No p art o f this manu al may be re produce d in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G1960-90079
Edition
Revision A, September 2011
Printed in USA
Agilent Technologies, Inc.
5301 Stevens Creek Blvd.
Santa Clara, CA 95051
Microsoft® is a U.S. registered trademark
of Microsoft Corporation.
Software Revision
This guide is valid for the B.04.02 SPI1 or
later revision of the Agilent
ChemStation software for the Agilent 6100
Series Quadrupole LC/MS systems, until
superseded.
If you have any comments about this guide,
please send an e-mail to
feedback_lcms@agilent.com.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user customers. Agilent provides this customary
commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 3
In This Guide...
The Concepts Guide presents an overview of the Agilent
6100 Series Quadrupole LC/MS systems, to help you
understand how the hardware and software work.
If you have any comments about this guide, please send an
e-mail to feedback_lcms@agilent.com.
1Overview
Learn how the hardware works in the Agilent 6100 Series
Quadrupole LC/MS systems, and get a brief introduction to
ChemStation software.
2Instrument Preparation
Learn the concepts you need to prepare the LC and column
for an analysis, and to tune the MS.
3Data Acquisition
Learn about setting up methods and running samples.
4Data Analysis
Learn the concepts you need for qualitative and quantitative
data analysis with ChemStation software.
5Reports
Learn about predefined results reports and about setting up
custom reports.
6Verification of Performance
Learn the concepts for Operational Qualification/
Performance Verification (OQ/PV) and system verification
with ChemStation software.
7Maintenance and Troubleshooting
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide3
Page 4
Learn about tools that are proved in ChemStation software
to help you maintain your system and diagnose and fix
problems.
4Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 5
Contents
1Overview of Hardware and Software9
How the Agilent quadrupole LC/MS systems work10
Overview10
Details11
Types of data you can acquire15
Scan versus selected ion monitoring (SIM)15
Generation of fragment ions: low versus high fragmentor16
Positive versus negative ions19
Multiple signal acquisition19
Purpose36
Summary of procedures36
Setting parameters for LC modules38
Column conditioning and equilibration39
Monitoring the stability of flow and pressure41
Preparation of the MS – tuning42
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide5
Page 6
Contents
Overview42
Ways to tune44
When to tune – Check Tune45
Autotune47
Manual tuning49
Tune reports51
Gain calibration53
3Data Acquisition57
Working with methods58
Method and Run Control View58
Loading, editing, saving and printing methods60
More on editing methods61
Running samples64
Running a single sample65
Running a sequence66
Flow injection analysis69
Monitoring analyses73
Online signal plots73
Quick method overview74
Logbooks74
Instrument shutdown76
4Data Analysis77
The Data Analysis View78
Loading and manipulating chromatograms80
Loading signals81
Removing signals from the chromatogram display85
Changing how chromatograms are displayed85
Working with spectra87
Displaying spectra88
6Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 7
Peak purity89
Performing quantification90
Integrating peaks90
Calibration92
Data review and sequence reprocessing94
The Navigation Table94
Batch review94
5Reports97
Using predefined reports98
Generating reports98
Report styles99
Defining custom reports101
Summary of process101
Example report templates101
The Report Layout View102
Contents
6Verification of Performance105
The Verification (OQ/PV) View106
Instrument verification107
Setting up and running instrument verification108
Available OQ/PV tests110
Verification logbook111
System verification112
Overview112
Setting up and running system verification113
7Maintenance and Troubleshooting115
The Diagnosis View116
Overview116
Instrument panel117
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide7
Page 8
Contents
Logbooks119
Maintenance120
Early maintenance feedback120
Maintenance logbook121
Maintenance procedures122
Venting and pumping down the MS122
Diagnosing and fixing problems124
Symptoms and causes124
Diagnostic tests for the MS125
Fixing problems126
8Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 9
Agilent 6100 Series Quadrupole LC/MS Systems
Concepts Guide
1
Overview of Hardware and Software
How the Agilent quadrupole LC/MS systems work 10
Overview 10
Details 11
Types of data you can acquire 15
Scan versus selected ion monitoring (SIM) 15
Generation of fragment ions: low versus high fragmentor 16
Positive versus negative ions 19
Multiple signal acquisition 19
This chapter provides an overview of the hardware and
software that comprises the Agilent 6100 Series Quadrupole
LC/MS systems. The family consists of three models: 6120B,
6130B, and 6150B.
Agilent Technologies
9
Page 10
1Overview of Hardware and Software
capillary
nebulizer
HPLC inlet
ion
source
rough
pump
split-flow
turbo pump
detector
quadrupole
mass filter
ion optics
How the Agilent quadrupole LC/MS systems work
How the Agilent quadrupole LC/MS systems work
Overview
Mass spectrometry (MS) is based on the analysis of ions
moving through a vacuum. The result is mass spectra, which
provide valuable information about the molecular weight,
structure, identity, quantity, and purity of a sample. MS adds
specificity to both qualitative and quantitative analyses.
A quadrupole mass analyzer is
sometimes called a quadrupole
mass filter or a quadrupole.
API – atmospheric pressure
ionization
Figure 1 shows a diagram of the Agilent 6100 Series Quadrupole
LC/MS systems. The ionization of a sample occurs at
atmospheric pressure in the ion source that is shown on the left.
The Agilent 6100 Series Quadrupole LC/MS systems are
compatible with a number of Agilent atmospheric pressure
ionization (API) sources.
Figure 1Block diagram for an Agilent quadrupole LC/MS system
A common atmospheric sampling interface introduces ions
from these ionization sources into the vacuum system of the
mass spectrometer. Various ion-optic elements focus and
guide the ions through a series of vacuum stages until they
reach the quadrupole mass analyzer, which separates the
ions. The ions then travel to the detector, where they are
recorded as signals.
10Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 11
Details
4321Vacuum stage:
Ion sourceIon transport and focusing region
quadrupole
capillary
nebulizer
HPLC inlet
fragmentation
zone (CID)
detector
skimmer
octopole
lenses
3 torr
5X10
-6
torr
Overview of Hardware and Software1
Details
Figure 2 and Figure 3 show more detailed schematics of the ion
paths in the Agilent 6100 Series Quadrupole LC/MS systems.
After the API source forms ions, the ion-optic elements in the
ion transport and focusing region of the system direct the ions
toward the quadrupole and the detector. During transit, the
ions move from atmospheric pressure (760 torr) at the source to
a vacuum in the 10
-6
torr range at the quadrupole and detector.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide11
Figure 2Ion path for Agilent 6130 and 6150 Quadrupole LC/MS sys-
tems
Page 12
1Overview of Hardware and Software
4321
Vacuum stage:
Ion sourceIon transport and focusing region
quadrupole
capillary
nebulizer
HPLC inlet
fragmentation
zone (CID)
detector
skimmers
octopole
lenses
2 torr
6X10
-6
torr
Details
Figure 3Ion path for Agilent 6120 Quadrupole LC/MS system
The ion transport and focusing region of the Agilent 6100 Series
Quadrupole LC/MS systems is enclosed in a vacuum manifold.
The function of the vacuum system is to evacuate regions of ion
focusing and transport and keep the quadrupole at low
pressure.
Because the nebulizer is at a right angle to the inlet capillary,
most of the solvent is vented from the spray chamber and never
reaches the capillary. Only ions, drying gas, and a small amount
By autotuning the instrument, you
automatically set most of the
voltages for the elements in the ion
path. See “Preparation of the MS –
tuning” on page 42.
of solvent are transmitted through the capillary.
The following discussion of the ion optics is organized
according to the stages of the ion path and the vacuum
stages of the mass spectrometer.
Ion transport and fragmentation (first vacuum stage)
Ions produced in the API source are electrostatically drawn
through a drying gas and then through a heated sampling
capillary into the first stage of the vacuum system. Near the exit
of the capillary is a metal skimmer with a small hole. Heavier
12Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 13
CID – collision-induced
dissociation
Overview of Hardware and Software1
Details
ions with greater momentum pass through the skimmer
aperture. Most of the lighter drying gas (nitrogen) molecules are
deflected by the skimmer and pumped away by a rough pump.
The ions that pass through the skimmer move into the second
stage of the vacuum system.
The atmospheric pressure ionization techniques are all
relatively “soft” techniques. They generate primarily:
+
• Molecular ions M
• Protonated molecules [M + H]
• Simple adduct ions [M + Na]
or M
-
+
+
• Ions representing simple losses, such as the loss of a
water molecule [M + H - H
+
O]
2
These types of ions give molecular weight information, but
you often need complementary structural information. To
gain structural information, you can fragment the analyte
ions in the first vacuum stage. To do that, you give them
extra energy and collide them with neutral molecules in a
process known as collision-induced dissociation (CID). A
voltage is applied at the end of the atmospheric sampling
capillary to add energy to the collisions and create more
fragmentation. For more information, see “Generation of
fragment ions: low versus high fragmentor” on page 16.
Ion transport (second and third vacuum stages)
An octopole ion guide is a set of
small parallel metal rods with a
common open axis through which
the ions can pass.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide13
Agilent 6130 and 6150 Quadrupole LC/MS systems In the second
vacuum stage, the ions are immediately focused by an octopole
ion guide that traverses two vacuum stages. The ions pass
through the octopole ion guide because of the momentum they
received from being drawn from atmospheric pressure through
the sampling capillary. Radio-frequency voltage applied to the
octopole rods repels ions above a particular mass range to the
open center of the rod set. The ions exit this ion guide and then
pass through two focusing lenses into the fourth stage of the
vacuum system.
Page 14
1Overview of Hardware and Software
From ion
source
To d e t ec t or
Details
Agilent 6120 Quadrupole LC/MS systemIn the second vacuum
stage, the ions are transported between skimmer 1 and
skimmer 2. They then enter the third vacuum stage, where
they pass through the octopole ion guide. The ions exit this
ion guide and then pass through two focusing lenses into the
fourth stage of the vacuum system.
Ion separation and detection (fourth vacuum stage)
In the fourth vacuum stage, the quadrupole mass analyzer
separates the ions by mass-to-charge ratio. An electron
multiplier then detects the ions.
m/z – mass/charge ratioThe quadrupole mass analyzer (Figure 4) consists of four
parallel rods to which specific direct-current (DC) and
radio-frequency (RF) voltages are applied. The analyte ions
are directed down the center of the rods. Voltages applied to
the rods generate electromagnetic fields. These fields
determine which mass-to-charge ratio of ions can pass
through the filter at a given time. The ions that pass through
are focused on the detector.
Figure 4Quadrupole mass analyzer
14Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 15
Types of data you can acquire
m/z
m/z
m/z
m/z
1 scan
1 scan
SIM
scan
discrete masses
mass range
abundance
abundance
time
time
Scan versus selected ion monitoring (SIM)
Overview of Hardware and Software1
Types of data you can acquire
You set up a scan or SIM analysis
in the Method and Run Control
view, described in Chapter 3.
As shown in Figure 5, quadrupole mass analyzers can operate
in two modes. To get the most from your analysis, it is
important to pick the appropriate mode. The discussion below
will help you choose.
Figure 5A quadrupole mass analyzer can operate in either scan mode
or selected ion monitoring (SIM) mode
Scan mode
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide15
In scan mode, a range of m/z values are analyzed, for
example, m/z 200 to 1000. The quadrupole sequentially
filters one mass after another, with an entire scan typically
taking about a second. (The exact time depends on mass
range and scan speed.) The MS firmware steps the
quadrupole through increasing DC and RF voltages, which
sequentially filters the corresponding m/z values across a
mass spectrum.
Page 16
1Overview of Hardware and Software
Generation of fragment ions: low versus high fragmentor
A full scan analysis is useful because it shows all of the ions in a
given mass range that are present in the ion source. Because it
provides a complete picture of all the ionized compounds that
occur above the detection limit in the chosen mass range, a full
scan analysis is often used for sample characterization,
structural elucidation, and impurity analysis. It is also the
starting point for development of methods for SIM data
acquisition (discussed next).
Selected ion monitoring (SIM) mode
To obtain the best sensitivity, the quadrupole is operated in
SIM mode. In SIM mode, the quadrupole analyzes the signals
of only a few specific m/z values. The required RF/DC
voltages are set to filter one mass at a time. Rather than
stepping through all the m/z values in a given mass range,
the quadrupole steps only among the values that the analyst
chooses. Because the quadrupole spends more time sampling
each of these chosen m/z values, the system can detect lower
levels of sample.
SIM mode is significantly more sensitive than scan mode but
provides information about fewer ions. Scan mode is
typically used for qualitative analyses or for quantitation
when analyte masses are not known in advance. SIM mode
is used for quantitation and monitoring of target compounds.
Generation of fragment ions: low versus high fragmentor
When you set up a method for data
acquisition, you can control the
amount of fragmentation with the
fragmentor setting. You set up a
method in the Method and Run
Control view, described in
Chapter 3.
16Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Fragment ions, also known as product ions, are formed by
breaking apart precursor ions. On the Agilent 6100 Series
Quadrupole LC/MS systems, the fragmentation region is
between the capillary exit and the skimmer, where the gas
pressure is about 2 to 3 torr. Depending on the voltage in this
region, precursor ions may pass through unchanged or they may
be fragmented.
Page 17
Overview of Hardware and Software1
m/z
100
200
300
0
50000
100000
150000
200000
250000
300000
350000
279.1
301.0
280.0
281.0
[M + Na]
+
[M + H]
+
S
NH
H
2
N
O
O
N
N
CH
3
CH
3
Generation of fragment ions: low versus high fragmentor
When a lower voltage is applied across this region, the ions
pass through unchanged. Even if these ions collide with the
gas molecules in this region, they usually do not have
enough energy to fragment. (See Figure 6.)
Figure 6Mass spectrum of sulfamethazine – low fragmentor
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide17
Page 18
m/z
100
200
300
0
20000
40000
60000
[M + H]
+
[M + Na]
+
80000
124.1
186.0
279.1
156.1
108.2
301.0
323.0
213.2
107.1
280.1
125.1
187.0
157.1
NH
H
2
N
O
O
N
N
CH
3
CH
3
S
m/z
156
m/z
186
m/z
124
m/z
213
m/z
108
1Overview of Hardware and Software
Generation of fragment ions: low versus high fragmentor
Figure 7Mass spectrum of sulfamethazine – high fragmentor
If the voltage is increased, the ions have more translational
energy. Then, if the ions collide with gas molecules, the
collisions convert the translational energy into molecular
vibrations that can cause the ions to fragment. This is called
collision-induced dissociation (CID). Figure 7 shows an
example. Even though this fragmentation does not occur where
the ions are formed at atmospheric pressure, it is a tradition to
call this type of fragmentation “in-source CID.” The ions from
molecular fragments are used for structural determination or
confirmation of the presence of a particular chemical species.
FIA – flow injection analysisThe ideal fragmentation voltage depends on the structure of
It is possible to produce both molecular ions and fragment
ions within the same spectrum by using an intermediate
fragmentation voltage.
the compound and the needs of the analysis. For target
compound analysis, it is good practice to determine in
advance the compound’s response to fragmentor setting. The
fastest way to accomplish this is with a flow injection analysis
18Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
(FIA) series. An FIA series allows you to inject the compound
multiple times within the same run, and to vary the fragmentor
Page 19
setting in different time windows. From the resulting data, you
can judge the best fragmentor setting. For more information on
FIA, see “Flow injection analysis” on page 69.
Positive versus negative ions
Overview of Hardware and Software1
Positive versus negative ions
You set the ion polarity when you
set up a method in the Method and
Run Control view, described in
Chapter 3.
Multiple signal acquisition
You establish the conditions for
multiple signal acquisition in the
Method and Run Control view,
described in Chapter 3.
Atmospheric pressure ionization techniques can produce
both positive and negative ions. For any given analysis, the
predominant ion type depends on the chemical structure of
the analyte and (particularly for electrospray ionization) the
pH of the solution. While either or both ion types may be
present in the ion source, the polarity of the ion optics in
the ion transport and focusing region determines which ion
type is detected.
Analyses of positive and negative ions require different
settings for the ion optics. The software-controlled autotune
process optimizes the settings for both positive and negative
ions, and stores them in a single tune file. During data
acquisition, the software accesses the tune file for the
appropriate settings.
The Agilent 6120, 6130 and 6150 LC/MS models allow you to
acquire multiple types of data during a single analysis.
Within a single analytical run, you can choose alternating
positive and negative ionization; alternating high and low
fragmentor settings; and alternating scan and SIM modes.
Because optimum MS conditions vary from compound to
compound, this multisignal capability enables you to analyze
more compounds, with greater sensitivity, within a single
run.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide19
Page 20
1Overview of Hardware and Software
Multiple signal acquisition
Polarity switching
The Agilent 6120, 6130 and 6150 LC/MS models allow you to
switch from scan to scan between analysis of positive ions
and analysis of negative ions. To switch polarities very quickly,
these models incorporate fast-switching power supplies for the
API source, the lens system, the quadrupole, and the detector.
The ability to switch polarities on the chromatographic time
scale is very useful for analysis of complete unknowns because
it obviates the need to run the sample twice to detect both types
of ions.
Alternating high/low fragmentor
With the Agilent 6120, 6130 and 6150 LC/MS models, you
can also alternate from scan to scan between high and low
fragmentation voltages. This capability allows you to acquire
scans at low fragmentor settings for molecular weight
information, and high fragmentor settings for structural
information.
Alternating SIM/scan
Many analyses require use of SIM mode to monitor and/or
quantitate target compounds at very low levels. Sometimes it is
also desirable to characterize the other sample components
with a scan analysis. The Agilent 6120, 6130 and 6150 LC/MS
models allow you to alternate between SIM and scan modes, so
you can accomplish both goals in a single analysis.
Putting it all together
The 6120, 6130 and 6150 LC/MS models can cycle through
four different user-selected acquisition modes on a
scan-by-scan basis within a single run. For example, you can
set up a single run to do the following:
• Positive ion scan with low fragmentor voltage
• Positive ion scan with high fragmentor voltage
• Negative ion scan with low fragmentor voltage
• Negative ion scan with high fragmentor voltage
20Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 21
Overview of Hardware and Software1
Multiple signal acquisition
Such an analysis is ideal for a mixture of compounds where
some respond better in positive mode and some respond
better in negative mode, and where you need both molecular
ions and fragment ions.
The time required for one cycle varies depending on the
number of modes chosen, the scan range, and the interscan
delay required for the switching. For separations with narrow
chromatographic peaks, it is important to ensure that total
cycle time is short enough that the instrument makes sufficient
measurements across the peak.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide21
Page 22
1Overview of Hardware and Software
NOTE
Ion sources
Ion sources
The Agilent 6100 Series Quadrupole LC/MS systems operate
with the following interchangeable atmospheric pressure
ionization (API) sources:
• ESI (electrospray ionization)
• ESI with Agilent Jet Stream technology
• APCI (atmospheric pressure chemical ionization)
• APPI (atmospheric pressure photoionization)
• MMI (multimode ionization)
The sources that are used on the 6100 Series LC/MS systems are the
B-type sources. The 6100 Series LC/MS systems are not compatible with
the A-type sources that were used on previous Agilent LC/MS models.
Electrospray ionization (ESI)
You control the spray chamber
parameters (nebulizer pressure,
drying gas flow and temperature,
and capillary voltage) when you set
up a method in the Method and
Run Control view, described in
Chapter 3.
22Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Electrospray ionization relies in part on chemistry to
generate analyte ions in solution before the analyte reaches
the mass spectrometer. As shown in Figure 8, the LC eluent is
sprayed (nebulized) into a spray chamber at atmospheric
pressure in the presence of a strong electrostatic field and
heated drying gas. The electrostatic field occurs between the
nebulizer, which is at ground in the Agilent design, and the
capillary, which is at high voltage.
The spray occurs at right angles to the capillary. This
patented Agilent design reduces background noise from
droplets, increases sensitivity, and keeps the capillary
cleaner for a longer period of time.
Page 23
Overview of Hardware and Software1
heated drying gas
capillary
nebulizer
HPLC inlet
solvent
spray
Electrospray ionization (ESI)
Figure 8Electrospray ion source
Electrospray ionization (ESI) consists of four steps:
1 Formation of ions
2 Nebulization
3 Desolvation
4 Ion evaporation
Formation of ions
Ion formation in API-electrospray occurs through more than
one mechanism. If the chemistry of analyte, solvents, and
buffers is correct, ions are generated in solution before
nebulization. This results in high analyte ion concentration
and good API-electrospray sensitivity.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide23
Page 24
1Overview of Hardware and Software
Electrospray ionization (ESI)
Preformed ions are not always required for ESI. Some
compounds that do not ionize in solution can still be
analyzed. The process of nebulization, desolvation, and ion
evaporation creates a strong electrical charge on the surface
of the spray droplets. This can induce ionization in analyte
molecules at the surface of the droplets.
Nebulization
Nebulization (aerosol generation) takes the sample solution
through these steps:
a Sample solution enters the spray chamber through a
grounded needle called a nebulizer.
b For high-flow electrospray, nebulizing gas enters the
spray chamber concentrically through a tube that
surrounds the needle.
c The combination of strong shear forces generated by the
nebulizing gas and the strong voltage (2–6 kV) in the spray
chamber draws out the sample solution and breaks it into
droplets.
d As the droplets disperse, ions of one polarity
preferentially migrate to the droplet surface due to
electrostatic forces.
e As a result, the sample is simultaneously charged and
dispersed into a fine spray of charged droplets, hence
the name electrospray.
Because the sample solution is not heated when the aerosol
is created, ESI does not thermally decompose most analytes.
Desolvation and ion evaporation
Before the ions can be mass analyzed, solvent must be removed
to yield a bare ion.
A counter-current of neutral, heated drying gas, typically
nitrogen, evaporates the solvent, decreasing the droplet
diameter and forcing the predominantly like surface-charges
closer together (see Figure 9).
24Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 25
Overview of Hardware and Software1
+
+
+
+
+
+
+
+
+
+
+
-
-
-
-
-
-
+
+
+
+
+
+
+
+
+
+
+
-
-
-
-
-
-
+
+
+
++
+
+
+
+++
-
-
-
-
-
-
+
+
+
+
+
+
-
-
-
-
+
evaporationanalyte ion ejected
Electrospray ionization (ESI)
Figure 9Desorption of ions from solution
Coulomb repulsion – repulsion
between charged species of the
same sign
When the force of the Coulomb repulsion equals that of the
surface tension of the droplet, the droplet explodes,
producing smaller charged droplets that are subject to
further evaporation. This process repeats itself, and droplets
with a high density of surface-charges are formed. When
charge density reaches approximately 10
evaporation occurs (direct ejection of bare ions from the
droplet surface). These ions are attracted to and pass
through a capillary sampling orifice into the ion optics and
mass analyzer.
The importance of solution chemistry
The choice of solvents and buffers is a key to successful
ionization with electrospray. Solvents like methanol that have
lower heat capacity, surface tension, and dielectric constant,
promote nebulization and desolvation. For best results in
electrospray mode:
• Adjust solvent pH according to the polarity of ions
desired and the pH of the sample.
• To enhance ion desorption, use solvents that have low
heats of vaporization and low surface tensions.
• Select solvents that do not neutralize ions through
gas-phase reactions such as proton transfer or ion pair
reactions.
• To reduce the buildup of salts in the ion source, select
more volatile buffers.
8
V/cm3, ion
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide25
Page 26
1Overview of Hardware and Software
Electrospray ionization (ESI)
Multiple charging
Electrospray is especially useful for analyzing large
biomolecules such as proteins, peptides, and
oligonucleotides, but can also analyze smaller molecules like
drugs and environmental contaminants. Large molecules
often acquire more than one charge. Because of this multiple
charging, you can use electrospray to analyze molecules as
large as 150,000 u even though the mass range (or more
accurately mass-to-charge range) for a typical quadrupole
LC/MS instrument is around 3000 m/z. For example:
100,000 u / 10 z = 1,000 m/z
The optional Agilent LC/MSD
Deconvolution & Bioanalysis
Software performs the calculations
to accomplish deconvolution.
When a large molecule acquires many charges, a
mathematical process called deconvolution is used to
determine the actual molecular weight of the analyte.
Agilent Jet Stream Technology
The Agilent Jet Stream technology is supported on compatible
Agilent 6100 Series LC/MS system.
Agilent Jet Stream Technology enhances analyte desolvation
by collimating the nebulizer spray and creating a
dramatically “brighter signal.” The addition of a collinear,
concentric, super-heated nitrogen sheath gas (Figure 10) to the
inlet assembly significantly improves ion drying from the
electrospray plume and leads to increased mass spectrometer
signal to noise allowing the triple quadrupole to surpass the
femtogram limit of detection. The Agilent Jet Stream
Technology is patent pending.
26Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 27
Overview of Hardware and Software1
Electrospray ionization (ESI)
Figure 10Electrospray Ion Source with Agilent Jet Stream Technology
Agilent Jet Stream thermal gradient focusing consists of a
superheated nitrogen sheath gas that is introduced collinear
and concentric to the pneumatically assisted electrospray.
Thermal energy from the superheated nitrogen sheath gas is
focused to the nebulizer spray producing the most efficient
desolvation and ion generation possible. The enhanced
molecular ion desolvation results in more ions entering the
sampling capillary as shown in Figure 10 and concomitant
improved signal to noise. Parameters for the Agilent Jet Stream
Technology are the superheated nitrogen sheath gas
temperature and flow rate and the nozzle voltage.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide27
Page 28
1Overview of Hardware and Software
+
+
+
+
+
+
+
++
++
++
++
++
++
++
drying gas
capillary
nebulizer (sprayer)
HPLC inlet
vaporizer
corona
(heater)
discharge
needle
Atmospheric pressure chemical ionization (APCI)
Atmospheric pressure chemical ionization (APCI)
APCI is a gas-phase chemical ionization process. The APCI
technique passes LC eluent through a nebulizing needle,
which creates a fine spray. The spray is passed through a heated
ceramic tube, where the droplets are fully vaporized
(Figure 11).
The resulting gas/vapor mixture is then passed over a
corona discharge needle, where the solvent vapor is ionized
to create reagent gas ions. These ions in turn ionize the
sample molecules via a chemical ionization process. The
sample ions are then introduced into the capillary.
Figure 11Atmospheric pressure chemical ionization (APCI) source
APCI requires that the analyte be in the gas phase for
ionization to occur. To vaporize the solvent and analyte, the
APCI source is typically operated at vaporizer temperatures
of 400 to 500 °C.
28Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 29
Overview of Hardware and Software1
Atmospheric pressure chemical ionization (APCI)
APCI is applicable across a wide range of molecular
polarities. It rarely results in multiple charging, so it is
typically used for molecules less than 1,500 u. Because of this
molecular weight limitation and use of high-temperature
vaporization, APCI is less well-suited than electrospray for
analysis of large biomolecules that may be thermally unstable.
APCI is well suited for ionization of the less polar compounds
that are typically analyzed by normal-phase chromatography.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide29
Page 30
1Overview of Hardware and Software
+
+
+
+
+
+
+
++
++
++
++
++
++
++
drying gas
capillary
nebulizer (sprayer)
HPLC inlet
vaporizer
UV lamp
(heater)
hν
Atmospheric pressure photoionization (APPI)
Atmospheric pressure photoionization (APPI)
With the APPI technique, LC eluent passes through a
nebulizing needle to create a fine spray. This spray is passed
through a heated ceramic tube, where the droplets are fully
vaporized. The resulting gas/vapor mixture passes through
the photon beam of a krypton lamp to ionize the sample
molecules (Figure 12). The sample ions are then introduced into
the capillary.
APPI and APCI are similar, with APPI substituting a lamp
for the corona needle for ionization. APPI often also uses an
additional solvent or mobile phase modifier, called a
“dopant”, to assist with the photoionization process.
APPI is applicable to many of the same compounds that are
typically analyzed by APCI. APPI has proven particularly
valuable for analysis of nonpolar compounds.
30Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 31
Multimode ionization (MMI)
drying gas
capillary
nebulizer
HPLC inlet
APCI
thermal container
ESI zone
zone
corona
discharge
needle
The multimode source is an ion source that can operate in
three different modes—APCI, ESI or simultaneous APCI/ESI.
The multimode source incorporates two electrically
separated, optimized zones—one for ESI and one for APCI.
During simultaneous APCI/ESI, ions from both ionization
modes enter the capillary and are analyzed simultaneously by
the mass spectrometer.
Overview of Hardware and Software1
Multimode ionization (MMI)
Figure 13Multimode source
Multimode ionization (MMI) is useful for screening of
unknowns, or whenever samples contain a mixture of
compounds where some respond by ESI and some respond
by APCI. In these cases, the multimode source obviates the need
to run the samples twice to accomplish a complete analysis.
Unlike the APCI and APPI sources where the temperature of
the vaporizer is monitored, in the multimode source the
actual vapor temperature is monitored. As a result, the
vaporizer is typically set to between 200 and 250 °C.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide31
Page 32
1Overview of Hardware and Software
Introduction to ChemStation software
Introduction to ChemStation software
Overview
ChemStation software for the Agilent 6100 Series Quadrupole
LC/MS systems is organized into views. Each view allows you to
do a specific set of tasks. The menus and toolbars change with
each view.
Figure 14These buttons allow you to switch among the six ChemSta-
tion views
The following summarizes the ChemStation views and their
functionality:
For more information about the
Method and Run Control view, see
Chapter 3.
32Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Method and Run Control
• Set up methods
• Change setpoints for the Agilent 1100/1200/1260/1290
Series LC Series LC modules, including the Chip Cube
• Change setpoints for the Agilent 6100 Series Quadrupole
LC/MS systems
• Change setpoints for the Agilent API sources
• Run single samples
• Run automated sequences
• Run an FIA series
• View data in real time, as it is acquired
Page 33
Overview of Hardware and Software1
Overview
For more information about the
Data Analysis view, see Chapter 4.
For more information about the
Report Layout view, see Chapter 5.
For more information about the
Verification view, see Chapter 6.
For more information about the
Diagnosis view, see Chapter 7.
Data Analysis
• View chromatograms and spectra from the MS and UV
detectors
• Integrate chromatographic peaks
• Perform quantitation
• Check peak purity
• Deconvolute multiply charged spectra
• Generate reports
• Reprocess data from sequences
Report Layout
• Design custom report templates
Verificat ion (OQ/PV)
• Verify system performance
Diagnosis
• Learn possible causes of instrument problems
• Run tests to diagnose instrument problems
• Receive notification when it is time to perform system
maintenance
• Pump down and vent the system
For more information about the
MSD Tune view, see Chapter 2.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide33
MSD Tune
• Optimize and calibrate the MS
Page 34
1Overview of Hardware and Software
Reviewing data remotely
Reviewing data remotely
There are two ways to set up a computer so you can review
ChemStation data remotely.
One way is to install a Data Analysis-only version of
ChemStation software on the remote computer. This
installation provides the same Data Analysis functionality
that you have on the ChemStation that controls your Agilent
6100 Series LC/MS system. It is ideal if you need full features
for in-depth data analysis.
Another way is to install the Analytical Studio Reviewer on
the remote computer. Analytical Studio Reviewer lets you
easily review ChemStation LC and LC/MS data files, but the
functionality is different than with the full ChemStation
Data Analysis. The Analytical Studio Reviewer software is
ideal for synthetic chemists and others who use the LC/MS
system for “walk-up” analysis.
34Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 35
Agilent 6100 Series Quadrupole LC/MS Systems
Concepts Guide
2
Instrument Preparation
Preparation of the LC system 36
Purpose 36
Summary of procedures 36
Setting parameters for LC modules 38
Column conditioning and equilibration 39
Monitoring the stability of flow and pressure 41
Preparation of the MS – tuning 42
Overview 42
Ways to tune 44
When to tune – Check Tune 45
Autotune 47
Manual tuning 49
Tune reports 51
Gain calibration 53
In this chapter, you learn the concepts that help you prepare
the instrument for an analysis. This chapter assumes that
the hardware and software are installed, the instrument is
configured and the performance verified. If this has not been
completed, see the Agilent 6100 Series Single Quad LC/MS System Installation Guide.
Agilent Technologies
35
Page 36
2Instrument Preparation
Preparation of the LC system
Preparation of the LC system
Purpose
To achieve good sensitivity, it is important to properly
prepare the LC and column prior to an LC/MS analysis.
For best signal-to-noise, the entire LC system must be free
of contamination from salts (such as nonvolatile buffers) and
unwanted organic compounds. Some contaminants that are
not bothersome for a UV detector can cause problems for
the MS. Contaminants may cause ion suppression and/or
high background, and these problems can seriously degrade
sensitivity.
To achieve a smooth baseline with little noise, the LC flow
must also be very stable.
Summary of procedures
The exact LC preparation steps depend on how the LC was
used previously and the type of analysis to be performed.
The following provides guidelines:
Typical preparation
Before beginning an analysis, the entire LC path should be
contaminant-free and the flow should be stable. Usually, you
can accomplish these goals by doing the following:
1 Purge the pump to remove air bubbles. Purge each
channel that you plan to use.
For instructions to purge the pump, search the online
Help for the keyword “purge” and scroll down the list of
topics until you see entries that begin with the word
“purge.”
2 Condition the column to remove impurities or residual
sample.
36Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 37
Instrument Preparation2
Summary of procedures
For more information, see “Column conditioning and
equilibration” on page 39.
3 Equilibrate the column at the initial mobile phase
composition.
For more information, see “Column conditioning and
equilibration” on page 39.
4 Ensure that the system f low and pressure are stable.
For more information, see “Monitoring the stability of flow
and pressure” on page 41.
More extensive preparation
While the four-step procedure that is outlined above works
well on a day-to-day basis, more extensive LC/column
flushing may be necessary if any of the following are true:
• You have not used this LC for MS.
• The column is new.
• You are changing to a different mobile phase composition.
• The LC was used to analyze dirty samples.
• The next analysis requires ultimate sensitivity.
A protocol for more thorough LC cleaning is given in the
Agilent 6100 Single Quad System Installation Guide. See
the section on conditioning the LC in the chapter on system
verification.
When you flush the LC, remember to flush all channels that
you plan to use. Also, flush the injector by making several
injections of the same solvent(s) that you use to flush the
system.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide37
Page 38
2Instrument Preparation
Injection
Pump
Mass spectrometer
Diode-array detector
Column thermostat,
Column switching valve
Solvent bottles
Setting parameters for LC modules
Setting parameters for LC modules
You set up the LC modules in the Method and Run Control
view. Within the system diagram, click each module to set
parameters.
Figure 15Example system diagram (yours may be different)
To access help for any system module, click Help on the
module context menu. To access help for a given dialog box,
click the Help button on the dialog box.
To set module control parameters
This procedure uses the pump module as an example.
1 Click More Pump > Control HPLC Pump on the
Instrument menu to open the Pump Control dialog box.
2 Set desired control parameters and click OK.
Alternate
method
38Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Select the desired control parameter such as Standby from
the Pump context menu.
Page 39
Column conditioning and equilibration
To set module setpoint parameters
This procedure uses the binary pump module as an example.
1 Click Set up Instrument Method on the Instrument menu
to open the Setup Method dialog box.
2 Click the BinPump tab.
3 Set desired setpoints and click OK.
To access other instrument parameters
1 Click to open the Instrument menu.
2 Click the desired command such as Select Injection
Source, Columns, or Instrument Configuration.
Column conditioning and equilibration
There are several ways to set parameters to condition and
equilibrate a column.
Instrument Preparation2
Conditioning
Column conditioning eliminates any previously separated
compounds or impurities from the column, particularly after
runs with solvent of a single composition (isocratic runs).
There are a number of ways to condition a column before a
sample run. One way is to pump the organic solvent that
you intend to use (100% solvent B) through the column for a
period of time. Another way is to run the gradient that you
intend to use, then extend the time at the final composition
until no further peaks elute.
When a column is new, “conditioning” may include injecting
a few samples or high-level standards until peak area and
retention time are stable.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide39
Page 40
2Instrument Preparation
Column conditioning and equilibration
Equilibration
Column equilibration returns column characteristics to their
initial state after a gradient run. To equilibrate a column
before a sample run, you pass the solvent of initial
composition through the column for a period of time.
Column conditioning and equilibration
You can condition and equilibrate a column in one of three
ways with ChemStation software.
• Interactively
You set the pump to the solvent composition for the end
of the run and higher-than-normal flow rates. You can
then immediately apply these setpoints to the pump. After
you pump about three column-volumes of solvent, then
set the pump to the solvent composition and flow rate for
the beginning of the run. With this procedure, you do not
store a data file.
If you use this procedure, you can tune the MS while you
condition and equilibrate the column. When you tune the
MS, the MS stream selection valve automatically diverts the
LC effluent to waste. For information on tuning, see
“Preparation of the MS – tuning” on page 42.
You set up a method or sequence in
the Method and Run Control view,
described in Chapter 3.
40Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
• With a method in an interactive run
You set up a method for your analysis and then run a
solvent blank. The run uses the method stop time. You
can also use a post-run time within the method to
equilibrate the column.
With this procedure, you store a data file.
• With a sequence
You set up a method for your analysis and then set up a
solvent blank as the first run in a sequence. The method
includes a post-run time to equilibrate the column.
With this procedure, you store a data file.
Page 41
Instrument Preparation2
Monitoring the stability of flow and pressure
Monitoring the stability of flow and pressure
The LC solvent flow and the system backpressure must be
stable to ensure a quiet baseline and best results for
API-MS. The best time to monitor the stability of flow and
pressure is after you have equilibrated the column, and
before you start the analysis.
Chapter 3 provides more
information about online signals,
which are also called online plots.
You can measure stability with ChemStation software. To do
this, you set up an isocratic method with the same solvent
composition as the initial composition you intend to use for
your analysis. During the run, you monitor the online signals
for flow and pressure.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide41
Page 42
2Instrument Preparation
Preparation of the MS – tuning
Preparation of the MS – tuning
Overview
Use the MSD Tune view for all
tasks that relate to tuning.
Tuning is the process of adjusting MS parameters to
generate high quality, accurate mass spectra. During tuning,
the MS is optimized to:
• Maximize sensitivity
• Maintain acceptable resolution
• Ensure accurate mass assignments
Parameters that are adjusted
The Agilent 6100 Series Quadrupole LC/MS systems have two
sets of parameters that can be adjusted. One set of parameters
is associated with the formation of ions. These parameters
control the spray chamber (for example, electrospray or APCI)
and fragmentor. The other set of parameters is associated with
the transmission, filtering, and detection of ions. These
parameters control the skimmer, octopole, lenses, quadrupole
mass filter, and high-energy dynode (HED) electron multiplier
(detector).
Tuning is primarily concerned with finding the correct
settings for the parameters that control the transmission,
filtering, and detection of ions. It is accomplished by
introducing a calibrant into the MS and generating ions.
Using these ions, the tune parameters are then adjusted to
achieve sensitivity, resolution, and mass assignment goals.
With a few exceptions, the parameters that control ion
formation are not adjusted. They are set to fixed values
known to be good for generating ions from the calibrant
solution.
42Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 43
Instrument Preparation2
Overview
Tune files and reports
The product of tuning is a tune file (actually a directory)
that contains parameter settings for both positive and
negative ionization. When data acquisition uses a tune file,
the settings appropriate for the ion polarity specified by the
data acquisition method are loaded automatically.
Autotune, the automated tuning program, also generates a
report. See page 51.
Use of tune files during data acquisition
During data acquisition, the parameters associated with ion
formation are controlled by the data acquisition method. The
parameters associated with ion transmission are controlled
by the tune file assigned to the data acquisition method.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide43
Page 44
2Instrument Preparation
NOTE
CAUTION
Ways to tune
Ways to tune
Access this functionality via the
Tune menu in the MSD Tune view.
ChemStation software provides the following two ways to
tune the MS:
• Autotune is an automated tuning program that tunes the
MS for good performance over the entire mass range. It
uses known compounds in a standard calibration mixture
that is introduced via the Calibrant Delivery System
(CDS). This is the tuning method that you use in most
cases.
• Manual Tune allows you to tune the MS by adjusting one
parameter at a time until you achieve the desired
performance. Manual tuning is most often used when you
need maximum sensitivity, when your analysis targets a
restricted mass range, or when you need a tuning
compound other than the standard calibrants.
In addition, a Check Tune program allows you to determine
whether you need to tune.
Check Tune, Autotune, and Manual Tune are discussed in
more detail in the next sections.
Frequent tuning is not required for normal operation. Once tuned, the
LC/MS is very stable. Tuning is generally not needed more often than
monthly, or at most weekly. If you suspect problems related to tuning, use
the Check Tune program to confirm that the MS is out of adjustment
before you retune it.
44Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Wait at least four hours after pump-down before tuning or operating
your Agilent 6100 Series LC/MS system. Ceramic components in the
quadrupole mass filter can absorb moisture from the air. Tuning or
acquiring data too soon may result in arcing. Further, it takes the
analyzer at least nine hours to reach thermal equilibrium. Tune files
created or data acquired before the MS is at thermal equilibrium may
have incorrect mass assignments and other inaccuracies.
Page 45
When to tune – Check Tune
Check Tune allows you to quickly determine whether the MS
is correctly tuned without performing a complete autotune.
It performs a single profile scan of the tune masses and
compares the peak widths and mass axes with target values.
If the values obtained by Check Tune are within acceptable
ranges, the tune report indicates that Check Tune passes.
(See Figure 16 on page 46.) If the values are outside of
acceptable ranges, Check Tune suggests that you adjust peak
widths or calibrate the mass axis.
A third parameter, gain, may be added to the parameters
checked by Check Tune. If so, Check Tune compares the
current gain value with the gain value from the most recent
autotune. For a discussion of gain, see page 53.
Instrument Preparation2
When to tune – Check Tune
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide45
Page 46
2Instrument Preparation
All masses pass the
mass axis and peak
width tests
When to tune – Check Tune
Figure 16Check Tune report
46Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 47
Autotune
Instrument Preparation2
Autotune
Autotune is a program that adjusts the MS for good
performance over the entire mass range.
Autotune files
A tune file, ATUNES.TUN is provided as a starting point for
autotune.
The results of autotune are saved to the current autotune
file. This makes autotune faster because each autotune
begins from the most recent good tune parameters. This
means, however, that previous tune values are lost. You may
want to occasionally save the autotune file to a different file
name before using autotune.
A single autotune file contains the results of both positive
ionization and negative ionization autotunes.
Autotune polarity and scan speed
You have up to six choices for running an autotune, depending
on 6100 model:
• Dual Polarity
• Positive Polarity
• Negative Polarity
• Dual Polarity Fast Scan
• Positive Polarity Fast Scan
• Negative Polarity Fast Scan
The Dual Polarity modes perform an autotune for both
positive ionization and negative ionization. Positive Polarity
and Negative Polarity modes perform autotunes only for the
specified polarity.
Gain checking in autotune
Gain checking occurs only in the positive polarity autotune
or the positive portion of a dual polarity autotune. For more
information on gain checking, see “Gain calibration” on page 53.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide47
Page 48
2Instrument Preparation
Autotune
Spray chamber parameters
Autotune adjusts the parameters that affect ion transmission.
The spray chamber parameters, which affect ion formation,
are not adjusted by autotune. They are set to default values
known to be good for generating ions of the calibrant
solution.
Because there may be slight variations from spray chamber
to spray chamber, the default parameters may not be
optimum for a particular spray chamber. Changes such as
adjustment of the nebulizer and normal aging of the LC/MS
system components can also result in a system where the
default values are not optimum. Therefore, it is possible to
manually set the spray chamber parameters. Typically, this is
necessary only when the instrument fails autotune or when
the results of autotune show a significant decrease in system
performance. Be aware that other factors can also decrease
system performance.
48Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 49
Manual tuning
CAUTION
The autotune program adjusts the MS for good performance
over the entire mass range, and this tune program is
sufficient for most applications. There are times, however,
when it is advantageous to tune the MS manually. These
include:
• When you want to achieve maximum sensitivity by
• When you want to tune specifically for the very low end
• When you want to tune with a compound other than the
The calibrant delivery system (CDS) is designed specifically for precise
delivery of the Agilent calibrants in a 90:10 acetonitrile:water solvent
mix. Other solvents and mixes may have different flow rates, which
may not be optimum for tuning. Further, the O-ring seals in the CDS are
ethylene-propylene. They can be damaged by aliphatic, aromatic, or
halogenated hydrocarbons. Never use these types of solvents in the
CDS. We recommend delivering non-standard calibrants through the
LC.
Instrument Preparation2
Manual tuning
sacrificing some resolution
of the mass range (< 150 u)
standard calibrants.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide49
Tips for manual tuning
Ionization
modes
PolaritySeparate tuning is required for positive and negative
Separate tuning is not required for electrospray, APCI/APPI
and multimode-ESI+APCI. Tuning involves adjusting the
parameters that control the transmission of ions. It does not
matter how those ions are generated. A tune file created in
electrospray mode often will provide good results for
APCI/APPI samples. The reverse is also true.
ionization. The MS must, at some time, be tuned twice —
once with positive ions and once with negative ions. The two
tunes can be stored under the same tune file name. Positive
ion and negative ion settings can be updated separately.
Page 50
2Instrument Preparation
Manual tuning
Mass rangeTuning is almost always done over the entire mass range.
The exception is if all the ions of interest are smaller than
approximately 150 u. Correct manual tuning over a reduced
mass range can improve transmission and detection of these
very low mass ions.
Steps to manual tune
The online Help includes step-by-step instructions for
manual tuning. The instructions assume the use of one of
the standard calibrants and tuning over the entire mass
range, but they can be altered for non-standard calibrants or
narrower mass ranges.
To access the manual tuning instructions:
1 Open the online Help.
2 Click to expand “How to Work with Your ChemStation.”
3 Click to expand “Basic Tasks.”
4 Click to expand “Method and Run Control.”
5 Click to expand “MSD Tune.”
6 Click the link to “Manually Tuning the LC/MSD or
CE/MSD.”
7 Scroll down until you see “Steps.”
50Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 51
Tune reports
Instrument Preparation2
Tune reports
At the end of every autotune, the system prints a tune
report. You can also manually print a tune report from the
File menu.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide51
Page 52
2Instrument Preparation
Tune reports
Figure 17Tune report (page 1 of 4)
52Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 53
Gain calibration
Gain calibration sets the electron multiplier voltage in a way
that simplifies method development and portability. Gain
calibration is performed automatically during autotune. You
can also check and adjust the gain calibration during manual
tune.
What is gain?
The HED electron multiplier receives an input current that is
generated by the ions that strike it, amplifies that current,
and generates a proportional output current.
Gain = Output current / input current
Gain is controlled by the electron multiplier voltage (EMV).
The higher the EMV, the higher the gain. The relationship
between EMV and gain is log linear. This linear relationship
is common for all multipliers. Since the slope of this line is
constant, to set the gain, the software needs only to adjust
the intercept. The gain calibration routine simply adjusts the
value of the intercept for the specific instrument. The
instrument-specific gain curve coefficients are then stored in
the MS.
Instrument Preparation2
Gain calibration
Gain calibration curve
A gain calibration curve is generated by taking readings of
the ion current generated at discreet multiplier settings
across the range of the multiplier. Separate gain calibration
curves are generated for positive and negative polarities,
since the ion current generated in these two modes varies to
some extent. A gain of 1 is set to be a multiplier value of
1e6. That is, one electron striking the input side of the
detector will generate 1e6 on the output side. A gain check
can be generated on both polarities or on the current
polarity.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide53
Page 54
2Instrument Preparation
Gain calibration
Gain and method portability
The use of gain enables the distribution of methods from
instrument to instrument. In principle, using the same gain
on two different instruments should result in the same
signal response, thereby easing method development and
instrument portability. Because the relationship between
EMV and gain is log linear, a gain of 2.0 should give twice
the abundance of a gain of 1.0.
More gain is not always better
In general, it is best to run the detector at the lowest gain
that still produces adequate abundance. High gains increase
noise as well as signal and often result in poorer
signal-to-noise ratios. Increasing the gain increases the EMV,
which shortens the life-span of the electron multiplier. The
maximum EMV is 3000 eV no matter how high you set the
gain; a gain of 70 or higher elicits the 3000 eV setting.
As an electron multiplier ages, it slowly becomes less
efficient. For a given ion current input, it generates a
smaller and smaller output current (abundance). Low
abundance caused by an aging electron multiplier is not
easily distinguished from low abundance caused by poor ion
generation or transmission (low ion current input). It may be
tempting to compensate for poor ion generation or
transmission by increasing the gain (thereby increasing the
EMV) even though the electron multiplier is actually
performing correctly at its previous gain. The increased gain
will improve abundance, but may decrease the
signal-to-noise ratio and will shorten the life-span of the
electron multiplier.
Check Gain Calibration
Access this functionality via the
Tune menu in the MSD Tune view.
54Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
During a Gain Calibration check, a reading of the ion
current is taken and checked against the current gain
calibration curve for the current polarity. If it is within
acceptable limits, no change is made to the curve for that
polarity. If the reading falls outside the limits, a new gain
curve is generated. A gain of 1 is set to be a multiplier of
Page 55
Instrument Preparation2
Gain calibration
1e6. That is, one electron striking the input side of the
detector will generate 1e6 on the output side. If that
multiplier setting exceeds 2600 volts, which indicates
decreasing electron multiplier performance or instrument
performance, you are warned that system maintenance may
be required.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide55
Page 56
2Instrument Preparation
Gain calibration
56Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 57
Agilent 6100 Series Quadrupole LC/MS Systems
Concepts Guide
3
Data Acquisition
Working with methods 58
Method and Run Control View 58
Loading, editing, saving and printing methods 60
More on editing methods 61
Running samples 64
Running a single sample 65
Running a sequence 66
Flow injection analysis 69
Monitoring analyses 73
Online signal plots 73
Quick method overview 74
Logbooks 74
Instrument shutdown 76
In this chapter, you learn the concepts that help you run
samples and acquire data. This chapter assumes that the
hardware and software are installed, the instrument is
configured and the performance verified. If this has not been
completed, see the Agilent 6100 Series Single Quad LC/MS System Installation Guide.
Agilent Technologies
57
Page 58
3Data Acquisition
Working with methods
Working with methods
Method and Run Control View
You set up methods and run analyses from the Method and
Run Control view, shown in Figure 18.
ChemStation methods control the instrument during data
acquisition. The easiest way to prepare a method is to load
a similar method (or load DEF_LC.M), save it with a new
name, modify (edit) it, and save it again. Then you can use
the method to run a single sample or multiple samples via a
sequence.
You can also set up a method for flow injection analysis
(FIA), where you make multiple injections in a single run.
58Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 59
Data Acquisition3
View Selection
Tool Set Selection: Sequence
or Single Sample
Message Line
Run Control
System Diagram
Top To ol ba r
Online Plot
Method Overview
File List
MethodSequence
Data
Analysis
Status Bar
MS, FIA
Disk
Space
Method and Run Control View
Figure 18In this view, you set up methods, establish instrument settings, and inject samples.
Status and Run Bars: Main TabRight-click the Method icon to see the following commands:
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide59
Run Time Checklist, Method Information, Edit Entire
Method, Method Audit Trail, Print Method, and Help.
Page 60
3Data Acquisition
Sequence
Single sample
Load a
Save
Edit entire
MethodMethodmethod
Select method from
list of recent methods
current
Loading, editing, saving and printing methods
Loading, editing, saving and printing methods
When you need to work with methods, there are several
ways to access the necessary functionality:
• Select from the Method menu, which allows you to load,
• Use the Method icon in the Run Control Bar.
• Use the toolbar or method drop-down list.
Figure 19Toolset selection
edit, save, and print methods.
Right-click the icon to access the context menu.
When you want to view all the tools for methods (or to run a
single sample interactively), you first click the icon for the
Single Sample Toolset, as shown in Figure 19.
Figure 20When you click the icon for the Single Sample Toolset, you access tools to work with methods.
• Use the ChemStation Explorer pane on the left side of the
Method and Run Control view.
Click the Method tab at the bottom to display the method
files.
Double-click a method to load it.
Right-click in the white area to view the context menu,
which contains options such as Load Method.
60Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 61
More on editing methods
NOTE
There are a number of ways to edit a method. If you need to
create a new method for a new analysis, then it is best to
review and edit the entire method, as described just below.
If you want to make only a few changes to an existing
method, then it is easiest to edit only the settings that need
to change, as described in “Editing a portion of the method” on
page 62.
Editing the entire method
You can edit the entire method from the Method menu or
the Edit entire Method icon in the Run Control Bar or
the Method icon in the Status Bar. If you elect to edit the
entire method, the appropriate dialog boxes are displayed in
succession. The first dialog box (Figure 21) lets you select the
parts of the method to view and edit. For more information as
you edit the method, click the Help button in each dialog box.
Data Acquisition3
More on editing methods
Figure 21Parts of the method that you can view and edit
The Run Time Checklist, the last item listed in Figure 21,
controls which parts of the method the software executes when
you run your analysis. As shown in Figure 22, a complete
method can include both data acquisition and data analysis.
Be sure to save the method after you edit it.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide61
Page 62
3Data Acquisition
Run a command or macro prior to an
analysis in a sequence
Run a command or macro at the end of
an analysis (for example, run an Excel macro)
Save a checksum-protected register file
for Good Laboratory Practices (GLP)
Save a complete record of the analysis
conditions
NOTE
More on editing methods
Figure 22Run Time Checklist
Editing a portion of the method
If you need to change only a few settings in a data
acquisition method, you can edit a subset of the method. You
can do that either from the system diagram (Figure 23) or the
Instrument menu.
62Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Figure 23You can right-click a module in the system diagram to access
context menus that allow you to change settings.
Be sure to save the method after you change the instrument settings.
Page 63
Tab l e 1Tasks for setting up the MSD
Data Acquisition3
More on editing methods
To access MSD parameters, refer to the table below.
If you want to:Click the MS icon and select this
menu item:
Set scan range or SIM ionsSet up MSD SignalsSet up MSD Signals
Set instrument polaritySet up MSD SignalsSet up MSD Signals
Set drying gas flow or temperature or
nebulizer pressure
Set the capillary voltageSpray ChamberMore MSD > Spray Chamber
Select MSD parameters to store during
a run
Spray ChamberMore MSD > Spray Chamber
Data CurvesMore MSD > Data Curves
Or choose the following from the
Instrument menu
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide63
Page 64
3Data Acquisition
Display Sequence Toolset
Display Single Sample Toolset
Running samples
Running samples
The LC/MS ChemStation provides three ways to run
samples:
• Run a single sample interactively and create a single data
file.
• Run sample(s) via a sequence and create a data file for
each sample.
FIA = flow injection analysis• Run multiple injections within a single data file, via FIA.
You can use the same method to run either a single sample
or samples in a sequence.
64Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
FIA icon
Figure 24To run samples, you first click the appropriate icons.
For FIA, you must modify the method to enable multiple
injections in the same run.
Page 65
Running a single sample
Single Sample
Click to enter sample information
Click to start the run
After you have set up and saved a method, you are ready to
run a single sample interactively as shown below. First you
click the icon to display the Single Sample Toolset.
Data Acquisition3
Running a single sample
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide65
Figure 25How to enter sample information and start the run.
Page 66
3Data Acquisition
Sequence
Running a sequence
Running a sequence
To run samples in an automated, unattended sequence, you
first click the icon to display the Sequence Toolset. The
Sequence icon on the left provides access to the Sequence
Toolset and the sampling diagram that shows the sample
tray.
Figure 26Sequence Toolset selection icon.
Loading, editing, saving and printing sequences
When you need to load, edit, save, and print sequences,
there are several ways to access the necessary functionality:
• Select from the Sequence menu.
Figure 27The Sequence menu allows you to load, edit, save, and print
sequences.
66Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 67
• Use the toolbar or sequence drop-down list.
LoadSaveSelect sequence from
list of recent sequences
SequenceSequence
NOTE
Figure 28Tools to work with sequences
• Use the Method and Run Control file list. If necessary,
• Click the sampling diagram to access the menu.
More on editing sequences
When you edit a sequence, the following dialog boxes are
applicable:
• Sequence Parameters – allows you to enter information
• Sequence Table – includes the sample list and any
• Sequence Output – controls printing of reports (sample
• Import Sequence – allows you to import a sample table
Data Acquisition3
Running a sequence
first click the Sequence tab at the bottom. You can
right-click in the white area to view the menu, or
double-click a sequence to load it.
like data file path, parts of the method to run, and a
post-sequence command/macro to shut down the system
calibration information
results and the sequence summary)
from a text (.txt) file or a delimited file, such as a
comma-separated value (.csv) file created in Microsoft®
Excel
After you have edited your sequence, be sure to save it.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide67
Page 68
3Data Acquisition
Click to start the sequence
Click Run Sequence
button in Sequence
Tab l e an d P ar t i al
Sequence dialog boxes
Click to start the sequence
Running a sequence
Starting the sequence
After you have set up and saved your sequence and any
methods that it uses, you are ready to run the sequence. As
shown in Figure 29, the software provides more than one way to
start the sequence.
68Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Figure 29Multiple ways to start the sequence.
Page 69
Flow injection analysis
Flow injection analysis (FIA) is injection of multiple samples
within the same run, and is performed without a column.
FIA is very useful when you want to optimize MS
parameters such as:
• Drying gas flow and temperature (which depend on the
mobile phase and flow rate, as well as the ionization
mode).
• Nebulizer pressure (which depends on the mobile phase
and flow rate, as well as the ionization mode).
• Fragmentor voltage (which depends on the structure of
the compound).
FIA is also useful to check quickly for compound sensitivity
or linearity, or any time you want to perform an analysis
without a separation.
Setting up FIA within a method
You can add FIA to a method either through the Method
menu or the FIA icon (see Figure 30).
Data Acquisition3
Flow injection analysis
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide69
Page 70
3Data Acquisition
Click to enable FIA or set up FIA series
FIA icon
NOTE
Flow injection analysis
Figure 30Two ways to enable FIA within a method
To optimize settings with a single sampleThe Edit FIA Series
dialog box has an autofill function so you can automatically
build an FIA table. You can select up to two MSD
parameters, and then automatically increment their settings.
This feature makes it easy to set up FIA to optimize settings,
as shown with the example in Figure 31. If you decide later to
add a setpoint, you can use the insert row/append row
functionality.
Some parameters, such as gas temperature, take time to reach their
setpoints. In these cases, perform multiple injections at each setpoint and
allow more time between injections.
70Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 71
Data Acquisition3
Automatically build an FIA table
Mark check box first
Flow injection analysis
Figure 31The FIA series is a convenient way to optimize MS parame-
ters. The example shows a test of fragmentor voltage.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide71
Page 72
3Data Acquisition
Add rows manually or automatically
Flow injection analysis
To inject multiple samplesIf you need to inject multiple
samples to check for sensitivity or linearity, you can also use
the autofill functionality within the Edit FIA Series dialog
box. In this situation, you keep the MS parameters constant
and increment only the vial numbers. Alternatively, you can use
the insert row/append row functionality to add samples.
Figure 32 shows an example of FIA for a linearity check.
72Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Figure 32For pure samples, the FIA series can be used to check sensi-
tivity or linearity. The example shows a linearity check.
Running multiple FIA methods
If you have multiple compounds to test or you need to
perform more complex method development, you can run a
succession of FIA methods, analogous to a sequence. You
access that capability via the RunControl menu.
Page 73
Monitoring analyses
Online signal plots
Data Acquisition3
Monitoring analyses
During an analysis, you can view online signals, as shown in
Figure 33. (Select View > Online Signals.)
You can click the little box with the arrows in the lower
left of the signal display to display the online plot as a
separate window. This feature is useful if you wish to
make the window larger so you can view more signals.
For the online plots, you can choose which signals to display,
and you can adjust the scales of the x and y axes.
Figure 33Online plot
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide73
Page 74
3Data Acquisition
Quick method overview
Quick method overview
Just to the right of the online plot, you can get a quick
overview of your method parameters. You can select which
type of settings to view, including LC, MS, and data analysis
settings. You can click each parameter to display the dialog
box to change that parameter.
Several logbooks are important in the Method and Run
Control view. You can view logbooks from the View menu or
the logbook icons as show in Figure 35.
Current logbook
The current logbook stores all error, system, and event
messages, and is helpful for general information and
troubleshooting. It logs events such as ChemStation startup
and shutdown, loading of sequences and methods, running of
sequences and methods, etc.
The log file exists as INSTRx.LOG (Where x is the instrument
number), in the temporary folder under the instrument
folder in CHEM32 (for example, C:\CHEM32\1\TEMP).
74Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 75
Data Acquisition3
Current
logbook
icon
Sequence
logbook
icon
Logbooks
Run logbook
RUN.LOG stores all messages related to analysis of a
particular sample. It resides in the data file folder. You can
use it to check that the sample was processed correctly.
Sequence logbook
The sequence logbook shows what happened as a sequence
ran. You can use it to troubleshoot errors that occurred
during unattended operation.
The log file exists as sequencename.LOG (where sequencename
is the name of the ChemStation sequence). Each time a
sequence runs, the software produces a log file, usually in
the DATA folder within the instrument folder (for example,
C:\CHEM32\1\DATA).
Figure 35How to open logbooks with icons.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide75
Page 76
3Data Acquisition
Instrument shutdown
Instrument shutdown
At the end of your analysis, if you do not plan to run more
samples until the next day, it is important to:
• Flush the system with a mobile phase without buffers.
• Put the system into standby mode.
For details, see the Basic Operation section of the Agilent
6100 Series Quad LC/MS Systems Quick Start Guide.
76Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 77
Agilent 6100 Series Quadrupole LC/MS Systems
Concepts Guide
4
Data Analysis
The Data Analysis View 78
Loading and manipulating chromatograms 80
Loading signals 81
Removing signals from the chromatogram display 85
Changing how chromatograms are displayed 85
Working with spectra 87
Displaying spectra 88
Peak purity 89
Performing quantification 90
Integrating peaks 90
Calibration 92
Data review and sequence reprocessing 94
The Navigation Table 94
Batch review 94
In this chapter, you learn the concepts that help you analyze
data. For more information on the concepts related to data
analysis, see these two manuals:
• Agilent ChemStation: Understanding your ChemStation
• Agilent ChemStation for LC 3D Systems: Understanding
Your Spectra Module
For more in-depth task-related information and specifics
about dialog boxes and toolbars, see the online Help. For
descriptions of all buttons on toolbars, double-click User Interface Reference, then double-click Toolbars in online
Help.
Agilent Technologies
77
Page 78
4Data Analysis
The Data Analysis View
The Data Analysis View
The Data Analysis view, shown in Figure 36, provides all the
tools for qualitative and quantitative data analysis. In this view,
you can:
• Evaluate chromatograms and spectra
• View and compare signals from both the MS and UV
• Integrate peaks and perform quantification
• Check peak purity
• Perform deconvolution (with the optional Agilent LC/MSD
• Generate reports from predefined report templates
• Reprocess data from sequences.
detectors
Deconvolution & Bioanalysis Software)
78Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 79
Data Analysis4
Title bar
LC/MS signal
Message line
Toolset selection
Top toolbar
DAD signal
File list
(ChemStation Explorer)
Cursor tools
View selection
Navigation
Ta bl e
Signal view selection
Navigation toolbarGraphics tools
The Data Analysis View
Figure 36The Data Analysis view provides menus and tools for qualitative and quantitative analysis.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide79
Page 80
4Data Analysis
Loading and manipulating chromatograms
Loading and manipulating chromatograms
In ChemStation software,
chromatograms are called signals.
TIC – total ion chromatogram
EIC – extracted ion chromatogram
BPC – base peak chromatogram
ChemStation software stores multiple signals in the same
data file. Examples of signals are MS chromatograms and UV
chromatograms.
For many analyses, it is important to be able to compare
results from different detectors or different instrument
settings, or to compare results from various samples.
ChemStation software allows you to load and compare
signals within a data file and between data files. For
example, within a data file, you can simultaneously load and
evaluate UV and MS data, or MS data at two different
fragmentor settings or ion polarities. You can load data files
from multiple samples to compare retention times and peak
heights.
Specialized chromatograms available for MS data include:
• Total ion chromatogram (TIC) – chromatogram that sums
all the ions at each time point
• Extracted ion chromatogram (EIC) – chromatogram of a
chosen m/z value or range of values
• Base peak chromatogram (BPC) – chromatogram that
represents the abundance of the largest peak of each
spectrum in the data file
Total ion chromatograms and base peak chromatograms are
useful to find all the chromatographic peaks in the sample.
Because of their specificity, extracted ion chromatograms are
ideal for quantification. You can also use them to check for
purity of chromatographic peaks. When chromatographic
peaks are composed of more than one component, the EICs
often fail to align.
80Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 81
Loading signals
View file informationSelect instrument curves to display
Specify integration
Select which signals to load
(CTRL-click for multiple signals)
Ways to load signals
You can load chromatograms (signals) in the following ways:
• Select from the File menu.
Data Analysis4
Loading signals
Figure 37The Load Signal dialog box lets you select which data files and signals to load.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide81
Page 82
4Data Analysis
Load
Signals
Load signals and overlay
with current signals
Batch
Single runs
Sequence
Loading signals
• Use the top toolbar.
Figure 38Tools to load signals.
• Use the Data file list in the ChemStation Explorer.
(If necessary, first click the Data tab at the bottom.)
Figure 39You can right-click the vials to view the menu, or double-click
the vials to load multiple data files.
82Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 83
Data Analysis4
Double-click a
signal to view it
Click overlay signals
to view signals from
multiple data files
Click the “+”
to expand
Loading signals
Refining your selection
Once you have loaded signals for a data file or files, you can
refine your selection as shown in Figure 40 and Figure 41.
Figure 40You can view specific signals by double-clicking a signal (on
left) or by right-clicking to display the menu at right.
Figure 41You can view specific signals by selecting from the signal list.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide83
Page 84
4Data Analysis
Overlay signal(s) – can be from different data files
Load signal(s) from a data file
Display extracted ion chromatograms
Overlay a base peak chromatogram
Subtract background spectrum from every point
in the data file (see online Help for BSB)
Loading signals
More about signals
In addition to loading signals from a single data file, the
File menu allows you to load signals from multiple files and
to display extracted ion chromatograms and backgroundsubtracted chromatograms.
Figure 42The File menu provides ways to load chromatograms and per-
form background subtraction.
In ChemStation software,
“overlay” means to simultaneously
load more than one signal.
In the File menu, Overlay Signal allows you to compare
signals from different samples or injections so you can
examine fine differences. Note that while the term “overlay”
may imply that you display the signals one on top the other
in the same window, you actually control the display using
buttons that you access from the Signal Toolset, the Purify
Toolset, or the Graphics Toolset. These buttons are:
Display overlaid
(In the Graphics Toolset, this
button toggles between
overlaid and separate.)
84Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Display separate
Page 85
Data Analysis4
Removing signals from the chromatogram display
Removing signals from the chromatogram display
If you display too may signals at the same time, the
individual displays can become too small or difficult to view.
If so, you can remove signals from the display in either of
the following ways:
• Select View > Window Functions > Delete Window.
• Use the tool within the Signal Toolset and the Purify
Toolset that allows you to delete signals from your
chromatogram display.
Changing how chromatograms are displayed
When you analyze data, it is very helpful to have control
over the data display. For example, you may want to display
chromatograms in separate windows or overlaid in a single
window. Separate windows are often the best choice if you
wish to view UV and MS chromatograms from the same
analysis. Overlaid windows may be better if you need to
compare fine differences between two complex samples.
It is also helpful to be able to zoom in and out, to add
annotations, to align chromatograms (for example, UV and
MS chromatograms), and to control whether or not retention
times and integration baselines are displayed.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide85
Page 86
4Data Analysis
Define how chromatograms are displayed
Zoom in and out
Add and edit annotations
Changing how chromatograms are displayed
The chromatogram display can be customized as follows:
• Select from the Graphics menu.
Figure 43The Graphics menu provides ways to change how the chro-
matograms are displayed.
• Use the toolsets that are described below.
• Graphics Toolset
The graphics tools provide ways to change your
signal display. You can add annotations and change
whether compound names, retention times, baselines,
titles, and axes are displayed. You can print windows
and copy them to the clipboard.
The icon to display the Graphics Toolset is located near
the middle of the Data Analysis window.
• Signal Toolset
The signal tools let you change the
chromatogram display. You can change from a separate
display to overlaid, align x- and y-axes, display signals
full-scale or on the same scale, smooth chromatograms,
subtract chromatograms (for example, subtract a blank
run) and perform other operations on chromatograms.
86Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 87
Working with spectra
Data Analysis4
Working with spectra
With ChemStation software, you can display both UV and
mass spectra.
In ChemStation software,
background spectra are called
reference spectra.
With MS data, each point in a chromatogram has an
associated mass spectrum. You can display single spectra,
averaged spectra, and background-subtracted spectra. For
LC/MS data, you generally subtract a background spectrum
from each analyte spectrum. When you subtract a
background spectrum, you remove the ions that are
attributable to mobile phase and other background.
For UV signals, depending on the storage mode, there may
be spectra only for the peak apexes or across the peaks.
There is a spectrum for each point in the chromatogram
only when you set Store to All when you set up the UV
detector portion of the method.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide87
Page 88
4Data Analysis
Select spectra
Select background spectra for subtraction (background subtraction must also be enabled in
Spectra Options)
Set options for spectral display and specify how
background subtraction is done (MS Reference)
Specify ion labeling
Examine peak purity by MS and/or UV
Displaying spectra
Displaying spectra
You can display spectra in the following different ways:
• Select from the Spectra menu.
Figure 44The Spectra menu provides ways to select and display spec-
tra.
• Use the toolsets that are described below.
• Spectrum Toolset
The spectrum tools help you to quickly
display and evaluate spectra. You can select spectra
(including background spectra), zoom in and out, save
spectra to a library, search a spectral library, set
spectral options, select a chromatographic peak for
purity analysis, and perform other tasks related to
spectra.
• Calculation Toolset
The calculation tools allow you to multiply, add,
and subtract spectra. You access them from the icon
shown here, which is part of the Spectrum Toolset.
88Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 89
Peak purity
Data Analysis4
Peak purity
For many analyses, it is important to know whether a
chromatographic peak represents a single component or
multiple components. ChemStation software allows you to
evaluate peak purity using UV and/or MS data.
To start, you select Spectra > Select Peak Purity. For
detailed concepts about peak purity, see Agilent
ChemStation for LC 3D Systems: Understanding Your
Spectra Module.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide89
Page 90
4Data Analysis
Performing quantification
Performing quantification
ESTD – external standard
calibration
ISTD – internal standard calibration
A calibration curve is a plot of peak
area or height on the y-axis versus
concentration on the x-axis.
Integrating peaks
Integration – the process of
measuring peak areas
Qualifier ions – ions that are used
to verify that the ion used for
quantification is attributable to the
analyte rather than an interference
Quantification is the process of determining how much of a
compound is present in a sample. You perform quantification
by comparing the peak areas or heights of analytes in your
sample with those in standards of known concentration. To
each standard and sample, you may add a constant amount
of an internal standard that you can use to correct for slight
variations in injection volume, detector sensitivity, etc.
For quantification, it is best to use SIM mode for data
acquisition because it generates more data points across a
peak, which provides the best precision and accuracy. When
you perform quantification, you typically analyze standards
at concentrations that bracket those of your samples. Then
generate a calibration curve for each analyte, and to
determine the amount of each analyte in your sample, using
ChemStation software.
Determination of peak areas is a basic component of
quantification. To avoid interferences, you use the peak
areas from extracted ion chromatograms (EICs) rather than
total ion chromatograms. You can use the EIC of the
molecular ion for quantification and the EIC(s) of one or
more fragment ions for confirmation. The latter are called
qualifier ions, and they must be present in the proper ratio
to confirm the presence of the analyte.
In the ChemStation software, the Signal Details specify
which extracted ion signals are used for quantification and
which (if any) are used for qualifier ions. You can access the
Signal Details under the Calibration menu, and you can save
the Signal Details within the method.
To ensure that peaks are properly integrated, you need to
establish the appropriate integration parameters. The
standard of lowest concentration usually presents the
greatest challenge for integration, so it is best to use it to
90Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 91
Data Analysis4
Perform integration
Establish integration events with an events table
Display tabular integration results
Establish manual integration events
Use manual integration events in method
NOTE
Integrating peaks
refine your integration settings. You can then save the
integration parameters with the method and apply them to
the higher-level standards and the samples.
Set integration parameters and perform peak integration in
the following different ways:
• Select from the Integration menu.
Figure 45The Integration menu
• Use the Integration Toolset.
The integration tools help you perform
integration and generate reports. You can define integration
events, perform automatic and manual integration, draw
baselines, zoom in and out, specify a report style, and view
and print reports.
When you are done, be sure to save the integration settings with your
method.
For more information about integration concepts, see Agilent
ChemStation: Understanding your ChemStation.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide91
Page 92
4Data Analysis
Calibration
Calibration
A calibration curve is a plot of peak
area or height on the y-axis versus
concentration on the x-axis.
When you perform calibration, you insert the peak
integrations from standards into a calibration table. The area
responses are used to generate a calibration curve.
In ChemStation software, each concentration of standard is
called a “level.” After you integrate the standard of lowest
concentration, you add it to the calibration table as level 1.
You then add the standard of second-lowest concentration as
level 2, and so on, until you have added all standards.
Tip: If you have saved the Signal Details (see the Calibration
menu) as part of the method, you can automatically load
and integrate the extracted ion chromatograms from each
standard. (See Figure 46.)
Figure 46The highlighted features make it faster to add levels to a cali-
bration curve.
92Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 93
Data Analysis4
Set up a new calibration table
Establish general settings like curve fit, ISTD/ESTD
Define the signals (such as EICs) to use for
quantification
Add a level to the calibration table
Set detailed options for the calibration table
NOTE
Calibration
Access to calibration tools and commands are provided in
the following ways in ChemStation software:
• Select from the Calibration menu.
Figure 47The Calibration menu
• Use the Calibration Toolset.
The calibration tools provide a quick way to
accomplish calibration tasks and generate reports. You can
create a new calibration table, add levels and peaks to the
table, recalibrate using the current chromatogram, and
perform other calibration tasks. You can also specify a
report style, and view and print reports.
When you are done, be sure to save the calibration table with your
method.
For more information about calibration concepts, see Agilent
ChemStation: Understanding your ChemStation.
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide93
Page 94
4Data Analysis
Sequence reprocessing toolsData review tools
Choose method for
data review
Click “+” for sample
details (signals, etc.)
Click for help with
Navigation Table
Data review and sequence reprocessing
Data review and sequence reprocessing
ChemStation software gives you two ways to quickly review
multiple data files: the Navigation Table and Batch review
described below.
The Navigation Table
The Navigation Table shows the data files from a given
folder and provides a convenient way to navigate between
samples. You can use it for data review or to reprocess
sequences.
Figure 48Navigation Table, shown with columns for sequence runs
Batch – a user-selected series of
data files from a sequence that is
94Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Batch review
processed using a single
user-defined method
Batch review enables fast review of sequences or selections
of analyses from a sequence. You can use batch review for
either quantitative or qualitative analysis. Whenever you run
a sequence, a batch file (with a .b extension) is generated
and saved in the same folder as the data files. This batch
file contains pointers to the data files in the batch review
itself.
Page 95
Data Analysis4
Preview and output report of results for the batch
Load batch file and select method and data files
to be included in batch review
View log of all actions related to the batch
Sort samples
Click for help
Batch table
Compound table
Batch review
You load a batch through the Batch menu (Figure 49). When you
load a batch, you have the opportunity to select the method to
use for the batch, and to individually select the desired data
files.
Figure 49The Batch menu
When you load a batch, the Batch Review toolbar, the batch
table, and the compound list are displayed automatically. For
help, click the question mark on the toolbar.
Figure 50Batch Review toolbar, batch table, and compound list
Agilent 6100 Series Quadrupole LC/MS System Concepts Guide95
Page 96
4Data Analysis
Batch review
96Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 97
Agilent 6100 Series Quadrupole LC/MS Systems
Concepts Guide
5
Reports
Using predefined reports 98
Generating reports 98
Report styles 99
Defining custom reports 101
Summary of process 101
Example report templates 101
The Report Layout View 102
In this chapter, you learn the concepts that help you
generate reports of results.
For more concepts related to reports, see these two manuals:
• Agilent ChemStation: Understanding your ChemStation
• Agilent ChemStation for LC 3D Systems: Understanding
Your Spectra Module
For in-depth task-related information and specifics about
dialog boxes and toolbars, see the online Help. For
descriptions of all buttons on toolbars, in the online Help
double-click User Interface Reference, then double-click
Toolbars.
Agilent Technologies
97
Page 98
5Reports
Specify report style, destination, etc.
Edit parameters for system performance reports
Specify how library search is done
Calculate and
print report
Calculate and
preview report
Specify report
Using predefined reports
Using predefined reports
A report provides qualitative or quantitative information
about a sample you have analyzed. You can print a report,
display it on your computer screen, or send it to a file.
ChemStation software includes a number of predefined
report styles, and you can generate new styles for custom
reports.
Generating reports
Ways to generate reports in Data Analysis view
• Select from the Report menu.
Figure 51The Report menu provides ways to print reports and set prop-
erties of reports.
• Use the report tools in the Integration
Toolset or the Calibration Toolset. You
access these toolsets by clicking one of
the buttons on the right.
Figure 52Tools to work with reports
98Agilent 6100 Series Quadrupole LC/MS System Concepts Guide
Page 99
Report styles
NOTE
Reports5
Report styles
System suitability reports
In the Report menu, System Suitability allows you to check
the performance of your system before and during your
analyses. This report style allows you to set up high and low
limits for retention time, peak height, peak width, theoretical
plates, resolution, library match, peak purity, and other
properties of calibrated peaks. When you choose one of the
Performance report styles, the report indicates when any of
the parameters are out of specification.
When you are done, be sure to save the report parameters with your
method.
ChemStation software provides a wide variety of report
styles. Most of them are described in one of these two
manuals:
• Agilent ChemStation: Understanding your ChemStation
• Agilent ChemStation for LC 3D Systems: Understanding
Your Spectra Module
Two report styles are specific to LC/MS, and are not
discussed in the above manuals.
• LC/MS – includes:
• Header information from the data file
• A total ion chromatogram (optional)
• Tabulated and graphical spectra for each integrated
peak
• Deconvolution results (available with the optional