Agilent 6611C Data Sheet

This series of linear-regulated 40-50 W dc power supplies is designed to maximize the throughput of DUTs through the manufacturing test process with fast programming and measurement, and also active downprogramming. It offers many advanced program­mable features including stored states and status reporting. Pro­gramming is done using industry standard SCPI commands via the GPIB or RS-232. Test system integration is further simplified by using the VXIPlug&Play drivers. The optional relays simplify system design and troubleshooting.
The half-rack size of the 6610A series makes it a convenient dc power supply for the R&D lab bench. The built-in microamp measurement system helps the engineer to easily and accurately monitor the output voltage and current without a complicated test setup.
Agilent Performance dc Power Supplies speed and accuracy for test optimization
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
25
Increase production throughput with fast programming speed
and fast downprogramming time Protect valuable prototypes with fast protection features Accurate and fast built-in measurement system
6611C - 6614C
Single-Output: 40-50 W GPIB
Specifications
(at 0˚to 55˚C unless otherwise specified)
6611C 6612C 6613C 6614C 6611C-
J05
Special Order Option
Number of outputs 11111 GPIB Yes Yes Yes Yes Yes Output Ratings
Voltage 0 to 8 V 0 to 20 V 0 to 50 V 0 to 100 V 0 to 10 V Current 0 to 5 A 0 to 2 A 0 to 1 A 0 to 0.5 A 0 to 5 A Programming accuracy(at 25°C ±5°C) Voltage 5 m V 10 m V 20 m V 50 m V 5 m V +Current 0.05% + 2 m A 1 m A 0.75 m A 0.5 m A 2 m A Ripple and noise 20 Hz to 20 MHz, with outputs
ungrounded or with either terminal grounded Voltage rms 0.5 mV 0.5 mV 0.5 mV 0.5 mV 0.5 mV
peak -to-peak 3 mV 3 mV 4 mV 5 mV 3 mV Normal mode rms 2 mA 1 mA 1 mA 1 mA 2 mA dc measurement accuracyvia GPIB or front-panel
meters respect to actual outputat 25°C ±5°C Voltage 0.03% + 2 mV 3 mV 6 mV 12 mV 2 mV Low current range -20 mA to + 20 mA 0.1% + 2.5 µA 2.5 µA 2.5 µA 2.5 µA 2.5 µA High current range +20 mA to + rated 1 0.2% + 0.5 mA 0.25 mA 0.2 mA 0.1 mA 0.5 mA
-20 mA to - rated 1 0.2% + 1.1 mA 0.85 mA 0.8 mA 0.7 mA 1.1 mA
Load regulation
Voltage 2 mV 2 mV 4 mV 5 mV 2 mV Current 1 mA 0.5 mA 0.5 mA 0.5 mA 1 mA
Line regulation
Voltage 0.5 mV 0.5 mV 1 mV 1 mV 0.5 mV Current 0.5 mA 0.5 mA 0.25 mA 0.25 mA 0.5 mA Transient response timeLess than 100 µs for the output to recover to its previous level (within 0.1% of the voltage
rating of the supply or 20 mV, whichever is greater) following any step change in load current of up to 50% of the output current rating of the supply
Supplemental Characteristics (Non-warranted characteristics determined by design and
useful in applying the product)
Average programming resolution
Voltage 2 mV 5 mV 12.5 mV 25 mV 3 mV Current 1.25 mA 0.5 mA 0.25 mA 0.125 mA 1.25 mA Sink current 3 A 1.2 A 0.6 A 0.3 A 3 A
26
Supplemental Characteristics for all model numbers
dc Floating Voltage: Output terminals
can be floated up to ±240 Vdc maximum from chassis ground
Remote Sensing: Up to two volts dropped in each load lead. Add 2 mV to the voltage load regulation specification for each one volt change in the postive output lead due to load current change.
Command Processing Time: Average time required for the output voltage to begin to change following receipt of digital date is 4 ms for the power supplies connected directly to the GPIB.
Output Programming Response Time: The rise and fall time (10/90% and 90/10%) of the output voltage is less than 2 ms. The output voltage change settles within 1 LSB (0.025% x rated voltage) of final value in less than 6 ms.
GPIB Interface Capabilities: IEEE-488.2, SCPI command set, and 6630A Series programming compatability
Input Power: (full load): 1.6 A, 100 W (6611C: 2.2 A, 120 W)
Regulatory Compliance: Complies with EMC directive 89/336/EEC (ISM 1B).
Warranty Period: Three years Size: 212.8 mm W x 88.1 mm H x
368.3 mm D (8.4 in x 3.5 in x 14.5 in) See page 102 for more details
Weight: 8.2 kg (18.16 lb) net;
10.6 kg (23.5 lb) shipping
Single-Output: 40-50 W GPIB (Continued)
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
Agilent Performance dc Power Supplies speed and accuracy for test optimization
Ordering Information
Opt 100 87 to 106 Vac, 47 to 63 Hz Opt 120 104 to 127 Vac, 47 to 63 Hz Opt 220 191 to 233 Vac, 47 to 63 Hz Opt 230 207 to 253 Vac, 47 to 63 Hz Opt 760 Isolation and Reversal relays
* Opt ICM Rack-mount Kit
(p/n 5063-9240)
* Opt AXS Rack-mount Kit
side-by-side mounting of two units, Lock-link Kit p/n 5061-9694; Flange Kit p/n 5062-3974
Opt 0L2 Additional standard documentation package
Opt 0B3 Service Manual
*Support rails required
Accessories
Rack-mount and slide for two side-by-side units of different lengths p/n 1494-0015, 5063-9255 and filler panel 5002-3999
Rack-mount slide and support for one instrument p/n 1494-0015, 5063-9255 and filler panel 5002-3999
E3663AC Support rails for Agilent rack cabinets
Loading...