The MN307 (Cat. #2116.23) is the late st in compact AC current pr obes. Designed to meet the most
stringent demands in industry and electrical contracting, it also meets the latest safety and performance
standards. The probe has a measurem ent range up to 12Arms w hich make it the perfect tool for
measurements w ith DMMs, recorders, pow er and har monic meters. The Model MN307 is compatible
with any AC v oltmeter, multimeter, or other v oltage measurement instrum ent that is capable of
displaying 100mV of output per amp of measured current, voltmeter accuracy of 0.75% or better and an
input impedance of 1M.
WARNING
The safety warnings are prov ided to ensure t he safety of personnel and proper operation of the
instrument. Read the instruction completely.
• Use caution on any circuit: potentially high voltages and currents may be present and may pose
a shock hazard.
• Do not use the probe if damaged. Always connect the current probe to the measuring device
before it is connected around the conductor
• Do not use on non-insulated conductor w ith a potential to ground greater than 600V C AT III
pollution 2. Use extreme caution when clamping around bare conductors or bus bars.
• Before each use, inspect the probe; look for cracks in housing or output cable insulation.
• Do not use clamp in wet environment or in locations that hazardous gases exist.
• Do not use the probe anywhere beyond the tactile barrier.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
CAT II: For measurements performed on circuits directly connected to the electrical distribution sy stem.
Examples are measurements on household appliances or portable tools.
CAT III: For measurements performed in the building installation at the distribution lev el such as on
hardwired equipment in fixed installation and circuit breakers.
CAT IV: For measurements performed at the primary el ectrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiv ing your shipment, make sure that the contents are consis tent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
2.56"
(65 mm)
.83"
(21 mm)
MN306 MN307
2.00"
(51 mm)
Tactical
Barrier
(keep hands below
this barrier)
5.47"
(139 mm)
(18.5 mm)
.1.18"
.73"
Black
Red
(30 mm)
ELECTRICAL SPECIFICATIONS
Nominal Range: 10A
Measurement Range: 0.1 to 12A
Transformation Ratio: Voltage output
Output Signal: 100mV/A (1V at 10A)
Accuracy and Phase Shift*:
Accuracy:
0.1 to 0.5A: 2.5% Reading ± 0.1mV
0.5 to 1A: 2.5% Reading
1 to 12A: 1% Reading
Phase Shift:
.01 to 0.1A: Not Specified
0.1 to 1A: 5°
1 to 5A: 3°
5 to 12A: 2.5°
*Reference conditions: 20°C to 26°C , 20 to 75% R H,
external magnetic field <40A/m, 48 to 65H z si ne w ave,
distortion f actor l ess than 1%, no DC component, no
external current carrying conductor, test sample centered.
Load impedance > 1M.
Frequency Range: 40 to 10kHz
Limit Operating Conditions:
20A permanently to 10kHz;
Derating above 10kHz: 20A x 0.1F (in kHz)
Crest Factor:
3 @ 10Arms with an error (due to CF) of 3%
Working Voltage:
600Vrms
Common Mode Voltage:
600Vrms
Influence of Adjacent Conductor:
< 15mA/A at 50Hz
Influence of DC on AC signal:
DC Current < 2A, < 5%
Influence of Conductor Position in Jaw:
0.5% of mV output @ 50/60Hz
Influence of Frequency:
40Hz to 1kHz: 0.5% of mV output
1kHz to 10kHz: 2% of mV output
Influence of Temperature:
150ppm/°K, or 0.15% of mV output per 10°K
Influence of Humidity (10 - 90% RH):
0.2% of mV output per 10°K @ 18° to 28°C
MECHANICAL SPECIFICATIONS
Operating Temperature:
14° to 131°F (-10° to +55°C)
Storage Temperature:
-40° to 158°F (-40° to +70°C)
Operating Relative Humidity:
10° to 35°C 85% RH
(without roll-off above 35°C)
Maximum Cable Diameter:
One Ø 0.78" (20mm), bus bar 20 x 5mm
Case Protection:
IP 40 (IEC 529)
Drop Test:
Test per IEC 68-2-32:
1.0m drop on 38mm of Oak on concrete.
Mechanical Shock:
Test per IEC 68-2-27
Vibration:
Test per IEC 68-2-6
Dimensions:
5.47 x 2.00 x 1.18" (139 x 51 x 30mm)
Weight:
180g (6.5oz)
Polycarbonate Material:
Jaws: Polycarbonate with 10% fiberglass charge,
Red UL 94 V0
Case: Polycarbonate 920A Gray
Opening Operations - Life:
> 50,000
Output:
Double/reinforced insulated 5 ft (1.5m) lead with
safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation between
the primary or secondary and the outer case of the
handle conforms to IEC 1010-2-32.
Dielectric Strength:
IEC 1010-2-32, Category III, Pollution Degree 2
5550V, 50/60 Hz between primary, secondary
and the outer case of the handle
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
E192383
ORDERING INFORMATION
AC Current Probe MN307 ................. Cat #2116.23
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model MN307
Connect the black lead of the current probe to “common” and the red lead to the AC v oltage input on
your DMM or other v oltage-measuring instrument. The AC current probe has an output of 100mV/A.
This means that for 10A AC in a conductor around which the probe is clamped, 1V AC will come out of
the probe leads to y our DMM or instrument. Se lect the range w hich corre sponds to the measured
current. If the current magnitude is unknow n, st art w ith the highest range and w ork dow n until the
appropriate range and resolution are reached. Clamp the probe around the conductor. Take the
reading on the meter and multiply it by 10 to obtain the measured current (e.g, 160mV reading = 160 x
10 = 1,600mA or 1.6A).
For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
When using a current probe w ith a meter, it is im portant to select the r ange that prov ides the best
resolution. Failure to do this may result in measurement errors.
Make sure that probe jaw mating surfaces are fr ee of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
For maintenance use only original factory replacement parts.
To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
To av oid electrical shock and/or damage to the instrument, do not get w ater or other foreign agents
into the probe
Cleaning
To ensure optimum performanc e, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaw s, use v ery fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
LI
MITED WARRANTY
The current probe is w arranted to the owner for a period of one y ear from the dat e of original purchase
against defects in manufacture. This limited w arranty is giv en by AEMC
distributor from whom it was purchased. This w arranty is void if the unit has been tampered w ith, abused
or if the defect is related to service not performed by AEMC
®
Instruments.
®
Instruments, not by the
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html
99-MAN 100158.v8 04/15