Agilent 6286A Spec Sheet

Hart Scientific 1502A
Specifications
Temperature Range†
Resistance Range
Probe
Characterizations ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 IPTS-68 R0, a, d, a4, and c4 Callendar-Van
Dusen Resistance
Accuracy(ppm of reading)
Temperature Accuracy†, typical (meter only)
Operating Temperature Range
–200°C to 962°C (–328°F to 1764°F)
RTD, PRT, or SPRT
R0, a, d, and b
0W to 20W: 0.0005W 20W to 400W: 25 ppm
±0.004°C at –100°C ±0.006°C at 0°C ±0.009°C at 100°C ±0.012°C at 200°C ±0.018°C at 400°C ±0.024°C at 600°C
16°C to 30°C
Resistance Resolution
Temperature Resolution
Excitation Current
Measurement Period
Digital Filter Exponential, 0 to 60 seconds time constant (user selectable) Probe
Connection
Communications Display 8-digit, 7-segment, yellow-green LED; 0.5-inch-high characters
Power Size 5.6" W x 7.1" D x 2.4" H (143 x 181 x 61 mm)
Weight 2.2 lb. (1.0 kg) †Temperature ranges and accuracy may be limited by the sensor you use.
0W to 20W: 0.0001W 20W to 400W: 0.001W
0.001°C
0.5 and 1 mA, user selectable, 2 Hz
1 second
4-wire with shield, 5-pin DIN connector RS-232 serial standard
IEEE-488 (GPIB) optional
115 VAC (±10%), 50/60 Hz, 10 A, nominal 230 VAC (±10%), 50/60 Hz, 10 A, nominal, specify