Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer
These safety warnings are provided to ensure the safety of personnel.
Please read and comply with these precautions:
• This instrument is protected from accidental voltages of not more than
50V with respect to earth in measurement CAT IV. The guaranteed level
of protection of this equipment may be compromised if used in a manner
not specied by the manufacturer.
• Safety is the responsibility of the operator.
• All metal objects or wires connected to the electrical system should be
assumed to be lethal until tested. Grounding systems are no exception.
• Never exceed the maximum rated voltage and current, and the mea-
surement category.
• Never exceed the protection limits, and always comply with the condi-
tions and place of use, indicated in the specications.
• Do not use the instrument or its accessories if they appear damaged.
• Use accessories that have overvoltage category and service voltages
greater than or equal to those of the instrument (CAT IV 50V). Use only
accessories that comply with safety standards (IEC 61010-2-031 & 32).
• Wear the appropriate protective gear (insulating boots and gloves).
• Check that no terminal is connected and the switch is set to OFF before
opening the device.
• Use only the charging unit supplied with the instrument to recharge the
battery.
• Troubleshooting and metrological verication procedures must only be
performed by qualied, approved personnel, or the factory.
• NOTE: The potentials on the various rods used for an earth measurement
may be dierent if a nearby electrical installation is defective or certain
weather conditions prevail (thunderstorms). It is up to the operator
to decide whether to continue or postpone measurements in these
situations.
Signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING that the operator must
refer to the user manual for instructions before operating the instrument. In
this manual, the symbol preceding instructions indicates that if the instructions
are not followed, bodily injury, installation/sample and/or product damage may
result.
Risk of electric shock. The voltage at the parts marked with this symbol may be
dangerous.
Ground/Earth symbol
In conformity with WEEE 2002/96/EC
1.2 Definition of Measurement Categories
CAT II: For measurements performed on circuits directly connected to the electrical
distribution system. Examples are measurements on household appliances or
portable tools.
CAT III:
CAT IV: For measurements performed at the primary electrical supply (<1000V) such
For measurements performed in the building installation at the distribution level
such as on hardwired equipment in xed installation and circuit breakers.
as on primary overcurrent protection devices, ripple control units, or meters.
1.3 Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are consistent with the
packing list. Notify your distributor of any missing items. If the equipment appears
to be damaged, le a claim immediately with the carrier and notify your distributor
at once, giving a detailed description of any damage. Save the damaged packing
container to substantiate your claim.
1.4 Ordering Information
Ground Resistance Tester Model 6472 .........................................Cat. #2135.51
Includes carrying bag, 110/240V power adapter with US power cord, optical USB cable,
rechargeable NiMH battery pack, and a USB stick with DataView® software, ground tester
workbook and user manual.
Includes meter, two carrying bags, two 150 ft color-coded leads on spools (red/blue), one 30
ft lead (green), two 5 ft color-coded leads (red/blue), 110/240V power adapter with US power
cord, optical USB cable, two T-shaped auxiliary ground electrodes, set of ve spaded lugs,
one 100 ft tape measure, rechargeable NiMH battery pack, and a USB stick with DataView®
software, ground tester workbook and user manual.
Includes meter, two carrying bags, two 300 ft color-coded leads on spools (red/blue), two 100
ft color-coded leads (hand-tied, green/black), two 5 ft color-coded leads (red/blue), 110/240V
power adapter with US power cord, optical USB cable, four T-shaped auxiliary ground
electrodes, set of ve spaded lugs, one 100 ft tape measure, rechargeable NiMH battery
pack, and a USB stick with DataView® software, ground tester workbook and user manual.
Includes meter, two carrying bags, two 500 ft color-coded leads on spools (red/blue), two
100 ft color-coded leads (hand-tied, green/black), one 30 ft lead (green), two 5 ft colorcoded leads (red/blue), 110/240V power adapter with US power cord, optical USB cable,
four T-shaped auxiliary ground electrodes, set of ve spaded lugs, one 100 ft tape measure,
rechargeable NiMH battery pack, and a USB stick with DataView® software, ground tester
workbook and user manual.
1.4.1 Kits, Accessories and Replacement Parts
Test Kit for 3-Pole (3-Point) Testing (150 ft) ................................. Cat. #2135.35
Includes two 150 ft color-coded leads on spools (red/blue), one 30 ft lead (green), two 5 ft
color-coded leads (red/blue), two T-shaped auxiliary ground electrodes, set of ve spaded
lugs, one 100 ft tape measure and carrying bag.
Test Kit for 4-Pole (4-Point) Testing (300 ft) ................................. Cat. #2135.36
Includes two 300 ft color-coded leads on spools (red/blue), two 100 ft color-coded leads
(hand-tied, green/black), two 5 ft color-coded leads (red/blue), four T-shaped auxiliary ground
electrodes, set of ve spaded lugs, one 100 ft tape measure and carrying bag.
Test Kit for 4-Pole (4-Point) Testing (500 ft) ................................. Cat. #2135.37
Includes two 500 ft color-coded leads on spools (red/blue), two 100 ft color-coded leads
(hand-tied, green/black), one 30 ft lead (green), two 5 ft color-coded leads (red/blue), four
T-shaped auxiliary ground electrodes, set of ve spaded lugs, one 100 ft tape measure and
carrying bag.
Test Kit for 3-Pole (3-Point) Testing
(Supplemental for 4-Pole (4-Point) Testing) .............................................Cat. #2135.38
Includes two 100 ft color-coded leads (hand-tied, green/black), one 30 ft lead (green), two
T-shaped auxiliary ground electrodes, set of ve spaded lugs, one 100 ft tape measure and
The Ground Resistance Tester Model 6472 is a portable measurement instrument
designed to measure:
• Bond/Connection Resistance (2-Pole and 4-Pole Kelvin sensing)
• Ground Resistance (3-Pole or 4-Pole)
• Ground Coupling Resistance
• Selective Ground Resistance
• Soil Resistivity (Wenner or Schlumberger method)
• Ground Potential
• Ground Resistance with 2 clamps (no auxiliary rods)
• Ground Resistance of Pylons with the GroundFlex® Adapter Model 6474
and GroundFlex® sensors allow the system to measure leakage current and
ground resistance of tower legs without disconnecting the overhead ground
conductor.
The Model 6472 measures from 0.01 to 99.99kΩ and is auto-ranging, automatically
seeking out the optimum measurement range, test frequency and test current.
Easy-to-use - Simply connect the leads, select the test mode, press Start and read
the results. Up to 512 test results can be stored in internal memory for later recall
to the display or downloaded to a PC via DataView® software.
The large LCD is easy-to-read and indicates ground electrode resistance, test
voltage, current and frequency as well as individual electrode resistance, battery
status and more.
The Model 6472 is CAT IV rated to 50V and is over voltage protected to more than
250VAC against accidental connection to live circuits. The voltage is also displayed
on screen. In the event of a system fault, the Model 6472 can withstand 250VAC.
Additional features of the Model 6472 include a heavy-duty eld case sealed
against dust and water when closed (the test button is also sealed against the elements); manual and automatic test frequency selection from 41 to 5078Hz; user
selectable 3-Pole Fall of Potential or 4-Pole Soil Resistivity test methods and user
selectable 2-Pole or 4-Pole Bond Resistance test method.
The Model 6472 is powered by 9.6V, 3.5 Ah NiMH rechargeable batteries. An external recharger powered from 120/230V, 50/60Hz is included and provides for
testing while recharging.
The Ground Resistance Tester Model 6472 is rugged, easy-to-use and ideal for
maintenance crews performing numerous tests. It exceeds mechanical and safety
specications for shock, vibration and drop tests per IEC standards. The adjustable test frequency provides for rejection of high levels of interference, allowing
it to be used under dicult conditions such as high stray currents that aect ac-
curacy.
2.2 Key Features
• Ground Resistance testing using the 2 clamp method (no auxiliary rods
needed)
• 2- and 4-Pole Bond Resistance/Continuity Measurement (DC Resistance)
with automatic polarity reversal
• 3-Pole Fall-of-Potential measurement with manual or automatic frequency
selection
• 4-Pole soil resistivity measurement with automatic calculation of Rho (ρ) and
user selection of the Wenner or Schlumberger test method
• 3-Pole earth/ground coupling measurement
• Manual and automatic frequency scan from 41 to 5078Hz for optimum test
accuracy in electrically noisy environments
• Selectable test voltage of 16 or 32V up to 250mA of test current
• Auto-o power management
• Automatic recognition of all electrode connections and their resistance value
• Stores up to 512 complete test results in internal memory
• Optically isolated USB communication
• Remote set up and operation of all measurements using DataView® software
• Automatic report generation including the fall of potential plot
• Rechargeable NiMH batteries from wall charger or vehicle power
• Rugged dustproof and rainproof eld case – IP53 rated in closed position
• Grounding standards IEC 61557 parts 4 and 5 compliant
• Includes DataView® software for data storage, real-time display, analysis,
NOTE: External voltages will be displayed only on the small displays (A2 and
A3) without the view of the main display (A1). This helps to rapidly recognize
that these are measurements of external voltages.
Backlight: The backlight turns ON automatically when entering a function and
turns OFF after the function is complete. Press the DISPLAY button to turn it back
on.
NOTE: In the Set-up mode, pressing the DISPLAY button also selects the
next configurable parameter.
A1
A2
A3
Top main display
Middle small display
Bottom small display
If this symbol blinks during a passive measurement, a voltage of more than
42V is present at the tester’s inputs.
If this symbol blinks during an active measurement, the limits of use have
been exceeded.
If the symbol appears constant during an active measurement, the values
to be measured are unstable.
A blinking symbol > (greater than) or < (less than) indicates that the
measurement range is exceeded.
If both symbols are blinking during a passive measurement, the values of
voltage and/or current are too small and are beyond the limits of use. The
resistance RPASS displayed by the tester is then highly uncertain.
If both symbols are blinking during an active measurement, the values
to be measured are strongly varying (remedy: switch on smoothing of
measured values with SMOOTH function).
Indicates the battery charge condition; the segments represent the energy.
Indicates whether the 3-Pole EARTH measurement or the EARTH COUPLING
measurement has been selected.
Indicates the sensitivity of the GroundFlex®: x1/10, x1, x10.
Display of the number of turns in GroundFlex® sensors (1 to 4).
Number of GroundFlex® sensors connected (1 to 4). If this symbol blinks,
the GroundFlex® may not be connected or the current ISEL measured by
the GroundFlex® sensor is too small and is outside the limits of use.
Object and test number for storage in memory.
Indicates lead compensation for 2-Pole measurement is active
If blinking, indicates the value is ready to be overwritten with a new value
or that a new value will be entered. If more than one value has to be
entered, use the CHANGE buttons to move to the next one.
Indicates the test frequency.
MANUAL
CONFIG
REMOTE
14
SWEEP
AUTO
MR
MEM
Indicates that the secondary function of a button will be used.
Indicates SWEEP mode - Used in 3-Pole and 4-Pole measurements.
Indicates MANUAL mode - the user has to stop the measurement,
otherwise the measurement keeps on running.
Indicates AUTO mode - measurement stops immediately after all results
are available.
If blinking, indicates the measurement had a disturbance when the measurement was started.
Indicates that the blinking value of AUTO/MANUAL mode, frequency,
voltage or DC current direction can be changed.
Memory recall - indicates the displaying of stored results is active
Indicates that there is recorded data in memory. When blinking, it indicates that the results need to be saved by pressing the MEM button.
Indicates the instrument is remotely controlled by the RS-232/USB
interface.
Indicates SMOOTHing of the displayed measurement results.
Indicates that a clamp is (steady) or should be (flashing) connected to ES.
ALARM
Measured value (R, U, I).
Indicates that the measurement is running.
Indicates the alarm is activated. An audible alarm will sound if the value
measured is above the limit defined in the SET-UP mode.
Distances to be entered between earth electrode and supplementary
electrodes or between earth electrodes.
Clamp should be connected to H.
If this clamp symbol blinks, the instrument is refusing measurement
because the clamp is not connected correctly.
Symbolizes the soil resistivity value (Rho): W or S indicates that Wenner
or Schlumberger method is used for the measurement (changed by the
Hz/OPTIONS button).
2.5 Button Functions
Starts a measurement and compensates for lead resistance (§ 5.2.1).
START/STOP
Measurement stops automatically (AUTO) or is stopped by pressing this
button again (MANUAL).
Selects the second function highlighted in yellow under the buttons.
Increases or decreases the value of the flashing parameter displayed or
MEM/MR
DISPLAY/
SMOOTH
Hz/OPTIONS/
DISTANCE
Ground Resistance Tester Model 6472
selects the next parameter. Press the button for a longer time to increase
the change of value speed at a faster rate.
Selects the parameter to be modified or moves the cursor to the right.
MEM: Stores measurements to memory.
MR: Retrieves measurements from memory.
DISPLAY: Displays all parameters of a measurement.
SMOOTH: Smooths the display of measurement providing a more
stable result.
Hz/OPTIONS: Enables configuration of measurement functions.
DISTANCE: Available for soil resistivity and V potential measurements. It
allows the setting of values of distances used for Rho (ρ) calculation.
Measuring method: Digital with a sampling frequency of 4028Hz, low pass, FFT.
The frequency of the strongest spectral component is displayed.
Measuring rate: Display updates approx. 3 times per s.
Measurement Range5 to 450Hz
Resolution1Hz
Operating error± 2Hz
Min. input voltage10mV
Min. current through a clamp0.5mA
Min. signal strength for GroundFlex
®
5mA
3.2.2 Voltage Measurements
Overvoltages up to 75Vrms are displayed as ">65V". Permanent overvoltages
between 70 and 75V at terminals H (Z) and E (X) can cause overheating of the
overvoltage protection varistor.
Voltages of more than 75Vrms lead to error message 31 (excessive external
voltage) or 32 (voltage measurement overrange).
If terminals H (Z) and E (X) are put into contact with the line voltage, the protection
fuse will blow.
(1): The Model 6472 cannot detect whether a current clamp SR182 or MN82 is connected. In the case
of the MN82 clamp, with currents > 10A and frequencies < 100Hz the instrument will not display any
warning messages. It is the operator’s responsibility to observe the limits of use when using the MN82
clamp.
0.00 to
9.99mA
± (15% + 2cts)± (7% + 2cts)± (5% + 2cts)
> 20%± (15% + 2cts)± (7% + 2cts)
10.0 to
99.9mA
(1)
100 to
999mA
50 to 99Hz
1.00 to
9.99A
(1)
10 to
40A
100 to 400Hz
(1)
3.2.4 DC Resistance Measurements
Measuring Method: Voltage/Current measurement (Standard EN 61557 part 4)
Nominal Output Voltage: 16VDC (if resistance < 22Ω the output voltage is
reduced to 10VDC)
Max Output Current: > 200mADC for resistances < 20Ω
Max Overload (permanent): 50Vrms (protection up to 250V is guaranteed)
Max Inductive Load: 2 H
Max Interfering Voltage: 60Vpeak > 10Hz
Time for Auto Range Selecting: approx 5 s
Measuring Time: 8 s with automatic polarity inversion
Measuring Rate: 3 per s in manual mode
Lead Compensation: Possible from 0 to 5Ω
Alarm Setting: “>” or “<” from 1 to 999Ω (max 3 digits)
Measuring Method: Voltage/Current measurement (EN 61557 part 5)
Open circuit voltage: 16 or 32Vrms square wave (if current > 240mA the output
voltage is reduced to 10Vrms)
Test frequency: Selectable from 41 to 5078Hz (see table in § 6.2)
Short circuit current: > 200mAAC
Noise suppression: > 80 dB for frequencies 20% or more above or below the
test frequency
Max. overload: 250Vrms
Max. value for RH & RS: 100kΩ
Measuring time:Short push on START: approx. 7 s for rst value of RE at 128Hz, then 3 measurements per s.
Long push on START: approx. 15 s for rst value of RE at
128Hz, then 3 measurements per s.
The following error indications refer to reference conditions with a test voltage of
32V, test frequency of 128Hz, RH and RS = 1kΩ, no external voltage.
The operating error of AC resistance measurements can be less than that speci-
ed for voltage or current because frequency characteristics of the voltage channel
are matched to those of the current channel.
Measurement of auxiliary electrodes RH, RS, RES, RE
Measurement
Range
Resolution0.1Ω0.1Ω1Ω10Ω100Ω
Operating error± (10% + 2cts)
0.14 to
9.99Ω
10.0 to
99.9Ω
100 to
999Ω
1.00 to
9.99kΩ
10.0 to
99.9kΩ
The START button must be pressed for more than 2 s. For frequencies between
41 and 256Hz, the resistance of the auxiliary earth electrodes is measured at the
test frequency set. At higher frequencies the resistance is measured at 256Hz.
The START button must be pressed for more than 2 s. For frequencies between
41 and 128Hz, the resistance of the auxiliary earth electrodes is measured at the
test frequency set. At higher frequencies the resistance is measured at 128Hz.
3.2.7 Earth/Ground Potential Measurements V Pot
Measuring method: Voltage/Current measurement
Open circuit voltage: 16 or 32Vrms square wave
Test frequency: Selectable from 41 to 5078Hz (see table in § 6.2)
Short circuit current: > 200mAAC
Noise suppression: > 80 dB for frequencies 20% or more above or below the
test frequency
Max. overload: 250Vrms
Max. value for RH & RS: 100kΩ
Measuring time:Short push on START: approx. 7 s for rst value of RE at
128Hz, then 3 measurements per s.
Long push on START: approx. 15 s for rst value of RE at
128Hz, then 3 measurements per s.
Measurement Range 0.00 to 99.99mV 100.0 to 999.9mV 1.000 to 9.999V 10.00 to 65.00V
Resolution0.01mV0.1mV1mV10mV
Intrinsic error U
S-E
± (5% + 1ct)
The intrinsic error specied refers to reference conditions with a test voltage of
32V, test frequency of 128Hz, RH, and RS = 1kΩ, no external voltage.
FunctionPower consumedTypical number of measurements
Device off< 5 mW—
Voltage Measurement1.5 W4500
(2)
mΩ
3-Poles, 4-Pole
V pot.4.9 W1500
2 Clamps3.7 W2000
GroundFlex
(1): Measurements in automatic mode at 25-s intervals - In sweep mode, the number of measurements is
divided by 3 or 4
(2): With R = 1Ω
(3): With RH + RE = 100Ω
(4): With RH + RS-ES = 100Ω
(3)
(4)
ρ
®
4.9 W1500
4.9 W1500
4.9 W1500
5.5 W1200
(1)
3.5 Mechanical
Dimensions: 10.7 x 9.76 x 5.12" (272 x 248 x 130mm)
Weight: 7 lbs (3.2kg) approx
Case Material: ULV0 Polypropylene
Terminals: 4mm recessed banana jacks
Case Protection: EN 60529 - IP53 (cover closed)
Drop Test: Per EN 61010-1
Vibration Test: Per EN 61557-1
3.6 Environmental
Operating Temperature: 32° to 113°F (0° to 45°C); 0 to 90% RH
Specied Operating Temperature
Storage Temperature: -40° to 158°F (-40° to 70°C); 0 to 90% RH
Altitude: < 3000m
(1): This range corresponds to the one defined by standard EN 61557, for which an operating error including the
quantities of influence is defined. When the device is used outside this range, 1.5%/10°C and 1.5% between
75 and 90% RH must be added to the operating error.
NOTE: For detailed information regarding ground resistance testing, see
the Understanding Ground Resistance Testing Workbook CD that was
included with the instrument.
4.1 General Operating Instructions
The Model 6472 has operating modes:
• Automatic mode for routine applications.
• Manual (Expert) mode in which the user can change the parameters of the
measurement functions.
NOTE: Fully charge the battery before the first use (see § 9.4)
4.1.1 Automatic Mode
• Set the switch to the desired function.
• Make the connections appropriate to the function.
• Press the START/STOP button. The device makes the measurement and
stops automatically.
• Read the measurement result on the display and the relevant parameters
using the DISPLAY button.
• To save the information in memory, use the MEM button (see § 7.1).
4.1.2 Manual (Expert) Mode
• Set the switch to the desired function.
• Make the connections appropriate to the function.
• Select MANUAL mode.
• Choose various measurement parameters using the Hz/OPTIONS button.
• Press the START/STOP button. The measurement frequency or the direction of the current (resistance measurement) can be changed during the
measurement. To view their impact upon the measurement and the parameters relevant to the measurement use the DISPLAY button.
• When the measurement results are acceptable, stop the measurement by
pressing the START/STOP button.
• View the result on the display and toggle through the relevant parameters
using the DISPLAY button.
• To record the information into memory, use the MEM button(see § 7).
The instrument also incorporates a “SWEEP” mode (automatic execution of several measurements at several dierent frequencies) It is used to indicate how the
measurement varies as a function of frequency. In this case, the device operates
as in the automatic mode and the measurement process stops after taking the
measurement at the nal frequency.
4.2 Instrument Configuration (SET-UP mode)
To congure the Model 6472 parameters, turn the rotary switch to SET-UP:
• All the segments on the display are activated for 1 second and then the
“PUSH button” message appears on the display.
• Various parameters are accessible by pressing the corresponding button.
• The number or symbol which can be modied ashes. These are changed
using the SELECT button [►]and CHANGE buttons [▲▼].
• To exit SET-UP, turn the rotary switch to another position.
• All of the parameters that are modied are permanently saved until a new
VOLTAGE CHECK: The tester first checks for possible interference. If there
is an external voltage of more than 42V, the warning triangle will be
displayed. If measurements are being performed in AUTO mode and a
disturbance frequency is detected, the tester will automatically look for a
frequency that is different from the default 128Hz.
5.1 Switching the Test Voltage
Available in all functions except mΩ and 2 Clamp.
If necessary, the test voltages can be switched to either 16 or 32V as follows:
• Press the Hz/OPTIONS button, then press it again to make the output volt-
age (Uout) blink.
• Use the ► button to switch to 16 or 32V, then press Hz/OPTIONS again.
This setting is retained when the unit is turned o.
5.2 Resistance Measurement mΩ (2-Wire)
WARNING: Before performing the resistance test, verify that the sample
under test is not energized.
1. Set the switch to
2. Connect the resistance to terminals H (Z) and E (X).
3. The 6472 makes a measurement with a positive current (R+), then reverses the
direction of the current and makes another measurement with negative (R-).
H
ALARM
E
AUTO
±
DC
Ω
Ω
Ω
R =
(R+) + (R-)
2
H
ALARM
E
AUTO
±
DC
Ω
Ω
Ω
4. To display the measurement parameters, press DISPLAY several times.
The device displays the following parameters:
R+, R-, +U
H-E
, +I
H-E
, -U
H-E
, -I
H-E
, U-Act (U
and its frequency) and R∆0 if there
H-E
is compensation for the measurement leads.
5.2.1 Lead Compensation Measurement
Lead compensation subtracts the resistance of the test leads from the measured
result.
• Short the two measurement leads connected to H (Z) and E (X) terminals.
• Press the 2nd button, then the START/STOP button to start the measure-
ment.
• This value will be deducted from all resistance values measured thereafter
until the rotary switch is turned to another function.
H
ALARM
E
appears on the display after the compensation value has been measured.
AUTO
±
DC
AC
Ω
R
0
Ω
NOTE: If the compensation resistance is > 5Ω, or if the leads are not shorted
when the measurement is started, the value of compensation will be canceled.
This function is active only for 2-Pole resistance measurements. By default, the
visual alarm (ALARM symbol ashes) and the audible alarm (buzzer sounds for a
few seconds) are triggered when R < 2Ω.
This threshold can be changed using the SET-UP function:
3. Start the measurement by pressing the START/STOP button.
H
S
ES
E
ALARM
AUTO
±
±
DC
DC
Ω
Ω
Ω
4. To display the measurement parameters, press DISPLAY several times.
The device displays the following parameters:
R+, R-, +U
H-E
, +I
H-E
, -U
H-E
, -I
H-E
, U-Act (U
and its frequency)
H-E
5.4 Earth/Ground Measurement (3-Pole)
This function is used to measure an earth (ground) resistance with 2 auxiliary
electrodes.
There are several measurement methods, the full and simplied (62% method) will
be discussed in this section.
5.4.1 62% Method
1. Set the switch to
2. Place electrodes H (Z) and S (Y) to form a straight line with the earth electrode
under test.
The distance between electrode S (Y) and the earth electrode is 62% of
the distance between electrode H (Z) and the earth electrode; the distance
between the electrodes H (Z) and E (X) should be 8 to 10 times the depth of
the rod you are testing.
Connect the cables to terminals H (Z) and S (Y) (red and blue respectively); dis-
connect the earth strap from the rod under test, then connect terminal E (green) to
the rod or electrode system to be tested.
3.
Press the 2nd button, then the DISTANCE button. The "hundredths" will blink.
H
S
A
AUTO
m
m
Ω
Ω
d
E
=
DISTANCE
=
4. To modify the hundredths (of meters or feet), press the ▲ button until the
desired digit is displayed (0-9).
5. To select and modify the tens, press the ► button and then the ▲ button until
the desired digit is displayed (0-9).
6. To select and modify the units and tenths of a unit, press the ► button and
then the ▲ button until the desired digits are displayed (0.0- 9.9).
7. To terminate the programming of distance, press the 2nd button and then the
DISTANCE button.
It is also necessary to program distance A. This is done in the same way as for
distance d.
8. Start the measurement by pressing the START/STOP button.
EARTH
H
S
A
d
E
=
=
AUTO
m
m
H
S
E
AUTO
Ω
mV
mA
9. To display the measurement parameters, press DISPLAY several times.
To measure the resistances of auxiliary electrodes H (Z) and S (Y), or if the resistance of the electrodes is too high(see § 10.1), start the measurement with a long
press of the START/STOP button.
R
, RS will be displayed.
H
EARTH
H
S
Ω
AUTO
kΩ
Ω
k
> 2s
E
NOTE: Repeat the process at 52% and at 72% for the S electrode.
5.4.2 Full Method
A more thorough measurement can be obtained by taking measurement by mov-
ing the S (Y) auxiliary electrode every 10% between H (Z) and E (X) and plotting
the results. A distinct plateau should result as shown below.
Move electrode S (Y) towards electrode H (Z) by a distance equal to 10% of d and
make another measurement. Then move electrode S (Y) again by a distance equal
to 10% of d, but towards the earth electrode.
H (Z)
S (Y)
52% d
62% d
72% d
d
The 3 measurement results should be the same to within a few percent. If this is
the case, the measurement is valid.
If not, electrode S (Y) is in the zone of inuence of the earth electrode. It will be
necessary to increase distance d for the H (Z) auxiliary electrode and repeat the
measurements.
Positioning of the Auxiliary Electrodes
To make sure that your earth measurements are not distorted by interference, we
recommend repeating the measurement with the auxiliary electrodes placed at a
dierent distance and in another direction (for example, rotated 90° from the rst
alignment).
H (Z)
S (Y)
H (Z)
S (Y)
E (X)
d1
If you nd the same values, your measurement is reliable. If the measured values
are substantially dierent, it is likely that they were inuenced by earth currents or
a groundwater source. It may be useful to drive the electrodes deeper and/or wet
the ground near them to reduce their contact resistance with the soil.
Avoid routing the connecting cables of the earth electrodes near or parallel to other
cables (transmission or supply), metallic conductors, rails, or fences: high test frequencies may cause cross-talk and aect the measurements.
This function is suited to the measurement of very low earth resistances. It
provides better resolution (10x better than 3P measurement) and there is no need
to compensate for the resistance of the measurement leads.
1. Set the switch to .
2. Place electrodes H (Z) and S (Y) at least 30m apart.
S (Y)
> 30 m
> 30 m
In order to avoid electromagnetic interference, it is best to unwind the full length of
each cable from the reel, to keep the cables as far apart as possible on the ground,
taking care not to form loops, and to avoid placing the cables near or parallel to
metallic conductors (cables, rails, fences, etc.). Connect the cables to terminals
H (Z) and S (Y); disconnect the earth strap from the rod under test, then connect
terminals E (X) and ES (Xv) to the earth electrode to be tested.
3. Start the measurement by pressing the START/STOP button.
H
S
ES
E
AUTO
Ω
mV
mA
4. To display the measurement parameters, press DISPLAY several times.
The device displays the following parameters:
RE, U
S-E
, I
H-E
, U-Act (U
and its frequency, U
H-E
and its frequency).
S-E
To measure the resistances of auxiliary electrodes H (Z) and S (Y), or if the resistance of the electrodes is too large(see § 10.1), start the measurement with a long
press of the START/STOP button.
We recommend repeating the measurement with the auxiliary electrodes placed at
a dierent distance and in another direction (see § 5.4.3).
5.5.2 Measurements with a Clamp
For this selective earth measurement you need a current clamp, either a Model
SR182 or Model MN82 (both available as an accessory). These two types of
current clamps are specially designed to work with the 6472 ground tester.
The SR182 is more precise, suited to the measurement of higher currents (up to
40Arms) and for use on thicker conductors, while the MN82 (which is easier to
handle) accepts currents up to 10Arms and can be placed on conductors up to
3/4" (20mm) in diameter.
1. Set the switch to .
2. Place electrodes H (Z) and S (Y) at least 30m (98 ft) apart so that there is no
interference between them.
H (Z)
> 30 m (98 ft)
S (Y)
> 30 m (98 ft)
R
R
R
E3
E1
E2
In order to avoid electromagnetic interference, it is best to unwind the full length of
each cable from the reel, to keep the cables as far apart as possible on the ground,
taking care not to form loops, and to avoid placing the cables near or parallel to
metallic conductors (cables, rails, fences, etc).
3. Connect the cables to terminals H (Z) and S (Y), connect a cable between
terminal E (X) and the earth electrode under test, then connect the clamp to
terminal ES (Xv). The device recognizes it automatically.
4. Clamp to the path of the earth electrode to be checked, then connect a lead
from the current clamp to this same point [connection to terminal ES (XV) ].
Take care not to place the cable of electrode H (Z) too close to the current
clamp in order to avoid any transmission of the AC signal to the clamp (espe-
cially when using an MN82 clamp).
5. Start the measurement by pressing the START/STOP button.
H
S
ES
E
AUTO
Ω
mV
R
= RE1
SEL
mA
6.You can now move the clamp and its lead to measure the other earth resist-
ances, RE2, RE3, etc.
7. To display the measurement parameters, press DISPLAY several times.
The device displays the following parameters:
R
, U
SEL
S-ES
frequency).
, I
H-E
, R-Act (R
), U-Act (U
PASS
and its frequency), I-Act (IES and its
H-E
To measure the resistances of auxiliary electrodes H (Z) and S (Y), or if the resistance of the electrodes is too large (see § 10.1), start the measurement with a long
press of the START/STOP button.
To measure the resistivity of the soil, two measurement methods are possible:
Wenner and Schlumberger
The dierence between the two methods lies in the positioning of the electrodes.
By default, the device selects the Wenner method. The Schlumberger method,
which allows you to move only 2 measurement electrodes rather than 3, is better
suited for measuring soil resistivity at multiple depths.
The soil resistivity measurement with dierent distances d, and therefore in different layers (in depth) of the soil, can be used to establish resistivity proles of
the soil in question, which can be useful for geological analysis, the exploration
of deposits, hydrological studies, etc., and to determine the location of an earth
electrode.
5.6.1 Advantages and Disadvantages of Wenner and Schlumberger
SchlumbergerWenner
AdvantageDisadvantageAdvantageDisadvantage
Need to move the two
potential electrodes ES
(Xv) and S (Y) only for
most readings. This can
significantly decrease the
time required to acquire
a sounding
Because the potential
electrode spacing is
small compared to
the current electrode
soakings, higher meter
sensitivity is required.
Because the potential
electrodes remain in
fixed locations, the effects of the near surface
lateral variations in
resistivity are reduced.
In general, interpretations
based on DC soundings
will be limited to simple,
horizontal; layered
structures
Ground Resistance Tester Model 6472
Potential electrode
spacing increases
as current electrode
spacing increases. Less
sensitive meters may
be used.
All four electrodes, the two
current E (X) and H (Z) and
the two potential ES (Xv)
and S (Y) must be moved
equally to acquire each
reading.
Because all electrodes
are moved for each
reading, this method can
be more susceptible to
near-surface, lateral, and
variations in resistivity.
These near surface lateral
variations could potentially
be misinterpreted in terms
of depth variations in
resistivity.
In general, interpretations
based on DC soundings
will be limited to simple,
horizontal; layered
structures.
5. Start the measurement by pressing the START/STOP button.
W
H
ρ
d
S
d
ES
d
E
ρ
= 2.π.d.R
W
AUTO
S-ES
k Ωm
Ω
m
To display the measurement parameters, press DISPLAY several times.
The device displays the following parameters:
ρW, R
S-ES
, d, U
S-ES
, I
H-E
, U-Act (U
and its frequency, U
S-ES
and its frequency).
H-E
To measure the resistances of the auxiliary electrodes H (Z), S (Y), ES (Xv) and E
(X), or if the resistance of the electrodes is too high (see § 10.1), start the measure-
To measure the resistances of the auxiliary electrodes H (Z), S (Y), ES (Xv),
and E (X), or if the resistance of the electrodes is too high (see § 10.1), start the
measurement by a long press of the START/STOP button.
R
and R
P-E
, then R
P-H
P-ES
and R
will be displayed.
P-S
> 2s
H
ρ
S
d
S
A
ES
d
E
AUTO
Ωm
kΩ
k
Ω
H
ρ
S
d
S
A
ES
d
E
AUTO
Ωm
kΩ
Ω
k
5.7 Potential Measurement (V pot)
The earth potential measurement is similar to the 3-pole earth measurement, but
rather than measuring the resistance, the device measures the potential U
auxiliary electrode S (Y) at dierent distances d from earth electrode E (X). The
relative potential U
S (Y) and the total applied voltage U
is the ratio between the voltage U
SrEL
; it is therefore a number between 0 and 1.
H-E
measured on electrode
S-E
After entering a set of values for dierent distances d, you can determine the varia-
tion of the potential around an earth.
1.
Set the switch to V pot.
2.
Place electrode H (Z) at a distance A from the earth and electrode S (Y) at a
distance d1 from the earth.
SrEL
on
H
S
d1
d2
d3
d4
d5
A
In order to avoid electromagnetic interference, it is best to unwind the full length of
each cable from the reel, to keep the cables as far apart as possible on the ground,
taking care not to form loops, and to avoid placing the cables near or parallel to
metallic conductors (cables, rails, fences, etc.). Connect the cables to terminals H
(Z) and S (Y) and connect terminal E (Z) to the earth.
This is a quick way to measure the value of the earth.
NOTE: The two clamp method typically yields a measurement higher than
the fall of potential method since it uses a higher frequency to perform the
test, resulting in an impedance measurement.
The current clamp connected to terminal H (Z) applies a voltage to the circuit to be
measured. The value of the resulting current is determined by the impedance of
the circuit to be measured. The current owing in the loop is measured using the
clamp connected to terminal ES (Xv). The device then calculates loop resistance
R
from these quantities.
LOOP
NOTE: Use only SR182 or MN82 current clamps, which are specially designed
to operate with the Model 6472.
1.Set the switch to.
2.Connect a clamp to terminal H (Z) and clamp it to a point down stream from
the rod under test that is a serial path to the earth.
3.Connect the other clamp to terminal ES (Xv) and clamp it to the rod or electrode system to be measured.
R
R
E1
R
E2
E3
R
E
Observe the following minimum distances between the two clamps in order to
avoid direct electromagnetic inuences between the transmitting clamp and the
receiving clamp:
Minimum separation
MN82SR182
Ground Resistance Tester Model 6472
46
Value measured
(Ω)
0 to 10.1m (4")0m (0")
1 to 50.4m (1' 04")0.1m (4")
1. Start the measurement by pressing the START/STOP button.
In the case of the diagram below, the earth impedance measured is equal to:
R
= RE + ( RE1 // RE2 // RE3).
LOOP
> 2s
H
ES
AUTO
m A
Hz
Ω
NOTE: In the AUTO mode, the measurement frequency is 1611Hz. To make
an earth measurement that is free of inductive effects, you must change to
MANUAL mode and choose a lower measurement frequency (see §6.2).
There are two user frequencies: one for resistivity and one for earth and soil potential measurements. These two values remain in memory even after the device
is switched o.
For soil resistivity measurements, the user frequency is limited to 128 Hz.
6.3 Manual Settings for mΩ Measurements
Pressing Hz/OPTIONS in MANUAL mode allows the following parameters to be
changed using the ► button:
• Terminal symbols H and E blinkH S ES E blink (4-Pole measurement)
• POS on H and DC+ blinks neg on H and DC- (reversal of polarity at terminal H)
In MANUAL mode the tester does not reverse polarity automatically during the
measurement; however, the polarity can be reversed during the measurement by
The 2-wire mΩ measurement gives a quick measurement result, accompanied by
an audible beep, during a continuity check.
The display is in the sensitivity range (0.5Ω to 1.99kΩ) and the terminal check is
limited to terminal H (Z) (a cable must be connected to it), making it possible to
start the measurement with an open circuit.
For a continuity check, the following settings are mandatory:
• The 2-wire mΩ measurement function must be selected.
• The device must be in manual mode.
• The alarm function must be active (On).
• The alarm threshold must be low (<).
• The buzzer must be activated (bEEP On).
6.4 3-Pole Earth/Ground Measurement & Coupling
6.4.1 3-Pole Earth/Ground Measurements
Pressing on Hz/OPTIONS in MANUAL mode allows the following parameters to
be changed using the ► button:
• EARTH blinks Earth Coupling Measurement
• 128 Hz blinks Test Frequency Measurement
• Test voltage blinks Test Voltage Selection
6.4.2 Earth/Ground Coupling Measurements
This measurement calls for making and storing 3 intermediate measurements (at
the same frequency). It is available only in MANUAL mode.
Press Hz/OPTIONS and use the ► button to switch from EARTH to EARTH COU-
PLING. Proceed as follows:
• If you want to eliminate the resistance of the measurement leads, you can
use lead compensation (2nd + START) before starting the actual coupling
measurement (see § 5.2.1).
• Turn the rotary switch to 3-Pole.
• Select a test frequency and a test voltage (if desired).
• The screen displays EARTH COUPLING 1. Make a 3-pole earth measure-
ment on the rst earth system (measurement of R1 in the connection diagram on previous page).
• Stop the measurement by pressing the START/STOP button. The MEM
symbol ashes to indicate that this result must be recorded in memory.
Press the MEM button twice. To save to another location, refer to §7.
• The screen now displays EARTH COUPLING 2. Make a 3-pole earth mea-
surement on the second earth system (measurement of R2). For this second
measurement, leave the H (Z) and S (Y) electrodes in the same positions as
for the rst measurement.
• Store this result in the same memory location as before by pressing the
MEM button twice.
• The screen now displays the message EARTH COUPLING 3. Disconnect
terminal S (Y) and make a 2-wire resistance measurement with terminal H (Z) connected to earth 1 and terminal E (X) to earth 2.
• Record this result by pressing the MEM button twice.
• EARTH COUPLING 4 and the results of the measurements are displayed.
The calculation is based on the following formulas:
RC = (R1 + R2 – R
C1 = RC/R1 and C2 = RC/R
Rb = R2 - R
C
)/2
1-2
= R1 - RC
2 RA
• To display all of the measurement parameters, press the MR button.
• To scroll through all of the measurements, use the ► and ▲ buttons.
The instrument has a total of 512 memory locations. Each of these locations is
dened by an object number (OBJ) from 01 to 99 and by a TEST number from 01
to 99.
For 3- and 4-pole earth measurements, and for measurements using the Ground-
Flex®, the sweep mode (SWEEP) can record several measurement results at the
same memory location, with the frequency as third addressing criterion.
During soil potential or resistivity measurements (Wenner or Schlumberger methods), several measurement results are recorded at the same memory location,
with the distance between electrodes as the third addressing criterion.
For earth coupling measurements (EARTH COUPLING 1, 2, 3, 4), the four mea-
surements provide the third addressing criterion for the same memory location.
None of the other measurements has an additional addressing criterion, so each
occupies only one memory address.
Since each measurement is dated, you must set the date and time of the device
before any storage in memory(see § 4.2).
7.1 Saving Measurements into Memory
Measurement results can be saved by performing the following:
• Press the MEM button. The tester will automatically suggest the next free
memory location ( message).
• Press the MEM button a second time to save to the OBJ/TEST location.
• If you decide not to save your results, press DISPLAY to exit MEM mode.
• To change the OBJ and TEST location, use the ► and ▲▼ buttons.
NOTE: If a memory location is already occupied, the message will be displayed. Pressing the MEM button will overwrite the measurement record.
When you change to SWEEP mode (§ 6.9), the device automatically activates the
MEM mode. It oers a memory location before starting the measurement. All of
the results obtained will be stored at this location at the end of the measurement.
For soil resistivity and potential measurements, if several measurements are made
with dierent distances d, you can store them under the same OBJ:TEST number,
with the distance as third addressing criterion.
It will be possible to overwrite values already stored with new ones having the
same distance d, or even to add new results having other values for the distance d
provided that all of the other measurement parameters are identical.
7.2 Recalling Measurements from Memory
To recall saved measurements, perform the following:
• Select a measurement function, then press the and MR buttons.
• Use the ► and ▲▼ buttons to choose the OBJ and TEST numbers.
• Press the DISPLAY button to display the time (tiME) and date (dAtE).
• Press DISPLAY again to display the measurement and its parameters.
• Press the MR button again to exit the recall mode.
If there are saved results in SWEEP mode, the frequency will be displayed as a
third address parameter below the OBJ:TEST number selected. Use the ► and
▲▼ buttons to select dierent frequencies.
The SET-UP function lets you read all memory addresses one after the other
independently of which measurement function is chosen.
Any measurement stored in the tester can be retrieved individually using Data-
View® software (see § 8).
7.3 Erasing Measurements from Memory
There are two ways to erase measurements - complete and selective.
7.3.1 Erasing All Measurements
• Set the switch to SET-UP.
• Push once on the MEM button to display the number of free and available
records.
• The display will show “dEL ALL”. Change the blinking NO to YES with the ▲
button, then perform a long press (>2s) on the MEM button.
WARNING: This will delete ALL saved records.
• To exit without erasing, perform a short press on the MEM button.
7.3.2 Erasing Selective Measurements
• Set the switch to SET-UP.
• Press the button.
• Press the MR button. The Object and Test numbers of the last stored test appear, with the Object number blinking.
• If you need to change the Object number, press the ▲ button until the desired
number is displayed. Then press the ► button; the Test number now blinks.
• Press the ▲ button until the desired test is displayed.
• Press the MEM button to display the dEL screen. By default, the blinking word
NO appears on the screen, indicating the test is not to be deleted. To delete
the test, press the ▲ button to change the blinking NO to YES, then press and
hold down (>2s) the MEM button. To exit without erasing, perform a short press
on the MEM button.
Note that if the test stored in this Object/Test number is a sweep test (which con-
sists of multiple measurements taken at dierent frequencies) this procedure only
deletes a single measurement. The remaining sweep measurements will still be
stored in this Object/Test. You must delete all individual measurements to completely remove a sweep test.
When a test is deleted, its Object/Test number is removed from memory. When you
subsequently view stored tests, this number will be skipped. This Object/Test number combination remains unavailable for storing tests until you completely erase
DO NOT CONNECT THE INSTRUMENT TO THE PC BEFORE INSTALLING THE
SOFTWARE AND DRIVERS.
NOTE: When installing, the user must have Administrative access rights
during the installation. The users access rights can be changed after the
installation is complete.
DataView
USB Flash Drive Install
1. Insert the USB stick into an available USB port (wait for driver to be installed).
2. If Autorun is enabled then an AutoPlay window should appear as shown.
®
must be reinstalled for each user in a multi-user system.
NOTE: If Autorun is disabled, it will be necessary to open Windows Explorer,
then locate and open the USB stick drive labeled “DataView” to view the files
on the drive.
3. In the AutoPlay window, select Open Folder to view Files.
4. Double-click on Setup.exe from the opened folder view to launch the Data-
11. In the Select Features window that appears, select the instrument’s control
panel that you want to install, then click Next.
NOTE: The PDF-XChange option must be selected to be able to generate
PDF reports from within DataView®.
Figure 8-2
12. In the Ready to Install the Program window, click on Install.
13. If the instrument selected for installation requires the use of a USB port, a warning box will appear, similar to Figure 8-3. Click OK.
Figure 8-3
NOTE: The installation of the drivers may take a few moments. Windows may
even indicate that it is not responding, however it is running. Please wait for
it to finish.
14. When the drivers are nished installing, the Installation Successful dialog
box will appear. Click on OK.
15.
Next, the Installation Wizard Complete window will appear. Click on Finish.
16.
A Question dialog box appears next. Click Yes to read the procedure for con-
necting the instrument to the USB port on the computer.
NOTE: The Set-up window remains open. You may now select another option
to download (e.g. Adobe® Reader), or close the window.
17. Restart your computer, then connect the instrument to the USB port on the
computer.
18.
Once connected, the Found New Hardware dialog box will appear. Windows
will complete the driver installation process automatically.
The DataView folder containing shortcuts for DataView® and each control
panel selected during installation has
been added to your desktop.
NOTE: If you connected your instrument to the computer before installing
the software and drivers, you may need to use the Add/Remove Hardware
utility to remove the instrument driver before repeating the process.
8.2 Connecting the Model 6472 to your Computer
The Model 6472 is supplied with an optically isolated USB interface cable required
for connecting the instrument to the computer. This cable (Cat. #2135.41) is
equipped with a USB type A on one end, and an optical connector on the other
end.
To connect the Model 6472 to your computer:
1. Connect the optical connector end of the cable to the serial port on the
front panel of the Model 6472 (see Figure 2-1, Item #4).
2. Connect the USB type A end of the cable, to an available USB port on
8.3.1 Establishing Communication to the Instrument
If the instrument is turned ON at the time the Ground Tester Control Panel is
opened, a communication link will happen automatically.
If the instrument is turned OFF at the time the Control Panel is opened, establish a
communication link by going to Instrument > Connect in the main menu.
•The Communication dialog box will appear. Make sure that the communication port displayed in the dialog box matches the port that the serial cable is
plugged into.
Figure 8-5
•Once the proper communication parameters have been specied, click OK.
NOTE: The connection status can be seen in the lower right-hand corner of
the Control Panel and in the Instrument window under status.
8.4 Configuring the Instrument
To congure the instrument, perform the following steps.
Double-click the Ground Tester icon in the DataView folder on your
1.
desktop.
2.
Go to Instrument >Congure from the main menu of the Control Panel.
NOTE: If a connection has not been previously established, the Communication dialog box will appear (see Fig. 8-5).
3.
Once identication is complete, a Conguration dialog window will appear
(if not already open).
The descriptions of the features below appear on several Conguration dialog
boxes, depending on the type of test selected.
• Load from File: Retrieves stored congurations to be used in programming
the instrument.
• Save to File: Saves the current conguration. This le will reside on the
computer’s disk drive. Saving dierent conguration setups can be useful
When 4-Pole Test Resistivity is selected from the tool bar, you can select:
• With or without Auxiliary Rod Resistance.
• Test voltage 16 or 32VAC.
• Auto frequency - selects the optimum frequency (from 41 to 128Hz), usually
128Hz, or select manually from the drop-down frequency list.
• Test Method - Wenner or Schlumberger.
• Distances - the distance “d” (for Wenner test method) or “d” and “a” (for Sch-
lumberger test method), between the test electrodes, needs to be entered
for resistivity calculation to read out directly in Rho.
Selecting a sweep frequency test from the drop-down list will bring you to the
sweep table selection.
The Input Noise option allows the viewing of noise for Z or Y inputs.
• Set number display as scientic (1000 as "1E+3") or letters (1000 as "1k").
• Set Label Terminals to be "X", "Xv", "Y", "Z", or "E", "ES", "S", H".
• Set Measurement units to meter or feet
• Set Rho units to Ohm-meter or Ohm-centimeter
8.5.2 Tools > Colors
Sets colors for measurement channels.
8.5.3 Tools > Recover Data
Every time the directory of recorded data is displayed by the Ground Tester Control
Panel, a backup copy of all the data is saved to a le. Use this option to recover
previous data.
8.6 Starting a Test
Press the START button to begin a test. Some tests congurations are automatic
and end when the data is stable. Others, generally when there is a manual fre-
quency, show a "Stop" button, which should be pressed when the data is available
on the screen and has stable values..
8.7 Downloading Data to Database
Select Recorded Data from the View menu to download the recorded data.
1. Select the data you want to download by clicking on the le name (see
Figure 8-14).
2. Select Create DataView Report. DataView® will be open with the newly
created report.
3. Alternatively, select Create Spreadsheet. Type a name for the download-
ed le. In the Save as type drop-down menu, select either .xls (saves it
as an Excel spreadsheet le), or .csv (saves it as a comma delimited le),
then click Save.
4. Alternatively, select View. After the download is complete, a window will
appear with a graph of the data, and some viewing or channel options. In
that window, you can select Create DataView Report, Create Spreadsheet, or Print.
The Session Properties dialog allows you to specify Operator, Site and Custom
parameters that are to be saved with recorded data. These parameters are used
when generating reports.
The Operator and Site tabs allow you to maintain lists of operators and sites,
saving you time when specifying parameters for reports.
On the left of the Operator and Site tabs is the list of previously dened Operators
and Sites. On the right of the Operator and Site tabs is the individual parameters
that will be saved in an associated database. Only a single set of operator and site
elds are saved in the recording database.
The Custom tab contains a list of user dened parameters. Along side each user
dened parameter is a check box. Items that are checked will be added to an
associated database. Only a single set of Custom parameters can be maintained
(unlike the Operator and Site lists). The Custom tab allows you to specify any user
dened parameters (in addition to the comments eld of the Site tab) that are to be
used in displaying a report.
NOTE: You can also set defaults for session properties within the Ground
Tester Control Panel by selecting File > Edit Session Properties.
8.8 Exporting the Report to a PDF File
• Select the data by clicking on the le name under Recorded Data from the in-
strument in the Instrument window on the left-hand side of the Control Panel.
• Click the Print button.
• In the Print dialog box that appears, select PDF-XChange 3.0 from the drop-
down menu, then click OK.
• When the PDF is complete, the Save As dialog box will appear.
• Choose a location to save the le and click Save.
This completes the process of conguring, running, retrieving data, and printing a
report using DataView® with your Model 6472.
In addition to the pre-designed report templates, DataView® allows you to
totally configure reports to your needs. Refer to the DataView® HELP file on
Please make sure that you have read and fully understand the WARNING
section on page 3.
• To avoid electrical shock, do not attempt to perform any servicing unless
you are qualied to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water
or other foreign agents into the case.
• Turn the instrument OFF and disconnect the unit from all circuits before
opening the case.
• Use specied spare parts only.
9.2 Cleaning
Disconnect the instrument from any source of electricity.
Use a soft cloth lightly moistened with soapy water. Wipe with a moist cloth and
then dry with a dry cloth. Never use alcohol, solvents or hydrocarbons.
9.3 Fuse Replacement
The unit is protected from overloads by two identical fuses:
• Fuse for terminal H (Z):
If this fuse is defective, the unit will no longer produce an output voltage, making it
impossible to perform resistance measurements.
To test this fuse, turn the switch to the mΩ position (2-Wire), connect measurement
leads to terminals H (Z) and E (X) and start a measurement. If the unit refuses to
operate and the symbol for terminal H (Z) blinks, the fuse must be replaced.
If this fuse is defective, the unit will no longer recognize a clamp that is connected
to terminal ES (Xv), making it impossible to perform a 4-Pole earth measurement
with a clamp or an earth measurement with 2 clamps.
To test this fuse, select the 4-Pole measurement function and connect a test clamp
to terminal ES (Xv). If a clamp symbol does not appear next to the symbol for terminal ES (Xv) in the display, the fuse must be replaced.
NOTE: For safety reasons this fuse must always be replaced by an identical
model: 0.63 A F 250V 5x20mm 1.5kA (Cat. #2135.81)
Procedure for replacing fuses:
1. Disconnect the unit completely, turn the rotary switch to OFF, and close
the lid.
2. Loosen the four permanent screws at the bottom of the unit, but do not
remove them.
3. Open the lid and carefully turn the housing upside-down, making sure the
front panel does not fall out. Then carefully pull the front panel and the unit
all the way out of the housing.
4. The fuse for terminal H (Z) is accessible at the back of the unit (in the
corner near the terminal for the battery charger).
5. If only the fuse for terminal H (Z) needs to be replaced, jump ahead to step
13. To replace the fuse for the clamp terminal ES (Xv), continue on.
6. Loosen the two screws on the lid of the battery case and remove the lid.
7. Pull the battery a short way out of the case without overstretching the con-
nection leads, and loosen the two screws at the bottom of the case. Then
put the battery back into the case.
8. Making sure the battery leads are not overstretched and that the battery
does not fall out, carefully lift o the back of the unit, turn it over and place
it next to the front panel with the electronic components.
9. The fuse for the clamp terminal ES (Xv) is now accessible on the printed
circuit board [in the corner near terminal E (X)]. When replacing the fuse,
avoid touching the circuitry or components.
10. Return the back of the unit to the front panel with the components, taking
care not to stretch the battery leads. Carefully lower the back of the unit
and make sure that it is aligned correctly (the four cylindrical holes on the
back must slip over the four mounting pins on the front panel). Also make
sure you do not pinch the battery leads or other wires or components.
11. Pull the battery a short way out of the case without overstretching the
connection leads, and retighten the two screws at the bottom of the case.
Then put the battery back into the case.
12. Put the lid back onto the battery case and screw it shut.
13. Wipe o any dirt on the seal and housing rim.
14. Put the unit back into the housing and screw it into place.
9.4 Recharging the Battery
NOTE: To make the best use of the battery and prolong its service life, observe the following rules.
• Use only the battery charger supplied with the unit; other chargers may be
hazardous!
• Recharge the battery only at temperatures between 32° and 104°F (0° and
+40°C).
• Observe the conditions of use and storage dened in the specications.
Even an NiMH battery can be recharged only a limited number of times. This number, and thus the service life of the battery, is highly dependent on the following
factors:
• the operating conditions
• the recharging conditions
As the earth tester might go some time without use and the battery has a natural
tendency to run down, you should check the battery level at regular intervals. To do
this, turn on the unit and check the battery level indicator at the top of the
display. If there is only one bar in the symbol or if there are no bars at all, the battery should be recharged.
If the unit is not used for a long time, the battery might run down all the way. In
this case, recharging can take several hours. In addition, the unit might not work
properly at the start of recharging.
The capacity and lifetime of the battery may also be diminished. After about ve
charging/discharging cycles the battery will recover its initial capacity.
To recharge the battery, plug the charging unit into the connector at the upper left
(connector in Fig. 2-1) and connect the charging unit to a 100 to 240VAC power
source (suitable for Europe and the U.S.). The frame of the battery symbol will
blink during charging. Charging goes faster when the tester is turned o. The battery voltage is shown on the large display next to Ubatt.
At the middle and bottom of the display you will see one of the following messages:
bAtt CHrGFast charging is active (normal state)
bAtt LOWBattery voltage is too low for a fast charge - charging at a lower current
bAttBattery voltage is too high for a fast charge - charging at a lower current
bAtt HOt
bAtt COLdBattery is too cold for a fast charge (< 0°C) - charging at a lower current
bAtt FULLBattery is full - switch to trickle charging
The 6472 tester can also be recharged from a 12VDC car outlet with a special
charging unit.
Battery is too warm for a fast charge (>40°C) - charging at a lower current
NOTE: In this case the low-potential of the vehicle’s 12V outlet is at the potential of terminals E (X) and ES (Xv) of the ground tester. For safety reasons,
do not connect or use the unit if there is a possibility that voltages at E (X) or
ES (Xv) could exceed 32V.
9.5 Replacing the Battery
The battery in this unit is a special product with customized safety features and
protective devices. It should be replaced only by the same model. If a dierent
model is used, there is a danger of re or explosion, leading to possible damage
or injury.
NOTE: For safety reasons, replace the battery with an identical model:
NiMH Custom Pack 9.6 V / 3.5 Ah (Cat. #2960.21)
Procedure for replacing the battery:
1. Disconnect the unit completely, turn the rotary switch to OFF, and close
the lid.
2. Loosen the four permanent screws at the bottom of the unit, but do not
remove them.
3. Open the lid and carefully turn the housing upside-down, making sure the
front panel does not fall out. Then carefully pull the front panel and the unit
all the way out of the housing.
4. Loosen the two screws on the lid of the battery case and remove the lid.
5. Pull the battery a short way out of the case without overstretching the con-
nection leads, and loosen the two screws at the bottom of the case. Then
put the battery back into the case.
6. Making sure the battery leads are not overstretched and that the battery
does not fall out, carefully lift o the back of the unit, turn it over and
place it next to the front panel with the electronic components.
7. Press the clip o the plug, and pull out the plug with the four battery
leads (in the corner, near the terminal for the optical interface). Avoid
touching the circuitry or components.
8. Remove the old battery from the case and put in a new one. Run the
leads with the plug through the largest slot in the battery case.
9. Insert the battery plug (in the corner, near the terminal for the optical
interface). The two pins should point towards the clip. Avoid touching the
circuitry or components.
10. Return the back of the unit (with the new battery in the case) to the front
panel with the components, taking care not to stretch the battery leads.
Carefully lower the back of the unit and make sure that it is aligned
correctly (the four cylindrical holes on the back must slip over the four
mounting pins on the front panel). Also make sure you do not pinch the
battery leads or other wires or components.
11. Pull the battery a short way out of the case without overstretching the
connection leads, and retighten the two screws at the bottom of the
case. Then put the battery back into the case.
12. Put the lid back onto the battery case and screw it shut.
13. Wipe o any dirt on the seal and housing rim.
14. Put the unit back into the housing and screw it into place.
15. Fully charge the new battery in the unit before using it.
16. Reset the time and date.
NOTE: If the instrument is not used for long periods of time (more than 6
months), it is recommended, in order to recover the initial capacity of the
battery, to make several charge-discharge cycles (3 to 5 times). Discharge
cycle (15 H) can be made with the instrument in MANUAL mode, DC 2-Pole
(2-Pole) resistance measurement and a short-circuit between H (Z) and E (X)
plugs.
This can happen in a 3- or 4-pole earth measurement, a resistivity measurement,
or an earth potential measurement.
This message is displayed when the measurement is triggered by a short press of
the START/STOP button and the resistances of the auxiliary electrodes are too high.
The measurement must then be started by a long press on the START/STOP but-
ton. The device then measures the values of the electrodes and compensates for
them to display the correct result.
10.1.2 Out of Range
>
<
or
<
and
The ashing > or < symbol indicates the measurement is
out of range.
If both symbols ash together, it means that the accuracy
of the value displayed is outside of the instrument
>
specication or is subject to large uctuations. Activating
the smoothing function (SMOOTH) may correct this.
Press the 2nd button, then the SMOOTH button to activate
smoothing.
Conditions that may give uncertain results are indicated on the display of the unit
NOISE
as follows:
FrequencyFunctionTriggering threshold
f > 513Hz
f ≤ 513Hz
f > 513Hz
3P, 4P, V pot.I
4Psel, GroundFlex
®
3P, 4P, ρ, V potI
4Psel, GroundFlex
®
All (except ρ and 2
clamps)
< 6 mA
H-E
I
’ < 6 mA
H-E
< 1 mA
H-E
I
’ < 1 mA
H-E
RS > 5 kΩ
(1)
(1)
f ≤ 513HzAllRS > 30 kΩ
4P selIES < 1 mA
GroundFlex
®
I
< 10 mA
SEL
Values measured (U,
All
I, R) unstable, varying
by more than 5% about
their mean values.
IES < 3 mA
R
PASS
I
SEL
U
S-ES
< 30 mA
< 10 mV
(2)
Indication on the
display unit
(3)
ashes
H
ashes
(3)
ashes
S
ashes
ashes
ashes
ashes
ashes
(3)
steady
ashes
ashes
(3)
(3)
R
PASS
AllU
S-ES
I
U
, U
< 3 mA
SEL
< 1 mV
S-ES
, U
S-E
> 42 V
H-E
-.--- (undened)
(3)
ashes
Stray voltage of which
IES < 0,3 mA
All
the frequency and/or
value is likely to inter-
(4)
fere with the measure-
ment.
(1) I
’: current I
H-E
(2) Not active if the SMOOTH function is selected.
(3) The symbol may also appear if there is an external voltage > 42 V on the terminals of the device.
(4) You should change to manual mode and modify the measurement voltage and/or frequency to make
a valid measurement (when the NOISE symbol is no longer illuminated).
When started up, the Model 6472 device automatically performs a self-test. If a
fault appears during this self-test or during a measurement, the device displays a
message in the form Err XX.
There are 3 categories of errors:
• Harmless Errors 6, 7, 11 and 17
The message appears for approximately 1 second to inform
the user. Consider a repair if the error recurs.
• Errors 6 and 7 are always preceded by an automatic reset.
• During Error 11, the tester automatically resets to the factory default settings.
• Correctable Errors 5, 9, 14, 15, 18, 19, 30, 31, 32 and 33
These errors concern the current measurement function and
disappear if the function is changed. The device can therefore
be used, but a repair will be necessary if the error persists.
• Error 18 indicates that the rechargeable battery in the
tester cannot be recharged. If this error occurs during
recharging, disconnect the tester from the charger and
follow the procedure described in the “fatal” errors.
• Error 19 can be corrected by clearing the entire memory of
the unit (see section 8.3).
• Errors 31, 32 and 33 indicate excessive voltage or
Explanations for some of the basic terms related to earth measurements follow:
Auxiliary electrode - H (Z)
A supplementary electrode through which a measurement current ows.
Earth conductor
The conductor which connects the installation to be earthed to the earth electrode.
Earth connection
A locally restricted group of electrically connected earth contacts. It also includes metal parts of installations such as pylon footings, bracings, cable
sheathing and earth conductors.
Earth electrode - E (X)
A conductor which is buried in the earth and makes electrical contact with it.
Earth potential
The potential dierence measured between the earth connection and a refer-
ence earth electrode.
Earth electrode - ES (Xv)
A connection on the earth electrode or earthing system used for measuring
the potential of the earth electrode.
Earth resistance
The resistance measured between the earth connection and a reference earth
electrode (cf. selective earth resistance).
Earth
The location of an earth connection (see also reference earth).
Earthing system
The set of all installations connected to the earth.
Electrode - S (Y)
An auxiliary electrode used for measuring the potential of the reference earth.
The voltage, which is proportional to the resistance of the earth connection, is
measured between this electrode and the earth electrode.
The area of earth (especially at the surface) which is outside the range of inu-
ence of the earth electrode or earthing system.
Resistivity of the earth (ρ)
Resistivity is dened as the resistance from wall to opposite wall of a cubic
meter of soil. It is measured in ohm-meters (Ωm).
Selective earth resistance
The parallel partial resistance of an earth connection or earthing system. It
is measured by selective measurements of current in the corresponding resistance branch. A selective earth resistance is always greater than the total
earth resistance (parallel connection).
Step voltage
The potential dierence bridged by a person who takes a step of 1m (approx.
3 ft), with the resulting current owing through the body from one foot to the
other (cf. touch voltage).
Supplementary electrode
An additional earth electrode (earth stake, earth rod).
Touch voltage
The potential dierence which a person is able to bridge with his or her body,
the resulting current being limited by the body and the local earth resistance
(cf. step voltage).
The general term earth measurement can refer to measurements of individual
earth electrodes, earth connections or entire earthing systems, depending on the
object of interest.
Glossary
This glossary lists the terms and abbreviations used on the instrument and the
digital display.
3-Poles : measurement of earth resistance with 2 auxiliary stakes (3-
Pole method)
4-Poles : 4-Pole measurement of low earth resistance with 2 auxiliary
stakes (4-Pole method)
GroundFlex® : selective earth measurement using the GroundFlex® Adapter
Model 6474 and GroundFlex® sensor
C1 : earth coupling coecient RA with earth Rb (C1= RC / R1 )
C2 : earth coupling coecient Rb with earth RA (C2= RC / R2 )
d, A : distances to be programmed to calculate resistivity according
: resistance of the earth loop measured with the "2 clamps"
LOOP
function
R
: value of the R-Act resistance (PASS for "passive"
PASS
measurement with interference current circulating in the
installation)
RS : resistance of the stake connected to terminal S
R
: selective earth resistance (measurement of the current with a
SEL
clamp or GroundFlex®)
(2)
R
: resistance between stakes S and ES (used for resistivity
S-ES
measurement)
R∆0 : measuring lead compensation resistance
S : S (Y) terminal (reference potential socket for calculating earth
resistance)
(1)
U-Act
U
U
: external voltage currently present on the instrument terminals
: voltage measured between terminals H and E
H-E
: voltage generated by the instrument between terminals H and
OUT
E (32V or 16V)
U
: voltage measured between terminals S and E
S-E
U
: voltage measured between terminals S and ES
S-ES
USR : frequency chosen by the user (from the English "user")
U
: voltage of terminal S (compared with E) expressed as a
SrEL
relative value (rEL); value without unit
V pot. : ground potential measurement
ρS : ground resistivity measured according to the Schlumberger
method (expressed as Ω.m)
ρW : ground resistivity measured according to the Wenner method
(expressed as Ω.m)
(1) The sux Act becomes In (for "Input" in English) when this value is recorded by the instru-
ment and then read, to make a distinction between the current and recorded values. In both
cases, this value on the display is associated with its frequency.
(2) In this case, the resistances of the 4 stakes used for measurement are indicated by R
The Model 6472 is warranted to the owner for a period of two years from
the date of original purchase against defects in manufacture. This limited
warranty is given by AEMC® Instruments, not by the distributor from whom it
was pur-chased. This warranty is void if the unit has been tampered with,
abused or if the defect is related to service not performed by AEMC®
Instruments.
Please print the online Warranty Coverage Information for your records.
Caution: To protect yourself against in-transit loss, we recommend you insure
your returned material.
NOTE: You must obtain a CSA# before returning any instrument.