Maximum Input
1 Mohm +/- 250 V (dc + ac) [ac< l0 kHz]
50 ohms: 5 V rms
Offset Range
Vertical Sensitivity Available Offset
1 mV to 50 mV/div ±2 V
>50 mV to 250 mV/div ±l0 V
>250 mV to 1.25 V/div ±50 V
>1.25 V/div to 5 V/div ±250 V
Offset Accuracy
±(l.25% of channel offset + 2% of full scale)
Voltage Measurement Accuracy
Dual Cursor ±[(1.25%)(full scale)+(0.032)(V/div)]
Single Cursor ±[(1.25%)(full scale)+(offset accuracy)+
(0.016)(V/div)]
Horizontal Specifications
Time Base Range
500 ps/div to 5 s/div
0.1 % of full scale)]
Real Time [(0.005%)(delta-t)+(0.2)(sample period)]
Peak Detect [(0.005%)(delta-t)+(l sample period)]
Time Tag
Resolution 100 ps
Accuracy ±[0.005%(reading)+l00ps]
Delay Range (posttrigger) 10E07 x sample period
Delay Range (pretrigger) 32K x (sample period)
Trigger Specifications
Sensitivity dc to 100 MHz 100 MHz to 500 MHz
Internal 0.5 div 1.0 div
Auxiliary dc to 50 MHz: 250 mVp-p
Pulse Width (minimum) 1 ns
Level Range
Internal ±1.5 X full scale from center screen
Auxiliary ±5V
Modes Edge, pattern, glitch, time qualified pattern,
line, state event-delayed, time-delayed, TV
(NTSC, PAL, and user-definable formats)
Trigger Coupling dc, ac, low frequency reject
Display Characteristics
Display Mode "A" models feature monochrome CRT (Cathode Ray Tube)
displays. "C" models feature flatpanel color TFT
(Thin Film Transistor) liquid-crystal displays.
Modes Averaging from 2 to 2048, envelope, infinite &
variable persistence from 500 ms to 10 s, connect
the dots, peak detect
Graticules full grid, axes, frame, or no graticule
Measurements
Automatic 23 measurements on front panel or over HP-IB:
Rise time V p-p V dc rms
Fall time V min Preshoot
Frequency V max Overshoot
Period V avg Voltage at time
- Width V base Time at min voltage
+ Width V amptd Time at max voltage
Duty cycle V top Time at voltage
Delta time V ac rms
User-Definable
Both upper and lower thresholds can be set from
-25% to 125% for all automatic measurements
Modes
Continuous, statistics, limit test, or waveform
compare
Math/Analysis Functions Operators
add, subtract, multiply, versus, integrate,
differentiate, invert magnify, and FFT
Additional Characteristics
Peak Detect Captures and displays glitches or other high-speed
events as narrow as 1 ns in real-time mode at sample
rates of 250 MSa/s or less with sequential single shot turned off.
Sequential Captures successive single-shot events without
Single Shot capturing the dead time in between, and stores up
to 400 Kbytes of waveform data.
Sequential Single-shot Throughput
Record Waveforms Stored Max Number of
Length Per Second Waveforms Stored
50 1333 4395
500 1111 739
8000 294 49
32,000 90 12
Waveform Store
4 nonvolatile, 2 pixel (volatile), and segmentable
memory for storing measurement failure waveforms or
sequential single-shot waveforms.
Screen Update Rate (typical at 500 ns/div)
record length (points)
Real Time 500 8K 16K 32K
updates/s. 150 110 84 58
Repetitive normal 8 avgs 128 avgs
updates/s: 150 91 91
Optional Capabilities
Optional Telecommunication-Mask Application
Make telecom mask-template measurements to ANSI, CCITT, and ISDN
standards using the HP 54520 or 54540 series oscilloscopes equipped
with Option 001. Option 001 features 21 standard masks stored on a
flexible disk. You can automatically trigger on positive isolated
ones in live traffic for many standard telecom signals. The scope
will automatically best-fit the test signal to many masks and give
automatic pass-fail comparisons of the mask to its corresponding
input signal.
For more information on Option 001, Telecommunications Masks
Applications ask your HP Sales representative for HP Product Overview
5963-1859E (international version) or 5963-1859 EUS (US version).
Optional Active Probing
The HP 1144A active probe features 800-MHz bandwidth, 2-pF input
capacitance, and 1-Mohm input resistance. The HP 1145A two-channel
active probe features 750-MHz bandwidth, 2-pF input capacitace, and
1 Mohm input resistance and is designed for easy connection to
surface mount devices.
I/O Characteristics
HP-IB Fully programmable, complies with IEEE 488.2
RS-232 Bidirectional serial communication
Centronics Talk-only parallel communication for HP PCL
compatible printers
Hardcopy Supported Printers
Color "C" Models Monochrome "A" Models
HP 540 Centronics Thinkjet HP-IB, RS232
HP 560C Centronics HP 540 Centronics
HP 560C, Centronics
B/W only
Supported Plotters:
HP Color Pro HP-IB HP 7440A HP-IB
HP 7475A HP-IB HP 7475A HP-IB
HP 7470A HP-IB HP 7470A HP-IB
HP 7550A HP-IB HP 7550A HP-IB
Disk Drive
DOS-compatible, 3-1/2" disk drive to store and recall
setups, waveforms, and screen images (TIFF, PCX, and
EPS). The disk drive is also utilized to download
the oscilloscope's operating-system software and can
be used to upgrade the scope's software as future
upgrades become available. A free software-upgrade
notification service is offered to registered users.
Front Panel Setups
9 (nonvolatile memories)
FFT Characteristics
Frequency Range dc to: 1 GHz
Frequency Resolution
Minimum (max samp rate)/(512 pts): 3.9 MHz
Maximum (min samp rate)/(32,768 pts): 0.305 MHz
Maximum (min samp rate)/(32,768 pts):
0.305 mHz
Horizontal Magnify
Specify the frequency that is displayed at center
screen, and magnify the frequency-domain display
about that point.
Freq. Accuracy
[1/2 (sample frequency) (1/32768)]+
(5 x 10E-5)( signal frequency)