Agilent 8563A Data Sheet

T.O. 33K3-4-2827-l
TECHNICAL MANUAL
CALIBRATION PROCEDURE
FOR
SPECTRUM ANALYZER
8563A (
(HEWLETT-PACKARD)
)
Distribution Statement - Distribution authorized to U. S. Government agencies and their contractors for official use or for administrative or operational purposes only, 15 August 1991. Other requests for this document shall be referred to AFMETCAL Detachment
Heath, OH 43056-6116.
Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent disclosure of the contents or reconstruction of the document.
Published under Authority of the Secretary of the Air Force
l/MLLW,
813 Irving-Wick Dr W, Suite 4M,
30NOVEMBER2000
CHANGE 4 - 28 FEBRUARY 2003
T.O. 33K3-4-2827-1
Date of issue for original and changed pages are:
Original.. Change.. Change Change.. Change..
TOTAL NUMBER OF PAGES IN THIS PUBLICATION IS 66, CONSISTING OF THE FOLLOWING:
Page
No.
Title..
.....................................
A
1- 2..
3
......................................
4 -
5
6-7.. 8
......................................
9
......................................
10 -15.. 16 -17..
18 -
20.............................
-
23.............................
21 24
....................................
25 -
29.............................
30
....................................
31
....................................
32 - 33.. 34 . .
..................................
35 ....................................
36 ....................................
37..
..................................
38
....................................
39
....................................
40 - 41
42
....................................
43
....................................
44 - 46 47
....................................
48
....................................
49
....................................
50
....................................
51
....................................
52
....................................
53
....................................
54-60.. 61
....................................
62
Blank..
...............O....................
................1............................
..................2......................28February 2002
...............
...............
...............................
...............................
.................................
...............................
...........................
...........................
...........................
.............................
.............................
...........................
........................
.3
.............................
.4
......................
*
Change
No.
4
4
0
1
0 2
0 3
0 2 0
1
2 0
3 2
0
4 0
1
3 0
1
0
1
0
1 4 2 3 2
1
0 2 0
1
0
30 November 2000
.30 June 2001
.30 July 2002
28
February 2003
Page
No.
A-l ..................................
A-2 Blank..
......................
*
Change
No.
0 0
Page
No.
* Change
No.
A Change 4
*Zero in this
c&mu
indicates an original page
USAI
1 CALIBRATION DESCRIPTION:
SPECTRUM ANALYZER
8563A ( )
(HEWLETT-PACKARD)
Table 1.
T.O.
33K3-4-2827-1
Test Instrument (TI) Performance Characteristics
Frequency Reference
Calibrator Amplitude
Frequency Readout and Marker Count
Display Average
Noise Level
Range: 10 MHz
Accuracy: Aging/year: <l X
Settability:
Temperature: <l X Range: -10
Accuracy: Range: Opt HO1 and S/N
9
H5 1,
9 Accuracy:
Frequency Readout: <*(frequency readout X frequency reference error + 5%
Xspan+ 15%XRESBW+ 10Hz)
Marker Count:
<*(marker
reference error + 50 Hz X N ** + 1 LSD)
Range: 9 and S/N <3 124); 9 Opt 026 and S/N 23 124)
Specifications
ti.3
kHz
to 22
Opt 026 and S/N 23 124,
kHz
to 26.5
frequency X frequency
kHz
to 22
kHz
to 26.5
~2.2
X
dBm
at 300 MHz
dB
GHz;
GHz
GHz
(Opt HO1
GHz
10-7;
10.‘; 10s8
(-10 to 55 “C) *
<3124,
(H51,
Test Method
Compared to Frequency
Standard
Measured with a
Power Meter
Compared to a known
frequency from a
Synthesized Sweeper
With no input to the TI, the Average Noise Level is measured on the TI CRT
Accuracy:
10 100 1 MHz to 2.9
2.9 to 6.46
6.46 to 13.0
13.0 to 19.7
19.7 to 22.0
(Opt HO1 and S/N ~3 124);
19.7 to 26.5 21 (H5 1, Opt 026 and S/N 23 124)
See footnotes at end of Table.
kHz, c-103 dBm;
kHz,
c-l 10
dBm;
GHz, c-130 dBm;
GHz, c-131 dBm;
GHz, c-120 dBm;
GHz, c-115 dBm;
GHz,
c-1 10
GHz,
c-1 10
kHz
offset from carrier)
dBm
dBm,
1
T.O. 33133-4-2827-l
Table 1. (Cont.)
Test Instrument (TI) Characteristics
Noise Sidebands
Resolution Bandwidth Switching
Resolution Bandwidth Accuracy and Selectivity
I
Performance Specifications
Range:
Accuracy: Offset:
>lO kHz,
20 Log N
>30 kHz,
20 Log N **)
>lOO kHz,
20 Log N
Accuracy:
Range: 10 Hz to 2 MHz
>lO, >30,
<(-86 <(-100
Range:
300
Accuracy:
10 Hz to 300
1 MHz, 2 MHz, Selectivity (RES BW
<15:1,60 dB
10 Hz to 1 MHz and 2 MHz
kHz
RES BW setting)
+25%;
+25% (+20%
and
dBc/Hz
**) dBc/Hz;
dBc/Hz
dBc/Hz;
<(-110
dBc/Hz
**) dBc/Hz
f0.5 dB
(referenced to
kHz, flO%;
to 3 dB BW
for Opt H51);
>lOO kHz
+
+
+
2300
Hz),
offsets
Test Method
Signal is applied to TI
and the Noise Sidebands
are measured
With an external signal
applied to the TI, the
amplitude reference is set with RES BW set to 300
kHz.
The remaining RES BW settings are compared to 300
amplitude Compared to a known
signal level
kHz
Input Attenuator
IF Gain
Scale Fidelity
See footnote at end of Table.
2
Range: ( referenced to 10
setting) 20 to 70
2.9
GHz
Accuracy: fl.8 dB maximum
Range: (10 dB Attenuator setting) 0 to -80
Accuracy:
Range: 0 to 90 dB with RES BW 2300 Hz setting; 0 to 100 dB with
RES BW setting 1100 Hz
Accuracy: LOG,
f1.5 dB over the 0 to -90 dB range
(0 to -100 dB with RES BW < 100 Hz); LINEAR:
dBm
+0.4 dB/4 dB
dB,
ti.6 dB/lO dB
<+l
.O dB
f3.0%
of Reference Level
dB
9
kHz
to
step,
to a maximum of
Table 1. (Cont.)
T.O.
33K3-4-2827-1
Test Instrument (TI)
Characteristics
Frequency Response
Performance Specifications
Range: Opt HO1 and S/N 9
kHz
to 22
GHz;
9
kHz
to 26.5 H51, Opt 026 and S/N 13124; 9
kHz
to 19.7
Accuracy: (Relative, WI 10 dB attenuator setting) 9
kHz
to 2.9
2.9 to 6.46
6.46 to 13.0
13.0 to 19.7
19.7 to 22.0
and S/N <3 124);
19.7 to 26.5
Opt 026 and S/N 13124)
Accuracy: (Absolute, referenced to CAL OUT w/ 10 dB attenuator setting)
9
kHz
to 2.9
9
kHz
to 6.46
9
kHz
to 13.0
9
kHz
to 19.7
9
kHz
to 22 and S/N <3 124); 9
kHz
to 26.5
Opt 026 and S/N 23124)
GHz,
GHz,
GHz, <+l .O dB;
GHz, <+1.5 dB;
GHz, d2.0 dB; GHz, c&3.0 dB;
GHz, <+3.0 dB
GHz, 43.0 dB
GHz, dl.5 dB;
GHz, c&2.5 dB; GHz, <+3.0 dB;
GHz, <*4.0 dB
GHz, cM.0 dB
GHz, <*4.0 dB
Opt
~3124,
H51
(Opt HO1
(H51,
(Opt HO1
(H51,
Test Method
Compared to a leveled signal
I
Frequency Span
Sweep Time
Range: 0 Hz, 100 X N ** to
19.25
GHz
(23.75
GHz
for
H51,Opt
Accuracy:
Range: 50 50 ms to 60 s (Span 1100 Hz X N **)
Accuracy: Opt H5 1); 130 ms,
(<50
Opt
ms:
EOl
026
and S/N23124)
+5%
l.ts
to 60 s (Zero Span);
~30
ms,
rt15% (+lO%
<fl%;
+lO%,
250 ms,
and
HOl)
for
+l%
Compared to a signal of known frequency
A modulated signal is
applied to the TI and the sweep time is determined
Change 1 3
T.O.
33K3-4-2827-1
Table 1. (Cont.)
Test Instrument (TI) Characteristics
Residual
Responses
Residual FM
Second Harmonic
Performance
SDecifications
Range: (0 dB of attenuation) 200
kHz
to 6.46
for Opt H5 1)
Accuracy: 200
c-90 dBm;
18
GHz
for Opt H5 1
Range: (Zero Span) Opt HO1 and S/N 9
kHz
to 22
9
kHz
to 26.5
Opt H5 1, Opt 026 and S/N 13 124
Accuracy: in 20 ms); 15 Hz X N ** p-p in 20 ms for Opt H5 1
Range: 10 MHz to
2.9
GHz
and
OptH51,9kHztolOMHz,
10 MHz to 1.45
1.45 to 11
GHz
(18
kHz
to 6.46
degradation to
<3124,
GHz;
GHz,
$10
Hz X N ** p-p
>2.75 GHz;
GHz,
GHz
GHz
GHz,
1-70 dBm
Test Method
With no input to TI
the residual response
are measured
up to
Signal is applied to TI and Residual FM is measured
Reference
is established and amplitude of harmonic is measured
Third Order Intermodulation Distortion
Accuracy: 10 MHz to 2.9
c-72 dBc
mixer; for -10
OptH51,9kHztolOMHz,
<-60 <-72
Range: 10 MHz to
>2.75 GHz;
9
>2.75 GHz
Accuracy: (two -30 signals at input mixer and separation) <-70 <-75
9 to 2.9
for -40
>2.75 GHz c-100 dBc
dBm
dBc; dBc;
kHz
to 10 MHz,
dBc, dBc, >2.75 GHz;
kHz
to 10 MHz, <-70
GHz, c-70 dBc; >2.75 GHz,
dBm
at input
at input mixer;
10 MHz to 1.45
1.45 to 11
Opt H5 1 and Opt 026,
10 MHz to 2.9
GHz,
2.9
10 MHz to
dBm
Opt H51 and Opt 026,
GHz,
dBc;
GHz,
GHz,
c-100
2.9
>50 kHz
GHz;
10 MHz
dBc
GHz,
<-75
With two separate signals
applied at the same time, a reference is established and distortion is measured
dBc
4
* Typical or Operational Specifications. Not calibrated. ** N = LO Harmonic (See below):
T.O.
33K3-4-2827-1
Frequency Band
9
kHz
to 2.9
GHz
2.75 to 6.46
5.86 to 13.0
12.4 to 19.7
19.1 to 22.0
19.1 to 26.5
GHz
GHz GHz
GHz
GHz
(Opt
(Opt
and S/N 23124)
2 EOUIPMENT REOUIREMENTS:
Minimum Use Calibration
Noun
2.1
FREQUENCY Range: 10 MHz
Specifications
STANDARD
Accuracy: 12.5 X
2.2 ELECTRONIC
Range: 0.1666 Hz to Hewlett-Packard
COUNTER 200
HO1
H51,
kHz;
at 10 MHz
and S/N
Opt 026
~3124)
10e8
Band
0
1 2 2
3
4 4
4
Equipment
Austron
2100F
5345A
LO Harmonic
1 1
3
4
Sub-
Item
2.3 SYNTHESIZED SWEEPER (26.5
2.4 POWER METER
2.5 POWER SENSOR
Accuracy:
+l
count of LSD
+0.25%,
at 10 MHz
Range:
GHz
S/N 23
+3
to -35
10 MHz
124A)
to22GHz
for H5 1, Opt 026 and 8340B
dBm
Accuracy: Frequency:
+O.OOl%;
SSB Phase Noise:
<-70 dBc
Range: -30 to
Accuracy:
at 50
+2.0%
kHz
offset
+20 dBm
Range: 300 MHz at
-20
dBm
Accuracy:
*2.5%
TAR 2.23: 1
Hewlett-Packard
Hewlett-Packard 436A
Hewlett-Packard
8481A
5
T.O.
33K3-4-2827-l
Minimum Use
Noun
2.6 POWER SENSOR
2.7 SPECTRUM
Specifications
Range: 50 MHz to 22 (26.5
GHz
for HS 1, OPT 026 and
S/N
23
124A)
Accuracy:
Range: 4.8
f5%
kHz
to 2.8
ANALYZER
Accuracy:
fl.5 dB
2.8 POWER SPLITTER Range: 10 MHz to 26.5
Accuracy: Tracking 9
kHz
to 18
GHz
10.25
2.9 SIGNAL
18 to 26.5
Range: 2.5
GHz
GHz
50.40
GENERATOR
Accuracy: SSB Phase Noise at 10 offset <- 118 dBc/I-Iz
GHz
GHz
GHz
dB
dB
kHz
Calibration
Hewlett-Packard
8485A
Hewlett-Packard
8566A/B
Hewlett-Packard
11667B
Hewlett-Packard 8663A
Sub-
2.10 FUNCTION GENERATOR
2.11 SYNTHESIZED LEVEL
GENERATOR
2.12 STEP ATTENUATOR
I
2.13 STEP
I
ATTENUATOR
2.14 TERMINATION
Range: 0.1 Hz to 500
Accuracy:
f0.25%
kHz
Range: 0 to -80 dB at 50 MHz
Accuracy:
Range: 9
+O.l dB
kHz
to 50 MHz
(Flatness)
Accuracy: f0.375
Range: 0 to 20
dB
dB
in 10 dB steps
Accuracy:
Range: 0 to 10
N/A
dB
in 1 dB steps
Accuracy:
Range: 50
Accuracy:
N/A
Q
N/A
Hewlett-Packard 3325B
Hewlett-Packard 3335A
Hewlett-Packard 8496B
Hewlett-Packard 8494B
Hewlett-Packard
908A
6 Change 2
T.O.
33K3-4-2827-1
Noun
2.15 SYNTHESIZED SIGNAL GENERATOR
2.16 DC BLOCKING CAPACITOR
2.17 POWER DIVIDER
2.18 LOW PASS FILTER
2.19 LOW PASS FILTERS
(2 EN
Minimum Use Calibration Specifications
Range: 2.7 to 5.1
+3
to -35
dBm
Accuracy: SSB Phase Noise
c-70 dBc
Range: N/A Accuracy : Range: 2.7 to 5.1 Accuracy: N/A Range: 50 MHz, rejection
>50 dB
Accuracy: Verify before use Range: 4.1
>50 dB
Accuracy: Verify before use
at 50
at 80 MHz
at 5.9
N/A
GHz,
GHz
kHz
offset
GHz
rejection
GHz
Equipment
Hewlett-Packard
8673D H15
Hewlett-Packard 08553-60169
Weinschel
1506
Hewlett-Packard 0955-0306
Hewlett-Packard 360D
Sub­Item
RCL Electronics
L1411
2.20 ATTENUATOR
2.21 PULSE GENERATOR
2.22 ATTENUATOR
(2 W
3 PRELIMINARY OPERATIONS:
3.1
Review and become familiar with entire procedure before beginning Calibration Process.
Unless otherwise designated, and prior to beginning the Calibration Process, ensure that all test equipment voltage and/or current outputs are set to zero (0) or turned off, where applicable. Ensure that all equipment switches are set to the proper position before making connections or applying power.
3.2 Connect TI and test equipment to appropriate power source and set all POWER switches to ON. Allow a 2 hour warm-up time for the TI and allow test equipment to warm-up as required by manufacturer of test equipment.
Range: 10 Accuracy: N/A Range: 0.1 Hz to 200 Accuracy: Range: 20
dB
N/A dB
kHz
Hewlett-Packard 8493C OPT-01 0
Hewlett-Packard
8116A
Hewlett-Packard 8491B OPT 020
I
3.3 After the TI has warmed up as specified, perform the Trace Alignment Procedure and Reference Level Calibration as outlined in Chapter 2 of the Spectrum Analyzer User’s Guide.
Change 2 7
T.O.
33K3-4-2827-l
3.4 Become familiar with the TI controls settings before proceeding with the calibration. The procedure will identify the Hardkeys as upper case and Bold and the Softkeys (Menu keys) as upper case only. Values shall be entered in by use of the DATA key pad. For example: Press TI AMPLITUDE, REF LVL and set to -10 The AMPLITUDE key is a DATA key pad, and then push the
3.5 The TI options and configurations are listed as follows:
Hardkey,
the REF LVL is a
MHzl-dBm/sec key which is also a part of the DATA keys.
Softkey
(Menu key), the 1 and 0 are entries using the
dBm.
a) Std TI with S/N range to 26.5
b) Opt EOl-configuration that includes external mixers. Frequency range is 22 c)
HOl-same
d)
H.51,
OPT e) OPT
f)
g) OPT extend the frequency range to 50 mixers do not require calibration.
13.6
Make copies of Table 6 and Table 8 for use as Calibration Worksheets.
3.7 Perform only that portion of the procedure that pertains to the TI being calibrated.
4 CALIBRATION PROCESS:
4.1 FREOUENCY REFERENCE CALIBRATION:
H.5
1- Frequency range is 26.5
OPT E5 1- Includes OPT H5 1 and two external mixers.
E36-
~3124
have a frequency range of 22
GHz.
configuration and frequency range of
026-
All models and serial numbers extend frequency range to 26.5
GHz.
consists of a special configuration Spectrum Analyzer 8563A OPT H36 and two mixers which
Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding.
GHz.
The OPT H36 TI has the same specs as the standard TI, and the
EOl.
NOTE
GHz.
Std TI with S/N 23124 have extended frequency
GHz.
GHz.
NOTE
Adjustment of the Time Base Oscillator is normal due to the Aging Rate of the crystal, This is common to all Quartz Oscillators. However, in order to ensure the reliability of the TI, the following action will be taken: If TI passes the following applicable steps, NO ADJUSTMENT ACTION should be entered into the Maintenance Data Collection System. If the TI failed, perform the applicable steps listed in applicable Maintenance Technical Order or Commercial Data and enter appropriate ADJUSTMENT ACTION into the Maintenance Data Collection System.
4.1 .lConnect Frequency Standard 10 MHz REF OUT to the Electronic Counter EXT FREQ STD INPUT (l-10 MHz). Set Electronic Counter INT
4.1.2 Connect TI 10 MHz REF OUTPUT to the Electronic Counter CH A Input. Set Electronic Counter
50$2/lMfl
8
switch to
5Oa.
The values in the following steps are derived from and multiplication of the
Aging Rate to determine the offset at twelve months specification. Use these calculated twelve month values (including Settability) regardless of length of the calibration interval for this TI in T.O.
STD/EXT
STD switch to EXT STD.
NOTE
and
adding the Settability
33K-l-100-1/2.
T.O. 33K3-4-2827-1
4.1.3 Adjust Electronic Counter controls for a stable indication and push RESET. Verify the Electronic Counter indication is 9 999 998.8 to 10 000 001.2 Hz.
4.1.4 Set Electronic Counter INT STD/EXT STD switch to INT STD and disconnect test setup.
4.2 CALIBRATOR AMPLITUDE CALIBRATION:
4.2.1 Standardize the Power Meter and Power Sensor (2.5) for a 300 MHz measurement.
4.2.2 Connect the Power Sensor (2.5) to the TI CAL OUTPUT.
4.2.3 The Power Meter must indicate between -9.7 to -10.3 dBm.
4.2.4 Disconnect the test setup.
4.3 FREQUENCY READOUT AND MARKER COUNT CALIBRATION:
4.3.1 Connect equipment as shown in Figure 1.
10 MHz REF
SYNTHESIZED SWEEPER
FREQUNECY
RFOUTPUT
STANDARD
-
0
EXT
INPUT 50
0
IN/OUT
LI
Figure
1.
4.3.2 Press the Synthesized Sweeper INSTR PRESET and set the controls as follows:
1.5
-10
GHz
dBm
cw POWER LEVEL Frequency Standard switch (rear panel) EXT
4.3.3 Press the TI PRESET key. Set controls as follows:
FREQUENCY
CENTERFREQ
SPAN
4.3.4 Press the TI RECALL, MORE 1 OF 2, FACTORY PRSEL PK, PEAK SEARCH.
1.5
GHz
IMHZ
4.3.5 Verify the TI MKR frequency indication is between the corresponding values listed in the Limits column of
Table 2.
4.3.6 Set the Synthesized Sweeper to the remaining Applied values listed in Table 2. Set TI controls to the remaining corresponding values in Table 2 and repeat steps 4.3.4 and 4.3.5.
Change3 9
T.O. 33K3-4-2827-1
Table 2.
Applied
1.5
1.5
1.5
1.5
1.5
1.5
4.0
4.0
4.0
4.0
4.0
(GHz)
TI SPAN (MHz)
1
10
20 50
100 1000 1 10
20 50
100
TI CENTJZR
FmQ
(G&l
1.5
1.5
1.5
1.5
1.5
1.5
4.0
4.0
4.0
4.0
4.0
Limits
1.499948 to 1.500052
1.49948 to 1.50052
1.49895 to 1.50105
1.49745 to 1.50255
1.4948 to 1.50520
1.4500 to 1.5500
3.999948 to 4.000052
3.99948 to 4.00052
3.99895 to 4.00105
3.99745 to 4.00255
3.9948 to 4.0052
(GHz)
4.0
9.0
9.0
9.0
9.0
9.0
9.0
16.0
16.0
16.0
16.0
16.0
16.0
1000 1 10
20 50
100 1000 1 10
20 50
100 1000
4.0
9.0
9.0
9.0
9.0 8.99745 to 9.00255
9.0
9.0
16.0
16.0
16.0
16.0
16.0
16.0
3.9500 to 4.0500
8.999948 to 9.000052
8.99948 to 9.00052
8.99895 to 9.00105
8.99480 to 9.00520
8.950 to 9.050
15.999948 to 16.000052
15.99948 to 16.00052
15.99895 to 16.00105
15.99745 to 16.00255
15.9948 to 16.0052
15.950 to 16.050
10
T.O.
33K3-4-2827-1
Applied
21.0
21.0
21.0
21.0 50 21.0
21.0 100 21.0
21.0
4.3.7 Set controls as follows:
(GHz)
SPAN
FREQ COUNT
COUNTER RES
Table 2.
TI SPAN (MHz) FREQ
1 10
20 21.0 20.99895 to 21.00105
1000
(Cont.)
TI CENTER
(GHz)
21.0
21.0
21.0
Limits
(GHz)
20.999948
20.99948 to 21.00052
20.99745 to 21.00255
20.9948 to 21.0052
20.950
1 MHz
1 Hz
to 21 BOO052
to 2
1.050
4.3.8 Press the TI PEAK SEARCH.
4.3.9 Verify the TI CNTR frequency indicates within
4.3.10 Set the Synthesized Sweeper to the remaining applied values listed in the Applied column of Table 3, set the TI controls to the remaining corresponding values in Table 3 and repeat steps 4.3.8 and 4.3.9.
TI
4.3.11
Applied
1.5
4.0
9.0
16.0
21.0
Set the Synthesized Sweeper for minimum output and disconnect the test setup.
(GHz)
SPAN (MHz)
1
1
1
1
1
the
values listed in the Limits column of Table 3.
Table 3.
TI CENTER FREQ
(GHz)
1.5
4.0
9.0 8.99999989 to 9.00000011
16.0 15.99999984 to 16.00000016
21.0 20.99999979
Limits
(GHz)
1.49999994 to 1.50000006
3.99999994 to 4.00000006
to 21 .oooooo21
11
T.O.
33K3-4-2827-1
4.4 DISPLAYED AVERAGE NOISE LEVEL CALIBRATION:
4.4.1 Connect the TI CAL OUTPUT to the TI INPUT 50 R connector. Press the TI PRESET and set the TI controls as follows:
FREQUENCY CENTER FREQ 300 MHz SPAN AMPLITUDE
REF LVL -10
AlTEN
LOG
dB/DIV
BW
RES BW 10 Hz
VIDEO BW 1 Hz
4.4.2 Press the TI PEAK SEARCH, CAL, and REF LVL ADJUST keys.
4.4.3 Use the TI knob to adjust the REF LVL CAL number until the TI MKR indicates -10.00 dB TI STORE REF LVL.
4.4.4 Disconnect the TI CAL OUTPUT from the TI INPUT 50 R connector. Connect the Termination to the TI INPUT 50 R connector.
4.4.5 Set the TI controls as follows:
FREQUENCY
100 Hz
dBm
OdB
1
dB
+0.17 dB.
Press
CENTER FREQ 10
SPAN AMPLITUDE
LOG
dB/DIV
REF LVL -50
MARKERS
MARKERS OFF
TRIG
4.4.6 Press the TI BW and VID AVG ON keys. Set to 10 Hz.
4.4.7 Press the TI TRACE and CLEAR WRITE A keys. Wait until VAVG 10 is displayed on the TI CRT.
4.4.8 Press the TI SGL SWP, MKR, and MKRNOISE ON keys. The TI MKR amplitude must indicate
c-1
10.7
dBm.
12
kHz
1170 Hz
10
dB
dBm
CON-r
7.7 dB is added to the TI marker amplitude indications to account for the TI normalization to a 1 Hz bandwidth and corrections for the log amplifiers and envelope detector.
4.4.9 Set the TI controls as follows:
NOTE
T.O.
33K3-4-2827-1
CENTER FREQ 99 MARKERS
TRIG
OFF
CONT
4.4.10 Repeat steps 4.4.7 and 4.4.8. The TI MKR amplitude must indicate c-l 17.7
4.4.11 Set the TI controls as follows:
FREQUENCY
CENTER START FREQ
FREQ
1 STOP FREQ 2.9 MARKERS
OFF
BW
RES
BW
1 VIDEO BW 10 VID AVG
OFF
kHz
dBm.
MHz
GHz
MHz
kHz
4.4.12 Press the TI SGL
4.4.13 Press the TI MKR, MKRNOISE ON, and PEAK SEARCH keys.
SWF’
key and wait for sweep to complete.
4.4.14 Press the TI MKR + and MARKER + CF keys. Set the TI controls as follows:
SPAN BW
RESBW
VIDEO BW 1 Hz VID AVG ON
TRIG
4.4.15 Repeat steps 4.4.7 and 4.4.8. The TI MKR amplitude must indicate e-137.7
dBm.
1170 Hz
10
Hz
CON-r
13
T.O.
33K3-4-2827-1
4.4.16 Set the TI controls as follows:
FREQUENCY
CENTERFREQ
START FREQ
2.9
GHz
STOP
FRBQ
MARKERS
BW
RF.5
BW
VIDEO BW
6.46 GHz
OFF
1
MHz
10
kHz
VID AVG OFF
4.4.17 Repeat steps 4.4.12 through 4.4.15. The TI MKR amplitude must indicate c-138.7
4.4.18 Set the TI controls as follows:
FREQUENCY
START FREQ
STOP
FREQ
MARKERS
BW
6.46 GHz
13.0
GHz
OFF
dBm.
1
10
OFF
MHz
kHz
RES BW
VIDEO BW
VID AVG
4.4.19 Repeat steps 4.4.12 through 4.4.15. The TI MKR amplitude must indicate c-127.7
4.4.20 Set the TI controls as follows:
FREQUENCY
13.0
START
STOP
MARKERS
PREQ
PREQ
GHz
19.7
GHz
OFF
14
dBm.
BW
T.O.
33K3-4-2827-1
RES BW
VIDEO BW 10
1 MHz
kHz
VID AVG OFF
4.4.21 Repeat steps 4.4.12 through 4.4.15. The TI MKR amplitude must indicate c-122.7
4.4.22 Set the TI controls as follows: FREQUENCY
START
STOP
PREQ
FREQ
19.7
22.0
GHz GHz
(Opt HO1 and S/N
26.5
GHz
(Opt H51, Opt 026 and S/N 23124)
MARKERS
OFF
BW RES BW 1 MHz
VIDEO BW 10
kHz
VID AVG OFF
dBm.
~3124)
4.4.23 Repeat steps 4.4.12 through 4.4.15. The TI MKR amplitude must indicate
c-1
17.7
dBm.
4.4.24 Disconnect the Termination from the TI INPUT 50 !Z connector.
4.5 NOISE SIDEBAND CALIBRATION:
4.5.1 Connect the Signal Generator RP OUTPUT connector to the TI INPUT 50 Q connector.
4.5.2 Set the Signal Generator controls as follows: FREQUENCY 2500 MHz AMPLITUDE -15
dBm
4.5.3 Press the TI PRESET key. Set controls as follows: FREQUENCY
2.5
CENTER FREQ CF STEP
SPAN
GHz
10kHz
1 MHz
15
T.O.
33K3-4-2827-1
AMPLITUDE
REF LVL
-10
dBm
A-ITEN
PEAK SEARCH
MKR
OdB
SIG TRK ON
10
SPAN
4.5.4
Wait for completion of two sweeps and then press TI controls as follows:
MKR
kHz
SIG TRK OFF
BW
RES BW
SPAN
1
kHz
ZERO SPAN
BW
1
VIDEO BW
Hz
4.5.5 Set the Signal Generator output level to place the peak of the signal on the TI CRT Reference Level.
4.5.6 Press the TI SGL SWP and SGL SWP. Wait for the sweep to complete and set TI controls as follows:
MKR
MKR NOISE ON
MARKER DELTA
NOTE
In the following steps, the TI A MKR will indicate in dB or
dBm
instead of
dBc.
4.5.7 Press TI FREQUENCY, CENTER FREQ,
1
TI A MKR amplitude indication is c-86
dBc/Hz.
fi,
and SGL SWP. Wait for sweep to complete and verify the
4.5.8 Press TI FREQUENCY, CENTER FREQ, 0, 4 and SGL SWP. Wait for sweep to complete and verify the
1
TI A MKR amplitude indication is c-86
4.5.9 Press TI FREQUENCY, CENTER FREQ, fi and CF STEP 30
dBc/Hz.
kHz.
16 Change 2
T.O.
33K3-4-2827-1
45.10 Press the TI SGL SWP and SGL
SWP.
Wait for the sweep to complete and set TI controls as follows:
MKR
MKR NOISE ON
MARKER DELTA
45.11 Press TI FREQUENCY CENTER FREQ, TI A MKR amplitude indication is <- 100
fi,
and SGL SWP. Wait for sweep to complete and verify the
dBc/Hz.
45.12 Press TI FREQUENCY, CENTER FREQ, u, u and SGL SWP. Wait for sweep to complete and verify the TI A MKR amplitude indication is <- 100
45.13 Press TI FREQUENCY, CENTER FREQ, n and CF STEP 100
dBc/I-Iz.
kHz.
4.5.14 Press the TI SGL SWP and SGL SWP. Wait for the sweep to complete and set TI controls as follows:
MKR
MKR NOISE ON
MARKER DELTA
4.5.15 Press TI A MKR amplitude indication is <- 110
‘II
FREQUENCY, CENTER FREQ, m, and SGL SWP. Wait for sweep to complete and verify the
dBc/Hz.
I
I
I
4.5.16 Press TI FREQUENCY, CENTER FREQ, 4, 4 and SGL SWP. Wait for sweep to complete and verify the TI A MKR amplitude indication is
<-1
10
dBc/I-Iz.
4.5.17 Set Signal Generator output to minimum and disconnect test setup.
4.6 RESOLUTION BANDWIDTH SWITCHING CALIBRATION:
4.6.1 Connect equipment as shown in Figure 2.
4()/N
MHz
1 SYNTHESIZED LEVEL
OUTPUT 50
Ci
TI
INPUT 50
Sit
10 MHz
IN/OUT
FIEF
Figure 2.
4.6.2 Set the Synthesized Level Generator controls for a 50 MHz at -5.00
dBm.
I
Change 2 17
T.O.
33K3-4-2827-1
4.6.3 Press the TI PRESET, CAL, and FULL IF ADJ keys. Wait for the IF ADJUST STATUS message to disappear, then set the TI controls as follows:
FREQUENCY
CENTER FREQ 50 MHz AMPLITUDE
LOG
dB/DIV
1
dB
SPAN
1 MHz
BW
RES BW
300
kHz
4.6.4 Press the TI CAL, IF ADJ OFF, PEAK SEARCH, MKR +, MKR REF LEVEL.
4.6.5 Press the TI PEAK SEARCH and MARKER DELTA keys.
4.6.6 Press the TI SPAN and set to the next value listed in the Span column of Table 4.
4.6.7 Press the TI BW, Press the TI
VBW/RBW
RES
BW and set to the corresponding value listed in the RES BW column of Table 4.
RATIO and set to 0.100.
4.6.8 Press the TI CAL and ADJ CURR IF STATE keys. Wait for the IF ADJUST STATUS message to
disappear.
4.6.9 Press the TI PEAK SEARCH key.
4.6.10 The TI A MKR amplitude must indicate between the corresponding values listed in the Limits column of Table 4.
4.6.11 Repeat steps 4.6.6 through 4.6.10 for the remaining values listed in Table 4.
Table 4.
18
TI SPAN TI RES BW Limits
1 MHz
10 MHz 5 MHz 500
kHz
100
kHz
50 kHz
1okHz
5lcHz
IkI-IZ
300 kHz 2
MHz
1
MHz
100
kHz
30
kHz
10
kHz
3kHz
1kHz
300
Hz
Reference
-0.5to+0.5
-0.5to+0.5
-0.5to+0.5
-0.5to+0.5
-0.5to+0.5
-0.5to+0.5
-0.5to+0.5
-0.5to+os
(dB)
Loading...
+ 44 hidden pages