AEMC MN03 Data sheet

AC Current Probe Model MN03

Small and compact, the AC Current Probe Model MN03 is the ideal complement for any multimeter to measure AC currents in low-power secondary or industrial applications.
Technical Assistance (800) 343-1391 www.aemc.com 1 of 4

Features

Small, compact size
Measurement range of 1 to 100A
Jaw opening accommodates
conductors up to 0.39'' diameter
Designed for DMMs, loggers, recorders and oscilloscopes
40Hz to 10kHz response
a c /Aa c output signals
1mV
Designed to EN 61010, 600V
CAT III safety standard
• DoubleInsulation
CE Mark

Applications

Measuring in breaker panels
a c
Industrial loads
HVAC
Residential and commercial sites
Data logging/recording
MN03_2129.18 (pdf) 0109 Rev.01

Specifications

MODEL MN03
ELECTRICAL
Range 100A Measurement Range 1 to 100A Accuracy in % 2% + 50mA Phase Shift Not specified Output Signal 1mV
(100mV at 100a)
Maximum Currents 1 < 150A permanent from 48Hz to 500Hz Influence of Temperature 0.2% per 10K Influence of Adjacent
Conductor Influence of Adjacent
Conductor Position in Jaws Influence of Frequency 2% from 65Hz to 500Hz
MECHANICAL
Operating Temperature 14° to 122°F (-10° to +50°C) Storage Temperature -40° to 176°F (-40° to 80°C) Operating Relative
Humidity
0 to 85% RH decreasing linearly above 35°C
Operating Altitude 0 to 2000m
Casing Protection Rating
(leak-proofing)
IP40
Fall Height 4.92 ft (1.5m) (IEC 68-2-32) Shock Protection 3.52 oz (100g)/6ms/half-period (IEC 68-2-27) Self-extinguishing Capability casing UL94 V2 Vibration Resistance IEC 68-2-6 Dimensions 4.43 x 1.48 x 1.02" (112.5 x 37.5 x 26mm) Weight approx. 6.35 oz (180g)
SAFETY
Electrical EN 61010-1, 600V CAT III, Pollution Degree 2
Double Insulation Yes
CE Mark Yes
(1) Conditions of reference: 23°C ± 3°K, 20 to 75% RH, sinusoidal signal with frequency of 48 to 500Hz, distortion factor < 1% with no DC component, external DC magnetic field < 40A/m, no external AC magnetic field, no external conductor with circulating current, conductor centered for measurement, measurement instrument load impedance 1MΩ (10A calibre) & 10kΩ (100A calibre).
(2) With clamp closed.
a c /aa c
2mA/A at 50Hz
0.1% at 50/60Hz
(2)
(EN 60529 Ed. 1995)
MN03_2129.18 (pdf) 0109 Rev.01