Agilent 85620A, 85629B Firmware Note

Firmware Note

Agilent Technologies 8560 Series, 85620A, and 85629B

Part Number 08560-90163 Supersedes: 08560-90150
Printed in USA December 1999
Notice.
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
© Copyright Agilent Technologies, Inc. 1997 1999 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. 1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA

Agilent 8560 Series, 85620A, and 85629B

Most of the instrument operation is controlled by firmware contained in read-only memory (ROM). The firmware is updated periodically to introduce new features and fix known bugs. The date code is displayed in the form YYMMDD.
This document lists the various firmware revisions for the following instruments:
Instrument Firmware Information
8560E 8561E 8562E 8563E
8564E 8565E
8560A 8561B 8563A
8561A 8562A 8562B
85620A Table 6 on page 11
85629B Table 8 on page 13
This firmware note also indicates which versions of the 85620A mass memory module and the 85629B test and adjustment module (TAM) can be used with the different portable spectrum analyzers.
Mass memory module compatibility information is found in Table 10 on page 15, and test and adjustment module compatibility information is found in Table 11 on page 16 and Table 12 on page 18. This compatibility information updates and supersedes similar information found in spectrum analyzer installation and verification manuals.
Table 1 on page 4
Table 2 on page 7
Table 3 on page 8 Table 4 on page 9
Table 5 on page 10
Table 7 on page 12
Table 9 on page 13
This note should be filed with either the 85620A documentation, 85629B documentation, or with the
8560 E-Series Spectrum Analyzer Calibration Guide.
Firmware Note 08560-90163 3

Agilent 8560E, 8561E, 8562E, 8563E, 8564E, and 8565E Firmware Revisions

The firmware revision date code installed in the 8560E, 8561E, 8562E, 8563E, 8564E, and 8565E is indicated in the active function block after turning the instrument on. The date code can also be found by pressing spectrum analyzer screen.
Table 1 8560E, 8561E, 8562E and 8563E Firmware Revision History
CONFIG, then DATECODE & OPTIONS. The firmware revision date code will be displayed on the
Firmware
Date
Code
990108
971024
970326
960830
960401
†‡§
†‡§
†§
†§
†§
ROM Part Number
U306 U307 U308 U309
08564­80057
08564­80046
08564­80042
08564­80038
08564­80027
08564­80058
08564­80047
08564­80043
08564­80039
08564­80028
08564­80059
08564­80048
08564­80044
08564­80040
08564­80029
08564­80060
08564­80049
08564­80045
08564­80041
08564­80030
U310
08564­80011
08564­80011
08564­80011
08564­80011
08564­80011
*
U311
08564­80012
08564­80012
08564­80012
08564­80012
08564­80012
*
Firmware Revision
History
Fixes FADC bug.
Fixes SIG ID On/Off and zero span amplitude jump. Year 2000 bug fixed.
Frac-N bug fix, allows larger Amp Cor corrections.
AC/DC coupling bug fix, weighting bug fix, linear log 0 bug fix, and sample unlock error bug fix.
Adds the 85620A memory module firmware to the main firmware in the spectrum analyzer and corrects some firmware errors. First revision to support 8562E and Option EMI.
941028
08564­80023
08564­80024
08564­80025
08564­80026
08564­80011
08564­80012
Eliminates intermittent sweep generator failures being reported by the TAM AFI routines.
941027
08564­80019
08564­80020
08564­80021
08564­80022
08564­80011
08564­80012
Improves amplitude repeatability.
4 Firmware Note 08560-90163
Firmware
Date
Code
ROM Part Number
U306 U307 U308 U309
U310
*
U311
*
Firmware Revision
History
940622
08564­80013
940210
08564­80007
930809 08563-
80072
08564­80014
08564­80008
08563­80073
08564­80015
08564­80009
08563­80074
08564­80016
08564­80010
08563­80075
08564­80011
08564­80011
08563­80076
08564­80012
08564­80012
08563­80077
Eliminates intermittent sweep generator failures being reported by the TAM AFI routines. Fixes battery problem, error 718, routine. Eliminates tracking generator problems (power level offset at IP and low-noise Frequency Control compatibility) for 8560E Option 002.
Provides additional power and adjacent channel power measurements and allows for amplitude correction to be added to the displayed signal. Also corrected were some firmware errors in remote power bandwidth.
This revision changes gain limits and eliminates unnecessary errors. It is also the latest revision that will function with an 8 MHz processor.
930715 08563-
80066
08563­80067
08563­80068
08563­80069
08563­80070
08563­80071
This revision eliminates unnecessary 500 series errors which appear during power-on in the 8560E and 8561E.
930226 08563-
80059
08563­80060
08563­80061
08563­80062
08563­80063
08563­80064
This revision supports Option 008 and the fast ADC (Option
007). It also supports 16 MHz controller operation.
Firmware Note 08560-90163 5
Firmware
Date
Code
ROM Part Number
U306 U307 U308 U309
U310
*
U311
*
Firmware Revision
History
920528 08563-
80032
*
Refer to Table 8 on pag e 13 for the date code of U310 and U311.
When upgrading to these firmware revisions, some spectrum analyzers require additional hardware changes:
8560E-Series spectrum analyzers have undergone two changes which affect firmware. The first was to change the A2 controller from an 8 MHz processor to a 16 MHz processor. The second was to change the A2U501 EEROM to a larger size to make room for additional flatness correction points.
8 MHz controller to 16 MHz controller (prefix changes from 3305A to 3310A).
A2U501 EEROM to A2U500 EEROM (prefix changes from 3410A to 3416A).
All instruments which contain 940210 or later firmware need only to have the Firmware Upgrade Kit
installed. Do not use the software included in the kit with these instruments.
The 08563-60058 Controller Upgrade Kit must be purchased if the analyzer has an 8 MHz controller. Analyzers which contain firmware revision 920528 have 8 MHz controllers. Some later analyzers may have been changed in the field to 16 MHz processors. To determine whether an analyzer has a 16 MHz controller, press PRESET, SPAN, 1 kHz. If the annotated sweep time is 477 ms, the CPU clock is 16 MHz. If the annotated sweep time is 740 ms, the clock speed is 8 MHz. Do not order the firmware upgrade or the
EEROM upgrade kit if the 08563-60058 is being ordered; it contains the current firmware and EEROM.
Analyzers with serial prefixes 3410A and below were originally shipped with the smaller EEROM A2U501. If the instrument has a 16 MHz controller, but has the smaller EEROM, the EEROM Upgrade Kit must be ordered in addition to the Firmware Upgrade Kit. To determine if the EEROM Upgrade Kit is required, the A2 assembly must be inspected. If the EEROM is in the A2U501 position, or if the A2 assembly part number is 08564-60010 or higher, the EEROM Upgrade Kit (5063-0669) must be ordered.
08563­80033
08563­80034
08563­80035
08563­80036
08563­80037
First version.
8561E analyzers must also have A14 frequency control assembly upgraded to the current version if it is 08561-60034 or earlier.
The part number for the firmware upgrade kit is 08560-60090.
§
Firmware revisions 960401 and later are compatible only with 85620A Mass Memory Modules with internal
firmware revision C (date code 910116).
6 Firmware Note 08560-90163
Table 2 8564E and 8565E Firmware Revision History
ROM Part Number
Firmware
Date Code
U306 U307 U308 U309
U310
*
U311
*
Firmware Revision
History
971024
†‡
08564­80046
970326
08564­80042
960830
08564­80038
960401
08564­80027
941028 08564-
80023
941027 08564-
80019
940622 08564-
80013
08564­80047
08564­80043
08564­80039
08564­80028
08564­80024
08564­80020
08564­80014
08564­80048
08564­80044
08564­80040
08564­80029
08564­80025
08564­80021
08564­80015
08564­80049
08564­80045
08564­80041
08564­80030
08564­80026
08564­80022
08564­80016
08564­80011
08564­80011
08564­80011
08564­80011
08564­80011
08564­80011
08564­80011
08564­80012
08564­80012
08564­80012
08564­80012
08564­80012
08564­80012
08564­80012
Fixes SIG ID On/Off and zero span amplitude jump. Year 2000 bug fixed.
Frac-N bug fix, allows larger Amp Cor corrections.
AC/DC coupling bug fix, weighting bug fix, linear log 0 bug fix, and sample unlock error bug fix.
Adds the 85620A mass memory module firmware to the main firmware in the spectrum analyzer.
Eliminates intermittent sweep generator failures being reported by the TAM AFI routines.
Improves amplitude repeatability.
Eliminates intermittent sweep generator failures being reported by the TAM AFI routines. Fixes battery problem, error 718, routine.
940210 08564-
80007
08564­80008
08564­80009
08564­80010
08564­80011
08564­80012
This revision changes the sequence of events in the LO/IF alignment routine to improve flatness repeatability.
931216 08564-
80003
*
Refer to Table 8 on page 13 for the date code of U310 and U311.
The part number for the firmware upgrade kit is 08560-60090.
Firmware revision 960401 and later is only compatible with Mass Memory Modules with internal
08564­80004
08564­80005
08564­80006
08563­80076
08563­80077
First version.
firmware revision C (date code 910116).
Firmware Note 08560-90163 7

Agilent 8560A, 8561B, and 8563A Firmware Revisions

The firmware revision date code installed in the 8560A, 8561B, and 8563A is indicated in the active function block after turning the instrument on. The date code can also be found by pressing
DATECODE & OPTIONS. The firmware revision date code will be displayed on the spectrum analyzer
then screen.
Starting with serial prefix 3207A for the 8560A, 3206A for the 8561B, and 3216A for the 8563A, the set of PROMS in each instrument consists of two 2 MBit PROMS instead of four 1 MBit PROMS (which shipped with earlier instruments).
Table 3 on pa ge 8 lists firmware revisions compatible with current instruments. Table 4 on pa ge 9 lists
firmware revisions compatible with earlier instruments requiring a set of four 1 MBit PROMS.
Table 3 8560A, 8561B, and 8563A Firmware Revision History (2 PROM Sets)
Firmware ROM Part Number
CONFIG,
Rev Date Code U306 U307
*
H
920504 08563-80041 08563-80040 This version fixes the tuning voltage problem,
Firmware Revision
History
related to 11974A compatibility.
G 920122 08563-80026 08563-80027 This is the first version of firmware that is
compatible with the new configuration of the A4 assembly found in current instruments. This version has the same feature set, however, as firmware revision 911108. The PROMS are simply 2 MBit instead of 1 MBit. This firmware revision has a compatibility problem relating to the 11974A.
*
The part number for the firmware upgrade kit is 5063-0270.
NOTE Instruments using A2 controller assembly 08560-60013 also require RAM
Expansion Kit 5062-7714.
8 Firmware Note 08560-90163
Table 4 8560A, 8561B, and 8563A Firmware Revision History (4 PROM Sets)
Firmware ROM Part Number
Rev Date Code U306 U307 U308 U309
*
920504 08563-
H
80043
F 911108 08563-
80016
E 900301 08563-
80006
D 891201 08560-
80023
C 890930 08563-
80002
08563­80042
08563­80017
08563­80007
08560­80024
08563­80003
08563­80044
08563­80018
08563­80008
08560­80025
08563­80004
08563­80045
08563­80019
08563­80009
08560­80026
08563­80005
Firmware Revision
History
This version fixes the tuning voltage problem related to 11974A compatibility. This is the last version of software that is compatible with the old A4/A16 configuration found in 8560As serial prefixed below 3207A, 8561B’s serial prefixed below 3206A, and 8563As serial prefixed below 3216A.
Fixes Meas. Uncal bug and adds adjacent channel power softkey. This version of firmware has a compatibility problem relating to the 11974A.
Fixes unnecessary OCXO error messages 905-907, Q band, U band flip for external mixers, MKBW command, and 3 dB and 6 dB BW keys.
Span change bug fix.
Oven controlled oscillator. (OCXO), 8563A, and 11974 series compatibility.
B 890720 08560-
80015
A 890704 08560-
80002
*
The part number for the firmware upgrade kit is 5063-0271.
08560­80016
08560­80003
08560­80017
08560­80004
08560­80018
08560­80005
Completely compatible with 85620A, revision 890524.
First version.
Firmware Note 08560-90163 9

Agilent 8561A, 8562A, and 8562B Firmware Revisions

The firmware revision date code installed in your particular 8561A, 8562A, or 8562B is indicated in the active function block after turning the instrument on.
Table 5 8561A, 8562A, and 8562B Firmware Revision History
ROM Part Number, 08562-
Firmware
Date Code
910415
900901 80182 80183 80184 80185 N/A N/A Fixed unnecessary OCXO
891012 80157 80158 80159 80160 80161 80162 Oven controlled oscillator
881031 80136 80137 80138 80139 80140 80141 8561A compatibility, FFT, 99%
880201 80120 80121 80122 80123 80124 80125 Printer dump fix, add 2 MHz
871115 80106 80107 80108 80109 80110 80111 Fixed roller oscillator/tuning
870824 80090 80091 80092 80093 80094 80095 Fixed cal-to-cal span variation
870728 80077 80078 80079 80080 80081 80082 85620A compatibility, 26.5
*
U306 U307 U308 U309 U310 U311
80198 80199 80200 80201 N/A N/A Fixed unnecessary error
Firmware Revision
History
messages 336 and 499.
errors.
(OCXO) compatibility.
power BW, and full span in single sweep.
RBW and Option 026 compatibility.
range bug.
bug.
GHz and 2.5 kHz upgrade (N=band #).
× N span
870312 80048 80049 80050 80051 80056 80057 Bug fixes only, first customer
version.
861218 80001 80002 80003 80004 80020 80021 First version.
*
The part number for the firmware upgrade kit is 08562-60163 for a four ROM set, and 5063-0241 for a six ROM set. Six ROM sets should be used with a controller board assembly with part number 08562-60110 or earlier.
10 Firmware Note 08560-90163

Agilent 85620A Firmware Revisions

Mass memory module operation is controlled by ROM (read-only memory) firmware. The module firmware version is displayed after pressing the
Table 10 on page 15 for firmware versions of the module that are compatible with the different portable
spectrum analyzers.
It is possible to get a different set of functions from a given mass memory module with revision C firmware, depending on the firmware revision of the host spectrum analyzer. In spectrum analyzers with firmware revision 960401 and later, the firmware that controls the mass memory module actually resides in the host spectrum analyzer and contains a more recent set of features and functionality. If that same mass memory module is installed on a host spectrum analyzer with firmware revision 941028 or earlier, the firmware that resides in the mass memory module (revision A, B, or C) will control the features and functionality of the module.
Mass memory modules with revision A or B firmware are incompatible with host spectrum analyzers with firmware revision 960401 and later.
NOTE DLPs, traces, and limit-lines stored on memory cards using a mass-memory
module with firmware date code 910116 or later cannot be read into a module with an earlier firmware date code (for example 890524). However, DLPs, traces, and limit-lines stored on cards using modules with firmware date code 890524 or earlier can be read into a module with firmware date code 910116 or later.
Table 6 85620A Firmware Revisions Shipped with Serial Prefix 3143A and Above
Firmware ROM Part Number
Rev
Date
Code
U7
MODULE key on the spectrum analyzer. Refer to
Firmware
Revision History
950829 Firmware is in spectrum
analyzer.
Adds the menu for formatting the memory card. Down-loadable program execution time is improved.
*
910116 85620-80021 CATALOG command, LIMIFAIL predefined
C
variable, and sorted directories added. Real arrays >75 elements now usable. EP can be used to change the value of a user-defined variable.
*
Mass memory modules with revision A or B firmware are incompatible with host spectrum
analyzers with firmware revision 960401 and later.
Refer to Table 1 on page 4 and Table 2 on page 7, for 856x E-Series Spectrum Analyzer upgrade
information.
Firmware Note 08560-90163 11
Table 7 85620A Firmware Revisions Shipped with Serial Prefix 3003A and Below
Firmware ROM Part Number
Firmware
Rev
Date Code
U1 U2 U3
Revision History
950829 Firmware is in the
spectrum analyzer
*
Adds the menu for formatting the memory card. Down-loadable program execution time is improved.
910116 85620-80023 85620-80024 85620-80025 CATALOG command,
C
LIMIFAIL predefined variable, and sorted directories added. Real arrays >75 elements now usable. EP can be used to change the value of a user-defined variable.
890524 85620-80014 85620-80015 85620-80016 First revision fully
B
compatible with 8560A, 8561B, and 8563A.
890214 85620-80005 85620-80007 85620-80009 First version.
A
* Refer to Table 1 o n page 4 and Table 2 on page 7 for 856x E-Series Spectrum Analyzer upgrade information.
The part number for the firmware upgrade kit is 85620-60013.
Mass memory modules with revision A or B firmware are incompatible with host spectrum
analyzers with firmware revision 960401 and later.
12 Firmware Note 08560-90163

Agilent 85629B Firmware Revisions

The firmware revision for the 85629B test and adjustment module (TAM) can be checked, when connected to a compatible 8560 Series spectrum analyzer, by pressing analyzer front panel. The firmware revision date code for the module will be listed at the top of the spectrum analyzer display. Each firmware revision is designed to test a specific version of spectrum analyzer. Table 9 on page 13 lists TAM firmware revisions A through E. Table 8 on page 13 lists TAM firmware revisions 920123 (F) and later. TAM firmware revisions A through E reside in the TAM. TAM firmware revisions 920123 and later are installed in the 8560 Series spectrum analyzer.
When testing a spectrum analyzer with TAM firmware revision 920123 or above, the TAM firmware installed in the spectrum analyzer will override the firmware in the TAM. When testing an earlier 8560 Series spectrum analyzer without TAM firmware installed, the firmware in the TAM will control the TAM. The firmware revision that is controlling the TAM can be easily identified using the procedure described above.
Table 8 ROM Part Numbers for TAM Firmware Revisions 920123 and Later
MODULE on the spectrum
Firmware
Date Code
940314
*
ROM Part Number
U310 U311
Firmware
Revision History
08564-80011 08564-80012 Compatible with controller board (08564-60010) and
the new Frequency Control board.
920504 08563-80046 08563-80047 Bug fix relating to automatic fault isolation linear
amplitude test.
920327 08563-80038 08563-80039 Fixes Preselector peak and YTO endpoint
adjustment. Previous version incompatible with 436A and 8902A power meters.
920123 08563-80030 08563-80031 PROMS upgraded from 1 MBit to 2 MBit and TAM
firmware now resides in the 8560 spectrum analyzer. This version is also compatible with new A4 log amp/cal osc.
*
The part number for the firmware upgrade kit is 5063-0260.
Table 9 ROM Part Numbers for TAM Firmware Revisions A through E
Firmware ROM Part Number
Rev Date Code U1 U3 U4
Firmware
Revision History
*
900301 85629-
E
80079
D 890930 85629-
80075
85629­80080
85629­80076
85629­80081
85629­80077
Compatible with new solid state switch and new CRT.
Compatible with 8563A, new A7 LO distribution amplifier adjustment, and new low end mixer.
C 890704 85629-
80071
85629­80072
85629­80073
Compatible with 8560A/61B and new oven controlled oscillator (OCX0).
Firmware Note 08560-90163 13
Firmware ROM Part Number
Rev Date Code U1 U3 U4
Firmware
Revision History
B 890621 85629-
80067
A 881031 85629-
80048
*
The part number for the firmware upgrade kit is 85629-60009.
85629­80068
85629­80049
85629­80069
85629­80050
Compatible with changes for A2 (-60010), A4 (-60109), and A14 (-60055).
First version.
14 Firmware Note 08560-90163

Compatibility of the 85620A and 8560 Series Spectrum Analyzers

The following table indicates the revisions of spectrum analyzer firmware that are compatible with each revision of 85620A mass memory module firmware.
Table 10 85620A and 8560 Series Firmware Compatibility
85620A Firmware Revision
Spectrum
Analyzer
890214
(Rev A)
890524
(Rev B)
910116
(Rev C)
950829
and later
8560A Not supported 890720 and later
firmware revisions
8560E Not supported 941028 and earlier
firmware revisions
8561A All revisions of
firmware
8561B Not supported 890720 and later
8561E Not supported 941028 and earlier
8562A 870728 and later
firmware revisions
8562B 870728 and later
firmware revisions
8562E Not supported Not supported All revisions of
All revisions of firmware
firmware revisions
firmware revisions
870728 and later firmware revisions
870728 and later firmware revisions
890720 and later firmware revisions
All revisions of firmware
All revisions of firmware
890720 and later firmware revisions
All revisions of firmware
870728 and later firmware revisions
870728 and later firmware revisions
firmware
Not supported
960401 and later firmware revisions
Not supported
Not supported
960401 and later firmware revisions
Not supported
Not supported
960401 and later firmware revisions
8563A Not supported All revisions of
firmware
8563E Not supported 941028 and earlier
firmware revisions
8564E Not supported 941028 and earlier
firmware revisions
8565E Not supported 941028 and earlier
firmware revisions
All revisions of firmware
All revisions of firmware
All revisions of firmware
All revisions of firmware
Not supported
960401 and later firmware revisions
960401 and later firmware revisions
960401 and later firmware revisions
Firmware Note 08560-90163 15

Compatibility of the 85629B and 8560 Series Spectrum Analyzers

The following two tables indicate the range of spectrum analyzer serial-number prefixes that are compatible with each revision of 85629B firmware. The TAM can be used with spectrum analyzers that have serial-number prefixes outside the range indicated in the following tables, but the TAM may not be fully compatible with the spectrum analyzer. The 8562E TAM interface software, available with 8562E Option 915, is required to use the TAM with the 8562E.
In general, firmware revisions are compatible with spectrum analyzers having lower serial-number prefixes than those shown (unless otherwise noted). The level of compatibility can be improved by replacing the TAM firmware with the latest version.
The TAM is not compatible with the 8564E or 8565E spectrum analyzer. The 8562E/TAM Interface Software is required to use the TAM with the 8562E. This software is provided with the 8562E Option
915.
Table 11 85629B and 8560 Series Compatibility (TAM Firmware Revisions A
through E)
*
900301
(Rev E)
Spectrum
Analyzer
881031
(Rev A)
85629B Firmware Revisions A through E
890621
(Rev B)†
890704
(Rev C)†
890930
(Rev D)†
8560A Not supported Not supported 2925A 2925A and
below
8560E
8561A
8561B
‡§
All serials
2841A to 2912A
*
All serials
2923A and below
Not supported Not supported 2925A 2925A and
*
All serials
2927A below
††
*
and
All serials
2929A and below
below
8561E
All serials
8562A 2912A and
below
8562B 2842A to
2912A
8562E
#
All serials
*
All serials
2923A and below
2923A and below
*
All serials
*
All serials
2927A below
2927A below
*
All serials
††
††
*
and
and
*
All serials
2929A and below
2929A and below
All serials
*
*
*
3147A and below
All serials
3147A and below
3147A and below
All serials
3147A and below
3147A and below
All serials
8563A Not supported Not supported Not supported 2925A 3147A and
below
8563E
All serials
*
All serials
*
All serials
*
All serials
*
All serials
*
*
*
*
8564E Not supported Not supported Not supported Not supported Not supported
16 Firmware Note 08560-90163
85629B Firmware Revisions A through E
*
Spectrum
Analyzer
881031
(Rev A)
890621
(Rev B)†
890704
(Rev C)†
890930
(Rev D)†
900301
(Rev E)
8565E Not supported Not supported Not supported Not supported Not supported
*
When the 85629B TAM is installed on an 8560 E-Series spectrum analyzer, the TAM firmware date code will be 920504 or above, even though the same TAM might use date code 920327 or below (in the TAM) when installed on an earlier (non-E-Series) spectrum analyzer.
The sampler bias manual probe test using TAM firmware revision D and earlier, may cause an error message, Manual probe revision code of board is NOT known by this module, when testing any instrument with serial prefix 3017A and above. Press Current Rev. then More Info, for the actual voltage levels and compare the readings to the current specification of >1.2 V and <3.0 V.
The manual probe check of A7J16 using TAM firmware revision D and earlier has incorrect test limits for testing an 8561A/B serial prefix 3009A and above. Press More Info for the actual voltage levels and compare the readings to the specification of 1.95 V and ±0.25 V.
§
When testing an 8561A spectrum analyzer and attempting to execute test 9 immediately after test 8, testing stops with no warning or error indication. These two tests should not be performed in this sequence. If this condition exists, press abort to return to the Test Menu, where a different sequence of tests can be selected. For example, after completing test 8, select test 10 and after it is completed go back and select test 9.
††
Except Option 003.
#
The 8562E/TAM Interface Software is required to use the TAM with the 8562E. This software is provided with the 8562E Option 915.
Firmware Note 08560-90163 17
Table 12 85629B and 8560 Series Compatibility (TAM Firmware in Spectrum
Analyzer)
Spectrum
Analyzer
920123 920327 920504 940314
8560A 3207A
85629B Firmware (in Spectrum Analyzer)
3221A and below All serials Not supported
*
and below
8560E N/A N/A 3410A and below All serials
8561A Not supported Not supported Not supported Not supported
8561B 3207A and below 3221A and below All serials Not supported
8561E N/A N/A 3410A and below All serials
8562A Not supported Not supported Not supported Not supported
8562B Not supported Not supported Not supported Not supported
8562E
N/A N/A 3410A and below All serials
8563A 3207A and below 3221A and below All serials Not supported
8563E N/A N/A 3410A and below All serials
8564E Not supported Not supported Not supported Not supported
8565E Not supported Not supported Not supported Not supported
*
When the 85629B TAM is installed on an 8560 E-Series spectrum analyzer, the TAM firmware date code will be 920504 or above, even though the same TAM might use date code 920327 or below (in the TAM) when installed on an earlier (non-E-Series) spectrum analyzer.
The 8562E/TAM Interface Software is required to use the TAM with the 8562E. This software is provided with the 8562E Option 915.
18 Firmware Note 08560-90163