
ADDENDUM
to the
Agilent Model 6813B
User's Guide and Service Manual
for the
2000 VA Output Capability Option 019
Agilent P/N 5964-8149 Printed in U.S.A.
Microfiche No 5964-8150 May, 2000

2000 VA Output Capability
The 2000 VA Output Capability Option 019 is available on the following ac source model:
Agilent 6813B Option 019 AC Power Source (2000VA)
Option 019 increases the output VA capability of the standard Agilent 6813B AC Power Source from
1750 VA to 2000 VA. To achieve the increased output VA capability, the output rms voltage, dc voltage
and rms current ratings have increased slightly, to the maximum ratings documented in Table A-1. The
primary restriction that this increased output imposes on Option 019 units as compared to standard units,
is the reduction in the output frequecy range to between 45 and 100 Hz.
Specifications for Option 019
Table A-1 lists the specifications changes for units with Option 019. Performance specifications are
warranted over the ambient temperature range of 0 to 40 oC. Unless otherwise noted, specifications are
for a sinewave with a resistive load at a frequency range of 45 Hz to 100 Hz, after a 30-minute warmup.
Table A-1. Performance Specifications
Parameter Agilent 6813B Option 019
Phases:
Maximum Output Ratings Power (VA):
dc Power (Watts):
rms Voltage:
dc Voltage:
rms Current (in real-time mode):
dc Current:
Repetitive peak Current:
Non-repetitive peak Current
(inrush):
Crest Factor (current):
Output Frequency Range:
Programming Accuracy
(rms detection mode @ 25oC±5oC),
±(% of output+offset)
Measurement Accuracy
(@25oC ±5oC), ± (% of output +
offset)
rms Voltage (45−100 Hz):
Frequency:
dc Voltage:
rms Voltage (45−100 Hz):
Frequency:
dc Voltage:
(crest factor higher at lower
dc; 45 Hz−100 Hz
0.01% + 10 µHz
0.03% + 100 mV
0.01% + 0.01 Hz
0.03% + 150 mV
1
2000 VA
1500 W
308 V
± 436 V
16 A
10 A
48 A
48 A
3 @16 A
currents)
0.15% + 0.3 V
0.5% + 0.3 V
Table A-2. Supplemental Characteristics
Parameter Agilent 6813B (019)
ac Input Voltage Range (Vac):
Copyright 1999, 2000 AgilentTechnologies, Inc.
2
191−254 Vac
(230 Vac nominal)

Verification and Calibration for Option 019
There are no changes to the output verification and calibration sections of the User's Guide for units with
Option 019.
Option 019 Factory Modifications
Make the following changes to the Agilent 6813B Service Manual (p/n 5962-0859) for Agilent 6813B
Option 019 units.
Initialization
For ROM revision A.00.07, change step 3 to “Scroll to 6813J02 and press Enter”.
NOTE: Option J02 and Option 019 are identical; the J02 option was renamed option 019. After
an Agilent 6813B Option 019 Rev A.00.07 has been re-initialized, the number <1002>
will be returned by a *OPT? query.
For ROM revision A.00.08 or higher, change step 3 to “Scroll to 6813B19 and press Enter”.
NOTE: An Agilent 6813B Option 019 Rev A.00.08 will return <019> for a *OPT? Query.
Replaceable Parts
Add or change the following part numbers in the parts list:
Reference Desig. Part Number Description
A4 5064-0081 AC Input Assembly Tested
A4 C1060 - C1067 0180-4963 Capacitor 3300uF 200V
A9 5064-0083 Output/Relay/Filter Assembly Tested
A9 R552 – R553 0698-3156 Resistor 14.7K 1%
A10 5064-0079 Control Assembly Tested
5964-8149 Appendix Option 019
9320-5540 Label-Information
3