No p art o f this manu al may be re produce d in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G1960-90079
Edition
Revision A, September 2011
Printed in USA
Agilent Technologies, Inc.
5301 Stevens Creek Blvd.
Santa Clara, CA 95051
Microsoft® is a U.S. registered trademark
of Microsoft Corporation.
Software Revision
This guide is valid for the B.04.02 SPI1 or
later revision of the Agilent
ChemStation software for the Agilent 6100
Series Quadrupole LC/MS systems, until
superseded.
If you have any comments about this guide,
please send an e-mail to
feedback_lcms@agilent.com.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user customers. Agilent provides this customary
commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
In This Guide...
The Concepts Guide presents an overview of the Agilent
6100 Series Quadrupole LC/MS systems, to help you
understand how the hardware and software work.
If you have any comments about this guide, please send an
e-mail to feedback_lcms@agilent.com.
1Overview
Learn how the hardware works in the Agilent 6100 Series
Quadrupole LC/MS systems, and get a brief introduction to
ChemStation software.
2Instrument Preparation
Learn the concepts you need to prepare the LC and column
for an analysis, and to tune the MS.
3Data Acquisition
Learn about setting up methods and running samples.
4Data Analysis
Learn the concepts you need for qualitative and quantitative
data analysis with ChemStation software.
5Reports
Learn about predefined results reports and about setting up
custom reports.
6Verification of Performance
Learn the concepts for Operational Qualification/
Performance Verification (OQ/PV) and system verification
with ChemStation software.
7Maintenance and Troubleshooting
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide3
Learn about tools that are proved in ChemStation software
to help you maintain your system and diagnose and fix
problems.
4Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Contents
1Overview of Hardware and Software9
How the Agilent quadrupole LC/MS systems work10
Overview10
Details11
Types of data you can acquire15
Scan versus selected ion monitoring (SIM)15
Generation of fragment ions: low versus high fragmentor16
Positive versus negative ions19
Multiple signal acquisition19
Purpose36
Summary of procedures36
Setting parameters for LC modules38
Column conditioning and equilibration39
Monitoring the stability of flow and pressure41
Preparation of the MS – tuning42
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide5
Contents
Overview42
Ways to tune44
When to tune – Check Tune45
Autotune47
Manual tuning49
Tune reports51
Gain calibration53
3Data Acquisition57
Working with methods58
Method and Run Control View58
Loading, editing, saving and printing methods60
More on editing methods61
Running samples64
Running a single sample65
Running a sequence66
Flow injection analysis69
Monitoring analyses73
Online signal plots73
Quick method overview74
Logbooks74
Instrument shutdown76
4Data Analysis77
The Data Analysis View78
Loading and manipulating chromatograms80
Loading signals81
Removing signals from the chromatogram display85
Changing how chromatograms are displayed85
Working with spectra87
Displaying spectra88
6Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Peak purity89
Performing quantification90
Integrating peaks90
Calibration92
Data review and sequence reprocessing94
The Navigation Table94
Batch review94
5Reports97
Using predefined reports98
Generating reports98
Report styles99
Defining custom reports101
Summary of process101
Example report templates101
The Report Layout View102
Contents
6Verification of Performance105
The Verification (OQ/PV) View106
Instrument verification107
Setting up and running instrument verification108
Available OQ/PV tests110
Verification logbook111
System verification112
Overview112
Setting up and running system verification113
7Maintenance and Troubleshooting115
The Diagnosis View116
Overview116
Instrument panel117
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide7
Contents
Logbooks119
Maintenance120
Early maintenance feedback120
Maintenance logbook121
Maintenance procedures122
Venting and pumping down the MS122
Diagnosing and fixing problems124
Symptoms and causes124
Diagnostic tests for the MS125
Fixing problems126
8Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Agilent 6100 Series Quadrupole LC/MS Systems
Concepts Guide
1
Overview of Hardware and Software
How the Agilent quadrupole LC/MS systems work 10
Overview 10
Details 11
Types of data you can acquire 15
Scan versus selected ion monitoring (SIM) 15
Generation of fragment ions: low versus high fragmentor 16
Positive versus negative ions 19
Multiple signal acquisition 19
This chapter provides an overview of the hardware and
software that comprises the Agilent 6100 Series Quadrupole
LC/MS systems. The family consists of three models: 6120B,
6130B, and 6150B.
Agilent Technologies
9
1Overview of Hardware and Software
capillary
nebulizer
HPLC inlet
ion
source
rough
pump
split-flow
turbo pump
detector
quadrupole
mass filter
ion optics
How the Agilent quadrupole LC/MS systems work
How the Agilent quadrupole LC/MS systems work
Overview
Mass spectrometry (MS) is based on the analysis of ions
moving through a vacuum. The result is mass spectra, which
provide valuable information about the molecular weight,
structure, identity, quantity, and purity of a sample. MS adds
specificity to both qualitative and quantitative analyses.
A quadrupole mass analyzer is
sometimes called a quadrupole
mass filter or a quadrupole.
API – atmospheric pressure
ionization
Figure 1 shows a diagram of the Agilent 6100 Series Quadrupole
LC/MS systems. The ionization of a sample occurs at
atmospheric pressure in the ion source that is shown on the left.
The Agilent 6100 Series Quadrupole LC/MS systems are
compatible with a number of Agilent atmospheric pressure
ionization (API) sources.
Figure 1Block diagram for an Agilent quadrupole LC/MS system
A common atmospheric sampling interface introduces ions
from these ionization sources into the vacuum system of the
mass spectrometer. Various ion-optic elements focus and
guide the ions through a series of vacuum stages until they
reach the quadrupole mass analyzer, which separates the
ions. The ions then travel to the detector, where they are
recorded as signals.
10Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Details
4321Vacuum stage:
Ion sourceIon transport and focusing region
quadrupole
capillary
nebulizer
HPLC inlet
fragmentation
zone (CID)
detector
skimmer
octopole
lenses
3 torr
5X10
-6
torr
Overview of Hardware and Software1
Details
Figure 2 and Figure 3 show more detailed schematics of the ion
paths in the Agilent 6100 Series Quadrupole LC/MS systems.
After the API source forms ions, the ion-optic elements in the
ion transport and focusing region of the system direct the ions
toward the quadrupole and the detector. During transit, the
ions move from atmospheric pressure (760 torr) at the source to
a vacuum in the 10
-6
torr range at the quadrupole and detector.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide11
Figure 2Ion path for Agilent 6130 and 6150 Quadrupole LC/MS sys-
tems
1Overview of Hardware and Software
4321
Vacuum stage:
Ion sourceIon transport and focusing region
quadrupole
capillary
nebulizer
HPLC inlet
fragmentation
zone (CID)
detector
skimmers
octopole
lenses
2 torr
6X10
-6
torr
Details
Figure 3Ion path for Agilent 6120 Quadrupole LC/MS system
The ion transport and focusing region of the Agilent 6100 Series
Quadrupole LC/MS systems is enclosed in a vacuum manifold.
The function of the vacuum system is to evacuate regions of ion
focusing and transport and keep the quadrupole at low
pressure.
Because the nebulizer is at a right angle to the inlet capillary,
most of the solvent is vented from the spray chamber and never
reaches the capillary. Only ions, drying gas, and a small amount
By autotuning the instrument, you
automatically set most of the
voltages for the elements in the ion
path. See “Preparation of the MS –
tuning” on page 42.
of solvent are transmitted through the capillary.
The following discussion of the ion optics is organized
according to the stages of the ion path and the vacuum
stages of the mass spectrometer.
Ion transport and fragmentation (first vacuum stage)
Ions produced in the API source are electrostatically drawn
through a drying gas and then through a heated sampling
capillary into the first stage of the vacuum system. Near the exit
of the capillary is a metal skimmer with a small hole. Heavier
12Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
CID – collision-induced
dissociation
Overview of Hardware and Software1
Details
ions with greater momentum pass through the skimmer
aperture. Most of the lighter drying gas (nitrogen) molecules are
deflected by the skimmer and pumped away by a rough pump.
The ions that pass through the skimmer move into the second
stage of the vacuum system.
The atmospheric pressure ionization techniques are all
relatively “soft” techniques. They generate primarily:
+
• Molecular ions M
• Protonated molecules [M + H]
• Simple adduct ions [M + Na]
or M
-
+
+
• Ions representing simple losses, such as the loss of a
water molecule [M + H - H
+
O]
2
These types of ions give molecular weight information, but
you often need complementary structural information. To
gain structural information, you can fragment the analyte
ions in the first vacuum stage. To do that, you give them
extra energy and collide them with neutral molecules in a
process known as collision-induced dissociation (CID). A
voltage is applied at the end of the atmospheric sampling
capillary to add energy to the collisions and create more
fragmentation. For more information, see “Generation of
fragment ions: low versus high fragmentor” on page 16.
Ion transport (second and third vacuum stages)
An octopole ion guide is a set of
small parallel metal rods with a
common open axis through which
the ions can pass.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide13
Agilent 6130 and 6150 Quadrupole LC/MS systems In the second
vacuum stage, the ions are immediately focused by an octopole
ion guide that traverses two vacuum stages. The ions pass
through the octopole ion guide because of the momentum they
received from being drawn from atmospheric pressure through
the sampling capillary. Radio-frequency voltage applied to the
octopole rods repels ions above a particular mass range to the
open center of the rod set. The ions exit this ion guide and then
pass through two focusing lenses into the fourth stage of the
vacuum system.
1Overview of Hardware and Software
From ion
source
To d e t ec t or
Details
Agilent 6120 Quadrupole LC/MS systemIn the second vacuum
stage, the ions are transported between skimmer 1 and
skimmer 2. They then enter the third vacuum stage, where
they pass through the octopole ion guide. The ions exit this
ion guide and then pass through two focusing lenses into the
fourth stage of the vacuum system.
Ion separation and detection (fourth vacuum stage)
In the fourth vacuum stage, the quadrupole mass analyzer
separates the ions by mass-to-charge ratio. An electron
multiplier then detects the ions.
m/z – mass/charge ratioThe quadrupole mass analyzer (Figure 4) consists of four
parallel rods to which specific direct-current (DC) and
radio-frequency (RF) voltages are applied. The analyte ions
are directed down the center of the rods. Voltages applied to
the rods generate electromagnetic fields. These fields
determine which mass-to-charge ratio of ions can pass
through the filter at a given time. The ions that pass through
are focused on the detector.
Figure 4Quadrupole mass analyzer
14Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Types of data you can acquire
m/z
m/z
m/z
m/z
1 scan
1 scan
SIM
scan
discrete masses
mass range
abundance
abundance
time
time
Scan versus selected ion monitoring (SIM)
Overview of Hardware and Software1
Types of data you can acquire
You set up a scan or SIM analysis
in the Method and Run Control
view, described in Chapter 3.
As shown in Figure 5, quadrupole mass analyzers can operate
in two modes. To get the most from your analysis, it is
important to pick the appropriate mode. The discussion below
will help you choose.
Figure 5A quadrupole mass analyzer can operate in either scan mode
or selected ion monitoring (SIM) mode
Scan mode
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide15
In scan mode, a range of m/z values are analyzed, for
example, m/z 200 to 1000. The quadrupole sequentially
filters one mass after another, with an entire scan typically
taking about a second. (The exact time depends on mass
range and scan speed.) The MS firmware steps the
quadrupole through increasing DC and RF voltages, which
sequentially filters the corresponding m/z values across a
mass spectrum.
1Overview of Hardware and Software
Generation of fragment ions: low versus high fragmentor
A full scan analysis is useful because it shows all of the ions in a
given mass range that are present in the ion source. Because it
provides a complete picture of all the ionized compounds that
occur above the detection limit in the chosen mass range, a full
scan analysis is often used for sample characterization,
structural elucidation, and impurity analysis. It is also the
starting point for development of methods for SIM data
acquisition (discussed next).
Selected ion monitoring (SIM) mode
To obtain the best sensitivity, the quadrupole is operated in
SIM mode. In SIM mode, the quadrupole analyzes the signals
of only a few specific m/z values. The required RF/DC
voltages are set to filter one mass at a time. Rather than
stepping through all the m/z values in a given mass range,
the quadrupole steps only among the values that the analyst
chooses. Because the quadrupole spends more time sampling
each of these chosen m/z values, the system can detect lower
levels of sample.
SIM mode is significantly more sensitive than scan mode but
provides information about fewer ions. Scan mode is
typically used for qualitative analyses or for quantitation
when analyte masses are not known in advance. SIM mode
is used for quantitation and monitoring of target compounds.
Generation of fragment ions: low versus high fragmentor
When you set up a method for data
acquisition, you can control the
amount of fragmentation with the
fragmentor setting. You set up a
method in the Method and Run
Control view, described in
Chapter 3.
16Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Fragment ions, also known as product ions, are formed by
breaking apart precursor ions. On the Agilent 6100 Series
Quadrupole LC/MS systems, the fragmentation region is
between the capillary exit and the skimmer, where the gas
pressure is about 2 to 3 torr. Depending on the voltage in this
region, precursor ions may pass through unchanged or they may
be fragmented.
Overview of Hardware and Software1
m/z
100
200
300
0
50000
100000
150000
200000
250000
300000
350000
279.1
301.0
280.0
281.0
[M + Na]
+
[M + H]
+
S
NH
H
2
N
O
O
N
N
CH
3
CH
3
Generation of fragment ions: low versus high fragmentor
When a lower voltage is applied across this region, the ions
pass through unchanged. Even if these ions collide with the
gas molecules in this region, they usually do not have
enough energy to fragment. (See Figure 6.)
Figure 6Mass spectrum of sulfamethazine – low fragmentor
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide17
m/z
100
200
300
0
20000
40000
60000
[M + H]
+
[M + Na]
+
80000
124.1
186.0
279.1
156.1
108.2
301.0
323.0
213.2
107.1
280.1
125.1
187.0
157.1
NH
H
2
N
O
O
N
N
CH
3
CH
3
S
m/z
156
m/z
186
m/z
124
m/z
213
m/z
108
1Overview of Hardware and Software
Generation of fragment ions: low versus high fragmentor
Figure 7Mass spectrum of sulfamethazine – high fragmentor
If the voltage is increased, the ions have more translational
energy. Then, if the ions collide with gas molecules, the
collisions convert the translational energy into molecular
vibrations that can cause the ions to fragment. This is called
collision-induced dissociation (CID). Figure 7 shows an
example. Even though this fragmentation does not occur where
the ions are formed at atmospheric pressure, it is a tradition to
call this type of fragmentation “in-source CID.” The ions from
molecular fragments are used for structural determination or
confirmation of the presence of a particular chemical species.
FIA – flow injection analysisThe ideal fragmentation voltage depends on the structure of
It is possible to produce both molecular ions and fragment
ions within the same spectrum by using an intermediate
fragmentation voltage.
the compound and the needs of the analysis. For target
compound analysis, it is good practice to determine in
advance the compound’s response to fragmentor setting. The
fastest way to accomplish this is with a flow injection analysis
18Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
(FIA) series. An FIA series allows you to inject the compound
multiple times within the same run, and to vary the fragmentor
setting in different time windows. From the resulting data, you
can judge the best fragmentor setting. For more information on
FIA, see “Flow injection analysis” on page 69.
Positive versus negative ions
Overview of Hardware and Software1
Positive versus negative ions
You set the ion polarity when you
set up a method in the Method and
Run Control view, described in
Chapter 3.
Multiple signal acquisition
You establish the conditions for
multiple signal acquisition in the
Method and Run Control view,
described in Chapter 3.
Atmospheric pressure ionization techniques can produce
both positive and negative ions. For any given analysis, the
predominant ion type depends on the chemical structure of
the analyte and (particularly for electrospray ionization) the
pH of the solution. While either or both ion types may be
present in the ion source, the polarity of the ion optics in
the ion transport and focusing region determines which ion
type is detected.
Analyses of positive and negative ions require different
settings for the ion optics. The software-controlled autotune
process optimizes the settings for both positive and negative
ions, and stores them in a single tune file. During data
acquisition, the software accesses the tune file for the
appropriate settings.
The Agilent 6120, 6130 and 6150 LC/MS models allow you to
acquire multiple types of data during a single analysis.
Within a single analytical run, you can choose alternating
positive and negative ionization; alternating high and low
fragmentor settings; and alternating scan and SIM modes.
Because optimum MS conditions vary from compound to
compound, this multisignal capability enables you to analyze
more compounds, with greater sensitivity, within a single
run.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide19
1Overview of Hardware and Software
Multiple signal acquisition
Polarity switching
The Agilent 6120, 6130 and 6150 LC/MS models allow you to
switch from scan to scan between analysis of positive ions
and analysis of negative ions. To switch polarities very quickly,
these models incorporate fast-switching power supplies for the
API source, the lens system, the quadrupole, and the detector.
The ability to switch polarities on the chromatographic time
scale is very useful for analysis of complete unknowns because
it obviates the need to run the sample twice to detect both types
of ions.
Alternating high/low fragmentor
With the Agilent 6120, 6130 and 6150 LC/MS models, you
can also alternate from scan to scan between high and low
fragmentation voltages. This capability allows you to acquire
scans at low fragmentor settings for molecular weight
information, and high fragmentor settings for structural
information.
Alternating SIM/scan
Many analyses require use of SIM mode to monitor and/or
quantitate target compounds at very low levels. Sometimes it is
also desirable to characterize the other sample components
with a scan analysis. The Agilent 6120, 6130 and 6150 LC/MS
models allow you to alternate between SIM and scan modes, so
you can accomplish both goals in a single analysis.
Putting it all together
The 6120, 6130 and 6150 LC/MS models can cycle through
four different user-selected acquisition modes on a
scan-by-scan basis within a single run. For example, you can
set up a single run to do the following:
• Positive ion scan with low fragmentor voltage
• Positive ion scan with high fragmentor voltage
• Negative ion scan with low fragmentor voltage
• Negative ion scan with high fragmentor voltage
20Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Overview of Hardware and Software1
Multiple signal acquisition
Such an analysis is ideal for a mixture of compounds where
some respond better in positive mode and some respond
better in negative mode, and where you need both molecular
ions and fragment ions.
The time required for one cycle varies depending on the
number of modes chosen, the scan range, and the interscan
delay required for the switching. For separations with narrow
chromatographic peaks, it is important to ensure that total
cycle time is short enough that the instrument makes sufficient
measurements across the peak.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide21
1Overview of Hardware and Software
NOTE
Ion sources
Ion sources
The Agilent 6100 Series Quadrupole LC/MS systems operate
with the following interchangeable atmospheric pressure
ionization (API) sources:
• ESI (electrospray ionization)
• ESI with Agilent Jet Stream technology
• APCI (atmospheric pressure chemical ionization)
• APPI (atmospheric pressure photoionization)
• MMI (multimode ionization)
The sources that are used on the 6100 Series LC/MS systems are the
B-type sources. The 6100 Series LC/MS systems are not compatible with
the A-type sources that were used on previous Agilent LC/MS models.
Electrospray ionization (ESI)
You control the spray chamber
parameters (nebulizer pressure,
drying gas flow and temperature,
and capillary voltage) when you set
up a method in the Method and
Run Control view, described in
Chapter 3.
22Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Electrospray ionization relies in part on chemistry to
generate analyte ions in solution before the analyte reaches
the mass spectrometer. As shown in Figure 8, the LC eluent is
sprayed (nebulized) into a spray chamber at atmospheric
pressure in the presence of a strong electrostatic field and
heated drying gas. The electrostatic field occurs between the
nebulizer, which is at ground in the Agilent design, and the
capillary, which is at high voltage.
The spray occurs at right angles to the capillary. This
patented Agilent design reduces background noise from
droplets, increases sensitivity, and keeps the capillary
cleaner for a longer period of time.
Overview of Hardware and Software1
heated drying gas
capillary
nebulizer
HPLC inlet
solvent
spray
Electrospray ionization (ESI)
Figure 8Electrospray ion source
Electrospray ionization (ESI) consists of four steps:
1 Formation of ions
2 Nebulization
3 Desolvation
4 Ion evaporation
Formation of ions
Ion formation in API-electrospray occurs through more than
one mechanism. If the chemistry of analyte, solvents, and
buffers is correct, ions are generated in solution before
nebulization. This results in high analyte ion concentration
and good API-electrospray sensitivity.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide23
1Overview of Hardware and Software
Electrospray ionization (ESI)
Preformed ions are not always required for ESI. Some
compounds that do not ionize in solution can still be
analyzed. The process of nebulization, desolvation, and ion
evaporation creates a strong electrical charge on the surface
of the spray droplets. This can induce ionization in analyte
molecules at the surface of the droplets.
Nebulization
Nebulization (aerosol generation) takes the sample solution
through these steps:
a Sample solution enters the spray chamber through a
grounded needle called a nebulizer.
b For high-flow electrospray, nebulizing gas enters the
spray chamber concentrically through a tube that
surrounds the needle.
c The combination of strong shear forces generated by the
nebulizing gas and the strong voltage (2–6 kV) in the spray
chamber draws out the sample solution and breaks it into
droplets.
d As the droplets disperse, ions of one polarity
preferentially migrate to the droplet surface due to
electrostatic forces.
e As a result, the sample is simultaneously charged and
dispersed into a fine spray of charged droplets, hence
the name electrospray.
Because the sample solution is not heated when the aerosol
is created, ESI does not thermally decompose most analytes.
Desolvation and ion evaporation
Before the ions can be mass analyzed, solvent must be removed
to yield a bare ion.
A counter-current of neutral, heated drying gas, typically
nitrogen, evaporates the solvent, decreasing the droplet
diameter and forcing the predominantly like surface-charges
closer together (see Figure 9).
24Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Overview of Hardware and Software1
+
+
+
+
+
+
+
+
+
+
+
-
-
-
-
-
-
+
+
+
+
+
+
+
+
+
+
+
-
-
-
-
-
-
+
+
+
++
+
+
+
+++
-
-
-
-
-
-
+
+
+
+
+
+
-
-
-
-
+
evaporationanalyte ion ejected
Electrospray ionization (ESI)
Figure 9Desorption of ions from solution
Coulomb repulsion – repulsion
between charged species of the
same sign
When the force of the Coulomb repulsion equals that of the
surface tension of the droplet, the droplet explodes,
producing smaller charged droplets that are subject to
further evaporation. This process repeats itself, and droplets
with a high density of surface-charges are formed. When
charge density reaches approximately 10
evaporation occurs (direct ejection of bare ions from the
droplet surface). These ions are attracted to and pass
through a capillary sampling orifice into the ion optics and
mass analyzer.
The importance of solution chemistry
The choice of solvents and buffers is a key to successful
ionization with electrospray. Solvents like methanol that have
lower heat capacity, surface tension, and dielectric constant,
promote nebulization and desolvation. For best results in
electrospray mode:
• Adjust solvent pH according to the polarity of ions
desired and the pH of the sample.
• To enhance ion desorption, use solvents that have low
heats of vaporization and low surface tensions.
• Select solvents that do not neutralize ions through
gas-phase reactions such as proton transfer or ion pair
reactions.
• To reduce the buildup of salts in the ion source, select
more volatile buffers.
8
V/cm3, ion
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide25
1Overview of Hardware and Software
Electrospray ionization (ESI)
Multiple charging
Electrospray is especially useful for analyzing large
biomolecules such as proteins, peptides, and
oligonucleotides, but can also analyze smaller molecules like
drugs and environmental contaminants. Large molecules
often acquire more than one charge. Because of this multiple
charging, you can use electrospray to analyze molecules as
large as 150,000 u even though the mass range (or more
accurately mass-to-charge range) for a typical quadrupole
LC/MS instrument is around 3000 m/z. For example:
100,000 u / 10 z = 1,000 m/z
The optional Agilent LC/MSD
Deconvolution & Bioanalysis
Software performs the calculations
to accomplish deconvolution.
When a large molecule acquires many charges, a
mathematical process called deconvolution is used to
determine the actual molecular weight of the analyte.
Agilent Jet Stream Technology
The Agilent Jet Stream technology is supported on compatible
Agilent 6100 Series LC/MS system.
Agilent Jet Stream Technology enhances analyte desolvation
by collimating the nebulizer spray and creating a
dramatically “brighter signal.” The addition of a collinear,
concentric, super-heated nitrogen sheath gas (Figure 10) to the
inlet assembly significantly improves ion drying from the
electrospray plume and leads to increased mass spectrometer
signal to noise allowing the triple quadrupole to surpass the
femtogram limit of detection. The Agilent Jet Stream
Technology is patent pending.
26Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Overview of Hardware and Software1
Electrospray ionization (ESI)
Figure 10Electrospray Ion Source with Agilent Jet Stream Technology
Agilent Jet Stream thermal gradient focusing consists of a
superheated nitrogen sheath gas that is introduced collinear
and concentric to the pneumatically assisted electrospray.
Thermal energy from the superheated nitrogen sheath gas is
focused to the nebulizer spray producing the most efficient
desolvation and ion generation possible. The enhanced
molecular ion desolvation results in more ions entering the
sampling capillary as shown in Figure 10 and concomitant
improved signal to noise. Parameters for the Agilent Jet Stream
Technology are the superheated nitrogen sheath gas
temperature and flow rate and the nozzle voltage.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide27
1Overview of Hardware and Software
+
+
+
+
+
+
+
++
++
++
++
++
++
++
drying gas
capillary
nebulizer (sprayer)
HPLC inlet
vaporizer
corona
(heater)
discharge
needle
Atmospheric pressure chemical ionization (APCI)
Atmospheric pressure chemical ionization (APCI)
APCI is a gas-phase chemical ionization process. The APCI
technique passes LC eluent through a nebulizing needle,
which creates a fine spray. The spray is passed through a heated
ceramic tube, where the droplets are fully vaporized
(Figure 11).
The resulting gas/vapor mixture is then passed over a
corona discharge needle, where the solvent vapor is ionized
to create reagent gas ions. These ions in turn ionize the
sample molecules via a chemical ionization process. The
sample ions are then introduced into the capillary.
Figure 11Atmospheric pressure chemical ionization (APCI) source
APCI requires that the analyte be in the gas phase for
ionization to occur. To vaporize the solvent and analyte, the
APCI source is typically operated at vaporizer temperatures
of 400 to 500 °C.
28Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Overview of Hardware and Software1
Atmospheric pressure chemical ionization (APCI)
APCI is applicable across a wide range of molecular
polarities. It rarely results in multiple charging, so it is
typically used for molecules less than 1,500 u. Because of this
molecular weight limitation and use of high-temperature
vaporization, APCI is less well-suited than electrospray for
analysis of large biomolecules that may be thermally unstable.
APCI is well suited for ionization of the less polar compounds
that are typically analyzed by normal-phase chromatography.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide29
1Overview of Hardware and Software
+
+
+
+
+
+
+
++
++
++
++
++
++
++
drying gas
capillary
nebulizer (sprayer)
HPLC inlet
vaporizer
UV lamp
(heater)
hν
Atmospheric pressure photoionization (APPI)
Atmospheric pressure photoionization (APPI)
With the APPI technique, LC eluent passes through a
nebulizing needle to create a fine spray. This spray is passed
through a heated ceramic tube, where the droplets are fully
vaporized. The resulting gas/vapor mixture passes through
the photon beam of a krypton lamp to ionize the sample
molecules (Figure 12). The sample ions are then introduced into
the capillary.
APPI and APCI are similar, with APPI substituting a lamp
for the corona needle for ionization. APPI often also uses an
additional solvent or mobile phase modifier, called a
“dopant”, to assist with the photoionization process.
APPI is applicable to many of the same compounds that are
typically analyzed by APCI. APPI has proven particularly
valuable for analysis of nonpolar compounds.