Agilent 37717C SDH / SONET Operating Manual

HP 37717C Communications Performance Analyzer
SDH / SONET Operating Manual
Copyright Hewlett­Packard Ltd.1997
All rights reserved. Reproduction, adaption, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Information in this document may apply to modules which use the VxWORKS TM software. The VxWORKS TM software was developed by Wind River Systems, Inc., which has copyright to it.
HP Part No. 37717-90284
First edition, September 97
Printed in U.K.
Warranty
The information contained in this document is subject to change without notice.
Hewlett-Packard mak es no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchanability and fitness for a particular purpose.
Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
WARNING
Warning Symbols Used on the Product
!
The product is marked with this symbol when the user should refer to the instruction manual in order to protect the apparatus against damage.
The product is marked with this symbol to indicate that hazardous voltages are present
The product is marked with this symbol to indicate that a laser is fitted. The user should refer to the laser safety information in the Calibration Manual.
Hewlett-Packard Limited Communications Measurements Division South Queensferry West Lothian, Scotland EH30 9TG
SDH / SONET Operating Manual
HP 37717C Communications Performance Analyzer
About This Book
“The HP 37717C SDH / SONET Operating Manual” explains the following:
How to select and use the SDH / SONET features provided with the SDH /
SONET options
How to make SDH / SONET measurements
For some operations and measurements, information from one of the following associated books may be required:
“The HP 37717C Mainframe Operating Manual” explains how to obtain the required display, how to use the front panel keys, how to interpret the status indicators, how to connect to external equipment and how to perform instrument tasks associated with the HP 37717C Communications Performance Analyzer irrespective of the option configuration.
“The HP 37717C PDH /DSn Operating Manual” describes the selection of PDH / DSn features and how to perform PDH / DSn tests with the HP 37717C Communications Performance Analyzer.
“The HP 37717C Jitter Operating Manual” describes the selection of Jitter features and how to perform Jitter tests with the HP 37717C Communications Performance Analyzer.
“The HP 37717C ATM Operating Manual” describes the selection of ATM features and how to perform ATM tests with the HP 37717C Communications Performance Analyzer.
iv
Contents
1 SDH / SONET Modules
Introduction to SDH / SONET 2 SDH / SONET Options 3 SDH / SONET Features 5
2 SDH / SONET Application Measurements
Add/Drop Multiplexer Testing 24 Alarm Stimulus/Response 29 DCC Testing 33 Desynchroniser Stress 36 Frame Synchronization 39 SDH Jitter Transfer 44 MSP Stimulus/Response 47 Optical Clock Recovery Stress 51 Payload Mapping/Demapping 53 Performance Monitor Stimulus/ Response 59 Selective Jitter Transfer Measurement 62
3 ETSI / ANSI Terminology
ETSI / ANSI Equivalent Terms 68
v
Contents
vi
1
Introduction to SDH / SONET page 2 SDH / SONET Options page 3 SDH / SONET Features page 5
SDH Settings page 5 SONET With ATM Payload, Settings page 6 STM-1 THRU Mode (Option A1T (A1U) only) page 7 STM-4 THRU Mode (Option A1T (A1U) and UKT or USN only) page 8 Transmit Static Overhead (Option A1T (A1U) only) page 9 Transmit Overhead Sequence page 13 Receive Overhead Monitor page 14 Receive Overhead Capture page 15 Errors and Alarms page 16 Overhead Bit Error Rate (A1T (A1U) only) page 17 Add Pointer Adjustments (A1T (A1U) only) page 17 Pointer Graph (A1T (A1U) only) page 19 MSP (Multiplexer Section / Transport Protection) Messages page 20 SYNC Messages page 21 Optical Clock Recovery Stress (A1T (A1U) only) page 21 DCC Insert and Drop (A1T (A1U) only) page 21 Optical Power Measurement (Options USN and UKT Only) page 22 STM-1/STM-4 Binary Interface (Option OYH with USN or UKT) page 22

1 SDH / SONET Modules

Information on the optional modules including selection of the features available.
SDH / SONET Modules

Introduction to SDH / SONET

Introduction to SDH / SONET
Synchronous Digital Hierarchy (SDH) is an international standard (ETSI) for high speed synchronous optical telecommunications networks. Synchronous Optical NETwork (SONET) is the U.S. (ANSI) equivalent of SDH defined by Bellcore.
The concept of a synchronous transport system, based on SDH / SONET standards, goes beyond the basic needs of a point to point transmission system. It includes the requirements of telecommunications networking - switching, transmission and network control. These capabilities, allow SDH / SONET to be used in all three traditional network application areas - Local Network, Inter-exchange Network and Long Haul Network - thus providing a unified telecommunication network structure.
The SDH / SONET standards are based on the principle of direct synchronous multiplexing. This means that individual tributary signals (Payload) may be multiplexed directly into a higher rate SDH / SONET signal without intermediate stages of multiplexing. SDH / SONET network elements, even those supplied by different manufacturers, can be interconnected directly giving cost and equipment savings.
There are minor differences between SDH and SONET, mainly in nomenclature. The most obvious technical difference occurs in the ‘SS’ bits in the H2 pointer byte of the Section OverHead (SOH) / Transport OverHead (TOH). The terminology on the instrument display is mainly ETSI. A table of ETSI / ANSI equivalent terms is given at the rear of this manual.
SDH / SONET is capable of transporting all the common tributary signals E1 (2.048 Mb/s), E3 (34.368 Mb/s), E4 (139.264 Mb/s), DS1 (1.544 Mb/s) and DS3 (44.736 Mb/s) currently in use. In addition SDH / SONET has the flexibility to readily accommodate any new types of service which are being introduced for example (ATM) or which may be required in the future. Approximately 5% of the SDH / SONET signal structure (Overhead) is reserved for network management and maintenance.
This means that SDH / SONET can be deployed as an overlay to the existing network thus providing enhanced network flexibility.
The HP 37717C provides comprehensive testing of both payload and overhead at electrical and optical SDH / SONET interfaces.
2
SDH / SONET Modules

SDH / SONET Options

SDH / SONET Options
SDH / SONET options are available according to your testing needs:
Option US1 (US5)
Provides an STM-1 Electrical interface. (When Option UH1 is fitted provides an STM-1 Optical interface. When Option UH2, URU, USN or UKT is fitted provides STM-1 and STM-4 optical interfaces.)
Provides frequency offset of the SDH signal of ±999 ppm.
Payloads of 140 Mb/s, 34 Mb/s and 2 Mb/s are available.(If Option UKJ U.S.A. Structured PDH is fitted, Structured Payloads are available and a 2Mb/s Insert port is provided.)
Bit errors can be added to the Payload.
Allows Errors & Alarms to be added to the SDH signal.
The K1K2 MSP bytes can be user programmed The S1 byte can be user programmed.
3
SDH / SONET Modules
SDH / SONET Options
Option A1T (A1U)
Provides an STM-1 Electrical interface. (When Option UH1 is fitted provides an STM-1 Optical interface. When Option UH2, URU, USN or UKT is fitted provides STM-1 and STM-4 optical interface.)
When Option UKZ is also fitted, provides an additional OC-3c transmit and receive facility.
Includes a THRU mode capability such that the Payload and the Section / Transport Overhead of the incoming STM-1 signal can be overwritten by the test set.
Provides frequency offset of the SDH / SONET signal of ±999 ppm. Payloads of 140 Mb/s, TU3, TU2 and TU12 are available.(If Option UKJ (USA) or UKN (USE) Structured PDH is fitted, Structured Payloads are available and a 2Mb/s, 34 Mb/sand 140 Mb/s Insert port is provided). Bit errors can be added to the Payload.
Allows Errors & Alarms to be added to the SDH / SONET signal.
Pointer Movements can be added to the SDH / SONET signal and a Graphical display of Pointer activities available.
Section / Transport and Path Overhead bytes are user programmable and can be monitored. Overhead byte activity can be captured and displayed.
Allows BER testing of Section / Transport and Path Overhead bytes.
DCC Drop and Insert capability is included.
Optical Clock stress capability is included.
Additional SDH interface capability is provided by the following Options: STM-1/STM-4 optical interface at 1550 nm, option URU. STM-1/STM-4 optical interface at 1330 nm and 1550 nm with access to STM-4
overhead and Optical power measurement, option USN. STM-1/STM-4 optical interface at 1330 nm and 1550 nm with access to STM-4
overhead and Optical power measurement, option UKT. SDH binary interface for Option UKT and USN only, option 0YH.
4
SDH / SONET Modules

SDH / SONET Features

SDH / SONET Features
Various SDH / SONET features are available with Options US1 (US5) and A1T (A1U). The use of these features is explained in the following pages:
SDH Settings
SIGNAL selection is normally STM-1. If Option UH1 is fitted STM-1 OPT is added. If Option UH2, URU, USN or UKT is fitted STM-1 OPT and STM-4 OPT are added.
If Option A1T (A1U) is fitted selection between INTERNAL and THRU is available.
CLOCK SYNC can be derived internally from the test set, from the SDH Master Timing Source or from the received SDH signal.
FREQUENCY OFFSET allows SDH line rate offset of ± 999 ppm).
PAYLOAD selection is 140 Mb/s, TU3 (34 Mb/s Option US1 (US5)), TU2 (Option A1T (A1U) only) or TU12 (ASYNC2 Mb/s Option US1 (US5)).
TU MODE selection sets the Tributary Mapping mode to ASYNC or FLOATING BYTE and is only available on Option A1T (A1U) when TU12 payload is selected.
TU CONCATENATION selection (A1T (A1U) only) determines the number of TU2 tributaries to be tested.
5
SDH / SONET Modules
SDH / SONET Features
SELECTED TU allows the selection of the Trib utary required for testing which in a TU12 payload is 1 of 63 Tributaries.
TU PAYLOAD sets the type of payload. If Unstructured PDH module, Option UKK (USB), is fitted only UNFRAMED is available. INSERT 34 Mb/s and INSERT 140 Mb/s are only available when Option A1T (A1U) is fitted.
P ATTERN allows selection of the T ributary test pattern.
OFFSET (Option A1T (A1U) only) allows the tributary payload to be frequency offset relative to its virtual container (VC-12, VC-3 or VC-4) in the range ± 100 ppm. 2M OFFSET if payload TU12, 34M OFFSET if payload TU3, 140M OFFSET if payload 140 Mb/s. This affects the stuff rate but does not cause pointer movements and can be used to test mapping Jitter.
PATTERN IN OTHER TU’s allows selection of the test pattern inserted in the non test tributaries.
SONET With ATM Payload, Settings
An ATM payload produced by Options UKN and UKZ may be transmitted at STM-1 and OC-3c by the SDH / SONET Option A1T (A1U). The ATM payload may then be multiplexed into an OC-12 signal with the dual wavelength (1310 / 1550 nm) Option USN.
NOTE: the OC-12 is not 622Mb/s ATM (OC-12c)
6
SDH / SONET Modules
SDH / SONET Features
The signal received by the ATM module (UKN or UKZ) may be retransmitted by selecting THRU mode. Jitter from the jitter generator module may be added to the retransmitted signal.
For ATM testing at STM-1, the ATM receiver of Option UKZ receives an ATM payload from Option A1T (A1U). For ATM testing at OC-3c or STM-1 within an OC-12, the dual wavelength (1310 / 1550 nm) module Option USN is required.
NOTE: the OC-12 is not 622Mb/s ATM (OC-12c)
STM-1 THRU Mode (Option A1T (A1U) only)
The primary purpose of THRU mode is to non-intrusively monitor SDH / SONET lines where no protected monitor points are available.
THRU Mode allows a ne w Payload and/or Section Overhead to be substituted in the received STM-1 SDH signal for test purposes. If PAYLOAD OVERWRITE [OFF] and SOH+POH CHANNEL OVERWRITE [OFF] are both selected the SDH signal is retransmitted without alteration.
The B1, B2 and B3 BIPS are not recalculated.
7
SDH / SONET Modules
SDH / SONET Features
When PAYLOAD OVERWRITE [AU4] is selected a user selectable Payload can be substituted for the received SDH payload.
A new path overhead is generated and the B3 BIP is recalculated.
The payload is selected as for SDH Internal mode.
When SOH+POH OVERWRITE [ON] is selected new Section / Transport and Path Overhead, is substituted for the received overhead. The B1, B2 and B3 Bips are recalculated.
The user can only modify certain overhead bytes by using one of the TRANSMIT test functions: Errors & Alarms, Sequences, Overhead BER, MSP Messages, DCC Insert and Stress Test.
The overhead cannot be modified using OVERHEAD SETUP.
STM-4 THRU Mode (Option A1T (A1U) and UKT or USN only)
Provides a non-intrusive monitor of the STM-4 signal where no protected monitor points are available. Neither the payload or overhead can be altered and the received signal is retransmitted unchanged.
8
SDH / SONET Modules
SDH / SONET Features
Transmit Static Overhead (Option A1T (A1U) only)
It can be desirable to set an overhead byte to a known static state to aid in troubleshooting, for example, to quickly check for “stuck bits” in path overhead bytes. This capability is provided under TRANSMIT OVERHEAD SETUP.
Path Overhead
The value of each bit of VC4 path overhead bytes: C2, F2, G1, H4, Z3, Z4 and Z5 can be set to 0 or 1. Byte B3 cannot be set. If TRANSMIT PAYLOAD [TU3] is selected the value of each bit of VC3 path overhead bytes C2, G1, F2, Z3, Z4 and Z5 can be set to 0 or 1.
An overhead byte cannot be set to a static value if a TEST FUNCTION is active in that byte, for example: If the ERROR & ALARMS Test Function is selected, and TU PATH UNEQUIPPED alarm is set, the value of C2 will be determined by the Alarm selection.
If TRANSMIT PAYLOAD [TU2] or [TU12] is selected H4 byte is set to carry a SEQUENCE. Selection between ; and is available.
COC1 SEQUENCE
The Signal Label value for VC-2 or VC-12 can also be set.
FULL SEQUENCE
REDUCED SEQUENCE
Path Trace The path trace capability allows:
A user selected data message to be inserted in the appropriate J1 byte to verify the VC-3 or VC-4 path connection
A user selected data message to be inserted in the appropriate J2 byte to verify the VC-2 or VC-12 path connection.
V C-3/VC-4 Path Trace
1 Press ; ;
OVERHEAD SETUP
TRANSMIT
SDH
and using
and the display softkeys set
up the display as shown
TRANSMIT
opposite.
2. Selection of
DEFAULT MESSAGE
programs the J1 byte to carry 64 ASCII Null characters.
9
SDH / SONET Modules
SDH / SONET Features
3. Selection of
TEST MESSAGE
programs the J1 byte to carry the displayed HP 37717C test message.
4. Selection of allows
USER MESSAGE
the user to program the message content of the J1 byte with up to 62 ASCII characters followed by CR LF.
5. Selection of allows
CRC7 MESSAGE
the user to program the message content of the J1 byte with up to 15 ASCII characters (padded out with ASCII Null characters if necessary) followed by a frame marker byte with CRC7.
10
SDH / SONET Modules
SDH / SONET Features
VC-2/VC-12 Path Trace
6. Selection of
DEFAULT MESSAGE
programs the J2 byte to carry 16 ASCII Null characters. Selection of programs
TEST MESSAGE
the J2 byte to carry the test set Serial number. Selection of allo ws the
USER MESSAGE
user to program the message content of the J2 byte with up to 14 ASCII characters followed by CR LF. The VC-2 or VC-12 Signal Label value can be set in the range 000 to 111.
Section / Transport Overhead
The value of each bit of Section / Transport overhead bytes in columns 1,4,7: A1,A2; C1; E1; F1; D1 - D3; K1,K2; D4 - D12; Z1; Z2 and E2 can be set to 0 or 1. Only bits 5 and 6 of byte H1 can be set to 0 or 1. Bytes B1, B2, H2 and H3 cannot be set at any time.
If columns 2,5,8 or 3,6,9 are selected only bytes A1,A2; H1 - H3; B2; Z1 and Z2 are labelled as the other overhead functions have not yet been defined. The v alue of each bit of Section / Transport ov erhead bytes: A1,A2; H1; H2; Z1; Z2 and all the bits of the unlabeled functions can be set to 0 or 1. Bytes B2 and H3 cannot be set at any time.
An overhead byte cannot be set to a static value if a TEST FUNCTION is active in that byte, for example: If the MSP MESSAGES Test Function is selected then K1,K2 value will be determined by the MSP MESSAGES selections.
Setting Undefined Overhead Functions For test purposes it may be important to be able set the value of those overhead functions which are presently undefined (Z2) or unlabeled as in columns 2,5,8 and 3,6,9.
11
SDH / SONET Modules
SDH / SONET Features
1. Press ; ;
OVERHEAD SETUP
TRANSMIT
SDH
and using
and set up the
TRANSMIT display as shown opposite
2. In this example the Z2 byte can be set in the range 00000000 to 11111111 using the display softkeys.
Static Overhead Known (Default) Conditions
After testing it may be desirable to return the static overhead to the known (default) state using one of the methods shown below.
1. Set up the ;
SETUP
TRANSMIT
OVERHEAD
display as shown opposite. This returns overhead, not set by a TRANSMIT test function to the known (default) state.
Any overhead byte set by a TEST FUNCTION will not be returned to the default state, for example: If the MSP MESSAGES Test Function is selected then K1,K2 value will be determined by the MSP MESSAGES selections.
2. Set up the ;
SETTINGS
OTHER
STORED
display as shown opposite.
This returns all settings, including Overhead and TRANSMIT T est Functions, to the known(default) state.
12
SDH / SONET Modules
SDH / SONET Features
Transmit Overhead Sequence
It can be desirable to insert a pattern into a functional group of overhead bytes for testing or troubleshooting purposes. This capability is provided under TRANSMIT SDH TEST FUNCTION SEQUENCE.
Overhead Bytes
Sequences can be inserted into 3XA13XA2 (Framing), C1, E1, F1, D1-D3 (Regenerator Section / Transport Overhead); K1K2, D4-D12, S1, Z1 column 2, Z1 column 3, Z2 column 4, Z2 column 5, E2 (Multiplexer Section / Transport Overhead) and J1, C2, G1, F2, H4, Z3, Z4 or Z5 (Path Overhead).
If TRANSMIT PAYLOAD [TU2] or [TU12] is selected sequences can only be inserted into the H4 byte using ; .
The sequence is derived from 5, user defined, hexadecimal blocks of data. Each block of data can be transmitted in up to 64000 frames.
The sequence can be run once only SINGLE RUN, or on a repetitiv e basis REPEAT RUN.
Transmit Frame Synchronization Sequence
1. Press ;
FUNCTION
TRANSMIT
and using
SDH TEST
and and the display softkeys set up the display as shown opposite.
TRANSMIT
OVERHEAD SETUP
Press to start the sequence. As
STARTED
a result of this sequence one OOF alarm and one LOF alarm should occur every two seconds.
13
SDH / SONET Modules
SDH / SONET Features
Receive Overhead Monitor
When first connecting to a network a start-up confidence check can be made by viewing the behavior of all the overhead bytes. If the network shows alarm indications, some diagnosis of the problem may be gained from viewing all the overhead bytes. This facility is provided on the RECEIVE display under
OVERHEAD MONITOR
All path overhead bytes are monitored and displayed. VC4 path overhead is shown opposite. If TRANSMIT PAYLOAD [TU3] is selected VC-3 path overhead can be monitored and displayed. If TRANSMIT PAYLOAD [TU12] is selected VC-12 path overhead can be monitored and displayed.
The display is updated once per second (once per 8000 frames approximately).
.
STATUS: Loss of frame alarm
All Section / Transport overhead bytes are monitored and displayed as shown opposite.
The display is updated once per second (once per 8000 frames approximately).
If any abnormal behavior is observed on a particular path or Section / Transport overhead byte, or an associated group of bytes (3XA1,3XA2; D1 - D3) the
RECEIVE
TEST FUNCTION
OVERHEAD CAPTURE
; display of can be used
to “zoom” in on the suspect byte, or bytes, on a frame by frame basis.
14
SDH / SONET Modules
SDH / SONET Features
Receive Overhead Capture
Regenerator Section / Transport, Multiplexer Section / Transport and Path overhead provide network support functions, responding dynamically to network conditions and needs. It is therefore desirable to be able to capture overhead activity on a frame by frame basis. This capability is provided under ; ;
TEST FUNCTION
;.
OVERHEAD CAPTURE
RECEIVE
Overhead Capture allows selection of the starting point of the capture by means of Trigger selection:
Trigger OFF start immediately the capture is initiated Trigger ON capture activity after a specified overhead state has occurred Trigger ON NOT capture activity after the first occurrence of a deviation from a
specified overhead state.
Trigger OFF can be used to provide a “frame by frame monitor” of the selected
byte, or bytes, immediately the capture is initiated.
Trigger ON and ON NOT can be used for “transient detection” from a specified
expected state.
SDH
The overhead capture feature provides up to 16 records of overhead state. Each
record will represent between 1 and 64,000 frames. A capture is started by
pressing and terminates when 16 records have been captured. The
capture can be terminated sooner by pressing .
STARTED
STOPPED
Frame by Frame Monitor of H1,H2
The frame by frame monitor capability provides a “zoom” in version of the Receiv e Overhead Monitor feature using the Trigger OFF condition of overhead capture.
1. Press ; ;
FUNCTION
RECEIVE
;
SDH TEST
OVERHEAD CAPTURE
and using and and the display softkeys set up the
RECEIVE
display as shown opposite. Press to start the capture.
STARTED
15
SDH / SONET Modules
SDH / SONET Features
Transient Detection on A1,A1,A1,A2,A2,A2
Under normal operating conditions the A1,A2 bytes will remain in a known stable state (F6F6F6282828). Using the Trigger ON NOT condition of the Overhead Capture, any transient deviations from that state can be detected.
1. Press ; ; ;
FUNCTION
RECEIVE
;
SDH TEST
OVERHEAD CAPTURE
and using and and the display softkeys set up the
RECEIVE
display as shown opposite.
Press to start the capture.
STARTED
Errors and Alarms
TRANSMIT
PAYLOAD
be added to the payload.
TRANSMIT
SDH ERRORS & ALARMS
Errors and Alarms to be added to the SDH Overhead.
;;
TEST FUNCTION PDH
ERRORS allows Bit errors to
;;
TEST FUNCTION
; allows
16
SDH / SONET Modules
SDH / SONET Features
Overhead Bit Error Rate (A1T (A1U) only)
1. The ;
TRANSMIT
OVERHEAD BER
test function inserts a 29-1 PRBS test pattern into the user selected byte of, Regenerator Section / Transport, Multiplexer Section / Transport or Path, overhead.
Single bit errors can be added to the PRBS test pattern.
2. The ; test
RECEIVE
OVERHEAD BER
function counts the errors present in the PRBS test pattern within the user selected overhead byte.
Results of Error Seconds, Error Ratio, Error Free Seconds,% Error Free Seconds, Pattern Loss Seconds and Elapsed Time are displayed.
Add Pointer Adjustments (A1T (A1U) only)
The ; test function allows the transmitted AU or
TRANSMIT
TU pointer value to be changed for testing purposes in four different ways:
ADJUST POINTER
1. provides a burst of pointers.
BURST
The size of the burst is determined by the number of PLACES selected. In this example 5 PLACES are selected and the pointer value will be stepped five times in unit steps e.g. 0 (start value), 1, 2, 3, 4, 5 (final value). The interval between steps is 4 frames (500 us). The direction of change is determined by the , ,
INCREMENT DECREMENT
ALTERNATE
If is selected the burst will be in the opposite direction to the
ALTERNATE
selection.
preceding burst. The burst is transmitted when ADJUST POINTER is selected.
ACTION
17
SDH / SONET Modules
SDH / SONET Features
2. allows a user defined
NEXT SUMMARY
pointer value to be transmitted in the range 0 to 782, with or without a New Data Flag (NDF). A NDF signifies a controlled change of Payload.
The current Pointer value is displayed for reference. The New Pointer is transmitted when ADJUST POINTER is
ACTION
selected.
3. allows either the line rate or
OFFSET
the VC/TU rate to be frequency offset, relative to each other, by up to ±100 ppm thus producing pointer movements.
When a VC-4 is offset relative to the line rate the 87:3 sequence of pointer movements is generated.
If the line rate is offset the TU rate is locked to the reference. If the TU rate is offset the line rate is locked to the reference.
4. provides pointer movement
G.783
sequences as per ITU G.783:
a. Periodic Single adjustments, each with opposite polarity to the preceding adjustment.
b. Periodic Single adjustments, with selectable polarity and added adjustment (1 extra). The added adjustment occurs 4 frames (500) after a periodic adjustment.The sequence has a repetition rate of approximately 30 seconds.
c. Periodic Single adjustments, with selectable polarity and cancelled adjustment (1 less). The sequence has a repetition rate of approximately 30 seconds.
18
SDH / SONET Modules
SDH / SONET Features
d. Periodic Double adjustments (pair of adjustments). The pair alternate in polarity. The spacing between pairs of adjustments, of like polarity, is 4 frames (500).
In all cases the interval between adjustments (between pairs in d) is user selectable. The sequence is started by selecting POINTER SEQUENCES .
STARTED
Pointer Graph (A1T (A1U) only)
The ; test
RECEIVE
function provides a graphical display of pointer movements.
The capture interval is user programmable.
POINTER GRAPH
19
SDH / SONET Modules
SDH / SONET Features
MSP (Multiplexer Section / Transport Protection) Messages
TRANSMIT
programming of the K1 and K2 bytes to exercise the MSP functions.K1 BITS 1->4 select the MSP Message to be transmitted.K1 BITS 5->8 select the channel in which the MSP Message is transmitted. NULL channel is the Protection channel.
K2 BITS 1->4 indicate the channel which
has been switched to protection.
K2 BIT 5 selects the MSP architecture, 1:1 or 1:N. 1:1 means the working line has an exclusive protection line. 1:N means the protection line is shared by several (up to 14) working lines.
K2 BITS 6->8 select the Reserved bits (Not Defined as yet).
The current K1/K2 Transmit and Receive states are displayed for reference.
To set the new K1/K2 values select TRANSMIT NEW K1/K2 [SELECT].
To transmit the new K1/K2 values select TRANSMIT NEW K1/K2 [DOWNLOAD].
; ; ; allows user
SDH TEST FUNCTION MSP MESSAGES
20
SDH / SONET Modules
SDH / SONET Features
SYNC Messages
TRANSMIT
SYNC MESSAGES
;; ;
SDH TEST FUNCTION
(Option US1 (US5) only) allows user programming of bits 5 to 8 of Multiplexer Section / Transport overhead byte S1. The current received value of S1 (Sync byte) is displayed for reference.
This function can be performed on Option A1T (A1U) using ;
OVERHEAD SETUP
TRANSMIT
.
Optical Clock Recovery Stress (A1T (A1U) only)
The ; test
TRANSMIT
function allows the STM-1 OPT payload to be overwritten with a block of zeros or ones, after scrambling, thus stressing any clock recovery circuits. Alternatively the user can select the G.958 test pattern which consists of consecutive blocks of four types of data: All 1’s; PRBS; All 0;s; a data block consisting of the first row of Section / Transport overhead bytes.
STRESS TEST
DCC Insert and Drop (A1T (A1U) only)
The ; test
TRANSMIT
function and the ; test function allow access to the Regenerator Section / Transport Overhead (192 kb/s) or Multiplexer Section / Transport Overhead (576 kb/s) DCC channel for protocol testing.f is selected the first bit
NORMAL
received in each byte will be the last bit dropped. Selection of or
REVERSED
both.
DCC INSERT
RECEIVE
DCC DROP
NORMAL
on either display will affect
21
SDH / SONET Modules
SDH / SONET Features
Optical Power Measurement (Options USN and UKT Only)
An optical power measurement is performed on the STM-1/STM-4 input signal if Option USN or UKT, Optical Interface Module is fitted.
Select the required input signal rate on the display.
RECEIVE
Select the required wavelength and view the result on the display.
RESULTS
STM-1/STM-4 Binary Interface (Option OYH with USN or UKT)
If Option OYH is fitted in conjunction with Option USN or UKT, STM-1/STM-4 Optical Interface, binary STM-1 and STM-4 signals can be generated and analyzed by the HP 37717B.
An STM-1 BINARY and STM-4 BINARY choice are added to the SIGNAL softkey menu.
The transmitter and receiver can be configured to operate with Normal or Inverted Clock and Data.
22
2
Add/Drop Multiplexer Testing page 24 Alarm Stimulus/Response page 29 DCC Testing page 33 Desynchroniser Stress page 36 Frame Synchronization page 39 SDH Jitter Transfer page 44 MSP Stimulus/Response page 47 Optical Clock Recovery Stress page 51 Payload Mapping/Demapping page 53 Performance Monitor Stimulus/ Response page 59 Selective Jitter Transfer Measurement page 62
2 SDH / SONET Application
Measurements
SDH / SONET Application Measurements

Add/Drop Multiplexer Testing

Add/Drop Multiplexer Testing
Application
The insertion of tributary signals into the Add/Drop multiplexer, which are then mapped into the SDH / SONET signal, should take place without introducing errors. The insertion and mapping process is tested by adding a test pattern to the tributary inserted at the tributary insert port. At the SDH / SONET side of the Add/Drop multiplexer the tributary is demapped by the HP 37717C Communications Performance Analyzer. By using the Optical Splitter, at the optical side of the Add/ Drop multiplexer, the Add/Drop multiplexer need not be taken out of service. A Bit error rate (BER) test is performed on the recovered tributary test pattern to determine whether errors have been introduced by the Add/Drop multiplexer.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Add/Drop Multiplexer Testing Test Setup Procedure
The following Options must be fitted to the HP 37717C to perform this test:
UKJ (USA)or UKK (USB) - PDH Module
US1 or A1T (A1U) - SDH / SONET Module
UH2 or URU- STM-1/4 Optical Interface
In this setup a 2 Mb/s payload, containing a test pattern, is inserted at the tributary insert port of the Add/Drop multiplexer. A portion of the STM4 Optical signal is tapped off by the Optical Splitter (approx. 10%) and the 2 Mb/s tributary is demapped by the HP 37717C Communications Performance Analyzer.
An Error measurement is performed on the demapped 2 Mb/s tributary test pattern. A SINGLE test period of 24 HOURS is used and the internal printer is enabled to
record results and alarms. The HP 37717C Communications Performance Analyzer GRAPHICS function is
enabled. The graphical results can be viewed on the display
24
GRAPH
SDH / SONET Application Measurements
Add/Drop Multiplexer Testing
1. Connect the HP 37717C to the network
equipment and set up the ;
SETTINGS CONTROL
OTHER
display as shown
opposite.
2. Set up the display as shown
TRANSMIT
opposite.
The PAYLOAD TYPE determines the Framing..
25
SDH / SONET Application Measurements
Add/Drop Multiplexer Testing
3. Set up the display as shown
RECEIVE
opposite.
4. Set up the display, function, as shown
OTHER
LOGGING
opposite.
Continuity Check
Before running the test carry out a continuity test to verify the measurement path.
1. Set up the display as shown
RESULTS
opposite.
2. Press to start a
RUN/STOP
measurement.
3. Press error add three times and
SINGLE
check that the errors are recorded on the
RESULTS
4. Press to stop the
display.
RUN/STOP
measurement.
26
SDH / SONET Application Measurements
Add/Drop Multiplexer Testing
Start the Add/Drop Multiplexer Test
1. Set up the display as shown
RESULTS
opposite.If you do not require stored graphics results select STORAGE [OFF].
2. Press to start the
RUN/STOP
measurement.
The measurement results and alarms are available on the display during
RESULTS
the test period. The graphical measurement results and alarms are stored in non volatile memory for viewing later on the GRAPH display.
The test can be halted at any time by pressing
RUN/STOP
At the End of the Test (Add/Drop Multiplexer Testing)
The Date and Time the test started and the instrument setup are logged on the
internal printer.
Results are logged on the internal printer at 1 hour intervals if the error count is
greater than 0.
Any alarms which occur during the test period will be logged on the internal
printer.
At the end of the test period a complete set of cumulative results are logged on
the internal printer.
A graphical record of the results during the test period can be viewed on the
GRAPH
display. If Remote Control option A3B, A3D, 1A8 or 1CW is fitted the graph results can be logged to an external printer, at a later date. See Graphics and External HP 550C DeskJet Printer.
27
SDH / SONET Application Measurements
Add/Drop Multiplexer Testing
Results and Alarm summaries can be viewed on the display.
GRAPH
The total graphics store capacity is normally 20,000 events. If GRAPH STORAGE RESOLUTION [FULL] is selected on the ; display
OTHER
MISCELLANEOUS
the capacity reduces to 10,000 events. The resolution, determined by the selection made under STORAGE on the
RESULTS
display, affects the ZOOM capability when viewing the bar graphs. If 1 SECOND is selected all resolutions are available under ZOOM. If 1 MIN is selected only 1 MIN/BAR, 15 MINS/BAR and 60 MINS/BAR are available. If 15 MINS is selected only 15 MINS/BAR and 60 MINS/BAR are available. If 1 HOUR is selected only 60 MINS/BAR is available.
Up to 10 sets of graphical results can be stored. If an attempt is made to store more than 10 sets of results, then a first in first out policy is operated and the oldest set of results will be lost. If graphics are enabled and a test is run which exceeds the remaining storage capacity, then some previously stored graphical results will be lost.
T o prev ent accidental overwriting of pre viously stored results the graphics capability should be disabled, when graphical results are not required, by selecting STORAGE [OFF] on the display.
RESULTS
28
SDH / SONET Application Measurements

Alarm Stimulus/Response

Alarm Stimulus/Response
Application
Network elements transmit alarms in response to certain error/alarm conditions to advise upstream and downstream equipment that these conditions exist. If these alarms are not transmitted in the proper manner, at the proper time, degradations in service will occur.
Alarm testing entails transmitting an alarm signal from the Communications Performance Analyzer and monitoring the network equipment alarm indicators and the upstream or downstream signal for the correct response.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Alarm Stimulus/Response Test Setup Procedure
The following options must be fitted to the HP 37717C to perform this test:
US1 or A1T (A1U) - SDH / SONET module
UH1, UH2, URU, USN or UKT - STM-1 Optical interface
In this setup the Communications Performance Analyzer transmits MS AIS Alarm (Line AIS) into the network. The network equipment alarm indicators are monitored for the appropriate alarms. The upstream signal is monitored for occurrences of MS FERF (Line FERF). The downstream signal can be monitored for occurrences of PATH AIS (AIS-P).
A similar procedure can be used for testing all other SDH / SONET alarms. See the following tables.
29
SDH / SONET Application Measurements
Alarm Stimulus/Response
SDH Alarms
Alarm RSTE MSTE PTE
Down Up Down Up Down Up
Loss Of Signal MS AIS N/A Path AIS MS FERF TU-Path AIS TU-Path FERF Loss Of Frame MS AIS N/A Path AIS MS FERF TU-Path AIS TU-Path FERF Loss Of Pointer N/A N/A Path AIS MS FERF TU-Path AIS TU-Path FERF MS AIS N/A N/A Path AIS MS FERF TU-Path AIS TU-Path FERF MS FERF N/A N/A N/A N/A N/A N/A
SONET Alarms
Alarm STE Maintenance Signals LTE Maintenance Signals
Down Up Down Up
Loss Of Signal AIS-L N/A AIS-P Line FERF Loss Of Frame AIS-L N/A AIS-P Line FERF Loss Of Pointer N/A N/A AIS-P Line FERF Line AIS (AIS-L) N/A N/A AIS-P Line FERF Line FERF N/A N/A N/A N/A
30
SDH / SONET Application Measurements
Alarm Stimulus/Response
1. Set up the ;
CONTROL
2. Set up the ; display
OTHER
SETTINGS
display as shown opposite.
TRANSMIT
SDH
as shown opposite.
3. Set up the ;
FUNCTION
TRANSMIT
TEST
display as shown opposite.
The ERROR ADD TYPE selected does not matter as long as RATE [OFF] is selected.
31
SDH / SONET Application Measurements
Alarm Stimulus/Response
4. Set up the display as shown
RESULTS
opposite.ALARM SECONDS are displayed but any of the other results can be selected from the softkey menu without affecting the measurement.
Start the Alarm Stimulus/Response Test
1. Connect the Communications Performance Analyzer to the upstream port of the network equipment and press on the HP 37717C.
RUN/STOP
2. Check that the network equipment registers MS AIS and that MS FERF alarm seconds are recorded on the display.
RESULTS
32
SDH / SONET Application Measurements

DCC Testing

DCC Testing
Application
The section overhead contains two Data Communication Channels (DCC), Regenerator Section DCC at 192 kb/s (overhead bytes D1- D3) and Multiplexer Section DCC at 576 kb/s (overhead bytes D4 - D12). The DCC communicates network management messages between network elements and the network controller via the operations support computer system.
If the DCC is not operating correctly these network management messages will be lost and degradations in network performance will pass unnoticed. This may result in a failure condition.
Full testing of the line and section DCC’s can be carried out using a protocol analyzer connected via the HP 37717C Communications Performance Analyzer to the appropriate overhead bytes. At the far end the HP 37717C Communications Performance Analyzer can drop the selected DCC to the protocol analyzer allowing the DCC integrity to be analyzed.
If you do not have access to a protocol analyzer capable of handling SDH DCC protocol, the DCC integrity can be verified by a BER test using an HP 37732A, Digital Telecomm/Datacomm Analyzer.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state prior to setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
DCC Test Setup Procedure
Alarm Stimulus/Response Test Setup Procedure
The following options must be fitted to the HP 37717C to perform this test:
A1T (A1U) - SDH / SONET module
UH1, UH2, URU, USN or UKT - STM-1 Optical interface
In this procedure the HP 37717C Communications Performance Analyzer accepts a 576 kb/s test pattern via the protocol analyzer port, inserting the test pattern in bytes D4 - D12 of the Multiplexer Section overhead and transmitting an STM-1 optical
33
SDH / SONET Application Measurements
DCC Testing
signal. The HP 37717C Communications Performance Analyzer receives the STM­1 optical signal and drops the Multiplexer Section DCC, via the protocol analyzer port, to the HP 37732A which performs the BER measurement.
1. Connect the HP 37732A and the HP 37717C to the network element, as shown and set up the ;
CONTROL
OTHER
SETTINGS
display as shown opposite.
2. Set up the SDH display as
TRANSMIT
shown opposite.
The CLOCK SYNC selection determines the synchronization source for the
TRANSMIT
clock. If
EXTERNAL MTS
is selected a 2 Mb/s reference must be connected to the front panel 2M REF IN port. The format can be CLOCK or DATA.
34
SDH / SONET Application Measurements
DCC Testing
3. Set up the ;
FUNCTION
4. Set up the ;
FUNCTION
TRANSMIT
display as shown opposite.
RECEIVE
display as shown opposite.
TEST
TEST
Start the DCC Test
1. Select TEST SELECT DATACOM on the HP 37732A.
2. Set TX Clock Source and RX Clock Source to [INTERF ACE] on the HP 37732A (Clock from HP 37717C protocol port).
3. Select the required pattern and monitor logic errors and frequency to verify the integrity of the DCC.
35
SDH / SONET Application Measurements

Desynchroniser Stress

Desynchroniser Stress
Application
At the boundary of the SDH / SONET network the 2 Mb/s or 140 Mb/s payload is demapped from the SDH / SONET signal. Pointer adjustments in the signal may cause high levels of tributary jitter in the output payload. Excessive amounts of tributary jitter will result in errors.
The desynchronizing phase lock loop of the network element should minimize the level of tributary jitter in the payload but correct operation under stress conditions must be verified. The desynchronizing phase lock loop can be stressed by adding pointer movement sequences (defined in CCITT standard G.783) to the SDH / SONET signal such that the test virtual container moves with respect to the SDH / SONET frame.
A jitter measurement is made to verify that the desynchroniser output jitter is within the required specification.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Desynchroniser Stress Test Setup Procedure
The following options must be fitted to the HP 37717C to perform this test:
UKK (USB) or UKJ (USA)- PDH module
UHN (US9) or A3N (A3P) - Jitter measurement module
A1T (A1U) - SDH / SONET module
UH1, UH2, URU, USN or UKT - STM-1/4 Optical interface
The HP 37717C Communications Performance Analyzer transmits an STM-4 optical signal carrying 2 Mb/s payload. Pointer movement sequences are added in a controlled manner. The desynchroniser output is returned to the HP 37717C and a jitter measurement is performed on the demapped 2 Mb/s signal.
36
SDH / SONET Application Measurements
Desynchroniser Stress
1. Connect the HP 37717C to the network equipment and set up the ;
display as shown opposite.
SDH
TRANSMIT
The CLOCK SYNC selection determines the synchronization source for the
TRANSMIT
clock. If
EXTERNAL MTS
is selected a 2 Mb/s reference must be connected to the 2M REF IN port. The format can be CLOCK or DATA.
37
SDH / SONET Application Measurements
Desynchroniser Stress
2. Set up the ;
FUNCTION
TRANSMIT
TEST
display as shown opposite.
Pointer adjustments are made every 10 ms with an extra ADDED adjustment as defined in CCITT standard G.783.
Pointer sequences are started by selecting
STARTED
3. Set up the ; display
.
RECEIVE
JITTER
as shown opposite.
4. Set up the display as shown
RESULTS
opposite. Jitter Hits can also be viewed without affecting the measurement.
Start the Desynchroniser Stress Test
1. Press to start the Jitter measurement.
RUN/STOP
38
SDH / SONET Application Measurements

Frame Synchronization

Frame Synchronization
Application
A network element should maintain synchronization even in the presence of some frame errors. If the number of frame errors exceeds the specified threshold for 3 ms, the network element will lose frame synchronization causing a new search for frame alignment to begin.
The frame synchronization process of the network element can be stressed by injecting frame errors, into the A1 and A2 framing bytes of the Regenerator Section overhead. As the frame error injection rate is increased to the frame synchronization threshold, the network element should indicate Out Of Frame (OOF) and Loss Of Frame (LOF) conditions. As the frame error injection rate is decreased again, the network element should regain frame synchronization.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Frame Synchronization Test Setup Procedure Frame Error Add Test Function In this setup the HP 37717C Communications
Performance Analyzer is used to insert frame errors in the A1 and A2, framing bytes of the Regenerator section overhead of an STM-1 optical signal. The STM-1 optical signal is transmitted to the network equipment. The network equipment OOF and LOF alarms are monitored as the frame error add rate is increased and decreased.
Sequence Generation Test Function
In this setup procedure the HP 37717C Communications Performance Analyzer generates a sequence of errored framing bytes to test the OOF and LOF alarm threshold criteria. The upstream STM-1 optical signal is monitored for occurrences of Multiplexer Section FERF. The downstream STM-1 optical signal can be monitored for AIS.
39
SDH / SONET Application Measurements
Frame Synchronization
1. Connect the HP 37717C to the network
equipment and set up the ;
SETTINGS CONTROL
OTHER
display as shown
opposite.
2. Set up the ; display
TRANSMIT
SDH
as shown opposite. The CLOCK SYNC selection determines
the synchronization source for the
TRANSMIT
clock. If
EXTERNAL MTS
is selected a 2 Mb/s reference must be connected to the front panel 2M REF IN port. The format can be CLOCK or DATA.
40
SDH / SONET Application Measurements
Frame Synchronization
3. Set up the display as shown
RESULTS
opposite.
4. Set up the ;
FUNCTION
TRANSMIT
TEST
display as shown opposite.
Start the Frame Synchronization Test (Frame Error Add)
1. Check that the Loss Of Frame (LOF) alarm indicator on the network element remains unlit and no occurrences of MS FERF are recorded.
2. Increase the Frame Error Add Rate to 2 IN 4 and check that the Loss Of Frame (LOF) alarm indicator on the network element remains unlit and no occurrences of MS FERF are recorded.
3. Increase the Frame Error Add Rate to 3 IN 4 and check that the Loss Of Frame (LOF) alarm indicator on the network element remains unlit and no occurrences of MS FERF are recorded.
4. Increase the Frame Error Add Rate to 4 IN 4 and check that the OOF and LOF alarm indicators on the network equipment are lit and occurrences of MS FERF are recorded.
5. Decrease the Frame Error Add Rate to 3 IN 4 and check that the OOF and LOF alarm indicators on the network equipment remain lit and occurrences of MS FERF are still being recorded.
41
SDH / SONET Application Measurements
Frame Synchronization
6. Decrease the Frame Error Add Rate to 2 IN 4 and check that the OOF and LOF alarm indicators on the network equipment go off, and no further occurrences of MS FERF are recorded.
Sequence Generation Test Function
1. Connect the HP 37717C to the network equipment and set up the
SETTINGS CONTROL
OTHER
display as shown
opposite.
2. Set up the ; display as
TRANSMIT
SDH
shown opposite.
The CLOCK SYNC selection determines the synchronization source for the clock. If is selected a 2
EXTERNAL MTS
TRANSMIT
Mb/s reference must be connected to the front panel 2M REF IN port. The format can be CLOCK or DATA.
3. Set up the display as shown
RESULTS
opposite.
42
SDH / SONET Application Measurements
Frame Synchronization
4. Set up the ; TEST
TRANSMIT
FUNCTION display as shown opposite.
Start the Frame Synchronization Test (Sequence Test)
1. Press on the ; display to start the
STARTED
TRANSMIT
TEST FUNCTION
sequence. As a result of this sequence one OOF alarm second and one LOF alarm second should occur every two seconds.
2. Check that the network element OOF and LOF alarm indicators cycle ON and OFF and that an occurrence of MS FERF is recorded every two seconds.
3. Press to stop the sequence and set up the ;
FUNCTION
4. Press on the ;
TEST FUNCTION
STOPPED
TRANSMIT
TEST
display as shown opposite.
STARTED
TRANSMIT
display to start the sequence. As a result of this sequence one OOF alarm second should occur every two seconds but LOF should not occur.
5. Check that the network element OOF alarm indicator cycles ON and OFF. The LOF alarm should not occur and no occurrences of MS FERF should be recorded.
43
SDH / SONET Application Measurements

SDH Jitter Transfer

SDH Jitter Transfer
Digital transmission systems use Regenerators to transport the signal over long distances. These Regenerators are cascaded together and it is important that each regenerator adds minimal amounts of jitter to the signal.
It is necessary during installation and maintenance to measure the degree to which jitter present at the input is amplified or attenuated by the network elements (Jitter Gain/Transfer).
The jitter transfer measurement entails measuring the input and output jitter at selected jitter frequencies within the jitter bandwidth. The jitter gain is calculated: Jitter Gain (dB) = 20 Log {Jitter out \\over Jitter in}
When the network equipment meets CCITT specification G.823 it should be possible to connect network elements without incurring bit errors.
Default (Known State) Settings
It can be advisable to set the HP 37717C to a known state prior to setting up to make a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Test Setup Procedure (Jitter Transfer Test)
The following Options must be fitted to the HP 37717C to perform this test:
UHK - Jitter Generation
44
SDH / SONET Application Measurements
SDH Jitter Transfer
A1N (A1R), A1P (A1S), A3L (A3M), A3V (A3W) or A3N (A3P) - SDH Jitter
Measurement
US1 or A1T (A1U) - SDH Module This setup procedure is based on 155.52 Mb/s (STM-1), 140 Mb/s payload, PRBS
test data with jitter. The Jitter frequenc y is v aried within the jitter bandwidth and the received jitter is measured to allow calculation of the jitter gain. The internal printer is enabled for recording of results and alarms.
1. Set up the OTHER SETTINGS CONTROL display as shown opposite.
Any SDH settings change made on the
TRANSMIT
or displays will
RECEIVE
automatically occur on the other.
2. Connect the HP 37717C to the line
equipment, select ; and
TRANSMIT
SDH
set up the display as shown opposite.
45
SDH / SONET Application Measurements
SDH Jitter Transfer
3. Select ; and set up
TRANSMIT
JITTER
the display as shown opposite. Select the required Jitter MODULATING
FREQUENCY and AMPLITUDE.
4. Setup the ;
RECEIVE
SDH JITTER
display as shown opposite. If Jitter filtering is required select from the
softkey menu.
Run the Test (Jitter Transfer)
1. Select and set up the display
RESULTS
as shown opposite.
Press to start the
RUN/STOP
measurement.
3. Record the Jitter Amplitude result from the display.
RESULTS
4. Select each jitter Modulating Frequency and Amplitude in turn on the display, press twice and record the Jitter Amplitude result from the
RUN/STOP
TRANSMIT
RESULTS display.
5. Calculate the Jitter gain for each frequency selected. Jitter Gain (dB) = 20 Log {Jitter out / Jitter in}. Where Jitter In is the
AMPLITUDE selected on the display.
TRANSMIT
46
SDH / SONET Application Measurements

MSP Stimulus/Response

MSP Stimulus/Response
Application
Multiplexer Section Protection (MSP) is an optional feature for SDH / SONET Multiplexer Section Terminating Equipment (MSTE). For those MSTE’s, in which it is provided, the MSP system is standardized to ensure the interworking of MSP between MSTE’s from different suppliers.
Standard messages, carried in the K1 and K2 bytes of the SDH / SONET signal transport overhead, indicate the state of the MSP.
Switching to the protection line occurs when one of the following conditions exists for a specified length of time:
Loss Of Signal (LOS)
Loss Of Frame (LOF)
Signal Fail - Bit Error Ratio > 1 X 10-
3
Signal Degrade - Bit Error Ratio programmable
MS AIS
The Signal Degrade Bit Error Ratio threshold is normally programmable in the range 1 X 10-5to 1 X 10-9.
The HP 37717C Communications Performance Analyzer can be used to test Multiplexer Section Protection switching by:
Generating the switching conditions listed above. Transmitting and monitoring the K1 K2 messages.
MSP Stimulus/Response 1+1 Architecture Test Setup Procedure
In this setup the HP 37717C PDH/SDH test set, inserted in the working line, generates MS B2 BIP errors in sufficient quantity to violate the Signal Degrade threshold of the Multiplexer Section Protection. The network equipment Service Terminal indicates that switching to the standby line has occurred. The activity on the K1 K2 bytes can be monitored on the TRANSMIT TEST FUNCTION MSP Messages display.
47
SDH / SONET Application Measurements
MSP Stimulus/Response
1. Set up the ; display
TRANSMIT
SDH
as shown opposite.
The CLOCK SYNC selection determines the synchronization source for the
TRANSMIT
clock. If
EXTERNAL MTS
is selected a 2 Mb/s reference must be connected to the front panel 2M REF IN port. The format can be CLOCK or DATA.
48
SDH / SONET Application Measurements
MSP Stimulus/Response
2. Set up the ;
FUNCTION
TRANSMIT
TEST
display as shown opposite.
1530 MS B2 BIP errors in 1 second corresponds to a BER of 1 in 10-5. The Service terminal should indicate switching to standby within 1 second.
MSP Stimulus/Response 1:N Architecture
The HP 37717C Communications Performance Analyzer ;
TEST FUNCTION MSP MESSAGES
; can be used to transmit and monitor the
TRANSMIT
K1 K2 messages.
The MSP Messages are transmitted when
DOWNLOAD
is pressed.
Two displays of K1 and K2 are provided:
1. Current TX - Values of K1 and K2 bytes which are currently being transmitted.
2. Current RX - Values of K1 and K2 bytes which are currently being received.
K1 Bits 1 ->4 Selects the MSP message to be transmitted.
Table 1 K1 Bits 1 - >4
Selection Message Selection Message
0000 NO REQUEST 1000 MANUAL SWITCH 0001 DO NOT REVERT 1001 NOT USED 0010 REVERSE REQUEST 1010 SD - Low Priority 0011 NOT USED 1011 SD - High Priority
49
SDH / SONET Application Measurements
MSP Stimulus/Response
Table 1 K1 Bits 1 - >4
Selection Message Selection Message
0100 EXERCISE 1100 SF - Low Priority 0101 NOT USED 1101 SF - High Priority 0110 WAIT TO RESTORE 1110 FORCED SWITCH 0111 NOT USED 1111 LOCKOUT OF PROT
SD - High Priority and SF - High Priority are only available when K2 bit 5 is set to 1
- 1: N architecture.
K1 Bits 5 ->8 Selects the channel used by the MSP Messages.
Table 2 K1 Bits 5 - >8
Selection Message Selection Message
0000 NULL CHANNEL 1000 WORKING CHANNEL #8 0001 WORKING CHANNEL #1 1001 WORKING CHANNEL #9 0010 WORKING CHANNEL #2 1010 WORKING CHANNEL #10 0011 WORKING CHANNEL #3 1011 WORKING CHANNEL #11 0100 WORKING CHANNEL #4 1100 WORKING CHANNEL #12 0101 WORKING CHANNEL #5 1101 WORKING CHANNEL #13 0110 WORKING CHANNEL #6 1110 WORKING CHANNEL #14 0111 WORKING CHANNEL #7 1111 EXTRA TRAFFIC CHANNEL
WORKING CHANNEL #2 through WORKING CHANNEL #14 and EXTRA TRAFFIC CHANNEL are only available when K2 Bit 5 is set to 1: N architecture. If K1 bits 1 >4 are set to 1111 LOCKOUT OF PR OT then K1 bits 5 ->8 are fixed at 0000 NULL CHANNEL.
K2 bits 1 - >4 Selects the bridged channel used by the MSP Messages. Can be
set in the range 0000 to 1111.
K2 bit 5 Determines the automatic protection switch architecture. 0 - 1 +
1 architecture 1 - 1: N architecture
50
SDH / SONET Application Measurements

Optical Clock Recovery Stress

K2 bits 6 ->8 Selects the reserved bits. Can be set in the range 000 to 101.
Optical Clock Recovery Stress
Application
Ideally the clock recovery circuits in the network equipment optical interfaces should recover a clock even in the presence of long strings of 0’s.
The optical clock recovery performance of the network equipment can be measured by increasing the length of a zero substitution block until errors occur.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Optical Clock Recovery Stress Test Setup Procedure
In this setup procedure the HP 37717C Communications Performance Analyzer transmits an STM-1 optical signal with zero’s substituted into the payload data pattern. The length of the block of zero’s is increased until the network equipment alarms are triggered.
51
SDH / SONET Application Measurements
Optical Clock Recovery Stress
1. Connect the HP 37717C to the network
equipment and set up the ;
display as shown opposite.
SDH
TRANSMIT
The CLOCK SYNC selection determines the synchronization source for the
TRANSMIT
clock. If
EXTERNAL MTS
is selected a 2 Mb/s reference must be connected to the front panel 2M REF IN port. The format can be CLOCK or DATA.
2. Set up the ;
FUNCTION
TRANSMIT
TEST
display as shown opposite
G.958 Test Pattern consists of consecutive blocks of four types of data: All 1’s PRBS All 0’s a data block consisting of the first row of section overhead bytes.
Start the Optical Clock Recovery Stress Test
Increase the Block Length until the network equipment alarms are triggered.
52
SDH / SONET Application Measurements

Payload Mapping/Demapping

Payload Mapping/Demapping
Application
The mapping and demapping of a 2 Mb/s or 140 Mb/s payload into/from the appropriate SDH / SONET containers should take place without introducing errors.
The mapping process is tested by inserting a test pattern in the 2 Mb/s or 140 Mb/s payload at the low-rate side of the terminal multiplexer. On the high-rate side of the terminal multiplexer , the payload is demapped from the SDH / SONET signal by the HP 37717C Communications Performance Analyzer.
The demapping process is tested by transmitting a SDH / SONET signal to the high­rate side of the multiplexer. On the low-rate side of the multiplexer the payload is received by the HP 37717C Communications Performance Analyzer.
A Bit error rate (BER) test is performed on the recovered payload test pattern to determine whether errors have been introduced by the mapping process.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Payload Mapping/Demapping Test Setup Procedure
The following Options must be fitted to the HP 37717C to perform this test:
US1 or A1T (A1U) - SDH / SONET Module
UH1, UH2, URU, USN or UKT - STM-1/STM-4 Optical Interface
For mapping a 140 Mb/s payload, containing a test pattern, is transmitted into the low-rate side of the terminal multiplexer. The 140 Mb/s payload is demapped from the STM-4 Optical signal at the high-rate side of the terminal multiplexer.
For demapping an STM-4 Optical signal is transmitted into the high-rate side of the Add Drop multiplexer. The 140 Mb/s signal, on the low-rate side of the Add Drop multiplexer , is recei v ed by the HP 37717C Communications Performance Analyzer.
A BER measurement is performed on the demapped 140 Mb/s payload test pattern. A SINGLE test period of 24 HOURS is used and the internal printer is enabled to
record results and alarms.
53
SDH / SONET Application Measurements
Payload Mapping/Demapping
The HP 37717C Communications Performance Analyzer GRAPHICS function is enabled. The graphical results can be viewed on the GRAPH display
Payload Mapping
54
Payload Demapping
t
tttt
SDH / SONET Application Measurements
Payload Mapping/Demapping
1. Connect the HP 37717C to the network equipment and set up the
SETTINGS CONTROL
OTHER
display as shown
opposite.
2. For Mapping set up the
TRANSMIT
display as shown opposite.
2a. For Demapping set up the display as shown opposite.
TRANSMIT
55
SDH / SONET Application Measurements
Payload Mapping/Demapping
3. For Mapping set up the
RECEIVE
display as shown opposite.
3a. For Demapping set up the display as shown opposite.
RECEIVE
4. Set up the display ,
OTHER
LOGGING
function, as shown opposite. All results are logged on the internal printer
at 1 hour intervals. Any alarms which occur during the test period will be logged on the internal printer.
56
SDH / SONET Application Measurements
Payload Mapping/Demapping
Start the Payload Mapping/Demapping Test
1. Set up the display as shown
RESUL TS
opposite. If you do not require stored graphics results select STORAGE [OFF].
2. Press to start the
RUN/STOP
measurement.
The measurement results and alarms are available on the RESULTS display during the test period.
The graphical measurement results and alarms are stored in non volatile memory for viewing later on the display.
The test can be halted at any time by pressing .
GRAPH
RUN/STOP
At the End of the Test (Payload Mapping/Demapping)
The Date and Time the test started and the instrument setup are logged on the
internal printer.
All results are logged on the internal printer at 1 hour intervals.
Any alarms which occur during the test period will be logged on the internal
printer.
At the end of the test period a complete set of cumulative results are logged on
the internal printer.
A graphical record of the results during the test period can be viewed on the
GRAPH display. If Remote Control option A3B, A3D, 1A8 or 1CW is fitted the graph results can be logged to an external printer, at a later date. See Graphics and External HP 550C DeskJet Printer.
Results and Alarm summaries can be viewed on the display.
GRAPH
The total graphics store capacity is normally 20,000 events. If GRAPH STORAGE RESOLUTION [FULL] is selected on the ; display
OTHER
MISCELLANEOUS
the capacity reduces to 10,000 events. The resolution, determined by the selection made under STORAGE on the
RESUL TS
display, affects the ZOOM capability when viewing the bar graphs. If 1
SECOND is selected all resolutions are available under ZOOM. If 1 MIN is selected
57
SDH / SONET Application Measurements
Payload Mapping/Demapping
only 1 MIN/BAR, 15 MINS/BAR and 60 MINS/BAR are available. If 15 MINS is selected only 15 MINS/BAR and 60 MINS/BAR are available. If 1 HOUR is selected only 60 MINS/BAR is available.
Up to 10 sets of graphical results can be stored. If an attempt is made to store more than 10 sets of results, then a first in first out policy is operated and the oldest set of results will be lost. If graphics are enabled and a test is run which exceeds the remaining storage capacity, then some previously stored graphical results will be lost.
T o prev ent accidental overwriting of pre viously stored results the graphics capability should be disabled, when graphical results are not required, by selecting STORAGE [OFF] on the display.
RESULTS
58
SDH / SONET Application Measurements

Performance Monitor Stimulus/ Response

Performance Monitor Stimulus/ Response
Application
Performance monitors built into the SDH / SONET network equipment count BIP errors, and communicate the results to the network controller via the Data Communication Channel (DCC). Performance monitors in Path Terminating Equipment (PTE) also communicate with the upstream equipment.
If the performance monitors are not operating correctly, degradations in network performance will pass unnoticed and may result in a failure condition.
The performance monitors can be tested by the Communications Performance Analyzer transmitting BIP errors in the appropriate byte of the overhead and monitoring upstream for the correct response:
Regenerator Section (RS) - B1 Byte of regenerator section overhead Multiplexer Section (MS) - B2 Bytes of multiplexer section overhead PATH - B3 Byte of path overhead
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a measurement. This clears all previous settings and provides a clearly defined instrument state. For a list of Default Settings and the procedure for accessing them see Stored Settings.
Performance Monitor Stimulus/Response Test Setup Procedure
The following Options must be fitted to the HP 37717C to perform this test:
US1 or A1T (A1U) - SDH / SONET Module
UH1, UH2, URU, USN or UKT - STM-1/STM-4 Optical Interface
In this setup the HP 37717C Communications Performance Analyzer inserts PATH BIP errors in byte B3 of the path overhead of the SDH / SONET signal. The upstream signal is monitored to provide a measure of the FEBE (Far End Block Error) count.
59
SDH / SONET Application Measurements
Performance Monitor Stimulus/ Response
1. Connect the HP 37717C to the network equipment and set up the ;
SETTINGS CONTROL
OTHER
display as shown
opposite.
2. Set up the ; display
TRANSMIT
SDH
as shown opposite.
The CLOCK SYNC selection determines the synchronization source for the
TRANSMIT
clock.
60
SDH / SONET Application Measurements
Performance Monitor Stimulus/ Response
3. Set up the display as shown
RESULTS
opposite.
PATH FEBE ERROR RESULTS are displayed but any of the other results can be selected from the softkey menu without affecting the measurement.
4. Set up the ; , display as shown
TEST
TRANSMIT
FUNCTION
opposite.
The ERROR RATE required can be selected from the softkey menu.
Start the Performance Monitor Stimulus/Response Test
1. Press on the HP 37717C Communications Performance Analyzer.
RUN/STOP
2. Check that the PATH FEBE error rate is the same as the generated PATH B3 BIP rate.
All the measurement results are available, throughout the test, on the
RESULTS
display.
At the end of the test:
the cumulative measurement results are available on the display.
RESULTS
61
SDH / SONET Application Measurements

Selective Jitter Transfer Measurement

Selective Jitter Transfer Measurement
The problem with many SDH jitter analyzers is the fact that their receivers are wideband receivers and are not able to measure within a sufficiently narrow bandwidth. The reason is that these instruments are designed to measure peak to peak jitter in the transmission network for troubleshooting purposes and are not designed to make selective jitter measurements. The jitter analyzer just measures the peak-peak value of the incoming jitter over a wide frequency range. The problem occurs when testing the jitter transfer of real network equipment i.e. SDH regenerators.
The regenerator produces intrinsic jitter and this disturbs the measurement as the jitter receiver cannot determine whether it is measuring the jitter produced by the jitter analyzers transmitter or the intrinsic jitter which is generated, at a different frequency , by the re generator . The problem is greatest at the higher jitter modulating frequencies when the amount of jitter generated, as per ITU-T G.958, is much smaller. The measurement is corrupted by the higher amplitude intrinsic jitter generated by the regenerator at lower frequencies and incorrectly measured by the analyzer.
The accurate method for measuring jitter transfer requires a selective measurement. One such method is to use a network analyzer in conjunction with the HP 37717C. The network analyzer provides the capability to measure jitter selectively and has increased sensitivity.
Default (Known State) Settings
It can be advisable to set the HP 37717C to a known state prior to setting up to make a measurement. This clears all previous settings and provides a clearly defined instrument state. The default settings are set by selecting
SETTINGS RECALL
Test Setup Procedure (Jitter Transfer Test)
The following Options must be fitted to the HP 37717C to perform this test:
A3K - Jitter Generation
A1M, A1N, A1P, A3L, A3V or A3N - SDH Jitter Measurement
US1 or A1T - SDH Module
62
STORED SETTING NUMBER 0 and pressing .
OTHER
STORED
SDH / SONET Application Measurements
Selective Jitter Transfer Measurement
This setup procedure is based on 155.52 Mb/s (STM-1), 140 Mb/s payload, PRBS test data with jitter. The jitter modulation is provided by the network analyzer. The HP 37717C demodulated jitter output is returned to the network analyzer for measurement. Before connecting to the regenerator to be tested the HP 37717C is looped back to back and the network analyzer is programmed to sweep over the required frequency range at the required amplitude. This provides a reference trace and removes the inaccuracies of the of the test configuration (inaccuracies of the HP 37717C and the Network Analyzer). The HP 37717C is connected to the regenerator and the network analyzer sweep is repeated. The difference between the two traces is the jitter transfer result.
TEST SET
Selective Jitter Transfer Test
1. Set up the OTHER SETTINGS CONTROL display as shown opposite.
Any SDH settings change made on the
TRANSMIT
or displays will
RECEIVE
automatically occur on the other.
63
SDH / SONET Application Measurements
Selective Jitter Transfer Measurement
2. Connect the HP 37717C to the network analyzer as shown. Connect STM-1/STM­4 IN to STM-1/STM-4 OUT. Select
TRANSMIT
SDH SDH
and set up
the display as shown opposite.
3. Select
TRANSMIT
SDH
JITTER
and set up the display as shown opposite.
4. Setup the
RECEIVE
SDH JITTER
display as shown opposite. If Jitter filtering is required select from
the softkey menu.
64
SDH / SONET Application Measurements
Selective Jitter Transfer Measurement
5. Select and set up the
RESULTS
display as shown opposite. Press to start the
RUN/STOP
measurement.
6. Adjust the network analyzer output level until the display. records the
RESULTS
required peak-peak jitter value.
7. Press to stop the
RUN/STOP
measurement.
0.01
0.01
0.01
8. Start the network analyzer sweep and store the resultant “reference trace”
9. Connect the HP 37717C to the regenerator as shown (loopback removed) and repeat the network analyzer sweep.
The difference between the two traces is the Jitter Transfer result.
65
SDH / SONET Application Measurements
Selective Jitter Transfer Measurement
66
3

3 ETSI / ANSI Terminology

A table of ETSI Terms with their ANSI equivalents.
ETSI / ANSI Terminology

ETSI / ANSI Equivalent Terms

ETSI / ANSI Equivalent Terms
The Terminology used on the instrument display is mainly ETSI terminology. The equivalent ANSI terminology is given in the following table
ETSI Term ANSI Term
I-n Inter Office, STM-n Intermediate Reach (IR) L-n.1 or L-n.2 long haul LR long reach Multiplexer Section (MS) Line MS-AIS Line AIS (AIS-L) MS-BIP Line BIP MS-DCC Line DCC MS FERF Line FERF MS-RDI RDI-L Multiplexer Section Overhead Line Overhead Network Node Interface Line Interface Path AIS AIS-P Path FERF RDI-P Regenerator Repeater Regenerator Section (RS) Section Remote Alarm Indicator Yellow Alarm Regenerator Section Overhead Section Overhead RS-DCC Section DCC S-n.1 or S-n.2 short haul Short Reach (SR) STM-n STS-n SOH TOH Section Overhead (SOH) Transport Overhead (TOH)
68
ETSI / ANSI Terminology
ETSI / ANSI Equivalent Terms
ETSI Term ANSI Term
Tributary Unit (TU) Virtual Tributary (VT) TU VT TU AIS VT AIS (AIS-V) TU FERF / TU RDI RDI-V / VT FERF TU REI VT FEBE VC SPE Virtual Container Payload Envelope Virtual Container (VC) Synchronous Payload Envelope (SPE) VP-FERF VP-RDI VC-FERF VC-RDI
NOTE: VC is an ETSI abbreviation for Virtual Container and an ETSI / ANSI abbreviation for (ATM) Virtual Channel. The context of VC must therefore be tak en into account when converting between standards.
69
ETSI / ANSI Terminology
ETSI / ANSI Equivalent Terms
70
Index
A
Add/Drop Multiplexer Testing, 24, 28 Alarm Stimulus/Response, 29, 32 Alarms
SDH, 30 SONET, 30
ANSI / ETSI Equivalent Terms, 68
B
Binary Interface STM-1/STM-4, 22
C
CLOCK SYNC, 5
D
DCC Drop & Insert, 21 DCC Testing, 33, 35 Desynchroniser Stress, 36, 38
E
ERROR ADD TYPE Selection, 31 Errors & Alarms, 16 ETSI / ANSI Equivalent Terms, 68
F
Frame by Frame Monitor of H1,H2, 15 Frame Synchronization, 39, 43 FREQUENCY OFFSET, 5
I
Introduction to SDH and SONET, 2
M
MSP Messages, 20 MSP Stimulus/Response, 47, 51
N
NetworkStress Testing
Desynchroniser Stress, 36
O
OC-3c Transmission, 6 OFFSET, 6 Optical Clock Recovery Stress, 51, 52 Optical Power Measurement, 22 Option A1T, 4 Option A1U, 4 Option OYH, 22
Option UKT, 22 Option US1, 3 Option US5, 3 Options USN, 22 Overhead BER, 17 Overhead Bytes, 13 Overhead Capture, 15 Overhead Monitor, 14 Overhead Sequence Generation, 13
P
Path Overhead, 9 Path Trace, 9 PATTERN, 6 PATTERN IN OTHER TU, 6 PAYLOAD, 5 Payload Mapping/Demapping, 53, 58 PAYLOAD OVERWRITE, 8 Performance Monitor Stimulus/Response,
61
PerformanceMonitor Stimulus/Response,
59 Pointer Adjustments, 17, 19 Pointer Graph, 19
R
Receive Overhead Capture, 15, 16 Receive Overhead Monitor, 14
S
SDH Alarms, 30 SDH Features
Options US1, US5, A1T and A1U, 5 SDH Introduction, 2 SDH Jitter Transfer, 44 SDH Options, 3, 4 SDH Settings, 5 Section / Transport Overhead, 11 SELECTED TU, 6 Selective Jitter Transfer Measurement, 62 Setting Undefined Overhead Functions, 11 SONET Alarms, 30 SONET Introduction, 2 SONET With ATM Payload, 6 Static Overhead Known, Default, Condi-
tions, 12 STM-1 THRU Mode Transmission, 7 STM-1/STM-4 Binary Interface, 22
STM-4 THRU Mode Transmission, 8 Stres Test, 21 SYNC Messages, 21
T
THRU Mode, 7, 8 Transient Detection, 16 Transmit Frame Synchronization Se-
quence, 13 Transmit Overhead Sequence, 13 Transmit Static Overhead, 9, 13 TU CONCATENATION, 5 TU MODE, 5 TU PAYLOAD, 6
71
Index
72
Hewlett-Packard Sales and Service Offices
Hewlett-Packard Sales and Service Offices
United States:
Hewlett-Packard Company 2101 Gaither Road Rockville MD 20850 (301) 258-2000
Hewlett-Packard Company 5201 Tollview Drive Rolling Meadows IL 60008 (708) 255-9800
Hewlett-Packard Company 1421 S. Manhattan Avenue Fullerton CA 92631 (714) 999-6700
Hewlett-Packard Company 2000 South Park Place Atlanta GA 30339 (404) 955-1500
Japan:
Yokogawa-Hewlett-Packard Ltd. Measurement Assistance Center 9-1, Takakura-Cho Hachioji-Shi Tokyo 192 Japan (81) 426 48 0722
Latin America:
Hewlett-Packard Latin America Region Headquarters 5200 Blue Lagoon Drive 9th Floor Miami Florida 33126 USA (305) 267 4245/4220
Australia/New Zealand:
Canada:
Hewlett-Packard Canada Ltd. 5150 Spectrum Way Mississauga Ontario L4W 5G1 (416) 206-4725
Europe:
Hewlett-Packard European Marketing Centre PO Box 999 1180 AZ Amstelveen The Netherlands
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn Victoria 3130 Australia Melbourne Caller 272 2555 (008) 13 1347
Far East:
Hewlett-Packard Pacific Ltd. 22-30/F Peregrine Tower Lippo Centre 89 Queensway Central Hong Kong (852) 848 7070
Learning Products Map
All of the learning products which apply to the HP 37717C Communications Performance Analyzer with ATM Services and LAN testing capability are shown below:
The HP 37717C Mainframe Operating Manual- 37717-90282
General operating information irrespective of option.
The HP 37717C PDH / DSn Operating Manual - 37717-90283
Information about the PDH / DSn modules, how to select the features available and measurement examples.
The HP 37717C SDH / SONET Operating Manual - 37717-90284
Information about the SDH / SONET modules, how to select the features available and measurement examples. This book also contains a table of ANSI / ETSI equivalent terms.
The HP 37717C Jitter Operating Manual - 37717-90285
Information about the Jitter modules, how to select the features available and measurement examples.
The HP 37717C ATM + LAN Operating Manual - 37717-90286
Information about the ATM and LAN modules, how to select the features available and measurement exam­ples. This book also contains tutorial information on some ATM and LAN measurements, Information on pre­stored sequences and a glossary of ATM and LAN terms.
Calibration Manual - 37717-90287:
Provides specifications and methods of testing that the instrument meets its specifications.
Remote Control Manual - 37717-90288:
Provides remote control information for instruments fitted with the RS232 and HP-IB remote control option modules.
About This Edition
This is the 1st edition of the 37717-90284 manual. It documents the product as of September 1997. Edition dates are as follows:
1st Edition, September 1997
Copyright Hewlett­Packard Ltd. 1997. All rights reserved. Reproduction, adaption, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
In This Book
This book provides information on HP 37717C modules with SDH / SONET capability when used with instruments which have ATM Services and LAN testing capability. It also provides applications associated with these modules. The individual applications contain techniques which may be of value for purposes other than those shown
Printed in U.K. 09/97 37717-90284
Loading...