No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Microsoft ® is a U.S. registered trademark
of Microsoft Corporation.
Windows ® and MS Windows ®
are U.S. registered trademarks of
Micr osoft Corp orat ion.
Pentium ®
Intel Corporation.
Adobe ® Acrobat ® Adobe, the Adobe
Logo, Acrobat and the Acrobat Logo are
trademarks of Adobe Systems
Incorporated.
is a U.S. registered trademark of
Manual Part Number
E5505--90003
Edition
First edition, June 2004
Printed in USA
Agilent Technologies, Inc.
1400 Fountaingrove Pkwy
Santa Rosa, CA 9540 3
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1- 2) (June 1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and met.
If software is for use in the performance of
a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227-701 4
(June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
To power on a racked system36
To power on a benchtop system36
Starting the Measurement Software37
Figure 4. Navigation to the E5500 user interface37
Figure 5. Phase noise measurement subsystem main screen38
Performing a Confidence Test39
Figure 6. Opening the file containing pre-stored parameters39
Figure 7. Navigating to the Define Measurement window40
Beginning a measurement40
Figure 8. Navigating to the New Measurement window40
Figure 9. Confirm new measurement41
Figure 10. Setup diagram displayed during the confidence test.41
Figure 11.Connect diagram example42
Making a measurement42
Figure 12. Typical phase noise curve for test set confidence test43
Sweep segments43
Agilent E5505A User’s Guid e3
Congratulations43
Learning more44
Table 3. Parameter data for the N5500A confidence test example44
Powering the System Off45
To power off a racked sy stem45
To power off a be nchtop system45
Using the E5500 Shutdown Utility45
Figure 13. Shutdown utility icon45
3Phase Noise Basics
What is Phase Noise?48
Figure 14.RF sideband spectrum49
Phase terms49
Figure 15 . CW signal sidebands viewed in the frequency domain50
Figure 16. Deriving L(f) from a RF analyzer display51
Figure 17.L(f) Described Logarithmically as a Function of Offset Frequency51
Figure 18. Region of validity of L(f)52
4 Expanding Your Measurement Experience
Starting the Measurement Software54
Figure 19. Navigate to E5500 user interface54
Using the Asset Manager55
Configuring an asset55
Figure 20. Navigate to Asset Manager55
Figure 21. Navigate to Add in Asset Manager56
Figure 22. Select source as asset type56
Figure 23. Choose source57
Figure 24. Select I/O library57
Figure 25. Enter asset and serial number58
Figure 26. Enter comment58
Figure 27. Click check-mark button59
Figure 28.Confirmation message59
Using the Server Hardware Connections to Specify the Source60
Figure 33 . Asset Manager on System menu64
Figure 34. Asset Manager window64
Figure 35 . GPIB address dialog box65
Testing the 8663A Internal/External 10 MHz66
Required equipment66
Defining the measurement66
Figure 36. Select the parameters definition file66
Figure 37. Enter Source Information67
Table 5. Tuning characteristics for various sources68
Selecting a reference source68
Figure 38.Selecting a reference source68
Selecting loop suppression verification69
Figure 39. Selecting loop suppression verification69
Setting up for the 8663A 10 MHz measurement69
Figure 40. Noise floor for the 8663 10 MHz measurement70
Figure 41.Noise floor example70
Beginning the measurement71
Figure 42. Selecting new measurement71
Figure 43. Confirm new measurement71
Figure 44. Connection diagram72
Table 6.Test set signal input limits and characteristics73
Sweep segments75
Figure 45. Oscilloscope display of beatnote from test set monitor port76
Making the measurement76
Figure 46. Selecting suppression77
Figure 47. Typical phase noise curve for an 8663A 10 MHz measurement78
Table 7. Parameter data for the 8663A 10 MHz measurement79
Testing the 8644B Internal/External 10 MHz81
Required equipment81
Defining the measurement81
Figure 48. Select the parameters definition file81
Figure 49. Sources tab in define measurement window82
Table 8. Tuning characteristics for various sources83
Selecting a reference source83
Figure 50.Selecting a reference source84
Selecting loop suppression verification84
Figure 51. Selecting loop suppression verification85
Setting up the 8663A 10 MHz measurement85
Figure 52. Noise floor for the 8644B 10 MHz measurement85
Figure 53.Noise floor example86
Beginning the measurement87
Agilent E5505A User’s Guid e5
Figure 54. Selecting a new measurement87
Figure 55. Confirm measurement dialog box87
Figure 56.Connect diagram dialog box88
Table 9.Test set signal input limits and characteristics89
Figure 57. Oscilloscope display of beatnote from test set monitor port91
Making the measurement92
Figure 58. Suppression selections92
Figure 59. Typical phase noise curve for an 8644B 10 MHz measurement.93
Table 10. Parameter data for the 8644B 10 MHz measurement94
Viewing Ma rkers96
Figure 60. Navigate to markers96
Figure 61 . Adding and deleting markers96
Omitting Spurs97
Figure 62. Navigate to display preferences97
Figure 63. Uncheck spurs97
Figure 64.Graph displayed without spurs98
Displaying the Parameter Summary99
Figure 65. Navigate to parameter summary99
Figure 66. Parameter summary100
Exporting Measurement Results101
Figure 67. Export results choices101
Exporting Trace Data102
Figure 68.Trace data results102
Exporting spur data103
Figure 69. Spur data results103
Exporting X-Y data104
Figure 70.X-Y data results104
5 Absolute Measurement Fundamentals
The Phase-Lock-Loop Technique106
Understanding the Phase-Lock-Loop Technique106
Figure 71. Simplified block diagram of the phase lock loop configuration106
The Phase-Lock-Loop Circuit106
Figure 72. Capture and drift-tracking range with tuning range of VCO107
Figure 73. Capture and drift-tracking ranges and beatnote frequency108
What Sets the Measurement Noise Floor?110
The System Noise Floor110
Table 11. Amplitude ranges for L and R ports110
Figure 74. Relationship between the R input level and system noise floor110
6Agilent E5505A User’s Guide
The Noise Level of the Reference Source111
Figure 75. Reference source noise approaches DUT noise111
Selecting a Reference112
Figure 76. DUT noise approaches reference noise112
Using a Similar Device112
Using a Signal Generator113
Tuning Requirements113
Table 12.Tuning Characteristics of Various VCO Source Options113
Figure 77.Voltage tuning range limits relative to center voltage of the VCO tuning
Figure 78. Peak tuning range118
The Tuning Qualifications118
Minimizing Injection Locking120
Adding Isolation120
Increasing the PLL Bandwidth120
Figure 79. Peak tuning range (PTR) Required by injection locking.121
Inserting a Device122
An attenuator122
Figure 80. Measurement noise floor relative to R-Port signal level122
An amplifier123
Figure 81.Measurement noise floor as a result of an added attenuator123
Evaluating Noise Above the Small Angle Line124
Determining the Phase-Lock-Loop bandwidth124
Figure 82. Phase lock loop bandwidth provided by the peak tuning range125
Figure 83. Graph of small angle line and spur limit126
Figure 84. Requirements for noise exceeding small angle limit127
6 Absolute Measurement Examples
Stable RF Oscillator130
Required equipment130
Defining the measurement130
Figure 85. Select the parameters definition file130
Figure 86. Enter source information131
Table 14.Tuning characteristics for various sources132
Agilent E5505A User’s Guid e7
Selecting a reference source132
Figure 87.Selecting a reference source133
Selecting Loop Suppression Verification133
Figure 88. Selecting loop suppression verification134
Setup considerations for stable RF oscillator measurement134
Figure 89. Noise floor for the stable RF oscillator measurement135
Figure 90.Noise floor calculation example135
Beginning the measurement136
Figure 91. Selecting a new measurement136
Figure 92. Confirm new measurement136
Figure 93.Connect diagram for the stable RF oscillator measurement137
Table 15.Test set signal input limits and characteristics138
Checking the beatnote139
Figure 94. Oscilloscope display of beatnote from test set Monitor port140
Making the measurement140
Figure 95. Selecting suppressions141
Figure 96.Typical phase noise curve for a stable RF oscillator142
Table 16. Parameter data for the stable RF oscillator measurement143
Free-Running RF Oscillator145
Required equipment145
Defining the measurement145
Figure 97. Select the parameters definition file146
Figure 98. Enter source information147
Table 17.Tuning characteristics for various sources147
Selecting a reference source148
Figure 99.Selecting a reference source148
Selecting Loop Suppression Verification148
Figure 100. Selecting loop suppression verification149
Setup considerations for the free-running RF oscillator measurement149
Figure 101. Noise floor for the free-running RF oscillator measurement150
Figure 102. Noise floor calculation example150
Beginning the measurement151
Figure 103. Selecting a new measurement151
Figure 104. Confirm measurement dialog box151
Figure 105. Connect diagram for the free-running RF oscillator measurement152
Table 18.Test set signal input limits and characteristics153
Checking the beatnote153
Figure 106. Oscilloscope display of beatnote from test set Monitor port154
Making the measurement155
Figure 107. Selecting suppressions156
Figure 108.Typical phase noise curve for a free-running RF oscillator157
8Agilent E5505A User’s Guide
Table 19. Parameter data for the free-running RF oscillator measurement158
RF Synthesizer Using DCFM160
Required equipment160
Defining the measurement160
Figure 109. Select the parameters definition file160
Figure 110. Enter source information161
Table 20.Tuning characteristics for various sources162
Selecting a reference source162
Figure 111. Selecting a reference source162
Selecting Loop Suppression Verification163
Figure 112. Selecting loop suppression verification163
Setup considerations for the RF synthesizer using DCFM measurement163
Figure 113. Noise floor for the RF synthesizer (DCFM) measurement164
Figure 114. Noise floor calculation example164
Beginning the measurement165
Figure 115. Selecting a new measurement165
Figure 116. Confirm measurement dialog box165
Figure 117. Connect diagram for the RF synthesizer (DCFM) measurement166
Table 21.Test set signal input limits and characteristics167
Checking the beatnote167
Figure 118. Oscilloscope display of beatnote from the test set Monitor port168
Making the measurement169
Figure 119. Selecting suppressions169
Figure 120. Typical phase noise curve for an RF synthesizer using DCFM170
Table 22. Parameter Data for the RF Synthesizer (DCFM) Measurement171
RF Synthesizer Using EFC173
Required equipment173
Defining the measurement173
Figure 121. Select the parameters definition file173
Figure 122. Enter Source Information175
Table 23.Tuning Characteristics for Various Sources175
Selecting a reference source176
Figure 123. Selecting a reference source176
Selecting Loop Suppression Verification176
Figure 124. Selecting Loop suppression verification177
Setup considerations for the RF synthesizer using EFC measurement177
Figure 125. Noise floor for the RF synthesizer (EFC) measurement178
Figure 126. Noise floor calculation example178
Beginning the measurement179
Figure 127. Selecting a new measurement179
Figure 128. Confirm measurement dialog box179
Agilent E5505A User’s Guid e9
Figure 129. Connect diagram for the RF synthesizer (EFC) measurement180
Table 24. Test set signal Input Limits and Characteristics181
Checking the beatnote181
Figure 130. Oscilloscope display of a beatnote from the test set Monitor port182
Making the measurement182
Figure 131. Selecting suppressions183
Figure 132.Typical phase noise curve for an RF synthesizer using EFC184
Table 25. Parameter data for the RF synthesizer (EFC) measurement185
Microwave Source187
Required equipment187
Defining the measurement187
Figure 133. Select the parameters definition file187
Figure 134. Enter source information189
Table 26.Tuning characteristics for various sources189
Selecting a reference source190
Figure 135. Selecting a reference source190
Selecting Loop Suppression Verification190
Figure 136. Selecting loop suppression verification191
Setup considerations for the microwave source measurement191
Figure 137. Noise characteristics for the microwave measurement191
Beginning the measurement192
Figure 138. Selecting a new measurement192
Figure 139. Confirm measurement dialog box192
Figure 140. Connect diagram for the microwave source measurement193
Table 27.Test set signal input limits and characteristics194
Checking the beatnote194
Figure 141. Oscilloscope display of a beatnote from the test set Monitor port195
Making the measurement196
Figure 142. Selecting suppressions197
Figure 143.Typical phase noise curve for a microwave source198
Table 28. Parameter data for the microwave source measurement199
Assumptions about Residual Phase Noise Measurements204
Figure 146. Setup for typical residual phase noise measurement204
Frequency translation devices205
10Agilent E5505A User’s Guide
Figure 147. Measurement setup for two similar DUTs205
Calibrating the Measurement206
Figure 148. General equipment setup for making residual phase noise
measurements206
Calibration and measurement guidelines207
Calibration options208
User entry of phase detector constant209
Figure 149. Measuring power at phase detector signal input port210
Table 29. Acceptable amplitude ranges for the phase detectors.210
Figure 150. Phase detector sensitivity211
Figure 151. Adjust for quadrature212
Figure 152. Measuring power at phase detector reference input port212
Measured ± DC peak voltage213
Figure 153. Connection to optional oscilloscope for determining voltage peaks214
Table 30. Acceptable Amplitude Ranges for the Phase Detectors214
Measured beatnote215
Table 31. Frequency ranges215
Procedure216
Figure 154. Measuring power from splitter216
Table 32. Acceptable amplitude ranges for the phase detectors216
Figure 155. Calibration source beatnote injection217
Synthesized residual measurement using beatnote cal217
Table 33. Frequency Ranges217
Procedure218
Figure 156. Synthesized residual measurement using beatnote cal218
Double-Sided spur218
Figure 157. Calibration setup219
Table 34. Acceptable amplitude ranges for the phase detectors220
Figure 15 8. Measuring carrier-to-sideband ratio of the modulated port220
Single-Sided spur221
Figure 159. Calibration setup for single-sided spur222
Table 35. Acceptable Amplitude Ranges for the Phase Detectors223
Figure 160. Carrier-to-spur ratio of modulated signal223
Figure 161. Carrier-to-spur ratio of non-modulated signal224
Measurement Difficulties225
System connections225
8 Residual Measurement Examples
Amplifier Measurement Example228
Required equipment228
Figure 162. Setup for residual phase noise measurement229
Agilent E5505A User’s Guid e11
Defining the measurement229
Figure 163. Select the parameters definition file229
Figure 164. Navigate to residual phase noise230
Figure 165. Enter frequencies into source tab231
Figure 166. Select constant in the cal tab231
Figure 167. Select parameters in the block diagram tab232
Figure 168. Select graph description on graph tab233
Setup considerations for amplifier measurement233
Beginning the measurement234
Figure 169. Select meter from view menu234
Figure 170. Selecting New Measurement234
Figure 171. Confirm new measurement235
Figure 172. Setup diagram for the 8349A amplifier measurement example235
Table 36.Test set signal input limits and characteristics236
Making the measurement237
Table 37. Acceptable amplitude ranges for the phase detectors237
Figure 173. Residual connect diagram example238
Figure 174. Connection to optional oscilloscope for determining voltage peaks238
Figure 175.Adjust phase difference at phase detector239
Figure 176. Adjust phase shifter until meter indicates 0 volts240
When the measurement is complete240
Figure 177. Typical phase noise curve for a residual measurement241
Table 38. Parameter data for the amplifier measurement example241
9 FM Discriminator Fundamentals
The Frequency Discriminator Method244
Figure 178. Basic delay line/mixer frequency discriminator method244
Basic theory244
The discriminator transfer response245
Figure 179. Nulls in sensitivity of delay line discriminator246
Table 39. Choosing a delay line248
10 FM Discriminator Measurement Examples
Introduction250
Figure 180.FM Discriminator measurement setup250
FM Discriminator Measurement using Double-Sided Spur Calibration251
Table 40. Required Equipment for the FM Discriminator Measurement Example251
Determining the discriminator (delay line) length251
Figure 181. Discriminator noise floor as a function of delay time252
Defining the measurement252
Figure 182. Select the parameters definition file252
12Agilent E5505A User’s Guide
Figure 183. Select measurement type253
Figure 184. Enter frequencies in source tab254
Figure 185. Enter parameters into the call tab255
Figure 186. Select parameters in the block diagram tab256
Figure 187. Select Graph Description on Graph Tab256
Setup considerations257
Beginning the measurement258
Figure 188. Select meter from view menu258
Figure 189. Selecting New Measurement258
Figure 190. Confirm new measurement259
Figure 191. Setup diagram for the FM discrimination measurement example259
Table 41. Test Set Signal Input Limits and Characteristics260
Figure 192. Connect diagram example261
Making the measurement261
Figure 193. Calibration measurement (1 of 5)262
Figure 194. Calibration measurement (2 of 5)262
Figure 195. Calibration measurement (3 of 5)263
Figure 196. Calibration measurement (4 of 5)263
Figure 197. Calibration measurement (5 of 5)263
When the measurement is complete264
Figure 198.Typical phase noise curve using double-sided spur calibration264
Table 42. Parameter data for the double-sided spur calibration example265
Discriminator Measurement using FM Rate and Deviation Calibration267
Required equipment267
Table 43. Required equipment for the FM discriminator measurement exam ple267
Determining the discriminator (delay line) length268
Figure 199. Discriminator noise floor as a function of delay time268
Defining the measurement268
Figure 200. Select the parameters definition file269
Figure 201. Select measurement type269
Figure 202. Enter frequencies in Source tab270
Figure 203. Enter parameters into the Cal tab271
Figure 204. Enter parameters in the Block Diagram tab272
Figure 205. Select graph description on Graph tab273
Setup considerations273
Beginning the measurement274
Figure 206. Select meter from the View menu274
Figure 207. Selecting New Measurement274
Figure 208. Confirm new measurement275
Figure 209. Setup diagram for the FM Discrimination measurement example275
Table 44.Test set signal input limits and characteristics276
Agilent E5505A User’s Guid e13
Figure 210. System connect diagram example277
Making the measurement277
Figure 211. Calibration measurement (1 of 5)278
Figure 212. Calibration measurement (2 of 5)278
Figure 213. Calibration measurement (3 of 5)279
Figure 214. Calibration measurement (4 of 5)279
Figure 215. Calibration measurement (5 of 5)279
When the measurement is complete280
Figure 216.Typical phase noise curve using rate and deviation calibration280
Table 45. Parameter data for the rate and deviation calibration example281
AM noise measurement block diagrams285
Figure 217. AM noise system with N5500A opt 001285
Figure 218. AM noise system with external detector285
Figure 219. AM Noise system with 70429A Opt K21 AM detector286
Figure 220. AM noise system with N5507A downconverter286
AM detector286
Figure 221. AM detector schematic286
Table 46. Maximum carrier offset frequency287
Calibration and Measurement General Guidelines289
Method 1: User Entry of Phase Detector Constant290
Method 1, example 1290
Figure 222. Phase detector constant AM noise setup (method1, example 1)290
Figure 223. AM noise calibration setup291
Figure 224. AM detector sensitivity graph291
Method 1, example 2292
Figure 225. Phase detector constant AM noise setup (method 1, example 2)292
Figure 226. Modulation sideband calibration setup293
Method 2: Double-Sided Spur294
Method 2, example 1294
Figure 227. Double-Sided spur AM noise setup (method 2, example 1)294
Figure 228. Measuring the carrier-to-sideband ratio295
Figure 229. Measuring the calibration constant295
Method 2, example 2296
Figure 230. Double-sided spur AM noise setup (method 2, example 2)296
14Agilent E5505A User’s Guide
Figure 231. Measuring power at the am detector296
Figure 232. Measuring carrier-to-sideband ratio297
Figure 233. Measuring the calibration constant297
Method 3: Single-Sided Spur299
Figure 234. AM noise measurement setup using single-sided spur299
Figure 235. Measuring relative spur level300
Figure 236. Measuring detector sensitivity300
12 AM Noise Measurement Examples
AM Noise with N5500A Option 001302
Required equipment302
Figure 237. AM noise measurement configuration302
Defining the measurement303
Figure 238. Select the parameters definition file303
Figure 239. Navigate to AM noise304
Figure 240. Enter Frequencies in Source Tab304
Figure 241. Enter parameters into the cal tab305
Figure 242. Select parameters in the block diagram tab305
Figure 243. Select graph description on graph tab306
Beginning the measurement307
Figure 244. Selecting a new measurement307
Figure 245. Confirm measurement dialog box307
Figure 246. Connect diagram for the AM noise measurement308
Table 47.Test set signal input limits and characteristics309
Figure 247. Connect diagram example309
Making the measurement310
When the measurement is complete310
Figure 248. Typical AM noise curve310
Table 48. Parameter data for the AM noise using an N5500A Option 001311
13 Baseband Noise Measurement Examples
Baseband Noise with Test Set Measurement Example314
Defining the measurement314
Figure 249. Select the parameters definition file314
Beginning the measurement315
Figure 250. Selecting a new measurement315
Figure 251. Confirm measurement dialog box315
Figure 252. Connect diagram dialog box316
Making the measurement316
Figure 25 3. Typical phase noise curve for a baseband using a test set
measurement.316
Agilent E5505A User’s Guid e15
Table 49. Parameter data for the baseband using a test set measurement317
Baseband Noise without Test Set Measurement Example318
Defining the measurement318
Figure 254. Select the parameters definition file318
Beginning the measurement319
Figure 255. Selecting a new measurement319
Figure 256. Confirm measurement dialog box319
Figure 257. Connect diagram for baseband without test set measurement320
Figure 258. Instrument connection dialog box320
Making the measurement320
Figure 259.Typical curve for a baseband without test set measurement.321
Table 50. Parameter data for the baseband without using a test set measurement321
14 Evaluating Your Measurement Results
Evaluating the Results324
Looking for obvious problems324
Figure 260. Noise plot showing obvious problems325
Comparing against expected data325
Figure 261. Compensation for added reference source noise326
Figure 262. Measurement results and reference source noise327
Gathering More Data328
Repeating the measurement328
Figure 263. Repeating a measurement328
Doing more research328
Outputting the Results329
Using a printer329
Graph of Results330
Marker330
Figure 264. Navigate to marker330
Figure 26 5. Add and delete markers331
Table 51. List of topics that discuss problem solving in this chapter335
16Agilent E5505A User’s Guide
Discontinuity in the graph335
Table 52. Potential causes of discontinuity in the graph335
Higher noise level336
Spurs on the graph336
Table 53. Spurs on the graph337
Table 54. Actions to eliminate spurs337
Small angle line338
Figure 271. L(f) Is only valid for noise levels below the small angle line339
Blanking Frequency and Amplitude Information on the Phase Noise Graph353
Security level procedure353
Figure 280. Navigate to security level353
Figure 281. Choosing levels of security354
Figure 282. Unsecured: all data is viewable354
Figure 283. Choosing levels of security355
Figure 284. Secured: frequencies cannot be found-1355
Figure 285. Secured: frequencies cannot be found-2356
Figure 286. Choosing levels of security356
Figure 287. Secured: frequencies and amplitudes cannot be viewed357
Agilent E5505A User’s Guid e17
16 Reference Graphs and Tables
Approximate System Noise Floor vs. R Port Signal Level360
Figure 288. Noise floor for R input port360
Phase Noise Floor and Region of Validity361
Figure 289. Region of validity361
Phase Noise Level of Various Agilent Sources362
Figure 290. Noise level for various reference sources362
Increase in Measured Noise as Ref Source Approaches DUT Noise363
Figure 291. Reference source and DUT noise levels363
Approximate Sensitivity of Delay Line Discriminator364
Figure 292. Delay line discriminator sensitivity364
AM Calibration365
Figure 293. AM detector sensitivity365
Voltage Controlled Source Tuning Requirements366
Figure 294. Tuning voltage required for phase lock366
Tune Range of VCO for Center Voltage367
Figure 295. Tune range of VCO for center voltage367
Peak Tuning Range Required by Noise Level368
Figure 296.Typical source noise level vs. minimum tuning range368
Phase Lock Loop Bandwidth vs. Peak Tuning Range369
Figure 297. PLL BW vs. peak tuning range369
Noise Floor Limits Due to Peak Tuning Range370
Figure 298. Noise at source’s peak tuning range370
Tuning Characteristics of Various VCO Source Options371
Table 55.Tuning parameters for several VCO options371
8643A Frequency Limits372
Table 56. 8643A frequency limits372
8643A mode keys372
Table 57. Operating characteristics for 8643A modes 1, 2, and 3373
How to access special functions373
Figure 299. 8643A special function keys373
Description of special functions 120 and 125373
8644B Frequency Limits375
Table 58. 8644B frequency limits375
8644B mode keys375
Table 59. Operating characteristics for 8644B modes 1, 2, and 3376
18Agilent E5505A User’s Guide
How to access special functions376
Figure 300. 8644B special functions keys376
Description of special function 120377
8664A Frequency Limits378
Table 60. 8664A frequency limits378
8664A mode keys378
Table 61. Operating characteristics for 8664A modes 2 and 3378
How to access special functions379
Figure 301. Special functions keys379
Description of special functions 120379
8665A Frequency Limits380
Table 62. 8665A frequency limits380
8665A mode keys380
Table 63. Operating characteristics for 8665A modes 2 and 3381
How to access special functions381
Figure 302. 8665A special functions keys381
Description of Special Functions 120 and 124381
8665B Frequency Limits383
Table 64. 8665B frequency limits383
8665B mode keys383
Table 65. Operating characteristics for 8665B modes 2 and 3384
How to access special functions384
Figure 303. 8665B Special functions keys384
Description of special functions 120 and 124385
Figure 304. Connect adapter to PC digitizer card397
Figure 305. PC to test set connection, standard model398
Figure 306. PC to test set (options 001 and 201) and downconverter connection399
Figure 307. E5505A system connections with standard test set401
Figure 308. E5505A system connections with test set option 001402
Figure 309. E5505A system connections with test set option 201403
19 PC Components Installation
Overview406
PC Digitizer Software: Phase 1407
To install the PC digitizer software 407
Hardware Installation408
Preparing for installation408
Figure 310. Remove screws from side of CPU408
Figure 311. Slide cover off409
Figure 312. Remove hold-down bar409
Accessing PC expansion slots410
Figure 313.Vertically-Mounted expansion slots410
Taki n g ESD prec a utio n s410
Installing the PC digitizer card411
Figure 314. PC digitizer card411
Figure 315. Insert PC digitizer card411
Figure 316. Secure card with screw412
Figure 317. Connect adapter to PC digitizer card412
Installing the GPIB interface card413
Figure 318. GPIB interface card413
Figure 319. Insert GPIB card413
Figure 320. Secure card with screw414
Figure 321. Replace cover414
System Interconnections415
Table 73. E5505A connectors and adapters415
Making Connections415
20Agilent E5505A User’s Guide
Figure 322. Test set connection, standard model416
Figure 323. Test set (options 001 and 201) and downconverter connection417
PC Digitizer Software: Phase 2418
Agilent I/O Libraries419
To install the Agilent I/O libraries 419
Measurement Software Installation424
To install the E5500 software 424
Asset Configuration426
Setting Up Asset Manager426
To set up Asset Manager 426
Configuring the Phase Noise Test Set428
Configuring the PC Digitizer432
Figure 324. Add assets432
Figure 325. Choose asset type433
Figure 326. Select supporting ACM433
Figure 327. Choose the interface and address for the PC digitizer434
Table 74. Default GPIB addresses435
Figure 328. Choose model and serial number436
Figure 329. Select (internal) in baseband source436
Figure 330. Enter a comment about the configured asset437
Figure 331. Asset manager screen showing configured PC Digitizer437
Configuring the Agilent E4411A/B (ESA-L1500A) Swept Analyzer438
Required equipment444
To verify the PC digitizer card input’s performance 444
PC Digitizer Card Input Performance Verification449
Required equipment449
To verify the PC digitizer card input’s performance 449
Agilent E5505A User’s Guid e21
21 Preventive Maintenance
Using, Inspecting, and Cleaning RF Connectors454
Repeatability454
RF Cable and Connector Care454
Proper Connector Torque455
Table 75. Proper Connector Torque455
Connector Wear and Damage455
SMA Connector Precautions456
Cleaning Procedure456
Table 76. Cleaning Supplies Available from Agilent457
General Procedures and Techniques458
Figure 338. GPIB, 3.5 mm, Type-N, power sensor, and BNC connectors458
Connector Removal459
Instrument Removal461
Standard instrument461
To remove an instrument from a rack 461
Half-Rack-Width Instrument462
To remove a half-width instrument from a system rack 462
Figure 339. Instrument lock links, front and rear463
Benchtop Instrument463
To remove an instrument from a benchtop system 463
Instrument Installation464
Standard rack instrument464
To install an instrument 464
Half-Rack-Width instrument465
To install the instrument in a rack 465
Benchtop instrument465
To install an instrument in a benchtop system 465
A Service, Support, and Safety Information
Safety and Regulatory Information468
Safety summary468
Equipment Installation468
Environmental conditions469
Before applying power469
Ground the instrument or system470
Fuses and Circuit Breakers470
Maintenance471
Safety symbols and instrument markings471
Table 77. Safety symbols and instrument markings471
22Agilent E5505A User’s Guide
Declaration of Conformity473
Compliance with German noise requirements473
Table 78. German noise requirements summary473
Compliance with Canadian EMC requirements473
Service and Support474
Agilent on the Web474
Return Procedure475
Determining your instrument’s serial number475
Figure 340. Serial number location475
Shipping the instrument476
To package the instrument for shipping 476
Agilent E5505A User’s Guid e23
24Agilent E5505A User’s Guide
E5505A Phase Noise Measurement System
User’s Guide
1
Getting Started
Introduction 26
Documentation Map 27
Additional Documentation28
System Overview 29
Agilent Technologies
25
1Getting Started
Introduction
This guide introduces you to the Agilent E5505A Phase Noise Measurement
System software and hardware. It provides procedures for configuring the
E5500 Phase Noise Measurement software, executing measurements,
evaluating results, and using the advanced software features. It also covers
phase noise basics and measurement fundamentals to get you started.
Use Table 1 on page 27as a guide to:
• Learning about the E5505A phase noise measurement system
• Learning about phase noise basics and measurement fundamentals
• Using the E5505A system to make specific phase noise measurements.
In this guide you’ll also find information on system connections and
specifications, and procedures for re-installing phase-noise-specific hardware
and software in the system PC.
NOTE
Installation information for your system is provided in the Agilent E5505A Phase Noise
Measurement System Installation Guide.
26Agilent E5505A User’s Guide
Documentation Map
Table 1E5505A user’s guide map
Getting Started1
Learning about the E5505A SystemLearning Phase Noise Basics &
Measurement Fundamentals
Chapter 1, “Getting Started”
Chapter 2, “Introduction and
Measurement”
Chapter 4, “Expanding Your
Measurement Experience”
Chapter 3, “Phase Noise Basics”
Chapter 5, “Absolute Measurement
Fundamentals”
Chapter 7, “Residual Measure ment
Fundamentals”
Chapter 9, “FM Discriminator
Fundamentals”
Chapter 11, “AM Noise Mea surement
Fundamentals”
Using the E5505A for Specific Phase
Noise Measurements
Chapter 6, “Absolute Measurement
Examples”
Chapter 8, “Residual Measurement
Examples”
Chapter 10, “FM Discriminator
Measurement Examples”
Chapter 12, “AM Noise Measurement
Examples”
Chapter 13, “Baseband Noise
Measurement Examples”
Chapter 14, “Evaluating Your
Measurement Results”
Chapter 15, “Advanced Software
Features”
Chapter 17, “System Specifications”
Chapter 18, “System
Interconnections”
Chapter 19, “PC Components
Installation”
Chapter 20, “PC Digitizer Performance
Verification”
Chapter 21, “Preventive
Maintenance”
Chapter A, “Service, Support, and
Safety Information”
Chapter 16, “Reference Graphs and
Ta b l es ”
Agilent E5505A User’s Guid e27
1Getting Started
Additional Documentation
You can access the complete set of PDF documents that support the E5505A
system through the system GUI. (Adobe
Acrobat Reader is supplied.)
Navigate the menu as shown in Figure 1. The files are stored on the system PC
hard drive and on the E5500A software CD. Be sure to explore the E5500 Help
menu for additional information.
The E5505A system documentation includes:
• Agilent E5505A Phase Noise Measurement System Installation Guide
• Agilent E5505A Phase Noise Measurement System User's Guide
• Agilent N5500A Phase Noise Test Set User's Guide
• Agilent E5500 Series Phase Noise Measurement Systems SCPI Command
Reference
• Agilent E5500 Phase Noise Measurement System Online Help
.
E5500_start_menu
04 Apr 04 rev 1
Figure 1Navigate to system documentation
28Agilent E5505A User’s Guide
System Overview
The E5505A Phase Noise Measurement System provides flexible sets of
measurements on one-port devices such as voltage controlled oscillators
(VCOs), dielectric resonator oscillators (DROs), crystal oscillators, and
synthesizers, and on two-port devices such as amplifiers and converters. The
E5505A system measures absolute and residual phase noise, AM noise, and
low-level spurious signals, as well as CW and pulsed signals. It operates in the
frequency range of 50 KHz to 26.5 GHz.
The E5505A phase noise measurement system combines standard
instruments, phase noise components, and PC software for maximum
flexibility and re-use of assets. The system PC operates under Windows® XP
Professional® and controls the system through the E5500 measurement
software. The E5500 software enables many stand-alone instruments to work
in the system. This standalone-instrument architecture easily configures for
various measurement techniques, including the phase-lock-loop
(PLL)/reference-source technique, and delay-line and FM-discriminator
methods.
Getting Started1
NOTE
The E5505A system is available as a one-bay wide, System II rack and as a
benchtop model. Due to the system’s flexibility, the hardware in the system
varies greatly with the options selected. You may be installing instruments you
already own in the system as well. A typical system includes these
components:
• Advantech custom PC with digitizer card assembly
• 15-inch display (flat-panel or standard), keyboard, and mouse
• Windows® XP Professional® operating system
• Agilent E5500 Phase Noise Measurement software
• Phase noise test set
• Downconverter
• RF source
Additional instruments may include a spectrum analyzer, oscilloscope,
RF counter, power meter, and power splitter.
For detailed information on the instruments in your E5505A phase noise measurement
system, refer to the individual instrument user guides (provided on CD-ROM).
Agilent E5505A User’s Guid e29
1Getting Started
Figure 2 shows a typical configuration of an E5505A benchtop system.
The E5505A replaces earlier Agilent E5500A/B series phase noise systems,
which are based on MMS technology. The E5505A system uses GPIB
communication and certain instruments have been redesigned with GPIB
functionality. However, the E5505A system and E5500 software are backwards
compatible with earlier systems and instruments, including the MMS
mainframe. You may easily integrate existing assets into your E5505A system.
Table 2 shows the E5505A instruments and earlier-model equivalents.
Tab l e 2Equivalent system/instrument model numbers
System or InstrumentNew NumberOld Number
Phase noise measurement system E5505 A E5501A, E5501B, E5502B,
E5503A, E5503B, E5504A, E5504B
Test setE5500A 70420A
6.6 GHz downconverterN5501A 70421A
18 GHz downconverterN5502A 70422A
26.5 GHz downconverterN5507A 70427A, 71707A
Microwave source N5508A70428A, 71708A
30Agilent E5505A User’s Guide
E5505A Phase Noise Measurement System
User’s Guide
2
Introduction and Measurement
Introducing the GUI 32
Designing to Meet Your Needs34
E5505A Operation: A Guided Tour 35
Powering the System On36
Performing a Confidence Test 39
Powering the System Off45
Agilent Technologies
31
2Introduction and Measurement
Introducing the GUI
The graphical user interface (GUI) gives the user instant access to all
measurement functions, making it easy to configure a system and define or
initiate measurements. The most frequently used functions are displayed as
icons on a toolbar, allowing quick and easy access to the measurement
information.
The forms-based graphical interaction helps you define your measurement
quickly and easily. Each form tab is labeled with its content, preventing you
from getting lost in the defining process.
The system provides three default segment tables. To obtain a quick look at
your data, select the “fast” quality level. If it is important to have more
frequency resolution to separate spurious signals, use the “normal” and “high
resolution” quality levels. If you need to customize the offset range beyond the
defaults provided, tailor the measurement segment tables to meet your needs
and save them as a custom selection.
You can place up to nine markers on the data trace that can be plotted with the
measured data.
Other features include:
• Plotting data without spurs
• Tabular listing of spurs
• Plotting in alternate bandwidths
• Parameter summary
• Color printouts to any supported color printer
Figure 3 on page 33 shows an example of the GUI.
32Agilent E5505A User’s Guide
System Requirements
Introduction and Measurement2
E5500_main_screen
24 Jun 04 rev 2
Figure 3E5500 graphical user interface (GUI)
Agilent E5505A User’s Guid e33
2Introduction and Measurement
Designing to Meet Your Needs
The E5505A Phase Noise Measurement System is a high performance
measurement tool that enables you to fully evaluate the noise characteristics
of your electronic instruments and components with unprecedented speed and
ease. The phase noise measurement system provides you with the flexibility
needed to meet today’s broad range of noise measurement requirements.
In order to use the phase noise system effectively, it is important that you have
a good understanding of the noise measurement you are making. This manual
is designed to help you gain that understanding and quickly progress from a
beginning user of the phase noise system to a proficient user of the system’s
basic measurement capabilities.
NOTE
Beginning
If you have just received your system or need help with connecting the hardware or loading
software, refer to your Agilent E5505A Phase Noise Measurement System Installation Guide now. Once you have completed the installation procedures, return to “E5505A
Operation: A Guided Tour" on page 35 to begin learning how to make noise measurements
with the system.
The section “E5505A Operation: A Guided Tour" on page 35 contains a
step-by-step procedure for completing a phase noise measurement. This
measurement demonstration introduces system operating fundamentals for
whatever type of device you plan to measure.
Once you are familiar with the information in this chapter, you should be
prepared to start Chapter 4, “Expanding Your Measurement Experience. After
you have completed that chapter, refer to Chapter 14, “Evaluating Your
Measurement Results for help in analyzing and verifying your test results.
34Agilent E5505A User’s Guide
E5505A Operation: A Guided Tour
This measurement demonstration introduces you to the system’s operat ion by
guiding you through an actual phase noise measurement.
You will be measuring the phase noise of the Agilent N5500A Phase Noise Test
Set’s low noise amplifier. (The measurement made in this demonstration is the
same measurement that is made to verify the system’s operation.)
As you step through the measurement procedures, you will soon discover that
the phase noise measurement system offers enormous flexibility for
measuring the noise characteristics of your signal sources and two-port
devices.
Required equipment
The equipment shipped with this system is all that is required to complete this
demonstration. (Refer to the E5505A Phase Noise Measurement System Installation Guide if you need information about setting up the hardware or
installing the software.)
Introduction and Measurement2
How to begin
NOTE
Follow the setup procedures beginning on the next page. The phase noise
measurement system displays a setup diagram that shows you the front panel
cable connections to make for this measurement.
If you need additional information about connecting instruments, refer to Chapter 18, “System Interconnections.”
Agilent E5505A User’s Guid e35
2Introduction and Measurement
Powering the System On
This section provides procedures for powering on a racked or benchtop
system. First connect your system to an appropriate AC power source, then
follow the steps below.
WARNING
Before applying power, make sure the AC power input and the location of the
system meet the requirements given in Ta b l e 4 on page 14. Failure to do so may
result in damage to the system or personal injury.
NOTE
Warm-up Time: The downconverter and RF source instruments contain ovenized
oscillators which must warm up for 30 minutes to produce accurate measurements.
Standby Mode: The RF source uses a standby mode to keep the ovenized oscillator warm
when the instrument is connected (plugged in) to AC power, even when the power switch is in the off position. To completely shut down the instrument, you must disconnect it from the
AC power supply.
To power on a racked system
1 Press the system power switch (front, top right of the rack) to the on
position.
2 Verify that all instrument power switches are on.
3 Allow the system to warm up for 30 minutes.
To power on a benchtop system
1 Press the power switch on each instrument to the on position.
2 If you have the system connected to a safety power strip, turn the strip’s
power switch to the on position.
3 Allow the system to warm up for 30 minutes.
36Agilent E5505A User’s Guide
Starting the Measurement Software
1 Place the E5500 phase noise measurement software disk in the CD-ROM
drive.
2 Using Windows® Start menu as in Figure 4, navigate to the E5500 User
Interface.
Introduction and Measurement2
E5500_start_menu
04 Apr 04 rev 1
Figure 4Navigation to the E5500 user interface
3 The phase noise measurement subsystem main screen appears (Figure 5 on
page 38).
Agilent E5505A User’s Guid e37
2Introduction and Measurement
E5500_main_screen
24 Jun 04 rev 2
NOTE
Figure 5Phase noise measurement subsystem main screen
The default background for the screen is gray. You can change the background color by
selecting View/Display Preferences and clicking on the Background Color button.
38Agilent E5505A User’s Guide
Performing a Confidence Test
This first measurement is a confidence test that functionally checks the
N5500A test set’s filters and low-noise amplifiers using the test set’s low noise
amplifier. The phase detectors are not tested. This confidence test also
confirms that the test set, PC, and analyzers are communicating with each
other. To conduct the test, use a file with pre-stored parameters named
Confidence.pnm.
1 On the E5500 GUI main menu, select File\Open.
2 If necessary, choose the drive or directory where the file you want is stored.
3 In the File Name box, select Confidence.pnm (Figure 6).
4 Click the Open button.
Introduction and Measurement2
E5500_open_win_conf
04 Apr 04 rev 1
Figure 6Opening the file containing pre-stored parameters
The appropriate measurement definition parameters for this example have
been pre-stored in this file. Table 3 on page 44 lists the parameter data that
has been entered for the N5500A confidence test example.
5 To view the parameter data in the software, navigate to the Define
Measurement window. Use Figure 7 on page 40 as a navigation guide. The
parameter data is entered using the tabbed windows. Select various tabs to
see the type of information entered behind each tab.
Agilent E5505A User’s Guid e39
2Introduction and Measurement
e5505a_users_nav_define_wind
24 Jun 04 rev 3
Figure 7 Navigating to the Define Measurement window
6 Click the Close button.
Beginning a measurement
1 From the Measure menu, choose New Measurement. See Figure 8.
E5500_new_measurement
04 Apr 04 rev 1
Figure 8Navigating to the New Measurement window
40Agilent E5505A User’s Guide
Introduction and Measurement2
2 When the Do you want to Perform a New Calibration and Measurement?
dialog box appears, click Yes. See Figure 9.
E5500_new_cali_meas
04 Apr 04 rev 1
Figure 9Confirm new measurement
3 When the Connect Diagram dialog box appears, connect the 50 Ω
termination, provided with your system, to the test set’s noise input
connector. (Refer to Figure 10 on page 41 for more information about the
correct placement of the 50 Ω termination.)
E5500_verify_con
23 Mar 04 rev 1
50 Ω termination
Figure 10Setup diagram displayed during the confidence test.
Agilent E5505A User’s Guid e41
2Introduction and Measurement
N5500A Standard Test Set
N5500A TEST SET
SIGNAL
50 kHz-1600 MHz
RF ANALYZER
INPUT
50 W
1V Pk
0.01 Hz-100 MHz +15 dBm MIN
PHASE DET OUTPUT
MONITOR
GPIB
NOISE
STATUS
50 kHz-1600MHz
REF INPUT
50 Ω
Termination
ANALYZERANALYZER
<100 kHz<100 MHz50 W 20mA MAX
100 W/1.25 LPS
N5500A Opt. 001 Test Set
N5500A OPT 001 TEST SET
SIGNAL
50 kHz-26.5 GHz
RF ANALYZER
ANALYZERANALYZER
<100 kHz<100 MHz50 W 20mA MAX
N5500A Option 201Test Set
N5500A OPT 201 TEST SET
SIGNAL
50 kHz-1600 MHz
RF ANALYZER
ANALYZERANALYZER
<100 kHz<100 MHz50 W 20mA MAX
e5505a_user_connect_diag.ai
rev2 10/20/03
INPUT
50 W
1V Pk
0.01 Hz-100 MHz +15 dBm MIN
PHASE DET OUTPUT
MONITOR
INPUT
50 W
1V Pk
0.01 Hz-100 MHz +15 dBm MIN
PHASE DET OUTPUT
MONITOR
GPIB
NOISE
100 W/1.25 LPS
GPIB
NOISE
100 W/1.25 LPS
TUNE VOLTAGE
OUT OF LOCK
STATUS
REF INPUT
50 kHz-1600MHz 1.2 - 26.5 GHz
+7 dBm MIN
FROM
DOWNCONVERTER
DOWNCONVERTER
TUNE VOLTAGE
OUT OF LOCK
STATUS
REF INPUT
50 kHz-1600MHz 1.2 - 26.5 GHz
+7 dBm MIN
mW SIGNAL
1.2-26.5 GHz
TUNE VOLTAGE
OUT OF LOCK
50 Ω
Termination
TO
50 Ω
Termination
Figure 11Connect diagram example
Making a measurement
1 Press the Continue button.
• Because you selected New Measurement to begin this measurement, the
system starts by running the routines required to calibrate the current
measurement setup.
• Figure 12 on page 43 shows a typical baseband phase noise plot for an
Figure 12Typical phase noise curve for test set confidence test
Sweep segments
When the system begins measuring noise, it places the noise graph on its
display. As you watch the graph, you see the system plot its measurement
results in frequency segments.
The system measures the noise level across its frequency offset range by
averaging the noise within smaller frequency segments. This technique enables
the system to optimize measurement speed while providing you with the
measurement resolution needed for most test applications.
Congratulations
You have completed a phase noise measurement. This measurement of the test
set’s low noise amplifier provides a convenient way to verify that the system
hardware and software are properly configured for making noise
measurements. If your graph looks like that in Figure 12, you can be confident
that your system is operating normally.
Agilent E5505A User’s Guid e43
2Introduction and Measurement
Learning more
Continue with this demonstration by turning to Chapter 4, “Expanding Your
Measurement Experience to” learn more about performing phase noise
measurements.
Tab l e 3Parameter data for the N5500A confidence test example
StepParametersData
1Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Minimum Number of Averages
FFT Quality
Swept Quality
2Cal Tab
Gain preceding noise input0 dB
• Baseband Noise (using a test set)
• 10 Hz
• 100 E + 6 Hz
1
• 4
• Fa st
• Fa st
3Block Diagram Tab
Noise SourceTest Set Noise Input
4Tes t Se t Tab
Input Attenuation
LNA Low Pass Filter
LNA Gain
DC Block
PLL I ntegra tor Attenuation
5Graph Tab
Title
Graph Type
X Scale Minimum
X Scale Maximum
Y Scale Minimum
Y Scale Maximum
Normalize trace data to
Scale trace data to a new carrier
frequency of:
Shift trace data DOWN by
Trace Smoothing Amount
Power present at input of DUT
• 0 dB
• 20 MHz (Auto checked)
• Auto Gain (Minimum Auto Gain –14 dB)
• Not checked
• 0 dBm
• Confidence Test, N5500A low noise amplifier.
• Base band noise (dBv/Hz)
• 10 Hz
• 100 E + 6 Hz
• 0 dBv/Hz
• –200 dBv/Hz
• 1 Hz bandwidth
• 1 times the current carrier frequency
• 0 dB
• 0
• 0 dB
1 The Stop Frequency depends on the analyzers configured in your phase noise system.
44Agilent E5505A User’s Guide
Powering the System Off
To power off a racked system
1 On the E5500 software menu, select File\Exit. Always shut down the E5500
software before powering off the E5505A system.
2 Press the system power switch (front, top right of the rack) to the off
position.
Introduction and Measurement2
CAUTION
Always shut down the E5500 software before powering off the E5505A system.
Failure to do so may produce errors in the stem, and result in an inoperable system
or inaccurate measurements. If you do receive errors during shutdown, startup, or
operation, use the E5500 Shutdown utility to restore functionality to the system.
To power off a benchtop system
1 On the E5500 software menu, select File\Exit.
2 Press the power switch on each instrument to the off position.
Using the E5500 Shutdown Utility
If you receive error messages during the power on or off procedures, or during
operation, use the E5500 Shutdown utility to shut down the system. This
utility automatically fixes most errors and restores functionality to the system.
If you still receive errors after running the E5500 Shutdown utility, call your
local Agilent Technologies Service Center.
To run the E5500 Shutdown utility
1 Double-Click on the E5500 Shutdown utility shortcut on the PC desktop
and follow the onscreen instructions. (You can also navigate to it using the
menu path Start/Agilent Subsystems/E5500 Phase Noise/Shutdown.)
Figure 13Shutdown utility icon
2 When the shutdown utility has finished, use the Start menu to shut down
the PC. Then power the system off.
Agilent E5505A User’s Guid e45
2Introduction and Measurement
46Agilent E5505A User’s Guide
E5505A Phase Noise Measurement System
User’s Guide
3
Phase Noise Basics
What is Phase Noise? 48
Phase terms49
Agilent Technologies
47
3Phase Noise Basics
f
What is Phase Noise?
Frequency stability can be defined as the degree to which an oscillating source
produces the same frequency throughout a specified period of time. Every RF
and microwave source exhibits some amount of frequency instability. This
stability can be broken down into two components:
• long-term stability
• short-term stability
Long-term stability describes the frequency variations that occur over long
time periods, expressed in parts per million per hour, day, month, or year.
Short-term stability contains all elements causing frequency changes about the
nominal frequency of less than a few seconds duration. The chapter deals with
short-term stability.
Mathematically, an ideal sinewave can be described by
Vt()V
Where = nominal amplitude,
But an actual signal is better modeled by
V
o
V
and = nominal frequency
2π fotsin
o
o
2πfotsin=
o
= linearly growing phase component,
Vt()Voε t()+2πfot∆φ t()+sin=
Where = amplitude fluctuations,
and = randomly fluctuating phase term or phase noise.
This randomly fluctuating phase term could be observed on an ideal RF
analyzer (one which has no sideband noise of its own) as in Figure 14 on
page 49.
ε t()
∆φ t()
48Agilent E5505A User’s Guide
Figure 14RF sideband spectrum
f()
Phase terms
There are two types of fluctuating phase terms:
e5505a_user_RF_sideband.ai
rev2 10/20/03
Phase Noise Basics3
• spurious signals
• phase noise
Spurious signals
The first are discrete signals appearing as distinct components in the spectral
density plot. These signals, commonly called spurious, can be related to known
phenomena in the signal source such as power line frequency, vibration
frequencies, or mixer products.
Phase noise
The second type of phase instability is random in nature, and is commonly
called phase noise. The sources of random sideband noise in an oscillator
include thermal noise, shot noise, and flicker noise.
Many terms exist to quantify the characteristic randomness of phase noise.
Essentially, all methods measure the frequency or phase deviation of the
source under test in the frequency or time domain. Since frequency and phase
are related to each other, all of these terms are also related.
Spectral densityOne fundamental description of phase instability or phase
noise is spectral density of phase fluctuations on a per-Hertz basis. The term
spectral density describes the energy distribution as a continuous function,
expressed in units of variance per unit bandwidth. Thus (Figure 15 on
Where BW (bandwidth is negligible with respect to any changes in versus
2
rad
------ --- ---==
Hz
φ
the fourier frequency or offset frequency (f).
L(f)Another useful measure of noise energy is L(f), which is then directly
related to by a simple approximation which has generally negligible error
φ
if the modulation sidebands are such that the total phase deviation are much
less than 1 radian (
∆φ
pk
<< radian).
L f()
1
-- -
S
f()=
∆φ
2
e5505a_user_CW_sidebands_freq.ai
rev2 10/20/03
Figure 15CW signal sidebands viewed in the frequency domain
L(f) is an indirect measurement of noise energy easily related to the RF power
spectrum observed on an RF analyzer. Figure 16 shows that the National
Institute of Standards and Technology (NIST) defines L(f) as the ratio of the
power--at an offset (
f) Hertz away from the carrier. The phase modulation
sideband is based on a per Hertz of bandwidth spectral density and or offset
frequency in one phase modulation sideband, on a per Hertz of bandwidth
spectral density and
(f) equals the Fourier frequency or offset frequency.
P
ssb
total signal power
------ ----- -==
P
s
= single sideband (SSB) phase noise to carrier ration (per Hertz)
50Agilent E5505A User’s Guide
Phase Noise Basics3
e5505a_user_derivingL_RF_display.ai
rev2 10/20/03
Figure 16Deriving L(f) from a RF analyzer display
is usually presented logarithmically as a spectral density plot of the
L f()
phase modulation sidebands in the frequency domain, expressed in dB relative
to the carrier per Hz (dBc/Hz) as shown in Figure 17. This chapter, except
where noted otherwise, uses the logarithmic form of as follows:
S∆ff()2f2L f()=
.
L f()
e5505a_user_log_offset_freq.ai
rev2 10/20/03
Figure 17L(f) Described Logarithmically as a Function of Offset Frequency
Agilent E5505A User’s Guid e51
3Phase Noise Basics
Caution must be exercised when is calculated from the spectral density of
the phase f luctuations because the calculation of is dependent on
Sφf()L f()
L f()
the small angle criterion. Figure 18, the measured phase noise of a free
running VCO described in units of , illustrates the erroneous results that
L f()
can occur if the instantaneous phase modulation exceeds a small angle line.
Approaching the carrier obviously increases in error as it indicates a
L f()
relative level of +45 dBc/Hz at a 1 Hz offset (45 dB more noise power at a 1 Hz
offset in a 1 Hz bandwidth than in the total power of the signal); which is of
course invalid.
Figure 18 shows a 10 dB/decade line drawn over the plot, indicating a peak
phase deviation of 0.2 radians integrated over any one decade of offset
frequency. At approximately 0.2 radians the power in the higher order
sidebands of the phase modulation is still insignificant compared to the power
in the first order sideband which insures that the calculation of remains
valid. Above the line the plot of becomes increasingly invalid, and
L f()Sφf()
L f()
must be used to represent the phase noise of the signal.
e5505a_user_region_validity_L.ai
rev2 10/20/03
Figure 18Region of validity of L(f)
52Agilent E5505A User’s Guide
E5505A Phase Noise Measurement System
User’s Guide
4
Expanding Your Measurement
Experience
Starting the Measurement Software54
Using the Asset Manager 55
Using the Server Hardware Connections to Specify the Source 60
Testing the 8663A Internal/External 10 MHz 66
Testing the 8644B Internal/External 10 MHz81
Viewing Ma rkers96
Omitting Spurs 97
Displaying the Parameter Summary99
Exporting Measurement Results 101
Agilent Technologies
53
4Expanding Your Measurement Experience
Starting the Measurement Software
1 Make sure your computer and monitor are turned on.
2 Place the E5500 Phase Noise Measurement System software disk in the disc
holder and insert in the CD-ROM drive.
3 Using Figure 19 as a guide, navigate to the E5500 User Interface.
E5500_start_menu
04 Apr 04 rev 1
Figure 19Navigate to E5500 user interface
54Agilent E5505A User’s Guide
Using the Asset Manager
Use the Asset Manager to add assets to your E5505A system. The process is
essentially the same for any asset, including reference sources. In fact, the
procedure in this section uses an Agilent 8663 source as an example. (The
procedure applies to all Agilent sources, including the 8257x series.)
Adding an asset involves two steps once the hardware connections have been
made:
• Configuring the asset
• Verifying the server hardware connections.
Expanding Your Measurement Experience4
WARNING
Configuring an asset
Be sure to power off the system before making all hardware connections other than GPIB.
Connecting assets with the power applied can result in personal injury and damage to the
hardware. (For more information on connecting assets, see Chapter 18, “System Interconnections)
This procedure demonstrates how to add an asset to the E5505A system.
1 Using Figure 20 as a guide, navigate to Asset Manager. For this example we
invoke the Asset Manager Wizard from the E5500 main screen. This is the
most common way to add assets.
E5500_asset_manager
23 Mar 04 rev 1
Figure 20Navigate to Asset Manager
Agilent E5505A User’s Guid e55
4Expanding Your Measurement Experience
2 Select Add in the Asset Manager window. See Figure 21.
E5500_add_source2
16 Apr 04 rev 1
Figure 21Navigate to Add in Asset Manager
3 From the Asset Type pull-down list in Choose Asset Role dialog box, select
Source, then click Next. See Figure 22.
.
E5500_add_source3
16 Apr 04 rev 1
Figure 22Select source as asset type
56Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
4 Click on the source to be added, then click the Next button (see Figure 23).
E5500_add_source4
16 Apr 04 rev 1
Figure 23Choose source
NOTE
5 From the Interface pull-down list, select GPIB0. See Figure 24.
6 In the Address box, type 19.
19 is the default address for a source. Ta b le 4 on page 63 shows the default GPIB address
for all system instruments.
7 In the Library pull-down list, select the Agilent Technologies VISA. Click
the Next button.
E5500_add_source5
16 Apr 04 rev 1
Figure 24Select I/O library
Agilent E5505A User’s Guid e57
4Expanding Your Measurement Experience
8 In the Set Model & Serial Numbers dialog box, type in your source name
and its corresponding serial number. Click the Next button. See Figure 25.
E5500_add_source6
16 Apr 04 rev 1
Figure 25Enter asset and serial number
9 In the Enter A Comment dialog box, you may type a comment that
associates itself with the asset you have just configured. Click Finish. See
Figure 26.
E5500_add_source7
16 Apr 04 rev 1
Figure 26Enter comment
58Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
10 In the Asset Manager window, select the source in the left window pane.
Click the check-mark button on the toolbar to verify connectivity. See
Figure 27.
E5500_add_source8
16 Apr 04 rev 1
Figure 27Click check-mark button
• The Asset Manager displays a message verifying the connection to your
asset. This indicates that you have successfully configured a source. (See
Figure 28.)
E5500_add_source9
16 Apr 04 rev 1
Figure 28Confirmation message
11 To exit the Asset Manager, on the menu select Server/Exit.
12 Perform the procedure “Using the Server Hardware Connec tions to Specify
the Source" on page 60.
Agilent E5505A User’s Guid e59
4Expanding Your Measurement Experience
Using the Server Hardware Connections to Specify the Source
1 From the System menu, choose Server Hardware Connections. See
Figure 29.
E5500_add_source10
16 Apr 04 rev 1
Figure 29Navigate to server hardware connections
2 Select the Sources tab shown in Figure 30.
E5500_add_source11
16 Apr 04 rev 1
Figure 30Select Sources tab
60Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
3 From the Reference Source pull-down list, select Agilent 8663A.
a A green check-mark appears after an automatic I/O check has been
successfully performed by the software. If nothing happens, click the
Check I/O button to manually initiate the check.
E5500_add_source12
16 Apr 04 rev 1
Figure 31Successful I/O check
b A red circle with a slash appears if the I/O check is unsuccessful.
E5500_add_source13
16 Apr 04 rev 1
Figure 32Failed I/O check
c If the I/O check fails, click the Asset Manager button to return to the
Asset Manager (see Figure 32).
Agilent E5505A User’s Guid e61
4Expanding Your Measurement Experience
d In the Asset Manager, verify that the 8663A is configured correctly. Do
the following:
• Check your system hardware connections.
• Click the green check-mark button on the Asset Manager’s toolbar to
verify connectivity.
• Return to Server Hardware Connections and click the Check I/O
button to re-check it.
NOTE
Use the same process to add additional assets to your E5505A system.
62Agilent E5505A User’s Guide
Setting GPIB Addresses
Table 4 shows the default GPIB addresses for the E5505A system instruments.
If you need to change a GPIB address to prevent a conflict between assets, use
the Asset Manager as shown in the easy procedure starting on page 64.
Tab l e 4Default GPIB addresses
InstrumentAddress
Te s t se t2 0
Downconverter28
Microwave downconverter28
RF analyzer17
FFT analyzer (PC digitizer card)1
FFT analyzer (89410A)18
Expanding Your Measurement Experience4
Source # 1 19
Source # 223
Counter3
Agilent E1430 VXI digitizer
Agilent E1437 VXI digitizer
Agilent E1420B VXI counter
Agilent E1441 VXI ARB
1 The E5500 software supports this instrument although it is not part of the standard E5505A system.
1
1
1
1
129
192
48
80
Agilent E5505A User’s Guid e63
4Expanding Your Measurement Experience
To change the GPIB address
1 On the E5500 main menu, select System/Asset Manager. See Figure 33.
E5500_asset_manager
23 Mar 04 rev 1
Figure 33Asset Manager on System menu
2 Double-Click on the desired instrument in the Asset Manager list (left
pane). See Figure 34.
E5500_asset_mgr_screen
23 Mar 04 rev 1
Figure 34Asset Manager window
64Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
3 Type the desired address in the dialog box. See Figure 35.
E5500_edit_asset_info
04 Apr 04 rev 1
Figure 35GPIB address dialog box
4 Click OK.
5 To exit the Asset Manager, on the menu select Server/Exit.
Next proceed to one of the following absolute measurements using either an
Agilent 8257x or an Agilent 8644B source:
• “Testing the 8663A Internal/External 10 MHz" on page 66.
• “Testing the 8644B Internal/External 10 MHz" on page 81.
Agilent E5505A User’s Guid e65
4Expanding Your Measurement Experience
Testing the 8663A Internal/External 10 MHz
This measurement example helps you measure the absolute phase noise of an
RF synthesizer.
CAUTION
To prevent damage to the test set’s hardware components, do not apply the input
signal to the signal input connector until the input attenuator has been correctly set
for the desired configuration, as shown in Tabl e 6 on page 73. Apply the input signal
when the connection diagram appears.
Required equipment
This measurement requires an Agilent 8663A in addition to the phase noise
test system and your device under test (DUT).
and adapters necessary to connect the DUT and reference source to the test
set.
NOTE
To ensure accurate measurements allow the DUT and measurement equipment to warm up
at least 30 minutes before making the noise measurement.
Defining the measurement
1 From the File menu in the E5500 User Interface,choose Open . If necessary,
choose the drive or directory where the file you want is stored.
2 In the File Name box, choose “Conf_SigGen_10MHz.pnm.” See Figure 36.
You also need the coaxial cables
.
E5500_open_win_conf
04 Apr 04 rev 1
Figure 36Selectthe parameters definition file
66Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
3 Click the Open button.
The appropriate measurement definition parameters for this example have
been pre-stored in this file. Table 7 on page 79 lists the parameter data that
has been entered for this measurement example.)
NOTE
Note that the source parameters entered for step 2 in Tab le 7 on page 79 may not be
appropriate for the reference source you are using. To change these values, refer to Ta b l e 5
on page 68, then continue with step 4 below. Otherwise, go to “Beginning the
measurement" on page 71.
4 Using Figure 37 as a guide, navigate to the Sources tab.
a Enter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz).
Enter the same frequency for the detector input frequency.
b Enter the VCO (Nominal) Tuning Constant (see Table 5 on page 68).
c Enter the Tune Range of VCO (see Table 5).
d Enter the Center Voltage of VCO (see Table 5).
e Enter the Input Resistance of VCO (see Table 5).
e5505a_user_enter_source_info
24 Jun 04 rev 3
Figure 37 Enter Source Information
Agilent E5505A User’s Guid e67
4Expanding Your Measurement Experience
Table 5Tuning characteristics for various sources
VCO SourceCarrier
Freq.
Tuning Constant
(Hz/V)
Center
Voltage
(V)
Voltage
Tuning Range
(± V)
Input
Resistance
(Ω)
Tuning
Calibration
Method
Agilent 8662/3A
EFC
DCFM
υ
0
5 E – 9 x υ
FM Deviation
0
0
0
10
10
1E + 6
1 K (8662)
600 (8663)
Measure
Com pute
Com pute
Agilent 8642A/BFM Deviation010600Compute
Agilent 8644BFM Deviation010600Compute
Other Signal Generator
DCFM Calibrated for ±1VFM Deviation010R
Other User VCO SourceEstimated within a
factor of 2
–10 to
+10
in
1 E + 6Measure
Com pute
Selecting a reference source
1 Using Figure 38 as a guide, navigate to the Block Diagram tab.
2 From the Reference Source pull-down list, select Agilent-8663.
3 When you have completed these operations, click the Close button.
Agilent-8257
e5505a_user_select_ref_source
24 Jun 04 rev 3
Figure 38 Selecting a reference source
68Agilent E5505A User’s Guide
Selecting loop suppression verification
1 Using Figure 39 as a guide, navigate to the Cal tab.
2 Check Verify calculated phase locked loop suppression and Always Show
Suppression Graph. Select If limit is exceeded: Show Loop Suppression
Graph.
3 When you have completed these operations, click the Close button.
Expanding Your Measurement Experience4
e5505a_user_select_loop
24 Jun 04 rev 3
Figure 39 Selecting loop suppression verification
Setting up for the 8663A 10 MHz measurement
The signal amplitude at the test set’s R input (Signal Input) port sets the
measurement noise floor level. Use Figure 40 on page 70 and Figure 41 on
page 70 to help determine the amplitude required to provide a noise f loor level
that is below the expected noise floor of your DUT. For more information
about this graph, refer to “Reference Graphs and Tables" on page 359.
Agilent E5505A User’s Guid e69
4Expanding Your Measurement Experience
+15
+5
-5
R Port signal level (dBm)
-15
-140-160-170-180-150
Expected phase noise floor of system (dBc/Hz)
n5505a_exp_phase_noise
25 Feb 04 rev 1
Figure 40Noise floor for the 8663 10 MHz measurement
L Port level +15dBm
f 10kHz
e5505a_user_noise_floor_ex
24 Jun 04 rev 3
Figure 41Noise floor example
70Agilent E5505A User’s Guide
If the output amplitude of your DUT is not sufficient to provide an adequate
measurement noise floor, it is necessary to insert a low-noise amplifier
between the DUT and the test set. Refer to “Inserting a Device" on page 122 for
details on determining the effect the amplifiers noise will have on the
measured noise floor.
Beginning the measurement
1 From the Measurement menu, choose New Measurement. See Figure 42.
Expanding Your Measurement Experience4
E5500_new_measurement
04 Apr 04 rev 1
Figure 42Selecting new measurement
2 When the Do you want to perform a New Calibration and Measurement?
prompt appears, click Ye s.
E5500_new_cali_meas
Figure 43Confirm new measurement
3 When the Connect Diagram dialog box appears, click on the hardware
pull-down arrow and select your hardware configuration from the list. See
Figure 44 on page 72.
Agilent E5505A User’s Guid e71
4Expanding Your Measurement Experience
CAUTION
E5500_verify_con
23 Mar 04 rev 1
Figure 44Connection diagram
4 Connect your DUT and reference sources to the test set at this time and
confirm your connections as shown in the appropriate connect diagram.
• The input attenuator (Option 001 only) is now correctly configured
based on your measurement definition.
The test set’s signal input is subject to the limits and characteristics contained in
Ta b l e 6 on page 73.
To prevent damage to the test set’s hardware components, do not apply the input
signal to the test set’s signal input connector until the input attenuator (Option 001)
has been set by the phase noise software, which occurs at the connection diagram.
5 Press Continue.
• As the system performs the calibration routines, various status messages
appear on the display. When the last message—Measuring PLL
suppression—appears, you can choose to continue the routine or stop it.
6 Press Continue using Adjusted Loop Suppression to continue making the
noise measurement, or press Abort to stop the measurement. (For
descriptions of the messages, see “Status messages" on page 74. You have
time to read through the descriptions while the system completes the
routines.)
72Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
:
Tab l e 6Test set signal input limits and characteristics
Limits
Frequency50 kHz to 26.5 GHz
Maximum Signal Input Power+30 dBm
At Attenuator Output, Operating Level Range:
• RF Phase Detectors
• Microwave Phase Detectors
• Internal AM Detector
Downconverters:
• N5502A/70422A
• N5507A/70427A
Characteristics
Input Impedance50 Ω nominal
AM NoiseDC coupled to 50 Ω load
0 to +23 dBm
0 to +5 dBm
0 to +20 dBm
0 to +30 dBm
+5 to +15 dBm
NOTE
Refer to the following system connect diagram examples in Chapter 18, “System
Interconnections” for more information about system interconnections.
•
Figure 307, “E5505A system connections with standard test set,” on page 401
• Figure 308, “E5505A system connections with test set option 001,” on page 402
• Figure 309, “E5505A system connections with test set option 201,” on page 403
Agilent E5505A User’s Guid e73
4Expanding Your Measurement Experience
Status messages
This section describes the status messages that appear on the display as the
system performs its calibration routines.
Determining Presence of Beat Note... An initial check is made to verify that a
beatnote is present within the system’s detection range.
Verifying Zero-Beat... The frequency of the beatnote is measured to see if it is
within 5% of the estimated Peak Tuning Range of the system. The system’s
Peak Tuning Range is the portion of the voltage-controlled-oscillator (VCO)
source’s tuning range being used for the measurement.
When the system measures the phase noise of a signal source using the Phase
Lock Loop technique (the technique being used in this example) it requires
that one of the two sources used in the setup is a VCO. As you will see later in
this demonstration, you are required to estimate the tuning range of the VCO
source you are using when you set up your own Phase Lock Loop
measurements.
Zero beating sources... The center frequencies of the sources are now
adjusted, if necessary, to position the beatnote within the 5% range. The
adjustment is made with the tune voltage applied to the VCO source set at its
nominal or center position.
Measuring the VCO Tuning Constant...The tuning sensitivity (Hz/V) of the VCO
source is now precisely determined by measuring the beatnote frequency at
four tune voltage settings across the tuning range of the VCO source. Linearity
across the tuning range is also verified
Measuring the Phase Detector Constant...The transfer characteristics (V/rad)
of the test set’s phase detector are now determined for the specific center
frequency and power level of the sources being measured.
Measuring PLL suppression...The required correction data is created to
compensate for the phase noise suppression which occurs within the
bandwidth of the phase lock loop created for this measurement. The computer
displays the PLL suppression curve and associated measurement values.
74Agilent E5505A User’s Guide
Sweep segments
When the system begins measuring noise, it places the noise graph on its
display. As you watch the graph, you see the system plot its measurement
results in frequency segments.
The system measures the noise level across its frequency offset range by
averaging the noise within smaller frequency segments. This technique enables
the system to optimize measurement speed while providing you with the
measurement resolution needed for most test applications.
When the measurement is complete, refer to Chapter 14, “Evaluating Your
Measurement Results for help in evaluating your measurement results. (If the
test system has problems completing the measurement, it informs you by
placing a message on the computer display.
Checking the beatnote
While the Connect Diagram is still displayed, Agilent recommends that you use
an oscilloscope (connected to the Monitor port on the test set) or a counter to
check the beatnote being created between the reference source and your DUT.
The objective of checking the beatnote is to ensure that the center frequencies
of the two sources are close enough in frequency to create a beatnote that is
within the Capture Range of the system.
Expanding Your Measurement Experience4
NOTE
The phase lock loop (PLL) Capture Range is 5% of the peak tuning range of the
VCO source you are using. (The peak tuning range for your VCO can be
estimated by multiplying the VCO tuning constant by the tune range of VCO.
Refer to Chapter 14, “Evaluating Your Measurement Results” if you are not
familiar with the relationship between the PLL capture range and the peak
tuning range of the VCO.)
If the center frequencies of the sources are not close enough to create a beatnote within
the capture range, the system is not able to complete its measurement.
The beatnote frequency is set by the relative frequency difference between the
two sources. If you have two very accurate sources set at the same frequency,
the resulting beatnote is very close to 0 Hz.
Searching for the beatnote requires that you adjust the center frequency of one
of the sources above and below the frequency of the other source until the
beatnote appears on the oscilloscope’s display. See Figure 45 on page 76.
If incrementing the frequency of one of the sources does not produce a
beatnote, you need to verify the presence of an output signal from each source
before proceeding.
Agilent E5505A User’s Guid e75
4Expanding Your Measurement Experience
0
V
E5505a_oscillo_disp_beatnote
25 Feb 04 rev 1
Figure 45Oscilloscope display of beatnote from test set monitor port
Making the measurement
1 Click the Continue button when you have completed the beatnote check
and are ready to make the measurement.
2 When the PLL Suppression Curve dialog box appears, check View Measured
Loop Suppression, View Smoothed Loop Suppression, and View Adjusted
Loop Suppression in the lower right of the dialog box.
See Figure 46 on page 77.
-1V/div
76Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
e5505a_user_select_suppression.ai
rev2 10/21/03
Figure 46Selecting suppression
There are four different curves for the this graph. (For more information about
loop suppression verification, refer to Chapter 15, “Advanced Software
Featu r e s ”).
• “Measured” loop suppression curve—this is the result of the loop
suppression measurement performed by the E5505A system.
• “Smoothed” measured suppression curve—this is a curve-fit representation
of the measured results, it is used to compare with the “theoretical” loop
suppression.
• “Theoretical” suppression curve—this is the predicted loop suppression
based on the initial loop parameters defined/selected for this particular
measurement (kphi, kvco, loop bandwidth, filters, gain, and others).
• “Adjusted” theoretical suppression curve—this is the new “adjusted”
theoretical value of suppression for this measurement. It is based on
changing loop parameters (in the theoretical response) to match the
“smoothed” measured curve as closely as possible.
When the measurement is complete, refer to Chapter 14, “Evaluating Your
Measurement Results
for help in evaluating your measurement results.
Figure 47 on page 78 shows a typical phase noise curve for an RF Synthesizer.
Agilent E5505A User’s Guid e77
4Expanding Your Measurement Experience
e5505a_user_typ_noise_curve_8663a
24 Jun 04 rev 3
Figure 47Typical phase noise curve for an 8663A 10 MHz measurement
Table 7 on page 79 contains the data stored in the parameter definitions file.
78Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
Tab l e 7Parameter data for the 8663A 10 MHz measurement
Step ParametersData
1Type and Range Tab
• Measurement Type
• Absolute Phase Noise (with phase locked loop)
• 10 Hz
• Start Frequency
• Stop Frequency
• Minimum Number of Averages
• 2 E + 6 Hz
• 4
• Fast
1
• FFT Quality
2Sources Tab
Carrier Source
• Frequency
• Power
• Carrier Source Output connected to
• Detector Input Frequency
• Reference Source Frequency
• Reference Source Power
• 10 E + 6 Hz
• 7 dBm
• Te st Se t
• 10 E +6 Hz
• 10 E +6 Hz (same as Carrier Source Frequency)
• 16 dBm
• VCO Tuning Parameters:
• Nominal Tune Constant
• Tune Range ±
• Center Voltage
• Input Resistance
• 1 E +3 Hz/V
• ± 10 Volts
• 0 Volts
• 600 ohms
3Cal Tab
• Phase Detector Constant
• VCO Tune Constant
• Phase Lock Loop Suppression
• If Limit is exceeded
• Measure Phase Detector Constant
• Calculate from expected VCO Tune Constant
• Verify calculated phase locked loop
suppression
• Show Suppression Graph
4Block Diagram Tab
• Carrier Source
• Downconverter
• Reference Source
• Timebas e
• Phase Detector
• Test Set Tune Voltage Destination
• VCO Tune Mode
5Test Set Tab
• Input Attenuation
• LNA Low Pass Filter
• LNA Gain
• DC Block
• PLL Integrator Attenuation
6Dowconverter TabThe downconverter parameters do not apply to
• Manual
• None
• Agilent 8663A
• None
• Automatic Detector Selection
• Reference Source
• DCFM
• 0 dB
• 20 MHz (Auto checked)
• Auto Gain (Minimum Auto Gain –14 dB)
• Not checked
• 0 dBm
this measurement example.
Agilent E5505A User’s Guid e79
4Expanding Your Measurement Experience
Tab l e 7Parameter data for the 8663A 10 MHz measurement (continued)
Step ParametersData
7Graph Tab
• Title
• Graph Type
• X Scale Minimum
• X Scale Maximum
• Y Scale Minimum
• Y Scale Maximum
• Normalize trace data to a:
• Scale trace data to a new carrier
frequency of:
• Shift trace data DOWN by
• Trace Smoothing Amount
• Power present at input of DUT
1 The Stop Frequency depends on the analyzers configured in your phase noise system.
• Confidence Test using Agilent 8663A Int vs Ext
10 MHz
• Single-sideband Noise (dBc/Hz)
• 10 Hz
• 4 E + 6 Hz
• 0 dBc/Hz
• –170 dBc/Hz
• 1 Hz bandwidth
• 1 times the current carrier frequency
• 0 dB
• 0
• 0 dB
80Agilent E5505A User’s Guide
Testing the 8644B Internal/External 10 MHz
This measurement example helps you measure the absolute phase noise of an
RF synthesizer.
Expanding Your Measurement Experience4
CAUTION
To prevent damage to the test set’s hardware components, do not apply the input signal to
the signal input connector until the input attenuator has been correctly set for the desired
configuration, as shown in Ta b l e 8 on page 83. Apply the input signal when the connection
diagram appears.
NOTE
To ensure accurate measurements allow the DUT and measurement equipment to warm up
at least 30 minutes before making the noise measurement.
Required equipment
This measurement requires an Agilent 8644B in addition to the phase noise
test system and your DUT.
necessary to connect the DUT and reference source to the test set.
Defining the measurement
1 From the file menu of the E5505A User Interface, choose Open. If
necessary, choose the drive or directory where the file you want is stored.
2 In the File Name box, choose “Conf_8644B_10MHz.pnm.” See Figure 42.
You also need the coaxial cables and adapters
E5500_open_win_conf
04 Apr 04 rev 1
Figure 48Select the parameters definition file
Agilent E5505A User’s Guid e81
4Expanding Your Measurement Experience
3 Click the Open button.
• The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table 10 on page 94 lists the parameter
data that has been entered for the RF Synthesizer using a DCFM
measurement example.
NOTE
The source parameters shown in Ta b l e 1 0 on page 94 may not be appropriate for the
reference source you are using. To change these values, refer toTab l e 8, “Tuning
characteristics for various sources,” on page 83, then continue with step 4 below.
Otherwise, go to “Beginning the measurement" on page 87.
4 Using Figure 49 on page 82 as a guide, navigate to the Sources tab
aEnter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz);
enter the same frequency for the detector input frequency.
bEnter the VCO (Nominal) Tuning Constant (see Table 8 on page 83).
cEnter the Tune Range of VCO (see Table 8 on page 83).
dEnter the Center Voltage of VCO (see Table 8 on page 83).
eEnter the Input Resistance of VCO (see Table 8 on page 83)
e5505a_user_enter_source_info
24 Jun 04 rev 3
Figure 49 Sources tab in define measurement window
82Agilent E5505A User’s Guide
Table 8Tuning characteristics for various sources
Expanding Your Measurement Experience4
VCO SourceCarrier
Freq.
Tuning Constant
(Hz/V)
Center
Voltage
(V)
Voltage
Tuning Range
(± V)
Input
Resistance (Ω)
Tuning
Calibration
Method
Agilent 8662/3A
EFC
DCFM
υ
0
5 E – 9 x υ
FM Deviation
0
0
0
10
10
1E + 6
1 K (8662)
600 (8663)
Measure
Com pute
Com pute
Agilent 8642A/BFM Deviation010600Compute
Agilent 8644BFM Deviation010600Compute
Other Signal Generator
DCFM Calibrated for
FM Deviation010R
in
Com pute
±1V
Other User VCO SourceEstimated within a
factor of 2
–10 to
+10
1 E + 6Measure
Selecting a reference source
1 From the Define menu, choose Measurement; then choose the Block
Diagram tab from the Define Measurement window. See Figure 50.
2 From the Reference Source pull-down list, select 8644.
Agilent E5505A User’s Guid e83
4Expanding Your Measurement Experience
Agilent-8644
e5505_user_select_ref_source8644
24 Jun 04 rev 3
Figure 50 Selecting a reference source
3 When you have completed these operations, click the Close button.
Selecting loop suppression verification
1 From the Define menu, choose Measurement; then choose the Cal tab from
the Define Measurement window.
2 In the Cal dialog box, check Verify calculated phase locked loop
suppression and Always Show Suppression Graph. Select If limit is
exceeded: Show Loop Suppression Graph. See Figure 51 on page 85.
3 When you have completed these operations, click the Close button.
84Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
e5505a_user_select_loop
24 Jun 04 rev 3
Figure 51 Selecting loop suppression verification
Setting up the 8663A 10 MHz measurement
The signal amplitude at the R input (Signal Input) port on the test set sets the
measurement noise floor level. Use the graph in Figure 52 and the example in
Figure 53 on page 86 to determine the amplitude. For more information, refer
to Chapter 16, “Reference Graphs and Tables.”
+15
+5
-5
R Port signal level (dBm)
-15
-140-160-170-180-150
Expected phase noise floor of system (dBc/Hz)
n5505a_exp_phase_noise
25 Feb 04 rev 1
Figure 52Noise floor for the 8644B 10 MHz measurement
L Port level +15dBm
f 10kHz
Agilent E5505A User’s Guid e85
4Expanding Your Measurement Experience
e5505a_user_noise_floor_ex
24 Jun 04 rev 3
Figure 53Noise floor example
If the output amplitude of your DUT is not sufficient to provide an adequate
measurement noise floor, it is necessary to insert a low-noise amplifier
between the DUT and the test set. Refer to the section “Inserting a Devic e" on
page 122 for details on determining the effect the amplifier’s noise wi ll have on
the measured noise floor.
86Agilent E5505A User’s Guide
Beginning the measurement
Expanding Your Measurement Experience4
CAUTION
To prevent damage to the test set’s hardware components, do not apply the input
signal to the signal input connector until the input attenuator has been correctly set
for the desired configuration, as shown in Ta bl e 9 on page 89. Apply the input
signals when the connection diagram appears, as in step 3 below.
1 From the Measurement menu, choose New Measurement. SeeFigure 54.
.
E5500_new_measurement
04 Apr 04 rev 1
Figure 54Selecting a new measurement
2 When the Do you want to perform a New Calibration and Measurement?
prompt appears, click Ye s.
E5500_new_cali_meas
04 Apr 04 rev 1
Figure 55Confirm measurement dialog box
3 When the Connect Diagram dialog box appears, click on the hardware
drop-down arrow and select your hardware configuration from the list. See
Figure 56 on page 88.
Agilent E5505A User’s Guid e87
4Expanding Your Measurement Experience
e5505a_conn_diag_dialog
rev 1 23 jun 04
Figure 56Connect diagram dialog box
CAUTION
4 Connect your DUT and reference sources to the test set at this time.
Confirm your connections as shown in the Connect Diagram (Figure 56).
• The input attenuator (Option 001 only) has now been correctly
configured based on your measurement definition.
The test set’s signal input is subject to the limits and characteristics in Ta b l e 9 on
page 89.
To prevent damage to the test set’s hardware components, do not apply the input
signal to the test set’s signal input connector until the input attenuator (Option 001)
has been set by the phase noise software, which occurs when the connection
diagram appears.
5 Press Continue.
• As the system performs the calibration routines, various status messages
appear on the display. When the last message—Measuring PLL
suppression—appears, you can choose to continue the routine or stop it.
6 Press Continue using Adjusted Loop Suppression to continue making the
noise measurement, or press Abort to stop the measurement. (For
descriptions of the messages, see “Status messages" on page 74. You have
time to read through the descriptions while the system completes the
routines.)
88Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
Tab l e 9Test set signal input limits and characteristics
Limits
Frequency50 kHz to 26.5 GHz
Maximum Signal Input Power+30 dBm
At Attenuator Output, Operating Level Range:
• RF Phase Detectors
• Microwave Phase Detectors
• Internal AM Detector
Downconverters:
• Agilent N5502A/70422A
• Agilent N5507A/70427A
Characteristics
Input Impedance50 Ω nominal
AM NoiseDC coupled to 50 Ω load
0 to +23 dBm
0 to +5 dBm
0 to +20 dBm
0 to +30 dBm
+5 to +15 dBm
NOTE
Refer to the following system connect diagram examples in Chapter 18, “System
Interconnections,” for more information about system interconnections.
•
Figure 307, “E5505A system connections with standard test set,” on page 401
• Figure 308, “E5505A system connections with test set option 001,” on page 402
• Figure 309, “E5505A system connections with test set option 201,” on page 403
Status messages
This section describes the status messages that appear on the display as the
system performs the calibration routines.
Determining Presence of Beat Note... An initial check is made to verify that a
beatnote is present within the system’s detection range.
Verifying zero-beat...The frequency of the beatnote is measured to see if it is
within 5% of the estimated Peak Tuning Range of the system. The system’s
Peak Tuning Range is the portion of the voltage-controlled-oscillator (VCO)
source’s tuning range being used for the measurement.
When the system measures the phase noise of a signal source using the Phase
Lock Loop technique (the technique being used in this example) it requires
that one of the two sources used in the setup is a VCO. As you see later in this
demonstration, you are required to estimate the tuning range of the VCO
source you are using when you set up your own Phase Lock Loop
measurements.
Agilent E5505A User’s Guid e89
4Expanding Your Measurement Experience
Zero beating sources...The center frequencies of the sources are now
adjusted, if necessary, to position the beatnote within the 5% range. The
adjustment is made with the tune voltage applied to the VCO source set at its
nominal or center position.
Measuring the VCO Tuning Constant...The tuning sensitivity (Hz/V) of the VCO
source is now precisely determined by measuring the beatnote frequency at
four tune voltage settings across the tuning range of the VCO source. Linearity
across the tuning range is also verified
Measuring the Phase Detector Constant...The transfer characteristics (V/rad)
of the test set’s phase detector are now determined for the specific center
frequency and power level of the sources being measured.
Measuring PLL suppression...The required correction data is created to
compensate for the phase noise suppression which occurs within the
bandwidth of the phase lock loop created for this measurement. The computer
displays the PLL suppression curve and associated measurement values.
Sweep segments
When the system begins measuring noise, it places the noise graph on its
display. As you watch the graph, you see the system plot its measurement
results in frequency segments.
The system measures the noise level across its frequency offset range by
averaging the noise within smaller frequency segments. This technique enables
the system to optimize measurement speed while providing you with the
measurement resolution needed for most test applications.
When the measurement is complete, refer to Chapter 14, “Evaluating Your
Measurement Results” for help in evaluating your measurement results. (If the
test system has problems completing the measurement, it informs you by a
message on the computer display.
Checking the beatnote
While the Connect Diagram is still displayed, Agilent recommends that you use
an oscilloscope (connected to the Monitor port on the test set) or a counter to
check the beatnote being created between the reference source and your DUT.
The objective of checking the beatnote is to ensure that the center frequencies
of the two sources are close enough in frequency to create a beatnote that is
within the Capture Range of the system.
The phase lock loop (PLL) Capture Range is 5% of the peak tuning range of the
VCO source you are using. (The peak tuning range for your VCO can be
estimated by multiplying the VCO tuning constant by the tune range of VCO.
90Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
Refer to Chapter 14, “Evaluating Your Measurement Results” if you are not
familiar with the relationship between the PLL capture range and the peak
tuning range of the VCO.)
NOTE
If the center frequencies of the sources are not close enough to create a beatnote within
the capture range, the system is not able to complete its measurement.
The beatnote frequency is set by the relative frequency difference between the
two sources. If you have two very accurate sources set at the same frequency,
the resulting beatnote is very close to 0 Hz.
Searching for the beatnote requires that you adjust the center frequency of one
of the sources above and below the frequency of the other source until the
beatnote appears on the oscilloscope’s display. See Figure 57.
If incrementing the frequency of one of the sources does not produce a
beatnote, you need to verify the presence of an output signal from each source
before proceeding.
0
V
E5505a_oscillo_disp_beatnote
25 Feb 04 rev 1
-1V/div
Figure 57Oscilloscope display of beatnote from test set monitor port
Agilent E5505A User’s Guid e91
4Expanding Your Measurement Experience
Making the measurement
1 Click the Continue button when you have completed the beatnote check
and are ready to make the measurement.
2 When the PLL Suppression Curve dialog box appears, select View
Measured Loop Suppression, View Smoothed Loop Suppression, and
View Adjusted Loop Suppress ion. See Figure 58 on page 92.
e5505a_user_select_suppression.ai
rev2 10/21/03
Figure 58Suppression selections
• There are four different curves for this graph. (For more information about
loop suppression verification, refer to Chapter 15, “Advanced Software
Features.”)
a “Measured” loop suppression curve—this is the result of the loop
suppression measurement performed by the E5505A system.
b “Smoothed” measured suppression curve—this is a curve-fit
representation of the measured results, it is used to compare with the
“theoretical” loop suppression.
c “Theoretical” suppression curve—this is the predicted loop suppression
based on the initial loop parameters defined/selected for this particular
measurement (kphi, kvco, loop bandwidth, filters, gain, and others).
d “Adjusted” theoretical suppression curve—this is the new “adjusted”
theoretical value of suppression for this measurement. It is based on
92Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
changing loop parameters (in the theoretical response) to match the
“smoothed” measured curve as closely as possible.
When the measurement is complete, refer to Chapter 14, “Evaluating Your
Measurement Results” for help in evaluating your measurement results.
Figure 59 on page 93 shows a typical phase noise curve for an RF Synthesizer.
E5505A_user_typ_noise_curve_8644B
24 Jun 04 rev 3
Figure 59Typical phase noise curve for an 8644B 10 MHz measurement.
Table 10 on page 94 contains the data stored in the parameter definitions file.
Agilent E5505A User’s Guid e93
4Expanding Your Measurement Experience
Table 10 Parameter data for the 8644B 10 MHz measurement
StepParametersData
1Type and Range Tab
• Measurement Type
• Start Frequency
• Stop Frequency
• Minimum Number of Averages
• FFT Quality
2Sources Tab
Carrier Source
• Frequency
• Power
• Carrier Source Output is connected to:
Detector Input
• Frequency
Reference Source
• Frequency
• Reference Source Power
VCO Tuning Parameters
• Nominal Tune Constant
• Tune Range ±
• Center Voltage
• Input Resistance
• Absolute Phase Noise (using a phase locked loop)
• 10 Hz
• 2 E + 6 Hz
• 4
• Fa st
• 10 E + 6 Hz
• 7 dBm
• Te st S e t
• 10 E +6 Hz
• 10 E +6 Hz (same as Carrier Source Frequency)
• 16 dBm
• 1 E +3 Hz/V
• ± 10 Volts
• 0 Volts
• 600 ohms
1
3Cal Tab
• Phase Detector Constant
• VCO Tune Constant
• Phase Lock Loop Suppression
• If Limit is exceeded
4Block Diagram Tab
• Carrier Source
• Downconverter
• Reference Source
• Timebase
• Phase Detector
• Test Set Tune Voltage Destination
• VCO Tune Mode
5Test Set Tab
• Input Attenuation
• LNA Low Pass Filter
• LNA Gain
• DC Block
• PLL Integrator Attenuation
• Measure Phase Detector Constant
• Calculate from expected VCO Tune Constant
• Verify calculated phase locked loop suppression
• Show Suppression Graph
• Manual
• None
• Agilent 8644B
• None
• Automatic Detector Selection
• Reference Source
• DC FM
• 0 dB
• 20 MHz (Auto checked)
• Auto Gain (Minimum Auto Gain –14 dB)
• Not checked
• 0 dBm
94Agilent E5505A User’s Guide
Expanding Your Measurement Experience4
Table 10 Parameter data for the 8644B 10 MHz measurement (continued)
StepParametersData
6Dowconverter TabThe downconverter parameters do not apply to this
measurement example.
7Graph Tab
• Title
• Graph Type
• X Scale Minimum
• X Scale Maximum
• Y Scale Minimum
• Y Scale Maximum
• Normalize trace data to a:
• Scale trace data to a new carrier frequency of:
• Shift trace data DOWN by:
• Trace Smoothing Amount
• Power present at input of DUT
• Confidence Test using Agilent 8644B Int vs Ext 10 MHz
• Single-sideband Noise (dBc/Hz)
• 10 Hz
• 4 E + 6 Hz
• 0 dBc/Hz
• - 170 dBc/Hz
• 1 Hz bandwidth
• 1 times the current carrier frequency
• 0 dB
• 0
• 0 dB
1 The stop frequency depends on the analyzers configured in your phase noise system.
Agilent E5505A User’s Guid e95
4Expanding Your Measurement Experience
Viewing Markers
The marker function allows you to display the exact frequency and amplitude
of any point on the results graph.
• To access the marker function, on the View menu, clic k Markers. See
Figure 60. In the dialog box containing Marker buttons, up to nine markers
may be added. To remove a highlighted marker, click the Delete button. See
Figure 61
Figure 60Navigate to markers
e5505a_user_add_delete_markers
24 Jun 04 rev 3
e5505a_user_nav_markers
24 Jun 04 rev 3
Figure 61Adding and deleting markers
96Agilent E5505A User’s Guide
Omitting Spurs
Expanding Your Measurement Experience4
The Omit Spurs function plots the currently loaded results without displaying
any spurs that may be present.
1 On the View menu, click Display Preferences. See Figure 62.
e5505a_user_nav_display_pref
24 Jun 04 rev 3
Figure 62Navigate to display preferences
2 In the Display Preferences dialog box, uncheck Spurs and click OK. See
Figure 63.
• The graph is displayed without spurs. See Figure 64 on page 98.
e5505a_user_uncheck_spurs.ai
rev2 10/21/03
Figure 63Uncheck spurs
3 To re-display the spurs, check Spurs in the Display Preferences dialog box.
Agilent E5505A User’s Guid e97
4Expanding Your Measurement Experience
e5505a_user_graph_without_spurs
24 Jun 04 rev 3
Figure 64Graph displayed without spurs
98Agilent E5505A User’s Guide
Displaying the Parameter Summary
The Parameter Summary function allows you to quickly review the
measurement parameter entries that were used for this measurement. The
parameter summary data is included when you print the graph.
1 On the View menu, click Parameter Summary. See Figure 65.
Expanding Your Measurement Experience4
e5505a_user_nav_param_sum
24 Jun 04 rev 3
Figure 65Navigate to parameter summary
2 The Parameter Summary Notepad dialog box appears. The data can be
printed or changed using standard Notepad functionality. See Figure 66 on
page 100.
Agilent E5505A User’s Guid e99
4Expanding Your Measurement Experience
Agilent 8644B Int vs Ext 10 MHz
Agilent 8644B; VCO tuned using DCFM.
Agilent N5502A
e5505a_user_text_param_sum
24 Jun 04 rev 3
Figure 66Parameter summary
100Agilent E5505A User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.