No p art o f this manu al may be re produce d in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G3430-90051
Edition
First edition, January 2013
Printed in USA and China
Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA
The material contained in this
document is provided “as is,” and
is subject to being changed,
without notice, in future editions.
Further, to the maximum extent
permitted by applicable law,
Agilent disclaims all warranties,
either express or implied, with
regard to this manual and any
information contained herein,
including but not limited to the
implied warranties of
merchantability and fitness for a
particular purpose. Agilent shall
not be liable for errors or for
incidental or consequential
damages in connection with the
furnishing, use, or performance of
this document or of any
information contained herein.
Should Agilent and the user have
a separate written agreement
with warranty terms covering the
material in this document that
conflict with these terms, the
warranty terms in the separate
agreement shall control.
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in damage to the product or
loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are
fully understood and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a
WARNING notice until the
indicated conditions are fully
understood and met.
27890 Series GC Safety Manual
Page 3
Agilent 7890 Series Gas Chromatograph
Safety Manual
1
Introduction
Important Safety Warnings 4
Hydrogen Safety 7
Micro-Electron Capture Detector (µECD) 14
Fuses and Batteries 16
Safety and Regulatory Certifications 17
Intended Use 21
Cleaning 21
Recycling the Product 21
Agilent Technologies
3
Page 4
Introduction
WARNING
WARNING
WARNING
Important Safety Warnings
Before moving on, there are several important safety notices that you
should always keep in mind when using the Agilent 7890 Series GC.
When handling/using chemicals for preparation or use within the GC, all applicable
local and national laboratory safety practices must be followed. This would include,
but is not limited to, correct use of Personal Protective Equipment (PPE), correct use
of storage vials, and correct handling of chemicals, as defined in the laboratory’s
internal safety analysis and standard operating procedures. Failure to adhere to
laboratory safety practices could lead to injury or death.
Many internal parts of the GC carry dangerous voltages
If the GC is connected to a power source, even if the power switch is off,
potentially dangerous voltages exist on:
• The wiring between the GC power cord and the AC power supply, the
AC power supply itself, and the wiring from the AC power supply to
the power switch.
47890 Series GC Safety Manual
With the power switch on, potentially dangerous voltages also exist on:
• All electronics boards in the instrument.
• The internal wires and cables connected to these boards.
• The wires for any heater (oven, detector, inlet, or valve box).
All these parts are shielded by covers. With the covers in place, it should be difficult
to accidentally make contact with dangerous voltages. Unless specifically
instructed to, never remove a cover unless the detector, inlet, or oven are turned off.
If the power cord insulation is frayed or worn, the cord must be replaced. Contact
your Agilent service representative.
Page 5
Do not use an Uninterruptable Power Supply (UPS) with a GC
WARNING
If the area where the GC is located suddenly looses power, an unsafe
condition can result if the GC remains powered on. Do not use the GC
with a UPS.
Electrostatic discharge is a threat to GC electronics
The printed circuit (PC) boards in the GC can be damaged by electrostatic
discharge. Do not touch any of the boards unless it is absolutely
necessary. If you must handle them, wear a grounded wrist strap and take
other antistatic precautions. Wear a grounded wrist strap any time you
must remove the GC right side cover.
Many parts are dangerously hot
Many parts of the GC operate at temperatures high enough to cause
serious burns. These parts include but are not limited to:
• The inlets
• The oven and its contents
• The detectors
• The column nuts attaching the column to an inlet or detector
• The valve box
Introduction
7890 Series GC Safety Manual5
You should always cool these areas of the GC to room temperature before
working on them. They will cool faster if you first set the temperature of
the heated zone to room temperature. Turn the zone off after it has
reached the setpoint. If you must perform maintenance on hot parts, use a
wrench and wear thermally protective gloves. Whenever possible, cool the
part of the instrument that you will be maintaining before you begin
working on it.
Be careful when working behind the instrument. During cool-down cycles, the GC
emits hot exhaust which can cause burns.
Page 6
Introduction
WARNING
WARNING
Oven thermal leaks
The insulation around the inlets, detectors, valve box, and the insulation cups is
made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend
the following safety procedures: ventilate your work area; wear long sleeves,
gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation
in a sealed plastic bag; wash your hands with mild soap and cold water after
handling the insulation.
Objects passing through the oven door seal can cause thermal leaks which create
hazardous hot spots which cause burns and melt equipment.
Do not allow wiring or temperature probes to pass through the oven door
jam. Agilent recommends using one of the access holes.
67890 Series GC Safety Manual
Page 7
Hydrogen Safety
WARNING
Hydrogen gas may be used as carrier gas, and/or as fuel for the FID, FPD,
and NPD. When mixed with air, hydrogen can form explosive mixtures.
When using hydrogen (H2) as the carrier gas or fuel gas, be aware that hydrogen
gas can flow into the GC oven and create an explosion hazard. Therefore, be sure
that the supply is turned off until all connections are made and ensure that the inlet
and detector column fittings are either connected to a column or capped at all times
when hydrogen gas is supplied to the instrument.
Hydrogen is flammable. Leaks, when confined in an enclosed space, may create a
fire or explosion hazard. In any application using hydrogen, leak test all
connections, lines, and valves before operating the instrument. Always turn off the
hydrogen supply at its source before working on the instrument.
Hydrogen is a commonly used GC carrier gas. Hydrogen is potentially
explosive and has other dangerous characteristics.
• Hydrogen is combustible over a wide range of concentrations. At
• Hydrogen has the highest burning velocity of any gas.
• Hydrogen has a very low ignition energy.
• Hydrogen that is allowed to expand rapidly from high pressure into the
• Hydrogen burns with a nonluminous flame which can be invisible under
Introduction
atmospheric pressure, hydrogen is combustible at concentrations from
4% to 74.2% by volume.
atmosphere can self- ignite due to an electrostatic spark.
bright light.
7890 Series GC Safety Manual7
GC precautions
When using hydrogen as a carrier gas, remove the large round plastic
cover for the MSD transfer line located on the GC left side panel. In the
unlikely event of an explosion, this cover may dislodge.
Page 8
Introduction
WARNING
Hydrogen shutdown
Hydrogen gas may be used as a carrier or as fuel for some detectors.
When mixed with air, hydrogen can form explosive mixtures.
The GC monitors inlet and auxiliary gas streams. If a stream shuts down
because it is unable to reach its flow or pressure setpoint and if that
stream is configured to use hydrogen, the GC assumes that a leak has
occurred and declares a hydrogen safety shutdown. The effects are:
• The offending channel and any associated channels (such as septum
purge) are set off.
• The split valves in the split/splitless and PTV inlets open.
• The oven (heater and fan) turns off.
• The small heated zones are turned off.
• An alarm tone sounds.
To recover from this state, fix the cause of the shutdown (tank valve
closed, serious leak, others). Turn the instrument off, then back on.
The GC cannot always detect leaks in inlet and/or detector gas streams. For this
reason, it is vital that column fittings should always be either connected to a
column, or have a cap or plug installed. The H
streams must be configured for
2
hydrogen so that the GC is aware of hydrogen use.
87890 Series GC Safety Manual
Page 9
Dangers unique to GC/MSD operation
WARNING
Hydrogen presents a number of dangers. Some are general, others are
unique to GC or GC/MSD operation. Dangers include, but are not limited
to:
• Combustion of leaking hydrogen.
• Combustion due to rapid expansion of hydrogen from a high- pressure
cylinder.
• Accumulation of hydrogen in the GC oven and subsequent combustion
(see your GC documentation and the label on the top edge of the GC
oven door).
• Accumulation of hydrogen in the MSD and subsequent combustion.
Hydrogen accumulation in a GC/MSD
The GC/MSD cannot always detect leaks in inlet and/or detector gas streams. For
this reason, it is vital that column fittings should always be either connected to a
column, or have a cap or plug installed. The H
hydrogen so that the GC is aware of hydrogen use.
Introduction
streams must be configured for
2
7890 Series GC Safety Manual9
All users should be aware of the mechanisms by which hydrogen can
accumulate (Table 1) and know what precautions to take if they know or
suspect that hydrogen has accumulated. Note that these mechanisms apply
to all mass spectrometers, including the GC/MSD.
Tab l e 1Potential hydrogen accumulation mechanisms in GC/MSD
MechanismResults
Mass spectrometer turned off A mass spectrometer can be shut down
deliberately. It can also be shut down accidentally
by an internal or external failure. A mass
spectrometer shutdown does not shut off the flow
of carrier gas. As a result, hydrogen may slowly
accumulate in the mass spectrometer.
Page 10
Introduction
Tab l e 1Potential hydrogen accumulation mechanisms in GC/MSD (continued)
MechanismResults
Mass spectrometer automated isolation
valves closed
Mass spectrometer manual isolation
valves closed
GC off A GC can be shut down deliberately. It can also be
Some mass spectrometers are equipped with
automated diffusion pump isolation valves. In these
instruments, deliberate operator action or various
failures can cause the isolation valves to close.
Isolation valve closure does not shut off the flow of
carrier gas. As a result, hydrogen may slowly
accumulate in the mass spectrometer.
Some mass spectrometers are equipped with
manual diffusion pump isolation valves. In these
instruments, the operator can close the isolation
valves. Closing the isolation valves does not shut
off the flow of carrier gas. As a result, hydrogen
may slowly accumulate in the mass spectrometer.
shut down accidentally by an internal or external
failure. Different GCs react in different ways. If a
7890 Series GC equipped with Electronic Pressure
Control (EPC) is shut off, the EPC stops the flow of
carrier gas. If the carrier flow is not under EPC
control, the flow increases to its maximum. This
flow may be more than some mass spectrometers
can pump away, resulting in the accumulation of
hydrogen in the mass spectrometer. If the mass
spectrometer is shut off at the same time, the
accumulation can be fairly rapid.
Power failureIf the power fails, both the GC and mass
spectrometer shut down. The carrier gas, however,
is not necessarily shut down. As described
previously, in some GCs a power failure may cause
the carrier gas flow to be set to maximum. As a
result, hydrogen may accumulate in the mass
spectrometer.
107890 Series GC Safety Manual
Page 11
Introduction
WARNING
WARNING
WARNING
Tab l e 1Potential hydrogen accumulation mechanisms in GC/MSD (continued)
MechanismResults
Major GC leaks or a column is missing or
broken
Once hydrogen has accumulated in a system, extreme caution must be used when
removing it. Incorrect startup of a system filled with hydrogen can cause an
explosion.
After a power failure, the mass spectrometer may start up and begin the pumpdown
process by itself. This does not guarantee that all hydrogen has been removed from
the system or that the explosion hazard has been removed.
Precautions
Take the following precautions when operating a GC/MSD system with
hydrogen carrier gas.
If the GC is off and configured for hydrogen, the
rear oven flap is open to vent out the hydrogen.
Proper lab venting, as described in the Site Prep
manual, is recommended. If there is a significant
leak, such as a missing column, the GC inlet, PCM,
and Aux are provided with flow limiting frits to
minimize any hydrogen accumulation within the
oven. When the GC is on, major leaks are detected
automatically.
7890 Series GC Safety Manual11
Equipment precaution
You MUST make sure the front side-plate thumbscrew is fastened
finger- tight. Do not overtighten the thumbscrew; it can cause air leaks.
Failure to secure your MSD as described above greatly increases the chance of
personal injury in the event of an explosion.
Page 12
Introduction
You must remove the plastic cover over the glass window on the front of
a 5975 MSD. In the unlikely event of an explosion, this cover may
dislodge.
General laboratory precautions
• Avoid leaks in the carrier gas lines. Use leak- checking equipment to
periodically check for hydrogen leaks.
• Eliminate from your laboratory as many ignition sources as possible
(open flames, devices that can spark, sources of static electricity, etc.).
• Do not allow hydrogen from a high pressure cylinder to vent directly to
atmosphere (danger of self- ignition).
• Use a hydrogen generator instead of bottled hydrogen.
• Provide proper system ventilation as described in the Site Prep manual.
Operating precautions
• Turn off the hydrogen at its source every time you shut down the GC
or MSD.
• Turn off the hydrogen at its source every time you vent the MSD (do
not heat the capillary column without carrier gas f low).
• Turn off the hydrogen at its source every time isolation valves in an
MSD are closed (do not heat the capillary column without carrier gas
flow).
• Turn off the hydrogen at its source if a power failure occurs.
• If a power failure occurs while the GC/MSD system is unattended, even
if the system has restarted by itself:
1 Immediately turn off the hydrogen at its source.
2 Turn off the GC.
3 Turn off the MSD and allow it to cool for 1 hour.
4 Eliminate all potential sources of ignition in the room.
5 Open the vacuum manifold of the MSD to atmosphere.
6 Wait at least 10 minutes to allow any hydrogen to dissipate.
7 Start up the GC and MSD as normal.
127890 Series GC Safety Manual
Page 13
When using hydrogen gas, check the system for leaks to prevent possible
WARNING
fire and explosion hazards based on local Environmental Health and
Safety (EHS) requirements. Always check for leaks after changing a tank
or servicing the gas lines. Always make sure the vent line is vented into a
fume hood.
Measuring hydrogen gas flows
Do not measure hydrogen together with air or oxygen. This can create explosive
mixtures that may be ignited by the automatic ignitor.
To avoid this hazard:
Turn the automatic ignitor off before you begin.
Always measure gases separately.
When measuring gas flows on a detector using hydrogen for the detector
flame or carrier gas, measure the hydrogen flow separately. Never allow
an air stream to enter when hydrogen is present in the flow meter.
Introduction
7890 Series GC Safety Manual13
Page 14
Introduction
WARNING
WARNING
Micro-Electron Capture Detector (µECD)
The µECD contains a cell plated with 63Ni, a radioactive isotope. The beta
particles released at the energy level in the detector have little penetrating
power—the surface layer of the skin or a few sheets of paper will stop
most of them—but they may be hazardous if the isotope is ingested or
inhaled. For this reason, handle the cell with care. Cap the detector inlet
and outlet fittings when the detector is not in use. Never introduce
corrosive chemicals into the detector. Vent detector exhaust outside the
laboratory environment.
Refer to the safety documentation provided with the detector for
important details about safety, maintenance, and compliance with local
government regulation.
Materials that may react with the 63Ni source, either to form volatile products or to
cause physical degradation of the plated film, must be avoided. These materials
include oxidizing compounds, acids, wet halogens, wet nitric acid, ammonium
hydroxide, hydrogen sulfide, PCPs, and carbon monoxide. This list is not exhaustive
but indicates the kinds of compounds that may cause damage to
63
Ni detectors.
147890 Series GC Safety Manual
In the extremely unlikely event that the oven or the detector-heated zone should go
into thermal runaway (maximum, uncontrolled heating in excess of 400 °C) and the
detector remains exposed to this condition for more than 12 hours, take the
following steps:
1 After turning off the main power and allowing the instrument to cool to room
temperature, cap the detector inlet and exhaust vent openings. Wear disposable
plastic gloves and observe normal laboratory safety precautions.
2 Contact your local Agilent Technologies sales office or distributor for ECD disposal
instructions
3 Include a letter stating the condition of abuse.
It is unlikely, even in this very unusual situation, that radioactive material will
escape the cell. However, permanent damage to the
63
Ni plating within the cell is
possible; therefore, the cell must be returned for exchange.
Page 15
Introduction
WARNING
WARNING
Do not use solvents to clean the µECD.
You may not open the µECD cell unless authorized to do so by your local nuclear
regulatory agency. Do not disturb the four socket-head bolts. These hold the cell
halves together. United States customers removing or disturbing them is a violation
of the terms of the exemption and could create a safety hazard.
When handling µECDs:
• Never eat, drink, or smoke.
• Always wear safety glasses when working with or near open µECDs.
• Wear protective clothing such as laboratory jackets, safety glasses, and
gloves, and follow good laboratory practices. Wash hands thoroughly
with a mild nonabrasive cleaner after handling µECDs.
• Cap the inlet and outlet fittings when the µECD is not in use.
• Connect the µECD exhaust vent to a fume hood or vent it to the
outside. See the latest revision of 10 CFR Part 20 (including Appendix
B), or the applicable state regulation. For other countries, consult with
the appropriate agency for equivalent requirements.
7890 Series GC Safety Manual15
Agilent Technologies recommends a vent line internal diameter of 6 mm
(1/4- inch) or greater. With a line of this diameter, the length is not
critical.
Page 16
Introduction
Fuses and Batteries
The GC requires fuses and batteries for proper operation. These must only
be accessed by Agilent trained service personnel.
Tab l e 2AC board fuses
Fuse designationLine voltageFuse rating and type
F1, F2All20 A, 250 Vac, IEC 127 type f (non-time delay),
F3, F4All8 A, 250 Vac, IEC type f (non-time delay),
Tab l e 3Logic board battery (7890A only)
Battery designationBattery rating and type
BT13-volt lithium-poly carbon battery, .048A-HR,
ceramic body
glass body
Panasonic model BR 1225
167890 Series GC Safety Manual
Page 17
Safety and Regulatory Certifications
The Agilent 7890 Series GC conforms to the following safety standards:
• Canadian Standards Association (CSA): C22.2 No. 61010- 1
• CSA/Nationally Recognized Test Laboratory (NRTL): UL 61010- 1
• International Electrotechnical Commission (IEC): 61010–1
• EuroNorm (EN): 61010–1
The Agilent 7890 Series GC conforms to the following regulations on
Electromagnetic Compatibility (EMC) and Radio Frequency Interference
(RFI):
• CISPR 11/EN 55011: Group 1, Class A
• IEC/EN 61326
• AUS/NZ
The 7890B GC complies with the following company standard:
• Q/YXFG27
This ISM device complies with Canadian ICES- 001. Cet appareil ISM est
conforme a la norme NMB—001 du Canada.
Introduction
The Agilent 7890 Series GC is designed and manufactured under a quality
system registered to ISO 9001. Declaration of Conformity available.
Instructions for Disposal of Waste Equipment by Users in the European
Union. This symbol on the product or its packaging indicates that this
product must not be disposed of with other waste. Instead, it is your
responsibility to dispose of your waste equipment by handing it over to a
designated collection point for the recycling of waste electrical and
electronic equipment. The separate collection and recycling of your waste
equipment at the time of disposal will help conserve natural resources and
ensure that it is recycled in a manner that protects human health and the
7890 Series GC Safety Manual17
Page 18
Introduction
Information
environment. For more information about where you can drop off your
waste equipment for recycling, please contact your local city recycling
office or the dealer from whom you originally purchased the product.
The Agilent Technologies 7890 Series Gas Chromatograph meets the
following IEC (International Electro- technical Commission) classifications:
Safety Class I, Transient Overvoltage Category II, Pollution Degree 2.
This unit has been designed and tested in accordance with recognized
safety standards and is designed for use indoors in non- classified
locations. If the instrument is used in a manner not specified by the
manufacturer, the protection provided by the instrument may be impaired.
Whenever the safety protection of the Agilent 7890 Series Gas
Chromatograph has been compromised, disconnect the unit from all power
sources and secure the unit against unintended operation.
Refer servicing to qualified service personnel. Substituting parts or
performing any unauthorized modification to the instrument may result in
a safety hazard.
187890 Series GC Safety Manual
Page 19
Symbols
Introduction
Warnings in the manual or on the instrument must be observed during all
phases of operation, service, and repair of this instrument. Failure to
comply with these precautions violates safety standards of design and the
intended use of the instrument. Agilent Technologies assumes no liability
for the customer’s failure to comply with these requirements.
See accompanying instructions for more
information.
Indicates a hot surface.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
Indicates potential explosion hazard.
Indicates radioactivity hazard.
Indicates electrostatic discharge hazard.
Indicates a hazard. See the Agilent 7890
Series GC user documentation for the
item labeled.
Indicates that you must not discard this
electrical/electronic product in domestic
household waste
7890 Series GC Safety Manual19
Page 20
Introduction
Electromagnetic compatibility
This device complies with the requirements of CISPR 11. Operation is
subject to the following two conditions:
• This device may not cause harmful interference.
• This device must accept any interference received, including
interference that may cause undesired operation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on,
the user is encouraged to try one or more of the following measures:
1 Relocate the radio or antenna.
2 Move the device away from the radio or television.
3 Plug the device into a different electrical outlet, so that the device and
the radio or television are on separate electrical circuits.
4 Make sure that all peripheral devices are also certified.
5 Make sure that appropriate cables are used to connect the device to
peripheral equipment.
6 Consult your equipment dealer, Agilent Technologies, or an experienced
technician for assistance.
7 Changes or modifications not expressly approved by Agilent
Technologies could void the user’s authority to operate the equipment.
Sound Emission Certification for Federal Republic of Germany
Sound pressure
Sound pressure Lp < 70 dB(A) according to DIN- EN 27779.
Schalldruckpegel
Schalldruckpegel LP < 70 dB(A) nach DIN- EN 27779.
207890 Series GC Safety Manual
Page 21
Intended Use
Agilent products must only be used in the manner described in the
Agilent product user guides. Any other use may result in damage to the
product or personal injury. Agilent is not responsible for any damages
caused, in whole or in part, by improper use of the products,
unauthorized alterations, adjustments or modifications to the products,
failure to comply with procedures in Agilent product user guides, or use
of the products in violation of applicable laws, rules or regulations.
Cleaning
To clean the unit, disconnect the power and wipe down with a damp,
lint- free cloth.
Recycling the Product
Introduction
For recycling, contact your local Agilent sales office.
7890 Series GC Safety Manual21
Page 22
Agilent Technologies
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.