Agilent 85133E Operating Guide

Operating and Service Manual
Agilent Technologies
85133E/F/H
NMD-2.4 mm -f- to 2.4 mm
Flexible
Test Port Return Cables
Agilent Part Number: 85133-90017
Printed in USA
Edition 5
Supersedes: October 2007
© Copyright 1990, 2006 - 2008 Agilent Technologies, Inc. All rights reserved.
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARA TE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
ii 85133E/F/H
CAUTION
The cable center conductor is fragile and will be seriously damaged if the cable is stretched, bent too tightly, or bent too oft en. Cables break when the bend radius is too small—less than 2.5” (6 cm) for 1.85 mm cables, and less than 4” (10 cm) for 2.4 mm cables. Use the following guidelines with cables:
• Never coil the cable too tightly.
• Never allow the cable to hang down from an instrument test port, or a storage fixture, or while holding it.
• Never bend the cable an excessive number of times.
85133E/F/H iii
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Agilent web site:
•Go to http://www.agilent.com.
• Enter the document’s part number (located on the title page) in the Search box.
• Click Search.
• Click on the hyperlink for the document.
• Click the printer icon located in the tool bar.
iv 85133E/F/H
1. General Information
The Cable Sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
85133E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
85133F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
85133H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Connector Designators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Clarifying the Terminology of a Connector Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Incoming Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Replaceable Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
2. Specifications
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Electrical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Supplemental Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Center Conductor Pin Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
3. Use, Maintenance, and Care of the Cables
Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Look for Obvious Defects and Damage First. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Inspect the Mating Plane Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Precision Slotless Connectors (2.4 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Cleaning Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Gaging Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Connector Gage Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
When to Gage Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Gaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
How to Make a Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
How to Separate a Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Avoiding Cable Movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Contents
4. Replaceable Par ts
Ordering Replaceable Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Ordering One Cable in a Cable Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Returning a Cable or Cable Set to Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Contacting Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Information About Network Analyzers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
A. Connector Care Quick Reference
Principles of Microwave Connector Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
85133E/F/H
1
Contents
2
85133E/F/H

1 General Information

85133E/F/H 1-1
General Information

The Cable Sets

The Cable Sets
The Agilent 85133E/F/H flexible cable sets are designed for use with Agilent network analyzers that have male NMD-2.4 mm test ports.

85133E

The 85133E set contains a single cable. Use it when a DUT is connected, with the appropriate adapter, directly to the test set test port. See Figure 1-1.

85133F

The 85133F set contains two cables. Each is shorter than the 85133E cabl e. Use the 85133F cables when a DUT is connected between cable ends. See Figure 1-1.
Figure 1-1 Cable Configurations

85133H

The 85133H set contains a single cable – the -f- to -m- cable from the 85133F cable set. It is shorter than the 85133E single cable. See Figure 1-1.
Connector Designators NMD Connectors
NMD denotes a connector that has larger than standard coupling threads for greater stability. NMD connectors are used on test cables, test ports, test port adapters, and test port cables.
Female slotted NMD test port connectors are used on the test set end of adapters and cables, and cannot be connected to standard male connectors.
Male NMD connectors are used on tes t sets (as test p orts), and on the DUT end of adapters and cables. Male connectors have both the larger threads (for use with test port adapters) and standard threads (for direct coupling to a DUT).
1-2 85133E/F/H
General Information

Incoming Inspection

PSC Connectors
PSC denotes a precision slotless connector. Precision slotless connectors are metrology grade connectors that have better electrical performance, better repeatability, and are more durable than slotted connectors.

Clarifying the Terminology of a Connector Interface

In this document and in the prompts of the PNA calibration wizard, the gender of cable connectors and adapters is referred to in terms of the center conductor. For example, a connector or device designated as 1.85 mm –f– has a 1.85 mm female center conductor.
8510-series , 872 x , a nd 87 5x ON LY: In contrast, during a measurement calibration, the network analyzer softkey menus label a 1.85 mm calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) refers to the short that is to be connected to the fem a le test port. This will be a male short from the calibration kit.
Table 1-1 Clarifying the Sex of Connectors: Examples
Terminology Meaning
Short
–f–
Short (f) Male short (male center conductor) to be connected to female port
A connector gage is referred to in terms of the connector that it measures. For instance, a male connector gage has a female connector on the gage so that it can measure male devices.
Female short (female center conductor)
Incoming Inspection
Verify that the shipment is complete by referring to Table 4-1 on page 4-4. If the case or any device appears damaged, or i f the shipment is incompl ete, not ify Agilent.
See “Contacting Agilent” on page 4-3. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company.
When you send the cable set to Agilent, include the following information:
• your company name and address
• the name of a technical contact person within your company, and the person’s complete phone number
• the model number and serial number of the cable set
85133E/F/H 1-3
General Information

Preventive Maintenance

• the part number and serial number of the device
• the type of ser v i ce re q u i red
•a detailed description of the problem
Preventive Maintenance
The best techniques for maintaining the integrity of the cables include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
• proper storage All of the above are described in Chapter 3. Failure to detect and remove dirt or metallic
particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections, resulting from center conductor values being out of limits (see Table 2 on page 2-4), or from bad connections, can also damage cable connectors.

Replaceable Parts

See Table 4-1 on page 4-4 for a complete list of cable set contents and their associated part numbers (you can order a single cable rather than the entire cable set), and for recommended items not included with the cable set. Refer to “Contacting Agilent” on
page 4-3 for information about communicating with Agilent.
1-4 85133E/F/H

2 Specifications

85133E/F/H 2-1
Specifications

Environmental Requirements

Environmental Requirements
Table 2-1 Environmental Requirements
Parameter Limits
Operating temperature +20 °C to +26 °C (+68 °F to +79 °F) Storage temperature −40 °C to +75 °C (−40 °F to +167 °F) Altitude
Operation < 4,500 meters (15,000 feet) Storage < 4,500 mete rs (15,000 feet)
Relative humidity Always non-condensing
Operation Up to 80% at 30°C Storage Up to 95% at 40°C

Electrical Specifications

Table 2-2 Electrical Specifications
Cable SWR Return Loss
(dB)
85133E ≤1.44 15 ≤0.1 + 0.57√f
85133F ≤0.1 + 0.38√f
85133H
a. f = frequency in GHz.
Insertion Loss (dB)
+ 0.02f
+ 0.017f
Frequency Range
a
(GHz)
DC to 50
2-2 85133E/F/H
Specifications

Supplemental Characteristics

Supplemental Characteristics
T able 2-3 lists supp lementa l perfor mance char acte ristic s. These are not specifi cati ons, but
are intended to provide additional information useful to your application. Supplemental characteristics are typical (but not warranted) performance parameters.
Table 2-3 Supplemental Characteristics (1 of 3)
Cable Cable Length Approximate
Electrical Length
cm in m ft cm in
85133E 97.2 38.25 1.125 3.690 <0.15 dB Change
85133F 62.9 24.75 0.7376 2.418 <0.08 dB Change
85133H
a. (f) = frequency in GHz. b. With a 90°, 2.5-inch bend radius. c. After three 90°, 2.5-inch bend radius/straighten cycles.
Magnitude and Phase Stability With a 90°
a,b
Bend
<0.17° (f) + 0.5°
<0.08° (f) + 0.5°
Random Use Magnitude and
Phase Stability
<0.07 dB Change <0.09° (f) + 0.5°
<0.04 dB Change <0.04° (f) + 0.5°
Table 2-3 Supplemental Characteristics (2 of 3)
Cable Set N umber
of Cables
85133E 1 NMD-2.4 mm -f- Slotted PSC-2.4 mm -f- Slotless
Test Set End Connector Typ e DUT End Connector Type
Minimum Recommended
a,c
Bend Radius
6.352 2.5
85133F 2 NMD-2.4 mm -f- Slotted NMD-2.4 mm -m- and
PSC-2.4 mm -f- Slotless
85133H 1 NMD-2.4 mm -f- Slotted NMD-2.4 mm -m-

Center Conductor Pin Depth

Center conductor pin depth is the distance the center conductor mating plane differs from being flush with the outer conductor mating plane. See Figure 2-1 The pin depth of a center conductor can be in one of two states: either protruding or recessed.
Protrusion is the condition in which the center conductor extends beyond the outer conductor mating plane. This c ondition will indi cate a pos itive value on the c onnector gage.
Recession is the condition in which the center conductor is set back from the outer conductor mating plane. This condition will indi cate a negative value on the connector gage.
85133E/F/H 2-3
Specifications
Supplemental Characteristics
Figure 2-1 Connector Center-Conductor Pin Depth
Table 2-3 Supplemental Characteristics (3 of 3)
Center-Conductor Pin Depth
Precision
Allowable Recession
a
Connector
mm in mm in
NMD-2.4 mm -f- -0.000 to -0.056 -0.0000 to -0.0022 0.0000 0.0000 NMD-2.4 mm -m- -0.0025 to -0.0254 -0.0001 to -0.001 PSC-2.4 mm -f- -0.0025 to -0.0254 -0.0001 to -0.001
a. Center conductor shoulder behind outer conductor mating plane.
Allowable Protrusion
2-4 85133E/F/H
3 Use, Maintenance, and Care of the
Cables
85133E/F/H 3-1
Use, Maintenance, and Care of the Cables

Electrostatic Discharge

Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static -sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static disc harges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test (DUTs), can also carry an ele ct rostatic charge. To pre v e n t da mage to the test se t, componen ts, and device s:
always wear a grounded wrist strap having a 1 MΩ resistor in series with it when handling components and devices or when making connections to the test set.
always use a grounded, conductive table mat while making connections.
always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or t est p ort. You ca n, for example , grasp t he ground ed oute r s hell of the test port or cable connector briefly.
always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the cent er
conducto r to th e ou ter conduct o r.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Refer to Chapter 4 for part numbers and instructions for ordering ESD protection devices.
Figure 3-1. ESD Protection Setup
3-2 85133E/F/H
Use, Maintenance, and Care of the Cables

Visual Inspection

Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles fro m the connector th reads may fall onto the mating plane surf ace of the connect o r w h e n it is di sconnected. One connect i on ma de with a dirty or damaged connector can damage both connectors beyond re pair.
Magnification is helpful when inspecting connectors, but it is not required and may actually be misleading. Defects and damage that cannot be seen without magnification generally have no effect on electrical or mechanical performance. Magnification is of great use in analyzing the nature and cause of the damage and in cleaning connectors, but it is not required for inspect ion. Use t he follo wing guidel ines when evaluating t he integrity of a connector.

Look for Obvious Defects and Damage First

Examine the connector first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the cable . The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are connected incorrectly.
Connector wear eventually degrades performance of the cable. Replace cables with worn connectors.
The test port connectors on the network analyzer test se t may have many connections each day, and are therefore also subject t o wear. It is recommended that an adapter be used a s a test port saver to minimize the wear on the test s e t’s test port connectors.

Inspect the Mating Plane Surfaces

Flat contact between the connectors at all points on their mating pl ane surfaces is required for a good connection. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to
excessive or uneven wear or misalignment. Light burnishing of the mating plane surfaces is normal, and is evident as light scratches
or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical per formance.
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Cables with damaged connectors should be repaired or discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration.
85133E/F/H 3-3
Use, Maintenance, and Care of the Cables

Precision Slotless Connectors (2.4 mm)

Precision Slotless Connectors (2.4 mm)
The female 2.4 mm connectors in the cable sets are metrology-grade, precision slotless connectors (PSC). Precision slot less connectors are used to improve accuracy. A characteristic of metrology-grade connectors is directly traceability to national measurement standards through the ir we ll-def ined mechanic al d imensions . With PSCs on test ports and standards, the accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
Conventional female center conducto rs are slott ed and, whe n mated, are fl ared by the male pin. Because physical dimensions determine connector impeda nce, this change in physical dimension affects electrical performance, making it very difficult to perform precision measurements with conventional slotted connectors .
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, the center conductor has an internal con tact that flexes to
accept the male pin.
3-4 85133E/F/H
Use, Maintenance, and Care of the Cables

Cleaning Connectors

Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
1. Use Compressed Air or Nitrogen
WARNING
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid cond ensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can caus e electrostatic effects when di rected into a connecto r. These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information.
2. Clean the Connector Threads
WARNING
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absor bed through the s kin. Wash thoroughly after handling.
Always use protective eyewear when using compressed air or nitrogen.
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container . It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to T able 4 -1 on page 4-4 for part numbers for isopropyl alcohol and cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab. b. Clean the connector threads. c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compresse d air or nitroge n . A lways comple te l y dr y a co n n e ctor before
85133E/F/H 3-5
Use, Maintenance, and Care of the Cables
Cleaning Connectors
you reassemble or use it.
3. Clean the Mating Plane Surfaces a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 2-1 on
page 2-4. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, t hen bl ow the co nnect or dry wit h a gentle stream o f clean,
low-pressure compresse d air or nitroge n . A lways comple te l y dr y a co n n e ctor before you reassemble or use it.
4. Reinspect Inspect the connector again to make sure that no particles or resi due are present.
3-6 85133E/F/H
Use, Maintenance, and Care of the Cables

Gaging Connectors

Gaging Connectors
The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. (See T able 4-1 on page 4-4 for part number information.) They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on test cables and other accessories, DUTs, and calibration kit devices. Do not use the gages for precise pin depth measurements.

Connector Gage Accuracy

The connector gages are only capable of performing coarse measurements. They do not provide the degree of accuracy necessary to pr ecisely measure the pin depth of the cable connectors. This is partially due to the repeatabil ity uncertain ties that are assoc iated with the measurement. Only the factory—through special gaging processes and electrical testing— can accurately verify the mechanical characteris tics of the cable connectors.
With prope r technique, however , the gages are us eful in detecting gross pin depth errors on cable connectors. To achieve maximum accuracy, random errors must be re duced by taking the average of at least three measurements having different gage orientations on the connector. Even the resultant average can be in error by as much as ± 0.0001 inch due to systematic (biasing) errors usually resulting from worn gages and gage masters. The information in Table 2-3 on page 2-4 assumes new gages and gage masters. As your gages undergo more use, their systematic errors can become more significant in the accuracy of the measurement.
NOTE
When measuring pin depth, the measured value (resultant average of three or more measurements) contains measurement uncertainty and is not necessarily the true value. Always compare the measured value with the observed pin depth limits (which account for measurement uncertainties) in
Table 2 on page 2-4 to evaluate the condition of cable connectors.

When to Gage Connectors

Gage a connector at the following times:
• Prior to using a cable fo r the firs t time: record the pin dep th measureme nt so that i t can be compared with future read ings. (It will s erve as a good t roubleshooting too l when you suspect damage may have occurred to the cable connector.)
• If either visual inspection or electrical performance suggests that the connector interface may be out of typical range (due to wear or damage, for example).
• If a cable is used by someone else or on another system or piece of equipment.
• Initially after every 100 connections, and after that as often as experience indicates.
85133E/F/H 3-7
Use, Maintenance, and Care of the Cables
Gaging Connectors

Gaging Procedures

Gaging 2.4 mm Connectors
NOTE
Always hold a connector gage by the gage barrel, below the dial indicator. This gives the best stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
1. Select th e p r oper gage for yo u r connector. Refer to Table 4-1 on page 4-4 for gage part numbers.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-2): a. While holding the gage by the barrel, and witho ut turning the gage or the device,
connect the gage to the gage master by interconnecting the male and female connectors. Connect the nut finger tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torq ue
wrench recommended for use to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a Torque Wre nch” on page 3-10 for additional information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set knob until the gage pointer lines up exactly with the zero mark.
d. Remove the gage master.
4. Gage the device connector (refer to Figure 3-2): a. While holding the gage by the barrel, and witho ut turning the gage or the device,
connect the gage to the device by interconnecting the male and female connectors. Connect the nut finger-tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torq ue
wrench included in the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a Torque Wre nch” on page 3-10 for additional
information. c. Gently tap the barrel of the gage with your finger to settle the gage reading. d. Read the gage indicator dial. Read only the black ± signs; not the red ± signs.
For maximum accuracy, measure the connector a minimum of three times and take
an average of the readings. After each measurement, rotate the gage a quarter-turn
to reduce measurement variations that result from the gage or t he connector face no t
being exactly perpendicular to the center axis. e. Compare the average reading with the observed pin depth limits in Table 2-3 on page
2-4.
3-8 85133E/F/H
Figure 3-2 Gaging 2.4 mm Connectors
Use, Maintenance, and Care of the Cables
Gaging Connectors
85133E/F/H 3-9
Use, Maintenance, and Care of the Cables

Making Connections

Making Connections
Good connections require a skilled operat or. The most common cause of meas ure men t erro r is bad connections. The following procedures illustrate how to make good connections.

How to Make a Connection

Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table mat. Refer to “Electrostatic Discharge” on page 3-2 for ESD precautions.
2. Visually ins pect the connectors. Refer to “Visual Inspectio n” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-5.
4. Use a connector gage to verify that all center conductors are within the observed pin depth values in Table 2-3 on page 2-4. Refer to “Gaging Connectors” on page 3-7.
5. Carefully align the connectors. The male connector center pin must slip concentrically into the contact finger of the female connector.
6. Push the connectors straight together and tight en the connector nut finger-tight.
CAUTION
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplis h this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables.
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection. Table 3-1 provides information about the torque wrench recommended for use with the calibration kit. Refer to Table 4-1 on page 4-4 for replacement part number and ordering information.
Table 3-1 Torque Wrench Information
Do not turn the device body. Only turn the connector nut. Damage to the center conductor can oc c u r i f th e device body is twisted.
Connector Type Torque Setting Torque Tolerance
2.4 mm 90 N-cm (8 in-lb) ±9.0 N-cm (±0.8 in-lb)
3-10 85133E/F/H
Use, Maintenance, and Care of the Cables
Making Connections
Using a torque wrench guarantees that t he co nnection is not too tight, preventi ng pos sible connector damage. It also guarantees that all connections are equa lly tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible to do this by hand if one of the connectors is fixed (as on a test port). However, it is recommended that you use an open-end wrench to keep the body of the device from turning.
1. Position both wrenches within 90 degrees of each other before applying force. See
Figure 3-3. Wrenches opposing each other (greater than 90 degrees apart) will cause a
lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together.
Figure 3-3 Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See
Figure 3-4.
Figure 3-4 Using the Torque Wrench
3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque.
85133E/F/H 3-11
Use, Maintenance, and Care of the Cables

Handling and Storage

4. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-4. Do not tighten the connection further.
CAUTION
You don’t have to fully break the handle of the torque wrench to reach the specified torque; doing so can cause the handle to kick back and loosen the connection. Any give at all in the handle is sufficient torque.

How to Separate a Connection

To avoid lateral (bending) force on the connec tor mating pla ne surfaces, always support the devices and connections.
CAUTION
1. Use an open-end wrench to prevent the device body from turning.
2. Use the torque wrench to loosen the connecting nut.
3. Complete the separation by hand, turning only the connect ing nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.
Do not turn the device body. Only turn the connector nut. Damage to the center conductor can oc c u r i f th e device body is twisted.
Handling and Storage
• Install the protective end caps and s tore the ca bles in the fo am-lined storage ca se when not in use.
• Never store cables loose in a box, desk, or bench drawer. This is the most common cause of cable connector damage during storage.
• Store cables in the same shape the y have when you use them; d o not eith er straighte n a cable or flex it more tightly. Even flexible cables last longer if you flex them as little as possible.
• Keep cable connectors clean.
• Do not touch connector mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove.
• Do not set cable connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.
3-12 85133E/F/H
Use, Maintenance, and Care of the Cables
Handling and Storage

Avoiding Cable Movement

When you use cables to m ake a precise calibration, you may have to fixture the cables to prevent excessive movement after the calibration. In some cases, unless you res trict cable movement, you may not be able to perform a verification after the calibration, especially if you use a precision calibration kit.
NOTE
After you perform a calibration, move the cables as little as possible. Every time you bend a cable, the phase changes slightly.
85133E/F/H 3-13
Use, Maintenance, and Care of the Cables
Handling and Storage
3-14 85133E/F/H

4 Replaceable Parts

85133E/F/H 4-1
Replaceable Parts

Ordering Replaceable Parts

Ordering Replaceable Parts
Table 4-1 on page 4-4 lists the replacement part numbers for items included in the
85133E/F/H cable sets. To order a listed part, note the description, the part number, and the quantity desired.
Telephone or send your order to Agilent Technologies (see “Contacting Agilent” on
page 4-3).

Ordering One Cable in a Cable Set

If you need only one of the cables in a cable set and don’t want to order an entire set, use the appropriate single cable part number listed in Table 4-1 on page 4-4. When you order using a single cable part number, you do not get a pair of cables; be sure you order the correct cable part number(s).
4-2 85133E/F/H
Replaceable Parts

Returning a Cable or Cable Set to Agilent

Returning a Cable or Cable Set to Agilent
If your cable set requires service, contact Agilent Technologies (see below). Include the following information:
• your company name and address
• a technical contact person within your company, and the person's complete phone number
• the model number and serial number of the cable set
• the part number and serial number of each device
• the type of ser v i ce re q u i red
•a detailed description of the problem and how the device was being used when the problem occurred (such as calibration or measurement)

Contacting Agilent

Assistance with test and measurement needs and information on finding a local Agilent office are available on the Web at:
http://www.agilent.com/find/assist
If you do not have access to the Internet, please contact your Agilent field engineer.
NOTE
In any correspondence or telephone conversation, ref er to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period.

Information About Network Analyzers

This document contains limited information about network analyzer system operation. For complete information, refer to the instrument documentation.
85133E/F/H 4-3
Replaceable Parts
Returning a Cable or Cable Set to Agilent
Table 4-1 Replaceable Parts
Descriptiona
Cables
85133E Single Cable:
NMD-2.4 mm –f– to PSC-2.4 mm –f– 1 85133-60015
85133F Cables:
NMD-2.4 mm –f– to NMD-2.4 mm –m– 1 85133-60017 NMD-2.4 mm –f– to PSC-2.4 mm –f– 1 85133-60016
85133H Single Cable:
NMD-2.4 mm –f– to NMD-2.4 mm –m– 1 85133-60017
Miscellaneous
Operating and Service Manual 1 85133-90017 Protective End Caps
STR 0.812-ID black as STR 0.234-ID red 1401-0202
Items NOT Included in a Cable Set:
Qty Agilent Part Number
1401-0214
necessary
ESD Protective Devices
Grounding wrist strap 1 9300-1367 5 ft grounding cord for wrist strap 1 9300-0980 2 ft by 4 ft conductive table mat with 15 ft grounding
wire ESD heel strap 1 9300-1308
Connector Cleaning Supplies
Isopropyl alcohol 30 ml 8500-5344 Foam tipped cleaning swabs 100 9301-1243
Wrenches
20 mm (8 in–lb) torque wrench 1 8710-1764 5/16 in., 90 N–cm (8 in–lb) torque wrench 1 8710-1765 7 mm open-end wrench 1 8710-1761
Miscellaneous
2.4 mm female gage set 1 85056-60017
2.4 mm male gage set 1 85056-60018
1 9300-0797
10x Magnifying Glass 1 1000-1114
4-4 85133E/F/H
Table 4-1 Replaceable Parts
Replaceable Parts
Returning a Cable or Cable Set to Agilent
Descriptiona
Qty Agilent Part Number
2.4 mm 50 Ω fixed termination -m- 1 00901-60001
2.4 mm 50 Ω fixed termination -f- 1 00901-60002
2.4 mm offset short -m- 1 85056-60001
2.4 mm offset short -f- 1 85056-60002
2.4 mm airline (5.0 cm) 1 85057-60001
a. Refer to “Clarifying the Terminology of a Connector Interface” on page 1-3.
85133E/F/H 4-5
Replaceable Parts
Returning a Cable or Cable Set to Agilent
4-6 85133E/F/H

A Connector Care Quick Reference

85133E/F/H A-1
Connector Care Quick Reference

Principles of Microwave Connector Care

Principles of Microwave Connector Care
Proper connector care and connection techniques are critical for accurate, repeatable measurements and for extending the life of your devices.
Prior to making connections to the network analyzer, carefully review the connector care information provided with your product.
See the following table for quick reference tips about connector care.
A-2 85133E/F/H
Connector Care Quick Reference
Principles of Microwave Connector Care
Table A-1 Connector Care Quick Reference
Handling and Storage
Do Do Not
Keep connectors clean Touch mating-plane surfaces Extend sleeve or connector nut Set connectors contact — end down Use plastic end-caps during storage
Visual Inspection
Do Do Not
Inspect all connectors carefully Use a damaged connector — ever Look for metal particles, scratches, and dents
Connector Cleaning
Do Do Not
Tr y compressed air first Use any abrasives Use isopropyl alcohol
a
Get liquid into plastic support beads
Clean connector threads
Gaging Connectors
Do Do Not
Clean and zero the gage before use Use an out-of-spec connector Use the correct gage type Use correct end of calibration block Gage all connectors before first use
Making Connections
Do Do Not
Align connectors carefully Apply bending force to connection Make preliminary connection lightly Over tighten preliminary connection Turn only the connector nut Twist or screw any connection Use a torque wrench for final connect Tighten past torque wrench “break” point
a. Use isopropyl alcohol in a well-ventilated area, allowing adequate time for moist
alcohol to evaporate and fumes to disperse prior to energizing equipment.
85133E/F/H A-3
Connector Care Quick Reference
Principles of Microwave Connector Care
A-4 85133E/F/H

Index

A
adapters, part numbers Agilent,contacting, 4-3 alcohol, isopropyl, as cleaning solvent, 3-5 altitude, 2-2 avoiding cable movement, 3-13
C
cable
, 2-3
length maintenance, 1-4 movement, 3-13 part numbers, 4-4
cable set
overview return to Agilent, 4-3
center conductor pin depth, 2-3 characteristics
supplemental
cleaning connectors, 3-5 cleaning supplies
ordering, 4-4 part numbers, 4-4
communicating with Agilent, 4-3 compressed air or nitrogen, 3-5 conductor
mating plane
connections, 3-2, 3-10
ESD protection, 3-10 final, 3-10 preliminary, 3-10 separating, 3-12 using torque wrench, 3-10
connector
care, A-2 cleaning, 3-5 cleaning supplies, 4-4 damage, 3-3 gage
accuracy handling, 3-8 when to do, 3-7
zeroing, 3-8 gaging, 3-8 life, 3-3 mating plane, 3-6 NMD type, 1-2 PSC type, 1-3 quick reference, A-3 sex, 1-3 slotless, 3-4 threads, 3-5 type
DUT end
test set end, 2-3 visual inspection, 3-3 wear, 3-3
connector terminology, 1-3 contacting Agilent, 4-3
, 1-2
, 2-3
, 2-3
, 3-7
, 2-3
, 4-4
D
damage
to connectors to shipment, 1-3
defective connectors, 3-3 description of cable sets, 1-2 device
conductor
mating plane
connecting, 3-10 disconnecting, 3-12 handling, 3-12 part numbers, 4-4 return to Agilent, 4-3 storage, 3-12 visual inspecti on, 3-3
dimensions
device
outer conductor
disconnections, 3-12 documentation warranty, -ii
E
electrical length electrical specifications, 2-2 electrostatic discharge, 3-2
supplies
part numbers
environmental requirements, 2-2 equipment
supplied, 4-2, 4-4
ESD, 3-2
precautions, 3-2, 3-5 supplies
part numbers
F
frequency range
specifications
G
gage
connector
handling master, 4-4
zeroing, 3-8
gage master
part numbers
gaging
connectors when to do, 3-7
gender, connector, 1-3
H
handling humidity, 2-2
, 3-3
, 2-3
, 2-3
, 2-3
, 4-4
, 4-4
, 2-2
, 3-8
, 4-4
, 3-8
, 3-12
85133E/F/H
Index-1
Index
I
incoming inspection insertion loss
specificatons
inspection
damage defects, 3-3 incoming, 1-3 mating plane, 3-3 visual, 3-3
isopropyl alcohol, as cleaning solvent, 3-5
L
length
cable electrical, 2-3
M
magnitude maintenance, 3-2
preventive, 1-4
making connections, 3-10
ESD protection, 3-10 precautions, 3-10
manual
part number, 4-4
mating plane
conductor, 2-3 connector, 3-6 inspection, 3-3 surfaces, 3-6
movement of cables, 3-13
N
nitrogen NMD connectors, 1-2
O
open-end wrench
part number, 4-4
ordering parts, 4-2
P
parts
numbers ordering, 4-2 replaceable, 4-2 required but not supplied, 4-4
phase stability, 2-3 pin depth, 2-3
definition of, 2-3 protrusion, 2-3 recession, 2-3
preventive maintenance, 1-4 protrusion
pin depth, 2-3
PSC connectors, 1-3
, 3-3
, 2-3
, 2-3
, 3-5
, 4-2
, 1-3
, 2-2
, 3-12
Q
quick reference, connector care
R
reaching Agilent recession
pin depth
regulations
environmental
replaceable parts, 4-2 requirements
environmental, 2-2
return loss
specifications, 2-2
return, set or cable, 4-3
S
separating connections set
overview return to Agilent, 4-3
sex, connector, 1-3 shipment
damage verifying complete, 1-3
specifications, 2-2
altitude, 2-2 electrical, 2-2 environmental, 2-2 frequency, 2-2 frequency range, 2-2 humidity, 2-2 insertion loss, 2-2 return loss, 2-2 temperature, 2-2 torque wrench, 3-10
stability, phase, 2-3 static discharge, 3-2 storage, 3-12
altitude, 2-2 humidity, 2-2 temperature, 2-2
supplemental characteristics, 2-3
T
temperature
operating range
threads
connector
torque wrench
part number specifications, 3-10
V
visual inspection, 3-3
W
warranty, documentation, -ii wrench
, 1-3
, 4-3
, 2-3
, 3-6
, 3-12
, 1-2
, 2-2
, 3-5
, 4-4
, A-3
Index-2
85133E/F/H
Index
open-end part numbers, 4-4 proper positioning of, 3-11 torque, 3-10, 3-11, 4-4
precautions for use of, 3-11
proper use of, 3-11
Z
zeroing
connector gage
, 3-11, 3-12, 4-4
, 3-8
85133E/F/H
Index-3
Loading...