THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED
"AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT
NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM
EXTENT PERMITTED BY APPLICABLE LAW, AGILENT
DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED
WITH REGARD TO THIS MANUAL AND ANY INFORMATION
CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE
A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN
THE SEPARATE AGREEMENT WILL CONTROL.
Printing Copies of Documentation from the
Web
To print copies of documentation from the Web, download the PDF file
from the Agilent web site:
• Go to http://www.agilent.com.
• Enter the document’s part number (located on the title page) in the
Quick Search box.
• Click GO.
• Click on the hyperlink for the document.
• Click the printer icon located in the tool bar.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime
contract or subcontract, Software is delivered and licensed as
“Commercial computer softwar e” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as defined in FAR 2. 101(a) or as
“Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or contract clause. Use,
duplication or disclosure of Software is subject to Agilent Technologies’
standard commercial license terms, and non-DOD Departments and
Agencies of the U.S. Government will receive no greater than Restricte d
Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S.
Government users will receive no greater than Limited Rights as
defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any technical data.
ii
Contacting Agilent
This information supersedes all prior HP contact information.
Figure 1 85052C Precision Calibration Kit Operating and Service
Package
B
A
v
INTRODUCTION
What's in This
Guide
Chapter 1This chapter introduces the components of the kit, and describes
Chapter 2In this chapter, detailed sequences describe how to set up the networ k
Chapter 3Since many applications require measurement of devices that are not
Chapter 4This chapter, entitled "Other Calibrat i ons," describes use of the
85052C 3.5 mm
Calibration Kit
Video
(Training CD)
This guide describes how to use the 85052C calibration kit with the
8510 vector network analyzer to perform the TRL 2-Port measurement
calibration technique , and how to evaluate the results of the
calibration.
important connection techniques.
analyzer system, and use the kit to perform the 3.5 mm TRL 2-Po rt
calibration procedure .
insertable, this chap ter de scribes the th eory of and how t o use the kit t o
perform the Adapter Removal procedure for measuremen t of
noninsertable devices.
network analyzer Modify Cal Kit function to redefi ne the calibration kit
and how to use the Lowband Reflection option
This guide is supplemented by a video presentation that shows how an
experienced operator uses the calibration kit parts . If you have access to
this CD, view it along with this manual first.
Before
Proceeding...
The parts in this kit are delicate and must be handled with care to
avoid damage that would affect the accuracy of the calibration. Check
all test device connectors carefully before connecting to the TRL
adapters.
The 85052C 3.5 mm Precision Calibration Kit is designed to be used in
TRL 2-Port measurement calib ration as implemented in vect or network
analyzers. TRL stands for Thru-Reflect-Line, naming the main
standards used in the accuracy enhancement pro cedure.
TRL represents a family of calibration techniques that measure various
combinations of transmission lines and reflection standards to
determine the 2-P ort 12-term er ror coeff ici ents. The specific calibration
technique described here uses measurements of the zero-length thru
connection, identical reflection standards at each port, and one or more
transmission lines of appropriate impedanc e and length for the
frequency range. Both the TRL 2-PORT and the Full 2-Port calibration
types use the same accuracy enhancement mathematics to correct the
measured data. TRL, as implemented with this calibration kit, brings
convenience, accuracy, and repeatability to the error corr ection process
because the typical TRL calibration requires fewer parts that are
simpler to connect and disconnect.
The 3.5 mm geometry connector is the most frequently used connector
when frequency coverage up to 26.5 GHz, reasonable commonality, and
durability is desired. Well constructed connectors and transmission
lines can work up to 34 GHz. Metrology grade versions of this connector
are used for high performance test ports and for cali bration standards.
Whether the device uses SMA, S MA-compatible, or 3.5 mm, the 3 .5 mm
connectors used on the test ports, adapters, cables, and calibration
standards provide the most accurate and repeatable solution.
All male connectors in this kit are precision with carefully controlled
dimensions. All female connecto rs are of the precision slot less (PSC-3.5)
type. These connector are designed for long repeatability when used
with appropriate technique. This means careful inspection of the kit
parts, familiarity with their use, and careful inspection of the Device
Under Test (DUT) connectors. Refer to the "Specifications" chapter in
the Agilent 85052C 3.5 mm Precision Calibration Kit User’ s and Service Manual, and “Precision Connectors” on page 9 for specific information
about precision slotless connectors.
2Chapter 1
Figure 1-1 85052C Precision Calibration Kit
Introduction to the Calibration Kit
Chapter 13
Introduction to the Calibration Kit
Calibration Kit Components
Calibration Kit Components
The 85052C 3.5 mm Precision Calibration Kit contains components
useful for several different calibration techniques. The parts used for
the TRL 2-Port calibration are:
• longer precision air line for 2 to 7 GHz
• shorter precision air line for 7 to 32 GHz
• male and female short circuits
• male and female fixed loads
Other parts included are:
• three precision TRL adapters
• two adapter anti-rotation clamps
• calibration kit definition disk
• male and female open circuits
• tools
Note the reference card in the box and the material list on the bottom of
the box.
4Chapter 1
Introduction to the Calibration Kit
Air Lines
Air Lines
These are precision air lines. Each one is always stored in its
installation tool to help prevent damage and preserve cleanliness.
The shorter air line is used for measurement calibrati on ov er the 7 to
32 GHz frequency range. The longer air line is used for measurement
calibration over the 2 to 7 GHz range. Each air line is fully specified at
all frequencies within its stated range.
Figure 1-2 Line Standards — 7 to 32 GHz and 2 to 7 GHz Air Lines Ins talled
in Tools with Storage Bottles
Figure 1-3 Polar Display Showing Phase Response of Longer Air Line Over
2 to 7 GHz Frequency Range
Chapter 15
Introduction to the Calibration Kit
Air Lines
Figure 1-4 Polar Display Showing Phase Response of Shorter Air Line
Over 7 to 26.5 GHz F requency Range
Each air line consists of a separate outer conductor and center
conductor and includes its own tool used for installing and removing the
parts from the test port. Both are ins ertable. Eac h has a male end and a
female end. The male end of the air line fits into the installation tool.
Figure 1-5 Air Lines Removed from Installation Tools
CAUTIONThe parts may be removed for inspection, if necessary. The tool is
always used to install th e parts for calibration. Use extreme care in
handling these parts. Use static dissipative finger sheaths, or "finger
cots," to prevent cont amination. Liquid or solid residu e on the connector
parts will degrade performance. In particular, the inner conductors can
be easily deformed by squeezing. Do not use metal tweezers or other
devices to pick up or hold the parts.
6Chapter 1
Figure 1-6 Installing the Air Lines in Installation Tools
Introduction to the Calibration Kit
Air Lines
0 to 2 GHz Loads
This implementation of the TRL 2-P o rt calibration allows measurement
of the loads to determine the error coef ficient s up to 2 GHz. Technically,
the measurement is treated as an infinite-length transmission line
whose input reflection is the refer ence impedance, Z
NOTEThe male and female broadband loads are manufactured to tolerances
that assure equivalent response at least below 2 GHz. If one is damaged
or otherwise becomes defective , calibration errors will result.
Figure 1-7 0 to 2 GHz Loads — Male and Female Fixed Broadband Loads
.
0
Reflect Standards
In this TRL 2-Port calibration, the reflect standards are identical short
circuits with one connected to Port 1 and one to Port 2.
The male short circuit and the female shor t circuit are manuf actured to
tolerances that assure equivalent responses. If one is damaged or
Chapter 17
Introduction to the Calibration Kit
Air Lines
otherwise becomes defective, calibration errors will result. Errors will
also result if the open circuits are used in the procedure instead of the
short circuits (unless the calibr ation kit standard assignment definition
is changed).
Figure 1-8 Reflection Standards — Male and Female Short Circ uits
Isolation Standards
The fixed loads used for the isolation part of the calibration are the
same broadband loads as used for the 0-2 loads measurement. Exact
equivalent responses are not required for this step.
Figure 1-9 Isolation Standards — Male and Female Brodband Loads
Precision Adapters
Three precision adapters are included in this kit. Two are always
recommended for the 3.5 mm TRL 2-Port measurement calibration. All
three can be used in the adapter removal calibration for measurement
of noninsertable devices . Complete performance verification assumes
use of the precision adapters in this kit as the test ports.
Figure 1-10 Precision TRL Test Port Adapters
8Chapter 1
Introduction to the Calibration Kit
Air Lines
Recession Limits The female receptacle recession and the male pin rec ession, along with
other mechanical specifications of Port 1 and Port 2, must be within
limits to meet complete measurement specificati ons. The precision
adapters supplied in this kit are manufactured to metrology stand ards
in order to obtain the best measurement accurac y specifications. If
other connectors than these precision adapters are used as the test
ports, they must be fully inspected. Using other than the highest
quality connectors as the test ports will reduce accuracy and
repeatability as well as possibly damaging the calibration standards.
Precision
Connectors
For general use, any 3.5 mm precision slotless female connector can be
used for the female test port. The 3.5 mm prec ision male connecto r used
on the male end of these adapters is alw a ys re quired fo r meas urements
between 7-32 GHz in order to connect the 7-32 GHz air line in this kit.
Any 3.5 mm precision male connector can be used at frequencies less
than 7 GHz.
Installation
Feature
The precision adapter male connector has a special feature to allow
installation of the 7-32 air line.
• Grasp the knurled sleeve and move the nut to its extended position
by pulling on it.
This is the position used for measurement of the 7-32 GH air line.
Figure 1-11 Extended Position
• Now move the sleeve forward and push the nut back into its
standard position.
This position is used to install the short air line, for connecting the
other calibration standards, and for connecting the device under test.
Figure 1-12 Standard Position
Chapter 19
Introduction to the Calibration Kit
Air Lines
Tools
A torque wrench and spanner wrench are included in the kit to loosen
and tighten the connectors. The small torque wrench is used for the
3.5-mm nuts. The large torque wrench is used for the 20-mm nuts on
the precision adapters. It is included with the test set. The spanner
wrench is used to hold the precision adapter while using the
appropriate torque wrench to tighten the connection.
Also included is the hex wrench for tightening and loosening the
adapter anti-rotation clamp securing screws.
CAUTIONWhen making connections, turn the nut on the device, never turn the
device itself. Turning the device will cause excessive wear of both
connector mating surfaces and cause debris to collect in the female
receptacle.
Figure 1-13 Torque Wrench
Figure 1-14 Spanner Wrench
Figure 1-15 Torque Wrench (not included)
Figure 1-16 Hex Wrench for Adapter Anti-Rotation Clamp
When making a connection:
1. Carefully engage the male and female contacts, and then ti ghten the
nut by hand.
2. If necessary, use the spanner to hold one device stationar y against
the direction of torque.
3. Hold the torque wrench with your thumb and index finger near the
end of the handle.
10Chapter 1
4. Tighten the connection until the ball in the hand le cre sts on the cam
(as the handle begins to "break," releasing tension).
It is not recommended or necessary to fully "break" the handle of the
torque wrench to reac h the f ully- specified t orque . As soon as the ha ndle
moves out of alignment, the connection is properly torqued.
Reverse the above procedure to disconnect the connectors.
Figure 1-17 Using the Spanner and Torque Wrench
Introduction to the Calibration Kit
Air Lines
Figure 1-18 Correct Torque Value
Chapter 111
Introduction to the Calibration Kit
Air Lines
12Chapter 1
2Performing a TRL 2-Port
Measurement Calibration
13
Performing a TRL 2-Port Measurement Calibration
Electrostatic Discharge
This example describes use of the 3.5 mm precision calibration kit to
measure an insertable device. When the test device is insertable, the
test ports can be connected together to establish the Thru connection
during calibration.
The procedure used for performing any measureme nt calibration is as
follows:
1. Clean and inspect all connectors.
2. Initiate measurement calibration procedure and measure standards.
3. Verify the calibration.
4. Connect the device under test.
The rest of this section explains the above steps.
Electrostatic Discharge
When you make connections to the network analyzer, either directly to
the test set port or indirec tl y t h ro u g h a ca b l e that is conne cted to the
test set port, protect the instrument against electrostatic discharge
(ESD). Always wear a grounding wrist strap connected to a conductive
bench mat when working near sensitive equipment.
CAUTIONThe human body almost always has some static charge. You are not
usually aware of this charge because the human threshold for the
perception of a static discharge shock is about 30 kV. ESD as low as 60
volts can damage sensitive microcircuits.
14Chapter 2
Performing a TRL 2-Port Measurement Calibration
Clean and Inspect
Clean and Inspect
The accuracy of error-corrected measurements using the network
analyzer depends upon the quality of the measurement calibrat ion. In
turn, the quality of the calibration depend s on the condition and
cleanliness of the test ports and calibration standards, and the
operator's ski in making connections. Repeatability and accuracy are
improved by removing contaminants from mating su rface during
cleaning.
Carefully inspect the following for cleanliness and damage. Look for
deformation of the male pin or the female receptacle, and for metallic
and organic residue on the mating surfaces .
• test set test por ts
• test port cables
• precision adapters
• calibration standards
• DUT Connectors
A low power magnifier, less than 6x, is adequate to inspect the parts . Do
not be concerned with minor defects such as scratches visible only with
the magnifier; these will probably not affect performance. If inspection
shows moisture, oil, or other res i due, clean the parts.
When examining the female connector, look at the inner contact to see
that the gripping fingers are not bent.
The greatest variable in the accuracy and repeatability for these
demanding measurements is the experience and attention to detail of
the person actually making the connections . Before going on, examine
all parts carefully.
Figure 2-1 Detail of Precision Slotless Female Connector
Chapter 215
Performing a TRL 2-Port Measurement Calibration
Cleaning
Cleaning
The parts must be clean in order to achieve best results. Use isopropyl
alcohol, dry air or nitrogen, and lint-free swabs to remove
contaminants.
In general, do not force cleaning materials into connector parts and do
not immerse the part in cleaning fluids. Just moisten the swab and
wipe the surface lightly. To dry the part, use a dry swab or light
pressure from a dry air or nitrogen source.
Do not leave lint or particles of the sw ab in the connector.
If there is doubt about the condition of the connect or, use the connector
gages to check the connector dimens ions. The same care should be used
for all connectors, es pecially the device under test.
16Chapter 2
Performing a TRL 2-Port Measurement Calibration
Using Connector Gages
Using Connector Gages
Gaging is optional for the precision connectors in this calibration kit if
the male and female center conductors are concentric and do not show
physical damage.
Before
Connecting
Figure 2-2 Connector Gage
However, due to the large range of dimensions found on most DUT
connectors, the DUT connect ors must always by inspected and gaged
before connecting them to the precision adapters. This is not usually a
problem for the DUT female connectors, but may be a problem for DUT
male connectors, especially the economical SMA and SMA-compatible
types.
Measurement
Errors
Connector
Damage
The gage measures the pin depth (recession) of the connectors. If the
pin depth to too far ne gative , errors will result from exces sive air gap in
the connect i on.
If the pin depth is positive (protrusion), both mating connectors will
almost certainly be damaged.
Also, if the male pin diameter is greater than the maximum 3.5 mm
specification, it will cause permanent damage to the female receptacle.
If you try to measure the pin depth and the gage will not fit over the
pin, it is an immediate indication that the pin is too large and must not
Chapter 217
Performing a TRL 2-Port Measurement Calibration
Using Connector Gages
be connected to the precision adapters. The male pin of the DUT
connector may fit into the gage and give a correct measurement, bu t
still be slightly too large for the precision adapter female receptacle.
Repair and
Replacement
In general, bent or deformed parts cannot be repaired in the field. It is
necessary t o re t u rn th e part to a service center for rep a i r o r
replacement. The air line center and outer conductors must be repl aced
as a set. Refer to the replaceable parts chapter in the service manual
supplied with th e ca li b ra tion kit for replacement part nu mb e rs and
ordering information.
Figure 2-3 Cross Section of Female Connector Showing Recession Limits
Figure 2-4 Cross Section of Male Connector Showing Recession Limits
18Chapter 2
Configure Port 1 and Port 2
Several configurations are possible, depending upon the frequency
range, available cables, operator convenience, and DUT requirements.
Following is an example hookup using the two-cable set and two
examples using a single cable. The appropriate precision adapters are
installed on the ends of the cables and become Port 1 and Port 2.
Figure 2-5 Dual Cable Set
Performing a TRL 2-Port Measurement Calibration
Configure Port 1 and Port 2
Figure 2-6 Single Cable, Port 1 Female
Figure 2-7 Single Cable, Port 1 Male
Chapter 219
Performing a TRL 2-Port Measurement Calibration
Install Test Port Return Cables
Install Test Port Return Cables
When all parts are r eady for us e , co nnect the t est p ort extensi on c ables.
To connect a cable to the test set:
• Loosen the test port anti-rotation clamp securing screw and slide the
clamp over the cable far enough to provide access to the cable
connector.
• Connect the cable to the test port and tighten using the torque
wrench.
• Slide the clamp toward the port, aligning it so that the flats on the
clamp mate with flats around the test port. Tighten the clamp
securing screw. This clamp keeps the cable from becoming loose from
the test set con n ector.
20Chapter 2
Performing a TRL 2-Port Measurement Calibration
Install Precision Adapters
To install the precision adapters and adapter anti-rotation clamp:
• Connect the adapter to the cable, tighten finger tight, and then use
the spanner and the torque wrench to achieve the final torque.
• Loosen the adapter anti-rotation clamp securing screws and slide
the clamp over the adapter. Align the clamp so that it can grip both
the cable connector body and the adapter body. Tighten the clamp
securing screws. The screw tightening order is not important.
The adapter anti-rotation c lamp assure s that t he TRL adapt er does not
become loose during calibration and measurement.
• Connect the second TRL adap ter and its anti-r otation cla mp to ser ve
as Port 2.
Figure 2-8 Two Cable Setup with All C omponents
Install Precision Adapters
Chapter 221
Performing a TRL 2-Port Measurement Calibration
Air Line Frequency Coverage
Air Line Frequency Coverage
This illustration shows the frequency range covered by the air line and
fixed loads for the Lines part of the TRL 2- P ort c alibration procedure . If
measuring over the entire frequency range , measure the 7-32 Line first,
then the 2-7 Line, then the 0-2 Loads. If measuring over a reduced
frequency range, only those standards for that frequency range need to
be measured. For example, if you are testing over the 3-22 GHz range,
only the 7-32 Line and the 2-7 Line must be measured.
The standard label for the 2- 7 Line would indicate that its data is valid
only from 2 to 7 GHz. However, you may choose to use data from the 2-7
Line down to the 1 GHz by simply not measuring the 0-2 Loads. If the
0-2 Loads are measured after the 2-7 Line, the 0-2 Loads dat a will
replace data from the 2-7 Line.
Figure 2-9 Frequency Coverage for Line Part of Measurement Calibration
22Chapter 2
Performing a TRL 2-Port Measurement Calibration
Load Calibration Kit Definition
Load Calibration Kit Definition
To begin measurement calibration, first load the calibration kit
definition from disk.
Press:
DISC
LOAD
CAL KIT 1-2
CAL KIT 1
FILE 1
When the calibration kit definition is loaded, the CAL 1 softkey label
will read 3.5 mm C. 1.
Chapter 223
Performing a TRL 2-Port Measurement Calibration
Measure Standards
Measure Standards
Now perform the TRL 2-Port measur ement cal ibration pr ocedure using
the 85052C 3.5 mm. Precision Calibration Kit.
NOTEThe standards are measured in the order: Thru, Reflect, Line, Isolation
in order to connect the load standards just once.
Press:
CAL
CAL 1
TRL 2-PORT
• Connect the TRL male adapter (Port 1) to the TRL female adapter
(Port 2).
•Press
S
THRU. The 8510 makes six measurements; they are S
, S22, and two specially redefined user parameters.
12
Figure 2-10 Thru Standards
• Connect the Short circuit to Port 1.
Figure 2-11 S
•Press
11
S
REFLECT SHORT. S11 is measured.
11
Reflection Short Standard
• Connect the Short circuit to Port 2.
S
•Press
REFLECT SHORT. S22 is measured.
22
, S21,
11
Figure 2-12 S
•Press
22
24Chapter 2
LINES.
Reflection Short Standard
When Stop Frequency Is Greater than 7 GHz
If the stop frequency is greater than 7 GHz, install the 7-32 GHz air
line as follows:
• Move the Port 1 nut to its standard position and careful insert the
air line into the male connector using the installation tool.
Figure 2-13 Insert Air Line
• Squeeze the installation tool (closing the slot) to release the center
conductor. Then pull the tool away from the connector, leaving the
outer and the center conductors place.
Performing a TRL 2-Port Measurement Calibration
Measure Standards
Figure 2-14 Squeeze Tool to Release, T hen Pull
• Examine the outer and center conductors to see that they are in
place and concentric. If the center conductor is no centered, use the
installation tool to align it.
Figure 2-15 Check Concentricity
• Move the adapter nut to its extended position.
Figure 2-16 Extended Position with Air Lines Installed
Chapter 225
Performing a TRL 2-Port Measurement Calibration
Measure Standards
• Carefully align and engage the Po rt 2 female contact with the Port 1
male contact.
• Tighten finger tight, then use the spanner and torque wrenches to
achieve the final torque.
Figure 2-17 Connect Port 2
•Press
7-32 LINE. The 8510 makes six measurements. All frequencies
of the current range are measured, b ut only data from 7 to 32 GHz is
used to compute error coefficients.
• To remove the 7-32 GHz air line: Loosen the Port 1 nut, and then
carefully move the Port 2 adapter away.
Figure 2-18 Disconnect Port 2
• Move the adapter nut to its standard position.
Figure 2-19 Standard Position with Air Line Installed
26Chapter 2
• Examine the outer and center conductors to see that they are in
place.
• Carefully press the insert ion tool onto the connector.
Figure 2-20 Connect the Insertion Tool
• Hold the tool close to the connector without squeezing the tool.
Carefully pull the assembly away from the connector.
Figure 2-21 Remove the 7-32 Line
Performing a TRL 2-Port Measurement Calibration
Measure Standards
• Check to see that both the inner and the outer conductor have been
removed and are secure in the installation tool.
If any part of the air line remains attached to the male port, first
press the tool bac k onto t he connec tor. Then, try to withdraw the tool
and the air line parts again without closing the slot.
If the air line center condu ctor remains with the fem ale port, us e the
installation tool to remove it as follows:
• Squeeze the tool and push it to engage the center conduct or.
• Hold the tool close to the connector, squeeze the tool without closing
the slot, and pull the assembly aw ay.
• Use your fingers to remove the center conduct or from the tool.
Figure 2-22 Remove Center Conductor from Port 2 (if necessary)
• Carefully insert the center conductor back into the installation tool,
male end first.
Chapter 227
Performing a TRL 2-Port Measurement Calibration
Measure Standards
When Start Frequency Is Less Than 7 GHz
If the start frequency is less than 7 GHz, install the 2-7 GHz air line as
follows:
• Move the Port 1 nut to its standard position. Carefully insert the air
line outer conductor and center pin into t he male connector using the
installation tool.
• Hold the air line outer conductor and tighten the Port 1 nut finger
tight. Use the torque wrench to achieve the final torque.
Figure 2-23 Connect the 2-7 Line
• Press the to o l l ightly to enga ge the center co n ductor.
Figure 2-24 Engage the Center Conductor
• Squeeze the tool to close the slot, then pull the tool away from the
connector, leaving the center conductor.
Figure 2-25 Remove the Tool
• Examine the outer and center co nduc tors t o see t hat t he are in plac e
and concentric. If the center conductor is no centered, use the tip of
the installation tool to align it.
Figure 2-26 Check the Concentricity
28Chapter 2
• Move the air line nut back to expose the center conductor. Carefully
engage the Port 2 female contact with the Port 1 male contact. Be
patient when making this connection because it is easy to move the
center conductor off center.
Figure 2-27 Move the Nut Back
• Move the nut forward to engag e the threads, tighten finger tight,
then use the spanner and the torque wrench to achieve the final
torque.
Figure 2-28 Connect Port 2
Performing a TRL 2-Port Measurement Calibration
Measure Standards
•Press
2-7 LINE. The 8510 makes six measurements. All frequencies of
the current range are measured, but data from 1 GHz to 7 GHz is
used to compute error coefficients.
• T o remove the 2-7 GHz air lin e: Loosen the air line nut and carefully
move Port 2 away from Port 1.
Figure 2-29 Disconnect Port 2
• Insert the tool and lightly press it to engage the center conductor.
Figure 2-30 Engage the Tool
Chapter 229
Performing a TRL 2-Port Measurement Calibration
Measure Standards
• Loosen the adapter nut and then disconnec t the outer conductor
from the adapter. The center conductor will be removed with the
outer conductor.
Figure 2-31 Remove 2-7 Line
NOTEIt is important to be very careful to move Port 2 away in a straight
motion because the center conductor may stay Port 2. If the center
conductor remains with the female port, use the installation tool to
remove it. Carefully ins er t the center co n ductor back in t o th e insertion
tool, male end first.
Figure 2-32 Remove Center Conductor from Port 2 (if necessary)
When Start Frequency Is Less than 2 GHz
If the start frequency is less than 2 GHz, measure the 0-2 Loads as
follows:
• Connect a fixed load to P ort 1 and a fixed load to Port 2.
•Press
are measured, but only data up to 2 GHz are used. These
measurements replace error coeffi cients obtained from measuremen t
of the 2-7 GHz air line below 2 GHz. Leave the loads connected for
the Isolation step.
0-2 LOADS. The 8510 makes six measurements. All frequencies
30Chapter 2
Figure 2-33 Connect the 0-2 Loads
Performing a TRL 2-Port Measurement Calibration
Measure Standards
•Press
•Press
LINES DONE.
ISOLATION.
If maximum transmission dynamic range is not required, skip the
isolation part of the calibration by pressing
then
ISOLATION DONE.
Or, to obtain maximum transmission dynamic range, perform the
isolation part of the calibration as follows:
• Connect a fixed load to P ort 1 and a fixed load to Port 2.
• Increase the
at least 128. Then press
•Press
•Press
REV. ISOL'N ISOL'N STD. S
ISOLA TION DONE. You may wish to reduce the ave raging fact or
AVERAGING FACTOR for the isolation measurement to
FWD. ISOL'N ISOL'N STD. S
for measurement of the device under test.
•Press
SAVE TRL 2-PORT.
Figure 2-34 Reverse Isolation Standards
is measured.
12
OMIT ISOLATION and
is measured.
21
When the computation of error coefficients is complete, press
CAL SET 1 (or any other cal set). Error coefficients derived from
measurement of the TRL standards are computed and saved, and
then 2-Port correction is turned on.
CAUTIONCarefully inspect the connectors on the device under tes t and measure
their center conduct or pin dept h before connect ing the devic e under test
to the system. Device under-test connectors with incorrect pin size, pin
depth, or pin alignment can damage the test port adapters.
Chapter 231
Performing a TRL 2-Port Measurement Calibration
Check the Calibration
Check the Calibration
A good first check of the calibration is to measure the transmiss ion and
reflection characteri st ics of the longe r air line. Connec t the air line
between Port 1 and Port 2 and view the S
the S
linear phase. The S
magnitude. The air line should exhibit low insertion loss and
11
should be constant over the entire frequency
11
range.
magnitude and phase and
21
Figure 2-35 2-7 Line S
Figure 2-36 2-7 Line S
Magnitude
11
Magnitude and Phase
21
For a second check, connect the 2-7 Line to Port 1, then connect the
Short circuit to the end of the air line, and view the reflection response.
32Chapter 2
Select:
S
11
DOMAIN
TIME BAND PASS
Figure 2-37 Line with Short Circuit
Performing a TRL 2-Port Measurement Calibration
Check the Calibration
View the S
response. Use the marker to measure the responses at 0
11
seconds (a rough measure of effective directivity) and the mismatch at
twice the length of the air line (a rough measure of effective sourc e
match). The trace should appear as shown below. If not, the calibration
has not been performed correctly or the calibration devices are
defective.
Figure 2-38 Plot of Time Band with Shorted Air Line
Chapter 233
Performing a TRL 2-Port Measurement Calibration
Check the Calibration
34Chapter 2
3Noninsertable Devices
35
Noninsertable Devices
General Theory
General Theory
The following sequence describes the adapter removal method of
calibration for measurement of a noninsertable device. This procedure
is the most complete and effective calibration procedure for
measurement of noninsertable devices. Although this technique does
require two separate 2-Port cali brations, it remai n s th e only traceable
method for minimizing the uncertainty in this m easurement. The three
precision adapters in this calibration ki t provide the parts required to
make the process relatively easy.
Detailed information about the adapter removal calibration is
contained in Product Note 8510-13, available from your local Sales
Office.
The previous calibration sequenc e assumed that the device und er test is
insertable; it has a male connector on one port and a female connector
on the other port. Thus, the measurement syst em can be calibrated and
then the test device inserted without changing the s ystem test port
connectors.
However, the majority of devices used in microwave systems are
noninsertable. Of inter est here are devices having either both male or
both female 3.5 mm connectors on Po rt 1 and Port 2.
36Chapter 3
Noninsertable Devices
Noninsertable Device Configurations
Noninsertable Device Configurations
Female to
Female
Figure 3-1 Female-Female Device Under Test
Male to MaleIf the device you are measuring has two male connectors, your setup
Figure 3-2 Male-Male Device Under Test
If the device you are measuring has two female connectors, your setup
would look like the figure below. The female-to-female precision adapter
is substituted for the device under test to accomplish this calibration.
would look like the figure below. The male-to-male precision adapter is
substituted for the device under test to accomplish this calibration.
Chapter 337
Noninsertable Devices
Adapter Removal Calibration Procedure
Adapter Removal Calibration Procedure
Create Cal Set
for Port 1
First create the Port 1 calibration set by performing the TR 2-Port
calibration between P ort 1 and th e adapter. Keep the adapter connected
to Port 2 during the entire procedure. Save the calibration in Cal Set 1.
Figure 3-3 Port 1 Cal Set
Create Cal Set
for Port 2
Next create the Port 2 calibr a tion set by performing the TRL 2-Port
calibration between P ort 2 and th e adapter. Keep the adapter connected
to Port 1 during the entire procedure. Save the calibration in Cal Set 2.
Figure 3-4 Port 2 Cal Set
Adapter
Removal
Mathematics
The Adapter Removal function mathematically combines the Port 1 cal
set with the Port 2 cal set to produce a third cal set having the effects of
the adapter removed. The resulting cal set is as if Port 1 and Port 2
could be connected together.
When the two calibrations have been saved, proceed with the adapter
removal sequence as follows:
Press:
CAL
MORE
MODIFY CAL SET
ADAPTER REMOVAL
•Press CAL SET for PORT 1, then CAL SET 1.
•Press
•Press
CAL SET for PORT 2, then CAL SET 2.
ADAPTER 3.5 mm C.1. This specifies the calibration kit that
includes the length specificatio n for the precision adapter.
•Press
MODIFY & SAVE, then CAL SET 3.
38Chapter 3
Adapter Removal Calibration Procedure
The new calibration set is computed and stored. 2-Port correction is
turned on.
Now remove the third adapter and connect the device under test.
The display should now show an accurate S -parameter measu rement of
the device.
Figure 3-5 Remove the Adapter and Connect the DUT
Noninsertable Devices
Chapter 339
Noninsertable Devices
Adapter Removal Calibration Procedure
40Chapter 3
4Other Calibrations
41
Other Calibrations
Changing the Air Line Frequency Range
Changing the Air Line Frequency Range
If your application requires a frequency range slightly beyond the
standard frequency range of one of the precision air lines, you may
extend the frequency range of one of the air lines in ord er to accomplish
the calibration by connecting only one of the air lines.
The error coefficients ar e determined wit h best accuracy when the Thru
and the Line phase response is separated by ± 90 degrees at
corresponding frequencies. When the transmission phase response of a
line standard is the same or gets ve ry cl ose to zer o o r 180 d egree s of the
Thru standard at the same frequency, the result becomes less certain.
The correct result cannot be determined when the phase response is
exactly equal or 180 degrees apart. This is why a particular length air
line covers a specific frequency range.
To evaluate the phase response, first do a simple transmission
frequency response calibration using the Thru standard (Port 1
connected directly to Port 2), then measure the S
The standard air line frequency specifications provide for greater than
20 degrees separation.
phase of the air line.
21
Experiment with this by using the Modify Cal Kit function to change
the Minimum Frequency or Maximum Frequency specification for the
air line, and then performing the TRL 2-Port calibration procedure.
With correction on , trace noise on the error-corrected trace ma y
increase at the points where the phase of the air line used for
calibration approaches the phase of the thru, often showing large
discontinuities at the point where the phase of both standards are
identical or separated by 180 degrees.
If this error is acceptable in your measurement, change the standard
label for the air line and save the redefined cal kit on disk for later use.
If the error is not acceptable, reload the standard cal kit definition.
42Chapter 4
Other Calibrations
Lowband Reflection Instead of 0-2 Loads or 2-7 Line
Lowband Reflection Instead of 0-2 Loads or 2-7
Line
The Lowband Reflection part of the TRL 2-P or t proced ure is optiona l. If
used, error coefficients obt a ined by this procedure replace
measurements of the 0-2 loads.
For this calibration kit, the lowband reflection part of the cali bration is
typically not used because measurement of the 0-2 Loads provides
better accuracy (given that the loads are in good condition and
repeatable).
If you do not select LOWBAND REFLECTION, the error coefficients
obtained from measurement of the TRL standards are used over the
complete current frequency range. If you do select LOWBAND
REFLECTION, err or coefficients obtained from these additional
calibration steps will be used up to 2 GHz.
In actual measurement applications , the Lowband Reflection option
will not produce greater accuracy than using the 0-2 Loads. If you are
uncertain, it is relatively simple to compare measurement results using
the different techniques.
Chapter 443
Other Calibrations
Lowband Reflection Procedure
Lowband Reflection Procedure
Selecting the Lowband Reflection part of the procedure guides you
through these additional steps to measur e an open circuit, short circui t,
and fixed load at Port 1 and an open circuit, short circuit, and fixed load
at Port 2. So, if start frequency is less than 2 GHz and you wish to use
the open/short/load technique for calibration below 2 GHz, then after
measuring the Thru, Reflect, Isolation and Line standards and before
pressing
sequence.
Measure Thru, Reflect, Isolation, and Line standards. If the start
frequency is less than 2 GHz, press
SAVE TRL 2-PORT, perform the following lowband reflection
LOWBAND REFLECTION.
•Press
• Connect the Short circuit to Port 1 and then press (S
• Connect the Open circuit to Port 1 and then press
• Connect the Fixed Load to Port 1 and then press
• Connect the Short circuit to Port 2 and then press
• Connect the Open circuit to Port 2 and then press
• Connect the fixed load to Port 2 and then press
•Press
•Press
When
LOWBAND REFLECTION
): SHORT.
11
(S11): OPEN.
(S11): LOAD.
(S22): SHORT.
(S22): OPEN.
(S22): LOAD.
REFLECT'N DONE.
SAVE TRL 2-PORT, and then CAL SET 1, (o r any other cal set).
SA VE is pressed, error coefficients from measurement of the
Open, Short, and Load are used for frequencies below GHz, and
coefficients from measurement of the Thru, Reflect and Line are used
for frequencies above 2 GHz.
44Chapter 4
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.