Agilent 6100 Familiarization Guide

Agilent 6100 Series Quadrupole LC/MS Systems
Familiarization Guide
Agilent Technologies
Notices
CAUTION
WARNING
© Agilent Technologies, Inc. 2011
Manual Part Number
G1960-90080
Edition
Revision A, September 2011
Printed in USA
Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051
This guide is valid for the B.03.01 or later revision of the Agilent ChemStation soft­ware for the Agilent 6100 Series Quadrupole LC/MS systems, until superseded.
If you have any comments about this guide, please send an e-mail to
feedback_lcms@agilent.com.
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Safety Notices
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
In this Guide...
This guide presents a series of exercises to help you learn the basic operation of your Agilent 6100 Series LC/MS system.
If you have any comments about this guide, please send an e-mail to feedback_lcms@agilent.com.
1 Prepare for the Analysis
Use these exercises to prepare the LC, to dilute a sulfa demonstration sample, and to check the tune on the MS.
2 Set Up and Run a Scan Method
Learn how to set up a scan method and acquire data for the sulfa demonstration mix.
3 Qualitative Data Analysis
Learn how to examine chromatograms and spectra to identify sample components. In these exercises, you review data from the sulfa sample you analyzed in Chapter 2, or from a data file that you received with your ChemStation software.
4 Set Up and Run a SIM Method
Learn how to set up a selected ion monitoring (SIM) method and acquire data for the sulfa demonstration mix.
5Set Up and Run a Sequence
Use these exercises to set up an automated sequence for SIM analyses of the sulfa mix at various concentrations, and to acquire data with that sequence.
6 Quantitative Data Analysis
Learn how to analyze data when you need to quantify sample components. These exercises use caffeine data files that you received with your ChemStation software.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 3
4 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Contents
1 Prepare for the Analysis 7
Exercise 1. Prepare the LC to run the sample 8
Task 1. Start up ChemStation 8 Tas k 2 . P ur ge t he pu mp 9 Task 3. Prepare the column for the analyses 10
Exercise 2. Prepare the samples for the analyses 12
Exercise 3. Check the current MS tune values and adjust if
necessary 14
2 Set Up and Run a Scan Method 15
Exercise 1. Set up a full-scan acquisition method 16
Task 1. Enter LC acquisition parameters 16 Task 2. Enter MS acquisition parameters 19
Exercise 2. Acquire data with the full-scan method 23
Task 1. Enter sample information 24 Task 2. Acquire the data 25
3 Qualitative Data Analysis 27
Exercise 1. Display and manipulate chromatograms 28
Exercise 2. Examine mass spectra 31
Exercise 3. Integrate the chromatogram 36
Exercise 4. Print a report 40
4 Set Up and Run a SIM Method 43
Exercise 1. Set up a SIM acquisition method 44
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 5
Contents
Task 1. Load the scan method you created previously 44 Task 2. Enter MS acquisition parameters 45
Exercise 2. Acquire data with the SIM method 48
Task 1. Enter sample information 49 Task 2. Acquire the data 50
5Set Up and Run a Sequence51
Exercise 1. Set up a sequence 52
Task 1. Prepare to create a new sequence 52 Task 2. Edit sequence parameters 53 Task 3. Set up the sequence table 55 Task 4. Set up the sequence output 58
Exercise 2. Run the sequence 60
6 Quantitative Data Analysis 61
Exercise 1. Create a method for quantification 62
Task 1. Create a new method 62 Task 2. Set up the signal for quantification 63 Task 3. Integrate the low-level standard 65 Task 4. Set general calibration parameters 67 Task 5. Set up the calibration curve 68 Task 6. Explore options to refine the calibration 72
Exercise 2. Process a sample and print a report 73
6 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System Familiarization Guide
1 Prepare for the Analysis
Exercise 1. Prepare the LC to run the sample 8
Task 1. Start up ChemStation 8 Task 2. Purge the pump 9
Task 3. Prepare the column for the analyses 10 Exercise 2. Prepare the samples for the analyses 12 Exercise 3. Check the current MS tune values and adjust if necessary 14
This chapter presents exercises to help you learn how to:
Prepare the LC and column for an analysis
Prepare the samples that you analyze in these exercises
Check the tune settings of the MS and adjust if necessary.
Before you start
Order the sample: Agilent Electrospray LC Demo Sample,
p/n 59987-20033.
Order the column: Agilent ZORBAX SB-C18, 2.1 mm x 30 mm,
3.5 µm, p/n 873700-902.
You may use another similar column, but you may need to adjust the HPLC conditions to obtain good separation.
Make sure that the electrospray source is installed.
Read the Agilent 6100 Series Quadrupole LC/MS Systems
Quick Start Guide and Chapter 2 of the Agilent 6100 Series Quadrupole LC/MS Systems Concepts Guide.
Agilent Technologies
7
1 Prepare for the Analysis
Exercise 1. Prepare the LC to run the sample
Exercise 1. Prepare the LC to run the sample
For the following tasks, try the steps in the first column. If you need more help, follow the detailed instructions in the middle column.
Task 1. Start up ChemStation
Steps Detailed Instructions Comments
1 Open the ChemStation window. Click the ChemStation icon on the
desktop.
Alternate method:
From the Start menu, select:
All Programs > Agilent ChemStation > Instrument 1 online.
8 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Prepare for the Analysis 1
Task 2. Purge the pump
Task 2. Purge the pump
Use these instructions with the binary and quaternary pumps. See the ChemStation online Help for instructions for the capillary and nanoflow pumps.
Steps Detailed Instructions Comments
1 Display the Method and Run Control
view.
2 Place the pump in standby mode. a Click More Pump > Control HPLC
3 Prepare solvents used in these
familiarization exercises.
A – 5 mM ammonium formate in
water
B – 5 mM ammonium formate in
methanol
4 Replace the solvent bottles with the
ones you just prepared.
5 Open the purge valve. a Turn the black purge valve on the
6 Enter a flow of 5 mL/min and
50% B, using water in channel A and methanol in channel B.
In the view selection area in the lower left, click Method and Run
Control.
Pump on the Instrument menu to
open the Pump Control dialog box.
b Select Standby and click OK.
a Into a 1-liter reservoir of
HPLC-grade water, add 1 mL of 5 M ammonium formate.
b Into a 1-liter reservoir of
HPLC-grade methanol, add 1 mL of 5 M ammonium formate.
Replace the bottles for channels A and B.
front of the pump counter-clockwise two turns.
b Place the tubing that exits the pump
into a 250-mL or larger beaker.
a Click the pump icon. b Select Set up Pump. c Enter the parameters in step 6 and
click OK.
Alternate method:
Select Standby from the Pump context menu.
The part number for ammonium formate is G1946-85021.
Each ampoule contains 2.2 mL of ammonium formate solution.
Be sure to use HPLC-grade solvents.
7 Turn the pump on and monitor the
tubing for bubbles.
a To turn the pump on, click the
little button to the lower right of the solvent delivery (pump) icon.
b Monitor for bubbles.
Purge for about 3 minutes to pass
3X the volume for the binary pump.
If you wish, you may purge each channel individually first, to ensure that neither is air-locked.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 9
1 Prepare for the Analysis
Task 3. Prepare the column for the analyses
Steps Detailed Instructions Comments
8 After the bubbles are gone and the
purge is complete, enter a flow of 1 mL/min and 100% B.
9 Purge a short while longer, and then
close the purge valve.
a Right-click the pump icon. b Select Set up pump. c Enter the new parameters in step 8,
and click OK.
a Continue to purge for a short while. b Close the black valve.
For more information on purging the pump, see the reference manual that you received with your pump.
Task 3. Prepare the column for the analyses
In the exercises in the next chapters, you analyze a mixture of four sulfonamide compounds. To perform the analyses in the following chapters, you must first condition and equilibrate your column.
Steps Detailed Instructions Comments
1 Disconnect the column from the
detector and MS.
a Turn the pump off by clicking
the little button to the lower right of the solvent delivery (pump) icon.
b Disconnect the column from the
detector and MS.
c Place the open end of the tubing
that exits the column into the beaker.
To prevent detector contamination, allow the column effluent to go directly to the waste beaker.
2 Flush the column with 100%
methanol at 1 mL/min (5 to 10 min).
ZORBAX SB-C18, 2.1 mm
3.5 µm, p/n 873700-902
× 30 mm,
a Tur n t h e pum p o n. b Flow methanol through the column
under the conditions used in Task 2,
step 8.
The data sheet shipped with the column cartridge recommends that you flush with 20 to 30 column­volumes of 100% methanol (approximately 5 to 7.5 mL).
10 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Prepare for the Analysis 1
Task 3. Prepare the column for the analyses
Steps Detailed Instructions Comments
3 Condition the column as follows,
using the solvents made up in Task 2,
step 6:
Flow rate – 0.4 mL/min
100% B for 1/2 hour
50% B for 1/2 hour
4 Equilibrate the column at the analysis
conditions:
12% B for 1/2 hour at 40 °C
5 While the column equilibrates, set
parameters for the MS spray chamber so it can heat and equilibrate as well.
Drying gas flow: 8 L/min
Nebulizer pressure: 35 psig
Drying gas temperature: 300
Capillary voltage: 3000 V
For 6150 with Agilent Jet Stream technology:
Sheath Gas Flow: 12 L/min
Sheath Gas Temp: 360°C
Nozzle Voltage: 0 V
o
C
a Click Set up Instrument Method on
the Instrument menu to open the Setup Method dialog box.
b Click the Pump tab. c Enter the flow rate in step 3. d For Solvent B, type 100 and click
Apply. e Wait 30 minutes. f For Solvent B, type 50 and click
Apply. g Wait 30 minutes.
a For Solvent B, type 12 and click
OK. b Click the TCC tab on the Setup
Method dialog box.
c For Temperature, type 40 and click
OK.
a Right-click the MSD
icon on the system
diagram and select
Spray Chamber. b Enter the parameters described in
step 5.
c Click OK. d Wait 10 minutes before you tune the
MS.
At a flow rate of 0.4 mL/min, the checkout column should produce about 70 to 80 bar pressure (measured without any fittings at the column exit).
If, after you perform these steps, the pump pressure through the column is too high, order a replacement SB-C18 column (p/n 873700-902).
If your column is not new, you can reduce the length of time that you condition the column.
While you condition and equilibrate the column, you may complete
step 5 in this exercise and then
work on the rest of the exercises in this chapter. Be sure to complete
step 6 before you go on to the next
chapter.
6 Reconnect the column to the DAD
and MS.
You can complete “Exercise 3.
Check the current MS tune values and adjust if necessary” either with
or without the column connected to the DAD and MS, but you do need to reconnect prior to the exercises in
Chapter 2, “Set Up and Run a Scan
Method.”
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 11
1 Prepare for the Analysis
N
S
O
OH
N
2
H
S
N
N
N
N
N
S
O
OH
N
2
H
N
N
N
S
O
OH
N
2
H
Cl
N
N
O
O
N
S
O
OH
N
2
H
Exercise 2. Prepare the samples for the analyses
Exercise 2. Prepare the samples for the analyses
In the exercises in the next chapters, you analyze a mixture of four sulfonamide compounds. The Electrospray LC Demo Sample (p/n 59987-20033), contains five ampoules with 100 ng/µL each of these compounds:
sulfamethizole (M+H)
sulfamethazine (M+H)
sulfachloropyridazine (M+H)
sulfadimethoxine (M+H)
+
= 271
+
= 279
+
= 311.
+
= 285
Sulfamethizole Sulfamethazine Sulfachloropyridazine Sulfadimethoxine
To perform the analyses in the following chapters, you must first prepare the sample at various dilutions. The final concentrations will be 1, 5 and 10 ng/µL. You will also prepare a solvent blank.
12 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Prepare for the Analysis 1
Exercise 2. Prepare the samples for the analyses
Steps Detailed Instructions Comments
1 Prepare a 1:10 dilution of the original
sample in a 1-mL autosampler vial.
Final concentration is 10 ng/µL
You will use this sample for the
scan analysis in Chapter 2, and for the SIM analyses in Chapter 4 and
Chapter 5.
2 Prepare a 1:20 dilution of the original
sample in a 1-mL autosampler vial.
Final concentration is 5 ng/µL
You will use this sample for the SIM
analysis in Chapter 5.
3 Prepare a 1:100 dilution of the original
sample in a 1-mL autosampler vial.
Final concentration is 1 ng/µL
You will use this sample for the SIM
analysis in Chapter 5.
4 Prepare a solvent blank in a 1-mL
autosampler vial.
You will use this sample for the SIM
analysis in Chapter 5.
a Transfer 100 µL of the sulfa mixture
into the autosampler vial.
b Add 900 µL of 90:10 water:methanol
that contains 5 mM ammonium formate (NH
c Cap the vial.
a Transfer 50 µL of the sulfa mixture
into the autosampler vial.
b Add 950 µL of 90:10 water:methanol
that contains 5 mM ammonium formate.
c Cap the vial.
a Transfer 10 µL of the sulfa mixture
into the autosampler vial.
b Add 990 µL of 90:10 water:methanol
that contains 5 mM ammonium formate.
c Cap the vial.
a Into the autosampler vial, transfer
990 µL of 90:10 water:methanol that contains 5 mM ammonium formate.
b Cap the vial.
HCO2).
4
The original sulfa mixture is dissolved in a solvent mixture of 70% water and 30% acetonitrile.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 13
1 Prepare for the Analysis
Exercise 3. Check the current MS tune values and adjust if necessary
Exercise 3. Check the current MS tune values and adjust if necessary
The MS is very stable and does not need to be tuned very often. You can usually tune just once a month, or once a week at most. You can use the Check Tune program described in this exercise to confirm that the MS is in adjustment.
Steps Detailed Instructions Comments
1 Switch to the MSD Tune view. • In the view selection area in the
lower left, click MSD Tune.
2 Select the tune file. a In the Select Tune File dialog box,
select ATU NE S.TU N.
b Keep the default of Positive Polarity
(Standard). c Click OK. d In the status bar near the top of the
MSD Tune view, verify that you see
the following:
Mode is API-ES
Source is ESI (electrospray)
3 Run a Check Tune. • From the Tune menu, select Check
Tune.
Note that Check Tune requires
values for comparison that are
determined from a previous
Autotune. Autotune is normally run
during installation.
4 If Check Tune report suggests that
you adjust peak widths or mass axis, then do that.
5 If the Check Tune report shows poor
sensitivity, which indicates that your MS settings are significantly out of adjustment, then run a full Autotune.
a From the Tune menu, select Adjust
Mass Peak Width. b From the Tune menu, select
Calibrate Mass Axis.
From the Tune menu, select
Autotune > Positive Polarity.
Make sure that you use an
appropriate calibrant with an appropriate source.
Check Tune is normally all that you need to do to confirm that the MS settings are correct.
If Check Tune indicates a problem with your MS settings, then proceed to step 4 and/or step 5.
The exercises in this manual use only the positive ion mode and standard scan speeds, so it is not necessary to tune for negative polarity or fast scan.
14 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System Familiarization Guide
2 Set Up and Run a Scan Method
Exercise 1. Set up a full-scan acquisition method 16
Task 1. Enter LC acquisition parameters 16 Task 2. Enter MS acquisition parameters 19
Exercise 2. Acquire data with the full-scan method 23
Task 1. Enter sample information 24 Task 2. Acquire the data 25
These exercises show you how to set up a scan data acquisition method for the demonstration sample (sulfa mix) and to acquire data with that method.
The LC parameters that you enter in these exercises are appropriate for the standard Agilent 1100/1200/1260/1290 Series liquid chromatography (LC) systems. You must enter LC parameters that are appropriate for your LC model.
To view the results of these exercises, see Chapter 3, “Qualitative Data Analysis.”
Before you start
Review the Agilent 6100 Series Quadrupole LC/MS Systems Quick Start Guide and Chapter 3 of the Agilent 6100 Series Quadrupole LC/MS Systems Concepts Guide.
Prepare the LC, column and sample as described in
Chapter 1, “Prepare for the Analysis.”
For the tasks on the following pages, try the steps on the left without the detailed instructions. If you need more help, follow the detailed instructions on the right.
Agilent Technologies
15
2 Set Up and Run a Scan Method
Exercise 1. Set up a full-scan acquisition method
Exercise 1. Set up a full-scan acquisition method
This exercise changes the default method and saves it as a new method. This exercise consists of the following tasks:
“Task 1. Enter LC acquisition parameters” on page 16
“Task 2. Enter MS acquisition parameters” on page 19
Task 1. Enter LC acquisition parameters
Steps Detailed Instructions Comments
1 Display the Method and Run Control
view.
2 Open the method DEF_LC.M. a Select File > Load > Method.
3 Save the method under a new name,
SULFAMSSCAN1.M.
4 Enter a volume of
injection.
1 µL for the
In the view selection area in the
lower left of the ChemStation window, click Method and Run
Control.
b If necessary, navigate to
C:\CHEM32\1\METHODS.
c Select DEF_LC.M and click OK.
a Select File > Save As > Method. b In the dialog box, for Name, type
SULFA MS SCAN 1.M.
c Click OK. d In the box for Comment for method
history, type a comment.
e Click OK.
a Click Set up Instrument Method on
the Instrument menu to open the Setup Method dialog box.
b Click the ALS tab. c Click Standard injection. d In the Injection volume box, type 1
for a 1-µL injection.
You save the method now with a new name to avoid inadvertently overwriting the default method later.
16 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method 2
Task 1. Enter LC acquisition parameters
Steps Detailed Instructions Comments
5 Enter pump parameters. a Click the Pump tab on the Setup
Method dialog box.
b Set the parameters as follows:
Flow=0.400mL/min StopTime=7.00 min PostTime=3.00min
Solvent A=Water 88% Solvent B=Methanol 12%
6 Set up the gradient timetable as it
appears in the figure below.
7 Enter a column compartment
temperature of 40
o
C.
8 Enter parameters for the diode-array
detector (DAD).
a Open the Timetable area in the
lower part of the tab, click Insert, and type the first line.
b Click Append and type the second
line.
c Repeat step b for lines 3 and 4.
a Click the TCC tab on the Setup
Method dialog box.
b Click the option button for oC. c Ty p e 40.0 for
o
C.
a Click the DAD tab on the Setup
Method dialog box.
b Enter the parameters shown below:
Use Signal A: Wavelength 272nm,
Bandwidth16 nm
Reference Wavelength: 360 nm,
Reference Bandwidth 100 nm
Spectrum Store: All in peak
Peakwidth: > 0.1 min
c Click OK to close the Setup Method
dialog box with the new setpoints.
Set the timetable parameters as follows:
Line 1 Time 1:00, %B=12, Flow=0.4 Line 2 Time 3:00, %B=100, Flow=0.4 Line 3 Time 6:00, %B=100, Flow=0.4 Line 4 Time 7:00, %B=12, Flow=0.4
The DAD is used in this example, but the variable wavelength detector (VWD) may be used analogously.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 17
2 Set Up and Run a Scan Method
Task 1. Enter LC acquisition parameters
Steps Detailed Instructions Comments
9 Select Data Acquisition only in the
Run Time Checklist.
10 Save the new parameters to the
method file, SULFA MS SCAN 1.M.
a Click Run Time Checklist on the
RunControl menu.
b Mark the Data Acquisition check
box.
c Click OK.
a Select File > Save > Method. b In the box for Comment for method
history, type a comment.
c Click OK.
While it is common to include Data
Analysis in the Run Time Checklist, for these exercises, you will view the results in Chapter 3, “Qualitative Data Analysis.”
18 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method 2
Set to 100 for 6120 Set to 125 for 6130 or 6150
Task 2. Enter MS acquisition parameters
Task 2. Enter MS acquisition parameters
Steps Detailed Instructions Comments
1 Enter parameters for the quadrupole
mass spectrometer (MS):
Signal 1, scan mode, positive
polarity
Scan range: 100 to 500
Fragmentor:
100 V for the Agilent 6120; 125 V for the 6130 or 6150
Gain: 1.00
Threshold: 150
Stepsize: 0.10
Peakwidth: 0.05 min
Scan data storage: Condensed
Active signals: 1 only
a Right-click the MSD
icon on the system diagram and select Set
up MSD Signals.
b Enter the parameters described in
step 1 and shown in the figure
below. Take care to enter the appropriate Fragmentor voltage for your MS model.
c Click OK.
To save disk space you usually acquire line spectra (Scan Data Storage = Condensed). However,
when you acquire spectra from intact proteins or protein digests/peptides, you must acquire and deconvolute profile spectra. (Scan Data Storage = Full).
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 19
2 Set Up and Run a Scan Method
Task 2. Enter MS acquisition parameters
Steps Detailed Instructions Comments
2 Enter parameters for the spray
chamber of the ion source:
Drying gas flow: 9 L/min
Nebulizer pressure: 40 psig
Drying gas temperature: 300
Capillary voltage: 3000 V
For 6150 with Agilent Jet Stream technology:
Nebulizer pressure: 30 psi
Drying gas flow: 7 L/min
Drying gas temperature: 350°C
Sheath gas flow: 12 L/min
Sheath gas temperature: 360 C°
Capillary voltage: 4000 V
Nozzle voltage: 0 V
Fragmentor: 200 V
Multiplier gain: 3
a Right-click the MSD
icon on the system diagram and select
o
C
Spray Chamber.
b Enter the parameters described in
step 2 and shown in the figure
below.
c Click OK.
For all models except 6150 with Agilent Jet Stream technology
For 6150 with Agilent Jet Stream technology
20 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method 2
Task 2. Enter MS acquisition parameters
Steps Detailed Instructions Comments
3 Set up to store the fragmentor voltage
throughout the run.
4 Save the method. a Select Method > Save Method to
a Right-click the MSD
icon on the system diagram and select
Data Curves. b Select Fragmentor - 1. c Click the Add button. d Click OK.
overwrite the method
SULFA MS SCAN 1.M. b In the box for Comment for method
history, type a comment. c Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 21
2 Set Up and Run a Scan Method
Task 2. Enter MS acquisition parameters
Steps Detailed Instructions Comments
5 Print the method. a Select Method > Print Method.
b Mark the check boxes as shown in
the figure below.
c Click the Print button.
22 Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method 2
Exercise 2. Acquire data with the full-scan method
Exercise 2. Acquire data with the full-scan method
Now you are ready to acquire data for the sulfa mix with the method you just created. This exercise consists of the following tasks:
“Task 1. Enter sample information” on page 24
“Task 2. Acquire the data” on page 25
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide 23
Loading...
+ 53 hidden pages