All rights reserved.
Reproduction, adaption,
or translation without prior
written permission is
prohibited, except as
allowed under the
copyright laws.
Information in this
document may apply to
modules which use the
VxWORKS TM software.
The VxWORKS TM
software was developed by
Wind River Systems, Inc.,
which has copyright to it.
HP Part No. 37717-90407
First edition, March 98
Printed in U.K.
Documentation
Warranty
The information contained
in this document is subject
to change without notice.
Hewlett-Packard mak es no
warranty of any kind with
regard to this material,
including, but not limited
to, the implied warranties
or merchantability and
fitness for a particular
purpose.
Hewlett-Packard shall not
be liable for errors
contained herein or for
incidental or
consequential damages in
connection with the
furnishing, performance,
or use of this material.
WARNING
Warning Symbols Used
on the Product
!
The product is marked
with this symbol when the
user should refer to the
instruction manual in order
to protect the apparatus
against damage.
The product is marked
with this symbol to
indicate that hazardous
voltages are present
The product is marked
with this symbol to
indicate that a laser is
fitted. The user should
refer to the laser safety
information in the
Calibration Manual.
Hewlett-Packard Limited
Telecommunications Networks Test Division
South Queensferry
West Lothian, Scotland EH30 9TG
PDH Concept Guide
HP 37717C Communications
Performance Analyzer
About This Book
The information on PDH testing in this book covers the following subjects:
• An Introduction to PDH, the PDH modules and their features.
• Measurement examples.
• Measurement result definitions
• Logging messages
iv
Contents
1 Introduction to PDH Testing
Introduction to PDH / DSn 2
Unstructured PDH Generation and Measurement. Option UKK (USB)
3
Structured PDH Generation and Measurement. Options UKJ (USA)
and UKN (USE) 4
Binary Interfaces. Option UH3 (US7) 5
Multiple PDH Outputs. Option UHC (US6) 6
DS1, DS3, E1 and E3 Transmit and Receive Interfaces for PDH/DSn
Payloads. Option 110 7
2 PDH / DSn Testing
Alarm Monitoring 10
Analysis of N x 64 kb/s 13
BERT Testing 17
Cross Multiplexer Testing 22
FAS Monitoring 28
Frequency Measurement 32
Frequency Offset Tolerance 35
Multiplexer Testing 39
Round Trip Delay 44
M.2100, M.2110, M.2120 Analysis 47
Introduction to PDH / DSn page 2
Unstructured PDH Generation and Measurement. Option UKK (USB) page 3
Structured PDH Generation and Measurement. Options UKJ (USA) and UKN (USE) page 4
Binary Interfaces. Option UH3 (US7) page 5
Multiple PDH Outputs. Option UHC (US6) page 6
DS1, DS3, E1 and E3 Transmit and Receive Interfaces for PDH/DSn Payloads. Option 110
page 7
1Introduction to PDH Testing
Information on PDH in general and the PDH test
features of the HP 37717C.
Introduction to PDH Testing
Introduction to PDH / DSn
Introduction to PDH / DSn
The Plesiochronous Digital Hierarchy (PDH / DSn) is still the dominant
technology in most existing telecommunications networks throughout
the world, although it is being replaced in many networks by
Synchronous Digital Hierarchy (SDH) or SONET networks. PDH / DSn
networks were developed at a time when point-to-point transmission was
the predominant network requirement. To support this requirement, the
standard approach to network management and maintenance was to use
manual distribution frames for access to individual signals. This is now
considered out of date and consequently SDH / SONET is now the
preferred network topology for new installations.
However PDH / DSn networks will exist for a long time to come even in
networks where SDH /SONET is the preferred technology. PDH is the
ETSI international standard, based on 2 Mb/s, defined by the ITU-T, and
covers the hierarchal transmission rates of 2 Mb/s, 8 Mb/s, 34 Mb/s and
140 Mb/s. DSn is the ANSI standard covering transmission rates of 1.544
Mb/s (DS1) and 44.736 Mb/s (DS3). PDH is asynchronous at 8 Mb/s, 34
Mb/s and 140 Mb/s. In order to access a signal, for rerouting or test
purposes, the whole line signal structure must be demultiplexed step by
step down to the 2 Mb/s level, because of the asynchronous nature of the
multiplexing.
At each multiplexing step, the bit rate of the individual tributary signals
is controlled within specified limits and is not synchronized with the
multiplex equipment. Because the bit rates of the individual tributaries
are controlled within specific limits this type of multiplexing is referred
to as Plesiochronous i.e. nearly synchronous. The individual tributaries
are synchronized with the equipment at each multiplex step by the
process of positive bit stuffing justification.
In new SDH / SONET networks, PDH / DSn signals are mapped into
virtual containers / tributaries before being transported as part of the
SDH / SONET payload. The SDH / SONET payload must then be
demapped into a PDH / DSn tributary signal.
Therefore all PDH / DSn, SDH / SONET and mixed PDH/SDH DSn /
SONET networks require test sets which have PDH / DSn interfaces and
PDH / DSn test capability.
2
Introduction to PDH Testing
Unstructured PDH Generation and Measurement. Option UKK (USB)
Unstructured PDH Generation and Measurement. Option
UKK (USB)
Option UKK (USB), Unstructured PDH, provides generation and
measurement of unstructured PDH at interface rates of 704 kb/s,
2.048 Mb/s, 8.448 Mb/s, 34.368 Mb/s and 139.264 Mb/s
Allows the addition of Frequency Offset to the PDH signal
Provides 75 Ω unbalanced or 120 Ω balanced input and output
interfaces.
Allows selection of PRBS, WORD or USER patterns
Allows selection of Line Code AMI, HDB3 or CMI
Provides an ECL pulse each time an error occurs (Error Out)
Option UKK [USB]
3
Introduction to PDH Testing
Structured PDH Generation and Measurement. Options UKJ (USA) and
UKN (USE)
Structured PDH Generation and Measurement. Options
UKJ (USA) and UKN (USE)
Options UKJ (USA) and UKN (USE), provide
generation and measurement of Structured or
Unstructured PDH at interface rates of 2.048 Mb/s (E1),
8.448 Mb/s (E2), 34.368 Mb/s (E3) and 139.264 Mb/s
(E4).
Structured test signal rates of 64 kb/s, 2.048 Mb/s, 8.448
Mb/s and 34.368 Mb/s are provided.
Frequency Offset can be added at the interface rates.
Line rate frequency can be measured.
Provides 75 Ω unbalanced or 120 Ω balanced input and
output interfaces.
Allows selection of PRBS, WORD or USER patterns
Option UKJ (USA)
Allows selection of Line Code AMI, HDB3 or CMI
Interface and test signal rates can be Framed or
Unframed
MUX connector allows a 2 Mb/s signal from external
equipment to be inserted into the HP 37717C test
signal.
DEMUX connector allows a 2 Mb/s signal from the HP
37717C to be Dropped to external equipment.
Option UKN (USE) also provides generation and
measurement of PDH signals with ATM payloads.
Information on the ATM capabilities of the UKN (USE)
module is given in the “Users Guide” and the “ATM
Concept Guide”.
4
Introduction to PDH Testing
Binary Interfaces. Option UH3 (US7)
Binary Interfaces. Option UH3 (US7)
Option UH3 (US7) provides binary NRZ interfaces for the structured
PDH module option UKJ (USA), and the unstructured PDH module
option UKK (USB).
The interfaces can operate at any of the standard rates ± 100 ppm.
When used with option UKK (USB), with an external binary clock input,
the interfaces can operate at any rate in the range 700 kb/s to
170 Mb/s.
TTL signal levels, at data and clock ports, and external clock port, are
valid from 700 kb/s to 50 Mb/s, 75Ω to ground.
ECL signals, at data and clock ports, and external clock port, are valid
from 700 kb/s to 170 Mb/s, 75Ω to -2V.
The external clock input may be used to clockcoded data from the PDH
/ DSn transmitter at standard rates. In addition if Option UKK (USB),
Unstructured PDH, is fitted can clock coded data in the range 700 kb/s
to 170 Mb/s.
Option UH3 (US7)
Simultaneous jitter of binary clock and binary data outputs is possible
(2, 8, 34 & 140 Mb/s) if Option UHK (USB) or A3K, Jitter Generator, is
fitted.
Jitter measurement on the binary clock input is possible (2, 8, 34 & 140
Mb/s ±;100 ppm) if Option UHN, A3L, A3V or A3N , Jitter Measurement,
is fitted.
Frequency measurement of the external clock input and binary clock
input is provided.
OUT 1 is delayed by 4 bits relative to the main PDH OUT.
OUT 2 is delayed by 8 bits relative to the main PDH OUT.
OUT 3 is delayed by 12 bits relative to the main PDH OUT.
6
Introduction to PDH Testing
DS1, DS3, E1 and E3 Transmit and Receive Interfaces for PDH/DSn
Payloads. Option 110
DS1, DS3, E1 and E3 Transmit and Receive Interfaces for PDH/
DSn Payloads. Option 110
Option 110, provides generation and measurement of
PDH /DSn signals at interface rates of: 1.544 Mb/s (DS1),
44.736 Mb/s (DS3), 2.048 Mb/s (E1), and 34.368 Mb/s (E3).
Frequency Offset can be added at the interface rates.
Line rate frequency can be measured.
Provides the following RZ input and output interfaces:
Allows selection of Line Code as:
AMI or B8ZS at DS1
AMI or HDB3 at 2Mb/s.
At DS3 the line code is fixed at B3ZS.
At 34 Mb/s the line code is fixed at HDB3.
Provides Input sensitivity level selection.
Provides output level control for DS1 and DS3.
DS1: DSX-1 or DS1-LO
DS3: DS3-HI, DSX-3 or DS3-900’.
Allows errors and alarms to be injected into the test signal.
An E1 Drop/Insert port allows an E1(2.048 Mb/s) signal to be
dropped/inserted from a E3 (34 Mb/s) signal. In conjunction
with the SONET/SDH test module (option 120), this port
allows the insertion of E1 into a TU-12/VT2.
A DS1 Drop/Insert port allows a DS1 signal to be dropped/
inserted from a DS3 or TU-11/VT1.5 when used in
conjunction with the SONET/SDH test module.
7
Introduction to PDH Testing
DS1, DS3, E1 and E3 Transmit and Receive Interfaces for PDH/DSn
Payloads. Option 110
8
2
Alarm Monitoring page 10
Analysis of N x 64 kb/s page 13
BERT Testing page 17
Cross Multiplexer Testing page 22
FAS Monitoring page 28
Frequency Measurement page 32
Frequency Offset Tolerance page 35
Multiplexer Testing page 39
Round Trip Delay page 44
M.2100, M.2110, M.2120 Analysis page 47
2PDH / DSn Testing
This Chapter gives examples of the instrument
operation in typical PDH test applications.
PDH / DSn Testing
Alarm Monitoring
Alarm Monitoring
Options Required Structured PDH UKJ (USA), 110 or UKN (USE)
Application
Problems in the network at all levels in the hierarchy can be detected by the
occurrence of alarms in each tributary of structured PDH systems.
Using the HP 37717C in a receive only mode, each tributary can be scanned and the
state of Frame Loss, Remote (RAI) and AIS alarms viewed on the
display.
Default (Known State) Settings
It can be advisable to set the HP 37717C to a known state prior to setting up to make
a measurement. This clears all previous settings and provides a clearly defined
instrument state. Select,, set STORED
SETTINGS NUMBER to 0 and ACTION to.
OTHER
STORED SETTINGS
RECALL
RESULTS
Test Setup Procedure (Alarm Monitoring)
This setup procedure is based on Structured PDH 140 Mb/s line traffic interfaced at
the line equipment 75 Ω protected Monitor point. If a protected Monitor Point is not
available then an HP 15510A Protective Probe may be used at an unprotected
Monitor point.
When 120 Ω Balanced alarm monitoring is desired at an unprotected Monitor point
use an HP 15511A Protective Monitor Probe. The instrument is used in a receiveonly mode to monitor Frame Loss, AIS and Remote alarms.
10
PDH / DSn Testing
Alarm Monitoring
1 Connect the HP 37717C to the line
terminal equipment protected Monitor
point and set up the display
MAIN SETTINGS
RECEIVE
as shown opposite.
The GAIN and EQUALIZER settings
should be set to optimize the received
signal.
2 Set up the display
STRUCTURED SETTINGS
RECEIVE
as shown
opposite.
2M PAYLOAD determines the Framing
and selection which should match that of
the network equipment.
11
PDH / DSn Testing
Alarm Monitoring
Start the Test (Alarm Monitoring)
1 Set up the display as shown
opposite and select
ON
RESULTS
PDH ALM SCAN
.
If any of the three alarms, Frame Loss,
RAI or AIS has occurred the appropriate
point in the hierarchy will be highlighted.
The test can be halted at any time by
selecting on the
PDH ALM SCAN
OFF
RESULTS
display.
2 T o determine which alarm has occurred
set the PDH ALM SCANand
then set the
STRUCTURED
RECEIVE
SETTINGS
OFF
display
to the tributary highlighted as shown
opposite and press.
RUN/STOP
Example 2Mb number 2 of 8Mb number 1
of 34 Mb number 1 is highlighted.
Many customer premises receive subrate signals for example 128 Kb/s or 384 kb/s.
The timeslots which make up these services may or may not be contiguous. Testing
these services requires that the test set to be able to insert a pattern across the
required timeslots.
T esting N x 64 kb/s channels structured within a 2 Mb/s signal can be carried out on
an End to End basis using two test sets.
Default (Known State) Settings
It can be advisable to set the HP 37717C to a known state prior to setting up to make
a measurement. This clears all previous settings and provides a clearly defined
instrument state. Select,, set STORED SETTINGS
NUMBER to 0 and ACTION to.
OTHER
STORED SETTINGS
RECALL
Test Setup Procedure (N x 64 kb/s Analysis)
This setup procedure is interfaced at 2 Mb/s, with a test signal of 6 X 64 kb/s noncontiguous timeslots. A PRBS test pattern is transmitted across the 6 timeslots. A
BER measurement is performed on the received test pattern.
13
PDH / DSn Testing
Analysis of N x 64 kb/s
HP 37717C #1
1. Set up the
CONTROL
OTHER
SETTINGS
display as shown opposite.
Any settings change made on the
or displays will
TRANSMIT
RECEIVE
automatically occur on the other.
2. Connect the HP 37717C to the line
equipment and set up the
MAIN SETTINGS
TRANSMIT
display as shown
opposite.
The settings of SIGNAL rate and LINE
CODE must match those of the network
equipment.
PAYLOAD TYPE selects the Framing
which must match that of the network
equipment.
3. Set up the display
STRUCTURED SETTINGS
TRANSMIT
as shown
opposite.
In this example timeslots 3, 5, 9, 25, 26
and 27 are selected.
14
PDH / DSn Testing
Analysis of N x 64 kb/s
4. Set up the display as shown
RESULTS
opposite.
The RESULTS type may be changed
during the measurement without
interrupting the test.
HP 37717C #2
1. Set up the
CONTROL
OTHER
display as shown opposite.
SETTINGS
Any settings change made on the
TRANSMIT
or displays will
RECEIVE
automatically occur on the other.
2. Set up the display
MAIN SETTINGS
RECEIVE
as shown opposite.
The settings of SIGNAL rate and LINE
CODE must match those of the network
equipment.
PAYLOAD TYPE selects the Framing
which must match that of the network
equipment.
15
PDH / DSn Testing
Analysis of N x 64 kb/s
3. Set up the
SETTINGS
RECEIVE
display as shown opposite.
STRUCTURED
In this example timeslots 3, 5, 9, 25, 26
and 27 are selected.
4. Set up the display as shown
RESULTS
opposite.
The RESULTS type may be changed
during the measurement without
interrupting the test.
Run the Test (N x 64 kb/s Analysis)
1. Press.
• The measurement results and alarms are available on the display
A transmission system must be specified for its overall error performance, measured
over a period of time. Conformance to these specifications ensures that an installed
system will meet the requirements of an Integrated Digital Network (IDN).
After troubleshooting, or during installation or commissioning, it is necessary to
check that the transmission link meets this error performance.
This can be performed in two ways:
• End T o End - Error performance measurements are made on an end-to-end basis
testing the Go and Return paths separately but simultaneously . The measurements
are often performed unattended and the results and other events, alarms for
example, logged on a printer or disc and timed by a real time clock facility.
Two HP 37717C's are required for this measurement, one at each end of the link.
End-to End Test
17
PDH / DSn Testing
BERT Testing
• Loopback - Error performance measurements are made via a loopback at the
remote end of the system testing the combined Go and Return paths. The
measurements are often performed unattended and the results and other events,
alarms for example, logged on a printer or disc and timed by a real time clock.
Loopback Test
Default (Known State) Settings
It can be advisable to set the HP 37717C to a known state prior to setting up to make
a measurement. This clears all previous settings and provides a clearly defined
instrument state. Select,, set STORED SETTINGS
OTHER
NUMBER to 0 and ACTION to.
STORED SETTINGS
RECALL
Test Setup Procedure (BERT Testing)
This setup procedure is based on 140 Mb/s, CMI, PRBS test data terminated in 75Ω .
A SINGLE test period of 24 HOURS is used and use of the a printer or disc for
recording of results and alarms is included. A graphical record of the results can be
viewed on the HP 37717C display at the end of the test period.
GRAPH
18
PDH / DSn Testing
BERT Testing
1. Set up the
CONTROL
OTHER
display as shown opposite
SETTINGS
(on both HP 37717C's if end to end).
Any settings change made on the
or displays will
TRANSMIT
RECEIVE
automatically occur on the other.
2. Connect the HP 37717C to the line
equipment and set up the
TRANSMIT
display as shown opposite (on both HP
37717C's if end to end).
3. Select the function as
LOGGING
shown opposite (on both HP 37717C's if
end to end).
A LOGGING PERIOD selection of
[USER PROGRAM] [10 MIN] provides
the following:
A complete set of period and a complete
set of cumulative results logged on the
printer or disc every 10 minutes.
A complete set of cumulative results logged on the printer or to disc at the end of the
test period.
19
PDH / DSn Testing
BERT Testing
4. Set the
TIMING CONTROL
RESULTS
as shown opposite
(on both HP 37717C's if end to end).
5. Select the
ERROR ANALYSIS
RESULTS
PDH RESULTS
display
The RESULTS type displayed can be
changed during the test period without
interrupting the test.
The RESULTS
TIMING CONTROL
STORAGE selection enables the graphics.
To disable graphics select STORAGE
[OFF].
Start the Test (BERT Testing)
1. Press (on both HP 37717C's if end to end).
The measurement results and alarms are available on the display during
RUN/STOP
RESULTS
the test period.
The graphical measurement results and alarms are stored in non volatile memory for
viewing later on the display.
The test can be halted at any time by pressing.
GRAPH
RUN/STOP
At the End of the Test (BERT Testing)
• The Date and Time the test started and the instrument setup are logged on the
selected logging device.
• All results are logged on the selected logging device at 10 minute intervals.
• Any alarms which occur during the test period will be logged on a printer or disc.
• At the end of the test period a complete set of cumulative results are logged on
the printer or disc.
• A graphical record of the results during the test period can be viewed on the
20
PDH / DSn Testing
BERT Testing
GRAPH
display. If Remote Control option A3B or A3D is fitted the graph results
can be logged to an external printer, at a later date.
• Results and Alarm summaries can be viewed on the display.
GRAPH
The total graphics store capacity is normally 20,000 events. If GRAPH STORAGE
RESOLUTION [FULL] is selected on the display
OTHER
MISCELLANEOUS
the capacity reduces to 10,000 events.
The resolution, determined by the selection made under STORAGE on the
RESULTS
TIMING CONTROL
display, affects the ZOOM capability when
viewing the bar graphs. If 1 SECOND is selected all resolutions are available under
ZOOM. If 1 MIN is selected only 1 MIN/BAR, 15 MINS/BAR and 60 MINS/BAR
are available.If 15 MINS is selected only 15 MINS/BAR and 60 MINS/BAR are
available. If 1 HOUR is selected only 60 MINS/BAR is available.
Up to 10 sets of graphical results can be stored. If an attempt is made to store more
than 10 sets of results, then a first in first out policy is operated and the oldest set of
results will be lost. If graphics are enabled and a test is run which exceeds the
remaining storage capacity, then some previously stored graphical results will be
lost.
T o prev ent accidental overwriting of pre viously stored results the graphics capability
should be disabled, when graphical results are not required, by selecting STORAGE
[OFF] on the display.
For comprehensive testing of network equipment it is essential that the test
equipment can multiplex/demultiplex the test signal.
The insertion of tributary signals into the PDH multiplexer, which are then
multiplexed into the 140 Mb/s PDH structure should take place without introducing
errors. The insertion and structuring process is tested by adding a test pattern to the
tributary inserted at the tributary insert port. At the high rate side of the PDH
multiplexer the tributary is destructured and a BER test performed.
By using a protected monitor point at the high rate side of the PDH multiplexer the
mux/demux need not be taken out of service.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a
measurement. This clears all previous settings and provides a clearly defined
instrument state. Select,, set STORED SETTINGS
NUMBER to 0 and ACTION to.
Cross Multiplexer Testing Test Setup Procedure
In this setup a 2 Mb/s Framed tributary, containing a test pattern, is inserted at the
tributary insert port of the PDH multiplexer. The 140 Mb/s structured PDH signal is
obtained from a protected monitor point. The 2 Mb/s tributary is destructured by the
HP 37717C test set and an Error measurement is performed on the 2 Mb/s tributary
test pattern.
A SINGLE test period of 24 HOURS is used and the printer is enabled to record
results and alarms.
The HP 37717C PDH/SDH test set GRAPHICS function is enabled. The graphical
results can be viewed on the display.
22
OTHER
STORED SETTINGS
RECALL
GRAPH
PDH / DSn Testing
Cross Multiplexer Testing
Cross Multiplexer Testing
1. Connect the HP 37717C to the network
equipment and set up the
SETTINGS CONTROL
OTHER
display as shown
opposite.
23
Loading...
+ 67 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.