Agilent 34401A Service Guide

Agilent 34401A 6 ½ Digit Multimeter
Service Guide
Agilent Technologies
Notices
© Agilent Technologies, Inc. 1991 - 2007
No p art o f this manual may be rep roduc ed in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
34401-90013
Edition
Seventh Editio n. August 2007
Printed in Malaysia
Agilent Technologies, Inc. 3501 Stevens Creek Blvd. Santa Clara, CA 95052 USA
Microsoft® and Windows® are U.S. regis­tered trademarks of Microsoft Corporation.
Software Revision
This guide is valid for the firmware that was installed in the instrument at the time of manufacture. However, upgrading the firm­ware may add or change product features. For the latest firmware and documentation, go to the product page at:
www.agilent.com/find/34401A
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Furt her, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a haz- ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
ii 34401A Service Guide
Safety Information
General
Do not use this product in any manner not specified by the manufacturer. The protec­tive features of this product may be impaired if it is used in a manner not speci­fied in the operation instructions.
Do not install substitute parts or perform any unauthorized modification to the prod­uct. Return the product to an Agilent Tech­nologi es Sales and S ervice Office for service and repair to ensure that safety features are maintained.
Ground the Instrument
If your product is provided with a ground­ing-type power plug, the instrument chassis and cover must be connected to an electri­cal ground to minimize shock hazard. The ground pin must be firmly connected to an electrical ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or discon­nection of the protective earth termin al will cause a potential shock hazard that could result in personal injury.
Cleaning
Clean the outside of the instrument with a soft, lint-free, slightly dampened cloth. Do not use detergent or chemical solvents.
Safety Symbols
Earth Ground
Chassis Ground
Risk of electric shock
Refer to manual for addi­tional safety information
Alternating Current
On supply
Off supply
‘In’ position of bi-stable push switch
‘Out’ position of bi-stable push switch
CAT II (300V)
IEC Measurement Category II. Inputs may be connected to mains (up to 300 VAC) under Category II overvoltage condi­tions.
WARNING
Main Power and Test Input Dis­connect: Unplug instrument from wall outlet, remove power cord, and remove all probes from all terminals before servicing. Only qualified, service-trained person­nel should remove the cover from the instrument.
WARNING
Line and Current Protection Fuses: For continued protection against fire, replace the line fuse and the current-protection fuse only with fuses of the specified type and rating.
WARNING
Front/Rear Switch: Do not change the position of the Front/Rear switch on the front panel while signals are present on either the front or rear set of ter­minals. The switch is not intended as an active multiplexer. Switch­ing while high voltages or cur­rents are present may cause instrument damage and lead to the risk of electric shock.
34401A Service Guide iii
WARNING
IEC Measurement Category II. The HI and LO input terminals may be connected to mains in IEC Cate­gory II installations for line volt­ages up to 300 VAC. To avoid the danger of electric shock, do not connect the inputs to mains for line voltages above 300 VAC. See "IEC Measurement Category II Overvoltage Protection" on the following page for further infor­mation.
WARNING
Protection Limits: To avoid instru­ment damage and the risk of elec­tric shock, do not exceed any of the Protection Limits defined in the following section.
Protection Limits
The Agilent 34401A Digital Multimeter pro­vides protection circuitry to prevent damage to the instrument and to protect against the danger of electric shock, provided the Pro­tection Limits are not exceeded. To ensure safe operation of the instrument, do not exceed the Protection Limits shown on the front and rear panel, and defined as follows:
Note: The front-panel terminals are shown above. The rear-panel terminals are identi­cal. The Front/Rear switch selects the ter­minal set to be used. Do not operate this
switch while signals are pr esent on the front or rear terminals. The current-protec-
tion fuse is on the rear panel.
Input Terminal Protection Limits
Protection Limits are defined for the input terminals:
Main Input (HI and LO) Terminals. The HI and LO input terminals are used for voltage, resistance, frequency (period), and diode test measurements. Two Protection Limits are defined for these terminals:
HI to LO Protection Limit. The Protection Limit from HI to LO (Input terminals) is 1000 VDC or 750 VAC, which is also the maximum voltage measurement. This limit can also be expressed as 1000 Vpk maximum.
LO to Ground Protection Limit. The LO input terminal can safely "float" a maxi­mum of 500 Vpk relative to ground.
As is implied by the above limits, the Protec­tion Limit for the HI input terminal is a maxi­mum of 1500 Vpk relative to ground.
Current Input Terminal. The current input ("I") terminal has a Protection Limit of 3A (rms) maximum current flowing from the LO input terminal. Note that the current input terminal will be at approximately the same voltage as the LO terminal.
Note: The current-protection circuitry includes a fuse on the rear panel. To main­tain protection, replace this fuse only with a fuse of the specified type and rating.
Sense Terminal Protection Limits
The HI and LO sense terminals are used only for four-wire resistance and tempera­ture measurements (" tion Limit is 200 Vpk for all of the terminal pairings:
LO sense to LO input
HI sense to LO input
HI sense to LO sense
Note: The 200 Vpk limit on the sense termi­nals is the Protection Limit. Operational voltages in resistance measurements are much lower — less than 10 V in normal operation.
4W"). The Protec-
IEC Measurement Category II Overvoltage Protection
To protect against the danger of electric shock, the Agilent 34401A Digital Multime­ter provides overvoltage protection for line-voltage mains connections meeting both of the following conditions:
The HI and LO input terminals are con­nected to the mains under Measurement Category II conditions, defined below,
The mains are limited to a maximum line voltage of 300 VAC.
and
iv 34401A Service Guide
IEC Measurement Category II includes elec­trical devices connected to mains at an out­let on a branch circuit. Such devices include most small appliances, test equipment, and other devices that plug into a branch outlet or socket. The 34401A may be used to make measurements with the HI and LO inputs connected to mains in such devices, or to the branch outlet itself (up to 300 VAC). However, the 34401A may not be used with its HI and LO inputs connected to mains in permanently installed electrical devices such as the main circuit-breaker panel, sub-panel disconnect boxes, or permanently wired motors. Such devices and circuits are subject to overvoltages that may exceed the protection limits of the 34401A.
Note: Voltages above 300 VAC may be mea­sured only in circuits that are isolated from mains. However, transient overvoltages are also present on circuits that are isolated from mains. The Agilent 34401A are designed to safely withstand occasional transient overvoltages up to 2500 Vpk. Do not use this equipment to measure circuits where transi ent overvoltages could exceed this level.
Additional Notices
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This product complies with the WEEE Direc­tive (2002/96/EC) marking requirement. The affixed product label (see below) indi­cates that you must not discard this electri­cal/electronic product in domestic household waste.
Product Category: With reference to the equipment types in the WEEE directive Annex 1, this product is classified as a "Monitoring and Contr ol instrumentation" product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or see www.agilent.com/environment/product for more information.
Agilent 34138A Test Lead Set
The Agilent 34401A is compatible with the Agilent 34138A Test Lead Set described below.
Test L e a d Ra t i n g s
Test Leads - 1000V, 15A
Fine Tip Probe Attachments - 300V, 3A
Mini Grabber Attachment - 300V, 3A
SMT Grabber Attachments - 300V, 3A
Operation
The Fine Tip, Mini Grabber, and SMT Grab­ber attachments plug onto the probe end of the Test Leads.
Maintenance
If any portion of the Test Lead Set is worn or damaged, do not use. Replace with a new Agilent 34138A Test Lead Set.
WARNING
If the Test Lead Set is used in a manner not specified by Agilent Technologies, the protection pro­vided by the Test Lead Set may be impaired. Also, do not use a dam­aged or worn Test Lead Set. Instrument damage or personal injury may result.
34401A Service Guide v
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name: Manufacturer’s Address:
Agilent Technologies, Incorporated 815 – 14th St. SW Loveland, Colorado 80537 USA
Declares, that the product
Product Name: Model Number:
Multimeter 34401A
Product Options: This declaration covers all options of the above product(s).
Conforms with the following European Directives:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.
Conforms with the following product standards: EMC Standard
IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998 CISPR 11:1990 / EN 55011:1991 IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 IEC 61000-4-3:1995 / EN 61000-4-3:1995 IEC 61000-4-4:1995 / EN 61000-4-4:1995 IEC 61000-4-5:1995 / EN 61000-4-5:1995 IEC 61000-4-6:1996 / EN 61000-4-6:1996 IEC 61000-4-11:1994 / EN 61000-4-11:1994
Canada: ICES-001:1998 Australia/New Zealand: AS/NZS 2064.1
Limit
Group 1 Class A 4kV CD, 8kV AD 3 V/m, 80-1000 MHz
0.5kV signal lines, 1kV power lines
0.5 kV line-line, 1 kV line-ground 3V, 0.15-80 MHz Dips: 30% 10ms; 60% 100ms Interrupt > 95%@5000ms
The product was tested in a typical configuration with Agilent Technologies test systems.
Safety
IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995 Canada: CSA C22.2 No. 1010.1:1992 UL 3111-1: 1994
18 July 2001
Date
Ray Corson
Product Regulations Program Manager
Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Straβe 130, D 71034 Böblingen, Germany
For further information, please contact your local Agilent Technologies sales office, agent or distributor.
Note: Unless otherwise indicated, this manual applies to all serial numbers.
The Agilent Technologies 34401A is a 6
1
-digit, high-performance
2
digital multimeter. Its co m bination o f bench-to p and syste m featur e s makes this multimeter a versa t ile s olu t ion for y ou r meas u re men t need s now and in the future.
Convenient Bench-Top Features
Highly visible vacuum-fluoresc en t disp l ay
Built-in math operations
Continuity and diode test functions
Hands-free, Reading Hold feature
Portable, ruggedized case with non-skid feet
Flexible System Features
GPIB (IEEE-488) interface and RS-232 interface
Standard programming languages: SCPI, Agilent 3478A, and
Fluke 8840
Reading rates up to 1000 readings per second
Storage for up to 512 readings
Warning
Limit testing with pass/fail signals
Optional 34812A IntuiLink/Meter Software
®
for Microsoft
Windows
TM
The procedures in this manual are intended for u se by qualified, service-trained personnel only.
Agilent 34401A Multimeter

The Front Pa nel at a Glance

1 Measurement Function keys 2 Math Operation keys 3 Single Trigger / Autotrigger / Reading Hold key 4 Shift / Local key
2
5 Front / Rear Input Terminal Switch 6 Range / Number of Digits Displayed keys 7 Menu Operation keys
The Front-Panel Men u at a Glance
The menu is organized in a top -down tree structure with three levels.
A: MEASurement MENU
1: AC FILTER > 2: CONTINUITY > 3: INPUT R > 4: RATIO FUNC > 5: RESOLUTION
B: MATH MENU
1: MIN-MAX > 2: NULL VALUE > 3: dB REL > 4: dBm REF R > 5: LI MIT TEST > 6: HIGH LIMIT > 7: LOW LIMIT
C: TRIGger ME N U
1: READ HOLD > 2: TRIG DELAY > 3: N SAMPLES
D: SYStem MENU
1: RDGS STORE > 2: SAVED RDGS > 3: ERROR > 4: TEST > 5: DISPLAY > 6 : BEEP > 7: COMMA > 8 : REVISION
E: Input / Output MENU
1: HP-IB ADDR > 2: INTERFACE > 3: BAUD RATE > 4: PARITY > 5: LANGUAGE
F: CALibration MENU
1: SECURED > [ 1: UNSECURED ] > [ 2: CALIBRATE ] > 3: CAL COUNT > 4: MESSAGE
The commands enclosed in square brackets ( [ ] ) are “hidden” unless the multimeter
*
is UNSECURED for calibration.
*
3
Display Annunciators
Adrs Rmt Man Trig Hold Mem Ratio Math ERROR Rear Shift 4W
Turns on during a measurement. Multimeter is addressed to listen or talk over the GPIB interface. Multimeter is in remote mode (remote interface). Multimeter is using manual ranging (autorange is disabled). Multimeter is waiting for a single trigger or external trigger. Reading Hold is enabled. Turns on when reading me mory is enabled. Multimeter is in dcv:dcv ratio function. A math operation is enabled (null, m in-max, dB, dBm, or limit test). Hardware or remote interface command errors are detected. Rear input terminals are selected. “Shift” key has been pressed. Multimeter is in 4-wire ohms function. Multimeter is in continuity test function. Multimeter is in diode test function.
Press “Shif t” again to tur n off.
To review the display annunciators, hold down the Shift key as you turn on the multimeter.
4
The Rear Panel at a Glance
1 Chassis Ground 2 Power-Line Fuse-Holder Assembly 3 Power-Line Voltage Setting 4 Front and Rear Current Input Fuse
5 Voltmeter Complete Output Terminal 6 External Trigger Input Terminal 7 GPIB (IEEE-488) Interface connector 8 RS-232 interface connector
Use the front-panel Input / Output Menu to:
Select the GPIB or RS-232 interface (see chapter 4 in the “User’s Guide”).
Set the GPIB bus address (see chapter 4 in the “User’s Guide”).
Set the RS-232 baud rate and parity (see chapter 4 in the “User’s Guide”).
5

In This Book

Specifications Chapter 1 lists the multimeter’s specifications and describes how to interpret these specifications.
Quick Start Chapter 2 prepares the multimeter for use and helps you get familiar with a few of its front-panel features.
Menu Tutorial Chapte r 3 introduces you to the front-panel menu and steps you through several si m p l e menu examples.
Calibration Procedures Chapter 4 provides a detailed desc ript ion of the multimeter’s calibrations and adjustments.
Theory of Operation Chapter 5 describes each functional block in the multimeter.
Service Chapter 6 provides guidelines for returning your multimeter to Agilent for servicing, or for servicing it yourself.
Replaceable Parts Chapter 7 contains a detailed parts list of the multimeter.
Backdating Chapter 8 describes the procedures involved with back issues of this manu al .
Schematics Chapter 9 provides the multimeter’s schematics.
If you have questions relating to the operation of the Agilent 34401A, call 1-800-452-4844 in the United States, or contact your nearest Agilent Sal es Of fic e .
If your 34401A fails within one year of purchase, Agilent will repair or replace it free of charge. Call 1-877-444-7278 (“ Agile nt Expr ess”) in th e United States, or c ontact you r neare st Agilen t Sales O ffice .
6

Contents

Chapter 1 Specifications
DC Characteristics 12 AC Characteristics 14 Frequency and Period Characteristics 16 Genera l I nformation 18 Produ ct Di mensions 19 To Calculate Total Me as urem en t Erro r 20 Interpreting Multimeter Specifications 22 Configuring for Highest Accuracy Measurements 25
Chapter 2 Quick Start
To Prepare the Multim eter for Use 29 If the Multimeter Does N ot Tu rn On 30 To Adjust the Carrying Handle 32 To Measure Voltage 33 To Measure Resistance 33 To Measure Current 34 To Measure Frequen cy (or Period) 34 To Test Continui ty 35 To Check Diodes 35 To Select a Range 36 To Set the Resolution 37 To Make Null (Relative) Measurements 38 To Store Minimum and Maximum Readings 39 To Make dB Measur emen t s 40 To Make dBm Measurements 41 To Trigger the Multimeter 42 To Make dcv:dcv Rat io Me as urem en ts 43 Front-Panel Display For mats 44 To Rack Mount the Multimeter 45
Contents
Chapter 3 Menu Tutorial
Front-Panel Menu Reference 49 A Front-Panel Menu Tutorial 51 Menu Examples 53
7
Contents
Contents
Chapter 4 Calibration Procedures Agilent Calibration Services 61
Calibration Interval 61 Time Required for Calibration 61 Automating Calibration Procedures 62 Recommended Test Equipment 63 Test Considerations 64 Performance Ver i ficati o n Test s 65 Zero Offset Ver ificat ion 67 Gain V eri fication 69 Optional AC Per for ma nce V erif icat ion Tes ts 72 Calib ra t i on Security C od e 73 Calibration Count 75 Calibration Mes sage 75 Calibration Procedures 76 Aborting a Calibration in Progress 76 Zero Ad j us t ment 77 Gain Adjustment 79 Optional Gain Calibration Procedures 82 Understanding the AC Signal Filter 85 Understanding Resolution 86 Error Message s 89
Chapter 5 Theory of Operation
Block Diagram 93 Front/Rear Selection 94 Function Switching 95 DC Amplifier 96 Ohms Current Source 98 AC Circuit 99 A-to-D Converter 101 Floating Log ic 103 Earth -R e ferenced Log i c 105 Power Supplies 106 Front Panel 107
8
Contents
Chapter 6 Service
Operating Checklist 111 Types of Service Available 112 Repackaging for Shipment 113 Electrostatic Discharge (ESD) Precautions 114 Surface Mount Repair 114 To Replace the Power-Li n e Fuse 114 To Replace The Current Input Fu s es 115 To Connect Pass/Fail Output Signals 115 Troubleshooting Hints 117 Self-Test Pro ced ur es 120
Chapter 7 Replaceable Parts
To Order Replaceable Parts 126 Backdating and Part Changes 126 Replaceable Parts: 34401-66501 (Main Assembly) 127 Replaceable Parts: 34401-66512 (Display Assembly) 133 Replaceable Parts: Agilent 34401A Mainframe 134 Manufacturer’s List 135
Chapter 8 Backdating 137
Chapter 9 Schematics
Mechanical Disassembly 9-3 Component Locator Diagram – Main Board (34401-66501) 9-5 Component Locator Diagram – Front Panel (34401-66512) 9-6 Agilent 34401A Block Diagram 9-7 Front/Rear Selection Schematic 9-8 Function Switching Schematic 9-9 DC Amplifier and Ohms Schematic 9-10 AC Circuit Schematic 9- 11 A/D Converter Schematic 9-12 Floating Logic Schematic 9-13 Earth -R e ferenced Log i c S ch e m a t i c 9-14 Power Supplies Sche matic 9-15 Front-Panel Display Schematic 9-16
Contents
9
Contents
10
1
1
Specifications
Chapter 1 Specific ations

DC Characteristics

DC Characteristics
Function
DC Voltage 100.000 0 mV
Range [ 3 ]
1.000000 V
10.00000 V
100.000 0 V
1000.00 0 V
Test Current or Burden Voltage
Accuracy Specifications ± ( % of reading + % of range ) [ 1 ]
24 Hour [ 2 ] 23°C
± 1°C
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0006
90 Day 23°C
± 5°C
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0010
1 Year 23°C ± 5°C
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0010
Temperature Coefficient /°C 0°C – 18°C 28°C – 55°C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
Resistance
[ 4 ]
DC Current 10.00000 mA
Continuity 1000.0 Diode Test
[12]
DC:DC Ratio 100 mV
Transfer Accuracy ( typical )
100.0000
1.000000 k
10.00000 k
100.000 0 k
1.000000 M
10.00000 M
100.000 0 M
100.000 0 mA
1.000000 A
3.000000 A
1.00 00 V 1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.001 + 0.002
to 1000 V
( 24 hour %
1 mA 0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 0.001 + 0.002
of range error )
1 mA 1 mA 100 10
µA
5
µA
500 nA 500 nA || 10 M
< 0.1 V < 0.6 V < 1 V < 2 V
2
µA
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.002 + 0.001
0.015 + 0.001
0.300 + 0.010
0.005 + 0.010
0.01 + 0.004
0.05 + 0.006
0.10 + 0.020
( Input Accuracy ) + ( Reference Accuracy )
Input Accuracy = accuracy specification for the HI-LO input signal. Reference Accuracy = accuracy specification for the HI-LO reference input signal.
Conditions:
 Within 10 minutes and ± 0.5°C.  Within  Following a 2-hour warm-up.  Fixed range b etween 10% and 100% of full scale.
 Using 6  Measurements are made using accepted metrology practices.
0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.020 + 0.001
0.800 + 0.010
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.120 + 0.020
±10% of initial value.
1
⁄2 digit s low resoluti on ( 100 PLC ).
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.040 + 0.001
0.800 + 0.010
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.120 + 0.020
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
0.002 + 0.0020
0.002 + 0.0005
0.005 + 0.0010
0.005 + 0.0020
12
Chapter 1 Specific ations
DC Characteristics
1
Measuring Characteristics
DC Voltage
Measurement Method: A/D Linearity:
Input Resistance:
0.1 V, 1 V, 10 V ranges 100 V, 1000 V ranges Input Bias Current: Inpu t Terminals: Input Protection:
Resistance
Measurement Method: Max. Lead Resistance:
(4-wire ohms) Input Protection:
DC Current
Shunt Resistor: Input Protection:
Continuity / Diode Test
Response Time: Continuity Threshold:
DC:DC Ratio Measurement Method: Input HI-LO Refere nc e HI -I np ut LO Input to Refe ren ce
Measurement Noise Rejection 60 Hz ( 50 Hz ) [ 5 ]
DC CMRR Integration Time
100 PLC / 1.67s (2s) 10 PLC / 167 ms (200 ms) 1 PLC / 16.7 ms (20 ms)
0.2 PLC / 3 ms (3 ms)
0.02 PLC / 400
µs (400 µs)
[12] Accuracy specifications ar e for th e voltage measured at the input terminals only. 1mA test current is typical. Variation in the current source will create some variation in the vol tage drop ac ross a diode junction.
Continuously integrating, multi-slope III A/D conv ert e r.
0.0002% of reading + 0.0001% of range Selectable 10 M
or >10 G[11]
10 MΩ ±1% < 30 pA at 25°C Copper alloy 1000 V on all ranges
Selectable 4-wire or 2-wire ohms. Current source referenced to LO input. 10% of range per lead for 100 ranges. 1 k
per lead on all other ranges.
, 1 k
1000 V on all ranges
for 1A, 3A. 5 for 10 mA, 100 mA
0.1 Externally accessible 3A, 250 V fuse Internal 7A, 250 V fuse
300 samples/sec with audible tone Adjustable from 1
to 1000
Input HI-LO / Reference HI-LO 100 mV to 1000 V ranges 100 mV to 10 V ranges (autoranged) Reference LO to Input LO voltage < 2 V Reference HI to Input LO voltage < 12V
140 dB
Normal Mode Rejection
60 dB [ 7 ] 60 dB [ 7 ] 60 dB [ 7 ] 0 dB 0 dB
Teflon is a registered trademark of E.I. duPont deNemours and Co.
[ 6 ]
Operating Characteristics [ 8 ]
Additional
Function
DCV, DCI, and Resistance
System Speeds
Digits
6 6 5 5 4
[ 9 ]
1
2
1
2
1
⁄2
1
2
1
⁄2
Readings/s
0.6 (0.5) 6 (5) 60 (50) 300 1000
Function Change Range C ha ng e Autorange Ti me ASCII readings to RS-232 ASCII readings to GPIB Max. Internal Trigger Rate Max. External Trigger Rate to Memory Max. External Trigger Rate to GPIB
Noise Error
0% of range 0% of range
0.001% of range
0.001% of range
0.01% of range
26/sec 50/sec <30 ms 55/sec 1000/sec 1000/sec 1000/sec 900/sec
[10] [10]
Autozero OFF Operation
Follow ing in st ru men t warm- u p at ca l ib rat i on temp er atu re ±1°C and <10 minutes, add 0.0002% range additional error + 5 µV.
Settling Considerations
Reading settling times are affected by source impedance, cable dielectric characteristi cs, and input signal changes.
Measurement Considerations
Agilent recommends the use of Teflon
â
or other high-impedance,
low-dielectric absorption wire insulation for these measurements.
[ 1 ] Specifications are for 1-hour warm-up at 61⁄2 digits. [ 2 ] Relative to calibration s tandards. [ 3 ] 20% overrange on all ranges, except 1000 Vdc, 3 A range. [ 4 ] Specifications are for 4-w ire ohms function, or 2-wire
ohms using Ma th Null. Without Math Null, add 0.2 additional error in 2-wire ohms function.
[ 5 ] For 1 k unbala nc e in LO lead . [ 6 ] For power-line frequency ± 0.1%. [ 7 ] For power-line frequency ± 1%, subtract 20 dB.
For
[ 8 ] Readings speeds for 60 Hz and ( 50 Hz ) operation,
± 3%, subtract 30 dB.
Autozero Off.
[ 9 ] Speeds are for 41⁄
digits, Dela y 0, A uto ze r o OF F,
2
and Display OFF. Includes measurement and data transfer over GPIB.
[ 10 ] Add 20 µV for dc volts, 4 µA for dc current, or
20 m
[ 11 ] For these ranges, inputs beyond ±17V are
clamped through 100 k
for resistance.
(typical).
13
Chapter 1 Specific ations

AC Characteristics

AC Characteristics
Accuracy Spe cificati ons ± ( % of reading + % of range ) [ 1 ]
Function Range [ 3 ] Frequency
True RMS AC Voltage
[ 4 ]
True RMS AC Current
[ 4 ]
Additional Low Frequency Errors ( % of reading )
100.000 0 mV
1.00 0000 V to
750.000 V
1.000000 A
3.00 000 A
Frequency
10 Hz – 20 Hz 20 Hz – 40 Hz 40 Hz – 100 Hz 100 Hz – 200 Hz 200 Hz – 1 kHz > 1 kHz
3 Hz – 5 Hz 5 Hz – 10 Hz 10 Hz – 20 kHz 20 kHz – 50 kHz 50 kHz – 100 kHz 100 kHz – 300 kHz
3 Hz – 5 Hz 5 Hz – 10 Hz 10 Hz – 20 kHz 20 kHz – 50 kHz 50 kHz – 100 kHz 100 kHz – 300 kHz [6]
3 Hz – 5 Hz 5 Hz – 10 Hz 10 Hz – 5 kHz
3 Hz – 5 Hz 5 Hz – 10 Hz 10 Hz – 5 kHz
AC Filter
Slow Medium Fast
0 0.74 –– 0 0.22 –– 0 0.06 0.73 0 0.01 0.22 0 0 0.18 0 0 0
[6]
[5]
24 Hour [ 2 ] 23°C
± 1°C
1.00 + 0.03
0.35 + 0.03
0.04 + 0.03
0.10 + 0.05
0.55 + 0.08
4.00 + 0.50
1.00 + 0.02
0.35 + 0.02
0.04 + 0.02
0.10 + 0.04
0.55 + 0.08
4.00 + 0.50
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
Additional Crest Factor Errors ( non-sinewave ) [ 7 ]
90 Day 23°C
± 5°C
1.00 + 0.04
0.35 + 0.04
0.05 + 0.04
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
Crest Factor
1 – 2 2 – 3 3 – 4 4 – 5
1 Year 23°C ± 5°C
1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
Erro r ( % of r eadi ng )
0.05%
0.15%
0.30%
0.40%
Temperature Coefficient/°C 0°C – 18°C 28°C – 55°C
0.10 0 + 0.004
0.03 5 + 0.004
0.00 5 + 0.004
0.01 1 + 0.005
0.06 0 + 0.008
0.20 + 0.02
0.10 0 + 0.003
0.03 5 + 0.003
0.00 5 + 0.003
0.01 1 + 0.005
0.06 0 + 0.008
0.20 + 0.02
0.10 0 + 0.006
0.03 5 + 0.006
0.01 5 + 0.006
0.10 0 + 0.006
0.03 5 + 0.006
0.01 5 + 0.006
Sinewave Transfer Accuracy ( typical )
Frequency
10 Hz – 50 kHz 50 kHz – 300 kHz
Error ( % of r ange )
0.002%
0.005%
14
Conditions:
 Sinewave input.  Within 10 minutes and ± 0.5°C.  Within ±10% of initia l volt ag e and ±1% of initial frequency.
 Following a 2-hour warm-u p.  Fixed range between 10% and 100% of full scale ( and <120 V ).  Using 6  Measur ements are made using accepted metrology practices.
1
⁄2 digit resolution.
Chapter 1 Specific ations
AC Characteristics
1
Measuring Characteristics
Measurement Noise Rejection
AC CMRR
True RMS AC Voltage
Measurement Method:
Crest Factor: AC Fil ter Bandwidth: Slow Medium Fast Input Impedance: Input Protection:
True RM S AC Curr ent
Measurement Method:
Shunt Resistor: Burden Voltage:
Input Protection:
Settling Conside ration s
Applying >300 V rms (or >1 A rms) will cause self-heating in signal-c on ditionin g co mp onents. Thes e err o r s are inc lu ded in the instrument specifications. Internal temperature changes due to s elf-heating may cause addit ional error on lower ac voltage ranges. The additional error will be less than 0.02% of reading and will generally dissipate within a few minutes.
[ 8 ]
70 dB
AC-c ou pled Tr ue R MS – measures the ac component of input with up to 400 Vdc of bias on any range. Maximum 5:1 at full scale
3 Hz – 300 kHz 20 Hz – 30 0 kHz 200 Hz – 300 kHz 1 M
Ω ± 2%, in parallel with 100 pF
750 V rms all ranges
Direct coupled to the fuse and shunt. AC-coupled True RMS measurement (measures the ac component only).
0.1
for 1 A and 3 A ranges
1 A range: < 1 V rms 3 A range: < 2 V rms Exter na lly ac ce ss ib l e 3A, 25 0 V f us e Internal 7A, 250 V fuse
Operating Characteristics [ 9 ]
[ 10 ] [ 11 ]
AC Filter
Slow Medium Fast Fast Fast
5/sec <0.8 sec 50/sec 50/sec 50/sec 50/sec 50/sec
1
⁄2 digits,
7
Volt-Hz.
Function
Digits
ACV, ACI 6
System Speeds
1
⁄2
1
6
1
6
⁄2
1
6
1
6
⁄2
[ 11 ] , [ 12 ]
2
2
Reading/s
7 sec/reading 1
1.6 10 50
Functio n or Ra ng e C ha ng e Autora ng e Tim e ASCII readings to RS-232 ASCII readings to GPIB Max. Internal Trigger Rate Max. External Trigger Rate to Memory Max. External Trigger Rate to GPIB/RS-232
[ 1 ] Specifications are for 1-hour warm-up at 6
Slow ac filter, sinewave input.
[ 2 ] Relative to calibration standards. [ 3 ] 20% overrange on all ranges, except 750 Vac, 3 A range. [ 4 ] Specifications are for sinewave input >5% of range.
For inputs from 1% to 5% of range and <50 kHz, add 0.1% of range additional error. For 50 kHz to 100 kHz, add 0.13% of range.
[ 5 ] 750 Vac ran ge limited to 100 kHz or 8x10 [ 6 ] Typically 30% of reading error at 1 MHz. [ 7 ] For frequencies below 100 Hz, slow AC filter specified
for sinewave input only.
[ 8 ] For 1 k unbalance in LO lead. [ 9 ] Maximum reading rates for 0.01% of ac step
addit ional error. Additional settling delay required when input dc level varies.
[ 10 ] For External Trigger or remote operation using default
settling delay ( Delay Auto ).
[ 11 ] Maximum useful limit with default settling delays defeated. [ 12 ] Speeds are for 4
1
⁄2 digits, Delay 0, Display OFF, and
Fast AC filter.
15
Chapter 1 Specific ations

Frequency and Period Characteristics

Frequency and Period Characteristics
Accuracy Specifications ± ( % of reading ) [ 1 ]
Function Range
Frequency, Period [ 4 ]
Additional Low-Frequency Errors ( % of reading )
Transfer Accuracy ( typical )
[ 3 ] Frequency
100 mV to 750 V
Frequency
3 Hz – 5 Hz 5 Hz – 10 Hz 10 Hz – 40 Hz 40 Hz – 100 Hz 100 Hz – 300 Hz 300 Hz – 1 kHz > 1 kHz
0.0005 % of reading Conditions:
3 Hz – 5 Hz 5 Hz – 10 Hz 10 Hz – 40 Hz 40 Hz – 300 kHz
Resolution
1
6
⁄2 51⁄2 41⁄2
0 0.12 0.12 0 0.17 0.17 0 0.2 0.2 0 0.06 0.21 0 0.03 0.21 0 0.01 0.07 0 0 0.02
90 Day
24 Hour 23°C ± 1°C
0.10
0.05
0.03
0.006
[ 4 ]
 Within 10 minutes and ± 0.5°C.  Within ±10% of initial value.  Following a 2-hour warm-up.
 For inputs > 1 kH z and > 100 mV.  Using 6  Measurements are made usi ng accepted metrology practi ces.
[ 2 ]
23°C
± 5°C
0.10
0.05
0.03
0.01
1
⁄2 digit slow resolution ( 1 second gate time ).
1 Year 23°C ± 5°C
0.10
0.05
0.03
0.01
Temperature Coefficient/°C 0°C – 18°C 28°C – 55°C
0.005
0.005
0.001
0.001
16
Chapter 1 Specific ations
Frequency and Period Characteristics
1
Measuring Characteristics
Frequency and Period
Measurement Method:
Voltage Ranges: Gate Time:
Settling Considerations
Errors wil l oc cu r wh en at t emp tin g t o meas ur e the f req ue nc y o r period of an i np ut f ol l owi ng a dc off se t v ol tag e ch an ge . Th e i np ut blocking RC time constant must be allowed to fully settle ( up to 1 sec ) before the most accurate measurements are possible.
Measurement Considerations
All frequency counters ar e susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors.
Reciprocal-counting technique. AC-coupled input using the ac voltage measurement function. 100 mV rms fu ll s ca le to 750 V rm s. Auto or manual rangin g. 10 ms, 100 ms, or 1 sec
Operating Characteristics [ 5 ]
Function
Frequency, Period
System Speeds
Digits
1
6
2
1
5
⁄2
1
4
⁄2
[ 5 ]
Configuration Rates Autora ng e Tim e ASCII readings to RS-232 ASCII readings to GPIB Max. Internal Trigger Rate Max. External Trigger Rate to Memory Max. External Trigger Rate to GPIB/RS-232
[ 1 ] Specifications are for 1- hour warm-up at 6 [ 2 ] Relative to calibration standards. [ 3 ] 20% overrange on all ranges, except 750 Vac range. [ 4 ] Input > 100 mV. For 10 mV to 100 mV inputs,
multiply % of reading error x10.
[ 5 ] Speeds are for 4
1
⁄2 digits, Delay 0, Display OFF,
and Fast AC filter.
Reading/s
1
9.8 80
14/sec <0.6 sec 55/sec 80/sec 80/sec 80/sec 80/sec
1
⁄2 digit s .
17
Chapter 1 Specific ations

Gene r a l Information

General In formation
General Specifications
Power Supply: Power Line Frequency:
Power Consumption: Operating Environment:
Storag e Env ironment: Rack Dimensions (HxWxD): Weight: Safety:
EMI: Vibration and Shoc k: Warranty:
Accessories Included
Test Lead Kit with probes, alligator, and grabber attachments. User’s Guide, Service Guide, test report, and power cord.
100 V / 12 0 V / 220 V / 240 V 45 Hz to 66 Hz and 360 Hz to 440 Hz. Automatically sensed at power-on. 25 VA peak ( 10 W avera ge ) Full accuracy for 0 Full accuracy to 80% R.H. at 40°C
-40°C to 70° C
88.5 mm x 212.6 mm x 348.3 mm
3.6 kg (8 lbs) Designed to CSA 231, UL 1244, IEC 1010-1 (1990) CISPR 11, IEC 801, MIL-461C (data on file) MIL-T-28800E Type III, Class 5 (data on file) 1 year
°C to 55°C
±10%.
Triggering and Memory
Reading
HOLD Sensitivity:
Samples per Trigge r: Trigger Delay: External Trigge r Delay: External Trigge r Jitter: Memory:
Math Functions
Null, Min/Max/Average, dB, dBm, Limit Test (with TTL output). dBm ref e ren ce resi s ta nc es : 50, 75 , 93, 11 0, 12 4, 12 5, 13 5, 15 0, 250, 300, 500, 600, 800, 900, 1000, 1200, or 8000 ohms.
Standard Programming Languages
SCPI (Standa rd C omm ands fo r Pr ogr amm ab le In st ru men t s) Agilent 3478A Language Emulati on Fluke 8840A, Fluke 8842A Language Emulation
Remote Interface
GPIB (IEEE-488.1, IEEE-488.2) and RS-232C
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-00 1
du Canada.
0.01 %, 0.1% , 1%, or 10% of read ing 1 to 50,000 0 to 36 00 sec ( 10 < 1 ms < 500
µs
512 readings
µs step size )
18
N10149
Chapter 1 Specific ations

Product Dimensions

Product Dimensions
Product Dimensions
1
103.8 mm
261.1 mm
212.6 mm
379.4 mm
88.5 mm
348.3 mm
TOP
All dimensions are shown in millimeter s.
19
Chapter 1 Specific ations

To Calculate Total Measurement Error

To Calculate Total Measurement Error
Each specifi ca t i on includes co rrection factors which account fo r err ors present due to operational limitations of the multimeter. This section explains these errors and shows how to a pply th em to you r measuremen ts. Refer to “Interpreting Multimeter Specifications,” starting on page 22, to get a better understanding of the terminology used and to help you interpret the multimeter’s specifications.
The multimeter’s accuracy specifications are expressed in the form: ( % of reading + % of range ). In addition to the reading erro r an d ran g e error, you may need to add additional errors for certain operating conditions. Check the list below to make sure you include all measurement errors for a given function. Also, make sure you apply the conditions as described in the f o otnotes on the specification pages.
If you are operatin g the m ultim e t er outsi de th e 23° C ± 5°C
temperatur e r ange s pec ified, apply an additional temperature coefficient error.
For dc voltage, dc current, and resistance measurements, you may
need to apply an additional reading speed error or autozero OFF error.
For ac voltage and ac current measurements, you may need to apply
an additional low frequency erro r or crest factor error.
Understanding the “ % of reading ” Error The reading error compensate s fo r i naccuracies that result from the functi on and range you select, as well as the input signal level. The reading error varies according to the input level on the selected range. This error is expressed in percent of readin g. The following table shows the reading error applied to the multimeter’s 24-hour dc voltage specification.
Range
10 Vdc 10 Vdc 10 Vdc
Input Level
10 Vdc 1 Vdc
0.1 Vdc
Reading Error (% of reading)
0.0015
0.0015
0.0015
Reading Error Voltage
150 µV
15 µV 1.5 µV
20
Chapter 1 Specific ations
To Calculate Total Measurement Error
Understandi ng th e “ % of rang e ” Error The range error compensates for inaccuracies that result from the function and range you select. The range erro r con trib u tes a cons t ant er r or , expr esse d as a percen t of range, independent of the input signal level. The fol low in g ta ble show s the range error applied to the mul timeter’s 24-hou r dc v oltage speci fication.
1
Rang e
10 Vdc 10 Vdc 10 Vdc
Input Level
10 Vdc 1 Vdc
0.1 Vdc
Range Error (% of range)
0.0004
0.0004
0.0004
Range
Error Voltage
40 µV
40 µV 40 µV

Total Measurement Error To compute the total measurement error, add the reading error and range error. You can then convert the total measurement error to a “percent of input” error or a “ppm (part-per­million) of input” error as shown below.

% of input error =
ppm of input error =
Total Measur em en t E rr or
Input Signal Level
Total Measur em en t E rr or
Input Signal Level
× 100
× 1,000,000
Error Example Assume that a 5 Vdc signal is input to the multimeter on the 10 Vdc range.
Compute the total measurement error using the 90-day accuracy specifications:
± (0.0020% of reading + 0.0005% of range).
Reading Error Range Error Total Error
= 0.0020% = 0.0005%
µV + 50 µV
= 100
× 5 Vdc × 10 Vdc
= 100 µV = 50
µV
=
± 150 µV ± 0.0030% of 5 Vdc
=
± 30 ppm of 5 Vdc
=
21
Chapter 1 Specific ations

Interpreting Multimeter Specifications

Interpreting Multimeter Specifications
This section is provided to give you a better understanding of the terminology used and wil l help you in terpre t the mu ltimete r’s spe cificatio ns.

Number of Digits and Overrange

The “number of dig its” specification is the most fundamental, and sometimes, the most confusing characteristic of a multimeter. The number of digits is equal to the maximum number of “9s” the multimeter can measure or display. This indicates the number of full digits. Most multimeters have the ability to overrange and add a partial
1
or “
” digit.
2
For example, the Agilent 34401A can measure 9.99999 Vdc on the 10 V range. This represents six full digits of resolution. The multimeter can also overrange on the 10 V range and measure up to a maximum of
12.00000 Vdc. This corresponds to a 6 overrange capability.
1
-digit measurement with 20%
2

Sensitivity

Sensitivity is the minimum level that the multimeter can detect for a given measurement. Sensitivity defines the ability of the multimeter to respond to small changes in the input level. For example, suppose you are monitoring a 1 mVdc signal and you want to adjust the level to within measurement w ould re quire a multim eter with a s ensitivit y of at least 1 You could use a 6 You could also use a 4
For ac voltage and ac current measurements, note that the smallest value that can be measured is different from the sensitivity. For the Agilent 34401A, these functions are specified to measure down to 1% of the selected range. For example, the multimeter can measure down to 1 mV on the 100 mV range.
22
±1 µV. To be able to respond to an adjustme nt this smal l, this
1
-digit multimeter if it has a 1 Vdc or smaller range.
2
1
-digit multimeter with a 10 mVdc range.
2
µV.
Chapter 1 Specific ations
Interpreting Multimeter Specifications

Resolution

Resolution is the numeric ratio of the maximum displayed value divided by the minimum dis pla yed valu e on a selec t ed ra nge. Reso lution is often expressed in percent, parts-per-million (ppm), counts, or bits.
For example, a 6
1
-digit multimeter with 20% overrange capability can
2
display a measurement with up to 1,200,000 counts of resolution. This corresponds to about 0.0001% (1 ppm) of full scale, or 21 bits including the sign bit. All four specifications are equivalent.

Accuracy

Accuracy is a measure of the “exactness” to which the multimeter’s measurement uncertainty can be determined relative to the calibration reference used. Absolute accuracy includes the multimeter’s relative accuracy specification plus the known error of the calibration reference relative to national standards (such as the U.S. National Institute of Standards and Technology). To be meaningful, the accuracy specifications must be accompanied with the conditions under which they are valid. These conditions should include temperature, humidity, and time.
1
There is no standar d co nv en ti on am on g mul tim eter man u fac t u re rs for the confidence limits at which specifications are set. The table be lo w shows the probability of non-conformance for each specificatio n with the given assumptions.
Specification Criteria
Mean
± 2 sigma
Mean
± 3 sigma
Mean
± 4 sigma
Probability of Failure
4.5%
0.3%
0.006%
Variations in performance from reading to reading, and instrument to instrument, decrease for increasing number of sigma for a given specification. This means that you c an a chi eve greater act ual m easurement precision for a specific accuracy specification number. The Agilent 34401A is designed and tested to meet performance better than mean
±4 sigma of the published accuracy specifications.
23
Chapter 1 Specific ations
Interpreting Multimeter Specifications

Transfer Accuracy

Transfer accuracy refer s to the error introduced by the multimeter due to noise and short-term drift. This error becomes apparent when comparing two nearly-equal signals for the purpose of “tran sf erring” the known ac cur a c y of one dev ic e to the ot h er .

24-Hour Accuracy

The 24-hour accuracy specification indicates the multimeter’s relative accuracy over its full measurement range for short time intervals and within a stable environment. Short-term accuracy is usually specified for a 24-hour period and for a
±1°C temperatu r e ran g e.

90-Day and 1-Year Accuracy

These long-term accuracy specifications are valid for a 23°C ± 5°C temperature range. These specifications include the initial calibratio n errors plus the multimeter’s long-term drift errors.

Temperature Coefficients

Accuracy is usually specified for a 23°C ± 5°C temperature range. This is a common temperature range for many operating environments. You must add additional temperature coefficient errors to the accuracy
specification if you are operating the multimeter outside a 23°C ± 5°C temperature range (the specification is per °C).
24
Loading...
+ 133 hidden pages