Agilent 8645A Product Note

Agilent 8645 Signal Generator Communication
Product Note 8645-2
A catalog of 8645A information
Table of contents
Operation related topics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Block diagram and theory of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Internal audio source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Frequency sweep capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Externally doubled outputs to 2060 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operation as a phase noise reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Programming with HP-SL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Command sequence independence using HP-SL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Performance related topics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Phase noise performance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Spurious performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Third order intermodulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Divided outputs below 515 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Stereo separation quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Minimizing fan noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Frequency agility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Functional description of frequency agile operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Faster frequency switching using multiple agile generators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Frequency accuracy of agile outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Relating phase error and frequency accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Amplitude dynamic range while frequency hopping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Amplitude shaping of agile outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
High rate, high deviation FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Simultaneous modulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Digitized FM operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
AC coupled FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Special capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Tailored operation through special functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Protecting classified instrument settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Storage registers and sequential recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Offsets and multipliers of frequency and amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Built-in calibration functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Finding failures with internal diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
2
This product note is actually a compilation of many brief prod­uct notes, each concerned with a particular aspect of the 8645A agile signal generator. Included in these pages are explanations of how this unique signal genera­tor operates, the capabilities it
has to offer and the performance it can provide. The objective of this product note is to be a refer­ence guide for the owner of a 8645A, to help maximize the use­fulness and performance of this agile signal generator in the intended application. While none
of the topics are covered in great detail and other literature may offer a more thorough treatment of a subject, these summaries should provide sufficient infor­mation to help in many situa­tions.
Operation related topics
Block diagram and theory of operation
The basis of the 8645A is a sin­gle fractional N loop controlling a VCO operating in the frequen­cy range of 515 to 1030 MHz. The reference signal for this phase lock loop originates from either an internal 10 MHz oscil­lator or an external input. An extensive divider section at the output of the phase lock loop provides coverage down to 252 kHz and a doubler in the output section extends the fre­quency range to 2060 MHz. All four modulation types are imple­mented in the 8645A with either the internal 400 kHz synthesizer integrated circuit providing the modulation waveform or an external input. Frequency modu­lation uses two techniques including an analog signal summed into the VCO tuning input and a digitized FM tech­nique that directly modifies the fractional N number of the phase lock loop. Phase modulation sig­nals are summed directly into the fractional N phase lock loop. Pulse modulation occurs directly after the divider section. Amplitude modulation is accom-
plished in the output section through control of the Automatic Level Control (ALC). The AM sig­nal is summed together with the level DAC which sets the ampli­tude level that reaches the atten­uators. The combination of the level DAC, the AM signal, and the attenuators (up to 120 dB of attenuation) determine the actu­al output level of the 8645A. The Reverse Power Protection (RPP) prevents the output circuits from damaging signals externally input through the generator’s output. Controlling all of this hardware in the many states the user can set up is a Motorola 68000 microprocessor.
The basic block diagram summa­rized above produces all the tra­ditional functions of a signal generator. For the applications intended for the 8645A, the phase noise and spurious signals must be very low at offsets greater than approximately 10 kHz. A major advantage of the block diagram is that a clean-up loop based on a delay line and a phase detector can be added in parallel to the fraction­al N phase lock loop. The 70 nsec delay line in the clean-up loop of the phase noise enhancement section decreases the phase noise and spurious signals to levels required by communica­tions hardware tests.
Besides high performance out­puts for traditional applications, the 8645A is designed to provide sequences of many frequencies in rapid order. Frequency switching is specified as fast as 15 usec between frequencies. To accomplish this switching speed, the fractional N phase lock loop is opened and replaced by a delay line frequency lock loop. Phase noise and spurious signals on the VCO output are again decreased by the delay line and phase detector in the fast hop enhancement section. VCO set­tings learned before fast hop operation begins are sent to the VCO through a pretune DAC in the order of the output frequen­cies the user wants and at the rate programmed. Amplitude information is simultaneously sent to the level DAC. A hard­ware state machine programmed by the microprocessor provides all the fast control signals need­ed while fast hop operation is underway.
Many of the operational areas briefly discussed on this page are covered more thoroughly in other parts of this product note. Refer to the table of contents for a listing of the topics.
3
4
Timebase configurations
The frequency stability of the 8645A depends a great deal on the reference oscillator in use. The standard internal timebase is a non-ovenized 10 MHz crystal oscillator with a typical aging rate of ±2 ppm per year. With this timebase, a 1 GHz output of the signal generator would not vary more than ±2 kHz in a year due to timebase aging. However, the frequency drift due to tem­perature changes may be twice this amount because this oscilla­tor is not ovenized. Although the 8645A has several design fea­tures to minimize internal tem­perature fluctuations, the standard timebase could drift by as much as ±4 ppm over a tem­perature range variation of 0 to +55 degrees centigrade.
Option 001 of the 8645A adds a more stable 10 MHz ovenized timebase to the instrument. The aging rate is specified to be bet­ter +0.0005 ppm or a 0.5 Hz vari­ation of a 1 GHz output in 24 hours after a 10 day warm-up. Frequency drift due to a ambient temperature change of 0 to +55 degrees centigrade is typically less than +0.006 ppm. The fre­quency of this timebase can be mechanically adjusted through a hole in the rear panel using a tweaker. Voltage control of the timebase frequency is available using the Electronic Frequency Control (EFC) input. The maxi­mum ±10 volt EFC input signal will produce a ±1 Hz frequency change of the 10 MHz output.
The output of this optional high stability timebase is only routed to the rear panel of the instru­ment as the oven ref output. An external jumper cable is used to input this reference signal at the ref in port for routing into the frequency synthesis circuits. When this jumper cable is con­nected, the instrument will sense the presence of a reference signal at the ref in input and uti­lize it automatically. Without a signal present at the ref in input, the 8645A will use the standard timebase as its refer­ence oscillator.
To allow other instruments to use the timebase signal from the 8645A, the rear panel 10 MHz ref out output provides an out­put of either the standard or optional timebase that is cur­rently in use. The signal genera­tor can also utilize an external 10 MHz timebase that would be input at the ref in input. Activating special function 161 will provide a readout indicting whether the 8645A is utilizing the standard timebase or a sig­nal entering the ref in input.
5
8645A Rear panel
Typical connection
Standard reference oscillator (10 MHz)
Optional high stability oscillator (10 MHz)
Switch control
To synthesizer
Ref in
Level detector
Oven ref output
EFC input
10 MHz ref out
8645A internal timebase configuration
Internal audio source
The internal audio source in the 8645A can generate four basic waveforms of sine, sawtooth, square, and white Gaussian noise. Waveforms are generated by a numerical synthesis tech­nique. The heart of the synthe­sizer is a Digital Waveform Synthesis Integrated Circuit (DWSIC). The DWSIC generates a continuous stream of numbers that represents instantaneous levels of the waveform. This “dig­ital” waveform is then converted to an analog signal by a digital-to-analog converter. The analog signal is conditioned by conventional analog circuitry and routed to various parts of the signal generator. The condi­tioning circuits include a sample and hold to remove DAC switch­ing noise, filters to remove quan­tization noise, and amplifiers to boost the output.
The internal audio source is used in the signal generator for modulation, sweeping, calibra­tion, and diagnostics. To the user, the source appears like an internal function generator used to modulate the carrier with the four basic waveforms. It is also used as a ramp voltage into the FM circuitry during phase con­tinuous sweep that disallows internal modulation being active this sweep mode. This source is used as an accurate DC refer­ence to calibrate FM deviation and AM depth when these modu­lations are active. The built-in diagnostics use the source for DC and AC signals to test vari­ous modules in the instrument. And of course the audio signal is available at the front panel audio output with programmable wave­forms, amplitude, and frequency.
The type of waveform produced can be selected by activating special function 130 or via GPIB with the command LFS:Waveform <type> where <type> is sin, square, saw, or WGN (for white Gaussian noise). The frequency can be selected over a range of 0.1 Hz to 400 kHz. Sawtooth and squarewave rates should be limited to less than 50 kHz because the output circuitry degrades the perform­ance at higher rates. Frequency accuracy is equal to the internal timebase accuracy of the instru­ment. Frequency switching speed of the source is typically less than 30 msec. Output level is programmable and ranges from 1 mV to 1 Vrms into a 600 ohm load with a specified accu­racy of ±20 mV. Adjusting the output level will effect the amount of internal modulation present such that a decrease in output level will proportionately decrease the amount of internal modulation. This feature can be used to increase the amount of external modulation allowed during simultaneous internal and external modulation. The sum of the internal and external voltages should not exceed
1.4 Vpeak during simultaneous modulation or clipping distor­tion may occur.
6
Frequency sweep capabilities
The 8645A was designed to have three different types of frequen­cy sweep operation to accommo­date a wide variety of applications. As is evident from the descrip­tions that follow, the wide devia­tion FM capabilities and the fast hop operation offer unique sweep capabilities not present in the typical RF signal generator.
The most useful sweep for find­ing the frequency response of narrowband devices is the phase continuous frequency sweep. The instrument uses the wide deviation FM circuitry to create a phase-continuous output over spans as wide as twice the maxi­mum FM deviation available for that carrier frequency range. In the main VCO band of 515 to 1030 MHz the maximum span is 20 MHz. This range is decreased by half for each divider band below this main carrier band. A sweeptime range of 10 msec to 10 seconds is allowed for any span that is chosen. Only a lin­ear frequency sweep is allowed. Another capability that offers very high accuracy of each fre­quency point of the sweep is the digitally stepped frequency sweep. The instrument will step the synthesizer across any span set by the user in a linear or log
frequency spacing. The number of discrete points output will depend on the span and sweep­time that is set. Sweeptime can range from 0.5 to 1000 seconds with each discrete point requir­ing typically 90 msec to com­plete. To reduce the amount of switching transients spurs due to each frequency change, the output level is reduced approxi­mately 60 dB between each dis­crete frequency. This amplitude blanking may cause dropouts on the displayed frequency response. Due to these dropouts it may be more useful to specify a fast hop sweep for wide fre­quency spans as the following describes.
A unique frequency sweep capa­bility of the 8645A is the fast hop sweep. Utilizing the frequen­cy agile capability, large frequen­cy spans with 1000 discrete frequency steps in as little as 100 msec per sweep. The num­ber of frequency steps varies according to the sweeptime and frequency range selected with each discrete step taking 30 microseconds for outputs from 128 to 2060 MHz. The user can set a sweeptime range from 10 msec to 100 seconds.
Although the output is blanked between each frequency step as in digitally stepped sweep, the duration of the blanking is so short that the detector used to measure the frequency response will typically not show the dropout on the oscilloscope or network analyzer. Either a linear and log distribution of frequency steps can be selected.
Each of the three types of fre­quency sweep described above can be operated in a continuous repetitive output or a single sweep output triggered by the press of a key or an HP-SL com­mand. Additionally the digitally stepped and fast hop sweep types can be operated manually using the front panel knob or up/down arrow keys. Up to three markers can be entered for out­put during a sweep. When the sweep reaches the marker fre­quency a 0 volt signal is output from the Z axis port on the rear panel. The Z axis output is +1 volt during a sweep and +5 volts during retrace to blank the CRT of an oscilloscope. The X axis output of 0 to 10 volts matched to the progress of the frequency sweep.
7
Externally doubled outputs to 2060 MHz
For applications requiring out­puts above the 1030 MHz maxi­mum frequency of the standard 8645A, consideration can be given to either ordering the 8645A with the optional internal doubler, installing the 11867A retrofit kit or using an external doubler. This technical brief summarizes the capabilities and performance the user can expect while using one such external doubler, the 11721A frequency doubler, to increase the frequency range of the 8645A to 2060 MHz. The 11721A frequency doubler is a passive, full-wave rectifying doubling circuit that was designed to minimize conversion loss over a wide frequency range. Its output frequency range is 10 MHz to 2560 MHz. At input levels above +13 dBm the
doubler has an almost constant conversion loss of approximately 11 dB. This typical conversion loss after the +16 to +18 dBm maximum output of the 8645A results in an output signal level of +5 to +7 dBm for the average 11721A external doubler. The harmonic and spurious content of the output is almost complete­ly a function of the input signal. Note that harmonics input to doubler (specified at <–30 dBc for the 8645A below 1030 MHz) will increase approximately 6 dB due to the doubling function. The same 6 dB increase will be present on the phase noise of the carrier. Any frequency mod­ulation at the input to the dou­bler will double in deviation also.
For frequency agile signals, the 11721A has no measurable affect on the frequency switching time up to the fastest time of 15 usec available on the 8645A.
By using special function 111 frequency multiplier with an entered multiplier of 2, the dis­play of the 8645A will represent the signal at the output of the doubler as a convenience. Simultaneously, the doubler’s conversion loss can be entered as an amplitude offset to cali­brate the display for the actual amplitude at the doubler’s out­put.
More general information about the 11721A frequency doubler in use with the 8662A synthesized signal generator is available in application note 283-2 (litera­ture number 5952-8217).
8
Operation as a phase noise measurement reference
Among several techniques for measuring the phase noise of a source is the method of using a second source to demodulate the phase instability using a phase detector. Commonly referred to as the phase detector method, this process requires that the second source or reference source have as good or better phase noise performance than the source being tested. It is also required that one of the sources have an FM capability in order to maintain phase quadrature at the output of the phase detector. These needs for good phase noise performance and FM capa­bility often result in a generic signal generator being used as the reference source of a phase noise measurement system. The subject of this brief product note is how to optimize use of the 8645A as a reference source for phase noise measurements. More information on the measurement technique itself can be found in literature related to products such as the 11729C carrier noise test set or 3048A phase noise measurement system.
Several features of the 8645A make it a good choice for use as a phase noise measurement source. These include the wide carrier frequency range, an out­put power of +16 dBm and a large FM deviation range. The phase noise of the 8645A’s out­put is very low at offsets greater than 10 kHz from the carrier, as is commonly required for testing channelized communication devices or systems. The 8645A has very few spurs on its output which simplifies the detection and interpretation of spurs from the test source. The typical phase noise and spurious per­formance is indicated in the graph included in the “phase noise performance” summary of this product note.
As with any reference source used in the phase detector method, only as much FM devia­tion as required to establish the phase lock loop for the measure­ment should be used. Minimizing the FM deviation decreases the noise contribution of the FM cir­cuits and reduces the potential for an unstable Phase Lock Loop (PLL). The design of the 8645A uses two different FM implemen­tations that the user should choose between according to the FM deviation range required. The standard FM is recommend­ed for phase noise measure­ments that use a PLL bandwidth of less than 1.6 kHz. Variations in the group delay of the FM cir­cuits for deviation settings to support more than 1.6 kHz could cause inaccurate measurements
or loop oscillations. For situa­tions that require more loop bandwidth, it is recommended that the fast hop mode be acti­vated for the measurement. In the fast hop mode the group delay of the FM is very low and remains constant at higher FM deviations. Although the phase noise at low offsets increases in this mode, it is generally accept­able as sources that require more FM range to maintain quadrature also have higher phase noise to be measured.
One other unique characteristic of the 8645A is that several cir­cuits internally are reset when­ever the center frequency setting is changed so the output is not phase continuous during these changes. The output is decreased by more than 60 dB during these resets so that the unspecified output during the transition will not affect the user’s device. This transition period lasts less than 85 msec typically. While this operational characteristic will not affect a phase noise meas­urement in progress, it will be apparent when the center fre­quency of the 8645A is being tuned as the beatnote disappears momentarily with each change. This signal interruption will cause the PLL to momentarily break lock. Activating special function 105 amplitude muting disables this amplitude blanking but the unspecified transitions of the output signal could still result in perturbations while tuning frequency.
9
Programming with HP-SL
Hewlett-Packard Systems Language (HP-SL) is the pro­gramming language for instru­mentation adopted by Agilent Technologies. This language uses standard GPIB hardware and will be used in many new Agilent products. The 8645A is the first signal generator to implement HP-SL. HP-SL uses self-explana­tory commands and is flexible for beginning and advanced pro­grammers. Programs written in HP-SL for the 8645A will be com­patible with the other generators with the exception of commands associated with unique functions of the signal generator such as fast hop capabilities. This is intended to minimize software modifications by the customer when hardware is upgraded or replaced.
Many Agilent divisions have con­tributed to the development of HP-SL and will use it as part of an interface system that con­forms to the new IEEE 488.2 standard. The advantage of the new IEEE standard is that it defines common global com­mands such as for the instru­ment preset function, as well as hardware and protocol that is compatible with previous stan­dards. In the short term HP-SL will be easier to learn and self documenting and in the long term, HP-SL will provide a more common language to reduce the cost of software support. A simple example shows how
HP-SL commands are self explanatory and what a typical program for the 8645A could look like. The following program lines will perform an instrument preset on the signal generator, set the RF frequency to 500 MHz and the amplitude to 10 dBm, and turn the RF output on.
100 Output 719; “*RST”
200 Output 719; “Frequency:CW 500 MHz”
300 Output 719; “Amplitude:Level 10 dBm”
400 Output 719; “Amplitude:State on”
This example programming can be further simplified because with HP-SL commands can be combined in a single output statement without regard for the order in which the instrument will execute the commands. This means that HP-SL instruments will take in the full command “message” of a single line of pro­gramming before executing any of the contents. The command message defines the final instru­ment state that is wanted with­out regard for the order of commands given. This eliminates the problem of programming an unallowed instrument state such as increasing FM deviation before increasing carrier fre­quency. However care must be taken that each individual mes­sage only defines one final instrument state and not several. With this HP-SL capability, the previous example can be changed to the following:
Since “*RST” defines a complete instrument state on its own, it cannot be combined with the other commands or it will be uncertain which state will result. This example also shows the use of the short form of commands as well as implied commands and implied units. The semicolon is used to separate commands in a single output, and the colon is used to separate words in a sin­gle command. The commands with asterisks are used with all IEEE 488.2 compatible instru­ments that can execute that function. In HP-SL commands, spaces should be between words and arguments but not before or after punctuation. Much more information on HP-SL program­ming with the 8645A is provided in the HP-SL Programming Guide (literature number 5951-6710).
10
100 Output 719; “*RST”
200 Output 719; “Freq 500 MHz;Ampl:Lev 10; Stat on”
Command sequence independence using HP-SL
A current problem with instru­ment programming is that each command that is received by an instrument is executed immedi­ately. When the user is trying to set up a complete instrument state, the order in which the commands are sent must be cor­rect so that each intermediate state is valid. Implementing HP-SL on the 8645A has elimi­nated this command order dependence through the creation of command “messages”. A mes­sage contains all of the instru­ment commands that will result in the desired instrument state. None of the commands are implemented by the instrument until the complete message is received. In this structure, the order of the commands in the message is irrelevant. The pro­grammer constructs messages to describe the final instrument state that is needed without wor­rying about the way the instru­ment gets to that state.
For example, suppose a signal generator had the following capability dependencies between carrier and FM deviation range:
Carrier range FM deviation range
100 MHz to 1 GHz 1 MHz to 10 MHz
10 MHz to 100 MHz 100 Hz to 1 MHz
With the previous control struc­ture, it is impossible to serially change the frequency and FM deviation because either com­mand to go to another range will cause an error as the other parameter is out of range. The programmer would have to cre­ate an intermediate state such as turn FM off before changing the frequency so that all intermedi­ate states were valid. The Performance Signal Generator (PSG) implementation of the IEEE 488.2 standard eliminates this problem because only the final state need be valid. The message of “Freq 200 MHz; FM 8 MHz” would put the instru­ment right to the new state that is wanted.
It is important that the user does not define an ambiguous state within a message by modi­fying the same function more than once in a single message. It is uncertain (and undefined) what the final instrument state would be if the 8645A received the following messages:
Freq:step 10 HZ;:Freq up;:Freq:step 100 HZ;:Freq down
As the frequency is repetitively changed in a single message the final frequency of the instru­ment will depend on the execu­tion order of the commands, which is not defined.
*RST;Freq 123 MHz;FM:State on
In this case the *RST command could be executed after the other commands, canceling their effects. The command *RST defines a complete instrument state by itself and so should be sent alone.
FM:State on;:AM:State on;:Mod:State off;:Freq 100 MHz;:Mod:State on
In this case the user has speci­fied conflicting states for the mod:state command.
If the execution order of a group of commands is important, the user must send a separate mes­sage for each command.
11
Performance related topics
Phase noise performance
The 8645A agile signal generator was designed to minimize the phase noise of its signal at off­sets corresponding to typical channel spacings of communica­tion systems. These offsets of interest are generally greater than 10 kHz from the carrier. Simultaneously, the close-in noise was reduced to assure low residual FM for receiver testing. The following summarizes the phase noise performance of the 8645A.
For offsets less than 100 Hz, the primary contributor of phase noise is the fractional N synthe­sis circuitry in a single phase lock loop if FM is not active. A typical level is –80 dBc (in a 1 Hz noise bandwidth) at 100 Hz offset for carriers in the main band of 515 to 1030 MHz. The phase noise at offsets between 100 Hz and 10 MHz is deter­mined primarily by a frequency discriminator inside a frequency locked loop. The typical phase noise level at a 20 kHz offset is –133 dBc in the main band. Beyond 10 MHz the phase noise is that of the VCO or output sec­tion divider noise floor at approximately –150 dBc. These phase noise levels at offsets less than 10 MHz will decrease by approximately 6 dB each time the carrier frequency is reduced by half due to the dividers in the block diagram. This phase noise reduction continues until the dividers’ noise floor of approxi­mately –150 dBc/Hz is reached.
At offset frequencies of 20 kHz or greater phase noise does not increase when FM is active as long as the deviation used is less than approximately 5% of the maximum FM deviation allowed at that carrier frequency. For example, in the main band of 515 to 1030 MHz the maximum available deviation is 10 MHz, but the phase noise performance at a 20 kHz offset remains the same as in CW operation if 500 kHz or less of FM deviation is set by the user. If the full 10 MHz deviation is used, the phase noise at this 20 kHz offset typi­cally increases by 17 dB to –116 dBc.
As with any signal generator, close-in phase noise of the 8645A goes up as the FM devia­tion increases. This is because the internal FM circuits con­tribute more noise as the devia­tion (gain) increases. For example, the phase noise level is approximately –80 dBc at a 10 Hz offset for a main band output from 515 to 1030 MHz with FM deviation set to 100 Hz. If FM deviation is set to 100 kHz the phase noise at this offset increases by 35 dB to typically –45 dBc. Using the full 10 MHz deviation, the maximum avail­able in this main band, phase noise will go up another 40 dB to approximately –5 dBc at a 10 Hz offset.
12
There is no degradation of phase noise at offsets greater than 10 kHz if special function 120 “Linear DCFM” is activated. At smaller offsets however ,linear FM operation will result in up to 35 dB less noise for operation using high FM deviations. For example, at a 10 Hz offset and with 10 MHz FM deviation set, the phase noise level in the default FM mode is typically –5 dBc while the linear FM level will be at –35 dBc. Similarly for a 1 kHz offset the levels are typi­cally –51 dBc and –77 dBc. At a FM deviation of approximately 20 kHz the level of phase noise is about equal for the two FM modes while at smaller devia­tions the default digitized FM actually exhibits lower phase noise. In general, the linear FM special function may improve phase noise performance at off­sets less than 10 kHz when FM deviations greater than 20 kHz are in use.
In summary, the 8645A will add the least amount of phase noise to the carrier if the lowest FM deviation necessary for the application is used.
Phase noise levels in fast hop operation are degraded approxi­mately 3 dB from non-agile lev­els due to a reduction of filtering on the VCO pre-tune lines. Less filtering is necessary, as the sig­nals on these lines require a higher bandwidth during fast hop operation. The typical level of phase noise for frequency agile signals between 515 and 1030 MHz is –130 dBc at a 20 kHz offset.
8645A signal generator typical phase noise and spurs at 1 GHz
13
Spurious performance
The spurious performance of the 8645A is quite good, but there are still spurs to be found. This product note describes the sources of potential spurs and where in the spectrum they can be found.
The harmonically related spurs are caused by nonlinear opera­tion of amplifiers in the RF path. The specification for harmonics of the carrier for carrier fre­quencies below 1030 MHz is –30 dBc. Typically, they are bet­ter than –35 dBc. Subharmonics typically are caused by a divider in the Phase Lock Loop (PLL) signal path that affects the main VCO output and amplification in the output section. The domi­nant subharmonic is at 0.5 * the VCO frequency. This spur in the main band (515 to 1030 MHz) is less than –70 dBc, in the dou­bled band (1030 to 2060 MHz) it is less than –45 dBc, and below 515 MHz it is practically non existent.
Nonharmonically related spurs are caused by a number of things. These include the power supply, microphonics, and digital circuits. The power supply spurs are all input line frequency relat­ed and are typically less than –60 dBc in the main band. A careful design of the regulators and power distribution circuits keeps the power supply ripple in the instrument very low. The front panel display circuitry can produce a spur at an offset of approximately 1.5 kHz. Its level is less than 20 dB above the sig­nal generator’s phase noise measured in a 1 Hz noise band­width. Microphonics is another source of spurs that depends on how severe the signal generator is being vibrated. One inherent source is the fan, with the loca­tion of the spur dependent on fan speed which in turn is a function of the instrument’s tem­perature. The fan spur is usually less than 20 dB above the noise.
A spur can be produced in the output when external modula­tion (i.e., FM) is enabled and the internal audio source is active. Its location will be at the audio source frequency. The level will depend on the amount of FM deviation programmed. For example, if 1 MHz deviation using an external FM source is set and the internal audio oscil­lator is at 100 kHz, it will cause a spur at approximately –80 dBc. It is recommended that the audio oscillator be turned off when not in use.
One final type of spur to be men­tioned is due to the fractional-N circuitry in the PLL. When the output frequency in the main band is not an integer multiple of 400 kHz, a spur will be pro­duced. This spur is caused by the PLL divider alternating between two divider numbers (integers) such that the average frequency is the desired frequen­cy. Compensation in the PLL cir­cuitry keeps these spurs to less than 25 dB above the phase noise of the output signal (in a 1 Hz noise bandwidth). The frac­tional-N spur frequency in the main band will be at half the dif­ference between the closest inte­ger multiple of 400 kHz and the instrument’s output frequency.
14
Third order intermodulation
Third Order Intermodulation (TOI) products result when the outputs of two signal generators are summed together in a com­bining network. These spurious signals occur at frequencies 2*F1-F2 and 2*F2-F1, where F1 and F2 are the output frequen­cies of signal generators 1 and 2 respectively. The unwanted intermodulation signals are the result of the Automatic Level Control (ALC) loops in the out­put sections of each generator “seeing” the other generator’s signal and responding to it as if it were unwanted modulation on the desired output signal. If the frequency difference between the desired output signal and the other generator’s signal is less than the bandwidth of the ALC loop, the loop can respond to the signal’s presence. In trying to remove this single sided “modu­lation”, the loop inadvertently produces modulation sidebands of its own. This unfortunate process is also occurring in the other generator’s ALC loop at the same time. The overall result is third order intermodulation products accompanying the two test signals at the output of the combiner.
Since signal generator outputs are usually combined to provide a stimulus to test the TOI per­formance of receiver front ends, it is important that the TOI products caused by the signal generators be well below those expected from the device under test. One way to reduce the TOI products from the signal genera­tors is to use a directional cou­pler rather than a resistive summer to combine the two sig­nal generator outputs. Another way to decrease TOI products is to reduce the bandwidths of the signal generator ALC loops well below the frequency spacing used. This approach can be taken to the extreme of entirely opening (or disabling) the ALC loops, since an open loop can be thought of as an infinitely nar­row (0 Hz) bandwidth because the ALC will not respond to a signal at any frequency.
The 8645A has five different ALC loop bandwidths: 200 kHz, 50 kHz, 5 kHz, 60 Hz, and 0 Hz. The instrument automatically selects the optimum bandwidth for lowest AM distortion, fastest amplitude switching speed, and lowest TOI. The three widest bandwidths are used only when AM is enabled. In that case, the bandwidth selected is a function of carrier frequency. The 60 Hz bandwidth is used whenever AM is turned off. At this bandwidth, the TOI level produced by two signal generators with a frequen­cy difference of 25 kHz and with output levels of +8 dBm is typi­cally less than –55 dB. For better performance with smaller fre­quency differences, the 0 Hz bandwidth can be selected using special function 104.
15
Divided outputs below 515 MHz
To create signals below the main VCO frequency range of 515 to 1030 MHz, the 8645A divides the VCO into lower frequency octaves using digital dividers switched into the signal path. A total of 11 divide bands extend the frequency coverage down to 252 kHz. This technique is very good for spectral purity as with each division of the signal the phase noise and spurs are decreased approximately 6 dB. This reduction in phase noise continues until the noise floor of the dividers is reached which is typically approximately –150 dBc. Residual FM is also reduced as the carrier frequency is divid­ed down. However, there are other consequences that must be dealt with as the following describes.
When FM is applied to the VCO the amount of deviation that is present in a divided output is divided by the same number as the carrier. For the lowest or 11th divider band this division equals 211 or a divisor of 2048. This large divisor is one of the reasons the 8645A has a maxi­mum of 10 MHz of FM deviation in the main band. With this large deviation available in main band there is still 10 MHz / 2048 or
4.8 kHz of deviation available in the 252 to 503 kHz band. FM rate also decreases with each successive divide band because each band has 2 half-octave low pass filters present to reduce the level of harmonics at the output.
Phase continuous frequency sweep is also reduced by the action of the divider circuits. The actual frequency change in this function is an FM operation using the full FM deviation avail­able in the main band to get a 20 MHz span (±10 MHz). The available span width is reduced by half with each successive divide in the same way FM deviation is.
The AM bandwidth is always limited to something much less than the carrier frequency because the level detector (which is designed to follow the AM envelope or any level varia­tion and not the RF) would start detecting the RF waveform if the bandwidth were too wide. In a divided output, the AM is applied to the divided RF output (not the main VCO signal) and therefore the AM bandwidth must be less than the band’s lowest RF signal so as not to react to the carrier. But as AM bandwidth is reduced, amplitude switching time gets longer and AM distortion is worse. In the 8645A three AM bandwidths are used to optimize the AM per­formance and still allow fast amplitude transitions as is nec­essary for fast hop operation. Over the frequency range of 128 to 2060 MHz a 100 kHz band­width is used, a 50 kHz band­width is active for signals down to 8 MHz and a 5 kHz bandwidth limits signals for outputs to 252 kHz. In fast hop operation, the amplitude is decreased by approximately 30 dB by the AM circuitry during each frequency change of the main VCO. The reduction of AM bandwidth for the lower divide bands which causes slower amplitude switch­ing time is the only reason the fast hop switching time is longer for low carrier frequencies.
16
Stereo separation quality
Stereo separation is a measure of a receiver’s ability to separate the left and right channel of a stereo signal. To a listener, this is a measure of the receiver’s ability to recreate the spatial impression of a stereo signal. In an FM system the audio informa­tion is received as a left + right signal and a left-right signal. The receiver decodes the left channel by adding the two signals and decodes the right channel by subtracting the two signals. The separation of the channels depends on the cancellation of the right channel during the addition and cancellation of the left channel during subtraction. For this to happen properly, the relative phase and amplitude of the two original signals must be kept equal.
In a signal generator the FM lin­earity and group delay flatness determines whether the relative phase and amplitude of the stereo signal is preserved. Typically, test signals near 1 kHz and 38 kHz are used to modulate the signal generator to test stereo separation in a receiver. The quality of the test signal, and therefore the measurement, will depend on the FM linearity and group delay at these fre­quencies.
Radio manufacturers specify stereo separation in dB as the amplitude difference between a desired signal in one channel and an undesired signal in the other channel. The desired signal is a known test signal used to stimulate one channel. The unde­sired signal is the unwanted “leakage” or coupling of the test signal into the other channel. Radio manufacturers typically specify 40 dB separation which is beyond most listener’s ability to detect distortion or “crosstalk” between the channels.
The 8645A in the fast hop opera­tion uses linear FM and has low group delay. Typical stereo sepa­ration is greater than 55 dB in this mode and is sufficient to test most consumer radio equip­ment. Using digitized FM in stan­dard operation of the 8645A will result in poor stereo separation due to variations in the group delay of the FM signal path. Activating special function 120 linear FM with AC coupling set provides separation similar to the fast hop mode but has the disadvantage that signals below 20 Hz cannot be used. Linear FM with DC coupling also has good stereo separation but poor cen­ter frequency resolution. Therefore, for the best stereo separation with the 8645A the user should activate the fast hop mode with DC coupling.
17
Minimizing fan noise
Increasing concern over the level of audio noise coming from test instruments has resulted in sev­eral design features to minimize noise from the Performance Signal Generator (PSG). The objectionable noise from test instruments comes from the fans used to create the internal air­flow to cool the electronics and prevent heat related failures. These modifications from what was done in the past are related to careful fan selection, fan speed and rear panel fan cover. An overriding consideration was to maintain the high reliability design goal for the PSG by ensur­ing sufficient airflow for cooling components.
A number of fans were evaluated for use in PSG. Along with being of the right physical size and pushing enough air, the noise level when running was consid­ered. Of the fans that would meet the cooling requirements, the one with lowest noise level was chosen. The next step was to evaluate the noise contribu­tion the fan cover was responsi­ble for. The shape of the grill work of the cover changes the noise level due to the fan blades passing close to it in their rota­tion. An analysis of noise sources of various grill shapes with the blade shape of the low noise fan led to making several grills to try out. The combination of the grill and low noise fan that produced the lowest noise was chosen for implementation in PSG.
The final step taken to reduce the noise of the PSG produced the biggest benefit for the aver­age user. Minimum airflow required for high reliability oper­ation is calculated assuming the ambient temperature at the max­imum operating temperature specified. For PSG, this tempera­ture is 55 degrees centigrade (131 degrees fahrenheit). Fan rotation speed is set to provide enough airflow at this high envi­ronmental temperature. At lower temperatures, less airflow is needed to keep internal compo­nents at their specified operat­ing temperature so the fan speed could be reduced. In most previ­ous instruments the fan speed is held constant at the highest air­flow needed for high ambient temperatures. In PSG instru­ments, the fan speed is con­trolled by a temperature sensor to vary airflow as needed to maintain as much as possible constant internal temperature over the full environmental range of 0 to 55 degrees centi­grade. Since the average user has the instrument in environ­ments much less than 55 degrees centigrade, the fan speed is much slower than the maximum it could do. As fan noise is directly related to the fan speed, in typical use the PSG instru­ments are much quieter than previous signal generators. This provides a much more pleasant environment for the operator of a performance signal generator.
18
Frequency agility
Functional description of frequency agile operation
The frequency agile operation of the 8645A is unique in both its capabilities and its operation. The following describes what the instrument is actually doing while in fast hop operation.
Entering the fast hop mode
Either pressing the fast hop mode select key or sending the counterpart HP-SL command will put the instrument in the fast hop synthesis mode. The instrument’s output frequency is no longer phase locked. Instead the frequency accuracy depends on an extremely stable VCO and a frequency locked loop. At this point the fast hop subsystem is set to idle allowing parameters such as output level and FM deviation to be programmed the same as in non-agile operation.
Entering channel information, hop rate, and dwell time
As the user enters each frequen­cy and amplitude into a channel location, the information is put into non-volatile memory for use during the learn operation that precedes frequency hopping. If a channel sequence is entered for the channel numbers, this infor­mation is also put in this memo­ry. A hop rate and dwell time are always in memory and are modified according to any new values entered. Any conflict between the hop rate and dwell time is not checked until the fre­quency learn operation is initiat­ed.
Initiating the learn operation
The learn operation recalls each frequency and amplitude stored in each channel location and sets the phase locked synthesizer and the ALC of the output to each value. At each setting the instrument records the VCO tun­ing voltage and the ALC amplifi­er gain. The output is turned off while this process is underway. The hop rate and dwell are also verified that they will not con­flict for the frequencies (and associated switching time) in the channel table. The only channels that are part of these operations are those in the current sequence table. If the user did not specify a sequence table, the 8645A creates one that reflects the number and order of the entries in the channel table. The instrument does not program frequencies and amplitudes of any channels that are repeated in the sequence table, rather the VCO and ALC settings already learned are copied into memory. The learn operation always lasts a minimum of 10 seconds to ensure that the hopping circuits are exercised sufficiently to sta­bilize any thermal changes in the transition from the idle state. As more unique frequencies are included in the channel table it takes longer to set up each state to record the settings and so the learn time required increases. For 2400 channels, learn time is approximately 1 minute. This time doubles to 2 minutes if FM is active.
Initiating hop operation
When the 8645A begins frequen­cy hopping a unique “fast con­troller” takes control of the VCO and ALC. The data contained in the fast hop memory is present­ed to the hardware to duplicate each channel in the order it appears in the sequence table. Depending on which fast hop mode is active, the fast con­troller may cycle through the sequence table at a programmed rate or enable external inputs to trigger a hop to the next channel or to select which sequence loca­tion to output based on the input at the fast hop bus. The instru­ment will remain in the hop state until the idle or learn oper­ation is selected or a function is changed that would invalidate the data in the fast hop memory (such as changing the FM devia­tion). Rate and dwell can be changed without having to learn again.
19
Faster frequency switching using multiple agile generators
The 8645A can provide frequen­cy agile outputs with hop rates of up to 50,000 hops/second depending on the carrier fre­quencies above 128 MHz, 11,000 hops/second above 8 MHz, and 2,000 hops/second above 252 kHz. While these hop rates are fast enough for the majority of agile applications, sometimes higher hop rates may be needed. The 8645A has been designed to make it easy to syn­chronize and combine the out­puts of multiple units in order to create agile signals at higher hop rates. This product note explains how to configure multiple units to work together.
Frequency agile operations that include interfacing with a radio usually involve control of the fol­lowing parameters: frequency selection, data valid, hop trigger­ing, dwell time, modulation and amplitude. Of these parameters only data valid and hop trigger­ing require extra attention when synchronizing the outputs of two or more 8645A’s. In a typical instrument set-up, each 8645A will have loaded into memory identical channel and sequence tables of all the frequencies and amplitudes to be output. Dwell time for each hop frequency will be constant and controlled by the instrument’s internal timers. The modulation waveform to be placed on the carrier would be input to both external FM inputs with the same FM deviation set on each generator. The frequen­cy control word to select each channel to be output according to its location in the sequence table would also be input to the fast hop bus of both instruments simultaneously. The RF output of each generator would be brought together with a combin­er for input to the device under test. All of these control inputs and instrument settings are identical to that required for operating a single 8645A.
By alternating which signal gen­erator receives the data valid and hop trigger signals it is pos­sible to give one 8645A time to switch frequencies while a sec­ond unit is producing the needed output. Then while the second unit is changing frequencies the first can provide the next out­put. In this way the combined agile output can be switched at much higher rates than are pos­sible with a single 8645A. Each unit gets a trigger signal at half the hop rate of the combined output will be. The data valid input to clock in each frequency word occurs at the same time as the hop trigger but is input to the opposite instrument. Consequently the same trigger signal (approximately +5 volts for 1 usec) can be used for both the hop trigger of one unit and the data valid input of the other unit since the two signals occur simultaneously but are just rout­ed to two different inputs.
Although both generators get the modulating signal, only the gen­erator that is presently out­putting a signal will carry the modulation since the output of the other generator is decreased by over 60 dB while it is chang­ing frequencies. Also the fre­quency word that goes to both fast hop bus inputs is ignored by the generator that does not also get a hop trigger to implement the word.
20
The diagram that follows illus­trates the connections that will alternate outputs from two agile signal generators to produce a hopped signal at rates above 50,000 hops/second for carrier frequencies above 128 MHz. Note that two pulse generators (such as an 8116A), one with a delayable trigger (such as an 8013B), are used to provide a hop trigger alternately to each generator. Frequency selection is controlled by the internal sequence table of each generator so the frequency control word and data valid inputs are not needed. The two outputs are combined using a power splitter such as the 11667A. With this configuration a maximum hop rate of 93 kHz can be produced with a switching time of 4.3 microseconds between channels
and a dwell time of 6.4 microsec­onds per frequency. By increas­ing the dwell time of each agile output the switching time can be reduced. To calculate the com­bined output switching time, it is only necessary to subtract the dwell time from the 8645A’s specified switching time for the carrier frequency range in use and divide by two. For example, if the signals were at carrier fre­quencies below 8 MHz with a specified switching time of 500 usec and the desired dwell time was 300 usec, the switching time of the combined output would be (500 usec minus 300 usec)/2 or 100 usec. These sig­nals would have the same speci­fications of ±2 ppm frequency accuracy and ±1 dB amplitude accuracy as an 8645A operating by itself would have.
21
Multiple signal generators provide faster frequency switching
Pulse generator
Delayed trigger
HOP trigger
8645A 8645A
Data
source
FM RF FM RF
Output
Pulse generator with delayable trigger
HOP
trigger
Power combiner
Frequency accuracy of agile outputs
The frequency accuracy and sta­bility of the 8645A is directly related to the 10 MHz timebase used as a reference for non-agile operation. The output accuracy is a direct multiple of the time­base error. For the high stability timebase specified at 0.0005 parts per million (ppm) aging rate per day, the worst case error of a 1 GHz output after 10 days would be 5 Hz assuming no initial inaccuracy.
For frequency agile outputs, the 8645A specifies a maximum error of ±2 ppm. There are two contributors to the frequency error: timebase error and tem­perature related drift. Timebase error is a factor because during the “learn” operation the 8645A briefly synthesizes each output frequency using the phase lock loop circuits and reads voltage levels of the VCO tuning line. While hopping this tuning volt­age is sent back to the VCO to create the output signal very rapidly. Any error of the time­base will be reflected in the tune voltage sent to the VCO used for fast hop signals.
Temperature related frequency errors result from a change in the operating temperature of the components in the agile signal path. Several steps have been taken to reduce the temperature variations within the instrument such as providing constant-tem­perature heating of the delay line and temperature regulating the fan speed that provides cool­ing. These design features and the large thermal mass of the instrument greatly reduce the sensitivity of the 8645A’s agile frequency accuracy to ambient temperature changes. In any case, it is recommended that the user re-learn the hop frequen­cies before beginning a frequen­cy agile test. Each learn operation will remove any tem­perature related offset between the fully synthesized calibrating signal and the agile output. Also, as noted in the specifications table, having the unit plugged-in for 24 hours and operating for a minimum of 2 hours before the learn operation and frequency hopping begin, will ensure the heating elements and the ther­mal mass of the instrument are at a stable operating tempera­ture.
The typical worst case frequency error of ±1 ppm for agile fre­quency outputs can be signifi­cantly improved if the test parameters for the application are within a certain criteria that minimizes the minute thermal variation of the agile compo­nents themselves. For example, tests conducted during the 8645A’s design show that the frequency error is reduced for agile test sequences of less than approximately 60 unique fre­quencies with hop rates of greater than 10 hops/second. The same improvement occurs for internally controlled sequences having up to the max­imum number of 2400 unique frequencies as long as the distribution of how often each frequency is output is pseudo­random. The average frequency error measured under these con­ditions was less than ±0.3 ppm.
22
Relating phase error and frequency accuracy
The quality of a signal can be specified in many ways including amplitude accuracy, spectral purity, output level, modulation distortion, etc. A common speci­fication of many sources is fre­quency accuracy or phase error. While many signal generators are specified in terms of frequency accuracy, other types of sources are just as commonly described with a phase error. Since the sig­nal generator is often used to simulate or substitute for a source specified in terms of phase error, it becomes neces­sary to convert between frequen­cy accuracy and phase error to determine if performance is suf­ficient. This calculation is of crit­ical importance for frequency agile sources that are specified to be within a stated frequency accuracy or phase error in a cer­tain amount of time after a fre­quency change is triggered. This product note discusses the con­version between phase error and frequency accuracy.
Converting a phase error specifi­cation to a frequency accuracy number is based on a key assumption: that the phase set­tling characteristics are approxi­mately linear. Since the rate of phase change of a signal is relat­ed to its frequency, knowing that a source’s output is uniformly approaching the desired final phase state allows a calculation of the corresponding frequency change. If the phase settling departs significantly from linear, the calculated frequency error will be too low for faster phase settling and too high for slower settling signals. The following examples use typical characteris­tics for a frequency agile local oscillator to illustrate the phase error to frequency accuracy con­version process.
Example 1: Converting phase error to frequency accuracy
A local oscillator switching in the range of 800 to 1000 MHz is specified to be within 0.1 radian of final phase 20 usec after the frequency change trigger is received. The signal settles to the final phase during the 5 usec duration of the output. Converting the accumulated phase error over the 5 usec duration results in a calculation of (0.1 radians)/(5 usec) equal­ing 20,000 radians/second that the signal is changing. This equals a frequency error of (20,000 rad/sec)/(2*pi rad/sec) or 3183 Hz. The frequency accu­racy of the 8645A is specified at ±2 ppm of the carrier which translates to a maximum of 2 kHz frequency error for the 800 to 1000 MHz frequency range. In this application the 8645A can substitute for the local oscillator as far as the frequency accuracy requirement is concerned.
If in the previous example the signal settled to its final phase faster than 5 usec, the calculated frequency error would be corre­spondingly higher. Another example follows which converts the frequency accuracy of the 8645A to its corresponding phase error for the previous example.
Example 2: Converting frequency accuracy to phase error
The frequency accuracy of the 8645A is specified at +2 ppm of the carrier frequency. For a 1 GHz output this equals a maximum error of 2 kHz. Converting 2 kHz to a phase error equals (2000 Hz)*(2*pi rad/sec) or 12,566 radians/sec. In the 5 usec dura­tion of the signal described in example 1, the maximum accu­mulated phase error is (12566 rad/sec)*(5 usec) or 0.0628 radians.
If the phase error is specified in terms of degrees the values given in radians in the above examples should be multiplied by (2*pi). In example 2 the 8645A’s phase error of 0.0628 radians translates to 0.3946 degrees.
23
Amplitude dynamic range while frequency hopping
The available output level dynamic range in fast hop mode is a function of the amplitude switching time required. The fre­quency switching time of the 8645A is not controllable by the user and is always typically 9 microseconds, the speed of the fundamental VCO. The specified switching times in the data sheet are actually amplitude switching times, that is the time it takes for the power to rise to 90% of its final value at the new fre­quency. This “power rise time” is also not directly controllable by the operator, but he can effect it by utilizing the amplitude hop­ping capability of the 8645A.
The 8645A hops with the ALC loop closed unless the operator selects [open loop hop “on”], special 202. For [open loop hop “on”], amplitude switching time does not degrade as a function of amplitude range utilized, but level accuracy does degrade. For [open loop hop “off”], special 202, the default mode, the avail­able amplitude range is a func­tion of the amplitude switching time required. This range can be used for either programming dif­ferent levels or for shaping the power envelope (using the AM port) while hopping. Both capa­bilities can be used simultane­ously as long as the total range is within the limit described in the graph below. Note that the amplitude accuracy is also a function of the range utilization and ALC special selected. The level accuracy degradation will occur at the lower amplitude output.
The typical, quantifiable, degra­dation in amplitude switching time and level accuracy is described in the graph below. Amplitude variations greater than 20 dB may be programmed, but amplitude inaccuracy could become much greater than 5 dB in some cases.
The reason for these interrela­tionships is, the same ALC loop is used simultaneously for DC coupled AM and level setting. To provide specified 90% AM with adequate design margin, 30 dB of ALC range is required at the worst case frequencies. The instrument was designed to give the user the option of using this “excess” range to control level while fast hopping. At the lower portions of this “excess” ALC range, the ALC loop bandwidth will decrease, increasing the amplitude settling time.
24
Amplitude shaping of agile outputs
As part of their frequency switching algorithms, many fre­quency agile radios reduce the RF carrier power when switch­ing between frequencies. Since controlling the characteristics of the amplitude transitions while switching is critical to proper hopped operation, the 8645A was designed to allow the user to shape the amplitude transi­tions of the RF carrier while in fast hop operation.
If the amplitude transitions are very sharp in nature, a frequen­cy agile carrier (when viewed at one specific frequency) will have the same sin (X/X) spectral sig­nature as a pulsed RF carrier. As with a pulsed RF carrier, the energy of the carrier will be dis­tributed throughout the lobes of the sin (X/X) envelope. This energy distribution is typically of no concern in pulsed applica­tions but in frequency agile applications the distributed energy can fall into adjacent communication channels, caus­ing disruption of communica­tions in those channels. To avoid this problem, frequency agile radios that reduce the RF carrier power when switching also shape the RF power transitions to minimize the spectral splatter associated with sharp (pulse like) amplitude transitions.
When in fast hop operation, the 8645A automatically ‘softens’ the amplitude transitions to decrease the spectral splatter. As power is being shut off at a spe­cific hop frequency, timers built into the instrument send a nega­tive step to the ALC loop so that the amplitude drops at a rate that is consistent with the ALC loop bandwidth. After several microseconds the ALC loop will have decreased the output ampli­tude by approximately 30 dB. At this point the pulse modulator is activated to get an additional 35 dB or more of amplitude decrease. This timing sequence is reversed when the power is being brought up at the new fre­quency, with the pulse modula­tor being turned off first and then the ALC loop being allowed to return to its pre-shutoff level. This sequence greatly reduces the spectral splatter from what it would be if only the pulse modu­lator were used. If the user acti­vates special function 202 ALC off, to get slightly faster frequen­cy switching speed, the ALC is not used to decrease the output power. Only the pulse modulator would be used so spectral splat­ter increases somewhat due to the more abrupt amplitude tran­sition. Additionally the power is only decreased approximately 35 dB between frequency hops.
The operator that would like to use the 8645A to emulate a transmitter that uses a rigorous amplitude shaping technique, or who needs to decrease spectral splatter for other reasons, can use the external DC AM port to shape the amplitude transitions during fast hop operation. The shaping signal, such as a raised cosine wave, when input into the external DC AM port controls the ALC loop to implement the amplitude shaping. To shape the amplitude rise and fall charac­teristics of a hopped signal, the shaping signal must be synchro­nized to the hop trigger or dwell time control (available on the rear panel of the 8645A). It should be noted that external shaping, as well as the automatic shaping previously described, use some of the available ALC range. This means that the amount of amplitude variation available while hopping is decreased. Elsewhere in this product note the interaction between amplitude hopping range, frequency switching speed, and amplitude shaping is discussed as “Agile Amplitude Dynamic Range”.
25
Modulation
High rate, high deviation frequency modulation
The 8645A has overcome many of the previous barriers to pro­vide FM with high deviation in an RF signal generator. By using digital techniques, deviations as high as 10 MHz are possible for carrier frequencies in the main carrier band of 515 to 1030 MHz. Accompanying this breakthrough in high deviation FM is a similar increase in the maximum rate of FM to 10 MHz. One reason for this extra FM performance is to ensure sufficient usable capabili­ty after the many divider stages that extend frequency coverage down to the minimum carrier output of 252 kHz.
In the 8645A, the amount of FM deviation selected determines the length of the delay line used in the delay line discriminator placed around the VCO to reduce phase noise. At higher selected FM deviations the delay line is shortened so that it does not reduce FM sensitivity in the deviation range that is set. With a shorter delay line the phase noise from the VCO increases, but generally this is acceptable for applications needing high FM deviations. The annunciator for the mode 1 key on the front panel lights up when the shorter delay line is in use. As FM devia­tion is reduced below approxi­mately 17% (1.76 MHz in the main band) of the maximum allowed at each carrier band, the extra delay line is automatically placed in the VCO signal path unless the user has specifically locked the instrument into mode 1 using the mode select keys. Phase noise and spurs are reduced for operation using the smaller FM deviation and the mode 2 indicator is lit. In this way, the best spectral purity is provided for any FM deviation that is selected.
The 8645A has two different types of FM referred to as digi­tized FM and linear FM, both of which can be used with the internal modulation oscillator and/or an external source at the front panel. Digitized FM is the default type and utilizes an A/D converter to translate the modu­lating waveform into digital information that is used to mod­ulate the fractional-N divider
number. This provides FM at rates from DC up to the phase locked loop bandwidth along with the capability for high devi­ations. A wideband, high slew rate analog path sums the modu­lation signal onto the VCO tune line to allow FM at rates from the PLL bandwidth up to
3.75 MHz and typically to 10 MHz while retaining the high deviations.
Linear FM is activated using spe­cial function 120. In this FM operation the digital path is switched out leaving just the analog path for improved flat­ness and stereo separation. In mode 2 with linear FM, the PLL is not used and linear DCFM is available through the frequency locked loop. During calibration, offsets are nulled in this path to improve frequency accuracy in fast hop operation which always uses linear FM. Group delay is less than 1 usec in linear FM with a typical value of 0.1 usec.
The maximum deviation avail­able is the divided result of the amount available in the main band of 515 to 1030 MHz. The maximum of 10 MHz deviation in this main band becomes 5 MHz in the first divide band of 207.5 to 515 MHz. The minimum devia­tion that can be set is 100 Hz in the main band and also gets halved by each divider band to a minimum of 1 Hz. The maximum available FM rate is also reduced by each divider section due to the half-octave filters present in each divider section to reduce the level of harmonics generated.
26
Simultaneous modulation
The 8645A has a wide range of combinations available for simul­taneous modulation. Various combinations of amplitude, fre­quency, phase and pulse modula­tion are provided as follows:
With AM: FM, phase, pulse, FM
and pulse, phase and pulse
With FM: AM, pulse, AM and
pulse
With phase: AM, pulse, AM and
pulse
With pulse: AM, FM, phase, AM
and FM, AM and phase
In addition to these combina­tions, the modulating waveform can be provided from either the internal modulation source or from an external source via a front panel input The internal/ external status of a given modu­lation type can be set independ­ent from any other modulation type that may also be active. The following combinations of inter­nal and external modulation waveform source are available for each modulation:
AM: Internal, external FM: Internal, external, internal
and external
Phase: Internal, external,
internal and external
Pulse: External only
In simultaneous internal and external FM or phase modula­tion, an external signal of typi­cally 30% or more of full scale input can be applied simultane­ously with a full scale internal signal without any limiting occurring.
When using the internal modula­tion source, rates from .1 Hz to 400 kHz are available with a res­olution of .1 Hz. Therefore the entire bandwidth of AM (100 kHz for carrier frequencies greater than 128 MHz) and phase modulation (150 Hz) as well as a major portion of the FM bandwidth can be covered with the internal modulation source. The level of the internal modulation signal can be adjust­ed with a resolution of .2% of full scale to provide improved reso­lution of level of various modula­tion types. The internal modulation source can also pro­vide complex waveforms such as sawtooth and squarewave at rates up to 50 kHz and white Gaussian noise of constant amplitude from .1 Hz to 400 kHz.
The external modulation source input of all modulation types except pulse can be set to AC coupling as an alternative to DC coupling. In AC coupling all DC drifts and biases up to ±10 volts from external sources are blocked without degradation of performance. The lower 3 dB bandwidth in AC coupling for all modulation types is typically 20 Hz.
The input impedance of all the external modulation ports is 600 ohms, except the FM port. The external FM input imped­ance is 50 ohms to allow exter­nal modulation sources to provide signals up to 10 MHz with low loss. By activating spe­cial function 123 the user can route modulation signals from the phase modulation input to the FM circuitry. Since the phase modulation input impedance is 600 ohms, this capability pro­vides higher input impedance for FM operation. The upper 3 dB FM bandwidth using this phase modulation input is approxi­mately 2 MHz.
27
Digitized FM operation
Synthesized signal generators have traditionally generated FM with a phase lock loop dedicated for this purpose. For AC coupled FM, this loop remains locked while the frequency modulating signal is injected into the loop to FM the output. For DC coupled FM, the loop must be unlocked so that low modulating frequen­cies are not canceled by the loop feedback. However, unlocking the loop for DC FM allows the output frequency to drift, which can be a problem for some appli­cations. The 8645A has this tra­ditional implementation of FM operation as a special function (120). The standard FM imple­mentation for the 8645A is one that removes the frequency drift problem of DC FM.
As standard operation, the 8645A uses a digital FM signal to modify the instantaneous syn­thesizer divide values in the fractional-N circuit. This method greatly reduces the frequency drift and offset that is usually associated with DC FM opera­tion. The only remaining drift is related to the Analog to Digital Converter (ADC) which is typi­cally much less than the drift of an open loop VCO. This modula­tion technique is also very accu­rate at low FM rates because the output of an ADC can be much more precise than the analog tuning curve of a VCO. A side benefit of major importance is that this technique does not introduce out-of-channel spurs. In a traditional signal generator block diagram, the FM loop needs to sum into the main loop with the unavoidable introduc­tion of crossing spurs.
There are some disadvantages to this digitized FM technique. One is that the digitization and sum­mation of the modulating signal takes time, resulting in approxi­mately 30 usec of group delay of the modulating signal at rates inside the PLL bandwidth. The non-flat group delay may cause distortion of the modulating sig­nal, and in some instances may actually cause the PLL to unlock. Another implication of this mod­ulation scheme is that when digi­tal FM is being used in a feedback loop, the quantization steps of the ADC may cause phase discontinuities. For these reasons some applications may require use of the traditional FM implementation available as spe­cial function 120.
The following describes the per­formance that results from acti­vating each type of FM in the 8645A.
Center frequency accuracy and temperature stability:
For the digitized FM mode, the initial center frequency accuracy is typically 0.1% of the FM devia­tion set (typically 1% if in fast hop mode). Non-digitized or lin­ear FM (special 120) has an off­set of typically 1 kHz for any deviation in the 515 to 1030 MHz carrier band (offset divides for lower carrier bands.) The carrier frequency temperature drift with digitized FM active is typically less than 0.1% of the set devia­tion over the full operating range of 0 to 55 ˚C. The center fre­quency accuracy in linear FM will vary approximately 1 kHz/ ˚C in the 515 to 1030 MHz carri­er range.
Phase noise variations at low offsets:
For deviations less than 5% of the maximum deviation available at any carrier frequency, digi­tized FM operation negligibly affects the phase noise perform­ance of the output. The noise increases with higher FM devia­tions. For linear FM, the phase noise at low offsets goes up by 20 dB when activated but does not increase as much for higher deviations as it does in digitized FM. In general, special 120 linear FM may improve phase noise performance over the digitized FM at offsets less than 10 kHz when deviations greater than 20 kHz are in use. The topic “Phase Noise Performance” in this product note has more information on this subject.
FM deviation accuracy:
For modulation rates <1 kHz, deviation accuracy for the digi­tized FM is a function of the ADC, typically <1% of the set deviation. For linear FM it is approximately 5%. At higher rates, the deviation accuracy is dependent on analog factors which makes it the same for either FM technique.
Square wave or digital modulation waveforms:
For linear FM the group delay is very flat and typical of other sig­nal generators. The group delay of digitized DC FM is a function of the modulating rate. For sin­gle-tone modulation signals or if the spectral energy is primarily below 10 kHz, group delay in dig­itized FM operation is constant and will not affect the output signal. For high rate or digital modulating signals, the variable group delay of digitized FM could cause serious distortion so linear DC FM should be selected.
28
AC coupled FM
This product note builds on the information given in the digi­tized FM operation product note to explain the operation and resulting performance of AC Coupled FM (ACFM).
The 8645A has digitized and lin­ear ACFM capabilities. Digitized ACFM is simply AC coupling of the digitized DCFM described in the “Digitized FM Operation” product note. The default digi­tized DC Coupled FM (DCFM) utilizes an Analog-to-Digital Converter (ADC) to digitize the incoming modulating signal as it occurs and uses the digitized information to modify the syn­thesis dividers. Low frequency 3 dB bandwidth of digitized
ACFM is approximately 7 Hz. This function should satisfy most applications if the slight frequency inaccuracy of approxi­mately .1% of the programmed deviation can be tolerated. Care must be exercised that the com­plexity of the modulating wave­form does not cause unlocking or distortion problems as explained in the “Digitized or Linear FM” product note. Due to the very low drift of the digitized DCFM operation in the 8645A, there is less use for digitized ACFM. It should be noted that the 50 ohm input impedance to ground of the external FM input will be still there for ACFM signals.
The 8645A also offers special function 120 to get linear ACFM. This mode is more typical of the ACFM in other signal generators with the modulating signal input directly to the VCO except that the phase lock loop that is mod­ulated has a much wider band­width than is typical. Because of this, the 3 dB low frequency cor­ner is either 300 Hz or 3 kHz, depending on whether mode 2 or 1 respectively of the mode select keys is lit. If the user can toler­ate this relatively high 3 dB fre­quency corner, the benefits this operation includes minimal group delay and precise center frequency. The applications where linear ACFM may be use­ful are when any frequency inac­curacy can’t be tolerated, or the group delay characteristics of the digitized FM can’t be tolerat­ed, or unlocking occurs due to square waves at the FM port.
29
Special Capabilities
Tailored operation through special functions
In addition to the features directly available to a user through the front panel keys, the 8645A has a number of capabili­ties called “special functions”. These functions are accessed with the [Special] key in the util­ity field on the front panel and allow a user to customize opera­tion of the instrument for a spe­cific application.
Special functions or “specials” are functionally grouped by number as follows:
To find a particular special in the list, the user can scroll through the list by pressing [Special] and either turning the knob or press­ing the up or down arrow keys. When the display shows the desired special, pressing [Enter] will access that special. A user may also press [Special] followed by the number of the special and [Enter] to access it. Pressing [Special] [Enter] will re-access the last special displayed. Once a special is accessed, it can be con­trolled with the same keys used for control of standard functions.
When displaying a special, the light above the special key will be lit if it is not at its default set­ting (the special is active). While scrolling through the specials as described above, the light will also come on when displaying the name of an active special. When no special is being dis­played and standard functions are being used or displayed, this light will be lit if there are any specials active. Pressing the [Display] key followed by the [Special] key will show a list of all active specials.
30
100-109: Amplitude & ALC functions 110-119: Carrier control (phase, frequency multiplier, sweep mode) 120-129: FM functions 130-159: Audio source control 160-169: Frequency reference 170-189: Tests, calibration, security, volt/power meter functions. 190-199: Serial number, display control 200-210: Fast hop functions
Protecting classified instrument settings
The 8645A incorporates a num­ber of functions to prevent the unauthorized exposure of classi­fied instrument settings and readouts. They range from blank­ing the displayed readouts, to an automatic memory erasure if power to the instrument is inter­rupted.
The simplest function which has been present on many instru­ments for some time now is dis­play blanking. With the 8645A, the entire display or specific functional readouts such as fre­quency or modulation can be blanked either from the front panel by activating special func­tions or using HP-SL commands via GPIB. This blanking includes all annunciator lights on the function specific keys as well as the alpha-numeric display. This will prevent the casual observer from seeing the instrument set­tings. To prevent an operator from unblanking the display by deactivating the special function another security-related capabil­ity, the local lockout can be initi­ated over GPIB. With local lockout active the instrument
keyboard will not respond to the user until the controller removes the local lockout over the GPIB or until power is cycled. If power is cycled the machine returns to whatever state it was in before power was turned off, except that local lockout will be dis­abled.
A function called RAM wipe was created to allow the user to erase all user entered parame­ters and operation specific cali­bration data from the internal RAM. When this special function 172 is activated, a power-on reset is performed on the instru­ment. All RAM locations are cleared and tested with checker­board patterns, and then cleared again. A side effect of this opera­tion is that the instrument must spend a few minutes re-calibrat­ing itself in order to restore the calibration coefficients in RAM. All storage registers are cleared, all fast hop sequence and chan­nel data are also cleared. This feature satisfies the require­ments set forth in Mil Std 380-380.
The three functions described so far, blanking, local lockout, and RAM wipe, can be used in com­bination with the security mode function for the highest level of security. Security mode special function 173, is like a one way operation in that no reduction of security is allowed, only increas­es. Activating security mode sets a flag in the instrument firmware such that all or part of the display can be blanked but not unblanked. If an instrument preset is performed, the instru­ment will remain in its current state of operation, with displays blanked and keyboard disabled if previously set, and the security mode still active. Security mode can be deactivated by turning special function 173 off or through an GPIB command. However, leaving security mode either way automatically and immediately initiates a RAM wipe operation. If power is cycled with security mode active a RAM wipe will be performed immediately when power is restored and the security mode will be deactivated. With securi­ty mode active the user can still gain control of the instrument, but not with any classified data parameters still present on the display or in memory.
31
Storage registers and sequential recall
Storage registers are used to store the current state of the machine for later recall. There are 50 storage registers provided in the 8645A. The storage regis­ters use non-volatile memory so power interruptions won’t affect the contents. Instrument preset has no effect on the storage reg­isters, but activating the RAM wipe special function to remove all user-entered data from the instrument will erase them. Storage registers 0 through 9 store the entire machine state except for the fast hop channel data. (The fast hop channel information is held in non-volatile memory also, so power interrupts won’t force a reload.) Registers 10 through 49 store only the current frequency and amplitude (frequency off­sets/multipliers or amplitude off­sets are not included). Any of the 50 storage registers can be recalled explicitly by specifying the register number.
Machine states are saved using the front panel keys by pressing [Save] <register number> (Enter]. To recall a register, press [Recall] <register number> [Enter]. All of the storage regis­ters can be erased at once by pressing [Clear all] [Enter].
Registers can also be recalled in a user-defined sequence that has the instrument repetitively step­ping from one register to anoth­er. Registers 0 through 9 can be included in the sequence list that can be up to 10 registers long including any repeated reg­isters. A sequence is entered by pressing [Set seq] <first register number.> [Enter] <second regis­ter number> [Enter] ... continue to ... <last register number> [Enter]. There are several ways to recall registers from a sequence list. To repetitively recall the entire sequence, press [Auto seq]. Pressing [Off] stops
this sequentially recall. To recall each sequence entry individual­ly, either press [Seq] on the front panel or provide a +5 volt trigger signal to the Seq input on the rear panel. The entire sequence list in memory can be displayed by pressing [Display] [Seq].
Each of the functions described above has its counterpart HP-SL command for control over GPIB.
32
Offsets and multipliers of frequency and amplitude
The 8645A has several features which allow the user to stream­line use of the instrument in applications where the frequen­cy or amplitude of the instru­ment is scaled or offset by the user. This change of amplitude or frequency may be due to an amplifier, attenuator, cable, mixer, multiplier, divider, etc. For some applications, it may be more useful to control and dis­play the frequency and ampli­tude in terms of a device’s output rather than the RF out­put of the signal generator.
Amplitude offset allows the user to program into the 8645A the amount of loss or gain in the connecting device. For example, to control and display the ampli­tude at a device’s input after a connecting cable loss of 1.2 dB, the user would press [Amptd ofs] <–1.2> [dB]. If the display previously read 0 dBm it will now read –1.2 dBm with an additional indicator “Offset” showing that an offset is in use. An offset to reflect the output of a 20 dB amplifier would be entered [Amptd ofs] <20> [dB]. The 0 dBm display would change to +20 dBm by these entries. All subsequent amplitude settings entered into the instrument will be displayed according to the new amplitude reference point. For example, entering an ampli­tude value of +5 dBm into the 8645A with a +20 dB offset active to reflect the output of an amplifier would actually produce –15 dBm at the RF output of the signal generator. In all cases, losses are entered as negative numbers and gains as positive entries. The maximum amplitude offset range is ±50 dB. Pressing [Amptd ofs] [Off] turns off an amplitude offset.
Frequency offset allows the user to enter a frequency shift between the RF output and the display of the 8645A to ref lect the result of the signal passing through an external device. The operation of a frequency offset is similar to that of an amplitude offset. For example, if the user wants to produce an output that will be translated by a mixing process upward by 2 GHz and have the display reflect the result of the frequency transla­tion, pressing [Freq ofs] <2> [GHz] will do it. All subse­quent frequency entries will reflect the offset, ie. entering a 3 GHz frequency setting with the 2 GHz offset active will result in a 1 GHz signal coming from the RF output of the signal genera­tor. A display that previously read 1 GHz will then read 3 GHz to reflect the output of the mix­ing process and the “Offset” indi­cator will be lit. Negative entries can be used to shift the RF out­put for a down-converting process. Frequency offsets are limited to ±50 GHz. To turn off a frequency offset, press [Freq ofs] [Off].
Frequency multiplier allows the user to enter a multiplier or divi­sor to modify the frequency out­put of the 8645A to reflect the action of a multiplier or divider connected to the signal genera­tor. Frequency multipliers are entered using special function
111. For example, if an external device multiplies the signal by four, then pressing [Special] <111> [Enter] <4> [Enter] will produce an RF output that when
multiplied by four will equal the entered frequency setting. The frequency display will show the frequency at the output of the multiplier. An “Offset” indicator is lit below the frequency read­out. Any subsequent frequency entries will be in terms of the entered multiplier and not the actual output of the generator, ie. entering 1 GHz with a multi­plier of 4 active will produce a 250 MHz signal at the RF output for multiplication to 1 GHz. Similarly, negative entries reflect a divider’s action. A multiplier range of ±10 can be entered and only integer values are allowed. To disable the multiplier func­tion, press [Special] <111> [Enter] <1> [Enter] which resets the frequency display for a mul­tiple of 1.
When using a frequency multipli­er, the FM deviation is scaled in the same way as the carrier fre­quency since an external multi­plier or divider also multiplies or divides the amount of FM deviation. Therefore, if a multi­plier of four was active in the 8645A, an entered FM deviation of 10 MHz would actually pro­duce a 2.5 MHz deviation on the actual output of the signal gener­ator and a displayed value of 10 MHz. Frequency offsets have no effect on FM.
Offsets and multipliers may be used simultaneously. When using both a frequency multipli­er and a frequency offset, the displayed frequency equals (the output frequency) times (the multiplier) plus (the offset).
33
Built-in calibration functions
The 8645A was designed with no manual internal adjustments. All adjustments are under the con­trol of the instrument firmware responding to a variety of sen­sors placed at strategic locations to monitor signal levels. This allows the instrument to adjust its signal parameters for an opti­mum output whenever the user wants. With this capability, the instrument can effectively cali­brate itself by making adjust­ments to signal levels and recording the settings that result in the optimum and specified output signal. This calibration data is stored in battery back-up RAM so that the user will not have to repeat the calibration whenever the instrument is turned on. The self-calibration process typically takes approxi­mately 2 minutes.
If the user suspects that the instrument is not performing correctly, a re-calibration of the instrument may be initiated by pressing [Special] <171> [Enter] [On] from the front panel or by sending the “*CAL?” HP-SL com­mon query over GPIB. The instrument will display a result code of 0 if no errors occurred during the calibration operation.
The internal temperature is con­stantly monitored by the instru­ment to insure the optimum calibrated output. Since portions of the calibration data are tem­perature sensitive, the instru­ment will store the current temperature when calibrating along with the calibration data. If the temperature changes enough (approximately 10 degrees centigrade) to potential­ly invalidate the calibration data, a warning will be placed in the message queue to notify the user. If the user feels that the instrument performance is not satisfactory, the instrument may be re-calibrated at the new tem­perature using the special func­tion.
Under certain conditions the cal­ibration data may be lost or become corrupt, such as due to the following events:
1. Disabling the RAM using a switch on the Digital Control Unit (DCU).
2. Executing the RAM wipe special function.
3. Turning off the security function (which executes the RAM wipe function).
4. Failing a self-calibration or diagnostic test.
5. Removing the DCU from the motherboard.
If the instrument powers up without valid calibration data, it will automatically initiate a self-calibration. The exception to this is when the service mode switch on the DCU is enabled to prevent nuisance calibrations while working on an incomplete instrument.
34
Finding failures with internal diagnostics
The 8645A contains diagnostic routines in the instrument firmware which test the instru­ment hardware and will report over 90% of all instrument fail­ures that could occur. These diagnostic routines will also iso­late the cause of the failure to a replaceable hardware module or a cable. The diagnostic routines are activated from the front panel using an instrument spe­cial function or via GPIB com­mands and only run when initiated by the user.
There is separate circuitry built into the instrument that continu­ally monitors high-level function­ality (such as the phase lock loop being locked) and will put a message in the instrument mes­sage buffer should a problem be detected. However, this message is not specific enough to deter­mine which module is the cause of the problem. When the user reads the message, the diagnos­tic routines can be activated to determine which hardware mod­ule is at fault.
There is special hardware in the instrument that makes these diagnostics possible. The digital controller contains a voltmeter with AC and DC measurement capability and each of the hard­ware modules contains at least one 8 channel multiplexer through which critical points on the modules can be measured by the voltmeter. A typical diagnos­tic routine would set a module to a specific operating mode and then set the multiplexer on that module so a critical point on the module can be measured. The routine would compare the volt­meter reading with the normal value. If the reading is not with­in preset limits, the routine will terminate testing and a result code is displayed on the front panel of the instrument indicat­ing where in the test sequence the failure occurred. If the read­ing is normal, the routine contin­ues to set the module to all its operating conditions and meas­ures critical circuit points for each set of conditions. A typical routine makes 100 measure­ments on a module.
The diagnostic routines can find almost all-functional failures, however, the diagnostics are lim­ited in their ability to find fail-
ures where a performance value is just out of spec. The internal voltmeter has an accuracy of approximately 2% and this is not accurate enough to measure some internal signals to deter­mine if the specification is being met. However, these ‘out of spec’ problems are historically a small part of total instrument failures. Another limitation is there are usually only 8 points on a mod­ule where measurements can be made. These 8 points were selected to maximize the amount of circuitry that can be tested, but there is still some circuitry on most modules that cannot be tested. In general the diagnostics will not find problems with phase noise, spurs, output level accuracy, high deviation FM and fast switching.
The service documentation avail­able with the PSG products con­sists of a diagnostics manual and a service manual. The diagnos­tics manual is needed to use the diagnostics. Its purpose is to guide the user to finding the bad module and replacing it with another module. The service manual contains schematic dia­grams and uses the result of the diagnostics to guide the user in repairing the bad module to the bad component.
35
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent's overall support policy: "Our Promise" and "Your Advantage."
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifica­tions and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product opera­tion, and provide basic measurement assistance for the use of specified capabil­ities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can pur­chase according to your unique technical and business needs. Solve problems effi­ciently and gain a competitive edge by contracting with us for calibration, extra­cost upgrades, out-of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional engi­neering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, opti­mize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
By internet, phone, or fax, get assistance with all your test & measurement needs
Online assistance:
www.agilent.com/find/assist
Phone or Fax United States:
(tel) 1 800 452 4844
Canada:
(tel) 1 877 894 4414 (fax) (905) 282-6495
China:
(tel) 800-810-0189 (fax) 1-0800-650-0121
Europe:
(tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (82-2) 2004-5004 (fax) (82-2) 2004-5115
Latin America:
(tel) (305) 269 7500 (fax) (305) 269 7599
Taiwan:
(tel) 080-004-7866 (fax) (886-2) 2545-6723
Other Asia Pacific Countries:
(tel) (65) 375-8100 (fax) (65) 836-0252 Email: tm_asia@agilent.com
Product specifications and descriptions in this document subject to change with­out notice.
© Agilent Technologies, Inc. 2001 Printed in USA July 18, 2001 5951-6712
Loading...