Agilent 53150A Product Overview

Full-featured CW Microwave Counters for Field, Factory or Lab
Product Overview
• Agilent 53150A 20 GHz Counter
• Agilent 53151A 26.5 GHz Counter
• Agilent 53152A 46 GHz Counter
High performance microwave counters: at home in the field, bench or ATE environment
• Simultaneous power and frequency measurement with analog peaking indicator
• GPIB and RS-232 standard
• Lightweight and rugged
• Battery optional
2
Convenience, portability and outstanding performance
The innovative designs of the Agilent 53150 Series microwave counters offer an uncluttered, feature­laden front panel. These designs present no-compromise performance and quality in a surprisingly small, light, battery operated product.
The convenience of a single microwave input
The Agilent 53150 Series has an advanced sampler that integrates a separate zero bias Schottky diode for the accurate measurement of input power. This allows measurement of both frequency and power with a single connection. No compromise in frequency coverage is required for this capability. The ultrawideband microwave input covers the entire RF and microwave spectrum, from intermediate frequencies IFs) of 50 MHz to millimeter waves.
The power measurement accuracy and repeatability of these counters rivals power meters with diode sensors. Since frequency and power of the input signal are measured simultaneously, adjustment for the diode’s frequency response is done automatically. And like the latest in diode sensors, compensation is also made for deviation from square law.
Field tough but ready for benchtop or ATE applications
The Agilent 53150 Series is as comfort­able in the field as in the laboratory. The rugged case with an integrated tilting handle can tolerate the vibration and shock expected in field use. The backlit LCD display ensures visibility in all environments, from dark to full sunlight, at distances exceeding 15 feet.
If ac power is unavailable, the internal, replaceable camcorder batteries pro­vide at least 2.5 hours of continuous operation. The unit can also be powered from an external 11-18 Vdc source.
For benchtop and ATE applications, the Agilent 53150 Series delivers full functionality and high measurement speed. The fully programmable RS-232 interface and high speed GPIB interface are standard features.
No compromise performance
The Agilent 53150 Series offers exceptional sensitivity by utilizing a single board design with low phase noise PLL circuitry. Despite their simple appearance, these counters retain all the powerful functions one expects in precision instrumentation: measurement averaging, arbitrary as well as nulling offsets for both frequency and power, display of power in either dBm or Watts.
Additional capabilities include full control of resolution, sampling rate, and GPIB address plus extensive self-diagnostics, fast acquisition times and full programmability. Performance surpasses the industry standard Agilent 5350 Series, in virtually every aspect, in a package less than half the weight and size.
3
Input characteristics Agilent 53150A Agilent 53151A Agilent 53152A
Frequency range
Channel 1 (Normal mode) 10 Hz - 125 MHz 10 Hz - 125 MHz 10 Hz - 125 MHz
(Low pass filter enabled) 10 Hz - 50 kHz 10 Hz - 50 kHz 10 Hz - 50 kHz
Channel 2 50 MHz - 20 GHz 50 MHz - 28.5 GHz 50 MHz - 26.5 GHz
Sensitivity
Channel 1
10-30 Hz 40 mV 40 mV 40 mV 30 Hz-125 MHz 25 mV 25 mV 25 mV
Channel 2
50-250 MHz –20 dBm –20 dBm –20 dBm
0.25-12.4 GHz –33 dBm –33 dBm –33 dBm
12.4-18 GHz –33 dBm –33 dBm –30 dBm 18-20 GHz –29 dBm –29 dBm –27 dBm 20-26.5 GHz N/A –25 dBm –27 dBm
26.5-40 GHz N/A N/A –23 dBm 40-46 GHz N/A N/A –17 dBm
Maximum input
Channel 1 2 Vrms 2 Vrms 2 Vrms Channel 2
50 MHz - 2 GHz +5 dBm +5 dBm +5 dBm 2-46 GHz +13 dBm +13 dBmq +13 dBm
Damage level
Channel 1 120 V (dc + ac pk) linearly 120 V (dc + ac pk) linearly 120 V (dc + ac pk) linearly Channel 2 derated to 5 Vrms at 125 MHz derated to 5 Vrms at 125 MHz derated to 5 Vrms at 125 MHz
+27 dBm +27 dBm +27 dBm
Impedance
(Nominal) Channel 1 1 M/60 pF 1 M/60 pF 1 M/60 pF Channel 2 50 50 50
Connector
Channel 1 BNC female BNC female BNC female Channel 2 SMA/APC-3.5 SMA/APC-3.5 2.92 mm removable,
compatible female compatible female SMA/APC-3.5 compatible female
SWR
Channel 2
50-250 MHz 1.5:1 typical 1.5:1 typical 1.5:1 typical
0.25-10 GHz 2.0:1 typical 2.0:1 typical 2.0:1 typical 10-20 GHz 3.0:1 typical 3.0:1 typical 3.0:1 typical 20-26.5 GHz N/A 3.0:1 typical 2.5:1 typical
26.5-46 GHz N/A N/A 2.5:1 typical
Coupling
Channel 1 ac ac ac Channel 2 ac ac ac
Acquisition time
(1 MHz FM rate) Channel 1 N/A N/A N/A Channel 2 (FM Auto/FM Off) 125 ms/100 ms 125 ms/100 ms 140 ms/115 ms
Resolution
Channel 1/Channel 2 1 Hz to 1 MHz 1 Hz to 1 MHz 1 Hz to 1 MHz
Measurement specifications and characteristics
All measurement specifications are over the full signal and temperature ranges unless otherwise noted.
4
Input characteristics Agilent 53150A Agilent 53151A Agilent 53152A
Emissions
(“kickback noise“) Channel 1 N/A N/A N/A Channel 2 (measuring/no input) –40 dBm/<–70 dBm –40 dBm/<–70 dBm –40 dBm/<–70 dBm
Residual stability*
Channel 1 N/A N/A N/A Channel 2 0.6 LSD rms 0.8 LSD rms 1.25 LSD rm
*Counter and source tied to
same timebase
Accuracy
Channel 1/Channel 2 ±1 LSD ± timebase error x ±1 LSD ± timebase error x ±1 LSD ± timebase error x
frequency frequency frequency
Measurement time
Channel 1 1/Resolution + 20 ms 1/Resolution + 20 ms 1/Resolution + 20 ms
Channel 2 1/Resolution + 1/Resolution + 1/Resolution +
Acquisition time + 20 ms Acquisition time + 20 ms Acquisition time + 20 ms
FM tolerance
Channel 1 N/A N/A N/A Channel 2 (FM Auto) 20 MHz p-p max @ 10 MHz rate 20 MHz p-p max @ 10 MHz rate 20 MHz p-p max to 26.5 GHz,
12 MHz p-p max above 26.5 GHz @ 10 MHz rate
(FM Off) 1 MHz p-p @ 10 MHz rate 1 MHz p-p @ 10 MHz rate 1 MHz p-p @ 10 MHz rate
AM tolerance
Channel 1 N/A N/A N/A Channel 2 Any index provided minimum Any index provided minimum Any index provided minimum
signal level is not less signal level is not less signal level is not less than sensitivity than sensitivity than sensitivity
Amplitude discrimination
Channel 1 N/A N/A N/A Channel 2
below 250 MHz N/A N/A N/A above 250 MHz Automatically measures the largest Automatically measures the largest Automatically measures the largest
signal present provided signal is signal present provided signal is signal present provided signal is >10 dB(typical) above any signal >10 dB(typical) above any signal >10 dB(typical) above any signal separated by less than 75 MHz; separated by less than 75 MHz; separated by less than 75 MHz; >20 dB (typical) above any signal >20 dB (typical) above any signal >20 dB (typical) above any signal separated by more than 75 MHz separated by more than 75 MHz separated by more than 75 MHz
Power measurement
Channel 1 N/A N/A N/A Channel 2
Range Counter sensitivity to +7 dBm Counter sensitivity to +7 dBm Counter sensitivity to +7 dBm Accuracy at input connector** (0 dBm to –20 dBm) ±1.5 dB ±1.5 dB ±1.0 dB
0.05-12.4 GHz ±1.5 dB ±1.5 dB ±1.5 dB
12.4-20 GHz N/A ±2.0 dB ±1.5 dB
26.5-46 GHz N/A N/A ±2.0 dB Resolution 0.01 dB 0.01 dB 0.01 dB Display dBm or millwatts/microwatts dBm or millwatts/microwatts dBm or millwatts/microwatts
**see graphs for typical data
5
Graph 1. –10 dBm input level at 25°C.
Graph 2. 0 dBm and –20 dBm input level at 25°C.
Typical* power measurement uncertainty at 25°C for various input levels
*Typical means approximately 2/3 of all units will meet these characteristics.
0.8
0.7
Agilent 53150A/151A Agilent 53152A
0.6
0.5
0.4
0.3
± dB Uncertainty
0.2
0.1
0
0.25 1 2 4 6 8 10 12 14 16 18 20 22 26.5 30 35 40 46
Frequency in GHz
0.8
0.7
0.6
0.5
0.4
0.3
± dB Uncertainty
0.2
0.1
Agilent 53150A/151A Agilent 53152A
0
0.25 1 2 4 6 8 10 12 14 16 18 20 22 26.5 30 35 40 46
Frequency in GHz
6
Graph 3. –25 dBm input level at 25°C.
Graph 4. –25 dBm input level from 0 to 55°C.
Typical* power measurement uncertainty at –25 dBm input level
*Typical means approximately 2/3 of all units will meet these characteristics.
1
0.9
0.8
0.7
0.6
0.5
0.4
± dB Uncertainty
0.3
0.2
0.1
0
0.25 1 2 4 6 8 10 12 14 16 18 20 22 26.5
Frequency in GHz
Agilent 53150A/151A Agilent 53152A
1.8
1.6
1.4
1.2
1.0
0.8
± dB Uncertainty
0.6
0.4
0.2
0
0.25 1 2 4 6 8 10 12 14 16 18 20 22 26.5
Agilent 53150A/151A Agilent 53152A
Frequency in GHz
Timebase
Frequency: 10 MHz
Output: 10 MHz sine wave, 1 Vp-p into 50
External timebase input: 1, 2, 5, 10 MHz;
1 to 5 Vp-p into 50
Connector: BNC female located on rear panel
Internal timebase stability
TCXO Oven
(Standard) (Option 001)
Aging rate
Per Day <5 x 10 Per Month <1 x 10–7<1.5 x 10
Short term <1 x 10–9<2 x 10 (1 sec. avg. time)
Line variation <1 x 10
–7
(±10%)
Warm-up <1 x 10
within 5 min. after turn-on
Temperature stability (0-55°C) <1 x 10–6<3 x 10
<1 x 10
at 25°C
–10
–10
–10
–8
–9
General information
Save and recall: Up to 9 complete instrument
setups may be saved and later recalled. These setups are retained when power is removed.
Sample rate: User-selectable
Fast (nominally 20 ms between readings), Medium (nominally 250 ms between readings), Slow (nominally 1 s between readings) and Hold.
Self test: Internal memory and count circuitry
automatically tested at startup, via menu selection, or remotely. Error messages
–8
displayed to indicate failed tests.
Size: 213 mm W x 88.5 mm H x 300 mm D
Operating temperature: 0-55°C
With battery option: 0-40°C
Weight: 4 kg without battery option,
6.4 kg with battery option
Warranty: 1 year
Programming: GPIB (IEEE-488.1-1987,
IEEE 488.2-1987) or RS-232C Language: SCPI-1992.0 (Standard Commands for Programmable Instruments) RS-232C rates: User-selectable 2400 to 19200 baud
Power supply:
ac: 90-132 Vac; 47.5-66 Hz or 360-440 Hz
216-264 Vac; 47.5-66 Hz line selection: automatic power requirements: 75 VA max. (25 W typ.)
dc: (Option 002 only): 11-18 Vdc; 2A max. Battery (Option 002):
Type: VHS camcorder, lead acid (2 each) Charge Time: 8 hours in unit Capacity: 2.5 hours min. at 25°C
Math functions:
Offset: Last reading and/or entered offset
to reading for either power or frequency Averaging: 1 to 99 measurement running average Cable Loss Compensation: Offsets power reading via linear interpolation of user-entered attenuations with up to 10 independent frequency points.
Display: Backlit LCD. Backlight can be
turned on or off via front panel control.
Sleep mode (Option 002 only):
Automatically activated if no input is present for 5 minutes.
Safety: Designed in compliance with
IEC-1010, CAN/CSA 1010.1
EMC: Designed in compliance with IEC-11,
EN50082-1, IEC801-2, -3, -4.
Accessories supplied
Operating, programming, and service manuals and ac power cord.
Accessories available
Battery charger 53150-60217
Spare battery 53150-80010
dc Power input cable 53150-60214
Ordering information
Agilent 53150A 20 GHz counter
Agilent 53151A 6.5 GHz counter
Agilent 53152A 46 GHz counter
Options
Opt 001 Oven Timebase
Opt 002 Battery and dc input
Opt A6J ANSI 7540 Compliant Calibration
Opt W30 Three years of Return Repair Service
7
www.agilent.com
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instru­ment and system we sell has a global warranty. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you receive your new Agilent equipment, we can help verify that it works properly and help with initial product operation.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and onsite education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
By internet, phone, or fax, get assistance with all your test & measurement needs
Online assistance: www.agilent.com/find/assist Phone or Fax
United States:
(tel) 800 829 4444 (fax) 800 829 4433
Canada:
(tel) 877 894 4414 (fax) 800 746 4866
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) (31 20) 547 2111 (fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115
Latin America:
(tel) (650) 752 5000
Taiwan:
(tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6836 0252 (e-mail) tm_asia@agilent.com
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Agilent Open Connectivity
Agilent simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent’s broad range of system-ready instruments, open industry software, PC-standard I/O and global support combine to accelerate test system development. More information is available at
www.agilent.com/find/openconnect.
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2005 Printed in the USA June 27, 2005 5966-3118EN