Agilent 6400 Concepts Guide

Agilent 6400 Series Triple Quadrupole LC/MS System
Concepts Guide
The Big Picture
Agilent Technologies
Notices
CAUTION
WARNING
No p art o f this manu al may be re produce d in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
G3335-90135
Edition
Revision A, November 2012
Printed in USA
Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051
Microsoft® and Windows® are U.S. regis­tered trademark s of Microsoft Corporation.
Software Revision
This guide applies to the Agilent MassHunter Workstation Software -- Data Acquisition for 6400 Series Triple Quadru­pole program version B.06.00 or higher until superseded.
If you have comments about this guide, please send an e-mail to
feedback_lcms@agilent.com.
.
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).Safety Notices
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
2 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
In This Guide...
The Concepts Guide presents “The Big Picture” behind the operation of the Agilent 6400 Series Triple Quadrupole LC/MS System by helping you understand how the hardware and software work.
1 Overview
Learn how the Agilent 6400 Series Triple Quadrupole helps you do your job.
2 Inner Workings – Triple Quadrupole MS versus Single
Quadrupole MS
Learn the concepts you need to understand how the Agilent triple quadrupole mass spectrometer works.
3 Agilent Triple Quadrupole MS and Sensitivity
Learn how the Agilent triple quadrupole mass spectrometer achieves high sensitivity.
4 Agilent MassHunter Workstation Software - Data Acquisition for
6400 Series Triple Quadrupole
Learn concepts behind the design of the Agilent MassHunter Workstation Software - Data Acquisition for Triple Quadrupole program.
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 3
4 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Contents
1 Overview 7
What kind of system do you have? 8 Help for applications 9 Help for acquisition 10 Help for data analysis 11
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole
MS 13
Single quadrupole MS operation 14
Design for a single quadrupole mass spectrometer 14 How a single quadrupole mass spectrometer works 15
Triple quadrupole MS operation 21
Design of the Agilent Triple Quadrupole MS 21 Innovative Enhancements in the 6490 Triple Quadrupole 23 Innovative Enhancements in the 6460 Triple Quadrupole 25 Innovative Enhancements in the 6430 Triple Quadrupole 27 Innovative Enhancements in the 6420 Triple Quadrupole 27 How a triple quadrupole mass spectrometer works 27 How Dynamic MRM works 30 How Triggered Dynamic MRM works 33
3 Agilent Triple Quadrupole MS and Sensitivity 39
How the Agilent Triple Quadrupole MS improves sensitivity 40
Noise reduction 40 Example of chemical noise reduction 43 Linearity of the Agilent 6400 Series Triple Quadrupole MS 45
How each component works to improve sensitivity 46
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 5
Agilent iFunnel Technology 46 Agilent Jet Stream Technology 47 LC/MS ion sources 49 Front-end ion optics 55 Collision cell 57 Detector 62 Pumping system 63
4 Agilent MassHunter Workstation Software - Data Acquisition for
6400 Series Triple Quadrupole 65
Tuning 67 Acquisition 69
6 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Agilent 6400 Series Triple Quadrupole LC/MS System Concepts Guide
1 Overview
What kind of system do you have? 8 Help for applications 9 Help for acquisition 10 Help for data analysis 11
This chapter provides an overview of the Agilent 6400 Series Triple Quadrupole LC/MS components and how they help get the job done.
Agilent Technologies
7
1Overview

What kind of system do you have?

What kind of system do you have?
You can set up a n A g ilent 6400 Series Triple Quadrupole LC/MS in several configurations:
ESI – Electrospray Ionization
APCI – Atmospheric Pressure
Chemical Ionization
APPI - Atmospheric Pressure
Photo Ionization
HPLC-Chip/MS – High Performance
Liquid Chromatography on a Chip
MMI - Multimode Ionization
For normal flow LC/MS with a binary pump, quaternary pump,
well-plate sampler (or autosampler or CTC PAL autosampler). The supported ion sources are ESI, APCI, APPI, and MMI.
For microflow LC/MS with a capillary pump, micro well-plate
sampler (or CTC PAL micro-plate autosampler) and ESI, APCI or MMI ion sources
For nanoflow LC/MS with a nanoflow pump, capillary pump,
micro well-plate sampler and HPLC-Chip/MS interface (used in place of a standard nanospray source) to increase reliability and boost performance with narrow peak dispersion and lower dead volumes.
Each Agilent combination has advantages for different applications. Each uses the same Data Acquisition program, Quantitative Analysis program and Qualitative Analysis program to enable these advantages.
The Agilent 6460 and 6490 Triple Quadrupole LC/MS systems are the only Triple Quadrupole that can use the Agilent Jet Stream Technology. This technology utilizes a super-heated sheath gas to collimate the nebulizer spray which dramatically increases the number of ions that enter the mass spectrometer.
The Agilent 6490 Triple Quadrupole LC/MS system also utilizes the iFunnel Technology which includes the Agilent Jet Stream Technology, shorter desolvation assembly with Hexabore Capillary, and the Dual Offset Ion Funnel.
8 Agilent 6400 Series Triple Quad LC/MS Concepts Guide

Help for applications

You can use one or more of the Agilent 6400 Triple Quadrupole LC/MS combinations to quantitate trace organic compounds in complex matrices:
Food safety studies
Environmental studies
Drug discovery
Toxicology
Forensics
Bioanalysis
Paired with Agilent’s 1260 and 1290 Infinity Series LCs, the 6400 Series Triple Quadrupole MS delivers sensitive, reproducible analyses of target compounds in complex matrices.
Femtogram-level limits of detection and quantitation for the
6430 and 6460
Zeptomole-level limits of detection and quantitation for the
6490
Overview 1
Help for applications
The dwell time is the amount of
time allotted for analyzing
each ion during a scan.
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 9
Minimized memory effects even at very short dwell times
Simplified operation with Agilent’s data analysis software
1Overview

Help for acquisition

Help for acquisition
To help you use the Agilent Triple Quadrupole LC/MS for these applications, the software lets you do these tasks in a single window with the Data Acquisition Program:
Prepare the instrument
To learn how to install the Agilent Triple Quadrupole LC/MS, see the Installation Guide.
To learn how to get started with the Agilent Triple Quadrupole LC/MS, see the Quick Start Guide.
To learn more about how to use the Agilent Triple Quadrupole LC/MS with real samples and data, see the
Familiarization Guide
To learn how to do individual tasks with the LC/MS, see the online Help.
To learn more about an Agilent 1260 or 1290 Infinity LC module, see the Agilent 1260 or 1290 Infinity LC User’s Guide for the module.
Start and stop the instruments from the software
Download settings to the Agilent 1260 or 1290 Infinity LC
and the Agilent 6400 Series Triple Quadrupole mass spectrometer in real time to control the instrument
Evaluate if the MS parameters are within the limits to
produce the specified mass accuracy and resolution with a
Checktune report
Optimize MS parameters automatically (Autotune) or
manually through Agilent tuning programs and print an Autotune report
Monitor the actual conditions of the instrument
View the real-time plot for chromatograms and instrument
parameters (both UV/Vis and MS) and print a real-time plot report
View the centroided line spectrum of a peak or the mass
range profile spectrum of a peak in real time
Set up acquisition methods
Enter and save parameter values for all LC modules and the
MS to an acquisition method
Select and label the total ion chromatograms or extracted ion
chromatograms that you want to appear in the real-time plot
Set up time segments for each scan type and analysis where
parameters change with the time segment or with the scans within the time segment
Print an acquisition method report
10 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Overview 1

Help for data analysis

Acquire data
Enter sample information and pre- or post-analysis programs
(scripts) and run single samples interactively
A worklist is a list of individual
samples and batches
(sequences) that you enter
and run automatically with
the Data Acquisition program.
Help for data analysis
Enter and automatically run both individual samples and
samples organized in a
Set up pre- and post-analysis scripts to run between samples
in a worklist
Set up and run a worklist to optimize MS acquisition
parameters
Print a worklist report
View system events, including start and stop times, run
events and errors and print an event log report
worklist (sequence of samples)
Quantitative Analysis Program
Agilent has designed the quantitative analysis program to help quantitate very low amounts of material with the following unique features:
Imports information directly from the acquisition method
Provides a curve-fit assistant to test all fits and statistics on
curve quality
Integrates with an automated, parameter-free integrator that
uses a novel algorithm, optimized for triple quadrupole data
Presents a Batch-at-a-Glance results window to help you
review and operate on an entire batch of data at once
Automatically detects outliers
Provides preconfigured templates for basic reporting and
enables the capability to create custom reports in Microsoft Excel
Please refer to the Agilent MassHunter Workstation Software - Quantitative Analysis Familiarization Guide or the online Help for the Quantitative Analysis program.
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 11
1Overview
Help for data analysis
Qualitative Analysis Program
For fast method development, this software is used to quickly review the qualitative aspects of the data, such as the optimum precursor to product ion transitions.
Agilent designed the Qualitative Analysis program to present large amounts of data for review in one central location. With the program you can do these operations for any type of mass spectrometer data that you open:
Extract chromatograms
View and extract peak spectra
Subtract background
Integrate the chromatogram
Find compounds
Identify compounds
Export results
You can also set up methods to automatically do the tasks in the list, as well as others, when you open the data files.
Please refer to the Agilent MassHunter Workstation Software - Qualitative Analysis Familiarization Guide or the online Help for the Qualitative Analysis program.
12 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Agilent 6400 Series Triple Quadrupole LC/MS System Concepts Guide
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS
Single quadrupole MS operation 14
Design for a single quadrupole mass spectrometer 14 How a single quadrupole mass spectrometer works 15
Triple quadrupole MS operation 21
Design of the Agilent Triple Quadrupole MS 21 How a triple quadrupole mass spectrometer works 27 Innovative Enhancements in the 6490 Triple Quadrupole 23 Innovative Enhancements in the 6460 Triple Quadrupole 25 Innovative Enhancements in the 6430 Triple Quadrupole 27 How Dynamic MRM works 30 How Triggered Dynamic MRM works 33
In this chapter you learn about concepts to help you understand the inner workings of the Agilent 6400 Series Triple Quadrupole LC/MS.
The foundation for understanding the operation of a triple quadrupole mass spectrometer is the operation of a single quadrupole mass spectrometer. Therefore, an explanation of the workings of a single quadrupole mass spectrometer is presented first.
Agilent Technologies
13
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS

Single quadrupole MS operation

Single quadrupole MS operation
To better understand the specific hardware features of the Agilent 6400 Series Triple Quadrupole Mass Spectrometer, this section first reviews the fundamental aspects of the single quadrupole mass spectrometer.

Design for a single quadrupole mass spectrometer

Mass spectrometry is based on the analysis of ions moving through a vacuum.
The ionization of a sample occurs in the ion source that is shown, schematically, on the left. The ions are analyzed by a mass analyzer (mass filter) that controls the motion of the ions as they travel to the detector to be converted into actual signals.
Figure 1 Schematic for single quadrupole mass spectrometer
m/z – mass/charge ratio The quadrupole mass analyzer consists of four parallel rods to
which specific DC and RF voltages are applied. These rods filter out all ions except those of one or more particular m/z values as determined by the voltages applied.
14 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS 2
External Ionization Source
Quadrupole Mass Filter
Detector

How a single quadrupole mass spectrometer works

The RF is applied to all four rods, but the negative (–) rods are 180 degrees out of phase with the positive (+) rods. The rods are labeled + and – in reference to the DC voltages applied to them.
All ions that comprise the sample are generated at the source. However, when a specific set of voltages is applied, only ions of the corresponding m/z value may pass through the quadrupole to reach the detector. As the voltages are increased to other values, ions with other m/z values are allowed to pass through. A full MS scan is obtained by increasing the DC and RF voltages applied to the four rods over an expanded range of values.
How a single quadrupole mass spectrometer works
A diagrammatic model can be used to illustrate the concept of how a single quadrupole instrument works. See Figure 2.
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 15
Figure 2 Conceptual model of a single quadrupole mass spectrometer
In the model,
All of the ions contained in a sample are formed in the
external ionization source and collected in a funnel. The balls of different colors and sizes represent different ions having different m/z values.
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS
How a single quadrupole mass spectrometer works
The quadrupole mass analyzer is represented by a moving
belt that serves to filter the ions as they pass through openings of various sizes. The ions pass from the funnel, through the filter, to the detector. Although in this image, ions that are smaller might fit through the openings, a quadrupole mass analyzer filters the ions so that only the “correct” ions pass through to the detector.
The detector is represented by the collecting funnel below
the filtering belt.
As the belt (the analyzer) moves, or the voltages on the rods are changed, ions with different m/z values are filtered through the mass spectrometer.
As the analyzer moves from a small m/z value to increasingly larger values, a full MS scan is created.
SIM – Selected Ion Monitoring If the belt does not move, the detector continues to monitor the
same single m/z value over the entire scan period. This type of analysis is known as SIM. It is the most sensitive operating mode for a single quadrupole mass spectrometer.
The scan period is selected (fixed) by the user. The user may set the dwell time to scan a specific mass range (e.g. m/z 50 to
1000) or to remain on one selected ion (SIM) or to move to several selected ions during the scan period. The quadrupole mass filter is not scanned in this mode. The required RF/DC voltages are often set to filter a single mass at one time.
For comparison, see “How a triple quadrupole mass
spectrometer works" on page 27.
Single quadrupole: SIM
To obtain the best sensitivity or quantitation, the single quadrupole is operated in SIM mode (Figure 3). The duty cycle is the measure of the instrument’s time actually devoted to measuring signals. In SIM mode, the single quadrupole analyzes the signal of a specific m/z ion almost all of the time. This results in nearly 100% acquisition during the duty cycle.
16 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS 2
API Source
Ion Guide
Detector
Quadrupole Mass Analyzer
How a single quadrupole mass spectrometer works
Figure 3 Single quadrupole: SIM
In this example,
1 All of the ions (+, -, and neutrals) are formed in the API
source.
2 Ion optics guide the ions to the quadrupole mass analyzer.
The Agilent Ion Guide is an octopole filter of eight equally spaced rods.
3 In the analyzer, only ions of a particular m/z value,
represented by blue balls, are allowed to pass through to the detector.
4 The detector completes the analysis.
This system has several advantages:
provides the best sensitivity for quantitation
increases selectivity
improves chromatographic specificity
provides no structural information
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 17
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS
API Source
Detector
Quadrupole Mass Analyzer
How a single quadrupole mass spectrometer works
Single quadrupole: Full Scan MS
In a full MS scan, the quadrupole serves as a mass filter over time, and a scan is carried out by stepping through increasing DC and RF voltages. This provides filtering through the corresponding m/z values across a mass spectrum. See Figure 4.
Figure 4 Single quadrupole: Full scan MS
The full scan MS mode is less sensitive because the duty cycle for each m/z is considerably less than 100%. The quadrupole mass analyzer scans sequentially, passing each m/z in the selected mass range to the detector.
A full scan MS is still a useful mode of operation because it shows all of the ions that are being formed in the ion source. This is useful for developing SIM acquisitions but also alerts analysts to other compounds co-eluting with compounds of interest.
18 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS 2
m/z 325 Analyte Precursor
m/z 325 Matrix Precursor
m/z 202 Matrix Precursor
m/z 184 Product Ion
m/z 124 Product Ion
Detector
API Source
How a single quadrupole mass spectrometer works
What about fragment ions?
Full scans with a single quadrupole instrument can also be used to study fragment ions. See Figure 5.
Figure 5 Fragment ions with single quadrupole MS
The diagram shows that fragment ions, also known as product ions, are formed by fragmenting or breaking apart precursor ions. Precursor ions formed in the ion source travel through the mass analyzer without change, unless extra energy is applied to their motion in a region where fragmentation can occur.
This fragmentation or collisionally induced dissociation (CID) can be carried out in a low pressure region between the ion source and the mass analyzer. The ion source is under atmospheric pressure, while the mass analyzer is at a much lower pressure because it has been evacuated of gas with a vacuum pump.
On the Agilent single quadrupole mass spectrometer, this region is between the capillary exit and the skimmer, where the gas pressure is about 2 Torr, or about three orders of magnitude below atmosphere pressure (760 Torr). Under normal operation, a voltage is applied across this region to keep the
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 19
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS
How a single quadrupole mass spectrometer works
ions passing through to continue on to the mass analyzer. Even if these ions collide with the gas molecules in this region, they usually do not have enough energy to fragment.
CID – Collisionally Induced
Dissociation
However, as the voltage is increased, the ions have more translational energy. Then, if the ions run into gas molecules, the collisions convert the translational energy into molecular vibrations that can cause the ions to fragment (Figure 6). This is collisionally induced dissociation (CID). Even though this fragmentation does not occur where the ions are formed at atmospheric pressure, it’s a tradition to call this type of fragmentation “In-source CID.”
Figure 6 Ion fragmentation caused by collision-induced dissociation
A single quadrupole mass spectrometer cannot be used to do MS/MS because all of the ions formed in the ion source are transferred to the quadrupole whether fragmented or not. At the end when the mass analyzer filters the ions, it is not possible to identify which product ions came from which precursor ions.
A triple quadrupole mass spectrometer can do MS/MS, with fragmentation within its collision cell as described in the next section.
20 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS 2

Triple quadrupole MS operation

Be sure to read the previous section on the concepts behind the operation of a single quadrupole mass spectrometer. Understanding these concepts helps you understand the operation of the triple quadrupole mass spectrometer.

Design of the Agilent Triple Quadrupole MS

The triple quadrupole mass spectrometer consists of an ion source, enhanced desolvation technology, followed by ion optics that transfer the ions to the first quadrupole positioned to the right of it. A diagram of some of the current Triple Quadrupole LC/MS products is shown in Figure 7 on page 22. The Agilent 6430 is shown in Figure 8 on page 23.
Triple quadrupole MS operation
Agilent 6400 Series Triple Quadrupole System
6420 Includes one turbo pump and one rough pump
6430 Adds an additional turbo pump
6460 Includes Agilent Jet Stream Technology
6490 Includes iFunnel technology (Agilent Jet
Highlights
Includes resistive capillary
Can upgrade to a 6430
Improves pumping in vacuum stage 2
Updates collision cell lenses
Can upgrade to a 6460
Includes 3,000 m/z Q1 and Q2 quadrupoles
Stream, hexabore capillary, and high pressure/low pressure ion funnels)
Adds additional rough pump for ion funnel
Includes high throughput quadrupole driver
electronics
Includes curved collision cell assembly
Has a smaller footprint
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 21
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS
Circled areas indicate enhancements
Design of the Agilent Triple Quadrupole MS
Figure 7 Innovative Enhancements in the 6420, 6430 and 6460
22 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Loading...
+ 50 hidden pages