Agilent P/N 5964-8237 Printed in U.S.A.
Microfiche No 5964-8238November, 2000
Page 2
Contents
Contents2
Auxiliary Output Capability3
Rear Panel Connections4
Installation4
Front Panel Programming4
Additional [Source] Commands5
Additional Calibration Commands5
Behavior Differences of Existing SCPI Commands6
Auxiliary Ouput Specifications7
Calibration for Option 0267
This document contains proprietary information protected by copyright. All rights are reserved. No part
of this document may be photocopied, reproduced, or translated into another language without the prior
consent of Agilent Technologies. The information contained in this document is subject to change
without notice.
Copyright 2000 Agilent Technologies, Inc.
2
Page 3
Auxiliary Output Capability
The Agilent Technologies Auxiliary Output Option 026 is available on the following ac sources:
Agilent 6812B AC Power Source (750VA)
Agilent 6813B AC Power Source (1750VA)
Option 026 consists of an auxiliary 26 Vrms output located on the rear panel of the ac source. This is a
programmable output with a maximum output current of 100mA rms. The output frequency tracks the
frequency of the main output of the ac source. Option 026 adds the following capabilities to the above
listed ac source models:
•AC voltage and phase are programmable, in immediate mode only.
•Software calibration is available for both voltage and phase. Calibration is performed at a single
frequency (400 Hz). Note that there is no compensation for frequency-dependent voltage variation or
phase shift.
•The auxiliary output waveform and frequency are always the same as the main output.
•For a sine wave output, the maximum programmed rms voltage value is 26.02 V. For any waveform,
the programmed rms voltage value may not result in a peak voltage of more than 36.8 V.
•Current limit is not programmable.
•DC voltage is not available.
Front Panel Menus - Additions
The following menu commands have been added to let you program the Auxilliary Output. In most cases
you will need to press the Function key multiple times to access the menu item. Press the
RRRR
Entry key to access the parameter.
Voltage
Shift
RRRR
R
RPhase
RR
VOLT:AUX <value>
PHASE:AUX <value>
Set auxiliary output voltage
Set auxiliary output phase
SCPI Programming Commands - Additions
The following SCPI commands have been added to let you access the auxiliary output.
CALibrate[SOURce]
:VOLTage:VOLTage
:AUXiliary[:AC]:AUXiliary <NRf+>
:PHAse:PHASe
:AUXiliary:AUXiliary <NRf+>
:DATA?
3
Page 4
Rear Panel Connections
Auxiliary Output.
Connector is removable.
L 2( N )
HP-IB
L
I
E
N
A
T
R
I
G
N
N
I
F
T
L
++-
. . . .
R
I
W
A
N
HI LO
I
N
1
. . . .
_
1
O
O
V
0
0
3
TRIGGER
SENSE
_
C
1
A
M
C
A
1
T
U
O
_
_
_
M
C
O
O
M
_
X
T
O
_
H
_
_
G
N
S
R
2
3
2
N
O
S
O
P
I
N
U
T
I
O
C
A
R
N
A
W
G
N
I
T
1
L
Rear Panel, Overall View
Installation
Turn the unit off before connecting any wires.
The Auxiliary Output has a HI and a LO connection. Two pins are designated HI and the other two pins
are designated LO. The maximum isolation voltage to ground is 300 Vac (±425 Vdc).
IMPORTANT: You should always use both pins of the HI and the LO outputs. This promotes current
sharing among the output pins.
Front Panel Programming
Setting the Auxiliary Output
To set the auxiliary output voltage and phase:
ActionDisplay
1.
To set the auxiliary output voltage, press Voltage. Then scroll to the VOLT:AUX
command. Use the numeric keypad to enter a voltage and press Enter.
2.
To set the auxiliary output phase, press Shift and Phase. Then scroll to the
PHASE:AUX command. Use the numeric keypad to enter a phase and press Enter.
3.
To enable the auxiliary output, press Output On.
4
VOLT:AUX 10
PHASE: AUX 25
Page 5
Additional [Source] Commands
The following additional source commands are provided with the Auxiliary Output Option 026.
VOLTage:AUXiliary
This command programs the ac rms output voltage level of the auxiliary output. The auxiliary output
waveform and frequency are always the same as the main output. The auxiliary output is enabled along
with the main output.
The maximum peak voltage of the auxiliary output is 36.8 V peak. This includes any combination of
voltage and function shape values. Therefore, the maximum value that can be programmed depends on
the peak-to-rms ratio of the selected waveform. For a sinewave, the maximum voltage that can be
programmed is 26.02 V rms.
Command Syntax
Parameters
Unit
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
[SOURce:]VOLTage:AUXiliary[:LEVel][:IMMediate][:AMPLitude] <NRf+>
For sinewaves: 0 to 26.02 | MAXimum | MINimum
V (rms voltage)
1
This command sets the phase of the auxiliary output voltage waveform relative to an internal reference.
The phase angle is programmed in degrees. Positive phase angles are used to program the leading phase,
and negative phase angles are used to program the lagging phase.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
[SOURce:]PHASe:AUXiliary[:ADJust|:IMMediate] <NRf+>
–360 through +360 (degrees) | MAXimum | MINimum
0 (degrees)
The following SCPI commands are used to calibrate the Auxiliary Output Option 026:
CALibrate:VOLTage:AUXiliary
Initiates calibration of the auxiliary output voltage.
Command Syntax
Parameters
Examples
CALibrate:VOLTage:AUXiliary[:AC]
None
CAL:VOLT:AUX
5
Page 6
CALibrate:PHASe:AUXiliary
Initiates calibration of the auxiliary output phase.
Command Syntax
Parameters
Examples
CALibrate:PHASe:AUXiliary
None
CAL:PHAS:AUX
CALibrate:DATA?
Lets you read the present value of the phase calibration. This command can only be used during the
auxiliary phase calibration procedure. It will return an error if it is used at any other time.
Query Syntax
Parameters
Examples
CALibrate:DATA?
<NR3>
CAL:DATA?
Behavior Differences of Existing SCPI Commands
This section documents the behavior differences of existing SCPI commands with the Auxiliary Output
Option 026.
OUTput:STATe
This command control the on/off state of both the main and auxiliary outputs.
*SAV and *RCL
These commands include the auxiliary output voltage and phase.
*RST
Sets the auxiliary output voltage and phase to zero.
*OPT?
Returns 26.
*IDN?
Except for the firmware revision number, *IDN? returns the same response as the standard
6812B/6813B.
In addition to the new calibration commands shown above, auxiliary voltage and phase calibration utilize
the standard commands CALibrate:LEVel and CALibrate:DATA.
6
Page 7
Auxiliary Ouput Specifications
Table A-1 lists the specifications of the Agilent Auxiliary Output Option 026. Performance specifications
are warranted over the ambient temperature range of 0 to 40 oC. Unless otherwise noted, specifications
are for a sinewave with a resistive load at a frequency range of 400Hz.
Table A-1. Option 026 Performance Specifications
ParameterValue
Maximu m Output Ratingsrms Voltage:
rms Current:
Frequency:
1
0-26 V
100 mA
400 Hz
Output Ripple and Noisevoltage p-p:
Programming Accuracy
1
The frequency of the auxiliary output tracks the main output.
This section includes calibration procedures for the Auxiliary Output Option 026. Add these procedures
to those described in Appendix B of the AC Source User's Guide.
Because the output of the ac source must be enabled during calibration, proceed with caution, since
voltages and currents will be active at both the auxiliary and the main output terminals.
ImportantPerform the verification procedures before calibrating your ac source. If the ac source
passes the verification procedures, the unit is operating within its calibration limits and
does not need to be re-calibrated.
WARNINGLETHAL VOLTAGES. Ac sources can supply 424 V peak at their output. DEATH
on contact may result if the output terminals or circuits connected to the output are
touched when power is applied. These procedures must be performed by a qualified
electronics technician or engineer trained on this equipment.
7
Page 8
Equipment Required
The equipment listed in the following table, or the equivalent to this equipment, is required for
verification and calibration.
Table B-1. Equipment Required
EquipmentCharacteristicsRecommended Model
4-pin connector plug
Digital Voltmeter
Universal Counter
10:1 probes (2)
GPIB Controller
This connector is supplied with the ac source
when ordered with Option 026.
Resolution: 10 nV @ 1 V
Readout: 8.5 digits
Accuracy: better than 20 ppm
Full GPIB capabilitiesHP Series 200/300 or equivalent
Summary of Front Panel Calibration Menu
The Entry keypad is used for calibration functions.
CalShift
Press this key to access the calibration menu.
DisplayCommand Function
CAL ON <value>Turns calibration mode on when the correct password value
is entered.
CAL OFFTurns calibration mode off
CAL:LEV <char>Advance to next step in sequence (P1, P2, P3, or P4).
CAL:DATA <value>Input a calibration measurement.
SHUNT <value>Enter a value for the external current shunt in ohms.
CAL:VOLT:OFFSETBegin voltage offset calibration
CAL:VOLT:DCBegin dc volt age calibrati on sequence
CAL:VOLT:ACBegin ac voltage calibrati on sequence
CAL:VOLT:AUXBegin auxiliary voltage calibration
CAL:VOLT:PROTBegin voltage protection calibration
CAL:VOLT:EXTBegin external voltage calibration sequence
CAL:CURR:ACBegin ac current calibration sequence
CAL:CURR:MEASBegin current measurement calibration sequence
CAL:CURR:EXTBegin external current calibration sequence
CAL:IMPBegin output impedance calibration sequence
CAL:PHAS:AUXBegin auxiliary phase calibration
CAL:SAVESaves the calibration constants in non-volatile memory.
CAL:PASS <value>Set new calibration password.
1252-1488
Agilent 3458A
Agilent 53132A or equivalent
8
Page 9
Enable Calibration Mode
ActionDisplay
1.
Reset the unit by selecting Shift, Output, scrolling to *RST, and pressing Enter.
2.
Press Shift Calibration, enter the calibration password from Entry keypad, and
press Enter. If the password is correct, the Cal annunciator will come on. (If no
password has been set, just press Enter.)
If the password is incorrect, an error occurs. If the active password is lost, the
calibration function can be recovered by setting an internal switch that defeats
password protection (see the Service Manual.)
If CAL DENIED appears, then an internal switch has been set to prevent the
calibration from being changed (see the Service Manual.)
Calibrating the Auxiliary Ouptut Voltage
ActionDisplay
3.Disconnect all loads from the ac source. Connect the DVM across the output
terminals of the auxiliary output using the 4-pin connector provided. Make sure the
DVM is operating in AC volts mode.
4.
Press Shift Calibration, scroll to CAL VOLT AUX, and press Enter.
5.
Press Shift Calibration, scroll to CAL LEV P1 and press Enter.
6.
Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to enter
the ac voltage value displayed on the DVM. Then press Enter
7.
Press Shift Calibration, scroll to CAL SAVE, and press Enter.
*RST
CAL ON 0.0
OUT OF RANGE
CAL DENIED
CAL:VOLT:AUX
CAL:LEV P1
CAL:DATA 0.00
CAL:SAVE
8.
Press Shift Calibration, select CAL OFF, and press Enter to exit Calibration
mode. *RST and *RCL will also set the calibration state to OFF.
Calibrating the Auxiliary Output Phase
ActionDisplay
1.
Disconnect all loads from the ac source. Reset the ac source by selecting Shift,Output, scrolling to *RST, and pressing Enter. Make sure that the Dis annunciator
on the front panel is on.
2.Use a clip lead and connect the LO terminal of the Auxiliary output to the ACC
teminal of the Main output.
3.Program the Main output of the ac source to 26Vrms @ 400 Hz. Program the
Auxiliary output of the ac source to 26Vrms. Press Output On to enable the outputs.
4.Set the 53132A controls as follows:
Other Meas – Phase 1 to 2
Channel 1 & 2 Impedance: 1 Megohm
Channel 1 & 2 Coup l ing: AC
Channel 1 & 2 Attenuator:10X
5.Connect input 2 of the 53132A to the Auxiliary output with a 10:1 probe, positive
lead to Hi out and negative lead to Low out (common to main output low).
6.Connect input 1 of the 53132A to the Main output with a 10:1 probe, positive lead to
AC out and negative lead to ACC out (common to auxiliary output low).
CAL OFF
*RST
9
Page 10
Note: The 10:1 probes must be properly compensated for the 1 Megohm inputs to
ensure that probes do not contribute a phase offset error. Compensation may
be easily done by connecting both probes to the same AC source output and
adjusting the compensation for one probe to produce a reading of 0.0 degrees
on the 53132A counter.
7.Enable calibration mode as previously described.
8.
Press Shift Calibration, scroll to CAL PHAS AUX, and press Enter.
9.
Press Shift Calibration, scroll to CAL LEV P1 and press Enter.
10.Using the universal counter (or e quivalent instrument), monitor the phase of the
auxiliary output with respect to the main output.
11.
Press Shift Calibration, scroll to CAL DATA, but do NOT press the Enter key.
12.Using the ENTR Y arrow keys (é and ê ), adjust the calibration parameter until the
measured phase difference is zero or as close to zero as possible. The actual phase
at the instrument output changes immediately in response to the arrow keys.
(This is the only time that the arrow keys operate in this “live” mode).
13.
When the measured phase difference has been adjusted to zero, press Enter.
14.
Press Shift Calibration, scroll to CAL SAVE, and press Enter.
15.
Press Shift Calibration, select CAL OFF, and press Enter to exit Calibration
mode. *RST and *RCL will also set the calibration state to OFF.
CAL:PHAS:AUX
CAL:LEV P1
CAL:DATA 0.00
CAL:SAVE
CAL OFF
10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.