AEMC JM875 Data sheet

AC Current Probes Model JM875 (pdf) Rev. 03 10/05
Technical Assistance (800) 343-1391 www.aemc.com1of 4
AC Current Probe Model JM875
The Model JM875 is a high performance AC current probe for higher currents. Excellent transformation and low phase shift, plus a broad frequency response, permit accurate measurements of current and power. The high quality magnetic cores and uniform windings provide for accurate measurements up to 30A
AC/300AAC/
3000A
AC. The “squared” jaw permits
multiple conductor or bus bar positioning. The Model JM875 provides excellent True RMS capabilities and reliable signal reproduction.
The Model JM875 provides an accurate voltage output proportional to the measured current (1mV/A, 10mV/A or 100mV/A). This voltage output enables instruments without current ranges to display or log currents on an AC voltage range.

Features

Nominal range: 30AAC, 300AAC,
3000A
AC (continuous cycle for the
full temperature range)
Output: 100mV
AC/AAC, 10mVAC/AAC,
1mV
AC/AAC
Designed for use on DMMs, power
and harmonic meters, recorders or
instruments with AC current ranges

Applications

Field testing
Electrical utility
HVAC
Industrial maintenance
Laboratory
Plant maintenance
AC Current Probes Model JM875 (pdf) Rev. 03 10/05
Technical Assistance (800) 343-1391 www.aemc.com2of 4
MODELS JM875
ELECTRICAL
Nominal Range 30A, 300A, 3000A Measurement Range 1 to 3200A Transformation Ratio Voltage output Output Signal 100mV
AC/AAC (3V @ 30A) 90A Peak
10mV
AC
/AAC (3V @ 300A) 900A Peak
1mVac/AAC (3V @ 3000A) 9000A Peak
Accuracy (30A Range)
1.5A 2% of Reading ± 10mV (0.5A) 6A 2% of Reading ± 10mV (0.5A)
30A 2% of Reading ± 10mV (0.5A)
Phase Shift (30A Range)
1.5A 15° 6A
30A
Accuracy (300A Range)
15A 2% of Reading ± 2mV (0.2mA) 60A 2% of Reading ± 2mV (0.2mA)
300A 2% of Reading ± 2mV (0.2mA)
Phase Shift (300A Range)
15A 3° 60A 1.5°
300A 1
Accuracy (3000A Range)
150A 2% of Reading ± 0.5mV 600A 2% of Reading ± 0.5mV
3000A 2% of Reading ± 0.5mV
Phase Shift (3000A Range)
150A 1.5° 600A
3000A 0.5°
Overload 3200A for 5 min Frequency Range 30 to 5kHz; current derating abouve 200Hz on the 3000A range Load Impedance >100kHz Working/Common
Mode Voltage
600V
AC
Output Termination Jack
MECHANICAL
Operating Temperature 14° to 122°F (-10° to 50°C) Storage Temperature -13° to 176°F (-25° to 80°C) Jaw Opening 3.54" (90mm) Maximum Conductor Size Cable: 2.52" Ø max (64mm); Bus bar: 1.97 x 5.31" (50 x 135mm); 2.52 x 3.94" (64 x 100mm) Dimensions 4.72 x 12.40 x 1.89" (120 x 315 x 48mm) Weight 2.65 lbs (1200g) Polycarbonate Material Handles: 10% fiberglass charged polycarbonate UL94 VO; Jaws: ABS UL V2
SAFETY
Electrical EN61010-2-32 Electromagnetic EN 50081-1 Class B; EN 50082-2 Electrostatic discharge IEC 1000-4-2;
Compatibility Radiated field IEC 1000-4-3; Fast transients IEC 1000-4-4; Magnetic field at 50/60Hz IEC 1000-4-4
Double Insulation Yes
CE Mark Yes
Note: Reference conditions:
23°C ± 3°K, 20 to 85% RH, 48 to 65Hz, external magnetic field <40A/m, no DC component, no external current carrying conductor, test sample centered, load impedance ≥ 1mΩ.