Agilent 6651A Data Sheet

This series of 500 W linear-regulated dc power supplies is designed to maximize the throughput of DUTs through the manufacturing test process with fast up and down programming time.
Valuable assemblies can be destroyed by a minor component failure that causes a surge of current to flow into the DUT. Fast protection features, including fast crowbar, mode crossover protection, and the ability to connect the protection circuitry of multiple power supplies can increase production yield.
Programming of the dc output and the protection features can be done either from the front panel or using industry standard SCPI commands, via the GPIB. Using the serial link, up to 16 power supplies can be connected through one GPIB address. Test system integration can be further simplified be using the VXIPlug&Play drivers. The out­put voltage and current can also be controlled with analog signals. This is helpful for certain types of noisy environments, and also immediate reactions to process changes.
Lab bench use is enhanced by the fan speed control, which helps to minimize the acoustic noise.
Agilent Performance dc Power Supplies speed and accuracy for test optimization
Specifications
(at 0˚to 55˚C unless otherwise specified)
6651A 6652A 6653A 6654A 6655A 6651A-
J01
Special Order Option
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
33
Increase test throughput with fast up and down programming time Protect valuable assemblies with fast protection features Proven reliability Low ripple and noise
6651A-6655A
Single-Output: 500 W GPIB
Number of outputs 111111 GPIB Yes Yes Yes Yes Yes Yes Output ratings
Output voltage 0 to 8 V 0 to 20 V 0 to 35 V 0 to 60 V 0 to 120 V 10 V Output current (40˚C) 0 to 50 A 0 to 25 A 0 to 15 A 0 to 9 A 0 to 4 A 50 A Maximum current (50˚C/55˚C)
45 A/42.5 A 22.5 A/21.3 A 13.5 A/12.8 A 8.1 A/7.7 A 3.6 A/3.4 A 45 A/42.5 A
Programming accuracy
at 25˚C ±5˚C Voltage 0.06% + 5 mV 10 mV 15 mV 26 mV 51 mV 6 mV Current 0.15% + 60 mA 25 mA 13 mA 8 mA 4 mA 60 mA
Ripple and noise
from 20 Hz to 20 MHz Voltage rms 300 µV 300 µV 400 µV 500 µV 700 µV 300 µV peak-peak 3 mV 3 mV 4 mV 5 mV 7 mV 3 mV Current rms 25 mA 10 mA 5 mA 3 mA 2 mA 2 5 mA Readback accuracy at 25˚C ±5˚C
(percent of reading plus fixed) System models only
Voltage 0.07% + 6 mV 15 mV 25 mV 40 mV 80 mV 7.5 mV +Current 0.15% + 67 mA 26 mA 15 mA 7 mA 3 mA 6 7 mA –Current 0.35% + 100 mA 44 mA 24 mA 15 mA 7 mA 100 mA
Load regulation
Voltage 1 mV 2 mV 3 mV 4 mV 5 mV 1 mV Current 2 mA 1 mA 0.5 mA 0.5 mA 0.5 mA 2 mA
Line regulation
Voltage 0.5 mV 0.5 mV 1 mV 1mV 2 mV 0.5 mV Current 2 mA 1 mA 0.75 mA 0.5 mA 0.5 mA 2 mA Transient response time Less than 100 µs for the output voltage to recover to its previous level
(within 0.1% of the voltage rating of the supply or 20 mV, whichever is greater) following any step change in load current of up to 50% of rated current
Supplemental Characteristics (Non-warranted characteristics determined by design and
useful in applying the product)
Average resolution
Voltage 2 mV 5 mV 10 mV 15 mV 30 mV 2.5 mV Current 15 mA 7 mA 4 mA 2.5 mA 1.25 mA 15 mA OVP 12 mV 30 mV 54 mV 93 mV 190 mV 16 mV OVP accuracy 160 mV 400 mV 700 mV 1.2 V 2.4 V 200 mV
34
Supplemental Characteristics for all model numbers
dc Floating Voltage: Output terminals can
be floated up to ±240 Vdc from chassis ground
Remote Sensing: Up to half the rated output voltage can be dropped in each load lead. The drop in the load leads subtracts from the voltage available for the load.
Command Processing Time: Average time required for the output voltage to begin to change following receipt of digital data is 20 ms for the power supplies connected directly to the GPIB
Output Programming Response Time:
The rise and fall time (10/90% and 90/10%) of the output voltage is less than 15 ms. The output voltage change settles within 1 LSB (0.025% x rated voltage) of final value in less than 60 ms.
Down Programming: An active down programmer sinks approximately 20% of the rated output current
Modulation: (Analog programming of output voltage and current)
Input signal: 0 to –5 V Input impedance: 10 k Ohm nominal
ac Input: (ac input frequency 47 to 63 Hz)
Voltage 100 Vac 120 V ac 220 Vac 240 Vac Current 12 A 10 A 5.7 A 5.3 A
Input Power: 1,380 VA, 1,100 W at full load;
120 W at no load GPIB Interface Capabilities: SH1, AH1, T6,
L4, SR1, RL1, PP0, DC1, DT1, E1, and C0. IEEE-488.2 and SCPI-compatible com­mand set.
Regulatory Compliance: Listed to UL 1244; conforms to IEC 61010-1.
Size: 425.5 mm W x 132.6 mm H x
497.8 mm D (16.75 in x 5.22 in x 19.6 in) See page 101 for more details
Weight: Net, 25 kg (54 lb); shipping, 28 kg (61 lb)
Warranty Period:One year
Single-Output: 500 W GPIB (Continued)
Specifications
(at 0˚to 55˚C unless otherwise specified)
6651A- 6651A- 6652A- 6653A- 6653A­J03 J09 J03 J04 J17
Special Order Special Order Special Order Special Order Special Order Option Option Option Option Option
Number of outputs 11111 GPIB Yes Yes Yes Yes Yes Output ratings
Output voltage 6 V 17V/20 V 27 V 40 V 30 V Output current (40˚C) 60 A 30 A/15 A 18.5 A 12.5 A 17.5 A Maximum current (50
˚
C/55˚C)
54 A/5 1A 27 A/25.5 A 16.65 A/15.72 A 11.25 A/10.6 A 15.75 A/14.87 A
13.5 A/12.75 A
Programming accuracy
at 25˚C ±5˚C
Voltage 0.06% + 5 mV 10 mV 13.5 mV 17.5 mV 15 mV Current 0.15% + 75 mA 36 mA 25 mA 13 mA 16 mA
Ripple and noise
from 20 Hz to 20 MHz Voltage rms 300 µV 300 µV 450 µV 1.6 mV 400 µV peak-peak 3 mV 4 mV 4.5 mV 5 mV 4 mV Current rms 30 mA 13 mA 10 mA 5 mA 6 mA Readback accuracy at 25˚C ±5˚C
(percent of reading plus fixed) System models only
Voltage 0.07% + 6 mV 15 mV 20.5 mV 30 mV 25 mV +Current 0.15% + 80 mA 40 mA 26 mA 15 mA 18 mA –Current 0.35% + 150 mA 55 mA 44 mA 24 mA 28 mA
Load regulation
Voltage 1 mV 2 mV 2 mV 3.5 mV 3 mV Current 6.5 mA 2 mA 1 mA 1 mA 0.5 mA
Line regulation
Voltage 0.5 mV 0.5 mV 0.5 mV 1 mV 1 mV Current 2 mA 2 mA 2 mA 0.75 mA 0.75 mA Transient response time Less than 100 µs for the output voltage to recover to its previous level
(within 0.1% of the voltage rating of the supply or 20 mV, whichever is greater) following any step change in load current of up to 50% of rated current
Supplemental Characteristics (Non-warranted characteristics determined by design and
useful in applying the product)
Average resolution
Voltage 2 mV 5 mV 6.75 mV 12mV 10 mV Current 18 mA 9 mA 7 mA 4 mA 5 mA OVP 12 mV 30 mV 30 mV 65 mV 54 mV OVP accuracy 160 mV 500 mV 400 mV 750 mV 700 mV
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
Agilent Performance dc Power Supplies speed and accuracy for test optimization