Agilent Technologies
8560 Series, 85620A, and 85629B
Part Number 08560-90163 Supersedes: 08560-90150
Printed in USA December 1999
Notice.
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited
to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies
shall not be liable for errors contained herein or for incidental or consequential damages in connection with
the furnishing, performance, or use of this material.
Most of the instrument operation is controlled by firmware contained in read-only memory (ROM). The
firmware is updated periodically to introduce new features and fix known bugs. The date code is
displayed in the form YYMMDD.
This document lists the various firmware revisions for the following instruments:
Instrument Firmware Information
8560E
8561E
8562E
8563E
8564E
8565E
8560A
8561B
8563A
8561A
8562A
8562B
85620ATable 6 on page 11
85629BTable 8 on page 13
This firmware note also indicates which versions of the 85620A mass memory module and the 85629B
test and adjustment module (TAM) can be used with the different portable spectrum analyzers.
Mass memory module compatibility information is found in Table 10 on page 15, and test and
adjustment module compatibility information is found in Table 11 on page 16 and Table 12 on page 18.
This compatibility information updates and supersedes similar information found in spectrum
analyzer installation and verification manuals.
Table 1 on page 4
Table 2 on page 7
Table 3 on page 8
Table 4 on page 9
Table 5 on page 10
Table 7 on page 12
Table 9 on page 13
This note should be filed with either the 85620A documentation, 85629B documentation, or with the
Agilent 8560E, 8561E, 8562E, 8563E, 8564E, and 8565E
Firmware Revisions
The firmware revision date code installed in the 8560E, 8561E, 8562E, 8563E, 8564E, and 8565E is
indicated in the active function block after turning the instrument on. The date code can also be found
by pressing
spectrum analyzer screen.
Table 1 8560E, 8561E, 8562E and 8563E Firmware Revision History
CONFIG, then DATECODE & OPTIONS. The firmware revision date code will be displayed on the
Firmware
Date
Code
990108
971024
970326
960830
960401
†‡§
†‡§
†§
†§
†§
ROM Part Number
U306U307U308U309
0856480057
0856480046
0856480042
0856480038
0856480027
0856480058
0856480047
0856480043
0856480039
0856480028
0856480059
0856480048
0856480044
0856480040
0856480029
0856480060
0856480049
0856480045
0856480041
0856480030
U310
0856480011
0856480011
0856480011
0856480011
0856480011
*
U311
0856480012
0856480012
0856480012
0856480012
0856480012
*
Firmware Revision
History
Fixes FADC bug.
Fixes SIG ID On/Off and zero
span amplitude jump. Year
2000 bug fixed.
AC/DC coupling bug fix,
weighting bug fix, linear log 0
bug fix, and sample unlock
error bug fix.
Adds the 85620A memory
module firmware to the main
firmware in the spectrum
analyzer and corrects some
firmware errors. First revision
to support 8562E and Option
EMI.
941028
†
0856480023
0856480024
0856480025
0856480026
0856480011
0856480012
Eliminates intermittent
sweep generator failures
being reported by the TAM
AFI routines.
941027
†
0856480019
0856480020
0856480021
0856480022
0856480011
0856480012
Improves amplitude
repeatability.
4 Firmware Note 08560-90163
Firmware
Date
Code
ROM Part Number
U306U307U308U309
U310
*
U311
*
Firmware Revision
History
940622
†
0856480013
940210
†
0856480007
93080908563-
80072
0856480014
0856480008
0856380073
0856480015
0856480009
0856380074
0856480016
0856480010
0856380075
0856480011
0856480011
0856380076
0856480012
0856480012
0856380077
Eliminates intermittent
sweep generator failures
being reported by the TAM
AFI routines. Fixes battery
problem, error 718, routine.
Eliminates tracking generator
problems (power level offset at
IP and low-noise Frequency
Control compatibility) for
8560E Option 002.
Provides additional power and
adjacent channel power
measurements and allows for
amplitude correction to be
added to the displayed signal.
Also corrected were some
firmware errors in remote
power bandwidth.
This revision changes gain
limits and eliminates
unnecessary errors. It is also
the latest revision that will
function with an 8 MHz
processor.
93071508563-
80066
0856380067
0856380068
0856380069
0856380070
0856380071
This revision eliminates
unnecessary 500 series errors
which appear during power-on
in the 8560E and 8561E.
93022608563-
80059
0856380060
0856380061
0856380062
0856380063
0856380064
This revision supports Option
008 and the fast ADC (Option
007). It also supports 16 MHz
controller operation.
Firmware Note 08560-90163 5
Firmware
Date
Code
ROM Part Number
U306U307U308U309
U310
*
U311
*
Firmware Revision
History
92052808563-
80032
*
Refer to Table 8 on pag e 13 for the date code of U310 and U311.
†
When upgrading to these firmware revisions, some spectrum analyzers require additional hardware changes:
• 8560E-Series spectrum analyzers have undergone two changes which affect firmware. The first was to
change the A2 controller from an 8 MHz processor to a 16 MHz processor. The second was to change the
A2U501 EEROM to a larger size to make room for additional flatness correction points.
• 8 MHz controller to 16 MHz controller (prefix changes from 3305A to 3310A).
• A2U501 EEROM to A2U500 EEROM (prefix changes from 3410A to 3416A).
• All instruments which contain 940210 or later firmware need only to have the Firmware Upgrade Kit
installed. Do not use the software included in the kit with these instruments.
• The 08563-60058 Controller Upgrade Kit must be purchased if the analyzer has an 8 MHz controller.
Analyzers which contain firmware revision 920528 have 8 MHz controllers. Some later analyzers may have
been changed in the field to 16 MHz processors. To determine whether an analyzer has a 16 MHz controller,
press PRESET, SPAN, 1 kHz. If the annotated sweep time is 477 ms, the CPU clock is 16 MHz. If the
annotated sweep time is 740 ms, the clock speed is 8 MHz. Do not order the firmware upgrade or the
EEROM upgrade kit if the 08563-60058 is being ordered; it contains the current firmware and
EEROM.
• Analyzers with serial prefixes 3410A and below were originally shipped with the smaller EEROM A2U501.
If the instrument has a 16 MHz controller, but has the smaller EEROM, the EEROM Upgrade Kit must be
ordered in addition to the Firmware Upgrade Kit. To determine if the EEROM Upgrade Kit is required, the
A2 assembly must be inspected. If the EEROM is in the A2U501 position, or if the A2 assembly part number
is 08564-60010 or higher, the EEROM Upgrade Kit (5063-0669) must be ordered.
0856380033
0856380034
0856380035
0856380036
0856380037
First version.
• 8561E analyzers must also have A14 frequency control assembly upgraded to the current version if it is
08561-60034 or earlier.
‡
The part number for the firmware upgrade kit is 08560-60090.
§
Firmware revisions 960401 and later are compatible only with 85620A Mass Memory Modules with internal
firmware revision C (date code 910116).
6 Firmware Note 08560-90163
Table 2 8564E and 8565E Firmware Revision History
ROM Part Number
Firmware
Date Code
U306U307U308U309
U310
*
U311
*
Firmware Revision
History
971024
†‡
0856480046
970326
‡
0856480042
960830
‡
0856480038
960401
‡
0856480027
94102808564-
80023
94102708564-
80019
94062208564-
80013
0856480047
0856480043
0856480039
0856480028
0856480024
0856480020
0856480014
0856480048
0856480044
0856480040
0856480029
0856480025
0856480021
0856480015
0856480049
0856480045
0856480041
0856480030
0856480026
0856480022
0856480016
0856480011
0856480011
0856480011
0856480011
0856480011
0856480011
0856480011
0856480012
0856480012
0856480012
0856480012
0856480012
0856480012
0856480012
Fixes SIG ID On/Off and
zero span amplitude jump.
Year 2000 bug fixed.
AC/DC coupling bug fix,
weighting bug fix, linear
log 0 bug fix, and sample
unlock error bug fix.
Adds the 85620A mass
memory module firmware
to the main firmware in the
spectrum analyzer.
Eliminates intermittent
sweep generator failures
being reported by the TAM
AFI routines.
Improves amplitude
repeatability.
Eliminates intermittent
sweep generator failures
being reported by the TAM
AFI routines. Fixes battery
problem, error 718, routine.
94021008564-
80007
0856480008
0856480009
0856480010
0856480011
0856480012
This revision changes the
sequence of events in the
LO/IF alignment routine to
improve flatness
repeatability.
93121608564-
80003
*
Refer to Table 8 on page 13 for the date code of U310 and U311.
†
The part number for the firmware upgrade kit is 08560-60090.
‡
Firmware revision 960401 and later is only compatible with Mass Memory Modules with internal
0856480004
0856480005
0856480006
0856380076
0856380077
First version.
firmware revision C (date code 910116).
Firmware Note 08560-90163 7
Agilent 8560A, 8561B, and 8563A Firmware Revisions
The firmware revision date code installed in the 8560A, 8561B, and 8563A is indicated in the active
function block after turning the instrument on. The date code can also be found by pressing
DATECODE & OPTIONS. The firmware revision date code will be displayed on the spectrum analyzer
then
screen.
Startingwith serial prefix 3207A for the 8560A, 3206A for the 8561B, and3216A for the 8563A, the set of PROMS in each instrument consists of two 2 MBitPROMS instead of four 1 MBit PROMS(whichshipped with earlier instruments).
Table 3 on pa ge 8 lists firmware revisions compatible with current instruments.Table4 on page9 lists
firmware revisions compatible with earlier instruments requiring a set of four 1 MBit PROMS.
Table 3 8560A, 8561B, and 8563A Firmware Revision History (2 PROM Sets)
Firmware ROM Part Number
CONFIG,
RevDate CodeU306U307
*
H
92050408563-8004108563-80040This version fixes the tuning voltage problem,
Firmware Revision
History
related to 11974A compatibility.
G92012208563-8002608563-80027This is the first version of firmware that is
compatible with the new configuration of the
A4 assembly found in current instruments.
This version has the same feature set,
however, as firmware revision 911108. The
PROMS are simply 2 MBit instead of 1 MBit.
This firmware revision has a compatibility
problem relating to the 11974A.
*
The part number for the firmware upgrade kit is 5063-0270.
NOTEInstruments using A2 controller assembly 08560-60013 also require RAM
Expansion Kit 5062-7714.
8 Firmware Note 08560-90163
Table 4 8560A, 8561B, and 8563A Firmware Revision History (4 PROM Sets)
Firmware ROM Part Number
Rev Date CodeU306U307U308U309
*
92050408563-
H
80043
F91110808563-
80016
E90030108563-
80006
D89120108560-
80023
C89093008563-
80002
0856380042
0856380017
0856380007
0856080024
0856380003
0856380044
0856380018
0856380008
0856080025
0856380004
0856380045
0856380019
0856380009
0856080026
0856380005
Firmware Revision
History
This versionfixes the tuning voltage problem related to 11974A compatibility.This isthe last version ofsoftwarethat is compatible withthe oldA4/A16 configuration found in 8560A’s serial prefixed below 3207A, 8561B’s
serial prefixed below 3206A, and
8563A’s serial prefixed below 3216A.
Fixes Meas. Uncal bug and adds
adjacent channel power softkey. This
version of firmware has a compatibility
problem relating to the 11974A.
Fixes unnecessary OCXO error messages905-907, Q band, U bandflipfor external mixers, MKBW command,and3 dB and 6 dB BW keys.
Span change bug fix.
Oven controlled oscillator. (OCXO),
8563A, and 11974 series compatibility.
B89072008560-
80015
A89070408560-
80002
*
The part number for thefirmware upgrade kit is 5063-0271.
0856080016
0856080003
0856080017
0856080004
0856080018
0856080005
Completely compatible with 85620A,
revision 890524.
First version.
Firmware Note 08560-90163 9
Agilent 8561A, 8562A, and 8562B Firmware Revisions
The firmware revision date code installed in your particular 8561A, 8562A, or 8562B is indicated in
the active function block after turning the instrument on.
Table 5 8561A, 8562A, and 8562B Firmware Revision History
The part number for the firmware upgrade kit is 08562-60163 for a four ROM set, and 5063-0241 for
a six ROM set. Six ROM sets should be used with a controller board assembly with part number
08562-60110 or earlier.
10 Firmware Note 08560-90163
Agilent 85620A Firmware Revisions
Mass memory module operation is controlled by ROM (read-only memory) firmware. The module
firmware version is displayed after pressing the
Table 10 on page 15 for firmware versions of the module that are compatible with the different portable
spectrum analyzers.
It is possible to get a different set of functions from a given mass memory module with revision C
firmware, depending on the firmware revision of the host spectrum analyzer. In spectrum analyzers
with firmware revision 960401 and later, the firmware that controls the mass memory module actually
resides in the host spectrum analyzer and contains a more recent set of features and functionality. If
that same mass memory module is installed on a host spectrum analyzer with firmware revision
941028 or earlier, the firmware that resides in the mass memory module (revision A, B, or C) will
control the features and functionality of the module.
Mass memory modules with revision A or B firmware are incompatible with host spectrum analyzers
with firmware revision 960401 and later.
NOTEDLPs, traces, and limit-lines stored on memory cards using a mass-memory
module with firmware date code 910116 or later cannot be read into a module with
an earlier firmware date code (for example 890524). However, DLPs, traces, and
limit-lines stored on cards using modules with firmware date code 890524 or
earlier can be read into a module with firmware date code 910116 or later.
Table 6 85620A Firmware Revisions Shipped with Serial Prefix 3143A and Above
Firmware ROM Part Number
Rev
Date
Code
U7
MODULE key on the spectrum analyzer. Refer to
Firmware
Revision History
950829Firmware is in spectrum
†
analyzer.
Adds the menu for formatting the memory card.
Down-loadable program execution time is
improved.
LIMIFAIL predefined
variable, and sorted
directories added. Real
arrays >75 elements now
usable. EP can be used to
change the value of a
user-defined variable.
* Refer to Table 1 o n page 4 and Table 2 on page 7 for 856x E-Series Spectrum Analyzer upgrade
information.
†
The part number for the firmware upgrade kit is 85620-60013.
‡
Mass memory modules with revision A or B firmware are incompatible with host spectrum
analyzers with firmware revision 960401 and later.
12 Firmware Note 08560-90163
Agilent 85629B Firmware Revisions
The firmware revision for the 85629B test and adjustment module (TAM) can be checked, when
connected to a compatible 8560 Series spectrum analyzer, by pressing
analyzer front panel. The firmware revision date code for the module will be listed at the top of the
spectrum analyzer display. Each firmware revision is designed to test a specific version of spectrum
analyzer. Table 9 on page 13 lists TAM firmware revisions A through E. Table 8 on page 13 lists TAM
firmware revisions 920123 (F) and later. TAM firmware revisions A through E reside in the TAM. TAM
firmware revisions 920123 and later are installed in the 8560 Series spectrum analyzer.
When testing a spectrum analyzer with TAM firmware revision 920123 or above, the TAM firmware
installed in the spectrum analyzer will override the firmware in the TAM. When testing an earlier
8560 Series spectrum analyzer without TAM firmware installed, the firmware in the TAM will control
the TAM. The firmware revision that is controlling the TAM can be easily identified using the
procedure described above.
Table 8ROM Part Numbers for TAM Firmware Revisions 920123 and Later
MODULE on the spectrum
Firmware
Date Code
940314
*
ROM Part Number
U310U311
Firmware
Revision History
08564-8001108564-80012Compatible with controller board (08564-60010) and
the new Frequency Control board.
92050408563-8004608563-80047Bug fix relating to automatic fault isolation linear
amplitude test.
92032708563-8003808563-80039Fixes Preselector peak and YTO endpoint
adjustment. Previous version incompatible with
436A and 8902A power meters.
92012308563-8003008563-80031PROMS upgraded from 1 MBit to 2 MBit and TAM
firmware now resides in the 8560 spectrum
analyzer. This version is also compatible with new
A4 log amp/cal osc.
*
The part number for the firmware upgrade kit is 5063-0260.
Table 9ROM Part Numbers for TAM Firmware Revisions A through E
Firmware ROM Part Number
RevDate CodeU1U3U4
Firmware
Revision History
*
90030185629-
E
80079
D89093085629-
80075
8562980080
8562980076
8562980081
8562980077
Compatible with new solid state switch and
new CRT.
Compatible with 8563A, new A7 LO
distribution amplifier adjustment, and new
low end mixer.
C89070485629-
80071
8562980072
8562980073
Compatible with 8560A/61B and new oven
controlled oscillator (OCX0).
Firmware Note 08560-90163 13
Firmware ROM Part Number
RevDate CodeU1U3U4
Firmware
Revision History
B89062185629-
80067
A88103185629-
80048
*
The part number for the firmware upgrade kit is 85629-60009.
8562980068
8562980049
8562980069
8562980050
Compatible with changes for A2 (-60010), A4
(-60109), and A14 (-60055).
First version.
14 Firmware Note 08560-90163
Compatibility of the 85620A and 8560 Series Spectrum
Analyzers
The following table indicates the revisions of spectrum analyzer firmware that are compatible with
each revision of 85620A mass memory module firmware.
Table 10 85620A and 8560 Series Firmware Compatibility
85620A Firmware Revision
Spectrum
Analyzer
890214
(Rev A)
890524
(Rev B)
910116
(Rev C)
950829
and later
8560ANot supported890720 and later
firmware
revisions
8560ENot supported941028 and earlier
firmware
revisions
8561AAll revisions of
firmware
8561BNot supported890720 and later
8561ENot supported941028 and earlier
8562A870728 and later
firmware
revisions
8562B870728 and later
firmware
revisions
8562ENot supportedNot supportedAll revisions of
All revisions of
firmware
firmware
revisions
firmware
revisions
870728 and later
firmware
revisions
870728 and later
firmware
revisions
890720 and later
firmware
revisions
All revisions of
firmware
All revisions of
firmware
890720 and later
firmware
revisions
All revisions of
firmware
870728 and later
firmware
revisions
870728 and later
firmware
revisions
firmware
Not supported
960401 and later
firmware
revisions
Not supported
Not supported
960401 and later
firmware
revisions
Not supported
Not supported
960401 and later
firmware
revisions
8563ANot supportedAll revisions of
firmware
8563ENot supported941028 and earlier
firmware
revisions
8564ENot supported941028 and earlier
firmware
revisions
8565ENot supported941028 and earlier
firmware
revisions
All revisions of
firmware
All revisions of
firmware
All revisions of
firmware
All revisions of
firmware
Not supported
960401 and later
firmware
revisions
960401 and later
firmware
revisions
960401 and later
firmware
revisions
Firmware Note 08560-90163 15
Compatibility of the 85629B and 8560 Series Spectrum
Analyzers
The following two tables indicate the range of spectrum analyzer serial-number prefixes that are
compatible with each revision of 85629B firmware. The TAM can be used with spectrum analyzers
that have serial-number prefixes outside the range indicated in the following tables, but the TAM may
not be fully compatible with the spectrum analyzer. The 8562E TAM interface software, available with
8562E Option 915, is required to use the TAM with the 8562E.
In general, firmware revisions are compatible with spectrum analyzers having lower serial-number
prefixes than those shown (unless otherwise noted). The level of compatibility can be improved by
replacing the TAM firmware with the latest version.
The TAM is not compatible with the 8564E or 8565E spectrum analyzer. The 8562E/TAM Interface
Software is required to use the TAM with the 8562E. This software is provided with the 8562E Option
915.
Table 11 85629B and 8560 Series Compatibility (TAM Firmware Revisions A
through E)
*
900301
(Rev E)
Spectrum
Analyzer
881031
(Rev A)
85629B Firmware Revisions A through E
890621
†
(Rev B)†
890704
(Rev C)†
890930
(Rev D)†
8560ANot supportedNot supported2925A2925A and
below
8560E
8561A
8561B
‡§
‡
All serials
2841A to
2912A
*
All serials
2923A and
below
Not supportedNot supported2925A2925A and
*
All serials
2927A
below
††
*
and
All serials
2929A and
below
below
8561E
All serials
8562A2912A and
below
8562B2842A to
2912A
8562E
#
All serials
*
All serials
2923A and
below
2923A and
below
*
All serials
*
All serials
2927A
below
2927A
below
*
All serials
††
††
*
and
and
*
All serials
2929A and
below
2929A and
below
All serials
*
*
*
3147A and
below
All serials
3147A and
below
3147A and
below
All serials
3147A and
below
3147A and
below
All serials
8563ANot supportedNot supportedNot supported2925A3147A and
When the 85629B TAM is installed on an 8560 E-Series spectrum analyzer, the TAM firmware date code will
be 920504 or above, even though the same TAM might use date code 920327 or below (in the TAM) when
installed on an earlier (non-E-Series) spectrum analyzer.
†
The sampler bias manual probe test using TAM firmware revision D and earlier, may cause an error message,
Manual probe revision code of board is NOT known by this module, when testing any instrument
with serial prefix 3017A and above. Press Current Rev. then More Info, for the actual voltage levels and compare
the readings to the current specification of >1.2 V and <3.0 V.
‡
The manual probe check of A7J16 using TAM firmware revision D and earlier has incorrect test limits for
testing an 8561A/B serial prefix 3009A and above. Press More Info for the actual voltage levels and compare the
readings to the specification of −1.95 V and ±0.25 V.
§
When testing an 8561A spectrum analyzer and attempting to execute test 9 immediately after test 8, testing
stops with no warning or error indication. These two tests should not be performed in this sequence. If this
condition exists, press abort to return to the Test Menu, where a different sequence of tests can be selected. For
example, after completing test 8, select test 10 and after it is completed go back and select test 9.
††
Except Option 003.
#
The 8562E/TAM Interface Software is required to use the TAM with the 8562E. This software is provided with
the 8562E Option 915.
Firmware Note 08560-90163 17
Table 12 85629B and 8560 Series Compatibility (TAM Firmware in Spectrum
When the 85629B TAM is installed on an 8560 E-Series spectrum analyzer, the TAM firmware
date code will be 920504 or above, even though the same TAM might use date code 920327 or below
(in the TAM) when installed on an earlier (non-E-Series) spectrum analyzer.
†
The 8562E/TAM Interface Software is required to use the TAM with the 8562E. This software is
provided with the 8562E Option 915.
18 Firmware Note 08560-90163