This guide describes how to service the Agilent 53147A, 53148A, and
53149A. The information in this guide applies to instruments having the
number prefix listed below, unless accompanied by a “Manual Updating
Changes” package indicating otherwise.
SERIAL PREFIX NUMBER:US4047 (53147A)
US4048 (53148A)
US4049 (53149A)
Agilent 53147A/148A/149A
Microwave Frequency Counter/
Power Meter/DVM
Copyright Agilent
Technologies, Inc. 2001, 2002
All Rights Reserved.
Reproduction, adaptation, or
translations without prior
written permission is
prohibited, except as allowed
under the copyright laws.
Printed: May 2002
Printed in U.S.A.
Manual part number
53147-90010
Certification and Warranty
Certification
Agilent Technologies, Inc.
certifies that this product met
its published specification at the
time of shipment from the
factory. Agilent further certifies
that its calibration measurements
are traceable to the United
States National Institute of
Standards and Technology
(formerly National Bureau of
Standards), to the extent
allowed by the Institute’s
calibration facility, and to the
calibration facilities of other
International Standards
Organization members.
Warranty
Agilent warrants Agilent
hardware, accessories and
supplies against defects in
materials and workmanship for
a period of one year from date of
shipment. If Agilent receives
notice of such defects during the
warranty period, Agilent will, at
its option, either repair or
replace products which prove to
be defective. Replacement
products may be either new or
like-new.
Agilent warrants that
Agilent software will not fail to
execute its programming
instructions, for the period
specified above, due to defects in
material and workmanship
when properly installed and
used. If Agilent receives notice
of such defects during the
warranty period, Agilent will
replace software media which
does not execute its
programming instructions due
to such defects.
For detailed warranty
information, see back matter.
Safety Considerations
General
This product and related
documentation must be
reviewed for familiarization
with this safety markings and
instructions before operation.
Before Cleaning
Disconnect the product from
operating power before cleaning.
Warning Symbols That May
Be Used In This Book
Instruction manual symbol; the
product will be marked with
this symbol when it is necessary
for the user to refer to the
instruction manual.
Indicates hazardous voltages.
Indicates earth (ground)
terminal.
or
Indicates terminal is connected
to chassis when such connection
is not apparent.
Indicates Alternating current.
Indicates Direct current.
Safety Considerations
(cont’d)
WARNING
BODILY INJURY OR DEATH
MAY RESULT FROM
FAILURE TO HEED A
WARNING. DO NOT
PROCEED BEYOND A
WARNING UNTIL THE
INDICATED CONDITIONS
ARE FULLY UNDERSTOOD
AND MET.
CAUTION
Damage to equipment, or
incorrect measurement data,
may result from failure to
heed a caution. Do not
proceed beyond a CAUTION
until the indicated conditions
are fully understood and met.
Safety Earth Ground
An uninterruptible safety earth
ground must be maintained
from the mains power source to
the product’s ground circuitry.
WARNING
WHEN MEASURING POWER
LINE SIGNALS, BE
EXTREMELY CAREFUL AND
ALWAYS USE A
STEP-DOWN ISOLATION
TRANSFORMER WHICH
OUTPUT IS COMPATIBLE
WITH THE INPUT
MEASUREMENT
CAPABILITIES OF THIS
PRODUCT. THIS PRODUCT’S
FRONT AND REAR PANELS
ARE TYPICALLY AT EARTH
GROUND. THUS, NEVER TRY
TO MEASURE AC POWER
LINE SIGNALS WITHOUT AN
ISOLATION TRANSFORMER.
For additional safety and
acoustic noise information,
see back matter.
Agilent Technologies, Inc.7.C.NL.03.22.01.R1.M.CW6FC
5301 Stevens Creek Boulevard
Santa Clara, California 95052-8059
Running the Self Test2-30
Self Test Error Messages2-31
ivAssembly Level Service Guide
Contents
3Replacing Assemblies
Introduction3-2
Tools Required3-3
Do This First3-3
Removing the Bumpers3-4
Removing the Handle3-5
Removing the Bumper Retainers3-6
Removing the Shroud3-7
Removing the Front Panel Assembly3-9
Removing the Rear Panel Assembly3-11
Removing the Cooling Fan3-14
Removing the Power Supply3-15
4Replaceable Parts
Introduction4-2
Replaceable Parts4-2
How To Order A Part4-3
Contacting Agilent Technologies4-4
Parts Identification4-5
Reference Designations4-5
Cabinet Parts and Hardware4-5
Accessories and Miscellaneous Items4-5
5Backdating
Introduction5-2
6Specifications
Introduction6-2
Measurement Specifications and Characteristics6-3
ARack Mounting the InstrumentA-2
Index
Assembly Level Service Guidev
Contents
viAssembly Level Service Guide
In This Guide
This guide provides assembly-level service information for the
Agilent 53147A, 53148A, and 53149A.
How to Use This Guide
Repair Strategy
This service guide is designed to isolate failures to the assembly-level only.
The Agilent 53147A/148A/149A instruments can be returned to Agilent for
all service work, including troubleshooting, and verifying specifications.
Contact your nearest Agilent Sales and Service Office for more details.
NOTEISD (Instrument Service Division) Emergency Response or Express
Calibration Service is available for Agilent customers in the USA.
If downtime is critical, you can receive your repaired instrument via
overnight shipment. Call 800-403-0801, and ask for Emergency Response or Express Calibration Service. When your instrument is repaired, it is
returned via overnight shipment at no extra charge.
If you decide to service the instrument yourself, use the troubleshooting
procedures in Chapter 2 (Service) and the disassembly and reassembly
procedures in Chapter 3 (Replacing Assemblies). Then use the calibration
instructions in Chapter 2 to calibrate the instrument for peakperformance operation, and finally, perform all of the performance tests in
Chapter 1 to verify that the instrument is operating to the specifications.
Assembly Level Service Guidevii
In This Guide
Instrument Identification
The instrument is identified by the serial number on the rear panel.
Agilent uses a two-part serial number with the first part (prefix)
identifying a series of instruments and the second part (suffix) identifying
a particular instrument within a series. Agilent-assigned alpha characters
before the suffix identify the country in which the instrument was
manufactured.
Instruments Covered by this Guide
This guide applies directly to Agilent 53147A, 53148A, and 53149A
instruments that have the same serial number prefix(es) shown on the
title page. If the serial number prefix of your instrument differs from that
listed on the title page of this guide, there may be differences between this
guide and your instrument.
Instruments having a higher serial prefix are covered (when required) by
one or more manual-change sheets included with this guide. If a required
change sheet is missing, contact your nearest Agilent Sales Office listed at
the back of this guide.
Assembly-Level Service Guide Organization
This Assembly-Level Service Guide consists of a table of contents,
a preface, six chapters, and an index. The page headers identify the
chapters and sections of this manual. The chapter contents are
summarized as follows:
Chapter 1 Performance Tests provides procedures that verify the
instrument operates properly and meets the Agilent 53147A/148A/149A
specifications given in Chapter 6, “Specifications,” in this guide.
Chapter 2 Service is divided into seven main sections that provide
instructions for returning the instrument to Agilent for service,
calibrating the instrument, and troubleshooting the assemblies in
the instrument.
Chapter 3 Replacing Assemblies provides procedures for replacing
defective assemblies and/or modules in the instrument.
viiiAssembly Level Service Guide
In This Guide
Chapter 4 Replaceable Parts lists the replaceable parts contained in
the instrument, and explains how to order replacement parts for your
instrument.
Chapter 5 Backdating contains information required to adapt this
manual for older instruments (to be provided when required).
Chapter 6 Specifications lists all the specifications and operating
characteristics for the Agilent 53147A/148A/149A.
How to Order Guides
The part number for this guide is listed on the Certification and Warranty
page (on the back of the title page) and on the back cover of this guide.
Assembly Level Service Guideix
In This Guide
Description of the Microwave Frequency
Counter/Power Meter/DVM
The Agilent 53147A, 53148A, and 53149A are capable of measuring
frequencies from 10 Hz to 125 MHz on Channel 1 and from 50 MHz to
20 GHz (53147A), 26.5 GHz (53148A), and 46 GHz (53149A) on Channel 2.
These instruments are also capable of measuring power (the power and
frequency ranges for power measurement are dependent on the powersensor model used) and DC voltages up to ±50 VDC. All three instruments
have a maximum frequency resolution of 1 Hz.
The Agilent 53147A, 53148A, and 53149A provide GPIB and RS-232 serial
interfaces and are suitable for field, bench-top, and ATE operation.
The basic measurement functions of the Agilent 53147A/148A/149A
include Frequency, Relative Frequency, Frequency Offset, and Power
(including Power Offset and Relative Power). All of these features are
accessible from the front panel and over the GPIB and RS-232 interfaces.
The Agilent 53147A/148A/149A includes the following additional
measurement functions and features that are designed specifically for
manufacturing and service applications:
•1, 2, 5, and 10 MHz external reference capability
•Optional high-stability oven oscillator for high-accuracy needs and
lengthened calibration cycles
•Frequency and power offset capabilities for relative measurements
•SCPI programming capability
•Battery and dc input option for operation in locations where ac
power is unavailable
•Optional soft carrying case for safe transportation and mobile use
Programmable control is performed via a GPIB or an RS-232 serial
interface. The GPIB and RS-232C ports are standard for the Agilent
53147A, 53148A, and 53149A.
xAssembly Level Service Guide
In This Guide
Options
The options available for the Agilent 53147A/148A/149A are listed
following this paragraph. Specifications for the options are listed in
Chapter 6, “Specifications.” Options ordered with the instrument are
installed at the factory and are ready for operation on delivery.
Hardware
•High Stability Oven Timebase, Option 001
•Battery/DC Power Input, Option 002
•Rack Mount Kit, Option 1CM
•Soft Carrying Case, Option 007
Retrofit
Options 001 and 002 can be installed only by authorized Agilent
Technologies Repair Centers.
Instructions for installing the Rack Mount Kit are provided in
Appendix A.
Assembly Level Service Guidexi
In This Guide
Accessories Supplied and Available
Accessories Supplied
•Power cord, 2.3 meters (Part number dependent upon destination country)
•Power sensor cable (Agilent P/N 11730A)
•DVM test leads (Agilent P/N 34132B)
Accessories Available
•Soft Carrying Case (Agilent P/N 53147-80016)
•Automotive Power Adapter (Agilent P/N 53150-60214)
•Battery (Agilent P/N 53150-80010)
•GPIB Cables (Agilent P/N 10833A/B/C/D)
•RS-232 Cable (Agilent P/N 53150-60215)
•Power Sensors (Agilent 8480 series)
Manuals Supplied
Agilent 53147A/148A/149A Operating and Programming Guide
(Agilent P/N 53147-90009)
Agilent 53147A/148A/149A Assembly-Level Service Guide
(Agilent P/N 53147-90010)
xiiAssembly Level Service Guide
1
Performance Tests
Verifying Specifications
Chapter 1 Performance Tests
Introduction
Introduction
1
This chapter provides procedures to test the electrical performance of the
Agilent 53147A, 53148A and 53149A. These procedures are based on the
specifications in Chapter 6, “Specifications.”
Two types of testing are provided:
•Operational Verification
•Complete Performance Tests
This chapter is organized as follows:
•Introductionpg. 1-2
•Equipment Requiredpg. 1-4
•Operational Verificationpg. 1-6
•Complete Performance Testspg. 1-17
•Performance Test Recordpg. 1-39
Operational Verification
Operational Verification is an abbreviated series of tests that you can
perform (instead of performing the Complete Performance Tests) to
provide a high degree of confidence that the instrument is operating
properly. Operational Verification is useful for incoming inspection,
routine maintenance, and after instrument repair.
1-2Assembly Level Service Guide
Chapter 1 Performance Tests
Introduction
Complete Performance Tests
The Complete Performance Tests verify the specifications listed in
Chapter 6, “Specifications.” All tests can be performed without opening
the instrument.
Recommended Calibration Cycle
The instrument requires periodic verification of operation. Depending on
the type of use, environmental conditions, aging, and measurement
accuracy required, the instrument should be checked using the operational
verification procedure at least once every year. A full Calibration and
Performance Test should be performed each time the instrument changes
environment or if an assembly or module has been replaced.
Test Record
The results of the Operational Verification and the Complete Performance
Tests should be recorded on a copy of the Performance Test Record, located
at the end of the Complete Performance Test section in this chapter.
1
Assembly Level Service Guide1-3
Chapter 1 Performance Tests
Equipment Required
Equipment Required
1
Table 1-1 lists the test equipment and accessories needed to perform the
tests in this chapter.
Table 1-1. Recommended Test Equipment and Accessories
Instrument TypeRequired Characteristics
Synthesized Signal
Generator
Microwave Synthesized
Signal Generator
Microwave Amplifier45 MHz to 50 GHzAgilent 83051A1P, T
Digital Multimeter6½ digit AC/DCAgilent 34401A 1OV, P
DC Power Supply25 VDCAgilent E3640A 1OV
Frequency Counter10 Hz to 125 MHzAgilent
Power MeterRange 1 mWAgilent 432A1P, C, T
Power Meter50 MHz to 50 GHz, +13 dBm to –50 dBmAgilent 437B1P, C, T
Power Meter100 kHz to 110 GHz, –70 dBm to +44 dBmAgilent E4418B1P, C, T
Thermistor MountSWR 1.05 at 50 MHz
10 Hz to 20 MHz, +7 dBm to –40 dBmAgilent 3325A/B2OV, P, T
10 MHz to 50 GHz, Accuracy >1 x 10
+7 dBm to –50 dBm
Accuracy ±0.5% at 50 MHz
-10
Model
Recommended *
Agilent 83650B1P, T
53150A/51A/52A
Agilent 478A-H75
or 478A-H76
Quantity
Required Use **
1P, C, T
1P, T
Power Sensor10 MHz to 18 GHz, -30 dBm to +20 dBmAgilent 8481A1P, T
Power Sensor50 MHz to 50 GHz, +13 dBm to –50 dBmAgilent 8487A1P, C, T
Power Sensor Cable2.4 mm to 2.4 mm, 5 ft.Agilent 11730A2
Power Sensor CableAgilent 8120-1082 1P, T
Range Calibrator3, 10, 30, 100, and 300 µW
1, 3, 10, 30, and 100 mW
* Equivalent equipment can be substituted for all instruments and accessories.
** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
† One supplied with Agilent 53147A/48A/49A
Agilent 11683A 1P
†
P, T
1-4Assembly Level Service Guide
Chapter 1 Performance Tests
Equipment Required
Table 1-1. Recommended Test Equipment and Accessories (continued)
Model
Instrument TypeRequired Characteristics
20 dB AttenuatorDC to 50 GHz
2.4 mm (m) to 2.4 mm (f)
Power SplitterDC to 50 GHz
2.4 mm (f3x)
CablesBNC (m) to BNC (m)
2.4 mm (m) to 2.4 mm (m)
AdaptersBNC (f) to Type N (m)
BNC (m) to SMA (f)
3.5 mm (f) to 2.92 mm (m)
2.92 mm (m) to 2.4 mm (m)
2.92 mm (m) to 2.4 mm (f)
2.4 mm (m) to 2.4 mm (m)
2.4 mm (f) to 2.4 mm (f)
* Equivalent equipment can be substituted for all instruments and accessories.
** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
† One supplied with Agilent 53147A/48A/49A
Operational Verification is an abbreviated series of tests that you can
perform (instead of performing the Complete Performance Tests) to
provide a high degree of confidence that the instrument is operating
properly. Operational Verification is useful for incoming inspection,
routine maintenance, and after instrument repair.
If you are not familiar with operating the instrument, you should review
the Chapter 1, “Getting Started,” in the Agilent 53147A/148A/149A Operating Guide. However, the procedures in this chapter are written
so that little experience is necessary. These procedures should be followed
in the order in which they appear.
Power-On Self Tests
NOTEThis test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Inspect the instrument for damage.
2Make sure no cables are connected to the instrument’s inputs.
3Connect the power cord to the instrument and the power source.
NOTES•Since the instrument’s power supply automatically senses the line
voltage, there is no AC input-voltage setting.
•It is normal for the fan in the instrument to run when the instrument is
in Standby mode. Power is supplied to the timebase whenever the
power cord is connected to maintain long-term measurement reliability,
and the fan helps to maintain the timebase’s temperature stability.
4Press and release the Power button on the front panel.
1-6Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
5Verify that the front-panel display shows the following:
•All segments of the front-panel display are temporarily activated.
•TESTING is displayed.
•SELF TEST OK is displayed.
•The model number of the instrument is displayed (a four-digit hex
number is also displayed).
•GPIB ADDR nn (nn = a two digit number from 0 to 30) is displayed.
•CH2 NO SIGNAL is displayed.
6If an error message is displayed, refer to the troubleshooting section in
Chapter 2, “Service.”
7Mark Pass or Fail in the Performance Test Record on page 1-39, Test 1.
NOTETo ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment and the
timebase to stabilize.
1
Assembly Level Service Guide1-7
Chapter 1 Performance Tests
Operational Verification
10 MHz Test
1Connect a BNC-to-BNC cable from the Reference 10MHz connector on
the instrument’s rear panel to the CHANNEL 1 input connector on the
1
front panel.
2Press and release the Chan Select key.
3Verify that Ch 1 is displayed in the upper-left corner of the display.
4Verify that the display reads 10,000,000 Hz.
5Mark Pass or Fail in the Performance Test Record on page 1-39, Test 2.
1-8Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
Run Self Test
NOTEThis test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1Disconnect all signal cables from the input and Reference 10MHz connectors.
2Press and release the Shift key, and then press and release the Menu
(Reset/Local) key.
3Press the up- and/or down-arrow key(s) as many times as necessary until
DO SELF TEST is displayed.
4Press and release the Enter key.
5If no errors are detected, SELF TEST OK is briefly displayed when the self
test is completed. If any error messages are displayed, refer to the
troubleshooting section in Chapter 2, “Service.”
6Mark Pass or Fail in the Performance Test Record on page 1-39, Test 3.
1
Assembly Level Service Guide1-9
Chapter 1 Performance Tests
Operational Verification
External Timebase Test
This test verifies the instrument’s 10 MHz external timebase specification
by verifying that setting REF OSC to EXT allows the instrument to be
1
NOTES•To ensure that the test results are valid, the instrument and the test
synchronized to an external reference signal.
Equipment Required
Agilent 3325A/B Synthesizer (2)
Agilent 10100C 50Ω Feedthrough Terminator
Agilent 10503 Series Coaxial Cables (BNC m to m) (2)
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect an Agilent 10503 series coaxial cable between the output of the
first Agilent 3325A/B Synthesizer and the Reference 10MHz connector on
the instrument’s rear panel (see Figure 1-1).
2Connect an Agilent 10100C 50Ω Feedthrough Terminator to the Counter’s
Channel 1 input connector.
3Connect an Agilent 10503 series coaxial cable between the output of the
second Synthesizer and the feedthrough terminator on the Counter’s
CHANNEL 1 input connector.
1-10Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
To Reference
10 MHz Connector
(Rear Panel)
Agilent 3325A/B Synthesizer
1
10 MHZ Output
Agilent
53147A/48A/49A
Agilent
Channel 1 (with 10100C 50Ω Feedthrough Attached)
Agilent 3325A/B Synthesizer
2 MHz Output
Figure 1-1. External Timebase Test Setup
4Cycle the POWER button to preset the instrument.
5Press the Shift key on the instrument’s front panel.
6Press the Menu key on the instrument’s front panel.
7If necessary, use the up- and down-arrow keys to cycle the display until
it reads REF OSC> INT.
8Press the right-arrow key once. INT begins to flash off and on.
9Press the up- or down-arrow key once. INT changes to EXT.
10Press the Enter key.
11Press the Chan Select key once to select the CHANNEL 1 input connector.
Assembly Level Service Guide1-11
Chapter 1 Performance Tests
Operational Verification
12Set the frequency of the first Synthesizer (connected to the instrument’s
rear-panel Reference 10MHz connector) to 10 MHz at 1 V rms.
1
13Set the frequency of the second Synthesizer (connected to the instrument’s
Channel 1 input connector) to 2 MHz at 100 mV rms.
14Press and release the instrument’s Reset/Local key.
15Verify that the Counter displays a reading of approximately 2 MHz and
that the Ext Ref annunciator is activated.
16Mark Pass or Fail for Test 4 in the Performance Test Record on page 1-39.
17Disconnect the test setup.
1-12Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
Power Meter Test
This test verifies the accuracy of the Power Meter.
Equipment Required
Agilent 11730A Power Sensor Cable (supplied with instrument)
Agilent 8481A, 8481D, 8482A, 8485A, or 8487A Power Sensor Head
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1Connect one end of the Agilent 11730A cable (supplied with instrument)
to the Power Meter INPUT connector on the front panel of the Agilent
53147A, 53148A, or 53149A.
1
2Connect the other end of the Agilent 11730A cable to a power sensor head.
3Press the Display Power key in the Power Meter area of the front panel.
The display should look like this (the reading will vary):
Ch 2
Freq
dB
dBm
Pwr
Watts
mW
uW
%
4If the power sensor head you are using is an Agilent 8481A, 8481D,
8482A, 8485A, or 8487A, select the power sensor head model in the HEAD
menu. If you are not using one of the power sensor heads listed above,
press the Cal Factor key, and enter the calibration factor for the head.
Assembly Level Service Guide1-13
Chapter 1 Performance Tests
Operational Verification
5Press the Zero key.
The display shows ZEROING for several seconds.
1
6Connect the power sensor head to the Power Meter OUTPUT connector.
7Press the Cal key.
The display briefly shows CALIBRATING.
8Turn on the power reference output in the instrument’s menu
(PWR REF > ON).
9Verify that the Counter displays a reading of 0 dBm ±.02 dB.
10Mark Pass or Fail for Test 5 in the Performance Test Record on page 1-39.
11Disconnect the test setup.
1-14Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
DVM Test
This test verifies the accuracy of the DVM.
Equipment Required
Agilent 34401A Digital Multimeter
Agilent E3640A DC Power Supply
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1Set the output of the E3640A DC Power Supply to +25 VDC.
2Set the Agilent 34401A Digital Multimeter to the lowest DC voltage range
capable of safely measuring 25 VDC.
1
3Connect the test leads to the Multimeter (observe correct polarity).
4Connect the test prods to the output terminals on the Agilent E3640A DC
Power Supply (observe correct polarity).
Agilent E3640A
Power Supply
Agilent
Agilent 34401A
Digital Multimeter
Agilent
Agilent
53147A/48A/49A
Agilent
Agilent E3640A
Power Supply
Agilent
Figure 1-2. DVM Test Setup
Assembly Level Service Guide1-15
Chapter 1 Performance Tests
Operational Verification
5Observe and record the voltage reading on the Agilent 34401A Multimeter.
6Disconnect the test leads from the Agilent E3640A DC Power Supply.
1
Be careful not to disturb the voltage setting.
7Connect the test leads to the Agilent 53147A/48A/49A (be sure to observe
the correct polarity).
8Press the Display DVM key on the Agilent 53147A/48A/49A front panel.
9Connect the test prods to the output terminals on the Agilent E3640A
DC Power Supply (be sure to observe the correct polarity).
10Observe and record the voltage reading on the Agilent 53147A/48A/49A.
11Compare the voltage readings you recorded in steps 5 and 10. If the
reading taken with the Agilent 53147A/48A/49A is within ±0.25% ±10 mV
of the reading taken with the Agilent 34401A, mark Pass for Test 6 in the
Performance Test Record on page 1-39. If the reading taken with the
Agilent 53147A/48A/49A is not within ±0.25% ±10 mV of the reading
taken with the Agilent 34401A, mark Fail for Test 6 in the Performance
Test Record on page 1-39.
This completes the Operational Verification.
1-16Assembly Level Service Guide
Chapter 1 Performance Tests
Complete Performance Tests
Complete Performance Tests
The Complete Performance Tests verify the specifications of the
Agilent 53147A/148A/149A listed in Chapter 6, “Specifications.” All of
these tests can be performed without opening the instrument. Table 1-2
lists a summary of the performance tests.
Record the results of the performance tests in the appropriate place on the
Performance Test Record, which starts on page 1-39.
NOTETo ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment and the
timebase to stabilize.
Table 1-2. Complete Performance Tests
Page NumberTest Description
Page 1-18Test 1: Channel 1 Frequency Sensitivity
Page 1-23Test 2: Channel 2 Frequency Sensitivity
Page 1-27Test 3: Power Measurement
Page 1-30Test 4: Power Reference Oscillator Frequency
Page 1-32Test 5: Power Meter Reference Level
1
NOTEOther instrument measurement functions (e.g., Averaging) are
mathematically derived by the microprocessor from the parameters
verified by these performance tests. If the instrument passes the
performance tests, the other measurement functions are also functioning
to specifications.
Assembly Level Service Guide1-17
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
Test 1: Channel 1 Frequency Sensitivity
1
This set of tests verifies the frequency-sensitivity specifications of
Channel 1 of the Agilent 53147A/148A/149A’s Frequency Counter.
Equipment Required
Agilent 3325A/B Synthesizer
Agilent 83650B Synthesizer
Agilent 437B Power Meter
Agilent 8487A/D Power Sensor
Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A)
Agilent 11667C Power Splitter
Agilent 8490D (Opt. 020) 20 dB Attenuator
Agilent 1250-2015 BNC (m) to SMA (f) Adapter
Agilent 11904A 2.92 mm (m) to 2.4 mm (m) Adapter
Agilent 10100C 50 Ω Feedthrough Terminator
Agilent 10503 Series Coaxial Cable (BNC m to m)
2 mm (m) to 2 mm (f) Cable (2)
Test 1a: 10 Hz to 20 MHz Frequency Sensitivity
NOTES•To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
•These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
•If the Agilent 3325A/B Synthesizer or the Agilent 83650B Synthesizer
reference output is not adequate to drive the 53147A/48A/49A, connect
the cable from the Synthesizer to the Reference 10 MHz connector on
the back of the 53147A/48A/49A, and set the 53147A/48A/49A to use
the internal oscillator.
1-18Assembly Level Service Guide
Loading...
+ 106 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.