passband
This example shows you the offset feature and the limit test failure indications that can appear
on the analyzer display.
ripple in filters with variable loss.
1. To offset all of the segments in the
limit
table by a llxed frequency, (for example, 50 MHz),
press:
The analyzer beeps and a FAIL notation appears on the analyzer display, as shown in
Figure
2-45.
Figure
2-45. Example Stimulus Offset of Limit Lines
2.
lb
return to 0 Hz offset, press:
,. :~-~:.::~,::~~;-:;
~~~~~~~~~~~~~~ @ lxlJ
”
.c,
.~......~.~.........~~..~.~.~
ii
‘..~~~..~~“,,“:~..:.
~.~.~...;;;....::.~...
.,.
..z ::..
3. To offset all of the segments in the limit table by a fixed amplitude, press:
The analyzer beeps and a FAIL notation appears on the analyzer display.
Making Measurements
2-51
Measuring Gain Compression
Gain compression occurs when the input power of an amplifier is increased to a level that
reduces the gain of the amplifier and causes a nonlinear increase in output power. The point at
which the gain is reduced by 1
will vary with frequency, so it is necessary to
the frequency band.
Once that point is identified, you can perform a power sweep of that CW frequency to measure
the input power at which the 1
compression. The following steps provide detailed instruction on how to apply various features
of the analyzer to accomplish these measurements.
dB
is called the 1
dI3
compression occurs and the absolute power out (in
dl3
compression point. The gain compression
find
the worst case point of gain compression in
dBm)
at
Input Power (dBm)
Figure 2-46. Diagram of Gain Compression
Note
1. Set up the stimulus and response parameters for your amplifier under test. To reduce the
effect of noise on the trace, press:
2. Perform the desired error correction procedure. Refer to Chapter 5, “Optimizing
Measurement Results,” for instructions on how to make a measurement correction.
3. Hook up the amplifier under test.
4.
lb
produce a normalized trace that represents gain compression, perform either step 5 or
step 6. Step 6 is optional
If the default output power of your analyzer is not high enough to force the
amplifier under test into compression, then the following procedure may have
to be performed with
the
addition of Option 007 or Option 085. Refer to “Hi
Power Measurements” for information on using Option 085.
2-62
Making Measurements
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.