Aemc instruments 3730, 3710 User Manual

Clamp-On
HOLD
+ —
A
R
1
HOLD
MEM
MR
NOISE
Model 3710
GROUN D TES TER
AEMC
I NS T R U M E NT S
A
ON
OFF
HOLD
+ —
A
R
1
HOLD
MEM
MR
NOISE
Model 3730
GROUN D TES TER
AEMC
I NS T R U M E NT S
MEM AL A
ON
OFF
Ground Resistance Testers Models 3710 and 3730
USER MANUAL
Owner’s Record
The serial number for the Model 3710 or Model 3730 is located inside the battery compartment of the instrument. Please record this number and purchase date for your records.
CLAMP-ON GROUND RESISTANCE TESTER MODEL 3710 CATALOG #1221.01 SERIAL #: PURCHASE DATE: DISTRIBUTOR:
CLAMP-ON GROUND RESISTANCE TESTER MODEL 3730 CATALOG #1221.02 SERIAL #: PURCHASE DATE: DISTRIBUTOR:
99-MAN-100061 5th Revision Printed 03/98
AEMC®Instruments by Chauvin Arnoux, Inc.

Table of Contents

Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Receiving Your Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Government Orders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Mechanical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Safety Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Model 3710 Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Model 3730 Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Digital Display Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Additional Features of the Model 3730 . . . . . . . . . . . . . . . . . . .12
Function Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Programming the Alarm Setpoint . . . . . . . . . . . . . . . . . . . . . . .17
Changing the Alarm Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Memory Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Clear Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Function Access Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Programming Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Principle of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Field Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Operating Reference Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Calibration Check Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Commonly Asked Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Table of Contents (cont.)
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Repair and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Technical and Sales Assistance . . . . . . . . . . . . . . . . . . . . . . . . . .36
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . .Inside back cover

Warning

These safety warnings are provided to ensure the safety of per­sonnel and the proper operation of the instrument.
• The instrument must not be operated beyond its specified oper­ating range.
• Safety is the responsibility of the operator.
• All metal objects or wires connected to the electrical system should be assumed to be lethal until tested. Grounding systems are no exception.
• Use extreme caution when using the instrument around ener­gized electrical equipment.
• Never attempt to use the instrument to twist or pry the ground electrode or ground wire away from the equipment being grounded.
• Open and close the clamp-on jaws slowly ensuring proper alignment.
• AEMC®Instruments considers the use of rubber gloves to be an excellent safety practice even if the equipment is properly oper­ated and correctly grounded.
Ground Resistance Tester Models 3710/3730
3

Receiving Your Shipment

Upon receiving your shipment, be sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, giving a detailed description of any damage. Save the damaged packing container to substantiate your claim.

Packaging

The Clamp-On Ground Resistance Testers Model 3710 (Cat. #1221.01) and Model 3730 (Cat. #1221.02) are shipped with a hard carrying case, one 9 V battery (not installed), an instruction manual with warranty registration and a calibration loop. The manual, warranty card, and cal­ibration loop are located behind the “egg crate” foam on the inside of the carrying case cover.

Government Orders

The Clamp-On Ground Resistance Tester Model 3710 (Cat. #1221.01) has been issued a National Stock Number. Orders for the Model 3710 should reference NSN 6625-01-377-8030.

Accessories

One multiple calibration loop, Cat. #1223.01
4
Ground Resistance Tester Models 3710/3730

Description

Outline
The ground resistance tester Models 3710 and 3730 measure ground rod and small grid resistance through any season, without the use of auxiliary ground rods. Clamp-on ground resistance testers are used in multi­grounded systems without disconnecting the ground under test. The Models 3710 and 3730 simply clamp around the ground conductor or rod and measure the resistance to ground. By performing measurements on intact ground systems, the user also verifies the quality of the grounding connections and bonds. Resistance and continuity of grounding loops around pads and buildings may also be measured.
Both models include a current measurement function. The probe’s high sensitivity enables measurement of leakage current flowing to ground or circulating in ground loops down to 1 mA and neutral currents to 30 A rms. This feature provides additional information which is becoming vital as distribution ground networks carry higher levels of noise and harmonics which affect power quality.
Both models offer battery life information at power-up and Auto-Off for power management. Additional features are also displayed on the large LCD to ensure precise measurements. The Buzzer and Auto-Off features may be disabled from the push-buttons at any time.
The Model 3730 offers an Alarm function and a Memory (logging) func­tion. In the Alarm mode, the probe will audibly and visually indicate if the reading is beyond an input set point. The user may also have the Alarm initiated above or below the set point. This Alarm feature permits quick field checks where only “pass” or “fail” readings will suffice.
The Memory function logs up to 99 measurements (Ohms or Amperes). This enables the user to conduct field surveys, and to retrieve and analyze the readings at a later time. The alarm settings and stored memory information are saved when the ground tester is turned off.
Product Construction
The Models 3710 and 3730 body cases are built of Lexan®for rugged use. The probe heads are encapsulated in a double-walled Lexan®shell
Ground Resistance Tester Models 3710/3730
5
for extra strength and are reinforced at the body interfaces for enhanced field reliability. Overall construction and mechanical design ratings such as drop test, shock and vibration, weatherproofing against water projections or dust, meet or exceed IEC (International Electrotechnical Commission) standards. The products have also been designed to meet IEC 1010, UL 1244, CSA and GS safety approvals .
Jaw Design
The probe head, or jaw, is a key component in the measurement and overall product performance. Considerable research and development were dedicated to ensuring its performance, reliability and versatility.
The large jaw has an inner opening diameter of 1.25" (32 mm) - double its predecessor’s diameter - while retaining the same overall jaw thickness, permitting use on tight ground conductors on poles and in manholes. The large opening accommodates not only ground rods, but larger ground conductors (up to 1000 MCM) typically found in telecom­munication or railroad applications.
The inner jaw is composed of two independent and individually shielded magnetic cores permitting the injection of a test signal and accurate return signal measurement without noise interference or cross talk common to separate probe instruments.
Thorough mechanical design, including small winglets, ensures repetitive jaw alignment for accuracy and prevents undesirable insertions into the jaw spring assembly.
Body Design
The ergonomic body design permits one-handed operation. The guard provides additional strength, and prevents the hand from slipping or coming into contact with conductors under test. The Lexan®body con­struction provides strength. The LCD lens cover may be easily replaced if scratched. The sealed push-buttons directly access all test functions and are easily operated even with gloved hands.
Lexan®is a registered trademark of General Electric Corporation.
6
Ground Resistance Tester Models 3710/3730

Electrical Specifications

Ground Resistance
Measurement Range
Autoranging
1.0 to 1200
Range
1.0 to 50.0
50.0 to 100.0 100 to 200 200 to 400 400 to 600
600 to 1200
Ground or Leakage Current
Autoranging
1 mA to 30.00 A rms
* Reference conditions: 23°C ± 3K, 50% RH ± 10%, battery at 8 V ± 0.2V, external magnetic field < 40 A/m, external electrical field < 1 V/m, conductor centered, loop resistance noninductive. Accuracy % of reading.
1 to 300 mA
0.300 A to 3.000 A
3.00 A to 30.00 A
Resistance Measurement Frequency: 1689 Hz Resistance Overload: OL displayed above 1200 Current Measurement Frequency: 47 to 800 Hz Current Overload: OL displayed above 30 A rms
Resolution
0.1
0.5 1 5
10 50
1 mA
0.001 A
0.01 A
Accuracy*
± (1.5% + 0.1 ) ± (2.0% + 0.5 )
± (3.0% + 1 ) ± (6.0% + 5 )
± (10% + 10 )
Approx. 25%
± (2.5% + 2 mA) ± (2.5% + 2 mA)
± (2.5% + 20 mA)
MECHANICAL SPECIFICATIONS Dimensions: 9.25 x 3.94 x 2.17" (235 x 100 x 55 mm) Weight: 2.2 lbs. (1 kg) Case Material: Lexan®920A (UL94V2)
Ground Resistance Tester Models 3710/3730
7
Jaw Cover Material:
Lexan®500R with 10% fiberglass charge (UL94V0)
LCD Cover Material: Lexan®920A (UL94V1) Color: Gray body, red jaws Jaw Window Diameter: 1.25” (32 mm) Jaw Opening: 1.38” (35 mm) Operating Temperature: -10 ° to 55 °C Operating Humidity: 0 to 90% RH @ -10 ° to 40 °C, 75% RH @ 55 °C Storage Temperature: -40 ° to 70 °C Power Supply: 9 V alkaline (IEC 6LF22 or NEDA 1604A) Battery Life:
typical: 8 hours or approximately 1000 measurements of 30 seconds
LCD: 3-3/4 digit, 1.73 x 1.10” (44 x 28 mm) US Design Patent: #362,639

SAFETY SPECIFICATIONS

UL, CSA approvals pending IEC 1010-1 Double Insulation
Environmental: IP30, IEC 359 Group III Vibration Test: IEC 68-2-6 Shock Test: IEC 68-2-27 Drop Test (1m): IEC 68-2-32 Dielectric Test: 2500 V AC Working Voltage: 150 V, Cat. III - Pollution Degree 2
300 V, Cat. III - Pollution Degree 1
Max A Overload: 100 A continuous, 200 A (< 5s) 50/60 Hz OL displayed above 30 A rms
Calibration Cycle: Calibration once a year is recommended
8
Ground Resistance Tester Models 3710/3730

Model 3710 Controls

HOLD
+ —
m
A
R
1
HOLD
MEM
MR
NOISE
Model 3710
GROUN D TES TER
AEMC
IN S T R U ME N T S
A
ON
OFF
1
2
3
4
7
56
1. Head Assembly: Consists of two individually shielded magnetic cores
2. Hold: Freezes last measured value on the display
3. Display: 3000 count LCD,
Ground Resistance Tester Models 3710/3730
function indicators
4. ON/OFF: Power ON or OFF, activates display self-test at power-up
5. : Resistance measurement
6. A: Current measurement
7. Lever: Opens or closes jaws
9

Model 3730 Controls

HOLD
+ —
m
A
R
1
HOLD
MEM
MR
NOISE
Model 3730
GROUN D TES TER
AEMC
IN S T R UM E N T S
MEM AL A
ON
OFF
1
2
3
49
7
568
1. Head Assembly: Consists of two individually shielded magnetic cores
2. Hold: Freezes last mea­sured value on the display
3. Display: 3000 count LCD, function indicators
4. ON/OFF: Power ON or OFF, activates display self test at power-up
5. : Resistance Measurement Increment the Alarm set-
point, increment Memory position in program mode
6. A: Current Measurement Decrement the Alarm
setpoint, decrement the Memory in program mode
7. Lever: Opens or closes jaws
8. AL: Activates/deactivates the Alarm function, access the value of the Alarm setpoint in program mode
9. MEM: Stores the reading into memory, reads Memory Recall (MR) values and clears the memory in recall mode
10
Ground Resistance Tester Models 3710/3730

Digital Display Features

mA, A
R
1
NOISE
HOLD
+ —
m
A
R
1
HOLD
MEM
MR
NOISE
Figure 3
Displayed when the beeper is active. To turn the beeper On/Off, push ON & together at power-up. The beeper is active when displayed.
Displayed when the HOLD function is activated. This function locks the present displayed value. During HOLD: A, , and Alarm functions are deactivated. The Memory function may still be used.
Indicates that the jaws of the instrument are not properly closed. The Open Jaw indication is present only in the Mode. Check for foreign matter in the jaw mating surface.
In the function, this symbol indicates the presence of exces­sive stray noise in the ground electrode under test. This symbol will be present if noise signals reach an amplitude of approx. 5 A or 50 V. Resistance measurement values are no longer valid. The beeper is also activated.
Displayed when measured resistance is below 1 . This may indicate that grounding electrode you are clamped onto is continuous with itself (the instrument is measuring a metallic loop and not the ground electrode resistance). In this situation, electrode resistance measurements may not be valid.
Current measurement mode.
Ground resistance measurement mode.
Ground Resistance Tester Models 3710/3730
11
Indicates the Auto-Off feature is deactivated and the instrument
MR
MEM
+ –
stays on permanently (until turned off by the user). With no symbol displayed, after 5 minutes of non-use the instrument will automatically shut itself off. The Auto-Off feature is turned On/Off by pressing Hold at power-up.
Display flashes this symbol (1/second) during a low battery condition. Measurements are still possible. When this symbol is displayed continuously, measurements are no longer possible. Battery replacement is necessary.
Indicates the approximate percentage of useful battery life remaining (0-100%). This function is displayed only upon initial power-up when the ON push-button is held down for more than 3 seconds.
The overload symbol will illuminate when the measurement reading is beyond the limits of the instrument:
Resistance > 1200 Current > 30.00 A rms The beeper is activated in current overload.

Additional Features of the Model 3730

Indicates that an alarm set point is set in memory. Alarm set points may be set from 1 to 199.
Arrows indicate whether the measured value is above or below alarm setpoint.
Memory function is activated. Ninety-nine (99) separate values may be logged for either resistance (), current (A) or a com­bination thereof. At the 99th value, display number will blink indicating that memory is full.
Memory recall register. Displays the numbered value being recalled from memory (1-99).
Note: Models 3710 and 3730 share the same LCD and display all segments at power-up. Functions not available on the 3710 are displayed at power-up.
12
Ground Resistance Tester Models 3710/3730

Function Controls

+ –
m
A
R
1
HOLD
MEM
MR
NOISE
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
+ –
m
A
R
1
HOLD
NOISE
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
Model 3710 and 3730 share common functions (On/Off, A, Ω, Auto-off, Buzzer disable, Hold). Even though the Model 3710 is shown in the examples for these functions, the controls will respond identically for the Model 3730.

On/Off

The ON/OFF push-button, when pushed actuates power on or power off with an audible control tone. Upon initial power-up, the instrument will be in the Ohms function () and will go to OL until clamped on. When the ON push-button is pushed and held down, after three seconds the instrument will beep and provide a self test featuring a full function dis­play (Fig. 4) followed by status indications for:
a. Status of remaining battery life (battery symbol blinks, 0-100%) (Fig. 5) b. Number of stored values in memory (MEM blinks, 0-99) (Model 3730
only) Release the ON/OFF push-button to initialize the instrument. A light chirping sound may be heard in the mode. This is a normal
condition generated by the instruments power management system.
Figure 4
Ground Resistance Tester Models 3710/3730
Figure 5
13

Auto-Off

HOLD

MR
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
MR
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
HOLD
A
ON
OFF
HOLD
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
The instrument will shut off automatically after five minutes of non-use. The instrument will issue a short beep 15 seconds prior to shutting down and the LCD flashes once per second. The Auto-Off may be disabled by turning on the meter while holding down the HOLD push-button. The indicator will appear on the display to indicate the Auto-Off feature has been disabled. (Fig. 6)
Hold
The HOLD push-button freezes the last measured value onto the display.
Figure 6
When the HOLD function is enabled,
is displayed on the LCD. (Fig. 7)
When the HOLD function is enabled, other control functions are dis­abled (except the memory function).
To exit HOLD, press the HOLD push-button and normal operation is resumed.
Note: When HOLD is pressed dur­ing power-up, the auto-off feature is disabled and is displayed on the LCD. See Auto-Off.
Figure 7

Beeper Disable

To disable the beeper, press the ON and Ω together at power-up. The indicator will disappear from the LCD. (Fig. 8)
14
Ground Resistance Tester Models 3710/3730
Figure 8

Resistance

R
1
HOLD
MR
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
R
1
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
MR
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
The function provides resistance measurements from 1.0 to 1200in six autoranges. (Fig. 9)
Range Resolution
1.0 to 50.0 0.1
50.0 to 100.0 0.5 100 to 200 1 200 to 400 5 400 to 600 10 600 to 1200 50
Figure 9
Upon initial power-on, the function is chosen by default.
The Models 3710 and 3730 will track down to 0.7 and indicate below 1 (Fig. 10). Below 0.7 the display stays locked on 0.7 . Accuracy is not defined below 1 . Measurements below 1 typically indicate the ground tester is clamped onto a closed loop and that the signal is not flowing through the ground under test.
Figure 10
If the resistance readings are above 1200 Ω, OL will be displayed on the LCD (Fig. 11). This may indicate high ground resistance but may also be caused by a disconnected or poorly bonded ground.
Figure 11
Ground Resistance Tester Models 3710/3730
15

Current

m
A
HOLD
MR
Model 3710
GROUND TESTER
AEMC
INSTRUMENTS
A
ON
OFF
Press A to enter the current function. True RMS current measurements are possible up to 30.00 A.(Fig. 12)
Above 30 Arms, OL will be displayed and the test must be discontinued. The beeper is also activated during OL.
Range Resolution
1 to 300 mA 1 mA
0.300 to 3.000 A 0.001 A (1 mA)
3.00 A to 30.00 A 0.01 A (10 mA)
Figure 12
16
Ground Resistance Tester Models 3710/3730

Alarm

Model 3730
GROUND TESTER
AEMC
INSTRUMENTS
MEM AL A
ON OFF
Model 3730
GROUND TESTER
AEMC
INSTRUMENTS
MEM AL A
ON
OFF
Turn the Alarm on/off by pressing the AL button. The alarm feature pro­vides a continuous audible warning (when beeper is active) and visual indications when a setpoint has been reached (Fig. 13, 14) (e.g. alarm indi­cation when ground electrode resis­tance is above the NEC limit of 25 .) If the alarm is not triggered, only will be displayed indicating that the alarm is active.
Figure 13

Programming the Alarm Setpoint

To program the alarm setpoint, push ON and AL together at power-up. The alarm setpoint value will blink. To program the proper setpoint value
(1 to 199 ), use the curser arrows ( down, up).
When the setpoint is reached, turn the instrument off. This will put the setpoint into memory. Turn the instrument back on (you may need to press AL to activate the alarm fea­ture).
Note: The alarm setpoint is preset at
Figure 14
25 . The alarm will be triggered when the setpoint is
exceeded
.
The user has the option of having the alarm triggered when a reading is
below
the programmed setpoint. This is accomplished with an internal
switch and outlined in the following section.
Ground Resistance Tester Models 3710/3730
17

Changing the Alarm Trigger

The Model 3730 alarm may be set to trigger above or below a setpoint. Follow the procedure below for setting the alarm trigger.
1. The instrument must be OFF and disconnected from any conductor.
2. Place the meter face down and remove the two phillips screws on the back of the instrument.
3. Remove the back cover by pulling down and away from the instrument.
4. To set the alarm to trigger above a setpoint, slide the alarm setpoint switch to the right. To set the alarm to trigger below a setpoint, slide the alarm setpoint switch to the left.
5. Replace the back cover while ensuring that it slips beneath the hold-down lip and fits securely into posi­tion. Use caution not to pinch the battery leads.
6. Install and tighten screws.
18
Alarm Setpoint Switch
Right - Alarm above the setpoint
Left - Alarm below the setpoint
Ground Resistance Tester Models 3710/3730
Battery
Compartment
Figure 15

Memory

MEM
Model 3730
GROUND TESTER
AEMC
INSTRUMENTS
MEM AL A
ON
OFF
MR
Model 3730
GROUND TESTER
AEMC
INSTRUMENTS
MEM AL A
ON
OFF
To activate the Memory feature, press the A and ON buttons together at power­up (Fig. 16). MEM will be displayed.
The Memory feature allows the user to log into memory 1 to 99 measured values (including OL). Measured values can now be logged for both resistance () and amps (A) by pushing the MEM button during measurement and during “HOLD”.
Figure 16
A long beep indicates the measurement was stored. After the 99th value is entered (MEM 99) will blink indicating memory is now full. Measurements may still be made but no longer logged. Stored values are retained when the instrument is turned off.
Pressing the A and ON buttons again at power-up will de-activate the Memory feature.

Memory Recall

To recall memory values, push ON and MEM together at power-up. MR will be dis­played with the memory number (Fig. 17). Memorized value will be displayed. To select stored values use the cursor arrows ( down, up). For rapid access of memory recall values, press the cursor push-button continuously.
Turn the instrument off to exit the Memory
Figure 17
Recall mode.

Clear Memory

To clear the memory, push ON & MEM together at power-up and hold them down. After a three second interval the “clear” message will be displayed (Fig. 18) along with four repetitive beeps and a fifth long tone. The memory is now cleared (Fig. 19).
Once cleared, releasing the ON & MEM buttons returns the instrument to Resistance measurement mode.
Ground Resistance Tester Models 3710/3730
19
MEM
Model 3730
GROUND TESTER
AEMC
INSTRUMENTS
MEM AL A
ON
OFF
Model 3730
AEMC
INSTRUMENTS
GROUND TESTER
MEM AL A
OFF
ON
Figure 18
Figure 19

Function Access Summary

Push Button(s) Function Setting
Function Used Saved @ Power Down
On/Off ON/OFF Function Default Setting A Function A No Alarm Function () AL Yes Display Hold HOLD No Auto-Off On/Off ON + HOLD No Beeper On/Off ON + No (3710), Yes (3730)

Programming Summary

Push Buttons Function Setting
Function Used Saved @ Power Down
Memory On/Off ON + A Yes Memory Recall (MR) ON + MEM (< 2 sec) No Clear Memory ON + MEM (> 6 sec) No Program Alarm ON + AL No
Setpoint
20
Ground Resistance Tester Models 3710/3730

Principle of Operation

Rx »
1
Σ
n i=1
1
Ri
V
I
= Rx +
1
Σ
n i=1
1
Ri
Typically grounded distribution system may be simulated by the basic circuit shown in Fig. 20 or an equivalent circuit, shown in Fig. 21. If volt­age (V) is applied to any measured grounding electrode Rx through a special transformer, current (I) flows through the circuit, thereby establishing the following equation:
where, usually
Therefore, V/I = Rx is established. If I is detected and measured with V kept constant, the measured grounding electrode resistance Rx can be obtained. A signal is fed to a special transformer via a power amplifier from a 1.689 kHz constant voltage oscillator. The resulting current is then sensed by a detection CT. An active filter is used to dampen earth current at commercial frequency and high-frequency noise.
Example: If we clamp around any grounding electrode in a multi-grounded system, the measured value of the electrode under test will be the resis­tance of that particular rod in series with the equivalent parallel resistance value that the rest of the multi-grounded system represents. If we had an electrical system that had 101 grounding electrodes and each had a resistance value of 25 , and we were to clamp around any electrode in the system, the measured value would be 25 in series with the equiva­lent parallel resistance or 0.25 . The displayed value would be 25.2 (instrument resolution to 0.1 ).
V/I = 25 + 0.25 Rx = 25.2
In most field applications, the number of electrodes that make up a multi-grounded system would be higher, therefore the equivalent parallel resistance is negligible with respect to the rod under test.
Ground Resistance Tester Models 3710/3730
21
I
V
Rx R1 R2 Rn-1 Rn
Rx R1 RnRn-1
RX
R1
R2
Figure 20
Rn-1
Rn
RX
22
R1
Figure 21
Ground Resistance Tester Models 3710/3730
Rn-1
Rn

Field Applications

Grounding conductor
Ground level
Ground rod
Butt plate
Utility
pole
Pole Ground Rods
Remove any molding covering the ground conductor, and provide suffi­cient room for the jaws of the clamp-on. The jaws must be able to close easily around the conductor. The jaws can be placed around the ground rod itself.
Note: The instrument must be placed so that the jaws close around an electrical path from the system neutral or ground wire to the ground rod or rods as the circuit provides.
Select the current range “A.” Clamp onto the ground conductor and mea­sure the ground current. The maximum range is 30.00 A Trms. However, if the ground current exceeds 5 A or noise exceeds 50 V, ground resis­tance measurements are not possible.
measurement.
Remove the instrument from the circuit noting the location
for maintenance and continue to the next test location. After noting the ground current, select the ground resistance range ()
and measure the resistance directly. The reading you measure with the clamp-on tester indicates not just the resistance of the rod, but of the connection to the system neutral and all bonding connections between the neutral and the rod.
Note that in Fig. 22 there is both a butt plate and a ground rod. In this type of circuit, it is necessary to place the instrument above the bond so that both grounds are includ­ed in the test. For future reference, note the date, ohms reading, current reading and pole number. Replace any molding you may have removed from the conductor.
Figure 22
Do not proceed further with the
Ground Resistance Tester Models 3710/3730
23
Note: A high reading or OL indicates one or more of the following:
Service box
Service meter
Pole-mounted transformer
Building wall
Ground
rods
Ground level
Water pipe
Service box
Service meter
Pole-mounted transformer
Building wall
A. Poor ground rod B. Open ground conductor C. High resistance bonds on the rod or splices on the conductor;
watch for buried split butts, clamps, and hammer-on connections.
Service Entrance or Meter
Follow the same procedure as for Pole Ground Rods. Notice that Fig. 24 shows multiple ground rods and in Fig. 23 the ground rods have
been replaced with a water pipe ground. You may also have both types acting as a ground. In these cases, it is necessary to make the mea-
Ground
Level
Building Wall
Pole-mounted
Transformer
Service Meter
Service Box
surements between the ser­vice neutral and both grounded points.
Figure 23
24
Ground
Rods
Building Wall
Pole-mounted
Transformer
Service Meter
Figure 24
Service Box
Water Pipe
Ground Resistance Tester Models 3710/3730
Pad Mounted Transformer
R
1
Ground rod(s)
Enclosure
Underground service
Concentric Neutral
Service
Buss
Open Door
Open Door
Ground
rods
Enclosure
Underground service
Observe all safety requirements — High voltage may be present!
Locate and number all rods (usually only a single rod is present). If the ground rods are inside the enclosure, refer to Fig. 25 and if they are outside the enclosure, refer to Fig. 26. If a single rod is found within the enclosure, the measurement should be taken on the con­ductor just before the bond on the ground rod. Often, more than one ground conductor is tied to this clamp, looping back to the enclosure or neutral.
In many cases, the best reading can be obtained by clamping the instrument onto the ground rod itself, below the point when the ground conductors are attached to the rod, so that you are measuring the ground circuit. Care must be taken to find the conductor with only one return path to the neutral.
Generally 0.7 indicates that you are on a closed loop. In Fig. 26, the ground rod is located outside the enclosure. Clamp at the indicated measuring point to obtain the cor­rect reading. If more than one rod exists at different corners of the enclosure, it is neces­sary to determine how they are connected to properly measure the ground resistance.
Ground Resistance Tester Models 3710/3730
Underground Service
Enclosure
Figure 25
Figure 26
Ground
Rods
25
Transmission Towers
Ground Rod
Support
Concrete Pad
Observe all safety requirements — Dangerously high voltage may be present!
Locate the ground conductor at the base of the tower. Many different configurations exist. Care should be taken when searching for the ground conductor. Fig. 27 shows a single leg mounted on a concrete pad with an external ground conductor. The point at which you clamp the instrument should be above all splices and connections which allow for multiple rods, butt wraps, or butt plates.
Note: Current flowing to ground may be high.
Central Office Locations
The main ground con­ductor from a ground window or a ground plane is usually the location to clamp the
Support
Concrete Pad
tester. Due to the wiring practices within the central office,
Ground Rod
Figure 27
there are many loca­tions at which you can look at the water pipe or counterpoise from with­in the building. An effective location is usually at the ground buss in the power room, or near the backup generator.
By measuring current and resistance at several points and comparing the readings, you will be able to identify neutral loops, utility grounds and central office grounds. The test is effective and accurate when the ground window is connected to the utility ground at only one point.
26
Ground Resistance Tester Models 3710/3730
Service Panel
GROUND BUS
NEUTRAL BUS
SERVICE
PANEL
PHONE
CATV
GROUND ELECTRODE
A B NEUTRAL
ELECTRIC
UTILITY
240 VAC
Bare
WH BK
120 VAC
If we clamp-on the grounding electrode conductor in Fig. 28, the multi­grounded system required to complete the circuit may be from different sources. These sources may include the following: Electric Neutral, Phone, CATV, and Water.
The Electric Utility Neutral should be the principle conductor of the low­impedance signal return path required to test the resistance of this ground electrode with the Model 3710/3730. This can be verified by removing other ground connections, noting if any changes appear in the readings. If by removing one of these signal return paths a much higher reading is noted, it could mean that the primary signal return path (the Electric Neutral) is defective or open.
Note: For safety reasons do not re­move any ground source without first de-energizing the service panel.
Ground Resistance Tester Models 3710/3730
Figure 28
27
Phone Pedestal
Provide sufficient room for the Model 3710/3730 jaws, which must be able to close easily around the conductor (Fig. 29). The jaws can be placed around the ground rod itself.
Note: If not enough room is available for this task refer to Fig. 30 and continue with this example. The clamp must be placed so that the jaws are in an electrical path from the sheath bond(s) or ground bar to the ground rod(s) as the circuit provides.
Select the current range “A.” Clamp onto the ground conductor and measure the ground current. The maximum current range is 30 Amps. If the ground current exceeds 5 Amps, ground resistance measure­ments are not possible. Do not proceed further with the measurement. Remove the clamp-on tester from the circuit, noting the location for maintenance, and continue to the next test location.
After noting the ground current, select the ground resistance range “” and measure the resistance directly. The reading you measure with the 3710/3730 indicates the resistance of not just the rod, but also the con­nection to the sheath ground and all bonding connections between the sheath bond and the rod.
A high reading indicates one or more of the following: A. Poor ground rod B. Open ground conductor C. High resistance bonds on the rod or splices on the conductor. Also
look for bad connections on the ground bar to the sheath(s).
Note: This same test can be done using the TIP instead of the sheath ground. However keep in mind that this is intended to be grounded only back at the central office.
28
Ground Resistance Tester Models 3710/3730
Above Telephone Space
Ground level
Sheath Connection
Ground rod
Phone
Pedestal
Ground Bar
Ground level
Sheath Connection
Ground rod
Phone
Pedestal
Ground Bar
1. Remove ground bar connection to pedestal rod
2. Extend pedestal rod lead to the ground bar using a temporary jumper
Select the current range “A.” Clamp onto the ground wire and measure the current (Fig. 31). The maximum current range is 30 Amps. If the ground current exceeds 5 Amps, resistance measurements are not possible. Do not proceed further with the measurement. Remove the clamp-on tester from the circuit, noting the location for attention, and continue to the next test location.
Ground Resistance Tester Models 3710/3730
Figure 30
Figure 29
29
After noting the current, select the ground resistance range “” and
R
1
R
1
Staples
Ground wire doubled under ground clamp
measure the resistance directly. The reading you measure with the 3710/3730 indicates the bonding connections between the Telephone ground and the Power company ground. One of the following two scenarios should be observed:
A. If “OL” or a high resistance reading is displayed this would indicate
no bond has been established or a poor connection exists between the Telephone and Power grounds.
B. If and/or a very low reading is displayed then the bond between
these two grounds has been verified.
General Measurement Notes
A reading of OL is most likely a cable which is not connected to the ground at both ends. It may be there is no ground rod at all, or that there is no path back to the system neutral.
A reading which is 0.7 generally indicates that the cable where you are clamped is continuous with itself. You may have located a ground loop, which can usually be confirmed by comparatively high current readings also obtained when readings are taken at multiple locations. A good low resistance ground may also have very high AC current flowing, depending upon the type of equipment being grounded.
30
Figure 31
Ground Resistance Tester Models 3710/3730

Operating Reference Card

AEMC
I N S T R U M E N T S
Clamp-on Ground Resistance Tester
Model 3710/3730
Operating Reference
!
ON/OFF
A
HOLD
HOLDONON
+ +
ON/OFF Measure
ground resistance []
Measure AC current [mA, A] Freeze (hold) the display Toggle the buzzer [ ] on/off Toggle auto power on/off
A ON
+
Toggle memory display on/off
AL
MEM
MEM
ON
ON
ON
+
+
+
AL
Turn alarm [ AL ] on/off
MEM
Store the reading in memory
Program alarm mode. Use to adjust
Short beep: On/off memory recall [MR]
Long beep (6s): Clear memory [CLR]
Up
Model 3730
Additional Features
Down
Included with the Model 3710 and 3730 you will find an operating refer­ence card like the one shown here.
This card summarizes all the features of the Model 3710 and 3730.
Ground Resistance Tester Models 3710/3730
Figure 32
31

Calibration Check Loop

®
INST R U M E N T S
CALIBRATION LOOP 25
The multiple calibration loop is now replaced by a single point 25 loop (Fig. 33). It is provided to perform a quick check of the instruments. Check the clamp-on ground resistance tester by clamping around the loop; the 3710 or 3730 should read between 24.2 and 25.8 .
* Value is for ambient temperature between 68° and 78°F. Value may differ a few counts under or above this temperature.
32
Figure 33
Ground Resistance Tester Models 3710/3730

Commonly Asked Questions for Clamp-on Ground Resistance Measurements

Q. Can the Clamp-on method be used effectively on high tension tow­ers despite the distance or spacing between them?
A. Yes. Extended distances will not adversely effect the measurement results. The important factor is that it be a multiple electrode system.
Q. Does the grounding electrode have to be disconnected and isolated as is required when using the “Fall-Of-Potential” test with auxiliary electrodes?
A. No. Actually the electrode must be connected to the system to provide the path for test signal injection as well as to provide the background impedance necessary as the reference. Additionally, if the tested rod is poorly bonded to the ground connector, a high reading will be present.
Q. Does this mean I am not able to test an independent electrode?
A. No, however as soon as the connection is made to a multiple elec­trode system, usually provided by the connection to the system neutral, you can clamp on and make a measurement.
Q. Must the instrument be clamped directly on the electrode?
A. No, the instrument will provide valid measurement results when clamped on to the electrode or the conductor leading to the grounding electrode. The reading through the conductor will also verify not only the grounding electrode resistance, but connections of the ground wire to the rest of the system.
Q. Does the system under test have to be energized or de-energized to perform the test?
A We are measuring the grounding network. The only requirements are that ground connections be made to the system under test and that the voltage-to-ground at the tested point not exceed the instrument rating.
Ground Resistance Tester Models 3710/3730
33
Q. Can the clamp-on method be used for substation grid resistance testing?
A. It depends. It can be used as long as the instrument can be clamped onto a point on the grounding system at a common point where it ties to the system neutral. Testing a grid must be treated as if it were a single electrode. Often, clamping on to points within the grid system will yield a closed loop reading.
Q. How does clamp-on ground testing compare with the standard “Fall-Of-Potential” test?
A. Empirical testing has validated that when performed correctly, both methods will provide accurate and repeatable readings for ground elec­trode resistance.
Q. How large a conductor can the jaw accommodate?
A. The inner diameter of the jaw is 1.25" and can accommodate cables up to 1000 MCM.

Battery Replacement Procedure

The Clamp-On Ground Resistance Testers are powered by a single 9 volt battery. The battery replacement indicator will display continuously when battery replacement is required.
Recommended replacement type is Alkaline (IEC 6LF22, 6LR61 or NEDA 1604A). Ni-Cad batteries may also be used. To change the bat­tery refer to figure 15, page 17 and follow this procedure.
1. The instrument must be OFF and disconnected from any conductor.
2. Place the meter face down and remove the two Phillips screws on the back of the instrument.
3. Remove back cover by pulling it down and away from the instrument.
4. Lift the battery away from the case and disconnect the battery.
5. Connect the new battery and position the leads so that they won’t get pinched when the cover is replaced.
34
Ground Resistance Tester Models 3710/3730
6. Replace the back cover noting that it slips beneath the hold down lip and fits securely into position.
7. Install and tighten screws.
Typical battery life is approximately 8 hours of use or about 1000 separate 30-second measurements.

Troubleshooting

Symptom Indication/
Instrument will not turn ON Low Battery
Recommended Action
Replacement required
Continuous audible tone Alarm setpoint tripped
Deactivate Alarm or Change Setpoint Trigger mode
Display is erratic Foreign matter in jaws
Clean jaw mating surfaces
Function controls inoperative Hold button on
Turn HOLD off
Light pulsing audible beep Normal operation
Battery power management feature
Ground Resistance Tester Models 3710/3730
35

Repair and Calibration

To guarantee that your instrument complies with the factory specifica­tions, we recommend that the Models 3710 & 3730 be submitted to our factory service center at one-year intervals for recalibration, or as required by other standards.
For instrument repair and/or calibration, please call our factory, toll-free, at (800) 945-AEMC (800-945-2362):
CHAUVIN ARNOUX, Inc. d.b.a. AEMC®Instruments 15 Faraday Drive Dover, NH 03820 USA Tel: (800) 954-2362
(603) 749-6434
Fax: (603) 742-2346 (Or contact your authorized distributor.) Estimates for repairs, normal recalibration, and calibration traceable to
N.I.S.T. are available upon request. Overseas customers must receive written authorization before returning any instrument.

Technical and Sales Assistance

If you are experiencing any technical problems, or require any assis­tance with the proper use or application of this instrument, please call our technical hotline:
CHAUVIN ARNOUX, Inc. d.b.a. AEMC® Instruments
200 Foxborough Blvd. Foxborough, MA 02035 USA
Tel: (800) 343-1391
(508) 698-2115
Fax: (508) 698-2118
www.aemc.com
36
Ground Resistance Tester Models 3710/3730

Limited Warranty

These Ground Resistance Testers Models 3710 & 3730 are warrantied to the owner for a period of 1 year from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC®Instruments, not by the distributor from whom it was purchased. This warranty is void if the Ground Resistance Testers Models 3710 & 3730 have been tampered with, abused, or if the defect is related to ser­vice not performed by AEMC®Instruments.
What AEMC®Instruments Will Do:
year period, you may return the Ground Resistance Testers Models 3710 & 3730 to us for repair or replacement free of charge, provided we have your REGISTRATION CARD on file. AEMC®Instruments will, at its option, repair or replace the faulty material.
Note: If a card is not on file, we will require a dated proof of purchase as well as your REGISTRATION CARD accompanied by the defective material.
What You Must Do:
from AEMC®Instruments, then return the Ground Resistance Testers Models 3710 & 3730, indicating place and date of purchase, with a writ­ten explanation of the reason for return. Return material, postage pre­paid to:
Chauvin Arnoux, Inc. d.b.a. AEMC®Instruments Service Department 15 Faraday Drive Dover, NH 03820 USA Tel: (800) 945-2362
Fax: (603) 742-2346
Caution: To protect against in-transit loss, we recommend that you insure your returned material.
For full warranty coverage, please read the Warranty Card which is affixed to the Warranty Registration Card. Please keep the Warranty Card with your records.
First obtain a return authorization by phone or by fax
(603) 749-6434
If a malfunction occurs within the 1
Clamp-On Ground Resistance Testers Models 3710 and 3730
USER MANUAL
Chauvin Arnoux, Inc. d.b.a. AEMC® Instruments
99-MAN-100061 Rev 05 03/98
Loading...