No p art o f this manu al may be re produce d in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G1960-90080
Edition
Revision A, September 2011
Printed in USA
Agilent Technologies, Inc.
5301 Stevens Creek Blvd.
Santa Clara, CA 95051
This guide is valid for the B.03.01 or later
revision of the Agilent ChemStation software for the Agilent 6100 Series Quadrupole
LC/MS systems, until superseded.
If you have any comments about this guide,
please send an e-mail to
feedback_lcms@agilent.com.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user customers. Agilent provides this customary
commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
In this Guide...
This guide presents a series of exercises to help you learn
the basic operation of your Agilent 6100 Series LC/MS
system.
If you have any comments about this guide, please send an
e-mail to feedback_lcms@agilent.com.
1Prepare for the Analysis
Use these exercises to prepare the LC, to dilute a sulfa
demonstration sample, and to check the tune on the MS.
2Set Up and Run a Scan Method
Learn how to set up a scan method and acquire data for the
sulfa demonstration mix.
3Qualitative Data Analysis
Learn how to examine chromatograms and spectra to
identify sample components. In these exercises, you review
data from the sulfa sample you analyzed in Chapter 2, or
from a data file that you received with your ChemStation
software.
4Set Up and Run a SIM Method
Learn how to set up a selected ion monitoring (SIM) method
and acquire data for the sulfa demonstration mix.
5Set Up and Run a Sequence
Use these exercises to set up an automated sequence for SIM
analyses of the sulfa mix at various concentrations, and to
acquire data with that sequence.
6Quantitative Data Analysis
Learn how to analyze data when you need to quantify
sample components. These exercises use caffeine data files
that you received with your ChemStation software.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide3
4Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Contents
1Prepare for the Analysis7
Exercise 1. Prepare the LC to run the sample8
Task 1. Start up ChemStation8
Tas k 2 . P ur ge t he pu mp9
Task 3. Prepare the column for the analyses10
Exercise 2. Prepare the samples for the analyses12
Exercise 3. Check the current MS tune values and adjust if
necessary14
2Set Up and Run a Scan Method15
Exercise 1. Set up a full-scan acquisition method16
Task 1. Enter LC acquisition parameters16
Task 2. Enter MS acquisition parameters19
Exercise 2. Acquire data with the full-scan method23
Task 1. Enter sample information24
Task 2. Acquire the data25
3Qualitative Data Analysis27
Exercise 1. Display and manipulate chromatograms28
Exercise 2. Examine mass spectra31
Exercise 3. Integrate the chromatogram36
Exercise 4. Print a report40
4Set Up and Run a SIM Method43
Exercise 1. Set up a SIM acquisition method44
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide5
Contents
Task 1. Load the scan method you created previously44
Task 2. Enter MS acquisition parameters45
Exercise 2. Acquire data with the SIM method48
Task 1. Enter sample information49
Task 2. Acquire the data50
5Set Up and Run a Sequence51
Exercise 1. Set up a sequence52
Task 1. Prepare to create a new sequence52
Task 2. Edit sequence parameters53
Task 3. Set up the sequence table55
Task 4. Set up the sequence output58
Exercise 2. Run the sequence60
6Quantitative Data Analysis61
Exercise 1. Create a method for quantification62
Task 1. Create a new method62
Task 2. Set up the signal for quantification63
Task 3. Integrate the low-level standard65
Task 4. Set general calibration parameters67
Task 5. Set up the calibration curve68
Task 6. Explore options to refine the calibration72
Exercise 2. Process a sample and print a report73
6Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System
Familiarization Guide
1
Prepare for the Analysis
Exercise 1. Prepare the LC to run the sample 8
Task 1. Start up ChemStation 8
Task 2. Purge the pump 9
Task 3. Prepare the column for the analyses 10
Exercise 2. Prepare the samples for the analyses 12
Exercise 3. Check the current MS tune values and adjust if necessary 14
This chapter presents exercises to help you learn how to:
• Prepare the LC and column for an analysis
• Prepare the samples that you analyze in these exercises
• Check the tune settings of the MS and adjust if necessary.
Before you start
• Order the sample: Agilent Electrospray LC Demo Sample,
p/n 59987-20033.
• Order the column: Agilent ZORBAX SB-C18, 2.1 mm x 30 mm,
3.5 µm, p/n 873700-902.
• You may use another similar column, but you may need to
adjust the HPLC conditions to obtain good separation.
• Make sure that the electrospray source is installed.
• Read the Agilent 6100 Series Quadrupole LC/MS Systems
Quick Start Guide and Chapter 2 of the Agilent 6100
Series Quadrupole LC/MS Systems Concepts Guide.
Agilent Technologies
7
1Prepare for the Analysis
Exercise 1. Prepare the LC to run the sample
Exercise 1. Prepare the LC to run the sample
For the following tasks, try the steps in the first column. If you
need more help, follow the detailed instructions in the middle
column.
Task 1. Start up ChemStation
StepsDetailed Instructions Comments
1 Open the ChemStation window.• Click the ChemStation icon on the
desktop.
Alternate method:
• From the Start menu, select:
All Programs > Agilent
ChemStation > Instrument 1
online.
8Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Prepare for the Analysis1
Task 2. Purge the pump
Task 2. Purge the pump
Use these instructions with the binary and quaternary
pumps. See the ChemStation online Help for instructions for
the capillary and nanoflow pumps.
StepsDetailed Instructions Comments
1 Display the Method and Run Control
view.
2 Place the pump in standby mode.a Click More Pump > Control HPLC
3 Prepare solvents used in these
familiarization exercises.
• A – 5 mM ammonium formate in
water
• B – 5 mM ammonium formate in
methanol
4 Replace the solvent bottles with the
ones you just prepared.
5 Open the purge valve.a Turn the black purge valve on the
6 Enter a flow of 5 mL/min and
50% B, using water in channel A and
methanol in channel B.
• In the view selection area in the
lower left, click Method and Run
Control.
Pump on the Instrument menu to
open the Pump Control dialog box.
b Select Standby and click OK.
a Into a 1-liter reservoir of
HPLC-grade water, add 1 mL of 5 M
ammonium formate.
b Into a 1-liter reservoir of
HPLC-grade methanol, add 1 mL of
5 M ammonium formate.
• Replace the bottles for channels A
and B.
front of the pump counter-clockwise
two turns.
b Place the tubing that exits the pump
into a 250-mL or larger beaker.
a Click the pump icon.
b Select Set up Pump.
c Enter the parameters in step 6 and
click OK.
Alternate method:
• Select Standby from the Pump
context menu.
• The part number for ammonium
formate is G1946-85021.
• Each ampoule contains 2.2 mL of
ammonium formate solution.
• Be sure to use HPLC-grade
solvents.
7 Turn the pump on and monitor the
tubing for bubbles.
a To turn the pump on, click the
little button to the lower right
of the solvent delivery (pump) icon.
b Monitor for bubbles.
• Purge for about 3 minutes to pass
3X the volume for the binary pump.
• If you wish, you may purge each
channel individually first, to ensure
that neither is air-locked.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide9
1Prepare for the Analysis
Task 3. Prepare the column for the analyses
StepsDetailed Instructions Comments
8 After the bubbles are gone and the
purge is complete, enter a flow of
1 mL/min and 100% B.
9 Purge a short while longer, and then
close the purge valve.
a Right-click the pump icon.
b Select Set up pump.
c Enter the new parameters in step 8,
and click OK.
a Continue to purge for a short while.
b Close the black valve.
For more information on purging the
pump, see the reference manual that
you received with your pump.
Task 3. Prepare the column for the analyses
In the exercises in the next chapters, you analyze a mixture
of four sulfonamide compounds. To perform the analyses in
the following chapters, you must first condition and
equilibrate your column.
StepsDetailed Instructions Comments
1 Disconnect the column from the
detector and MS.
a Turn the pump off by clicking
the little button to the lower
right of the solvent delivery
(pump) icon.
b Disconnect the column from the
detector and MS.
c Place the open end of the tubing
that exits the column into the
beaker.
• To prevent detector contamination,
allow the column effluent to go
directly to the waste beaker.
2 Flush the column with 100%
methanol at 1 mL/min (5 to 10 min).
• ZORBAX SB-C18, 2.1 mm
3.5 µm, p/n 873700-902
× 30 mm,
a Tur n t h e pum p o n.
b Flow methanol through the column
under the conditions used in Task 2,
step 8.
• The data sheet shipped with the
column cartridge recommends that
you flush with 20 to 30 columnvolumes of 100% methanol
(approximately 5 to 7.5 mL).
10Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Prepare for the Analysis1
Task 3. Prepare the column for the analyses
StepsDetailed Instructions Comments
3 Condition the column as follows,
using the solvents made up in Task 2,
step 6:
• Flow rate – 0.4 mL/min
• 100% B for 1/2 hour
• 50% B for 1/2 hour
4 Equilibrate the column at the analysis
conditions:
• 12% B for 1/2 hour at 40 °C
5 While the column equilibrates, set
parameters for the MS spray chamber
so it can heat and equilibrate as well.
• Drying gas flow: 8 L/min
• Nebulizer pressure: 35 psig
• Drying gas temperature: 300
• Capillary voltage: 3000 V
For 6150 with Agilent Jet Stream
technology:
• Sheath Gas Flow: 12 L/min
• Sheath Gas Temp: 360°C
• Nozzle Voltage: 0 V
o
C
a Click Set up Instrument Method on
the Instrument menu to open the
Setup Method dialog box.
b Click the Pump tab.
c Enter the flow rate in step 3.
d For Solvent B, type 100 and click
Apply.
e Wait 30 minutes.
f For Solvent B, type 50 and click
Apply.
g Wait 30 minutes.
a For Solvent B, type 12 and click
OK.
b Click the TCC tab on the Setup
Method dialog box.
c For Temperature, type 40 and click
OK.
a Right-click the MSD
icon on the system
diagram and select
Spray Chamber.
b Enter the parameters described in
step 5.
c Click OK.
d Wait 10 minutes before you tune the
MS.
• At a flow rate of 0.4 mL/min, the
checkout column should produce
about 70 to 80 bar pressure
(measured without any fittings at
the column exit).
• If, after you perform these steps, the
pump pressure through the column
is too high, order a replacement
SB-C18 column (p/n 873700-902).
• If your column is not new, you can
reduce the length of time that you
condition the column.
• While you condition and equilibrate
the column, you may complete
step 5 in this exercise and then
work on the rest of the exercises in
this chapter. Be sure to complete
step 6 before you go on to the next
chapter.
6 Reconnect the column to the DAD
and MS.
• You can complete “Exercise 3.
Check the current MS tune values
and adjust if necessary” either with
or without the column connected to
the DAD and MS, but you do need to
reconnect prior to the exercises in
Chapter 2, “Set Up and Run a Scan
Method.”
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide11
1Prepare for the Analysis
N
S
O
OH
N
2
H
S
N
N
N
N
N
S
O
OH
N
2
H
N
N
N
S
O
OH
N
2
H
Cl
N
N
O
O
N
S
O
OH
N
2
H
Exercise 2. Prepare the samples for the analyses
Exercise 2. Prepare the samples for the analyses
In the exercises in the next chapters, you analyze a mixture
of four sulfonamide compounds. The Electrospray LC Demo
Sample (p/n 59987-20033), contains five ampoules with
100 ng/µL each of these compounds:
To perform the analyses in the following chapters, you must
first prepare the sample at various dilutions. The final
concentrations will be 1, 5 and 10 ng/µL. You will also
prepare a solvent blank.
12Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Prepare for the Analysis1
Exercise 2. Prepare the samples for the analyses
StepsDetailed Instructions Comments
1 Prepare a 1:10 dilution of the original
sample in a 1-mL autosampler vial.
• Final concentration is 10 ng/µL
• You will use this sample for the
scan analysis in Chapter 2, and for
the SIM analyses in Chapter 4 and
Chapter 5.
2 Prepare a 1:20 dilution of the original
sample in a 1-mL autosampler vial.
• Final concentration is 5 ng/µL
• You will use this sample for the SIM
analysis in Chapter 5.
3 Prepare a 1:100 dilution of the original
sample in a 1-mL autosampler vial.
• Final concentration is 1 ng/µL
• You will use this sample for the SIM
analysis in Chapter 5.
4 Prepare a solvent blank in a 1-mL
autosampler vial.
• You will use this sample for the SIM
analysis in Chapter 5.
a Transfer 100 µL of the sulfa mixture
into the autosampler vial.
b Add 900 µL of 90:10 water:methanol
that contains 5 mM ammonium
formate (NH
c Cap the vial.
a Transfer 50 µL of the sulfa mixture
into the autosampler vial.
b Add 950 µL of 90:10 water:methanol
that contains 5 mM ammonium
formate.
c Cap the vial.
a Transfer 10 µL of the sulfa mixture
into the autosampler vial.
b Add 990 µL of 90:10 water:methanol
that contains 5 mM ammonium
formate.
c Cap the vial.
a Into the autosampler vial, transfer
990 µL of 90:10 water:methanol that
contains 5 mM ammonium formate.
b Cap the vial.
HCO2).
4
• The original sulfa mixture is
dissolved in a solvent mixture of
70% water and 30% acetonitrile.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide13
1Prepare for the Analysis
Exercise 3. Check the current MS tune values and adjust if necessary
Exercise 3. Check the current MS tune values and adjust if necessary
The MS is very stable and does not need to be tuned very
often. You can usually tune just once a month, or once a
week at most. You can use the Check Tune program
described in this exercise to confirm that the MS is in
adjustment.
StepsDetailed InstructionsComments
1 Switch to the MSD Tune view.• In the view selection area in the
lower left, click MSD Tune.
2 Select the tune file.a In the Select Tune File dialog box,
select ATU NE S.TU N.
b Keep the default of Positive Polarity
(Standard).
c Click OK.
d In the status bar near the top of the
MSD Tune view, verify that you see
the following:
• Mode is API-ES
• Source is ESI (electrospray)
3 Run a Check Tune.• From the Tune menu, select Check
Tune.
Note that Check Tune requires
values for comparison that are
determined from a previous
Autotune. Autotune is normally run
during installation.
4 If Check Tune report suggests that
you adjust peak widths or mass axis,
then do that.
5 If the Check Tune report shows poor
sensitivity, which indicates that your
MS settings are significantly out of
adjustment, then run a full Autotune.
a From the Tune menu, select Adjust
Mass Peak Width.
b From the Tune menu, select
Calibrate Mass Axis.
• From the Tune menu, select
Autotune > Positive Polarity.
• Make sure that you use an
appropriate calibrant with an
appropriate source.
• Check Tune is normally all that you
need to do to confirm that the MS
settings are correct.
• If Check Tune indicates a problem
with your MS settings, then
proceed to step 4 and/or step 5.
• The exercises in this manual use
only the positive ion mode and
standard scan speeds, so it is not
necessary to tune for negative
polarity or fast scan.
14Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System
Familiarization Guide
2
Set Up and Run a Scan Method
Exercise 1. Set up a full-scan acquisition method 16
Task 1. Enter LC acquisition parameters 16
Task 2. Enter MS acquisition parameters 19
Exercise 2. Acquire data with the full-scan method 23
Task 1. Enter sample information 24
Task 2. Acquire the data 25
These exercises show you how to set up a scan data
acquisition method for the demonstration sample (sulfa mix)
and to acquire data with that method.
The LC parameters that you enter in these exercises are
appropriate for the standard Agilent 1100/1200/1260/1290
Series liquid chromatography (LC) systems. You must enter LC
parameters that are appropriate for your LC model.
To view the results of these exercises, see Chapter 3,
“Qualitative Data Analysis.”
Before you start
• Review the Agilent 6100 Series Quadrupole LC/MS
Systems Quick Start Guide and Chapter 3 of the Agilent
6100 Series Quadrupole LC/MS Systems Concepts Guide.
• Prepare the LC, column and sample as described in
Chapter 1, “Prepare for the Analysis.”
For the tasks on the following pages, try the steps on the left
without the detailed instructions. If you need more help, follow
the detailed instructions on the right.
Agilent Technologies
15
2Set Up and Run a Scan Method
Exercise 1. Set up a full-scan acquisition method
Exercise 1. Set up a full-scan acquisition method
This exercise changes the default method and saves it as a new
method. This exercise consists of the following tasks:
• “Task 1. Enter LC acquisition parameters” on page 16
• “Task 2. Enter MS acquisition parameters” on page 19
Task 1. Enter LC acquisition parameters
StepsDetailed InstructionsComments
1 Display the Method and Run Control
view.
2 Open the method DEF_LC.M.a Select File > Load > Method.
3 Save the method under a new name,
SULFAMSSCAN1.M.
4 Enter a volume of
injection.
1 µL for the
• In the view selection area in the
lower left of the ChemStation
window, click Method and Run
Control.
b If necessary, navigate to
C:\CHEM32\1\METHODS.
c Select DEF_LC.M and click OK.
a Select File > Save As > Method.
b In the dialog box, for Name, type
SULFA MS SCAN 1.M.
c Click OK.
d In the box for Comment for method
history, type a comment.
e Click OK.
a Click Set up Instrument Method on
the Instrument menu to open the
Setup Method dialog box.
b Click the ALS tab.
c Click Standard injection.
d In the Injection volume box, type 1
for a 1-µL injection.
• You save the method now with a
new name to avoid inadvertently
overwriting the default method
later.
16Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method2
Task 1. Enter LC acquisition parameters
StepsDetailed InstructionsComments
5 Enter pump parameters.a Click the Pump tab on the Setup
Method dialog box.
b Set the parameters as follows:
Flow=0.400mL/min
StopTime=7.00 min
PostTime=3.00min
Solvent A=Water 88%
Solvent B=Methanol 12%
6 Set up the gradient timetable as it
appears in the figure below.
7 Enter a column compartment
temperature of 40
o
C.
8 Enter parameters for the diode-array
detector (DAD).
a Open the Timetable area in the
lower part of the tab, click Insert,
and type the first line.
b Click Append and type the second
line.
c Repeat step b for lines 3 and 4.
a Click the TCC tab on the Setup
Method dialog box.
b Click the option button for oC.
c Ty p e 40.0 for
o
C.
a Click the DAD tab on the Setup
Method dialog box.
b Enter the parameters shown below:
• Use Signal A: Wavelength 272nm,
Bandwidth16 nm
• Reference Wavelength: 360 nm,
Reference Bandwidth 100 nm
• Spectrum Store: All in peak
• Peakwidth: > 0.1 min
c Click OK to close the Setup Method
dialog box with the new setpoints.
Set the timetable parameters as
follows:
Line 1 Time 1:00, %B=12, Flow=0.4
Line 2 Time 3:00, %B=100,
Flow=0.4
Line 3 Time 6:00, %B=100,
Flow=0.4
Line 4 Time 7:00, %B=12, Flow=0.4
• The DAD is used in this example,
but the variable wavelength
detector (VWD) may be used
analogously.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide17
2Set Up and Run a Scan Method
Task 1. Enter LC acquisition parameters
StepsDetailed InstructionsComments
9 Select Data Acquisition only in the
Run Time Checklist.
10 Save the new parameters to the
method file, SULFA MS SCAN 1.M.
a Click Run Time Checklist on the
RunControl menu.
b Mark the Data Acquisition check
box.
c Click OK.
a Select File > Save > Method.
b In the box for Comment for method
history, type a comment.
c Click OK.
• While it is common to include Data
Analysis in the Run Time Checklist,
for these exercises, you will view
the results in Chapter 3,
“Qualitative Data Analysis.”
18Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method2
Set to 100 for 6120
Set to 125 for 6130 or 6150
Task 2. Enter MS acquisition parameters
Task 2. Enter MS acquisition parameters
StepsDetailed InstructionsComments
1 Enter parameters for the quadrupole
mass spectrometer (MS):
• Signal 1, scan mode, positive
polarity
• Scan range: 100 to 500
• Fragmentor:
100 V for the Agilent 6120; 125 V for
the 6130 or 6150
• Gain: 1.00
• Threshold: 150
• Stepsize: 0.10
• Peakwidth: 0.05 min
• Scan data storage: Condensed
• Active signals: 1 only
a Right-click the MSD
icon on the system
diagram and select Set
up MSD Signals.
b Enter the parameters described in
step 1 and shown in the figure
below. Take care to enter the
appropriate Fragmentor voltage for
your MS model.
c Click OK.
• To save disk space you usually acquire line spectra (Scan Data
Storage = Condensed). However,
when you acquire spectra from
intact proteins or protein
digests/peptides, you must acquire
and deconvolute profile spectra.
(Scan Data Storage = Full).
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide19
2Set Up and Run a Scan Method
Task 2. Enter MS acquisition parameters
StepsDetailed InstructionsComments
2 Enter parameters for the spray
chamber of the ion source:
• Drying gas flow: 9 L/min
• Nebulizer pressure: 40 psig
• Drying gas temperature: 300
• Capillary voltage: 3000 V
For 6150 with Agilent Jet Stream
technology:
• Nebulizer pressure: 30 psi
• Drying gas flow: 7 L/min
• Drying gas temperature: 350°C
• Sheath gas flow: 12 L/min
• Sheath gas temperature: 360 C°
• Capillary voltage: 4000 V
• Nozzle voltage: 0 V
• Fragmentor: 200 V
• Multiplier gain: 3
a Right-click the MSD
icon on the system
diagram and select
o
C
Spray Chamber.
b Enter the parameters described in
step 2 and shown in the figure
below.
c Click OK.
For all models except 6150 with Agilent Jet Stream technology
For 6150 with Agilent Jet Stream technology
20Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method2
Task 2. Enter MS acquisition parameters
StepsDetailed InstructionsComments
3 Set up to store the fragmentor voltage
throughout the run.
4 Save the method.a Select Method > Save Method to
a Right-click the MSD
icon on the system
diagram and select
Data Curves.
b Select Fragmentor - 1.
c Click the Add button.
d Click OK.
overwrite the method
SULFA MS SCAN 1.M.
b In the box for Comment for method
history, type a comment.
c Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide21
2Set Up and Run a Scan Method
Task 2. Enter MS acquisition parameters
StepsDetailed InstructionsComments
5 Print the method.a Select Method > Print Method.
b Mark the check boxes as shown in
the figure below.
c Click the Print button.
22Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method2
Exercise 2. Acquire data with the full-scan method
Exercise 2. Acquire data with the full-scan method
Now you are ready to acquire data for the sulfa mix with the
method you just created. This exercise consists of the following
tasks:
• “Task 1. Enter sample information” on page 24
• “Task 2. Acquire the data” on page 25
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide23
2Set Up and Run a Scan Method
Task 1. Enter sample information
Task 1. Enter sample information
StepsDetailed InstructionsComments
1 Display the Single Sample toolbar.• In the top toolbar, click
the single sample icon.
1 Display the Sample Information dialog
box.
2 Enter the sample information:
• Operator name
• Subdirectory: Sulfas
• Prefix: Sulfa_scan
• Location: Vial 1
• Sample Name: Sulfas 10 ng/µ
• Comment: Scan familiarization
exercise
L
a Click Sample Info on the
RunControl menu.
a Enter the parameters described in
step 2 and shown in the figure
below.
b Click OK.
• If you select Prefix/Counter, the
file names increment automatically
from one run to the next.
24Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Scan Method2
Task 2. Acquire the data
Task 2. Acquire the data
StepsDetailed InstructionsComments
1 Place the vial of sulfa sample you
prepared at 10 ng/µL into position 1
in the autosampler.
2 Inject the sulfa mix sample.• Click the Single Sample button to
start the run.
3 Monitor the total ion chromatogram
and the UV chromatogram during
data acquisition.
a From the Online Plot window, click
the Change button.
b In the list of Available Signals,
select DAD A: Signal=272,16
Reference=360,100 and click Add.
c In the list of available signals, select
MSD: Signal 1 and click Add.
d Monitor the MS signal to ensure a
stable baseline.
• You prepared this sample in
“Exercise 2. Prepare the samples for
the analyses” on page 12.
This button is present only when you
have selected Single Sample mode
from the top toolbar.
• If the baseline fluctuation for the
MS signal is greater than 10%, the
nebulizer and source chamber may
require maintenance. See the
Agilent 6100 Series Single Quad
LC/MS System Maintenance Guide.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide25
2Set Up and Run a Scan Method
Task 2. Acquire the data
StepsDetailed InstructionsComments
4 Save the signals for the Online Plot
window.
5 When the analysis is done, view the
results.
a In the Edit Signal Plot dialog box,
click the Apply to Method button.
b Save the method.
• To view the results, go to the next
exercise.
• The C18 column may require one or
two injections of the sample to be
fully conditioned. During these
initial injections, everything may be
eluted from the column in the void
volume. Repeat the process and
separation will occur.
26Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System
Familiarization Guide
3
Qualitative Data Analysis
Exercise 1. Display and manipulate chromatograms 28
Exercise 2. Examine mass spectra 31
Exercise 3. Integrate the chromatogram 36
Exercise 4. Print a report 40
This chapter shows you how to analyze data when you need
to identify or confirm sample components.
These exercises use the data file
you generated in Chapter 2.
Alternatively, you can use the sulfa
demo data file that you received
with the ChemStation software.
Before you start
• Read the Agilent 6100 Series Quadrupole LC/MS Systems
Quick Start Guide.
• Read the chapter on Data Analysis in the Agilent 6100
Series Quadrupole LC/MS Systems Concepts Guide.
• Set up and run the acquisition method in Chapter 2, “Set Up
and Run a Scan Method” or that you have the mssulfas.d data
file in the MSDEMO data folder on your system.
For the tasks on the following pages, perform the exercises
in the order they appear. Try the steps on the left without
the detailed instructions. If you need more help, follow the
detailed instructions on the right.
Agilent Technologies
27
3Qualitative Data Analysis
Exercise 1. Display and manipulate chromatograms
Exercise 1. Display and manipulate chromatograms
In this exercise, you load chromatograms and change the
chromatographic display.
StepsDetailed InstructionsComments
1 Display Data Analysis view.• In the view selection area of the
ChemStation window, click Data
Analysis.
2 Load the method
SULFAMSSCAN1.M.
3 Display the Signal Toolset.• Click the Signal icon,
a Select File > Load > Method.
b Navigate to the folder
C:\CHEM32\1\METHODS.
c Select the method file and click OK.
which is near the middle
of the window.
• If you just completed the previous
exercise, that method is already
loaded.
28Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Qualitative Data Analysis3
Exercise 1. Display and manipulate chromatograms
StepsDetailed InstructionsComments
4 Do one of the following:
• Open the data file,
SULFA_SCAN00001.D, which you
acquired in Chapter 2.
• Open the data file mssulfas.d,
located in the MSDEMO folder.
a Select File > Load Signal.
b Navigate to the appropriate folder,
either:
• C:\CHEM32\1\DATA\SULFAS,
or
• C:\CHEM32\1\DATA\MSDEMO.
c Select the data file.
d Set other parameters as shown
below and click OK.
• For other ways to load signals, see
the Data Analysis chapter in the
Concepts Guide.
• If you wish to complete Chapter 4,
“Set Up and Run a SIM Method”,
then you need to process the data
file you generated in Chapter 2. You
need the report from that data file to
set up your SIM groups.
5 Verify that you see the DAD and MS
chromatograms.
a Check that you see a display that is
similar to the one shown below.
b Verify that you see the DAD signal
in the top chromatogram.
c Confirm that you see the MSD
signal in the bottom chromatogram.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide29
3Qualitative Data Analysis
a
b
Exercise 1. Display and manipulate chromatograms
StepsDetailed InstructionsComments
6 Change the chromatogram view so
that the MS and UV chromatograms
are overlaid in the display.
7 Remove the DAD signal from the
display.
a In the Signal Toolset near
the middle of the window,
click the icon to display
overlaid signals.
b Check that you see the overlaid
chromatograms, as shown below.
c Click the icon to display
separate signals.
a In the Navigation Table, click the +
to display more information.
b Under the Signals tab, double-click
the signal labeled MSD1 TIC.
c When you see the message about
the method, click OK.
d Verify that the DAD window is gone
and only the TIC is displayed.
• The icon in step a is also
available in the Graphics
Toolset, but in that toolset
it toggles overlaid/separate. You
click the icon shown above to turn
on the display of the Graphics
Toolset.
• If you do not see the
Navigation Table shown
below, in the top toolbar,
click the icon shown above.
• For other ways to remove signals,
see the chapter on Data Analysis in
the Concepts Guide.
30Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Qualitative Data Analysis3
Exercise 2. Examine mass spectra
Exercise 2. Examine mass spectra
In this exercise, you learn to display mass spectra. You
choose background (reference) spectra that you can later
subtract from the spectra of the peaks of interest. You learn
how to display a single spectrum and an averaged spectrum
for a peak.
StepsDetailed InstructionsComments
1 Zoom in on the first peak in the
chromatogram.
2 Display the Spectrum Toolset.a Click the Spectrum
a Click the icon to zoom in.
b Use the mouse pointer to
draw a rectangle around
the peak. Take care to include the
chromatographic baseline.
c Check that your peak looks similar
to the one below.
d Note the width of the peak at half
height. You will need this
information to set up the SIM
analysis in Chapter 4.
icon, which is near
the middle of the window.
b If there is not room under your
chromatogram window to display
spectra, use your mouse pointer to
reduce the height of the
chromatogram window.
• If you want to try again, you can
zoom back out. Do one of the
following:
• Double-click the chromatogram
window.
• Click the icon to zoom
out.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide31
3Qualitative Data Analysis
Exercise 2. Examine mass spectra
StepsDetailed InstructionsComments
3 Get the first reference spectrum, to
the left of the peak.
4 Get a second reference spectrum, to
the right of the peak.
5 View your reference spectra. a If you cannot see the spectra, adjust
a To s e le ct th e f ir st
reference spectrum,
click the icon that is
highlighted here.
b In the chromatogram window, do
one of the following at the
chromatographic baseline just
before the peak:
• Click to select a single spectrum.
• Click and drag to select an
average spectrum.
a To select the second
reference spectrum,
click the icon that is
highlighted here.
b In the chromatogram window, do
one of the following at the
chromatographic baseline just after
the peak:
• Click to select a single spectrum.
• Click and drag to select an
average spectrum.
the size and location of the window
labeled Reference Mass
Spectrum(a).
b Note the two background spectra
— one before the peak and one
after.
32Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Qualitative Data Analysis3
Exercise 2. Examine mass spectra
StepsDetailed InstructionsComments
6 Set the spectral options to do manual
background subtraction.
a Click the icon to display the
Spectral Options dialog
box.
b Click the MS Reference tab.
c Under Reference Spectrum, click
Manual.
d Mark the check boxes for Ref1 and
Ref2. Note that the time ranges of
the reference spectra that you just
selected are specified there.
e Click OK.
• The spectral options apply to all
subsequent spectra until you
change the options.
• If the chromatographic baseline
changes over the course of the run,
select new reference spectra that
are close in time to each peak of
interest.
• Near the middle of the Data
Analysis window, you can view and
change the setting for background
subtraction.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide33
3Qualitative Data Analysis
Exercise 2. Examine mass spectra
StepsDetailed InstructionsComments
7 Get a single background-subtracted
spectrum for the first LC peak.
a Click the icon to get a mass
spectrum at any time point.
b In the chromatogram
window, click somewhere on the
peak to get the spectrum.
c If necessary for easier viewing,
adjust the size and location of the
window labeled MS Spectrum.
d Verify that the spectrum is similar to
the one shown below.
• Under the conditions used to
acquire the demo data file
(mssulfas.d), the compounds elute
in the following order:
• Depending on the organic mobile
phase and the modifiers, the elution
order for the 279 and 285 may
change.
34Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Qualitative Data Analysis3
Exercise 2. Examine mass spectra
StepsDetailed InstructionsComments
8 Get a average background-subtracted
spectrum for the first LC peak.
9 Be sure to see step 6 in “Exercise 3.
Integrate the chromatogram” for an
easier, faster way to display spectra.
a Click the icon to get an
averaged mass spectrum.
b In the chromatogram
window, click and drag the mouse
across the peak, as shown below.
c View the average spectrum in the
window labeled MS Spectrum.
• When a chromatographic peak
consists of a single compound, an
average spectrum is usually more
accurate.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide35
3Qualitative Data Analysis
Exercise 3. Integrate the chromatogram
Exercise 3. Integrate the chromatogram
In this exercise, you learn to set integration events and
integrate the chromatogram. Even if you do not care about
quantitation, integration helps locate peaks for other
purposes. For example, after integration, mass spectra of
each peak can be printed with a report.
StepsDetailed InstructionsComments
1 Display the total ion chromatogram in
its entirety.
2 Display the Integration Toolset.• Click the Integration
a Minimize any spectral windows that
hide the chromatogram window.
b Click the icon to zoom out.
icon, which is near
the middle of the window.
36Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Qualitative Data Analysis3
Exercise 3. Integrate the chromatogram
StepsDetailed InstructionsComments
3 Integrate the chromatogram.a Click the Auto Integrate
icon, which is near the
middle of the window.
b Verify that the results are similar to
those shown below.
• Auto Integrate estimates initial
integration parameters.
• If you do not see the
retention times, in the
graphics tools, click the
icon to display retention times.
• If you do not see the pink
integration baseline, in the
graphics tools, click the
icon to display baselines.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide37
3Qualitative Data Analysis
Exercise 3. Integrate the chromatogram
StepsDetailed InstructionsComments
4 Adjust the integration parameters to
get only four integrated peaks.
a Click the icon for Edit/Set
Integration Events Table.
b In the Integration Events
table, for Baseline Correction,
select Advanced.
c For Height Reject, type 500000.
d Click the Integrate current
Chromatogram icon.
e Verify that your results are
similar to those shown below.
• For detailed information about
integration events, see Agilent
ChemStation: Understanding Your
ChemStation.
38Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Qualitative Data Analysis3
Exercise 3. Integrate the chromatogram
StepsDetailed InstructionsComments
5 Save the integration events to the
method in memory.
6 Use the integrated chromatogram as
the basis for a faster way to display
background-subtracted spectra.
• Click the icon to exit and
save the integration results.
a Click the Spectrum
icon.
b Click the icon to display the
Spectral Options dialog
box.
c Click the MS Reference tab.
d Under Reference Spectrum, click
Automatic.
e Click OK.
f Click the icon to get a mass
spectrum at the peak apex.
g In the chromatogram
window, click somewhere on the
fourth peak to get the spectrum.
h Verify that the spectrum is similar to
the one shown below.
• To save the events to the method on
disk, you also need to save the
method to disk, as described in
step 3 on page 41.
• When you set Reference Spectrum,
to Automatic, the software
automatically takes the reference
spectra for each peak, as described
in the Spectral Options dialog box.
• The icon to get the mass spectrum
at the peak apex is available only if
you have integrated the
chromatogram. No matter where
you click on the peak, it gets the
spectrum at the apex. With this tool,
you may not need to zoom in on the
chromatogram to achieve a precise
location for the spectrum.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide39
3Qualitative Data Analysis
Exercise 4. Print a report
Exercise 4. Print a report
In this exercise, you print a report, which you will use in
Chapter 4, “Set Up and Run a SIM Method.”
StepsDetailed InstructionsComments
1 Specify the LCMS Qualitative report
style, with the report printed to the
screen.
a Select Report > Specify Report.
b In the Specify Report dialog box,
under Destination, mark the check
box for Screen.
c For Report Style, select LCMS
Qualitative.
d Check that other settings are as
shown below.
e Click OK.
40Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Qualitative Data Analysis3
Exercise 4. Print a report
StepsDetailed InstructionsComments
2 Print the report.a Select Report > Print Report.
b After a short wait, view the Report
window.
c Verify that page 1 of the report
contains header information and an
integrated chromatogram.
d At the bottom of the report window,
click the Next button.
e Verify that page 2 of the report
shows extracted ion
chromatograms and a mass
spectrum of the first
chromatographic peak.
f Continue to click the Next button to
view results for the three additional
chromatographic peaks.
g At the bottom of the report window,
click the Print button. This prints a
hard copy of the report.
h At the bottom of the report window,
click the Close button.
3 Save the method.a Select File > Save > Method to
overwrite the method
SULFA MS SCAN 1.M.
b In the box for Comment for method
history, type a comment.
c Click OK.
• If you wish to complete Chapter 4,
“Set Up and Run a SIM Method”,
then save the hard copy and refer to
it when you set up your SIM groups.
• The extracted ion chromatograms
are indicators of peak purity; if the
retention times fail to coincide, the
peak likely represents more than
one compound.
• You save the method now so that
your integration parameters,
spectral display options, report
settings, and other data analysis
settings become part of the method.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide41
3Qualitative Data Analysis
Exercise 4. Print a report
42Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System
Familiarization Guide
4
Set Up and Run a SIM Method
Exercise 1. Set up a SIM acquisition method 44
Task 1. Load the scan method you created previously 44
Task 2. Enter MS acquisition parameters 45
Exercise 2. Acquire data with the SIM method 48
Task 1. Enter sample information 49
Task 2. Acquire the data 50
These exercises show you how to set up a data acquisition
method that uses selected ion monitoring (SIM). You set up
the method for the demonstration sample (sulfa mix) and
then run the sample with that method.
To set up the SIM method, you modify the scan method that
you created in Chapter 2. To set up the SIM acquisition, you
need the following for each of the four sulfa compounds:
• The LC retention time
• The masses of ions in the spectrum
You get that information from the report you generated in
Chapter 3.
Before you start
• Complete the previous exercises in this manual.
Agilent Technologies
43
4Set Up and Run a SIM Method
Exercise 1. Set up a SIM acquisition method
Exercise 1. Set up a SIM acquisition method
In this exercise, you start with your existing scan method
and modify it for SIM analysis. You keep the same LC
conditions and modify only the MS conditions. This exercise
consists of the following tasks:
• “Task 1. Load the scan method you created previously”
(below)
• “Task 2. Enter MS acquisition parameters” on page 45
Task 1. Load the scan method you created previously
StepsDetailed InstructionsComments
4 Display Method and Run Control
view.
5 Open the method
SULFAMSSCAN1.M.
6 Save the method under a new name,
SULFA MS SIM 1.M.
• In the view selection area of the
ChemStation window, click Method
and Run Control.
a Select File > Load > Method.
b If necessary, navigate to
C:\CHEM32\1\METHODS.
c Select SULFA MS SCAN 1.M and
click OK.
a Select File > Save As > Method.
b In the dialog box, for Name, type
SULFA MS SIM 1.M.
c Click OK.
d In the box for Comment for method
history, type a comment.
e Click OK.
• You save the method now to avoid
inadvertent overwrites of your scan
method.
44Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a SIM Method4
Task 2. Enter MS acquisition parameters
Task 2. Enter MS acquisition parameters
StepsDetailed InstructionsComments
1 Enter the chromatographic peak
width for the SIM analysis.
a Right-click the MSD
icon on the system
diagram and select Set
up MSD Signals.
b When the Set Up MSD Signals
dialog box is displayed, type 0.05
For Peakwidth.
c Click OK.
• The peak width is an important
setting. It is used to calculate the
appropriate SIM dwell times to
deliver sufficient points across a
chromatographic peak to give good
quantitation.
• Peak width is defined as the full
width at half maximum (FWHM),
the width at 50% of the peak height.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide45
4Set Up and Run a SIM Method
Task 2. Enter MS acquisition parameters
StepsDetailed InstructionsComments
2 Set up the first SIM ions using the
masses (to the nearest 0.1) that you
observed in the spectra from your
scan analysis:
• Sulfamethizole: Time 0,
SIM Ions 271 and 156.
a Under MSD Signal Settings, Signal
1, for Mode, select SIM.
b In the table, for Fragmentor, type
one of the following:
• 150 for the Agilent 6120
• 200 for the Agilent 6130 or 6150
c In the table, change Group 1 to
Sulfamethizole, and for SIM Ion,
refer to the spectrum on your
printout and type the mass (to the
nearest 0.1) for the 271 ion.
d Click Add Ion, and type the mass for
the sulfamethizole 156 ion.
• In this example, each SIM group
includes a pseudo-molecular ion
and one fragment ion for
confirmation.
• Note that the figure below does not
show the fourth sulfa drug.
46Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a SIM Method4
Task 2. Enter MS acquisition parameters
StepsDetailed InstructionsComments
3 Set up the remaining SIM ions, using
the masses (to the nearest 0.1) that
you observed in the spectra from your
scan analysis:
• Sulfachloropyridazine: Time 1.3,
SIM Ions 285, 287, and 156.
• Sulfamethazine: Time 2.3,
SIM Ions 279 and 186.
• Sulfadimethoxine: Time 3.3,
SIM Ions 311 and 156.
4 Save the method.a Select Method > Save Method to
a Click Add Grp, and type the name,
start time and mass (approximately
285) for sulfachloropyridazine.
b Click Add Ion, and type the mass for
the sulfachloropyridazine 156 ion.
c Click Add Ion, and type the mass for
the sulfachloropyridazine 287 ion.
d Repeat these steps until you have
entered two or three ions for each of
the remaining compounds.
e Click OK to close the Set Up MSD
Signals dialog box.
overwrite the method
SULFA MS SIM 1.M.
b In the box for Comment for method
history, type a comment.
c Click OK.
• Alternatively, instead of making
separate groups for each compound
as described here, all of the SIM
ions could be entered into “Group
1", which could be re-named
“Sulfonamides”. The first SIM
group can contain up to 100 ions.
• You may need to adjust the start
time for each SIM group. Refer to
your printout from Chapter 3 to
determine a start time so that each
group change occurs about midway
between the chromatographic
peaks.
• If the retention time difference
between sulfachloropyridazine and
sulfamethazine is less than 0.3
minutes, merge these ions into one
group.
• The sulfachloropyridazine
additionally includes the chlorine
isotope at m/z 287.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide47
4Set Up and Run a SIM Method
Exercise 2. Acquire data with the SIM method
Exercise 2. Acquire data with the SIM method
Now you are ready to acquire data for the sulfa mix with
the method you just created. This exercise consists of the
following tasks:
• “Task 1. Enter sample information” on page 49
• “Task 2. Acquire the data” on page 50
48Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a SIM Method4
Task 1. Enter sample information
Task 1. Enter sample information
StepsDetailed InstructionsComments
1 Display the Single Sample toolbar.• In the top toolbar, click
the single sample icon.
1 Display the Sample Information dialog
box.
2 Enter the sample information:
• Operator name
• Subdirectory: Sulfas
• Prefix: Sulfa_SIM
• Location: Vial 1
• Sample Name: Sulfas 10 ng/µL
• Comment: SIM familiarization
exercise
a Click Sample Info on the
RunControl menu.
a Enter the parameters described in
step 2 and shown in the figure
below.
b Click OK.
• If you select Prefix/Counter, the
file names increment automatically
from one run to the next.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide49
4Set Up and Run a SIM Method
Task 2. Acquire the data
Task 2. Acquire the data
StepsDetailed InstructionsComments
1 Place the vial of sulfa sample you
prepared at 10 ng/µL into position 1
in the autosampler.
2 Inject the sulfa mix sample.• Click the Single Sample start
button.
3 Monitor the total ion chromatogram
and the UV chromatogram during
data acquisition.
4 When the analysis is done, view the
results.
a Activate the Online Plot window.
b Monitor the MS signal to ensure a
stable baseline.
a Display Data Analysis view.
b Load the data file you just created.
c Examine the DAD and MS
chromatograms.
• You prepared this sample in
“Exercise 2. Prepare the samples for
the analyses” on page 12.
This button is present only when you
have selected Single Sample mode
from the top toolbar.
• If the baseline fluctuation for the
MS signal is greater than 10%, the
nebulizer and source chamber may
require maintenance. See the
Agilent 6100 Series Single Quad
LC/MS System Maintenance Guide.
• If you need help, follow the general
procedure in “Exercise 1. Display
and manipulate chromatograms” on
page 28 in Chapter 3.
50Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System
Familiarization Guide
5
Set Up and Run a Sequence
Exercise 1. Set up a sequence 52
Task 1. Prepare to create a new sequence 52
Task 2. Edit sequence parameters 53
Task 3. Set up the sequence table 55
Task 4. Set up the sequence output 58
Exercise 2. Run the sequence 60
These exercises show you how to set up a sequence for the
SIM analysis of the demonstration sample (sulfa mix), and to
acquire data with that sequence.
In the sequence, you run the sulfa mix at three
concentrations: 1, 5 and 10 ng/µL. You also run a solvent
blank.
Before you start
• Read the Agilent 6100 Series Quadrupole LC/MS Systems
Quick Start Guide and Chapter 3 of the Agilent 6100
Series Quadrupole LC/MS Systems Concepts Guide.
• Complete the previous exercises in this manual.
For details about sequences, see the automation chapter in
Agilent ChemStation: Understanding Your ChemStation.
Agilent Technologies
51
5Set Up and Run a Sequence
Exercise 1. Set up a sequence
Exercise 1. Set up a sequence
Task 1. Prepare to create a new sequence
StepsDetailed InstructionsComments
1 Display Method and Run Control
view.
2 Display the Sequence Toolset.• In the top toolbar, click
3 Display the Autosampler Tray
diagram.
• In the view selection area of the
ChemStation window, click Method and Run Control.
the icon to display the
Sequence Toolset.
• Click Sampling Diagram on the View menu.
4 Initiate setup of a new sequence.• Select Sequence> New sequence.The empty default sequence file,
DEF_LC.S, is loaded automatically.
5 Save the sequence under a new
name, SULFA MS SIM 1.S
a Select Sequence > Save Sequence As.
b For Name, type SULFA MS SIM 1.S.
c Click OK.
52Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Sequence5
Task 2. Edit sequence parameters
Task 2. Edit sequence parameters
StepsDetailed InstructionsComments
1 Open Sequence Parameters dialog
box.
2 Enter the sequence parameters for
Operator Name and Data File.
• Click Sequence > Sequence
Parameters.
a Enter the following parameters,
shown in step 1.
• Operator name: You r na m e
• Subdirectory: Sulfas
• Prefix: Sulfa_seq
• The sequence parameters are
settings that are common to all the
samples in the sequence.
• To avoid overwrite of data files, type
a new subdirectory for each
sequence. The directory will be
created if it doesn’t already exist on
your computer.
• Unique file names are automatically
created for each data file within the
subdirectory.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide53
5Set Up and Run a Sequence
Task 2. Edit sequence parameters
StepsDetailed InstructionsComments
3 Enter the rest of the sequence
parameters:
a Enter the following parameters
shown in the figure below.
• Parts of methods to run:
According to Runtime Checklist
• Wait: 10 minutes after loading a
new method
• Shutdown: STANDBY
• Not Ready Timeout: 15 minutes
• Sequence Comment: Sequence
familiarization exercise
b Click OK.
• If you wanted to run only
reprocessing (data analysis), you
would set that in Part of methods to
run.
• The Wait allows the instrument to
equilibrate when a new method is
loaded.
Post-Sequence Command/Macros are
a convenient way to turn off lamps,
pumps, etc. The command or macro is
run at the end of the sequence or in the
event of an error.
Two examples of Post-Sequence
Command/Macros are:
• MSSetState is a command that can
change the MS state to standby.
See the online Help for commands.
• SHUTDOWN.MAC is a macro that
will shut down the system, but you
must customize it before using it.
54Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Sequence5
Task 3. Set up the sequence table
Task 3. Set up the sequence table
StepsDetailed InstructionsComments
1 Set up the sequence table to:
• Run duplicate injections of a blank.
• Run duplicate injections of the
sulfa mix at three concentrations:
1, 5 and 10 ng/µL.
• Use the method
SULFA MS SIM 1.M, that you
created in Chapter 4, “Set Up and
Run a SIM Method”.
a Click Sequence > Sequence Table.
b Select the first line of the sequence
table. In the sequence table, under
Line, click the number 1.
c Click the Cut button to delete the
line.
d Click the button for the
Insert/Filldown Wizard, shown
below.
e Fill in the values and click OK.
• In this step, you set up the parts of
the sequence table that are
common to all the samples.
• You will specify the sample names
later in this exercise.
• There are a number of ways to add
samples to the sequence table. This
exercise illustrates just one of the
ways — use of the Insert/Filldown
Wizard.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide55
5Set Up and Run a Sequence
Task 3. Set up the sequence table
StepsDetailed InstructionsComments
2 View the sequence table that you
have created so far.
3 (Optional) Customize the sequence
table to match the format in step 2.
a Compare your table with the one
below.
b Note any differences, such as
columns that are included and
column widths.
a In the lower right-hand
corner of the dialog box,
click the icon to customize
the sequence table.
b Clear the check boxes for any
unnecessary columns, as shown
below.
c Increase the width of the column for
the sample name, as shown below.
d Decrease the width of the column
for the method name, as shown
below.
e Click OK.
• Your results will likely differ, but in
the next step you may recreate the
table format below.
• For descriptions of any columns you
removed, see the online Help.
56Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Sequence5
Task 3. Set up the sequence table
StepsDetailed InstructionsComments
4 Type the following sample names into
the table:
• Vial 1 – blank
• Remaining vials – sulfa mix at 1, 5
and 10 ng/µL
5 Save the sequence.• Click the Save Sequence
a Modify the Sample Name for each
sample, as shown below.
b Click OK.
button in the Sequence
toolset.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide57
5Set Up and Run a Sequence
Task 4. Set up the sequence output
Task 4. Set up the sequence output
StepsDetailed InstructionsComments
1 Set up the sequence to print a short
sequence summary to the printer.
a Click Sequence > Sequence
Output.
b Mark the check box for Print
Sequence Summary Report.
c Mark the check box for Report to
Printer.
d Click the Setup button.
e Fill in the dialog box as shown
below.
f Click OK in the Sequence Summary
Parameters dialog box.
g Click OK in the Sequence Output
dialog box.
• In addition to the sequence
summary report, you can print
individual sample reports, as
specified in your method. (You do
not print individual reports in this
exercise.)
• For details about sequence reports,
see the chapter on ChemStation
reports in Agilent ChemStation: Understanding Your ChemStation.
• The setup shown in the dialog box
below prints the simplest summary
report.
2 Save the sequence.• Click the Save Sequence
button in the Sequence
toolset.
58Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Set Up and Run a Sequence5
Task 4. Set up the sequence output
StepsDetailed InstructionsComments
3 Print the sequence.a Select Sequence > Print Sequence.
b Mark the check boxes as shown in
the figure below.
c Click the Print button.
If you click the Print All button, you
print all the parts of the sequence
rather than the items you just
specified.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide59
5Set Up and Run a Sequence
Exercise 2. Run the sequence
Exercise 2. Run the sequence
Now you are ready to acquire data with the sequence you
just created.
StepsDetailed InstructionsComments
1 Confirm that your sequence includes
four samples.
2 Place the samples you prepared in
Chapter 1 into the appropriate
positions in the autosampler.
3 Inject the samples.• Click the Sequence
4 (Optional) For the first blank analysis,
monitor the total ion chromatogram
and the UV chromatogram during
data acquisition.
5 When the sequence is done, view the
Sequence Summary Report.
6 When the sequence is finished, view
the results.
• Verify that the Autosampler Tray
diagram shows four samples.
start button on the
Run Control Bar.
a Activate the Online Plot window.
b Monitor the MS signal to ensure a
stable baseline.
a Retrieve the report from the printer.
b Examine the report to confirm that
all the samples ran.
a Display Data Analysis view.
b Load the first data file you just
created.
c Examine the DAD and MS
chromatograms.
d Repeat step b and step c for the
other data files.
You prepared the samples in “Exercise
2. Prepare the samples for the
analyses” on page 12.
This button is only available if you
have selected Sequence mode on the
main toolbar.
As the sequence progresses, the
Autosampler Tray diagram is
color-coded as follows:
Gray - samples that have been
analyzed.
White - samples not yet analyzed.
Blue - the current sample.
• If you need help, follow the general
procedure in “Exercise 1. Display
and manipulate chromatograms” on
page 28 in Chapter 3.
• When you analyze your own
samples, you can set up your
method to automatically generate a
data analysis report for each sample
in the sequence.
60Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Agilent 6100 Series Quadrupole LC/MS System
Familiarization Guide
6
Quantitative Data Analysis
Exercise 1. Create a method for quantification 62
Task 1. Create a new method 62
Task 2. Set up the signal for quantification 63
Task 3. Integrate the low-level standard 65
Task 4. Set general calibration parameters 67
Task 5. Set up the calibration curve 68
Task 6. Explore options to refine the calibration 72
Exercise 2. Process a sample and print a report 73
This chapter shows you how to use the ChemStation Data
Analysis to perform quantification. The exercises in this
chapter illustrate a simple calibration that uses data files
that you received with your ChemStation software.
Before you start
• Read the Agilent 6100 Series Quadrupole LC/MS Systems
Quick Start Guide.
• Read the chapter on Data Analysis in the Agilent 6100
Series Quadrupole LC/MS Systems Concepts Guide.
• Make sure that you have the caffeine data files on your
ChemStation. Check for the files under
C:\CHEM32\1\DATA\MSDEMO. The file names are
CAFCAL0X.D, where x is a number from 1 to 5.
Agilent Technologies
61
6Quantitative Data Analysis
Exercise 1. Create a method for quantification
Exercise 1. Create a method for quantification
In this exercise, you create a calibrated method that you can
use to quantify caffeine in the demo data.
Task 1. Create a new method
In this task, you load a default method and save it to a new
name. You later modify the new method to create a
calibrated method.
StepsDetailed InstructionsComments
1 Display Data Analysis view.• In the view selection area in the
lower left of the ChemStation
window, click Data Analysis.
a Select File > Save As > Method.
b Navigate to the folder
C:\CHEM32\1\METHODS.
c In the dialog box, for Name, type
CAFFEINE CAL.M.
d Click OK.
e In the box for Comment for method
history, type a comment.
f Click OK.
62Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Quantitative Data Analysis6
Task 2. Set up the signal for quantification
Task 2. Set up the signal for quantification
In this exercise, you add an extracted ion chromatogram
(EIC) to the list of available signals for the method. Then
you add this EIC to the Signal Details, so you can
automatically load and integrate signals for the rest of the
caffeine standards.
StepsDetailed InstructionsComments
1 Open the data file CAFCAL01.D,
located in the MSDEMO folder.
2 Extract the major ion of caffeine. a Select File > Extract Ions.
3 Display the Calibration Toolset.• Click the Calibration
a Select File > Load Signal.
b Navigate to the folder:
C:\CHEM32\1\DATA\MSDEMO.
c Select the data file CAFCAL01.D.
d If necessary, clear the check box for
Load using Signal Details.
e In the Signals box, click the signal
that begins with MSD1 TIC.
f Click OK.
b For Ion 1, type 195.1.
c For Ion 2, type 195.1.
d Click OK.
icon, which is near
the middle of the window.
• For other ways to load signals, see
the chapter on Data Analysis in the
Concepts Guide.
+
• The 195 ion is the (M+H)
ion.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide63
6Quantitative Data Analysis
Task 2. Set up the signal for quantification
StepsDetailed InstructionsComments
4 Set up the signal for quantification.a Do one of the following:
• Click the icon to Edit
current method signals.
• Select Calibration >
Signal Details.
b From the list of Available Signals,
select MSD1 195, EIC=195.1:195.1.
c Click Add to Method.
d Click OK.
5 (Optional) Save the method under the
same name (CAFFEINE CAL.M).
a Select File > Save > Method.
b In the box for Comment for method
history, type a comment.
c Click OK.
• The EIC signal is available only
because you loaded the 195 EIC in
step 2.
• For these exercises, you save the
method frequently, but you could
wait instead until you had
established all the method settings.
64Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Quantitative Data Analysis6
Task 3. Integrate the low-level standard
Task 3. Integrate the low-level standard
In this exercise, you establish integration parameters for
your calibrated method. You use the low-level standard
because it is usually the most difficult to integrate.
StepsDetailed InstructionsComments
1 Display the Integration Toolset.• Click the
Integration icon,
which is near the middle of the
window.
2 Integrate the chromatogram.a Click the Auto Integrate
icon, which is near the
middle of the window.
b Check that you have five integrated
peaks with these initial settings.
• Auto Integrate estimates initial
integration parameters and then
performs the integration.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide65
6Quantitative Data Analysis
Task 3. Integrate the low-level standard
StepsDetailed InstructionsComments
3 Adjust the integration parameters to
get one integrated peak.
a Click the icon to Edit/Set
Integration Events Table.
b In the integration events for
all signals, for Baseline Correction,
select Advanced.
c Click the Auto Integrate
icon.
d When you are prompted to
save the events table, click Yes .
e Verify that your results are the same
or very similar to those shown
below.
• For detailed information about
integration events, see Agilent
ChemStation: Understanding Your
ChemStation.
4 Save the integration events with the
method in memory.
5 (Optional) Save the method under the
same name (CAFFEINE CAL.M).
• Click the icon to exit and
save the integration results.
a Select File > Save > Method.
b In the box for Comment for method
history, type a comment.
c Click OK.
66Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
a Select File > Save > Method.
b In the box for Comment for method
history, type a comment.
c Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide67
6Quantitative Data Analysis
Task 5. Set up the calibration curve
Task 5. Set up the calibration curve
In this exercise, you integrate the rest of the standards and
add all standards to the calibration curve.
StepsDetailed InstructionsComments
1 Display the Calibration Toolset.• Click the Calibration
icon, which is near
the middle of the window.
2 Add the low-level standard to the
calibration curve.
a Do one of the following:
• Click the New
Calibration Table icon.
• Select Calibration >
New Calibration Table.
b Click Automatic Setup Level 1.
c Click OK.
d In the Calibration Table pane
(shown below), under Compound,
type caffeine and under Amt
(amount), type 0.5.
• Do not worry at this point if your
calibration curve displays a
message that says the curve is
invalid.
68Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Quantitative Data Analysis6
Task 5. Set up the calibration curve
StepsDetailed InstructionsComments
3 Load and integrate the second
standard.
a Select File > Load Signal.
b Under File name, select
CAFCAL02.D.
c Mark the check box for Load using
Signal Details.
d Mark the check box for Integrate
after load.
e Check that your dialog box looks like
the one below.
f Click OK.
• These settings enable you in a
single step to load the appropriate
signal(s) and integrate them.
4 Add the second standard to the
calibration curve.
a Click the icon to Add new
level.
b In the dialog box, for
Default Amount, type 1 and click
OK.
c Verify that the calibration table now
has two entries, and the calibration
curve contains two points.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide69
6Quantitative Data Analysis
Task 5. Set up the calibration curve
StepsDetailed InstructionsComments
5 Add the remaining three standards to
the calibration table:
• CAFCAL03.D: 5 ng/µL
• CAFCAL04.D: 25 ng/µL
• CAFCAL05.D: 50 ng/µL
a Select File > Load Signal.
b Under File name, select the next
data file.
c Verify that the chromatogram is
properly integrated.
d Click the icon to Add new
level.
e In the dialog box, for
Default Amount, type the amount shown in step 5 and click OK.
f Verify that the calibration table and
the calibration curve contain the
new entry.
g Repeat step a through step f until
you have added all the standards.
h Confirm that your calibration table
and calibration curve look like the
ones below.
• If multiple peaks are integrated in a
chromatogram, retention time is
used to find the correct peak for the
calibration curve.
70Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Quantitative Data Analysis6
Task 5. Set up the calibration curve
StepsDetailed InstructionsComments
6 Refine the calibration curve.a Select Calibration > Calibration
Settings.
b Under Default Calibration Curve,
for Ty p e , select Quadratic.
c Click OK.
d Verify that your calibration curve
now looks like the one below.
7 (Optional) Save the method under the
same name (CAFFEINE CAL.M).
a Select File > Save > Method.
b In the box for Comment for method
history, type a comment.
c Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide71
6Quantitative Data Analysis
Task 6. Explore options to refine the calibration
Task 6. Explore options to refine the calibration
This exercise describes additional calibration table layouts
that give you more calibration options. You do not need
these options to process the caffeine demonstration data, but
you may need them when you process your own samples.
Peak Details.
b Verif y that you see these columns in
the calibration table:
• Curve Type
• Origin
• Weight
Identification Details.
b Verif y that you see these columns in
the calibration table:
• Resp % (response percent)
• +- (window for the response
percent)
• Pk Usage (peak usage)
a Select Calibration Table Options >
Overview.
b Verify that the calibration table
looks the same as in step 5 on
page 70.
• Note that this calibration table
layout lets you change:
• Curve Type: The type of
calibration curve (linear,
quadratic, etc.)
• Origin: How the origin (zero
point) is treated.
• Weight: The relative weights of
the data points.
• Note that this calibration table
layout lets you define:
• Pk Usage: How the calibration
uses the peak, for example, as a
main calibration ion or a qualifier
ion
• Resp %: The expected response
of the qualifier ion, as a
percentage of the main peak
• +-: A window for the expected
percentage.
72Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Quantitative Data Analysis6
Exercise 2. Process a sample and print a report
Exercise 2. Process a sample and print a report
In this exercise, you specify a report and test your
calibration method by processing one of the standards as if
it were a sample. You print a report of the results.
StepsDetailed InstructionsComments
1 Specify a report with the following
settings:
• Report destination: Screen
• External standard (ESTD)
calculation, based on area
• Report style: Detail
a Do one of the following:
• Select Report > Specify Report.
• Click the Specify Report
icon.
b Enter parameters as
described in step 1 and shown in
the figure below.
c Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide73
6Quantitative Data Analysis
Exercise 2. Process a sample and print a report
StepsDetailed InstructionsComments
2 Save the method under the same
name (CAFFEINE CAL.M).
3 Load the standard of medium
concentration.
4 Process the medium-level standard
and print the report.
a Select File > Save > Method.
b In the box for Comment for method
history, type a comment.
c Click OK.
a Select File > Load Signal.
b Under File name, select
CAFCAL03.D.
a Do one of the following:
• Select Report > Print Report.
• Click the icon to preview
results.
b Verify that page 1 of the
report contains header information,
an integrated chromatogram, and an
external standard report.
c Check that the caffeine amount is
about 5 ng/µL.
d At the bottom of the report window,
click the Next button.
e Verify that page 2 of the report
shows the calibration curve with
the measured point identified with
dotted lines.
f (Optional) At the bottom of the
report window, click the Print
button so you get a hard copy.
g At the bottom of the report window,
click the Close button.
• Another way to generate a
hard copy is to click the
Print Report icon.
74Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
www.agilent.com
In This Book
When you do the
exercises in this book,
you learn how to:
• Prepare your LC/MS
system for an analysis
•Set up methods for
scan and selected ion
monitoring analyses