Agilent 53310A Data Sheet

Agilent 53310A
Modulation Domain Analyzer
Data Sheet
2
Product Description
This section offers a brief description of the Agilent Technologies 53310A Modulation Domain Analyzer. It follows the menu structure of the prod­uct and describes its operation. For specifications, see the Product Specifications and Characteristics section.
Function
Frequency
Channel: A, B, or C (Optional)
Time Interval
Measurement: +TI or ±TI Channel: A and B (Separate) or A only (Common) Start: or Stop: or
Input
Channels A and B
Voltage Threshold Range:
Manual: +10 V to –10 V, settable in 2.5 mV steps Auto: Find 50% Threshold
Coupling: ac or dc Impedance: 1 Mor 50 Hysteresis: Min to Max, settable in ten steps
Channel C (Optional)
Voltage Threshold: 0 volts Coupling: ac Impedance: 50 Hysteresis: Minimum
External Arm
Voltage Threshold: 0V, TTL (1.5 Volts), or ECL (–1.3 Volts)
Vertical/Histogram Display Range
Display measurement range is settable as Center and Span, or Minimum and Maximum. For frequency measurements, the maximum display value can be selected no larger than eight times the minimum display value.
Timebase
Main Timebase
Range:
Panorama Off: 1 µs/div to 1 s/div Panorama On: 1 to 20 times the window timebase setting
(standard); 1 to 80 times the window timebase (Option 001)
Window Timebase
Range: 1 µs/div to 1 s/div Position: Adjustable throughout the panorama
Trigger
Display Position: Left, Center, or Right Delay: Adjustable
Dual timebases allow you to capture all the information you want, as shown in the panorama, while analyzing measurement details in the window.
3
Trigger
Mode: Auto or Triggered
Triggered:
Edge Trigger
Slope: or of External
Value Trigger
Source: Frequency or Time Interval measurements, depending on measurement function
Range
Frequency: 80% of minimum display value to 120% of maximum display value Time Interval: full measurement range Slope: or HF Reject: On or Off
Display
Type: “vs Time” or Histogram
Frame type: Frame, Axes, Grid, or Off
vs Time
Acquisition Mode: Real Time or Repetitive Real Time Persistence: Single or Infinite Repetitive Averaging: On or Off
Histogram
Probability Scale (percent): Log or Linear Probability Scaling: Auto or Manual Accumulate: On or Off
Markers
Horizontal and vertical markers are available in “vs Time” and Histogram displays. Markers can be used in conjuction with automated analysis. They may also be used to delimit a portion of the meas­urement data for detailed analysis.
Histogram
Histogram Type: Histogram from “vs Time” or Fast Histogram Accumulate: On or Off Fast Histogram:
Acquire Start: Auto, or External Start on or edge Number of Measurements: 1 to 16 million per acquisition, >1014can be accumulated.
Sampling
Modes: Auto, Edge, or Time
Auto: Constant event sampling based on Timebase settings Edge:
Channel A or B: Every 1 to 256 events External: Every event
Time: 400 ns – 1 second
Utility
GPIB/Print setup
Mode: Addressed or Talk Only
Clicker: On or Off Screen Saver: On or Off Calibration and Diagnostic routines
Autoscale
Setup parameters are automatically determined to display the dynamics of the input signal. Measurement function and input conditioning should be selected prior to pressing Autoscale.
Automated Analysis
Built-in analysis functions include: Minimum, Maximum, Peak to Peak, Mean, 1/Mean, Standard Deviation, (Modulation) Rate, 1/Rate, and Probability.
Save/Recall
Ten measurement setups can be stored and recalled.
4
Product Specifications and Characteristics
Both warranted specifications and operating char­acteristics of the Agilent 53310A are discussed in this section. To distinguish warranted specifica­tions from operating characteristics, specifications
are highlighted throughout in italics.
Frequency Measurements
Range
Channel A: 10 Hz to 200 MHz Channel B: 10 Hz to 100 MHz Channel C: 50 MHz to 2.5 GHz
Maximum Measurement Rate
Fast Histogram: 1.5 MHz Other Modes: 1 MHz
Resolution
Maximum available measurement resolution or display resolution, whichever is greater
Maximum Available Measurement Resolution (Auto Sampling):
See Graph 1 for Channels A and B See Graph 2 for Option 030 Channel C
Note: Option 031 High Resolution 2.5 GHz Channel C offers superior measurement resolution. Refer to “Dynamic Frequency and Jitter Analysis at the Touch of a Button,” Agilent 53310A brochure.
Display Resolution: vs Time, or Histogram of vs Time
Window Off: Display Span/256 Window On: Display Span/224 Fast Histogram: Display Span/450
Accuracy: ±[Resolution + (Frequency x Reference Error)†]
† Refer to Graph 3.
Legend for Graphs 1 and 2
Graph 1. Maximum available frequency resolution for Channels A
and B. Larger timebase settings and averaging will reduce the effects of random noise and improve resolution.
100 Hz
10 µHz
100 µHz
1 mHz
10 mHz
100 mHz
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
Frequency Resolution (rms)
Measured Frequency (1 V rms sinewave, 100 µV rms noise)
A
B
C
D
E
G
F
Graph 2. Maximum available frequency resolution for Option 030
2.5 GHz Channel C. Larger timebase settings and averaging will reduce the effects of random noise and improve resolution.
10 MHz
100 µHz
1 mHz
10 mHz
100 mHz
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
100 MHz
10 GHz
Frequency Resolution (rms)
Measured Frequency (0 dBm sinewave, 100 µV rms noise)
1 GHz
A
B
C
D
E
F
G
50 MHz
2.5 GHz
G
F
E
D
C
B
A
Interval at Center Setting
(Interval at Center: Manual)
Timebase Setting
(Interval at Center: Auto)
20 µs/Div
200 µs/Div
2 ms/Div
20 ms/Div
200 ms/Div
1 µs
10 µs
100 µs
1 ms
10 ms
100 ms
0.5 s
5
Time Interval Measurements
+ Time Interval
Range: +20 ns to +1 second Maximum Measurement Rate:
Fast Histogram: 2.5 MHz Other Modes: 1.25 MHz
± Time Interval
Range: –0.5 s to +0.5 second Maximum Measurement Rate:
Fast Histogram: 2.0 MHz Other Modes: 1.25 MHz
Resolution
Maximum available measurement resolution or display resolution, whichever is greater
Maximum Available Measurement Resolution:
Notes
1) Threshold trigger errors are usually negligible for input slew rates >5V/µs.
2) Through averaging, maximum available measurement resolution can be significantly improved.
Display Resolution: vs Time, or Histogram of vs Time
Panorama Off: Display Span/256 Panorama On: Display Span/224 Fast Histogram: Display Span/450
Accuracy: ± Resolution ±(Time Interval x Reference Error)
††
±Start Threshold Level Timing Error
±Stop Threshold Level Timing Error
±1 ns Systematic Error
* 125 ps rms typical †† Refer to Graph 4. ††† Refer to Graph 5. ‡ Refer to Graph 6.
100 Hz
1 µHz
10 µHz
100 µHz
1 mHz
10 mHz
100 mHz
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
10 GHz
Frequency Error
C
BFA
E
D
Measured Frequency
Graph 3. Frequency Reference Error. Crystal aging will affect your frequency accuracy.
Legend for Graphs 3 and 4
F
E
D
C
B
A
Standard
Reference
Option 010 Reference
1 Month
1 Year
10 Years
Time Since Last Calibration
Graph 4. Time Interval Reference Error. Crystal aging will affect your time interval accuracy.
1 µs
Measured Time Interval
1 ps
10 ps
100 ps
1 ns
10 ns
100 ns
1 µs
10 µs
100 µs
10 µs
100 µs
1 ms 10 ms
100 ms
1 s
Time Interval Error
C
B
F
A
E
D
200
2
2
ps rms * Threshold Trigger Errors
†††
()
+
 
 
6
Time Axis in vs Time
Resolution
Main Timebase Setting/45, with panorama off Window Timebase Setting/45, with panorama on
Accuracy: Resolution
Note: Time aliasing can occur when time between meas-
urements exceeds 360 divisions.
Inputs
Channel A and B
Sensitivity (minimum hysteresis): 20 mV rms sine wave to
100 MHz (25 mV rms sine wave for Freq A from 100 MHz to 200 MHz)
Minimum Pulse Width: 5 ns at 60 mV p-p (2.5 ns at 75 mV
p-p for Frequency A measurements above 50 MHz)
Input Amplifier Noise: 600 µV rms Threshold Drift: ±3 mV after warm-up at 25°C Voltage Threshold Accuracy: ±(25 mV + 1% of Threshold
Value)
Maximum Hysteresis: Increases the minimum input signal
amplitude required by a factor of three, providing addi­tional noise immunity.
Impedance: 50 or 1 M(500 kin common) ac Coupling: 100 Hz cutoff frequency Capacitance (1 M): <20 pF (<30 pF in common) Dynamic Range (ac): 60 mV p-p to 5 V p-p Signal Operating Range (dc):
1 M: ±10 volts 50 Ω: ±5 volts
Damage Level:
1 MΩ: 40 V rms for <5 kHz, 5 V rms for >5 kHz 50 : 5 V rms
10 µV
1 ps
10 ps
100 ps
1 ns
10 ns
100 ns
1 µs
100 µV
1 mV 10 mV
100 mV
1 V
RMS Threshold Trigger Error (per edge)
Input Signal Noise (rms)
0.1 V/µ
s Input
1 V/µs Input
10 V/µ
s Input
0.1 V/ns Input
1 V/ns Input
100 mV/µs
1 ps
10 ps
100 ps
1 ns
10 ns
100 ns
1 µs
1 V/µs
10 mV/ns 100 mV/ns
1 V/ns
10 V/ns
Timing Error (per edge)
Input Signal Slew Rate
Graph 5. Noise on the input signal will add error to time-interval measurements. Error is associated with both start and stop edges.
Graph 6. Voltage Threshold Level Timing Error varies with input slew rate. Error is associated with both start and stop edges.
Channel C (Option 030)
Sensitivity: –25 dBm to 1.5 GHz, –20 dBm from >1.5 GHz to
2.0 GHz, –15 dBm from >2.0 GHz to 2.5 GHz
Maximum Input Level: +7 dBm Damage Level: +15 dBm Note: For specifications on the Option 031 High Resolution
2.5 GHz Channel C, refer to “Dynamic Frequency and Jitter Analysis at the Touch of a Button,” 53310A brochure.
External Arm
Impedance: 1 M Delay: <10 ns Note: Sensitivity, Minimum Pulse Width, Signal Operating Range, and Damage Level of the External input are the same as that of Channel B.
Frequency Reference
Standard Crystal
Temperature Stability: <8 x 10–6, referenced to 25°C Short Term Stability: <4 x 10–9for 1 second average Aging Rate: <3 x 10–7/month
Option 010 High Stability Oven Reference
Temperature Stability: <7 x 10–9, referenced to 25°C Short Term Stability: <4 x 10
–11
for 1 second average
Aging Rate: <5 x 10
–10
/day, <1 x 10–7/year
Warmup: Within 5 x 10–9of final value*, 10 minutes after
turn-on**
When:
1) 53310A is operated in a 25°C environment
2) Oscillator off-time** was less than 24 hours
3) Oscillator aging rate was <5 x 10
–10
per day prior to
turn-off**
Rear Panel Connectors
GPIB
Controls: The Agilent 53310A provides full pro­grammability. All instrument settings and operat­ing modes, except specific self-test routines, may be remotely programmed via GPIB (IEEE Std
488.1-1987). The programming codes and formats comply with IEEE Standard Codes, Formats, Protocols, and Common Commands (IEEE Std
488.2-1987), and Standard Commands for Programmable Instruments (SCPI) Standard, Version 1990.0.
* “Final value” is defined as oscillator frequency 24 hours after turn-on. ** “Turn-off”,“turn-on”, and “off-time” apply to periods when power is disconnected
from the 53310A rear panel.
Data Acquisition and Transfer Rate: A 450 point data record can be acquired and transferred to a computer at a rate ~17 times per second, as tested with an HP 9000, Series 300 controller. For this test, a 1-MHz carrier was applied to the analyzer with a timebase setting of 40 µs/division.
Data Transfer Rates: ~175 kByte/s
Interface Capabilities: SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2
Test Limit Output
The Test Limit Output will go high when measure­ments fall outside the display range of the Fast Histogram. In “vs Time” mode, it will indicate each time the Value Trigger condition is met following the pre-trigger delay.
Operating Range: Low <0.6 V and High >1.5 V into 50
Frequency Standard Input
Frequency: 5 MHz or 10 MHz Operating Range: 1 V p-p to 5 V p-p, into 1 k Damage Level: 10 V rms
Frequency Standard Output
Frequency: 10 MHz, or the External Reference if the
Frequency Standard Input is used.
Operating Range (ac coupled):
50 Ω: >1 V p-p square wave 1 M: >2 V p-p square wave
Power Requirements
Voltage: 115/230 Vac (–25% to +15%) Frequency: 48 to 66 Hz Maximum Power: 300 VA
General
Operating Temperature: 0 to 55°C Weight: 10 kg net, 18 kg shipping Dimensions: 425 mm W x 194 mm H x 363 mm D (440 mm D
with handle extended)
Warranty
The Agilent 53310A Modulation Domain Analyzer comes with a one-year return-for-service warranty.
7
8
Specifications — Option 031
Both warranted specifications and operating char­acteristics of the Agilent 53310A are discussed in this section. To distinguish warranted specifica­tions from operating characteristics, specifications
are highlighted throughout in italics.
Input Conditions
Range: 50 MHz to 2.5 GHz Sensitivity:
–13 dBm 50 MHz to 200 MHz –17 dBm 200 MHz to 2 GHz –12 dBm 2 GHz to 2.5 GHz
Maximum Input Level: +20 dBm Damage Level:+23 dBm Impedance: 50 Coupling: ac RF Burst Width: 50 µs to CW
Maximum Measurement Rate
Fast Histogram: 1.5 MHz Other Modes: 1 MHz Fast Sampling: 8 MHz (repetitive)
External Local Oscillator Input
Level: +6dBm (±1dB) Impedance: 50 Frequency Range: 150 MHz to 2.5 GHz
Resolution
Maximum available measurement resolution or display resolution, whichever is greater
Maximum Available Measurement Resolution
(Auto Sampling): see Graph A Display Resolution: vs Time or Histogram of vs Time
Window Off: Display Span/256
Window On: Display Span/224
Fast Histogram: Display Span/450
Accuracy: ±[Resolution + (Frequency x Reference Error*)]
RF Envelope Trigger
Level: Adjustable in 100 steps (e.g., –25 dBm to 0 dBm @ 1 GHz) Output: 0 to 0.4 volt into 50 or TTL level into high impedance
Maximum Available Frequency Resolution
Frequency Resolution (rms) Internal LO
1 mHz
10 mHz
100 mHz
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
100 MHz
10 GHz
1 GHz
200 MHz
2.5 GHz
B
D
E
F
G
Measured Frequency
(0 dBm sine wave, 100 µV rms noise)
C
A
1
Timebase Setting Interval @ center:
Auto
G
F
E
D
C
B
A
20 µs/Div
200 µs/Div
2 ms/Div
20 ms/Div
200 ms/Div
1 µs
10 µs
100 µs
1 ms
10 ms
100 ms
0.5 s
Interval @ Center
Setting (Interval
@ Center: Manual)
Frequency Resolution
(Ext. LO)†
7 kHz 1 kHz 20 Hz
5 Hz 2 Hz
50 mHz
5 mHz
2 mHz
1
125 ns††
Graph A. Maximum available frequency resolution for Option 031 Channel C. Larger timebase setting and averaging will reduce the effects of random noise and improve resolution. Please refer to Graph 2 on page 4 for resolution over the 50 MHz – 200 MHz band.
† Typical resolution over the 200 MHz to 2.5 GHz range using an Agilent 8663A
Synthesizer as an external LO
†† Fast Sampling mode
Specifications — Option 305
Minimum System Requirements
Agilent 53310A:
Works with any option, but Fast (repetitive) Sample Rate
is not supported. IBM-compatible PC:
386-based with coprocessor
4 Mbyte RAM
2 Mbyte hard disk space
MS-DOS®5.0
Windows®3.1
IEEE 488.2 Interface:
HP 82335A or HP 82341A National Instruments AT-GPIB or AT-GPIB/TNT with NI-488.2 version 2.1.1 drivers for Windows
Performance Characteristics
Typical Performance (1 GHz Carrier)
Off. Freq IF Noise Floor
<10 Hz 20 Hz –180 dB 100 Hz 200 Hz –170 dB 1 kHz 2 kHz –160 dB 10 kHz 20 kHz –150 dB 100 kHz 200 kHz –140 dB 500 kHz 1 MHz –130 dB
MS-DOS®and Windows®are U.S. registered trademarks of Microsoft Corporation.
9
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the produc­tion life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
“Our Promise” means your Agilent test and measurement equip­ment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifica­tions and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabili­ties, at no extra cost upon request. Many self-help tools are available.
Your Advantage
“Your Advantage” means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out­of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional services. Experienced Agilent engineers and tech­nicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
By internet, phone, or fax, get assistance with all your test and measurement needs.
Online Assistance
www.agilent.com/find/assist
Phone or Fax
United States: (tel) 1 800 452 4844
Canada: (tel) 1 877 894 4414 (fax) (905) 206 4120
Europe: (tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599
Australia: (tel) 1 800 629 485 (fax) (61 3) 9272 0749
New Zealand: (tel) 0 800 738 378 (fax) (64 4) 495 8950
Asia Pacific: (tel) (852) 3197 7777 (fax) (852) 2506 9284
Product specifications and descriptions in this document subject to change without notice.
Copyright © 1998, 2000 Agilent Technologies Printed in U.S.A. 8/00 5966-4092E
Loading...