Agilent 4000, 5000, 6000i, 7000 User Manual

Agilent Cary UV-Vis (NIR) Spectrophotometers
Models 4000, 5000, 6000i and 7000
Long Term Shut Down and Start Up Instructions

Shut down the system

To shut down the system:

1 Turn off the instrument purge from the supply (if applicable). 2 Turn off the instrument power from the panel switch. 3 Disconnect the mains power lead. 4 Remove any installed accessories and store them in their original packaging. 5 Fit the sample compartment door. 6 Close the sample compartment lid. 7 Shutdown the controlling PC.

Start up the system

To start up the system:

1 Turn on the instrument purge from the supply (if applicable). 2 Connect the instrument power lead. 3 Switch on the instrument. 4 Switch on the controlling PC. 5 Allow a 2-hour instrument warmup period. 6 Open the Validate application and run the Instrument Performance Tests.
www.agilent.com
© Agilent Technologies, Inc. 2020
Edition 1 04/20