Agilent 6400 Quick Start Guide

Agilent 6400 Series Triple Quadrupole LC/MS System
Quick Start Guide
Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3. Prepare the Triple Quad instrument 17 Step 4. Set up and run an acquisition method 21 Step 5. Review results with the Qualitative Analysis program 29 Step 6. Set up and run quantitation 29
Use this guide for your first steps with the Agilent 6400 Series Triple Quad LC/MS, and as a road map for your user information.

What is the Agilent 6400 Series Triple Quad LC/MS system?

The Agilent Triple Quadrupole LC/MS is a liquid chromatograph triple quadrupole mass spectrometer that performs MS/MS using three sets of parallel rods (in this case, quadrupole, hexapole, quadrupole). The first quadrupole separates ions into precursor ions that are fragmented in the hexapole into product ions, which are separated by the second quadrupole. Often, two or more precursor ions and their product ions are monitored in sequence in MRM (multiple reaction monitoring) mode. You can monitor up to 4000 MRM transitions by using Dynamic MRM.
The Agilent 6490 Triple Quadrupole contains the iFunnel Technology which provides more efficient sampling of ions into the mass spectrometer. The iFunnel Technology includes the Agilent Jet Stream Technology, shorter desolvation assembly with Hexabore Capillary, and Dual Offset Ion Funnel.

What’s New in Data Acquisition

The Agilent 6490 and 6460 Triple Quad LC/MS are shipped by default with the Agilent Jet Stream Technology that utilizes a super-heated sheath gas to collimate the nebulizer spray which dramatically increases the number of ions that enter the mass spectrometer.
You can set up an Agilent Triple Quad LC/MS with the Agilent 1100, 1200, 1260 or 1290 Infinity LC modules.
Also, the Agilent Triple Quad LC/MS comes with Agilent MassHunter Workstation Software that includes three major programs:
Data Acquisition – From one screen you can tune the mass spectrometer, control and monitor instrument parameters, set up acquisition methods and worklists containing multiple samples and monitor real-time run plots.
Quantitative Analysis – From one screen you can set up a batch of data files and quantify, evaluate and requantify the results. From this screen you have access to the Method Editor for setting up and editing the quantitation methods.
Qualitative Analysis – From one screen you can extract and integrate chromatograms, subtract background, extract peak spectra, and compare data from different types of data files.
What’s New in Data Acquisition
The Data Acquisition program for the 6400 Series Triple Quadrupole has many new features in this revision.
in B.06.00
The supported models are 6410B, 6420A, 6430A, 6460A, and 6490A.
Fast Scan is supported on all models.
The Agilent G7100A Capillary Electrophoresis system is supported.
The study types Worklist-Only and Worklist Import both support creating a
Quantitative Analysis method.
The Autotune program has the ability to tune only the unit mass.
The Autotune program can report m/z below 100 for positive and negative
autotunes.
You can adjust the EMV separately in the Manual Tune > Detector tab.
The Tune report contains the instrument serial number and the firmware
revision.
Triggered MRM databases are supported in Optimizer.
Additional options for TMRM and compound naming are supported when
searching, filtering and importing to an acquisition method.
2 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
What’s New in Data Acquisition
The Source and iFunnel Optimizer program is available to help optimize some source and ion funnel parameters.
In Triggered MRM you can select two different triggers for each compound.
In SIM, MRM, and DMRM methods you can specify a Compound Group.
You can specify a trigger entrance delay, a trigger time delay and a trigger
window to allow you to acquire triggered scans closer to the apex of the peak.
Polarity switching is supported for TMRM. Different compounds can have different polarities.
If the method is a Triggered MRM method, then you can enable or disable triggering for all compounds in a method for a specified sample type. For example, if the Sample Type is Calibration, all of the primary transitions are acquired, but the secondary transitions are never triggered.
You can update the threshold value in a method using a Dynamic MRM or Triggered MRM data file.
When you switch to Triggered Dynamic MRM, you can automatically set the default trigger to the transition with the highest m/z (the product ion).
When you switch from a Triggered Dynamic MRM run to a Dynamic MRM run, the secondary MRMs are grayed out and are not acquired during the run.
The Update MRM tool can also fill in the Trigger Window values from a data file or a Quantitation report. A variety of options are available.
You can create an MS Indexed data file which supports faster operation in the Quantitative Analysis program.
The user interface of the Instrument Status window has been updated.
The user interface of the DMRM Viewer has been updated.
The DMRM Viewer supports tMRM methods.
The DMRM Viewer has added the ability to filter on primaries only or on all
transitions.
The DMRM Viewer has added Compound and Compound Group navigation.
In the Study Manager, the Worklist-Only study and the Worklist Import study
include a page to enable quantitation and report generation.
You can select a post-analysis script that will run after every method in a worklist; it runs after the Data Analysis portion of the method is complete.
The 1290 Quaternary Pump (G4204A) is supported.
The FlexCube (G4227A) is supported in standalone mode.
UIB 2 (G1390B) is supported.
For a list of updates made to previous versions, refer to the online Help.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 3

Where to find information

Help

Where to find information
Help
Press F1 To get more information about a pane, window or dialog box, place the
cursor on the part of the pane, window or dialog box of interest and press F1.
Help menu From the Help menu, access “How-to” help and reference help.

Documents

You can find these manuals delivered with the Triple Quad hardware or software. You can also find a PDF version on the installation disk, in the Manuals folder. Many of these manuals are also available on the www.chem.agilent.com
Installation Guide This guide is used to install the hardware and software,
configure the instrument, and verify performance.
Upgrade Guide This guide is used to upgrade your MassHunter Workstation
software from a previous version.
web site.
Maintenance Guide Use this guide to help maintain and troubleshoot your
Agilent Triple Quad LC/MS.
Concepts Guide - The Big Picture Learn the background information to help you
understand operation of the hardware and software.
Data Acquisition for 6400 Series Triple Quad Familiarization Guide Do the
exercises to learn to use the Triple Quad LC/MS hardware and Data Acquisition program for 6400 Series Triple Quad.
Study Manager Quick Start Guide Use this guide to learn to use the MassHunter
Study Manager software.
MassHunter Optimizer Software Quick Start Guide Use this guide to learn about
the MassHunter Optimizer program. The MassHunter Optimizer program provides a way to automatically optimize the data acquisition parameters for MRM mode (multiple-reaction monitoring) on a Triple Quad instrument for each individual compound analyzed.
4 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide

Training

Where to find information
Training
MassHunter QQQ Compliance Quick Start Guide Use this guide to learn about the
MassHunter QQQ Compliance program.
MassHunter Quant Compliance Quick Start Guide Use this guide to learn about
the MassHunter Quant Compliance program. (This guide is on the Quantitative Analysis disk.)
Qualitative Analysis Familiarization Guide Do the exercises to learn to use the
Qualitative Analysis program. (This guide is on the Qualitative Analysis disk.)
For information on what is new in the Qualitative Analysis program, refer to the Qualitative Analysis Familiarization Guide.
Quantitative Analysis Familiarization Guide Do the exercises to learn to use the
Quantitative Analysis program. (This guide is on the Quantitative Analysis disk.)
For information on what is new in the Quantitative Analysis program, refer to the Quantitative Analysis Familiarization Guide.
Familiarization Guide Use all three familiarization guides to get to know the
software.
Quick Start Guide Use the quick start guides for Study Manager, Optimizer and
Compliance to get to know these programs.
Training Courses Visit www.chem.agilent.com to view a listing of training courses
for the Agilent Triple Quad LC/MS.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 5

Getting Started

Start Here!
Set up, run
and analyze
samples
3 Set up and
run an acquisition method
1 Start the
Data Acquisition software
2 Prepare the
LC/MS
4 Review results
with Qualitative Analysis
5 Set up and run
quantitation
Getting Started
Install the Triple Quadrupole LC/MS hardware and software
Use the Installation Guide to install the hardware and software and verify performance.
Set up, run and analyze samples
The roadmap below shows you the steps to set up and run a batch of samples from start to finish. Follow the instructions on the next pages to get started and to learn where to find the information to help you with each step in this roadmap.
6 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide

Step 1. Start the Data Acquisition software

NOTE
Step 1. Start the Data Acquisition software
The instructions below include the following assumptions:
The hardware and software are installed.
The instrument is configured.
Use instructions in the Installation Guide to configure the instrument for the first time and any time you change the LC configuration.
The LC modules and the Triple Quad MS are turned on, but the LC pump is not running.
After installation, you see all of the Agilent MassHunter Workstation Software icons on your Desktop.To start the Data Acquisition program, double-click the Data Acquisition icon.
The Data Acquisition window appears.
Getting Started
When Data Acquisition opens, the software engines automatically start. If you need to restart them, right-click the Acq System Launcher icon in the system tray, and click Start Engines.
If you have recently changed LC modules, remember to configure the instrument again. See the Installation Guide for instructions.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 7
Getting Started
Instrument Status window
Chromatogram Plot window
Method Editor window
Spectrum Plot window
Actuals window
The Sample Run and Worklist windows are tabbed here. These three windows are “sharing” this space. You click the tab to switch to a different window.
Step 1. Start the Data Acquisition software
Windows—where you do most of your work
When you first start the Data Acquisition program, the main window appears. You do almost all of your work within the eight windows of this main window. These windows provide the tools to set up acquisition methods, run samples interactively or automatically, monitor instrument status, monitor runs and tune the instrument.
Figure 1 Main window of the Data Acquisition software program
8 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
Getting Started
Step 1. Start the Data Acquisition software
Show/hide the windows You can show one window at a time on the screen or up
to seven windows. You can never hide all of the windows. To show or hide a window, you click the commands in the View menu. You can also hide a window by clicking the X icon in the upper right corner of the window.
When you click a window, the title of the active window changes to a different color. Press F1 to obtain help on the active window. You can also drag a window border to resize the window. If you double-click the title of the window, the window “floats” outside of the main window. You can double-click the title bar again to “dock” the window. You can also float and dock the window by right-clicking the title of the window and clicking Floating.
Instrument Status window With this window you view the status of each device
configured with the instrument- Error, Not ready, Pre-run, Post-r un, Running, Injecting, Idle, Offline, or Standby. You also set non-method control and
configuration parameters for the LC devices and the MS instrument.
This window displays each device's current status both as text and by its color-coding:
Color Coding in the Instrument Status Window
Color Status
Red Error
Yellow Not ready
Purple Pre-run, post-run, Waiting
Blue Running, injecting
Green Idle
Dark gray Offline
Light gray Standby (for example, lamps off)
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 9
Getting Started
Step 1. Start the Data Acquisition software
Actuals window With this window you view the current value of selected
instrument parameters.
Chromatogram Plot window With this window you monitor the chromatogram
plots in real time. These plots can be user-defined signals and/or instrument parameters.
Spectrum Plot window With this window you monitor the spectral plot in real
time.
10 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
Getting Started
Step 1. Start the Data Acquisition software
Method Editor window With this window you enter acquisition parameters for
the method. If you click the button, then you can see the tune values in the Tune Parameters dialog box.
Sample Run window With this window you enter sample information to run
individual samples interactively, and you can start a single sample run. You can also specify an Override DA Method and select either Both Acquisition and DA or DA Only for the Method Type, and then Data Analysis is run as part of the method.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 11
Getting Started
Step 1. Start the Data Acquisition software
Worklist window With this window you enter sample information for multiple
samples. When you run the worklist, the samples are automatically run in the order listed in the worklist. You can select whether to run Acquisition Only, to run Both Acquisition and DA, or to run DA only by selecting one of these options for the Part of method to run in the Worklist Run Parameters dialog box.
Tune window With this window you tune the mass spectrometer. You can use the
automatic tuning algorithms that are provided, or you can manually tune the instrument.
12 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide

Step 2. Prepare the LC modules

Read and follow the instructions in the online Help for each of the tasks in the checklist described on the following pages.
1 Switch LC stream to Was t e .
While you condition or equilibrate the column, you can tune the Triple Quad MS. During this time you do not want pump effluent streaming into the Triple Quad MS, so you switch the direction of the LC stream away from the MS ion source and to waste.
If you have the LC connected to a VWD or DAD, you can still monitor the fluctuations of the VWD or DAD real-time chromatogram before a run.
a Right-click the QQQ device in the Instrument Status window.
Getting Started
Step 2. Prepare the LC modules
b Click LC > Waste.
2 Purge the LC pump.
Follow the directions for purging the pump in the User Guide for your pump.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 13
Getting Started
Step 2. Prepare the LC modules
3
Set up to condition or equilibrate the column.
a Type LC parameters, and click the Apply button in the toolbar to download
them to the LC.
b Right-click an LC module in the Instrument Status window to change any
non-method control parameters, if necessary.
c Monitor the baseline and adjust the plot to make sure the column is
equilibrated and the baseline stable. (See step 4 and step 5 on page 15.)
14 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
Getting Started
Step 2. Prepare the LC modules
4
Set up to view real-time parameter values (actuals).
As you prepare for a run and during a run, you want to see the actual values of the instrument parameters. You can do this in the Instrument Status window.
a Right-click the Actuals list to see the Setup command.
b Click Setup to bring up the list of Actuals available for monitoring.
c Add all the parameter values you intend to monitor, and click the OK
button. You can set the background and text color for each parameter. You can also set a range for the parameters which are numbers. If the value of the parameter is not within the limits which you entered, then the background of the parameter is set to red.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 15
Getting Started
Step 2. Prepare the LC modules
5
Set up real-time plot displays.
As you condition the column, you set up the displays to monitor the effluent.
Right-click the chromatogram plot, and click Change.
In the Edit Signal Plot dialog box, you can select the type of display signal.
16 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide

Step 3. Prepare the Triple Quad instrument

Step 3. Prepare the Triple Quad instrument
Do a Checktune, Autotune or Manual Tune
1 From the Context list, click Tune.
You can see the Instrument Status window, the Actuals window and the Tune window when you switch to the Tune context. Click Tune in the View menu if the Tune window is not visible.
Getting Started
2 Click Checktune to evaluate if the MS parameters are within the limits to
produce the specified mass accuracy and resolution. Checktune takes up to 15 minutes to run.
If the current tune file was last tuned with the Fast Scan Autotune, then the Checktune algorithm only checks MS2. If the current tune file was last tuned with Autotune, then the Checktune algorithm checks both MS1 and MS2.
If the current tune file was last tuned with the Fast Scan Autotune, then the Fast Scan check box is marked on the Manual Tune tab.
Do a Checktune regularly.
If Checktune passes, then skip to step 5.
If Checktune fails, then you can try using the Manual Tune tab to fix the problem.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 17
Getting Started
Step 3. Prepare the Triple Quad instrument
3
Try the following quick changes to get Checktune to pass.
a Click the Manual Tune tab.
b Click the MS1 tab and then click the Adjust Gain & Offset button.
c Click the MS2 tab and then click the Adjust Gain & Offset button.
d Click the Autotune tab.
e Click the Checktune button.
If Checktune fails again, you need to do an Autotune, which is described next.
4 Click Autotune to tune the MS automatically (15 to 20 minutes for all models
except the 6490. The 6490 takes approximately 45 minutes per ionization mode.). The system automatically changes different tune parameters to tune the MS. You only do an Autotune when it is necessary.
You only mark t h e Start from Factory Defaults check box after an instrument shutdown and restart. Then, if you click Autotune, the instrument is tuned starting from the factory defaults (20 to 25 minutes for all models except the
6490. The 6490 takes approximately 45 minutes per ionization mode.).
Checktune and Autotune reports are automatically generated after Autotune completes successfully. If Autotune fails, no reports are printed.
If Autotune fails or you assess that the Triple Quadrupole MS needs custom values entered for its tune parameters, you can manually tune the instrument.
18 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
Getting Started
Step 3. Prepare the Triple Quad instrument
5
(optional) Click the Fast Scan Autotune Only button if you want to tune for Fast Scan. You only do this step if the Autotune results are acceptable.
Fast Scan acquisition is only supported on the 6490 with two of the scan types: MS2 Scan and Product Ion scan. Fast Scan autotune only tunes with the Wide (1.2) resolution, Samples set to 1 (on the Manual Tune > Acquisition Tab), and the step size must be 0.2. The scan time in the Acquisition tab is generated automatically based on the step size and the mass range.
6 From the Context list, click Acquisition.
Switch LC stream to MS
After you condition the column and tune the Triple Quad MS, you switch the LC stream from Waste to MS. See “Switch LC stream to Waste.” on page 13 for how to do this.
Monitor MS baseline and spectral displays
If you did not monitor the LC baseline with a VWD or DAD, make sure that the Triple Quadrupole baseline is stable and no spectra of interfering intensity appear in the display.
If you did monitor the LC baseline with a VWD or DAD, change back to the default Triple Quad displays.
1 Right-click the chromatogram plot, and click Change. 2 Select the MS signal, and click OK.
View the system logbook for events and errors
As you prepare the instrument, you may run into an error that you want to troubleshoot. You do this through the System Logbook Viewer.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 19
Getting Started
Step 3. Prepare the Triple Quad instrument
Click the Log icon ( ) in the toolbar of the Data Acquisition window, and view the logged events.
Or click Tools > System Logbook Viewer.
Or right-click the icon in the system taskbar. First, click Enable Notification. Then, right-click the LOG icon and click Configure. The system
can notify you of new errors and warning by showing messages from the taskbar.
When the System Logbook Viewer is open, you can select the time period to keep System Logbook entries. You can set the value from 1 week to 1 year. To do this, you click Tools > Purge Settings.
20 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide

Step 4. Set up and run an acquisition method

NOTE
Step 4. Set up and run an acquisition method
Read and follow the instructions in the online Help for each of the tasks described on the following pages.
Also, do Exercise 1 of the Data Acquisition Familiarization Guide to learn how to set up and run an acquisition method.
1 Set up the method:
a Type the values and settings for each of the tabs below.
b (optional) If you want to download the settings to the instrument, click
Apply.
c To save the method, click Method > Save As.
d Name the method and click OK.
2 Enter values for all of the LC modules configured for the instrument.
Getting Started
Make sure when you type the MS parameters on the next page that the tune file is the one that you want to use with the acquisition.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 21
Getting Started
Step 4. Set up and run an acquisition method
3
Enter the Triple Quadrupole parameter values.
a Select the Scan Type from the list in the Time segments table. The Scan
segments table is cleared when you change the Scan Type. The parameters available on the right change depending on the Scan Type.
If you are changing the Scan Type from MRM to Dynamic MRM or to Triggered MRM, you can copy and paste the transitions from the Clipboard to the new Scan segments table. See the online Help for more information.
b Type in any Acquisition values you want to change. You can enter multiple
Scan segments.
You cannot set the fragmentor voltage in Acquisition if the instrument type is an Agilent 6490. It always uses the value in the tune file.
4 Set up to change Triple Quad MS parameters with segments and scans:
a To add a segment, right-click anywhere in the Scan segments section to
bring up the Scan Segments shortcut menu, and click Add Row.
b Type the parameters for each Scan segment.
22 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
Getting Started
Step 4. Set up and run an acquisition method
5
Set up signals for the Chromatogram plot:
a Click the Chromatogram tab.
b Select the Chromatogram Type, and type other plot values.
6 Set up the Stored instrument curves in the Instrument tab. In the Qualitative
Analysis program, you can display these values in the MS Actuals window for the current spectrum. With the Triple Quadrupole, the values in the MS Actuals window in the Qualitative Analysis program are the values that you save in the Instrument tab.
a Click the Instrument tab.
b Select the Stored instrument curves. These values will be shown in the MS
Actuals window in the Qualitative Analysis program for each spectrum.
7 Set up the data analysis (DA) parameters.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 23
Getting Started
Step 4. Set up and run an acquisition method
A method can contain qualitative analysis parameters, quantitative analysis parameters or both. A Data Analysis method is a method that contains data acquisition parameters with either the Qual Automation check box marked on the Qual tab or the Quant Automation check box marked on the Quant tab.
a Click the DA tab.
b (optional) Mark the Qual Automation check box. The name of the current
Qualitative Analysis method is shown in the list. If you want to change the Qualitative Analysis method that is connected, click the “...” button to select a different method. When the Data Acquisition method is saved, the Qualitative Analysis method that you selected is copied or linked to the Data Acquisition method.
c (optional) Click the Quant tab. Mark the Quant Automation check box. The
name of the current Quantitative Analysis method is shown in the list. If you want to change the Quantitative Analysis method that is connected, click the “...” button to select a different method. When the Data Acquisition method is saved, the Quantitative Analysis method that you selected is copied or linked to the Data Acquisition method.
d (optional) Mark the Generate Report check box on the Quant tab. Then,
you select the Template F ile to use. If you want to print the report, mark the Print to default printer check box. You can also mark the Publish check box to create a CSV file, a TXT file, or a PDF file.
24 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
Getting Started
Step 4. Set up and run an acquisition method
8
Save the method.
a Click Method > Save As or Method > Save.
b If necessary, name the method and click OK.
9 Set up and run interactive samples:
a Click the Sample Run window. By default, it is tabbed with the Worklist
and Method Editor windows.
b Enter the information such as the Sample Name, the Data File Name and
the Path.
c Enter the Additional Information. You can change the value of the
parameters in the Additional Information list.
You can run a Data Analysis method from this window by selecting Both Acquisition and DA or DA Only for the Method Type. In addition, you have to set Override DA method to indicate which Data Acquisition method contains the DA (Data Analysis) method to execute. You always have to do this.
d To start the single sample run, click the Run button ( ) in the Sample
Run window or the Run button ( ) in the main toolbar.
You can run the single sample in either locked or unlocked mode. When the mode is locked, no one can change the method or sample parameters during a run. You also cannot overwrite this data file in the Data Acquisition program. The Lock button ( ) in the main toolbar indicates that locked mode is on.
10 Set up and run worklists
a Click the Worklist tab to show the Worklist window. If the Worklist window
is not visible, click View > Worklist.
b Right-click the upper left corner of the worklist.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 25
Getting Started
Step 4. Set up and run an acquisition method
c
Click Add Multiple Samples. The Add Multiple Samples dialog box opens.
d Enter all the information on the Sample Information tab.
e Click the Sample Position tab to specify the sample vial locations (make
sure the specific sample tray type has been configured by right-clicking the autosampler device image).
26 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide
Getting Started
Step 4. Set up and run an acquisition method
f
Specify the locations, and click the OK button.
g To set up the worklist run, right-click the upper left corner of the worklist,
and click Worklist Run Parameters.
h Click the Page 1 tab.
i Type the paths for the method and data files.
j Click the Page 2 tab.
k Review the information and click the OK button.
l To start the run, click the Run Worklist button ( ) in the main toolbar
or the Run button ( ) in the Worklist window
You can run the worklist in either locked or unlocked mode. When the mode is locked, no one can change the method or the worklist while the worklist is running.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 27
Getting Started
NOTE
Step 4. Set up and run an acquisition method
To use an acquisition method that has a different DA method than the method entered in the worklist, show the column called Override DA Method in the worklist by using the Show/Hide/Order Columns dialog box. In this column, type the name of another method containing the DA parameters you want to use for the sample. The DA part of this method is used instead of the DA part of the current method.
You can also type the name of this method in the Add Multiple Samples dialog box.
28 Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide

Step 5. Review results with the Qualitative Analysis program

Step 5. Review results with the Qualitative Analysis program
Use the Qualitative Analysis program to do these tasks and more:
Review results for acquisition method development
Select the most appropriate precursor and product ions for MRM analyses
Find compounds
Identify compounds
Do molecular feature extraction
Do the exercises in the Qualitative Analysis Familiarization Guide to help you learn how to use the Qualitative Analysis program.
Do Exercise 1 of the Data Acquisition for Triple Quad Familiarization Guide to help you learn how to use the Qualitative Analysis program to develop acquisition methods.
Also, refer to the online Help for the Qualitative Analysis program to learn how to do more operations to analyze your data.

Step 6. Set up and run quantitation

Another primary tool for analyzing and reporting Triple Quad LC/MS results is the Quantitative Analysis program.
Do the exercises in the Quantitative Analysis Familiarization Guide to learn how to do these operations to quantitate the acquired data files:
Set up a batch and a method to automatically quantitate a set of samples
Review results by learning how to view and use the Batch-at-a-Glance results
screen
Identify and use outliers to change the method and requantitate the data
using a better calibration curve fit or other more appropriate settings
Getting Started
Also, refer to the online Help for the Quantitative Analysis program to learn how to do more operations to analyze your data.
Agilent 6400 Series Triple Quadrupole LC/MS Quick Start Guide 29
www.agilent.com
In this book
This book contains brief instructions to help you get started with your Agilent Triple Quadrupole LC/MS. This book takes a quick look at using the Data Acquisition program to:
Prepare the instrument for a run.
Set up acquisition methods.
Set up and run worklists.
Microsoft®, Windows®, Windows 7® and Excel® are registered trademarks of Microsoft Corporation in the US and/or other countries.
© Agilent Technologies, Inc. 2012
Printed in USA Revision A, November 2012
*G3335-90137*
G3335-90137
Loading...