Agilent 8563E User’s and Service Guide

$JLOHQW7HFKQRORJLHV (2SWLRQ. 8VHU·VDQG6HUYLFH*XLGH
Agilent Technologies
Model 8563E Option K35
User’s and Service Guide
Use this manual with the following documents:
Agilent 8563E CDMA Test Set User’s Guide
Agilent Technologies 8560 E-Series Firmware Upgrade Kit Installation Note
Agilent Part Number: 08563-90235
Printed in USA April 2001
Notice
The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incide ntal or conseq uentia l damages in connect ion with the furnishing, performance, or use of this material.
Agilent Technologies assumes no responsibility for the use or reliability of its software on equipment that is not furnished by Agilent Technologies.
Restricted R ig hts Legend
Use, duplication, or disclosure by the U.S. Government is subject to rest ri ct ions as set forth in subparagraph (c)(1)(ii) of the Righ ts in Technical Data and Computer Software clause at DFARS 252.227-7013 for DOD agencies, and subparagraphs (c)(1) and (c)(2) of the Commercia l Computer Software Restricted Rights cl ause at FAR 52.227-19 for other agencies.
Agilent Technologies 1212 Valley House Drive Rohnert Park, CA 94928-4999, U.S.A. © Copyright 2000, 2001 Agilent Technologies
ii Agilent 8563E Option K35 User’s and Service Guide
Warranty Information
IMPORTANT Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
NOTE The actual warran ty on your instrument depends on the date it was ordered as well
as whether or not any warranty options were purchased at that time. To determine the exact warranty on your instrument, contact Agilent Technologies with the model and serial number of your instrument. Refer to “Contacting Agilent” on
page 18.
This Agilent Technologies instrument product is warranted against def ects in material and workmanship for the warranty period. During the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective.
If the warranty covers repair or service to be performed at the Buyer’s facility , then the service or repair will be performed at the Buyer’s facility at no charge within Agilent’s service travel areas. Ou tside Agilen t’s service travel areas, warranty service will be performed at Buyer’s facility only upon Agilent’s prior agreement, and Buyer shall pay Agilent’s round-trip travel expenses. In all other areas, products must be returned to a service facility designated by Agilent.
If the product is to be retu rned to Agilent Technologies for service or r epair , it mu st be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when p rop erly insta lled on tha t ins trument. Agi lent Technologies does not warrant that the operation of the instrument, software, or firmware will be uninterrupted or error-free.
Agilent 8563E Option K35 User’s and Service Guide iii
Limitation of Warranty. The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT , OR ANY OTHER LEGAL THEORY.
iv Agilent 8563E Option K35 User’s and Service Guide
Contents
Agilent Technologies Model 8563E Option K35
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Part 1: Option K35 Quick Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Equipment Needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Reference Level Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Known Bugs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Part 2: Option K35 ACPR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Typical W-CDMA ACPR Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Front Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Enabling Option K35 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Service Center Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Remove the Analyzer Cover Assembly. . . . . . . . . . . . . . . . . . . . . . 10
Download the Serial Number and ID String . . . . . . . . . . . . . . . . . 10
Replace the Analyzer Cover Assembly . . . . . . . . . . . . . . . . . . . . . . 11
Verifying the Option K35 Operation . . . . . . . . . . . . . . . . . . . . . . . . . 12
Equipment Needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Safety Earth Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Contacting Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Appendix A:
Firmware Installation Note Modifications . . . . . . . . . . . . . . . . . . . . . . 19
Remove the Analyzer Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . 19
Download the Serial Number and ID String. . . . . . . . . . . . . . . . . . . 20
Replace the Analyzer Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . 21
Contents-1
Contents-2
Agilent Technologies
8563E Option K35
This User’s and Service Guide contains information required for installing, operating and servicing the Agilent 8563E Option K35 W-CDMA test set.
Agilent 8563E Option K35 User’s and Service Guide 1
Agilent Technologi es 8563E Option K35
Description
Description
The Agilent 8563E Option K35 provides increased dynamic range of at least 73 dBc for W-CDMA ACPR measurements when used with an 856xE/EC Option H35. Adjacent Channel Power Ratio (ACPR) measurements with at least 900 kHz of guard band are optimized in the Option K35. The Option K35 also provides control menus integrated into the softkeys of the 856xE/EC for ACPR measurements through firmware. With a frequency range of 1.2 GHz to 3.0 GHz, the Option K35 meets the needs for the emerging W-CDMA specifications.
Specifications
Frequency Range: 1.2 GHz to 3.0 GHz W-CDMA ACPR: 73 dBc Dynamic Range* Conversion Loss: 15 dB Nominal Maximum Input Power: 0 dBm Unmodulated CW, 0 Vdc,
Atten = 0 dB Damage Input Power: +30 dBm CW, 0 Vdc LO Input Power: 3. 0 GHz to 6.8 GHz LO Input Power Range: +9 dBm to +20 dBm Environmental Temperature Range: 25° C ±5° C
*RF Input power = –5 dBm, Channel Width = 4.096 MHz, Channel Spacing = 5.000 MHz
2 Agilent 8563E Option K35 User’s and Service Guide
Agilent Technologies 8563E Option K35
Description
In the following graph (Figure 1), th e optimum W -CDMA ACPR dynamic range is calculated to be 73 dB with an optimum input power of –5 dBm at the input fo r the mixer. This data is calculated from typical performance data. The W-CDMA ACPR dynamic range is specified to be 70 dB at a mi xer i nput l eve l of –4.8 dBm.
NOTE The left (solid) curve indicates the dynamic range as limited by noise. The right
(dotted) curve indicates the dynamic range when it is limited by third-order distortion products.
Figure 1 ACP Dynamic Range (dB) vs. Input Mixer Level (dBm)
Agilent 8563E Option K35 User’s and Service Guide 3
Agilent Technologi es 8563E Option K35
Installation
Installation
The following installation instructions are specific to the Agilent 8561E/EC, 8562E/EC, 8563E/EC, 8564E/EC or 8565E/EC spectrum analyzer. Refer to Figure 2.
NOTE The following key conventions are used throughout this document.
HARDKEYS] are labeled front panel keys
•[
SOFTKEYS are display defined keys (in the menus)
1. Position the spectrum analyzer on top of the Agilent 8563E Option K35.
2. Connect one flex cable (pa rt number 5062-66 62) fro m the LO OUT port on t he spectrum analyzer to the LO IN port on the Option K35.
3. Connect one flex cable (part number 5062-6662) from the IF IN port on the spectrum analyzer to the IF OUT port on the Option K35.
4. Connect the DUT RF signal to the RF INPUT port of the Option K35 for ACPR measurements.
Figure 2 Option K35 Installation
Flex Cable (5062-6662)
Cable to DUT
4 Agilent 8563E Option K35 User’s and Service Guide
Agilent Technologies 8563E Option K35
Getting St arted
Getting Started
Part 1: Option K35 Quick Check
Equipment Needed
Signal Source, Frequency Range of 1.2 to 3.0 GHz
Agilent Technologies 856xE/EC Option H35
Agilent Technologies 8563E Option K35
Procedure
1. Set both the signal source and the spectrum analyzer to 2 GHz.
2. Connect the RF OUTPUT from the signal source to the RF IN of the signal analyzer. Observe the signal on t he screen of the signal analyzer.
3. Move the RF signal from the signal analyzer and connect it to the RF IN port of the Option K35.
4. Select the Option K35 on the signal analyzer using the following instructions: a. If using front panel commands to turn on the Option K35, press the
following keys:
[Meas / User]
ACP MENU
K35 T/S
ON OFF
b. If using GPIB commands to turn on the Option K35 mode, enter the
following sequence:
ACPRTS ON
or
ACPRTS 1
NOTE The command to turn the test set off is ACPRTS OFF or ACPRTS 0. The command
to query the current state is ACPRTS ?.
5. Observe the signal on the screen again and confirm that the Option K35 is functioning.
Agilent 8563E Option K35 User’s and Service Guide 5
Agilent Technologi es 8563E Option K35
Getting Started
6. Using the mixer conversion efficiency information supplied with the instrument, enter the value of the Conversion Loss for the given frequency by pressing the following keys:
[AUX CTRL]
EXT MXR
AMPTD CORRECT
AVG CNV LOSS
Enter the value and then press Enter (Hz).
At this point, the signal analyzer will read in absolute amplitude.
Reference Level Settings
The maximum allowable Reference Level is:
RL = –30 dBm + Avg Conv Loss (dB)
The IF input maximum level is –30 dBm. Therefore, the combination of the A ver ag e Conver sion Los s (se en as CL o n scre en) an d the Re feren ce Le vel ( seen as RL on screen) will total –30 dB.
For example, if the Average Conversion Loss entered is 16.2 dB (CL = 16.2 dB) and the maximum IF input is –30 dBm, then the maximum Reference Level is –13.8 dB.
RL = –13.8 dB = –30 dBm + 16.2 dB
Known Bugs
Each time
[Instrument Preset] (IP) is presse d, the Average Conversion Loss must
be re-entered. Alt hough t he on- scre en CL va lue h as not c hanged from t he pr evious settings, the internal Average Conversion Loss has been reset to an unknown value.
6 Agilent 8563E Option K35 User’s and Service Guide
Agilent Technologies 8563E Option K35
Getting St arted
Part 2: Option K35 ACPR
1. With a CW signal connected to the RF INPUT port of the Option K35, set the center frequency of the spectrum analyzer to the same frequency.
2. Turn the K35 optio n on using the same key or GPIB sequenc e outlined in step 4 of Part 1.
The CW signal will b e viewed on th e s creen with the appropr iate set tings to t he reference level and span.
Typical W-CDMA ACPR Setup
Residual Bandwidth = 30 kHz Video Bandwidth = 300 kHz DET = SAMPLE Channel Width = 4.096 MHz Channel Spacing = 5.00 MHz
Agilent 8563E Option K35 User’s and Service Guide 7
Agilent Technologi es 8563E Option K35
Front Panel
Front Panel
Figure 3 Option K35 Front Panel Features
IF OUTPUT SMA Connector
LO INPUT, SMA Connector
Attenuator Adjustment
Control
RF INPUT,
Type-N Connector
8 Agilent 8563E Option K35 User’s and Service Guide
Theory of Operation
The following block diagram illustrates the theory of operation for the Agilent 8563E Option K35.
Figure 4 Option K35 Block Diagram
Agilent Technologies 8563E Option K35
Theory of Operation
AGILENT 8563EC OPTION H35
SPECTRUM ANALYZER
LO
OUT
RANGE
3 GHz to 6.8 GHz
LO INPUT IF OUTPUT
DIVIDE
BY 2
1.5 GHz to 3.4 GHz
.
.
310.7 MHz
2
DUT
RANGE
1.2 GHz to 3.0 GHz
2nd
IF IN
RF INPUT
MIXER
BUFFER
AMPLIFIER
1 dB STEP
ATTENUATOR
8563E
OPTION K35
Agilent 8563E Option K35 User’s and Service Guide 9
0 -11 dB
Agilent Technologi es 8563E Option K35
Service Information
Service Information
There are no adjustable parts in th is instrumen t.
If your instrument requires service or if you need to remove or replace a component, contact Agilent Technologies. Refer to “Contacting Agilent” on
page 18.
All TORX-drive screws are metric-threaded. All pozi-drive screws are inch-threaded.
Enabling Option K35
NOTE Option Enable must be performed at an Agilent Service Center.
To enable Option K35 to work with the 8560 E/EC series spectrum analyzer, Option H35 must be e nabled i n the i nstrument . Send th e 8560 E/ EC serie s analyzer to be upgraded, the firmware and this manual to an Agilent Service Center to add Option H35 to the 8560 E/EC series analyzer.
Service Center Instructions
Remove the Analyzer Cover Assembly
1. Disconnect the line-power cord, remove any adapters from the front panel connectors, and place the analyzer on its front panel.
2. If an 85620A mass memory module is mounted on the rear panel, remove it. Loosen (but do not remove) the four rear bumper screws using a 4 mm hex wrench. Pu ll the cover assembly off tow ards the rear of the instrument.
Download the Serial Number and ID String
1. Move the WR PROT/WR ENA jumper on the A2 controller assembly to the WR ENA position.
2. Set up the computer as described in the operation verification chapter of the calibration guide (or the installation and verification manual, shipped with earlier instrume nt s). Connect a GPIB (HP-IB) cable between the comput er a nd the spectrum analyzer.
3. Connect a power cord t o the sp ectru m anal yzer a nd set the LINE s witch t o ON. Wait f or the power-on adjustments to complete. Ignore any er ror messages.
4. On the spectrum analyzer, press current address is not 18, press
.
ADR
[CONFIG] ANALYZER ADDRESS. If the
[18] ENTER (the Hz key), and STORE HPIB
10 Agilent 8563E Option K35 User’s and Service Guide
Agilent Technologies 8563E Option K35
Service Information
5. Download the LD_SERID progra m and r un the program. This program is used to download the spectrum analyzer serial number and ID string into its EEROM. This information must be present for the s pectrum analyzer to function correctly. With th e 856xE/EC series analyze r connected via GPIB, add the new option “H35” t o the exis ting list o f options . This will turn on th e option to work with the 8563E Option K35 test set.
a. The software will prompt you to enter the se rial number pre fix and the ser ial
number suffix. For example, if the complete serial number is 3221A00129, enter
3321 for the prefix (the program will automatically append the A) and
enter
00129 for the suffix.
b. Next, you will be prompted to enter th e spectrum analyzer model number . A
list of valid model numbers will be displayed. Enter the model number exactly as it appears in the list. All alphabetic characters must be upper-case.
c. You will then be prompted to indicate if there are any options installed.
Answer “Y” (yes). Enter Option H35. If the serial pref ix is 3305A or belo w , enter Option 008; these analyzers have the signal identification hardware. Enter a Q (qu it) if there are no more options to add to the ID string.
6. After successfully downloading the new ID string, turn the spectrum analyzer power off and on. Some error messages will still appear; these will be cleared later.
7. Press
[CONFIG] DATECODE &OPTIONS. Under Options: should appear
the spectrum analyzer model number and 008 (if the serial prefix is 3305A or less) along with any other installed options.
Replace the Analyzer Cover Assembly
1. Move the WR PROT/WR ENA jumper on the A2 controller assembly to the WR PROT position.
2. Disconnect the power cord and GPIB cable from the spectrum analyzer.
3. Slide the cover onto th e instrument from the rea r. Be sure to locate the cover air vent holes on the bottom side of the spectrum analyzer.
CAUTION When replacing the spectrum analyzer cover, use caution to avoid damaging any
cables.
4. Attach the cover with the four screws loosened earlier, and tighten the four screws gradually to ensure that the cover is seated in the front frame gasket groove.
5. Torque each screw to 40 to 50 in-lb. to ensure proper EMI gasket compression.
Agilent 8563E Option K35 User’s and Service Guide 11
Agilent Technologi es 8563E Option K35
Service Information
Verifying the Option K35 Operation
Equipment Needed
Power Meter and Sensor
Signal Generator
Miscellaneous RF cables
Procedure
1. Set the frequency of the signal generator to a level within the range of the Agilent 8563E Option K35.
2. Set the power level of the signal generator to –20 dBm.
3. Measure the power level using the power meter and sensor. Note the power level.
4. Set the Agilent 856xEC spectrum analyzer to the same frequency as the signal generator by pressing the following softkeys:
ZERO SPAN
K35 T/S ON
5. Connect the signal source to the Option K35 RF IN. Make sure that the LO OUT of the spectrum analyzer is connected to the LO IN of the Option K35.
6. Connect the power meter and sens or to t he IF OUT of the Opti on K35. Note the power level.
The difference between the two measured power levels should be approximately the same as the value in the Mixer Conversion Efficiency information supplied with the Option K35.
12 Agilent 8563E Option K35 User’s and Service Guide
Troubleshooting
Use the following guide to troubleshoot through the listed possible problems involving the operation of the Agilent 8563E Option K35. If problems persist or cannot be corrected using thi s in formati on provi ded, co ntact Agilent Technologies.
Table 1 Troubleshooting Guide
Problem Check for the following:
No signal on screen 1. The standard configuration as described in the
Agilent Technologies 8563E Option K35
Service Information
installation section is used
2. The RF signal is between 1.2 GHz and
3.0 GHz
3. Option K35 menu item is turned ON
[User Menu]>ACP>K35 ON
4. Use the following configuration:
Use two separate RF sources, one for the LO IN and the other for RF IN. Set the LO source frequency to (RF IN Frequency +
310.7 MHz)*2
Amplitude is low, signal is on the screen, but amplitude is low
Power ON LED is not illuminated
Center frequency reads >3 GHz after pressing K35 ON OFF
The screen should show approximately 15 dB insertion loss thru Option K35
1. Connections are at their proper torque settings
2. Attenuator is set to 0 dB
3. The standard config uration is us ed (as des cribed in the installation section)
1. Power is applied to the Option K35
2. Fuse is OK
1. Ensure that the 856xEC is in the unpreselected mode:
[CONFIG]>EXT MXR>UNPR
Agilent 8563E Option K35 User’s and Service Guide 13
Agilent Technologi es 8563E Option K35
Service Information
Replaceable Parts
The following table shows the replaceable parts for this instrument.
Table 2 Replaceable Parts List
Description Agilent
Part Number
Front panel knob 0370-3215 1 Screw- rear frame 0515-0169 3 Screw - heatsink to deck 0515-0372 7 Screw - pc board to deck 0515-0374 5 Screw - PC board to heatsink 0515-1044 8 Screw - power supply to deck 0515-1079 3 Screw - bumper 0515-1146 8 Screw - attenuator bracket to deck 0515-1227 2 Screw - front/rear frame 0515-1367 8 Screw - power supply cover 0515-1459 2 Screw - power switch 0515-1460 2 Screw - power supply 0515-1585 4 Nut - dress panel 0535-0082 9
Quantity
LO Connector-front panel 1250-1666 2 RF Output Connector 1250-2191 1 Tie wrap 1400-0249 2 Cable clip - line switch cable 1400-0774 4 Cap - conductive, SMA 1401-0245 2 Ca p - conductive, Type-N 1401-0247 1 Washer - star interior 2190-0014 4 Washer - LO connector, rear 2190-0421 1 Flat Washers - heatsink 2190-0572 8 Washer - LO connector, front 2190-0761 1 Screw - attenuator to bracket 2200-0105 2 Set Screw - for knobs and shaft 3030-0007 4 Cover - power supply 5002-0611 1 Front Frame 5021-8693 1
14 Agilent 8563E Option K35 User’s and Service Guide
Table 2 Replaceable Parts List
Agilent Technologies 8563E Option K35
Service Information
Description Agilent
Part Number
Rear Frame 5021-8694 1 Adapter 5021-8695 1 Bumpers - corners 5041-8928 4 Feet 5041-8929 4 Cable Assembly - line switch 5062-4853 1 Instrument Cover Assembly 5062-4854 1 RF Cables 5062-6656 1 RF Cables 5062-6662 6 Hole Plug - chrome SMA 6960-0023 2 Hole Plug - black BNC 6960-0149 3 Warning Label - power supply 7120-4293 3 Warning Label - metric 7121-2527 1 Gasket - metallized 8160-0520 1
Quantity
Cable, LED power ON 08563-60126 1 Downconverter Board 08563-60129 1 ROM - U306 (Option K37 Only) 08563-80057 1 ROM - U307 (Option K37 Only) 08563-80058 1 ROM - U308 (Option K37 Only) 08563-80059 1 Panel - dress, rear 85640-00002 1 Panel - dress front 85640-00005 1 Bracket - attenuator 85640-00006 1 Center Deck 85640-00007 1 Heatsink 85640-20005 1 Power Supply Assembly 85640-60006 1 Attenuator 08494-60012 1
Calibration Certificate 9320-6288 1 User’s and Service Guide 08563-90235 1
Agilent 8563E Option K35 User’s and Service Guide 15
Agilent Technologi es 8563E Option K35
Safety and Regulatory Information
Safety and Regulatory Information
General
The following safety symbols are used throughout this manual. Familiarize yourself with the symbols and their meaning before operating this instrument.
WARNING Warning denotes a hazard. It calls attention to a procedure which, if not
correctly p erformed or adhe r ed to, could r e sult in inj ury or lo ss of lif e. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not correctly
performed or adhered to, could result in damage to or destruction of the instrument. Do not proce ed beyond a caution s ign until the in dicated condi tions are fully understood and met.
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
This symbol is used to mark the on position of the power line switch.
This symbol is used to mark the standby position of the power line switch.
This symbol indicates that the input power required is ac.
16 Agilent 8563E Option K35 User’s and Service Guide
Agilent Technologies 8563E Option K35
Safety and Regulatory Information
Safety Earth Ground
WARNING This is a Safety Class 1 Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted into a socket outlet pr ovided with a protected earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
WARNING If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
Before Applying Power
Verify that the product is configured to match the available main power source as described in the input po wer configu ration instructions in the standard m anual.
If this product is to be powered by an autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.
Agilent 8563E Option K35 User’s and Service Guide 17
Agilent Technologi es 8563E Option K35
Contacting Agilent
Contacting Agilent
Online assistance: www.agilent.com/find/assist United States
(tel) 1 800 452 4844
New Zealand
(tel) 0 800 738 378 (fax) (+64) 4 495 8950
Malaysia
(tel) 1 800 828 848 (fax) 1 800 801 664
Taiwan
(tel) 0800-047-866 (fax) (886) 2 25456723
Latin America
(tel) (305) 269 7500 (fax) (305) 269 7599
Japan
(tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Philippines
(tel) (632) 8426802
PLDT subscr iber only):
(tel) (
1 800 16510170
(fax) (632) 8426809
PLDT subscriber only):
(fax) (
1 800 16510288
People’s Re p ublic of China
(tel) (preferred): 800-810-0189 (tel) (alternate): 10800-650-0021 (fax) 10800-650-0 12 1
Canada
(tel) 1 877 894 4414 (fax) (905) 282-6495
Australia
(tel) 1 800 629 485 (fax) (+61) 3 9210 5947
Thailand
(tel) outside Bangkok:
(088) 226 008
(tel) within Bangkok:
(662) 661 3999 (fax) (66) 1 661 3714
India
(tel) 1-600-11-2929 (fax) 000-800-650-1101
Europe
(tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
Singapore
(tel) 1 800 375 8100 (fax) (65) 836 0252
Hong Kong
(tel) 800 930 871 (fax) (852) 2506 9233
18 Agilent 8563E Option K35 User’s and Service Guide
Agilent Technologies 8563E Option K35
Appendix A: Firmware Installation Note Modifications
Appendix A: Firmware Installation Note Modifications
The following modifications are made to the Agilent Technologies 8560 E-Series Firmware Upgrade Kit Installation Note with instruments with firmware revision
prior to 971024.
NOTE The Agilent Technologies 8560 E-Series Firmware Upgrade Kit Installation Note,
is sent only when Agilent 8560 E/EC series spectrum analyzers have not had Option K35 compatibility firmware installed at the factory (as in Options H35) The modifications to the Agilent 856xE/EC must be performed at an Agilent Technologies approved service center for those instrument with firmware revision prior to 971024.
Replace Table 1-1. Firmware Upgrade Kit Parts List with the following tab le:
Description Agilent
Part Number
EPROM, programmed U306 08564-80057 1 EPROM, programmed U307 08564-80058 1 EPROM, programmed U308 08564-80059 1 EPROM, programmed U309 08564-80060 1 8560 E-Series Firmware Upgrade Kit Installation Note
08560-90164 1
Quantity
Remove the Analyzer Cover Assembly
1. Disconnect the line-power cord, remove any adapters from the front panel connectors, and place the analyzer on its front panel.
2. If an 85620A mass memory module is mounted on the rear panel, remove it. Loosen (but do not remove) the four rear bumper screws using a 4 mm hex wrench. Pull the cover assembly off towards the rear o f the instrument.
Agilent 8563E Option K35 User’s and Service Guide 19
Agilent Technologi es 8563E Option K35
Appendix A: Firmware Installation Note Modifications
Download the Serial Number and ID String
The following modifications are made to the Agilent Technologies 8560 E/EC-Series Firmware Upgrade Kit Installation Note with instuments with
firmware revisions after 971024..
1. Move the WR PROT/WR ENA jumper on the A2 controller assembly to the WR ENA position.
2. Set up the computer as described in the operation verification chapter of the calibration guide (or the installation and verification manual, shipped with earlier instruments). Connect a GPIB cable between the computer and the spectrum analyzer.
3. Connect a power cord t o the sp ectru m anal yzer a nd set the LINE s witch t o ON. Wait f or the power-on adjustments to complete. Ignore any er ror messages.
4. On the spectrum analyzer, press current address is not 18, press
.
ADR
[CONFIG] ANALYZER ADDRESS. If the
[18] ENTER (the Hz key), and STORE HPIB
5. Download the LD_SERID progra m and r un the program. This program is used to download the spectrum analyzer serial number and ID string into its EEROM. This information must be present for the s pectrum analyzer to function correctly. With th e 856xE/EC series analyze r connected via GPIB, add the new option “H35” t o the exis ting list o f options . This will turn on th e option to work with the 8563E Option K35 test set.
a. The software will prompt you to enter the se rial number pre fix and the ser ial
number suffix. For example, if the complete serial number is 3221A00129, enter
3321 for the prefix (the program will automatically append the A) and
enter
00129 for the suffix.
b. Next, you will be prompted to enter th e spectrum analyzer model number . A
list of valid model numbers will be displayed. Enter the model number exactly as it appears in the list. All alphabetic characters must be upper-case.
c. You will then be prompted to indicate if there are any options installed.
Answer “Y” (yes). Enter Option H35. If the serial pref ix is 3305A or belo w , enter Option 008; these analyzers have the signal identification hardware. Enter a Q (qu it) if there are no more options to add to the ID string.
6. After successfully downloading the new ID string, turn the spectrum analyzer power off and on. Some error messages will still appear; these will be cleared later.
7. Press
[CONFIG] DATECODE &OPTIONS. Under Options: should appear
the spectrum analyzer model number and 008 (if the serial prefix is 3305A or less) along with any other installed options.
20 Agilent 8563E Option K35 User’s and Service Guide
Agilent Technologies 8563E Option K35
Appendix A: Firmware Installation Note Modifications
Replace the Analyzer Cover Assembly
1. Move the WR PROT/WR ENA jumper on the A2 controller assembly to the WR PROT position.
2. Disconnect the power cord and GPIB cable from the spectrum analyzer.
3. Slide the cover onto th e instrument from the rea r. Be sure to locate the cover air vent holes on the bottom side of the spectrum analyzer.
CAUTION When replacing the spectrum analyzer cover, use caution to avoid damaging any
cables.
4. Attach the cover with the four screws loosened earlier, and tighten the four screws gradually to ensure that the cover is seated in the front frame gasket groove.
5. Torque each screw to 40 to 50 in-lb. to ensure proper EMI gasket compression.
Agilent 8563E Option K35 User’s and Service Guide 21