Agilent 8712ET User Manual

User’s Guide
Agilent Technologies
8712ET and 8714ET
RF Network Analyzers
Part No. 08714-90011
Printed in USA
Print Date: June 2000
Supersedes: October 1999
Notice
Softkey
The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Key Conventions
This manual uses the following conventions:
FRONT PANEL KEY
This represents a key physically located on the instrument (a “hardkey”).
This indicates a “softkey,” a key whose label is determined by the instrument’s firmware, and is displayed on the right side of the instrument’s screen next to the eight unlabeled keys.
Screen Text
This indicates text displayed on the instrument’s screen.
Safety Information
For safety and regulatory information see Chapter 10, “Safety and
Regulatory Information.” For warranty and assistance information see Chapter 9, “Specifications.”
Firmware Revision
This manual documents analyzers with firmware revisions E.06.00 and above.
ii ET User’s Guide
Acknowledgments
Lotus® 1-2-3® are U.S. registered trademarks of Lotus Development Corporation.
Windows, Word97, and Excel97 are registered trademarks of Microsoft Corp.
Portions of the software include source code from the Info–ZIP group. This code is freely available on the Internet by anonymous ftp asftp.uu.net:/pub/archiving/zip/unzip51/.tar.Z, and from CompuServe asunz51.zip in the IBMPRO forum, library 10, (data compression).
ET User’s Guide iii

Introducing the Analyzer

The Agilent Technologies 8712ET and 8714ET are easy-to-use RF network analyzers optimized for production measurements of reflection and transmission parameters. The instrument integrates an RF synthesized source, transmission/reflection test set, multi-mode receivers, and display in one compact box.
The source features 1 Hz resolution, 40 ms (or faster) sweep time, and up to +16 dBm output power.
The three-channel, dual mode receivers provide dynamic range of greater than 100 dB in narrowband-detection measurement mode. For measurements of frequency-translating devices, the network analyzer features broadband internal detectors and external detector inputs. The receivers incorporate digital signal processing and microprocessor control to speed operation and measurement throughput.
Two independent measurement channels and a large display show the measured results of one or two receiver channels in several user-selectable formats. An external VGA monitor can be connected to the rear panel for enhanced measurement viewing in color.
Measurement functions are selected with front panel hardkey and softkey menus. Measurements can be printed or plotted directly with a compatible peripheral. Instrument states can be saved to the internal floppy disk, internal non-volatile memory, or internal volatile memory. Built-in service diagnostics are available to simplify troubleshooting procedures
Measurement calibrations and data averaging provide performance improvement and flexibility. Measurement calibrations consist of normalizing data, utilizing the internal factory calibration, or calibrating with external standards. Measurement calibration reduces errors associated with crosstalk, directivity, frequency response, and source match. Refer to Chapter 9, “Specifications,” for error correction specifications.
iv ET User’s Guide
How to Use This Guide
The first 6 chapters of this guide explain how to perform measurements, calibrate the instrument, and use the most common instrument functions.
Chapters 7 through 11 are reference material. Use these chapters to look up information such as front panel features, specific key functions, and specifications.
ET User’s Guide v
8712ET and 8714ET Network Analyzer Documentation Map
The CDROM provides the contents of all of the documents listed below.
The User’s Guide shows how to make measurements, explains commonly-used features, and tells you how to get the most performance from the analyzer.
The LAN Interface User’s Guide Supplement shows how to use a local area network (LAN) for programming and remote operation of the analyzer.
The Automating Measurements User’s Guide Supplement provides information on how to configure and control test systems for automation of test processes.
The Programmer’s Guide provides programming information including GPIB and SCPI command references, as well as short programming examples.
The Example Programs Guide provides a tutorial introduction using BASIC programming examples to demonstrate the remote operation of the analyzer
.
vi ET User’s Guide
The Service Guide provides the information needed to adjust, troubleshoot, repair , and verify analyzer conformance to published specifications.
The HP Instrument BASIC User’s Handbook describes programming and interfacing techniques using HP Instrument BASIC, and includes a language reference.
The HP Instrument BASIC User’s Handbook Supplement shows how to use HP Instrument BASIC to program the analyzer.
The Option 100 Fault Location and Structural Return Loss Measurements User’s Guide Supplement provides theory and measurement examples for making fault location and SRL measurements. (Shipped only with Option 100 analyzers.)
The CATV Quick Start Guide provides abbreviated instructions for testing the quality of coaxial cables. (Shipped only with Option 100 analyzers.)
The Cellular Antenna Quick Start Guide provides abbreviated instructions for verifying the performance of cellular antenna systems. (Shipped only with Option 100 analyzers.)
ET User’s Guide vii
Contents
1. Installing the Analyzer
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Step 1. Check the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Step 2. Meet Electrical and Environmental Requirements. . . . . . . . . . . . . . . . . . . . . . 1-4
Step 3. Check the Analyzer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Step 4. Configure the Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Connecting Peripherals and Controllers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Installing the Analyzer in a Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Clean the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Check the RF Front Panel Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
2. Getting Started
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Front Panel Tour . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Entering Measurement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Presetting the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Entering Frequency Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Entering Source Power Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Scaling the Measurement Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Entering Active Measurement Channel and Measurement Type. . . . . . . . . . . . . . .2-7
Viewing Measurement Channels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Performing the Operator's Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Equipment List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Make a Transmission Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Make a Broadband Power Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Make a Reflection Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
If the Analyzer Fails the Operator's Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
3. Making Measurements
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Measuring Devices with Your Network Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Attenuation and Amplification in a Measurement Setup . . . . . . . . . . . . . . . . . . . . . 3-8
When to Change the System Impedance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
The Typical Measurement Sequence. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Using the BEGIN Key to Make Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Contents-ix
Contents
BEGIN Key Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Using the BEGIN Key to Configure Measurements . . . . . . . . . . . . . . . . . . . . . . . . .3-13
AUTOST Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
The User BEGIN Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
Measuring Transmission Response. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16
Enter the Measurement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16
Perform an Enhanced Response Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
Measuring Reflection Response. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-33
Enter the Measurement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-33
Perform a One-Port Reflection Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-34
Connect the DUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-36
View and Interpret the Reflection Measurement Results . . . . . . . . . . . . . . . . . . . .3-38
Making a Power Measurement using Broadband Detection . . . . . . . . . . . . . . . . . . . .3-40
Enter the Measurement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-41
Perform a Normalization Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-42
Connect the DUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-43
View and Interpret the Power Measurement Results. . . . . . . . . . . . . . . . . . . . . . . .3-44
Measuring Conversion Loss. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-46
Enter the Measurement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-47
Perform a Normalization Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-47
Connect the DUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-49
View and Interpret the Conversion Loss Results . . . . . . . . . . . . . . . . . . . . . . . . . . .3-50
Making Measurements with the Auxiliary Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-52
Auxiliary Input Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-53
Measuring Group Delay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-54
Enter the Measurement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-55
Perform an Enhanced Response Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-55
Connect the DUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-56
View and Interpret the Group Delay Measurement Results . . . . . . . . . . . . . . . . . .3-56
Measuring Impedance using the Smith Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-58
Enter the Measurement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-58
Perform a One-Port Reflection Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-59
Connect the DUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-59
View and Interpret the Impedance Measurement Results. . . . . . . . . . . . . . . . . . . .3-60
Measuring Impedance Magnitude. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-64
How the Reflection Measurement Works. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-64
How the Transmission Measurement Works. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-65
Contents-x
Contents
4. Using Instrument Functions
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Using Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
To Activate Markers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
To Turn Markers Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
To Use Marker Search Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
To Use Marker Math Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
To Use Delta () Marker Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
To Use Other Marker Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
To Use Polar Format Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
To Use Smith Chart Markers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Using Limit Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
To Create a Flat Limit Line. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
To Create a Sloping Limit Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
To Create a Single Point Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
To Use Marker Limit Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
To Use Relative Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Other Limit Line Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
Additional Notes on Limit Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Using Reference Tracking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-42
To Track the Peak Point. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
To Track a Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-44
Customizing the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Using the Split Display Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-46
Enabling/Disabling Display Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
Modifying Display Annotation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
Expanding the Displayed Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-52
Using Memory Traces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-55
To Store a Data Trace to the Display Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-55
To View the Measurement Data and Memory Trace . . . . . . . . . . . . . . . . . . . . . . . . 4-55
To Divide Measurement Data by the Memory Trace . . . . . . . . . . . . . . . . . . . . . . . . 4-55
Saving and Recalling Measurement Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-57
Saving Instrument Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-59
To Recall from a Floppy Disk or Internal Memory. . . . . . . . . . . . . . . . . . . . . . . . . . 4-62
Other File Utilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-65
To Use Directory Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-67
Formatting a Floppy Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-68
Contents-xi
Contents
Connecting and Configuring Printers and Plotters . . . . . . . . . . . . . . . . . . . . . . . . . . .4-69
Select a Compatible Plotter or Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-69
Select an Appropriate Interface Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-70
Connect the Printer or Plotter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-71
Configure the Hardcopy Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-72
Define the Printer or Plotter Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-74
Printing and Plotting Measurement Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-79
To Select the Copy Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-79
To Define the Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-80
Using a Keyboard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-84
To Connect the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-84
To Use the Keyboard to Edit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-84
Front Panel Control using a Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-85
Using an External VGA Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-88
Customizing Color on an External Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-88
Synchronizing and Positioning the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-90
5. Optimizing Measurements
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Increasing Sweep Speed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
To Increase the Start Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
To Set the Sweep Time to AUTO Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
To Widen the System Bandwidth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
To Reduce the Amount of Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
To Reduce the Number of Measurement Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
To View a Single Measurement Channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
To Turn Off Alternate Sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-7
To Turn Off Markers and Marker Tracking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-7
To Turn Off Spur Avoidance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
To Avoid Frequency Bandcrossings (8714ET only) . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
Increasing Network Analyzer Dynamic Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
To Increase the Receiver Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
To Reduce the Receiver Noise Floor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Reducing Trace Noise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
To Activate Averaging for Reducing Trace Noise . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
To Change System Bandwidth for Reducing Trace Noise . . . . . . . . . . . . . . . . . . . .5-12
Contents-xii
Contents
To Eliminate Receiver Spurious Responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Reducing Mismatch Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Reducing Mismatch Errors in a Reflection Measurement. . . . . . . . . . . . . . . . . . . . 5-15
Reducing Mismatch Errors in a Transmission Measurement. . . . . . . . . . . . . . . . .5-16
Reducing Mismatch Errors when Measuring Both Reflection
and Transmission. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Compensating for Phase Shift in Measurement Setups . . . . . . . . . . . . . . . . . . . . . . . 5-17
Port Extensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Electrical Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Measuring Devices with Long Electrical Delay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
6. Calibrating for Increased Measurement Accuracy
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Measurement Calibration Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
The Calibration Reference Plane. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Default versus User-Defined Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
When to Use a Default Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
When to Perform a User-Defined Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Calibration Choices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Retrieving Previous User-Defined Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Presetting the Analyzer: How Calibration Is Affected. . . . . . . . . . . . . . . . . . . . . . . 6-10
To Perform a Normalization Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
To Perform a Transmission Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-12
To Perform a Reflection Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
To Perform a Conversion Loss Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
Calibration Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Selecting a Calibration Kit Stored in the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Creating a User-Defined Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-18
Saving and Recalling the Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Saving the Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Recalling the Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Checking the Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Using Calibration Check for Analysis and Troubleshooting . . . . . . . . . . . . . . . . . . 6-28
To Perform a Calibration Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29
Error Term Descriptions and Typical Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-31
Contents-xiii
Contents
7. Front/Rear Panel
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
BNC Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Multi-pin Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
RF Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-16
Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-18
Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-19
Display Intensity Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-21
Disk Drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-22
Line Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-23
Power Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-23
The Line Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-25
The Voltage Selector Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-26
8. Hardkey/Softkey Reference
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
Numeric Entries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-6
B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-12
C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-15
D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-21
E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-28
F. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-31
G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-36
H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-39
I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-42
K . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-45
L. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-46
M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-52
N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-60
O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-63
P. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-64
R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-69
S. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-76
Contents-xiv
Contents
T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-89
U . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-94
V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-96
W. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-98
X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-99
Y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-100
Z . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-102
9. Specifications
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
System Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Test Port Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
Test Port Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-29
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-35
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-35
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-37
Data Hardcopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-37
Automation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-38
Measurement Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-40
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-43
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-45
Limitation of Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-46
Exclusive Remedies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-46
Agilent Technologies Sales and Service Offices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-47
10. Safety and Regulatory Information
Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-2
Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-4
Statement of Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-4
Cleaning Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Shipping Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Instrument Markings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Regulatory Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Contents-xv
Contents
Notice for Germany: Noise Declaration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-6
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-6
11. Factory Preset State and Memory Allocation
Factory Preset and Peripheral States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-2
Factory Preset State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-2
Peripheral State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-8
Volatile Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-12
Save/Recall Memory Allocation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-13
Types of Storage Disks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-13
Types of Storable Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-15
How to Determine the Size of Disk Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-16
Memory Usage Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-18
Contents-xvi

1 Installing the Analyzer

ET User’s Guide 1-1
Installing the Analyzer

Introduction

Introduction
This chapter will guide you through the four steps needed to correctly and safely install your network analyzer. The four steps are:
“Step 1. Check the Shipment” on page 1-3
“Step 2. Meet Electrical and Environmental Requirements” on
page 1-4
“Step 3. Check the Analyzer Operation” on page 1-9
“Step 3. Check the Analyzer Operation” on page 1-9
1-2 ET User’s Guide
Installing the Analyzer

Step 1. Check the Shipment

Step 1. Check the Shipment
After you have unpacked your instrument, it is recommended that you keep the packaging materials so they may be used if your instrument should need to be returned for maintenance or repair.
NOTE The packaging material is designed to protect the analyzer from damage
that can happen during shipping. Returning the analyzer in anything other than the original packaging may result in non-warranted damage.
Check the items received against the Product Checklist (included in your shipment) to make sure that you received everything.
Inspect the analyzer and all accessories for any signs of damage that may have occurred during shipment. If your analyzer or any accessories appear to be damaged or missing, call your nearest Agilent Technologies sales or service office. Refer to Table 9-8, “Agilent Technologies Sales and
Service Offices,” on page 9-48 for the nearest office.
ET User’s Guide 1-3
Installing the Analyzer

Step 2. Meet Electrical and Environmental Requirements

Step 2. Meet Electrical and Environmental Requirements
1. Set the line voltage selector to the position that corresponds to the ac power source you will be using.
CAUTION Before switching on this instrument, make sure that the line voltage
selector switch is set to the voltage of the mains supply and the correct fuse (T 5 A 250 V) is installed. Assure the supply voltage is in the specified range.
NOTE The working fuse and a spare are located in the power cable receptacle.
See Figure 7-13 on page 7-25.
Figure 1-1 Voltage Selector Switch Location
1-4 ET User’s Guide
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
2. Ensure the available ac power source meets the following requirements:
Nominal
Setting
115 V 230 V
If the ac line voltage does not fall within these ranges, an autotransformer that provides third-wire continuity to ground should be used.
3. Ensure the operating environment meets the following requirements for safety:
• indoor use
• altitude up to 15,000 feet (4,572 meters)
• temperature 0 °C to 55 °C
• maximum relative humidity 5 to 95 percent relative at +40 °C
(non-condensing)
CAUTION This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.
NOTE The above requirements are for safety only. Separate conditions that
must be met for specified performance are noted in Chapter 9,
“Specifications.”
AC Line Power
90 to 132 Vac (47 to 63 Hz)
198 to 264 Vac (47 to 63 Hz)
ET User’s Guide 1-5
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
4. V erify that the power cable is not damaged, and that the power source outlet provides a protective earth ground contact. Note that the following illustration depicts only one type of power source outlet. Refer to Figure 7-12 on page 7-24 to see the different types of power cord plugs that can be used with your analyzer.
Figure 1-2 Protective Earth Ground
WARNING This is a Safety Class I product (provided with a protective
earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor , inside or outside the instrument, is likely to make the product dangerous. Intentional interruption is prohibited.
1-6 ET User’s Guide
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
WARNING If this instrument is to be energized via an external
autotransformer for voltage reduction, make sure that its common terminal is connected to a neutral (earthed pole) of the power supply.
5. Install the analyzer so that the detachable power cord is readily identifiable and is easily reached by the operator. The detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument. The front panel switch is only a standby switch and not a LINE switch. Alternatively, an externally installed switch or circuit breaker (which is readily identifiable and is easily reached by the operator) may be used as a disconnecting device.
6. Install the analyzer according to the enclosure protection provided. This instrument does not protect against the ingress of water. It does protect against finger access to hazardous parts within the enclosure.
7. Ensure there are at least two inches of clearance around the sides and back of either the stand-alone analyzer or the system cabinet.
Figure 1-3 Ventilation Clearance Requirements
ET User’s Guide 1-7
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
8. Set up a static-safe workstation. Electrostatic discharge (ESD) can damage or destroy components.
• table mat with earth ground wire: part number 9300-0797
• wrist-strap cord with 1 Meg Ohm resistor: part number 9300-0980
• wrist-strap: part number 9300-1367
• heel straps: part number 9300-1308
• floor mat
1-8 ET User’s Guide
Installing the Analyzer

Step 3. Check the Analyzer Operation

Step 3. Check the Analyzer Operation
1. Turn on the line switch of the analyzer. After approximately 30 seconds, a display box should appear on the screen with the following information:
• the model number of your analyzer (either 8712ET or 8714ET)
• the firmware revision
• the serial number of your analyzer
• installed options
2. Verify that the serial number and options displayed on the screen match the information on the rear panel serial label.
3. The operator's check should be performed on the analyzer to provide a high degree of confidence that the analyzer is working properly. Refer to Chapter 2, “Getting Started,” for instructions on how to perform the operator's check.
ET User’s Guide 1-9
Installing the Analyzer
Step 4. Configure the Analyzer
Step 4. Configure the Analyzer
You can begin making measurements by simply connecting your analyzer to an appropriate power source and turning it on. This section, however, will explain how to connect common peripherals and controllers, and how to install your analyzer into a rack system.
1-10 ET User’s Guide
Installing the Analyzer
Step 4. Configure the Analyzer

Connecting Peripherals and Controllers

Figure 1-4 Analyzer Rear Panel Line Module and Selected Connectors
Refer to Figure 1-4:
• The GPIB port is for use with computers and peripherals (printers, plotters, etc.).
• The parallel and RS-232 (serial) ports are also for peripherals. The parallel and serial ports can also be programmed via IBASIC for general I/O control. See the HP Instrument BASIC User's Handbook for information on using IBASIC.
ET User’s Guide 1-11
Installing the Analyzer
Step 4. Configure the Analyzer
• The VIDEO OUT COLOR VGA port allows you to connect a color VGA monitor for enhanced viewing. See “Using an External VGA
Monitor” on page 4-88 for more information.
• The LAN ETHERTWIST connector is for connecting your analyzer to a LAN (local area network) for control and access. See The LAN Interface User’s Guide Supplement for information on how to use your analyzer in a LAN.
GPIB Connections An GPIB system may be connected in any configuration as long as the
following rules are observed:
• The total number of devices is less than or equal to 15.
• The total length of all the cables used is less than or equal to 2 meters times the number of devices connected together up to an absolute maximum of 20 meters. For example, the maximum cable length is 4 meters if only 2 devices are involved. The length between adjacent devices is not critical as long as the overall restriction is met.
See Figure 1-5 for different connection configurations.
1-12 ET User’s Guide
Figure 1-5 GPIB Connection Configurations
Installing the Analyzer
Step 4. Configure the Analyzer
Table 1-1 Maximum GPIB Cable Lengths
Instruments/Peripherals in System
Two
Fifteen (max)
ET User’s Guide 1-13
Maximum GPIB Cable Length between Each Pair of Devices
4 m
20 m (total)
Installing the Analyzer
Select Copy Port
Select
Print/Plot GPIB Addr
Enter
Select Copy Port
Select
Print/Plot GPIB Addr
Enter
Step 4. Configure the Analyzer
Parallel and Serial Connections
Other Connections If you plan to use a keyboard, barcode reader, external video monitor, or
To Set GPIB Addresses
Parallel and serial devices often require specific cables—check their manuals for details. Parallel cable length should not exceed 25 feet. The analyzer may experience problems talking to a printer if this length is exceeded. Connect the required control cables and secure them. (Tighten the knurled screws or comparable fasteners.)
external detectors, connect them to the appropriate rear panel connectors. See Figure 1-4, “Analyzer Rear Panel Line Module and
Selected Connectors.” Also see Chapter 7, “Front/Rear Panel,” for more
information on front and rear panel connectors. To communicate via GPIB, each external device must have a unique
address and the network analyzer must recognize each address. T o check or set each external device's actual address, refer to the device's manual (most addresses are set with switches).
The following are examples of how to check or set the device's recognized address on the network analyzer:
Printer: Press . Use the
front panel knob to highlight the line that reads HP Printer PCL GPIB. Press . The second line of the screen displays settings: in this case the address. The default address is 5, however most printers are factory set to address 1 (one). To change the recognized address, press
HARDCOPY
number
.
NOTE Only one hardcopy GPIB address can be set at a time. Changing the
Plotter: Press . Use the
front panel knob to highlight the line that reads HP Plotter HPGL GPIB. Press . The second line of the screen displays settings: in this case the address. The default address is 5 and most plotters are factory set to address 5, so changing the address is probably not necessary. To change the recognized address, press
printer address, for example, changes the plotter to the same address.
1-14 ET User’s Guide
HARDCOPY
number
.
Installing the Analyzer
GPIB
8712ET Address
8714ET Address
Enter
Select Copy Port
Select
Select Copy Port
Select
Select Copy Port
Select
LAN Printr IP Addr
Select Copy Port
Step 4. Configure the Analyzer
To Configure Peripheral Settings
Analyzer: Press ,
network analyzer's address will appear (the default is
16). To change the address, press , .
If your system uses serial or parallel peripherals, follow the guidelines below to configure the system. Refer to the peripheral's manual for correct cables and settings. The parallel and serial ports have standard Centronics DB-25 and RS232 pinouts respectively as explained in
Chapter 7, “Front/Rear Panel.”
Serial Devices: Press . Use the
entry controls to highlight your type of printer or plotter, and press . If the baud rate or handshake at the top of the screen are incorrect, use the softkeys to change them.
Parallel Devices: Press . Use the
entry controls to highlight your type of printer or plotter, and press .
LAN Printer: Press . Use the
entry controls to highlight HP LaserJet PCL5/6 PCL5 LAN, and press . If the printer IP
address at the top of the screen is incorrect, press
SYSTEM OPTIONS
or . The
number
HARDCOPY
HARDCOPY
HARDCOPY
to enter the correct IP address.
NOTE When is selected, the first two lines in the box at
NOTE Use a PCL5 printer for fastest hardcopies. See “Configure the Hardcopy
the top of the display screen show the current settings.
Port” on page 4-72 for more information.
ET User’s Guide 1-15
Installing the Analyzer
Step 4. Configure the Analyzer

Installing the Analyzer in a Rack

Use only the recommended rack mount kit (Option 1CM when ordered with the analyzer or part number 08712-60036 when ordered separately) with this instrument; it needs side support rails. Do not attempt to mount it by the front panel (handles) only. This rack mount kit allows you to mount the analyzer with or without handles.
To install the network analyzer in an HP/Agilent 85043D rack, follow the instructions in the rack manual.
CAUTION To install the network analyzer in other racks, note that they may
promote shock hazards, overheating, dust contamination, and inferior system performance. Consult your Agilent customer engineer about installation, warranty, and support details.
CAUTION VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
Place other system instruments (computer, printer, plotter) where convenient, within the GPIB cable length limits (see Table 1-1,
“Maximum GPIB Cable Lengths,” on page 1-13) or other interface
cabling limits.
1-16 ET User’s Guide
Installing the Analyzer

Preventive Maintenance

Preventive Maintenance
Preventive maintenance consists of two tasks. It should be performed at least every six months—more often if the instrument is used daily on a production line or in a harsh environment.

Clean the Display Use a soft cloth and, if necessary, a mild cleaning solution to clean the

display.

Check the RF Front Panel Connectors

Figure 1-6 Maximum and Minimum Protrusion of Center Conductor from
Visually inspect the front panel connectors. The most important connectors are those to which the DUT is connected, typically the RF cable end or the RF IN connector. All connectors should be clean and the center pins centered. The fingers of female connectors should be unbroken and uniform in appearance. If you are unsure whether the connectors are good, gauge the RF IN and RF OUT connectors to confirm that their dimensions are correct.
Mating Plane
ET User’s Guide 1-17

2 Getting Started

ET User’s Guide 2-1
Getting Started

Introduction

Introduction
The 8712ET and 8714ET are easy-to-use, fully integrated RF component test systems.Each instrument includes a synthesized source, a wide dynamic range receiver, and a built-in test set. Controls are grouped by functional block, and settings are displayed on the instrument screen. This section familiarizes new users with the layout of the front panel and the process of entering measurement parameters into the analyzer.
Figure 2-1 Network Analyzer Front Panel Features
2-2 ET User’s Guide
Getting Started
Softkeys
Sweep Time

Front Panel Tour

Front Panel Tour
1 The CRT Display The analyzer's large CRT displays data, markers, limit lines, Instrument BASIC
(IBASIC) programming code, softkey menus, and measurement parameters quickly and clearly. Refer to “Display” on page 7-16 for more information.
2 The key simplifies measurement setups. The key allows quick
BEGIN BEGIN BEGIN
and easy selection of basic measurement parameters for a user-selected class of devices (e.g., filters, amplifiers, or mixers). For example, when making a transmission measurement, selecting Filter as your device type puts the analyzer into narrowband detection mode, maximizing measurement dynamic range. In comparison, selecting Mixer as your device type puts the analyzer into broadband detection mode, enabling frequency-translation measurements. This capability allows new users to start making measurements with as few as four keystrokes.
3 MEAS The measure keys select the measurements for each measurement channel. The
analyzer's measurement capabilities include transmission, reflection, power, conversion loss, and multiport selection (for use with multiport test sets).
4 SOURCE The source keys select the desired source output signal to the device under test, for
example, selecting source frequency range or output power. The source keys also control sweep time, number of points, and sweep triggering.
5 CONFIGURE The configure keys control receiver and display parameters. These parameters include
receiver bandwidth and averaging, display scaling and format, marker functions, and instrument calibration.
6 SYSTEM The system keys control system level functions. These include instrument preset,
save/recall, and hardcopy output. GPIB parameters and IBASIC are also controlled with these system keys.
7 The Numeric Keypad Use the number keys to enter a specific numeric value for a chosen parameter. Use the
key or the softkeys to terminate the numeric entry with the appropriate
.
8
HARDKEYS
9
ENTER
units. You can also use the front panel knob for making continuous adjustments to parameter values, while the keys allow you to change values in steps. Hardkeys are front panel keys physically located on the instrument front panel. In text,
these keys will be represented by the key name with a box around it such as:
PRESET
Softkeys are keys whose labels are determined by the analyzer's firmware. The labels are on the screen next to the 8 blank keys, which are located along the right edge of the analyzer’s display. In text, these softkeys will be represented by the key name with
shading behind it such as: .
ET User’s Guide 2-3
Getting Started
42.5
.

Entering Measurement Parameters

Entering Measurement Parameters
This section describes how to input measurement parameter information into the network analyzer.
NOTE When entering parameters, you can use the numeric key pad, as
described in each example, or you can use the keys or the front panel knob to enter data.
NOTE When you are instructed to enter numeric values in this manual, it often
can get cluttered and confusing to depict each key stroke. So in this manual, numbers (no matter how many characters) are depicted inside one keycap. F or example, if you are instructed to enter the number42.5, it will be depicted inside one keycap like this: . To enter this number , the following keys need to be pressed in succession:
.
5
You can follow along with these examples by connecting the filter and cable that were supplied with your instrument as shown in Figure 2-2.
4 2
Figure 2-2 Connect the Filter to the Analyzer
2-4 ET User’s Guide
Getting Started
User Preset
Factory Preset
User Preset
Factory Preset
Entering Measurement Parameters

Presetting the Analyzer

Press the key and then press either or
PRESET
. The key returns the analyzer to user-defined preset settings you may have saved as file UPRESET.STA. If this file doesn’t exist, you can create it by following the instructions in the pop-up message that appears when the key is pressed. When the
key is pressed, the following major default conditions apply:
Frequency range (8712ET) 0.3 to 1300 MHz Frequency range (8714ET) 0.3 to 3000 MHz Power level Measurement Channel 1
1
0 dBm Transmission
measurement Measurement Channel 2
Off
measurement Format Log Magnitude Number of points 201 Sweep time Auto
NOTE The measurement parameters that you enter will be retained in the
NOTE Refer to Chapter 11, “Factory Preset State and Memory Allocation, ” for a
Scale 10 dB/div Reference 0 dB System Bandwidth Medium wide
1. Preset power level can be set to other than 0 dBm if desired. See “Entering Source Power Level” on page 2-6 for more information.
analyzer's memory when the power is turned off, and will be restored when the power is turned back on.
comprehensive table of factory preset conditions.
ET User’s Guide 2-5
Getting Started
Start
MHz
Stop
MHz
Center
Span
Disp Freq Resolution
Level
dBm
Level
1.6
dBm
Pwr Level at Preset
2.5
dBm
Autoscale
Scale/Div
Enter
Reference Position
Enter
Reference Level
Enter
Entering Measurement Parameters

Entering Frequency Range

1. Press the key to access the frequency softkey menu.
FREQ
2. To change the low end of the frequency range to 10 MHz, press
10
.
3. To change the high end of the frequency range to 900 MHz, press
900
.
4. You can also set the frequency range by using the and softkeys. For instance, if you set the center frequency to
160 MHz and the span to 300 MHz, the resulting frequency range would be 10 to 310 MHz.
NOTE When entering frequencies, be sure to terminate your numeric entry
with the appropriate softkey to obtain the correct units. If you use the
ENTER
key to terminate a frequency entry, the units default to Hz.
The default displayed frequency resolution is kHz. You can change the resolution by pressing , and then
FREQ
selecting a new resolution.

Entering Source Power Level

1. Press the key to access the power level softkey menu.
2. To change the power level to 3 dBm, press and or
ENTER
3. To change the power level to1.6 dBm, press
or .
ENTER
POWER
3
.

Scaling the Measurement Trace

4. To change the power level that will always be restored when you
preset the analyzer, press and or . This entry does not affect the current power level.
ENTER
1. Press the key to access the scale menu.
SCALE
2. To view the complete measurement trace on the display, press
.
3. To change the scale per division to 5 dB/division, press
5
.
4. To move the reference position (indicated by the symbol on the left
side of the display) to the first division down from the top of the display, press . Figure 2-3 shows
9
how each reference position is identified.
5. To change the reference level to 0 dB, press
0
.
2-6 ET User’s Guide
Figure 2-3 Reference Positions
User Preset
Factory Preset
Transmissn
Reflection
Getting Started
Entering Measurement Parameters

Entering Active Measurement Channel and Measurement Type

The and keys allow you to choose which
MEAS 1 MEAS 2
measurement channels are active, and measurement parameters for the channels. When a particular measurement channel is selected, its display is brighter than the other channel, and any changes made to measurement parameters will affect only the selected measurement channel. (Some measurement parameters cannot be independently set on each measurement channel. F or these parameters , both c hannels will be affected regardless of selected channel status.)
1. To measure transmission on measurement channel 1 and reflection on measurement channel 2, press the following keys:
PRESET
MEAS 2
2. Both channels' measurements are now visible on the analyzer's display screen. Note that the selected measurement channel's (channel 2) measurement trace is brighter than the other measurement channel's trace. Refer to Figure 2-4.
or
MEAS 1
ET User’s Guide 2-7
Getting Started
Meas OFF
More Display
Split Disp FULL split
Entering Measurement Parameters
Figure 2-4 Both Measurement Channels Active

Viewing Measurement Channels

3. To view only the measurement channel 2 reflection measurement, press .
4. To view both measurement channels again, press .
5. To view both measurement channels separately on a split screen, press . Refer to
Figure 2-5, “Split Display.”
2-8 ET User’s Guide
MEAS 1
MEAS 1
DISPLAY
Figure 2-5 Split Display
Getting Started
Entering Measurement Parameters
You have now learned how to enter common measurement parameters and how to manipulate the display for optimum viewing of your measurement. You can now proceed on to performing the operator's check, or refer to Chapter 3, “Making Measurements,” for detailed information on making specific types of measurements.
ET User’s Guide 2-9
Getting Started

Performing the Operator's Check

Performing the Operator's Check
The operator's check should be performed when you receive your instrument, and any time you wish to have confidence that the analyzer is working properly. The operator's check does not verify performance to specifications, but should give you a high degree of confidence that the instrument is performing properly if it passes.
The operator's check consists of making the following measurements with the cable that was supplied with your analyzer:
• transmission
• broadband power
• reflection
• reflection (with a 50 or 75 load, instead of the cable)

Equipment List

To perform the operator's check, you will need the following:
• A known good cable such as the one that was supplied with your analyzer. The cable you use should have 0.5 dB of insertion loss up to 1.3 GHz and 0.75 dB of insertion loss from 1.3 to 3.0 GHz.
• A known good load (> 40 dB return loss) that matches the test port impedance of your analyzer such as one from calibration kit 85032B/E (50 ) or 85036B/E (75 ).
2-10 ET User’s Guide
Getting Started
User Preset
Factory Preset
Enter
dBm
Default Response
Performing the Operator's Check

Make a Transmission Measurement

1. Connect the equipment as shown in Figure 2-6. Use a known good cable such as the one that was supplied with your analyzer.
NOTE The quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in “Equipment List” on
page 2-10.
Figure 2-6 Equipment Setup for Performing the Operator’s Check
2. Press or
3. Press .
4. Press .
5. Verify that the data trace falls within ±0.5 dB of 0 dB. See Figure 2-7 for a typical 8714ET result. The 8712ET should look similar, but end at 1300 MHz.
ET User’s Guide 2-11
PRESET
.
POWER 0 CAL
SCALE .1
Getting Started
Performing the Operator's Check
Figure 2-7 Verify Transmission Measurement
2-12 ET User’s Guide
Getting Started
Power
Start
MHz
Enter
dBm
Performing the Operator's Check

Make a Broadband Power Measurement

1. Leave the cable connected to the analyzer as shown in Figure 2-6.
NOTE The quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in “Equipment List” on
page 2-10.
2. Press
3. Press (unless done in the previous measurement).
4. V erify that the data trace is within±2 dB of 0 dBm. See Figure 2-8 for a typical 8714ET result. The 8712ET should look similar, but end at 1300 MHz.
Figure 2-8 Verify Broadband Power Measurement
1
MEAS 1
.
POWER 0
FREQ
10
SCALE
ET User’s Guide 2-13
Getting Started
User Preset
Factory Preset
Reflection
Enter
dBm
Default 1-Port
Performing the Operator's Check
Make a Reflection Measurement
1. Leave the cable connected to the analyzer as shown in Figure 2-6.
NOTE The quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in “Equipment List” on
page 2-10.
2.
Press or
PRESET
SCALE 10
MEAS 1
.
3. Press .
4. Press .
5. Verify that the data trace falls completely below 16 dB. See Figure
2-9 for a typical 8714ET result. The 8712ET should look similar, but
end at 1300 MHz.
POWER 0 CAL
2-14 ET User’s Guide
Figure 2-9 Verify Reflection Measurement
Getting Started
Performing the Operator's Check
6. Disconnect the cable and connect a known good load to the RF OUT port as shown in Figure 2-10.
ET User’s Guide 2-15
Getting Started
Reference Level
Trigger
Hold
Performing the Operator's Check
Figure 2-10 Connect the Load
7. Verify that the data trace falls below 30 dB. If the data trace is off the screen, press and the key until the trace moves up onto the screen.
SCALE
This concludes the operator's check. However, further confidence can be obtained by performing the following:
• Measure a known filter to verify that its measured response is the same as is expected. (A 175 MHz filter is supplied with the analyzer.) Verify both the frequency accuracy and noise floor.
• Check broadband response with the filter using conversion-loss mode (same as B*/R*).
• If the analyzer's frequency accuracy is critical for your application, verify a CW frequency using a frequency counter. Verify to ±.005% accuracy (for example, ± 2500 Hz at 500 MHz). Ensure that the analyzer is placed in trigger-hold mode (press
) to measure frequencies.
2-16 ET User’s Guide
MENU
Getting Started
Performing the Operator's Check

If the Analyzer Fails the Operator's Check

First, repeat the operator's check using a different cable and load to eliminate these as a possible cause of failure. If your analyzer does not meet the criteria in the operator's check, your analyzer may need adjustment or servicing. Contact any Agilent Technologies sales or service office for assistance. Refer to Table 9-8, “Agilent Technologies
Sales and Service Offices,” on page 9-48 for the nearest office. Before
shipping your analyzer, fill out and attach the blue repair tag, located at the back of the analyzer’s Service Guide.
ET User’s Guide 2-17

3 Making Measurements

ET User’s Guide 3-1
Making Measurements

Introduction

Introduction
This chapter provides an overview of basic network analyzer measurement theory, a section explaining the typical measurement sequence, a segment describing the use of the key, and detailed examples of the following measurements:
“Measuring Transmission Response” on page 3-16
“Measuring Reflection Response” on page 3-33
“Making a Power Measurement using Broadband Detection” on
page 3-40
“Measuring Conversion Loss” on page 3-46
“Making Measurements with the Auxiliary Input” on page 3-52
“Measuring Group Delay” on page 3-54
“Measuring Impedance using the Smith Chart” on page 3-58
“Measuring Impedance Magnitude” on page 3-64
BEGIN
3-2 ET User’s Guide

Measuring Devices with Your Network Analyzer

This section provides a basic overview of how the network analyzer measures devices. The analyzer has an RF signal source that produces an incident signal that is used as a stimulus to the device under test. Your device responds by reflecting a portion of the incident signal and transmitting the remaining signal. If the device is passive, some of the transmitted signal will be absorbed, indicating a “lossy” device. If the device is active, the transmitted signal may be amplified, indicating that the device has gain. Figure 3-1 shows how a device under test (DUT) responds to an RF source stimulus.
Figure 3-1 DUT Response to an RF Signal
Making Measurements
Measuring Devices with Your Network Analyzer
Refer to Figure 3-2 for the following discussion regarding detection schemes and modes. The transmitted signal (routed to input B) and the reflected signal (input A) are measured by comparison to the incident signal. The analyzer couples off a small portion of the incident signal to use as a reference signal (routed to input R). Sweeping the source frequencies, the analyzer measures and displays the response of your test device. Figure 3-2 shows the transmitted, reflected, and reference signal inputs.
ET User’s Guide 3-3
Making Measurements
Measuring Devices with Your Network Analyzer
Figure 3-2 Simplified Block Diagram
3-4 ET User’s Guide
This page intentionally left blank.
Making Measurements
Measuring Devices with Your Network Analyzer
ET User’s Guide 3-5
Making Measurements
Measuring Devices with Your Network Analyzer
Refer to Figure 3-3 for the following discussion. The network analyzer receiver has two signal detection modes:
• broadband detection mode
• narrowband detection mode
There are two internal broadband detectors: B* and R*. External broadband detectors can also be used when connected to the X and Y ports on the rear panel of the analyzer. When the network analyzer is in the broadband detection mode, it measures the total power of all signals present at these measurement ports, independent of signal frequency. This enables the characterization of frequency translation devices such as mixers, receivers, and tuners, where the RF input and output frequencies are not the same. Figure 3-3 labels the transmitted signal for broadband detection input as B*, and the reference signal as R*.
When the network analyzer is in the narrowband detection mode, the receiver is tuned to the source frequency. This technique provides greater dynamic range by decreasing the receiver's bandwidth. Figure 3-3 shows receiver B as the narrowband detection input for the transmitted signal, and receiver R as the narrowband detection input for the reference signal.
3-6 ET User’s Guide
Figure 3-3 Block Diagram
Making Measurements
Measuring Devices with Your Network Analyzer
ET User’s Guide 3-7
Making Measurements
Measuring Devices with Your Network Analyzer
The following table shows the correlation between different types of measurements, input channels, and signals.
Measurement Detection Mode Input Channels Input Signals
Transmission Reflection Power Conversion Loss
Narrowband Narrowband Broadband Broadband
B/R A/R
B*
B*/R*
transmitted/incident reflected/incident transmitted transmitted/incident
Attenuation and Amplification in a Measurement Setup
The measurement setup that you use may require attenuation or amplification. The following sections describe when to use them.
When to Use Attenuation
• For accurate measurements, use external attenuation to limit the power at the RF IN port to +10 dBm (for narrowband-detection measurements) or +16 dBm (for broadband measurements).
CAUTION Always use attenuation on the TRANSMISSION RF IN port if your test
device's output power exceeds the receiver damage limit of +20 dBm or ±30 Vdc.
• Use attenuation on the RF IN port to reduce mismatch errors. See
“Reducing Mismatch Errors” on page 5-15 for more information.
3-8 ET User’s Guide
Making Measurements
More Cal
System Z0
50Ω 75
Measuring Devices with Your Network Analyzer
When to Use Amplification
• For accurate measurements, amplification may be needed on the analyzer's RF OUT port. Use amplification when your test device requires input power that exceeds the analyzer's maximum specified output power.
NOTE If you use an amplifier between the analyzer’s output port and your DUT,
you won’t be able to measure the reflection response of your DUT.
The maximum specified output power is dependent upon the option configuration of your analyzer as well as the frequency range of your test setup. It ranges from +6 to +16 dBm. See Chapter 9,
“Specifications,” to determine the maximum specified output power of
your analyzer.

When to Change the System Impedance

Your analyzer has a system characteristic impedance of either 50 or 75 ohms, yet may be changed to the alternate impedance. If using minimum-loss pads for impedance conversions, the alternate impedance should be selected so that the measurement results are displayed relative to the conversion impedance.
For example, if you have a 50 ohm instrument and are making 75 ohm measurements, you may be using 50 to 75 ohm minimum-loss pads. Measurement results can be reported relative to 75 ohms, not 50 ohms, if the alternate system impedance is selected. This includes marker readouts, Smith chart results, or SRL impedance computations (Option 100).
To change the system impedance, press the following keys on the analyzer:
CAL
οr
The built-in cal kit selections will be converted to the selected system impedance.
ET User’s Guide 3-9
Making Measurements
Measuring Devices with Your Network Analyzer

The Typical Measurement Sequence

A typical measurement consists of performing four major steps:
Step 1. Enter the Measurement Parameters
Step 2. Calibrate the Analyzer
Step 3. Connect the Equipment
Step 4. View and Interpret the Measurement
The easiest way to set up the analyzer's parameters for a simple measurement is to use the key. When this key is selected, the analyzer automatically sets up a generic set of parameters to match the device type you choose. (See “Using the BEGIN Key to Make
Measurements” on page 3-11.)
For measurements that require you to enter your own specific measurement parameters (such as frequency range, source power level, number of points, and sweeptime), use the instrument’s keys to enter your selections rather than using the key. See the measurement examples, located later in this chapter.
Your analyzer can provide highly accurate measurements without performing any additional user-defined calibrations if certain conditions are met. Chapter 6, “Calibrating for Increased Measurement Accuracy,” explains when additional user-defined calibration is necessary.
Connect the DUT and any other required test equipment. See the measurement examples later in this chapter for typical equipment setup configurations.
Use the , , and functions to optimize viewing of the measurement results.
Markers, limit lines, and hard copies of the display are common means of interpreting measurement results.
SCALE DISPLAY FORMAT
BEGIN
BEGIN
See Chapter 4, “Using Instrument Functions,” for detailed information on using instrument functions to view and interpret your measurements.
3-10 ET User’s Guide

Using the BEGIN Key to Make Measurements

Figure 3-4 The BEGIN Key
Making Measurements
Using the BEGIN Key to Make Measurements
The key allows you to quickly and easily configure the analyzer
BEGIN
(from the condition) to measure the following:
• amplifiers
• filters
• broadband passive devices (a cable, for example)
• mixers
• cable fault location and structural return loss (Option 100 only)
Configuring basic measurements from the key helps you ensure correct instrument setup. The analyzer guides you through the initial steps and configures itself for the device type you select.
ET User’s Guide 3-11
PRESET
BEGIN
Making Measurements
Using the BEGIN Key to Make Measurements

BEGIN Key Overview

The key sets up a generic instrument state for the testing of
BEGIN
various types of devices. The key has two different behaviors, depending on whether
BEGIN
you are selecting a new device type, or a new measurement type.
Selecting a New Device
Selecting a New Measurement
When you use the key to select a new device type and measurement, the analyzer does the following:
• presets the analyzer (except for external reference parameters, and trigger mode)
• takes a sweep
• autoscales the measurement
• places a marker on the maximum or minimum point (depending on the type of measurement)
• makes the marker active
• modifies the sweep time (Option 100 only)
See Table 3-1, “Measurement Configurations from the BEGIN Key,” on
page 3-14 for a table of parameters for each measurement type.
Once you have selected your device, you can use the softkeys to select the measurement you wish to make. When you select a new measurement, a preset is not done. It is assumed that you are simply changing measurement types and that you may have already changed some of the analyzer's parameters (such as frequency, power, etc.) for your DUT. In this situation, you would probably not want these parameters changed for subsequent measurements.
BEGIN
NOTE If the new measurement selected is a broadband measurement such as
power, or conversion loss, the start frequency is limited to at least 10 MHz. Therefore, if your customized setup contains a start frequency below 10 MHz and you choose power, or conversion loss, the start frequency will be changed to 10 MHz. The stop frequency will remain unchanged, unless it was set to below 10 MHz.
3-12 ET User’s Guide
Making Measurements
User Preset
Factory Preset
Transmissn
Reflection
Power
Power
Amplifier
Conversion Loss
Conversion Loss
Mixer
Using the BEGIN Key to Make Measurements
The BEGIN Key and Measurement Channels
The key is designed to work when measurement channel 1 is
BEGIN
active. However, it does change the mode of measurement channel 2 as well.
If measurement channel 2 is active when the key is used to select a new device type, measurement channel 2 is turned off, and measurement channel 1 is made active.
If measurement channel 2 is active when the key is used to select a new measurement type, measurement channel 2 will be left on and active. However, the analyzer then proceeds to set up channel 1 for the requested measurement type, even though channel 2 is the active channel.
BEGIN
BEGIN
Using the BEGIN Key to Configure Measurements
This procedure shows you how to configure the network analyzer for measurements.
1. Press ( or ). Presetting the instruments puts it into a known state with predefined parameters.
2. Press and then use a softkey to select the type of device that you will be measuring (amplifier, filter, broadband passive device, mixer, or cable—Option 100 only).
PRESET
BEGIN
3. Connect your test device to the network analyzer.
4. Use the softkeys to select the type of measurement you want to make:
• Press if you want to measure the transmission
characteristics of an amplifier, filter, or broadband passive device.
• Press if you want to measure the reflection
characteristics of your device.
• Press if you want to measure the RF power of a device.
(The selection is under the menu.)
• Press if you want to measure the conversion
loss of a device. (The selection is under the
menu.)
ET User’s Guide 3-13
Making Measurements
SRL
SRL
Cable
Fault Location
Fault Location
Cable
SRL
Fault Location
Using the BEGIN Key to Make Measurements
• Press (Option 100 only) if you want to measure the structural return loss of a cable. (The selection is under the
menu.)
• Press (Option 100 only) if you want to measure the cable fault location. (The selection is under the menu.)
Depending on your selection, the analyzer is set to one of the following configurations. (The and configurations are discussed in the Option 100 User's Guide Supplement.)
Table 3-1 Measurement Configurations from the BEGIN Key
Transmission Reflection Power Conversion Loss
Frequency Range (8712ET)
Frequency Range (8714ET)
Measurement Channel 1
Detection Mode Narrowband Narrowband Broadband Internal Broadband Internal Measurement
Paths
0.300 MHz to 1300 MHz
0.300 MHz to 3000 MHz
Transmission Reflection Power Conversion Loss
B/R A/R B* B*/R*
0.300 MHz to 1300 MHz
0.300 MHz to 3000 MHz
10 MHz to 1300 MHz
10 MHz to 3000 MHz
10 MHz to 1300 MHz
10 MHz to 3000 MHz
The following measurement configurations apply to all of the measurement types in the previous table:
Power Level Factory preset or user preset power
level
1
Measurement Channel 2 Off Format Log Mag Number of Points 201 Sweep Time Mode Auto Sweep Triggering Continuous Averaging Off System Bandwidth Medium Wide
1. Factory preset power level is user-defined by using the “Pwr Level at Preset” softkey. The fac­tory default is 0 dBm.
3-14 ET User’s Guide
Making Measurements
Autost
User BEGIN
User BEGIN
Recall Program
User BEGIN on OFF
User BEGIN
User BEGIN
User BEGIN
User BEGIN
Using the BEGIN Key to Make Measurements

AUTOST Files

When the analyzer's power is turned on, it first checks for an IBASIC autostart file (AUTOSTART.BAS) on the non-volatile RAM disk and then on the 3.5” disk. If found, the file is loaded and run. This feature simplifies the task of turning on an automated test station at the beginning of a working day or test session. To manually load and run an autostart file, press .
BEGIN

The User BEGIN Function

The softkey gives you the capability to redefine the
BEGIN
key to define macros such as:
• softkeys to implement fast save/recall
• softkeys to implement most-used functions or features
• softkeys to implement often-used features that involve a number of
Macros must be defined with an IBASIC program. If no program is currently installed (either by AUTOST or
program.
key menu and install user-defined macro functions. Use this
steps
), the analyzer will automatically create a default
selects the key menu to "user" mode
when on, and to normal operation when off. Once you have changed the mode to on, the same menu
will be displayed for subsequent key presses of the hardkey. (This is not true if your IBASIC program has changed. If the program has changed, the mode is reset to off.)
Use of the function does not restrict access to any normally available instrument feature (such as marker functions), nor does this key affect sweep update rates.
Refer to the IBASIC example programs provided on the Example Programs Disk for implementation requirements. Keystroke recording may be used to modify or update programs.
See the Automating Measurements User’s Guide Supplement for more information.
ET User’s Guide 3-15
BEGIN
BEGIN
Making Measurements
User Preset
Factory Preset

Measuring Transmission Response

Measuring Transmission Response
This section uses an example measurement to describe how to calibrate for and make a basic transmission response measurement. This example uses an enhanced response calibration. See the table below for a list of all possible calibration choices for a transmission response measurement.
TIP An optimum calibration is critical for achieving best measurement
accuracy. Refer to Chapter 6, “Calibrating for Increased Measurement
Accuracy,” for detailed information about all of the various aspects of
calibration.
Table 3-2 Calibration Choices for a Transmission Response Measurement
Measurement Type Calibration Choices
Transmission Default Response
Response Response & Isolation Enhanced Response Normalize
NOTE This example measurement uses the default instrument parameters for
In this example, a bandpass filter like the one that was supplied with your network analyzer is used.

Enter the Measurement Parameters

Press ( or ) on the analyzer to set the analyzer to the default mode which includes measuring transmission on measurement channel 1.
a transmission measurement. If your particular transmission measurement requires specific parameters (such as frequency range, source power level, number of data points, and sweeptime), enter them now.
3-16 ET User’s Guide
PRESET
Making Measurements
Measuring Transmission Response

Perform an Enhanced Response Calibration

An enhanced response calibration uses known standards to correct for source match and frequency response errors when using narrowband detection. When you perform an enhanced transmission response calibration, the analyzer performs correction at each data point across the selected frequency band. The default number of data points per sweep is 201, but you can select any number between 3 and 1601. Interpolation recalculates the error correction array for reduced frequency spans. If the frequency span is increased, the calibration is invalidated, and the default response calibration is automatically restored.
To perform an enhanced transmission response calibration, you will need one of the following calibration kits depending on the nominal impedance of your analyzer:
Cal Kit Model
Number
85039B 75 type-F ✓✓ 85036B 75 type-N ✓✓ 85036E 75 type-N 85032B 50 type-N ✓✓ 85032E 50 type-N 85031B 50 APC-7 N/A N/A 85033D 50 3.5 mm ✓✓ 85038A 50 7-16 ✓✓ 85038F 50 7-16
85038M 50 7-16
Impedance
Connector
Type
For Use with Male Test Ports
For Use
with Female
Test Ports
ET User’s Guide 3-17
Making Measurements
More Cal
Cal Kit
Enhanced Response
Measuring Transmission Response
NOTE By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the analyzer is type-N female because the front panel RF ports are female (the calibration standards, in turn, have male connectors).
Calibrating for Insertable Devices
When measuring a through-standard during calibration, the test ports normally mate directly together. For example, two cables with the appropriate connectors can be connected directly without a through-adapter, resulting in a zero-length through-standard. An insertable device is a device that can be inserted in place of a zero-length through-standard. It has the same type of connector on each port, but with a different sex, or the same type of sexless connector on each port (APC-7, for example).
Chapter 6 provides detail about when this calibration is necessary, and
information about other calibrations available for transmission measurements. If you wish to perform an enhanced response calibration on your insertable device, perform the following steps:
NOTE If you are going to be using calibration standards other than the default
(female type-N), you must select the connector type by pressing
and then selecting the appropriate type.
1. Press .
2. The instrument prompts you to connect four standards—open, short, load, and through cable—as shown in the following graphic.
3-18 ET User’s Guide
CAL
CAL
Making Measurements
Measure Standard
Measuring Transmission Response
Open, Short, Load Connections Through-Cable Connection
3. Press after connecting each standard.
4. The analyzer will measure each standard and then calculate the new calibration coefficients. The message "Calibration complete." will appear for a few seconds when the analyzer is done calculating the new error correction array.
NOTE Changing sweep frequencies (and other source parameters) may affect
your calibration. See Chapter 6 for more information.
ET User’s Guide 3-19
Making Measurements
Measuring Transmission Response
Connect the DUT.
Figure 3-5 Equipment Setup for a Transmission Response Measurement
3-20 ET User’s Guide
Making Measurements
Autoscale
Marker Search
Max Search
Mkr−> Max
Measuring Transmission Response
View and Interpret the Transmission Measurement Results.
1. To view the entire measurement trace on the display, press
.
2. To interpret the transmission measurement, refer to Figure 3-6,
“Example of a Transmission Response Measurement Display,” or your
analyzer's display if you are making this measurement on your instrument.
a. The values shown on the horizontal axis are the frequencies in
MHz. The values shown on the vertical axis are the power ratios in decibels (dB) of the transmitted signal through the device divided by the incident power. To display the result in logarithmic magnitude format (designated by "Log Mag" at the top of the measurement screen), the analyzer computes the measurement trace using the following formula:
P
trans

Transmission dB() 10 log
where P where P
b. A level of 0 dB would indicate a lossless through cable or device
(no loss or gain). Values greater than 0 dB indicate that the DUT has gain. Values less than 0 dB indicate loss.
= the power transmitted through the device and
trans
= the incident power.
inc
=
------------- -

P
inc
SCALE
3. To quickly determine the filter's minimum insertion loss, press
MARKER
4. Note the marker readout in Figure 3-6 provides the frequency and amplitude of the minimum insertion loss point.
ET User’s Guide 3-21
.
Making Measurements
Measuring Transmission Response
Figure 3-6 Example of a Transmission Response Measurement Display
5. See “Using Markers” on page 4-3 for more detailed information on using markers to interpret measurements.
NOTE For the measurement to be valid, input signals must fall within the
dynamic range of the analyzer. See Chapter 5, “Optimizing
Measurements,” for techniques to increase the dynamic range of the
analyzer.
3-22 ET User’s Guide
Making Measurements
More Cal
Cal Kit
Type-N(m)
Measuring Transmission Response
Calibrating for Noninsertable Devices
A noninsertable device is one that cannot be inserted in place of a zero-length through-standard. It has the same connectors on each port (type and sex) or has a different type of connector on each port (SMA on one port, and type-N on the other, for example). The following two calibration methods are available for noninsertable devices:
• swap equal adapters
• modify the calibration kit definition of the through standard In the following examples, the noninsertable device is type-N, with
female input and output connectors.
Method A: Swap Equal Adapters.
With this method, you use two precision matched adapters which are “equal.” To be equal, the adapters must have the same match, characteristic impedance (Z0), insertion loss, and electrical delay. The adapters in most HP/Agilent calibration kits have matched electrical lengths, even if the physical lengths appear different.
NOTE For analyzers with 50 ohm input impedance only: in the 50 ohm type-N
calibration kit (85032B), there are four equal adapters: two APC-7 to type-N(f), and two APC-7 to type-N(m). To create adapter A in the following example, connect an APC-7-to-type-N(f) adapter with an APC-7 to type-N(m) adapter. To create adapter B, connect two APC-7 to type-N(m) adapters.
NOTE By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the analyzer is type-N female because the front panel RF ports are female (the calibration standards, in turn, have male connectors).
1. Connect a test cable with type-N(m) connectors to the TRANSMISSION port.
2. To select a calibration kit for the REFLECTION port, press
, and use the front panel knob or the
keys to highlight the REFLECTION port in the display’s table.
Press .
3. Connect adapter A (type-N(m) to type-N(f)) to the REFLECTION port.
ET User’s Guide 3-23
CAL
Making Measurements
Prior Menu
Prior Menu
Enhanced Response
Measure Standard
Measuring Transmission Response
4. Press .
5. When the instrument prompts you to connect a through cable, connect the TRANSMISSION port test cable to adapter A on the REFLECTION port as shown in Figure 3-7. Press
.
Figure 3-7 Through-Cable Connection
6. Remove adapter A, and place adapter B (type-N(m) to type-N(m)) on the REFLECTION port. Adapter B becomes the effective test port.
7. When the instrument prompts you, connect three standards (open, short, and load) to adapter B—as shown in Figure 3-8.
3-24 ET User’s Guide
Figure 3-8 Open, Short, Load Connections
Measure Standard
Making Measurements
Measuring Transmission Response
NOTE Changing sweep frequencies (and other source parameters) may affect
8. Press after connecting each standard.
9. The analyzer will measure each standard and then calculate the new calibration coefficients. The message "Calibration complete." will appear for a few seconds when the analyzer is done calculating the new error correction array.
your calibration. See Chapter 6 for more information.
ET User’s Guide 3-25
Making Measurements
Measuring Transmission Response
Connect the DUT.
10.Measure the test device with adapter B in place as shown in
Figure 3-9.
Figure 3-9 Equipment Setup for a Transmission Response Measurement of
a Noninsertable Two-Port Device
3-26 ET User’s Guide
Making Measurements
User Preset
Factory Preset
More Cal
Cal Kit
Type-N(m)
Reflection
Measuring Transmission Response
Method B: Modify the Calibration Kit Definition of the Through Standard.
With this method, it is only necessary to use one adapter. The DUT has type-N(f) input and output connectors. The calibration kit through-standard definition is modified to compensate for the adapter and then saved as a user kit. However, the electrical delay of the adapter must first be found.
NOTE By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the analyzer is type-N female because the front panel RF ports are female (the calibration standards, in turn, have male connectors).
NOTE For analyzers with 50 ohm input impedance only: in the 50 ohm type-N
calibration kit (85032B), there are four equal adapters: two APC-7-to-type-N(f), and two APC-7-to-type-N(m). To create the adapter in the following example, connect two APC-7-to-type-N(f) adapters.
1. Connect a test cable with type-N(m) connectors to the REFLECTION port.
2. Connect a test cable with type-N(m) connectors to the TRANSMISSION port.
3. To select a calibration kit for the REFLECTION port, press
PRESET
highlight the REFLECTION port in the display’s table. Press
4. Press . Perform a one-port reflection calibration using type-N(f) standards on the type-N(m) open-end of the REFLECTION port test cable. Refer to “Perform a One-Port
Reflection Calibration” on page 3-34.
5. Connect the type-N female-to-female adapter to the type-N(m) open­end of the REFLECTION port test cable.
6. Add a type-N(m) short to the open end of the adapter as shown in
Figure 3-10.
ET User’s Guide 3-27
( or )
, and use the front panel knob or the keys to
.
MEAS 1
CAL
Making Measurements
Delay
2:
Marker Functions
Marker Math
Statistics
More Cal
Cal Kit
Type-N(f)
Measuring Transmission Response
Figure 3-10 Short Connection to the Adapter
7. Select the delay format for your measurement by pressing
FORMAT
8. Press to automatically select marker 1 as the active marker. Position marker 1 to the left edge of the measurement trace.
9. Press to select marker 2 as the active marker. Position marker 2 to the right edge of the measurement trace.
10.Press . Find the mean value, located in the upper right corner of the display.
11.Divide the mean value by 2. Record this value for use later in the procedure.
12.Select a type-N(f) calibration kit for the REFLECTION port. This is necessary because the type-N(m) short (connected to the adapter on port 2) is a component in a type-N(f) calibration kit. Press
Press .
3-28 ET User’s Guide
MARKER
keys to highlight the REFLECTION port in the display’s table.
.
CAL
, and use the front panel knob or the
Measuring Transmission Response
Operating Parameters
Next Screen
More Cal
Cal Kit
Type-N(m)
More Cal
Cal Kit
Modify Type-N(m)
Thru
Delay
Prior Menu
Mod Kit SaveRecall
Save
Add Kit Description
Select Char
Enter
Enhanced Response
13.Determine the offset delay of the calibration short:
Making Measurements
• Press
• Press until the Cal Kit screen is displayed.
• In the “CAL KIT: Type N(m)” display, find the numeric value for the delay of the short.
14.Subtract this value from the value calculated in Step 11. This corresponds to the delay of the adapter.
15.Remove the short from the adapter.
16.Reselect a type-N(m) calibration kit for the REFLECTION port. This is necessary because the type-N(m) short has been removed from the adapter. Press , and use the front panel knob or the keys to highlight the REFLECTION port in the display’s table. Press .
17.Modify the calibration kit definition of the through-standard by entering the electrical delay of the adapter. To do so:
• Press
18.Save (and automatically assign) the modified cal kit buffer as a user-defined cal kit for the REFLECTION port. To do so:
• Press . Use the front panel
knob or the keys to highlight the desired User Kit in the display’s table. (In this example, highlight Cal Kit A.) Press
SYSTEM OPTIONS
CAL
CAL
. Enter the delay value with the keypad and softkeys.
.
If you want to enter a cal kit description in the display’s table, press . Use a keyboard connected to the analyzer’s DIN KEYBOARD connector to enter the description. (If you don’t have a keyboard, you can use the analyzer’s
keys or the front panel knob and the key to select and enter characters from the top of the analyzer’s display. Press when done.)
19.Start the calibration process by pressing
CAL
ET User’s Guide 3-29
.
MEAS 1
Transmissn
Making Measurements
Measure Standard
Measuring Transmission Response
20.When the instrument prompts you, connect three standards (open, short, and load) to the REFLECTION port test cable as shown in
Figure 3-11.
21.Press after connecting each standard.
Figure 3-11 Open, Short, Load Connections
3-30 ET User’s Guide
22.When the instrument prompts you to connect a through cable, insert
Measure Standard
the female-to-female adapter between the test cables connected to the REFLECTION port and the TRANSMISSION port as shown in
Figure 3-12. Press .
Figure 3-12 Through-Cable Connection
Making Measurements
Measuring Transmission Response
NOTE Changing sweep frequencies (and other source parameters) may affect
23.The analyzer will measure each standard and then calculate new calibration coefficients. The message "Calibration complete." will appear for a few seconds when the analyzer is done calculating the new error correction array.
24.Remove the adapter between the test cables connected to the REFLECTION port and the TRANSMISSION port.
your calibration. See Chapter 6 for more information.
ET User’s Guide 3-31
Making Measurements
Measuring Transmission Response
Connect the DUT.
25.Measure the test device using the setup shown in Figure 3-13.
Figure 3-13 Equipment Setup for a Transmission Response Measurement of
a Type-N(f) Two-Port Noninsertable Device
3-32 ET User’s Guide
Making Measurements
User Preset
Factory Preset
Reflection
Measuring Reflection Response
Measuring Reflection Response
This section uses an example measurement to describe how to calibrate for and make a basic reflection response measurement. In this example, a user-defined one-port calibration is used. See the table below for a list of all possible calibration choices for a reflection response measurement.
TIP An optimum calibration is critical for achieving best measurement
accuracy. Refer to Chapter 6, “Calibrating for Increased Measurement
Accuracy,” for detailed information about all of the various aspects of
calibration.
Table 3-3 Calibration Choices for a Reflection Response Measurement
Measurement Type Calibration Choices
Reflection Default 1-Port
1-Port
In this example, a bandpass filter like the one that was supplied with your network analyzer is used.
NOTE This example measurement uses the default instrument parameters for

Enter the Measurement Parameters

Press the following keys on the analyzer:
PRESET
( or )
MEAS 1
a reflection response measurement. If your particular reflection measurement requires specific parameters (such as frequency range, source power level, number of data points, and sweep time), enter them now.
ET User’s Guide 3-33
Making Measurements
Measuring Reflection Response
Perform a One-Port Reflection Calibration
A one-port reflection calibration uses known standards to correct for directivity, source match, and frequency response errors when using narrowband detection. When you perform a one-port reflection calibration, the analyzer performs correction at each data point across the selected frequency band. The default number of data points per sweep is 201, but you can select any number between 3 and 1601. Interpolation recalculates the error correction array for reduced frequency spans. If the frequency span is increased, the calibration is invalidated, and the default one-port calibration is automatically restored.
To perform a one-port reflection calibration, you will need one of the following calibration kits depending on the nominal impedance of your analyzer:
Cal Kit Model
Number
85039B 75 type-F ✓✓ 85036B 75 type-N ✓✓ 85036E 75 type-N 85032B 50 type-N ✓✓ 85032E 50 type-N 85031B 50 APC-7 N/A N/A 85033D 50 3.5 mm ✓✓ 85038A 50 7-16 ✓✓ 85038F 50 7-16
85038M 50 7-16
Impedance
Connector
Type
For Use with Male Test Ports
For Use
with Female
Test Ports
3-34 ET User’s Guide
Making Measurements
Cal Kit
1-Port
Measure Standard
Measuring Reflection Response
NOTE By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the analyzer is type-N female because the front panel RF ports are female (the calibration standards, in turn, have male connectors).
NOTE If you are going to be using calibration standards other than the default
(female type-N), you must select the connector type by pressing
and then selecting the appropriate type.
Chapter 6 provides detail about when this calibration is necessary. If you
wish to perform a one-port reflection calibration on your instrument for a reflection response measurement, execute the following steps:
CAL
1. Press .
2. The instrument will prompt you to connect three standards (open, short and load) and measure them. See Figure 3-14.
Figure 3-14 Open, Short, Load Connections
3. Press after connecting each standard.
CAL
4. The analyzer will measure each standard and then calculate new calibration coefficients. The message "Calibration complete." will appear for a few seconds when the analyzer is done calculating the new error correction array.
ET User’s Guide 3-35
Making Measurements
Measuring Reflection Response
NOTE Changing sweep frequencies (and other source parameters) may affect
your calibration. See Chapter 6 for more information.

Connect the DUT

Figure 3-15 Equipment Setup for a Reflection Measurement of a Two-Port
Device
NOTE If you connect the DUT between both analyzer ports, it is recommended
that you use a 10 dB pad on the output of the DUT to improve measurement accuracy. If you connect a two-port DUT to the REFLECTION port only, it is recommended that you use a high-quality load (such as a calibration standard) on the output of the DUT.
3-36 ET User’s Guide
Making Measurements
Measuring Reflection Response
Figure 3-16 Equipment Setup for a Reflection Measurement of a One-Port
Device
ET User’s Guide 3-37
Making Measurements
Autoscale
Measuring Reflection Response
View and Interpret the Reflection Measurement Results
1. To view the entire measurement trace on the display, press ,
.
2. To interpret the reflection measurement, refer to Figure 3-17,
“Example of a Reflection Measurement Display,” or your analyzer's
display if you are making this measurement on your instrument. a. The values shown on the horizontal axis are the frequencies in
MHz. The values shown on the vertical axis are the power ratios in decibels (dB) of the reflected signal divided by the incident power. To display the result in logarithmic magnitude format (designated by Log Mag at the top of the measurement screen), the analyzer computes the measurement trace using the following formula:
P
refl

Reflection dB() 10 log
where P where P
b. A level of 0 dB indicates that all of the power applied to the DUT is
reflected back, and that none of it passes through the DUT or is absorbed by the DUT.
c. Values less than 0 dB indicate that power is either absorbed or
transmitted by the DUT. Although they are not typically seen, values greater than 0 dB do occur under certain circumstances such as when the measurement needs to be enhanced by calibration, or when the device is active (an amplifier for instance) and perhaps oscillating.
= the power of the signal reflected from the device and
refl
= the incident power.
inc
=
----------

P
inc
SCALE
3-38 ET User’s Guide
Measuring Reflection Response
Figure 3-17 Example of a Reflection Measurement Display
Making Measurements
3. To quickly determine the filter's return loss, press and then use the front panel knob, the keys, or the numeric keypad to read the value of return loss at the desired frequency.
4. See “Using Instrument Functions” on page 4-1 for detailed information on using markers to interpret measurements.
ET User’s Guide 3-39
MARKER
Making Measurements
Detection Options
Narrowband Internal
B

Making a Power Measurement using Broadband Detection

Making a Power Measurement using Broadband Detection
Power measurements can be made using either narrowband or broadband detection. The example in this section is of a broadband power measurement. If you are only interested in the output power of your device at the same frequency as the analyzer's source, you can select for a narrowband-detection power measurement. A narrowband-detection power measurement only measures the power within the tuned receiver's bandwidth, centered at the source frequency.
When you measure a device for absolute output power, the network analyzer uses the broadband detection mode and measures the total power of all frequencies present in the transmitted signal (B*). This signal may contain frequencies other than the source frequency such as when the DUT is a mixer.
This section uses an example measurement to describe how to normalize the data and measure the total output power of an amplifier.
MEAS 1
NOTE Broadband power measurements are only specified for measurements
with a start frequency of 10 MHz.
3-40 ET User’s Guide
Making Measurements
User Preset
Factory Preset
More
Power
Making a Power Measurement using Broadband Detection

Enter the Measurement Parameters

Press the following keys on the analyzer:
PRESET
( or )
MEAS 1
NOTE This example measurement uses the default instrument parameters for
a power measurement. If your particular power measurement requires different parameters (such as source power level, number of data points, and sweep time), enter them now.
CAUTION Damage to your analyzer will occur if the receiver input power exceeds
+20 dBm or ±30 Vdc. The analyzer's source cannot significantly exceed this level, however if your DUT has gain, then attenuation on the analyzer’s RF IN port may be necessary. See “Attenuation and
Amplification in a Measurement Setup” on page 3-8 for more
information.
ET User’s Guide 3-41
Making Measurements
Start
MHz
Normalize
Normalize on OFF
Making a Power Measurement using Broadband Detection

Perform a Normalization Calibration

Normalization is the simplest type of calibration. The analyzer stores a measurement trace into memory and divides subsequent measurements by the stored data to remove unwanted frequency response errors. This calibration is used for the power measurement to remove the insertion-loss error of the test cable. Changing the frequency span or number of measurement points will invalidate a normalization calibration.
Execute the following steps to perform a normalization calibration:
1. Connect the equipment as shown in Figure 3-18.
Figure 3-18 Through-Cable Connection
2. Set the following frequency parameters:
FREQ
10
3. Press or . This stores the data into memory and divides subsequent
measurements by the stored data to remove frequency response errors.
3-42 ET User’s Guide
DISPLAY
CAL
Making a Power Measurement using Broadband Detection
4. Insert the amplifier as shown in Figure 3-19.

Connect the DUT

Figure 3-19 Equipment Setup for a Power Measurement
Making Measurements
ET User’s Guide 3-43
Making Measurements
Autoscale
Making a Power Measurement using Broadband Detection

View and Interpret the Power Measurement Results

1. To view the measurement trace, press .
2. To interpret the power measurement, refer to Figure 3-20 or your analyzer's display if you are making this measurement on your instrument.
a. When making a power measurement, the display shows the power
measured at the analyzer's RF IN connector. This power is absolute power, as opposed to a power ratio.
b. Note that when making a power measurement, the values
associated with the vertical axis are in units of dBm, which is the power measured in reference to 1 mW.
0 dBm = 1 mW
10 dBm = 100 µW
• +10 dBm = 10 mW Other units available for a power measurement include dBW,
dBµW, W, mW, and µW.
SCALE
3-44 ET User’s Guide
Making a Power Measurement using Broadband Detection
Figure 3-20 Example of a Power Measurement
Making Measurements
CAUTION If the analyzer's RF output power level is set to higher than the specified
output power for your analyzer, the source could go unleveled. See
Chapter 9, “Specifications,” for source and receiver specifications. If your
device requires input power greater than your analyzer's specified output power, you may need to use a preamplifier in your measurement setup. However, remember to not exceed the receiver damage limit of +20 dBm.
ET User’s Guide 3-45
Making Measurements

Measuring Conversion Loss

Measuring Conversion Loss
Conversion loss is the ratio of IF output power to RF input power expressed in dB. This section uses an example measurement to describe how to measure the conversion loss of a broadband mixer.
When characterizing a device's conversion loss, the analyzer uses broadband detection to compare the transmitted signal (B*) to the reference signal (R*). This is because the input and output signals of a frequency-translating device may be different. Since broadband detection measures signals at all frequencies, you may want to use a filter to remove unwanted signals such as LO feedthrough when performing this measurement.
For example, an RF signal at 900 MHz mixed with an LO signal at 200 MHz, results in mixing products at 700 MHz and 1100 MHz, as well as the original 900 MHz and 200 MHz RF and LO signals.
Figure 3-21 Filtering Out the Unwanted Mixing Products
3-46 ET User’s Guide
Making Measurements
User Preset
Factory Preset
More
Conversion Loss
Measuring Conversion Loss
Inserting a 700 MHz bandpass filter in the measurement setup removes the unwanted signals at 200 MHz, 900 MHz, and 1100 MHz, providing an accurate measurement of the desired IF signal at 700 MHz.
In the following example, the conversion loss of a mixer will be measured with RF input frequencies over a 15 MHz span centered at 900 MHz. With an LO frequency of 200 MHz, the mixer IF frequency will sweep over a 15 MHz span centered at 700 MHz.

Enter the Measurement Parameters

Press the following keys on the analyzer:
PRESET
( or )
MEAS 1
NOTE This example measurement uses the default instrument parameters for
a conversion loss measurement. If your particular conversion loss measurement requires different parameters (such as source power level, number of data points, and sweeptime), enter them now.

Perform a Normalization Calibration

Normalization is the simplest type of calibration. The analyzer stores a measurement trace into memory and divides subsequent measurements by the stored data to remove unwanted frequency-response errors. This calibration is used for this measurement to remove the insertion loss errors of the IF filter and the interconnect cables. Changing the frequency span or number of measurement points will invalidate a normalization calibration.
Perform the following steps to perform a normalization calibration:
1. Connect the equipment as shown in Figure 3-22.
ET User’s Guide 3-47
Making Measurements
Center
700
MHz
Span
15
MHz
Normalize
Normalize on OFF
Center
MHz
Measuring Conversion Loss
Figure 3-22 Through-Cable Connection
2. Set the following frequency parameters:
FREQ
This sets the analyzer frequency range to sweep over the passband of the IF filter (700 MHz).
3. Press or . This stores the filter response passband into memory, and sets up a
normalized trace so that the filter response magnitude is removed from the measurement.
4. Insert the mixer as shown in Figure 3-23.
5. Press to change the center frequency so that the mixing product of the mixer is in the passband of the IF filter.
3-48 ET User’s Guide
DISPLAY
FREQ
CAL
900
Making Measurements
Measuring Conversion Loss

Connect the DUT

Figure 3-23 Equipment Setup for a Conversion Loss Measurement
ET User’s Guide 3-49
Making Measurements
Autoscale
Marker Search
Max Search
Mkr−> Max
Measuring Conversion Loss

View and Interpret the Conversion Loss Results

1. If necessary to view the measurement trace, press
.
2. To interpret the conversion loss measurement, refer toFigure 3-24 or your analyzer's display if you are making this measurement on your instrument.
a. The values shown on the horizontal axis represent the source RF
output. The values shown on the vertical axis are the power ratio in decibels (dB) of the transmitted signal through the device divided by the incident power. To display the result in logarithmic magnitude format (designated by Log Mag at the top of the measurement screen), the analyzer computes the measurement trace using the following formula:
Conversion Loss dB() 10 log
where P and where P
b. A level of 0 dB would indicate a perfect device (no loss or gain).
Values greater than 0 dB indicate that the mixer has gain. Values less than 0 dB indicate mixer conversion loss.
3. If you wish, you can quickly determine the mixer's minimum conversion loss by pressing
= the power measured at the IF output of the mixer
trans
= the incident power at the RF input.
inc
MARKER
.
=
SCALE
P
trans

------------- -

P
inc
3-50 ET User’s Guide
Figure 3-24 Example of a Conversion Loss Measurement
Making Measurements
Measuring Conversion Loss
NOTE For the measurement to be valid, input signals must fall within the
dynamic range and frequency range of the analyzer. See Chapter 5 for techniques to increase the dynamic range of the analyzer.
ET User’s Guide 3-51
Loading...