Agilent 6060B Data Sheet

The 6060B and 6063B each provides one load input. This is more conve­nient for single input applications than a mainframe product.
These electronic loads are particu­larly suited for the lab bench. Entering commands manually using the front panel keypad is simpler because the channel does not need to be specified, as in a mainframe configuration. The keypad entry is further simplified because these products do not have the download­able LIST feature of the N3300A Series, which helps to maximize production throughput. Extensive protection is included to help protect your valuable prototypes under test. This includes overvolt­age, overcurrent, overtemperature, overpower, and reverse polarity.
These loads are suitable for manufacturing test systems where maximizing speed is not critical. They use industry standard SCPI instructions, and also have VXIPlug&Play drivers to simplify system design. For the greatest speed and accuracy in programming and measurement, see the N3300A Series of DC electronic loads.
DC Electronic Loads maximize throughput with real-life loading conditions
Specifications
6060B 6063B
More detailed specifications at
www.agilent.com/find/6060
104
Cost-effective for single input applications Convenient optional front panel input connection
6060B and 6063B
Single-Input
250 W to 300 W
Amperes 0 to 60 A 0 to 10 A Volts 3 to 60 V 3 to 240 V Maximum power (at 40° C) 300 W 250 W Constant current mode
Ranges 0 to 6 A, 0 to 60 A 0 to 1 A, 0 to 10 A Accuracy 0.1% ±75 mA 0.15% ±10 mA Regulation 10 mA 8 mA
Constant voltage mode
Accuracy 0.1% ±50 mV 0.12% ±120 mV Regulation (w/remote sense) 10 mV 10 mV Constant resistance mode 0.033 to 1.0 0.20 to 24.0 Ranges 1 to 1,000 24 to 10,000
10 to 10,000 240 to 50,000
Accuracy
1 Ω: 0.8% ±8 mΩ (with ≥6 A at input) 24 : 0.8% ±200 mΩ (with ≥1 A at input) 1 K: 0.3% ±8 mS (with 6 V at input) 10 K/: 0.3% ±0.3 mS (with 24 V at input) 10 KΩ: 0.3% ±8 mS (with ≥6 V at input) 50 KΩ: 0.3% ±0.3 mS (with ≥24 V at input)
Transient generator
Frequency range 0.25 Hz to 10 kHz 0.25 Hz to 10 kHz Accuracy 3% 3%
Duty cycle range 3 to 97% (0.25 Hz to 1 kHz) 3 to 97% (0.25 Hz to 1 kHz)
6 to 94% (1 to 10 kHz) 6 to 94% (1 to 10 kHz)
Accuracy 6% of setting ±2% 6% of setting ±2% Current level high range 60-A range: 10-A range:
Accuracy 0.1% ±350 mA 0.18% ±50 mA Current level low range 6-A range: 1-A range:
Accuracy 0.1% ±80 mA 0.18% ±13 mA Voltage level 3 to 60 V 3 to 240 V
Voltage level accuracy 0.1% ±300 mV 0.15% ±1.1 V
Readback specifications
Current readback accuracy 0.05% ±65 mA 0.12% ±10 mA Voltage readback accuracy ±(0.05% + 45 mV) ±(0.1% + 150 mV)
Ripple and noise
(20-Hz to 10-MHz noise) 4 mA rms 1 mA rms Current 40 mA peak-to-peak 10 mA peak-to-peak Voltage 6 mV rms 6 mV rms
105
Notes:
1. Operating temperature range is 0° to 55°C. All specifications apply for 25°C ±5°C, except as noted.
2. Maximum continuous power available is derated linearly from 40°C to 75% of maximum at 55°C.
3. DC current accuracy specifications apply 30 seconds after input is applied.
Single-Input: 250 W to 300 W (Continued)
Specifications
More detailed specifications at
www.agilent.com/find/6060
DC Electronic Loads maximize throughput with real-life loading conditions
6060B 6063B
Supplemental Characteristics (Non-warranted characteristics determined by design that are
useful in applying the product) Constant current mode 60-A range: 16 mA 10-A range: 2.6 mA Resolution 6-A range: 1.6 mA 1-A range: 0.26 mA Temperature coefficient 100 ppm/°C ±5 mA/°C 150 ppm/°C ±1 mA/°C
Constant voltage mode
Resolution 16 mV 64 mV Temperature coefficient 100 ppm/°C ±5 mV/°C 120 ppm/°C ±10 mV/°C Constant resistance mode 1 Ω : 0.27 m 24 : 6 m
Resolution 1 K: 0.27 mS 10 K: 0.011 mS
10 K: 0.027 mS 50 K: 0.001 mS
Temperature coefficient 1 : 800 ppm/°C ±0.4 m/°C 24 Ω: 800 ppm/°C ±10 mΩ/°C
1 K: 300 ppm/°C ±0.6 mS/°C 10 K: 300 ppm/°C ±0.03 mS/°C 10 K: 300 ppm/°C ±0.6 mS/°C 50 K: 300 ppm/°C ±0.03 mS/°C
Transient generator
Frequency range 0.25 Hz to 10 kHz 0.25 Hz to 10 kHz Resolution 4% or less 4% or less
Duty cycle range 3 to 97% (0.25 Hz to 1 kHz) 3 to 97% (0.25 Hz to 1 kHz)
6 to 94% (1 to 10 kHz) 6 to 94% (1 to 10 kHz)
Resolution 4% 4% Current level high range 60-A range: 10-A range:
Resolution 260 mA 43 mA Current level low range 6-A range: 1-A range:
Resolution 26 mA 4 mA Current temperature coefficient 100 ppm/°C ±7 mA/°C 180 ppm/°C ±1.2 mA/°C Voltage level resolution 260 mV 1 V Voltage temperature coefficient 150 ppm/°C ±5 mV/°C 120 ppm/°C ±10 mV/°C Programmable slew rate 60-A range: 1 A/ms to 5 A/µs 10-A range: 0.17 A/ms to 0.83 A/µs
6-A range: 0.1 A/ms to 0.5 A/µs 1-A range: 17 A/ms to 83 A/ms
Rise/fall time 12 µs to 8 ms 16 µs to 8 ms
Analog programming bandwidth 10 kHz (–3 dB frequency) 10 kHz (–3 dB frequency) Analog programming accuracy
Current (low range) 4.5% ±75 mA 3% ±8 mA Current (high range) 4.5% ±250 mA 3% ±20 mA Temperature coefficient 100 ppm/°C ±6 mA/°C 150 ppm/°C ±1 mA/°C Voltage 0.8% ±200 mV 0.5% ±150 mV Temperature coefficient 100 ppm/°C ±1 mV/°C 120 ppm/°C ±10 mV/°C
Analog programming voltage 0 to 10 V 0 to 10 V Readback specifications 17 mA (via GPIB) 2.7 mA (via GPIB)
Current readback resolution 20 mA (front panel) 10 mA (front panel) Temperature coefficient 50 ppm/°C ±5 mA/°C 100 ppm/°C ±1 mA/°C Voltage readback resolution 17 mV (via GPIB) 67 mV (via GPIB)
20 mV (front panel) 100 mV (front panel)
Temperature coefficient 50 ppm/°C ±1.2 mV/°C 100 ppm/°C ±8 mV/°C
Loading...