AC Current Probe
Model MN02
The MN02 is a small, compact AC
current probe. Designed to meet
the most stringent demands in
industry and electrical contracting,
they also meet the latest safety
and performance standards.
The probe has a measurement
range up to 100Arms which
makes it a perfect tool for
measurement with DMMs and
recorders.
The Model MN02 is compatible
with any AC ammeter, multimeter,
or other current measurement
instrument with an input
impedance lower than 10Ω.
To achieve the stated accuracy,
use the probe with an ammeter
having an accuracy of 0.75% or
better.
Technical Assistance (800) 343-1391 www.aemc.com 1 of 4
Features
• Small, compact size
• Measurement ranges of
50mA to 100A (1Ω load)
50mA to 90A (10Ω load)
• Jaw opening accommodates
conductors up to 0.39" diameter
• Designed for DMMs, loggers,
recorders and oscilloscopes
• 48 to 10,000Hz frequency range
• 1mA/A from 1Ω to 10Ω
output signals
• Designed to EN61326-1,
300V CAT IV, 600V CAT III
safety standard
• DoubleInsulation
• CE Mark
Applications
• Measuring in breaker panels
• Industrial loads
• HVAC
• Residential and commercial sites
• Data logging/recording
• General AC current measurements
MN02_2129.20 (pdf) 0710 Rev.01
![](/html/32/3251/3251689acbe2903d22b440c7eb22add795d9f865740654a216143ad6002d2749/bg2.png)
Specifications
MODEL MN02
ELECTRICAL
Range 100A
Measurement Range 50mA to 100A (1Ω load)
50mA to 90A (10Ω load)
Accuracy in % ≤ 1% Reading ± 0.02A (1Ω load)
≤ 1.5% Reading ± 0.01A (10Ω load)
Phase Shift Not specified
Output Signal 1mA/A from 1Ω to 10Ω
Load Impedance ≤10Ω
Maximum Currents 1 < 150A permanent from 48Hz to 500Hz
Influence of Temperature ≤ 0.2% per 10˚K
Influence of Adjacent
Conductor
Influence of Adjacent
Conductor Position in Jaws
Influence of Frequency ≤ 2% from 65Hz to 500Hz
Maximum Output Voltage
(secondary open)
MECHANICAL
Operating Temperature 14° to 122°F (-10° to +50°C)
Storage Temperature -40° to 176°F (-40° to +80°C)
Operating Relative
Humidity
0 to 85% RH decreasing linearly above 35°C
Operating Altitude 0 to 2000m
Casing Protection Rating
(weatherproofing)
Drop 4.92 ft (1.5m) (IEC 68-2-32)
Shock Protection 3.52 oz (100g)/6ms/half-period (IEC 68-2-27)
Self-extinguishing Capability casing UL94 V2
Vibration Resistance IEC 68-2-6
Dimensions 4.43 x 1.48 x 1.02" (112.5 x 37.5 x 26mm)
Weight 6.5 oz (180g)
SAFETY
Electrical EN61326-1, 300V CAT IV, 600V CAT III, Pollution Degree 2
Double Insulation Yes
CE Mark Yes
(1) Conditions of reference: 18°C to 28°C, 20 to 75% RH, external magnetic field <40 A/m, 48 to 65 Hz sine wave, distortion factor less
than 1%, no DC component, no external current carrying conductor, test sample centered. Load impedance 1Ω or 10Ω.
≤ 2mA/A at 50Hz
≤ 0.1% at 50/60Hz
≤ 30V
IP 40 (IEC 529)
MN02_2129.20 (pdf) 0710 Rev.01
Technical Assistance (800) 343-1391 www.aemc.com 2 of 4