This series of linear-regulated
40-50 W DC power supplies is
designed to maximize the
throughput of DUTs through the
manufacturing test process with
fast programming and measurement,
and also active downprogramming.
It offers many advanced programmable features including stored
states and status reporting. Programming is done using industry
standard SCPI commands via the
GPIB or RS-232. Test system
integration is further simplified by
using the VXIPlug&Play drivers.
The optional relays simplify system
design and troubleshooting.
The half-rack size of the 6610A
series makes it a convenient
DC power supply for the R&D lab
bench. The built-in microamp
measurement system helps the
engineer to easily and accurately
monitor the output voltage and
current without a complicated
test setup.
Application Notes:
10 Practical Tips You Need to
Know About Your Power Products
5965-8239E
10 Hints for Using Your Power Supply
to Decrease Test Time
5968-6359E
Understanding Linear Power Supply
Operation (AN1554)
5989-2291EN
High Performance DC Power Supplies speed and accuracy for test optimization
More detailed specifications at
www.agilent.com/find/6610
34
Small, compact size for bench and system use
Fast, low-noise outputs
Dual-range, precision low current measurement
Built-in measurements and advanced programmable features
Protection features to ensure DUT safety
6611C - 6614C
Single-Output
40-50 W GPIB
Specifications
(at 0˚to 55˚C unless
otherwise specified)
6611C 6612C 6613C 6614C 6611C-
J05
Special Order
Option
Number of outputs 11111
GPIB Yes Yes Yes Yes Ye s
Output Ratings
Voltage 0 to 8 V 0 to 20 V 0 to 50 V 0 to 100 V 0 to 10 V
Current 0 to 5 A 0 to 2 A 0 to 1 A 0 to 0.5 A 0 to 5 A
Programming accuracy (at 25°C ±5°C)
Voltage 5 m V 10 m V 20 m V 50 m V 5 m V
+Current 0.05% + 2 m A 1 m A 0.75 m A 0.5 m A 2 m A
Ripple and noise 20 Hz to 20 MHz, with outputs
ungrounded or with either terminal grounded
Voltage rms 0.5 mV 0.5 mV 0.5 mV 0.5 mV 0.5 mV
peak -to-peak 3 mV 3 mV 4 mV 5 mV 3 mV
Normal mode rms 2 mA 1 mA 1 mA 1 mA 2 mA
DC measurement accuracy via GPIB or front-panel
meters with respect to actual output at 25°C ±5°C
Voltage 0.03% + 2 mV 3 mV 6 mV 12 mV 2 mV
Low current range -20 mA to + 20 mA 0.1% + 2.5 µA 2.5 µA 2.5 µA 2.5 µA 2.5 µA
High current range +20 mA to + rated 1 0.2% + 0.5 mA 0.25 mA 0.2 mA 0.1 mA 0.5 mA
-20 mA to - rated 1 0.2% + 1.1 mA 0.85 mA 0.8 mA 0.7 mA 1.1 mA
Load regulation
Voltage 2 mV 2 mV 4 mV 5 mV 2 mV
Current 1 mA 0.5 mA 0.5 mA 0.5 mA 1 mA
Line regulation
Voltage 0.5 mV 0.5 mV 1 mV 1 mV 0.5 mV
Current 0.5 mA 0.5 mA 0.25 mA 0.25 mA 0.5 mA
Transient response time Less than 100 µs for the output to recover to its previous level (within 0.1% of the voltage
rating of the supply or 20 mV, whichever is greater) following any step change in load current of up to 50% of the
output current rating of the supply
Supplemental Characteristics (Non-warranted characteristics determined by design and
useful in applying the product)
Average programming resolution
Voltage 2 mV 5 mV 12.5 mV 25 mV 3 mV
Current 1.25 mA 0.5 mA 0.25 mA 0.125 mA 1.25 mA
Sink current 3 A 1.2 A 0.6 A 0.3 A 3 A