Agilent 53147A Service Guide

Assembly Level Service Guide
This guide describes how to service the Agilent 53147A, 53148A, and 53149A. The information in this guide applies to instruments having the number prefix listed below, unless accompanied by a “Manual Updating Changes” package indicating otherwise.
SERIAL PREFIX NUMBER: US4047 (53147A)
US4048 (53148A) US4049 (53149A)
Agilent 53147A/148A/149A Microwave Frequency Counter/ Power Meter/DVM
Copyright Agilent Technologies, Inc. 2001, 2002
All Rights Reserved. Reproduction, adaptation, or translations without prior written permission is prohibited, except as allowed under the copyright laws.
Printed: May 2002
Printed in U.S.A.
Manual part number 53147-90010
Certification and Warranty
Certification
Agilent Technologies, Inc. certifies that this product met its published specification at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (formerly National Bureau of Standards), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Warranty
Agilent warrants Agilent hardware, accessories and supplies against defects in materials and workmanship for a period of one year from date of shipment. If Agilent receives notice of such defects during the warranty period, Agilent will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
Agilent warrants that Agilent software will not fail to execute its programming instructions, for the period specified above, due to defects in material and workmanship when properly installed and used. If Agilent receives notice of such defects during the warranty period, Agilent will replace software media which does not execute its programming instructions due to such defects.
For detailed warranty information, see back matter.
Safety Considerations
General
This product and related documentation must be reviewed for familiarization with this safety markings and instructions before operation.
Before Cleaning
Disconnect the product from operating power before cleaning.
Warning Symbols That May Be Used In This Book
Instruction manual symbol; the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
or
Indicates terminal is connected to chassis when such connection is not apparent.
Indicates Alternating current.
Indicates Direct current.
Safety Considerations (cont’d)
WARNING
BODILY INJURY OR DEATH MAY RESULT FROM FAILURE TO HEED A WARNING. DO NOT PROCEED BEYOND A WARNING UNTIL THE INDICATED CONDITIONS ARE FULLY UNDERSTOOD AND MET.
CAUTION
Damage to equipment, or incorrect measurement data, may result from failure to heed a caution. Do not proceed beyond a CAUTION until the indicated conditions are fully understood and met.
Safety Earth Ground
An uninterruptible safety earth ground must be maintained from the mains power source to the product’s ground circuitry.
WARNING
WHEN MEASURING POWER LINE SIGNALS, BE EXTREMELY CAREFUL AND ALWAYS USE A STEP-DOWN ISOLATION TRANSFORMER WHICH OUTPUT IS COMPATIBLE WITH THE INPUT MEASUREMENT CAPABILITIES OF THIS PRODUCT. THIS PRODUCT’S FRONT AND REAR PANELS ARE TYPICALLY AT EARTH GROUND. THUS, NEVER TRY
TO MEASURE AC POWER LINE SIGNALS WITHOUT AN ISOLATION TRANSFORMER.
For additional safety and acoustic noise information, see back matter.
Agilent Technologies, Inc. 7.C.NL.03.22.01.R1.M.CW6FC 5301 Stevens Creek Boulevard Santa Clara, California 95052-8059
Contents
1 Performance Tests
Introduction 1-2
Operational Verification 1-2 Complete Performance Tests 1-3 Recommended Calibration Cycle 1-3
Test Record 1-3 Equipment Required 1-4 Operational Verification 1-6
Power-On Self Tests 1-6
10 MHz Test 1-8
Run Self Test 1-9
External Timebase Test 1-10
Power Meter Test 1-13
DVM Test 1-15 Complete Performance Tests 1-17 Test 1: Channel 1 Frequency Sensitivity 1-18
Test 1a: 10 Hz to 20 MHz Frequency Sensitivity 1-18
Test 1b: 50 MHz to 125 MHz Frequency Sensitivity 1-20 Test 2: Channel 2 Frequency Sensitivity 1-23
Test 2a: 50 MHz to 20 GHz Frequency Sensitivity 1-23
Test 2b: 22 GHz to 26.5 GHz Frequency Sensitivity 1-25
Test 2c: 30 GHz to 46 GHz Frequency Sensitivity 1-26 Test 3: Power Measurement 1-27
Power Measurement Accuracy Test 1-27 Test 4: Power Reference Oscillator Frequency 1-30
Power Reference Oscillator Frequency Test 1-30 Test 5: Power Meter Reference Level 1-32
Power Meter Reference Level: Method 1 1-32
Power Meter Reference Level: Method 2 1-34
Performance Test Record (Page 1 of 3) 1-39
Performance Test Record (Page 2 of 3) 1-40
Performance Test Record (Page 3 of 3) 1-41
Assembly Level Service Guide iii
Contents
2 Service
Introduction 2-2 Returning the Instrument to
Agilent Technologies for Service 2-3
Providing Repair Information 2-3 Packing the Instrument in the Original
Packaging Materials 2-4
Packing the Instrument in Commercially Available
Packaging Materials 2-4
Calibration Procedures 2-6
Equipment Required 2-6 10 MHz Timebase Calibration 2-7 Power Reference Oscillator Frequency Calibration 2-9
Power Reference Oscillator Level Adjustment 2-13
DVM Calibration 2-20
Pre-Troubleshooting Information 2-23
Safety Considerations 2-23 Recommended Test Equipment 2-24 Repair Considerations 2-24
Electrostatic Discharge 2-24 Surface Mount Repair 2-25 Disassembly and Reassembly Specifics 2-25
After Service Considerations 2-25
Product Safety Checks 2-25 Product Performance Checks 2-26
Assembly Identification and Location 2-26
Troubleshooting the Instrument 2-28
Power Supply Check 2-28
Adjusting the +5 VDC Output 2-29
Self-Test 2-30
Running the Self Test 2-30 Self Test Error Messages 2-31
iv Assembly Level Service Guide
Contents
3 Replacing Assemblies
Introduction 3-2 Tools Required 3-3 Do This First 3-3 Removing the Bumpers 3-4 Removing the Handle 3-5 Removing the Bumper Retainers 3-6 Removing the Shroud 3-7 Removing the Front Panel Assembly 3-9 Removing the Rear Panel Assembly 3-11 Removing the Cooling Fan 3-14 Removing the Power Supply 3-15
4 Replaceable Parts
Introduction 4-2 Replaceable Parts 4-2 How To Order A Part 4-3
Contacting Agilent Technologies 4-4 Parts Identification 4-5
Reference Designations 4-5
Cabinet Parts and Hardware 4-5
Accessories and Miscellaneous Items 4-5
5 Backdating
Introduction 5-2
6 Specifications
Introduction 6-2
Measurement Specifications and Characteristics 6-3
A Rack Mounting the Instrument A-2
Index
Assembly Level Service Guide v
Contents
vi Assembly Level Service Guide

In This Guide

This guide provides assembly-level service information for the Agilent 53147A, 53148A, and 53149A.

How to Use This Guide

Repair Strategy
This service guide is designed to isolate failures to the assembly-level only.
The Agilent 53147A/148A/149A instruments can be returned to Agilent for all service work, including troubleshooting, and verifying specifications. Contact your nearest Agilent Sales and Service Office for more details.
NOTE ISD (Instrument Service Division) Emergency Response or Express
Calibration Service is available for Agilent customers in the USA. If downtime is critical, you can receive your repaired instrument via overnight shipment. Call 800-403-0801, and ask for Emergency Response or Express Calibration Service. When your instrument is repaired, it is returned via overnight shipment at no extra charge.
If you decide to service the instrument yourself, use the troubleshooting procedures in Chapter 2 (Service) and the disassembly and reassembly procedures in Chapter 3 (Replacing Assemblies). Then use the calibration instructions in Chapter 2 to calibrate the instrument for peak­performance operation, and finally, perform all of the performance tests in Chapter 1 to verify that the instrument is operating to the specifications.
Assembly Level Service Guide vii
In This Guide
Instrument Identification
The instrument is identified by the serial number on the rear panel. Agilent uses a two-part serial number with the first part (prefix) identifying a series of instruments and the second part (suffix) identifying a particular instrument within a series. Agilent-assigned alpha characters before the suffix identify the country in which the instrument was manufactured.
Instruments Covered by this Guide
This guide applies directly to Agilent 53147A, 53148A, and 53149A instruments that have the same serial number prefix(es) shown on the title page. If the serial number prefix of your instrument differs from that listed on the title page of this guide, there may be differences between this guide and your instrument.
Instruments having a higher serial prefix are covered (when required) by one or more manual-change sheets included with this guide. If a required change sheet is missing, contact your nearest Agilent Sales Office listed at the back of this guide.
Assembly-Level Service Guide Organization
This Assembly-Level Service Guide consists of a table of contents, a preface, six chapters, and an index. The page headers identify the chapters and sections of this manual. The chapter contents are summarized as follows:
Chapter 1 Performance Tests provides procedures that verify the instrument operates properly and meets the Agilent 53147A/148A/149A specifications given in Chapter 6, “Specifications,” in this guide.
Chapter 2 Service is divided into seven main sections that provide instructions for returning the instrument to Agilent for service, calibrating the instrument, and troubleshooting the assemblies in the instrument.
Chapter 3 Replacing Assemblies provides procedures for replacing defective assemblies and/or modules in the instrument.
viii Assembly Level Service Guide
In This Guide
Chapter 4 Replaceable Parts lists the replaceable parts contained in the instrument, and explains how to order replacement parts for your instrument.
Chapter 5 Backdating contains information required to adapt this manual for older instruments (to be provided when required).
Chapter 6 Specifications lists all the specifications and operating characteristics for the Agilent 53147A/148A/149A.
How to Order Guides
The part number for this guide is listed on the Certification and Warranty page (on the back of the title page) and on the back cover of this guide.
Assembly Level Service Guide ix
In This Guide
Description of the Microwave Frequency Counter/Power Meter/DVM
The Agilent 53147A, 53148A, and 53149A are capable of measuring frequencies from 10 Hz to 125 MHz on Channel 1 and from 50 MHz to 20 GHz (53147A), 26.5 GHz (53148A), and 46 GHz (53149A) on Channel 2. These instruments are also capable of measuring power (the power and frequency ranges for power measurement are dependent on the power­sensor model used) and DC voltages up to ±50 VDC. All three instruments have a maximum frequency resolution of 1 Hz.
The Agilent 53147A, 53148A, and 53149A provide GPIB and RS-232 serial interfaces and are suitable for field, bench-top, and ATE operation.
The basic measurement functions of the Agilent 53147A/148A/149A include Frequency, Relative Frequency, Frequency Offset, and Power (including Power Offset and Relative Power). All of these features are accessible from the front panel and over the GPIB and RS-232 interfaces.
The Agilent 53147A/148A/149A includes the following additional measurement functions and features that are designed specifically for manufacturing and service applications:
1, 2, 5, and 10 MHz external reference capability
Optional high-stability oven oscillator for high-accuracy needs and
lengthened calibration cycles
Frequency and power offset capabilities for relative measurements
SCPI programming capability
Battery and dc input option for operation in locations where ac
power is unavailable
Optional soft carrying case for safe transportation and mobile use
Programmable control is performed via a GPIB or an RS-232 serial interface. The GPIB and RS-232C ports are standard for the Agilent 53147A, 53148A, and 53149A.
x Assembly Level Service Guide
In This Guide

Options

The options available for the Agilent 53147A/148A/149A are listed following this paragraph. Specifications for the options are listed in Chapter 6, “Specifications.” Options ordered with the instrument are installed at the factory and are ready for operation on delivery.
Hardware
High Stability Oven Timebase, Option 001
Battery/DC Power Input, Option 002
Rack Mount Kit, Option 1CM
Soft Carrying Case, Option 007
Retrofit
Options 001 and 002 can be installed only by authorized Agilent Technologies Repair Centers.
Instructions for installing the Rack Mount Kit are provided in Appendix A.
Assembly Level Service Guide xi
In This Guide

Accessories Supplied and Available

Accessories Supplied
Power cord, 2.3 meters (Part number dependent upon destination country)
Power sensor cable (Agilent P/N 11730A)
DVM test leads (Agilent P/N 34132B)
Accessories Available
Soft Carrying Case (Agilent P/N 53147-80016)
Automotive Power Adapter (Agilent P/N 53150-60214)
Battery (Agilent P/N 53150-80010)
GPIB Cables (Agilent P/N 10833A/B/C/D)
RS-232 Cable (Agilent P/N 53150-60215)
Power Sensors (Agilent 8480 series)
Manuals Supplied
Agilent 53147A/148A/149A Operating and Programming Guide (Agilent P/N 53147-90009)
Agilent 53147A/148A/149A Assembly-Level Service Guide
(Agilent P/N 53147-90010)
xii Assembly Level Service Guide
1
Performance Tests
Verifying Specifications
Chapter 1 Performance Tests

Introduction

Introduction
1
This chapter provides procedures to test the electrical performance of the Agilent 53147A, 53148A and 53149A. These procedures are based on the specifications in Chapter 6, “Specifications.”
Two types of testing are provided:
Operational Verification
Complete Performance Tests
This chapter is organized as follows:
Introduction pg. 1-2
Equipment Required pg. 1-4
Operational Verification pg. 1-6
Complete Performance Tests pg. 1-17
Performance Test Record pg. 1-39

Operational Verification

Operational Verification is an abbreviated series of tests that you can perform (instead of performing the Complete Performance Tests) to provide a high degree of confidence that the instrument is operating properly. Operational Verification is useful for incoming inspection, routine maintenance, and after instrument repair.
1-2 Assembly Level Service Guide
Chapter 1 Performance Tests
Introduction

Complete Performance Tests

The Complete Performance Tests verify the specifications listed in Chapter 6, “Specifications.” All tests can be performed without opening the instrument.

Recommended Calibration Cycle

The instrument requires periodic verification of operation. Depending on the type of use, environmental conditions, aging, and measurement accuracy required, the instrument should be checked using the operational verification procedure at least once every year. A full Calibration and Performance Test should be performed each time the instrument changes environment or if an assembly or module has been replaced.

Test Record

The results of the Operational Verification and the Complete Performance Tests should be recorded on a copy of the Performance Test Record, located at the end of the Complete Performance Test section in this chapter.
1
Assembly Level Service Guide 1-3
Chapter 1 Performance Tests

Equipment Required

Equipment Required
1
Table 1-1 lists the test equipment and accessories needed to perform the tests in this chapter.
Table 1-1. Recommended Test Equipment and Accessories
Instrument Type Required Characteristics
Synthesized Signal Generator
Microwave Synthesized Signal Generator
Microwave Amplifier 45 MHz to 50 GHz Agilent 83051A 1 P, T
Digital Multimeter 6½ digit AC/DC Agilent 34401A 1 OV, P
DC Power Supply 25 VDC Agilent E3640A 1 OV
Frequency Counter 10 Hz to 125 MHz Agilent
Power Meter Range 1 mW Agilent 432A 1 P, C, T
Power Meter 50 MHz to 50 GHz, +13 dBm to –50 dBm Agilent 437B 1 P, C, T
Power Meter 100 kHz to 110 GHz, –70 dBm to +44 dBm Agilent E4418B 1 P, C, T
Thermistor Mount SWR 1.05 at 50 MHz
10 Hz to 20 MHz, +7 dBm to –40 dBm Agilent 3325A/B 2 OV, P, T
10 MHz to 50 GHz, Accuracy >1 x 10 +7 dBm to –50 dBm
Accuracy ±0.5% at 50 MHz
-10
Model Recommended *
Agilent 83650B 1 P, T
53150A/51A/52A
Agilent 478A-H75 or 478A-H76
Quantity Required Use **
1P, C, T
1P, T
Power Sensor 10 MHz to 18 GHz, -30 dBm to +20 dBm Agilent 8481A 1 P, T
Power Sensor 50 MHz to 50 GHz, +13 dBm to –50 dBm Agilent 8487A 1 P, C, T
Power Sensor Cable 2.4 mm to 2.4 mm, 5 ft. Agilent 11730A 2
Power Sensor Cable Agilent 8120-1082 1 P, T
Range Calibrator 3, 10, 30, 100, and 300 µW
1, 3, 10, 30, and 100 mW
* Equivalent equipment can be substituted for all instruments and accessories. ** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting † One supplied with Agilent 53147A/48A/49A
Agilent 11683A 1 P
P, T
1-4 Assembly Level Service Guide
Chapter 1 Performance Tests
Equipment Required
Table 1-1. Recommended Test Equipment and Accessories (continued)
Model
Instrument Type Required Characteristics
20 dB Attenuator DC to 50 GHz
2.4 mm (m) to 2.4 mm (f)
Power Splitter DC to 50 GHz
2.4 mm (f3x)
Cables BNC (m) to BNC (m)
2.4 mm (m) to 2.4 mm (m)
Adapters BNC (f) to Type N (m)
BNC (m) to SMA (f)
3.5 mm (f) to 2.92 mm (m)
2.92 mm (m) to 2.4 mm (m)
2.92 mm (m) to 2.4 mm (f)
2.4 mm (m) to 2.4 mm (m)
2.4 mm (f) to 2.4 mm (f)
* Equivalent equipment can be substituted for all instruments and accessories. ** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting † One supplied with Agilent 53147A/48A/49A
Recommended *
Agilent 8490D (Opt. 020)
Agilent 11667C/D 1 P, T
Agilent 10503 series
Agilent E9635A Agilent 1250-2015 — Agilent 11904A Agilent 11904D Agilent 11900A Agilent 11900B
Quantity Required Use **
1P, T
2OV, P,
C, T
1 1 1 1 1 1 1
P, C , T
1
Assembly Level Service Guide 1-5
Chapter 1 Performance Tests

Operational Verification

Operational Verification
1
Operational Verification is an abbreviated series of tests that you can perform (instead of performing the Complete Performance Tests) to provide a high degree of confidence that the instrument is operating properly. Operational Verification is useful for incoming inspection, routine maintenance, and after instrument repair.
If you are not familiar with operating the instrument, you should review the Chapter 1, “Getting Started,” in the Agilent 53147A/148A/149A Operating Guide. However, the procedures in this chapter are written so that little experience is necessary. These procedures should be followed in the order in which they appear.

Power-On Self Tests

NOTE This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Inspect the instrument for damage.
2 Make sure no cables are connected to the instrument’s inputs.
3 Connect the power cord to the instrument and the power source.
NOTES Since the instrument’s power supply automatically senses the line
voltage, there is no AC input-voltage setting.
It is normal for the fan in the instrument to run when the instrument is in Standby mode. Power is supplied to the timebase whenever the power cord is connected to maintain long-term measurement reliability, and the fan helps to maintain the timebase’s temperature stability.
4 Press and release the Power button on the front panel.
1-6 Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
5 Verify that the front-panel display shows the following:
All segments of the front-panel display are temporarily activated.
TESTING is displayed.
SELF TEST OK is displayed.
The model number of the instrument is displayed (a four-digit hex
number is also displayed).
GPIB ADDR nn (nn = a two digit number from 0 to 30) is displayed.
CH2 NO SIGNAL is displayed.
6 If an error message is displayed, refer to the troubleshooting section in
Chapter 2, “Service.”
7 Mark Pass or Fail in the Performance Test Record on page 1-39, Test 1.
NOTE To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment and the timebase to stabilize.
1
Assembly Level Service Guide 1-7
Chapter 1 Performance Tests
Operational Verification

10 MHz Test

1 Connect a BNC-to-BNC cable from the Reference 10MHz connector on
the instrument’s rear panel to the CHANNEL 1 input connector on the
1
front panel.
2 Press and release the Chan Select key.
3 Verify that Ch 1 is displayed in the upper-left corner of the display.
4 Verify that the display reads 10,000,000 Hz.
5 Mark Pass or Fail in the Performance Test Record on page 1-39, Test 2.
1-8 Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification

Run Self Test

NOTE This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Disconnect all signal cables from the input and Reference 10MHz connectors.
2 Press and release the Shift key, and then press and release the Menu
(Reset/Local) key.
3 Press the up- and/or down-arrow key(s) as many times as necessary until
DO SELF TEST is displayed.
4 Press and release the Enter key.
5 If no errors are detected, SELF TEST OK is briefly displayed when the self
test is completed. If any error messages are displayed, refer to the troubleshooting section in Chapter 2, “Service.”
6 Mark Pass or Fail in the Performance Test Record on page 1-39, Test 3.
1
Assembly Level Service Guide 1-9
Chapter 1 Performance Tests
Operational Verification

External Timebase Test

This test verifies the instrument’s 10 MHz external timebase specification by verifying that setting REF OSC to EXT allows the instrument to be
1
NOTES To ensure that the test results are valid, the instrument and the test
synchronized to an external reference signal.
Equipment Required
Agilent 3325A/B Synthesizer (2) Agilent 10100C 50 Feedthrough Terminator Agilent 10503 Series Coaxial Cables (BNC m to m) (2)
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Connect an Agilent 10503 series coaxial cable between the output of the
first Agilent 3325A/B Synthesizer and the Reference 10MHz connector on the instrument’s rear panel (see Figure 1-1).
2 Connect an Agilent 10100C 50Ω Feedthrough Terminator to the Counter’s
Channel 1 input connector.
3 Connect an Agilent 10503 series coaxial cable between the output of the
second Synthesizer and the feedthrough terminator on the Counter’s CHANNEL 1 input connector.
1-10 Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification
To Reference
10 MHz Connector
(Rear Panel)
Agilent 3325A/B Synthesizer
1
10 MHZ Output
Agilent 53147A/48A/49A
Agilent
Channel 1 (with 10100C 50 Feedthrough Attached)
Agilent 3325A/B Synthesizer
2 MHz Output
Figure 1-1. External Timebase Test Setup
4 Cycle the POWER button to preset the instrument.
5 Press the Shift key on the instrument’s front panel.
6 Press the Menu key on the instrument’s front panel.
7 If necessary, use the up- and down-arrow keys to cycle the display until
it reads REF OSC> INT.
8 Press the right-arrow key once. INT begins to flash off and on.
9 Press the up- or down-arrow key once. INT changes to EXT.
10 Press the Enter key.
11 Press the Chan Select key once to select the CHANNEL 1 input connector.
Assembly Level Service Guide 1-11
Chapter 1 Performance Tests
Operational Verification
12 Set the frequency of the first Synthesizer (connected to the instrument’s
rear-panel Reference 10MHz connector) to 10 MHz at 1 V rms.
1
13 Set the frequency of the second Synthesizer (connected to the instrument’s
Channel 1 input connector) to 2 MHz at 100 mV rms.
14 Press and release the instrument’s Reset/Local key.
15 Verify that the Counter displays a reading of approximately 2 MHz and
that the Ext Ref annunciator is activated.
16 Mark Pass or Fail for Test 4 in the Performance Test Record on page 1-39.
17 Disconnect the test setup.
1-12 Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification

Power Meter Test

This test verifies the accuracy of the Power Meter.
Equipment Required
Agilent 11730A Power Sensor Cable (supplied with instrument) Agilent 8481A, 8481D, 8482A, 8485A, or 8487A Power Sensor Head
NOTES To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Connect one end of the Agilent 11730A cable (supplied with instrument)
to the Power Meter INPUT connector on the front panel of the Agilent 53147A, 53148A, or 53149A.
1
2 Connect the other end of the Agilent 11730A cable to a power sensor head.
3 Press the Display Power key in the Power Meter area of the front panel.
The display should look like this (the reading will vary):
Ch 2
Freq
dB
dBm
Pwr
Watts mW uW %
4 If the power sensor head you are using is an Agilent 8481A, 8481D,
8482A, 8485A, or 8487A, select the power sensor head model in the HEAD menu. If you are not using one of the power sensor heads listed above, press the Cal Factor key, and enter the calibration factor for the head.
Assembly Level Service Guide 1-13
Chapter 1 Performance Tests
Operational Verification
5 Press the Zero key.
The display shows ZEROING for several seconds.
1
6 Connect the power sensor head to the Power Meter OUTPUT connector.
7 Press the Cal key.
The display briefly shows CALIBRATING.
8 Turn on the power reference output in the instrument’s menu
(PWR REF > ON).
9 Verify that the Counter displays a reading of 0 dBm ±.02 dB.
10 Mark Pass or Fail for Test 5 in the Performance Test Record on page 1-39.
11 Disconnect the test setup.
1-14 Assembly Level Service Guide
Chapter 1 Performance Tests
Operational Verification

DVM Test

This test verifies the accuracy of the DVM.
Equipment Required
Agilent 34401A Digital Multimeter Agilent E3640A DC Power Supply
NOTES To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Set the output of the E3640A DC Power Supply to +25 VDC.
2 Set the Agilent 34401A Digital Multimeter to the lowest DC voltage range
capable of safely measuring 25 VDC.
1
3 Connect the test leads to the Multimeter (observe correct polarity).
4 Connect the test prods to the output terminals on the Agilent E3640A DC
Power Supply (observe correct polarity).
Agilent E3640A Power Supply
Agilent
Agilent 34401A Digital Multimeter
Agilent
Agilent 53147A/48A/49A
Agilent
Agilent E3640A Power Supply
Agilent
Figure 1-2. DVM Test Setup
Assembly Level Service Guide 1-15
Chapter 1 Performance Tests
Operational Verification
5 Observe and record the voltage reading on the Agilent 34401A Multimeter.
6 Disconnect the test leads from the Agilent E3640A DC Power Supply.
1
Be careful not to disturb the voltage setting.
7 Connect the test leads to the Agilent 53147A/48A/49A (be sure to observe
the correct polarity).
8 Press the Display DVM key on the Agilent 53147A/48A/49A front panel.
9 Connect the test prods to the output terminals on the Agilent E3640A
DC Power Supply (be sure to observe the correct polarity).
10 Observe and record the voltage reading on the Agilent 53147A/48A/49A.
11 Compare the voltage readings you recorded in steps 5 and 10. If the
reading taken with the Agilent 53147A/48A/49A is within ±0.25% ±10 mV of the reading taken with the Agilent 34401A, mark Pass for Test 6 in the Performance Test Record on page 1-39. If the reading taken with the Agilent 53147A/48A/49A is not within ±0.25% ±10 mV of the reading taken with the Agilent 34401A, mark Fail for Test 6 in the Performance Test Record on page 1-39.
This completes the Operational Verification.
1-16 Assembly Level Service Guide
Chapter 1 Performance Tests

Complete Performance Tests

Complete Performance Tests
The Complete Performance Tests verify the specifications of the Agilent 53147A/148A/149A listed in Chapter 6, “Specifications.” All of these tests can be performed without opening the instrument. Table 1-2 lists a summary of the performance tests.
Record the results of the performance tests in the appropriate place on the Performance Test Record, which starts on page 1-39.
NOTE To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment and the timebase to stabilize.
Table 1-2. Complete Performance Tests
Page Number Test Description
Page 1-18 Test 1: Channel 1 Frequency Sensitivity
Page 1-23 Test 2: Channel 2 Frequency Sensitivity
Page 1-27 Test 3: Power Measurement
Page 1-30 Test 4: Power Reference Oscillator Frequency
Page 1-32 Test 5: Power Meter Reference Level
1
NOTE Other instrument measurement functions (e.g., Averaging) are
mathematically derived by the microprocessor from the parameters verified by these performance tests. If the instrument passes the performance tests, the other measurement functions are also functioning to specifications.
Assembly Level Service Guide 1-17
Chapter 1 Performance Tests

Test 1: Channel 1 Frequency Sensitivity

Test 1: Channel 1 Frequency Sensitivity
1
This set of tests verifies the frequency-sensitivity specifications of Channel 1 of the Agilent 53147A/148A/149A’s Frequency Counter.
Equipment Required
Agilent 3325A/B Synthesizer Agilent 83650B Synthesizer Agilent 437B Power Meter Agilent 8487A/D Power Sensor Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A) Agilent 11667C Power Splitter Agilent 8490D (Opt. 020) 20 dB Attenuator Agilent 1250-2015 BNC (m) to SMA (f) Adapter Agilent 11904A 2.92 mm (m) to 2.4 mm (m) Adapter Agilent 10100C 50 Feedthrough Terminator Agilent 10503 Series Coaxial Cable (BNC m to m) 2 mm (m) to 2 mm (f) Cable (2)

Test 1a: 10 Hz to 20 MHz Frequency Sensitivity

NOTES To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
If the Agilent 3325A/B Synthesizer or the Agilent 83650B Synthesizer
reference output is not adequate to drive the 53147A/48A/49A, connect the cable from the Synthesizer to the Reference 10 MHz connector on the back of the 53147A/48A/49A, and set the 53147A/48A/49A to use the internal oscillator.
1-18 Assembly Level Service Guide
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
1 Connect an Agilent 10503 Series Coaxial Cable between the reference
output connector on the Agilent 3325A/B Synthesizer and the instrument’s Reference 10MHz connector.
2 Connect an Agilent 10100C 50Ω Feedthrough Terminator to the Counter’s
CHANNEL 1 connector.
3 Connect an Agilent 10503 Series Coaxial Cable between the RF output
connector on the Agilent 3325A/B Synthesizer and the 50 feedthrough terminator on the Counter’s CHANNEL 1 connector.
1
To Reference Output (Rear Panel)
Agilent 3325A/B Synthesizer
To Reference
Output
(Rear
Agilent 53147A/48A/49A
Agilent
Panel)
Channel 1
Output
Figure 1-3. 10 Hz to 20 MHz Frequency Sensitivity Test Setup
4 Verify that the Counter is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-11).
5 Set the output of the Synthesizer to the first frequency specified for
Test 1a in the Performance Test Record on page 1-39 (10 Hz).
6 Set the power output of the Synthesizer to the first power value specified
for Test 1a in the Performance Test Record on page 1-39 (–14.9 dBm/40 mV).
7 Record the frequency value read on the Counter (±1 count) for Test 1a in
the Performance Test Record on page 1-39.
Assembly Level Service Guide 1-19
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
8 Repeat steps 5, 6, and 7 for each of the remaining frequencies specified for
Test 1a. Don’t forget to change the power level when performing the 1 KHz test (–19.2 dBm/25 mV).
1
9 If all of the frequency values you recorded for Test 1a are correct, mark
Pass in the Performance Test Record for Test 1a. If any of the frequency readings you recorded are incorrect, mark Fail in the Performance Test Record.
NOTES Do not disconnect the cable from the instrument’s Reference 10MHz
connector. This connection is used in the following test.
Do not turn off the instrument. Turning the instrument off and back on resets the reference oscillator (REF OSC) selection to internal (INT) and the channel selection to CHANNEL 2.

Test 1b: 50 MHz to 125 MHz Frequency Sensitivity

1 Connect an Agilent 10100C 50 Feedthrough Terminator to the
CHANNEL 1 input connector on the instrument’s front panel (refer to Figure 1-4 for steps 1 through 9).
Agilent 83650B Synthesizer
Reference
Output
(Rear Panel)
Agilent 437B Power Meter
Agilent 53147A/48A/49A
Agilent
To Reference 10 MHz
Connector
(Rear Panel)
Agilent
Feedthrough
Channel 1
RF Output
10100C 50
Agilent
Agilent
Agilent
1250-2015 BNC to SMA Adapter
8490D 20 dB Attenuator
11904A 2.92 mm to 2.4 mm Adapter
Agilent
11667C
Power Splitter
Agilent
Power Sensor
Input
8487A/D
Figure 1-4. 50 MHz to 125 MHz Frequency Sensitivity Test Setup
1-20 Assembly Level Service Guide
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
2 Connect the Agilent 1250-2015 BNC (m) to SMA (f) Adapter to the
Agilent 10100C Feedthrough Terminator on the Counter’s CHANNEL 1 connector.
3 Connect the Agilent 11904A 2.92 mm (m) to 2.4 mm (m) Adapter to the
Agilent 1250-2015 Adapter.
4 Connect the Agilent 8490D 20 dB Attenuator to the Agilent 11904A
Adapter.
5 Connect one of the output connectors on the Agilent 11667C Power
Splitter to the Agilent 8490D Attenuator
6 Connect the Agilent 11900B 2.4 mm (f) to 2.4 mm (f) Adapter to the RF
output of the Synthesizer.
7 Connect a 2.4 mm (m) to 2.4 mm (m) cable between the Adapter on the
Synthesizer’s RF output connector and the input connector on the Agilent 11667C Power Splitter.
1
8 Connect the Agilent 8487A/D Power Sensor to the remaining output
connector on the Agilent 11667C Power Splitter.
9 Connect the Agilent 11730A Power Sensor Cable between the Power
Sensor and the sensor connector on the Agilent 437B Power Meter.
10 Verify that the Counter is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-11).
11 Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 1b in the Performance Test Record on page 1-39 (50 MHz).
12 Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 1b in the Performance Test Record on page 1-39 (+.8 dBm). (The added 20 dBm compensates for the 20 dB Attenuator.)
Assembly Level Service Guide 1-21
Chapter 1 Performance Tests
Test 1: Channel 1 Frequency Sensitivity
13 Record the frequency value read on the Counter (±1 count) for Test 1b in
the Performance Test Record on page 1-39.
1
NOTES Do not disconnect the rest of the test setup, as these connections are
14 Set the output of the Synthesizer to 125 MHz, and repeat steps 11
through 13.
15 If both of the frequency values you recorded for Test 1b are correct, mark
Pass in the Performance Test Record for Test 1b. If either of the frequency readings you recorded are incorrect, mark Fail in the Performance Test Record.
16 Disconnect the Agilent 10100C Feedthrough Terminator, the
Agilent 1250-2015 Adapter, and the Agilent 11904A Adapter from the Counter’s CHANNEL 1 connector.
also used in the Channel 2 Frequency Sensitivity test.
Do not turn off the instrument. Turning the instrument off and back on resets the reference oscillator (REF OSC) selection to internal (INT).
1-22 Assembly Level Service Guide
Chapter 1 Performance Tests

Test 2: Channel 2 Frequency Sensitivity

Test 2: Channel 2 Frequency Sensitivity
This set of tests verifies the frequency-sensitivity specifications of Channel 2 of the Agilent 53147A/148A/149A’s Frequency Counter.
Equipment Required
Agilent 83650B Synthesizer Agilent 437B Power Meter Agilent 8487A/D Power Sensor Agilent 11667C Power Splitter Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A) Agilent 8490D (Opt. 020) 20 dB Attenuator Agilent 11904D 2.92 mm (m) to 2.4 mm (f) Adapter Agilent 10503 series Coaxial Cable (BNC m to m)
2.4 mm (m) to 2.4 mm (m) Cable (2)

Test 2a: 50 MHz to 20 GHz Frequency Sensitivity

NOTES To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Connect the Agilent 11904D 2.92 mm (m) to 2.4 mm (f) Adapter to the
Counter’s CHANNEL 2 input connector (refer to Figure 1-5 for steps 1 through 4).
1
2 Connect the Agilent 8490D Attenuator to the Agilent 11904D Adapter,
as shown in Figure 1-5 (the other end of the Attenuator should remain connected to the Agilent 11667C Power Splitter).
3 Verify that the cable (with Adapter) is still connected between the
RF output of the Agilent 83650B Synthesizer and the Power Splitter input connector.
Assembly Level Service Guide 1-23
Chapter 1 Performance Tests
Test 2: Channel 2 Frequency Sensitivity
4 Verify that the Power Sensor is still connected to the Power Splitter and
that the Power Sensor Cable is still connected between the Power Sensor and the sensor connector on the Power Meter.
1
Agilent 83650B Synthesizer
To Reference
Output
(Rear Panel)
Agilent 53147A/48A/49A
Agilent
To Reference 10 MHz
Connector
(Rear Panel)
Channel 2
RF Output
HP 8490D 20 dB Attenuator (50 MHz to 40 GHz tests only)
HP 11904D Adapter
Figure 1-5. 50 MHz to 20 GHz Frequency Sensitivity Test Setup
5 Verify that the instrument is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-11).
6 Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 2a in the Performance Test Record on page 1-40 (50 MHz).
Agilent 437B Power Meter
Input
HP 11667C Power Splitter
HP 8487A/D Power Sensor
7 Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 2a in the Performance Test Record on page 1-40 (0 dBm). (The added 20 dBm compensates for the 20 dB Attenuator.)
8 Record the frequency value read on the Counter (±1 count) for Test 2a in
the Performance Test Record on page 1-40.
1-24 Assembly Level Service Guide
Chapter 1 Performance Tests
Test 2: Channel 2 Frequency Sensitivity
9 Repeat steps 6, through 8 for each of the remaining frequencies specified
for Test 2a. Don’t forget to change the power level when performing the 300 MHz and 19 GHz tests (53147A, 53148A, and 53149A) and the 16 GHz and 19 GHz tests (53149A only).
NOTE It may be necessary to repeat Step 5.
10 If all of the frequency values you recorded for Test 2a are correct, mark
Pass in the Performance Test Record. If any of the frequency readings you recorded are incorrect, mark Fail in the Performance Test Record.

Test 2b: 22 GHz to 26.5 GHz Frequency Sensitivity

NOTE This test is appropriate for the Agilent 53148A and 53149A.
1 Use the same test setup as described for Test 2a on page 1-23.
2 Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 2b in the Performance Test Record on page 1-40 (22 GHz).
1
3 Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 2b in the Performance Test Record on page 1-40 (–5 dBm for the Agilent 53148A; –7 dBm for the Agilent 53149A). (The added 20 dBm compensates for the 20 dB Attenuator.)
4 Record the frequency value read on the Counter (±1 count) for Test 2b in
the Performance Test Record on page 1-40.
5 Repeat steps 2, through 4 for each of the remaining frequencies specified
for Test 2b.
6 If all of the frequency values you recorded for Test 2b are correct, mark
Pass in the Performance Test Record for Test 2b. If any of the frequency readings you recorded are incorrect, mark Fail in the Performance Test Record.
Assembly Level Service Guide 1-25
Chapter 1 Performance Tests
Test 2: Channel 2 Frequency Sensitivity

Test 2c: 30 GHz to 46 GHz Frequency Sensitivity

NOTE This test is appropriate for the 53149A only.
1
1 Use the same test setup as described for Test 2a on page 1-23.
2 Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 2c in the Performance Test Record on page 1-40 (30 GHz).
3 Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 2c in the Performance Test Record on page 1-40 (–3 dBm). (The added 20 dBm compensates for the 20 dB Attenuator.)
4 Record the frequency value read on the Counter (±1 count) for Test 2c in
the Performance Test Record on page 1-40.
5 Repeat steps 2, through 4 for each of the remaining frequencies specified
for Test 2c. Don’t forget to change the power level for the 42, 44, and 46 GHz tests.
6 If all of the frequency values you recorded for Test 2c are correct, mark
Pass in the Performance Test Record for Test 2c. If any of the frequency readings you recorded are incorrect, mark Fail in the Performance Test Record.
NOTES Do not disconnect the Power Sensor or the Power Sensor Cable from
the power meter or the Power Splitter. These connections are reused in the following tests.
Do not turn off the instrument. Turning the instrument off and back on resets the reference oscillator (REF OSC) selection to internal (INT).
1-26 Assembly Level Service Guide
Chapter 1 Performance Tests

Test 3: Power Measurement

Test 3: Power Measurement
This set of tests verifies the power-measurement specifications of Agilent 53147A/148A/149A’s Power Meter.
Equipment Required
Agilent 11683A Range Calibrator Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A)

Power Measurement Accuracy Test

NOTES To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Connect the Agilent 11730A Power Sensor Cable between Agilent 11683A
Range Calibrator’s output connector and the Agilent 53147A, 53148A, or 53149A’s Power Meter INPUT connector.
Agilent 53147A/48A/49A
Agilent
Agilent 11683A
Range Calibrator
1
Power Input
Output
Figure 1-6. Power Measurement Accuracy Test Setup
2 Press the Reset/Local key on the Agilent 53147A, 53148A, or 53149A.
Assembly Level Service Guide 1-27
Chapter 1 Performance Tests
Test 3: Power Measurement
3 Press the Shift + dBm/W key on the Agilent 53147A, 53148A, or 53149A to
set the Power Meter to display readings in Watts, mW, and/or µW.
1
NOTE The Range Calibrator output level is adjustable in 5 dB increments.
4 Set the Range Calibrator switches as follows:
FUNCTION: STANDBY POLARITY: NORMAL RANGE: 1 mW LINE: ON
5 Press the Zero key on the Agilent 53147A, 53148A, or 53149A. When the
reading appears, verify that it is 0.00 ±0.06 µW.
6 Change the Range Calibrator’s FUNCTION switch to CALIBRATE.
7 Press the Cal key on the Agilent 53147A, 53148A, or 53149A. Wait for the
Power Meter to complete calibration.
8 Verify that the Agilent 53147A, 53148A, or 53149A display reads
1.000 ±0.006 mW.
Thus, the 3 µW, 30 µW, 300 µW, 3 mW, and 30 mW legends on the RANGE switch are approximations. The true values for these settings are 3.16 µW,
31.6 µW, 316 µW, 3.16 mW, and 31.6 mW.
9 Set the Range Calibrator RANGE switch to each of the positions shown in
Table 1-3. For each setting, verify that the Agilent 53147A, 53148A, or 53149A autoranges properly and that the reading displayed is within the limits shown in the table.
NOTES The lower the setting of the RANGE switch, the longer it takes for
the reading to stabilize.
You may have to repeat Step 5 (re-zero the instrument) for each measurement.
1-28 Assembly Level Service Guide
Chapter 1 Performance Tests
Test 3: Power Measurement
Table 1-3. Range Calibrator Settings and Expected Results
Range Calibrator Agilent 53147A/48A/49A
RANGE Switch Setting Minimum Actual Maximum
3 µW 3.10 µW 3.23 µW
10 µW 9.90 µW 10.10 µW
30 µW 31.4 µW 31.8 µW
100 µW 99.5 µW 100.5 µW
300 µW 0.314 mW 0.318 mW
1 mW 0.995 mW 1.005 mW
3 mW 3.14 mW 3.18 mW
10 mW 9.95 mW 10.05 mW
30 mW 31.4 mW 31.8 mW
100 mW 99.5 mW 100.5 mW
1
10 Record the actual power readings in the Performance Test Record on
page 1-41.
11 Press the Shift + dBm/W key on the Agilent 53147A, 53148A, or 53149A
again to set the Power Meter to display readings in dBm.
12 Verify that the annunciator changes to dBm and that the indication is
within 20.00 ±00.04 dBm with the Range Calibrator’s RANGE switch in the 100 mw / 20 dBm position.
13 Record the actual power reading in the Performance Test Record.
14 Set the Range Calibrator’s RANGE switch to –10 dBm.
15 Verify that the Power Meter displays a reading of –10.00 ±0.04 dBm.
16 Record the actual power reading in the Performance Test Record.
17 Review all of the results for Test 3, and mark Pass or Fail for Test 3 in the
Performance Test Record.
Assembly Level Service Guide 1-29
Chapter 1 Performance Tests

Test 4: Power Reference Oscillator Frequency

Test 4: Power Reference Oscillator Frequency
1
The internal power reference oscillator is used to calibrate the Power Meter to the power sensor head. This test verifies that the frequency output of the power reference oscillator is 50.0 MHz ±0.5 MHz.
Equipment Required
Agilent 53150A Microwave Frequency Counter Agilent 10503 Series Coaxial Cables (BNC m to m) Agilent 1250-1535 BNC (f) to Type N (m) Adapter

Power Reference Oscillator Frequency Test

NOTES The instrument and the test equipment must be powered on for at least
30 minutes prior to measuring the power reference oscillator output. This allows the internal temperatures of the equipment to stabilize.
This procedure specifies the Agilent 53150A for measuring the frequency output of the 53147A/48A/49A’s power reference oscillator. If the 53147A/48A/49A’s Frequency Counter is known to be in calibration (see “Test 1: Channel 1 Frequency Sensitivity” on page 1-18), you can use it in place of the 53150A in this procedure.
1 Connect an Agilent E9635A adapter to the Agilent 53147A/48A/49A’s
Power Meter OUTPUT connector as shown in Figure 1-7.
2 Connect an Agilent 10503 series cable between the Agilent E9635A
adapter on the Agilent 53147A/148A/149A’s Power Meter OUTPUT connector and the CHANNEL 1 input connector on the Agilent 53150A/51A/52A Frequency Counter.
1-30 Assembly Level Service Guide
Chapter 1 Performance Tests
Test 4: Power Reference Oscillator Frequency
Agilent E9635A Adapter on Power Reference Oscillator
Output Connector Agilent 53147A/48A/49A
Agilent
Agilent 53150A/51A/52A Microwave Frequency Counter
53150A
Agilent
POWER
Rel Freq
Avg On Rel Pwr
Standby
Freq
Menu
Offset
Rate
Reset/ Local
On/Off
Shift
Clear
GHz MHz kHz Hz
Ch 12
Offset
Offset
MODIFY
Pwr
HPIB
Offset
Resol
Avg
Enter
+/-
20 GHz Counter
CHANNEL 1
10 Hz to 125 MHz
1M
Ext Rel Rate Rmt SRQ
Error Shift
Gate
Chan Select
Rel Freq
Offset On/Off
Hold
DAMAGE
+30 dBm
CHANNEL 2POWERFREQ
Channel 2 50 MHz to 20 GHz
dBm/ W
DAMAGE
+27 dBm
Display Power
Rel Pwr
Offset On/Off
50
dB dBm Watts mW uW %
Channel 1
Input
Figure 1-7. Reference Oscillator Frequency Adjustment Setup
3 Set the Agilent 53150A/51A/52A Frequency Counter to measure frequency
on Channel 1.
4 Turn on the Agilent 53147A/48A/49A’s internal power reference oscillator:
1
Press Shift + Menu, use the up and/or down arrow keys to cycle through the menu to PWR REF, press the right-arrow key, then use the up or down arrow key to change the setting to ON, and press the Enter key. The Pwr Ref annunciator should now be activated.
NOTE The reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete. Setting PWR REF in the 53147A/148A/149A’s menu to ON turns the power reference oscillator on, and it remains on until the PWR REF menu setting is returned to OFF. When you calibrate the Power Meter with PWR REF set to ON, the power reference oscillator remains on after calibration is complete.
5 Record the power reading shown on the Agilent 53150A/51A/52A for
Test 4 in the Performance Test Record on page 1-41.
6 Mark Pass or Fail for Test 4 in the Performance Test Record.
Assembly Level Service Guide 1-31
Chapter 1 Performance Tests

Test 5: Power Meter Reference Level

Test 5: Power Meter Reference Level
1
Two methods for verifying the level of the Power Meter reference signal are provided in this section. Both methods provide acceptable results, and either one can be used.

Power Meter Reference Level: Method 1

Equipment Required
Agilent E4418B Power Meter Agilent 8481A Power Sensor Agilent 11730A Power Sensor Cable (supplied with 53147A/48A/49A)
NOTES To ensure that the test results are valid, the Agilent 53147A/48A/49A
and the test equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
The calibration of the Agilent E4418B Power Meter must be no more than one level removed from NIST (National Institute of Standards and Technology). If the E4418B Power Meter’s calibration is more than one level removed from NIST, the test results will be invalid.
This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Connect the Agilent 11730A Power Sensor Cable to the Agilent 8481A
Power Sensor (see Figure 1-8).
Output connector (with Agilent 8481A
Agilent E4418B Power Meter
Figure 1-8. E4418B Power Meter Zeroing and Calibration Test Setup
1-32 Assembly Level Service Guide
Power Sensor Attached)
Input
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
2 Connect the free end of the Agilent 11730A Power Sensor Cable to the
input connector on the Agilent E4418B Power Meter.
3 Connect the Agilent 8481A Power Sensor to the power reference output
connector of the Agilent E4418B Power Meter.
4 Zero the E4418B Power Meter.
5 Calibrate the E4418B Power Meter.
6 Disconnect the Agilent 8481A Power Sensor from the Agilent E4418B
Power Meter, and immediately connect it to the Power OUTPUT connector on the Agilent 53147A/48A/49A (see Figure 1-9).
Output connector (with Agilent 8481A
Power Sensor Attached)
Agilent 53147A/48A/49A
Agilent
Agilent E4418B Power Meter
Input
Figure 1-9. Power Meter Reference Level Test Setup
7 Record the power reading shown on the Agilent E4418B Power Meter for
Test 5 in the Performance Test Record on page 1-41.
1
8 Mark Pass or Fail for Test 5 in the Performance Test Record.
Assembly Level Service Guide 1-33
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level

Power Meter Reference Level: Method 2

Equipment Required
1
Agilent 432A Power Meter Agilent 478A-H75 or 478A-H76 Thermistor Mount Agilent 34401A Digital Multimeter Agilent 8120-1082 Power Sensor Cable
NOTES To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to beginning the tests. This allows the internal temperatures of the equipment to stabilize.
This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Turn the Agilent 432A Power Meter off. Do not turn it on until instructed
to do so.
2 Set the Agilent 34401A Digital Multimeter to measure resistance, and
connect the DVM test prods between the center pin of the V
RF
on the Agilent 432A Power Meter rear panel and pin 1 on the thermistor­mount end of the Agilent 8120-1082 Power Sensor Cable (see Figure 1-10 to locate pin 1).
2
1
connector
6
3
5
4
Figure 1-10. Agilent 8120-1082 Power Sensor Cable End View
3 Round off the resistance reading to two decimal places and keep a record
of this value as R. It is the internal bridge resistance (R) of the Agilent 432A Power Meter (this should be approximately 200.3 ohms).
1-34 Assembly Level Service Guide
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
4 Connect the Agilent 478A-H75 or 478A-H76 Thermistor Mount to the
Agilent 53147A/148A/149A Power Meter OUTPUT connector, as shown in Figure 1-11.
NOTE Do not use a cable between the Thermistor Mount and the
53147A/148A/149A Power Meter OUTPUT connector. Connect the Thermistor Mount directly to the OUTPUT connector.
Agilent 478A-H75 or
478A-H76 Thermistor Mount
(connected to Power
Meter OUTPUT connector)
Agilent 53147A/48A/49A
Agilent
Agilent 8120-1082 Power Sensor Cable
Agilent 432A
Power Meter
V
RF
V
COMP
Agilent 34401A Digital Multimeter
-
+
Agilent
1
+
-
Figure 1-11. Power Meter Reference Level Test Setup
5 Connect the Agilent 8120-1082 Power Sensor Cable between the Thermistor
Mount and the input connector on the Agilent 432A Power Meter.
6 Turn on the 432A Power Meter. Wait 30 minutes for the Power Meter and
the Thermistor Mount to stabilize before proceeding to the next step.
7 Set the 432A Power Meter’s RANGE switch to Coarse Zero.
8 Adjust the COARSE ZERO control on the 432A Power Meter’s front panel
to obtain a zero reading on the meter.
9 Set the 432A Power Meter’s RANGE switch to .01 mW / –20 dBm.
10 Press the FINE ZERO switch on the 432A Power Meter’s front panel down
to obtain a zero reading on the meter.
11 Set the 432A Power Meter’s RANGE switch to 1 mW.
Assembly Level Service Guide 1-35
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
12 Set the Digital Multimeter to measure microvolts.
CAUTION Ensure that the Agilent 34401A Digital Multimeter’s input leads are
1
isolated from chassis ground before performing the next step.
13 Connect the positive test prod from the Agilent 34401A Multimeter to
the V
connector on the Agilent 432A Power Meter rear panel.
COMP
NOTE The V
and VRF connectors are BNC female. You can use two BNC-to-
COMP
banana jack adapters and two BNC cables to make it easier to connect the Multimeter to the Power Meter.
14 Connect the negative test prod from the Agilent 34401A Digital
Multimeter to the V
connector on the 432A Power Meter’s rear panel.
RF
15 If the reading on the Multimeter is less than 400 microvolts, proceed to
step 16. If the reading is 400 microvolts or more, press the FINE ZERO toggle switch on the 432A Power Meter’s front panel down.
16 Observe the reading on the Multimeter, round off the reading to the
nearest microvolt, and keep a record of this value as V
.
0
17 Turn on the Agilent 53147A/148A/149A’s internal power reference
oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through the menu to PWR REF, press the right-arrow key, then use the up or down arrow key to change the setting to ON, and press the Enter key. The Pwr Ref annunciator should now be activated.
NOTE The reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete. When PWR REF in the 53147A/148A/149A’s menu is set to ON, the power reference oscillator is turned on, and it is left on after calibration.
18 Observe the reading on the Multimeter, and keep a record of this value
.
as V
1
1-36 Assembly Level Service Guide
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
19 Disconnect the Multimeter’s negative test lead from the 432A Power Meter’s
V
connector, and connect it to the 432A Power Meter’s chassis ground.
RF
20 Observe the reading on the Multimeter, and keep a record of this value
as V
COMP
.
21 Calculate the output level of the 53147A/148A/149A’s power reference
oscillator (P
) using the following formula:
rf
Prf =
2 V
( V1 – V0 ) + V
comp
4R (Calibration Factor)
0
2
– V
2
1
Where:
= Power reference oscillator output level
P
rf
V
comp
= previously recorded value = previously recorded value
V
1
V
= previously recorded value
0
R = previously recorded value
Calibration Factor = value for thermistor mount at 50 MHz (traceable to NIST)
1
Assembly Level Service Guide 1-37
Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
NOTE The following is an example showing how a calculation of the Power
Reference Output Level is performed:
1
V0=170 µV or 170 x 10–6 V
= 78,107 µV or 78,107 x 10–6 or .078 V
V
1
V
comp
= 5.1583 V
R = 200
R = previously recorded value
Calibration Factor = 99.68%
2
– V
2
1
Note that V
Prf =
2 V
is eventually dropped out of the equation in the following
0
( V1 – V0 ) + V
comp
0
4R (Calibration Factor)
steps, because its value is very small (effectively zero).
2
P
=
rf
4(200) (99.68%)
0.80469 – 0.00608
=
2 (5.1583) (.078 – 0)
+ (.000170)2 – (0.78)
0.79861
=
797.44 797.44
1.00147 mW
=
22 Verify that Prf is between 0.988 mW and 1.012 mW.
23 Record the value for P
in the Performance Test Record on page 1-41.
rf
24 Mark Pass or Fail for Test 5 in the Performance Test Record.
1-38 Assembly Level Service Guide
Chapter 1 Performance Tests

Performance Test Record (Page 1 of 3)

Performance Test Record (Page 1 of 3)
Agilent 53147A/148A/149A Frequency Counter/Power Meter/DVM
Serial Number: Test Performed By: Date:
Notes: Pre Calibration Test Post Calibration Test
Test
Number
1 Power on Self Tests N/A
2 10 MHz Test 10,000,000 Hz
3 Run Self Test N/A
4 10 MHz External Timebase Input Test 2 MHz 5 Power Meter Test 0 dBm ±.02 dB
6 DVM Test +25 Vdc ±0.25% ±10 mV
Test
Number
1a 10 Hz
Operational Verification Specifications
Complete
Performance Tests Specifications
Channel 1 Frequency Sensitivity
30 Hz 1 kHz 500 kHz 5 MHz 10 MHz 20 MHz
Repair/Work Order No. Temperature: Relative Humidity:
(+24.365 Vdc to +25.635 Vdc)
10 Hz @ 40 mV rms (–14.9 dBm) 30 Hz @ 40 mV rms (–14.9 dBm) 1 kHz @ 25 mV rms (–19.2 dBm)
500 kHz @ 25 mV rms (–19.2 dBm)
5 MHz @ 25 mV rms (–19.2 dBm) 10 MHz @ 25 mV rms (–19.2 dBm) 20 MHz @ 25 mV rms (–19.2 dBm)
1
Test Results Pass Fail
❏❏
❏❏
❏❏
❏❏
❏❏
❏❏
Test Results Pass Fail
❏❏
Channel 1 Frequency Sensitivity
1b 50 MHz
125 MHz
❏❏
50 MHz @ 25 mV rms (–19.2 dBm)
125 MHz @ 25 mV rms (–19.2 dBm)
Assembly Level Service Guide 1-39
Chapter 1 Performance Tests

Performance Test Record (Page 2 of 3)

Performance Test Record (Page 2 of 3)
1
Test
Number
Channel 2 Frequency Sensitivity
2a 50 MHz
100 MHz 250 MHz 300 MHz 500 MHz 1 GHz
2.5 GHz 5 GHz 10 GHz
12.4 GHz 16 GHz 18 GHz 19 GHz 20 GHz
Channel 2 Frequency Sensitivity
2b 22 GHz
24 GHz
26.5 GHz
Channel 2 Frequency Sensitivity
2c 30 GHz
34 GHz 40 GHz 42 GHz 44 GHz 46 GHz
Complete
Performance Tests Specifications
53147A 53148A 53149A ❏❏
–20 dBm
" "
–33 dBm
" " " " " " " "
–29 dBm
"
–20 dBm
" "
–33 dBm
" " " " " " " "
–29 dBm
"
53147A 53148A 53149A ❏❏
N/A N/A N/A
–25 dBm
" "
53147A 53148A 53149A ❏❏
N/A N/A N/A N/A N/A N/A
N/A N/A N/A N/A N/A N/A
–20 dBm
" "
–33 dBm
" " " " " "
–30 dBm
"
–27 dBm
"
–27 dBm
" "
–23 dBm
" "
–17 dBm
" "
Test Results Pass Fail
1-40 Assembly Level Service Guide
Chapter 1 Performance Tests

Performance Test Record (Page 3 of 3)

Performance Test Record (Page 3 of 3)
Test
Number
Power Measurement:
33 µW
100 µW 300 µW
10 mW 30 mW
100 mW
20.00 dBm
–10.00 dBm
4
5
Power Reference Frequency:
50.0 MHz
Power Reference Level:
1.0 mW (0.00 dBm)
Complete
Performance Tests Specifications
10 µW 30 µW
1 mW 3 mW
± 0.02 dB or ± 0.5%
" " " " " " " "
“ ±0.04 dBm ±0.04 dBm
50.0 MHz ± 1%
(49.5 MHz to 50.5 MHz)
1.0 mW ± 1.2%
0.988 mW to 1.012 mW
1
Test Results Pass Fail
❏❏
❏❏
❏❏
Assembly Level Service Guide 1-41
Chapter 1 Performance Tests
Performance Test Record (Page 3 of 3)
1
1-42 Assembly Level Service Guide
2
Service
Chapter 2 Service

Introduction

Introduction
This chapter provides service information for your Agilent 53147A/ 148A/149A. It is divided into four major sections:
Returning the Instrument to Agilent Technologies for Service (page 2-3). This section provides you with step-by-step instructions on how to return the instrument for service.
Calibration Procedures (page 2-6). This section provides step-by-
2
step procedures for calibrating the Agilent 53147A/148A/149A.
Pre-Troubleshooting Information (page 2-23). This section provides you with pertinent information such as safety considerations, recommended test equipment, repair and after-service considerations, and assembly identification and location.
Troubleshooting the Instrument (page 2-28). This section provides you with troubleshooting procedures that isolate the faulty assembly or module. Replacement and recalibration of most modules can only be performed at an authorized Agilent Technologies Service Center.
If the instrument is under warranty, return it to Agilent for service. Refer to “Returning the Instrument to Agilent Technologies for Service” on page 2-3. If you decide to troubleshoot the instrument yourself, refer to the section titled “Troubleshooting the Instrument” on page 2-28.
2-2 Assembly Level Service Guide
Chapter 2 Service

Returning the Instrument to Agilent Technologies for Service

Returning the Instrument to Agilent Technologies for Service

Providing Repair Information

Before shipping the instrument to an Agilent Technologies office for service or repair, call the nearest Agilent Sales Office to make arrangements. Then, tag and package the Agilent 53147A/148A/149A for shipment.
NOTE ISD (Instrument Service Division) Emergency Response or Express
Calibration Service is available for Agilent customers in the USA. If downtime is critical, you can receive your repaired instrument via overnight shipment. Call 800-403-0801, and ask for Emergency Response or Express Calibration Service. When your instrument is repaired, it is returned via overnight shipment at no extra charge.
1 Write the following information on a tag:
Owner’s name and address
Instrument model number
Complete serial number
Description of service required or failure indications
2 Attach the tag to the instrument.
3 Pack the instrument.
If the original packaging materials are available, use the procedure titled “Packing the Instrument in the Original Packaging Materials.” If the original packaging materials are not available, you can order new packaging materials from an Agilent Sales Office. The new packaging materials are identical to those used by the factory when packaging new instruments. To use commercially available packaging materials, use the procedure titled “Packing the Instrument in Commercially Available Packaging Materials.” Both procedures are on the following page.
2
Assembly Level Service Guide 2-3
Chapter 2 Service
Returning the Instrument to Agilent Technologies for Service

Packing the Instrument in the Original Packaging Materials

To prevent shipping damage, it is best to repack the instrument in its original packaging. In any correspondence, refer to the instrument by the model number and the complete serial number.
1 Disconnect the power cord, probes, cables, or other accessories attached to
the instrument.
2 Make sure the folded corrugated spacer (which normally contains the
manuals) is in the box to ensure proper fitting.
2
3 Make sure one of the polystyrene blocks is in its proper position in the box.
4 Place the instrument, front panel end first, on the polystyrene block.
5 Place the other polystyrene block on top of the rear end of the instrument
to secure it.
6 Do not return the manuals with the instrument. Return an accessory only
when it is a part of the failure symptoms.
7 Seal the shipping container securely.
2-4 Assembly Level Service Guide
Chapter 2 Service
Returning the Instrument to Agilent Technologies for Service

Packing the Instrument in Commercially Available Packaging Materials

If the original factory packaging materials are not available, use the following procedure to pack the instrument in commercially available shipping materials.
1 Wrap the instrument in heavy paper or plastic.
2 Place the instrument in a strong shipping container. Use a double-walled
carton made of at least 350 lb. test material.
3 Protect the control panel with cardboard.
4 Add a layer of shock-absorbing material that is 3 to 4 inches (70 to 100 mm)
thick around all sides of the instrument cushion it and to prevent it from moving within the container
5 Seal the shipping container securely.
6 Mark the shipping container FRAGILE.
2
Assembly Level Service Guide 2-5
Chapter 2 Service

Calibration Procedures

Calibration Procedures
There are four calibration procedures that can be performed outside of an Agilent authorized service facility. The following procedures describe how to calibrate the frequency of the instrument’s timebase, the frequency and output level of the power reference oscillator, and the DVM.

Equipment Required

Table 2-1 lists the test equipment and accessories needed to perform the
2
Table 2-1. Recommended Test Equipment and Accessories
calibration procedures in this chapter.
Model
Instrument Type Required Characteristics
Digital Multimeter 6½ digit AC/DC Agilent 34401A 1
Power Meter Range 1 mW Agilent 432A 1
Power Meter 100 kHz to 110 GHz, –70 dBm to +44 dBm Agilent E4418B 1
Frequency Counter 10 Hz to 125 MHz Agilent 53150A/51A/52A 1
Thermistor Mount SWR 1.05 at 50 MHz
Accuracy ±0.5% at 50 MHz
Power Sensor 10 MHz to 18 GHz, -30 dBm to +20 dBm Agilent 8481A 1
Power Sensor Cable 5 ft. Agilent 11730A 1
Range Calibrator 3, 10, 30, 100, and 300 µW
1, 3, 10, 30, and 100 mW
Serial Cable RS-232, DB-25 to RJ-45 Agilent 53150-60215 1
Precision DC Power Supply + 50 V / – 50 V Agilent 6613C or 6614C 1
Cable BNC (m) to BNC (m) Agilent 10503 series 1
Adapter Type N (m) to BNC (f) Agilent E9635A
* Equivalent equipment can be substituted for all instruments and accessories. ** OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting † One supplied with Agilent 53147A/48A/49A
Recommended *
Agilent 478A-H75 or 478A-H76
Agilent 11683A 1
Quantity Required
1
1
2-6 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures

10 MHz Timebase Calibration

Equipment Required
10 MHz Frequency Reference Standard Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A) Agilent 10503 series BNC (m) to BNC (m) Coaxial Cable Small, non-metallic adjustment tool for slotted screw heads
NOTE The instrument and the test equipment must be powered on for at least
30 minutes prior to beginning this calibration procedure. This allows the internal temperatures of the equipment to stabilize.
1 Connect an Agilent 10503 series coaxial cable from the output of your
10 MHz frequency standard to the Counter’s CHANNEL 1 input on the instrument’s front panel.
2 Press the Resol key, and verify that the resolution is set to 1 HZ.
If necessary change the resolution to 1 HZ.
3 If there is a label covering the calibration adjustment access hole,
remove it (see Figure 2-1).
2
Assembly Level Service Guide 2-7
Chapter 2 Service
Calibration Procedures
2
Calibration Access Hole
Figure 2-1. Calibrating the 10 MHz Frequency Reference Oscillator
4 Insert a non-metallic, slot-type adjustment tool through the hole near the
bottom of the instrument’s left side and into the slot in the timebase calibration screw.
5 Turn the calibration screw clockwise to increase the Counter’s reading or
counter-clockwise to decrease the reading until the frequency value displayed on the Counter’s front panel is exactly 10,000,000 Hz.
2-8 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures

Power Reference Oscillator Frequency Calibration

The internal power reference oscillator is used to calibrate the Power Meter to the power sensor head. This adjustment calibrates the frequency output of the power reference oscillator to 50.0 MHz ±0.5 MHz.
Equipment Required
Agilent 53150A Microwave Frequency Counter Agilent 10503 series BNC (m) to BNC (m) Coaxial Cable Agilent E9635A Type N (m) to BNC (f) Adapter Small, non-metallic adjustment tool for slotted screw heads
NOTES The instrument and the test equipment must be powered on for at least
30 minutes prior to adjusting the power reference oscillator output. This allows the internal temperatures of the equipment to stabilize.
This procedure specifies the Agilent 53150A for measuring the frequency output of the 53147A/48A/49A’s power reference oscillator. If the 53147A/48A/49A’s Frequency Counter is known to be in calibration (see “Test 1: Channel 1 Frequency Sensitivity” on page 1-18), you can use it in place of the 53150A in this procedure.
2
Adjusting the power reference oscillator frequency may also affect the output level of the oscillator. Therefore, after you adjust the frequency to 50.0 MHz ± 0.05 MHz, always check the output level and adjust it, if necessary, as described in “Power Reference Oscillator Level Adjustment” on page 2-13.
1 Turn off the Agilent 53147A/48A/49A, and disconnect the power cord.
2 Remove the shroud from the Agilent 53147A/48A/49A (see “Removing the
Shroud” on page 3-7).
3 Position the Agilent 53147A/48A/49A on its right side so that you have
access to the front panel and the left side of the main circuit board (see Figure 2-2).
Assembly Level Service Guide 2-9
Chapter 2 Service
p
Calibration Procedures
Agilent 53147A/48A/49A
Agilent
Agilent E9635A Adapter on Power Reference Oscillator
Output Connector
Agilent 53150A/51A/52A Microwave Frequency Counter
53150A
Agilent
GHz MHz kHz Hz
POWER
Ch 12
Rel Freq
Offset Avg On Rel Pwr
2
Standby
Freq
Menu
Offset
Reset/
Rate
Local
On/Off
Shift
Clear
dB dBm Watts mW uW
Offset
%
MODIFY
Pwr
HPIB
Offset
Resol
Avg
Enter
+/-
20 GHz Counter
Ext Rel Rate Rmt SRQ
Error Shift
Gate
Chan Select
Rel Freq
Offset On/Off
CHANNEL 1
10 Hz to 125 MHz
Hold
DAMAGE
+30 dBm
CHANNEL 2POWERFREQ
Channel 2 50 MHz to 20 GHz
dBm/ W
DAMAGE
+27 dBm
Display Power
Rel Pwr
Offset On/Off
1M
50
Channel 1
ut
In
Figure 2-2. Reference Oscillator Frequency Adjustment Setup
4 Connect the power cord to the Agilent 53147A/48A/49A and to a
power outlet.
WARNING DANGEROUS VOLTAGES ARE PRESENT INSIDE THE AGILENT
53147A/148A/149A WHEN IT IS CONNECTED TO A POWER SOURCE. WORK VERY CAREFULLY AND AVOID CONTACT WITH INTERNAL COMPONENTS WHEN THE SHROUD IS REMOVED AND POWER IS APPLIED TO PREVENT INJURY OR DEATH FROM ELECTRICAL SHOCK.
CAUTION Some internal components of the Agilent 53147A/48A/49A are subject to
damage from electrostatic discharge (ESD), physical contact, and/or electrical short circuits. Work very carefully when the shroud is removed, and avoid all unnecessary contact with the internal components.
2-10 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
5 Turn on the Agilent 53147A/48A/49A and then wait for at least
30 minutes to allow the internal temperature of the instrument to stabilize.
6 Connect an Agilent E9635A adapter to the Power Meter OUTPUT connector
as shown in Figure 2-2.
7 Connect an Agilent 10503 series cable from the Agilent E9635A adapter
on the Agilent 53147A/148A/149A’s Power Meter OUTPUT connector to the CHANNEL 1 input connector on the Agilent 53150A/51A/52A Frequency Counter.
8 Set the Agilent 53150A/51A/52A Frequency Counter to measure frequency
on Channel 1.
9 Turn on the Agilent 53147A/148A/149A’s internal power reference oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through the menu to PWR REF, press the right-arrow key, then use the up or down arrow key to change the setting to ON, and press the Enter key. The Pwr Ref annunciator should now be activated.
NOTE The reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete. Setting PWR REF in the 53147A/148A/149A’s menu to ON turns the power reference oscillator on, and it remains on until the PWR REF menu setting is returned to OFF. When you calibrate the Power Meter with PWR REF set to ON, the power reference oscillator remains on after calibration is complete.
10 Insert a small, non-conductive, slot-type adjustment tool into the hole
in the side of the power reference oscillator shield (see Figure 2-3). This opening allows access to L9, the adjustable inductor that is used to adjust the frequency of the power reference oscillator.
2
Assembly Level Service Guide 2-11
Chapter 2 Service
Calibration Procedures
L
(Frequency Adjustment)
R
2
98
(Power Adjustment)
9
Figure 2-3. Power Reference Oscillator Adjustment Locations
11 Set the frequency to 50.0 MHz ±0.05 MHz (49.95 MHz to 50.05 MHz).
Turn the adjustment tool counter-clockwise to increase the frequency or clockwise to decrease the frequency.
2-12 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
Power Reference Oscillator Level Adjustment
Two methods for calibrating the level of the Power Meter reference signal are provided in this section. Both methods provide acceptable results, and either one can be used.
Method 1
Equipment Required
Agilent E4418B Power Meter Agilent 8481A Power Sensor Agilent 11730A Power Sensor Cable (supplied with 53147A/48A/49A)
NOTES To ensure that the results are valid, the Agilent 53147A/48A/49A and
the test equipment should be powered on for at least 30 minutes prior to beginning the calibration procedure. This allows the internal temperatures of the equipment to stabilize.
The calibration of the Agilent E4418B Power Meter must be no more than one level removed from NIST (National Institute of Standards and Technology). If the E4418B Power Meter’s calibration is more than one level removed from NIST, the test results will be invalid.
2
This test is appropriate for the Agilent 53147A, 53148A, and 53149A.
1 Connect the Agilent 11730A Power Sensor Cable to the Agilent 8481A
Power Sensor (see Figure 2-4).
Output connector (with Agilent 8481A
Agilent E4418B Power Meter
Power Sensor Attached)
Input
Figure 2-4. E4418B Power Meter Zeroing and Calibration Test Setup
Assembly Level Service Guide 2-13
Chapter 2 Service
Calibration Procedures
2 Connect the free end of the Agilent 11730A Power Sensor Cable to the
input connector on the Agilent E4418B Power Meter.
3 Connect the Agilent 8481A Power Sensor to the power reference output
connector of the Agilent E4418B Power Meter.
4 Zero the E4418B Power Meter.
5 Calibrate the E4418B Power Meter.
6 Disconnect the Agilent 8481A Power Sensor from the Agilent E4418B
Power Meter, and immediately connect it to the Power OUTPUT connector
2
on the Agilent 53147A/48A/49A (see Figure 2-5).
Output connector (with Agilent 8481A
Power Sensor Attached)
Agilent 53147A/48A/49A
Agilent
Agilent E4418B Power Meter
Input
Figure 2-5. Power Meter Reference Level Test Setup
7 Use a non-metallic, slot-type adjustment tool to adjust the setting of R98
on the main circuit board (see Figure 2-3) until the power level is as close to 1.000 mW as possible (the level must be between 0.988 mW and
1.012 mW). Turn the adjustment tool counter-clockwise to increase the power level or clockwise to decrease the power level.
2-14 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
Method 2
Equipment Required
Agilent 432A Power Meter Agilent 478A-H75 or 478A-H76 Thermistor Mount Agilent 34401A Digital Multimeter Agilent 8120-1082 Power Sensor Cable
NOTE To ensure that the results are valid, the Agilent 53147A/48A/49A and the
test equipment should be powered on for at least 30 minutes prior to beginning the calibration procedure. This allows the internal temperatures of the equipment to stabilize.
1 Make sure the Agilent 432A Power Meter is off. Do not turn it on until
instructed to do so.
2 Set the Agilent 34401A Digital Multimeter to measure resistance, and
connect the DVM test prods between the center pin of the V
connector
RF
on the Agilent 432A Power Meter rear panel and pin 1 on the thermistor­mount end of the Agilent 8120-1082 Power Sensor Cable (see Figure 2-6 to locate pin 1).
2
2
6
3
1
5
4
Figure 2-6. Agilent 8120-1082 Power Sensor Cable End View
3 Round off the resistance reading to two decimal places and keep a record
of this value as R. It is the internal bridge resistance (R) of the Agilent 432A Power Meter (this should be approximately 200.3 ohms).
4 Connect the Agilent 478A-H75 or 478A-H76 Thermistor Mount to the
Agilent 53147A/148A/149A Power Meter OUTPUT connector, as shown in Figure 2-7.
Assembly Level Service Guide 2-15
Chapter 2 Service
Calibration Procedures
NOTE Do not use a cable between the Thermistor Mount and the 53147A/
148A/149A Power Meter OUTPUT connector. Connect the Thermistor Mount directly to the OUTPUT connector.
Agilent 478A-H75 or
478A-H76 Thermistor Mount
(connected to Power
Meter OUTPUT connector)
Agilent 53147A/48A/49A
Agilent
Agilent 432A
Power Meter
Agilent 34401A
V
-
RF
Digital Multimeter
Agilent
2
V
COMP
Agilent 8120-1082 Power Sensor Cable
Figure 2-7. Power Meter Reference Level Calibration Setup
5 Connect the Agilent 8120-1082 Power Sensor Cable between the
Thermistor Mount and the input connector on the Agilent 432A Power Meter.
+
+
-
6 Turn on the 432A Power Meter. Wait 30 minutes for the Power Meter and
the Thermistor Mount to stabilize before proceeding to the next step.
7 Set the 432A Power Meter’s RANGE switch to Coarse Zero.
8 Adjust the COARSE ZERO control on the 432A Power Meter’s front panel
to obtain a zero reading on the meter.
9 Set the 432A Power Meter’s RANGE switch to .01 mW / –20 dBm.
10 Press the FINE ZERO control on the 432A Power Meter’s front panel down
to obtain a zero reading on the meter.
11 Set the 432A Power Meter’s RANGE switch to 1 mW.
12 Set the Digital Multimeter to measure microvolts.
2-16 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
CAUTION Ensure that the Agilent 34401A Digital Multimeter’s input leads are
isolated from chassis ground before performing the next step.
13 Connect the positive test prod from the Agilent 34401A Multimeter to the
V
connector on the Agilent 432A Power Meter rear panel.
COMP
NOTE The V
and VRF connectors are BNC female. You can use two BNC-to-
COMP
banana jack adapters and two BNC cables to make it easier to connect the Multimeter to the Power Meter.
14 Connect the negative test prod from the Agilent 34401A Multimeter to the
connector on the 432A Power Meter’s rear panel.
V
RF
15 If the reading on the Multimeter is less than 400 microvolts, proceed to
step 16. If the reading is 400 microvolts or more, press the FINE ZERO toggle switch on the 432A Power Meter’s front panel down.
16 Observe the reading on the Multimeter, round off the reading to the
nearest microvolt, and keep a record of this value as V
.
0
17 Turn on the Agilent 53147A/148A/149A’s internal power reference oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through the menu to PWR REF, press the right-arrow key, then use the up or down arrow key to change the setting to ON, and press the Enter key. The Pwr Ref annunciator should now be activated.
NOTE The reference oscillator is normally off. It is turned on automatically
during calibration and then turned back off when calibration is complete. Setting PWR REF in the 53147A/148A/149A’s menu to ON turns the power reference oscillator on, and it remains on until the PWR REF menu setting is returned to OFF. When you calibrate the Power Meter with PWR REF set to ON, the power reference oscillator remains on after calibration is complete.
2
18 Observe the reading on the Multimeter, and keep a record of this value
as V
.
1
Assembly Level Service Guide 2-17
Chapter 2 Service
Calibration Procedures
19 Disconnect the Multimeter’s negative test lead from the 432A Power Meter’s
V
connector, and connect it to the 432A Power Meter’s chassis ground.
RF
20 Observe the reading on the Multimeter, and keep a record of this value
as V
COMP
.
21 Calculate the output level of the 53147A/148A/149A’s power reference
oscillator (P
2
Where:
) using the following formula:
rf
( V1 – V0 ) + V
comp
4R (Calibration Factor)
V
comp
Prf =
2 V
= Power reference oscillator output level
P
rf
= previously recorded value = previously recorded value
V
1
V
= previously recorded value
0
0
2
– V
2
1
R = previously recorded value
Calibration Factor = value for thermistor mount at 50 MHz (traceable to NIST)
2-18 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
NOTE The following is an example showing how a calculation of the Power
Reference Output Level is performed:
R = 200
V0=170 µV or 170 x 10–6 V
= 78,107 µV or 78,107 x 10–6 or .078 V
V
1
V
comp
= 5.1583 V
R = previously recorded value
Calibration Factor = 99.68%
2
– V
2
1
Note that V
Prf =
2 V
is eventually dropped out of the equation in the following
0
( V1 – V0 ) + V
comp
0
4R (Calibration Factor)
steps, because its value is very small (effectively zero).
2
P
=
rf
4(200) (99.68%)
0.80469 – 0.00608
=
2 (5.1583) (.078 – 0)
+ (.000170)2 – (0.78)
0.79861
=
797.44 797.44
1.00147 mW
=
2
22 Use a non-metallic, slot-type adjustment tool to adjust the setting of R98
on the main circuit board (see Figure 2-3) until P
is between 0.988 mW
rf
and 1.012 mW. Turn the adjustment tool counter-clockwise to increase the power level or clockwise to decrease the power level.
Assembly Level Service Guide 2-19
Chapter 2 Service
Calibration Procedures

DVM Calibration

The DVM calibration procedure is semi-automatic in that it is accomplished via software embedded in the Agilent 53147A/48A/49A. A computer and an RS-232 serial cable are required to access the calibration software.
Equipment Required
A computer with an RS-232 serial interface and terminal software Agilent 53150-60215 RS-232 Serial Cable Agilent 6613C or 6614C Precision DC Power Supply
NOTE To ensure that the results are valid, the Agilent 53147A/48A/49A and
2
the Agilent 6613C/6614C Precision DC Power Supply should be powered on for at least 30 minutes prior to beginning the calibration procedure. This allows the internal temperatures of the equipment to stabilize.
1 Press the Display DVM key on the Agilent 53147A/48A/49A.
2 Connect the DVM test leads to the DVM + and – connectors on the Agilent
53147A/48A/49A (red to + and black to –).
3 Connect the test prods to the + and – output connectors on the Agilent
6613C/6614C Precision DC Power Supply.
4 Set the 6613C/6614C Precision DC Power Supply to output exactly
+ 50.0 V.
5 Connect the Agilent 53150-60215 RS-232 Serial Cable between the
computer’s serial port (DB-25) and the Agilent 53147A/48A/49A’s serial port (RJ-45).
6 Start the terminal software on the computer.
7 Send “!acq 60” from the terminal program to the Agilent 53147A/48A/49A
(be sure to type lowercase letters when typing “!acq 60”).
8 Press “5” on the computer to choose option 5 from the DMM Tests menu
(see Figure 2-8).
2-20 Assembly Level Service Guide
Chapter 2 Service
Calibration Procedures
*********************** DVM TESTS ***********************
Select one of the following DVM tests:
1. Measure Fz - Frequency in Zero Mode
2. Measure Fr - Frequency in Positive Reference Mode
3. Measure Fn - Frequency in Negative Reference Mode
4. Measure Fm - Frequency in Input mode
5. Calculate and Save Calibration Constant.
6. Measure Voltage
7. Set the Gate time: 10ms - 1000 ms.
8. Choose Gated mode
9. Choose Reciprocal mode
A. Choose Single Measurement mode. B. Choose Continuous Measurement mode. E. Exit DVM tests.
Choose one of the above? (1 - 9, E) 5
Figure 2-8. DVM Tests Main Menu
9 When the calibration program displays:
2
Apply 50.000 Volts Reference to the DVM. Ready? (y/n) y
type “y”.
10 When the calibration program displays:
Update from old 10974.12 to new 10974.1 Cal. Constant? (y/n) y
type “y” (the values in the display line above are simulated).
11 When the calibration program displays:
Apply –50.000 Volts Reference to the DVM. Ready? (y/n) y
12 Reverse the connection of the test prods in the + and – output connectors
on the Agilent 6613C/6614C Precision DC Power Supply (red to – output, black to + output), or reverse the polarity of the power outputs.
13 Type “y”.
Assembly Level Service Guide 2-21
Chapter 2 Service
Calibration Procedures
14 When the calibration program displays:
Update from old 10974.01 to new 10974.96 Cal. Constant? (y/n) y
type “y” (the values in the display line above are simulated).
15 When the calibration program redisplays its menu (as shown in
Figure 2-8), type “e” to exit the program.
16 Close the terminal program and disconnect the serial cable from the
computer and the instrument.
2
2-22 Assembly Level Service Guide
Chapter 2 Service

Pre-Troubleshooting Information

Pre-Troubleshooting Information
This section contains the following pertinent troubleshooting information:
Safety Considerations
Recommended Test Equipment
Repair Considerations
After Service Considerations
Assembly Identification and Location

Safety Considerations

Although this instrument has been designed in accordance with international safety standards, this guide contains information, cautions, and warnings which must be followed to ensure safe operation and to retain the instrument in a safe condition. Service instructions, and adjustment procedures requiring removal of the instrument cover, are for use by service-trained personnel only. To avoid dangerous electric shock, do not perform any servicing or make any adjustments with the cover removed, unless qualified to do so.
2
WARNING BEFORE APPLYING AC POWER, THE INSTRUMENT AND ALL
PROTECTIVE EARTH TERMINALS, EXTENSION CORDS, AUTO TRANSFORMERS, AND DEVICES CONNECTED TO THE INSTRUMENT SHOULD BE CONNECTED TO A PROTECTIVE EARTH GROUNDED SOCKET.
ANY INTERRUPTION OF THE PROTECTIVE GROUNDING CONDUCTOR INSIDE OR OUTSIDE THE INSTRUMENT OR DISCONNECTION OF THE PROTECTIVE EARTH TERMINAL WILL CAUSE A POTENTIAL SHOCK HAZARD THAT COULD RESULT IN PERSONAL INJURY. INTENTIONAL INTERRUPTION IS PROHIBITED.
Assembly Level Service Guide 2-23
Chapter 2 Service
Pre-Troubleshooting Information
Any maintenance and repair of the opened instrument under voltage should be avoided as much as possible and, if necessary, should be carried out only by a skilled person who is aware of the hazards involved (for example, fire and electric shock).

Recommended Test Equipment

Test equipment recommended for testing and troubleshooting the Agilent 53147A/148A/149A is listed in Chapter 1, “Performance Tests.” Substitute equipment may be used if it meets or exceeds the required characteristics listed in Table 1-1.
2

Repair Considerations

Electrostatic Discharge
Electronic components and assemblies in the Agilent 53147A/148A/149A can be permanently degraded or damaged by electrostatic discharge. Use the following precautions when servicing the instrument:
1 ENSURE that static-sensitive devices or assemblies are serviced at
static-safe work stations providing proper grounding for service personnel.
2 ENSURE that static-sensitive devices or assemblies are stored in static
shielding bags or containers.
3 DO NOT wear clothing subject to static-charge buildup, such as wool or
synthetic materials.
4 DO NOT handle components or assemblies in carpeted areas.
5 DO NOT remove an assembly or component from its static shielding
protection until you are ready to install it.
6 AVOID touching component leads. (Handle by packaging only.)
2-24 Assembly Level Service Guide
Chapter 2 Service
Pre-Troubleshooting Information
Surface Mount Repair
None of the assemblies in the instrument that use surface-mount components can be repaired outside of authorized Agilent Service Centers. Any attempt to repair these assemblies voids the Agilent factory warranty.
Disassembly and Reassembly Specifics
Refer to Chapter 3 of this guide, “Replacing Assemblies,” for complete disassembly and reassembly instructions, and Chapter 4, “Replaceable Parts” for an exploded view of the instrument’s parts.

After Service Considerations

Product Safety Checks
The following safety checks must be performed after any troubleshooting and repair procedures have been completed to ensure the safe operation of the instrument.
WARNING THE RESISTANCE CHECKS DESCRIBED IN THE FOLLOWING
TEXT REQUIRE THAT THE POWER CORD BE CONNECTED TO THE INSTRUMENT AND THAT AC POWER BE DISCONNECTED. BE SURE THAT THE POWER CORD IS NOT CONNECTED TO AN AC POWER SOURCE BEFORE PERFORMING ANY SAFETY CHECKS.
1 VISUAL INSPECTION. Visually inspect the interior of the instrument
for any signs of abnormal internally generated heat, such as discolored printed circuit boards or components, damaged insulation, or evidence of arcing. Determine and remedy the cause of any such condition.
2 GROUND CONTINUITY TEST. Plug the power cord into the rear- panel
power receptacle. (DO NOT connect the instrument to AC power at this time.) Using a suitable ohmmeter, check resistance from the instrument's metallic connection (such as the rear panel or BNC ground collar) to the ground pin on the power cord plug. The reading must be less than 1Ω. Flex the power cord while making this measurement to determine whether intermittent discontinuities exist.
2
Assembly Level Service Guide 2-25
Chapter 2 Service
Pre-Troubleshooting Information
3 Check any indicated front- or rear-panel ground terminals marked,
using the above procedure.
4 INSULATION RESISTANCE TEST. Tie the line and neutral pins of the
power cord plug together. Measure the resistance from the instrument enclosure (chassis) to the line and neutral pins of the power cord plug. The minimum acceptable resistance is 2 MΩ. Replace any component which results in a failure.
Product Performance Checks
After replacement of any functional assembly, perform the operational
2
verification test in Chapter 1 of this guide, “Performance Tests.”

Assembly Identification and Location

The assembly number, name, and Agilent part number of each of the Agilent 53147A, 53148A, and 53149A assemblies are listed in Table 2-2. Figure 2-9 illustrates the replaceable assemblies in the Agilent 53147A/148A/149A.
Table 2-2. Agilent 53147A/148A/149A Assembly Identification
Assembly Name Agilent Part No.
A1 Adapter (53149A only) 53152-80001
A2 Front Panel Assembly (53147A) A2 Front Panel Assembly (53148A) A2 Front Panel Assembly (53149A)
A3 Front Panel Power Switch Assembly (Power/Standby Switch) 53147-20203
A4 Battery Housing Assembly * 53150-00002
A5 DC/DC Converter PCB * 53150-60004
A6 Battery Switch Assembly * 53147-60223
A7 Battery * 53150-80010
A8 Battery Sled Assembly * 53150-00004
A9 Battery Cover Assembly 53150-00003
A10 Power Entry Module Assembly 53147-40007
A11 Rear Panel Assembly 53147-60219
A12 Cooling Fan Assembly 53147-80012
A13 Power Supply Assembly 0950-3299
* With Option 002 (Battery Option) only.
53147-60207 53148-60207 53149-60207
2-26 Assembly Level Service Guide
Chapter 2 Service
Pre-Troubleshooting Information
A5
A4
A3
A13
A12
A6
A11
A7
A9
A10
A8
2
A2
A1
Figure 2-9. Replaceable Assembly Locations — Top Internal View
Assembly Level Service Guide 2-27
Chapter 2 Service

Troubleshooting the Instrument

Troubleshooting the Instrument

Power Supply Check

WARNING HAZARDOUS VOLTAGES ARE PRESENT ON THE POWER
SUPPLY ASSEMBLY. ONLY TRAINED AND QUALIFIED SERVICE PERSONNEL SHOULD PERFORM THE FOLLOWING PROCEDURE.
2
NOTE The DC Power Supply Assembly is fused for safety reasons. It must be
1 Remove the power cord from the back of the instrument.
2 Remove the cover as described in Chapter 3.
assumed that, if the fuses are blown, the power supply has probably suffered damage and has therefore become unreliable. If the power supply fails, it must be replaced, even if the fuse in the power supply appears to be all that has failed.
3Connect the instrument to the power source.
4 Connect the negative lead of a voltmeter to the chassis and measure each
of the voltages listed in Table 2-3 on Power Supply connectors TB1 and TB2 (see Figure 2-10).
Pin 8
Neutral
TB1 (AC In)
Line
TB2 (DC Out)
Pin 1
Approx. location of R23
Figure 2-10. A4 DC Power Supply Test-Point Locations
2-28 Assembly Level Service Guide
Chapter 2 Service
Troubleshooting the Instrument
Table 2-3 lists A4 Power Supply’s input and output voltages, the corresponding wire color for each input and output, and the tolerances specified for these voltages.
Table 2-3. A4 Power Supply Inputs and Outputs
Input/Output Voltage Connector and Pin Designation Wire Color Tolerance
AC-input (neutral) 0 VAC* TB1, pin N blue
AC-input (line in) 85 - 264 VAC* TB1, pin L brown
–15.0 VDC @ 0.7 A max. TB2, pin 1 lavender ±3%
0.0 VDC (ground for –15 VDC) TB2, pin 2 brown
N/A (unused) TB2, pin position 3 (no pin) N/A
N/A (unused) TB2, pin 4
+15.0 VDC @ 2 A max. TB2, pin 5 blue ±5%
0.0 VDC (ground for +15 VDC) TB2, pin 6 brown
+5.0 VDC @ 5A max. TB2, pin 7 red ±1%
0.0 VDC (ground for +5 VDC) TB2, pin 8 black
* AC-input values are referenced to chassis ground.
Adjusting the +5 VDC Output
The +5 VDC output of the power supply is adjustable. If this voltage is found to be outside of the tolerance range listed in Table 2-3, it can be adjusted using R23, which is visible on the top surface of the power supply’s circuit board. When adjusting the +5 VDC output, use a nonmetallic or insulated adjustment tool.
2
Assembly Level Service Guide 2-29
Chapter 2 Service
Troubleshooting the Instrument

Self-Test

The instrument performs a Self-Test procedure when it is powered on. The Self-Test can also be invoked from the instrument’s Menu.
Running the Self Test
1 Disconnect all cables connected to the instrument’s front and rear signal
connectors.
2 Connect the power cord to the power input connector on the instrument’s
rear panel and to an appropriate power source.
2
3 Press and release the Power button on the front panel.
4 Observe the front-panel display and note any error messages displayed.
5 When CH 2 NO SIGNAL is displayed, press and release the Shift key,
and then press and release the Menu key.
6 Press the up-arrow key or the down-arrow key repeatedly until
DO SELF TEST is displayed, and then press and release the Enter key.
7 Observe the front-panel display and note any error messages displayed.
2-30 Assembly Level Service Guide
Chapter 2 Service
Troubleshooting the Instrument
Self Test Error Messages
Table 2-4 lists and describes messages that are generated by the instrument during Self-Test to indicate whether a component passed or failed its test. These messages are sent via the RS-232 serial output only— they do not appear on the instrument’s front-panel display.
Table 2-4. Self-Test Messages
Message Description
ROM TEST FAIL ROM TEST OK
RAM DATA LINES OK RAM DATA ERROR RAM ADDR LINES OK RAM ADDR ERROR RAM TEST OK
EEPROM FAIL - CONFIGURATION DATA
ROM FAIL; Computed checkbyte does not match the value stored in EEPROM.
EEPROM FAIL - CONFIGURATION DATA; Needs to be (re)initialized.
EEPROM FAIL - POWER CAL DATA; Using default data
EEPROM FAIL - SAVED SETTINGS; Using default data
EEPROM FAIL - SAVED SETTINGS; Invalid EEPROM SAV n Data.
GPIB FAIL; Conf. Test The GPIB hardware failed its confidence test.
ROM failed read test. ROM passed read test.
RAM data lines passed test. RAM data lines failed test. RAM address lines passed test. RAM address lines failed test. RAM tests completed with no errors detected.
The configuration data saved in EEPROM memory is defective.
The checksum of the ROM data does not match the value stored in EEPROM.
The EEPROM org code does not verify with current revision of ROM code.
The checksum of the EEPROM power­calibration table is bad. Factory default calibration data will be used.
The checksum of the user settings stored in EEPROM is bad. Factory default settings will be used.
The checksum of one set of user settings (1 – 9) stored in EEPROM is bad.
2
Assembly Level Service Guide 2-31
Chapter 2 Service
Troubleshooting the Instrument
Tables 2-5 and 2-6 list and describe the messages that are generated by the instrument during Self-Test or during operation to indicate that a problem has been detected. These messages are displayed on the instrument’s front-panel display and are also sent via the RS-232 serial output (in many cases, the exact message text that is displayed on the front panel is a condensed form of the message that is sent via RS-232).
Table 2-5. Front Panel Display Error Messages
Message Description
12V FAIL The +12 VDC output from the power supply is not within specifications.
–12V FAIL The –12 VDC output from the power supply is not within specifications.
2
–5V FAIL The –5 VDC output from the power supply is not within specifications.
ADC FAIL Failure detected in the ADC.
PATH FA I L Failure detected in the Channel 1 signal path.
THRS FAIL Failure detected in the Channel 1 threshold circuit.
THRS FAIL Failure detected in the Channel 2 RF threshold circuit.
THRS THRU Failure detected in the Channel 2 through-path threshold circuit.
THRS HET Failure detected in the Channel 2 heterodyne-path threshold circuit.
EEP WRT FAIL Failure detected while writing to EEPROM.
FPANEL FAIL The front panel or its interconnecting cable is defective or not properly connected.
FPGA FAIL Failure detected in the FPGA (Field Programmable Gate Array).
GPIB FAIL Failure detected in the GPIB hardware.
HET PATH FAIL Failure detected in the heterodyne-path circuit.
IIC FAIL An attempt to write to the LCD display failed.
INSTCFG FAIL The instrument’s configuration data is missing or has become corrupted.
OVERTEMP The instrument’s internal temperature is above the acceptable limit.
PWR CAL FAIL The instrument’s power-calibration data is missing or corrupted.
PWR CKT FAIL Failure detected in the Channel 2 power-measurement circuit.
RAM FAIL Failure detected in RAM.
ROM FAIL Failure detected in ROM.
SAV SET FAIL One or more of the sets of user settings is missing or is corrupted.
SERVICE FAIL The instrument’s service data is missing or has become corrupted.
THRU FAIL Failure detected in the through-path circuit of Channel 2.
VCO/CNT FAIL Failure detected in the VCO or the Count Chain.
2-32 Assembly Level Service Guide
Chapter 2 Service
Troubleshooting the Instrument
Table 2-6. RS-232 Error Messages
Message Description
12V FAIL The +12 VDC output from the power supply is not within specifications.
–12V FAIL The –12 VDC output from the power supply is not within specifications.
–5V FAIL The –5 VDC output from the power supply is not within specifications.
ADC FAIL Failure detected in the ADC.
B1 SIGNAL PATH FAIL Failure detected in the Channel 1 signal path.
B1 THRESHOLD FAIL Failure detected in the Channel 1 threshold circuit.
B2 RF THRESHOLD FAIL Failure detected in the Channel 2 RF threshold circuit.
B2 THROUGH-PATH THRESHOLD FAIL
B2 HETERODYNE PATH THRESHOLD FAIL
EEPROM FAIL - WRITE Failure detected while writing to EEPROM.
FRONT PANEL FAIL The front panel or its interconnecting cable are defective or not properly connected.
FPGA FAIL Failure detected in the FPGA (Field Programmable Gate Array).
GPIB FAIL Failure detected in the GPIB hardware.
HETERODYNE PATH FA IL
IIC FAIL An attempt to write to the LCD display failed.
INSTCFG FAIL The instrument’s configuration data is missing or corrupted.
OVER TEMPERATURE The instrument’s internal temperature is above the limit.
PWR CAL FAIL The instrument’s power-calibration data is missing or corrupted.
POWER METER FAIL Failure detected in the Channel 2 power-measurement circuit.
RAM FAIL Failure detected in RAM.
ROM FAIL Failure detected in ROM.
SAV SET FAIL One or more of the sets of user settings is missing corrupted.
SERVICE FAIL The instrument’s service data is missing or corrupted.
THRU PATH FAIL Failure detected in the through-path circuit of Channel 2.
VCO/COUNT CHAIN FAIL Failure detected in the VCO or the Count Chain.
Failure detected in the Channel 2 through-path threshold circuit.
Failure detected in the Channel 2 heterodyne-path threshold circuit.
Failure detected in the heterodyne-path circuit.
2
Assembly Level Service Guide 2-33
Chapter 2 Service
Troubleshooting the Instrument
2
2-34 Assembly Level Service Guide
3
Replacing Assemblies
Disassembly and Reassembly
Chapter 3 Replacing Assemblies

Introduction

Introduction
This chapter provides instructions for removing major assemblies in the Agilent 53147A/148A/149A. To install an assembly, apply the removal instructions in reverse.
The following disassembly and repair procedures are provided:
Removing the Bumpers pg. 3-4
Removing the Handle pg. 3-5
Removing the Bumper Retainers pg. 3-6
Removing the Shroud pg. 3-7
Removing the Front Panel Assembly pg. 3-8
Removing the Rear Panel Assembly pg. 3-11
Removing the Cooling Fan pg. 3-13
3
Removing the Power Supply pg. 3-14
WARNING HAZARDOUS VOLTAGES ARE PRESENT ON THE POWER
SUPPLY ASSEMBLY. DISCONNECT THE INSTRUMENT FROM THE POWER SOURCE, AND WAIT AT LEAST SIX MINUTES BEFORE WORKING INSIDE THE INSTRUMENT.
CAUTION Do not replace assemblies when the instrument is operating to prevent
damage to components.
NOTE Unless otherwise stated, directional referents in the procedures in this
chapter are accurate when viewing the instrument from the front.
Part numbers for all replaceable parts and assemblies are provided in Chapter 5, “Replaceable Parts.”
3-2 Assembly Level Service Guide
Chapter 3 Replacing Assemblies

Tools Required

Tools Required
The following tools are required for the removal/replacement procedures in this chapter:
#1 Phillips screwdriver
#2 Phillips screwdriver
TORX
1/4-inch open-end or box wrench
1/4-inch nut driver
5/16-inch nut driver
7 mm nut driver
Needle-nose pliers
®
15 (T15) hand screwdriver

Do This First

Perform the following steps before beginning any of the removal and replacement procedures:
1 If the instrument is on, press and release the POWER button on the
front panel.
2 Disconnect the AC power cord from the AC input socket on the rear panel.
3 If the instrument has the Battery Option (002) installed, remove the
battery sled, and disconnect any DC input cable from the EXT DC connector on the rear panel.
CAUTION While most of the hardware used in the instrument is SAE, the six
TORX screws that attach the handle mounts to the instrument are metric. To avoid damage to the threaded holes in the chassis that handle mount screws fit in, be sure to use the correct screws.
Assembly Level Service Guide 3-3
3
Chapter 3 Replacing Assemblies

Removing the Bumpers

Removing the Bumpers
Use the following procedure to remove the front and rear protective bumpers:
1 Remove the rear bumper by lifting the inner edge of the bumper away
from the shroud near one corner at the top and one side and pulling that corner away from the instrument. Repeat the process with the remaining corners until the bumper is free of the instrument.
Auxiliary
Reference 10 MHz
WARNING:
90-130 VAC 50/60/400 Hz 190-230 VAC 50/60 Hz
Main Power
In or Out
FUSE
AC POWER
1.5A T 250V
RS-
232
GPIB
To avoid electrical shock:
No user serviceable parts inside. Refer all servicing to qualified personnel.
Do not remove covers.
Counter/ Power Meter/ DVM
Agilent
!
!
OPTIONS
001 Oven Time Base 002 Battery
Made in U.S.A.
This unit must be earth grounded.
of domestic and foreign content
20 GHz Counter/Power Meter/DVM
GHz
MHz kHz Hz
POWER
Ch 12
Rel Freq
Offset
dB
Avg On
dBm Watts mW
Rel Pwr
uW
Standby
Offset
%
MODIFY
Cal
Pwr
Freq
Menu
Factor
Offset
Offset
Reset/
Avg
Freq
Resol
Local
On/Off
Rate
Enter
+ / -
Clear
Shift
COUNTER
Gate
Offset On/Off
Select
COUNTER
POWER METER
CHANNEL 1
Counter
Power
Ext Rel
Hold
10 Hz to 125 MHz
OUTPUT
Meter
Rate Rmt SRQ
Error
1M
DVM
Shift
POWER METER
DVM
DAMAGE
Power Ref.
+
+30 dBm
dBm/W
1.00mW 50MHz
Display
Zero
Power
COUNTER
CHANNEL 2
INPUT
Store
Rel Pwr
50MHz to 26.5 GHz
Cal
Offset
Do NOT Exceed
DAMAGE
On/Off
Sensor Max Power
+27 dBm
50
DAMAGE
Display
Rel FreqChan
DVM
±60 VDC
3
Figure 3-1. Removing the Bumpers
2 Use the same procedure (as in Step 1) to remove the front bumper.
NOTE The bumpers can be used to support and protect the instrument after the
shroud is removed. Place both bumpers on their inner edges on your work surface, and then place the instrument on top of them. The rear bumper can also serve as a holder for the instrument (once the bumper retainers are removed) when you need to position it upright on either end.
3-4 Assembly Level Service Guide
Chapter 3 Replacing Assemblies

Removing the Handle

Removing the Handle
Use the following procedure to remove the carrying handle:
1 Remove the front bumper (see “Removing the Bumpers” on page 3-4).
2 Stand the instrument on end on its back bumper.
3 Position the handle so that it is against the top surface of the shroud
(see Figure 3-2).
Figure 3-2. Removing the Handle
4 Use a T15 TORX screwdriver to remove two of the 8-32 x 1/4-inch handle
mounting screws (remove the two screws that are closest to the back of the instrument).
5 Repeat Step 4 for the other handle mount.
6 While holding the handle to support it, remove the remaining two screws
(one on each handle mount).
7 Remove the handle.
Assembly Level Service Guide 3-5
3
Chapter 3 Replacing Assemblies

Removing the Bumper Retainers

Removing the Bumper Retainers
Use the following procedure to remove the bumper retainers:
1 Remove the front and rear bumpers (see “Removing the Bumpers” on
page 3-4).
2 Use a #1 Phillips screwdriver to remove the two 4-40 x 7/16-inch flathead
machine screws that attach each bumper retainer to the chassis (see Figure 3-3).
3
Figure 3-3. Removing the Bumper Retainers
3 Repeat Step 2 for each bumper retainer you need to remove.
CAUTION When the all of the bumper-retainer screws and the handle are removed,
the shroud is not attached to the chassis. Exercise caution when moving the instrument in this condition to prevent the chassis from sliding out of the shroud. If you do not intend to remove the shroud, insert the bumper retainer screws in the holes they were removed from (the holes are countersunk to accept the screws, and the screws can safely be inserted fully into the holes).
3-6 Assembly Level Service Guide
Chapter 3 Replacing Assemblies

Removing the Shroud

Removing the Shroud
Use the following procedure to remove the shroud:
WARNING WHEN THE SHROUD IS REMOVED FROM THE COUNTER,
DANGEROUS LINE VOLTAGES THAT CAN CAUSE SERIOUS INJURY ARE EXPOSED. ALWAYS DISCONNECT THE POWER CORD BEFORE REMOVING THE SHROUD.
1 Remove all detachable connector fittings and/or adapters from the
connectors on the instrument’s front panel.
2 Remove the front and rear bumpers (see “Removing the Bumpers” on
page 3-4).
NOTE The bumpers can be used to support and protect the instrument after the
shroud is removed. Place both bumpers on their inner edges on your work surface, and then place the instrument on top of them. The rear bumper can also serve as a holder for the instrument (once the bumper retainers are removed) when you need to position it upright on either end.
3
3 Remove the handle (see “Removing the Handle” on page 3-5).
4 Remove the four bumper retainers from the front end of the instrument
(see Figure 3-3 and “Removing the Bumper Retainers” on page 3-6).
5 Remove the six bumper retainers at the rear end of the chassis.
6 Slide the shroud off the chassis to the rear (see Figure 3-4), and place it
aside (if the shroud does not slide easily off the chassis, press on the rear panel with one hand while pulling the shroud with the other hand).
Assembly Level Service Guide 3-7
Chapter 3 Replacing Assemblies

Removing the Front Panel Assembly

3
Figure 3-4. Removing the Shroud
Removing the Front Panel Assembly
Use the following procedure to remove the front panel assembly:
1 Remove the shroud (see “Removing the Shroud” on page 3-7).
2 Use a 5/16-inch wrench to loosen the nut that attaches the Power Meter
output cable to the back of the Power Meter OUTPUT connector on the front panel assembly.
3 Disconnect the Power Meter output cable from the back of the Power
Meter OUTPUT connector.
3-8 Assembly Level Service Guide
Chapter 3 Replacing Assemblies
Removing the Front Panel Assembly
4 Disengage the Power Meter output cable from the cable clamp on the
chassis floor near the right front corner of the chassis.
5 Disconnect the Power Meter input cable from J22, which is accessible
through the rectangular opening in the chassis floor near the right-front corner of the chassis.
6 Disengage the Power Meter input cable from the cable clamp on the
chassis floor near the right front corner of the chassis.
7 Disconnect the Power Meter INPUT connector’s ground wire (black) from
the ground lug on the chassis floor.
8 Disconnect the Channel 1 input cable from J15, which is accessible
through a hole in the chassis floor that is at the front of the chassis, behind the Chan Select key.
9 Disconnect the DVM input connector from J23, which is accessible
through the square hole in the chassis floor approximately 1/3 of the way back from the front panel in line with the Display DVM key.
10 Use a microwave torque wrench to partially loosen the nut that connects
the Channel 2 input connector to the sampler on the main circuit board.
3
CAUTION Do not attempt to fully loosen the nut at this time. As soon as the nut
turns freely, stop turning it, and proceed to the next step.
11 Remove the two 6-32 x ¼ flathead machine screws from the countersunk
holes at the top-front corners of the chassis (see Figure 3-5).
Assembly Level Service Guide 3-9
Chapter 3 Replacing Assemblies
Removing the Front Panel Assembly
3
Figure 3-5. Removing the Front Panel Assembly
12 Remove the five 4-40 x ¼-inch flathead machine screws that attach the
front-panel assembly to the bottom of the chassis.
13 Remove the two 6-32 x ¼-inch flathead machine screws from the
countersunk holes at each side of the front the end of the chassis.
14 While holding the front panel assembly with one hand, use the fingers of your
other hand to continue to loosen the nut on the Channel 2 input connector. Gently pull the front panel assembly away from the chassis as you loosen the nut until the front panel assembly is completely disconnected.
CAUTION When reattaching the front panel assembly, reverse the procedure above.
You must use a microwave torque wrench to tighten the nut that attaches the Channel 2 input connector to the sampler. Using any other tool to tighten the coupling nut will damage the connector, the sampler, or both.
3-10 Assembly Level Service Guide
Chapter 3 Replacing Assemblies

Removing the Rear Panel Assembly

Removing the Rear Panel Assembly
Use the following procedure to remove the rear panel assembly:
1 Remove the shroud (see “Removing the Shroud” on page 3-7).
2 If the Battery Option is not installed, remove the four 4-40 self-locking hex
nuts that attach the battery opening cover to the rear panel (these nuts are accessible from inside the chassis). Remove the battery opening cover, and skip to step 5.
3 If the Battery Option is installed, remove the battery sled from the
Counter (for instructions on removing the battery sled, see Appendix C of the Agilent 53147A/148A/149A Operating Guide).
4 If the Battery Option is installed, remove the four 4-40 x 3/16-inch
self-locking flathead machine screws that attach the rear panel assembly to the Battery Option chassis as shown in Figure 3-6 (two screws on each side of the opening).
Auxiliary
RS-
232
Reference 10 MHz
In or Out
FUSE
1.5A T 250V
No user serviceable parts inside. Refer all servicing to qualified personnel.
!
OPTIONS
001 Oven Time Base 002 Battery
GPIB
!
Agilent
Battery
option
screws
This unit must be earth grounded.
Counter/ Power Meter/ DVM
Battery
Battery Charging
EXT DC
11 to 18 VDC
+
Made in U.S.A.
of domestic and foreign content
Rear panel screws
Rear panel screws
90-130 VAC 50/60/400 Hz 190-230 VAC 50/60 Hz
WARNING:
AC POWER
Main Power
To avoid electrical shock: Do not remove covers.
Rear panel screws
Figure 3-6. Rear Panel Assembly Screws
5 Remove the nine 4-40 x 3/16-inch self-locking flathead machine screws
that attach the rear panel assembly to the chassis (see Figure 3-6).
NOTE The screw in the lower-left corner of the rear panel assembly (as viewed
from the rear of the Counter) may be covered by the serial-number sticker. If it is, carefully cut away the lower-left corner of the sticker to expose the screw head.
Assembly Level Service Guide 3-11
3
Chapter 3 Replacing Assemblies
Removing the Rear Panel Assembly
6 Disconnect the external-reference input/output cable (coaxial) from
connector J8 on the main board. J8 is directly under the back edge of the power supply near the chassis wall.
7 A two-wire AC-input cable (brown and blue wires) runs between the
AC input module on the rear-panel assembly and the connector that is accessible through the cutout in the front of the power supply enclosure (TB1). Disconnect this cable from the power supply.
8 If the Battery Option is installed, there is a two-wire cable (brown and
white wires) that runs between the AC input module on the rear panel assembly and connector J3 on the Battery Option circuit board (J3 is the third connector from the back edge of the board). Disconnect this cable from J3 on the Battery Option circuit board.
9 Use a 5/16-inch nut driver to remove the self-locking hex nut that attaches
the rear panel assembly grounding wire (green/yellow) to the right wall of the chassis. Slide the grounding lug off the welded stud on the chassis wall, and thread the lock nut back onto the stud a few turns until it is finger-tight.
10 Use a 7 mm nut driver to remove the two hex standoff screws that attach
3
the GPIB connector to the rear panel assembly.
WARNING AN ELECTRICAL SHOCK HAZARD EXISTS IF THE COUNTER IS
OPERATED WITH THE REAR PANEL ASSEMBLY GROUNDING WIRE NOT PROPERLY CONNECTED TO THE CHASSIS. TO PREVENT SERIOUS INJURY AND/OR DAMAGE TO THE EQUIPMENT, NEVER OPERATE THE COUNTER UNLESS THIS GROUND WIRE IS SECURELY CONNECTED TO THE CHASSIS WALL AND THE REAR PANEL ASSEMBLY.
11 Carefully pull the rear panel assembly away from the back end of the
chassis. Guide the cables that remain attached to the rear panel assembly as you remove it to prevent them from catching on other cables or other components of the Counter.
CAUTION When reinstalling the rear panel assembly, position it carefully to avoid
damage to the Battery Charging LED, the serial-port connector (RJ-12), the GPIB connector, and the main board. If these components are not correctly aligned with the openings in the back panel, they may be damaged.
3-12 Assembly Level Service Guide
Loading...