Agilent 8560EC 30 Hz to 2.9 GHz
Agilent 8561EC 30 Hz to 6.5 GHz
Agilent 8562EC 30 Hz to 13.2 GHz
Agilent 8563EC 30 Hz to 26.5 GHz
Agilent 8564EC 30 Hz to 40 GHz
Agilent 8565EC 30 Hz to 50 GHz
1. The 8561EC has been discontinued and replaced
with 8562EC.
1
The Agilent 8560 EC Series spectrum analyzers have a color display,
offer standard digitized fast time domain sweeps (Option 856xE-007
on the 8560 E Series), and are Class 3 MIL-rugged. The 8560 EC
Series is identical to the 8560E Series in all other respects.
Frequency Specifications, Agilent 8560 EC Series
Unless noted, all specifications describe the instrument’s warranted performance under the following conditions: 5-minute warm-up from ambient conditions,
autocoupled controls, digital display, IF ADJ ON, REF LVL CAL adjusted, SECOND IF OUTPUT and 1ST LO OUTPUT terminated in 50 Ω. After a 30-minute warm-up,
and over a temperature range of 20 °C to 30 °C, the preselector does not have to be peaked at each signal of interest; under these conditions factory preselector
peak values are sufficient to meet all specifications. Typical performance is non-warranted. Supplemental characteristics are denoted by “nominal” and
“approximately”; these constitute non-warranted functional performance information derived during the design process and are not tested on a continuing basis.
Frequency range
8560EC 8561EC 8562EC 8563EC 8564EC 8565EC
Internal 30 Hz
mixing 2.9 GHz 6.5 GHz 13.2 GHz 26.5 GHz 40 GHz 50 GHz
External 18 GHz to 18 GHz to 18 GHz to 18 GHz to 18 GHz to 18 GHz to
mixing 325 GHz 325 GHz 325 GHz 325 GHz 325 GHz 325 GHz
Frequency band Harmonic mixing mode (N)
30 Hz to 2.9 GHz 1
2.75 GHz to 6.46 GHz 1
5.86 GHz to 13.2 GHz 2
12.4 GHz to 26.8 GHz 4
26.4 GHz to 31.15 GHz 4
31.0 GHz to 50 GHz 8
2
to 30 Hz
2
to 30 Hz
2
to 30 Hz
1
to 30 Hz
1
to 30 Hz
1
to
Frequency reference Option 856xEC-103
Temperature stabili ty 3 ±1 x 10
–8
±1x10
Aging (per year) ±1 x 10–7 ±2x10
(per day nom.) ±5 x 10
Initial achievable accuracy ±2.2 x 10
–10 (4)
–8
±1 x 10
Short-term warmup accuracy factors (nominal)
5 minute ±1 x 10
15 minute ±1 x 10
–7
–8
–6
–6
–6
Frequency readout accuracy
(Star t, stop, center and marker frequency functions)
Span > 2 MHz x N
Span ≤ 2 MHz x N
5
±(freq readout x freq ref accuracy
5
±(freq readout x freq ref accuracy
6
+5% x span +15% x RBW +10 Hz)
6
+1% x span +15% x RBW +10 Hz)
Frequency counter accuracy
Marker count accuracy ±(marker freq x freq ref accuracy
(S/N ≥ 25 dB) +1 LSD of counter)
Accuracy at 1 GHz ±225 Hz (5-minute warm-up)
(25 °C, 1 yr aging, marker resolution = 1 Hz) ±135 Hz (15-minute warm-up)
7
7
±3003 Hz (Option 856xEC-103)
Delta count accuracy ±(delta freq x freq ref accuracy
+ 4 Hz x N
5
(S/N ≥ 25 dB) +2 LSD)
Counter resolution Selectable from 1 Hz to 1 MHz
7
+2 Hz x N
6
5
Frequency span
Range0, 100 Hz to full span(100 Hz x N
Accuracy
Span > 2 MHz x N
Span ≤ 2 MHz x N
2.9 GHz to 6.46 GHz –30 dBm each ≤ –90 dBc +15 dBm
6.46 GHz to 26.8 GHz –30 dBm each ≤ –75 dBc +7.5 dBm
26.8 GHz to 50 GHz – 30 dBm each ≤ – 85 dBc (nominal) +12.5 dBm (nominal)
Image responses Mixer level
10 MHz to 26.8 GHz –10 dBm –80 dBc
26.8 GHz to 50 GHz –30 dBm –60 dBc
Multiple and out-of-band responses
Mixer level
10 MHz to 26.8 GHz –10 dBm –80 dBc
26.8 GHz to 50 GHz –30 dBm – 55 dBc
Residual responses
≤ –90 dBm, for the range from 200 kHz to 6.46 GHz, no input signal, 0 dB input attenuation
Display range
Viewing area Color display, approximately 9.6 cm (v) x 13 cm (h)
Scale calibration 10 x 10 divisions
Log scale 10, 5, 2, 1 dB per division
Linear scale 10% of reference level per division
Scale fidelity
Incremental Maximum
Log range 0 to –90 dB 0 to –90 dB
RBW ≥ 300 Hz ±0.1 dB/dB ±0.85 dB
RBW ≤ 100 Hz ±0.2 dB/2 dB ±0.85 dB
5
Linear range ±3% of reference level
Reference level range
Log, adjustable in 0.1 dB steps
30 Hz to 31.15 GHz –120 to +30 dBm
31.15 GHz to 50 GHz –115 to +30 dBm
Linear, adjustable in 1% steps
30 Hz to 31.15 GHz 2.2 mV to 7.07 V
31.15 GHz to 50 GHz 3.98 mV to 7.07 V
1. Excluding display-related sidebands at multiples of 60 Hz
2. Agilent 85 61EC: distortion –72 dBc, SHI +32dBm
2.9 GHz to 6.46.GHz 1.5/1.1/2 .5/1.5 1.5/1.1/2.5/1.5 1.5/1.0/2.4/1.5 1.7/1.4/2.6/1.8
6.46 to 13.2 GHz 2.2/1.5/2.9/2.0 2.2/1.5/2.9/2.0 2.6/2.2/3.0/2.8
13.2 to 22 GHz 2.5/1.5/4.0/2.5 2.5/2.5/4.0/3.5
22 to 26.8 GHz 3.3/2.2/4.0/2.5 3.3/2.2/4.5/4.0
26.8 to 31.15 GHz 3.1/2.9/4.0/3.0
31.15 GHz to 40 GHz (Agilent 8564EC) 2.6/2.4/4.0/3.2
31.15 GHz to 50 GHz (Agilent 8565EC) 3.2/3.0/4.0/4.0
2
/ typical absolute
3
Band switching uncertainty
±1 dB (added to relative frequency response for between-band measurements)
Calibrator output
300 MHz x (1 ± frequency reference accuracy
4
) at –10 dBm ±0.3 dB
Input attenuator
Switching uncertainty (referenced to 10 dB attenuation)
30 Hz to 2.9 GHz for 20 to 70 dB settings of input attenuator:
±0.6 dB/10 dB step, 1.8 dB maximum
Repeatability ±0.1 dB (nominal)
IF gain uncertainty
±1 dB (0 to –80 dBm reference levels with 10 dB input attenuation)
IF alignment uncertainty
±0.5 dB (additional uncertainty only when using 300 Hz RBW)
Resolution bandwidth switching uncertainty
±0.5 dB (relative to 300 kHz RBW)
Pulse digitization uncertainty
(pulse response mode, PRF > 720/sweep time)
Log Linear
RBW ≤ 1 MHz < 1.25 dB pk-pk < 4% of ref level
RBW = 2 MHz < 3 dB pk-pk < 12% of ref level
Standard Deviation (RBW < 1 MHz) < 0.2 dB (nominal)
Time-gated spectrum analysis
Gate delay
Range 3 µs to 65.535 ms ≤ 0.5 µs
Resolution 1 µs
Accuracy ±1 µs
(From GATE TRIGGER INPUT to positive edge of GATE OUTPUT)
Gate length
Range 1 µs to 65.535 ms
Resolution 1 µs
Accuracy ±1 µs
(From positive edge to negative edge of GATE OUTPUT)
2. Absolute flatness values referenced to 300 MHz CAL OUT
3. Typical values at 25 °C
4. Frequency reference accuracy = aging x time since last adjustment + initial achievable accuracy + temperature stability
5. Up to 1 µs jitter due to 1 µs resolution of gate delay clock
7
Amplitude Specifications, continued
Delayed sweep
Trigger modes Free run, line, external, video
Range
Sweep time < 30 ms –9.9 ms to +65.535 ms
Sweep time ≥ 30 ms +2 µs to +65.535 ms
Resolution 1 µs
Accuracy ±1 µs
Demodulation
Spectrum demodulation
Modulation type AM and FM
Audio output Speaker and phone jack with volume control
Marker pause time 100 ms to 60 s (nominal)
Inputs/Outputs, Agilent 8560 EC Series
(All values are nominal)
Front panel connectors
RF input
Agilent 8560EC, 8561EC, 8562EC, 8563EC Type N female, 50 Ω
(Option 856xEC-026, Agilent 8563EC only) APC 3.5 mm male, 50 Ω
Agilent 8564EC, 8565EC APC 2.4 mm male, 50 Ω
VSWR (≥ 10 dB atten)
30 Hz to 2.9 GHz < 1.5:1 dB
2.9 GHz to 50 GHz < 2.3:1 dB
LO emission level
(average with 10 dB atten) ≤ –80 dBm
IF input SMA female, 50 Ω
Frequency 310.7 MHz
Full screen level –30 dBm
Gain compression –23 dB
First LO output SMA female, 50 Ω
Frequency 3.000 - 6.8107 GHz
Amplitude +16.5 dBm ±2.0 dB
Cal output BNC female, 50 Ω
Probe power +15 Vdc, –12.6 Vdc, and Gnd (150 mA max each)
Rear panel connectors
Earphone Sub-miniature mono jack, 0.2 W into 4 Ω
10 MHz REF in/out Shared BNC female, 50 Ω
Output freq accuracy ±(10 MHz x freq ref accuracy)
Output amplitude 0 dBm
Input amplitude –2 to +10 dBm
Video output BNC, 50 Ω
Amplitude (RBW ≥ 300 Hz) 0 to +1 V full scale
LO sweep frequency analog voltage output
(LO sweep or V/GHz function selectable from the
front panel, BNC female, 120 Ω)
LO sweep output 0 to 10 V (no load)Frequency analog voltage output (internal mixer mode)
Output ramp voltage proportional to start and stop frequencies.
Transfer function: 0.5 V/GHz.
0.5 V/GHz output (external mixer mode)
Output ramp voltage proportional to LO frequency: (LO = 3 to 6.8107 GHz).
Transfer function: (1.5 V/GHz x LO frequency (GHz) –0.2054) ±50 mV (typ).
8
Inputs/Outputs, Agilent 8560 EC Series, continued
Blanking/gate
Output Shared BNC female, 50 Ω
Blanking mode
During sweep Low T TL level
During retrace High TTL level
Gate mode
Gate on High TTL level
Gate off Low TTL level
External/gate
Trigger inpu t Shared BNC female, > 10 kΩ
Settable to high TTL or low TTL
GPIB IEEE-488 bus connector
Inter face functions SH1, AH1, T 6, L4, LE0, RL1, PP1, DC1, DT1, C1, C2 8, TE0, SR1
Direct printer output Supports HP 3630A PaintJet printer, HP 2225A ThinkJet printer
Direct plotter output Supports HP 7225A/7440A/7470A/7475A/7550A
Options
Option 856xEC-001 second IF output, Agilent 8560 EC Series
(all values are nominal)
3 dB bandwidth 8560EC 8561EC 8562EC 8563EC 8564EC, 8565EC
NF
conversion gain
30 Hz to 2.9 GHz
24 dB 25 dB 20 dB 25 dB 28 dB
1.2 dB –6.5 dB –1.2 dB –1.2 dB –1.2 dB
2.9 GHz to 6.5 GHz > 30 MHz > 30 MHz > 30 MHz > 30 MHz
26 dB 22 dB 22 dB 23 dB
–1 dB –3 dB –1 dB –1 dB
6.5 GHz to 13.2 GHz > 37 MHz > 37 MHz > 37 MHz
26 dB 26 dB 28 dB
–5.7 dB –5.7 dB –5.7 dB
13.2 GHz to 22 GHz > 4 5 MHz > 45 MHz
30 dB 32 dB
–8 dB –8 dB
22 GHz to 26.8 GHz > 4 5 MHz > 45 MHz
32 dB 35 dB
–8 dB –8 dB
26.8 GHz to31.15 GHz > 25 MHz
28 dB
–9 dB
31.15 GHz to 40 GHz > 25 MHz
38 dB
–19 dB
40 GHz to 50 GHz > 25 MHz
42 dB
–23 dB
Option 8560EC-002 Built-in tracking generator
1
> 25 MHz > 25 MHz > 25 dB > 25 MHz > 25 MHz
2
Frequency specifications
Frequency range 300 kHz to 2.9 GHz
Accuracy
After peaking ±(frequency reference accuracy x tuned frequency + 5% x span + 295 Hz)
Tracking drift (nominal) Usable in 1 kHz RBW after 5 minutes warm-up. Usable in 300 Hz RBW after
30-minute warm-up.
Minimum RBW 300 Hz
1. DC coupled for frequencies below 100 kHz. Option 8 56xEC-006 required for operation below 9 kHz in Agilent 8563EC, 8564EC, 8565EC.
2. Option 8560EC-002 is no longer offered.
3. Tracking generator not usable with resolution bandwidths ≤ 100 Hz.
3
9
Options, continued
Amplitude specifications
Output level –10 dBm to +1 dBm
10 dBm to +2.8 dBm (typical)
Resolution 0.1 dB
Accuracy
Vernier ±0.20 dB, ±0.5 dBm max (25 °C ±10 °C)
Absolute ±0.75 dB
Level flatness ±2.0 dB
Effective source match 1.92:1 (nominal)
Total absolute accuracy ±3.25 dB
External ALC input (rear panel) BNC female use with negative detector
1. Leakage measured with maximum power into 50 Ω, with 50 Ω loads on the
TG output and RF input.
2. Difference between maximum power output and tracking generator feedthrough.
10
Environmental Specifications, Agilent 8560 EC Series
Per MIL- PRF-28800F, Class 3
Calibration interval
Agilent 8560EC, 8561EC, 8562EC, 8563EC: 2 years
Agilent 8564EC, 8565EC: 1 year
Warm-up time 5 minutes in ambient conditions
Temperature 0 °C to +55 °C (operating); –40 °C to +75 °C (not operating)
Humidity 95% @ 40 °C for 5 days
Rain resistance Drip-proof at 16 liters/hour/sq. ft.
Altitude 15,000 ft. (operating), 50,000 ft. (non-operating)
Pulse shock (half sine) 30 g for 11 ms duration
Transit drop 8-inch drop on six faces and eight corners
Electromagnetic compatibility: Conducted and radiated interference in compliance with CISPR Pub. 11
(1990). Meets Mil-STD-461C, part 2, with certain exceptions.
Power requirements: 115 VAC operation: 90 to 140 V rms, 3.2 A rms max, 47 to 440 Hz
230 VAC operation: 180 to 250 V rms, 1.8 A rms max, 47 to 66 Hz
Maximum power dissipation
Agilent 8560EC, 8561EC, 8562EC, 8563EC 180 W
Agilent 8564EC, 8565EC: 260 W
Audible noise (nominal): < 5.0 Bels power at room temp (ISO DP7779)
Dimensions (w/o handle, cover): 337 mm W x 187 mm H x 461 mm D
Weight (nominal)
Agilent 8560EC, 8561EC, 8562EC, 8563EC: 16.5 kg (36 lbs)
Agilent 8564EC, 8565EC: 17.3 kg (38 lbs)
11
Agilent Email Updates
t
www.agilent.com/fi nd/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direc
www.agilent.com/fi nd/agilentdirect
Quickly choose and use your test
equipment solutions with confi dence.
Agilent
Open
www.agilent.com/fi nd/open
Agilent Open simplifi es the process
of connecting and programming
test systems to help engineers
design, validate and manufacture
electronic products. Agilent offers
open connectivity for a broad range
of system-ready instruments, open
industry software, PC-standard I/O
and global support, which are
combined to more easily integrate
test system development.
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confi dence in your measurements.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/fi nd/removealldoubt
www.agilent.com
For more information on Agilent Technologies’
products, applications or services, please
contact your local Agilent offi ce. The complete
list is available at:
www.agilent.com/fi nd/contactus
Americas
Canada (877) 894-4414
Latin America 305 269 7500
United States (800) 829-4444
Asia Pacifi c
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Thailand 1 800 226 008