This guide describes how to service the Agilent 53147A, 53148A, and
53149A. The information in this guide applies to instruments having the
number prefix listed below, unless accompanied by a “Manual Updating
Changes” package indicating otherwise.
SERIAL PREFIX NUMBER:US4047 (53147A)
US4048 (53148A)
US4049 (53149A)
Agilent 53147A/148A/149A
Microwave Frequency Counter/
Power Meter/DVM
Copyright Agilent
Technologies, Inc. 2001, 2002
All Rights Reserved.
Reproduction, adaptation, or
translations without prior
written permission is
prohibited, except as allowed
under the copyright laws.
Printed: May 2002
Printed in U.S.A.
Manual part number
53147-90010
Certification and Warranty
Certification
Agilent Technologies, Inc.
certifies that this product met
its published specification at the
time of shipment from the
factory. Agilent further certifies
that its calibration measurements
are traceable to the United
States National Institute of
Standards and Technology
(formerly National Bureau of
Standards), to the extent
allowed by the Institute’s
calibration facility, and to the
calibration facilities of other
International Standards
Organization members.
Warranty
Agilent warrants Agilent
hardware, accessories and
supplies against defects in
materials and workmanship for
a period of one year from date of
shipment. If Agilent receives
notice of such defects during the
warranty period, Agilent will, at
its option, either repair or
replace products which prove to
be defective. Replacement
products may be either new or
like-new.
Agilent warrants that
Agilent software will not fail to
execute its programming
instructions, for the period
specified above, due to defects in
material and workmanship
when properly installed and
used. If Agilent receives notice
of such defects during the
warranty period, Agilent will
replace software media which
does not execute its
programming instructions due
to such defects.
For detailed warranty
information, see back matter.
Safety Considerations
General
This product and related
documentation must be
reviewed for familiarization
with this safety markings and
instructions before operation.
Before Cleaning
Disconnect the product from
operating power before cleaning.
Warning Symbols That May
Be Used In This Book
Instruction manual symbol; the
product will be marked with
this symbol when it is necessary
for the user to refer to the
instruction manual.
Indicates hazardous voltages.
Indicates earth (ground)
terminal.
or
Indicates terminal is connected
to chassis when such connection
is not apparent.
Indicates Alternating current.
Indicates Direct current.
Safety Considerations
(cont’d)
WARNING
BODILY INJURY OR DEATH
MAY RESULT FROM
FAILURE TO HEED A
WARNING. DO NOT
PROCEED BEYOND A
WARNING UNTIL THE
INDICATED CONDITIONS
ARE FULLY UNDERSTOOD
AND MET.
CAUTION
Damage to equipment, or
incorrect measurement data,
may result from failure to
heed a caution. Do not
proceed beyond a CAUTION
until the indicated conditions
are fully understood and met.
Safety Earth Ground
An uninterruptible safety earth
ground must be maintained
from the mains power source to
the product’s ground circuitry.
WARNING
WHEN MEASURING POWER
LINE SIGNALS, BE
EXTREMELY CAREFUL AND
ALWAYS USE A
STEP-DOWN ISOLATION
TRANSFORMER WHICH
OUTPUT IS COMPATIBLE
WITH THE INPUT
MEASUREMENT
CAPABILITIES OF THIS
PRODUCT. THIS PRODUCT’S
FRONT AND REAR PANELS
ARE TYPICALLY AT EARTH
GROUND. THUS, NEVER TRY
TO MEASURE AC POWER
LINE SIGNALS WITHOUT AN
ISOLATION TRANSFORMER.
For additional safety and
acoustic noise information,
see back matter.
Agilent Technologies, Inc.7.C.NL.03.22.01.R1.M.CW6FC
5301 Stevens Creek Boulevard
Santa Clara, California 95052-8059
Running the Self Test2-30
Self Test Error Messages2-31
ivAssembly Level Service Guide
Contents
3Replacing Assemblies
Introduction3-2
Tools Required3-3
Do This First3-3
Removing the Bumpers3-4
Removing the Handle3-5
Removing the Bumper Retainers3-6
Removing the Shroud3-7
Removing the Front Panel Assembly3-9
Removing the Rear Panel Assembly3-11
Removing the Cooling Fan3-14
Removing the Power Supply3-15
4Replaceable Parts
Introduction4-2
Replaceable Parts4-2
How To Order A Part4-3
Contacting Agilent Technologies4-4
Parts Identification4-5
Reference Designations4-5
Cabinet Parts and Hardware4-5
Accessories and Miscellaneous Items4-5
5Backdating
Introduction5-2
6Specifications
Introduction6-2
Measurement Specifications and Characteristics6-3
ARack Mounting the InstrumentA-2
Index
Assembly Level Service Guidev
Contents
viAssembly Level Service Guide
In This Guide
This guide provides assembly-level service information for the
Agilent 53147A, 53148A, and 53149A.
How to Use This Guide
Repair Strategy
This service guide is designed to isolate failures to the assembly-level only.
The Agilent 53147A/148A/149A instruments can be returned to Agilent for
all service work, including troubleshooting, and verifying specifications.
Contact your nearest Agilent Sales and Service Office for more details.
NOTEISD (Instrument Service Division) Emergency Response or Express
Calibration Service is available for Agilent customers in the USA.
If downtime is critical, you can receive your repaired instrument via
overnight shipment. Call 800-403-0801, and ask for Emergency Response or Express Calibration Service. When your instrument is repaired, it is
returned via overnight shipment at no extra charge.
If you decide to service the instrument yourself, use the troubleshooting
procedures in Chapter 2 (Service) and the disassembly and reassembly
procedures in Chapter 3 (Replacing Assemblies). Then use the calibration
instructions in Chapter 2 to calibrate the instrument for peakperformance operation, and finally, perform all of the performance tests in
Chapter 1 to verify that the instrument is operating to the specifications.
Assembly Level Service Guidevii
In This Guide
Instrument Identification
The instrument is identified by the serial number on the rear panel.
Agilent uses a two-part serial number with the first part (prefix)
identifying a series of instruments and the second part (suffix) identifying
a particular instrument within a series. Agilent-assigned alpha characters
before the suffix identify the country in which the instrument was
manufactured.
Instruments Covered by this Guide
This guide applies directly to Agilent 53147A, 53148A, and 53149A
instruments that have the same serial number prefix(es) shown on the
title page. If the serial number prefix of your instrument differs from that
listed on the title page of this guide, there may be differences between this
guide and your instrument.
Instruments having a higher serial prefix are covered (when required) by
one or more manual-change sheets included with this guide. If a required
change sheet is missing, contact your nearest Agilent Sales Office listed at
the back of this guide.
Assembly-Level Service Guide Organization
This Assembly-Level Service Guide consists of a table of contents,
a preface, six chapters, and an index. The page headers identify the
chapters and sections of this manual. The chapter contents are
summarized as follows:
Chapter 1 Performance Tests provides procedures that verify the
instrument operates properly and meets the Agilent 53147A/148A/149A
specifications given in Chapter 6, “Specifications,” in this guide.
Chapter 2 Service is divided into seven main sections that provide
instructions for returning the instrument to Agilent for service,
calibrating the instrument, and troubleshooting the assemblies in
the instrument.
Chapter 3 Replacing Assemblies provides procedures for replacing
defective assemblies and/or modules in the instrument.
viiiAssembly Level Service Guide
In This Guide
Chapter 4 Replaceable Parts lists the replaceable parts contained in
the instrument, and explains how to order replacement parts for your
instrument.
Chapter 5 Backdating contains information required to adapt this
manual for older instruments (to be provided when required).
Chapter 6 Specifications lists all the specifications and operating
characteristics for the Agilent 53147A/148A/149A.
How to Order Guides
The part number for this guide is listed on the Certification and Warranty
page (on the back of the title page) and on the back cover of this guide.
Assembly Level Service Guideix
In This Guide
Description of the Microwave Frequency
Counter/Power Meter/DVM
The Agilent 53147A, 53148A, and 53149A are capable of measuring
frequencies from 10 Hz to 125 MHz on Channel 1 and from 50 MHz to
20 GHz (53147A), 26.5 GHz (53148A), and 46 GHz (53149A) on Channel 2.
These instruments are also capable of measuring power (the power and
frequency ranges for power measurement are dependent on the powersensor model used) and DC voltages up to ±50 VDC. All three instruments
have a maximum frequency resolution of 1 Hz.
The Agilent 53147A, 53148A, and 53149A provide GPIB and RS-232 serial
interfaces and are suitable for field, bench-top, and ATE operation.
The basic measurement functions of the Agilent 53147A/148A/149A
include Frequency, Relative Frequency, Frequency Offset, and Power
(including Power Offset and Relative Power). All of these features are
accessible from the front panel and over the GPIB and RS-232 interfaces.
The Agilent 53147A/148A/149A includes the following additional
measurement functions and features that are designed specifically for
manufacturing and service applications:
•1, 2, 5, and 10 MHz external reference capability
•Optional high-stability oven oscillator for high-accuracy needs and
lengthened calibration cycles
•Frequency and power offset capabilities for relative measurements
•SCPI programming capability
•Battery and dc input option for operation in locations where ac
power is unavailable
•Optional soft carrying case for safe transportation and mobile use
Programmable control is performed via a GPIB or an RS-232 serial
interface. The GPIB and RS-232C ports are standard for the Agilent
53147A, 53148A, and 53149A.
xAssembly Level Service Guide
In This Guide
Options
The options available for the Agilent 53147A/148A/149A are listed
following this paragraph. Specifications for the options are listed in
Chapter 6, “Specifications.” Options ordered with the instrument are
installed at the factory and are ready for operation on delivery.
Hardware
•High Stability Oven Timebase, Option 001
•Battery/DC Power Input, Option 002
•Rack Mount Kit, Option 1CM
•Soft Carrying Case, Option 007
Retrofit
Options 001 and 002 can be installed only by authorized Agilent
Technologies Repair Centers.
Instructions for installing the Rack Mount Kit are provided in
Appendix A.
Assembly Level Service Guidexi
In This Guide
Accessories Supplied and Available
Accessories Supplied
•Power cord, 2.3 meters (Part number dependent upon destination country)
•Power sensor cable (Agilent P/N 11730A)
•DVM test leads (Agilent P/N 34132B)
Accessories Available
•Soft Carrying Case (Agilent P/N 53147-80016)
•Automotive Power Adapter (Agilent P/N 53150-60214)
•Battery (Agilent P/N 53150-80010)
•GPIB Cables (Agilent P/N 10833A/B/C/D)
•RS-232 Cable (Agilent P/N 53150-60215)
•Power Sensors (Agilent 8480 series)
Manuals Supplied
Agilent 53147A/148A/149A Operating and Programming Guide
(Agilent P/N 53147-90009)
Agilent 53147A/148A/149A Assembly-Level Service Guide
(Agilent P/N 53147-90010)
xiiAssembly Level Service Guide
1
Performance Tests
Verifying Specifications
Chapter 1 Performance Tests
Introduction
Introduction
1
This chapter provides procedures to test the electrical performance of the
Agilent 53147A, 53148A and 53149A. These procedures are based on the
specifications in Chapter 6, “Specifications.”
Two types of testing are provided:
•Operational Verification
•Complete Performance Tests
This chapter is organized as follows:
•Introductionpg. 1-2
•Equipment Requiredpg. 1-4
•Operational Verificationpg. 1-6
•Complete Performance Testspg. 1-17
•Performance Test Recordpg. 1-39
Operational Verification
Operational Verification is an abbreviated series of tests that you can
perform (instead of performing the Complete Performance Tests) to
provide a high degree of confidence that the instrument is operating
properly. Operational Verification is useful for incoming inspection,
routine maintenance, and after instrument repair.
1-2Assembly Level Service Guide
Chapter 1 Performance Tests
Introduction
Complete Performance Tests
The Complete Performance Tests verify the specifications listed in
Chapter 6, “Specifications.” All tests can be performed without opening
the instrument.
Recommended Calibration Cycle
The instrument requires periodic verification of operation. Depending on
the type of use, environmental conditions, aging, and measurement
accuracy required, the instrument should be checked using the operational
verification procedure at least once every year. A full Calibration and
Performance Test should be performed each time the instrument changes
environment or if an assembly or module has been replaced.
Test Record
The results of the Operational Verification and the Complete Performance
Tests should be recorded on a copy of the Performance Test Record, located
at the end of the Complete Performance Test section in this chapter.
1
Assembly Level Service Guide1-3
Chapter 1 Performance Tests
Equipment Required
Equipment Required
1
Table 1-1 lists the test equipment and accessories needed to perform the
tests in this chapter.
Table 1-1. Recommended Test Equipment and Accessories
Instrument TypeRequired Characteristics
Synthesized Signal
Generator
Microwave Synthesized
Signal Generator
Microwave Amplifier45 MHz to 50 GHzAgilent 83051A1P, T
Digital Multimeter6½ digit AC/DCAgilent 34401A 1OV, P
DC Power Supply25 VDCAgilent E3640A 1OV
Frequency Counter10 Hz to 125 MHzAgilent
Power MeterRange 1 mWAgilent 432A1P, C, T
Power Meter50 MHz to 50 GHz, +13 dBm to –50 dBmAgilent 437B1P, C, T
Power Meter100 kHz to 110 GHz, –70 dBm to +44 dBmAgilent E4418B1P, C, T
Thermistor MountSWR 1.05 at 50 MHz
10 Hz to 20 MHz, +7 dBm to –40 dBmAgilent 3325A/B2OV, P, T
10 MHz to 50 GHz, Accuracy >1 x 10
+7 dBm to –50 dBm
Accuracy ±0.5% at 50 MHz
-10
Model
Recommended *
Agilent 83650B1P, T
53150A/51A/52A
Agilent 478A-H75
or 478A-H76
Quantity
Required Use **
1P, C, T
1P, T
Power Sensor10 MHz to 18 GHz, -30 dBm to +20 dBmAgilent 8481A1P, T
Power Sensor50 MHz to 50 GHz, +13 dBm to –50 dBmAgilent 8487A1P, C, T
Power Sensor Cable2.4 mm to 2.4 mm, 5 ft.Agilent 11730A2
Power Sensor CableAgilent 8120-1082 1P, T
Range Calibrator3, 10, 30, 100, and 300 µW
1, 3, 10, 30, and 100 mW
* Equivalent equipment can be substituted for all instruments and accessories.
** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
† One supplied with Agilent 53147A/48A/49A
Agilent 11683A 1P
†
P, T
1-4Assembly Level Service Guide
Chapter 1 Performance Tests
Equipment Required
Table 1-1. Recommended Test Equipment and Accessories (continued)
Model
Instrument TypeRequired Characteristics
20 dB AttenuatorDC to 50 GHz
2.4 mm (m) to 2.4 mm (f)
Power SplitterDC to 50 GHz
2.4 mm (f3x)
CablesBNC (m) to BNC (m)
2.4 mm (m) to 2.4 mm (m)
AdaptersBNC (f) to Type N (m)
BNC (m) to SMA (f)
3.5 mm (f) to 2.92 mm (m)
2.92 mm (m) to 2.4 mm (m)
2.92 mm (m) to 2.4 mm (f)
2.4 mm (m) to 2.4 mm (m)
2.4 mm (f) to 2.4 mm (f)
* Equivalent equipment can be substituted for all instruments and accessories.
** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
† One supplied with Agilent 53147A/48A/49A
Operational Verification is an abbreviated series of tests that you can
perform (instead of performing the Complete Performance Tests) to
provide a high degree of confidence that the instrument is operating
properly. Operational Verification is useful for incoming inspection,
routine maintenance, and after instrument repair.
If you are not familiar with operating the instrument, you should review
the Chapter 1, “Getting Started,” in the Agilent 53147A/148A/149A Operating Guide. However, the procedures in this chapter are written
so that little experience is necessary. These procedures should be followed
in the order in which they appear.
Power-On Self Tests
NOTEThis test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Inspect the instrument for damage.
2Make sure no cables are connected to the instrument’s inputs.
3Connect the power cord to the instrument and the power source.
NOTES•Since the instrument’s power supply automatically senses the line
voltage, there is no AC input-voltage setting.
•It is normal for the fan in the instrument to run when the instrument is
in Standby mode. Power is supplied to the timebase whenever the
power cord is connected to maintain long-term measurement reliability,
and the fan helps to maintain the timebase’s temperature stability.
4Press and release the Power button on the front panel.
1-6Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
5Verify that the front-panel display shows the following:
•All segments of the front-panel display are temporarily activated.
•TESTING is displayed.
•SELF TEST OK is displayed.
•The model number of the instrument is displayed (a four-digit hex
number is also displayed).
•GPIB ADDR nn (nn = a two digit number from 0 to 30) is displayed.
•CH2 NO SIGNAL is displayed.
6If an error message is displayed, refer to the troubleshooting section in
Chapter 2, “Service.”
7Mark Pass or Fail in the Performance Test Record on page 1-39, Test 1.
NOTETo ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment and the
timebase to stabilize.
1
Assembly Level Service Guide1-7
Chapter 1 Performance Tests
Operational Verification
10 MHz Test
1Connect a BNC-to-BNC cable from the Reference 10MHz connector on
the instrument’s rear panel to the CHANNEL 1 input connector on the
1
front panel.
2Press and release the Chan Select key.
3Verify that Ch 1 is displayed in the upper-left corner of the display.
4Verify that the display reads 10,000,000 Hz.
5Mark Pass or Fail in the Performance Test Record on page 1-39, Test 2.
1-8Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
Run Self Test
NOTEThis test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Disconnect all signal cables from the input and Reference 10MHz connectors.
2Press and release the Shift key, and then press and release the Menu
(Reset/Local) key.
3Press the up- and/or down-arrow key(s) as many times as necessary until
DO SELF TEST is displayed.
4Press and release the Enter key.
5If no errors are detected, SELF TEST OK is briefly displayed when the self
test is completed. If any error messages are displayed, refer to the
troubleshooting section in Chapter 2, “Service.”
6Mark Pass or Fail in the Performance Test Record on page 1-39, Test 3.
1
Assembly Level Service Guide1-9
Chapter 1 Performance Tests
Operational Verification
External Timebase Test
This test verifies the instrument’s 10 MHz external timebase specification
by verifying that setting REF OSC to EXT allows the instrument to be
1
NOTES•To ensure that the test results are valid, the instrument and the test
synchronized to an external reference signal.
Equipment Required
Agilent 3325A/B Synthesizer (2)
Agilent 10100C 50Ω Feedthrough Terminator
Agilent 10503 Series Coaxial Cables (BNC m to m) (2)
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect an Agilent 10503 series coaxial cable between the output of the
first Agilent 3325A/B Synthesizer and the Reference 10MHz connector on
the instrument’s rear panel (see Figure 1-1).
2Connect an Agilent 10100C 50Ω Feedthrough Terminator to the Counter’s
Channel 1 input connector.
3Connect an Agilent 10503 series coaxial cable between the output of the
second Synthesizer and the feedthrough terminator on the Counter’s
CHANNEL 1 input connector.
1-10Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
To Reference
10 MHz Connector
(Rear Panel)
Agilent 3325A/B Synthesizer
1
10 MHZ Output
Agilent
53147A/48A/49A
Agilent
Channel 1 (with 10100C 50Ω Feedthrough Attached)
Agilent 3325A/B Synthesizer
2 MHz Output
Figure 1-1. External Timebase Test Setup
4Cycle the POWER button to preset the instrument.
5Press the Shift key on the instrument’s front panel.
6Press the Menu key on the instrument’s front panel.
7If necessary, use the up- and down-arrow keys to cycle the display until
it reads REF OSC> INT.
8Press the right-arrow key once. INT begins to flash off and on.
9Press the up- or down-arrow key once. INT changes to EXT.
10Press the Enter key.
11Press the Chan Select key once to select the CHANNEL 1 input connector.
Assembly Level Service Guide1-11
Chapter 1 Performance Tests
Operational Verification
12Set the frequency of the first Synthesizer (connected to the instrument’s
rear-panel Reference 10MHz connector) to 10 MHz at 1 V rms.
1
13Set the frequency of the second Synthesizer (connected to the instrument’s
Channel 1 input connector) to 2 MHz at 100 mV rms.
14Press and release the instrument’s Reset/Local key.
15Verify that the Counter displays a reading of approximately 2 MHz and
that the Ext Ref annunciator is activated.
16Mark Pass or Fail for Test 4 in the Performance Test Record on page 1-39.
17Disconnect the test setup.
1-12Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
Power Meter Test
This test verifies the accuracy of the Power Meter.
Equipment Required
Agilent 11730A Power Sensor Cable (supplied with instrument)
Agilent 8481A, 8481D, 8482A, 8485A, or 8487A Power Sensor Head
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect one end of the Agilent 11730A cable (supplied with instrument)
to the Power Meter INPUT connector on the front panel of the Agilent
53147A, 53148A, or 53149A.
1
2Connect the other end of the Agilent 11730A cable to a power sensor head.
3Press the Display Power key in the Power Meter area of the front panel.
The display should look like this (the reading will vary):
Ch 2
Freq
dB
dBm
Pwr
Watts
mW
uW
%
4If the power sensor head you are using is an Agilent 8481A, 8481D,
8482A, 8485A, or 8487A, select the power sensor head model in the HEAD
menu. If you are not using one of the power sensor heads listed above,
press the Cal Factor key, and enter the calibration factor for the head.
Assembly Level Service Guide1-13
Chapter 1 Performance Tests
Operational Verification
5Press the Zero key.
The display shows ZEROING for several seconds.
1
6Connect the power sensor head to the Power Meter OUTPUT connector.
7Press the Cal key.
The display briefly shows CALIBRATING.
8Turn on the power reference output in the instrument’s menu
(PWR REF > ON).
9Verify that the Counter displays a reading of 0 dBm ±.02 dB.
10Mark Pass or Fail for Test 5 in the Performance Test Record on page 1-39.
11Disconnect the test setup.
1-14Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
DVM Test
This test verifies the accuracy of the DVM.
Equipment Required
Agilent 34401A Digital Multimeter
Agilent E3640A DC Power Supply
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1Set the output of the E3640A DC Power Supply to +25 VDC.
2Set the Agilent 34401A Digital Multimeter to the lowest DC voltage range
capable of safely measuring 25 VDC.
1
3Connect the test leads to the Multimeter (observe correct polarity).
4Connect the test prods to the output terminals on the Agilent E3640A DC
Power Supply (observe correct polarity).
Agilent E3640A
Power Supply
Agilent
Agilent 34401A
Digital Multimeter
Agilent
Agilent
53147A/48A/49A
Agilent
Agilent E3640A
Power Supply
Agilent
Figure 1-2. DVM Test Setup
Assembly Level Service Guide1-15
Chapter 1 Performance Tests
Operational Verification
5Observe and record the voltage reading on the Agilent 34401A Multimeter.
6Disconnect the test leads from the Agilent E3640A DC Power Supply.
1
Be careful not to disturb the voltage setting.
7Connect the test leads to the Agilent 53147A/48A/49A (be sure to observe
the correct polarity).
8Press the Display DVM key on the Agilent 53147A/48A/49A front panel.
9Connect the test prods to the output terminals on the Agilent E3640A
DC Power Supply (be sure to observe the correct polarity).
10Observe and record the voltage reading on the Agilent 53147A/48A/49A.
11Compare the voltage readings you recorded in steps 5 and 10. If the
reading taken with the Agilent 53147A/48A/49A is within ±0.25% ±10 mV
of the reading taken with the Agilent 34401A, mark Pass for Test 6 in the
Performance Test Record on page 1-39. If the reading taken with the
Agilent 53147A/48A/49A is not within ±0.25% ±10 mV of the reading
taken with the Agilent 34401A, mark Fail for Test 6 in the Performance
Test Record on page 1-39.
This completes the Operational Verification.
1-16Assembly Level Service Guide
Chapter 1 Performance Tests
Complete Performance Tests
Complete Performance Tests
The Complete Performance Tests verify the specifications of the
Agilent 53147A/148A/149A listed in Chapter 6, “Specifications.” All of
these tests can be performed without opening the instrument. Table 1-2
lists a summary of the performance tests.
Record the results of the performance tests in the appropriate place on the
Performance Test Record, which starts on page 1-39.
NOTETo ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment and the
timebase to stabilize.
Table 1-2. Complete Performance Tests
Page NumberTest Description
Page 1-18Test 1: Channel 1 Frequency Sensitivity
Page 1-23Test 2: Channel 2 Frequency Sensitivity
Page 1-27Test 3: Power Measurement
Page 1-30Test 4: Power Reference Oscillator Frequency
Page 1-32Test 5: Power Meter Reference Level
1
NOTEOther instrument measurement functions (e.g., Averaging) are
mathematically derived by the microprocessor from the parameters
verified by these performance tests. If the instrument passes the
performance tests, the other measurement functions are also functioning
to specifications.
Assembly Level Service Guide1-17
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
Test 1: Channel 1 Frequency Sensitivity
1
This set of tests verifies the frequency-sensitivity specifications of
Channel 1 of the Agilent 53147A/148A/149A’s Frequency Counter.
Equipment Required
Agilent 3325A/B Synthesizer
Agilent 83650B Synthesizer
Agilent 437B Power Meter
Agilent 8487A/D Power Sensor
Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A)
Agilent 11667C Power Splitter
Agilent 8490D (Opt. 020) 20 dB Attenuator
Agilent 1250-2015 BNC (m) to SMA (f) Adapter
Agilent 11904A 2.92 mm (m) to 2.4 mm (m) Adapter
Agilent 10100C 50 Ω Feedthrough Terminator
Agilent 10503 Series Coaxial Cable (BNC m to m)
2 mm (m) to 2 mm (f) Cable (2)
Test 1a: 10 Hz to 20 MHz Frequency Sensitivity
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
•If the Agilent 3325A/B Synthesizer or the Agilent 83650B Synthesizer
reference output is not adequate to drive the 53147A/48A/49A, connect
the cable from the Synthesizer to the Reference 10 MHz connector on
the back of the 53147A/48A/49A, and set the 53147A/48A/49A to use
the internal oscillator.
1-18Assembly Level Service Guide
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
1Connect an Agilent 10503 Series Coaxial Cable between the reference
output connector on the Agilent 3325A/B Synthesizer and the
instrument’s Reference 10MHz connector.
2Connect an Agilent 10100C 50Ω Feedthrough Terminator to the Counter’s
CHANNEL 1 connector.
3Connect an Agilent 10503 Series Coaxial Cable between the RF output
connector on the Agilent 3325A/B Synthesizer and the 50Ω feedthrough
terminator on the Counter’s CHANNEL 1 connector.
1
To Reference
Output
(Rear Panel)
Agilent 3325A/B Synthesizer
To Reference
Output
(Rear
Agilent
53147A/48A/49A
Agilent
Panel)
Channel 1
Output
Figure 1-3. 10 Hz to 20 MHz Frequency Sensitivity Test Setup
4Verify that the Counter is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-11).
5Set the output of the Synthesizer to the first frequency specified for
Test 1a in the Performance Test Record on page 1-39 (10 Hz).
6Set the power output of the Synthesizer to the first power value specified
for Test 1a in the Performance Test Record on page 1-39
(–14.9 dBm/40 mV).
7Record the frequency value read on the Counter (±1 count) for Test 1a in
the Performance Test Record on page 1-39.
Assembly Level Service Guide1-19
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
8Repeat steps 5, 6, and 7 for each of the remaining frequencies specified for
Test 1a. Don’t forget to change the power level when performing the
1 KHz test (–19.2 dBm/25 mV).
1
9If all of the frequency values you recorded for Test 1a are correct, mark
Pass in the Performance Test Record for Test 1a. If any of the frequency
readings you recorded are incorrect, mark Fail in the Performance
Test Record.
NOTES•Do not disconnect the cable from the instrument’s Reference 10MHz
connector. This connection is used in the following test.
•Do not turn off the instrument. Turning the instrument off and back on
resets the reference oscillator (REF OSC) selection to internal (INT)
and the channel selection to CHANNEL 2.
Test 1b: 50 MHz to 125 MHz Frequency Sensitivity
1Connect an Agilent 10100C 50 Ω Feedthrough Terminator to the
CHANNEL 1 input connector on the instrument’s front panel (refer to
Figure 1-4 for steps 1 through 9).
Agilent 83650B Synthesizer
Reference
Output
(Rear Panel)
Agilent 437B Power Meter
Agilent
53147A/48A/49A
Agilent
To Reference 10 MHz
Connector
(Rear Panel)
Agilent
Feedthrough
Channel 1
RF Output
10100C 50
Agilent
Agilent
Agilent
1250-2015 BNC to SMA Adapter
8490D 20 dB Attenuator
11904A 2.92 mm to 2.4 mm Adapter
Agilent
11667C
Power Splitter
Agilent
Power Sensor
Input
8487A/D
Figure 1-4. 50 MHz to 125 MHz Frequency Sensitivity Test Setup
1-20Assembly Level Service Guide
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
2Connect the Agilent 1250-2015 BNC (m) to SMA (f) Adapter to the
Agilent 10100C Feedthrough Terminator on the Counter’s CHANNEL 1
connector.
3Connect the Agilent 11904A 2.92 mm (m) to 2.4 mm (m) Adapter to the
Agilent 1250-2015 Adapter.
4Connect the Agilent 8490D 20 dB Attenuator to the Agilent 11904A
Adapter.
5Connect one of the output connectors on the Agilent 11667C Power
Splitter to the Agilent 8490D Attenuator
6Connect the Agilent 11900B 2.4 mm (f) to 2.4 mm (f) Adapter to the RF
output of the Synthesizer.
7Connect a 2.4 mm (m) to 2.4 mm (m) cable between the Adapter on the
Synthesizer’s RF output connector and the input connector on the
Agilent 11667C Power Splitter.
1
8Connect the Agilent 8487A/D Power Sensor to the remaining output
connector on the Agilent 11667C Power Splitter.
9Connect the Agilent 11730A Power Sensor Cable between the Power
Sensor and the sensor connector on the Agilent 437B Power Meter.
10Verify that the Counter is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-11).
11Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 1b in the Performance Test Record on page 1-39
(50 MHz).
12Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 1b in the
Performance Test Record on page 1-39 (+.8 dBm). (The added 20 dBm
compensates for the 20 dB Attenuator.)
Assembly Level Service Guide1-21
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
13Record the frequency value read on the Counter (±1 count) for Test 1b in
the Performance Test Record on page 1-39.
1
NOTES•Do not disconnect the rest of the test setup, as these connections are
14Set the output of the Synthesizer to 125 MHz, and repeat steps 11
through 13.
15If both of the frequency values you recorded for Test 1b are correct, mark
Pass in the Performance Test Record for Test 1b. If either of the frequency
readings you recorded are incorrect, mark Fail in the Performance Test
Record.
16Disconnect the Agilent 10100C Feedthrough Terminator, the
Agilent 1250-2015 Adapter, and the Agilent 11904A Adapter from the
Counter’s CHANNEL 1 connector.
also used in the Channel 2 Frequency Sensitivity test.
•Do not turn off the instrument. Turning the instrument off and back on
resets the reference oscillator (REF OSC) selection to internal (INT).
1-22Assembly Level Service Guide
Chapter 1 Performance Tests
Test 2: Channel 2 Frequency Sensitivity
Test 2: Channel 2 Frequency Sensitivity
This set of tests verifies the frequency-sensitivity specifications of
Channel 2 of the Agilent 53147A/148A/149A’s Frequency Counter.
Equipment Required
Agilent 83650B Synthesizer
Agilent 437B Power Meter
Agilent 8487A/D Power Sensor
Agilent 11667C Power Splitter
Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A)
Agilent 8490D (Opt. 020) 20 dB Attenuator
Agilent 11904D 2.92 mm (m) to 2.4 mm (f) Adapter
Agilent 10503 series Coaxial Cable (BNC m to m)
2.4 mm (m) to 2.4 mm (m) Cable (2)
Test 2a: 50 MHz to 20 GHz Frequency Sensitivity
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect the Agilent 11904D 2.92 mm (m) to 2.4 mm (f) Adapter to the
Counter’s CHANNEL 2 input connector (refer to Figure 1-5 for steps 1
through 4).
1
2Connect the Agilent 8490D Attenuator to the Agilent 11904D Adapter,
as shown in Figure 1-5 (the other end of the Attenuator should remain
connected to the Agilent 11667C Power Splitter).
3Verify that the cable (with Adapter) is still connected between the
RF output of the Agilent 83650B Synthesizer and the Power Splitter
input connector.
Assembly Level Service Guide1-23
Chapter 1 Performance Tests
Test 2: Channel 2 Frequency Sensitivity
4Verify that the Power Sensor is still connected to the Power Splitter and
that the Power Sensor Cable is still connected between the Power Sensor
and the sensor connector on the Power Meter.
1
Agilent 83650B Synthesizer
To Reference
Output
(Rear Panel)
Agilent
53147A/48A/49A
Agilent
To Reference 10 MHz
Connector
(Rear Panel)
Channel 2
RF Output
HP 8490D 20 dB Attenuator
(50 MHz to 40 GHz tests only)
HP 11904D Adapter
Figure 1-5. 50 MHz to 20 GHz Frequency Sensitivity Test Setup
5Verify that the instrument is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-11).
6Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 2a in the Performance Test Record on page 1-40
(50 MHz).
Agilent 437B Power Meter
Input
HP 11667C Power Splitter
HP 8487A/D Power Sensor
7Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 2a in the
Performance Test Record on page 1-40 (0 dBm). (The added 20 dBm
compensates for the 20 dB Attenuator.)
8Record the frequency value read on the Counter (±1 count) for Test 2a in
the Performance Test Record on page 1-40.
1-24Assembly Level Service Guide
Chapter 1 Performance Tests
Test 2: Channel 2 Frequency Sensitivity
9Repeat steps 6, through 8 for each of the remaining frequencies specified
for Test 2a. Don’t forget to change the power level when performing the
300 MHz and 19 GHz tests (53147A, 53148A, and 53149A) and the
16 GHz and 19 GHz tests (53149A only).
NOTEIt may be necessary to repeat Step 5.
10If all of the frequency values you recorded for Test 2a are correct, mark
Pass in the Performance Test Record. If any of the frequency readings you
recorded are incorrect, mark Fail in the Performance Test Record.
Test 2b: 22 GHz to 26.5 GHz Frequency Sensitivity
NOTEThis test is appropriate for the Agilent 53148A and 53149A.
1Use the same test setup as described for Test 2a on page 1-23.
2Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 2b in the Performance Test Record on page 1-40
(22 GHz).
1
3Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 2b in the
Performance Test Record on page 1-40 (–5 dBm for the Agilent 53148A;
–7 dBm for the Agilent 53149A). (The added 20 dBm compensates for the
20 dB Attenuator.)
4Record the frequency value read on the Counter (±1 count) for Test 2b in
the Performance Test Record on page 1-40.
5Repeat steps 2, through 4 for each of the remaining frequencies specified
for Test 2b.
6If all of the frequency values you recorded for Test 2b are correct, mark
Pass in the Performance Test Record for Test 2b. If any of the frequency
readings you recorded are incorrect, mark Fail in the Performance Test
Record.
Assembly Level Service Guide1-25
Chapter 1 Performance Tests
Test 2: Channel 2 Frequency Sensitivity
Test 2c: 30 GHz to 46 GHz Frequency Sensitivity
NOTEThis test is appropriate for the 53149A only.
1
1Use the same test setup as described for Test 2a on page 1-23.
2Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 2c in the Performance Test Record on page 1-40
(30 GHz).
3Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 2c in the
Performance Test Record on page 1-40 (–3 dBm). (The added 20 dBm
compensates for the 20 dB Attenuator.)
4Record the frequency value read on the Counter (±1 count) for Test 2c in
the Performance Test Record on page 1-40.
5Repeat steps 2, through 4 for each of the remaining frequencies specified
for Test 2c. Don’t forget to change the power level for the 42, 44, and
46 GHz tests.
6If all of the frequency values you recorded for Test 2c are correct, mark
Pass in the Performance Test Record for Test 2c. If any of the frequency
readings you recorded are incorrect, mark Fail in the Performance Test
Record.
NOTES•Do not disconnect the Power Sensor or the Power Sensor Cable from
the power meter or the Power Splitter. These connections are reused in
the following tests.
•Do not turn off the instrument. Turning the instrument off and back on
resets the reference oscillator (REF OSC) selection to internal (INT).
1-26Assembly Level Service Guide
Chapter 1 Performance Tests
Test 3: Power Measurement
Test 3: Power Measurement
This set of tests verifies the power-measurement specifications of
Agilent 53147A/148A/149A’s Power Meter.
Equipment Required
Agilent 11683A Range Calibrator
Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A)
Power Measurement Accuracy Test
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect the Agilent 11730A Power Sensor Cable between Agilent 11683A
Range Calibrator’s output connector and the Agilent 53147A, 53148A, or
53149A’s Power Meter INPUT connector.
Agilent
53147A/48A/49A
Agilent
Agilent 11683A
Range Calibrator
1
Power Input
Output
Figure 1-6. Power Measurement Accuracy Test Setup
2Press the Reset/Local key on the Agilent 53147A, 53148A, or 53149A.
Assembly Level Service Guide1-27
Chapter 1 Performance Tests
Test 3: Power Measurement
3Press the Shift + dBm/W key on the Agilent 53147A, 53148A, or 53149A to
set the Power Meter to display readings in Watts, mW, and/or µW.
1
NOTEThe Range Calibrator output level is adjustable in 5 dB increments.
5Press the Zero key on the Agilent 53147A, 53148A, or 53149A. When the
reading appears, verify that it is 0.00 ±0.06 µW.
6Change the Range Calibrator’s FUNCTION switch to CALIBRATE.
7Press the Cal key on the Agilent 53147A, 53148A, or 53149A. Wait for the
Power Meter to complete calibration.
8Verify that the Agilent 53147A, 53148A, or 53149A display reads
1.000 ±0.006 mW.
Thus, the 3 µW, 30 µW, 300 µW, 3 mW, and 30 mW legends on the RANGE
switch are approximations. The true values for these settings are 3.16 µW,
31.6 µW, 316 µW, 3.16 mW, and 31.6 mW.
9Set the Range Calibrator RANGE switch to each of the positions shown in
Table 1-3. For each setting, verify that the Agilent 53147A, 53148A, or
53149A autoranges properly and that the reading displayed is within the
limits shown in the table.
NOTES•The lower the setting of the RANGE switch, the longer it takes for
the reading to stabilize.
•You may have to repeat Step 5 (re-zero the instrument) for each
measurement.
1-28Assembly Level Service Guide
Chapter 1 Performance Tests
Test 3: Power Measurement
Table 1-3. Range Calibrator Settings and Expected Results
Range CalibratorAgilent 53147A/48A/49A
RANGE Switch SettingMinimumActualMaximum
3 µW3.10 µW3.23 µW
10 µW9.90 µW10.10 µW
30 µW31.4 µW31.8 µW
100 µW99.5 µW100.5 µW
300 µW0.314 mW0.318 mW
1 mW0.995 mW1.005 mW
3 mW3.14 mW3.18 mW
10 mW9.95 mW10.05 mW
30 mW31.4 mW31.8 mW
100 mW99.5 mW100.5 mW
1
10Record the actual power readings in the Performance Test Record on
page 1-41.
11Press the Shift + dBm/W key on the Agilent 53147A, 53148A, or 53149A
again to set the Power Meter to display readings in dBm.
12Verify that the annunciator changes to dBm and that the indication is
within 20.00 ±00.04 dBm with the Range Calibrator’s RANGE switch in
the 100 mw / 20 dBm position.
13Record the actual power reading in the Performance Test Record.
14Set the Range Calibrator’s RANGE switch to –10 dBm.
15Verify that the Power Meter displays a reading of –10.00 ±0.04 dBm.
16Record the actual power reading in the Performance Test Record.
17Review all of the results for Test 3, and mark Pass or Fail for Test 3 in the
Performance Test Record.
Assembly Level Service Guide1-29
Chapter 1 Performance Tests
Test 4: Power Reference Oscillator Frequency
Test 4: Power Reference Oscillator Frequency
1
The internal power reference oscillator is used to calibrate the Power
Meter to the power sensor head. This test verifies that the frequency
output of the power reference oscillator is 50.0 MHz ±0.5 MHz.
Equipment Required
Agilent 53150A Microwave Frequency Counter
Agilent 10503 Series Coaxial Cables (BNC m to m)
Agilent 1250-1535 BNC (f) to Type N (m) Adapter
Power Reference Oscillator Frequency Test
NOTES•The instrument and the test equipment must be powered on for at least
30 minutes prior to measuring the power reference oscillator output.
This allows the internal temperatures of the equipment to stabilize.
•This procedure specifies the Agilent 53150A for measuring the
frequency output of the 53147A/48A/49A’s power reference oscillator.
If the 53147A/48A/49A’s Frequency Counter is known to be in calibration
(see “Test 1: Channel 1 Frequency Sensitivity” on page 1-18), you can
use it in place of the 53150A in this procedure.
1Connect an Agilent E9635A adapter to the Agilent 53147A/48A/49A’s
Power Meter OUTPUT connector as shown in Figure 1-7.
2Connect an Agilent 10503 series cable between the Agilent E9635A
adapter on the Agilent 53147A/148A/149A’s Power Meter OUTPUT
connector and the CHANNEL 1 input connector on the Agilent
53150A/51A/52A Frequency Counter.
1-30Assembly Level Service Guide
Chapter 1 Performance Tests
Test 4: Power Reference Oscillator Frequency
Agilent E9635A Adapter on
Power Reference Oscillator
Output Connector
Agilent
53147A/48A/49A
Agilent
Agilent
53150A/51A/52A
Microwave Frequency Counter
53150A
Agilent
POWER
Rel Freq
Avg On
Rel Pwr
Standby
Freq
Menu
Offset
Rate
Reset/
Local
On/Off
Shift
Clear
GHzMHzkHzHz
Ch 12
Offset
Offset
MODIFY
Pwr
HPIB
Offset
Resol
Avg
Enter
+/-
20 GHz Counter
CHANNEL 1
10 Hz to 125 MHz
1M Ω
Ext Rel
Rate Rmt SRQ
Error
Shift
Gate
Chan
Select
Rel Freq
Offset
On/Off
Hold
DAMAGE
+30 dBm
CHANNEL 2POWERFREQ
Channel 2 50 MHz to 20 GHz
dBm/ W
DAMAGE
+27 dBm
Display
Power
Rel Pwr
Offset
On/Off
50 Ω
dB
dBm
Watts
mW
uW
%
Channel 1
Input
Figure 1-7. Reference Oscillator Frequency Adjustment Setup
3Set the Agilent 53150A/51A/52A Frequency Counter to measure frequency
on Channel 1.
4Turn on the Agilent 53147A/48A/49A’s internal power reference oscillator:
1
Press Shift + Menu, use the up and/or down arrow keys to cycle through
the menu to PWR REF, press the right-arrow key, then use the up or
down arrow key to change the setting to ON, and press the Enter key.
The Pwr Ref annunciator should now be activated.
NOTEThe reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete.
Setting PWR REF in the 53147A/148A/149A’s menu to ON turns the power
reference oscillator on, and it remains on until the PWR REF menu setting
is returned to OFF. When you calibrate the Power Meter with PWR REF
set to ON, the power reference oscillator remains on after calibration is
complete.
5Record the power reading shown on the Agilent 53150A/51A/52A for
Test 4 in the Performance Test Record on page 1-41.
6Mark Pass or Fail for Test 4 in the Performance Test Record.
Assembly Level Service Guide1-31
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
Test 5: Power Meter Reference Level
1
Two methods for verifying the level of the Power Meter reference signal
are provided in this section. Both methods provide acceptable results, and
either one can be used.
Power Meter Reference Level: Method 1
Equipment Required
Agilent E4418B Power Meter
Agilent 8481A Power Sensor
Agilent 11730A Power Sensor Cable (supplied with 53147A/48A/49A)
NOTES•To ensure that the test results are valid, the Agilent 53147A/48A/49A
and the test equipment should be powered on for at least 30 minutes
prior to beginning the tests. This allows the internal temperatures of
the equipment to stabilize.
•The calibration of the Agilent E4418B Power Meter must be no more
than one level removed from NIST (National Institute of Standards
and Technology). If the E4418B Power Meter’s calibration is more than
one level removed from NIST, the test results will be invalid.
•This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect the Agilent 11730A Power Sensor Cable to the Agilent 8481A
Power Sensor (see Figure 1-8).
Output connector (with Agilent 8481A
Agilent E4418B
Power Meter
Figure 1-8. E4418B Power Meter Zeroing and Calibration Test Setup
1-32Assembly Level Service Guide
Power Sensor Attached)
Input
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
2Connect the free end of the Agilent 11730A Power Sensor Cable to the
input connector on the Agilent E4418B Power Meter.
3Connect the Agilent 8481A Power Sensor to the power reference output
connector of the Agilent E4418B Power Meter.
4Zero the E4418B Power Meter.
5Calibrate the E4418B Power Meter.
6Disconnect the Agilent 8481A Power Sensor from the Agilent E4418B
Power Meter, and immediately connect it to the Power OUTPUT connector
on the Agilent 53147A/48A/49A (see Figure 1-9).
Output connector (with Agilent 8481A
Power Sensor Attached)
Agilent
53147A/48A/49A
Agilent
Agilent E4418B
Power Meter
Input
Figure 1-9. Power Meter Reference Level Test Setup
7Record the power reading shown on the Agilent E4418B Power Meter for
Test 5 in the Performance Test Record on page 1-41.
1
8Mark Pass or Fail for Test 5 in the Performance Test Record.
Assembly Level Service Guide1-33
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
Power Meter Reference Level: Method 2
Equipment Required
1
Agilent 432A Power Meter
Agilent 478A-H75 or 478A-H76 Thermistor Mount
Agilent 34401A Digital Multimeter
Agilent 8120-1082 Power Sensor Cable
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Turn the Agilent 432A Power Meter off. Do not turn it on until instructed
to do so.
2Set the Agilent 34401A Digital Multimeter to measure resistance, and
connect the DVM test prods between the center pin of the V
RF
on the Agilent 432A Power Meter rear panel and pin 1 on the thermistormount end of the Agilent 8120-1082 Power Sensor Cable (see Figure 1-10
to locate pin 1).
2
1
connector
6
3
5
4
Figure 1-10. Agilent 8120-1082 Power Sensor Cable End View
3Round off the resistance reading to two decimal places and keep a record
of this value as R. It is the internal bridge resistance (R) of the Agilent
432A Power Meter (this should be approximately 200.3 ohms).
1-34Assembly Level Service Guide
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
4Connect the Agilent 478A-H75 or 478A-H76 Thermistor Mount to the
Agilent 53147A/148A/149A Power Meter OUTPUT connector, as shown in
Figure 1-11.
NOTEDo not use a cable between the Thermistor Mount and the
53147A/148A/149A Power Meter OUTPUT connector. Connect the
Thermistor Mount directly to the OUTPUT connector.
Agilent 478A-H75 or
478A-H76 Thermistor Mount
(connected to Power
Meter OUTPUT connector)
Agilent
53147A/48A/49A
Agilent
Agilent 8120-1082
Power Sensor Cable
Agilent 432A
Power Meter
V
RF
V
COMP
Agilent 34401A
Digital Multimeter
-
+
Agilent
1
+
-
Figure 1-11. Power Meter Reference Level Test Setup
5Connect the Agilent 8120-1082 Power Sensor Cable between the Thermistor
Mount and the input connector on the Agilent 432A Power Meter.
6Turn on the 432A Power Meter. Wait 30 minutes for the Power Meter and
the Thermistor Mount to stabilize before proceeding to the next step.
7Set the 432A Power Meter’s RANGE switch to Coarse Zero.
8Adjust the COARSE ZERO control on the 432A Power Meter’s front panel
to obtain a zero reading on the meter.
9Set the 432A Power Meter’s RANGE switch to .01 mW / –20 dBm.
10Press the FINE ZERO switch on the 432A Power Meter’s front panel down
to obtain a zero reading on the meter.
11Set the 432A Power Meter’s RANGE switch to 1 mW.
Assembly Level Service Guide1-35
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
12Set the Digital Multimeter to measure microvolts.
CAUTIONEnsure that the Agilent 34401A Digital Multimeter’s input leads are
1
isolated from chassis ground before performing the next step.
13Connect the positive test prod from the Agilent 34401A Multimeter to
the V
connector on the Agilent 432A Power Meter rear panel.
COMP
NOTEThe V
and VRF connectors are BNC female. You can use two BNC-to-
COMP
banana jack adapters and two BNC cables to make it easier to connect the
Multimeter to the Power Meter.
14Connect the negative test prod from the Agilent 34401A Digital
Multimeter to the V
connector on the 432A Power Meter’s rear panel.
RF
15If the reading on the Multimeter is less than 400 microvolts, proceed to
step 16. If the reading is 400 microvolts or more, press the FINE ZERO
toggle switch on the 432A Power Meter’s front panel down.
16Observe the reading on the Multimeter, round off the reading to the
nearest microvolt, and keep a record of this value as V
.
0
17Turn on the Agilent 53147A/148A/149A’s internal power reference
oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through
the menu to PWR REF, press the right-arrow key, then use the up or down
arrow key to change the setting to ON, and press the Enter key. The Pwr Ref annunciator should now be activated.
NOTEThe reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete.
When PWR REF in the 53147A/148A/149A’s menu is set to ON, the power
reference oscillator is turned on, and it is left on after calibration.
18Observe the reading on the Multimeter, and keep a record of this value
.
as V
1
1-36Assembly Level Service Guide
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
19Disconnect the Multimeter’s negative test lead from the 432A Power Meter’s
V
connector, and connect it to the 432A Power Meter’s chassis ground.
RF
20Observe the reading on the Multimeter, and keep a record of this value
as V
COMP
.
21Calculate the output level of the 53147A/148A/149A’s power reference
oscillator (P
) using the following formula:
rf
Prf =
2 V
( V1 – V0 ) + V
comp
4R (Calibration Factor)
0
2
– V
2
1
Where:
= Power reference oscillator output level
P
rf
V
comp
= previously recorded value
= previously recorded value
V
1
V
= previously recorded value
0
R = previously recorded value
Calibration Factor = value for thermistor mount at 50 MHz (traceable to NIST)
1
Assembly Level Service Guide1-37
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
NOTEThe following is an example showing how a calculation of the Power
Reference Output Level is performed:
1
V0=170 µV or 170 x 10–6 V
= 78,107 µV or 78,107 x 10–6 or .078 V
V
1
V
comp
= 5.1583 V
R = 200
Ω
R = previously recorded value
Calibration Factor = 99.68%
2
– V
2
1
Note that V
Prf =
2 V
is eventually dropped out of the equation in the following
0
( V1 – V0 ) + V
comp
0
4R (Calibration Factor)
steps, because its value is very small (effectively zero).
2
P
=
rf
4(200) (99.68%)
0.80469 – 0.00608
=
2 (5.1583) (.078 – 0)
+ (.000170)2 – (0.78)
0.79861
=
797.44797.44
1.00147 mW
=
22Verify that Prf is between 0.988 mW and 1.012 mW.
23Record the value for P
in the Performance Test Record on page 1-41.
rf
24Mark Pass or Fail for Test 5 in the Performance Test Record.
1-38Assembly Level Service Guide
Chapter 1 Performance Tests
Performance Test Record (Page 1 of 3)
Performance Test Record (Page 1 of 3)
Agilent 53147A/148A/149A Frequency Counter/Power Meter/DVM
Serial Number:
Test Performed By:
Date:
Notes:
Pre Calibration Test ❏Post Calibration Test❏
Test
Number
1Power on Self TestsN/A
210 MHz Test10,000,000 Hz
3Run Self TestN/A
410 MHz External Timebase Input Test2 MHz
5Power Meter Test0 dBm ±.02 dB
6DVM Test+25 Vdc ±0.25% ±10 mV
Test
Number
1a10 Hz
Operational VerificationSpecifications
Complete
Performance TestsSpecifications
Channel 1 Frequency Sensitivity
30 Hz
1 kHz
500 kHz
5 MHz
10 MHz
20 MHz
Repair/Work Order No.
Temperature:
Relative Humidity:
This chapter provides service information for your Agilent 53147A/
148A/149A. It is divided into four major sections:
•Returning the Instrument to Agilent Technologies for Service
(page 2-3). This section provides you with step-by-step instructions on
how to return the instrument for service.
•Calibration Procedures (page 2-6). This section provides step-by-
2
step procedures for calibrating the Agilent 53147A/148A/149A.
•Pre-Troubleshooting Information (page 2-23). This section
provides you with pertinent information such as safety considerations,
recommended test equipment, repair and after-service considerations,
and assembly identification and location.
•Troubleshooting the Instrument (page 2-28). This section provides
you with troubleshooting procedures that isolate the faulty assembly or
module. Replacement and recalibration of most modules can only be
performed at an authorized Agilent Technologies Service Center.
If the instrument is under warranty, return it to Agilent for service.
Refer to “Returning the Instrument to Agilent Technologies for Service”
on page 2-3. If you decide to troubleshoot the instrument yourself, refer to
the section titled “Troubleshooting the Instrument” on page 2-28.
2-2Assembly Level Service Guide
Chapter 2 Service
Returning the Instrument to Agilent Technologies for Service
Returning the Instrument to
Agilent Technologies for Service
Providing Repair Information
Before shipping the instrument to an Agilent Technologies office for service
or repair, call the nearest Agilent Sales Office to make arrangements.
Then, tag and package the Agilent 53147A/148A/149A for shipment.
NOTEISD (Instrument Service Division) Emergency Response or Express
Calibration Service is available for Agilent customers in the USA.
If downtime is critical, you can receive your repaired instrument via
overnight shipment. Call 800-403-0801, and ask for Emergency Response or Express Calibration Service. When your instrument is repaired, it is
returned via overnight shipment at no extra charge.
1Write the following information on a tag:
•Owner’s name and address
•Instrument model number
•Complete serial number
•Description of service required or failure indications
2Attach the tag to the instrument.
3Pack the instrument.
If the original packaging materials are available, use the procedure titled
“Packing the Instrument in the Original Packaging Materials.” If the
original packaging materials are not available, you can order new
packaging materials from an Agilent Sales Office. The new packaging
materials are identical to those used by the factory when packaging new
instruments. To use commercially available packaging materials, use the
procedure titled “Packing the Instrument in Commercially Available
Packaging Materials.” Both procedures are on the following page.
2
Assembly Level Service Guide2-3
Chapter 2 Service
Returning the Instrument to Agilent Technologies for Service
Packing the Instrument in the Original
Packaging Materials
To prevent shipping damage, it is best to repack the instrument in its
original packaging. In any correspondence, refer to the instrument by the
model number and the complete serial number.
1Disconnect the power cord, probes, cables, or other accessories attached to
the instrument.
2Make sure the folded corrugated spacer (which normally contains the
manuals) is in the box to ensure proper fitting.
2
3Make sure one of the polystyrene blocks is in its proper position in the box.
4Place the instrument, front panel end first, on the polystyrene block.
5Place the other polystyrene block on top of the rear end of the instrument
to secure it.
6Do not return the manuals with the instrument. Return an accessory only
when it is a part of the failure symptoms.
7Seal the shipping container securely.
2-4Assembly Level Service Guide
Chapter 2 Service
Returning the Instrument to Agilent Technologies for Service
Packing the Instrument in Commercially Available
Packaging Materials
If the original factory packaging materials are not available, use the
following procedure to pack the instrument in commercially available
shipping materials.
1Wrap the instrument in heavy paper or plastic.
2Place the instrument in a strong shipping container. Use a double-walled
carton made of at least 350 lb. test material.
3Protect the control panel with cardboard.
4Add a layer of shock-absorbing material that is 3 to 4 inches (70 to 100 mm)
thick around all sides of the instrument cushion it and to prevent it from
moving within the container
5Seal the shipping container securely.
6Mark the shipping container FRAGILE.
2
Assembly Level Service Guide2-5
Chapter 2 Service
Calibration Procedures
Calibration Procedures
There are four calibration procedures that can be performed outside of an
Agilent authorized service facility. The following procedures describe how
to calibrate the frequency of the instrument’s timebase, the frequency and
output level of the power reference oscillator, and the DVM.
Equipment Required
Table 2-1 lists the test equipment and accessories needed to perform the
2
Table 2-1. Recommended Test Equipment and Accessories
calibration procedures in this chapter.
Model
Instrument TypeRequired Characteristics
Digital Multimeter6½ digit AC/DCAgilent 34401A 1
Power MeterRange 1 mWAgilent 432A1
Power Meter100 kHz to 110 GHz, –70 dBm to +44 dBmAgilent E4418B1
Frequency Counter10 Hz to 125 MHzAgilent 53150A/51A/52A1
Thermistor MountSWR 1.05 at 50 MHz
Accuracy ±0.5% at 50 MHz
Power Sensor10 MHz to 18 GHz, -30 dBm to +20 dBmAgilent 8481A1
Power Sensor Cable5 ft.Agilent 11730A1
Range Calibrator3, 10, 30, 100, and 300 µW
1, 3, 10, 30, and 100 mW
Serial CableRS-232, DB-25 to RJ-45Agilent 53150-602151
Precision DC Power Supply + 50 V / – 50 VAgilent 6613C or 6614C1
CableBNC (m) to BNC (m)Agilent 10503 series1
AdapterType N (m) to BNC (f)Agilent E9635A
* Equivalent equipment can be substituted for all instruments and accessories.
** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
† One supplied with Agilent 53147A/48A/49A
Recommended *
Agilent 478A-H75 or
478A-H76
Agilent 11683A 1
Quantity
Required
1
†
1
2-6Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
10 MHz Timebase Calibration
Equipment Required
10 MHz Frequency Reference Standard
Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A)
Agilent 10503 series BNC (m) to BNC (m) Coaxial Cable
Small, non-metallic adjustment tool for slotted screw heads
NOTEThe instrument and the test equipment must be powered on for at least
30 minutes prior to beginning this calibration procedure. This allows the
internal temperatures of the equipment to stabilize.
1Connect an Agilent 10503 series coaxial cable from the output of your
10 MHz frequency standard to the Counter’s CHANNEL 1 input on the
instrument’s front panel.
2Press the Resol key, and verify that the resolution is set to 1 HZ.
If necessary change the resolution to 1 HZ.
3If there is a label covering the calibration adjustment access hole,
remove it (see Figure 2-1).
2
Assembly Level Service Guide2-7
Chapter 2 Service
Calibration Procedures
2
Calibration Access Hole
Figure 2-1. Calibrating the 10 MHz Frequency Reference Oscillator
4Insert a non-metallic, slot-type adjustment tool through the hole near the
bottom of the instrument’s left side and into the slot in the timebase
calibration screw.
5Turn the calibration screw clockwise to increase the Counter’s reading or
counter-clockwise to decrease the reading until the frequency value
displayed on the Counter’s front panel is exactly 10,000,000 Hz.
2-8Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
Power Reference Oscillator Frequency Calibration
The internal power reference oscillator is used to calibrate the Power
Meter to the power sensor head. This adjustment calibrates the frequency
output of the power reference oscillator to 50.0 MHz ±0.5 MHz.
Equipment Required
Agilent 53150A Microwave Frequency Counter
Agilent 10503 series BNC (m) to BNC (m) Coaxial Cable
Agilent E9635A Type N (m) to BNC (f) Adapter
Small, non-metallic adjustment tool for slotted screw heads
NOTES•The instrument and the test equipment must be powered on for at least
30 minutes prior to adjusting the power reference oscillator output.
This allows the internal temperatures of the equipment to stabilize.
•This procedure specifies the Agilent 53150A for measuring the
frequency output of the 53147A/48A/49A’s power reference oscillator.
If the 53147A/48A/49A’s Frequency Counter is known to be in calibration
(see “Test 1: Channel 1 Frequency Sensitivity” on page 1-18), you can
use it in place of the 53150A in this procedure.
2
•Adjusting the power reference oscillator frequency may also affect the
output level of the oscillator. Therefore, after you adjust the frequency
to 50.0 MHz ± 0.05 MHz, always check the output level and adjust it,
if necessary, as described in “Power Reference Oscillator Level
Adjustment” on page 2-13.
1Turn off the Agilent 53147A/48A/49A, and disconnect the power cord.
2Remove the shroud from the Agilent 53147A/48A/49A (see “Removing the
Shroud” on page 3-7).
3Position the Agilent 53147A/48A/49A on its right side so that you have
access to the front panel and the left side of the main circuit board
(see Figure 2-2).
Assembly Level Service Guide2-9
Chapter 2 Service
p
Calibration Procedures
Agilent
53147A/48A/49A
Agilent
Agilent E9635A Adapter on
Power Reference Oscillator
Output Connector
Agilent
53150A/51A/52A
Microwave Frequency Counter
53150A
Agilent
GHzMHzkHzHz
POWER
Ch 12
Rel Freq
Offset
Avg On
Rel Pwr
2
Standby
Freq
Menu
Offset
Reset/
Rate
Local
On/Off
Shift
Clear
dB
dBm
Watts
mW
uW
Offset
%
MODIFY
Pwr
HPIB
Offset
Resol
Avg
Enter
+/-
20 GHz Counter
Ext Rel
Rate Rmt SRQ
Error
Shift
Gate
Chan
Select
Rel Freq
Offset
On/Off
CHANNEL 1
10 Hz to 125 MHz
Hold
DAMAGE
+30 dBm
CHANNEL 2POWERFREQ
Channel 2 50 MHz to 20 GHz
dBm/ W
DAMAGE
+27 dBm
Display
Power
Rel Pwr
Offset
On/Off
1M Ω
50 Ω
Channel 1
ut
In
Figure 2-2. Reference Oscillator Frequency Adjustment Setup
4Connect the power cord to the Agilent 53147A/48A/49A and to a
power outlet.
WARNINGDANGEROUS VOLTAGES ARE PRESENT INSIDE THE AGILENT
53147A/148A/149A WHEN IT IS CONNECTED TO A POWER
SOURCE. WORK VERY CAREFULLY AND AVOID CONTACT
WITH INTERNAL COMPONENTS WHEN THE SHROUD IS
REMOVED AND POWER IS APPLIED TO PREVENT INJURY
OR DEATH FROM ELECTRICAL SHOCK.
CAUTIONSome internal components of the Agilent 53147A/48A/49A are subject to
damage from electrostatic discharge (ESD), physical contact, and/or
electrical short circuits. Work very carefully when the shroud is removed,
and avoid all unnecessary contact with the internal components.
2-10Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
5Turn on the Agilent 53147A/48A/49A and then wait for at least
30 minutes to allow the internal temperature of the instrument
to stabilize.
6Connect an Agilent E9635A adapter to the Power Meter OUTPUT connector
as shown in Figure 2-2.
7Connect an Agilent 10503 series cable from the Agilent E9635A adapter
on the Agilent 53147A/148A/149A’s Power Meter OUTPUT connector to
the CHANNEL 1 input connector on the Agilent 53150A/51A/52A
Frequency Counter.
8Set the Agilent 53150A/51A/52A Frequency Counter to measure frequency
on Channel 1.
9Turn on the Agilent 53147A/148A/149A’s internal power reference oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through
the menu to PWR REF, press the right-arrow key, then use the up or down
arrow key to change the setting to ON, and press the Enter key. The
Pwr Ref annunciator should now be activated.
NOTEThe reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete.
Setting PWR REF in the 53147A/148A/149A’s menu to ON turns the power
reference oscillator on, and it remains on until the PWR REF menu setting
is returned to OFF. When you calibrate the Power Meter with PWR REF
set to ON, the power reference oscillator remains on after calibration is
complete.
10Insert a small, non-conductive, slot-type adjustment tool into the hole
in the side of the power reference oscillator shield (see Figure 2-3).
This opening allows access to L9, the adjustable inductor that is used to
adjust the frequency of the power reference oscillator.
2
Assembly Level Service Guide2-11
Chapter 2 Service
Calibration Procedures
L
(Frequency Adjustment)
R
2
98
(Power Adjustment)
9
Figure 2-3. Power Reference Oscillator Adjustment Locations
11Set the frequency to 50.0 MHz ±0.05 MHz (49.95 MHz to 50.05 MHz).
Turn the adjustment tool counter-clockwise to increase the frequency or
clockwise to decrease the frequency.
2-12Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
Power Reference Oscillator Level Adjustment
Two methods for calibrating the level of the Power Meter reference signal
are provided in this section. Both methods provide acceptable results, and
either one can be used.
Method 1
Equipment Required
Agilent E4418B Power Meter
Agilent 8481A Power Sensor
Agilent 11730A Power Sensor Cable (supplied with 53147A/48A/49A)
NOTES•To ensure that the results are valid, the Agilent 53147A/48A/49A and
the test equipment should be powered on for at least 30 minutes prior
to beginning the calibration procedure. This allows the internal
temperatures of the equipment to stabilize.
•The calibration of the Agilent E4418B Power Meter must be no more
than one level removed from NIST (National Institute of Standards
and Technology). If the E4418B Power Meter’s calibration is more than
one level removed from NIST, the test results will be invalid.
2
•This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect the Agilent 11730A Power Sensor Cable to the Agilent 8481A
Power Sensor (see Figure 2-4).
Output connector (with Agilent 8481A
Agilent E4418B
Power Meter
Power Sensor Attached)
Input
Figure 2-4. E4418B Power Meter Zeroing and Calibration Test Setup
Assembly Level Service Guide2-13
Chapter 2 Service
Calibration Procedures
2Connect the free end of the Agilent 11730A Power Sensor Cable to the
input connector on the Agilent E4418B Power Meter.
3Connect the Agilent 8481A Power Sensor to the power reference output
connector of the Agilent E4418B Power Meter.
4Zero the E4418B Power Meter.
5Calibrate the E4418B Power Meter.
6Disconnect the Agilent 8481A Power Sensor from the Agilent E4418B
Power Meter, and immediately connect it to the Power OUTPUT connector
2
on the Agilent 53147A/48A/49A (see Figure 2-5).
Output connector (with Agilent 8481A
Power Sensor Attached)
Agilent
53147A/48A/49A
Agilent
Agilent E4418B
Power Meter
Input
Figure 2-5. Power Meter Reference Level Test Setup
7Use a non-metallic, slot-type adjustment tool to adjust the setting of R98
on the main circuit board (see Figure 2-3) until the power level is as close
to 1.000 mW as possible (the level must be between 0.988 mW and
1.012 mW). Turn the adjustment tool counter-clockwise to increase the
power level or clockwise to decrease the power level.
2-14Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
Method 2
Equipment Required
Agilent 432A Power Meter
Agilent 478A-H75 or 478A-H76 Thermistor Mount
Agilent 34401A Digital Multimeter
Agilent 8120-1082 Power Sensor Cable
NOTETo ensure that the results are valid, the Agilent 53147A/48A/49A and the
test equipment should be powered on for at least 30 minutes prior to
beginning the calibration procedure. This allows the internal temperatures
of the equipment to stabilize.
1Make sure the Agilent 432A Power Meter is off. Do not turn it on until
instructed to do so.
2Set the Agilent 34401A Digital Multimeter to measure resistance, and
connect the DVM test prods between the center pin of the V
connector
RF
on the Agilent 432A Power Meter rear panel and pin 1 on the thermistormount end of the Agilent 8120-1082 Power Sensor Cable (see Figure 2-6 to
locate pin 1).
2
2
6
3
1
5
4
Figure 2-6. Agilent 8120-1082 Power Sensor Cable End View
3Round off the resistance reading to two decimal places and keep a record
of this value as R. It is the internal bridge resistance (R) of the Agilent
432A Power Meter (this should be approximately 200.3 ohms).
4Connect the Agilent 478A-H75 or 478A-H76 Thermistor Mount to the
Agilent 53147A/148A/149A Power Meter OUTPUT connector, as shown in
Figure 2-7.
Assembly Level Service Guide2-15
Chapter 2 Service
Calibration Procedures
NOTEDo not use a cable between the Thermistor Mount and the 53147A/
148A/149A Power Meter OUTPUT connector. Connect the Thermistor
Mount directly to the OUTPUT connector.
Agilent 478A-H75 or
478A-H76 Thermistor Mount
(connected to Power
Meter OUTPUT connector)
Agilent
53147A/48A/49A
Agilent
Agilent 432A
Power Meter
Agilent 34401A
V
-
RF
Digital Multimeter
Agilent
2
V
COMP
Agilent 8120-1082
Power Sensor Cable
Figure 2-7. Power Meter Reference Level Calibration Setup
5Connect the Agilent 8120-1082 Power Sensor Cable between the
Thermistor Mount and the input connector on the Agilent 432A Power
Meter.
+
+
-
6Turn on the 432A Power Meter. Wait 30 minutes for the Power Meter and
the Thermistor Mount to stabilize before proceeding to the next step.
7Set the 432A Power Meter’s RANGE switch to Coarse Zero.
8Adjust the COARSE ZERO control on the 432A Power Meter’s front panel
to obtain a zero reading on the meter.
9Set the 432A Power Meter’s RANGE switch to .01 mW / –20 dBm.
10Press the FINE ZERO control on the 432A Power Meter’s front panel down
to obtain a zero reading on the meter.
11Set the 432A Power Meter’s RANGE switch to 1 mW.
12Set the Digital Multimeter to measure microvolts.
2-16Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
CAUTIONEnsure that the Agilent 34401A Digital Multimeter’s input leads are
isolated from chassis ground before performing the next step.
13Connect the positive test prod from the Agilent 34401A Multimeter to the
V
connector on the Agilent 432A Power Meter rear panel.
COMP
NOTEThe V
and VRF connectors are BNC female. You can use two BNC-to-
COMP
banana jack adapters and two BNC cables to make it easier to connect the
Multimeter to the Power Meter.
14Connect the negative test prod from the Agilent 34401A Multimeter to the
connector on the 432A Power Meter’s rear panel.
V
RF
15If the reading on the Multimeter is less than 400 microvolts, proceed to
step 16. If the reading is 400 microvolts or more, press the FINE ZERO
toggle switch on the 432A Power Meter’s front panel down.
16Observe the reading on the Multimeter, round off the reading to the
nearest microvolt, and keep a record of this value as V
.
0
17Turn on the Agilent 53147A/148A/149A’s internal power reference oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through
the menu to PWR REF, press the right-arrow key, then use the up or down
arrow key to change the setting to ON, and press the Enter key. The Pwr Ref annunciator should now be activated.
NOTEThe reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete.
Setting PWR REF in the 53147A/148A/149A’s menu to ON turns the power
reference oscillator on, and it remains on until the PWR REF menu setting
is returned to OFF. When you calibrate the Power Meter with PWR REF
set to ON, the power reference oscillator remains on after calibration
is complete.
2
18Observe the reading on the Multimeter, and keep a record of this value
as V
.
1
Assembly Level Service Guide2-17
Chapter 2 Service
Calibration Procedures
19Disconnect the Multimeter’s negative test lead from the 432A Power Meter’s
V
connector, and connect it to the 432A Power Meter’s chassis ground.
RF
20Observe the reading on the Multimeter, and keep a record of this value
as V
COMP
.
21Calculate the output level of the 53147A/148A/149A’s power reference
oscillator (P
2
Where:
) using the following formula:
rf
( V1 – V0 ) + V
comp
4R (Calibration Factor)
V
comp
Prf =
2 V
= Power reference oscillator output level
P
rf
= previously recorded value
= previously recorded value
V
1
V
= previously recorded value
0
0
2
– V
2
1
R = previously recorded value
Calibration Factor = value for thermistor mount at 50 MHz (traceable to NIST)
2-18Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
NOTEThe following is an example showing how a calculation of the Power
Reference Output Level is performed:
R = 200
Ω
V0=170 µV or 170 x 10–6 V
= 78,107 µV or 78,107 x 10–6 or .078 V
V
1
V
comp
= 5.1583 V
R = previously recorded value
Calibration Factor = 99.68%
2
– V
2
1
Note that V
Prf =
2 V
is eventually dropped out of the equation in the following
0
( V1 – V0 ) + V
comp
0
4R (Calibration Factor)
steps, because its value is very small (effectively zero).
2
P
=
rf
4(200) (99.68%)
0.80469 – 0.00608
=
2 (5.1583) (.078 – 0)
+ (.000170)2 – (0.78)
0.79861
=
797.44797.44
1.00147 mW
=
2
22Use a non-metallic, slot-type adjustment tool to adjust the setting of R98
on the main circuit board (see Figure 2-3) until P
is between 0.988 mW
rf
and 1.012 mW. Turn the adjustment tool counter-clockwise to increase the
power level or clockwise to decrease the power level.
Assembly Level Service Guide2-19
Chapter 2 Service
Calibration Procedures
DVM Calibration
The DVM calibration procedure is semi-automatic in that it is accomplished
via software embedded in the Agilent 53147A/48A/49A. A computer and
an RS-232 serial cable are required to access the calibration software.
Equipment Required
A computer with an RS-232 serial interface and terminal software
Agilent 53150-60215 RS-232 Serial Cable
Agilent 6613C or 6614C Precision DC Power Supply
NOTETo ensure that the results are valid, the Agilent 53147A/48A/49A and
2
the Agilent 6613C/6614C Precision DC Power Supply should be powered
on for at least 30 minutes prior to beginning the calibration procedure.
This allows the internal temperatures of the equipment to stabilize.
1Press the Display DVM key on the Agilent 53147A/48A/49A.
2Connect the DVM test leads to the DVM + and – connectors on the Agilent
53147A/48A/49A (red to + and black to –).
3Connect the test prods to the + and – output connectors on the Agilent
6613C/6614C Precision DC Power Supply.
4Set the 6613C/6614C Precision DC Power Supply to output exactly
+ 50.0 V.
5Connect the Agilent 53150-60215 RS-232 Serial Cable between the
computer’s serial port (DB-25) and the Agilent 53147A/48A/49A’s serial
port (RJ-45).
6Start the terminal software on the computer.
7Send “!acq 60” from the terminal program to the Agilent 53147A/48A/49A
(be sure to type lowercase letters when typing “!acq 60”).
8Press “5” on the computer to choose option 5 from the DMM Tests menu
2. Measure Fr - Frequency in Positive Reference Mode
3. Measure Fn - Frequency in Negative Reference Mode
4. Measure Fm - Frequency in Input mode
5. Calculate and Save Calibration Constant.
6. Measure Voltage
7. Set the Gate time: 10ms - 1000 ms.
8. Choose Gated mode
9. Choose Reciprocal mode
A. Choose Single Measurement mode.
B. Choose Continuous Measurement mode.
E. Exit DVM tests.
Choose one of the above? (1 - 9, E) 5
Figure 2-8. DVM Tests Main Menu
9When the calibration program displays:
2
Apply 50.000 Volts Reference to the DVM. Ready? (y/n) y
type “y”.
10When the calibration program displays:
Update from old 10974.12 to new 10974.1 Cal. Constant? (y/n) y
type “y” (the values in the display line above are simulated).
11When the calibration program displays:
Apply –50.000 Volts Reference to the DVM. Ready? (y/n) y
12Reverse the connection of the test prods in the + and – output connectors
on the Agilent 6613C/6614C Precision DC Power Supply (red to – output,
black to + output), or reverse the polarity of the power outputs.
13Type “y”.
Assembly Level Service Guide2-21
Chapter 2 Service
Calibration Procedures
14When the calibration program displays:
Update from old 10974.01 to new 10974.96 Cal. Constant? (y/n) y
type “y” (the values in the display line above are simulated).
15When the calibration program redisplays its menu (as shown in
Figure 2-8), type “e” to exit the program.
16Close the terminal program and disconnect the serial cable from the
computer and the instrument.
2
2-22Assembly Level Service Guide
Chapter 2 Service
Pre-Troubleshooting Information
Pre-Troubleshooting Information
This section contains the following pertinent troubleshooting information:
•Safety Considerations
•Recommended Test Equipment
•Repair Considerations
•After Service Considerations
•Assembly Identification and Location
Safety Considerations
Although this instrument has been designed in accordance with
international safety standards, this guide contains information, cautions,
and warnings which must be followed to ensure safe operation and to
retain the instrument in a safe condition. Service instructions, and
adjustment procedures requiring removal of the instrument cover, are for
use by service-trained personnel only. To avoid dangerous electric shock,
do not perform any servicing or make any adjustments with the cover
removed, unless qualified to do so.
2
WARNINGBEFORE APPLYING AC POWER, THE INSTRUMENT AND ALL
PROTECTIVE EARTH TERMINALS, EXTENSION CORDS, AUTO
TRANSFORMERS, AND DEVICES CONNECTED TO THE
INSTRUMENT SHOULD BE CONNECTED TO A PROTECTIVE
EARTH GROUNDED SOCKET.
ANY INTERRUPTION OF THE PROTECTIVE GROUNDING
CONDUCTOR INSIDE OR OUTSIDE THE INSTRUMENT OR
DISCONNECTION OF THE PROTECTIVE EARTH TERMINAL
WILL CAUSE A POTENTIAL SHOCK HAZARD THAT COULD
RESULT IN PERSONAL INJURY. INTENTIONAL
INTERRUPTION IS PROHIBITED.
Assembly Level Service Guide2-23
Chapter 2 Service
Pre-Troubleshooting Information
Any maintenance and repair of the opened instrument under voltage should
be avoided as much as possible and, if necessary, should be carried out only
by a skilled person who is aware of the hazards involved (for example, fire
and electric shock).
Recommended Test Equipment
Test equipment recommended for testing and troubleshooting the
Agilent 53147A/148A/149A is listed in Chapter 1, “Performance Tests.”
Substitute equipment may be used if it meets or exceeds the required
characteristics listed in Table 1-1.
2
Repair Considerations
Electrostatic Discharge
Electronic components and assemblies in the Agilent 53147A/148A/149A
can be permanently degraded or damaged by electrostatic discharge.
Use the following precautions when servicing the instrument:
1ENSURE that static-sensitive devices or assemblies are serviced at
static-safe work stations providing proper grounding for service personnel.
2ENSURE that static-sensitive devices or assemblies are stored in static
shielding bags or containers.
3DO NOT wear clothing subject to static-charge buildup, such as wool or
synthetic materials.
4DO NOT handle components or assemblies in carpeted areas.
5DO NOT remove an assembly or component from its static shielding
protection until you are ready to install it.
6AVOID touching component leads. (Handle by packaging only.)
2-24Assembly Level Service Guide
Chapter 2 Service
Pre-Troubleshooting Information
Surface Mount Repair
None of the assemblies in the instrument that use surface-mount
components can be repaired outside of authorized Agilent Service Centers.
Any attempt to repair these assemblies voids the Agilent factory warranty.
Disassembly and Reassembly Specifics
Refer to Chapter 3 of this guide, “Replacing Assemblies,” for complete
disassembly and reassembly instructions, and Chapter 4, “Replaceable
Parts” for an exploded view of the instrument’s parts.
After Service Considerations
Product Safety Checks
The following safety checks must be performed after any troubleshooting
and repair procedures have been completed to ensure the safe operation of
the instrument.
WARNINGTHE RESISTANCE CHECKS DESCRIBED IN THE FOLLOWING
TEXT REQUIRE THAT THE POWER CORD BE CONNECTED TO
THE INSTRUMENT AND THAT AC POWER BE DISCONNECTED.
BE SURE THAT THE POWER CORD IS NOT CONNECTED TO AN
AC POWER SOURCE BEFORE PERFORMING ANY SAFETY
CHECKS.
1VISUAL INSPECTION. Visually inspect the interior of the instrument
for any signs of abnormal internally generated heat, such as discolored
printed circuit boards or components, damaged insulation, or evidence
of arcing. Determine and remedy the cause of any such condition.
2GROUND CONTINUITY TEST. Plug the power cord into the rear- panel
power receptacle. (DO NOT connect the instrument to AC power at this
time.) Using a suitable ohmmeter, check resistance from the instrument's
metallic connection (such as the rear panel or BNC ground collar) to the
ground pin on the power cord plug. The reading must be less than 1Ω.
Flex the power cord while making this measurement to determine
whether intermittent discontinuities exist.
2
Assembly Level Service Guide2-25
Chapter 2 Service
Pre-Troubleshooting Information
3Check any indicated front- or rear-panel ground terminals marked,
using the above procedure.
4INSULATION RESISTANCE TEST. Tie the line and neutral pins of the
power cord plug together. Measure the resistance from the instrument
enclosure (chassis) to the line and neutral pins of the power cord plug.
The minimum acceptable resistance is 2 MΩ. Replace any component
which results in a failure.
Product Performance Checks
After replacement of any functional assembly, perform the operational
2
verification test in Chapter 1 of this guide, “Performance Tests.”
Assembly Identification and Location
The assembly number, name, and Agilent part number of each of the
Agilent 53147A, 53148A, and 53149A assemblies are listed in Table 2-2.
Figure 2-9 illustrates the replaceable assemblies in the Agilent
53147A/148A/149A.
A2 Front Panel Assembly (53147A)
A2 Front Panel Assembly (53148A)
A2 Front Panel Assembly (53149A)
A3 Front Panel Power Switch Assembly (Power/Standby Switch)53147-20203
A4 Battery Housing Assembly *53150-00002
A5 DC/DC Converter PCB *53150-60004
A6 Battery Switch Assembly *53147-60223
A7 Battery *53150-80010
A8 Battery Sled Assembly *53150-00004
A9 Battery Cover Assembly53150-00003
A10 Power Entry Module Assembly53147-40007
A11 Rear Panel Assembly53147-60219
A12 Cooling Fan Assembly53147-80012
A13 Power Supply Assembly0950-3299
* With Option 002 (Battery Option) only.
53147-60207
53148-60207
53149-60207
2-26Assembly Level Service Guide
Chapter 2 Service
Pre-Troubleshooting Information
A5
A4
A3
A13
A12
A6
A11
A7
A9
A10
A8
2
A2
A1
Figure 2-9. Replaceable Assembly Locations — Top Internal View
Assembly Level Service Guide2-27
Chapter 2 Service
Troubleshooting the Instrument
Troubleshooting the Instrument
Power Supply Check
WARNINGHAZARDOUS VOLTAGES ARE PRESENT ON THE POWER
SUPPLY ASSEMBLY. ONLY TRAINED AND QUALIFIED
SERVICE PERSONNEL SHOULD PERFORM THE FOLLOWING
PROCEDURE.
2
NOTEThe DC Power Supply Assembly is fused for safety reasons. It must be
1Remove the power cord from the back of the instrument.
2Remove the cover as described in Chapter 3.
assumed that, if the fuses are blown, the power supply has probably
suffered damage and has therefore become unreliable. If the power supply
fails, it must be replaced, even if the fuse in the power supply appears to
be all that has failed.
3Connect the instrument to the power source.
4Connect the negative lead of a voltmeter to the chassis and measure each
of the voltages listed in Table 2-3 on Power Supply connectors TB1 and
TB2 (see Figure 2-10).
Pin 8
Neutral
TB1 (AC In)
Line
TB2 (DC Out)
Pin 1
Approx. location of R23
Figure 2-10. A4 DC Power Supply Test-Point Locations
2-28Assembly Level Service Guide
Chapter 2 Service
Troubleshooting the Instrument
Table 2-3 lists A4 Power Supply’s input and output voltages, the
corresponding wire color for each input and output, and the tolerances
specified for these voltages.
Table 2-3. A4 Power Supply Inputs and Outputs
Input/Output VoltageConnector and Pin Designation Wire Color Tolerance
AC-input (neutral) 0 VAC*TB1, pin Nblue—
AC-input (line in)
85 - 264 VAC*TB1, pin Lbrown—
–15.0 VDC @ 0.7 A max.TB2, pin 1lavender±3%
0.0 VDC (ground for –15 VDC) TB2, pin 2brown—
N/A (unused)TB2, pin position 3 (no pin)N/A—
N/A (unused)TB2, pin 4
+15.0 VDC @ 2 A max.TB2, pin 5blue±5%
0.0 VDC (ground for +15 VDC) TB2, pin 6brown—
+5.0 VDC @ 5A max.TB2, pin 7red±1%
0.0 VDC (ground for +5 VDC)TB2, pin 8black—
* AC-input values are referenced to chassis ground.
Adjusting the +5 VDC Output
The +5 VDC output of the power supply is adjustable. If this voltage is
found to be outside of the tolerance range listed in Table 2-3, it can be
adjusted using R23, which is visible on the top surface of the power supply’s
circuit board. When adjusting the +5 VDC output, use a nonmetallic or
insulated adjustment tool.
2
Assembly Level Service Guide2-29
Chapter 2 Service
Troubleshooting the Instrument
Self-Test
The instrument performs a Self-Test procedure when it is powered on.
The Self-Test can also be invoked from the instrument’s Menu.
Running the Self Test
1Disconnect all cables connected to the instrument’s front and rear signal
connectors.
2Connect the power cord to the power input connector on the instrument’s
rear panel and to an appropriate power source.
2
3Press and release the Power button on the front panel.
4Observe the front-panel display and note any error messages displayed.
5When CH 2 NO SIGNAL is displayed, press and release the Shift key,
and then press and release the Menu key.
6Press the up-arrow key or the down-arrow key repeatedly until
DO SELF TEST is displayed, and then press and release the Enter key.
7Observe the front-panel display and note any error messages displayed.
2-30Assembly Level Service Guide
Chapter 2 Service
Troubleshooting the Instrument
Self Test Error Messages
Table 2-4 lists and describes messages that are generated by the
instrument during Self-Test to indicate whether a component passed or
failed its test. These messages are sent via the RS-232 serial output only—
they do not appear on the instrument’s front-panel display.
Table 2-4. Self-Test Messages
MessageDescription
ROM TEST FAIL
ROM TEST OK
RAM DATA LINES OK
RAM DATA ERROR
RAM ADDR LINES OK
RAM ADDR ERROR
RAM TEST OK
EEPROM FAIL - CONFIGURATION
DATA
ROM FAIL; Computed checkbyte does
not match the value stored in EEPROM.
EEPROM FAIL - CONFIGURATION
DATA; Needs to be (re)initialized.
EEPROM FAIL - POWER CAL DATA;
Using default data
EEPROM FAIL - SAVED SETTINGS;
Using default data
EEPROM FAIL - SAVED SETTINGS;
Invalid EEPROM SAV n Data.
GPIB FAIL; Conf. TestThe GPIB hardware failed its confidence test.
ROM failed read test.
ROM passed read test.
RAM data lines passed test.
RAM data lines failed test.
RAM address lines passed test.
RAM address lines failed test.
RAM tests completed with no errors detected.
The configuration data saved in EEPROM
memory is defective.
The checksum of the ROM data does not match
the value stored in EEPROM.
The EEPROM org code does not verify with
current revision of ROM code.
The checksum of the EEPROM powercalibration table is bad. Factory default
calibration data will be used.
The checksum of the user settings stored in
EEPROM is bad. Factory default settings will
be used.
The checksum of one set of user settings (1 – 9)
stored in EEPROM is bad.
2
Assembly Level Service Guide2-31
Chapter 2 Service
Troubleshooting the Instrument
Tables 2-5 and 2-6 list and describe the messages that are generated by
the instrument during Self-Test or during operation to indicate that a
problem has been detected. These messages are displayed on the
instrument’s front-panel display and are also sent via the RS-232 serial
output (in many cases, the exact message text that is displayed on the
front panel is a condensed form of the message that is sent via RS-232).
Table 2-5. Front Panel Display Error Messages
MessageDescription
12V FAILThe +12 VDC output from the power supply is not within specifications.
–12V FAILThe –12 VDC output from the power supply is not within specifications.
2
–5V FAILThe –5 VDC output from the power supply is not within specifications.
ADC FAILFailure detected in the ADC.
PATH FA I LFailure detected in the Channel 1 signal path.
THRS FAILFailure detected in the Channel 1 threshold circuit.
THRS FAILFailure detected in the Channel 2 RF threshold circuit.
THRS THRUFailure detected in the Channel 2 through-path threshold circuit.
THRS HETFailure detected in the Channel 2 heterodyne-path threshold circuit.
EEP WRT FAILFailure detected while writing to EEPROM.
FPANEL FAILThe front panel or its interconnecting cable is defective or not properly connected.
FPGA FAILFailure detected in the FPGA (Field Programmable Gate Array).
GPIB FAILFailure detected in the GPIB hardware.
HET PATH FAILFailure detected in the heterodyne-path circuit.
IIC FAILAn attempt to write to the LCD display failed.
INSTCFG FAILThe instrument’s configuration data is missing or has become corrupted.
OVERTEMPThe instrument’s internal temperature is above the acceptable limit.
PWR CAL FAILThe instrument’s power-calibration data is missing or corrupted.
PWR CKT FAILFailure detected in the Channel 2 power-measurement circuit.
RAM FAILFailure detected in RAM.
ROM FAILFailure detected in ROM.
SAV SET FAILOne or more of the sets of user settings is missing or is corrupted.
SERVICE FAILThe instrument’s service data is missing or has become corrupted.
THRU FAILFailure detected in the through-path circuit of Channel 2.
VCO/CNT FAILFailure detected in the VCO or the Count Chain.
2-32Assembly Level Service Guide
Chapter 2 Service
Troubleshooting the Instrument
Table 2-6. RS-232 Error Messages
MessageDescription
12V FAILThe +12 VDC output from the power supply is not within specifications.
–12V FAILThe –12 VDC output from the power supply is not within specifications.
–5V FAILThe –5 VDC output from the power supply is not within specifications.
ADC FAILFailure detected in the ADC.
B1 SIGNAL PATH FAILFailure detected in the Channel 1 signal path.
B1 THRESHOLD FAILFailure detected in the Channel 1 threshold circuit.
B2 RF THRESHOLD FAILFailure detected in the Channel 2 RF threshold circuit.
B2 THROUGH-PATH
THRESHOLD FAIL
B2 HETERODYNE PATH
THRESHOLD FAIL
EEPROM FAIL - WRITEFailure detected while writing to EEPROM.
FRONT PANEL FAILThe front panel or its interconnecting cable are defective or not properly connected.
FPGA FAILFailure detected in the FPGA (Field Programmable Gate Array).
GPIB FAILFailure detected in the GPIB hardware.
HETERODYNE
PATH FA IL
IIC FAILAn attempt to write to the LCD display failed.
INSTCFG FAILThe instrument’s configuration data is missing or corrupted.
OVER TEMPERATUREThe instrument’s internal temperature is above the limit.
PWR CAL FAILThe instrument’s power-calibration data is missing or corrupted.
POWER METER FAILFailure detected in the Channel 2 power-measurement circuit.
RAM FAILFailure detected in RAM.
ROM FAILFailure detected in ROM.
SAV SET FAILOne or more of the sets of user settings is missing corrupted.
SERVICE FAILThe instrument’s service data is missing or corrupted.
THRU PATH FAILFailure detected in the through-path circuit of Channel 2.
VCO/COUNT CHAIN FAILFailure detected in the VCO or the Count Chain.
Failure detected in the Channel 2 through-path threshold circuit.
Failure detected in the Channel 2 heterodyne-path threshold circuit.
Failure detected in the heterodyne-path circuit.
2
Assembly Level Service Guide2-33
Chapter 2 Service
Troubleshooting the Instrument
2
2-34Assembly Level Service Guide
3
Replacing Assemblies
Disassembly and Reassembly
Chapter 3 Replacing Assemblies
Introduction
Introduction
This chapter provides instructions for removing major assemblies in the
Agilent 53147A/148A/149A. To install an assembly, apply the removal
instructions in reverse.
The following disassembly and repair procedures are provided:
•Removing the Bumperspg. 3-4
•Removing the Handlepg. 3-5
•Removing the Bumper Retainerspg. 3-6
•Removing the Shroudpg. 3-7
•Removing the Front Panel Assemblypg. 3-8
•Removing the Rear Panel Assemblypg. 3-11
•Removing the Cooling Fanpg. 3-13
3
•Removing the Power Supplypg. 3-14
WARNINGHAZARDOUS VOLTAGES ARE PRESENT ON THE POWER
SUPPLY ASSEMBLY. DISCONNECT THE INSTRUMENT FROM
THE POWER SOURCE, AND WAIT AT LEAST SIX MINUTES
BEFORE WORKING INSIDE THE INSTRUMENT.
CAUTIONDo not replace assemblies when the instrument is operating to prevent
damage to components.
NOTEUnless otherwise stated, directional referents in the procedures in this
chapter are accurate when viewing the instrument from the front.
Part numbers for all replaceable parts and assemblies are provided in
Chapter 5, “Replaceable Parts.”
3-2Assembly Level Service Guide
Chapter 3 Replacing Assemblies
Tools Required
Tools Required
The following tools are required for the removal/replacement procedures
in this chapter:
•#1 Phillips screwdriver
•#2 Phillips screwdriver
•TORX
•1/4-inch open-end or box wrench
•1/4-inch nut driver
•5/16-inch nut driver
•7 mm nut driver
•Needle-nose pliers
®
15 (T15) hand screwdriver
Do This First
Perform the following steps before beginning any of the removal and
replacement procedures:
1If the instrument is on, press and release the POWER button on the
front panel.
2Disconnect the AC power cord from the AC input socket on the rear panel.
3If the instrument has the Battery Option (002) installed, remove the
battery sled, and disconnect any DC input cable from the EXT DC
connector on the rear panel.
CAUTIONWhile most of the hardware used in the instrument is SAE, the six
TORX screws that attach the handle mounts to the instrument are metric.
To avoid damage to the threaded holes in the chassis that handle mount
screws fit in, be sure to use the correct screws.
Assembly Level Service Guide3-3
3
Chapter 3 Replacing Assemblies
Removing the Bumpers
Removing the Bumpers
Use the following procedure to remove the front and rear protective
bumpers:
1Remove the rear bumper by lifting the inner edge of the bumper away
from the shroud near one corner at the top and one side and pulling that
corner away from the instrument. Repeat the process with the remaining
corners until the bumper is free of the instrument.
Auxiliary
Reference 10 MHz
WARNING:
90-130 VAC 50/60/400 Hz
190-230 VAC 50/60 Hz
Main Power
In or Out
FUSE
AC POWER
1.5A T
250V
RS-
232
GPIB
To avoid electrical shock:
No user serviceable parts inside.
Refer all servicing to qualified personnel.
Do not remove covers.
Counter/ Power Meter/ DVM
Agilent
!
!
OPTIONS
001 Oven Time Base
002 Battery
Made in U.S.A.
This unit must be earth grounded.
of domestic and foreign content
20 GHz Counter/Power Meter/DVM
GHz
MHz kHz Hz
POWER
Ch 12
Rel Freq
Offset
dB
Avg On
dBm
Watts
mW
Rel Pwr
uW
Standby
Offset
%
MODIFY
Cal
Pwr
Freq
Menu
Factor
Offset
Offset
Reset/
Avg
Freq
Resol
Local
On/Off
Rate
Enter
+ / -
Clear
Shift
COUNTER
Gate
Offset
On/Off
Select
COUNTER
POWER METER
CHANNEL 1
Counter
Power
Ext Rel
Hold
10 Hz to 125 MHz
OUTPUT
Meter
Rate Rmt SRQ
Error
1MΩ
DVM
Shift
POWER METER
DVM
DAMAGE
Power Ref.
+
+30 dBm
dBm/W
1.00mW 50MHz
Display
Zero
Power
COUNTER
CHANNEL 2
INPUT
Store
Rel Pwr
50MHz to 26.5 GHz
Cal
Offset
Do NOT Exceed
–
DAMAGE
On/Off
Sensor Max Power
+27 dBm
50Ω
DAMAGE
Display
Rel FreqChan
DVM
±60 VDC
3
Figure 3-1. Removing the Bumpers
2Use the same procedure (as in Step 1) to remove the front bumper.
NOTEThe bumpers can be used to support and protect the instrument after the
shroud is removed. Place both bumpers on their inner edges on your work
surface, and then place the instrument on top of them. The rear bumper
can also serve as a holder for the instrument (once the bumper retainers
are removed) when you need to position it upright on either end.
3-4Assembly Level Service Guide
Chapter 3 Replacing Assemblies
Removing the Handle
Removing the Handle
Use the following procedure to remove the carrying handle:
1Remove the front bumper (see “Removing the Bumpers” on page 3-4).
2Stand the instrument on end on its back bumper.
3Position the handle so that it is against the top surface of the shroud
(see Figure 3-2).
Figure 3-2. Removing the Handle
4Use a T15 TORX screwdriver to remove two of the 8-32 x 1/4-inch handle
mounting screws (remove the two screws that are closest to the back of
the instrument).
5Repeat Step 4 for the other handle mount.
6While holding the handle to support it, remove the remaining two screws
(one on each handle mount).
7Remove the handle.
Assembly Level Service Guide3-5
3
Chapter 3 Replacing Assemblies
Removing the Bumper Retainers
Removing the Bumper Retainers
Use the following procedure to remove the bumper retainers:
1Remove the front and rear bumpers (see “Removing the Bumpers” on
page 3-4).
2Use a #1 Phillips screwdriver to remove the two 4-40 x 7/16-inch flathead
machine screws that attach each bumper retainer to the chassis (see
Figure 3-3).
3
Figure 3-3. Removing the Bumper Retainers
3Repeat Step 2 for each bumper retainer you need to remove.
CAUTIONWhen the all of the bumper-retainer screws and the handle are removed,
the shroud is not attached to the chassis. Exercise caution when moving
the instrument in this condition to prevent the chassis from sliding out of
the shroud. If you do not intend to remove the shroud, insert the bumper
retainer screws in the holes they were removed from (the holes are
countersunk to accept the screws, and the screws can safely be inserted
fully into the holes).
3-6Assembly Level Service Guide
Chapter 3 Replacing Assemblies
Removing the Shroud
Removing the Shroud
Use the following procedure to remove the shroud:
WARNINGWHEN THE SHROUD IS REMOVED FROM THE COUNTER,
DANGEROUS LINE VOLTAGES THAT CAN CAUSE SERIOUS
INJURY ARE EXPOSED. ALWAYS DISCONNECT THE POWER
CORD BEFORE REMOVING THE SHROUD.
1Remove all detachable connector fittings and/or adapters from the
connectors on the instrument’s front panel.
2Remove the front and rear bumpers (see “Removing the Bumpers” on
page 3-4).
NOTEThe bumpers can be used to support and protect the instrument after the
shroud is removed. Place both bumpers on their inner edges on your work
surface, and then place the instrument on top of them. The rear bumper
can also serve as a holder for the instrument (once the bumper retainers
are removed) when you need to position it upright on either end.
3
3Remove the handle (see “Removing the Handle” on page 3-5).
4Remove the four bumper retainers from the front end of the instrument
(see Figure 3-3 and “Removing the Bumper Retainers” on page 3-6).
5Remove the six bumper retainers at the rear end of the chassis.
6Slide the shroud off the chassis to the rear (see Figure 3-4), and place it
aside (if the shroud does not slide easily off the chassis, press on the rear
panel with one hand while pulling the shroud with the other hand).
Assembly Level Service Guide3-7
Chapter 3 Replacing Assemblies
Removing the Front Panel Assembly
3
Figure 3-4. Removing the Shroud
Removing the Front Panel Assembly
Use the following procedure to remove the front panel assembly:
1Remove the shroud (see “Removing the Shroud” on page 3-7).
2Use a 5/16-inch wrench to loosen the nut that attaches the Power Meter
output cable to the back of the Power Meter OUTPUT connector on the
front panel assembly.
3Disconnect the Power Meter output cable from the back of the Power
Meter OUTPUT connector.
3-8Assembly Level Service Guide
Chapter 3 Replacing Assemblies
Removing the Front Panel Assembly
4Disengage the Power Meter output cable from the cable clamp on the
chassis floor near the right front corner of the chassis.
5Disconnect the Power Meter input cable from J22, which is accessible
through the rectangular opening in the chassis floor near the right-front
corner of the chassis.
6Disengage the Power Meter input cable from the cable clamp on the
chassis floor near the right front corner of the chassis.
7Disconnect the Power Meter INPUT connector’s ground wire (black) from
the ground lug on the chassis floor.
8Disconnect the Channel 1 input cable from J15, which is accessible
through a hole in the chassis floor that is at the front of the chassis,
behind the Chan Select key.
9Disconnect the DVM input connector from J23, which is accessible
through the square hole in the chassis floor approximately 1/3 of the way
back from the front panel in line with the Display DVM key.
10Use a microwave torque wrench to partially loosen the nut that connects
the Channel 2 input connector to the sampler on the main circuit board.
3
CAUTIONDo not attempt to fully loosen the nut at this time. As soon as the nut
turns freely, stop turning it, and proceed to the next step.
11Remove the two 6-32 x ¼ flathead machine screws from the countersunk
holes at the top-front corners of the chassis (see Figure 3-5).
Assembly Level Service Guide3-9
Chapter 3 Replacing Assemblies
Removing the Front Panel Assembly
3
Figure 3-5. Removing the Front Panel Assembly
12Remove the five 4-40 x ¼-inch flathead machine screws that attach the
front-panel assembly to the bottom of the chassis.
13Remove the two 6-32 x ¼-inch flathead machine screws from the
countersunk holes at each side of the front the end of the chassis.
14While holding the front panel assembly with one hand, use the fingers of your
other hand to continue to loosen the nut on the Channel 2 input connector.
Gently pull the front panel assembly away from the chassis as you loosen the
nut until the front panel assembly is completely disconnected.
CAUTIONWhen reattaching the front panel assembly, reverse the procedure above.
You must use a microwave torque wrench to tighten the nut that attaches
the Channel 2 input connector to the sampler. Using any other tool to
tighten the coupling nut will damage the connector, the sampler, or both.
3-10Assembly Level Service Guide
Chapter 3 Replacing Assemblies
Removing the Rear Panel Assembly
Removing the Rear Panel Assembly
Use the following procedure to remove the rear panel assembly:
1Remove the shroud (see “Removing the Shroud” on page 3-7).
2If the Battery Option is not installed, remove the four 4-40 self-locking hex
nuts that attach the battery opening cover to the rear panel (these nuts
are accessible from inside the chassis). Remove the battery opening cover,
and skip to step 5.
3If the Battery Option is installed, remove the battery sled from the
Counter (for instructions on removing the battery sled, see Appendix C
of the Agilent 53147A/148A/149A Operating Guide).
4If the Battery Option is installed, remove the four 4-40 x 3/16-inch
self-locking flathead machine screws that attach the rear panel assembly
to the Battery Option chassis as shown in Figure 3-6 (two screws on each
side of the opening).
Auxiliary
RS-
232
Reference 10 MHz
In or Out
FUSE
1.5A T
250V
No user serviceable parts inside.
Refer all servicing to qualified personnel.
!
OPTIONS
001 Oven Time Base
002 Battery
GPIB
!
Agilent
Battery
option
screws
This unit must be earth grounded.
Counter/ Power Meter/ DVM
Battery
Battery
Charging
EXT DC
11 to 18 VDC
+
Made in U.S.A.
of domestic and foreign content
Rear panel
screws
Rear panel
screws
90-130 VAC 50/60/400 Hz
190-230 VAC 50/60 Hz
WARNING:
AC POWER
Main Power
To avoid electrical shock:
Do not remove covers.
Rear panel screws
Figure 3-6. Rear Panel Assembly Screws
5Remove the nine 4-40 x 3/16-inch self-locking flathead machine screws
that attach the rear panel assembly to the chassis (see Figure 3-6).
NOTEThe screw in the lower-left corner of the rear panel assembly (as viewed
from the rear of the Counter) may be covered by the serial-number sticker.
If it is, carefully cut away the lower-left corner of the sticker to expose the
screw head.
Assembly Level Service Guide3-11
3
Chapter 3 Replacing Assemblies
Removing the Rear Panel Assembly
6Disconnect the external-reference input/output cable (coaxial) from
connector J8 on the main board. J8 is directly under the back edge of the
power supply near the chassis wall.
7A two-wire AC-input cable (brown and blue wires) runs between the
AC input module on the rear-panel assembly and the connector that is
accessible through the cutout in the front of the power supply enclosure
(TB1). Disconnect this cable from the power supply.
8If the Battery Option is installed, there is a two-wire cable (brown and
white wires) that runs between the AC input module on the rear panel
assembly and connector J3 on the Battery Option circuit board (J3 is the
third connector from the back edge of the board). Disconnect this cable
from J3 on the Battery Option circuit board.
9Use a 5/16-inch nut driver to remove the self-locking hex nut that attaches
the rear panel assembly grounding wire (green/yellow) to the right wall of
the chassis. Slide the grounding lug off the welded stud on the chassis
wall, and thread the lock nut back onto the stud a few turns until it is
finger-tight.
10Use a 7 mm nut driver to remove the two hex standoff screws that attach
3
the GPIB connector to the rear panel assembly.
WARNINGAN ELECTRICAL SHOCK HAZARD EXISTS IF THE COUNTER IS
OPERATED WITH THE REAR PANEL ASSEMBLY GROUNDING
WIRE NOT PROPERLY CONNECTED TO THE CHASSIS. TO
PREVENT SERIOUS INJURY AND/OR DAMAGE TO THE
EQUIPMENT, NEVER OPERATE THE COUNTER UNLESS THIS
GROUND WIRE IS SECURELY CONNECTED TO THE CHASSIS
WALL AND THE REAR PANEL ASSEMBLY.
11Carefully pull the rear panel assembly away from the back end of the
chassis. Guide the cables that remain attached to the rear panel assembly
as you remove it to prevent them from catching on other cables or other
components of the Counter.
CAUTIONWhen reinstalling the rear panel assembly, position it carefully to avoid
damage to the Battery Charging LED, the serial-port connector (RJ-12),
the GPIB connector, and the main board. If these components are not
correctly aligned with the openings in the back panel, they may be
damaged.
3-12Assembly Level Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.