Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: ______________________________
Model #: 6471
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC® Instruments
www.aemc.com
READ CAREFULLY BEFORE
USING FOR THE FIRST TIME
Your instrument is equipped with a NiMH battery. This technology oers several
advantages:
• Long battery charge life for a limited volume and weight.
• Possibility of quickly recharging your battery.
• Signicantly reduced memory eect: you can recharge your
battery even if it is not fully discharged.
• Respect for the environment: no pollutant materials such as lead
or cadmium, in compliance with the applicable regulations.
After prolonged storage, the battery may be completely discharged. If so, it must
be completely recharged.
Your instrument may not function during part of this recharging operation.
Full recharging of a completely discharged battery may take several hours.
NOTE: In this case, at least 5 charge/discharge cycles will be necessary
for your battery to recover 95% of its capacity.
To make the best possible use of your battery and extend its eective service life:
• Only use the charger supplied with your instrument. Use of
another charger may be dangerous.
• Only charge your instrument at temperatures between 0° and 40°C.
• Comply with the conditions of use dened in the operating manual.
• Comply with the storage conditions specied in the operating
manual.
NiMH technology allows a limited number of charge/discharge cycles depending
signicantly on:
• The conditions of use.
• The charging conditions.
Please refer to § 9.5 for correct replacement of the battery.
Do not dispose of the battery pack with other solid waste. Used batteries
must be entrusted to a qualied recycling company or to a company
These safety warnings are provided to ensure the safety of personnel.
Please read and comply with these precautions:
• This instrument is protected from accidental voltages of not more than
50V with respect to earth in measurement CAT IV. The guaranteed level
of protection of this equipment may be compromised if used in a manner
not specied by the manufacturer.
• Safety is the responsibility of the operator.
• All metal objects or wires connected to the electrical system should be
assumed to be lethal until tested. Grounding systems are no exception.
• Never exceed the maximum rated voltage and current, and the mea-
surement category.
• Never exceed the protection limits, and always comply with the condi-
tions and place of use, indicated in the specications.
• Do not use the instrument or its accessories if they appear damaged.
• Use accessories that have overvoltage category and service voltages
greater than or equal to those of the instrument (CAT IV 50V). Use only
accessories that comply with safety standards (IEC 61010-2-031 & 32).
• Wear the appropriate protective gear (insulating boots and gloves).
• Check that no terminal is connected and the switch is set to OFF before
opening the device.
• Use only the charging unit supplied with the instrument to recharge the
battery.
• Troubleshooting and metrological verication procedures must only be
performed by qualied, approved personnel, or the factory.
• NOTE: The potentials on the various rods used for an earth measurement
may be dierent if a nearby electrical installation is defective or certain
weather conditions prevail (thunderstorms). It is up to the operator
to decide whether to continue or postpone measurements in these
situations.
6
Ground Resistance Tester Model 6471
1.1 International Electrical Symbols
Signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING that the operator must
refer to the user manual for instructions before operating the instrument. In
this manual, the symbol preceding instructions indicates that if the instructions
are not followed, bodily injury, installation/sample and/or product damage may
result.
Risk of electric shock. The voltage at the parts marked with this symbol may be
dangerous.
Ground/Earth symbol
In conformity with WEEE 2002/96/EC
1.2 Definition of Measurement Categories
CAT II: For measurements performed on circuits directly connected to the electri-
cal distribution system. Exa
ances or portable tools.
CAT III: For measurements performed in the building installation at the distribu-
tion level such as on hardwired equipment in xed installation and circuit
breakers.
CAT IV: For measurements performed at the primary electrical supply (<1000V)
such as on primary overcurren
meters.
mples are measurements on household appli-
t protection devices, ripple control units, or
1.3 Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are consistent with the
packing list. Notify your distributor of any missing items. If the equipment appears
to be damaged, le a claim immediately with the carrier and notify your distributor
at once, giving a detailed description of any damage. Save the damaged packing
container to substantiate your claim.
Ground Resistance Tester Model 6471
7
1.4 Ordering Information
Ground Resistance Tester Model 6471 (no probes included) .... Cat. #2135.48
Includes carrying bag 110/240V power adapter with US power cord, optical USB cable,
rechargeable NiMH battery, and a USB stick with DataView® software, ground tester
workbook and user manual.
Ground Tester Model 6471 (probes included) .............................Cat. #2135.49
Includes carrying bag, set of two SR182 current probes, 110/240V power adapter with US
power cord, optical USB cable, rechargeable NiMH battery, and a USB stick with DataView
software, ground tester workbook and user manual.
Ground Tester Model 6471 Kit (300 ft - probes included) ........... Cat. #2135.50
Includes meter, two carrying bags, two 300 ft color-coded leads on spools (red/blue), two
100 ft color-coded leads (hand-tied, green/black), two 5 ft color-coded leads (red/blue), set
of two SR182 current probes, 110/240V power adapter with US power cord, optical USB
cable, four T-shaped auxiliary ground electrodes, set of ve spaded lugs, one 100 ft tape
measure, rechargeable NiMH battery, and a USB stick with DataView® software, ground
tester workbook and user manual.
Ground Tester Model 6471 Kit (300 ft - no probes included) ......Cat. #2135.60
Includes meter, two carrying bags, two 300 ft color-coded leads on spools (red/blue), two
100 ft color-coded leads (hand-tied, green/black), two 5 ft color-coded leads (red/blue),
110/240V power adapter with US power cord, optical USB cable, four T-shaped auxiliary
ground electrodes, set of ve spaded lugs, one 100 ft tape measure, rechargeable NiMH
battery, and a USB stick with DataView
®
software, ground tester workbook and user manual.
Ground Tester Model 6471 Kit (500 ft - no probes included) ......Cat. #2135.61
Includes meter, two carrying bags, two 500 ft color-coded leads on spools (red/blue), two
100 ft color-coded leads (hand-tied, green/black), two 5 ft color-coded leads (red/blue),
110/240V power adapter with US power cord, optical USB cable, four T-shaped auxiliary
ground electrodes, set of ve spaded lugs, one 100 ft tape measure, rechargeable NiMH
battery, and a USB stick with DataView® software, ground tester workbook and user manual.
®
1.4.1 Kits, Accessories and Replacement Parts
Test Kit for 3-Pole (3-Point) Testing (150 ft) ................................. Cat. #2135.35
Includes carrying bag, two 150 ft color-coded leads on spools (red/blue), one 30 ft lead
(green), two 5 ft color-coded leads (red/blue), two T-shaped auxiliary ground electrodes, set
of ve spaded lugs and one 100 ft tape measure.
Test Kit for 4-Pole (4-Point) Testing (300 ft) ................................. Cat. #2135.36
Includes carrying bag, two 300 ft color-coded leads on spools (red/blue), two 100 ft colorcoded leads (hand-tied, green/black), two 5 ft color-coded leads (red/blue), four T-shaped
auxiliary ground electrodes, set of ve spaded lugs and one 100 ft tape measure.
Test Kit for 4-Pole (4-Point) Testing (500 ft) ................................. Cat. #2135.37
Includes carrying bag, two 500 ft color-coded leads on spools (red/blue), two 100 ft colorcoded leads (hand-tied, green/black), one 30 ft lead (green), two 5 ft color-coded leads (red/
blue), four T-shaped auxiliary ground electrodes, set of ve spaded lugs and one 100 ft tape
measure.
8
Ground Resistance Tester Model 6471
Test Kit for 3-Pole (3-Point) Testing
(Supplemental for 4-Pole (4-Point) Testing) .............................................Cat. #2135.38
Includes carrying bag, two 100 ft color-coded leads (hand-tied, green/black), one 30 ft lead
(green), two T-shaped auxiliary ground electrodes, set of ve spaded lugs and one 100 ft tape
measure.
Extra Large Classic Tool Bag ...........................................................Cat. #2133.73
Replacement - Power Adapter 110/240V & Power Cord 115V US ...Cat. #5000.13
Replacement - Power Cord 115V US Plug .......................................Cat. #5000.14
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com/store for availability
Ground Resistance Tester Model 6471
9
CHAPTER 2
PRODUCT FEATURES
2.1 Description
The Ground Resistance Tester Model 6471 is a portable measurement instrument
designed to measure:
• Ground Resistance with 2 clamps (no auxiliary rods needed)
• Bond/Connection Resistance (2-Pole and 4-Pole Kelvin sensing)
• Ground Resistance (3-Pole or 4-Pole)
• Ground Coupling Resistance
• Selective Ground Resistance
• Soil Resistivity (Wenner or Schlumberger method)
The Model 6471 measures from 0.01 to 99.99kΩ and is auto-ranging, automatically
seeking out the optimum measurement range, test frequency and test current.
Easy-to-use - Simply connect the leads, select the test mode, press Start and read
the results. Up to 512 test results can be stored in internal memory for later recall
to the display or downloaded to a PC via DataView® software.
The large LCD is easy-to-read and indicates ground electrode resistance, test
voltage, current and frequency as well as individual electrode resistance, battery
status and more.
The Model 6471 is CAT IV rated to 50V and is over voltage protected to more than
250VAC against accidental connection to live circuits. The voltage is also displayed
on screen. In the event of a system fault, the Model 6471 can withstand 250VAC.
Additional features of the Model 6471 include a heavy-duty eld case sealed
against dust and water when closed (the test button is also sealed against the
elements); manual and automatic test frequency selection from 41 to 513Hz;
4-Pole Soil Resistivity test methods and user selectable 2-Pole or 4-Pole Bond
Resistance test method.
The Model 6471 is powered by 9.6V, 3.5 Ah NiMH rechargeable batteries. An ex-
ternal recharger powered from 120/230V, 50/60Hz is included and provides for
testing while recharging.
The Ground Resistance Tester Model 6471 is rugged, easy-to-use and ideal for
maintenance crews performing numerous tests. It exceeds mechanical and safety
specications for shock, vibration and drop tests per IEC standards. The adjustable
test frequency provides for rejection of high levels of interference, allowing it to be
used under dicult conditions such as high stray currents that aect accuracy.
10
Ground Resistance Tester Model 6471
2.2 Key Features
• Ground Resistance testing using the 2 clamp method (no auxiliary rods
needed)
• 2- and 4-Pole Bond Resistance/Continuity measurement (DC Resistance)
with automatic polarity reversal
• 4-Pole Soil Resistivity measurement with automatic calculation of Rho (ρ) and
user selection of the Wenner or Schlumberger test method
• 3-Pole Earth Coupling measurement
• Manual and Automatic frequency scan from 41 to 513Hz for optimum test ac-
curacy in electrically noisy environments
• Selectable test voltage of 16 or 32V up to 250mA of test current
• Auto-o power management
• Automatic recognition of all electrode connections and their resistance value
• Stores up to 512 complete test results in internal memory
• Optically isolated USB communication
• Remote set up and operation of all measurements using DataView® software
• Rechargeable NiMH batteries from wall charger or vehicle power
• Rugged dustproof and rainproof eld case – IP53 rated in closed position
• Grounding standards IEC 61557 parts 4 and 5 compliant
• Includes DataView® software for data storage, real-time display, analysis, re-
port generation and system conguration
Ground Resistance Tester Model 6471
11
2.3 Control Features
7
23
Ω
1
Figure 2-1
4
GROUND TESTER
MODEL 6471
5
6
1. Four terminals: H (Z) (auxiliary electrode), S (Y) (electrode), ES (Xv) (earth/
ground electrode) and E (X) (earth/ground electrode)
All terminals accept 4mm Ø banana plugs. Terminals H (Z) and ES (Xv) also
accept special plugs for current clamps. S (Y) will take a shielded cable.
2. Connector for charging the battery (see § 9.4).
3. 256 segment multi-line backlit LCD (see § 2.4).
4. Connector for an optical interface to a PC. Either an RS-232 or USB connection can be used (see § 8 - DataView®).
5. Rotary switch: OFF position, 5 measurement functions (see § 5 and 6) and
SET-UP function (see § 4.2).
6. START/STOP button: Starts the measurement and compensates for the leads
(in the mΩ measurement function - see § 5.2.1).
7. Six function buttons (see § 2.5).
12
Ground Resistance Tester Model 6471
2.4 Display
A1
A2
A3
NOTE: External voltages will be displayed only on the small displays (A2 and
A3) without the view of the main display (A1). This helps to rapidly recognize
that these are measurements of external voltages.
Backlight: The backlight turns ON automatically when entering a function and
turns OFF after the function is complete. Press the DISPLAY button to turn it back
on.
NOTE: In the Set-up mode, pressing the DISPLAY button also selects the
next configurable parameter.
A1
A2
A3
Ground Resistance Tester Model 6471
Top main display
Middle small display
Bottom small display
If this symbol blinks during a passive measurement, a voltage of more than
42V is present at the tester’s inputs.
If this symbol blinks during an active measurement, the limits of use have
been exceeded.
If the symbol appears constant during an active measurement, the values
to be measured are unstable.
Warning buzzer is activated
Figure 2-2
13
> or <
NOISE
EARTH
COUPLING
S x 1/10
A blinking symbol > (greater than) or < (less than) indicates that the
measurement range is exceeded.
If both symbols are blinking during a passive measurement, the values of
voltage and/or current are too small and are beyond the limits of use. The
resistance RPASS displayed by the tester is then highly uncertain.
If both symbols are blinking during an active measurement, the values
to be measured are strongly varying (remedy: switch on smoothing of
measured values with SMOOTH function).
Indicates the battery charge condition; the segments represent the energy.
Indicates whether the 3-Pole EARTH measurement or the EARTH COUPLING
measurement has been selected.
Symbol not used in Model 6471
Symbol not used in Model 6471
Symbol not used in Model 6471
OBJ:TEST
DISTANCE
FREQUENCY
SWEEP
MANUAL
AUTO
CONFIG
MR
MEM
REMOTE
Object and test number for storage in memory.
Indicates lead compensation for 2-Pole measurement is active
If blinking, indicates the value is ready to be overwritten with a new value
or that a new value will be entered. If more than one value has to be
entered, use the CHANGE buttons to move to the next one.
Indicates the test frequency.
Indicates that the secondary function of a button will be used.
Symbol not used in Model 6471
Indicates MANUAL mode - the user has to stop the measurement,
otherwise the measurement keeps on running.
Indicates AUTO mode - measurement stops immediately after all results
are available.
If blinking, indicates the measurement had a disturbance when the measurement was started.
Indicates that the blinking value of AUTO/MANUAL mode, frequency,
voltage or DC current direction can be changed.
Memory recall - indicates the displaying of stored results is active
Indicates that there is recorded data in memory. When blinking, it indicates that the results need to be saved by pressing the MEM button.
Indicates the instrument is remotely controlled by the RS-232/USB
interface.
14
Ground Resistance Tester Model 6471
SMOOTH
Indicates SMOOTHing of the displayed measurement results.
Indicates that a clamp is (steady) or should be (flashing) connected to ES.
ALARM
Measured value (R, U, I).
Indicates that the measurement is running.
Indicates the alarm is activated. An audible alarm will sound if the value
measured is above the limit defined in the SET-UP mode.
Distances to be entered between earth electrode and supplementary
electrodes or between earth electrodes.
Clamp should be connected to H.
If this clamp symbol blinks, the instrument is refusing measurement
because the clamp is not connected correctly.
Symbolizes the soil resistivity value (Rho): W or S indicates that Wenner
or Schlumberger method is used for the measurement (changed by the
Hz/OPTIONS button).
2.5 Button Functions
Starts a measurement and compensates for lead resistance (§ 5.2.1).
START/STOP
Measurement stops automatically (AUTO) or is stopped by pressing this
button again (MANUAL).
Selects the second function highlighted in yellow under the buttons.
Increases or decreases the value of the flashing parameter displayed or
MEM/MR
DISPLAY/
SMOOTH
Hz/OPTIONS/
DISTANCE
Ground Resistance Tester Model 6471
selects the next parameter. Press the button for a longer time to increase
the change of value speed at a faster rate.
Selects the parameter to be modified or moves the cursor to the right.
MEM: Stores measurements to memory.
MR: Retrieves measurements from memory.
DISPLAY: Displays all parameters of a measurement.
SMOOTH: Smooths the display of measurement providing a more
stable result.
Hz/OPTIONS: Enables configuration of measurement functions.
DISTANCE: Available for soil resistivity and V potential measurements. It
allows the setting of values of distances used for Rho (ρ) calculation.
15
CHAPTER 3
SPECIFICATIONS
3.1 Reference Conditions
Influencing ParametersReference Values
Temperature68°F ± 5.5°F (20°C ± 3°C)
Relative Humidity45 to 55% RH
Power Supply9 to 11.2V
Frequency Range of Input Signal0 to 440Hz
Capacitance Parallel to Input Resistance0 uF
Electric Field<1 V/m
Magnetic Field< 40 A/m
3.2 Electrical
3.2.1 Frequency Measurements
Measuring method: Digital with a sampling frequency of 4028Hz, low pass, FFT.
The frequency of the strongest spectral component is displayed.
Measuring rate: Display updates approx. 3 times per s.
Measurement Range5 to 450Hz
Resolution1Hz
Operating error± 2Hz
Min. input voltage10mV
Min. current through a clamp0.5mA
3.2.2 Voltage Measurements
Overvoltages up to 75Vrms are displayed as ">65V". Permanent overvoltages
between 70 and 75V at terminals H (Z) and E (X) can cause overheating of the
overvoltage protection varistor.
Voltages of more than 75Vrms lead to error message 31 (excessive external
voltage) or 32 (voltage measurement overrange).
If terminals H (Z) and E (X) are put into contact with the line voltage, the protection
fuse will blow.
16
Ground Resistance Tester Model 6471
External Voltage Measurements
Measuring Method: Digital by sampling at 4028Hz, low pass lter, FFT.
The frequency of the strongest spectral component is displayed.
Measuring Rate: Display updates approx. 3 times per s.
Signal Conversion: TRMS or sum of all harmonics 10 to 450Hz during selective
earth measurements with a clamp.
Measured Range0.00 to 9.99V10 to 65V
Resolution0.01V0.1V
Intrinsic error± (2% + 1ct)
Operating error± (5% + 1ct)
Input impedance Z
, Z
(Z
H-E
) 1.2MΩ
S-E
S-ES
Operating frequencyDC and 15 to 440Hz
3.2.3 Current Measurements
External Current Measurements
Measuring Method: Digital by sampling at 4028Hz, low pass lter, FFT.
Measuring Rate: Display updates approx. 3 times per s.
Signal Conversion: Sum of all harmonics 10 to 450Hz
With Current Probe SR182
Measurement Range
0.00 to
9.99mA
10.0 to
99.9mA
100 to
999mA
1.00 to
9.99A
Resolution0.01mA0.1mA1mA0.01A0.1A
Frequency range16 to 49Hz50 to 99Hz100 to 400Hz
Operating error from
0.5 to 100mA
Operating error from
0.1 to 40.0A
Ground Resistance Tester Model 6471
± (10% + 2cts)± (5% + 2cts)± (3% + 2cts)
> 20%± (10% + 2cts)± (5% + 2cts)
10 to
40A
17
With Current Probe MN82
Measurement Range
0.00 to
9.99mA
10.0 to
99.9mA
100 to
999mA
1.00 to
9.99A
10 to
40A
(1)
Resolution0.01mA0.1mA1mA0.01A0.1A
Frequency range16 to 49Hz
Operating error from
0.5 to 100mA
± (15% + 2cts)± (7% + 2cts)± (5% + 2cts)
Operating error from
0.1 to 40.0A
(1): The Model 6471 cannot detect whether a current clamp SR182 or MN82 is connected. In the case
of the MN82 clamp, with currents > 10A and frequencies < 100Hz the instrument will not display
any warning messages. It is the operator’s responsibility to observe the limits of use when using the
MN82 clamp.
(1)
50 to 99Hz
> 20%± (15% + 2cts)± (7% + 2cts)
(1)
100 to 400Hz
3.2.4 DC Resistance Measurements
Measuring method: Voltage/Current measurement (Standard EN 61557 part 4)
Nominal output voltage: 16VDC (if resistance < 22Ω the output voltage is reduced
to 10VDC)
Max output current: > 200mADC for resistances < 20Ω
Max overload (permanent): 50Vrms (protection up to 250V is guaranteed)
Max inductive load: 2 H
Max interfering voltage: 60Vpeak > 10Hz
Time for auto range selecting: approx 5 s
Measuring time: 8 s with automatic polarity inversion
Measuring rate: 3 per s in manual mode
Lead compensation: Possible from 0 to 5Ω
Alarm setting: “>” or “<” from 1 to 999Ω (max 3 digits)
2-Pole mΩ measurement
Measurement
Range
0.12 to
9.99Ω
10.0 to
99.9Ω
100 to
999Ω
1.00 to
9.99kΩ
10.0 to
99.9kΩ
Resolution0.01Ω0.1Ω1Ω10Ω100Ω
Intrinsic error± (2% + 2cts)
Operating error± (5% + 3cts)
18
Ground Resistance Tester Model 6471
4-Pole mΩ measurement
Measurement
Range
Resolution0.001Ω0.01Ω0.1Ω1Ω10Ω
Intrinsic error± (2% + 2cts)
Operating error± (5% + 5cts)
0.020 to
9.999Ω
10.00 to
99.99Ω
100.0 to
999.9Ω
1.000 to
9.999kΩ
10.00 to
99.99kΩ
3.2.5 AC Earth/Ground Resistance Measurements
Measuring method: Voltage/Current measurement (EN 61557 part 5)
Open circuit voltage: 16 or 32Vrms square wave (if current > 240mA the output
voltage is reduced to 10Vrms)
Test frequency: Selectable from 41 to 513Hz (see table in § 6.2)
Short circuit current: > 200mAAC
Noise suppression: > 80 dB for frequencies 20% or more above or below the
test frequency
Max. overload: 250Vrms
Max. value for RH & RS: 100kΩ
Measuring time:Short push on START: approx. 7 s for rst value of RE at 128Hz, then 3 measurements per s.
Long push on START: approx. 15 s for rst value of RE at
128Hz, then 3 measurements per s.
The following error indications refer to reference conditions with a test voltage of
32V, test frequency of 128Hz, RH and RS = 1kΩ, no external voltage.
The operating error of AC resistance measurements can be less than that speci-
ed for voltage or current because frequency characteristics of the voltage channel
are matched to those of the current channel.
Measurement of auxiliary electrodes RH, RS, RES, RE
Measurement
Range
Resolution0.1Ω0.1Ω1Ω10Ω100Ω
Operating error± (10% + 2cts)
0.14 to
9.99Ω
10.0 to
99.9Ω
100 to
999Ω
1.00 to
9.99kΩ
10.0 to
99.9kΩ
The START button must be pressed for more than 2 s. For frequencies between
41 and 256Hz, the resistance of the auxiliary earth electrodes is measured at the
test frequency set. At higher frequencies the resistance is measured at 256Hz.
Ground Resistance Tester Model 6471
19
3-Pole earth resistance measurement RE
Measurement
Range
0.09 to
9.99Ω
10.0 to
99.9Ω
100 to
999Ω
1.00 to
9.99kΩ
10.0 to
99.9kΩ
Resolution0.01Ω0.1Ω1Ω10Ω100Ω
Intrinsic error± (2% + 1ct)
Operating conditions: RE < 3 x RH, U
Values for RH, RS and RE Frequency (Hz)
(RH + RS) / RE
< 3000
(RH + RS) / RE
< 5000
Note: For a test voltage U
RH ≥ 0Ω, Rs ≤ 3kΩ41 to 513± (3% + 2cts)
RH > 3kΩ, Rs ≤ 30kΩ41 to 513± (10% + 2cts)
RH > 30kΩ, Rs < 100kΩ41 to 128± (10% + 3cts)
of 16V, halve the value for RH.
OUT
4-Pole earth resistance measurement R
Measurement
Range
0.01 to
9,999Ω
10.00 to
99.99Ω
OUT
= 32V
E
100.0 to
999.9Ω
Operating error for R
1.000 to
9.999kΩ
10.00 to
99.99kΩ
Resolution0.001Ω0.01Ω0.1Ω1Ω10Ω
Intrinsic error± (2% + 1ct)
Operating conditions: RE < 3 x RH, U
Values for RH, RS and RE Frequency (Hz)
(RH + RS) / RE
< 3000
(RH + RS) / RE
< 5000
RH ≥ 0Ω, Rs ≤ 3kΩ41 to 513± (3% + 2cts)
RH > 3kΩ, Rs ≤ 30kΩ41 to 513± (10% + 2cts)
RH > 30kΩ, Rs < 100kΩ41 to 128± (10% + 3cts)
Note: With a test voltage U
of 16V, halve the value of RH.
OUT
OUT
= 32V
Operating error for R
E
E
Selective 4-Pole earth resistance measurement with SR182 or MN82 clamp
Same characteristics as 4-Pole earth measurements with the following particular
conditions:
Minimum current: SR182, IES >0.5mA
MN82, IES >2mA
Maximum R
MN82, (R
ratio: SR182, (R
SEL/RE
SEL/RE
SEL/RE
) <500
) <120
with RH + RE <20Ω
20
Ground Resistance Tester Model 6471
3.2.6 Soil Resistivity Measurement ρ
Measuring method: Voltage/Current measurement (EN 61557 part 5)
Open circuit voltage: 16 or 32Vrms square wave
Test frequency: Selectable from 41 to 128Hz (see table in § 6.2)
Short circuit current: > 200mAAC
Noise suppression: > 80 dB for frequencies 20% or more above or below the
test frequency
Max. overload: 250Vrms
Max. value for RH, RS, RES, RE: 100kΩ
Calculation formula, Wenner: ρW = 2p d R
Calculation formula, Schlumberger: ρS = (p (d2 - (A/2)²) / A) RMaximum value of ρ: 999kΩm (display in kΩft is not possible)
Measuring time: Short push on START: approx. 8 s for rst value of R
128 Hz, then 3 measurements per s.
Long push on START: approx. 20 s for rst value of R
128 Hz, then 3 measurements per s.
S-ES
S-ES
S-ES
S-ES
at
at
Measurement
Range
0.00 to
9.99Ω
10.0 to
99.9Ω
100 to
999Ω
1.00 to
9.99kΩ
10.0 to
99.9kΩ
Resolution0.01Ω0.1Ω1Ω10Ω100Ω
Intrinsic error± (2% + 1ct)
The intrinsic error specied refers to reference conditions with a test voltage of
32V, test frequency of 128Hz, R
Operating conditions : R
R
R
R
R
R
R
R
is the resistance of the earth rods R
rod
cal.
electrode
electrode
electrode
S-ES
electrode
/ R
electrode
/ R
electrode
/ R
Note: With a test voltage U
Ground Resistance Tester Model 6471
, R
, R
ρ-H
< 3 x R
ρ-S
and:Operating error for R
ρ
-H
ρ-ES
and R
≤ 100kΩ
≤ 2000
S-ES
≤ 50kΩ
≤ 10,000
S-ES
≤ 10kΩ
≤ 20,000
S-ES
, R
ρ
of 16V, halve the value of R
OUT
, R
ρ
-E
-ES
= 1kΩ, no external voltage.
ρ-E
S-ES
± (7% + 2cts)
± (15% + 3cts)
± (20% + 1ct)
, R
ρ
assumed to be identi-
ρ
-E
-H
.
rod
21
Measurement of auxiliary electrodes R
Measurement
Range
0.14 to
9.99Ω
10.0 to
99.9Ω
, R
, R
, R
ρ
ρ
-H
ρ
-S
100 to
999Ω
-ES
ρ
-E
1.00 to
9.99kΩ
10.0 to
99.9kΩ
Resolution0.1Ω0.1Ω1Ω10Ω100Ω
Operating error± (10% + 2cts)
The START button must be pressed for more than 2 s. For frequencies between
41 and 128Hz, the resistance of the auxiliary earth electrodes is measured at the
test frequency set. At higher frequencies the resistance is measured at 128Hz.
3.2.7 Earth/Ground Measurements with 2 Clamps
Measuring method: Voltage/Current measurement with a rectangular AC signal
Signal frequency: Automatic: 1611Hz - Manual: 128, 1367, 1611 or 1758Hz
Noise suppression: > 80 dB at frequencies diering by 20% or more from the
test frequency
Max. interfering current: 20A
Max. value for RH, RS: 100kΩ
Measuring time: approx. 7 s for the rst value of R
Measurement Range0.10 to 9.99Ω10.0 to 99.9Ω100 to 500Ω
Measurement frequency1367Hz - 1611Hz - 1758Hz
Resolution0.01Ω0.1Ω1Ω
Operating error
(1)
(without external current)
peak
, then 3 meas. per s.
Loop
SR182 ± (10% + 1cts)
MN82 ± (20% + 2cts)
Measurement Range0.10 to 9.99Ω10 to 30Ω
Measurement frequency128Hz
Resolution0.01Ω0.1Ω
Operating error
(1)
(without external current)
(1): Observe the minimum distance between the two clamps according to table in § 5.7.
SR182 ± (20% + 2cts)
MN82not specified
3.3 Data Storage
Memory Capacity: 512 test results (64kB)
Communication: Optically isolated USB
22
Ground Resistance Tester Model 6471
3.4 Power
Power Source: Rechargeable 9.6V, 3.5Ah NiMH Battery Pack
Power Supply: 110/220, 50/60Hz external charger with 18VDC, 1.9A output or
12VDC vehicle power
Fuse: 0.63A, 250V, 5x20mm, 1.5kA
Battery Life:
FunctionPower consumedTypical number of measurements
Device off< 5 mW—
Voltage Measurement1.5 W4500
(2)
mΩ
3-Poles, 4-Pole
2 Clamps3.7 W2000
(1): Measurements in automatic mode at 25-s intervals
(2): With R = 1Ω
(3): With RH + RE = 100Ω
(4): With RH + RS-ES = 100Ω
(3)
(4)
ρ
4.9 W1500
4.9 W1500
4.9 W1500
3.5 Mechanical
Dimensions: 10.7 x 9.76 x 5.12" (272 x 248 x 130mm)
Weight: 7 lbs (3.2kg) approx
Case Material: ULV0 Polypropylene
Terminals: 4mm recessed banana jacks
Case Protection: EN 60529 - IP53 (cover closed)
Drop Test: Per EN 61010-1
Vibration Test: Per EN 61557-1
(1)
Ground Resistance Tester Model 6471
23
3.6 Environmental
Operating Temperature: 32° to 113°F (0° to 45°C); 0 to 90% RH
Specied Operating Temperature
Storage Temperature: -40° to 158°F (-40° to 70°C); 0 to 90% RH
Altitude: < 3000m
(1): This range corresponds to the one defined by standard EN 61557, for which an operating error including the
quantities of influence is defined. When the device is used outside this range, 1.5%/10°C and 1.5% between
75 and 90% RH must be added to the operating error.
(1)
: 0° to 95°F (0° to 35°C); 0 to 75% RH
3.7 Safety
Electromagnetic Compatibility
This instrument satises the EMC and LVD directives required for the CE marking
and product standard IEC 61326-1 (Ed. 97) + A1 (Ed. 98).
• Immunity in industrial environment
• Emissions in residential environment.
Electrical safety according to EN 61010-1 (Ed. 2 of 2001)
Measurement according to EN 61557 (Ed. 2 of 2007) parts 1, 4 and 5.
CAT IV, 50V
Pollution Degree 2
*Specications are subject to change at any time without notice
24
Ground Resistance Tester Model 6471
CHAPTER 4
OPERATION
NOTE: For detailed information regarding ground resistance testing, see
the Understanding Ground Resistance Testing Workbook CD that was
included with the instrument.
4.1 General Operating Instructions
The Model 6471 has operating modes:
• Automatic mode for routine applications.
• Manual (Expert) mode in which the user can change the parameters of the
measurement functions.
NOTE: Fully charge the battery before the first use (see § 9.4)
4.1.1 Automatic Mode
• Set the switch to the desired function.
• Make the connections appropriate to the function.
• Press the START/STOP button. The device makes the measurement and
stops automatically.
• Read the measurement result on the display and the relevant parameters
using the DISPLAY button.
• To save the information in memory, use the MEM button (see § 7.1).
4.1.2 Manual (Expert) Mode
• Set the switch to the desired function.
• Make the connections appropriate to the function.
• Select MANUAL mode.
• Choose various measurement parameters using the Hz/OPTIONS button.
• Press the START/STOP button. The measurement frequency or the direc-
tion of the current (resistance measurement) can be changed during the
measurement. To view their impact upon the measurement and the parameters relevant to the measurement use the DISPLAY button.
Ground Resistance Tester Model 6471
25