The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequent ial damages in connect ion with the furnishing , performance, or
use of this material.
Warranty Information
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS.
FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW,
AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT
SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE
OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD
AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Certification
Agilent Technologies Company certifies that this product met its published specifications
at the time of shipment from the factory. Agilent Technologies further certifies that its
calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institute’s calibration facility, and
to the calibration facilities of other International Standards Organization members.
Regulatory Inf ormation
The regulatory information is located in Chapter 15 , “Safety and Regulatory Inf ormation.”
ii
Assistance
Front-Panel Key
SOFTKEY
Product maintenance agre ements and other customer as sista nce a greement s are available
for Agilent Technologies products. For any assistance, contact your nearest Agilent
Technologies sales or service office. See Table 15-1 on pa ge 15-3 for the nearest office.
Safety Note s
The following safety notes are used throughout this manual. Familiarize your self with
each of the notes and its meaning before operating this instrument. All pertinent safety
notes for using this product are located in Chapter 15 , “Safety and Regulatory
Information.”
WARNINGWarning denotes a hazard. It calls attention to a procedure w hich, if
not correctly performed or adhered to, could result in injury or loss
of life. Do not proceed beyond a warning note until the indicated
conditions are fully under stood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to , would result in damage to or destructi on of
the instrument. Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met.
How to Use This Guide
This guide uses the following conventions:
This represents a key physically located on the
instrument.
This represents a “softkey,” a key whose label is
determined by the instrument’s firmware.
Screen TextThis represents text displayed on the instrument’s screen.
iii
Documentation Map
The Installation an d Quick Start Guide pr ovides procedures for
installing, configuring, and verifying the operation of the analyzer. It
also will help you familiarize yourself with the basic operation of the
analyzer.
The User’s Guide shows how to make measurements, explains
commonly-used features, and tells you how to get the most
perform ance from yo u r an alyzer.
The Reference Guide provides refe ren ce informat ion, such as
specifications, menu maps, and key definitions.
The Programm er’s Guide provides general GPIB programming
informat ion, a comm and reference, and example prog rams. The
Programmer’s Guide contains a CD-ROM with example programs.
The CD-ROM provides the Installa tio n and Qu ick Start Guide, the
User’s Guid e, the Referen ce Guide, and the Progra mmer’ s Gu ide in
PDF format for vie wing or prin t ing from a PC.
The Service Guide pr ovides information on calibrating,
troubleshooting, and servicing your analyzer . The Serv ic e Gu ide i s not
part of a standard shipment and is available only as Option 0BW, or
by ordering part number 08720-90397. A CD-ROM with the Service Guide in PDF format is included for viewing or printing from a PC.
Service Equipment and Analyzer Options8719ET/20ET/22ET
Information in This Chapter8719ES/20ES/22ES
RF Network Analyzers
Information in This Chapter
This chapter contains information on the required equipment used to test and repair the
network analyzer. A list of analyzer options and service support options is included at the
end of the chapter.
Sections in This Chapter
•Service Test Equipment on page 1-3
•Principles of Microwave Connector Care on page 1-6
•Analyzer Options on page 1-7
•Service and Support Options on page 1-9
1-2Service Guide
8719ET/20ET/22ET Service Equipment and Analyzer Options
8719ES/20ES/22ESService Test Equipment
RF Network Analyzers
Service Test Equipment
Required Equipment
Frequency Counte rFreq: 0.050 to 20 GHz
Accuracy:3ppm max
Frequency Counte rFreq: 0.050 to 26.5 GHz
Accuracy:3ppm max
Spectrum AnalyzerMax Freq: 1.8 GHz RBW: 300 Hz8591EA,T
Power MeterAccura cy : ±0.02 dB436A, 437B, 438A
Power Sensor Freq: 0.050 to 20.05 GHz
Range: –30 to +20 dBm
Power Sensor Freq: 0.050 to 40 G Hz
Range: –30 to +20 dBm
Digital VoltmeterResolution: 10 mVAnyT
OscilloscopeBandwidth: 100 MHzAnyT
PrinterRaster graphics capabilityAnyP
Critical Specifications Recommended Model or
HP/Agilent Part Number
Test Instruments
5350B, 53150AP,A,T
5351B, 53151AP,A,T
E4418A (EPM-441A)
E4419A (EPM-442A)
8485AP,A,T
8487AP,A,T
Use
P,A,T
Calibration and Verification Kits
2.4 mm Calibration Kit No substitute85056AP,A,T
2.4 mm Verification Kit No substitute85057BP
3.5 mm Calibration KitNo substitute85052BP,A,T
3.5 mm Verification Kit
7 mm Calibration KitNo substitute85050BP,T
7 mm Ve rification Kit
Type-N Calibration KitNo substitut e85054BP,T
Type-N Verification Kit
1
2
3
No substitute85053BP
No substitute85051BP
No substitute85055AP
P - Performance TestsA - AdjustmentsT - Troubleshooting
1. Verification can only be done up to 26.5 GHz on the Agilent 8722ET/ES while using the 3.5
mm Verification Kit.
2. Verification can only be done up to 18 GHz, while using the 7mm Verification Kit.
3. Verification can only be done up to 18 GHz, while using the Type-N Verification Kit.
Service Guide1-3
Service Equipment and Analyzer Options8719ET/20ET/22ET
Service Test Equipment8719E S/20ES/22ES
RF Network Analyzers
Required EquipmentCritical Specificat ions Recommended Model or
HP/Agilent Part Number
Adapt ers and Loads
RF Load3.5 mm(f), 50 ohm 00902-60004
(Part of 85052B)
RF Load2.4 mm(f), 50 ohm 00901-60004
(Part of 85056A)
RF Short3.5 mm (f) 85052-60007
(Part of 85052B)
RF Short2.4mm (f)85056-60021
(Part of 85056A)
RF Open3.5 mm(f) 85052-60009
(Part of 85052B)
RF Open2.4 mm(f) 85056-60023
(Part of 85056A)
RF Adapter3.5 mm(f) to 3.5 mm(f) 85052-60012
(Part of 85052B)
RF Adapter2.4 mm(f) to 2.4 mm(f) 85056-60006
(Part of 85056A)
RF Power Splitter3.5 mm 11667BT
RF Power Splitter2.4 mm11667CT
Use
P
P
P
P
A
A
P,A
P,A
RF Fixed Attenuator3.5 mm 8493C Opt. 010A
RF Fixed Attenuator2.4 mm8490D Opt. 010A
RF Adapter2.4 mm (f) to Type-N (f)11903BA
RF Adapter3.5 mm (f) to Type-N (f)1250-1745 (2 total)A
24-inch
RF Cable Set3.5 mm connectors85131FP,A,T
RF Cable Set7 mm connectors to 3.5 mm85132FP
RF Cable Set 2.4 mm connectors85133FP,A,T
Exten si o n Ca b l esSMB (f ) to BNC (m) (4 ft . le n g t h)8120–5048T
Extension CablesSMB (m) to SMB (f)8120–5040A,T
Coax CableBNC (m) to BNC (m), 50W10503AA
GPIB Cables1 meter length10833AA
P - Performance TestsA - AdjustmentsT - Troubleshooting
11500C (2 total)P,A
1-4Service Guide
8719ET/20ET/22ET Service Equipment and Analyzer Options
8719ES/20ES/22ESService Test Equipment
RF Network Analyzers
Required EquipmentCritical Specifications Recommended Model or
HP/AgilentPart Number
Tools and Static Safety Par ts
Non-Metallic Adjust Tool8830-0024A
Tool Kit No substitute 08722-60018P,A,T
Anti-Static Wrist Strap9300-1367A,T
Anti-Static Wrist Strap Cord9300-0980A,T
Static Control Table Ma t and
P - Performance TestsA - AdjustmentsT - Troubleshooting
9300-0797A,T
Use
Service Guide1-5
Service Equipment and Analyzer Options8719ET/20ET/22ET
Principles of Microwave Connector Care8719ES/20ES/22ES
RF Network Analyzers
Principles of Microwave Connector Care
Proper connector care and connection technique are critical for accurate and repeatable
measurements.
Refer to the calibration kit documentation for connector care infor mati on. Prior to making
connections to the network analyzer, carefully review the information about inspecting,
cleaning, and gauging connectors.
Practicing good connector care and connection technique extends the life of these devices.
In addition, you obtain the most accurate measurements.
For additional connector care instruction, contact your nearest Agilent Technologies sales
or service office about course numbers HP/Agilent 85050A+24A and HP/Agilent
85050A+24D.
See the following table for quick reference tips about connector care.
Extend sleeve or connector nutSet connectors contact- end down
Use plastic end -c a ps du r i ng sto rag e
Visual Inspecti on
DoInspect all connectors carefullyDo NotUse a damaged connector - ever
Look for metal particles, scratches, and dents
Connector Cleaning
DoTry compressed air firstDo NotUse any abrasi ve s
Use isopropyl alcoholGet liquid into plastic support beads
Clean connect or threads
Gauging Conne c tors
DoClean and zero the gauge bef ore useDo NotUse an out-o f-spe c connec tor
Use the correct gauge type
Use correct end of calibration block
Gauge all connectors before first use
Making Connec tions
DoAlign conne ct ors carefull yDo NotApply bending force to connection
Make preliminary connection lightlyOver tighten preliminary connection
Turn only the connec tor nutTwist or screw any connection
Use a torque wrench for final connectTighten past torque wrench “break” point
1-6Service Guide
8719ET/20ET/22ET Service Equipment and Analyzer Options
8719ES/20ES/22ESAnalyzer Options
RF Network Analyzers
Analyzer Options
Option 1D5, High Stability Frequency Reference
Option 1D5 offers ±0.05 ppm temperature stability from 0 to 55 °C (referenced to 25 °C).
Option 004, Step Attenuator (ET Models)
This option adds a 55 dB step attenuator into the RF output path.
Option 007, Mechanical Transfer Switch (ES Models)
This option replaces the solid state transfer switch with a mechanical transfer switch,
providing the analyzer with greater power handling capability. Because the mechanical
transfer switch has less loss than the solid state transfer switch, the output power of an
Option 007 analyzer is 5 dB higher.
Option 010, Time Domain
This option allows the analyzer to display the time domain response of a network by
computing the inverse Fourier transform of the frequency domain response. The analyzer
shows the response of a test device as a function of time or distance. Displaying the
reflection coefficient of a network versus time determines the magnitude and location of
each discontinuity. Displaying the transmission coefficient of a network versus time
determines the characteristics of individual transmiss ion paths . Ti me domain operation
retains all accuracy inherent with the active error correction.
Option 012, Direct Access Receiver Configuration (ES Models)
This option provides front panel access to the A and B samplers. This allows direct access
to the sampler inputs for improved sensitivity in applicat ions s uc h as antenna tests, or for
the insertion of attenuators between the couplers and samplers. Direct access to the
B (A) sampler provides increased dynamic range in the forward (reverse) direction.
Service Guide1-7
Service Equipment and Analyzer Options8719ET/20ET/22ET
Analyzer Options8719ES/20ES/22ES
RF Network Analyzers
Option 085, High Power System (ES Models)
This option is designed to permit the measurement of high power devices. With an
external power amplifier, this config ur ation will al low up to 20 Watts (+43 dBm) of output
at the test ports. The maximum test port input power is 1 Watt (+30 dBm) CW, but
jumpers on the front panel allow the insertion of high power attenuators or isolators. This
allows test device output levels up to the power limits of the inserted components.
Additionally, there is an external reference input that allows the external amplifier’s
frequency response and drift to be ratioed out, and there are internally controlled step
attenuators between the couplers and samplers to prevent overload. A network analyzer
with this option can be configured to operate as a normal instrument (with slightly
degraded output power level and accuracy) or as an instrument capable of making single
connection multiple measurements. Bias tees are not part of the signal separation
circuitry of the Option 085. Because of high output power, Option 085 is only available
with a mechanical transfer switch similar to Option 007. Option 085 also includes direct
access to the receiver (Option 012).
Option 089, Frequency Offset Mode (ES Models)
This option adds the ability to offset the source and receiver frequencies for frequency
translation measurements . This provides the analyzer with mixer measurement capabil ity.
It also provides a graphical setup that allows easy configuration of your frequency
translation measurement.
Option 400, Four-Sampler Test Set (ES Models)
This option reconfigures the analyzer’s test set to ratio out the characteristics of the test
port transfer switch, and to include a second reference channel that allows full accuracy
with a TRL measurement calibration.
NOTEThe 8722ES option 400 is optimized to operate with a tes t port cable length of
one meter. It may be necessary to set up the 8722ES Option 400 in the step
sweep mode if you are operating the analyzer with test port cables or
fixturing with an electronic length greater than one meter.
Option 1CP, Rack Mount Flange Kit With Handles
Option 1CP is a rack mount kit containing a pair of flanges, with handles, and the
necessary hardware to mount the analyzer in an equipment rack wit h 482.6 mm
(19 inches) horizontal spacing.
Option 1CM, Rack Mount Flange Kit Without Handles
Option 1CM is a rack mount kit containing a pair of flanges, without handles, and the
necessary hardware to mount the analyzer in an equipment rack wit h 482.6 mm
(19 inches) horizontal spacing.
1-8Service Guide
8719ET/20ET/22ET Service Equipment and Analyzer Options
8719ES/20ES/22ESService and Support Options
RF Network Analyzers
Service and Support Options
The anal yz e r’s stan dard warranty is a three-year re tu rn to A gilent servic e war ranty.
The following service and support options are available at the time you purchase an
Agilent 8 719ET/ES, Agilent 8720ET/ES, or an Agilent 8722ET/ES network analyzer.
Option W01
This option converts the standard three-year return to Agilent Technologies service
warranty t o a one-year on-site service warranty. This option may not be available in all
areas.
Option W31
This option converts the standard three-year return to Agilent Technologies service
warranty t o a three-year on-site service warranty. This option may not be available in all
areas.
Option W50
This option adds two years to the standard three-year return to Agilent Technologies
warranty for a total of a five-year return to Agilent Technologies service warranty.
Option W51
This option converts the standard three-year return to Agilent Technologies service
warranty t o a fiv e-y ear on-site se rvice warrant y. This opt i on may not be available in all
areas.
Option W32
This option provides yearly return to Ag ilent Technologies for a commercial calibration for
a period of three years. The calibration provided is traceable to national standards.
Option W52
This option provides yearly return to Ag ilent Technologies for a commercial calibration for
a period of five years. The calibration provided is traceable to national standards.
Option W34
This option provides yearly return to Agilent Technologies for a Standards Compliant
Calibration for a period of three years. This type of calibration meets the ANSI/NCSL
Z540–1–1994 standard.
Service Guide1-9
Service Equipment and Analyzer Options8719ET/20ET/22ET
Service and Support Options8719ES/20ES/22ES
RF Network Analyzers
Option W54
This option provides yearly return to Agilent Technologies for a Standards Compliant
Calibration for a period of five years. This type of calibration meets the ANSI/NCSL
Z540–1–1 994 standard.
NOTEIf the previous servi ce and support optio ns were not purchas ed along with the
analyzer, there are many other repair and calibration options available from
Agilent Technologies’ support organization. These options cover a range of
on-site services and agreements with varying response times as well as
return to Agilent Technologies agreements and per-incident pricing. Contact
the nearest Agilent Technologies sales or service office for details. Refer to
Chapter 15 , “Safety and Regulatory Information.”
1-10Service Guide
2System Verification and
Performance Tests
2-1
System Verification and Per formance Tests 8719ET/20ET/22ET
Sections in This Chapter 8719ES/20ES/22ES
RF Network Analyzers
Sections in This Chapter
This chapter consists of five sections.
How to Test the Performance of Your Analyzer
•ANSI/NCSL Z540–1–1994 Verification‚ on page 2-3
•Non-AN S I/ NC SL Z540–1–1994 Verification‚ on page 2-3
•Instrum ent Verification C ycl e, on page 2-4
•ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart, on page 2-4
•Non–ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart, on page 2-5
Preliminary Tests
•Check the Temperature and Humidity, on page 2-6
•Clean and Gauge All Connectors, on page 2-7
•Perform the Internal Test, on page 2-7
•Procedure to Perform Operator’s Check, on page 2-8
•Check the Test Port Cables, on page 2-9
System Verification
•Equipment Initialization, on page 2-20
•Measurement Calibration, on page 2-22
•Verification Device Measurements, on page 2-25
•Interpreting the Verification Results, on page 2-31
Performance Tests
•Frequency Accuracy Performance Test, on page 2-34
•Level Accuracy Performance Test, on page 2-36
•Source Linearity Performance Test, on page 2-38
•Dynamic Range Performance Test, on page 2-42
Performance Test Records
•Performance Test Record, on page 2-46
2-2Service Guide
8719ET/20ET/22ET System Ver ificati on and Performance T ests
8719ES/20ES/22ESHow to Test the Performance of Your Analyzer
RF Network Analyzers
How to Test the Performance of Your Analyzer
There are two different way s to verif y the perfor mance of your a nalyzer. One method meets
ANSI/NCSL Z540–1–1994 standards, and the other method does not. To determine which
type of verification you wish to perform, refer to the follo wing descr ipti ons and flow charts.
ANSI/NCSL Z540–1–1994 Verification
This type of veri f icat ion co nsis ts of c onduc ting the preliminary checks , system verifica tion,
and the performance tests w ithout stoppi ng to troubleshoot . Exceptions will only be made
in case of catastrophic fa ilure or ca ble connector damage . In order to obtain da ta of how the
analyzer was performing at the time of verification, these tests must be done even if you
are aware that the analyzer will not pass. Obtaining th e data ( s ystem veri fi c ation p r int out
and performance test record) at this point is necessary so that you will understand that
your measurements may not have been accurate. You must wait until after the
ANSI/NCSL Z540–1–1994 verification is complete before troubleshooting and repairing
any problems. After troubleshooting , co nsult Table 14-1 o n page 14-35 to find the
necessary adjustment procedures. Then repeat the system verification and performance
tests, generating a new set of data.
Non-ANSI/NCSL Z540–1–1994 Verification
This type of veri f icat ion co nsis ts of c onduc ting the preliminary checks , system verifica tion,
and perfor mance tests, but stoppin g at any point if the analyzer fails a test. You will
troubleshoot and repair the first problem encountered without continuing to other tests.
After troubleshooting, consult Table 1 4-1 on page 1 4-35 to find the necessary adjustment
procedures. Then repeat the system verification and performance tests. As the analyzer
passes the system verification and all the tests, you will print out the system verification
results and fill in the performance test record.
Service Guide2-3
System Verification and Per formance Tests 8719ET/20ET/22ET
How to Test the Perf or mance of Your Analyzer 8719ES/20ES/22ES
RF Network Analyzers
Instrument Verification Cycle
The performance of the network analyzer should be verified at least once per year. The
following flowcharts illustrate the test path for both types of verifications.
Figure 2-1 ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
2-4Service Guide
8719ET/20ET/22ET System Ver ificati on and Performance T ests
8719ES/20ES/22ESHow to Test the Performance of Your Analyzer
RF Network Analyzers
Figure 2-2 Non–ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
Service Guide2-5
System Verification and Per formance Tests 8719ET/20ET/22ET
Preliminary Tests 8719ES/20ES/22ES
RF Network Analyzers
Preliminary Te sts
CAUTIONUse an antistatic work surface and wrist strap to lessen the chance of
electrostatic discharge.
Required Equipment and Tools
Equipment Description8719/20 8722
Calibration Kit 85052B85056A
RF Cable Set 85131F85133F
Verification Kit 85053B 85057B
Check the Temperature and Humidity
1. Measure the temperature and humidity of the environment and record the values in
Table 2-14, “Performance Test Record for All Models,” on page 2-46. The performance is
specified at an ambient temperature of +23 ±3 °C. Therefore, the environmental
temperature must remain in the range of +20 °C to +26 °C. Once the measurement
calibration has been done, the ambient temperature must be held to range of ± 1 °C.
2. Open the calibration and verifica tio n kits a nd pla ce all the d evi c es on top of the foam so
they will reach room temperatur e. The temperature of the devices is important because
device dimensions and therefore the electrical chara ct eri stic s ch ange with temperature .
3. Switch on the power to the analyzer.
NOTETo achieve the maximum system stability, allow the analyzer to warm up for
at least 30 minutes.
2-6Service Guide
8719ET/20ET/22ET System Ver ificati on and Performance T ests
CAUTIONTo prevent damage to the calibration kit and verification kit devices always
replace or repair any damaged connectors before proceeding with these tests.
NOTEAlways use adapters when verifying a system with SMA connectors.
1. Visually inspect all the connectors f or any burr s, gold flakes, or places where the gold is
worn.
2. Clean all the connectors with alcohol and foam-ti pped sw a bs. Dry the connectors with a
dry foam-tipped swab.
3. Visually inspect the calibration block and the end of the connector gauge before any
measurements are made.
4. Gauge all devices, cables, and test port connectors . Ref er to the calibrat ion kit manuals
for the correct use of gauges.
Perform the Internal Test
This quick, automated internal test may save time by indicating an instrument fault
before doing the performance tests. The tests performed are described in
Chapter 10 , “Service Key Menus and Error Messages.”
NOTETo achieve the maximum system stability, allow the analyzer to warm up for
at least 30 minutes.
To run the analyzer internal test, press the following:
Preset
System
A PASS message will appear on the display if the instrument passes this test.
Preset
Service Guide2-7
System Verification and Per formance Tests 8719ET/20ET/22ET
PRESET : FACT ORY
SERVICE MENU
TESTS
EXTERNAL TESTS
EXECUTE TEST
CONTI NUE
CONTI NUE
PRESET : FACT ORY
SERVICE MENU
TESTS
EXECUTE TEST
CONTI NUE
CONTI NUE
PRESET : FACT ORY
SERVICE MENU
TESTS
EXECUTE TEST
CONTI NUE
Preliminary Tests 8719ES/20ES/22ES
RF Network Analyzers
Procedure to Perform Operator’s Check
PORT 1 or REFLECTION port Check
1. Connect a short to PORT 1 (REFLECTION port on ET models).
2. Press the following:
Preset
System
Preset
• If the message PRESS[CONTI NUE] appears on the display during the test, that
particular attenuator setting check has failed. Press to check the
other attenuator settings .
• If the message FAIL appears on the analyzer display, the analyzer has failed the
PORT 1 operation check. Refer to the flow charts on page 2-4 and page 2-5.
• If the message DONE appears on the analyzer display, the analyzer has passed the
PORT 1 operation check.
PORT 2 Check (E S Mode ls)
3. Connect the short to PORT 2.
4. Press the following:
Preset
System
22x1
Preset
• If the message PRESS[CONTI NUE] appears on the display during the test, that
particular attenuator setting check has failed. Press to check the
other attenuator settings .
• If the message FAIL appears on the analyzer display, the analyzer has failed the
PORT 2 operation check. Refer to the flow charts on page 2-4 and page 2-5.
• If the message DONE appears on the analyzer display, the analyzer has passed the
PORT 2 operation check.
Transmission Port Check (ET Models)
5. Connect an RF cable from the REFLECTION port to the TRANSMISSION port.
6. Press the following:
Preset
System
22x1
2-8Service Guide
Preset
Loading...
+ 452 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.