Agilent 85052D Service Manual

User’s and Service Guide
Agilent Technologies 85052D
3.5 mm Economy Calibration Kit
This manual directly applies to 85052D calibration kits with serial number prefix
3106A.
Agilent Part Number: 85052-90079
Printed in USA
Print Date: September 2007
© Copyright 1994–1995, 2000–2002, 2004, 2007 Agilent Technologies, Inc. All rights reserved.
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FUR­THER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CON­NECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCU­MENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
ii
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Agilent web site:
•Go to http://www.agilent.com.
Enter the product model number in the search function and click Search.
Click on the Manuals hyperlink.
Open the PDF of your choice and print the document.
iii
iv
Contents
1 General Information
Calibration Kit Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Calibration Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Equipment Required but Not Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Recording the Device Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Calibration Kits Documented in This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Calibration Kit History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
85052D Kits with Serial Prefix 3027A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Precision Slotless Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Clarifying the Terminology of a Connector Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
2 Specifications
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Temperature—What to Watch Out For . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Mechanical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Pin Depth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Residual Errors after Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
3 Use, Maintenance, and Care of the Devices
Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Look for Obvious Defects and Damage First . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Inspect the Mating Plane Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Inspect Female Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Cleaning Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Gaging Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Connector Gage Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
When to Gage Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Gaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
How to Make a Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
How to Separate a Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
4 Performance Verification
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
How Agilent Verifies the Devices in Your Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Recertification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
How Often to Recertify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Where to Send a Kit for Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
5 Troubleshooting
Troubleshooting Process. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
85052D Contents-v
Contents
Where to Look for More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Returning a Kit or Device to Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Contacting Agilent. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
6 Replaceable Parts
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
A Standard Definitions
Class Assignments and Standard Definitions Values are Available on the Web. . . . . . . . . .A-2
Contents-vi 85052D

1 General Information

85052D 1-1

General Information

Calibration Kit Overview

Calibration Kit Overview

The Agilent 85052D 3.5 mm calibration kit is used to calibrate Agilent network analyzers up to 26.5 GHz for measurements of components with 3.5-mm connectors.

Kit Contents

The 85052D calibration kit includes the following items:
• user’s and service guide
• offset opens and shorts, and broadband loads
• three 3.5 mm adapters
• 5/16 in, 90 N-cm (8 in-lb) torque wrench
• 7 mm open-end wrench
• data disk that contains the calibration definitions of the devices in the calibration kit for 8510 systems.
• data disk set that contains the specifications and performance verification data for 8510 systems.
NOTE
For measurement convenience, the kit also contains three 3.5 mm adapters. The adapters are primarily intended for use in measuring non-insertable devices, but can also be used as a connector saver.
Refer to Chapter 6 for a complete list of kit contents and their associated part numbers.
Broadband Loads
The broadband loads are metrology-grade terminations that have been optimized for performance up to 26.5 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.
Offset Opens and Shorts
The offset opens and shorts are built from parts that are machined to the current state-of-the-art in precision machining.
A backup copy of each data disk and printout should be made immediately upon receipt of the calibration kit. Refer to your analyzer user’s guide or embedded Help system for instructions on duplicating a disk.
The offset short’s inner conductors have a one-piece construction, common with the shorting plane. The construction provides for extremely repeatable connections.
The offset opens have inner conductors that are supported by a strong, low-dielectric constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts are designed so that the difference in phase of their reflection coefficients is approximately
1-2 85052D
General Information
Calibration Kit Overview
180 degrees at all frequencies.
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance and to ensure stable, repeatable connections.
The adapters are designed so that their nominal electrical lengths are the same, which allows them to be used in calibration procedures for non-insertable devices.

Calibration Definitions

The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration.
The calibration definitions can be:
• resident within the analyzer
• downloaded from the Web at http:na.tm.agilent.com/pna/caldefs/stddefs.html.
• loaded from the provided disk (8510 series network analyzers only)
• manually entered from the front panel
Refer to your network analyzer user’s guide or embedded Help for instructions on manually entering calibration definitions, selecting the calibration kit, and performing a calibration.

Equipment Required but Not Supplied

Some items are required or recommended for successful operation of your kit, but are not supplied with the kit. Refer to Table 6-2 on page 6-3 for a list of these items and for ordering information.
85052D 1-3
General Information

Incoming Inspection

Incoming Inspection

Refer to Chapter 6 to verify a complete shipment. Use Table 1-1 to record the serial numbers of all serialized devices in your kit.

Check for damage. The foam-lined storage case provides protection during shipping. If the case or any device appears damaged, or if the shipment is incomplete, refer to “Contacting

Agilent” on page 5-5. Agilent will arrange for repair or replacement of incomplete or
damaged shipments without waiting for a settlement from the transportation company. See "Returning a Kit or Device to Agilent," on page 5-4.

Recording the Device Serial Numbers

In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in

Table . Recording the serial numbers will prevent confusing the devices in this kit with
similar devices from other kits.

The adapters included in the kit are for measurement convenience only and are not serialized.

Table 1-1 Serial Number Record for the 85052D
Device Serial Number
Calibration kit
Broadband load –m–
Broadband load –f–
Open –m–
Open –f–
Short –m–
Short –f–
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
1-4 85052D
General Information

Calibration Kits Documented in This Manual

Calibration Kits Documented in This Manual

This manual applies to any 8552B calibration kit with serial number prefix 3106A. If your calibration kit has a different serial number prefix, refer to the next section for information on how this manual applies.

Calibration Kit History

This section describes calibration kits with serial number prefixes lower that the ones listed on the title page.

85052D Kits with Serial Prefix 3027A

These calibration kits did not have the calibration constants disk to support the Agilent 8510C network analyzer. The part numbers provided in this manual are the recommended replacement parts for these kits. The devices in these kits should meet the specifications published in this manual.
85052D 1-5
General Information

Precision Slotless Connectors

Precision Slotless Connectors

The female 3.5 mm connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC). A characteristic of metrology-grade connectors is direct traceability to national measurement standards through their well-defined mechanical dimensions.
Conventional female center conductors are slotted. When mated, the female center conductor is flared by the male pin. Because physical dimensions determine connector impedance, electrical characteristics of the female connector (and connection pair) are dependent upon the mechanical dimensions of the male pin. While connectors are used in pairs, their male and female halves are always specified separately as part of a standard, instrument, or device under test. Because of these facts, making precision measurements with the conventional slotted connector is very difficult, and establishing a direct traceability path to primary dimensional standards is nearly impossible.
The precision slotless connector was developed to eliminate these problems. All PSCs are female. A PSC incorporates a center conductor with a solid cylindrical shell that defines the outside diameter of the female center pin. Its outside diameter and, therefore, the impedance in its region does not change. The inner part provides an internal contact that flexes to accept the allowed range of male pin diameters.
The calibration of a network analyzer having a conventional slotted female connector on the test port remains valid only when the device under test and all calibration standards have identical male pin diameters. For this reason PSC test port adapters are supplied in most calibration kits.

Precision slotless connectors have the following characteristics:

• There is no loss of traceable calibration on test ports when the male pin diameter of the connector on the device under test is different from the male pin diameter of the calibration standard.
• The female PSC and its mating male connector can be measured and specified separately as part of the device either is attached to.
• All female connectors can have a known, stable impedance based only on the diameters of their inner and outer conductors.
• Female calibration standards can be fully specified. Their specifications and traceability are unaffected by the diameter of the male mating pin.
• A fully traceable performance verification is made using a precision 50 ohm airline having a PSC.
• Measurement repeatability is enhanced due to non-changing connector characteristics with various pin diameters.
With PSCs on test ports and standards, the percentage of accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
1-6 85052D
General Information

Clarifying the Terminology of a Connector Interface

Clarifying the Terminology of a Connector Interface

In this document and in the prompts of the PNA calibration wizard, the gender of cable connectors and adapters is referred to in terms of the center conductor. For example, a connector or device designated as 1.85 mm –f– has a 1.85 mm female center conductor.
8510-series, 872x, and 875x ONLY: In contrast, during a measurement calibration, the network analyzer softkey menus label a 1.85 mm calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) refers to the short that is to be connected to the female test port. This will be a male short from the calibration kit.
Table 1-2 Clarifying the Sex of Connectors: Examples
Terminology Meaning
Short
–f–
Short (f) Male short (male center conductor) to be connected to female port

A connector gage is referred to in terms of the connector that it measures. For instance, a male connector gage has a female connector on the gage so that it can measure male devices.

Female short (female center conductor)

Preventive Maintenance

The best techniques for maintaining the integrity of the devices in the kit include:

• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques

All of these are described in Chapter 3. Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections, resulting from pin depth values being out of the observed limits (see Table 2-2 on page 2-4) or from bad connection techniques, can also damage these devices.

85052D 1-7
General Information
Clarifying the Terminology of a Connector Interface
1-8 85052D

2 Specifications

85052D 2-1
Specifications

Environmental Requirements

Environmental Requirements

NOTE
Samples of this product have been type-tested in accordance with the Agilent Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation and end-use; those stresses include but are not limited to temperature, humidity, shock, vibration, altitude and power-line conditions. Test methods are aligned with IEC
60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Table 2-1 Environmental Requirements
Parameter Limits
Temp er at ur e
Operating
Storage 40 °C to +75 °C
Error-corrected range
Altitude
Operating < 4,500 meters (15,000 feet)
Storage < 4,500 meters (15,000 feet)
Relative humidity Type tested, 0% to 95% at 40 °C, non-condensing
a
b
+20 °C to +26 °C
± 1 °C of measurement calibration temperature
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range shown in
Table 2-1.
IMPORTANT
Avoid unnecessary handling of the devices during calibration because your fingers are a heat source.
2-2 85052D
Specifications

Mechanical Characteristics

Mechanical Characteristics

Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Agilent Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion or improper pin depth when the kit leaves the factory.
“Gaging Connectors” on page 3-6 explains how to use gages to determine if the kit devices
have maintained their mechanical integrity. Refer to Table 2-2 for typical and observed pin depth limits.

Pin Depth

Pin depth is the distance the center conductor mating plane differs from being flush with the outer conductor mating plane. See Figure 2-1. The pin depth of a connector can be in one of two states: either protruding or recessed.
Protrusion is the condition in which the center conductor extends beyond the outer conductor mating plane. This condition will indicate a positive value on the connector gage.
Recession is the condition in which the center conductor is set back from the outer conductor mating plane. This condition will indicate a negative value on the connector gage.
Figure 2-1 Connector Pin Depth
85052D 2-3
Specifications
Mechanical Characteristics
The pin depth value of each calibration device in the kit is not specified, but is an important mechanical parameter. The electrical performance of the device depends, to some extent, on its pin depth. The electrical specifications for each device in the kit take into account the effect of pin depth on the device’s performance. Table 2-2 lists the typical pin depths and measurement uncertainties, and provides observed pin depth limits for the devices in the kit. If the pin depth of a device does not measure within the observed pin depth limits, it may be an indication that the device fails to meet electrical specifications. Refer to Figure 2-1 for a visual representation of proper pin depth (slightly recessed).
Table 2-2 Pin Depth Limits
Device
Opens 0 to 0.0127 mm
Shorts 0 to 0.0127 mm
Fixed loads 0.0025 to 0.0254 mm
Adapter 0.0025 to 0.0254 mm
a. Approximately +2 sigma to 2 sigma of gage uncertainty based on studies done at the
factory according to recommended procedures.
b. Observed pin depth limits are the range of observation limits seen on the gage reading due
to measurement uncertainty. The depth could still be within specifications.
Typical Pin Depth
0 to 0.00050 in
0 to 0.00050 in
0.0001 to 0.0010 in
0.0001 to 0.0010 in
Measurement Uncertainty
+0.0064 to 0.0064 mm +0.00025 to 0.00025 in
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in
a
Observed Pin Depth Limits
+0.0064 to 0.0191 mm +0.00025 to 0.00075 in
+0.0041 to 0.0168 mm +0.00016 to 0.00066 in
+0.0016 to 0.0295 mm +0.0006 to 0.00116 in
+0.0016 to 0.0295 mm +0.0006 to 0.00116 in
b
2-4 85052D
Specifications

Electrical Specifications

Electrical Specifications

The electrical specifications in Table 2-3 apply to the devices in your calibration kit when connected with an Agilent precision interface.

Table 2-3 Electrical Specifications for 85052D 3.5 mm Devices
Device Specification Frequency (GHz)
Broadband loads Return loss ≥ 46 dΒ (ρ ≤ 0.00501) dc to ≤ 2
(male and female) Return loss 44 dΒ (ρ ≤ 0.00631) > 2 to ≤ 3
Return loss 38 dB (ρ ≤ 0.01259) > 3 to ≤ 8
Return loss 36 dB (ρ ≤ 0.01585) > 8 to ≤ 20
Return loss 34 dB (ρ ≤ 0.01995) > 20 to ≤ 26.5
Offset opens
(male and female) ±1.20° deviation from nominal > 3 to ≤ 8
Offset shorts
(male and female) ±1.00° deviation from nominal > 3 to ≤ 8
Adapters Return loss ≥ 30 dΒ (ρ ≤ 0.03162) dc to ≤ 8
a. The specifications for the opens and shorts are given as allowed deviation from the
a
a
nominal model as defined in the standard definitions.
±0.65° deviation from nominal
±2.00° deviation from nominal > 8 to 20
±2.00° deviation from nominal > 20 to 26.5
±0.50° deviation from nominal
±1.75° deviation from nominal > 8 to 20
±1.75° deviation from nominal > 20 to 26.5
Return loss 28 dΒ (ρ ≤ 0.03981)
Return loss 26 dΒ (ρ ≤ 0.05012)
dc to ≤ 3
dc to ≤ 3
> 8 to 18
> 18 to 26.5
85052D 2-5
Specifications
Electrical Specifications

Residual Errors after Calibration

The 8510 “Specifications and Performance Verification” software can be used to obtain a printout of the residual errors after a calibration has been performed. Refer to the “Specifications and Performance Verification” section of the 8510C On-Site Service Manual for information on how to use the software.

Certification

Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Agilent Verifies the
Devices in Your Kit” on page 4-2 for more information.
2-6 85052D
3 Use, Maintenance, and Care of the
Devices
85052D 3-1
Use, Maintenance, and Care of the Devices

Electrostatic Discharge

Electrostatic Discharge

Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).

Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test (DUT), can also carry an electrostatic charge. To prevent damage to the test set, components, and devices:
always wear a grounded wrist strap having a 1 MΩ resistor in series with it when handling components and devices or when making connections to the test set.
always use a grounded, conductive table mat while making connections.
always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center
conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Figure 3-1 shows a typical ESD protection setup using a grounded mat and wrist strap.
Refer to Table 6-2 on page 6-3 for information on ordering supplies for ESD protection.
Figure 3-1 ESD Protection Setup
3-2 85052D
Use, Maintenance, and Care of the Devices

Visual Inspection

Visual Inspection

Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.
In some cases, magnification is necessary to see damage to a connector; a magnifying device with a magnification of 10x is recommended. However, not all defects that are visible only under magnification will affect the electrical performance of the connector. Use the following guidelines when evaluating the integrity of a connector.

Look for Obvious Defects and Damage First

Examine the connectors first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are not connected properly.
Connector wear eventually degrades performance of the device. Calibration devices should have a long life if their use is on the order of a few times per week. Replace devices with worn connectors.
The test port connectors on the network analyzer test set may have many connections each day, and are, therefore, more subject to wear. It is recommended that an adapter be used as a test port saver to minimize the wear on the test set’s test port connectors.

Inspect the Mating Plane Surfaces

Flat contact between the connectors at all points on their mating plane surfaces is required for a good connection. See Figure 2-1 on page 2-3. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance.
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration.
85052D 3-3
Use, Maintenance, and Care of the Devices

Cleaning Connectors

Inspect Female Connectors

Inspect the contact fingers in the female center conductor carefully. These can be bent or broken, and damage to them is not always easy to see. A connector with damaged contact fingers will not make good electrical contact and must be replaced.
NOTE
The female 3.5 mm connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC). Precision slotless connectors are used to improve accuracy. With PSCs on test ports and standards, the accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
Conventional female center conductors are slotted and, when mated, are flared by the male pin. Because physical dimensions determine connector impedance, this change in physical dimension affects electrical performance, making it very difficult to perform precision measurements with conventional slotted connectors.
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, this center conductor has an internal contact that flexes to accept the male pin.
This is particularly important when mating nonprecision to precision devices.

Cleaning Connectors

Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.

1. Use Compressed Air or Nitrogen

WARNING
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces. Clean air cannot damage a connector or leave particles or residues behind.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause electrostatic effects when directed into a connector. These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information.
3-4 85052D
Always use protective eyewear when using compressed air or nitrogen.
2. Clean the Connector Threads
Use, Maintenance, and Care of the Devices
Cleaning Connectors
WARNING
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-2 on page 6-3 for a part number for cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 2-1 on
page 2-3. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
4. Inspect
Inspect the connector to make sure that no particles or residue remain. Refer to “Visual
Inspection” on page 3-3.
85052D 3-5
Use, Maintenance, and Care of the Devices

Gaging Connectors

Gaging Connectors

The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices. Do not use the gages for precise pin depth measurements. See
Table 6-2 on page 6-3 for part number information.

Connector Gage Accuracy

The connector gages are only capable of performing coarse measurements. They do not provide the degree of accuracy necessary to precisely measure the pin depth of the kit devices. This is partially due to the repeatability uncertainties that are associated with the measurement. Only the factory—through special gaging processes and electrical testing— can accurately verify the mechanical characteristics of the devices.
With proper technique, the gages are useful in detecting gross pin depth errors on device connectors. To achieve maximum accuracy, random errors must be reduced by taking the average of at least three measurements having different gage orientations on the connector. Even the resultant average can be in error by as much as ± 0.0001 inch due to systematic (biasing) errors usually resulting from worn gages and gage masters. The information in Table 2-2 on page 2-4 assumes new gages and gage masters. Therefore, these systematic errors were not included in the uncertainty analysis. As the gages undergo more use, the systematic errors can become more significant in the accuracy of the measurement.
The measurement uncertainties in Table 2-2 are primarily a function of the assembly materials and design, and the unique interaction each device type has with the gage. Therefore, these uncertainties can vary among the different devices. For example, note the difference between the uncertainties of the opens and shorts.
The observed pin depth limits in Table 2-2 add these uncertainties to the typical factory pin depth values to provide practical limits that can be referenced when using the gages. See “Pin Depth” on page 2-3. Refer to “Kit Contents” on page 1-2 for more information on the design of the calibration devices in the kit.
NOTE
When measuring pin depth, the measured value (resultant average of three or more measurements) contains measurement uncertainty and is not necessarily the true value. Always compare the measured value with the observed pin depth limits (which account for measurement uncertainties) in
Table 2-2 on page 2-4 to evaluate the condition of device connectors.
3-6 85052D
Use, Maintenance, and Care of the Devices
Gaging Connectors

When to Gage Connectors

Gage a connector at the following times:
• Prior to using a device for the first time, record the pin depth measurement so that it can be compared with future readings. (It will serve as a good troubleshooting tool when you suspect damage may have occurred to the device.)
• If either visual inspection or electrical performance suggests that the connector interface may be out of typical range (due to wear or damage, for example).
• If a calibration device is used by someone else or on another system or piece of equipment.
• Initially, after every 100 connections, and after that, as often as experience indicates.
85052D 3-7
Use, Maintenance, and Care of the Devices
Gaging Connectors

Gaging Procedures

Gaging 3.5 mm Connectors
NOTE
Always hold a connector gage by the gage barrel, below the dial indicator. This gives the best stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
1. Select the proper gage for your connector. Refer to Table 6-1 on page 6-2 for gage part numbers.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage or the device,
connect the gage to the gage master by interconnecting the male and female connectors. Connect the nut finger tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque
wrench included in the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a Torque Wrench” on page 3-10 for additional information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set knob until the gage pointer lines up exactly with the zero mark.
d. Remove the gage master.
4. Gage the device connector (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage or the device,
connect the gage to the device by interconnecting the male and female connectors. Connect the nut finger-tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque
wrench included in the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a Torque Wrench” on page 3-10 for additional information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. Read only the black ± signs; not the red ± signs.
For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis.
e. Compare the average reading with the observed pin depth limits in Table 2-2 on page
2-4.
3-8 85052D
Figure 3-2 Gaging 3.5 mm Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85052D 3-9
Use, Maintenance, and Care of the Devices

Connections

Connections

Good connections require a skilled operator. The most common cause of measurement error is bad connections. The following procedures illustrate how to make good connections.

How to Make a Connection

Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table mat. Refer to “Electrostatic Discharge” on page 3-2 for ESD precautions.
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-4.
4. Use a connector gage to verify that all center conductors are within the observed pin depth values in Table 2-2 on page 2-4. Refer to “Gaging Connectors” on page 3-6.
5. Carefully align the connectors. The male connector center pin must slip concentrically into the contact finger of the female connector.
6. Push the connectors straight together and tighten the connector nut finger tight.
CAUTION
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables.
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection. Table 3-1 provides information about the torque wrench recommended for use with the calibration kit. A torque wrench is included in the calibration kit. Refer to Table 6-1 on page 6-2 for replacement part number and ordering information.
Do not turn the device body. Only turn the connector nut. Damage to the center conductor can occur if the device body is twisted.
Table 3-1 Torque Wrench Information
Connector Type Torque Setting Torque Tolerance
3.5 mm 90 N-cm (8 in-lb) ±9.0 N-cm (±0.8 in-lb)
3-10 85052D
Use, Maintenance, and Care of the Devices
Connections
Using a torque wrench guarantees that the connection is not too tight, preventing possible connector damage. It also guarantees that all connections are equally tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible to do this by hand if one of the connectors is fixed (as on a test port). However, it is recommended that you use an open-end wrench to keep the body of the device from turning.
1. Position both wrenches within 90 degrees of each other before applying force. See
Figure 3-3. Wrenches opposing each other (greater than 90 degrees apart) will cause a
lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together.
Figure 3-3 Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See
Figure 3-4.
Figure 3-4 Using the Torque Wrench
3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it; otherwise, you apply an unknown amount of torque.
85052D 3-11
Use, Maintenance, and Care of the Devices

Handling and Storage

4. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-4 on page 3-11. Do not tighten the connection further.
CAUTION
You don’t have to fully break the handle of the torque wrench to reach the specified torque; doing so can cause the handle to kick back and loosen the connection. Any give at all in the handle is sufficient torque.

How to Separate a Connection

To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and connections.
CAUTION
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connecting nut.
3. Complete the separation by hand, turning only the connecting nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.
Do not turn the device body. Only turn the connector nut. Damage to the center conductor can occur if the device body is twisted.

Handling and Storage

• Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use.
• Never store connectors loose in a box, or in a desk or bench drawer. This is the most common cause of connector damage during storage.
• Keep connectors clean.
• Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove.
• Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.
3-12 85052D

4 Performance Verification

85052D 4-1
Performance Verification

Introduction

Introduction
The performance of your calibration kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.

How Agilent Verifies the Devices in Your Kit

Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to the National Institute of Standards and Technology (NIST). The airline and short characteristics are developed from mechanical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation. The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) testing of the calibration devices, Agilent includes the test measurement uncertainty as a guardband to guarantee each device meets the published specification. For recertifications (after sale), no guardband is used and the measured data is compared directly with the specification to determine the pass or fail status. The measurement uncertainty for each device is, however, recorded in the calibration report that accompanies recertified kits.
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Agilent Technologies.
4-2 85052D
Performance Verification

Recertification

Recertification
The following will be provided with a recertified kit:
• a new calibration sticker affixed to the case
• a certificate of calibration
• a calibration report for each device in the kit listing measured values, specifications, and uncertainties
NOTE
Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-5 for a list of offices.
A list of NIST traceable numbers may be purchased upon request to be included in the calibration report.
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a different recertification interval that reflects the usage and wear of the kit.
NOTE
The recertification interval should begin on the date the kit is first used after the recertification date.
Where to Send a Kit for Recertification
Contact Agilent Technologies for information on where to send your kit for recertification. Contact information is listed on page 5-5. Refer to “Returning a Kit or Device to Agilent”
on page 5-4 for details on sending your kit.
85052D 4-3
Performance Verification
Recertification
4-4 85052D

5 Troubleshooting

85052D 5-1
Troubleshooting

Troubleshooting Process

Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.
Figure 5-1 Troubleshooting Flowchart
5-2 85052D
Troubleshooting

Where to Look for More Information

Where to Look for More Information
This manual contains limited information about network analyzer system operation. For detailed information on using a VNA, ENA or PNA series network analyzer, refer to the appropriate user guide or online Help.
• To view the ENA or PNA online Help, press the Help key on the front panel of the network analyzer.
• To view an online VNA user guide, use the following steps:
1. Go to www.agilent.com.
2. Enter your VNA model number (Ex: 8753ES) in the Search box and click Search.
3. Under the heading Manuals & Guides, click on the title/hyperlink for the
document PDF you want to view.
If you need additional information, see “Contacting Agilent” on page 5-5.
85052D 5-3
Troubleshooting

Returning a Kit or Device to Agilent

Returning a Kit or Device to Agilent
If your kit or device requires service, contact Agilent Technologies for information on where to send it. See “Contacting Agilent” on page 5-5 for contact information. Include a service tag (located near the end of this manual) on which you provide the following information:
• your company name and address
• a technical contact person within your company, and the person's complete telephone number
• the model number and serial number of the kit
• the part number and serial number of each device
• the type of service required
•a detailed description of the problem and how the device was being used when the problem occurred (such as calibration or measurement)
5-4 85052D
Troubleshooting

Contacting Agilent

Contacting Agilent
Assistance with test and measurements needs and information on finding a local Agilent office are available on the Web at:
www.agilent.com/find/assist
If you do not have access to the Internet, please contact your Agilent field engineer.
NOTE
In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period.
85052D 5-5
Troubleshooting
Contacting Agilent
5-6 85052D

6 Replaceable Parts

85052D 6-1
Replaceable Parts

Introduction

Introduction
Table 6-1 lists the replacement part numbers for the 85052D calibration kit. Table 6-2 lists
the replacement part numbers for items not included in the calibration kit that are either required or recommended for successful operation of the kit.
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Agilent Technologies. See “Contacting Agilent” on page 5-5 for contact information.
Table 6-1 Replaceable Parts for the 85052D Calibration Kit
Description Qty per
kit
Calibration Devices (3.5 mm)
Broadband load –m–
Broadband load –f–
Offset short –m– 1 85052-60006
Offset short –f– 1 85052-60007
Offset open –m– 1 85052-60008
Offset open –f– 1 85052-60009
–f– to –f– 1 85052-60012
–m– to –f– 1 85052-60013
–m– to –m– 1 85052-60014
For 3.5 mm –f– connectors As
b
b
Adapters (3.5 mm)
Protective End Caps for Connectors
a
1 00902-60003
1 00902-60004
Required
Agilent Part
Number
1401-0202
For 3.5 mm –m– connectors As
Required
Wre nch
5/16 in, 90 N-cm (8 in-lb) torque wrench 1 8710-1765
7 mm open-end wrench 1 8710-1761
Calibration Kit Storage Case
Storage box assembly 1 85052-60029
Box (without foam pads) 1 5180-7862
Foam pad (for lid) 1 5181-5544
1401-0208
6-2 85052D
Table 6-1 Replaceable Parts for the 85052D Calibration Kit
Replaceable Parts
Introduction
Description Qty per
kit
Foam pad (for lower case) 1 85056-80023
Disk holder 1 5180-8491
Miscellaneous Items
Calibration constants disk (8510 series) 1 85052-10012
Specifications and performance verification disk set (8510 series)
User’s and service guided
a. See “Clarifying the Terminology of a Connector Interface” on page 1-7 b. Broadband load has replaced lowband load. c. See the 8510C On-Site Service Manual for instructions on using the disk. d. See “Printing Copies of Documentation from the Web” on page -iii.
c
1 08510-10033
1 85052-90079
.
Agilent Part
Number
Table 6-2 Items Not Included in the Calibration Kit
Description Qty Agilent Part
Number
ESD Protection Devices
Grounding wrist strap 1 9300-1367
5 ft grounding cord for wrist strap 1 9300-0980
2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300-0797
ESD heel strap 1 9300-1308
Connector Cleaning Supplies
Anhydrous isopropyl alcohol (>92% pure)
Foam-tipped cleaning swabs 100 9301-1243
3.5 mm –f– pin-depth gage
3.5 mm –m– pin-depth gage
3.5 mm slotless connector contact repair kit
b
b
a
Miscellaneous Items
c
-- --
1 11752-60105
1 11752-60106
1 85052-60049
a. Agilent can no longer safely ship isopropyl alcohol, so customers should purchase it
locally.
b. Refer to "Clarifying the Terminology of a Connector Interface" on page 1-7.
c. All female connectors on the precision devices in this kit are slotless connectors. Refer to
"Inspect Female Connectors" on page 3-4.
85052D 6-3
Replaceable Parts
Introduction
Figure 6-1 Replaceable Parts for the 85052D Calibration Kit
6-4 85052D
Figure 6-2 Replaceable Parts for the 85052D Calibration Kit
Replaceable Parts
Introduction
85052D 6-5
Replaceable Parts
Introduction
6-6 85052D

A Standard Definitions

85052D A-1
Standard Definitions

Class Assignments and Standard Definitions Values are Available on the Web

Class Assignments and Standard Definitions Values are Available on the Web

Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Agilent’s Calibration Kit Definitions Web page at
http://na.tm.agilent.com/pna/caldefs/stddefs.html.
For a detailed discussion of calibration kits, refer to the Agilent Application Note, “Specifying Calibration Standards and Kits for Agilent Vector Network Analyzers.” This application note covers calibration standard definitions, calibration kit content and its structure requirements for Agilent vector network analyzers. It also provides some examples of how to set up a new calibration kit and how to modify an existing calibration kit definition file. To download a free copy, go to www.agilent.com and enter literature number 5989-4840EN in the Search window.
A-2 85052D

Index

A
accuracy
connector gage
adapters
part numbers, 6-2 specifications
Agilent Technologies
application note contacting
alcohol
isopropyl
as cleaning solvent precautions for use of
altitude, specifications
, 3-6
, 1-3
, 2-5
, A-2
, 5-4, 5-5
, 3-5
, 3-5
, 2-2
B
box, part number, 6-2 broadband loads
part numbers specifications, 2-5
, 1-2
, 6-2
C
cal kit
contents history verifying, 4-2
calibration
bad certificate of, 4-3 constants, See calibration
definitions
definitions, entering report temperature
calibration kit
Agilent Application Note contents
history modifying definition files overview performance
case, part number certificate of calibration certification
of device specifications
characteristics
mechanical
class assignments
downloading from Agilent Web
, 1-2
, 1-5
, 5-2
definitions
, 1-3
disk part number
, 6-3
, 4-3
, 2-2
, 1-2
drawing of
, 6-4, 6-5
, 1-5
, 1-2
how Agilent verifies
, 6-2
, 2-3
, A-2
site
, 1-3
, A-2
, A-2
, 4-2
, 4-3
, 2-6
cleaning connectors cleaning supplies
part number
compressed air
for cleaning
conductive mat, part number
6-3
connections
ESD protection final, 3-10 preliminary separating using torque wrench
connector
cleaning damage, 3-3 defects female gage
accuracy handling use of, 3-6 zeroing
gaging
to determine pin depth, 3-6
when to do gender mating plane surfaces
cleaning slotless terminology, 1-7 threads
cleaning
inspecting, 3-3 visual inspection
, 3-3
wear
affect on electrical
constants, calibration, See
calibration definitions
contacting Agilent
Technologies
contents
calibration kit drawing of incomplete
what to do
, 6-3
, 3-4
, 3-2, 3-10
, 3-10
, 3-12
, 3-4
, 3-3
, 3-4
, 3-6
, 3-6, 3-8
, 3-6, 3-8
, 3-6, 3-8
, 3-7
, 1-7
, 3-5
, 1-6, 3-4, 6-3
, 3-5
performance
, 6-4, 6-5
, 1-4
, 3-4
, 1-3
, 3-2
, 3-10
, 3-3
, 3-3
, 5-4
, 1-2
D
damage
caused by electrostatic
discharge
, 3-3
device inspecting for to connectors, 3-3
data
, 3-2
, 3-3
recertification
defective connectors definitions
calibration
,
entering permanently stored
part numbers, 6-2
deviation from nominal
phase
device
cleaning connecting damage disconnecting handling maintenance, 1-7 performance
verifying
specifications, 2-5
certification of electrical traceability, 4-2, 4-3
storage temperature visual inspection, 3-3
disconnections disk
calibration definitions
part number
specifications and performance
part number
documentation, part number
6-3
downloading class
assignments & std definitions from the Web
A-2
, 4-3
, 1-3
, 1-3
, 2-5
, 3-4
, 3-10
, 3-3
, 3-12
, 3-12
, 4-2
, 2-6
, 2-5
, 3-12
, 2-2
, 3-12
, 6-3
, 6-3
E
electrical characteristics
effects of temperature
electrical specifications electrostatic discharge, See
ESD end cap, part numbers environmental
regulations requirements specifications
equipment required
, 3-2
ESD
protection, 3-2 supplies
part, numbers
, 3-5
, 2-2 , 2-2
, 6-3
, 3-3
, 1-3
,
,
, 2-2
, 2-5
, 6-2
, 1-3
85052D Index-i
Index
F
female connectors, 3-4
inspection of
flowchart, troubleshooting,
5-2
frequency, specifications
, 3-4
, 2-5
G
gage
connector, 1-3
handling
zeroing
gaging
connectors
when to do
procedures, 3-8 to determine pin depth
gender, connector general information, 1-1
, 3-8
, 3-8
, 3-6, 3-8
, 3-7
, 3-6
, 1-7
H
handling, 3-12 heel strap, part number humidity, specifications
, 6-3
, 2-2
I
incoming inspection, 1-4 information, troubleshooting,
5-3
inspection
damage defects female connectors incoming mating plane surfaces visual
isopropyl alcohol
as cleaning solvent precautions for use of
, 3-3
, 3-3
, 3-4
, 1-4
, 3-3
, 3-3
, 3-5
, 3-5
K
kit
contents
overview
, 1-2, 6-4, 6-5
drawing of
, 1-2
, 6-4, 6-5
L
limits
pin depth, 2-4
loads
broadband broadband, part numbers, 6-2
, 1-2
M
maintenance, 3-2
of devices preventive
making connections, 3-10
ESD protection precautions
manual
copyright dates part number print date, 1-i printing title
mat
conductive
part number
mating plane surfaces
cleaning connector inspection of, 3-3
mechanical characteristics
2-3
affect on electrical performance,
verifying
modifying calibration kit
definition files
, 1-7
, 1-7
, 3-2
, 3-10
, 1-i
, 1-i, 6-2
, 1-iii
, 1-i
, 6-3
, 3-5
, 3-5
2-3
, 3-6
, A-2
N
National Institute of
Standards and Technology
(NIST) nitrogen, for cleaning numbers
replaceable parts serial
, 2-6, 4-2
, 3-4
, 6-2, 6-3
, 1-4
O
observed limits
pin depth
offset opens and shorts
part numbers specifications
open-end wrench, 1-2, 3-12
part number
opens
part numbers, 6-2 specifications
ordering, parts
, 2-4
, 1-2
, 6-2
, 2-5
, 6-2
, 1-2
, 2-5
, 6-2, 6-3
P
part numbers, 6-2
of items in kit of items not in kit, 6-3
parts
included in kit not included in kit, 6-2, 6-3 ordering replaceable
performance
and specifications disk
part number
performance verification
pin depth
precision adapters
,
precision slotless connectors,
preventive maintenance procedures
protective end caps, part
protrusion
, 5-2
failure
affect on electrical performance,
2-4
gaging to determine observed limits, 2-4, 3-6 protrusion recession typical values, 2-4
part numbers
1-6
cleaning
, 3-8
gaging zeroing, 3-8
numbers
pin depth
, 6-2
, 1-2, 6-2
, 6-2
, 6-2
, 6-3
, 2-3
, 3-6
, 2-3
, 2-3
, 1-3
, 6-2
, 1-7
, 3-4
, 6-2
, 2-3
R
recertification
how to order interval locations services included
recession
pin depth
regulations
environmental
replaceable parts
drawing of
report, calibration requirements
environmental
return kit or device to Agilent,
1-4, 5-4
return loss specifications
, 4-3
, 4-3
, 4-3
, 4-3
, 2-3
, 2-2, 3-5
, 6-2, 6-3
, 6-4, 6-5
, 4-3
, 2-2
, 2-5
Index-ii 85052D
Index
S
separating connections, 3-12 serial numbers
devices, 1-4 recording
service service tag, 4-3, 5-4 shorts
specifications
standard definitions
standards
static
storage storage case, part number strap
supplies
swabs
, 5-4
, 1-2
part numbers specifications, 2-5
adapters altitude
operating storage
and performance disk
part number
broadband loads certification of, 2-6 deviation from nominal phase
2-5
device, 2-5 electrical environmental frequency, 2-5 humidity
operating storage, 2-2
offset opens and shorts return loss sliding loads, 2-5 temperature torque wrench traceability, 4-2, 4-3
downloading from Agilent Web
, A-2
site
international National Institute of Standards
and Technology (NIST)
4-2
discharge electricity
, 3-12
heel and wrist, part number
cleaning
part number, 6-3
cleaning
, 1-4
, 1-4
, 6-2
, 2-2
, 2-5
, 2-2
, 2-2
, 6-3
, 2-5
, 2-5
, 2-2
, 2-2
, 2-5
, 2-5
, 2-2
, 3-10
, 2-6
, 2-6,
, 3-2
, 3-2
, 6-2
, 6-3
, 1-3
, 3-5
tag
temperature
test data threads
,
torque wrench
traceability
troubleshooting, flowchart
user’s and service guide
verification
visual inspection
wear
wrench
wrist strap, part number
, 4-3, 5-4
service
affect on electrical performance
2-2
calibration, 2-2
cautions about
changes in
device, 2-2
error-corrected
measurement
specifications
operating storage
verification and measurement,
, 2-2
, 2-2
, 2-2
2-2
, 4-3
connector
cleaning inspecting, 3-3
specifications
of device specifications
5-2
part number
performance
temperature
connector
affect on electrical
open-end
proper positioning of
torque
precautions for use of, 3-11 proper use of
, 3-5
, 6-3
, 4-3 , 2-2
, 3-3
performance
, 1-3, 3-11, 3-12
, 1-3, 3-10, 3-11
T
, 2-2
, 2-2
, 2-2
, 2-2
, 1-3
, 3-10
U
V
W
, 3-11
, 3-3
, 3-3
, 4-2, 4-3
, 3-11
,
, 6-3
,
zeroing
connector gage
Z
, 3-8
85052D Index-iii
Index
Index-iv 85052D
Loading...