Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
certifies that this instrument has been calibrated using
standards and instruments traceable to international
standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be requested at
the time of purchase, or obtained by returning the
instrument to our repair and calibration facility, for a
nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair and
calibration section at www.aemc.com.
No part of this documentation may be reproduced in any form or by any means (including electronic storage and retrieval
or translation into any other language) without prior agreement and written consent from Chauvin Arnoux®, Inc., as
governed by United States and International copyright laws.
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Tel: (800) 945-2362 or (603) 749-6434 • Fax: (603) 742-2346
This documentation is provided “as is,” without warranty of any kind, express, implied, or otherwise. Chauvin Arnoux®,
Inc. has made every reasonable effort to ensure that this documentation is accurate; but does not warrant the accuracy or
completeness of the text, graphics, or other information contained in this documentation. Chauvin Arnoux®, Inc. shall not
be liable for any damages, special, indirect, incidental, or inconsequential; including (but not limited to) physical, emotional
or monetary damages due to lost revenues or lost prots that may result from the use of this documentation, whether or
not the user of the documentation has been advised of the possibility of such damages.
Chauvin Arnoux®, Inc and AEMC® are registered trademarks of AEMC® Instruments.
Page 4
Thank you for purchasing the Megohmmeter Model 6536.
For best results from your instrument:
■ Read these operating instructions carefully
■ Comply with the precautions for use
WARNING, risk of DANGER! The operator must refer to these
instructions whenever this danger symbol appears
WARNING, risk of electric shock. The voltage applied to parts
marked with this symbol may be hazardous
Equipment is protected by double insulation
> 700V
The voltage on the terminals must not exceed 700V
Ground/Earth
Battery
Information or useful tip
Remote test probe
The product is recyclable in accordance with standard ISO14040
This instrument exceeds regulatory requirements with respect to
recycling and reuse
Indicates conformity with European directives, in particular LVD and
EMC
Indicates that, in the European Union, the instrument must undergo
selective disposal in compliance with Directive WEEE 2002/96/EC.
This instrument must not be treated as household waste
Denition of Measurement Categories (CAT)
■ CAT II corresponds to measurements taken on circuits directly connected to
low-voltage installations.
Example: power supply to electro-domestic devices and portable tools.
■ CAT III corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or xed industrial
devices.
■ CAT IV corresponds to measurements taken at the source of low-voltage
installations.
Example: power feeders and protection devices.
Page 5
PRECAUTIONS FOR USE
This instrument is compliant with safety standard IEC 61010-2-030, and the leads
are compliant with IEC 61010-031, for voltages up to 600V in CAT IV or 1000V in
CAT III.
Failure to observe the following safety instructions may result in electric shock,
re, explosion, and damage to the instrument and installation.
■ Carefully read and understand all precautions for use.
■ Be aware of all electrical hazards when using this instrument.
■ Using this instrument other than as specied may compromise its user
protection features.
■ The safety of any system in which this instrument is incorporated is the
responsibility of the integrator of the system.
■ This instrument can be used on CAT IV installations, for voltages
not exceeding 600V
RMS
with respect to ground or 700V
between terminals.
■ Do not use the instrument on networks whose voltage or category
exceeds those specied in this manual.
■ Observe all environmental conditions of use (see § 3).
■ Except for voltage measurements, take no measurements on
electrically “live” systems.
■ Do not use the instrument if it appears damaged, incomplete, or poorly
closed.
■ Before each use, check the condition of the insulation on the leads,
housing, and accessories. Any part on which the insulation is
deteriorated (even partially) must be set aside for repair or scrapping.
■ Using the instrument without its battery compartment cover may result
in electric shock to the user.
■ Before using your instrument, ensure it is completely dry.
■ Use only the leads and accessories supplied. The use of leads (or
other accessories) of a lower voltage rating or category limits the use
of the instrument/leads (or accessories) combination to the lowest
category and service voltage.
■ When handling the leads, test probes, and alligator clips, keep your
ngers behind the physical guards.
■ Before removing the battery compartment cover, ensure all
measurement leads and accessories are disconnected. Replace all
batteries at once. Use alkaline batteries.
■ Use personal protection equipment where appropriate.
■ All troubleshooting and metrological checks must be done by
Upon receiving your megohmmeter product package, ensure the contents are
consistent with the packing list. Notify your distributor of any missing items. If
the equipment appears to be damaged, le a claim immediately with the carrier
and notify your distributor at once, providing a detailed description. Save the
damaged packing container to substantiate your claim.
Ordering Information:
Megohmmeter Model 6536.............................................................. Cat. #2155.56
Megohmmeter Model 6536 ESD Floor Kit ....................................... Cat. #2155.57
Shipping Contents:
Megohmmeter Model 6536
Two color-coded (red/black) 1.5m test leads,
two alligator clips, two grip probes and
one test probe (black)
Also includes 6 AA batteries and a user manual.
4
Soft Carrying Case
ESD Floor Kit
(only with Cat. #2155.57)
Megohmmeter Model 6536
Page 9
1.2 Accessories
Megohmmeter Test Probe (600V CAT IV) ........................................Cat. #2155.75
Case - Field Case (Waterproof - Replacement for ESD Floor Kit) ... Cat. #2118.98
Case - Hands Free Carrying Case ................................................... Cat. #2118.99
Continuity Pole .................................................................................Cat. #2138.54
1.3 Replacement Parts
Lead - Set of 2, Color-coded 5 ft (Red/Black) Silicone Leads, Test Probes &
(Replacement for ESD Floor Kit) ...................................................... Cat. #1017.45
For accessories and replacement parts, visit our store at www.aemc.com.
1.4 Description
The Megohmmeter Model 6536 is a portable measuring instruments with digital
displays, and is powered by batteries. The instrument measures insulation with
test voltages from 10 to 100V in one-volt steps. Other features include:
■ Continuity measurement
■ Resistance measurement
■ Programmable alarms
The Model 6536 is also available as a oor resistance testing kit (see § 2.2.5)
for testing both point-to-ground and point-to-point electrostatic discharge (ESD).
In addition to the instrument, the kit contains two 5 lb (2.3 kg) oor weights
(also referred to as NFPA probes) that are in compliance with EOS/ESD 11.11
specications. Each oor weight is coated with insulating paint and a conductive
rubber base. The instrument, oor weights, and test leads are packaged in a
rugged eld case. The kit provides all the necessary elements for compliance
with the ANSI/ESD STM97.2-2006 standard for testing ESD ooring.
Megohmmeter Model 6536
5
Page 10
1.5 Front of Instrument
VAR
>
<
ALARM
HOLD
REL
G
M mAµA
k
G
G
M
k
P
1
M
G
2
DC
AC
V
3
4
k
HOLD
V
Ω
SET-UP
MEM
TEST
OFF
Ω
CLR
0
REL
MΩ
VAR
5
MEGOHMMETER
MODEL 6536
Figure 1
1. Input terminals
2. Blue backlit LCD
3. Six function buttons (see § 1.8)
4. TEST button to start insulation measurements (see § 2.2.2.1)
5. Five-position rotary switch to choose the function or to turn the instrument
OFF
6
Megohmmeter Model 6536
Page 11
1.6 Back of Instrument
1
2
3
4
5
Figure 2
1. Captive quarter-turn screw
2. Battery compartment cover
3. Mounting magnets, molded into instrument case
4. Non-skid pads
5. Stand
Megohmmeter Model 6536
7
Page 12
1.7 Terminals
0
> 700V
600V CAT IV
Figure 3
The instrument has one positive ( + ) terminal and one negative ( - ) terminal. The
negative terminal also supports the remote probe accessory (see § 2.2.2.3).
1.8 Function Buttons
Several keys have two functions. One is marked on the key, and is enabled via
a short press. The second function is marked under the key, and is enabled by a
long (>2 seconds) press.
BUTTON
HOLD
SET-UP
∆Rel
DESCRIPTION
Selects the and functions (§ 2.2.2).
Toggles backlighting ON and OFF (§ 2.1.5).
Freezes/unfreezes the displayed measurement on the LCD (§
2.1.4).
Accesses the instrument’s setup parameters and information (§
2.1.1).
Applies lead compensation in continuity testing (§ 2.2.3.1).
Activates/deactivates alarms (§ 2.1.2).
The and keys allow you to:
■ Modify the display and program the durations of
insulation measurements (§ 2.2.2.2).
■ Choose the continuity test current (§ 2.2.3).
■ Program the alarm thresholds (§ 2.1.2).
Displays the difference between the present measurement and a
stored reference measurement (§ 2.1.3).
8
Megohmmeter Model 6536
Page 13
1.9 LCD Display
G
M
V
DC
AC
k
<
>
ALARM
VAR
k
M mAµA
G
G
M
HOLD
REL
P
1
2
3
4
Figure 4
1. Logarithmic bar graph displays insulation measurements
2. Secondary display area
3. Main display area
4. Icons/indicators
When the measured value is below the minimum, the instrument displays - - - - .
When measuring voltage, if the reading falls outside the range dened by the
positive and negative limits, the instrument displays OL or – OL.
Megohmmeter Model 6536
9
Page 14
2. OPERATION
SET-UP
Except when measuring voltage, all measurements must be made on
powered-off systems. Therefore check to ensure there is no voltage on
the system under test before making a non-voltage measurement. When
the rotary switch is set to the voltage or an insulation testing position,
the instrument measures and displays any voltage present at the input
terminals prior to the user pressing the test button.
2.1 Setting Up the Instrument
2.1.1 Configuration Settings
A >2 second press of the SET-UP button enables you to
change conguration settings on the instrument. You can
then use the ▲ and ► buttons to scroll through and modify
> 2s
parameters.
In Set-Up mode, the ▲ button performs the following functions:
1st press
on
10
The alarm buzzer is active. To
deactivate it:
1. Press . On will blink to
indicate it is selected.
2. Press to change the
setting to OFF.
3. Press to validate the
change.
The symbol disappears from the
display when you exit Set-Up.
Note that this setting reverts to On
when you turn OFF the instrument.
Megohmmeter Model 6536
Page 15
2nd press
P
on
3rd press
on
Automatic switching to standby
mode is activated. To deactivate it:
1. Press to select OFF (the
setting blinks).
2. Press to change the
setting to On.
3. Press to validate the
change.
The P symbol appears on the
display when you exit Set-Up.
Note that this setting reverts to OFF
when you turn OFF the instrument.
Displays the instrument model
number.
4th press
on
5th press
on
6th press
on
Displays the instrument rmware
version.
Displays the instrument hardware
version.
Return to the rst press.
2.1.2 Alarms
The instrument includes an alarm function that sounds an audible buzzer when a
dened alarm condition is measured.
2.1.2.1 Activating/Deactivating the Alarm Function
The alarm function is available in insulation, resistance, and
continuity measurement modes. Pressing the button
activates the alarm. The
with the threshold value.
TEST
symbol appears on the LCD, along
Megohmmeter Model 6536
11
Page 16
ALARM
To turn OFF the alarm buzzer while it is sounding, press the HOLD button. To
deactivate an active alarm function, press the
<
button.
2.1.2.2 Setting an Alarm Threshold
While is displayed indicating the alarm function is active, you can change
the alarm threshold by pressing the button (except during insulation
measurements). For each testing mode, there are three pre-dened threshold
values:
■ Continuity: < 2Ω, <1Ω and <0.5Ω
■ Resistance: >50kΩ, >100kΩ and >200kΩ
■ Insulation: <10kΩ, <50kΩ and <100kΩ
In each measurement mode, the third threshold can be replaced by a userdened value. To do this:
1. Press the ► button while the threshold value is displayed.
2. The > symbol starts blinking; you can change it to < by pressing the
button. This symbol indicates the direction of the alarm threshold: < for a
low threshold and > for a high threshold.
3. To change the threshold setting, press the ► button to navigate to the
rst digit, and then use the button to change its value.
4. Use the ► and buttons to select and change the other digits in the
threshold value, as well as the units of measurement.
5. When nished setting the threshold, press the ► button to validate the
setting.
12
Megohmmeter Model 6536
Page 17
2.1.2.3 Viewing Alarms
ALARM
<
M
ALARM
k
M
>
CLR
When the measurement is below a low alarm threshold or above a high alarm
threshold, the instrument emits a continuous audible signal and the LCD
indicates the threshold crossed:
When checking continuity, this enables you to determine whether or not the
continuity measurement is less than 2Ω simply by listening, without looking at the
display. You can similarly check insulation quality.
2.1.3 ∆REL Function
For an insulation or resistance measurement, you can
congure the instrument to subtract a reference value from the
measured value and display the difference.
To activate this function, press the ∆REL button during
REL
Megohmmeter Model 6536
an active measurement. This measurement becomes the
reference (Rref) and will be stored and subtracted from
subsequent measurement values (Rmeas). The ∆REL symbol
appears on the LCD while this function is activated.
13
Page 18
M
REL
V
G
If the measured value is less than the stored value, the display becomes
HOLD
negative.
You can display the difference as a percentage of the reference (x
100) by pressing the button until the % sign appears:
%
G
M
REL
For insulation measurements, only the digital display is modied by ∆REL.
The bar graph continues to display the true measured value.
To deactivate the ∆REL function, press the ∆REL button during an active
measurement, or turn the rotary switch to another setting.
2.1.4 HOLD Function
14
Pressing the HOLD button freezes the display of the
measurement. This can be done in all functions except during a
timed measurement ().
To unfreeze the display, press the HOLD button again.
Megohmmeter Model 6536
Page 19
2.1.5 Backlighting
0
Pressing the button turns ON backlighting for the LCD.
To switch it OFF, press the button again. Otherwise,
backlighting goes OFF automatically at the end of one
minute.
2.1.6 Standby Mode
After 5 minutes of operation with no user activity, the instrument automatically
switches to standby mode. To restore normal operation, simply press any button.
The instrument returns to the state it was in prior to entering standby mode, with
no loss of information (value of the last measurement, compensation of the leads,
∆REL, timed mode, alarm, etc.).
Automatic switching to standby mode is disabled during:
■ insulation measurements in mode and in mode.
■ continuity measurements, for as long as measurements are made.
Automatic switching to standby is disabled via the SET-UP button (see § 2.1.1).
2.2 Taking Measurements
2.2.1 Voltage Measurement
To ensure proper and accurate operation of the instrument, we recommend
measuring a known voltage (such as an electrical outlet) before measuring
unknown voltages.
1. Set the switch to V or MΩ VAR.
Megohmmeter Model 6536
15
Page 20
V
k
Ω
OFF
Ω
MΩ
VAR
2. Using the leads, connect the system to be tested to the instrument’s
V
DC
AC
terminals.
G VHz
DC
AC
V
The instrument displays the voltage on the terminals. It detects whether the
voltage is AC or DC.
16
Megohmmeter Model 6536
Page 21
In the MΩ VAR setting, the symbol indicates that the voltage is too high
V
k
Ω
OFF
Ω
MΩ
VAR
(>25V) and that insulation measurements are prohibited.
If the voltage is >15V, continuity and resistance measurements are prohibited.
2.2.2 Insulation Measurement
Insulation measurement results can be affected by the impedances of
additional circuits connected in parallel or by transient currents.
Do not start any measurement while the symbol is displayed.
1. Set the rotary switch to the MΩ VAR position.
2. The instrument displays the congured test voltage. To modify this
setting, press the ► key to enter edit mode (The rst digit of the voltage
setting blinks). You can change this by pressing the ▲ key. Then use the
► and ▲ keys to navigate to and change the remaining digits. Allowable
test voltages are 10 through 100V. When nished, press ► to validate
the change.
3. Use the leads to connect the system to be tested to the instrument’s
terminals. The system under test must be powered down and
discharged. When testing insulation, the typical connection is negative
(black) lead to conductor and positive (red) lead to ground or the outer
insulation of the device under test.
Megohmmeter Model 6536
17
Page 22
DC
V
3. (Optional) Press the button to display the current or the elapsed time.
(You can also do this during the measurement.)
4. Press the TEST button and hold it down until the displayed
measurement is stable. Note that if the instrument detects a voltage
greater than 25V in the system under test, pressing the TEST button
has no effect because the test will be prohibited. (An error screen will
appear.)
5. The resistance measurement is displayed on the LCD’s main display
area and on the bar graph. The secondary display area indicates the test
voltage generated by the instrument.
M
G V
> 1s
TEST
The symbol indicates that the instrument is generating a hazardous
voltage (>70V).
6. At the end of the measurement, release the TEST button. The
instrument stops generating the test voltage and discharges the device
being tested. The symbol is displayed until the voltage on the
system under test has fallen below 70V.
18
M
Megohmmeter Model 6536
Page 23
NOTE: Do not disconnect the leads and do not start any measurement while the
symbol is displayed.
When you release the TEST button, the measurement results remain displayed
until the next measurement, or the HOLD button is pressed, or the instrument is
turned OFF.
2.2.2.1 TEST Button Operation
Pressing the TEST button starts an insulation measurement. In normal mode, the
test voltage is generated for as long as the button is pressed. When the button is
released, the measurement stops.
In mode, press the test button once to start the measurement, then press it a
second time to stop; there is no need to keep the button pressed. However, if you
do not stop the measurement, it will stop automatically after 15 minutes.
In timed test mode () press the TEST button once to start the measurement.
The test will stop automatically at the end of the dened test duration time.
2.2.2.2 Timed Tests
The TIMER button activates timed test mode. This button is active only for
insulation measurements.
1st pressThis locks the TEST button. After you start the
measurement, it continues to run without requiring you to keep the TEST button pressed. The
test will run until you stop it, or when 15 minutes
have passed.
2nd pressThis activates timed test mode. You can set a test
duration between 1 and 39:59 minutes. Use the
and buttons to modify the value displayed.
When the time duration is displayed, press the
button to enter edit mode. When the rst digit blinks, you can change it using the button.
Press to go to the next digit and to change
it. Then press to validate.
3th pressExits timed test mode.
When is activated, pressing the TEST button starts the test. The LCD displays
the measurement, along with a “countdown” timer showing the time remaining in
the test. The test automatically stops when the duration end time is reached and
the result is displayed.
Megohmmeter Model 6536
19
Page 24
DAR
HOLD
TEST
Successive presses on the button display intermediate
values. These include:
■ programmed time
■ voltage at the end of the measurement
■ current at the end of the measurement
Press the TEST button to return to voltage measurement.
2.2.2.3 Remote Control Probe (Optional)
The optional remote control probe is used to trigger the measurement using the
TEST button on the probe. To use this accessory, refer to its separate operating
instructions.
M
M
DAR
When the probe is connected, the symbol is displayed on the
instrument’s LCD.
20
DC
V
Megohmmeter Model 6536
Page 25
2.2.3 Continuity Measurement
0
> 2s
0
Continuity measurement measures a low resistance (<10 or 100Ω depending on
the current) at a high current (200 or 20mA).
A current of 20mA reduces the power consumption of the instrument,
increasing its battery life. However, the standard IEC 61557 requires
200mA current for continuity testing.
If an external voltage >15V is detected in the system under test during the
continuity measurement, the instrument is protected without a fuse. The
continuity measurement is stopped and the instrument reports an error
until the voltage disappears.
2.2.3.1 Lead Compensation
Before checking continuity, you should compensate for the
resistance of the measurement leads. This ensures that the
resistance measurement excludes the resistance in the leads.
To do this, set the rotary switch to Ω. Then short-circuit the
measurement leads and press the
button for >2 seconds.
M mA
The display changes to zero and the symbol is displayed. The resistance
of the leads will be systematically subtracted from all continuity measurements. If
the resistance of the leads is >10Ω, there is no compensation. The compensation
remains in memory until the instrument is turned OFF.
If the leads are changed with no change of compensation, the display may
become negative. The instrument reports that the compensation must be redone
by displaying a blinking
Megohmmeter Model 6536
symbol.
21
Page 26
To remove the compensation of the leads, leave the leads open and press the
0
V
k
Ω
OFF
Ω
MΩ
VAR
button for >2 seconds. The LCD displays the resistance of the leads and
the symbol goes off.
2.2.3.2 Continuity Measuring
1. Set the rotary switch to Ω.
2. Press the ► button to display the measurement current. The
measurement current appears blinking on the LCD. You can change the
current by pressing the ► button.
3. Use the leads to connect the instrument to the system to be tested. The
system to be tested must be powered down.
M mA
R
22
Megohmmeter Model 6536
Page 27
M mA
The instrument displays the
V
k
Ω
OFF
Ω
MΩ
VAR
k
resistance and the current used in
the test.
To obtain a continuity value per
standard IEC 61557:
1. Take a measurement at
200 mA and note its value,
R1.
2. Reverse the leads and
note the value R2.
3. Calculate the mean:
2.2.4 Resistance Measurement
Resistance measurements are made with a low current.
1. Set the rotary switch to kΩ.
2. Connect the system to be tested to the instrument. The device to be
tested must be powered down.
3. The instrument displays the results.
Megohmmeter Model 6536
23
Page 28
2.2.5 Floor Resistance Testing
ANSI/ESD STM97.2-2006 is the standard test method for measuring the
electrostatic voltage on a person in combination with oor materials and footwear
as a system. ESD STM 97.2 can be used to characterize ESD protective oor
materials including oor coverings (tiles, carpets, epoxies, and laminated
structures), mats, paints/coatings, and oor nishes. Electrostatic voltage must
be less than +/-100V.
The Model 6536 can be purchased as part of a kit speci
testing oor resistance in areas where electrostatic discharge is a concern, such
as computer rooms, electronic assembly rooms, and hospitals. In addition to the
instrument, the kit includes two 5 pound (2.3kg) floor weights (also called NFPA
probes) for consistent and repeatable contact with the surface under test. The
floor weights are in compliance with EOS/ESD 11.11 specifications. Each floor
weight is coated with insulating paint and rubber base.
cally designed for
24
Megohmmeter Model 6536
Page 29
The instrument, oor weights, and test leads are packaged in a rugged eld
case. The kit provides all the necessary elements for compliance with the ANSI/
ESD STM97.2-2006 standard for testing ESD ooring. The kit can test the
effectiveness of conductive carpets, mats, tables, chairs, and other items.
The following simplied procedure is an extrapolation from the ANSI/ESD
STM97.2-2006 standard. You may also refer to other sources for alternate
procedures.
2.2.5.1 Setup
■ The oor sample should consist of a section covering 48 x 48” (122 x
122cm) in area.
■ When the sample is to be tested after installation, the test area
dimensions should not exceed a section of oor 20 x 20’ (6.1 x 6.1m).
■ Whenever possible, condition the test area at least 24 hours at 73.4°F
±1.8°F (23°C ±1°C) and 50% ±5% RH and test under the same
conditions.
2.2.5.2 Test Procedure
1. For uninstalled panels: Place the dry specimen on a nonconductive
surface and lightly wipe with a lint-free cloth to remove any foreign
material prior to placing the weights. Place the weights 2” (5.1cm) from
an edge of the sample and 36” (91cm) apart. Apply 100V and take a
reading ve seconds after application of the voltage.
2. For installed panels: Place the weights 36” (91cm) apart and at least
36” (91cm) from any ground connections or grounded object resting
on the oor sample. Apply 100V and take a reading ve seconds after
application of the voltage.
3. Unless otherwise specied, make ve measurements on each oor
section with the weights at different locations for each measurement,
and record the value to two signicant gures (see Figure 5 below).
Megohmmeter Model 6536
25
Page 30
36 "
Table Top
Electrodes
> 700V
600V CAT IV
Figure 5
4. Repeat this procedure with the Model 6536 connected between one
weight and a known electrical ground (Figure 6).
26
Megohmmeter Model 6536
Page 31
Groundable Points
Table Top
Electrode
> 700V
600V CAT IV
Figure 6
For an uninstalled oor sample (step 1) the ground should be part of
the sample panel. For an installed oor sample (step 2) the ground
should be a water pipe or known electrical ground. Swap the leads at
the instrument for each measurement and record the average value
obtained as the value for that measurement.
5. When nished, turn off the Model 6536 and return it to the case.
If the resistance changes with time during a measurement, the value observed
after about 5 seconds should be considered the measured value.
Megohmmeter Model 6536
27
Page 32
2.2.5.3 Measurement Results
The following values may serve as a minimum resistivity level to provide an
appropriate level of protection in manufacturing environments.
■ Resistance from electrode to electrode (Figure 5): Measured value: ≥
1MΩ
■ Resistance from test electrode to a groundable point (Figure 6): Measured value: ≤ 10,000MΩ (10GΩ)
28
Megohmmeter Model 6536
Page 33
3. SPECIFICATIONS
3.1 General Reference Conditions
Quantity of InuenceReference Values
Temperature23° ± 3°C (73.4° ± 5.4°F)
Relative humidity45 to 55% RH
FrequencyDC and 45 to 65Hz
Supply voltage
Electric eld0V/m
Magnetic eld< 40A/m
■ The intrinsic uncertainty is the error specied for the reference
conditions.
■ The operating uncertainty includes the intrinsic uncertainty plus
variations of the quantities of inuence (position, supply voltage,
temperature, etc.) as dened in standard IEC 61557.
In this section, uncertainties are typically expressed as % of the reading (R)
plus number of display counts (ct).
Capacitance
in parallel on
resistance to
be measured
Common
mode
rejection in
AC 50/60 Hz
0 to 5µF at
1mA
0 to 2µF
0 to 1µF
0 to 600VacV50dB40dB
MΩ1% R + 1 ct
VT = 10V and
25V
from 10kΩ to 1
GΩ
VT = 50V and
100V
from 10kΩ to 3
GΩ
VT = 50V
≤ 5GΩ
2% R + 1 ct3% R + 2 ct
6% R + 2 ct10% R + 2 ct
6% R + 2 ct10% R + 2 ct
3.3.3 Resistance and Continuity Measurement
Inuencing
Parameter
Temperature
Relative
humidity
Supply voltage6.6 to 9.6V
50/60Hz
AC voltage
superposed on
the test voltage
(VT)
Common mode
rejection in AC
50/60 Hz
Range of
Inuence
-20 to + 55°C
(-4 to 131°F)
20 to 80% RH
0.5Vacat 200mA
For R ≥ 10Ω:
0.4Vac
Accepts no
perturbations
0 to 600Vac
Quantity
Inuenced
at 200mA2% R/10°C + 2 ct
at 20mA2% R/10°C + 2 ct
R1% R/10°C + 2 ct
at 200mA4% R + 2 ct
at 20mA4% R + 2 ct
R3% R + 2 ct
at 200mA
at 20mA
R
at 20mA
R
at 200mA
at 20mA
R
TypicalMaximum
Inuence
0.1% R + 2 ct
5% R + 10 ct
50dB40dB
34
Megohmmeter Model 6536
Page 39
3.3.4 Intrinsic Uncertainty and Operating Uncertainty
These megohmmeters comply with standard IEC 61557, which requires that the
operating uncertainty (called B) must be less than 30%.
In insulation and continuity measurements:
where:
A = intrinsic uncertainty
E1 = inuence of the reference position ± 90°
E2 = inuence of the supply voltage within the limits indicated by the
manufacturer
E3 = inuence of the temperature between 0 and 35°C (32 and 95°F)
3.4 Power Supply
The instrument is powered by six 1.5V alkaline AA (LR6) batteries.
The voltage range ensuring correct operation is from 6.6V to 9.6V.
Typical life between charges:
Megohmmeter Model 6536
35
Page 40
■ Insulation: 6000 5-second measurements at 100V for R = 100kΩ, at the
rate of one measurement per minute
■ Continuity: 3000 5-second measurements, at the rate of one
measurement per minute
3.5 Environmental Conditions
Indoor use
Range of operation: -20 to +55°C (-4 to 131°F) and 20 to 80% RH
Range of storage (without batteries): -30 to +80°C (-22 to 176°F) and 10 to 90%
RH without condensation
Altitude: <2000m (6562ft)
Degree of pollution: 2
3.6 Mechanical Specifications
Dimensions (L x W x H): 211 x 108 x 60mm (8.31 x 4.25 x 2.36”)
Weight: approximately 850g (1.87lb)
Ingress protection:
■ IP 54 per IEC 60529, not in operation
■ IK 04 per IEC 50102
Drop test: per IEC 61010-1
3.7 Safety Standards
Safety according to: EN 61010-2-30 : 2010
Insulation Class: 2
Pollution Degree: 2
Overvoltage Category: 600V CAT IV
Immunity according to: EN 61326-1 : 2013
Emission according to: EN 61326-1 : 2013
Specications are subject to change without notice.
36
Megohmmeter Model 6536
Page 41
4. MAINTENANCE & TROUBLESHOOTING
G VHz %
Except for the batteries, the instrument contains no parts that can
be replaced by personnel who have not been specially trained and
accredited. Any unauthorized repair or replacement of a part by an
“equivalent” may impair safety.
4.1 Maintenance
4.1.1 Cleaning
Disconnect the unit completely and turn the rotary switch to OFF.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and
dry rapidly with a dry cloth or forced air. Do not use alcohol, solvents, or
hydrocarbons.
Do not use the instrument again until it is completely dry.
4.1.2 Replacing the Batteries
At start-up, the instrument displays the remaining battery life:
If the battery voltage is too low to ensure correct operation of the instrument, a
“low battery” message appears on the LCD and the symbol blinks:
Megohmmeter Model 6536
37
Page 42
This indicates the batteries must be replaced. All batteries must be replaced at
the same time. To do this:
1. Disconnect any attached leads or accessories from the instrument and
turn the rotary switch to OFF.
2. Use a tool or a coin to turn the quarter-turn screw of the battery
compartment cover.
3. Remove the battery compartment cover.
4. Remove the batteries from the compartment.
Do not treat spent batteries as ordinary household waste. Take them to
the appropriate collection facility for recycling.
38
Megohmmeter Model 6536
Page 43
5. Place the new batteries in the compartment, ensuring that each battery’s
G
<
M mAµA
G
polarity is correct.
6. Put the battery compartment cover in place and screw the quarter-turn
screw back in.
4.2 Troubleshooting
4.2.1 Errors
During instrument operation, errors may be displayed on the LCD. The causes
of any errors must be corrected before the instrument can resume normal
operation.
4.2.1.1 Voltage present before an insulation measurement
Before taking an insulation
measurement, the instrument
measures voltage on the system under
test. If it detects voltage in excess
of 25V and you attempt to take a
measurement, the instrument displays
the message shown to the left, and no
measurement is taken.
You must eliminate the voltage to
resume taking the measurement.
4.2.1.2 Range exceeded during an insulation measurement
If during an insulation measurement
the value to be measured exceeds the
measurement range (which depends
on the test voltage), the instrument
reports this condition. For example, the
screen to the left is displayed when the
range is exceeded while measuring in
the 100V range.
Megohmmeter Model 6536
39
Page 44
4.2.1.3 Voltage present during a continuity or resistance measurement
If during a continuity or resistance
measurement the instrument detects
an external voltage in excess of
15V (AC or DC), it interrupts the
measurement and displays the screen
show to the left.
You must eliminate the voltage to
resume the measurement.
4.2.2 Resetting the Instrument
You can reset your instrument at any time. To do this:
1. Press the ▲ and buttons simultaneously.
2. Turn the rotary switch to any setting other than OFF.
3. The instrument reboots.
4.3 Repair and Calibration
To ensure that your instrument meets factory specications, we recommend
that it be scheduled to be sent back to our factory Service Center at one-
year intervals for recalibration, or as required by other standards or internal
procedures.
For instrument repair and calibration: You must contact our Service Center
for a Customer Service Authorization Number (CSA#). This will ensure that
when your instrument arrives, it will be tracked and processed promptly. Please
write the CSA# on the outside of the shipping container. If the instrument is
returned for calibration, we need to know if you want a standard calibration, or
a calibration traceable to N.I.S.T. (Includes calibration certicate plus recorded
calibration data.)
Ship To: Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
(Or contact your authorized distributor.) Costs for repair, standard calibration, and
calibration traceable to N.I.S.T. are available.
NOTE: You must obtain a CSA# before returning any instrument.
40
Megohmmeter Model 6536
Page 45
4.4 Technical and Sales Assistance
If you are experiencing any technical problems, or require any assistance with
the proper operation or application of your instrument, please call, fax, or e-mail
our technical support team:
Phone: (800) 343-1391 • (508) 698-2115
E-mail: techsupport@aemc.com
4.5 Limited Warranty
The Model 6536 is warranted to the owner for a period of two years from the date
of original purchase against defects in manufacture. This limited warranty is
given by AEMC® Instruments, not by the distributor from whom it was purchased.
This warranty is void if the unit has been tampered with or abused, or if the
defect is related to service not performed by AEMC® Instruments.
Full warranty coverage and product registration is available on our website at
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
What AEMC® Instruments will do: If a malfunction occurs within the warranty
period, you may return the instrument to us for repair, provided we have your
warranty registration information on file or a proof of purchase. AEMC®
Instruments will, at its option, repair or replace the faulty material.
4.6 Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone or by
fax from our Service Department (see address below), then return the instrument
along with the signed CSA Form. Please write the CSA# on the outside of the
shipping container. Return the instrument, postage or shipment pre-paid to:
Ship To: Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments