The information contained in this document is subject to change without
notice.
Hewlett-Packard makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and
fitness for a particular purpose. Hewlett-Packard shah not be liable for errors
contained herein or for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.
For safety and regulatory information see Chapter 11. For warranty and
assistance information see Chapter 10.
This manual documents analyzers with firmware revisions
C.04.50
and
above. Some features will not be available or will require different keystrokes
in analyzers with earlier hrmware revisions. For full compatibility, you
can upgrade your firmware to the latest version. Contact your nearest
Hewlett-Packard sales or service office for information.
ExcelTM
Lotus@l-2-3@
is a product of Microsoft Corp.
are U.S. registered trademarks of Lotus Development
Corporation.
Microsoft@ is a U.S. registered trademark of Microsoft Corp.
QuickBasicTM
is a product of Microsoft Corp.
Windows@ is a registered trademark of Microsoft Corp.
Portions of the software include source code from the Info-ZIP
code is freely available on the Internet by anonymous ftp
tar.Z,
and from CompuServe
in the
IBMPRO
form, library 10 (data compression).
@Copyright Hewlett-Packard Company 1996, 1997, 1998
All Rights Reserved. Reproduction, adaptation, or translation without prior
written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA
HP
871X
and HP
8714C
RF Network Analyzers
The HP
for production measurements
instrument integrates an RF synthesized source, transmission/reflection test
set, multi-mode receivers, and display in one compact box.
The source features 1 Hz resolution, 50 ms (or faster) sweep time, and up
to + 16
The three-channel, dual mode receivers provide dynamic range of greater
than 100 dB in narrowband measurement mode. For measurements of
frequency-translating devices, the network analyzer features broadband
internal and external detector inputs. The receivers incorporate digital signal
processing and microprocessor control to speed operation and measurement
throughput.
Two independent measurement channels and a large CRT display the
measured results of one or two receiver channels in several user-selectable
formats. An external VGA monitor can be connected to the rear panel for
enhanced measurement viewing in color.
Measurement functions are selected with front panel hardkeys and
menus. Measurements can be printed or plotted directly with a compatible
peripheral. Instrument states can be saved to the internal floppy disk,
internal non-volatile memory, or internal volatile memory. Built-in service
diagnostics are available to simplify troubleshooting procedures.
Measurement calibrations and data averaging provide performance
improvement and flexibility. Measurement calibrations consist of normalizing
data, utilizing the internal factory calibration, or calibrating with external
standards. Measurement calibration reduces errors associated with directivity,
frequency response, and source match. Directivity is corrected to 40 dB and
source match to 30 dB for unproved measurements.
8712C
and HP
dBm
output power.
8714C
are easy-to-use RF network analyzers optimized
,of
reflection and transmission parameters. The
softkey
. . .
111
How to Use This Guide
The first 7 chapters of this guide explain how to perform measurements,
calibrate the instrument, and use the most common instrument functions.
Chapters 8 through 12 are reference material. Use these chapters to look
up information such as front panel features, specific key functions and
specifications.
iv
Contents
1. Installing the Analyzer
Step 1. Check the Shipment
Step 2. Meet Electrical and Environmental Requirements .
Step 3. Check the Analyzer Operation
Step 4. Configure the Analyzer
Connecting Peripherals and Controllers
Installing the Analyzer In a Rack
Preventive Maintenance...............
2. Getting Started
Front Panel Tour
Entering Measurement Parameters
Performing the Operator’s Check
Equipment List ..................
Make a Transmission Measurement
Make a Reflection Measurement
If the Analyzer Fails the Operator’s Check
.............
.........
............
.......
..........
..................
...........
...........
.........
...........
......
l-3
l-4
l-9
l-10
l-11
1-17
1-18
2-3
2-4
2-13
2-14
2-15
2-17
2-19
3. Making Measurements
Measuring Devices with Your Network Analyzer
When to Use Attenuation and Amplification in a
Measurement Setup ...............
When to Change the System Impedance
The Typical Measurement Sequence
Using the BEGIN Key to Make Measurements
lJiZi!i]
Using the
The User BEGIN Function (Option
Measuring Transmission Response
Enter the Measurement Parameters
Calibrate For a Transmission Response Measurement
Connect the DUT .................
View and Interpret the Transmission Measurement
Measuring Reflection Response
Enter the Measurement Parameters
Key Overview
................
Results....................
[m)
....
Key To Configure Measurements
...........
............
.....
.......
.........
.....
lC2
only)
.........
.........
....
3-3
3-9
3-10
3-11
3-12
3-13
3-15
3-17
3-18
3-18
.
3-19
3-21
3-22
3-24
3-24
Contents-l
Calibrate For a Reflection Response Measurement ...
Connect the DUT ...................
View and Interpret the Reflection Measurement Results
Making a Power Measurement using Broadband Detection
Enter the Measurement Parameters.........
Connect the DUT .................
View and Interpret the Power Measurement Results . .
Measuring Conversion Loss ..............
Enter the Measurement Parameters.........
Perform a Normalization Calibration .........
Connect the DUT .................
View and Interpret the Conversion Loss Results ....
Measuring AM Delay (Option 1DA or 1DB)......
Enter the Measurement Parameters.........
Calibrate For an AM Delay Measurement......
Connect the DUT .................
View and Interpret the AM Delay Results......
Making Measurements with the Auxiliary Input .....
Auxiliary Input Characteristics...........
Measuring Group Delay...............
Enter the Measurement Parameters.........
Calibrate For a Transmission Response Measurement
Connect the DUT .................
View and Interpret the Group Delay Measurement
Results....................
Measuring Impedance Using the Smith Chart ......
Enter the Measurement Parameters.........
Calibrate For a Reflection Response Measurement ...
To Avoid Frequency Bandcrossings by Minimizing the
Span (HP 8714C only).............
Increasing Network Analyzer Dynamic Range ......
To Increase the Receiver Input Power........
To Reduce the Receiver Noise Floor .........
Reducing Trace Noise................
To Activate Averaging for Reducing Trace Noise...
To Change System Bandwidth for Reducing Trace Noise
To Eliminate Receiver Spurious Responses......
Reducing Mismatch Errors ..............
Reducing Mismatch Errors in a Reflection Measurement
Reducing Mismatch Errors in a Transmission
Measurement...
Reducing Mismatch. Errors When’
Reflection and Transmission ...........
Compensating for Phase Shift in Measurement Setups . .
Port Extensions ..................
Electrical Delay
Measuring Devices with Long Electrical Delay.....
..................
.
Measuring’Bbth
4-97
4-97
4-98
4-99
4-101
4-102
4-104
5-3
5-3
5-4
5-4
5-5
5-5
5-6
5-7
5-7
5-8
5-9
5-10
5-10
5-11
5-13
5-13
5-14
5-14
5-17
5-17
5-18
...
5-18
5-19
5-19
5-20
5-21
Contents-4
6.
Calibrating for Increased Measurement Accuracy
Measurement Calibration Overview..........
The Calibration Reference Plane..........
Determine if a Calibration is Necessary
When a Calibration Is Not Necessary.........
When a Calibration Is Necessary ...........
Choose an Appropriate Calibration Method .......
To Perform a Normalization Calibration
To Perform a Transmission Calibration ........
To Perform a Reflection Calibration .........
To Perform a Conversion Loss Calibration......
To Perform an AM Delay Calibration (Option
1DB
only)
To Perform a Calibration With Non-Standard
Connectors..................
Writing or Editing Your Own Cal Kit File......
Save the Calibration .................
Check the Calibration................
Using Calibration Check for Analysis and
Troubleshooting
To Perform a Calibration Check ...........
Error Term Descriptions and Typical Values.....
7. Automating Measurements
Configuring Your Test System
Measurement System Topology
Expandability and Large Systems
Throughput Considerations
Selecting a Measurement Controller
Selecting a Programming Language
Operator Interaction
Prompting the Operator
Using Graphics to Create On-Screen Diagrams
User-Defined
Data Entry Using a Barcode Reader
Data Entry Using an External Keyboard
Using the Analyzer’s Title Feature
Hot Keys on External Keyboard For Common Functions
User-Defined TTL Input/Output
Using a Foot Switch or Button Box
Limit Test Pass/Fail TTL Input/Output
Equipment Setup For a Transmission Response Measurement
3-5.
Example of a Transmission Measurement Display
3-6.
Equipment Setup For a Reflection Response Calibration
3-7.
Equipment Setup For a Reflection Measurement of a Two-Port
3-8.
Device
Equipment Setup For a Reflection Measurement of a One-Port
3-9.
Device
Example of a Reflection Measurement Display
3-10.
Equipment Setup For a Power Measurement
3-11.
Example of a Power Measurement
3-12.
Filtering Out the Unwanted Mixing Product
3-13.
Equipment Setup For a Conversion Loss Measurement
3-14.
Example of a Conversion Loss Measurement
3-15.
Equipment Setup For an AM Delay Response Calibration
3-16.
Equipment Setup For an AM Delay Measurement
3-17.
Example of an AM Delay Measurement
3-18.
Equipment Setup For a Group Delay Measurement
3-19.
......................
....................
.....................
Key
.......................
.......................
.................
....................
...................
.................
.............
............
.............
..........
............
............
......
.............
..............
..............
......
.......
........
.............
........
........
......
..........
.....
...
...
. .
.
l-4
l-6
l-7
l-11
1-13
1-18
2-2
2-5
2-9
2-11
2-12
2-15
2-16
2-17
2-18
3-3
3-5
3-7
3-12
3-21
3-23
3-26
3-27
3-27
3-29
3-32
3-34
3-36
3-39
3-41
3-44
3-45
3-47
3-52
Contents-10
Contents
3-20. Example of a Phase-Derived Delay Measurement Display
. .
3-21. Equipment Setup For a Reflection Measurement of a Two-Port
Device
.......................
3-22. Equipment Setup For a Reflection Measurement of a One-Port
Device
3-23. Interpreting the Smith Chart
..:....................
...............
3-24. Determining the Magnitude and Phase of the Reflection
Coefficient
3-25. Example of an Impedance Measurement
3-26. Impedance Calculation for Reflection Measurements
3-27. Impedance Calculation for Transmission Measurements
4-l. The
(jMARKER)
.....................
4-2. Connect the Filter to the Analyzer
4-3. Markers at Minimum and
4-4. Peak and Minimum Search Criteria
4-5. Peak and Minimum Search Criteria at Display Endpoints
4-6. -6 dB Bandwidth Marker Search
4-7. -6 dB Notch Marker Search
4-8. Peak and Minimum Search Criteria
4-9. Peak and Minimum Search Criteria at Display Endpoints
4-10. Multi-Peak Search Mode
4-
11. Multi-Notch Search Mode
4-
12. Marker Statistics Function
4-13. Marker Flatness Function
4-14. RF Filter Statistics Function
4-15. Delta Marker Mode
4-16. Limit Lines
4-17. Limit Lines Example 1
4-18. Limit Lines Example 2
4-
19. Reference Positions
4-20. Split Display
4-2 1. Display Features
4-22. The Display Annotation
4-23. Normal Display
4-24. Expanded Display
4-25. Peripheral Connections
4-26. Hardcopy Components and Formats Available
4-27. Trace List Values
......................
..........
.....
...
Key
............
Maximum
Values
.........
............
...
.............
...............
............
...
.................
.................
................
................
...............
...................
.......................
..................
..................
...................
......................
....................
.................
.....................
....................
..................
.......
....................
5-l. Relationship Between Frequency Span, Sweep Time, and
Number of Points
5-2. Compensating for Test Fixture Delay
..................
...........
3-54
3-57
3-57
3-59
3-60
3-61
3-63
3-64
4-3
4-4
4-9
4-10
4-11
4-14
4-16
4-17
4-18
4-19
4-20
4-22
4-24
4-26
4-28
4-36
4-48
4-49
4-51
4-54
4-55
4-58
4-63
4-64
4-81
4-93
4-94
5-6
5-19
Contents- 11
Contents
6-l. Sources of Errors
6-2. Mismatch Errors
....................
......
6-3. The Calibration Reference Plane ’ 1 1 1 : 1 1 1 1 1 1 1 1
6-4. Typical Directivity Error Term
6-5. Typical Source Match (Corrected) Error Term
6-6. Typical Source Match (Uncorrected) Error
6-7. Typical Load Match Error
7-3. Network Analyzer Without BASIC, Controlled by a Computer
7-4. Network Analyzer Running
7-5. Example
Test
System Setup
7-6. Connect a Switch to the USER TTL IN/OUT Connector
7-7. Connect a Switch to the USER TTL IN/OUT Connector
7-8. Measurement Control
7-9. Writing to the Parallel Port
7-10. Digital Latch Circuit
7-
11. Customized Annotation
7-12. Paper Numbering
....................
8-l. Analyzer Connectors - Front Panel
8-2. Analyzer Connectors - Rear Panel
8-3.
HP-lB
Connector and Cable
8-4. Parallel Port Pm-outs
8-5.
RS-232
Connector Pin-out
8-6. VIDEO OUT Connector Pin-out
BASIC,
Controlled by a Computer
...............
...
...
...................
................
...................
.................
............
.............
..............’..
...................
................
..............
8-7. Probe Power Connector
8-8. The Analyzer Line Power Switch’ : : 1. : : : : : : : : :
8-9. Display Intensity Control
8-10. Disk Drive
.......................
8-l 1. Power Cable and Line (Mains) Plug Part Numbers
8-12. Location of Line Fuses
8-13. Voltage Selector Switch Location
10-l.
Receiver Dynamic Accuracy (narrowband)
10-2. Absolute Power Accuracy (broadband)
.................
......
..................
.............
.........
...........
1
:
6-3
6-4
6-6
6-30
6-31
6-32
6-33
6-34
6-35
6-36
7-6
7-8
7-10
7-11
7-19
7-28
7-43
7-44
7-57
7-58
7-62
7-78
8-3
8-4
8-7
8-9
8-10
8-11
8-12
8-17
8-19
8-20
8-22
8-23
8-24
10-9
10-10
Contents-12
lhbles
l-l. Maximum HP-IB Cable Lengths
3-l. Measurement Configurations from the
4-l.
Disk Access
4-2. Typical Print Times
5-l. Relationship Between System Bandwidth and Sweep Speed
6-l. Calibration Types
6-2. Calibration Check Error
7-l. Keyboard
7-2. Writeable Ports
7-3. Readable Ports
7-4. Writeable Ports
7-5. Readable Ports
7-6. Parallel Port Pins
8-1. General Bus Management Lines
8-2. VGA Compatible Monitor Characteristics
10-l.
Hewlett-Packard Safes and Service Offices
12-
1. Disk Capacities
12-2. Maximum Number of Files and Directories
12-3. Sizes of Instrument State Components
.......................
...................
....................
Terms
Template
Definition
.....................
.....................
.....................
.....................
....................
.....................
..............
(BEGIN)
..............
...............
..............
Key
......
..........
.........
.........
...........
.
12-14
12-15
1-14
3-16
4-74
4-96
5-4
6-9
6-27
7-33
7-39
7-40
7-55
7-55
7-57
8-8
8-11
lo-24
12-18
Contents-13
1
-
Installing the Analyzer
Installing the Analyzer
This chapter will guide you through the four steps needed to correctly and
safely install your network analyzer. The four steps are:
1. Check the Shipment
2. Meet Electrical and Environmental Requirements
3. Check the Analyzer Operation
4. Configure the Analyzer
l-2
Step 1. Check the Shipment
After you have unpacked your instrument, it is recommended that you keep
the packaging materials so they may be used if your instrument should need
to be returned for maintenance or repair.
Check the items received against the Product Checklist (included in your
shipment) to make sure that you received everything.
Inspect the analyzer and all accessories for any signs of damage that may
have occurred during shipment. If your analyzer or any accessories appear
to be damaged or missing, call your nearest Hewlett-Packard Sales or Service
office. Refer to
‘fable 10-l
in Chapter 10 for the nearest office.
l-3
Step 2. Meet Electrical and Environmental
Requirements
1. Set the line voltage selector to the position that corresponds to the ac
power source you will be using.
CAUTION
Before switching on this instrument, make sure that the line voltage selector
switch is set to the voltage of the power supply and the correct fuse
(T 5A 250 V) is installed. Assure the supply voltage is in the
NOTE
The working fuse and a spare are located in the power cable receptacle. See figure 8-12.
I
specsed
range.
I
POWER CAB:E
RECEPTACLE
Figure l-l. Voltage Selector Switch location
l-4
VOLTAGE
SELECTOR
SWITCH
PP~~C
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
2. Ensure the available ac power source meets the following requirements:
115
v
90 to 132 Vat
I
230
V1 198 to 254 Vat
147
to 66 Hzl
147
to 66 Hzl
I
If the ac line voltage does not fall within these ranges, an autotransformer
that provides third wire continuity to ground should be used.
3. Ensure the operating environment meets the following requirements for
safety:
l
indoor use
l
altitude up to 15,000 feet (4,572 meters)
0
temperature 0 “C to 55
OC
CAUTION
l
maximum relative humidity 5 to 95 percent relative at +40
“C
(non-condensing)
l
this product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2
This product is designed for use in Installation Category II and Pollution
Degree 2.
NOTE
The above requirements are for safety only Separate conditions that must be met for specified
performance are noted in Chapter 10.
l-5
I
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
I
-
4. Verify that the power cable is not damaged, and that the power source
outlet provides a protective earth ground contact. Note that the following
illustration depicts only one type of power source outlet. Refer to
Figure 8- 11 to see the different types of power cord plugs that can be used
with your analyzer.
PROTECTIVE
EARTH GROUND
WARNING
Figure
l-2.
Protective Earth Ground
This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. Any
interruption of the protective conductor, inside or outside the instrument,
is likely to make the product dangerous. Intentional interruption is
prohibited.
l-6
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
WARNING
If this instrument is to be energized via an external autotransformer for
voltage reduction, make sure that its common terminal is connected to a
neutral (earthed pole) of the power supply.
5. Install the analyzer so that the detachable power cord is readily
identifiable
and is easily reached by the operator. The detachable power
cord is the instrument disconnecting device. It disconnects the mains
circuits from the mains supply before other parts of the instrument.
The front panel switch is only a standby switch and not a LINE switch.
Alternatively, an externally installed switch or circuit breaker (which is
readily identifiable and is easily reached by the operator) may be used as a
disconnecting device.
6. Ensure there are at least two inches of clearance around the sides and
back of either the stand-alone analyzer or the system cabinet.
TWO INCH CLEARANCE
SIDES AND
REAR
RACK
TWO INCH CLEARANCE
SIDES
AND
Rgp
po64b
/
EDGE OF BENCH
Figure 1.3. Ventilation Clearance Requirements
l-7
Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
7. Set up a static-safe workstation. Electrostatic discharge (ESD) can damage
or destroy components.
t MegOhm
Resistor
l table mat with earth ground wire:
HP part number 9300-0797
l wrist-strap cord with 1 Meg Ohm resistor:
HP part number
0
wrist-strap:
HP part number 9300-1367
l heal straps:
HP part number 9300-1308
0
floor mat:
part number
9300-0960
1664R
l-8
qg653d
Step 3. Check the Analyzer Operation
1. Turn on the line switch of the analyzer. After approximately 30 seconds, a
display box should appear on the screen with the following information:
l
The model number of your analyzer (either HP
l
The firmware revision
l
The serial number of your analyzer
l
Installed options
2. Verify that the serial number and options displayed on the screen match
the information on the rear panel serial label.
3. The operator’s check should be performed on the analyzer to provide a
high degree of confidence that the analyzer is working properly. Refer to
Chapter 2 for instructions on how to perform the operator’s check.
8712C
or HP
8714C)
l-9
Step 4. Configure the Analyzer
You can begin making measurements by simply connecting your analyzer
to an appropriate power source and turning it on. This section, however,
will explain how to connect common peripherals and controllers, and how to
install your analyzer into a rack system.
l-10
Step 4. Configure the Analyzer
Connecting Peripherals and Controllers
CONTROL PORTS
A
f
LAN
ETHERTWIST
SERIALCOLOR VGA
7
Installing the Analyzer
VIDEO OUT
EXiERNAL
DETECTORS
Figure
HP- I B
L I NE
l-4.
Analyzer Rear Panel line Module and Selected Connectors
VOLiAGE
PARALLEL
VOiTAGE
SELECTOR
SWITCH
KEYBOARD
PP~~C
Refer to Figure
l
The HP-IB port is for use with computers and peripherals (printers,
l-4:
plotters, etc.)
l
The parallel and
lC2
the parallel and serial ports can also be programmed via
RS-232
(serial) ports are
also
for peripherals. With Option
IBASIC
for
l-11
Installing the Analyzer
Step 4. Configure the Analyzer
general I/O control. See the BASICUser’s Handbook for information on
using
BASIC.
l
The VIDEO OUT COLOR VGA port allows you to connect a color VGA
monitor for enhanced viewing. See “Using an External VGA Monitor” in
Chapter 4 for more information.
l
The LAN ETHERTWIST connector is for connecting your analyzer to a
LAN (local area network) for control and access. You must have Option
lF7 to utilize this port. See the Option
lF7
User’s Guide Supplement for
information on how to use your analyzer in a LAN.
1-12
Installing the Analyzer
Step 4. Configure the Analyzer
HP-IB Connections
An HP-IB system may be connected in any
conEguration
as long as the
following rules are observed:
l
The total number of devices is less than or equal to 15.
l
The total length of all the cables used is less than or equal to 2 meters
times the number of devices connected together up to an absolute
maximum of 20 meters. For example, the maximum cable length is
4 meters if only 2 devices are involved. The length between adjacent
devices is not critical as long as the overall restriction is met.
See Figure l-5 for different connection configurations.
STARLINEAR
NETWORK
ANALYZER
NETWORK
ANALYZER
Figure
l-5.
HP-IB Connection Configurations
1-13
Installing the Analyzer
Step 4. Configure the Analyzer
Table l-l. Maximum HP-IB Cable lengths
Parallel and Serial
Connections
Other Connections
InstrumentslPeriphvrals
in SystamBatwrrn Each Pair
L,,.,
I
Maximum HP-IB
of Dwicas
“,“.
ltotall
Cable Length
I
Parallel and serial devices often require specific cables-check their manuals
for details. Parallel cable length should not exceed 25 feet. The analyzer may
experience problems talking to a printer if this length is exceeded. Connect
the required control cables and secure them. (Tighten the knurled screws or
comparable fasteners.)
If you plan to use a keyboard, external video monitor, or external detectors,
connect them to the appropriate rear panel connectors. See Figure 1-4. Also
see Chapter 8 for more information on front and rear panel connectors.
1-14
Installing the Analyzer
Step 4. Configure the Analyzer
To Set HP-IB Addresses
lb communicate via HP-& each external device must have a unique address
and the network analyzer must recognize each address. ‘lb check or set each
external device’s actual address, refer to the device’s manual (most addresses
are set with switches).
The following are examples of how to check or set the device’s recognized
address on the network analyzer:
l Printer: press
to highlight the line that reads HP Printer PCL HP-IB. Press
(j-1
S~t&$ ?Zupy
Port . Use the front panel knob
S&&t
.
The second line of the screen displays settings: in this case the address.
The default address is 5, however most printers are factory set to address
1 (one). ‘lb change the recognized address, press
[ptYiGJ Erlter
l Plotter: press
.
[HARDCOPYI) Seloc% CuPy
Port . Use the front panel knob
to highlight the line that reads HP Plotter HPGL HP-IB. Press
PrinWr
HP-SBAzldr
Select
.
The second line of the screen displays settings: in this case the address.
The default address is 5 and most plotters are factory set to address 5
(Eve), so changing the address is probably not necessary. ‘lb change the
recognized address, press
PMnt@r B&LB
A&lr (number)
Entsz:
.
NOTE
Only one hardcopy HP-IB address can be set at a time. Changing the printer address, for example,
changes the plotter to the same address.
l HP
8712C
or HP
8714C:
XP 872X &d&&s& OT
press (SYSTEM OPTIONS)
BP
S”ft$\C Addrsls&
HP-II3
. The network analyzer’s
address will appear (the default is 16). To change the address, press
(iGzq Enter.
1-15
Installing the Analyzer
Step 4. Configure the Analyzer
To Configure Peripheral
Settings
If your system uses serial or parallel peripherals, follow the guidelines below
to configure the system. Refer to the peripheral’s manual for correct cables
and settings. The parallel and serial ports have standard Centronics
and
RS232
pinouts respectively as explained in Chapter 8, “Front/Rear
DB-25
Panel.’
l Serial Devices: press
(m)
Select
Copy
Port
, use the entry
controls to highlight your type of printer or plotter and press Select . If
the baud rate or handshake at the top of the screen are incorrect, use the
softkeys
l Parallel Devices: press
to change them.
(m)
Sefect CuPy Port
, use the entry
controls to highlight your type of printer or plotter and press Select .
l
LAN Printer (Option
use the entry controls to highlight HP LaserJet
lF7
only): press [HARDCOPY)
S@X%ct
Copy Port ,
PCL5/6 PCL5
LAN,
and press S&,o~t . If the printer IP address at the top of the screen is
incorrect, press
NOTE
When
S+l+clt &yy
of the display screen show the current settings for your convenience.
X,AN PWrtsr IP addr
Port is selected, the first two lines in the box that appears at the top
to enter the correct IP address.
NOTE
Use a
information.
1-16
PCL5
printer for fastest hardcopies. See “Configure the Hardcopy Port” in Chapter 4 for more
Installing the Analyzer In a Rack
Installing the Analyzer
Step 4. Configure the Analyzer
CAUTION
CAUTION
Use only the recommended rack mount kit (Option
1CM
when ordered with
the analyzer or HP part number 08712-60036 when ordered separately) with
this instrument; it needs side support rails. Do not attempt to mount it by
the front panel (handles) only. This rack mount kit allows you to mount the
analyzer with or without handles.
lb install the network analyzer in an HP
85043D
rack, follow the instructions
in the rack manual.
lb install the network analyzer in other racks, note that they may promote
shock hazards, overheating, dust contamination, and inferior system
performance. Consult your HP customer engineer about installation,
warranty, and support details.
When installing the product in a cabinet, the convection into and out of
the instrument must not be restricted. The ambient temperature (outside
the cabinet) must be less than the maximum operating temperature of the
instrument by 4 OC for every 100 watts dissipated in the cabinet. If the
total power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
Place other system instruments (computer, printer, plotter) where convenient,
within the HP-IB cable length limits (see
‘Ihble
l-l) or other interface cabling
limits.
1-17
Preventive Maintenance
Preventive maintenance consists of two tasks. It should be performed at least
every six months-more often if the instrument is used daily on a production
line or in a harsh environment.
Clean the CRT
Check the RF Front
Panel Connectors
Use a soft cloth and, if necessary, a mild cleaning solution.
Visually inspect the front panel connectors. The most important connectors
are those to which the DUT is connected, typically the RF cable end or
the RF IN connector. All connectors should be clean and the center pins
centered. The
tigers
of female connectors should be unbroken and uniform
in appearance. If you are unsure whether the connectors are good, gauge the
RF IN and RF OUT connectors to
W4 b-c
Figure 1-6. Maximum and Minimum Protrusion of Center Conductor From Mating Plane
confirm
that their dimensions are correct.
Mating plane
Max = 0.207 in.
Min =
0.204
in.
l-18
2
Getting Started
Getting Started
The HP
8712C
and HP
8714C
are easy-to-use, fully integrated RF component
test systems. Each instrument includes a synthesized source, a wide dynamic
range receiver and a built-in test set. Controls are grouped by functional
block, and settings are displayed on the instrument CRT. This section
familiarizes new users with the layout of the front panel and the process of
entering measurement parameters into the analyzer.
2-2
Dob48b
Figure 2-1. Network Analyzer Front Panel Features
Front Panel
I
The CRT DisplayThe analyzer’s large
e
(m]
lbur
programming coda, softkay menus and measurement parameters quickly and clearly. Refer to
“Display’ in Chapter 8 for more information.
The
(w)
selection of basic measurement parameters for a user-specified class of devices
amplifiers, or
Filter as Your device type puts the analyrer into narrowband detection mode,
measurement dynamic range. In comparison, selecting Mixer as Your device type puts the
analyzer into broadband detection mode, enabling frequency translation measurements. This
capability allows new users to start making measurements with as few as four keystrokes.
key simpliiias measurement setups. The begin key allows quick and easy
For example, when making a transmission measurement, selecting
IIBASICI
[e.g.,
filters,
maximizin!
3
MEAS
4
SOURCE
6
CONFIGUREThe configure keys control receiver and display parameters. These parameters include
B SYSTEMThe system keys control system level functions. These include instrument preset, save/recall,
7
The Numeric KeypadUse the number keys to enter a specific numeric value for a chosen parameter. Use the
~I(JJARDKEYS)
3
Softkep;e
The measure keys select the measurements for each channel. The analyzer’s measurement
capabilities include transmission, reflection, power, conversion loss, and AM delay
IDA and
The source keys select the desired source output signal to the device under test, for
example, selecting source frequency range or output power. The source keys also control
sweep time, number of points, and sweep triggering.
receiver bandwidth and averaging, display scaling and format, marker functions, and
instrument calibration.
and hardcopy output.
keys.
[ENTER)
You can also use the front panel knob for making continuous adjustments to parameter
values, while the m and 0 keys allow You to change values in steps.
Hardkays are front panel keys physically located on the instrument front panel. In text,
these keys will be represented by the key name with a box around it
such as:
Softkeys are keys whose labels are determined by the analyzer’s firmware. The labels are
displayed on the screen next to the 8 blank keys next to the display screen on the analyzer.
In text, these keys will be represented by the key name with shading behind it such es:
Snsep
IDB onlyl.
HP-18
parameters end
key or the softkays to terminate the numeric entry with the appropriate units.
L-j.
IBASIC
are also controlled with these system
Tim.
IOptions
2-3
Entering Measurement Parameters
This section describes how to input measurement parameter information into
the network analyzer.
NOTE
When entering parameters, you can use the numeric key pad, as described in each example, or you
can use the m
NOTE
(IJ
keys or the front panel knob to enter data.
When you are instructed to enter numeric values in this manual, it often can get cluttered and
confusing to depict each key stroke. So in this manual, numbers
are depicted inside one keycap. For example, if you are instructed to enter the number -42.5, it will
be depicted inside one keycap like this:
pressed in succession: a @ @ fJ
(--42.5).
0.
To enter this number, the following
(no
matter how many characters)
keys
need to
You can follow along with these examples by connecting the filter and cable
that were supplied with your instrument as shown in Figure 2-2.
2-4
Getting Started
Entering Measurement Parameters
NETWORK
Figure 2-2. Connect the Filter to the Analyzer
ANALYZER
2-5
Getting Started
Entering Measurement Parameters
Presetting the Analyzer Press the
it reverts to a known operating condition. When this key is pressed, the
following major default conditions apply:
(FEEEQ
key. When the analyzer is preset with the
Frequency
Frequency range*0.3 to 3000 MHz
Power level30
Measurement Channel 1
measurement
Measurement Channel 2
measurement
Format
Number of points
Sweep time
Scale
Reference0
System Bandwidth
1 HP
2 HP
3 Preset power level can be set to other then 0
rangel
87121:
only
8714C
only
Source Power Level,’ later in this chapter for more information.
0.3 to 1300 MHz
dBm
Transmission
Off
leg
Magnitude
201
Auto
10 dB/div
dB
Medium Wide
dBm
if desired. See ‘Entering
@SiZi]
key,
NOTE
The measurement parameters that you enter will be retained in the analyzer’s memory when the
power is turned off, and will be restored when the power is turned back on.
See Also
Refer to Chapter 12, “Preset State and Memory Allocation,” for a comprehensive table of preset
conditions.
2-6
Getting Started
Entering Measurement Parameters
Entering Frequency
Range
1. Press the
2.
lb change the low end of the frequency range to 10 MHz, press
LFREQ)
key to access the frequency
softkey
menu.
S~;W&
Ilo]m.
3.
lb change the high end of the frequency range to 900 MHz, press
Stop
(EiJMHz.
4+
You can also set the frequency range by using the
CerSer
and Span
softkeys. For instance if you set the center frequency to 160 MHz and the
span to 300 MHz, the resulting frequency range would be 10 to 310 MHz.
NOTE
When entering frequencies, be sure to terminate Your numeric entry with the appropriate softkey to
obtain the correct units. If you use the
default to Hz.
The default displayed frequency resolution is kHz. You can change the resolution by pressing
Disp
Fmq
Re&oLutictn
CENTER)
, and then selecting a new resolution.
key to terminate a frequency entry, the units
[FREQI
2-7
Getting Started
Entering Measurement Parameters
Entering Source Power
Level
1. Press the
2.
lb change the power level to 3
CPOWER)
key to access the power level softkey menu.
dBm,
press L&e1 0, and
@YU
(ENTERI).
3.
‘lb change the power level to - 1.6
dE!m,
press
LlsoeI (--1.6) #$!!I
piE-).
NOTE
Your analyzer’s power level (depending upon its option configuration) may not be
0
dam.
4.
‘lb
change the power level that will always be restored when you preset
the analyzer, press
PXYMSIZ Pwr
Level
12.51
and
dEsm
settable
or
(ENTER).
to below
entry does not affect the current power level.
or
or
This
2-8
Getting Started
Entering Measurement Parameters
Scaling the
Measurement Trace
1. Press the
~ZKCEJ
key to access the scale menu.
2. lb view the complete measurement trace on the display, press
Autoscals
3.
lb change the scale per division to 5
.
dB/division
press
Scals/Ri~ Q
Enter.
4. ‘RI move the reference position (indicated by the ) symbol on the left side
of the display) to the first division down from the top of the display, press
Refex*ace Pusitioa @
Enter.
Figure 2-3 shows how each reference
position is identified.
5. To
change the reference level to 0
l&rfz%X
.
10
9
8
7
6
5
4
3
2
1
0
dE%,
press
R~&XE%WXI L%v&t
po’352b-c
@
Figure 2-3. Reference Positions
2-9
Getting Started
Entering Measurement Parameters
Entering the Active
Measurement Channel
and Type of
Measurement
The [MEASj and
(jMEAs]
keys allow you to choose which measurement
channel is active, and measurement parameters for that channel. When a
particular measurement channel is active, its display is brighter than the
inactive channel, and any changes made to measurement parameters will
affect only the active measurement channel. (Some measurement parameters
cannot be independently set on each measurement channel. For these
parameters, both channels will be affected regardless of active channel
status.)
1. ‘lb measure transmission on measurement channel 1 and reflection on
measurement channel 2, press the following keys:
2. Both channels’ measurements are now visible on the analyzer’s display
screen. Note that the active measurement channel’s (channel 2)
measurement trace is brighter than the other measurement channel’s
trace.
2-10
Getting Started
Entering Measurement Parameters
Dl:
Transmlsslon
Dl:
Transmlsslon
>2:
Reflection
>2:
Reflection
-30ux
-30
-40-40
AbsAbs
Start 0.300 MHz
Log MagLog Mag10.0
Log MagLog Mag10.0
dB/
dB/
dB/
dB/
Ref10.0
Ref
Ref10.0
Ref
Stoo
3 000.000 MHz
Figure 2-4. Both Measurement Channels Active
0.00
0.00
0.00
0.00
dB
dB
dB
dB
44
1
2-11
Getting Started
Entering Measurement Parameters
Viewing Measurement
Channels
1. ‘lb view only the measurement channel 2 reflection measurement press
(MEAS- Maa&
2. lb view both measurement channels again, press
OFF.
[MEAsJ
3. lb view both measurement channels separately on a split screen, press
[D’SPLAV) Mtors Display Split Diep FULL spltt .
)I: Transmission
IIII
dB
30
2
1
-1
-2
-3
Abs
Start 0.300
D2:
dB
30
20
10
I
-10
-20
-30
Abs
Start 0.300 MHzstop 3 000.000 MHZ
II
I
I
!I
ReflectionLog Mag10.0
I
II
MHz
Log Mag10.0
I
II
I
I
III
Figure 2-5. Split Display
dB/
Ref
IIII
II
stop3000.000
dB/
Ref
0.00
I
I
0.00
dB
dB
I
I
_
I
MHZ1
You have now learned how to enter common measurement parameters and
how to manipulate the display for optimum viewing of your measurement.
You can now proceed on to performing the operator’s check, or refer to
Chapter 3 for detailed information on making specific types of measurements.
2-12
Performing the Operator’s Check
The operator’s check should be performed when you receive your instrument,
and any tune you wish to have confidence that the analyzer is working
properly. The operator’s check does not verify performance to speciiications,
but should give you a high degree of confidence that the instrument is
performing properly if it passes.
The operator’s check consists of making a transmission measurement with the
cable that was supplied with your analyzer, and a reflection measurement
with the cable and again with a 50 fl or 75 tI termination (load).
2-13
Getting Started
Performing the Operator’s Check
Equipment List
‘Ib
perform the operator’s check you will need the following:
l
A known good cable such as the one that was supplied with your analyzer.
The cable you use should have
and 50.75 dB of insertion loss from 1.3 to 3.0
l
A known good load (> 40 dB return loss) that matches the test
port impedance of your analyzer such as one from calibration kit
HP
85032BE
(50 Q) or HP
SO.5 dB
85036B/E
of insertion loss up to 1.3
GHz.
(75 0).
GHz
2-14
Getting Started
Performing the Operator’s Check
Make a Transmission Measurement
1. Connect the equipment as shown in Figure 2-6. Use a known good cable
such as the one that was supplied with your analyzer.
NETWORK ANALYZER
RF OUT
paSl,b
2.6. Equipment Setup for Transmission Measurement
Figure
2.
Press
3. Press
4. Verify that the data trace falls within
(j-1
m(TiJ
typical result.
[ZKF) .i Eiter
@iYET].
.
3~0.5 dB
U
of 0
RF IN
dB.
See Figure 2-7 for a
2-15
Getting Started
Performing the Operator’s Check
h: Transmission
b2: Off
I
+
0.5
L
Start 0.300 MHz
Figure 2-7. Verify Transmission Measurement
Log Mag
0.i dB/ Aef
stop 3 000.0~
0.00
dS
00 MH
po653b-c
7
NOTE
The quality of the cable will affect this measurement; make sure you use a cable with the
characteristics described in “Equipment List.”
I
2-16
I
Make a Reflection Measurement
1.
Leave the cable connected to the analyzer.
2.
Press
@EXi1)
Reflection
(?Ei@ Ilo]
Enter .
Getting Started
Performing the Operator’s Check
Verify that the data trace falls completely below -16
3.
a typical result.
Fl:
dB
Dz:
40
Reflection
Off
Log Mag
Results
must fall
below
this line
-16
dB
Chl
-30
-40
Abs
Start 0.300 MHz
Figure 2-8. Verify Reflection Measurement
dB.
See Figure 2-8 for
10.0
dB/
Ref 0.00
Stop 3 000.000 MHz
po654bhc
dB
4. Disconnect the cable and connect a known good load to the RF OUT port
as shown in Figure 2-9.
2-17
Getting Started
Performing the Operator’s Check
NETWORK ANALYZER
LOAD
Figure 2-9. Connect the load
5. Verify that the data trace falls below -30
screen, press
[EiKF]
R~~Yxw~
L%xr&
moves up onto the screen.
2-18
dB.
If the data trace is off the
and the
(IJ
key until the trace
Getting Started
If the Analyzer F’ails the Operator’s Check
First, repeat the operator’s check using a different cable and load to eliminate
these as a possible cause of failure.
If your analyzer does not meet the criteria in the operator’s check, your
analyzer may need adjustment or servicing. Have a qualified service
technician check the instrument or contact any Hewlett-Packard Sales or
Service Office for assistance. Refer to %ble
office.
10-l
in Chapter 10 for the nearest
2-19
3
Making Measurements
Making Measurements
This chapter provides an overview of basic network analyzer measurement
theory, a section explaining the typical measurement sequence, a segment
describing the use of the
measurements:
l
Measuring Transmission Response
l
Measuring Reflection Response
l
Making a Power Measurement using Broadband Detection
l
Measuring Conversion Loss
l
Measuring AM Delay (Option
l
Making Measurements with the Auxiliary Input
l
Measuring Group Delay
l
Measuring Impedance Using the Smith Chart
l
Measuring Impedance Magnitude
[BEGIN]
key, and detailed examples of the following
1DA
or
1DB)
3-2
Measuring Devices with Your Network
Analyzer
This section provides a basic overview of how the network analyzer measures
devices. The analyzer has an RF signal source that produces an incident
signal that is used as a stimulus to the device under test. Your device
responds by reflecting a portion of the incident signal and transmitting (or
perhaps altering and transmitting) the remaining signal. Figure 3-l shows
how a device under test (DUT) responds to an RF source stimulus.
Source
INCIDENT
TRANSMITTED
Figure 3.1. DUT Response to an RF Signal
3-3
Making Measurements
Measuring Devices with Your Network Analyzer
Refer to Figure 3-2 for the following discussion regarding detection schemes
and modes. The transmitted signal (routed to input B) and the reflected
signal (input A) are measured by comparison to the incident signal. The
network analyzer couples off a
small
portion of the incident signal to use as
a reference signal (routed to input R). The network analyzer sweeps the
source frequencies, resulting in a measured and displayed response of your
test device. Figure 3-2 shows the transmitted, reflected, and reference signal
inputs.
3-4
R-
A------
v
Making Measurements
Measuring Devices with Your Network Analyzer
\i
Figure 3-2. Simplified Block Diagram
B
i on
pp615c
Making Measurements
Measuring Devices with Your Network Analyzer
Refer to Figure 3-3 for the following discussion. The network analyzer
receiver has two signal detection modes:
l
broadband detection mode
l
narrowband detection mode
There are two internal broadband detector inputs: B* and
R*.
External
broadband detectors can also be used when connected to the X and Y
ports on the rear panel of the analyzer. When the network analyzer is in
the broadband detection mode, it measures the total power of all signals
present at these measurement ports, independent of signal frequency. This
enables the characterization of frequency translation devices such as mixers,
receivers, and tuners, where the RF input and output frequencies are not the
same. Figure 3-3 labels the transmitted signal for broadband detection input
as
B*,
and the reference signal as R*.
When the network analyzer is in the narrowband detection mode, the
receiver is tuned to the source frequency. This technique provides greater
dynamic range by decreasing the receiver’s bandwidth. Figure 3-3 shows the
transmitted signal for narrowband detection input as B, and the reference
signal as R.
3-6
REAR PANEL
External Detectors
AUX Input
Y-‘
Input A
Reflected;
Making Measurements
Measuring Devices with Your Network Analyzer
Y
Y
RF
SOUrCe
I
I
I
FRONT PANEL
Reflection
Figure 3-3. Block Diagram
ss
ion
3-7
Making Measurements
Measuring Devices with Your Network Analyzer
The following table shows the correlation between different types of
measurements, input channels and signals.
When to Use Attenuation and Amplification
in a Measurement Setup
l
For accurate measurements, use external attenuation to limit the power at
the RF IN port to + 10
(for broadband measurements).
dBm
(for narrowband measurements) or + 16
dBm
CAUTION
When to Use
Amplification
Always
device’s output power exceeds the receiver damage limit of +23
f25
l
use attenuation on the TRANSMISSION RF IN port if your test
dBm
or
Vdc.
Use attenuation on the RF IN port to reduce mismatch errors. See
“Reducing Mismatch Errors” in Chapter 5 for more information.
l
In an AM delay measurement (Options
directly
the - 10
before
the DUT if the device’s input power must be less than
dBm
minimum
specihed
1DA
and
1DB
only), use attenuation
detector level. (If you reduce the input
power to the DUT by lowering the analyzer’s source power, the reference
detector, X, will be below its specified range.)
Use attenuation directly after the DUT if the device’s output power is
greater than the + 13
l
For accurate measurements, ampltication may be needed on the analyzer’s
dBm
maximum
speciiied
detector level.
RF OUT port. Use amplification when your test device requires input
power that exceeds the analyzer’s maximum specified output power.
The maximum specified output power is highly dependent upon the
model and option configuration of your analyzer as well as the frequency
range of your test setup. It ranges from +4 to + 16
dBm.
See “Source
Specifications” in Chapter 10 to determine the maximum spectied output
power of your analyzer.
l
In an AM delay measurement (Options
amplification
directly before the DUT if the device’s input power must be
1DA
and
1DB
only), use
greater than the maximum power available at the power splitter output.
Use
ampliiication
than the -10
directly after the DUT it the device’s output power is less
dBm
minimum
speciiied
level needed by the test detector
(Y).
3-9
Making Measurements
Measuring Devices with Your Network Analyzer
When to Change the System Impedance
Analyzers with a system characteristic impedance of 50 or 75 ohms, may
be switched to the alternate impedance. If using minimum loss pads for
impedance conversions, the alternate impedance should be selected so that
the measurement results are displayed relative to the conversion impedance.
For example, if you have a 50 ohm instrument and are making 75 ohm
measurements, you may be using a 50 to 75 ohm minimum loss pad.
Measurement results can be reported relative to 75 ohms, not 50 ohms, if the
alternate system impedance is selected. This may include marker readouts,
Smith chart results, or SRL impedance computations (Option 100).
lb
change the system impedance, press the following keys on the analyzer:
(CAL)
-
The built-in cal kit selections will be converted to the selected system
impedance.
3-10
Making Measurements
Measuring Devices with Your Network Analyzer
The Typical Measurement Sequence
A typical measurement consists of performing four major steps:
Step 1. Enter the
Measurement
Parameters
Step 2. Calibrate the
Analyzer
Step 3. Connect the
Equipment
Step 4. View and
Interpret the
Measurement
The easiest way to set up the analyzer’s parameters for a simple
measurement is to use the (BEGIN) key. (See “Using the BEGIN Key to Make
Measurements,” next, in this chapter.)
For some measurements you may wish to enter your own specific
measurement parameters. Use the instrument’s keys to input your
parameters.
This step may be omitted under certain conditions. Your analyzer can
provide highly accurate measurements without performing any additional
user-calibrations if certain conditions are met. Chapter 6 explains when
additional calibration is necessary.
Connect the DUT and any other required test equipment. See the
measurement examples later in this chapter for typical equipment setup
conhgurations.
Use the (SCALE), [DISPLAY), and
[FORMAT)
functions to optimize viewing of the
measurement results.
Markers, limit lines, and hard copies of the display are common means of
interpreting measurement results.
See Chapter 4 for detailed information on using instrument functions to view
and interpret your measurements.
3-11
Using the BEGIN Key to Make Measurements
BEGIN
I
I
NETWORK ANALYZER
W~lOb
Figure 3-4. The
The
@EEi@
the
@Zi?ZF]
0 ampliGers
0
filters
l
broadband passive devices
0
mixers
l
cables (Option 100 only)
Conf@uring
correct instrument set up. The analyzer guides you through the initial steps
and
conf@res
3-12
key allows you to quickly and easily configure the analyzer (from
condition) to measure the following devices:
basic measurements from the
itself for the device type you select.
(EiZiK) Key
(El
key helps you ensure
Making Measurements
Using the BEGIN Kay to Make Measurements
(BEGIN) Key Overview
The
@E%i]
types of devices.
The (BEGIN) key has two different behaviors, depending on whether you are
selecting a new device type, or a new meusurementtype.
key sets up a generic instrument state for the testing of various
Selecting a New Device
Selecting a New
Measurement
When you use the
[WI
key to select a new device type and measurement,
the analyzer does the following:
l
presets the analyzer (except for external reference parameters, and trigger
mode)
0
takes a sweep
l autos&es the measurement
l
places a marker on the maximum or minimum point (depending on the
type of measurement)
l
makes the marker active
l
displays the AM delay connection diagram (when AM delay measurement is
chosen; Option
l
modifles
See
the sweep time (Option 100 only)
Table
3-l for a table of parameters for each measurement type.
Once you have selected your device, you can use the
1DA
or
1DB
only)
softkeys
to select the
measurement you wish to make. When you select a new measurement, a
preset is not done. It is assumed that you are simply changing measurement
types and that you may have changed some of the analyzer’s parameters
(such as frequency, power, etc.) for your DUT, and that you would not want
these parameters changed for subsequent measurements.
3-13
Making Measurements
Using the BEGIN Key to Make Measurements
NOTE
If the new measurement selected is a broadband measurement such as power, conversion loss, or AM
delay, the start frequency is limited to at least 10 MHz. Therefore, if your customized setup contains
a start frequency below 10 MHz and you choose power, conversion loss, or AM delay, the start
frequency will be changed to 10 MHz. The stop frequency will remain unchanged, unless it was sat to
below 10 MHz.
The (BEGIN) Kay and
Measurement Channels
The (BEGIN) key is designed to work when measurement channel 1 is active.
However, it does change the measurement mode of measurement channel 2
as well.
If measurement channel 2 is active when the
[BEGIN)
key is used to select a
new devicetype, measurement channel 2 is turned off, and measurement
channel 1 is made active.
If measurement channel 2 is active when the
(EiZi@
key is used to select a
new measurement type, measurement channel 2 will be left on and active.
However, the analyzer then proceeds to setup channel 1 for the requested
measurement type, even though channel 2 is the active channel.
3-14
Making Measurements
Using the BEGIN Key to Make Measurements
Using the (k%) Key To Configure Measurements
This procedure shows you how to configure the network analyzer for
measurements.
1.
Press
(j-1.
predehned parameters.
Press
2.
(jBEGlNJ
will be measuring (amplifier,
cable).
Connect your test device to the network analyzer.
3.
Presetting the instruments puts it into a known state with
and then use a
softkey
to select the type of device that you
hlter,
broadband passive device, mixer, or
Use the
4.
l Press
l
softkeys
T~xn~r&%sn
to select the type of measurement you want to make:
if you want to measure the transmission
characteristics of an
Press
Rsflectian
if you want to measure the reflection characteristics
amplifier,
filter, or broadband passive device.
of your device.
l
Press
PQWW
if you want to measure the RF power of a device. (The
Power
selection is under the
l Press
l Press
Canvsrsian Lass
a device. (The
I$M
Delay (Option
Conversion
delay of a device. (The AM Delay selection is under the
l
Press
SRf,
(Option 100 only) if you want to measure the structural
return loss of a cable. (The SRL selection is under the
l Press Fault Location (Option 100 only) if you want to measure the
&Gif%&r
menu.)
if you want to measure the conversion loss of
Loss selection is under the
1DA
or
1DB
only) if you want to measure the
Mfi&
Mix~
Cablo
menu.)
cable fault location. (The Fault Location selection is under the
menu.
)
menu.)
menu.)
CabXs
3-15
Making Measurements
Using the BEGIN Kay to Make Measurements
Depending on your selection, the analyzer is set to one of the following
configurations. (The SRL and Fault Location
confIgurations
are discussed in
the Option 100 User’s Guide Supplement.)
Table 3-1. Measurement Configurations from the
Transmission
Frequrncy
Frequency
Powor Levrl
Measurement Channel 1
Maasurmaont Channel
Format
Number of Points
Swoop Time Mode
Sweep Triggering
Dotaction Mode
Measurement Paths
I
Avoraaina
I
System Bandwidth
I
1 Options
2 HP
3 HP
4 Preset power level is user-defined by using the
5 Maximum power is dependent upon the option configuration of your analyzer. See Chapter 10 to determine the maximum specified power for your analyzer.
Rango0.300
Rango’
1OA
and
871X
8714C
MHz-1300
0.300
MHz-3000
preset power
Transmission
2
Log Mag
Continuous
Narrowbend
I
I
Medium Wide
I
10B
only
MHz 0.300
MHz 0.300
level4
off
201201
AutoAutoAutoAutoAuto
B/R
Off
Prmt
Pwr Level
Reflection
MHz-1300
MHz-3000
preset power
ReflectionPowerConversion
Continuous
Narrowbend
I
I
Medium Wide
I
softkqc
MHz 10
MHz 10
level4
preset power level4
Broadband Internal
Off
I
I
The factory default is 0
MHz-1300
MHz-3000
Medium WideMedium WideNarrow
Power
Off
log Mag
201
Continuous
Off
darn.
@iZiQ
Kay
Conversion Loss
MHj
10
MHz-1300
MHI
10
MHz-3000
oreset
Dower
MHz
1
10
MHz 10
level4 maximum
loss
off
h Mag
Continuous
Broadband Internal Broadband External
AM Delay’
MHz-1300
MHz-3000
soeciiied5
AM Delay
N/A
MHz
MHz
3-16
Making Measurements
Using the BEGIN Key to Make Measurements
The User BEGIN Function (Option lC2 only)
The
User
BEGIN softkey gives you the capability to
menu and
install
user-defined macro functions. The
available if your analyzer has
define macros such as:
l
Softkeys
l
Softkeys
l
Softkeys
to implement
fast
save/recall
to implement most used functions or features
to implement often-used features that involve a number of steps
BASIC
(Option
lC2)
redefine
the
USW REGIN
installed. Use this key to
C-1
key
key is only
Macros must be
program is currently installed (either by AUTOST or
deiined
within an
IBASIC
program. If no
VWW
BEGIN
RwxXI, Program)
the
analyzer will automatically create a default program.
USB~
BEGIN
oft
OFF selects the
[BEGIN]
key menu to
%ser”
mode when
ON, and to normal operation when OFF.
Once you have changed the
be displayed for subsequent key presses of the
true if your
User BEGJN
BASIC
program has changed. If the program has changed, the
mode is reset to OFF.)
U%W RRGI#
mode to ON, the same menu will
(-1
hardkey.
(This is not
Use of the User BEGIN function does not restrict access to any normally
available instrument feature such as marker functions, etc., nor does this key
affect sweep update rates.
Refer to example programs provided on the
IBASIC
programs disk for
implementation requirements. Keystroke recording may be used to modify or
update
UWW REGIN
programs.
See Chapter 7, “Automating Measurements,” for more information.
3-17
Measuring Transmission Response
This section uses an example measurement to describe how to calibrate for
and make a basic transmission response measurement. In this example, a
bandpass
used.
Enter the Measurement Parameters
Press
includes measuring transmission on measurement channel 1.
c-1
titer
like the one that was supplied with your network analyzer is
on the analyzer to set the analyzer to the default mode which
NOTE
This example measurement uses the default instrument parameters far a transmission measurement. If
your particular transmission measurement requires specific parameters (such as frequency range, source
power level, number of data points, and
sweeptime)
enter them now
3-18
Making Measurements
Measuring Transmission Response
Calibrate For a Transmission Response Measurement
Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform an enhanced transmission response calibration.
When you perform an enhanced transmission response calibration, the
analyzer performs correction at the selected number of data points across
the selected frequency band. Interpolation recalculates the error correction
array for reduced frequency spans. If the frequency span is increased the
calibration is invalidated, and the default response calibration is automatically
restored.
Chapter 6 provides detail about when additional calibration is necessary, and
information about other calibrations available for transmission measurements.
If you wish to calibrate your instrument for a transmission response
measurement, perform the following steps:
1.
Press
@ l&hated
Rasponso .
2. The instrument prompts you to connect four standards-open, short, load,
and through cable-as shown below.
NETWORK ANALYZER
j
RF OUT
.-.--.---------i---------------,
OPEN
Open, Short, load Connections
3. Press
SHORT
pbsata~~~ Standard
LOAD
after connecting each standard.
NETWORK ANALYZER
RF OUT
po611b
Through
Cable,Connection
RF IN
3-19
Making Measurements
Measuring Transmission Response
4. The analyzer will measure each standard and then calculate the new
calibration coefficients. The message “Calibration complete. ’ will
appear for a few seconds when the analyzer is done calculating the new
error correction array.
5. The calibration may be saved in memory or on a disk for later use if you
wish. However, the current calibration for each measurement channel is
always saved in nonvolatile (battery-backed) memory and will be used
the next time the analyzer is preset (see note below) or turned on. See
Chapter 6 for information on saving calibrations to the analyzer’s internal
memory, the analyzer’s random access memory (RAM), or to a floppy disk.
NOTE
Changing sweep frequencies land other source parameters) may affect your calibration. See Chapter 6,
“Calibrating for Increased Measurement Accuracy,” for more information.
I
I
3-20
Connect the DUT
NETWORK ANALYZER
Making Measurements
Measuring Transmission Response
TEST
Figure 3-5. Equipment Setup For e Transmission Response Measurement
3-21
Making Measurements
Measuring Transmission Response
View and Interpret the Transmission Measurement
Results
1. RI view the entire measurement trace on the display, press
@Xii--
2. ‘lb interpret the transmission measurement, refer to Figure 3-6 or
your analyzer’s display if you are making this measurement on your
instrument.
a. The values shown on the horizontal axis are the frequencies in MHz.
Awmiw~~s .
The values shown on the vertical axis are the power ratios in decibels
(dES)
of the transmitted signal through the device divided by the
incident power. ‘lb display the result in logarithmic magnitude format
(designated by “Log Mag” at the top of the measurement screen), the
analyzer computes the measurement trace using the following formula:
Transmission
where
where
Ptrans
Pin=
(&I)
= 10 log
the power transmitted through the device and
=
= the incident power.
P
F
(
znc
>
b. A level of 0 dB would indicate a perfect through cable or device (no
loss or gain). Values greater than 0 dB indicate that the DUT has gain.
Values less than 0 dB indicate loss.
3. lb quickly determine the filter’s minimum insertion loss, press
Mwk~r
Search Max
Search Mlm -> Max.
@iEKK)
4. Note the marker readout in Figure 3-6 provides the frequency and
amplitude of the minimum insertion loss point.
3-22
)I: Transmlsslon
D2:
Off
Log Mag
Making Measurements
Measuring Transmission Response
2n
n dR/
RPf
-60 00 rlfl
c
-140
Abs
Start 0.300 MHz
Stop 3 000.000 MHz
Figure 3-6. Example of a Transmission Measurement Display
5. See “Using Markers” in Chapter 4 for more detailed information on using
markers to interpret measurements.
NOTE
For the measurement to be valid, input signals must fall within the dynamic range of the analyzer. See
Chapter 5 for techniques to increase the dynamic range of the analyzer.
3-23
Measuring Reflection Response
This section uses an example measurement to describe how to calibrate
for and make a basic reflection response measurement. In this example, a
bandpass
used.
Enter the Measurement Parameters
Press the following keys on the analyzer:
pEFY]
(jKiiG--
titer
like the one that was supplied with your network analyzer is
NOTE
I
This example measurement uses the default instrument parameters for a reflection response
measurement. If your particular reflection measurement requires specific parameters (such as frequency
range, source power level, number of data points, and sweep time), enter them now
3-24
Making Measurements
Measuring Reflection Response
Calibrate For a Reflection Response Measurement
Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform a reflection one-port calibration. A one port
calibration uses known standards to correct for directivity, source match, and
frequency response errors in narrowband measurements.
To perform a reflection one-port calibration you will need one of the following
calibration kits depending on the nominal impedance of your analyzer:
HP 850323
HP
85032B
HP 850363
HP
85036B
HP
85033D
for 50 !I type-N female connector calibrations
for type-N female or type-N male 50 Q connector calibrations
for 75 0 type-N female connector calibrations
for type-N female or type-N male 75 6) connector calibrations
for 3.5 mm female or 3.5 mm male 50 tI connector
calibrations
HP
85039B
NOTE
If you are going to be using calibration standards other than the default (female
select the type of connector type by pressing
type.
for Type-F 75 61 connector calibrations
ICALl Ca3, Kf%
and then selecting the appropriate
type-N),
you must
3-25
Making Measurements
Measuring Reflection Response
Chapter 6 provides detail about when this calibration is necessary. If you
wish to calibrate your instrument for a reflection one-port measurement,
perform the following steps:
1.
Press (CAL)
Dria POW
2. The instrument will prompt you to connect three standards (open, short
and load) and measure them. See Figure 3-7.
NETWORK ANALYZER
LOAD
3. Press
Figure
Ffeas~~
OPENSHORT
3.7.
Equipment Setup For a Reflection Response Calibration
Standard after connecting each standard.
4. The analyzer will measure each standard and then calculate new
calibration coefficients. The message “Calibration complete. n will
appear for a few seconds when the analyzer is done calculating the new
error correction array.
5. The calibration may be saved in memory for later use if you wish. See
Chapter 6 for information on saving calibrations.
3-26
Connect the DUT
Making Measurements
Measuring Reflection Response
NETWORK ANALYZER
/
P II
IN
Figure 3-8. Equipment Setup For a Reflection Measurement of a Two-Port Device
NETWORK ANALYZER
NETWORK ANALYZER
,
LT
Q ’
LOAD
0
RF OUT
I
PD*1w
Figure 3-8. Equipment Setup For a Reflection Measurement of e One-Port Device
DEVICE UNDER TEST
III
3-27
Making Measurements
Measuring Reflection Response
h:Reflectlon
D2:
off
-30
-35
Start 0.300 MHz
Log Mag
5.0
Ref -15.00 dB C
dB/
stol3
3 000.000 MHZ
Figure 3.10. Example of a Reflection Measurement Display
3. To quickly determine the
titer’s
return loss, press
[““ARKER)
and then use
the front panel knob, the m a keys, or the numeric keypad to read the
value of return loss at the desired frequency.
4. See “Using Markers” in Chapter 4 for detailed information on using
markers to interpret measurements.
3-29
Making a Power Measurement using Broadband
Detection
Power measurements can be made using either narrowband or broadband
detection. The example in this section is of a broadband power measurement,
If
you are only interested in the output power of your device at the
same frequency as the analyzer’s source, you can select
D#to~tion Options #mawband Internal B
for a narrowband power
measurement. A narrowband power measurement only measures the power
within the tuned receiver’s bandwidth centered at the source frequency.
When you measure a device for absolute output power, the network analyzer
uses the broadband detection mode and measures the total power of
frequencies present in the transmitted signal (B*). This signal may contain
frequencies other than the source frequency such as when the DUT is a
mixer.
This section uses an example measurement to describe how to normalize the
data and measure the total output power of an ampltier.
@EEi)
ah
NOTE
Broadband power measurements are only specified for measurements with a start frequency
of
210
MHz.
3-30
Making Measurements
Making a Power Measurement using Broadband Detection
Enter the Measurement Parameters
Press the following keys on the analyzer:
c-1
(jMEAs-
POW%@
NOTE
This example measurement uses the default instrument parameters for a power measurement. If your
particular power measurement requires specific parameters (such as frequency range, source power
level, number of data points, and sweep time1 enter them now
CAUTION
Damage to your analyzer will occur if the receiver input power exceeds
+23
dBm
or 25 Vdc. The analyzer’s source cannot
signilicantly
exceed this
level, however if your DUT has gain, then attenuation on the RF IN port
may be necessary. See “When to Use Attenuation and
Amplilication
in a
Measurement Setup” earlier in this chapter for more information.
3-31
Making Measurements
Making a Power Measurement using Broadband Detection
Connect the DUT
NETWORK ANALYZER
TEST
3-32
Figure
3-l
1. Equipment Setup For a Power Measurement
Making Measurements
Making a Power Measurement using Broadband Detection
View and Interpret the Power Measurement Results
1.
To view the measurement trace, press
2. Figure 3-12 shows the results of an example power measurement.
3. lb interpret the power measurement, refer to Figure 3-12 or your
analyzer’s display if you are making this measurement on your
instrument.
a. When making a power measurement, the display shows the output
power measured at the analyzer’s RF IN connector. This power is
absolute power, as opposed to a power ratio.
b. Note that when making a power measurement, the values associated
with the vertical axis are in units of
in reference to 1
mW.
CsCALEJ M~~%cale
dBm,
which is the power measured
.
OdBm=
-10
dBm
= 100
+lOdBm =
1mW
PW
1OmW
3-33
Making Measurements
Making a Power Measurement using Broadband Detection
)I:
Power
)I:
Power
D2:
Off
D2:
Off
Log MagLog Mag
0.5
dB/
dB/
Ref0.5
Ref
5.00
5.00
dBm
dBm
CAUTION
3
AbsAbs
Stop 1 300.000 MHz
Start 10.000 MHz
Start 10.000 MHz
Figure
Figure
3-l
2. Example of a Power Measurement
3-l
2. Example of a Power Measurement
Stop 1 300.000 MHz
If the analyzer’s RF output power level is set to higher than the specified
output power for your analyzer, the source could go unleveled. See
Chapter 10 for source and receiver specifications. If your device requires
input power greater than your analyzer’s specified output power, you may
need to use a preamplifier in your measurement setup. However, remember
to not exceed the receiver damage limit of +23
dRm.
3-34
Measuring Conversion Loss
Conversion loss is the ratio of IF output power to RF input power expressed
in
dB.
This section uses an example measurement to describe how to measure
the conversion loss of a broadband mixer.
When characterizing a device’s conversion loss, the analyzer uses broadband
detection to compare the transmitted signal (B*) to the reference signal
(R*). This is because the input and output signals of a frequency-translating
device may be different. Since broadband detection measures signals at all
frequencies, you may want to use a hlter to remove unwanted signals such as
LO feedthrough when performing this measurement.
For example, an RF signal at 900 MHz mixed with an LO signal at 200 MHz,
results in mixing product signals at 700 MHz and 1100 MHz, as well as the
original 900 MHz and 200 MHz RF and LO signals.
3-35
Making Measurements
Measuring Conversion loss
a;-
200
E,;
200
MHz
700 MHz
Bandpass Filter
700900
MHzMHz MHz
(RF-LO) (RF) (RF+LO)
1100
po650b-c
Figure 3-13. Filtering Out the Unwanted Mixing Product
Inserting a 700 MHz bandpass filter in the measurement setup removes the
unwanted signals at 200 MHz, 900 MHz and 1100 MHz, providing an accurate
measurement of the desired IF signal at 700 MHz.
In the following example, the conversion loss of a mixer will be measured
with RF input frequencies over a 15 MHz span centered at 900 MHz. With an
LO frequency of 200 MHz, the mixer IF frequency will sweep over a 15 MHz
span centered at 700 MHz.
3-36
Making Measurements
Measuring Conversion loss
Enter the Measurement Parameters
Press the following keys on the analyzer:
(PRESET)
(MEAS)
NOTE
This example measurement uses the default instrument parameters for a conversion loss measurement.
If your particular conversion loss measurement requires specific parameters
source power level, number of data points, and sweeptimel enter them now
isuch
as frequency range,
I
3-37
Making Measurements
Measuring Conversion Loss
Perform a Normalization Calibration
Normalization is the simplest type of calibration. The analyzer stores
normalized data into memory and divides subsequent measurements by the
stored data to remove unwanted frequency response errors. This calibration
is used for this measurement to remove the insertion loss error of the IF
flter. Changing the frequency span or number of measurement points will
invalidate a normalization calibration.
Perform the following steps to perform a normalization calibration:
1. Connect the equipment as shown in Figure 3-14, except replace the mixer
with a through cable.
2. Set the following frequency parameters:
(FREQ)
This sets the analyzer frequency range to sweep over the passband of the
IF hlter
3.
Press
(700 MHz).
(EiKEj
lkwmalize
.
This stores the filter response passband into memory, and sets up a
normalized trace so that the filter response magnitude is removed from the
measurement.
4. Replace the through cable with the mixer as shown in Figure 3-14.
5.
Press
IFREQl Cen%er 1900)
MHz to change the center frequency so that the
mixing product of the mixer is in the passband of the IF filter.
3-38
Connect the DUT
NETWORK ANALYZER
Making Measurements
Measuring Conversion Loss
OSCILLATOR
Figure 3-14. Equipment Setup For a Conversion Loss Measurement
3-39
Making Measurements
Measuring Conversion Loss
View and Interpret the Conversion Loss Results
1.
If necessary to view the measurement trace, press
2. To interpret the conversion loss measurement, refer to Figure 3-15 or
your analyzer’s display if you are making this measurement on your
instrument.
a. The values shown on the horizontal axis represent the source RF
output. The values shown on the vertical axis are the power ratio in
decibels
(dB)
of the transmitted signal through the device divided by
the incident power. ‘lb display the result in logarithmic magnitude
format (designated by Log Mag at the top of the measurement screen),
the analyzer computes the measurement trace using the following
formula:
Conversion Loss
where
where
Ptrans
=
= the incident power at the RF input.
Pine
(dB)
= 10 log
the power measured at the IF output of the mixer and
P
F
(
*nc
>
c-1
Au7xjscale
.
b. A level of 0 dR would indicate a perfect device (no loss or gain). Values
greater than 0 dB indicate that the mixer has gain. Values less than
0 dR indicate mixer conversion loss.
3. If you wish, you can quickly determine the mixer’s minimum conversion
loss by pressing [MARKER) MarkerSsaxxh
3-40
Nax: SCWX~ Hkx” --) Max.
Making Measurements
Measuring Conversion Loss
W.: Conv Loss
data
? D2:
off
I
-7.6
Center 900.000 MHz
/
Figure 3-15. Example of a Conversion Loss Measurement
/M
Log Mag
III
0.2
dB/
Ref
-6.80 d8
I
Span 15.000 MHz
NOTE
For the measurement to be valid, input signals must fall within the dynamic range and frequency range
of the analyzer. See Chapter 5 for techniques to increase the dynamic range of the analyzer.
I
1
3-41
Measuring AM Delay (Option
An AM delay measurement characterizes the group delay (or envelope delay)
of a device. lb perform this measurement you must have ordered either
Option
options include internal instrument hardware and firmware, two external
scalar detectors and a power splitter.
Group delay flatness can be a key specification for many components and
systems. Distortionless transmission of a signal requires constant amplitude
and group delay response over the frequency bandwidth. Group delay is the
measurement of signal transmission time through a device. It is
derivative of the phase characteristic with respect to frequency.
If the device under test is a frequency translator, the device input and
output frequencies will by deiinition be different. This generally makes the
measurement of the device phase response (and therefore group delay) very
difhcult.
1DA
(AM Delay, 50 ohm) or Option
1DB
IDA
or
1DB)
(AM Delay, 75 ohm). These
defined
as the
The AM Delay option overcomes this
modulation technique to measure group delay. In this technique, a small
amount of amplitude modulation is applied to the RF output of the analyzer.
Scalar detectors are used to detect this modulation both before and after
the device under test. The group delay can then be calculated from the
phase difference between these two signals (modulation envelopes). Since
broadband detection is used, the Option
through nearly any device, including frequency translators.
There are several important considerations in an AM delay
the device is a limiter or has AGC (automatic gain control), this will tend to
distort or remove the amplitude modulation used for the measurement. Any
limiting or AGC in the device should be disabled before making an AM delay
measurement. The broadband detection used for AM delay is susceptible to
spurious signals and noise. High-level spurious signals should be removed
with altering. The signal levels at both the reference and test detectors
should be kept as high as possible. The
both detectors in an AM delay measurement is - 10 to + 13
device input power must be outside this range, amplification or attenuation
must be used directly before the device. If the device output power is outside
the - 10 to + 13
after the device.
3-42
dBm
range, attenuation or amplihcation must be used directly
dilhculty
lDA/lDB
speciiied
by using an amplitude
analyzer can measure delay
measurementIf
incident power range for
dBm.
If the
Making Measurements
Measuring AM Delay (Option
1DA
or 1
DB)
Enter the Measurement Parameters
Connect the detectors and power splitter to the analyzer as shown in
Figure 3-16 and then press the following keys on the analyzer:
[W)
@EXE-)
$M
Dolay
You may also press the following keys to access AM delay. Pressing these
keys will result in a connection diagram being displayed on the screen of the
analyzer.
fpizm-)
@zig
MiPW
AM Dolay
NOTE
This example measurement uses the default instrument parameters for an AM delay measurement. If
your particular AM delay measurement requires specific parameters
power level, number of data points, and sweeptimel enter them now
[such
as frequency range, source
3-43
Making Measurements
Measuring AM Delay (Option 1 DA or
1DBI
Calibrate For an AM Delay Measurement
1. Connect the equipment as shown:
NETWORK
ANALYZER
2.
Press
Figure
a
Response
1
3-l
6. Equipment Setup For an AM Delay Response Calibration
Maamm
un
Standard .
3-44
Connect the DUT
Making Measurements
Measuring AM Delay (Option
1DA
or
1DBl
Figure
3.17.
Equipment Setup For an AM Delay Measurement
3-45
Making Measurements
Measuring AM Delay (Option
1DA
or
1DBI
View and Interpret the AM Delay Results
This example is an AM delay measurement of a frequency converter.
1. To
view the measurement trace, press
2. lb interpret the AM delay measurement, refer to Figure 3-18.
a. Note that the vertical axis is displaying time rather than power as
in previous example measurements. The AM delay measurement
measures the time required for power to travel through the DUT at
various frequencies. The measurement trace will be noisier as the
power level is attenuated by the DUT.
b. Since delay is proportional to the derivative of phase, flat (constant)
delay indicates linear phase. Delay measurements are typically
performed to measure the deviation from linear phase. Deviation from
linear phase (flat delay) would indicate that the DUT is distorting the
signal.
@?XZj An%orrcale
.
3-46
)I:
AM Delay
D2:
off
55.6
Measuring AM Delay (Option
kHz
20
ns/
Fief
Making Measurements
1DA
or 1
DBI
1.46
/JS
C
1.381.38
AbsAbs
Center 233.800 MHzSpan 8.000 MHz
Figure 3-16. Example of an AM Delay Measurement
3. Use the marker flatness function to determine the maximum deviation.
See Chapter 4 for how to use the marker functions.
3-47
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.