This guide describes how to use the HP 53181A 225 MHz Frequency
Counter. The information in this guide applies to instruments having the
number prefix listed below, unless accompanied by a “Manual Updating
Changes” package indicating otherwise.
SERIAL PREFIX NUMBER:3409 AND ABOVE
HP 53181A 225 MHz Frequency
Counter
Copyright Hewlett- P ac ka rd
Company 1994, 1995, 1997,
1999
All Rights Reserved.
Reproduction, adaptation, or
translations without prior
written permission is
prohibited, except as allowed
under the copyright laws.
Printed: January 1999
Printed in USA
Manual part number
53181-90001
Certification
and Warranty
Certification
Hewlett-Packard Company
certifies that this product met
its published specification at the
time of shipment from the
factory. Hewlett-Packard
further certifies that its
calibration measurements are
traceable to the United States
National Institute of Standards
and Technol ogy (formerly
National Bureau of Standards),
to the extent allowed by the
Institute’s calibration facility,
and to the calibration facilities
of other International
Standards Organization
members.
Warranty
HP warrants HP hardware,
accessories and supplies against
defects in materials and
workmanship for a period of
three years from date of
shipment. If HP receives notice
of such defects during the
warranty period, HP will, at its
option, either repair or replace
products which prove to be
defective. Replacement products
may be either new or like-new.
HP warrants that HP software
will not fail to execute its
programming instructions, for
the period specified above, due
to defects in material and
workmanship when properly
installed and used. If HP
receives notice of such defects
during the warranty period, HP
will replace software media
which does not execute its
programming instructions due
to such defects.
For detailed warra nty
information, see back matter.
Safety Considerations
General
This product and related
documentation must be
reviewed for familiarizat ion
with this safety marki ngs and
instructions before operation.
Before Cleaning
Disconnect the product from
operating power before
cleaning.
Warning Symbols That May
Be Used In This Book
Instruction manual symbol; the
product will be marked with
this symbol when it is necessary
for the user to refer to the
instruction manual.
Indicates hazardous voltages.
Indicates earth (ground)
terminal.
or
Indicates terminal is con n ected
to chassis when such connection
is not apparent.
Indicates Alternating current.
Indicates Direct current.
Safety Considerations
(contd)
WARNING
BODILY INJUR Y OR DEATH
MAY RESULT FROM
FAILURE TO HEED A
WARNING. DO NOT
PROCEED BEYOND A
WARNING UNTIL THE
INDICATED CONDITIONS
ARE FULLY UNDERSTOOD
AND MET .
CAUTION
Damage to equipment, or
incorrect measurement data,
may result from failure to
heed a caution. Do not
proceed beyond a CAUTION
until the indicated conditions
are fully understood and met.
Safety Earth Ground
An uninterruptible safety earth
ground must be maintained
from the mains power source to
the product’s ground circuitry.
WARNING
WHEN MEASURING POWER
LINE SIGNALS, BE
EXTREMELY CAREFUL AND
ALWAYS USE A
STEP-DOWN ISOLATION
TRANSFORMER WHICH
OUTPUT IS COMPATIBLE
WITH THE INPUT
MEASUREMENT
CAPABILITIES OF THIS
PRODUCT. THIS PRODUCT’S
FRONT AND REAR PANELS
ARE TYPCIALLY AT EARTH
GROUND.
TO MEASURE AC POWER
LINE SIGNALS WITHOUT AN
ISOLATION TRANSFORMER.
For additional safety and
acoustic noise information, see
back matter.
THUS, NEVER TRY
Hewlett-Packard Company7.NC.NL.A.11.03.97.R1.P.CW6FC
Santa Clara Division
5301 Stevens Creek Boulevard
Santa Clara, California 95052-8059
Contents
In This Guide
Contents and Organizationxii
Related Documentsxiii
Types of Service Available if Your Instrument Failsxiv
Standard Repair Services (Worldwide)xiv
Express Repair/Performance Calibration Service
(USA Only)xiv
Repair Instrument Yourselfxiv
Repackaging for Shipmentxv
Description of the 225 MHz Frequency Counterxvi
Optionsxvii
The Front Panel at a Glance1-2
The Front Panel Indicators at a Glance1-3
The Front Panel Indicators at a Glance (Cont.) 1-4
The Front Panel Menus at a Glance1-5
The Front Panel Menus at a Glance (Cont.)1-6
The Front Panel Menus at a Glance (Cont.)1-7
The Display Annunciators at a Glance1-8
The Display Special Characters at a Glance1-9
The Underscore Placeholder1-9
The Asterisk Placeholder1-9
Operating Guideiii
Contents
The Limit Test Graph Characters at a Glance1-10
The Rear Panel at a Glance1-11
Making Measurements 1-12
To Measure Frequency1-14
To Select Input Coupling and Impedance1-15
Selecting the Type of Statistics (Stats)1-29
Computing Stats on Filtered Data Only1-31
Displaying Stats After Filtering Data of Input
Signal1-33
Disabling Stats and Math1-34
To Control Measurement1-35
ivOperating Guide
Contents
2Operating Your Frequency Counter
Introduction2-2
Chapter Summary2-2
Where to Find Some Key Working Examples2-3
How this Counter Wor k s for You2-4
Summary of the Measurement Sequence2-5
Using the Measurement Control Keys (Run and Stop/
Single)2-6
Overview of the Measurement Control Keys2-6
To Use the Measurement Control Keys2-7
Using Entry/Select (Arrow) Keys2-9
To Use During Numeric Entry2-9
To Use When Sequencing Through the Other Meas and
Recall Menus2-9
To Use During State Changing (ON/OFF, LO/MED/
HI, etc.)2-10
To Use on Prompted Help Messages (MATH HELP: NO/
YES, PRT HELP: NO/YES)2-10
Using the MEASURE Menu Keys2-11
Overview of the MEASURE Menus2-11
To Measure Frequency2-12
To Measure Frequency Ratio (Optional)2-13
To Measure Period2-13
To Measure Positive/Negative Volt Peaks2-14
Using the Gate & External Arm Menu Key2-15
Overview of Gate/External Arming Functions2-15
TIME Arming2-15
AUTO Arming2-15
EXTERNAL Arming2-15
DIGITS Arming2-16
To Use the Gate and External Arm2-17
Example Procedure for Gate and External Arm2-17
Operating Guidev
Contents
Example Procedure for Changing the Number of
Digits of Resolution Displayed for More
Precise Measurements2-19
Using the MATH Menu Keys2-21
Overview of Scale/Offset Math Menu2-21
To Use the Scale/Offset Math Menu2-22
Example Procedure for Scale Function2-22
Example Procedure for Offset Function2-23
Example Procedure for Turning Off Math Mode2-25
Example Procedure for Set Reference Function—Offset
From the Last Measurement Value (Seeing Digits Not
Visible on the 12-Digit Display)2-25
Overview of Statistics (Stats) Menu2-26
To Use the Stats Menu for Automatic and Continuous
Statistical Analysis2-27
Example Procedure for Computing Stats2-27
Example Procedure for Easy Viewing of Stats2-28
Example Procedure for Filtering Data (Using Limits)
During Stats2-29
Example Procedure for Configuring SINGLE to Initiate
N Measurements2-30
Example Procedure for Turning Off Stats Mode2-30
Using the LIMITS Menu Keys2-31
Overview of Limits Menus2-31
To Set and Use Automatic Limit Testing2-32
Limits Testing Example 1—Flag and Stop Measuring
On L imits2-32
Limits Testing Example 2—Flag On Limits but
Continue Measuring2-34
Limits Testing Example 3—Use Analog Graph Display
While Adjusting/Aligning Input Signal2-34
Limits Testing Example 4—Selecting Filtering
Conditions of Stats Computation2-36
viOperating Guide
Contents
Limits Testing Example 5—Sending the Limit-Detect
Output to the RS-232 Serial Port2-38
Using the DISPLAY Digits Keys2-39
Overview of the DISPLAY Digits Keys2-39
To Use the DISPLAY Digits Keys2-40
Using CHANNEL 1 Input Conditioning Keys2-41
Overview of Trigger/Sensitivity Menu2-41
To Use the Trigger/Sensitivity Key to Adjust Counter’s
Trigger Voltage and Sensitivity Levels2-44
Overview of Input Conditioning Toggle Keys2-47
Using the Save and Recall Menus2-48
Overview of Save and Recall Functions2-48
To Use the Save Function2-49
To Use the Recall Function2-50
To Unsave a Measurement Setup2-51
Using the Print Menu2-52
Overview of the Print Menu2-52
To Use the Print Menu2-52
Using the Utility Menu2-53
Overview of the Utility Menu2-53
To Set the HP-IB Address2-54
Measurement Specifications (Continued)3-12
Measurement Definitions3-15
Measurement Arming and Processing3-16
General Information3-17
Operating Guideix
Contents
xOperating Guide
In This Guide
This book is the operating guide for the HP 53181A 225 MHz Frequency
Counter. It consists of a table of contents, this preface, a quick reference
guide, three chapters, and an index.
This preface contains the following information:
•Contents and Organization
•Related Documents
•Types of Service Available if Your Instrument Fails
•Repackaging for Shipment
•Description of the 225 MHz Frequency Counter
•Options
•Accessories Supplied and Available
page xii
page xiii
page xiv
page xv
page xvi
page xvii
page xviii
Operating Guidexi
In This Guide
Contents and Organization
The Table of Contents lists the beginning of each chapter in the guide,
helping you locate information.
The Preface introduces you to the operating guide, the Frequency
Counter, and provides general information about the Counter.
The Quick Reference Guide consists of a Menu Tree (tear-out sheet)
that serves as a tool to trigger your memory or get you quickly
reacquainted with the instrument, and Menu Roadmaps that illustrate
how to navigate through the menus. It is located after this preface.
Chapter 1, “Getting Started,” is a quick start guide that gives you a brief
overview of the Counter’s keys, indicators, menus, display, and
connectors. Last, a graphical procedure for performing a complete
measurement is provided.
Chapter 2, “Operating Your Frequency Counter,” is an operator’s
reference. You are given an overview of each group of front-panel keys,
operating functions, and menus followed by a series of exercises that guide
you through the operation of the Counter.
Chapter 3, “Specifications,” lists the specifications and characteristics of
the Counter.
The Index helps you find items easily by arranging information in a
sequence different from that found in the Table of Contents, disclosing
relationships among items.
xiiOperating Guide
In This Guide
Related Documents
For more information on counters refer to the following Series 200
Application Notes:
•Fundamentals of Electronic Frequency Counters Application
Note 200—HP part number 02-5952-7506.
•Understanding Frequency Counter Specifications Application
Note 200-4—HP part number 02-5952-7522.
Operating Guidexiii
In This Guide
Types of Service Available if Your Instrument
Fails
If your Counter fails within three years of original purchase, HP will
repair it free of charge. If your instrument fails after your 3-year warranty
expires HP will repair it, or you can repair it yourself by ordering the
service guide.
There are three types of repair services:
•Standard repair service—if downtime is not critical.
•Express Repair/Performance Calibration Service—if downtime is
critical.
•Use the Assembly-Level Service Guide and repair the unit yourself.
Standard Repair Services (Worldwide)
Contact your nearest HP Service Center. They will arrange to have your
Counter repaired.
Express Repair/Performance Calibration Service
(USA Only)
If downtime is critical, you can receive your repaired Counter via
overnight shipment. Just call 1-800-403-0801 and ask for Express Repair/Performance Calibration Service. When your Counter is repaired,
it will be returned via overnight shipment.
Repair Instrument Yourself
If your Counter 3-year warranty has expired and you choose to repair the
instrument yourself or would like more details on self test and calibration,
use the procedures in the Assembly-Level Service Guide.
xivOperating Guide
In This Guide
Repackaging for Shipment
For the Express Repair/Performance Calibration Service described above,
return your failed Counter to the designated HP Service Center, using the
shipping carton of the instrument. HP will notify you when your failed
instrument has been received.
If the instrument is to be shipped to HP for service or repair, be sure you
do the following:
•Attach a tag to the instrument identifying the owner and indicating
the required service or repair. Include the instrument model number
and full serial number.
•Place the instrument in its original container (if available) with
appropriate packaging material.
•Secure the container with strong tape or metal bands.
If the original shipping container is not available, place your unit in a
container which will ensure at least 4 inches of compressible packaging
material around all sides of the unit. Use static free packaging materials
to avoid additional damage to your unit.
HP suggests that you always insure shipments.
Operating Guidexv
In This Guide
Description of the 225 MHz Frequency
Counter
The HP 53181A is a frequency counter capable of measuring frequencies
to 225 MHz on Channel 1. Depending on which optional Channel 2 the
counter contains, this capability is extended to 1.5 GHz (Option 015),
3.0 GHz (Option 030), 5.0 GHz (Option 050) or 12.5 GHz (Option 124).
The HP 53181A has a frequency resolution of 10 digits in one second.
The HP 53181A provides users with a HP-IB measuring speed of up to
200 measurements per second, and is suitable for bench-top and
ATE operation.
The HP 53181A basic measurement functions include Frequency, Period,
Peak Voltage, and Frequency Ratio (if Channel 2 is installed).
The HP 53181A includes additional measurement functions and features
that are designed specifically for manufacturing and service applications:
•1, 5, 10 MHz external reference capability—to match customer’s house
standard,
•optional ultra high, high, or medium stability oven oscillators for high
accuracy needs and lengthened calibration cycles,
•external gating,
•full math and statistics,
•automatic limit testing,
•analog display mode limit testing, and
•SCPI programming capability.
Programmable control is performed via an HP-IB. The HP-IB and a
talk-only RS-232C serial port are standard for the HP 53181A. The serial
port is for printing measured and analyzed data on serial printers, or for
outputting an out-of-limit signal.
xviOperating Guide
In This Guide
Options
The options available for the HP 53181A 225 MHz Frequency Counter are
listed following this paragraph. Specifications for the options are listed in
Chapter 3, “Specifications.” If you’ve purchased an option with the initial
order, it will be installed at the factory and ready for operation at delivery.
Refer to the “Retrofitting Options” chapter in the Assembly-Level Service
Guide for instructions on field installation of the options.
NOTE
The “0s” and “1s” in the following option numbers are numeric characters
(that is, they are not letters).
Channel 1 will have both front and rear terminals. Option 015 and Option 124 Channel 2 will have a front
terminal only.
Operating Guidexvii
In This Guide
Accessories Supplied and Available
Accessories Supplied
•Power cord, 2.3 meters
Accessories Available
•HP 34161A Accessory Pouch
•HP 34131A Transit Case
•Printer RS-232 Interface cables, HP 24542G or HP 24542H
•HP-IB cables, HP 10833A/B/C/D
xviiiOperating Guide
HP 53181A Quick Reference Guide
The Quick Reference Guide is designed for experienced users of the
HP 53181A Frequency Counter. It is intended to be used as a tool to
trigger your memory. If you are using the HP 53181A for the first time,
HP recommends that you at least read Chapter 1, “Getting Started,”
in the Operating Guide first.
The Quick Reference Guide follows this page, and consists of the following
items:
•A Menu Tree which may be torn out of the guide for external use
(pages 1 an d 2).
•Menu Roadmaps which illustrate via key-press sequences how to
navigate through the menus under the menu keys (pages 3
through 10). Key-press sequences are provided for the following
menu keys:
1 Measurement function menu keys
2 Limits menu keys
3 Math menu keys
4 Sign (+ or −) selection toggle ke y
5 Data Entry/Select (or arrow) keys
6 Enter numeric data (terminate) key.
MEASURELIMITSMATHDISPLAYCHANNEL 1
Freq
Ch 1
Other
Meas
Recall
Freq
Ch 2
Gate &
ExtArm
LocalUtility
Save &
Print
9
Also tranverses a menu and ex i ts at the
end of a menu.
23
3 GHz
Uppr &
Lower
Limit
Modes
Run
10
11
Scale &
Offset
Stats
Stop/
Single
4
MHz
µs
Limit
Gate
More Digits
Fewer Digits
12
8 Utility menu key (Hold during power-up to
access Utility functions.)
9 Recall, Save and Print menu keys
10 Gate and External Arm menu key
11 Measurement control keys
12 Display Digits keys
13 Channel 1 Trigger/Sensitivity menu key and
input conditioning keys
+/–
5
67
Enter
Trigger
Sensitivity
50Ω
1MΩ
DC
AC
X10
Attenuate
CHANNEL 2
100 MHz to MAX
Damage Lvl:
5V rms MAX.50Ω
0.1 Hz − 225 MHz
Damage Lvl:
5V rms MAX.50Ω
100kHz
Filter
13
!
!
7 Channel 2 optional RF input
NOTE
It is normal operation for the fan in the Counter to continue to run after
the Counter is placed in Standby mode. Power to the timebase is
continuous to maintain long term measurement reliability, and the fan
helps maintain timebase temperature stability.
1-2Operating Guide
Chapter 1 Getting Started
The Front Panel Indicators at a Glance
The Front Panel Indicators at a Glance
There are eight different gro ups of indicato rs or LEDs. They are listed and
described in the following table.
IndicatorsDescription of the Indicators
When the indicator for the Freq Ch 1 or Freq Ch 2
Freq
Ch 1
Other
Meas
Scale &
Offset
Stats
+/–
Freq
Ch 2
Limit
Modes
Local
Save &
Print
Trigger
Sensitivity
Enter
keys is lit, it indicates that the Frequency function
is enabled for the corresponding input channel.
When the indicator for the Other Meas key is li t , i t
simultaneously indicates that the key’s menu and
its menu item (for example, PERIOD 1) is enabled.
When these indicators are lit, the key’s “enable”
menu item (that is, Limit Modes/LIM TEST,
Scale & Offset/MATH, Stats/STATS, and Save &
Print/PRINT) is enabled.
When this indicator is lit, it indicates that
auto-triggering is enabled.
When this indicator flashes, it indicates that the
arrow keys can be used to modify or enter data.
The Enter key is used to enter numeric values or
parameters. It can also be used to traverse through
a menu. Pressing the Enter key at the end or
bottom of a menu will cause the Counter to exit the
menu and return to the measurement mode.
1
When one of these indicators is lit, it indicates that
Run
Stop/
Single
the Run or Single function is enabled.
Operating Guide1-3
Chapter 1 Getting Started
The Front Panel Indicators at a Glance (Cont.)
The Front Panel Indicators at a Glance (Cont.)
1
IndicatorsDescription of the Indicators
When this indicator flashes, it indicates that the
Counter is triggering on the input signal. If the
input signal is too high, this indicator remains ON.
If the input signal is too low, this indicator is OFF.
When one of these indicators is lit, it indicates that
50Ω
1MΩ
X10
Attenuate
Remote
Error (Flashing)
SRQ
DC
AC
100kHz
Filter
the adjacent choice (that is, 50Ω, DC, X10, or
100kHz Filter) is enabled or active. Note that when
these indicators are not lit, then the other choice
(that is, 1MΩ, AC, X1, or no filter) is active.
A lit Remote indicator indicates that the Counter is
in remote mode (Note: In the remote mode, the
Save & Print key becomes the Local key.)
If (while in remote) an error occurs, the Remote
indicator will flash. The indi cator will continue
flashing until the controller has read or cleared the
error queue, or until the front panel returns to local
mode.
An unlit Remote indicator indicates that the
Counter is in local mode.
The SRQ indicator indicates that the Counter has
requested service from th e controller. The SRQ
indicator will remain lit until the controller has
recognized the service request and serial poll ed the
Counter, or taken specific action to cancel the
request (for example, *CLS command).
1-4Operating Guide
Chapter 1 Getting Started
The Front Panel Menus at a Glance
The Front Panel Menus at a Glance
Freq
Ch 1
Other
Meas
Gate &
ExtArm
1
FREQUENCY 1
Freq
Ch 2
FREQUENCY 2
1
Uppr &
Lower
UPPR: 0.000000
LOWR: 0.000000
PERIOD 1
1
RATIO 1 TO 2
1
RATIO 2 TO 1
VOLT PEAKS 1
Limit
Modes
LIM TEST: OFF
LIM TEST: ON
ON FAIL:GO ON
ON FAIL: STOP
SHOW: NUMBER
SHOW: GRAPH
2
GATE: TIME
TIME: .100 s
Scale &
Offset
GATE: AUTO
SCALE:1.000000
OFFS: 0.000000
SET REF: NO
SET REF: YES
SET REF: HELP
MATH: OFF
MATH: ON
MATH HELP: NO
MATH HELP: YES
GATE: EXTERNL
START: POS
START: NEG
STOP: AUTO
STOP: NEG
STOP: POS
STOP: TIME
TIME: .100 s
Stats
SHOW: MEAS
SHOW: STD DEV
SHOW: MEAN
SHOW: MAX
SHOW: MIN
N: 100
STATS: OFF
STATS: ON
USE: ALL MEAS
USE: IN LIMIT
ON SINGLE: 1
ON SINGLE: N
GATE: DIGITS
DIGITS: 4
1
These menu items appear only if your Counter contains the optional 1.5 GHz, 3.0 GHz,
or 5.0 GHz RF Input (Channel 2).
2
Refer to Chapter 2 in this guide for details on the Gate & ExtArm menu.
Only registers which can be recalled will appear in this menu.
5
This menu item only appears if an instrument setup has been saved.
Save &
Print
SAVE:
5
UNSAVE:
PRINT: OFF
PRINT: ON
PRT HELP: NO
PRT HELP: YES
DC
AC
CH 1: DC
CH 1: AC
Trigger
Sensitivity
AUTO TRG: ON
LEVEL: 50 PCT
SLOPE: POS
SLOPE: NEG
SENSTVTY: HI
SENSTVTY: LO
SENSTVTY:MED
X10
Attenuate
CH 1: X10 ATT
CH 1: X1 ATT
AUTO TRG: OFF
LEVEL: 0.000V
100kHz
Filter
CH 1: LP FILT
CH 1: NO FILT
1-6Operating Guide
Chapter 1 Getting Started
The Front Panel Menus at a Glance (Cont.)
The Front Panel Menus at a Glance (Cont.)
NOTE
Turn power off, press and hold Recall (Utility) key, then press
POWER key to access this menu.
POWER
On / Stby
REV: X X X X
HP-IB: 3
TIMEBAS: AUTO
TIMEBAS: INT
TIMEBAS: EXT
CAL HELP: NO
CAL HELP: YES
TEST LOOP: OFF
TEST LOOP: ON
TST PRINT:OFF
TST PRINT: ON
Utility
Recall
TEST: NO
6
TEST: DISP
TEST: CPU
TEST: ROM
TEST: RAM
TEST: EEPROM
TEST: HP-IB
TEST: QSPI
TEST: FPGA
TEST: FR END
TEST: MEAS
TEST: INTERP
TEST: KEYPAD
TEST: PRINT
BAUD: 9600
BAUD: 19200
BAUD: 300
BAUD: 1200
BAUD: 2400
PARITY: OFF
PARITY: EVEN
PARITY: ODD
SW PACE: XON
SW PACE: NONE
DTR: HIGH
DTR: LIMIT
DTR: HW PACE
SHOW 9 AS: 9.0
SHOW 9 AS: 9,0
1
6
These menu items appear only if TEST LOOP is OFF.
Operating Guide1-7
Chapter 1 Getting Started
The Display Annunciators at a Glance
The Display Annunciators at a Glance
1
MHz
µs
Gate
Period FreqCh 1 Ch 2Limit ExtRef
AnnunciatorIndication
PeriodCounter is set to measure Period.
Freq Counter is set to measure Frequency.
Ch 1Counter’s channel 1 is selected to measure an input signal.
Ch 2Counter’s channel 2 is selected to measure an input signal.
LimitCounter is limit testing and the current measurement
exceeds the user-entered limits.
ExtRefCounter is set to use the signal connected at rear panel
Ref In connector as the timebase (TIMEBAS: EXT);
or Counter is set to automatically (TIMEBAS: AUTO) select
the timebase and ha s chosen the signal con nected at th e rear
panel Ref In connector.
HzThe displayed data is in units of Hertz.
MThe prefix for the units of the displayed data is mega (10
µThe prefix for the units of the displayed data is micro (10
sThe displayed data is in units of seconds.
GateThe gate is open. Before a measurement cycle starts, this
annunciator is OFF, indicating the gate is closed. During a
measurement cycle, the annunciator is ON, indicating the
gate is open.
6
).
−
6
).
1-8Operating Guide
Chapter 1 Getting Started
The Display Special Characters at a Glance
The Display Special Characters at a Glance
The Underscore Placeholder
FreqCh 2
Special CharacterDescription
A placeholder that indicates the digit has been
blanked using the Display Digits keys.
The Asterisk Placeholder
FreqCh 1
1
MHz
Hz
Special CharacterDescription
A placeholder that indicates this digit is not
significant because of the measurement resolution.
(The appearance of this asterisk is determined by the
Counter; thus, these cannot be removed by the
Display Digits keys.)
Operating Guide1-9
Chapter 1 Getting Started
The Limit Te st Graph Characters at a Glance
The Limit Test Graph Characters at a Glance
1
FreqCh 1
Special CharacterDescription
The colons represent the lower and upper limits.
The asterisk represents the last measurement.
1-10Operating Guide
Chapter 1 Getting Started
The Rear Panel at a Glance
The Rear Panel at a Glance
12
1
3
WARNING:
1
I
N
P
U
T
S
WARNING:
NO OPERATOR SERVICEABLE PARTS INSIDE, REFER SERVICING TO SERVICE TRAINED PERSONNEL.
!
60 VA
10 MHz Out
AC LINE:
50/60 Hz
50/60/400 Hz
100 - 120 VAC
200 - 240 VAC
2
3
Ext
Arm
4
Ref
In
FOR CONTINUED FIRE PROTECTION, USE SPECIFIED ~ LINE FUSE.
5689
1 Rear-panel input connectors
(optional)
2 Power module (Senses incomin g
voltage and automa tically
selects proper setup.)
3 Fan
4 External Arm input connector
5 External Reference Input connector
ISM 1-A
FOR LABORATORY USE BY
QUALIFIED PERSONNEL
FOUR USAGE EN LABORATOIRE
PAR PERSONNEL QUALIFIE
8 Oscillator Adjust potentiometer
9 RS-232 Interface or Limit Output
connector
NOTE
It is normal operation for the fan in the Counter to continue to run after
the Counter is placed in Standby mode. Power to the timebase is
continuous to maintain long term measurement reliability, and the fan
helps maintain timebase temperature stability.
Operating Guide1-11
Chapter 1 Getting Started
Making Measurements
Making Measurements
1
One of the first things you will want to do with your Counter is to become
acquainted with its front panel. Therefore, we have written the
procedures in this section to familiarize you with some of its controls.
The following procedures are provided:
•First you are shown how to turn on the Counter and measure the
frequency of a signal applied to the Counter’s input channels.
•Second, you are shown how to use the input coupling, impedance, and
trigger/sensitivity keys to set the input conditions of Channel 1 to
match the signal being measured.
•Third, you are shown how to scale and offset the measurement result.
•Fourth, you are shown how to set upper and lower limits for
measurements.
•Fifth, you are shown how to enable the Counter to compute statistics
(such as standard deviation) and display statistics of measurements.
•Last, you are shown how to use the Run and Stop/Single keys to
control measurements.
The order of the procedures in this chapter is the recommended order for
making measurements with this Counter.
1-12Operating Guide
Chapter 1 Getting Started
Making Measurements
Study and refer to the following legend, as needed, to understand the
meaning of the icons which are used throughout this chapter.
12
34
1 Press key one time
and release
2 Press key two times
and release
3 Repeated key presses
4 Press and hold
5 Result
6 Auto operation
7 Connect signal
8 Disconnect signal
Legend
5
6
1
78
91110
9 Indicator off
10 Indicator on
11 Indicator flashing
Operating Guide1-13
Chapter 1 Getting Started
Making Measurements
To Measure Frequency
1
POWER
On / Stby
Period FreqCh 1 Ch 2Limit ExtRef
FreqCh 1
FreqCh 1
MHz
µs
Gate
Connect (for demonstration purposes) the Counter’s rear-panel
10 MHz Out signal to CHANNEL 1 input as shown in the illustrated
procedure, below.
CHANNEL 1
MHz
Gate
Damage Lvl:
5V rms MAX.50Ω
!
FreqCh 1
The Counter will automatically display the measured frequency of the
input signal.
1-14Operating Guide
Chapter 1 Getting Started
CHANNEL 1
Trigger
Sensitivity
DC
AC
50Ω
1MΩ
X10
Attenuate
100kHz
Filter
Damage Lvl:
5V rms MAX. 50Ω
!
Making Measurements
If you need or want to change CHANNEL 1’s coupling, impedance, and
triggering conditions to match the input signal you are trying to measure,
the next procedures “To Select Input Coupling and Impedance” and
“To Set Input Channel Trigger Level/Sensitivity” demonstrate this.
Perform these procedures whether or not you want to customize the
Counter’s input conditions to measure your signal; doing this will help you
become familiar with the DC/AC, 50Ω/1MΩ, and Trigger/Sensitivity
keys.
To Select Input Coupling and Impedance
Selecting Input Coupling
DC
AC
Channel 1’s input coupling is now set to dc.
1
NOTE
If you want to change the coupling back to the default ac coupling, perform
the following step.
DC
AC
Selecting Input Impedance
50Ω
1MΩ
Channel 1’s input impedance is now set to 50Ω.
The “arrow” keys can also be used to toggle the state of toggle keys
(DC/AC, 50Ω/1MΩ, etc.) as indicated by the flashing indicator within the
arrow keys. However, for simplicity in this procedure, use the
corresponding toggle key to change states.
Operating Guide1-15
Chapter 1 Getting Started
Enter
+/–
Making Measurements
If you want to change the input impedance back to the default
1MΩimpedance, perform the following step.
1
50Ω
1MΩ
FreqCh 1
MHz
Gate
To Set Input Channel Trigger Level/Sensitivity
Changing Trigger Mode
Trigger
Sensitivity
Press any one of these arrow keys to toggle to the next state of Auto
Trigger.
1-16Operating Guide
Chapter 1 Getting Started
Making Measurements
Modifying Input Trigger Level
Trigger
Sensitivity
The leftmost “0” digit in the LEVEL display is highlighted, indicating that
if you press the d key once the displayed value will increase to 1.000 volt
as shown in the following step.
Enter
Channel 1’s trigger level is now set to +1V.
To set the trigger level to −2V, perform the following steps.
1
+/–
Enter
Channel 1’s trigger level is now set to −2V.
Operating Guide1-17
Chapter 1 Getting Started
Making Measurements
Selecting Input Trigger Slope
1
Trigger
Sensitivity
Display 1, 7/13/92
Display 1, 7/13/92
Refer to the sub-section titled “To Use the Trigger/Sensitivity Key to
Adjust Counter’s Trigger Voltage and Sensitivity Levels” in Chapter 2 for
more information about trigger slope, if needed.
Selecting Input Sensitivity
Trigger
Sensitivity
Display 1, 7/13/92
Display 1, 7/13/92
Display 1, 7/13/92
Refer to the sub-section titled “To Use the Trigger/Sensitivity Key to
Adjust Counter’s Trigger Voltage and Sensitivity Levels” in Chapter 2 for
more information about sensitivity, if needed.
1-18Operating Guide
Chapter 1 Getting Started
Stats
Scale &
Offset
MATH
Making Measurements
Starting the Measurement
Run
FreqCh 1
MHz
Gate
The Run key initiates repetitive measurements, and is described in the
section titled “To Control Measurement” at the end of this chapter.
To Select Scale and Offset
The Scale & Offset key allows for multiplication and addition,
respectively, of the measurement by user-specified cons tants. Modification
of the displayed mea su rement by th ese Ma th op erat io ns is re pres ent ed by
the following equation:
The Scale and Offset Math operations can be used, for example, to
subtract systematic errors or display the percentage difference between
signals.
Entering the Scale Value
Scale &
Offset
1
To demonstrate the Scale Math operation, set Scale to 10 as shown in the
following steps.
Operating Guide1-19
Chapter 1 Getting Started
Making Measurements
Press and hold the d key until the value of Scale is 10 as shown in the
following step.
1
Enter
NOTE
BE SURE to press the Enter key to enter the value of 10.
The Scale is now set to 10, and MATH has been enabled.
The Scale & Offset indicator is now lit to show that MATH is enabled.
Since MATH is enabled, the results are being scaled and offset.
Entering the Offset Value
To demonstrate the Offset Math operation, set the Offset to 1 MHz as
shown in the following steps.
Scale &
Offset
At this point, pressing the s key will cause the Counter to display the full
display of the Offset value as shown in the following step.
1-20Operating Guide
Chapter 1 Getting Started
Making Measurements
Press the s key six more times to cause the Counter to display your entry
in Mega units as shown in the following step.
NOTE
M
The leftmost “0” digit in the OFFSet display is highlighted, indicating that
if you press d key once the displayed value will increase to 1 Mega
(that is, 1E6) as shown in the followi ng step.
M
Enter
M
BE SURE to press the Enter key to enter the 1 Mega value.
The Offset is now set to 1 Mega.
Displaying the Math Results
1
Run
FreqCh 1
M
The Counter displays th e modi fi ed measu rement res ults , whi ch are bas ed
on the scale and offset values that you selected in the previous steps.
(For more details and real applications of the Ma th Scale and Offset
operations, refer to the appropriate section in Chapter2, “Operating Your
Frequency Counter.”)
Operating Guide1-21
Chapter 1 Getting Started
Making Measurements
Disabling Math
1
Scale &
Offset
Display 1, 7/13/92
Display 1, 7/13/92
Note that the Scale & Offset key indicator is now off.
MHz
Gate
NOTE
Run
FreqCh 1
DO NOT cycle POWER because you will need to use these Scale and
Offset values in the follow ing procedure “To Set Limits of Measu rements.”
Continue to the following procedure.
1-22Operating Guide
Chapter 1 Getting Started
Making Measurements
To Set Limits of Measurements
To demonstrate how Math and Limits work together, use the Scale (10)
and Offset (1 Mega) values selected in the previous procedure “To Select
Scale and Offset.” Enable Math by performing the following steps.
Scale &
Offset
Display 1, 7/13/92
Display 1, 7/13/92
1
Run
FreqCh 1
M
The result of this Math operation is a measurement of 101 MHz.
(Measurement × Scale) + Offset = Result
(10 MHz × 10) + 1 Mega = 101 Mega
Now, set the upper limit to 102 Mega and the lower limit to100 Mega by
performing the following procedures. (Figure 1-1 and Figure 1-2 illustrate
the limits settings.)
Operating Guide1-23
Chapter 1 Getting Started
Limit
Modes
Uppr &
Lower
LIMITS
Making Measurements
Setting the Upper Limit
1
Uppr &
Lower
Press the s key nine times to cause the Counter to display your entry in
hundreds of Mega units as shown in the following step.
M
The leftmost “0” digit in the UPPR display is highlighted as shown above,
indicating that each press of the d key will increase the displayed value.
M
M
M
NOTE
Enter
M
BE SURE to press the Enter key to enter the 102 Mega value.
1-24Operating Guide
Chapter 1 Getting Started
Making Measurements
1
2
1 102 Mega Upper Limit
2 101 Mega Scale/Of fset Measurement
Figure 1-1. 102 Mega Upper Limit Setting
Setting the Lower Limit
Uppr &
Lower
1
NOTE
Press the arrow keys as shown in th e foll owing steps to set th e lower limi t
value.
Press the s key nine times to cause the Counter to display your entry in
hundreds of Mega units as shown in the following step.
M
M
Enter
M
BE SURE to press the Enter key to enter the 100 Mega value.
Operating Guide1-25
Chapter 1 Getting Started
Making Measurements
Limits should now be set as shown in Figure 1-2.
1
1 102 Mega Upper Limit
2 101 Mega Scale/Of fset Measuremen t
3 100 Mega Lower Limit
1
2
3
Figure 1-2. 100 Mega Lower and 102 Mega Upper Limits Settings
Figure 1-3 represents what transpired during this Math and Limits
procedure.
1
234
1 Raw Measurements
2 Math
3 Limits
4 Measurements (Scale/Offset Results)
within Limits
Figure 1-3. Math and Limits Results
1-26Operating Guide
Chapter 1 Getting Started
Making Measurements
Setting the Counter to Flag and Stop Measuring On
Out-of-Limit Measurements
If you want the Counter to stop measuring when the signal exceeds the
limits (102 to 100 Mega) that you entered in the previous procedure,
perform the following steps to select the STOP choice in the ON FAIL
display. (Note that ON FAIL: GO ON is the default state after power-up.)
Limit
Modes
Run
FreqCh 1
M
Gate
The current modified measurement of the input signal applied to
CHANNEL 1 is displayed.
Since the Counter is now in the stop-on-fail mode, the Limit annunciator
in the display will light and the Counter will stop making measurements
when a measurement exceeds the limits you se t. The mea sur ement result
that exceeded the entered limits will be held on the display for
observation.
1
Operating Guide1-27
Chapter 1 Getting Started
Making Measurements
Setting the Counter to Flag On Limits But Continue
Measuring
Perform the following steps to select the GO ON choice in the ON FAIL
1
display if you want the Counter to continue measuring even though an
measurement result exceeds the limits previously entered.
Limit
Modes
Run
FreqCh 1
M
Gate
The current modified measurement of the input signal applied to
CHANNEL 1 is displayed.
Since the Counter is now in the go-on-fail mode, the Limit annunciator in
the display will light each time a measurement exceeds the limits you set.
However, the Counter will continue to make measurements.
Disabling Limit Testing
Limit
Modes
The Counter is now making measurements without limit testing.
1-28Operating Guide
Chapter 1 Getting Started
Stats
Scale &
Offset
MATH
Making Measurements
Disabling Math
Scale &
Offset
Display 1, 7/13/92
Display 1, 7/13/92
Run
FreqCh 1
MHz
Gate
The Counter is now making measurements without the scale/offset values
calculated into the measurements.
To Perform Statistics on Measurements
Selecting the Type of Statistics (Stats)
Suppose you want the Counter to compute and display the standard
deviation of the current input data (which is the 10 MHz signal applied
to CHANNEL 1). Also, you want the Counter to make 20 measurements
before it computes the standard deviation. Perform the following steps.
1
Stats
Operating Guide1-29
Chapter 1 Getting Started
Making Measurements
Updating the SHOW configuration caused Stats to be enabled. The Stats
indicator is now lit.
1
NOTE
BE SURE to press the Enter key to enter the value of 20.
Stats
Enter
1-30Operating Guide
Chapter 1 Getting Started
Making Measurements
The Counter is now set to make statistics based on 20 measurements.
Run
FreqCh 1
FreqCh 1
FreqCh 1
Hz
Gate
Hz
Gate
In this case, the displayed standard deviation value is computed on all
measurements of the 10 MHz signal since no limits were set.
Computing Stats on Filtered Data Only
A special feature of the Counter allows you to use the upper and lower
limits to filter data before statistical processing or computation as shown
in Figure 1-4.
1
1
1 Raw Measurements
2 Limits
2
3 Filtered data (USE: IN LIMIT)
4 Statistics
3
4
Figure 1-4. Filtering Data Before Statistical Computation
Operating Guide1-31
Chapter 1 Getting Started
Making Measurements
Perform the following steps to select the IN LIMIT choice in the USE
display if you want the Counter to compute statistics on only frequency
measurements within the limits you set.
1
Stats
Display 1, 7/13/92
Display 1, 7/13/92
Since the Limits were set to 100 Mega and 102 Mega values that are
based on a scale of 10 and offset of 1 Mega, you must re-enable Math now
to get the measurements to be within the limits. Perform the following
steps.
Scale &
Offset
Display 1, 7/13/92
Display 1, 7/13/92
1-32Operating Guide
Chapter 1 Getting Started
Making Measurements
Displaying Stats After Filtering Data of Input Signal
Let’s assume you have set the upper and lower limits for the input signal,
and selected the IN LIMIT (filtering) choice. Now, perform the following
steps to display the standard deviation of the filtered measurements.
(Note that the first step in the following procedure is optional since you
should have already set Stats to show standard deviation at the beginning
of this Stats procedure. But, you may want to perform the step anyway to
verify that the Counter is displaying the standard deviation of the
measurement.)
Stats
Display 1, 7/13/92
Run
FreqCh 1
1
FreqCh 1
FreqCh 1
Gate
Gate
The standard deviation value shown in the previous illustration is for
demonstration purposes . The statistic is computed using only those
measurements which fell within the limits you set.
(For more details on the Stats and Limits functions, refer to the
appropriate sections in Chapter 2, “Operating Your Frequency Counter.”)
Now, disable Math and Stats as shown in the following procedure.
Operating Guide1-33
Chapter 1 Getting Started
Making Measurements
Disabling Stats and Math
1
Stats
Scale &
Offset
Display 1, 7/13/92
Display 1, 7/13/92
Run
FreqCh 1
MHz
Gate
The Counter is now making and displaying normal measurements
(that is, the Counter is not showing statistics or scale/offset results).
1-34Operating Guide
Chapter 1 Getting Started
RunSingle/
Stop
Making Measurements
To Control Measurement
Use these two keys to control the measurement of the Counter.
The Run key provides repetitive measurements, whereas the
Stop/Single key allows you to make one measurement.
With the 10 MHz signal still connected to CHANNEL 1, perform the
following steps so you can better understand the Run and Stop/Single
operations.
1
Run
FreqCh 1
MHz
Gate
The Counter is now making repetitive measurements (continuously
making “live” measurements).
Stop/
Single
FreqCh 1
MHz
The Counter stopped making measurement s. The Gate annunciator is not
lit. Hence, pressing the Stop/Single key while the Counter is making
measurements (in Run) causes the Counter to stop after the
measurement in progress is completed. If you pres s the Stop/Single key
again while the Counter is stopped, the Counter will make a single
measurement and then stop—the Gate annunciator will light one time,
momentarily.
If you press the Stop/Single key while the Counter is stopped and when
the Stats menu item ON SINGLE is set to N, the Counter will make
N measurements and then stop. This enables a set of statistics to be
computed.
While the Counter is still stopped, perform the following step.
Run
FreqCh 1
MHz
Gate
The Counter is making repetitive measurements again.
Operating Guide1-35
Chapter 1 Getting Started
Making Measurements
1
1-36Operating Guide
2
Operating Your Frequency Counter
Operator’s Reference
Chapter 2 Operating Your Frequency Counter
Introduction
Introduction
This is the operator’s reference chapter which contains information and
procedures for the front-panel keys, operating functions, and menus of the
HP 53181A 225 MHz Frequency Counter.
Chapter Summary
•How this Counter Works for Youpage 2-4
2
•Summary of the Measurement Sequencepage 2-5
•Using the Measurement Control Keys
(Run and Stop/Single)
•Using Entry/Select (Arrow) Keyspage 2-9
•Using the MEASURE Menu Keyspage 2-11
•Using the Gate & External Arm Menu Keypage 2-15
•Using the MATH Menu Keyspage 2-21
•Using the LIMITS Menu Keyspage 2-31
•Using the DISPLAY Digits Keyspage 2-39
•Using CHANNEL 1 Input Conditioning Keyspage 2-41
•Using the Save and Recall Menuspage 2-48
•Using the Print Menupage 2-52
•Using the Utility Menupage 2-53
•Front Panel Display Messagespage 2-63
•Preset Values After Power-up and *RSTpage 2-68
•Common Questionspage 2-74
page 2-6
2-2Operating Guide
Chapter 2 Operating Your Frequency Counter
Introduction
Where to Find Some Key Working Examples
•Example Procedure for Gate and External Armpage 2-17
•Example Procedure for Changing the Number of
Digits of Resolution Displayed for More
Precise Measurements
•Example Procedure for Scale Functionpage 2-22
•Example Procedure for Offset Functionpage 2-23
•Example Procedure for Set Reference Function—Offset
From the Last Measurement Value (Seeing Digits Not
Visible on the 12-Digit Display)
•Example Procedure for Computing Statspage 2-27
•Limits Testing Example 1—Flag and Stop Measuring
On Limits
•Limits Testing Example 2—Flag On Limits but
Continue Measuring
•Limits Testing Example 3—Use Analog Graph Display
While Adjusting/Aligning Input Signal
•Limits Testing Example 4—Selecting Filtering
Conditions of StatsComputation
•Limits Testing Example 5—Sending the Limit-Detect
Output to the RS-232 Serial Port
page 2-19
page 2-25
page 2-32
page 2-34
page 2-34
page 2-36
spage 2-38
2
Operating Guide2-3
Chapter 2 Operating Your Frequency Counter
How this Counter Works for You
How this Counter Works for You
The followin g text lists some of the key things the Counter does for you.
•The Counter presets the menus to default states and values at
power-up (refer to Table 2-6 for a detailed list of the preset values).
Cycling the POWER key presets the Counter.
•The Counter’s Utility menu allows you to select such things as
timebase source, HP-IB configuration, and RS-232 serial port
2
configuration. After your selections, the Counter automatically stores
all these selections in non-volatile memory (except the timebase source); thus, these settings (except timebase source) will not change
when power has been off or after a remote interface reset.
•The Counter automatically displays measurement(s) in a couple of
seconds after you select a measurement function (e.g., Period).
•The Counter accepts your numeric entry for a menu item after you
press the Enter key. You MUST press the Enter key to complete
numeric entry.
•The Counter automatically enables Limits when you enter a numeric
value for upper and/or lower limits, or update any menu item in the
Limit Modes menu.
•The Counter automatically enables Math after you enter a numeric
value for scale and offset.
•The Counter performs Statistical analyses for you, and will enable
Stats when you update any item in the Stats menu.
•The Counter automatically disables Limits, Math, and Stats when you
select another measurement function.
•The Counter saves measurement setup(s).
•The Counter instantly recalls the measurement setup you want to use.
•The Counter prints out your measurement and analysis data.
2-4Operating Guide
Chapter 2 Operating Your Frequency Counter
Summary of the Measurement Sequence
Summary of the Measurement Sequence
Use Utility menu to:
•choose the timebase source.
•configure the HP-IB if you intend to operate the Counter remotely.
•set RS-232 serial port if you intend for the Counter to perform
printing and/or limit-detecting.
Use MEASURE menu keys to select the measurement function.
Use CHANNEL 1 keys to set up input conditioning.
Use Gate & ExtArm key to:
•set the gate time and resolution.
•set arming.
Use Scale & Offset key to set up math operations.
Use Uppr & Lower key to set limits.
Use Limit Modes key to set up limit testing.
Use Stats key to set up statistics and limit filtering.
Use Save & Print key to enable or disable printing.
Use Run and Stop/Single keys to control measurements.
2
Operating Guide2-5
Chapter 2 Operating Your Frequency Counter
RunStop/
Single
Using the Measurement Control Keys (Run and Stop/Single)
Using the Measurement Control Keys
(Run and Stop/Single)
Overview of the Measurement Control Keys
Two measurement control keys are provided on the Counter: Run and
Stop/Single. In general, the Run key provides repetitive measurements
while the Stop/Single key allows you to make single-shot measurements.
The Run key allows you to:
2
•put the Counter into a continuous measurement loop.
•exit any measurement after your selection of another measurement
function.
•abort the current measurement by starting a new measurement
(if already in Run mode or in Single mode with a measurement in
progress). This also clears any statistics being processed.
The Stop/Single key allows you to:
•put the Counter into Single mode (if the Counter was in Run mode)
where one measurement is taken with each Stop/Single keypress.
(Stats menu item ON SINGLE should be set to 1.)
•put the Counter into Single mode (if the Counter was in Run mode)
where N measurements are taken with each Stop/Single keypress.
(Stats menu item ON SINGLE should be set to N.)
•stop (abort) the current measurement (if the Counter was in Single
mode).
2-6Operating Guide
Chapter 2 Operating Your Frequency Counter
Using the Measurement Control Keys (Run and Stop/Single)
To Use the Measurement Control Keys
The following proced ure demonstrates how these keys function.
1Connect power source to Counter, and turn on Counter.
All segments of the front-panel display will light up while the Counter
performs its power-on self-test, and then dashes are displayed.
The Counter is now ready to measure frequency of a signal applied to
CHANNEL 1 input. Note that the Freq and Ch1 annunciators light.
2Connect an input signal to CHANNEL 1.
The Counter automatically displays the measured frequency of the input
signal. Note that the Run key indicator is ON.
3Press Stop/Single key.
The Counter is put into Single mode and the current measurement is
completed. One measurement is taken with each press of Stop /Single key.
Note that the Stop/Single key indicator is ON.
4Press Run key.
The Counter aborts any measurement in progress, and starts making
repetitive measurements. That is, the Counter is in a continuous
measurement loop.
5Press Stop/Single key.
The Counter is put into Single mode.
2
Operating Guide2-7
Chapter 2 Operating Your Frequency Counter
Using the Measurement Control Keys (Run and Stop/Single)
6Press Stats key until
any one of the arrow keys to set
The Counter is set to take N meas urements on each Stop/ Single keypress
(note: N = 100 at power-up). The state of
the Stop/Single key operation while statistics are enabled.
Observe that the Stats indicator is ON. This indicates statistical
processing is enabled.
7Press Stop/Single key.
2
One set of N frequency measurements is taken with each press of the
Stop/Single key. One set of statistics is computed with each press of the
Stop/Single key.
8Press Stats key until
9Press any one of the arrow keys to select a statistical result to
display, and press Freq Ch 1 key.
The Counter is now displaying one of the statistical results.
10Press Stop/Single key.
ON SINGLE
SHOW
menu item appears.
menu item appears, then press
ON SINGLE
ON SINGLE
menu item to N.
menu item affects
One set of N frequency measurements is taken with each press of the
Stop/Single key. At the completion of th e Nth meas ur eme nt, the Cou nte r
displays the statistical result selected in step 9.
11Press Stats key until
any one of the arrow keys to set
The Counter is set to take one measurement on each Stop/Sing le
keypress.
Disabling the statistics (with the
to cause the Stop/Single key to initiate one measurement at a time.
2-8Operating Guide
ON SINGLE
STATS: OFF
menu item appears, then press
ON SINGLE
menu item to 1.
menu item) is another way
Chapter 2 Operating Your Frequency Counter
Enter
+/–
Using Entry/Select (Arrow) Keys
Using Entry/Select (Arrow) Keys
There are six entry/select keys of which four are “arrow” keys.
The function of the four arrow keys and the Enter key depends on the
Counter’s operating mode (that is, numeric entry, state changing,
sequencing through choices in a menu, etc.). This section describes how
the keys perform in the different operating modes.
To Use During Numeric Entry
•Use the s key to move left to select adjustable digits (it does not wrap
around).
•Use the g key to move right to select ad justable digi ts (it doe s not wrap
around).
•Use the d key to increment the selected (highlighted) digit of the
displayed value.
•Use the f key to decrement the selected (highlighted) digit of the
displayed value.
•Use the +/− key to change the sign of the numeric value.
•Use the Enter key to complete a numeric entry. (If a numeric entry is
changed and the Enter key is NOT pressed, then the value is not
changed.)
To Use When Sequencing Through the Other Meas and
Recall Menus
•Use the d or s key to go back to the previous choice in a function or
recall menu. Repeatedly pressing this key will cycle through (and loop
around) the list of choi ces.
•Use the f or g key to go forward or to the next choice in a function or
recall menu. Repeatedly pressing this key will cycle through (and loop
around) the list of choi ces.
2
•Use the Enter key to select the function or execute the recall, or wait
for the menu to time out for automatic selection/execution.
Operating Guide2-9
Chapter 2 Operating Your Frequency Counter
Using Entry/Select (Arrow) Keys
To Use During State Changing (ON/OFF,
LO/MED/HI, etc.)
•Press any of the arrow keys to toggle or ch ange to th e next stat e of the
parameters found in the following menus:
•The +/− key is ignored during the st ate changing operating mode.
Ω
To Use on Prompted Help Messages
(MATH HELP: NO/YES, PRT HELP: NO/YES)
•When
•When
MATH HELP: YES
“
(MEAS x SCALE) + OFFS = RESULT
PRT HELP: YES
“
CONFIGURE PORT ON UTILITY MENU; REMEMBER TO SAVE
SETUP FIRST
” instruction.
is displayed, press Enter key to display the
” message.
is displayed, press Enter key to display the
2-10Operating Guide
Chapter 2 Operating Your Frequency Counter
Other
Meas
Freq
Ch 2
Gate &
ExtArm
MEASURE
Freq
Ch 1
Using the MEASURE Menu Keys
Using the MEASURE Menu Keys
Overview of the MEASURE Menus
The measurement function menus are divided into three categories:
•Frequency measurements on Channel 1.
•Frequency measurements on Channel 2 (optional).
•Other measurements (period, ratio (optional), and peak voltage).
Table 2-1. Menus of the MEASUREment Keys
Freq Ch 1 key Freq Ch 2 keyOther Meas key
2
FREQUENCY 1FREQUENCY 2
(optional)
PERIOD 1
RATIO 1 TO 2 (optional)
RATIO 2 TO 1 (optional)
VOLT PEAKS 1
To sequence through the Other Meas menu, you simply need to repeate dly
press the Other Meas key to cycle through (and loop around) the items (or
measurement functions) under the key. Each press of a Other Meas key
will advance the Counter to the next measurement function in the menu.
NOTE
The menus under the MEASURE keys will terminate (select function and
return to measurement result display) after a short timeout.
WARNINGWhen measuring power line frequencies, be extremely careful and always
use a step-down isolation transformer of which output voltage is
compatible with the input measurement capabilities of this product.
Since the shell of the BNC input connector is grounded to the chassis of
the instrument, the hot side of an ac power line could be connected to the
BNC shell and create a direct short to ground. Therefore, NEVER TRY
TO MEASURE AC POWER LINE FREQUENCIES WITHOUT AN
ISOLATION TRANSFORMER.
Operating Guide2-11
Chapter 2 Operating Your Frequency Counter
Using the MEASURE Menu Keys
To Measure Frequency
1Connect power source to Counter, and turn on Counter.
All segments of the front-panel display will light up while the Counter
performs its power-on self-test, and then dashes are displayed.
The Counter is now ready to measure frequency of a signal applied to
CHANNEL 1 input. Note that the Freq and Ch1 annunciators light.
2Connect an input signal to CHANNEL 1.
The Counter should automatically display the measured frequency of the
2
input signal. The trigger levels are set to the 50% points of the signal.
However, you may disable auto triggering, and change the trigger levels
and slopes.
To set up CHANNEL 1’s coupling, impedance, and triggering conditions to
match the input signal you are trying to measure, refer to the procedures
“To Select Input Coupling and Impedance” and “To Set Input Channel
Trigger Level/Sensitivity” in Chapter 1, “Getting Started,” or the section
titled “Using CHANNEL 1 Input Conditioning Keys” in this chapter.
3To measure the frequency of a signal applied to the optional
CHANNEL 2 input, press Freq Ch2 key.
FREQUENCY 2
annunciators will light. The Counter will be ready to measure the
frequency of a signal applied to CHANNEL 2 input.
Note that the Frequency 2 menu item is not available in Counters that do
not contain the optional Channel 2.
2-12Operating Guide
will be momentarily displayed and the Freq and Ch2
Chapter 2 Operating Your Frequency Counter
Using the MEASURE Menu Keys
To Measure Frequency Ratio (Optional)
1Connect signals to be measured to channels 1 and 2.
2Press Other Meas key until
the frequency ratio between a signal appl ied to CHANNEL 1 input
and a signal applied to the optional CHANNEL 2 input.
RA TIO 1 T O 2
annunciators will light. The Counter will be ready to measure the
frequency ratio of a signal applied to CHANNEL 1 in relation to a signal
applied to CHANNEL 2 (Ch1/Ch2).
RATIO 2 TO 1 is also available (Ch2/Ch1).
Note that Ratio 1 to 2 and Ratio 2 to 1 menu items are not available in
Counters that do not con tain the optional Channel 2.
Note the result is not scaled by 100; it is not a percentage.
The front-panel “arrow” (or entry/select) keys can also be used when
sequencing through the Other Meas menu. Refer to the section titled
“Using Entry/Select (Arrow) Keys” for details.
will be momentarily displayed and the Freq, Ch1, and Ch2
RATIO 1 TO 2
is displayed to measure
To Measure Period
Press Other Meas key until
PERIOD 1
light, and the Counter is ready to measure the period of a signal applied to
CHANNEL 1.
is momentarily displayed, the Period and Ch1 annunciators
PERIOD 1
is displayed.
2
Operating Guide2-13
Chapter 2 Operating Your Frequency Counter
Using the MEASURE Menu Keys
To Measure Positive/Negative Volt Peaks
Press Other Meas key until
VOLT PEAKS 1
and the Counter is ready to measure minimum and maximum voltage
peaks of signals applied to CHANNEL 1.
NOTE
2
NOTE
In Volt Peaks function, the Counter measures the input signal after the
signal has been conditioned by the input settings (impedance, coupling,
attenuation, and filter). One reminder of this is the (
message, which is displayed when Voltage Peaks is selected and the
measurement channel is ac coupled.
The result is corrected for X10 attenuation.
Arming is not used for Voltage Peaks measurements. Also, Limits, Math,
and Display Digits are not available for Voltage Peaks measurements.
is momentarily displayed, the Ch1 annunciator lights,
VOLT PEAKS 1
is displayed.
AC COUPLED
)
2-14Operating Guide
Chapter 2 Operating Your Frequency Counter
Other
Meas
Freq
Ch 2
Gate &
ExtArm
MEASURE
Freq
Ch 1
Using the Gate & External Arm Menu Key
Using the Gate & External Arm Menu Key
Overview of Gate/External Arming Functions
Table 2-2 lists the menus for all the arming configurations, and shows
how the Gate and External Arming menu changes as a function of the
arming mode.
The Counter has four arming modes: auto, external, digits and time.
The arming modes are defined in the following paragraphs.
TIME Arming
For frequency, period, and frequency rati o, time arming means you can
set the gate time—the length of time for which your signal is measured.
2
NOTE
NOTE
Gate time and resolution are linked: the longer the gate time, the greater
the resolution. With an increase in resolution, you get a larger number of
digits on the display.
Time arming is demonstrated in the sub-section titled “Example
Procedure for Gate and External Arm.”
AUTO Arming
For frequency, period, and frequency rati o, auto arming means make
individual measurements as fast as possible.
EXTERNAL Arming
For frequency, period, and frequency rati o, external arming means you
can start a measurement by using an external signal.
When external arming mode is enabled, a signal must be connected to the
Counter’s rear-panel Ext Arm connector.
Operating Guide2-15
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
DIGITS Arming
For frequency, period, and frequency ratio, digits arming means you can
set the number of digits of resolution that you require y our result to provide.
Digits arming is demonstrated in the sub-section titled “Example
Procedure for Changing the Number of Digits Displayed.”
*Table 2-2. The Gate & External Arm Key Menus as a Function of the
Arming Mode
Frequency, Period, Ratio
2
Time Arming:
GATE:TIME
TIME:<time>
Auto Arm i ng:
GATE:AUTO
Digits Arming:
GATE:DIGITS
DIGITS: <digits>
External Arming (Refer to page 2-17 or the Menu Roadmap in the Quick Reference
Guide, which follows the Preface, for mo r e in fo rm ation.):
GATE:EXTERNL
START: POS
NEG
STOP:AUTO
NEG
POS
TIME
TIME:<time>
* There is no Gate and External Arm menu available for Voltage Peaks measurements.
2-16Operating Guide
Chapter 2 Operating Your Frequency Counter
Ext
Arm
Using the Gate & External Arm Menu Key
To Use the Gate and External Arm
Example Procedure for Gate and External Arm
For demonstration purposes, you want to set up the Counter so that it
uses an external arm to start a measurement. You want to make a
Frequency measurement on a signal applied to channel 1. Also, you want
to perform the measurement during an interval of 5 milliseconds as shown
in the following figure.
NOTE
As shown in the following figure, the Counter is always synchronized to
the input signal. Arming occurs on the rising edge of external arm
(ExtArm), and the measurement gate is synchronized to the trigger event
on CHANNEL 1.
CHANNEL 1
Ext
Arm
Gate
5 ms
Perform the following procedure to accomplish this task.
1Turn on the Counter, and press Stop/Single to put the Counter
into the single measurement mode.
2Connect the signal to channel 1 input, and press F req Ch 1 key.
3Connect an external TTL signal to the Counter’s rear-panel
Ext Arm connector.
2
4Press Gate & ExtArm key.
GATE: TIME
is displayed.
Operating Guide2-17
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
5Press any one of the arrow keys until
displayed.
6Press Gate & ExtArm or Enter key.
START: POS
press any one of the arrow keys to toggle to the next state—
7Press Gate & ExtArm or Enter key.
STOP: NEG
2
8Press any one of the arrow keys until
9Set the Gate Time to 5 ms by performing the following steps:
a. Press Gate & Ex tArm or Enter key.
TIME: .100
b. Press the appropriate arrow keys until
c. Press Enter key.
NOTE
BE SURE to press the Enter key to complete the numeric entry.
is displayed. (If
is displayed.
seconds (the default time) is displayed.
START: NEG
GATE: EXTERNL
is displayed instead, then
STOP: TIME
TIME: .00500
is
POS
).
is displayed.
s is displayed.
The counter will now gate for a duration of 5 ms. The 5 ms interval will
begin after the external edge is detected.
10Press Stop/Single key to start the measurement.
The measurement is started after the external arm pulse occurs.
2-18Operating Guide
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
Example Procedu re for Chan g i ng th e Nu m be r of
Digits of Resolution Displayed for More
Precise Measurements
1Turn POWER key OFF then ON again (cycle POWER key) to preset
the Counter.
2Connect (for demonstration purposes) the 10 MHz Out signal to
the channel 1 input.
3Press Other Meas key until
After a few seconds,
number of digits (or resolution) displays a precise measurement.
4To demonstrate how to use “digits arming” to change the number
of digits displayed, perform the following steps first.
a. Press Gate & ExtArm key
GATE: TIME
b. Press any one of the arrow keys until
c. Press Enter or Run key.
0.100,0
In “auto arming” you produce results quickly. The short gate time
produces a lower-resolution result.
µs is displayed. Note that fewer digits are displayed.
0.100,000,000,0
is displayed.
PERIOD 1
is displayed.
µs is displayed. Hence, the default
AUTO
is displayed.
2
Operating Guide2-19
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
5To change the display to the better measurement resolution using
digits arming, perform the following steps:
a. Press Gate & Ex tArm key
GATE: AUTO
b. Press any one of the arrow keys until
c. Press Gate & ExtArm or Enter key.
DIGITS: 4
2
d. To set the number of digits displayed to 8, press the appropriate arrow
keys until
e. Press Enter key.
f.Press Run key.
0.100,000,000
Note that 8 digits of resolution are displayed (
For the convenience of the user, the leading “1” does not count as a
digit when 8 digits are selected; thus, a result of
alternate with a result of
no change of settings is required.
is displayed.
DIGITS
is displayed (the default number of digits)
DIGITS: 8
µs is displayed.
is displayed.
is displayed.
1 2 3 4 5 6 7 8
0.100,000,000
0.099,999,999
0.100,000,000
µs with a stable display, and
µs).
µs can
2-20Operating Guide
Chapter 2 Operating Your Frequency Counter
Stats
Scale &
Offset
MATH
Using the MATH Menu Keys
Using the MATH Menu Keys
Note that Math and Limits are not available for Voltage Peaks
measurements.
Overview of Scale/Offset Math Menu
The Scale and Offset functions within the Math menu allow you to
perform simple mathematical operations on the measurement result
before it is displayed. The Scale and Offset functions allow for
multiplication and addition. Modification of the displayed measurement
by these math operations is represented by the following equation:
(Measurement × Scale) + Offset = Displayed Result
The math operations can be used, for example, to subtract systematic
errors or display the percentage difference between signals. Also, the
overall math operation can be disabled and then re-established without
having to re-enter constants.
2
The menu items of the Scale and Offset Math functions allow you to:
•enter a desired multiplication factor for a measurement (SCAL:).
•enter a desired addition or subtraction value for a measurement
(OFFS:).
•get and display the last measurement value, and use it as a negative
offset (SET REF: NO or YES).
•disable or enable the Math mode (MATH: OFF or ON)—Note that
the Math mode is automatically enabled when you enter the scale
and/or offset value(s); thus, the primary function of this menu item is
to allow you to turn the Math mode off.
•get Help, which displays the equation “
RESULT
” to tell you how the scale and offset results are computed by
(MEAS X SCALE) + OFFS =
the Counter (MATH HELP: NO or YES).
Operating Guide2-21
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
To Use the Scale/Offset Math Menu
Example Procedure for Scale Function
For demonstration purposes, you have a motor that has a tachometer
attached to its rotating shaft that generates a pulse for every revolution of
the shaft.
Your task is to measure the speed or revolutions per minute (rpm) of
the tachometer’s output, and make the Counter directly display the
measurement in rpm (which is fundamentally a frequency measurement).
The equation for this task is:
2
1Connect the output to channel 1 input, and press Freq Ch 1 key.
2Press Scale & Offset key until
3Set the Scale to 60 (since there are sixty seconds in a minute) by
NOTE
Frequency (Revolution/Sec) × Scale (60 Sec/Min) =
Results (Revolution/Min)
Perform the following procedure to accomplish this task.
The frequency of the signal is displayed.
SCAL: 1.000000
performing the following steps:
a. Press appropriate arrow keys to set Scale to 60.
b. Press Enter key.
BE SURE to press the Enter key to complete the numeric entry.
The scale is now set to 60.
When the scale factor is set, both scale and offset are applied to the
measurement result. Verify that the offset value is zero for applications
requiring only scale by performing the following step.
is displayed.
2-22Operating Guide
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
NOTE
4Press Scale & Offset key, if required, until
OFFS: 0.000000
“0.000000”:
a. Press the appropriate arrow keys to set the Offset t o “0.000000.”
b. Press Enter key.
BE SURE to press the Enter key to complete the numeric entry.
The offset is now set to 0.
5Press Run key.
The rpm measurement is displa yed.
Note that units indicators
should be displayed. If the Offset value is not set to
Hz
and s do not appear when MATH is ON.
OFFS:
is displayed.
Example Procedure for Offset Function
The offset function can be use d to add or subtract a fixed value from
measurement results. This can be very useful, for example, when tuning
or aligning odd frequencies of a local oscillator (LO). The target frequency
can be entered as a negative offset, then the Counter can display the
differences between the LO’s frequenc y and the target freq ue ncy. You can
then adjust the LO until zero is displayed in the Counter’s display.
2
Your task is to use the offset function to tune or align an LO to the target
frequency of 222.127 MHz. Perform the following procedure to accomplish
this task.
1Cycle POWER key to preset the Counter.
2Connect the LO to channel 1, and press Fre q Ch 1 key to measure
the frequency of the LO.
3Press Scale & Offset key until
4Set Scale to “1” by pressing the appropriate arrow keys to enter
the value 1.000000 if Scale is not already set to “1.”
Operating Guide2-23
SCAL:
is displayed.
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
5Press Enter key.
NOTE
2
NOTE
BE SURE to press the Enter key to complete the numeric entry.
6Press Scale & Offset key, if required, until
OFFS: 0.000000
7Set the Offset to -222.127 MHz.
Refer to the sub-section titled “Entering the Offset Value” in Chapter 1,
“Getting Started,” for details on how to perform numeric entry for Offset.
a. Use the appropriate combination of the arrow keys to display a value
of 222.127 MHz.
b. Press +/− key, which is located below the arrow keys, to change the
plus sign to minus.
OFFS: -222.127 MHz
c. Press Enter key.
BE SURE to press the Enter key to complete numeric entry.
8Press Run key.
should be displayed. This is the default value.
is displayed.
OFFS:
is displayed.
The resultant frequency is displayed with -222.127 MHz subtracted from
each measurement result.
The Counter displays the error or differences from the targeted
222.127 MHz.
For example, the displa y might indicate
9Adjust the LO as necessary until the Counter displays zero
(
00.000,000,0
Your local oscillator is now properly aligned.
2-24Operating Guide
).
+000.023 MHz
.
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
Example Procedure for Turning Off Math Mode
1Press Scale & Offset key until
2Press any one of the arrow keys.
MATH: OFF
modify the displayed result.
3Press Run key to return the Counter to its normal operation.
is displayed. Now, Scale and Offset no l onger are used to
MATH: ON
is displayed.
Example Procedure for Set Reference Function—Offset From
the Last Measurement Value (See in g Digits Not Visi ble on
the 12-Digit Display)
The
SET REF: YES
value, rounded to elev en digits , to set the OFFSET. It also re sets t he scal e
value to 1.0. Thus, any diffe rence in the offset (o r now the reference va lue)
and the current measurement value is displayed.
Let’s assume you are measuring the frequency of a stable source, but you
are only interested in the least significant digi ts of the measurement.
Use the
value.
Your task is to measure a stable source to obtain the changes in the least
significant digits.
SET REF: YES
menu item uses the negative of the last measurement
menu item to configure the appropriate OFFSET
2
1Cycle POWER key to preset the Counter.
GATE
2Change gate time to 15.0 seconds on
3To demonstrate SET REF:, connect a 10 MHz sine wave signal to
channel 1 of the Counter, and press Freq Ch 1 key.
(Note that since the gate time is set to 15 seconds, it will take about
15 seconds to for the Counter to display the measurement.)
10.000,000,000,1 MHz
4Press Scale & Offset key until
Operating Guide2-25
is displayed.
SET REF:
menu.
is displayed.
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
5Press the any one of the arrow keys to select
press Enter key.
OFFS: −10.00000 M
Note that this valu e represents the last measurement of the input signal
rounded to eleven digi ts. The Counter us es it as the ref erence off set val ue,
and reverses the sign from + to − .
6Press Run key.
−
12*
µ (120 µs) is displayed, for example, at the completion of the next
2
measurement.
This value (−0.00012) is the small difference between the signal being
measured and the reference value (
Using the reference allows you to observe small differences in the
13th–15th digits which would not be observable in the 12-digit display.
is displayed, for example.
OFFS: −10.00000 M
SET REF: YES
) obtained in step 5.
, and
Overview of Statistics (Stats) Menu
The Stats menu allows you to:
•select which computed statistics you want displayed
(SHOW:STD DEV, MEAN , MAX, or MIN).
•choose between having the actual measurement or statistical result
displayed (SHOW: MEAS).
•select the number of measurements you want to make for each
computed statistic (N:).
•enable or disable the Stats mode (STATS: ON or OFF).
•enable the Counter to perform statistical computation on only
measurements that are within the user-entered limits; that is, filter
data (USE: IN LIMIT), or
•enable the Counter to perform statistical computation on all
measurements, whether they fall in or out of the user-entered limits
(USE: ALL MEAS).
•configure Stop/Single key to initiate a set of N measurements
(ON SINGLE: 1 or N).
2-26Operating Guide
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
To Use the Stats Menu for Automatic and Continuous
Statistical Analysis
Example Procedure for Computing Stats
For demonstration purposes, let’s say you need to know the mean
(average) and minimum frequency values of a signal. Also, you want the
Counter to make 20 measurements before it performs these statistical
computations.
Perform the following procedure to accomplish this task.
NOTE
1Connect the signal to channel 1 input, and press Freq Ch 1 key.
2Press Stats key until
N: 100
3Set N to 20 (since you want t he Counter t o make 20 measurements
before computing the mean and minimum values) by pressing the
appropriate arrow keys.
N: 20
4Press Enter key.
BE SURE to press the Enter key to complete numeric entry.
The Counter is now set to make 20 measurements before computing
statistics.
5Press Stats key until
6Press any one of the arrow keys until
is displayed (100 is the default value).
is displayed.
N:
is displayed.
SHOW:
is displayed.
SHOW: MEAN
is displayed.
2
Operating Guide2-27
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
7Press Run key.
Immediately after the Run key is pressed, the Counter momentarily displays
DOING STATS
displays the mean frequency value of the input signal.
. When the Coun te r ha s ma de 2 0 v ali d me asu rem en ts , it t he n
NOTE
The Counter will display updated statistics results every 20 measurements.
8Press Stop/Single key after you have a mean value displayed.
9To display the minimum frequency value, perform the following:
2
a. Press Stats key.
b. Press any one of the arrow keys until
c. Press Freq Ch 1 key.
CAUTIONPressing the Run key now would clea r the pr esent statistics results and
calculate a new set of results after “N” measurements.
Example Procedure for Easy Viewing of Stats
The d and f arrow keys provide a short cut to viewing each of the stats
(that is, standard deviation, mean , maximum, an d minimu m) wh en the
Counter is already displaying a statistic measurement. These keys allow you
to cycle through the different Stats results and automatically view them.
While the Counter is displaying a statistics measurement, press
either the d or f arrow key to cycle through and display each of
the four computed statistics, and the current measurement.
SHOW: MIN
is displayed.
NOTE
The Display Digits keys (that is, More Digits and Fewer Digits)
do NOT affect standard deviation display resul ts.
2-28Operating Guide
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
Example Procedure for Filtering Data (Using Limits)
During Stats
1Using the procedure in the previous sub-section titled “Example
Procedure for Computing Stats,” set up the Counter to display the
minimum frequency value of an input signal.
2Set the upper and lower limits. (Refer to the section in Chapter 1 titled
“To Set Limits of Measurements” if needed.)
3Press Stats key until
4Press one of the arrow keys to toggle to
USE: IN LIMIT
5Press Run key.
The Counter will now perform the statistical computations (that is,
standard deviation, mean, maximum, and minimum) on only those
measurements that are within the user-entered limits. The Counter then
displays the statistics you picked (minimum, in this case).
6Press Stats key until
7Press any one of the arrow keys to toggle to
is displayed.
USE: ALL MEAS
8Press Run key.
The Counter will now perform the statistical computations (that is,
standard deviation, mean, maximum, and minimum) on all
measurements, whether they fall in or out of the user-entered limits.
The Counter then displays the statistic you picked (minimum, in this
case).
USE:
is displayed.
USE:
is displayed.
is displayed.
IN LIMIT
ALL MEAS
state.
state.
2
Operating Guide2-29
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Keys
Example Procedure for Configuring SINGLE to Initiate N
Measurements
Please refer to the procedure titled “To Use the Measurement Control
Keys” on page 2-7 in this chapter for an example use of ON SINGLE.
Example Procedure for Turning Off Stats Mode
1Press Stats key until
2Press any one of the arrow keys.
2
STATS : OFF
3Press Run key to return the Counter to its normal operation.
is displayed.
STATS: ON
is displayed.
2-30Operating Guide
Chapter 2 Operating Your Frequency Counter
Limit
Modes
Uppr &
Lower
LIMITS
Using the LIMITS Menu Keys
Using the LIMITS Menu Keys
Note that Math and Limits menus are not available for Voltage Peaks
measurements.
Overview of Limits Menus
The menu items under the Limits keys allow you to:
•select the desired upper and lower measurement limits
(UPPR:, LOWR:).
•disable or enable Limit Testing (LIM TEST: OFF or ON)—Note that
the Limit Testing is automatically enabled when you set the upper
and lower limits, or update any menu item in the Limit Modes menu;
thus, the primary function of the LIM TEST menu item is to allow you
to turn Limit Testing off.
•set the Counter to stop making measurements when a measurement
exceeds the user-entered limits (ON FAIL: STOP). (Stop on limits
does not have any effect if the Counter is in the single measurement
mode.)
2
•enable the Counter to continue taking measurements after a
measurement exceeds the user-entered limits (ON FAIL: GO ON).
•change the Counter display from numeric to analog graph to quickly
visualize changes in the signal (or to quickly see when measurements
are in or out of limits) (SHOW: NUMBER or GRAPH).
Operating Guide2-31
Chapter 2 Operating Your Frequency Counter
Using the LIMITS Menu Keys
To Set and Use Automatic Limit Testing
Limits Testing Example 1—Flag and Stop Measuring
On Limits
Automatic limit testing can be used to monitor a test signal to assure that
it does not drif t outside of a user-entered set of limi ts.
For demonstration purposes, you want to first measure the frequency of a
10 MHz signal applied to channel 1 of the Counter. Next, you want to set
limits that would cause the Counter to flag (turn on the Limits
annunciator in the front-panel display) and stop making measurements if
2
this signal drifts more than ±1 kHz.
1Connect the 10 MHz input signal to channel 1 of the Counter,
and press Freq Ch 1 key.
Assume that the frequency measured is 10 MHz. If the drift is not to
be larger than ±1 kHz, then the upper limit should be set to 10.001 MHz
and the lower limit to 9.999 MHz. Let’s set these values in the following
steps.
NOTE
2Press Uppr/Lower key until
UPPR: 0.000000
3SetUPPR to 10.001 MHz. (Refer to the section in Chapter 1 titled
“To Set Limits of Measurements” if needed.)
After you enter the value 10.001 MHz, BE SURE to press the
Enter key to c omplete the en try.
4Press Uppr/Lower key, if required, until
LOWR: 0.000000
2-32Operating Guide
is displayed (0 is the default value).
is displayed (0 is the default value).
UPPR:
is displayed.
LOWR:
is displayed.
Chapter 2 Operating Your Frequency Counter
Using the LIMITS Menu Keys
NOTE
5Set
6Press Limit Modes key until
7Press Limit Modes key.
8Press any one of the arrow keys to toggle to the next state.
9Press Run key to start making measurements with this
LOWR
After you enter the value 9.999 MHz, BE SURE to press the
Enter key to c omplete the en try.
Note that once either the upper or lower limits have been set, the limit
testing is automatically enabled as indicated by the
display, and the Limit Modes indicator.
ON FAIL: GO ON
ON FAIL: STOP
Counter to stop making measurements when the input signal drifts more
than ±1 kHz.
limits-testing choice selected.
to 9.999 MHz.
is displayed.
is displayed. The
LIM TEST: ON
ON FAIL: STOP
is displayed.
LIM TEST: ON
function tells the
2
Now, if a measurement drifts out of the user-entered lim its the Limit
annunciator in the display will light. The Counter will stop making
measurements and hold the out-of-limit measurement in the display.
Operating Guide2-33
Chapter 2 Operating Your Frequency Counter
Using the LIMITS Menu Keys
Limits Testing Example 2—Flag On Limits but
Continue Measuring
If you decide that you want the Counter to flag measurements that are out
of limits but to continue taking measurements, then perform the following
steps.
1Press Limit Modes key until
2Press any one of the arrow keys to toggle to the next state.
ON FAIL: GO ON
is displayed.
ON FAIL: STOP
is displayed.
2
3Press Run key to start making measurements with this
limits-testing choice selected.
Now, each time a measurement drifts out of the user-entered limits the
Limit annunciator in the display will light, but the Counter will continue
to make measurements.
The Limit annunciator will go on whenever a result is out of limits.
As soon as a result is within limits, the Limit annunciator will go off.
Limits Testing Example 3—Use Analog Graph Display While
Adjusting/A l ig n i ng In pu t Si g n a l
1For demonstration purposes, connect a 10 MHz signal to
channel 1, and press Freq Ch 1 key.
2Using the Uppr & Lower key, set the upper limit to 10.2 MHz and
the lower limit to 9.8 MHz. (Refer to the section in Chapter 1 titled
“To Set Limits of Measurements” if needed.)
3Press Limit Modes key until
4Press any one of the arrow keys to toggle to the next state.
SHOW: GRAPH
2-34Operating Guide
is displayed.
SHOW: NUMBER
is displayed.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.