Agilent 70420A Users Guide

Agilent E5505A Phase Noise Measurement System
User’s Guide
First edition, June 2004
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2004
Microsoft ® is a U.S. registered trademark of Microsoft Corporation.
Windows ® and MS Windows ® are U.S. registered trademarks of Micr osoft Corp orat ion.
Pentium ® Intel Corporation.
Adobe ® Acrobat ® Adobe, the Adobe
Logo, Acrobat and the Acrobat Logo are
trademarks of Adobe Systems
Incorporated.
is a U.S. registered trademark of
Manual Part Number
E5505--90003
Edition
First edition, June 2004
Printed in USA
Agilent Technologies, Inc. 1400 Fountaingrove Pkwy Santa Rosa, CA 9540 3
Warranty
The material contained in this docu­ment is provided as is, and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Gov­ernment will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1- 2) (June 1987). U.S. Govern­ment users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
If software is for use in the performance of a U.S. Government prime contract or sub­contract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-701 4 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent

Contents

1 Getting Started
2 Introduction and Measurement
Introduction 26
Documentation Map 27
Table 1. E5505A users guide map 27
Additional Documentation 28
Figure 1. Navigate to system documentation 28
System Overview 29
Figure 2. E5505A benchtop system, typical configuration 30 Table 2. Equivalent system/instrument model numbers 30
Introducing the GUI 32
Figure 3. E5500 graphical user interface (GUI) 33
Designing to Meet Your Needs 34
Beginning 34
E5505A Operation: A Guided Tour 35
Required equipment 35 How to begin 35
Powering the System On 36
To power on a racked system 36 To power on a benchtop system 36
Starting the Measurement Software 37
Figure 4. Navigation to the E5500 user interface 37 Figure 5. Phase noise measurement subsystem main screen 38
Performing a Confidence Test 39
Figure 6. Opening the file containing pre-stored parameters 39 Figure 7. Navigating to the Define Measurement window 40 Beginning a measurement 40 Figure 8. Navigating to the New Measurement window 40 Figure 9. Confirm new measurement 41 Figure 10. Setup diagram displayed during the confidence test. 41 Figure 11.Connect diagram example 42 Making a measurement 42 Figure 12. Typical phase noise curve for test set confidence test 43 Sweep segments 43
Agilent E5505A Users Guid e 3
Congratulations 43 Learning more 44 Table 3. Parameter data for the N5500A confidence test example 44
Powering the System Off 45
To power off a racked sy stem 45 To power off a be nchtop system 45 Using the E5500 Shutdown Utility 45 Figure 13. Shutdown utility icon 45
3Phase Noise Basics
What is Phase Noise? 48
Figure 14.RF sideband spectrum 49 Phase terms 49 Figure 15 . CW signal sidebands viewed in the frequency domain 50 Figure 16. Deriving L(f) from a RF analyzer display 51 Figure 17.L(f) Described Logarithmically as a Function of Offset Frequency 51 Figure 18. Region of validity of L(f) 52
4 Expanding Your Measurement Experience
Starting the Measurement Software 54
Figure 19. Navigate to E5500 user interface 54
Using the Asset Manager 55
Configuring an asset 55 Figure 20. Navigate to Asset Manager 55 Figure 21. Navigate to Add in Asset Manager 56 Figure 22. Select source as asset type 56 Figure 23. Choose source 57 Figure 24. Select I/O library 57 Figure 25. Enter asset and serial number 58 Figure 26. Enter comment 58 Figure 27. Click check-mark button 59 Figure 28.Confirmation message 59
Using the Server Hardware Connections to Specify the Source 60
Figure 29. Navigate to server hardware connections 60 Figure 30. Select Sources tab 60 Figure 31. Successful I/O check 61 Figure 32.Failed I/O check 61
Setting GPIB Addresses 63
Table 4. Default GPIB addresses 63
4 Agilent E5505A Users Guide
Figure 33 . Asset Manager on System menu 64 Figure 34. Asset Manager window 64 Figure 35 . GPIB address dialog box 65
Testing the 8663A Internal/External 10 MHz 66
Required equipment 66 Defining the measurement 66 Figure 36. Select the parameters definition file 66 Figure 37. Enter Source Information 67 Table 5. Tuning characteristics for various sources 68 Selecting a reference source 68 Figure 38.Selecting a reference source 68 Selecting loop suppression verification 69 Figure 39. Selecting loop suppression verification 69 Setting up for the 8663A 10 MHz measurement 69 Figure 40. Noise floor for the 8663 10 MHz measurement 70 Figure 41.Noise floor example 70 Beginning the measurement 71 Figure 42. Selecting new measurement 71 Figure 43. Confirm new measurement 71 Figure 44. Connection diagram 72 Table 6.Test set signal input limits and characteristics 73 Sweep segments 75 Figure 45. Oscilloscope display of beatnote from test set monitor port 76 Making the measurement 76 Figure 46. Selecting suppression 77 Figure 47. Typical phase noise curve for an 8663A 10 MHz measurement 78 Table 7. Parameter data for the 8663A 10 MHz measurement 79
Testing the 8644B Internal/External 10 MHz 81
Required equipment 81 Defining the measurement 81 Figure 48. Select the parameters definition file 81 Figure 49. Sources tab in define measurement window 82 Table 8. Tuning characteristics for various sources 83 Selecting a reference source 83 Figure 50.Selecting a reference source 84 Selecting loop suppression verification 84 Figure 51. Selecting loop suppression verification 85 Setting up the 8663A 10 MHz measurement 85 Figure 52. Noise floor for the 8644B 10 MHz measurement 85 Figure 53.Noise floor example 86 Beginning the measurement 87
Agilent E5505A Users Guid e 5
Figure 54. Selecting a new measurement 87 Figure 55. Confirm measurement dialog box 87 Figure 56.Connect diagram dialog box 88 Table 9.Test set signal input limits and characteristics 89 Figure 57. Oscilloscope display of beatnote from test set monitor port 91 Making the measurement 92 Figure 58. Suppression selections 92 Figure 59. Typical phase noise curve for an 8644B 10 MHz measurement. 93 Table 10. Parameter data for the 8644B 10 MHz measurement 94
Viewing Ma rkers 96
Figure 60. Navigate to markers 96 Figure 61 . Adding and deleting markers 96
Omitting Spurs 97
Figure 62. Navigate to display preferences 97 Figure 63. Uncheck spurs 97 Figure 64.Graph displayed without spurs 98
Displaying the Parameter Summary 99
Figure 65. Navigate to parameter summary 99 Figure 66. Parameter summary 100
Exporting Measurement Results 101
Figure 67. Export results choices 101 Exporting Trace Data 102 Figure 68.Trace data results 102 Exporting spur data 103 Figure 69. Spur data results 103 Exporting X-Y data 104 Figure 70.X-Y data results 104
5 Absolute Measurement Fundamentals
The Phase-Lock-Loop Technique 106
Understanding the Phase-Lock-Loop Technique 106 Figure 71. Simplified block diagram of the phase lock loop configuration 106 The Phase-Lock-Loop Circuit 106 Figure 72. Capture and drift-tracking range with tuning range of VCO 107 Figure 73. Capture and drift-tracking ranges and beatnote frequency 108
What Sets the Measurement Noise Floor? 110
The System Noise Floor 110 Table 11. Amplitude ranges for L and R ports 110 Figure 74. Relationship between the R input level and system noise floor 110
6 Agilent E5505A Users Guide
The Noise Level of the Reference Source 111 Figure 75. Reference source noise approaches DUT noise 111
Selecting a Reference 112
Figure 76. DUT noise approaches reference noise 112 Using a Similar Device 112 Using a Signal Generator 113 Tuning Requirements 113 Table 12.Tuning Characteristics of Various VCO Source Options 113 Figure 77.Voltage tuning range limits relative to center voltage of the VCO tuning
curve 114
Estimating the Tuning Constant 115
Table 13. VCO tuning constant calibration method 115
Tracking Frequency Drift 116
Evaluating beatnote drift 116
Changing the PTR 118
Figure 78. Peak tuning range 118 The Tuning Qualifications 118
Minimizing Injection Locking 120
Adding Isolation 120 Increasing the PLL Bandwidth 120 Figure 79. Peak tuning range (PTR) Required by injection locking. 121
Inserting a Device 122
An attenuator 122 Figure 80. Measurement noise floor relative to R-Port signal level 122 An amplifier 123 Figure 81.Measurement noise floor as a result of an added attenuator 123
Evaluating Noise Above the Small Angle Line 124
Determining the Phase-Lock-Loop bandwidth 124 Figure 82. Phase lock loop bandwidth provided by the peak tuning range 125 Figure 83. Graph of small angle line and spur limit 126 Figure 84. Requirements for noise exceeding small angle limit 127
6 Absolute Measurement Examples
Stable RF Oscillator 130
Required equipment 130 Defining the measurement 130 Figure 85. Select the parameters definition file 130 Figure 86. Enter source information 131 Table 14.Tuning characteristics for various sources 132
Agilent E5505A Users Guid e 7
Selecting a reference source 132 Figure 87.Selecting a reference source 133 Selecting Loop Suppression Verification 133 Figure 88. Selecting loop suppression verification 134 Setup considerations for stable RF oscillator measurement 134 Figure 89. Noise floor for the stable RF oscillator measurement 135 Figure 90.Noise floor calculation example 135 Beginning the measurement 136 Figure 91. Selecting a new measurement 136 Figure 92. Confirm new measurement 136 Figure 93.Connect diagram for the stable RF oscillator measurement 137 Table 15.Test set signal input limits and characteristics 138 Checking the beatnote 139 Figure 94. Oscilloscope display of beatnote from test set Monitor port 140 Making the measurement 140 Figure 95. Selecting suppressions 141 Figure 96.Typical phase noise curve for a stable RF oscillator 142 Table 16. Parameter data for the stable RF oscillator measurement 143
Free-Running RF Oscillator 145
Required equipment 145 Defining the measurement 145 Figure 97. Select the parameters definition file 146 Figure 98. Enter source information 147 Table 17.Tuning characteristics for various sources 147 Selecting a reference source 148 Figure 99.Selecting a reference source 148 Selecting Loop Suppression Verification 148 Figure 100. Selecting loop suppression verification 149 Setup considerations for the free-running RF oscillator measurement 149 Figure 101. Noise floor for the free-running RF oscillator measurement 150 Figure 102. Noise floor calculation example 150 Beginning the measurement 151 Figure 103. Selecting a new measurement 151 Figure 104. Confirm measurement dialog box 151 Figure 105. Connect diagram for the free-running RF oscillator measurement 152 Table 18.Test set signal input limits and characteristics 153 Checking the beatnote 153 Figure 106. Oscilloscope display of beatnote from test set Monitor port 154 Making the measurement 155 Figure 107. Selecting suppressions 156 Figure 108.Typical phase noise curve for a free-running RF oscillator 157
8 Agilent E5505A Users Guide
Table 19. Parameter data for the free-running RF oscillator measurement 158
RF Synthesizer Using DCFM 160
Required equipment 160 Defining the measurement 160 Figure 109. Select the parameters definition file 160 Figure 110. Enter source information 161 Table 20.Tuning characteristics for various sources 162 Selecting a reference source 162 Figure 111. Selecting a reference source 162 Selecting Loop Suppression Verification 163 Figure 112. Selecting loop suppression verification 163 Setup considerations for the RF synthesizer using DCFM measurement 163 Figure 113. Noise floor for the RF synthesizer (DCFM) measurement 164 Figure 114. Noise floor calculation example 164 Beginning the measurement 165 Figure 115. Selecting a new measurement 165 Figure 116. Confirm measurement dialog box 165 Figure 117. Connect diagram for the RF synthesizer (DCFM) measurement 166 Table 21.Test set signal input limits and characteristics 167 Checking the beatnote 167 Figure 118. Oscilloscope display of beatnote from the test set Monitor port 168 Making the measurement 169 Figure 119. Selecting suppressions 169 Figure 120. Typical phase noise curve for an RF synthesizer using DCFM 170 Table 22. Parameter Data for the RF Synthesizer (DCFM) Measurement 171
RF Synthesizer Using EFC 173
Required equipment 173 Defining the measurement 173 Figure 121. Select the parameters definition file 173 Figure 122. Enter Source Information 175 Table 23.Tuning Characteristics for Various Sources 175 Selecting a reference source 176 Figure 123. Selecting a reference source 176 Selecting Loop Suppression Verification 176 Figure 124. Selecting Loop suppression verification 177 Setup considerations for the RF synthesizer using EFC measurement 177 Figure 125. Noise floor for the RF synthesizer (EFC) measurement 178 Figure 126. Noise floor calculation example 178 Beginning the measurement 179 Figure 127. Selecting a new measurement 179 Figure 128. Confirm measurement dialog box 179
Agilent E5505A Users Guid e 9
Figure 129. Connect diagram for the RF synthesizer (EFC) measurement 180 Table 24. Test set signal Input Limits and Characteristics 181 Checking the beatnote 181 Figure 130. Oscilloscope display of a beatnote from the test set Monitor port 182 Making the measurement 182 Figure 131. Selecting suppressions 183 Figure 132.Typical phase noise curve for an RF synthesizer using EFC 184 Table 25. Parameter data for the RF synthesizer (EFC) measurement 185
Microwave Source 187
Required equipment 187 Defining the measurement 187 Figure 133. Select the parameters definition file 187 Figure 134. Enter source information 189 Table 26.Tuning characteristics for various sources 189 Selecting a reference source 190 Figure 135. Selecting a reference source 190 Selecting Loop Suppression Verification 190 Figure 136. Selecting loop suppression verification 191 Setup considerations for the microwave source measurement 191 Figure 137. Noise characteristics for the microwave measurement 191 Beginning the measurement 192 Figure 138. Selecting a new measurement 192 Figure 139. Confirm measurement dialog box 192 Figure 140. Connect diagram for the microwave source measurement 193 Table 27.Test set signal input limits and characteristics 194 Checking the beatnote 194 Figure 141. Oscilloscope display of a beatnote from the test set Monitor port 195 Making the measurement 196 Figure 142. Selecting suppressions 197 Figure 143.Typical phase noise curve for a microwave source 198 Table 28. Parameter data for the microwave source measurement 199
7 Residual Measurement Fundamentals
What is Residual Noise? 202
The noise mechanisms 202 Figure 144. Additive noise components 202 Figure 145. Multiplicative noise components 203
Assumptions about Residual Phase Noise Measurements 204
Figure 146. Setup for typical residual phase noise measurement 204 Frequency translation devices 205
10 Agilent E5505A Users Guide
Figure 147. Measurement setup for two similar DUTs 205
Calibrating the Measurement 206
Figure 148. General equipment setup for making residual phase noise measurements 206
Calibration and measurement guidelines 207 Calibration options 208 User entry of phase detector constant 209 Figure 149. Measuring power at phase detector signal input port 210 Table 29. Acceptable amplitude ranges for the phase detectors. 210 Figure 150. Phase detector sensitivity 211 Figure 151. Adjust for quadrature 212 Figure 152. Measuring power at phase detector reference input port 212 Measured ± DC peak voltage 213 Figure 153. Connection to optional oscilloscope for determining voltage peaks 214 Table 30. Acceptable Amplitude Ranges for the Phase Detectors 214 Measured beatnote 215 Table 31. Frequency ranges 215 Procedure 216 Figure 154. Measuring power from splitter 216 Table 32. Acceptable amplitude ranges for the phase detectors 216 Figure 155. Calibration source beatnote injection 217 Synthesized residual measurement using beatnote cal 217 Table 33. Frequency Ranges 217 Procedure 218 Figure 156. Synthesized residual measurement using beatnote cal 218 Double-Sided spur 218 Figure 157. Calibration setup 219 Table 34. Acceptable amplitude ranges for the phase detectors 220 Figure 15 8. Measuring carrier-to-sideband ratio of the modulated port 220 Single-Sided spur 221 Figure 159. Calibration setup for single-sided spur 222 Table 35. Acceptable Amplitude Ranges for the Phase Detectors 223 Figure 160. Carrier-to-spur ratio of modulated signal 223 Figure 161. Carrier-to-spur ratio of non-modulated signal 224
Measurement Difficulties 225
System connections 225
8 Residual Measurement Examples
Amplifier Measurement Example 228
Required equipment 228 Figure 162. Setup for residual phase noise measurement 229
Agilent E5505A Users Guid e 11
Defining the measurement 229 Figure 163. Select the parameters definition file 229 Figure 164. Navigate to residual phase noise 230 Figure 165. Enter frequencies into source tab 231 Figure 166. Select constant in the cal tab 231 Figure 167. Select parameters in the block diagram tab 232 Figure 168. Select graph description on graph tab 233 Setup considerations for amplifier measurement 233 Beginning the measurement 234 Figure 169. Select meter from view menu 234 Figure 170. Selecting New Measurement 234 Figure 171. Confirm new measurement 235 Figure 172. Setup diagram for the 8349A amplifier measurement example 235 Table 36.Test set signal input limits and characteristics 236 Making the measurement 237 Table 37. Acceptable amplitude ranges for the phase detectors 237 Figure 173. Residual connect diagram example 238 Figure 174. Connection to optional oscilloscope for determining voltage peaks 238 Figure 175.Adjust phase difference at phase detector 239 Figure 176. Adjust phase shifter until meter indicates 0 volts 240 When the measurement is complete 240 Figure 177. Typical phase noise curve for a residual measurement 241 Table 38. Parameter data for the amplifier measurement example 241
9 FM Discriminator Fundamentals
The Frequency Discriminator Method 244
Figure 178. Basic delay line/mixer frequency discriminator method 244 Basic theory 244 The discriminator transfer response 245 Figure 179. Nulls in sensitivity of delay line discriminator 246 Table 39. Choosing a delay line 248
10 FM Discriminator Measurement Examples
Introduction 250
Figure 180.FM Discriminator measurement setup 250
FM Discriminator Measurement using Double-Sided Spur Calibration 251
Table 40. Required Equipment for the FM Discriminator Measurement Example 251 Determining the discriminator (delay line) length 251 Figure 181. Discriminator noise floor as a function of delay time 252 Defining the measurement 252 Figure 182. Select the parameters definition file 252
12 Agilent E5505A Users Guide
Figure 183. Select measurement type 253 Figure 184. Enter frequencies in source tab 254 Figure 185. Enter parameters into the call tab 255 Figure 186. Select parameters in the block diagram tab 256 Figure 187. Select Graph Description on Graph Tab 256 Setup considerations 257 Beginning the measurement 258 Figure 188. Select meter from view menu 258 Figure 189. Selecting New Measurement 258 Figure 190. Confirm new measurement 259 Figure 191. Setup diagram for the FM discrimination measurement example 259 Table 41. Test Set Signal Input Limits and Characteristics 260 Figure 192. Connect diagram example 261 Making the measurement 261 Figure 193. Calibration measurement (1 of 5) 262 Figure 194. Calibration measurement (2 of 5) 262 Figure 195. Calibration measurement (3 of 5) 263 Figure 196. Calibration measurement (4 of 5) 263 Figure 197. Calibration measurement (5 of 5) 263 When the measurement is complete 264 Figure 198.Typical phase noise curve using double-sided spur calibration 264 Table 42. Parameter data for the double-sided spur calibration example 265
Discriminator Measurement using FM Rate and Deviation Calibration 267
Required equipment 267 Table 43. Required equipment for the FM discriminator measurement exam ple 267 Determining the discriminator (delay line) length 268 Figure 199. Discriminator noise floor as a function of delay time 268 Defining the measurement 268 Figure 200. Select the parameters definition file 269 Figure 201. Select measurement type 269 Figure 202. Enter frequencies in Source tab 270 Figure 203. Enter parameters into the Cal tab 271 Figure 204. Enter parameters in the Block Diagram tab 272 Figure 205. Select graph description on Graph tab 273 Setup considerations 273 Beginning the measurement 274 Figure 206. Select meter from the View menu 274 Figure 207. Selecting New Measurement 274 Figure 208. Confirm new measurement 275 Figure 209. Setup diagram for the FM Discrimination measurement example 275 Table 44.Test set signal input limits and characteristics 276
Agilent E5505A Users Guid e 13
Figure 210. System connect diagram example 277 Making the measurement 277 Figure 211. Calibration measurement (1 of 5) 278 Figure 212. Calibration measurement (2 of 5) 278 Figure 213. Calibration measurement (3 of 5) 279 Figure 214. Calibration measurement (4 of 5) 279 Figure 215. Calibration measurement (5 of 5) 279 When the measurement is complete 280 Figure 216.Typical phase noise curve using rate and deviation calibration 280 Table 45. Parameter data for the rate and deviation calibration example 281
11 AM Noise Measurement Fundamentals
AM-Noise Measurement Theory of Operation 284
Basic noise measurement 284 Phase noise measurement 284
Amplitude Noise Measurement 285
AM noise measurement block diagrams 285 Figure 217. AM noise system with N5500A opt 001 285 Figure 218. AM noise system with external detector 285 Figure 219. AM Noise system with 70429A Opt K21 AM detector 286 Figure 220. AM noise system with N5507A downconverter 286 AM detector 286 Figure 221. AM detector schematic 286 Table 46. Maximum carrier offset frequency 287
Calibration and Measurement General Guidelines 289
Method 1: User Entry of Phase Detector Constant 290
Method 1, example 1 290 Figure 222. Phase detector constant AM noise setup (method1, example 1) 290 Figure 223. AM noise calibration setup 291 Figure 224. AM detector sensitivity graph 291 Method 1, example 2 292 Figure 225. Phase detector constant AM noise setup (method 1, example 2) 292 Figure 226. Modulation sideband calibration setup 293
Method 2: Double-Sided Spur 294
Method 2, example 1 294 Figure 227. Double-Sided spur AM noise setup (method 2, example 1) 294 Figure 228. Measuring the carrier-to-sideband ratio 295 Figure 229. Measuring the calibration constant 295 Method 2, example 2 296 Figure 230. Double-sided spur AM noise setup (method 2, example 2) 296
14 Agilent E5505A Users Guide
Figure 231. Measuring power at the am detector 296 Figure 232. Measuring carrier-to-sideband ratio 297 Figure 233. Measuring the calibration constant 297
Method 3: Single-Sided Spur 299
Figure 234. AM noise measurement setup using single-sided spur 299 Figure 235. Measuring relative spur level 300 Figure 236. Measuring detector sensitivity 300
12 AM Noise Measurement Examples
AM Noise with N5500A Option 001 302
Required equipment 302 Figure 237. AM noise measurement configuration 302 Defining the measurement 303 Figure 238. Select the parameters definition file 303 Figure 239. Navigate to AM noise 304 Figure 240. Enter Frequencies in Source Tab 304 Figure 241. Enter parameters into the cal tab 305 Figure 242. Select parameters in the block diagram tab 305 Figure 243. Select graph description on graph tab 306 Beginning the measurement 307 Figure 244. Selecting a new measurement 307 Figure 245. Confirm measurement dialog box 307 Figure 246. Connect diagram for the AM noise measurement 308 Table 47.Test set signal input limits and characteristics 309 Figure 247. Connect diagram example 309 Making the measurement 310 When the measurement is complete 310 Figure 248. Typical AM noise curve 310 Table 48. Parameter data for the AM noise using an N5500A Option 001 311
13 Baseband Noise Measurement Examples
Baseband Noise with Test Set Measurement Example 314
Defining the measurement 314 Figure 249. Select the parameters definition file 314 Beginning the measurement 315 Figure 250. Selecting a new measurement 315 Figure 251. Confirm measurement dialog box 315 Figure 252. Connect diagram dialog box 316 Making the measurement 316 Figure 25 3. Typical phase noise curve for a baseband using a test set
measurement. 316
Agilent E5505A Users Guid e 15
Table 49. Parameter data for the baseband using a test set measurement 317
Baseband Noise without Test Set Measurement Example 318
Defining the measurement 318 Figure 254. Select the parameters definition file 318 Beginning the measurement 319 Figure 255. Selecting a new measurement 319 Figure 256. Confirm measurement dialog box 319 Figure 257. Connect diagram for baseband without test set measurement 320 Figure 258. Instrument connection dialog box 320 Making the measurement 320 Figure 259.Typical curve for a baseband without test set measurement. 321 Table 50. Parameter data for the baseband without using a test set measurement 321
14 Evaluating Your Measurement Results
Evaluating the Results 324
Looking for obvious problems 324 Figure 260. Noise plot showing obvious problems 325 Comparing against expected data 325 Figure 261. Compensation for added reference source noise 326 Figure 262. Measurement results and reference source noise 327
Gathering More Data 328
Repeating the measurement 328 Figure 263. Repeating a measurement 328 Doing more research 328
Outputting the Results 329
Using a printer 329
Graph of Results 330
Marker 330 Figure 264. Navigate to marker 330 Figure 26 5. Add and delete markers 331
Omit Spurs 332
Figure 266. Select display preferences 332 Figure 267. Uncheck spurs 332 Figure 268. Graph without spurs 333 Parameter summary 333 Figure 269. Navigate to parameter summary 333 Figure 270. Parameter summary notepad 334
Problem Solving 335
Table 51. List of topics that discuss problem solving in this chapter 335
16 Agilent E5505A Users Guide
Discontinuity in the graph 335 Table 52. Potential causes of discontinuity in the graph 335 Higher noise level 336 Spurs on the graph 336 Table 53. Spurs on the graph 337 Table 54. Actions to eliminate spurs 337 Small angle line 338 Figure 271. L(f) Is only valid for noise levels below the small angle line 339
15 Advanced Software Features
Introduction 342
Phase-Lock-Loop Suppression 343
Figure 272. PLL suppression verification graph 343 PLL suppression parameters 343
Ignore-Out-Of-Lock Mode 346
PLL Suppression Verification Process 347
PLL suppression information 347 Figure 273. Default PLL suppression verification graph 347 Figure 274. Measured loop suppression curve 348 Figure 275. Smoothed loop suppression curve 349 Figure 276.Theoretical loop suppression curve 349 Figure 277. Smoothed vs. theoretical loop suppression curve 350 Figure 278. Smoothed vs. Adjusted theoretical loop suppression curve 350 Figure 279. Adjusted theoretical vs. theoretical loop suppression curve 351 PLL gain change 352 Maximum error 352 Accuracy degradation 352
Blanking Frequency and Amplitude Information on the Phase Noise Graph 353
Security level procedure 353 Figure 280. Navigate to security level 353 Figure 281. Choosing levels of security 354 Figure 282. Unsecured: all data is viewable 354 Figure 283. Choosing levels of security 355 Figure 284. Secured: frequencies cannot be found-1 355 Figure 285. Secured: frequencies cannot be found-2 356 Figure 286. Choosing levels of security 356 Figure 287. Secured: frequencies and amplitudes cannot be viewed 357
Agilent E5505A Users Guid e 17
16 Reference Graphs and Tables
Approximate System Noise Floor vs. R Port Signal Level 360
Figure 288. Noise floor for R input port 360
Phase Noise Floor and Region of Validity 361
Figure 289. Region of validity 361
Phase Noise Level of Various Agilent Sources 362
Figure 290. Noise level for various reference sources 362
Increase in Measured Noise as Ref Source Approaches DUT Noise 363
Figure 291. Reference source and DUT noise levels 363
Approximate Sensitivity of Delay Line Discriminator 364
Figure 292. Delay line discriminator sensitivity 364
AM Calibration 365
Figure 293. AM detector sensitivity 365
Voltage Controlled Source Tuning Requirements 366
Figure 294. Tuning voltage required for phase lock 366
Tune Range of VCO for Center Voltage 367
Figure 295. Tune range of VCO for center voltage 367
Peak Tuning Range Required by Noise Level 368
Figure 296.Typical source noise level vs. minimum tuning range 368
Phase Lock Loop Bandwidth vs. Peak Tuning Range 369
Figure 297. PLL BW vs. peak tuning range 369
Noise Floor Limits Due to Peak Tuning Range 370
Figure 298. Noise at sources peak tuning range 370
Tuning Characteristics of Various VCO Source Options 371
Table 55.Tuning parameters for several VCO options 371
8643A Frequency Limits 372
Table 56. 8643A frequency limits 372 8643A mode keys 372 Table 57. Operating characteristics for 8643A modes 1, 2, and 3 373 How to access special functions 373 Figure 299. 8643A special function keys 373 Description of special functions 120 and 125 373
8644B Frequency Limits 375
Table 58. 8644B frequency limits 375 8644B mode keys 375 Table 59. Operating characteristics for 8644B modes 1, 2, and 3 376
18 Agilent E5505A Users Guide
How to access special functions 376 Figure 300. 8644B special functions keys 376
Description of special function 120 377
8664A Frequency Limits 378
Table 60. 8664A frequency limits 378 8664A mode keys 378 Table 61. Operating characteristics for 8664A modes 2 and 3 378 How to access special functions 379 Figure 301. Special functions keys 379 Description of special functions 120 379
8665A Frequency Limits 380
Table 62. 8665A frequency limits 380 8665A mode keys 380 Table 63. Operating characteristics for 8665A modes 2 and 3 381 How to access special functions 381 Figure 302. 8665A special functions keys 381 Description of Special Functions 120 and 124 381
8665B Frequency Limits 383
Table 64. 8665B frequency limits 383 8665B mode keys 383 Table 65. Operating characteristics for 8665B modes 2 and 3 384 How to access special functions 384 Figure 303. 8665B Special functions keys 384 Description of special functions 120 and 124 385
17 System Specifications
Specifications 388
Table 66. Mechanical and environmental specifications 388 Table 67. Operating characteristics 388 Reliable accuracy 389 Table 68. RF phase detector accuracy 389 Table 69. AM detector accuracy 389 Measurement qualifications 389 Tuning 390 Computer 390
Power Requirements 391
Table 70. E5505A maximum AC power requirements 391
Agilent E5505A Users Guid e 19
18 System Interconnections
Making Connections 394
System Connectors 395
Table 71. E5505A connectors and adapters 395
System Cables 396
Table 72. E5505A cables and connections 396
Connecting Instruments 397
Figure 304. Connect adapter to PC digitizer card 397 Figure 305. PC to test set connection, standard model 398 Figure 306. PC to test set (options 001 and 201) and downconverter connection 399 Figure 307. E5505A system connections with standard test set 401 Figure 308. E5505A system connections with test set option 001 402 Figure 309. E5505A system connections with test set option 201 403
19 PC Components Installation
Overview 406
PC Digitizer Software: Phase 1 407
To install the PC digitizer software 407
Hardware Installation 408
Preparing for installation 408 Figure 310. Remove screws from side of CPU 408 Figure 311. Slide cover off 409 Figure 312. Remove hold-down bar 409 Accessing PC expansion slots 410 Figure 313.Vertically-Mounted expansion slots 410 Taki n g ESD prec a utio n s 410 Installing the PC digitizer card 411 Figure 314. PC digitizer card 411 Figure 315. Insert PC digitizer card 411 Figure 316. Secure card with screw 412 Figure 317. Connect adapter to PC digitizer card 412 Installing the GPIB interface card 413 Figure 318. GPIB interface card 413 Figure 319. Insert GPIB card 413 Figure 320. Secure card with screw 414 Figure 321. Replace cover 414
System Interconnections 415
Table 73. E5505A connectors and adapters 415 Making Connections 415
20 Agilent E5505A Users Guide
Figure 322. Test set connection, standard model 416 Figure 323. Test set (options 001 and 201) and downconverter connection 417
PC Digitizer Software: Phase 2 418
Agilent I/O Libraries 419
To install the Agilent I/O libraries 419
Measurement Software Installation 424
To install the E5500 software 424
Asset Configuration 426
Setting Up Asset Manager 426 To set up Asset Manager 426 Configuring the Phase Noise Test Set 428 Configuring the PC Digitizer 432 Figure 324. Add assets 432 Figure 325. Choose asset type 433 Figure 326. Select supporting ACM 433 Figure 327. Choose the interface and address for the PC digitizer 434 Table 74. Default GPIB addresses 435 Figure 328. Choose model and serial number 436 Figure 329. Select (internal) in baseband source 436 Figure 330. Enter a comment about the configured asset 437 Figure 331. Asset manager screen showing configured PC Digitizer 437 Configuring the Agilent E4411A/B (ESA-L1500A) Swept Analyzer 438
License Key for the Phase Noise Test Set 439
Figure 332. Navigate to E5500 asset manager 439 Figure 333. Navigate to license keys 440 Figure 334. License_key.txt 441 Figure 33 5. Copy keyword into license key field 441 Figure 336. Licensing confirmation 442 Figure 337. Licensing error 442
20 PC Digitizer Performance Verification
Verifying PC Digitizer Card Output Performance 444
Required equipment 444 To verify the PC digitizer card inputs performance 444
PC Digitizer Card Input Performance Verification 449
Required equipment 449 To verify the PC digitizer card inputs performance 449
Agilent E5505A Users Guid e 21
21 Preventive Maintenance
Using, Inspecting, and Cleaning RF Connectors 454
Repeatability 454 RF Cable and Connector Care 454 Proper Connector Torque 455 Table 75. Proper Connector Torque 455 Connector Wear and Damage 455 SMA Connector Precautions 456 Cleaning Procedure 456 Table 76. Cleaning Supplies Available from Agilent 457
General Procedures and Techniques 458
Figure 338. GPIB, 3.5 mm, Type-N, power sensor, and BNC connectors 458 Connector Removal 459
Instrument Removal 461
Standard instrument 461 To remove an instrument from a rack 461 Half-Rack-Width Instrument 462 To remove a half-width instrument from a system rack 462 Figure 339. Instrument lock links, front and rear 463 Benchtop Instrument 463 To remove an instrument from a benchtop system 463
Instrument Installation 464
Standard rack instrument 464 To install an instrument 464 Half-Rack-Width instrument 465 To install the instrument in a rack 465 Benchtop instrument 465 To install an instrument in a benchtop system 465
A Service, Support, and Safety Information
Safety and Regulatory Information 468
Safety summary 468 Equipment Installation 468 Environmental conditions 469 Before applying power 469 Ground the instrument or system 470 Fuses and Circuit Breakers 470 Maintenance 471 Safety symbols and instrument markings 471 Table 77. Safety symbols and instrument markings 471
22 Agilent E5505A Users Guide
Declaration of Conformity 473 Compliance with German noise requirements 473 Table 78. German noise requirements summary 473 Compliance with Canadian EMC requirements 473
Service and Support 474
Agilent on the Web 474
Return Procedure 475
Determining your instruments serial number 475 Figure 340. Serial number location 475 Shipping the instrument 476 To package the instrument for shipping 476
Agilent E5505A Users Guid e 23
24 Agilent E5505A Users Guide
E5505A Phase Noise Measurement System Users Guide
1 Getting Started
Introduction 26 Documentation Map 27 Additional Documentation 28 System Overview 29
Agilent Technologies
25
1 Getting Started

Introduction

This guide introduces you to the Agilent E5505A Phase Noise Measurement System software and hardware. It provides procedures for configuring the E5500 Phase Noise Measurement software, executing measurements, evaluating results, and using the advanced software features. It also covers phase noise basics and measurement fundamentals to get you started.
Use Table 1 on page 27 as a guide to:
Learning about the E5505A phase noise measurement system
Learning about phase noise basics and measurement fundamentals
Using the E5505A system to make specific phase noise measurements.
In this guide you’ll also find information on system connections and specifications, and procedures for re-installing phase-noise-specific hardware and software in the system PC.
NOTE
Installation information for your system is provided in the Agilent E5505A Phase Noise Measurement System Installation Guide.
26 Agilent E5505A Users Guide

Documentation Map

Table 1 E5505A users guide map

Getting Started 1
Learning about the E5505A System Learning Phase Noise Basics &
Measurement Fundamentals
Chapter 1, Getting Started
Chapter 2, Introduction and
Measurement
Chapter 4, Expanding Your
Measurement Experience
Chapter 3, Phase Noise Basics
Chapter 5, “Absolute Measurement
Fundamentals
Chapter 7, Residual Measure ment
Fundamentals
Chapter 9, FM Discriminator
Fundamentals
Chapter 11, “AM Noise Mea surement
Fundamentals
Using the E5505A for Specific Phase Noise Measurements
Chapter 6, “Absolute Measurement
Examples
Chapter 8, Residual Measurement
Examples
Chapter 10, “FM Discriminator
Measurement Examples
Chapter 12, “AM Noise Measurement
Examples
Chapter 13, “Baseband Noise
Measurement Examples
Chapter 14, “Evaluating Your
Measurement Results
Chapter 15, “Advanced Software
Features
Chapter 17, “System Specifications”
Chapter 18, “System
Interconnections
Chapter 19, “PC Components
Installation
Chapter 20, “PC Digitizer Performance
Verification
Chapter 21, “Preventive
Maintenance
Chapter A, Service, Support, and
Safety Information
Chapter 16, “Reference Graphs and
Ta b l es
Agilent E5505A Users Guid e 27
1 Getting Started

Additional Documentation

You can access the complete set of PDF documents that support the E5505A system through the system GUI. (Adobe
Acrobat Reader is supplied.) Navigate the menu as shown in Figure 1. The files are stored on the system PC hard drive and on the E5500A software CD. Be sure to explore the E5500 Help menu for additional information.
The E5505A system documentation includes:
Agilent E5505A Phase Noise Measurement System Installation Guide
Agilent E5505A Phase Noise Measurement System User's Guide
Agilent N5501A/N5502A Phase Noise Downconverter User's Guide
Agilent N5507A Phase Noise Downconverter User's Guide
Agilent N5500A Phase Noise Test Set User's Guide
Agilent E5500 Series Phase Noise Measurement Systems SCPI Command
Reference
Agilent E5500 Phase Noise Measurement System Online Help
.
E5500_start_menu 04 Apr 04 rev 1

Figure 1 Navigate to system documentation

28 Agilent E5505A Users Guide

System Overview

The E5505A Phase Noise Measurement System provides flexible sets of measurements on one-port devices such as voltage controlled oscillators (VCOs), dielectric resonator oscillators (DROs), crystal oscillators, and synthesizers, and on two-port devices such as amplifiers and converters. The E5505A system measures absolute and residual phase noise, AM noise, and low-level spurious signals, as well as CW and pulsed signals. It operates in the frequency range of 50 KHz to 26.5 GHz.
The E5505A phase noise measurement system combines standard instruments, phase noise components, and PC software for maximum flexibility and re-use of assets. The system PC operates under Windows® XP Professional® and controls the system through the E5500 measurement software. The E5500 software enables many stand-alone instruments to work in the system. This standalone-instrument architecture easily configures for various measurement techniques, including the phase-lock-loop (PLL)/reference-source technique, and delay-line and FM-discriminator methods.
Getting Started 1
NOTE
The E5505A system is available as a one-bay wide, System II rack and as a benchtop model. Due to the systems flexibility, the hardware in the system varies greatly with the options selected. You may be installing instruments you already own in the system as well. A typical system includes these components:
Advantech custom PC with digitizer card assembly
15-inch display (flat-panel or standard), keyboard, and mouse
Windows® XP Professional® operating system
Agilent E5500 Phase Noise Measurement software
Phase noise test set
Downconverter
RF source
Additional instruments may include a spectrum analyzer, oscilloscope, RF counter, power meter, and power splitter.
For detailed information on the instruments in your E5505A phase noise measurement system, refer to the individual instrument user guides (provided on CD-ROM).
Agilent E5505A Users Guid e 29
1 Getting Started
Figure 2 shows a typical configuration of an E5505A benchtop system.

Figure 2 E5505A benchtop system, typical configuration

The E5505A replaces earlier Agilent E5500A/B series phase noise systems, which are based on MMS technology. The E5505A system uses GPIB communication and certain instruments have been redesigned with GPIB functionality. However, the E5505A system and E5500 software are backwards compatible with earlier systems and instruments, including the MMS mainframe. You may easily integrate existing assets into your E5505A system.
Table 2 shows the E5505A instruments and earlier-model equivalents.

Tab l e 2 Equivalent system/instrument model numbers

System or Instrument New Number Old Number
Phase noise measurement system E5505 A E5501A, E5501B, E5502B,
E5503A, E5503B, E5504A, E5504B
Test set E5500A 70420A
6.6 GHz downconverter N5501A 70421A
18 GHz downconverter N5502A 70422A
26.5 GHz downconverter N5507A 70427A, 71707A
Microwave source N5508A 70428A, 71708A
30 Agilent E5505A Users Guide
E5505A Phase Noise Measurement System Users Guide
2 Introduction and Measurement
Introducing the GUI 32 Designing to Meet Your Needs 34 E5505A Operation: A Guided Tour 35 Powering the System On 36 Performing a Confidence Test 39 Powering the System Off 45
Agilent Technologies
31
2 Introduction and Measurement

Introducing the GUI

The graphical user interface (GUI) gives the user instant access to all measurement functions, making it easy to configure a system and define or initiate measurements. The most frequently used functions are displayed as icons on a toolbar, allowing quick and easy access to the measurement information.
The forms-based graphical interaction helps you define your measurement quickly and easily. Each form tab is labeled with its content, preventing you from getting lost in the defining process.
The system provides three default segment tables. To obtain a quick look at your data, select the “fast quality level. If it is important to have more frequency resolution to separate spurious signals, use the “normal and high resolution quality levels. If you need to customize the offset range beyond the defaults provided, tailor the measurement segment tables to meet your needs and save them as a custom selection.
You can place up to nine markers on the data trace that can be plotted with the measured data.
Other features include:
Plotting data without spurs
Tabular listing of spurs
Plotting in alternate bandwidths
Parameter summary
Color printouts to any supported color printer
Figure 3 on page 33 shows an example of the GUI.
32 Agilent E5505A Users Guide
System Requirements
Introduction and Measurement 2
E5500_main_screen 24 Jun 04 rev 2

Figure 3 E5500 graphical user interface (GUI)

Agilent E5505A Users Guid e 33
2 Introduction and Measurement

Designing to Meet Your Needs

The E5505A Phase Noise Measurement System is a high performance measurement tool that enables you to fully evaluate the noise characteristics of your electronic instruments and components with unprecedented speed and ease. The phase noise measurement system provides you with the flexibility needed to meet todays broad range of noise measurement requirements.
In order to use the phase noise system effectively, it is important that you have a good understanding of the noise measurement you are making. This manual is designed to help you gain that understanding and quickly progress from a beginning user of the phase noise system to a proficient user of the system’s basic measurement capabilities.
NOTE

Beginning

If you have just received your system or need help with connecting the hardware or loading software, refer to your Agilent E5505A Phase Noise Measurement System Installation Guide now. Once you have completed the installation procedures, return to “E5505A
Operation: A Guided Tour" on page 35 to begin learning how to make noise measurements
with the system.
The section “E5505A Operation: A Guided Tour" on page 35 contains a step-by-step procedure for completing a phase noise measurement. This measurement demonstration introduces system operating fundamentals for whatever type of device you plan to measure.
Once you are familiar with the information in this chapter, you should be prepared to start Chapter 4, Expanding Your Measurement Experience. After you have completed that chapter, refer to Chapter 14, “Evaluating Your Measurement Results for help in analyzing and verifying your test results.
34 Agilent E5505A Users Guide

E5505A Operation: A Guided Tour

This measurement demonstration introduces you to the systems operat ion by guiding you through an actual phase noise measurement.
You will be measuring the phase noise of the Agilent N5500A Phase Noise Test Sets low noise amplifier. (The measurement made in this demonstration is the same measurement that is made to verify the system’s operation.)
As you step through the measurement procedures, you will soon discover that the phase noise measurement system offers enormous flexibility for measuring the noise characteristics of your signal sources and two-port devices.

Required equipment

The equipment shipped with this system is all that is required to complete this demonstration. (Refer to the E5505A Phase Noise Measurement System Installation Guide if you need information about setting up the hardware or installing the software.)
Introduction and Measurement 2

How to begin

NOTE
Follow the setup procedures beginning on the next page. The phase noise measurement system displays a setup diagram that shows you the front panel cable connections to make for this measurement.
If you need additional information about connecting instruments, refer to Chapter 18, System Interconnections.
Agilent E5505A Users Guid e 35
2 Introduction and Measurement

Powering the System On

This section provides procedures for powering on a racked or benchtop system. First connect your system to an appropriate AC power source, then follow the steps below.
WARNING
Before applying power, make sure the AC power input and the location of the system meet the requirements given in Ta b l e 4 on page 14. Failure to do so may result in damage to the system or personal injury.
NOTE
Warm-up Time: The downconverter and RF source instruments contain ovenized
oscillators which must warm up for 30 minutes to produce accurate measurements.
Standby Mode: The RF source uses a standby mode to keep the ovenized oscillator warm when the instrument is connected (plugged in) to AC power, even when the power switch is in the off position. To completely shut down the instrument, you must disconnect it from the AC power supply.

To power on a racked system

1 Press the system power switch (front, top right of the rack) to the on
position.
2 Verify that all instrument power switches are on.
3 Allow the system to warm up for 30 minutes.

To power on a benchtop system

1 Press the power switch on each instrument to the on position.
2 If you have the system connected to a safety power strip, turn the strips
power switch to the on position.
3 Allow the system to warm up for 30 minutes.
36 Agilent E5505A Users Guide

Starting the Measurement Software

1 Place the E5500 phase noise measurement software disk in the CD-ROM
drive.
2 Using Windows® Start menu as in Figure 4, navigate to the E5500 User
Interface.
Introduction and Measurement 2
E5500_start_menu 04 Apr 04 rev 1

Figure 4 Navigation to the E5500 user interface

3 The phase noise measurement subsystem main screen appears (Figure 5 on
page 38).
Agilent E5505A Users Guid e 37
2 Introduction and Measurement
E5500_main_screen 24 Jun 04 rev 2
NOTE

Figure 5 Phase noise measurement subsystem main screen

The default background for the screen is gray. You can change the background color by selecting View/Display Preferences and clicking on the Background Color button.
38 Agilent E5505A Users Guide

Performing a Confidence Test

This first measurement is a confidence test that functionally checks the N5500A test set’s filters and low-noise amplifiers using the test sets low noise amplifier. The phase detectors are not tested. This confidence test also confirms that the test set, PC, and analyzers are communicating with each other. To conduct the test, use a file with pre-stored parameters named Confidence.pnm.
1 On the E5500 GUI main menu, select File\Open.
2 If necessary, choose the drive or directory where the file you want is stored.
3 In the File Name box, select Confidence.pnm (Figure 6).
4 Click the Open button.
Introduction and Measurement 2
E5500_open_win_conf 04 Apr 04 rev 1

Figure 6 Opening the file containing pre-stored parameters

The appropriate measurement definition parameters for this example have been pre-stored in this file. Table 3 on page 44 lists the parameter data that has been entered for the N5500A confidence test example.
5 To view the parameter data in the software, navigate to the Define
Measurement window. Use Figure 7 on page 40 as a navigation guide. The parameter data is entered using the tabbed windows. Select various tabs to see the type of information entered behind each tab.
Agilent E5505A Users Guid e 39
2 Introduction and Measurement
e5505a_users_nav_define_wind 24 Jun 04 rev 3

Figure 7 Navigating to the Define Measurement window

6 Click the Close button.

Beginning a measurement

1 From the Measure menu, choose New Measurement. See Figure 8.
E5500_new_measurement 04 Apr 04 rev 1
Figure 8 Navigating to the New Measurement window
40 Agilent E5505A Users Guide
Introduction and Measurement 2
2 When the Do you want to Perform a New Calibration and Measurement?
dialog box appears, click Yes. See Figure 9.
E5500_new_cali_meas 04 Apr 04 rev 1
Figure 9 Confirm new measurement
3 When the Connect Diagram dialog box appears, connect the 50
termination, provided with your system, to the test sets noise input connector. (Refer to Figure 10 on page 41 for more information about the correct placement of the 50 termination.)
E5500_verify_con 23 Mar 04 rev 1
50 termination
Figure 10 Setup diagram displayed during the confidence test.
Agilent E5505A Users Guid e 41
2 Introduction and Measurement
N5500A Standard Test Set
N5500A TEST SET
SIGNAL
50 kHz-1600 MHz
RF ANALYZER
INPUT
50 W 1V Pk
0.01 Hz-100 MHz +15 dBm MIN
PHASE DET OUTPUT
MONITOR
GPIB
NOISE
STATUS
50 kHz-1600MHz
REF INPUT
50 Termination
ANALYZER ANALYZER
<100 kHz <100 MHz 50 W 20mA MAX
100 W/1.25 LPS
N5500A Opt. 001 Test Set
N5500A OPT 001 TEST SET
SIGNAL
50 kHz-26.5 GHz
RF ANALYZER
ANALYZER ANALYZER
<100 kHz <100 MHz 50 W 20mA MAX
N5500A Option 201Test Set
N5500A OPT 201 TEST SET
SIGNAL
50 kHz-1600 MHz
RF ANALYZER
ANALYZER ANALYZER
<100 kHz <100 MHz 50 W 20mA MAX
e5505a_user_connect_diag.ai rev2 10/20/03
INPUT
50 W 1V Pk
0.01 Hz-100 MHz +15 dBm MIN
PHASE DET OUTPUT
MONITOR
INPUT
50 W
1V Pk
0.01 Hz-100 MHz +15 dBm MIN
PHASE DET OUTPUT
MONITOR
GPIB
NOISE
100 W/1.25 LPS
GPIB
NOISE
100 W/1.25 LPS
TUNE VOLTAGE
OUT OF LOCK
STATUS
REF INPUT
50 kHz-1600MHz 1.2 - 26.5 GHz
+7 dBm MIN
FROM
DOWNCONVERTER
DOWNCONVERTER
TUNE VOLTAGE
OUT OF LOCK
STATUS
REF INPUT
50 kHz-1600MHz 1.2 - 26.5 GHz
+7 dBm MIN
mW SIGNAL
1.2-26.5 GHz
TUNE VOLTAGE
OUT OF LOCK
50 Termination
TO
50 Termination
Figure 11 Connect diagram example

Making a measurement

1 Press the Continue button.
Because you selected New Measurement to begin this measurement, the
system starts by running the routines required to calibrate the current measurement setup.
Figure 12 on page 43 shows a typical baseband phase noise plot for an
phase noise test set.
42 Agilent E5505A Users Guide
Introduction and Measurement 2
e5500_install_curve_sys_confidence_test rev2 10/10/03
Figure 12 Typical phase noise curve for test set confidence test

Sweep segments

When the system begins measuring noise, it places the noise graph on its display. As you watch the graph, you see the system plot its measurement results in frequency segments.
The system measures the noise level across its frequency offset range by averaging the noise within smaller frequency segments. This technique enables the system to optimize measurement speed while providing you with the measurement resolution needed for most test applications.

Congratulations

You have completed a phase noise measurement. This measurement of the test sets low noise amplifier provides a convenient way to verify that the system hardware and software are properly configured for making noise measurements. If your graph looks like that in Figure 12, you can be confident that your system is operating normally.
Agilent E5505A Users Guid e 43
2 Introduction and Measurement

Learning more

Continue with this demonstration by turning to Chapter 4, Expanding Your Measurement Experience to learn more about performing phase noise measurements.
Tab l e 3 Parameter data for the N5500A confidence test example
Step Parameters Data
1 Type and Range Tab
Measurement Type Start Frequency Stop Frequency Minimum Number of Averages FFT Quality Swept Quality
2 Cal Tab
Gain preceding noise input 0 dB
Baseband Noise (using a test set)
10 Hz
100 E + 6 Hz
1
4
Fa st
Fa st
3 Block Diagram Tab
Noise Source Test Set Noise Input
4 Tes t Se t Tab
Input Attenuation LNA Low Pass Filter LNA Gain DC Block PLL I ntegra tor Attenuation
5 Graph Tab
Title
Graph Type X Scale Minimum X Scale Maximum Y Scale Minimum Y Scale Maximum Normalize trace data to Scale trace data to a new carrier
frequency of: Shift trace data DOWN by Trace Smoothing Amount Power present at input of DUT
0 dB
20 MHz (Auto checked)
Auto Gain (Minimum Auto Gain –14 dB)
Not checked
0 dBm
Confidence Test, N5500A low noise amplifier.
Base band noise (dBv/Hz)
10 Hz
100 E + 6 Hz
0 dBv/Hz
• –200 dBv/Hz
1 Hz bandwidth
1 times the current carrier frequency
0 dB
0
0 dB
1 The Stop Frequency depends on the analyzers configured in your phase noise system.
44 Agilent E5505A Users Guide

Powering the System Off

To power off a racked system

1 On the E5500 software menu, select File\Exit. Always shut down the E5500
software before powering off the E5505A system.
2 Press the system power switch (front, top right of the rack) to the off
position.
Introduction and Measurement 2
CAUTION
Always shut down the E5500 software before powering off the E5505A system.
Failure to do so may produce errors in the stem, and result in an inoperable system or inaccurate measurements. If you do receive errors during shutdown, startup, or operation, use the E5500 Shutdown utility to restore functionality to the system.

To power off a benchtop system

1 On the E5500 software menu, select File\Exit.
2 Press the power switch on each instrument to the off position.

Using the E5500 Shutdown Utility

If you receive error messages during the power on or off procedures, or during operation, use the E5500 Shutdown utility to shut down the system. This utility automatically fixes most errors and restores functionality to the system. If you still receive errors after running the E5500 Shutdown utility, call your local Agilent Technologies Service Center.
To run the E5500 Shutdown utility
1 Double-Click on the E5500 Shutdown utility shortcut on the PC desktop
and follow the onscreen instructions. (You can also navigate to it using the menu path Start/Agilent Subsystems/E5500 Phase Noise/Shutdown.)
Figure 13 Shutdown utility icon
2 When the shutdown utility has finished, use the Start menu to shut down
the PC. Then power the system off.
Agilent E5505A Users Guid e 45
2 Introduction and Measurement
46 Agilent E5505A Users Guide
E5505A Phase Noise Measurement System Users Guide
3 Phase Noise Basics
What is Phase Noise? 48 Phase terms 49
Agilent Technologies
47
3 Phase Noise Basics
f

What is Phase Noise?

Frequency stability can be defined as the degree to which an oscillating source produces the same frequency throughout a specified period of time. Every RF and microwave source exhibits some amount of frequency instability. This stability can be broken down into two components:
long-term stability
short-term stability
Long-term stability describes the frequency variations that occur over long time periods, expressed in parts per million per hour, day, month, or year.
Short-term stability contains all elements causing frequency changes about the nominal frequency of less than a few seconds duration. The chapter deals with short-term stability.
Mathematically, an ideal sinewave can be described by
Vt() V
Where = nominal amplitude,
But an actual signal is better modeled by
V
o
V
and = nominal frequency
2π fotsin
o
o
2πfotsin=
o
= linearly growing phase component,
Vt() Vo ε t()+2πfot ∆φ t()+sin=
Where = amplitude fluctuations,
and = randomly fluctuating phase term or phase noise.
This randomly fluctuating phase term could be observed on an ideal RF analyzer (one which has no sideband noise of its own) as in Figure 14 on page 49.
ε t()
∆φ t()
48 Agilent E5505A Users Guide

Figure 14 RF sideband spectrum

f()

Phase terms

There are two types of fluctuating phase terms:
e5505a_user_RF_sideband.ai rev2 10/20/03
Phase Noise Basics 3
spurious signals
phase noise
Spurious signals
The first are discrete signals appearing as distinct components in the spectral density plot. These signals, commonly called spurious, can be related to known phenomena in the signal source such as power line frequency, vibration frequencies, or mixer products.
Phase noise
The second type of phase instability is random in nature, and is commonly called phase noise. The sources of random sideband noise in an oscillator include thermal noise, shot noise, and flicker noise.
Many terms exist to quantify the characteristic randomness of phase noise. Essentially, all methods measure the frequency or phase deviation of the source under test in the frequency or time domain. Since frequency and phase are related to each other, all of these terms are also related.
Spectral density One fundamental description of phase instability or phase
noise is spectral density of phase fluctuations on a per-Hertz basis. The term spectral density describes the energy distribution as a continuous function, expressed in units of variance per unit bandwidth. Thus (Figure 15 on
φ
page 50) may be considered as:
Agilent E5505A Users Guid e 49
3 Phase Noise Basics
S
f()
Sφ f()
------ --- ------ ------- --- ------ --- ------ --- ------ ------ --- ------ --- -------
BW used to measure ∆φ
∆φ
rms
f()
rms
2
Where BW (bandwidth is negligible with respect to any changes in versus
2
rad
------ --- ---== Hz
φ
the fourier frequency or offset frequency (f).
L(f) Another useful measure of noise energy is L(f), which is then directly
related to by a simple approximation which has generally negligible error
φ
if the modulation sidebands are such that the total phase deviation are much less than 1 radian (
∆φ
pk
<< radian).
L f()
1
-- -
S
f()=
∆φ
2
e5505a_user_CW_sidebands_freq.ai rev2 10/20/03
Figure 15 CW signal sidebands viewed in the frequency domain
L(f) is an indirect measurement of noise energy easily related to the RF power spectrum observed on an RF analyzer. Figure 16 shows that the National Institute of Standards and Technology (NIST) defines L(f) as the ratio of the power--at an offset (
f) Hertz away from the carrier. The phase modulation
sideband is based on a per Hertz of bandwidth spectral density and or offset frequency in one phase modulation sideband, on a per Hertz of bandwidth spectral density and
L f()
power density in one phase modulation sideband()
------ --------- -------- --------- --------- --------- -------- --------- --------- --------- --------- -------- --------- --------- ------------ --- -
(f) equals the Fourier frequency or offset frequency.
P
ssb
total signal power
------ ----- -== P
s
= single sideband (SSB) phase noise to carrier ration (per Hertz)
50 Agilent E5505A Users Guide
Phase Noise Basics 3
e5505a_user_derivingL_RF_display.ai
rev2 10/20/03
Figure 16 Deriving L(f) from a RF analyzer display
is usually presented logarithmically as a spectral density plot of the
L f()
phase modulation sidebands in the frequency domain, expressed in dB relative to the carrier per Hz (dBc/Hz) as shown in Figure 17. This chapter, except where noted otherwise, uses the logarithmic form of as follows:
S∆ff() 2f2L f()=
.
L f()
e5505a_user_log_offset_freq.ai
rev2 10/20/03
Figure 17 L(f) Described Logarithmically as a Function of Offset Frequency
Agilent E5505A Users Guid e 51
3 Phase Noise Basics
Caution must be exercised when is calculated from the spectral density of the phase f luctuations because the calculation of is dependent on
Sφf() L f()
L f()
the small angle criterion. Figure 18, the measured phase noise of a free running VCO described in units of , illustrates the erroneous results that
L f()
can occur if the instantaneous phase modulation exceeds a small angle line. Approaching the carrier obviously increases in error as it indicates a
L f()
relative level of +45 dBc/Hz at a 1 Hz offset (45 dB more noise power at a 1 Hz offset in a 1 Hz bandwidth than in the total power of the signal); which is of course invalid.
Figure 18 shows a 10 dB/decade line drawn over the plot, indicating a peak
phase deviation of 0.2 radians integrated over any one decade of offset frequency. At approximately 0.2 radians the power in the higher order sidebands of the phase modulation is still insignificant compared to the power in the first order sideband which insures that the calculation of remains valid. Above the line the plot of becomes increasingly invalid, and
L f() Sφf()
L f()
must be used to represent the phase noise of the signal.
e5505a_user_region_validity_L.ai
rev2 10/20/03
Figure 18 Region of validity of L(f)
52 Agilent E5505A Users Guide
E5505A Phase Noise Measurement System Users Guide
4 Expanding Your Measurement Experience
Starting the Measurement Software 54 Using the Asset Manager 55 Using the Server Hardware Connections to Specify the Source 60 Testing the 8663A Internal/External 10 MHz 66 Testing the 8644B Internal/External 10 MHz 81 Viewing Ma rkers 96 Omitting Spurs 97 Displaying the Parameter Summary 99 Exporting Measurement Results 101
Agilent Technologies
53
4 Expanding Your Measurement Experience

Starting the Measurement Software

1 Make sure your computer and monitor are turned on.
2 Place the E5500 Phase Noise Measurement System software disk in the disc
holder and insert in the CD-ROM drive.
3 Using Figure 19 as a guide, navigate to the E5500 User Interface.
E5500_start_menu 04 Apr 04 rev 1

Figure 19 Navigate to E5500 user interface

54 Agilent E5505A Users Guide

Using the Asset Manager

Use the Asset Manager to add assets to your E5505A system. The process is essentially the same for any asset, including reference sources. In fact, the procedure in this section uses an Agilent 8663 source as an example. (The procedure applies to all Agilent sources, including the 8257x series.)
Adding an asset involves two steps once the hardware connections have been made:
Configuring the asset
Verifying the server hardware connections.
Expanding Your Measurement Experience 4
WARNING

Configuring an asset

Be sure to power off the system before making all hardware connections other than GPIB. Connecting assets with the power applied can result in personal injury and damage to the hardware. (For more information on connecting assets, see Chapter 18, System Interconnections)
This procedure demonstrates how to add an asset to the E5505A system.
1 Using Figure 20 as a guide, navigate to Asset Manager. For this example we
invoke the Asset Manager Wizard from the E5500 main screen. This is the most common way to add assets.
E5500_asset_manager 23 Mar 04 rev 1
Figure 20 Navigate to Asset Manager
Agilent E5505A Users Guid e 55
4 Expanding Your Measurement Experience
2 Select Add in the Asset Manager window. See Figure 21.
E5500_add_source2 16 Apr 04 rev 1
Figure 21 Navigate to Add in Asset Manager
3 From the Asset Type pull-down list in Choose Asset Role dialog box, select
Source, then click Next. See Figure 22.
.
E5500_add_source3 16 Apr 04 rev 1
Figure 22 Select source as asset type
56 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
4 Click on the source to be added, then click the Next button (see Figure 23).
E5500_add_source4 16 Apr 04 rev 1
Figure 23 Choose source
NOTE
5 From the Interface pull-down list, select GPIB0. See Figure 24.
6 In the Address box, type 19.
19 is the default address for a source. Ta b le 4 on page 63 shows the default GPIB address for all system instruments.
7 In the Library pull-down list, select the Agilent Technologies VISA. Click
the Next button.
E5500_add_source5 16 Apr 04 rev 1
Figure 24 Select I/O library
Agilent E5505A Users Guid e 57
4 Expanding Your Measurement Experience
8 In the Set Model & Serial Numbers dialog box, type in your source name
and its corresponding serial number. Click the Next button. See Figure 25.
E5500_add_source6 16 Apr 04 rev 1
Figure 25 Enter asset and serial number
9 In the Enter A Comment dialog box, you may type a comment that
associates itself with the asset you have just configured. Click Finish. See
Figure 26.
E5500_add_source7 16 Apr 04 rev 1
Figure 26 Enter comment
58 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
10 In the Asset Manager window, select the source in the left window pane.
Click the check-mark button on the toolbar to verify connectivity. See
Figure 27.
E5500_add_source8 16 Apr 04 rev 1
Figure 27 Click check-mark button
The Asset Manager displays a message verifying the connection to your
asset. This indicates that you have successfully configured a source. (See
Figure 28.)
E5500_add_source9 16 Apr 04 rev 1
Figure 28 Confirmation message
11 To exit the Asset Manager, on the menu select Server/Exit.
12 Perform the procedure Using the Server Hardware Connec tions to Specify
the Source" on page 60.
Agilent E5505A Users Guid e 59
4 Expanding Your Measurement Experience

Using the Server Hardware Connections to Specify the Source

1 From the System menu, choose Server Hardware Connections. See
Figure 29.
E5500_add_source10 16 Apr 04 rev 1

Figure 29 Navigate to server hardware connections

2 Select the Sources tab shown in Figure 30.
E5500_add_source11 16 Apr 04 rev 1

Figure 30 Select Sources tab

60 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
3 From the Reference Source pull-down list, select Agilent 8663A.
a A green check-mark appears after an automatic I/O check has been
successfully performed by the software. If nothing happens, click the Check I/O button to manually initiate the check.
E5500_add_source12 16 Apr 04 rev 1

Figure 31 Successful I/O check

b A red circle with a slash appears if the I/O check is unsuccessful.
E5500_add_source13 16 Apr 04 rev 1

Figure 32 Failed I/O check

c If the I/O check fails, click the Asset Manager button to return to the
Asset Manager (see Figure 32).
Agilent E5505A Users Guid e 61
4 Expanding Your Measurement Experience
d In the Asset Manager, verify that the 8663A is configured correctly. Do
the following:
Check your system hardware connections.
Click the green check-mark button on the Asset Managers toolbar to
verify connectivity.
Return to Server Hardware Connections and click the Check I/O
button to re-check it.
NOTE
Use the same process to add additional assets to your E5505A system.
62 Agilent E5505A Users Guide

Setting GPIB Addresses

Table 4 shows the default GPIB addresses for the E5505A system instruments.
If you need to change a GPIB address to prevent a conflict between assets, use the Asset Manager as shown in the easy procedure starting on page 64.

Tab l e 4 Default GPIB addresses

Instrument Address
Te s t se t 2 0
Downconverter 28
Microwave downconverter 28
RF analyzer 17
FFT analyzer (PC digitizer card) 1
FFT analyzer (89410A) 18
Expanding Your Measurement Experience 4
Source # 1 19
Source # 2 23
Counter 3
Agilent E1430 VXI digitizer
Agilent E1437 VXI digitizer
Agilent E1420B VXI counter
Agilent E1441 VXI ARB
1 The E5500 software supports this instrument although it is not part of the standard E5505A system.
1
1
1
1
129
192
48
80
Agilent E5505A Users Guid e 63
4 Expanding Your Measurement Experience

To change the GPIB address

1 On the E5500 main menu, select System/Asset Manager. See Figure 33.
E5500_asset_manager 23 Mar 04 rev 1

Figure 33 Asset Manager on System menu

2 Double-Click on the desired instrument in the Asset Manager list (left
pane). See Figure 34.
E5500_asset_mgr_screen 23 Mar 04 rev 1

Figure 34 Asset Manager window

64 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
3 Type the desired address in the dialog box. See Figure 35.
E5500_edit_asset_info 04 Apr 04 rev 1

Figure 35 GPIB address dialog box

4 Click OK.
5 To exit the Asset Manager, on the menu select Server/Exit.
Next proceed to one of the following absolute measurements using either an Agilent 8257x or an Agilent 8644B source:
Testing the 8663A Internal/External 10 MHz" on page 66.
Testing the 8644B Internal/External 10 MHz" on page 81.
Agilent E5505A Users Guid e 65
4 Expanding Your Measurement Experience

Testing the 8663A Internal/External 10 MHz

This measurement example helps you measure the absolute phase noise of an RF synthesizer.
CAUTION
To prevent damage to the test sets hardware components, do not apply the input signal to the signal input connector until the input attenuator has been correctly set for the desired configuration, as shown in Tabl e 6 on page 73. Apply the input signal when the connection diagram appears.

Required equipment

This measurement requires an Agilent 8663A in addition to the phase noise test system and your device under test (DUT). and adapters necessary to connect the DUT and reference source to the test set.
NOTE
To ensure accurate measurements allow the DUT and measurement equipment to warm up at least 30 minutes before making the noise measurement.

Defining the measurement

1 From the File menu in the E5500 User Interface, choose Open . If necessary,
choose the drive or directory where the file you want is stored.
2 In the File Name box, choose Conf_SigGen_10MHz.pnm. See Figure 36.
You also need the coaxial cables
.
E5500_open_win_conf 04 Apr 04 rev 1
Figure 36 Select the parameters definition file
66 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
3 Click the Open button.
The appropriate measurement definition parameters for this example have been pre-stored in this file. Table 7 on page 79 lists the parameter data that has been entered for this measurement example.)
NOTE
Note that the source parameters entered for step 2 in Tab le 7 on page 79 may not be appropriate for the reference source you are using. To change these values, refer to Ta b l e 5 on page 68, then continue with step 4 below. Otherwise, go to “Beginning the
measurement" on page 71.
4 Using Figure 37 as a guide, navigate to the Sources tab.
a Enter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz).
Enter the same frequency for the detector input frequency.
b Enter the VCO (Nominal) Tuning Constant (see Table 5 on page 68).
c Enter the Tune Range of VCO (see Table 5).
d Enter the Center Voltage of VCO (see Table 5).
e Enter the Input Resistance of VCO (see Table 5).
e5505a_user_enter_source_info
24 Jun 04 rev 3
Figure 37 Enter Source Information
Agilent E5505A Users Guid e 67
4 Expanding Your Measurement Experience
Table 5 Tuning characteristics for various sources
VCO Source Carrier
Freq.
Tuning Constant (Hz/V)
Center Voltage (V)
Voltage Tuning Range (± V)
Input Resistance (Ω)
Tuning Calibration Method
Agilent 8662/3A
EFC DCFM
υ
0
5 E – 9 x υ
FM Deviation
0
0 0
10 10
1E + 6 1 K (8662) 600 (8663)
Measure Com pute Com pute
Agilent 8642A/B FM Deviation 0 10 600 Compute
Agilent 8644B FM Deviation 0 10 600 Compute
Other Signal Generator DCFM Calibrated for ±1V FM Deviation 0 10 R
Other User VCO Source Estimated within a
factor of 2
–10 to +10
in
1 E + 6 Measure
Com pute

Selecting a reference source

1 Using Figure 38 as a guide, navigate to the Block Diagram tab.
2 From the Reference Source pull-down list, select Agilent-8663.
3 When you have completed these operations, click the Close button.
Agilent-8257
e5505a_user_select_ref_source 24 Jun 04 rev 3
Figure 38 Selecting a reference source
68 Agilent E5505A Users Guide

Selecting loop suppression verification

1 Using Figure 39 as a guide, navigate to the Cal tab.
2 Check Verify calculated phase locked loop suppression and Always Show
Suppression Graph. Select If limit is exceeded: Show Loop Suppression Graph.
3 When you have completed these operations, click the Close button.
Expanding Your Measurement Experience 4
e5505a_user_select_loop 24 Jun 04 rev 3
Figure 39 Selecting loop suppression verification

Setting up for the 8663A 10 MHz measurement

The signal amplitude at the test sets R input (Signal Input) port sets the measurement noise floor level. Use Figure 40 on page 70 and Figure 41 on page 70 to help determine the amplitude required to provide a noise f loor level that is below the expected noise floor of your DUT. For more information about this graph, refer to “Reference Graphs and Tables" on page 359.
Agilent E5505A Users Guid e 69
4 Expanding Your Measurement Experience
+15
+5
-5
R Port signal level (dBm)
-15
-140 -160 -170 -180-150 Expected phase noise floor of system (dBc/Hz)
n5505a_exp_phase_noise 25 Feb 04 rev 1
Figure 40 Noise floor for the 8663 10 MHz measurement
L Port level +15dBm
f 10kHz
e5505a_user_noise_floor_ex 24 Jun 04 rev 3
Figure 41 Noise floor example
70 Agilent E5505A Users Guide
If the output amplitude of your DUT is not sufficient to provide an adequate measurement noise floor, it is necessary to insert a low-noise amplifier between the DUT and the test set. Refer to “Inserting a Device" on page 122 for details on determining the effect the amplifiers noise will have on the measured noise floor.

Beginning the measurement

1 From the Measurement menu, choose New Measurement. See Figure 42.
Expanding Your Measurement Experience 4
E5500_new_measurement 04 Apr 04 rev 1
Figure 42 Selecting new measurement
2 When the Do you want to perform a New Calibration and Measurement?
prompt appears, click Ye s.
E5500_new_cali_meas
Figure 43 Confirm new measurement
3 When the Connect Diagram dialog box appears, click on the hardware
pull-down arrow and select your hardware configuration from the list. See
Figure 44 on page 72.
Agilent E5505A Users Guid e 71
4 Expanding Your Measurement Experience
CAUTION
E5500_verify_con 23 Mar 04 rev 1
Figure 44 Connection diagram
4 Connect your DUT and reference sources to the test set at this time and
confirm your connections as shown in the appropriate connect diagram.
The input attenuator (Option 001 only) is now correctly configured
based on your measurement definition.
The test sets signal input is subject to the limits and characteristics contained in
Ta b l e 6 on page 73.
To prevent damage to the test sets hardware components, do not apply the input signal to the test set’s signal input connector until the input attenuator (Option 001) has been set by the phase noise software, which occurs at the connection diagram.
5 Press Continue.
As the system performs the calibration routines, various status messages
appear on the display. When the last message—Measuring PLL suppressionappears, you can choose to continue the routine or stop it.
6 Press Continue using Adjusted Loop Suppression to continue making the
noise measurement, or press Abort to stop the measurement. (For descriptions of the messages, see “Status messages" on page 74. You have time to read through the descriptions while the system completes the routines.)
72 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
:
Tab l e 6 Test set signal input limits and characteristics
Limits
Frequency 50 kHz to 26.5 GHz
Maximum Signal Input Power +30 dBm
At Attenuator Output, Operating Level Range:
RF Phase Detectors
Microwave Phase Detectors
Internal AM Detector
Downconverters:
N5502A/70422A
N5507A/70427A
Characteristics
Input Impedance 50 nominal
AM Noise DC coupled to 50 Ω load
0 to +23 dBm 0 to +5 dBm 0 to +20 dBm
0 to +30 dBm +5 to +15 dBm
NOTE
Refer to the following system connect diagram examples in Chapter 18, System Interconnections for more information about system interconnections.
Figure 307, E5505A system connections with standard test set, on page 401
Figure 308, E5505A system connections with test set option 001, on page 402
Figure 309, E5505A system connections with test set option 201, on page 403
Agilent E5505A Users Guid e 73
4 Expanding Your Measurement Experience
Status messages
This section describes the status messages that appear on the display as the system performs its calibration routines.
Determining Presence of Beat Note... An initial check is made to verify that a
beatnote is present within the systems detection range.
Verifying Zero-Beat... The frequency of the beatnote is measured to see if it is
within 5% of the estimated Peak Tuning Range of the system. The system’s Peak Tuning Range is the portion of the voltage-controlled-oscillator (VCO) sources tuning range being used for the measurement.
When the system measures the phase noise of a signal source using the Phase Lock Loop technique (the technique being used in this example) it requires that one of the two sources used in the setup is a VCO. As you will see later in this demonstration, you are required to estimate the tuning range of the VCO source you are using when you set up your own Phase Lock Loop measurements.
Zero beating sources... The center frequencies of the sources are now
adjusted, if necessary, to position the beatnote within the 5% range. The adjustment is made with the tune voltage applied to the VCO source set at its nominal or center position.
Measuring the VCO Tuning Constant... The tuning sensitivity (Hz/V) of the VCO
source is now precisely determined by measuring the beatnote frequency at four tune voltage settings across the tuning range of the VCO source. Linearity across the tuning range is also verified
Measuring the Phase Detector Constant... The transfer characteristics (V/rad)
of the test sets phase detector are now determined for the specific center frequency and power level of the sources being measured.
Measuring PLL suppression... The required correction data is created to
compensate for the phase noise suppression which occurs within the bandwidth of the phase lock loop created for this measurement. The computer displays the PLL suppression curve and associated measurement values.
74 Agilent E5505A Users Guide

Sweep segments

When the system begins measuring noise, it places the noise graph on its display. As you watch the graph, you see the system plot its measurement results in frequency segments.
The system measures the noise level across its frequency offset range by averaging the noise within smaller frequency segments. This technique enables the system to optimize measurement speed while providing you with the measurement resolution needed for most test applications.
When the measurement is complete, refer to Chapter 14, Evaluating Your Measurement Results for help in evaluating your measurement results. (If the test system has problems completing the measurement, it informs you by placing a message on the computer display.
Checking the beatnote
While the Connect Diagram is still displayed, Agilent recommends that you use an oscilloscope (connected to the Monitor port on the test set) or a counter to check the beatnote being created between the reference source and your DUT. The objective of checking the beatnote is to ensure that the center frequencies of the two sources are close enough in frequency to create a beatnote that is within the Capture Range of the system.
Expanding Your Measurement Experience 4
NOTE
The phase lock loop (PLL) Capture Range is 5% of the peak tuning range of the VCO source you are using. (The peak tuning range for your VCO can be estimated by multiplying the VCO tuning constant by the tune range of VCO. Refer to Chapter 14, Evaluating Your Measurement Results if you are not familiar with the relationship between the PLL capture range and the peak tuning range of the VCO.)
If the center frequencies of the sources are not close enough to create a beatnote within the capture range, the system is not able to complete its measurement.
The beatnote frequency is set by the relative frequency difference between the two sources. If you have two very accurate sources set at the same frequency, the resulting beatnote is very close to 0 Hz.
Searching for the beatnote requires that you adjust the center frequency of one of the sources above and below the frequency of the other source until the beatnote appears on the oscilloscopes display. See Figure 45 on page 76.
If incrementing the frequency of one of the sources does not produce a beatnote, you need to verify the presence of an output signal from each source before proceeding.
Agilent E5505A Users Guid e 75
4 Expanding Your Measurement Experience
0
V
E5505a_oscillo_disp_beatnote 25 Feb 04 rev 1
Figure 45 Oscilloscope display of beatnote from test set monitor port

Making the measurement

1 Click the Continue button when you have completed the beatnote check
and are ready to make the measurement.
2 When the PLL Suppression Curve dialog box appears, check View Measured
Loop Suppression, View Smoothed Loop Suppression, and View Adjusted Loop Suppression in the lower right of the dialog box. See Figure 46 on page 77.
-1V/div
76 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
e5505a_user_select_suppression.ai rev2 10/21/03
Figure 46 Selecting suppression
There are four different curves for the this graph. (For more information about loop suppression verification, refer to Chapter 15, Advanced Software Featu r e s ).
“Measured loop suppression curvethis is the result of the loop
suppression measurement performed by the E5505A system.
“Smoothed measured suppression curvethis is a curve-fit representation
of the measured results, it is used to compare with the “theoretical” loop suppression.
“Theoretical suppression curve—this is the predicted loop suppression
based on the initial loop parameters defined/selected for this particular measurement (kphi, kvco, loop bandwidth, filters, gain, and others).
“Adjusted theoretical suppression curvethis is the new “adjusted
theoretical value of suppression for this measurement. It is based on changing loop parameters (in the theoretical response) to match the smoothed measured curve as closely as possible.
When the measurement is complete, refer to Chapter 14, Evaluating Your Measurement Results
for help in evaluating your measurement results.
Figure 47 on page 78 shows a typical phase noise curve for an RF Synthesizer.
Agilent E5505A Users Guid e 77
4 Expanding Your Measurement Experience
e5505a_user_typ_noise_curve_8663a 24 Jun 04 rev 3
Figure 47 Typical phase noise curve for an 8663A 10 MHz measurement
Table 7 on page 79 contains the data stored in the parameter definitions file.
78 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
Tab l e 7 Parameter data for the 8663A 10 MHz measurement
Step Parameters Data
1 Type and Range Tab
Measurement Type
Absolute Phase Noise (with phase locked loop)
10 Hz
Start Frequency
Stop Frequency
Minimum Number of Averages
2 E + 6 Hz
4
Fast
1
FFT Quality
2Sources Tab
Carrier Source
Frequency
Power
Carrier Source Output connected to
Detector Input Frequency
Reference Source Frequency
Reference Source Power
10 E + 6 Hz
7 dBm
Te st Se t
10 E +6 Hz
10 E +6 Hz (same as Carrier Source Frequency)
16 dBm
VCO Tuning Parameters:
Nominal Tune Constant
Tune Range ±
Center Voltage
Input Resistance
1 E +3 Hz/V
± 10 Volts
0 Volts
600 ohms
3Cal Tab
Phase Detector Constant
VCO Tune Constant
Phase Lock Loop Suppression
If Limit is exceeded
Measure Phase Detector Constant
Calculate from expected VCO Tune Constant
Verify calculated phase locked loop
suppression
Show Suppression Graph
4Block Diagram Tab
Carrier Source
Downconverter
Reference Source
Timebas e
Phase Detector
Test Set Tune Voltage Destination
VCO Tune Mode
5 Test Set Tab
Input Attenuation
LNA Low Pass Filter
LNA Gain
DC Block
PLL Integrator Attenuation
6 Dowconverter Tab The downconverter parameters do not apply to
Manual
None
Agilent 8663A
None
Automatic Detector Selection
Reference Source
DCFM
0 dB
20 MHz (Auto checked)
Auto Gain (Minimum Auto Gain –14 dB)
Not checked
0 dBm
this measurement example.
Agilent E5505A Users Guid e 79
4 Expanding Your Measurement Experience
Tab l e 7 Parameter data for the 8663A 10 MHz measurement (continued)
Step Parameters Data
7Graph Tab
Title
Graph Type
X Scale Minimum
X Scale Maximum
Y Scale Minimum
Y Scale Maximum
Normalize trace data to a:
Scale trace data to a new carrier
frequency of:
Shift trace data DOWN by
Trace Smoothing Amount
Power present at input of DUT
1 The Stop Frequency depends on the analyzers configured in your phase noise system.
Confidence Test using Agilent 8663A Int vs Ext
10 MHz
Single-sideband Noise (dBc/Hz)
10 Hz
4 E + 6 Hz
0 dBc/Hz
• –170 dBc/Hz
1 Hz bandwidth
1 times the current carrier frequency
0 dB
0
0 dB
80 Agilent E5505A Users Guide

Testing the 8644B Internal/External 10 MHz

This measurement example helps you measure the absolute phase noise of an RF synthesizer.
Expanding Your Measurement Experience 4
CAUTION
To prevent damage to the test sets hardware components, do not apply the input signal to the signal input connector until the input attenuator has been correctly set for the desired configuration, as shown in Ta b l e 8 on page 83. Apply the input signal when the connection diagram appears.
NOTE
To ensure accurate measurements allow the DUT and measurement equipment to warm up at least 30 minutes before making the noise measurement.

Required equipment

This measurement requires an Agilent 8644B in addition to the phase noise test system and your DUT. necessary to connect the DUT and reference source to the test set.

Defining the measurement

1 From the file menu of the E5505A User Interface, choose Open. If
necessary, choose the drive or directory where the file you want is stored.
2 In the File Name box, choose Conf_8644B_10MHz.pnm.” See Figure 42.
You also need the coaxial cables and adapters
E5500_open_win_conf 04 Apr 04 rev 1
Figure 48 Select the parameters definition file
Agilent E5505A Users Guid e 81
4 Expanding Your Measurement Experience
3 Click the Open button.
The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table 10 on page 94 lists the parameter data that has been entered for the RF Synthesizer using a DCFM measurement example.
NOTE
The source parameters shown in Ta b l e 1 0 on page 94 may not be appropriate for the reference source you are using. To change these values, refer toTab l e 8, Tuning characteristics for various sources,” on page 83, then continue with step 4 below. Otherwise, go to “Beginning the measurement" on page 87.
4 Using Figure 49 on page 82 as a guide, navigate to the Sources tab
a Enter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz);
enter the same frequency for the detector input frequency.
b Enter the VCO (Nominal) Tuning Constant (see Table 8 on page 83).
c Enter the Tune Range of VCO (see Table 8 on page 83).
d Enter the Center Voltage of VCO (see Table 8 on page 83).
e Enter the Input Resistance of VCO (see Table 8 on page 83)
e5505a_user_enter_source_info
24 Jun 04 rev 3
Figure 49 Sources tab in define measurement window
82 Agilent E5505A Users Guide
Table 8 Tuning characteristics for various sources
Expanding Your Measurement Experience 4
VCO Source Carrier
Freq.
Tuning Constant (Hz/V)
Center Voltage (V)
Voltage Tuning Range (± V)
Input Resistance (Ω)
Tuning Calibration Method
Agilent 8662/3A
EFC DCFM
υ
0
5 E – 9 x υ
FM Deviation
0
0 0
10 10
1E + 6 1 K (8662) 600 (8663)
Measure Com pute Com pute
Agilent 8642A/B FM Deviation 0 10 600 Compute
Agilent 8644B FM Deviation 0 10 600 Compute
Other Signal Generator
DCFM Calibrated for
FM Deviation 0 10 R
in
Com pute
±1V
Other User VCO Source Estimated within a
factor of 2
–10 to +10
1 E + 6 Measure

Selecting a reference source

1 From the Define menu, choose Measurement; then choose the Block
Diagram tab from the Define Measurement window. See Figure 50.
2 From the Reference Source pull-down list, select 8644.
Agilent E5505A Users Guid e 83
4 Expanding Your Measurement Experience
Agilent-8644
e5505_user_select_ref_source8644 24 Jun 04 rev 3
Figure 50 Selecting a reference source
3 When you have completed these operations, click the Close button.

Selecting loop suppression verification

1 From the Define menu, choose Measurement; then choose the Cal tab from
the Define Measurement window.
2 In the Cal dialog box, check Verify calculated phase locked loop
suppression and Always Show Suppression Graph. Select If limit is exceeded: Show Loop Suppression Graph. See Figure 51 on page 85.
3 When you have completed these operations, click the Close button.
84 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
e5505a_user_select_loop 24 Jun 04 rev 3
Figure 51 Selecting loop suppression verification

Setting up the 8663A 10 MHz measurement

The signal amplitude at the R input (Signal Input) port on the test set sets the measurement noise floor level. Use the graph in Figure 52 and the example in
Figure 53 on page 86 to determine the amplitude. For more information, refer
to Chapter 16, Reference Graphs and Tables.
+15
+5
-5
R Port signal level (dBm)
-15
-140 -160 -170 -180-150 Expected phase noise floor of system (dBc/Hz)
n5505a_exp_phase_noise 25 Feb 04 rev 1
Figure 52 Noise floor for the 8644B 10 MHz measurement
L Port level +15dBm
f 10kHz
Agilent E5505A Users Guid e 85
4 Expanding Your Measurement Experience
e5505a_user_noise_floor_ex 24 Jun 04 rev 3
Figure 53 Noise floor example
If the output amplitude of your DUT is not sufficient to provide an adequate measurement noise floor, it is necessary to insert a low-noise amplifier between the DUT and the test set. Refer to the section “Inserting a Devic e" on page 122 for details on determining the effect the amplifiers noise wi ll have on the measured noise floor.
86 Agilent E5505A Users Guide

Beginning the measurement

Expanding Your Measurement Experience 4
CAUTION
To prevent damage to the test sets hardware components, do not apply the input signal to the signal input connector until the input attenuator has been correctly set for the desired configuration, as shown in Ta bl e 9 on page 89. Apply the input signals when the connection diagram appears, as in step 3 below.
1 From the Measurement menu, choose New Measurement. See Figure 54.
.
E5500_new_measurement 04 Apr 04 rev 1
Figure 54 Selecting a new measurement
2 When the Do you want to perform a New Calibration and Measurement?
prompt appears, click Ye s.
E5500_new_cali_meas 04 Apr 04 rev 1
Figure 55 Confirm measurement dialog box
3 When the Connect Diagram dialog box appears, click on the hardware
drop-down arrow and select your hardware configuration from the list. See
Figure 56 on page 88.
Agilent E5505A Users Guid e 87
4 Expanding Your Measurement Experience
e5505a_conn_diag_dialog rev 1 23 jun 04
Figure 56 Connect diagram dialog box
CAUTION
4 Connect your DUT and reference sources to the test set at this time.
Confirm your connections as shown in the Connect Diagram (Figure 56).
The input attenuator (Option 001 only) has now been correctly
configured based on your measurement definition.
The test sets signal input is subject to the limits and characteristics in Ta b l e 9 on page 89.
To prevent damage to the test sets hardware components, do not apply the input signal to the test set’s signal input connector until the input attenuator (Option 001) has been set by the phase noise software, which occurs when the connection diagram appears.
5 Press Continue.
As the system performs the calibration routines, various status messages
appear on the display. When the last message—Measuring PLL suppressionappears, you can choose to continue the routine or stop it.
6 Press Continue using Adjusted Loop Suppression to continue making the
noise measurement, or press Abort to stop the measurement. (For descriptions of the messages, see “Status messages" on page 74. You have time to read through the descriptions while the system completes the routines.)
88 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
Tab l e 9 Test set signal input limits and characteristics
Limits
Frequency 50 kHz to 26.5 GHz
Maximum Signal Input Power +30 dBm
At Attenuator Output, Operating Level Range:
RF Phase Detectors
Microwave Phase Detectors
Internal AM Detector
Downconverters:
Agilent N5502A/70422A
Agilent N5507A/70427A
Characteristics
Input Impedance 50 nominal
AM Noise DC coupled to 50 load
0 to +23 dBm 0 to +5 dBm 0 to +20 dBm
0 to +30 dBm +5 to +15 dBm
NOTE
Refer to the following system connect diagram examples in Chapter 18, System Interconnections, for more information about system interconnections.
Figure 307, E5505A system connections with standard test set, on page 401
Figure 308, E5505A system connections with test set option 001, on page 402
Figure 309, E5505A system connections with test set option 201, on page 403
Status messages
This section describes the status messages that appear on the display as the system performs the calibration routines.
Determining Presence of Beat Note... An initial check is made to verify that a
beatnote is present within the systems detection range.
Verifying zero-beat... The frequency of the beatnote is measured to see if it is
within 5% of the estimated Peak Tuning Range of the system. The system’s Peak Tuning Range is the portion of the voltage-controlled-oscillator (VCO) sources tuning range being used for the measurement.
When the system measures the phase noise of a signal source using the Phase Lock Loop technique (the technique being used in this example) it requires that one of the two sources used in the setup is a VCO. As you see later in this demonstration, you are required to estimate the tuning range of the VCO source you are using when you set up your own Phase Lock Loop measurements.
Agilent E5505A Users Guid e 89
4 Expanding Your Measurement Experience
Zero beating sources... The center frequencies of the sources are now
adjusted, if necessary, to position the beatnote within the 5% range. The adjustment is made with the tune voltage applied to the VCO source set at its nominal or center position.
Measuring the VCO Tuning Constant... The tuning sensitivity (Hz/V) of the VCO
source is now precisely determined by measuring the beatnote frequency at four tune voltage settings across the tuning range of the VCO source. Linearity across the tuning range is also verified
Measuring the Phase Detector Constant... The transfer characteristics (V/rad)
of the test sets phase detector are now determined for the specific center frequency and power level of the sources being measured.
Measuring PLL suppression... The required correction data is created to
compensate for the phase noise suppression which occurs within the bandwidth of the phase lock loop created for this measurement. The computer displays the PLL suppression curve and associated measurement values.
Sweep segments
When the system begins measuring noise, it places the noise graph on its display. As you watch the graph, you see the system plot its measurement results in frequency segments.
The system measures the noise level across its frequency offset range by averaging the noise within smaller frequency segments. This technique enables the system to optimize measurement speed while providing you with the measurement resolution needed for most test applications.
When the measurement is complete, refer to Chapter 14, Evaluating Your Measurement Results” for help in evaluating your measurement results. (If the test system has problems completing the measurement, it informs you by a message on the computer display.
Checking the beatnote
While the Connect Diagram is still displayed, Agilent recommends that you use an oscilloscope (connected to the Monitor port on the test set) or a counter to check the beatnote being created between the reference source and your DUT. The objective of checking the beatnote is to ensure that the center frequencies of the two sources are close enough in frequency to create a beatnote that is within the Capture Range of the system.
The phase lock loop (PLL) Capture Range is 5% of the peak tuning range of the VCO source you are using. (The peak tuning range for your VCO can be estimated by multiplying the VCO tuning constant by the tune range of VCO.
90 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
Refer to Chapter 14, Evaluating Your Measurement Results if you are not familiar with the relationship between the PLL capture range and the peak tuning range of the VCO.)
NOTE
If the center frequencies of the sources are not close enough to create a beatnote within the capture range, the system is not able to complete its measurement.
The beatnote frequency is set by the relative frequency difference between the two sources. If you have two very accurate sources set at the same frequency, the resulting beatnote is very close to 0 Hz.
Searching for the beatnote requires that you adjust the center frequency of one of the sources above and below the frequency of the other source until the beatnote appears on the oscilloscopes display. See Figure 57.
If incrementing the frequency of one of the sources does not produce a beatnote, you need to verify the presence of an output signal from each source before proceeding.
0
V
E5505a_oscillo_disp_beatnote 25 Feb 04 rev 1
-1V/div
Figure 57 Oscilloscope display of beatnote from test set monitor port
Agilent E5505A Users Guid e 91
4 Expanding Your Measurement Experience

Making the measurement

1 Click the Continue button when you have completed the beatnote check
and are ready to make the measurement.
2 When the PLL Suppression Curve dialog box appears, select View
Measured Loop Suppression, View Smoothed Loop Suppression, and View Adjusted Loop Suppress ion. See Figure 58 on page 92.
e5505a_user_select_suppression.ai rev2 10/21/03
Figure 58 Suppression selections
There are four different curves for this graph. (For more information about
loop suppression verification, refer to Chapter 15, Advanced Software Features.”)
a “Measured loop suppression curvethis is the result of the loop
suppression measurement performed by the E5505A system.
b “Smoothed measured suppression curvethis is a curve-fit
representation of the measured results, it is used to compare with the theoretical loop suppression.
c “Theoretical suppression curvethis is the predicted loop suppression
based on the initial loop parameters defined/selected for this particular measurement (kphi, kvco, loop bandwidth, filters, gain, and others).
d “Adjusted theoretical suppression curvethis is the new “adjusted
theoretical value of suppression for this measurement. It is based on
92 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
changing loop parameters (in the theoretical response) to match the smoothed measured curve as closely as possible.
When the measurement is complete, refer to Chapter 14, Evaluating Your Measurement Results for help in evaluating your measurement results.
Figure 59 on page 93 shows a typical phase noise curve for an RF Synthesizer.
E5505A_user_typ_noise_curve_8644B 24 Jun 04 rev 3
Figure 59 Typical phase noise curve for an 8644B 10 MHz measurement.
Table 10 on page 94 contains the data stored in the parameter definitions file.
Agilent E5505A Users Guid e 93
4 Expanding Your Measurement Experience
Table 10 Parameter data for the 8644B 10 MHz measurement
Step Parameters Data
1 Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Minimum Number of Averages
FFT Quality
2Sources Tab
Carrier Source
Frequency
Power
Carrier Source Output is connected to:
Detector Input
Frequency
Reference Source
Frequency
Reference Source Power
VCO Tuning Parameters
Nominal Tune Constant
Tune Range ±
Center Voltage
Input Resistance
Absolute Phase Noise (using a phase locked loop)
10 Hz
2 E + 6 Hz
4
Fa st
10 E + 6 Hz
7 dBm
Te st S e t
10 E +6 Hz
10 E +6 Hz (same as Carrier Source Frequency)
16 dBm
1 E +3 Hz/V
± 10 Volts
0 Volts
600 ohms
1
3Cal Tab
Phase Detector Constant
VCO Tune Constant
Phase Lock Loop Suppression
If Limit is exceeded
4Block Diagram Tab
Carrier Source
Downconverter
Reference Source
Timebase
Phase Detector
Test Set Tune Voltage Destination
VCO Tune Mode
5 Test Set Tab
Input Attenuation
LNA Low Pass Filter
LNA Gain
DC Block
PLL Integrator Attenuation
Measure Phase Detector Constant
Calculate from expected VCO Tune Constant
Verify calculated phase locked loop suppression
Show Suppression Graph
Manual
None
Agilent 8644B
None
Automatic Detector Selection
Reference Source
DC FM
0 dB
20 MHz (Auto checked)
Auto Gain (Minimum Auto Gain –14 dB)
Not checked
0 dBm
94 Agilent E5505A Users Guide
Expanding Your Measurement Experience 4
Table 10 Parameter data for the 8644B 10 MHz measurement (continued)
Step Parameters Data
6 Dowconverter Tab The downconverter parameters do not apply to this
measurement example.
7Graph Tab
Title
Graph Type
X Scale Minimum
X Scale Maximum
Y Scale Minimum
Y Scale Maximum
Normalize trace data to a:
Scale trace data to a new carrier frequency of:
Shift trace data DOWN by:
Trace Smoothing Amount
Power present at input of DUT
Confidence Test using Agilent 8644B Int vs Ext 10 MHz
Single-sideband Noise (dBc/Hz)
10 Hz
4 E + 6 Hz
0 dBc/Hz
- 170 dBc/Hz
1 Hz bandwidth
1 times the current carrier frequency
0 dB
0
0 dB
1 The stop frequency depends on the analyzers configured in your phase noise system.
Agilent E5505A Users Guid e 95
4 Expanding Your Measurement Experience

Viewing Markers

The marker function allows you to display the exact frequency and amplitude of any point on the results graph.
To access the marker function, on the View menu, clic k Markers. See
Figure 60. In the dialog box containing Marker buttons, up to nine markers
may be added. To remove a highlighted marker, click the Delete button. See
Figure 61

Figure 60 Navigate to markers

e5505a_user_add_delete_markers 24 Jun 04 rev 3
e5505a_user_nav_markers 24 Jun 04 rev 3

Figure 61 Adding and deleting markers

96 Agilent E5505A Users Guide

Omitting Spurs

Expanding Your Measurement Experience 4
The Omit Spurs function plots the currently loaded results without displaying any spurs that may be present.
1 On the View menu, click Display Preferences. See Figure 62.
e5505a_user_nav_display_pref 24 Jun 04 rev 3

Figure 62 Navigate to display preferences

2 In the Display Preferences dialog box, uncheck Spurs and click OK. See
Figure 63.
The graph is displayed without spurs. See Figure 64 on page 98.
e5505a_user_uncheck_spurs.ai rev2 10/21/03

Figure 63 Uncheck spurs

3 To re-display the spurs, check Spurs in the Display Preferences dialog box.
Agilent E5505A Users Guid e 97
4 Expanding Your Measurement Experience
e5505a_user_graph_without_spurs 24 Jun 04 rev 3

Figure 64 Graph displayed without spurs

98 Agilent E5505A Users Guide

Displaying the Parameter Summary

The Parameter Summary function allows you to quickly review the measurement parameter entries that were used for this measurement. The parameter summary data is included when you print the graph.
1 On the View menu, click Parameter Summary. See Figure 65.
Expanding Your Measurement Experience 4
e5505a_user_nav_param_sum 24 Jun 04 rev 3

Figure 65 Navigate to parameter summary

2 The Parameter Summary Notepad dialog box appears. The data can be
printed or changed using standard Notepad functionality. See Figure 66 on page 100.
Agilent E5505A Users Guid e 99
4 Expanding Your Measurement Experience
Agilent 8644B Int vs Ext 10 MHz
Agilent 8644B; VCO tuned using DCFM.
Agilent N5502A
e5505a_user_text_param_sum 24 Jun 04 rev 3

Figure 66 Parameter summary

100 Agilent E5505A Users Guide
Loading...